Cell migration is central to a variety of physiologic

Size: px
Start display at page:

Download "Cell migration is central to a variety of physiologic"

Transcription

1 A Quantitative, Facile, and High-Throughput Image-Based Cell Migration Method Is a Robust Alternative to the Scratch Assay Wendy Gough, 1 Keren I. Hulkower, 3 Renee Lynch, 1 Patrick Mcglynn, 2 Mark Uhlik, 2 Lei yan, 2 and Jonathan A. Lee 1 Cell migration is a key phenotype for a number of therapeutically important biological responses, including angiogenesis. A commonly used method to assess cell migration is the scratch assay, which measures the movement of cells into a wound made by physically scoring a confluent cell monolayer to create an area devoid of cells. Although this method has been adequate for qualitative characterization of migration inhibitors, it does not provide the highly reproducible results required for quantitative compound structure-activity relationship evaluation because of the inconsistent size and placement of the wound area within the microplate well. The Oris Cell Migration Assay presents a superior alternative to the scratch assay, permitting formation of precisely placed and homogeneously sized cell-free areas into which migration can occur without releasing factors from wounded or dead cells or damaging the underlying extracellular matrix. Herein the authors compare results from the scratch and Oris cell migration assays using an endothelial progenitor cell line and the Src kinase inhibitor dasatinib. They find that using the Acumen Explorer laser microplate cytometer in combination with the Oris Cell Migration Assay plate provides a robust, efficient, and cost-effective cell migration assay exhibiting excellent signal to noise, plate uniformity, and statistical validation metrics. (Journal of Biomolecular Screening 2011;16: ) Key words: cellular imaging, cell migration, dasatinib, high-content imaging, high-throughput screening, multiplexed assay, phenotypic assay, phenotypic mechanism of action, scratch assay Introduction Cell migration is central to a variety of physiologic processes such as tissue repair and regeneration as well as pathologic processes such as cancer. Abnormal cell migration is attributed to the failure of wounds to heal as well as to the capabilities of tumor cells to metastasize. 1 As such, mechanisms involved in cellular migration are attractive targets for therapeutic intervention. We have been measuring cell migration as it relates to therapeutically important biological responses, including angiogenesis. We initially employed an image-based scratch migration assay read on the Acumen Explorer. In this method, a wound area that is devoid of cells is created within a confluent cell 1 Departments of Quantitative and Structural Biology. 2 Angiogenesis-Tumor Microenvironment, Lilly Research Laboratories, Eli Lilly and Company, Indianapolis, IN, USA. 3 Platypus Technologies, LLC, Madison, WI, USA. Received Apr 23, 2010, and in revised form Nov 1, Accepted for publication Nov 4, Journal of Biomolecular Screening 16(2); 2011 DOI: / monolayer in 96-well plates. Cells are stained with a vital dye, imaged, treated with drug, and reimaged after 24 h to determine the extent of cellular migration and its inhibition by compounds. Although this method is useful for qualitative characterization of migration inhibitors, it does not provide the highly reproducible results required by the pharmaceutical industry for quantitative compound testing. The variability in this scratch assay format reflects the difficulty in creating a reproducible cell-free area with uniform dimensions and placement within a microplate well. 2 An additional caveat of the scratch assay is that the dead and dying cells resulting from the wounding process may release apoptotic factors and other cytokines that could potentially affect signal transduction pathways in the cells that close the wound. Finally, the process of creating a cellular scratch area could also physically damage the underlying extracellular matrix and/or plate surface, thereby causing an adverse effect on cell migration. 3 Recently, the Oris Cell Migration Assay was developed to overcome these shortcomings associated with the scratch assay. The Oris assays use silicone stoppers as barriers that contact the bottom of an optically clear 96-well black skirted fluorescence microplate to form precisely positioned 2-mm central cell-free zones. Cells are seeded and allowed to attach and 2011 Society for Laboratory Automation and Screening 155

2 Gough et al. spread in the presence of the stoppers. Subsequent removal of the stoppers reveals the central cell-free zones surrounded by a monolayer of cells into which migration can occur. This assay format allows an unobstructed view of cell motility throughout the duration of the experiment without physically damaging cells or underlying extracellular matrices. It also provides decreased variability and increased robustness over woundhealing/scratch assays and is compatible with high-content imaging (HCI) techniques. Unlike Boyden chamber-based assays that use membrane filters or motility assays that track cells via displacement of fluorescent beads, Oris Cell Migration Assays allow for direct visualization of the cells at any point during the assay with the ability to use multiplexed staining to discern phenotypic features of migrating cells. Hci assays have recently begun to provide information-rich, multiparameter readouts with fixed or live-cell assays amendable for medium- to high-throughput compound testing. 4,5 In addition, Hci assays allow the investigator to define specific regions within the microplate well for analysis with minimal additional cost per assay data point. The Acumen cytometer sequentially scans a microplate with up to 3 lasers and can thus provide basic cell migration information in addition to information on various cellular processes such as changes in signal transduction or cellular differentiation states. The application of laser scanning over the entire microplate surface means that analysis is not performed by a well basis, which allows for the Acumen instrument to perform and analyze single-color scans of 96-, 384-, or 1536-well plates in less than 10 min. Herein we compare the scratch assay to the Oris Cell Migration Assay to measure migration of endothelial colonyforming cells (ECFCs) and its inhibition by dasatinib, a Src kinase inhibitor. 6 Our results using the Acumen Explorer demonstrate that the Oris Cell Migration Assay yields better plate uniformity, more reproducible concentration-response curves, and more robust Z factors than the scratch assay. In addition, the Oris Cell Migration Assay requires fewer handling steps and eliminates the need for collecting pre- and postmigration images from the test wells, making it a more cost-effective and superior alternative to the scratch assay. Enablement of a robust and statistically rigorous cell migration assay allows quantitative evaluation of complex cell biological processes to support target-directed lead optimization 7 and phenotypic drug discovery. 8 Materials Materials and Methods Chemicals and materials were purchased from the following vendors and used without modification: dasatinib (US Biological, Swampscott, ma), ecfc (human endothelial colonyforming cells; EndGenitor Technologies, Inc., Indianapolis, IN), EGM -2MV Microvascular Endothelial Cell Growth Medium-2 (Lonza, Walkersville, MD), Angiokit Optimised Growth Medium Pack (TCS CellWorks, Buckingham, UK), CellTracker Green CMFDA and fetal bovine serum (Invitrogen, Carlsbad, CA), 96-well poly-d-lysine-coated cell culture plates (BD, Franklin Lakes, NJ), Prefer (Anatech Limited, Battle Creek, MI), propidium iodide (Molecular Probes, Eugene, OR), and Oris Cell Migration Collagen I Coated plates (Platypus Technologies, LLC, Madison WI). Scratch assay ECFCs (40,000/well) were plated in EGM -2MV media containing 10% fetal bovine serum (FBS) on poly-d-lysinecoated 96-well plates. Following an overnight incubation, the media were removed from the cells and replaced with serumfree media containing 3.6 µm CellTracker Green CMFDA. The CellTracker Green CMFDA solution was removed after a 30-min incubation at 37 C and replaced with complete media containing FBS. The cells were allowed to recover for at least 30 min, and scratches were made in the monolayers using a pin tool attachment on a Beckman Coulter Multimek instrument (Beckman Coulter, Brea, ca). The cells were rinsed with 200 µl phosphate-buffered saline (PBS) and premigration images collected on the Acumen Explorer. PBS was then removed, and cells were treated with various concentrations of dasatinib in Angiokit Optimised Growth Medium containing 10% FBS, 0.4% DMSO. After a 24-h incubation with the compound, the media was removed and replaced with 200 µl PBS for postmigration image acquisition. The Acumen Explorer setup includes 488-nm laser excitation, a 500- to 530-nm emission filter, and a software-defined rectangle surrounding the cellfree scratch area (the migration area). Mean intensity measurements were used to gate out cells with dim signals from debris, and total fluorescent area (µm 2 ) within the migration area was used for analysis. Oris Cell Migration Assay ECFCs (30,000/well) were plated in EGM -2MV media containing 10% FBS on Oris Cell Migration Collagen I Coated plates containing cell seeding stoppers. The cells were allowed to attach and spread for 16 to 20 h prior to manual removal of the stoppers. The cells were washed and media replaced with Angiokit Optimised Growth Medium containing 10% FBS, 0.4% DMSO, and the indicated concentrations of dasatinib. After a 24-h incubation with the compound, the media were removed and the cells were incubated in 100 µl of Prefer fixative. After 30 min, the cells were permeabilized by treating with PBS containing 0.1% Triton X-100 for 15 min. Following TX-100 treatment, the cells were washed twice with PBS, and 50 µl of 15 µm propidium iodide in RNase-free PBS was added to the cells. Images were captured on the Acumen Explorer using a 488-nm laser and >655-nm filter. A square Journal of Biomolecular Screening 16(2); 2011

3 Quantitative Cell Migration Assay around the cell-free zone, the migration area, was defined. In the operational version of the assay, a rectangle encompassing the confluent cell monolayer from the same well defined the cytotoxicity region. Mean intensity measurements were used to gate out cells with dim signals from debris. Total fluorescent cellular area (µm 2 ) in the migration and cytotoxicity areas were used to calculate percent inhibition of migration (equation 1) and percent cell death (equation 2) where Area Dmso, Area compound treated, and Area max dasatinib are the total cellular areas within the assay s analysis region in the presence of 0.4% DMSO (solvent control), test compound, and a maximum inhibitory dose of dasatinib (40 µm), respectively. ( Area DMSO Areacompound treated ) % Inhibition migration = 100 (1) ( Area Area ) Data analysis DMSO maxdasatinib ( Area DMSO Areacompound treated ) % Cell death = 100 (2) Area DMSO Plate uniformity plots and concentration-response curves were generated for the data obtained using the scratch and Oris Cell Migration Assays using a plate uniformity calculator 9 and GraphPad Prism (GraphPad Software, La Jolla, Ca). Three-day plate uniformity and replication of potency studies were used to statistically validate the results of the Oris Cell Migration Assay using the Eli Lilly and National Institutes of Health (NIH) Chemical Genomics Center (NCGC) quantitative biology guidelines. 9 Results and Discussion Evaluation of a scratch assay for cell migration We evaluated the effect of dasatinib dose on ECFC migration in a traditional scratch assay. Premigration images of labeled cells were captured following wound formation within the ECFC monolayer (Fig. 1a). Red rectangles in the images define a specific analysis region that is identically placed within each well. Variation in the size and placement of the scratch in the premigration reference images illustrates the nonuniform nature and inconsistent dimensions of the scratch wound area (Fig. 1a). Following 24 h of incubation, ECFCs fill in and close the wound in control samples, whereas dasatinib inhibits wound closure. Data analysis from a plate uniformity experiment using high, medium, and low dasatinib concentrations results in a Z factor of 0.20 (Fig. 1b) without obvious edge effects. These results indicate poor assay performance. 9 A dasatinib dose-response experiment using this assay format shows a qualitative decrease in wound closure with increasing inhibitor concentrations, but extensive replicates are required to resolve a dose-response curve (Fig. 1b). Analysis of the dose-response curve using 8 replicates per data point results in an IC 50 of 1.5 µm in agreement with previous findings. 6 Taken together, these results illustrate the issues related to the statistical robustness and efficient operation of the traditional scratch assay and indicate the unsuitability of this assay format for efficient quantitative compound evaluation. It is important to note that the statistical robustness of the scratch assay is highly dependent on the method and automation used to initiate the monolayer wound. In contrast with our results, Simpson et al. 10 have shown that the scratch assay was effective in the assessment of sirna molecules that acted as inhibitors and activators of wound healing. These investigators observed different Z factors for accelerators and inhibitors of cell migration such as RHOA sirna (0.41) and RAC1 sirna (0.29), respectively. 10 Evaluation of the Oris Cell Migration Assay The Oris Cell Migration Assay differs from a scratch assay by use of silicone stoppers to act as nondamaging barriers during the cell seeding process. Following cell adhesion and stopper removal, the monolayer exhibits circular cell-free regions of uniform size. Superposition of the migration region boundaries onto the cell-free regions in the cell monolayer illustrates the reproducible placement and size of this region within each microplate well (Fig. 2a, top row). During assay operations, compounds are added to test plates immediately following the stopper removal and washing steps and then incubated for an additional 24 h. Dasatinib inhibits the closure of the cell-free zone in a dose-dependent manner. Data analysis from a plate uniformity experiment using high, medium, and low dasatinib concentrations results in a Z factor of 0.70 without obvious edge effects (Fig. 2b), a significant improvement over the scratch assay performance (signal window of 1.52 and a Z factor of 0.20; Fig. 1b). Preliminary doseresponse data indicate a significantly smaller standard deviation using 4 replicates per data point in the Oris assay (Fig. 2b) compared to results obtained with the scratch assay using 8 replicates per data point (Fig. 1b). Taken together, the Oris assay provides high-quality data by making cell-free regions of reproducible size and uniform well placement. Validation of the Oris Cell Migration Assay for high-throughput screening To investigate whether the Oris Cell Migration Assay was suitable for high-throughput screening (HTS), we followed the 3-day plate uniformity and replication of potency validation protocols established by the Eli Lilly and NIH Chemical Genomics Center. 9 The plate uniformity assessment tests 3 plates with interleaved high, mid, and low dasatinib concentrations per day for 3 days. Row or column view from a typical plate is shown in Figure 2b; no evidence of systematic variation Journal of Biomolecular Screening 16(2);

4 Gough et al. FIG. 1. Scratch assay: effect of dasatinib on endothelial colony-forming cell (ECFC) migration. (a) Analysis of premigration and 24-h migration images from untreated and dasatinib-treated cells using an Acumen Explorer. A detection zone (red) was defined in the same location for all wells and the area covered with cells determined under the indicated conditions. (b) Plate uniformity and dasatinib concentration response results of ECFC migration. Values from each well from three 96-well plates configured with an interleaved high, mid, and low concentration of dasatinib corresponding to 0, 1, and 20 µm, respectively (left). The resulting Z factor was Concentration-response curve of dasatinib titration; each point represents the average of 8 assay wells ± SD Journal of Biomolecular Screening 16(2); 2011

5 Quantitative Cell Migration Assay FIG. 2. Oris Cell Migration Assay: effect of dasatinib on endothelial colony-forming cell (ECFC) migration. (a) Analysis of ECFC-covered area following 24-h migration with the indicated concentration of dasatinib or immediately following stopper removal (pre-migration). A detection zone (green square) was defined in the same location for all wells and the area covered with cells determined under the indicated conditions. (b) Plate uniformity and dasatinib concentration-response results of ECFC migration. Values from each well from three 96-well plates configured with an interleaved high, mid, and low concentration of dasatinib corresponding to 40, 2, and 0 µm, respectively (left). The resulting Z factor was Concentration-response curve of dasatinib titration; each point represents the average of 4 assay wells ± SD. or edge effects is observed. Aggregation of the high, mid, and low results and corresponding Z scores for the validation experiment (9 plates over 3 days) are shown Fig 3a. Z scores ranged from 0.54 to 0.77 well above the Z score of 0.4 suggested for cell-based screening. 9 This analysis indicates that Oris migration assay passes Eli Lilly NCGC screening Journal of Biomolecular Screening 16(2);

6 Gough et al. FIG. 3. Oris Cell Migration Assay: statistical validation of plate uniformity and reproducibility of compound potency. (a) Aggregated plate uniformity data (3 plates/day for 3 days) from interleaved high, mid, and low concentrations of dasatinib corresponding to 40, 2, and 0 µm, respectively. Summary of Z scores obtained from plate uniformity experiment, 3 plates/day for 3 days. (b) Replication of compound potency. Twenty-nine compounds were tested in 10-point dose-response experiments in 2 independent runs on separate days using a single measurement per data point. Compound potency reproducibility is expressed as a ratio of IC 50 values versus geometric mean of IC 50 value and log IC 50 versus log IC 50 values obtained between runs. The calculated within-run minimum significant ratio (MSR) value of 1.97 indicates that compounds with IC 50 ratios >2.0, if determined within the same experiment, are significantly different. guidelines and has sufficient reproducibility and plate uniformity to test compounds using single-point determinations. In addition to plate uniformity, assays supporting quantitative analysis of compound structure-activity relationships must provide reliable and reproducible measures of compound potency. The reproducibility of potency measurements was evaluated by comparing the IC 50 values of 29 migration inhibitors of varying potency in 2 independent experiments (Fig. 3b) and by calculation of the minimum significant ratio (MSR). 11 The pass criteria for potency measurements require an MSR value of <3, indicating that 2 compounds that have an efficacy ratio greater than the MSR are statistically significant. 11 Graphical comparison of IC 50 value variation in the Oris migration assay indicates very good day-to-day reproducibility. The calculated within-run MSR value of 1.97 indicates that compounds with IC 50 values that differ more than a factor of 2 (in the same experiment) are statistically significant. 11 Taken together, the plate uniformity and compound potency data (Fig. 3) indicate that the Oris -Acumen Explorer migration assay is a robust and statistically validated assay capable of supporting quantitative compound evaluation using a single determination per data point. Although statistically validated, this stopperbased migration assay is limited for HTS operations by the need for manual stopper removal. Preliminary results using a stopper-free version of the Oris Cell Migration Assay resulted in Z scores above 0.5 (unpublished data). The stopper-free Oris Pro Cell Migration Assay employs a self-dissolving biocompatible gel to form the exclusion zone, which does not require manual intervention and may be useful for HTS operations with further optimization. Phenotypic multiplexing with the Oris Cell Migration Assay In addition to compound screening and structure-activity relationship support, quantitative biologists frequently need to Journal of Biomolecular Screening 16(2); 2011

7 Quantitative Cell Migration Assay FIG. 4. Multiplexed assay simultaneously monitors endothelial colony-forming cell (ECFC) migration and cytotoxicity, which defines phenotypic mechanism. The cell-covered areas of 2 regions of interest are monitored after 24 h of drug exposure. Increase in cell-covered area within the green area monitors cell migration; decrease in cell-covered area of the blue area monitors cytotoxicity. Cell-covered area (red) and area outside of well (yellow) are excluded from analysis. Concentration-response curves of cell migration inhibition (black line) and cell death (blue line) for a true migration inhibitor, dasatinib, and a false positive, compound A, using a single measurement per data point. devise testing paradigms to investigate compound mechanism of action or differentiate test compounds from standard of care compounds. In the case of a putative cell migration inhibitor, a trivial and unwanted mechanism of action is compound-mediated cytotoxicity. To address this issue, the operational version of the assay defines 2 regions of interest within each migration plate well (Fig. 4). The first analysis area encompasses the circular space, which is created after stopper removal and is devoid of cells. Closure of this region by surrounding cells is a measure of cell migration and has been statistically validated (Fig. 3A,B). The second analysis region, a portion of the same microplate well seeded with cells, provides a means to determine changes in cellular area, a nonvalidated, relative measure of compound-mediated monolayer toxicity (Fig. 4). The use of multiplexed assay readouts simultaneously provides information on compound potency in ECFC migration and cytotoxicity, which differentiates the phenotype of a true migration inhibitor (dasatinib; Fig. 4) from a false positive (compound A; Fig. 4). The Oris Cell Migration Assay has previously been used to examine the promigratory effects of anthocyanins on fibroblasts and keratinocytes, 2 models of wound healing, where statistically significant dose- and time-dependent stimulation of migration was observed for both cell types. 12 In agreement with our results, a recent preliminary communication demonstrated the ability to achieve robust Z factors and signal windows using the Oris Cell Migration Assay in conjunction with an Acumen ex3 microplate cytometer to analyze HT-1080 and MDA-231 cells labeled with nuclear, whole-cell, or actin-specific dyes. 13 The current study extends the observations of Wylie et al. 13 by incorporation of (1) an alternative measure of cell count (area rather than cell number), (2) rigorous statistical assay validation procedures, (3) a direct comparison of the Oris Cell Migration Assay with the traditional scratch assay, and (4) a multiplexed phenotypic readout that identifies true migration actives (Fig 4). The Oris Cell Migration Assay differs from other migration assay formats in regard to the overall biology that is potentially Journal of Biomolecular Screening 16(2);

8 Gough et al. FIG. 5. Comparison of scratch and Oris Cell Migration Assays: compounds that modulate tubulin polymerization or inhibit SRC, RAC1, ROCK1, and myosin II activity were tested in the endothelial colony-forming cell (ECFC) assay in both formats. Inhibitors of ROCK1 and myosin II were inactive in both formats. A plot of log IC 50 (scratch), a nonvalidated assay, versus log IC 50 (Oris ), a validated assay, of the remaining compounds results in R 2 = 0.6, which increased to 0.8 after elimination of 1 outlier. interrogated. In addition to cell migration, the scratch assay may be sensitive to factors related to cell damage or wound healing. Similarly, Boyden chamber-based assays may be more sensitive with biological systems where compound gradients are important. The Oris migration assay measures the biology directly relevant to cell motility independent of the various chemosensing functions that are components of other cell migration assays. Direct comparison of 17 compounds that modulate tubulin polymerization or inhibit src, rac1, rock1, and myosin ii resulted in 12 compounds with ic 50 values with less than a 3-fold difference between the Oris and scratch assays. Significantly, both the rock1 inhibitors, Y and fasudil, and the myosin ii inhibitor blebbistatin were inactive in both formats, consistent with the observation that Rho-kinase and myosin ii inhibitors decrease endothelial cell migration in 3D but not 2D cell culture 14 (data not shown). Log ic 50 (scratch) versus log ic 50 (Oris ) plots of the remaining 14 compounds resulted in R 2 = 0.6, which increased to 0.8 after elimination of 1 outlier (Fig. 5). Although it is difficult to compare results between validated and nonvalidated assays, the data indicate an overall correspondence between the 2 assay formats for the compounds tested. The development of a reliable and quantitative assay to measure cellular migration has greatly facilitated our ability to classify compounds by biological phenotype in early drug discovery. The Oris Cell Migration Assay, when optimized for use with fixed-cell imaging, is more stable, has fewer steps, and yields results that are more reproducible than the scratch assay. Many of the time-consuming and error-prone steps of the livecell scratch assay were eliminated by using the Oris Cell Migration Assay with fixed cells. These modifications, along with the uniform migration areas and the versatility of the Acumen Explorer, have allowed us to obtain results that exceed the required signal-to-noise, plate uniformity, and compound potency reproducibility criteria established by Eli Lilly and ncgc for assay validation. 9 In addition, incorporation of phenotypic multiplexing in the assay provides an effective means to readily identify and eliminate migration inhibitors working through cytotoxic mechanisms. Given the analyte multiplexing potential of the Acumen Explorer ex3, 13 additional cellular parameters such as biomarkers of specific signal transduction pathway activities or cellular differentiation states can be easily added to future assays. Use of the Oris Cell Migration Assay will allow us to make sure and rapid progress in targetdirected lead optimization 7 or phenotypic drug discovery 8 efforts. References 1. Horwitz R, Webb D: Cell migration. Curr Biol 2003;13:R756-R Keese CR, Wegener J, Walker SR, Giaever I: Electrical wound healing assay for cells in vitro. Proc Natl Acad Sci U S A 2004;101: Kam Y, Guess C, Estrada L, Weidow B, Quaranta V: A novel circular invasion assay mimics in vivo invasive behavior of cancer cell lines and distinguishes single-cell motility in vitro. BMC Cancer 2008;8: Giuliano KA, Haskins JR, Taylor DL: Advances in high content screening for drug discovery. Assay Drug Dev Technol 2003;1: Yarrow JC, Feng Y, Perlman ZE, Kirchhausen T, Mitchison TJ: Phenotypic screening of small molecule libraries by high throughput cell imaging. Comb Chem High Throughput Screen 2003;4: Shor AC, Keschman EA, Lee FY, Muro-Cacho C, Letson GD, Trent JC, et al: Dasatinib inhibits migration and invasion in diverse human sarcoma cell lines and induces apoptosis in bone sarcoma cells dependent on Src kinase for survival. Cancer Res 2007;67: Lee JA, Cox K, Kriauciunas A, Chu S: The roles of high content cellular imaging. In SA Hanley H (ed): Lead Optimization in High Content Screening. New York: John Wiley, 2008: The Lilly Phenotypic Assay Initiative, PD2 [Online]. Retrieved from 9. Inglese J, Auld D, Sittamapalam S: Eli Lilly-NCGC Assay Guidance Manual [Online]. Retrieved from manual_toc.html 10. Simpson KJ, Selfors LM, Bui J, Reynolds A, Leake D, Khvorova A, et al: Identification of genes that regulate epithelial cell migration using an sirna screening approach. Nat Cell Biol 2008;10: Eastwood BJ, Farmen MW, Iversen PW, Craft TJ, Smallwood JK, Garbison KE, et al: The minimum significant ratio: a statistical parameter to characterize the reproducibility of potency estimates from concentrationresponse assays and estimation by replicate-experiment studies. J Biomol Screen 2006;11: Nizamutdinova IT, Kim YM, Chung JI, Shin SC, Jeong Y, Seo HG, et al: Anthocyanins from black soybean seed coats stimulate wound healing in Journal of Biomolecular Screening 16(2); 2011

9 Quantitative Cell Migration Assay fibroblasts and keratinocytes and prevent inflammation in endothelial cells. Food Chem Toxicol 2009;47: Wylie P, Gehler S, Hulkower K: A novel multiplexed-compatible high throughput cell migration screening assay using an Acumen ex3. Paper presented at the annual meeting of the Association of Laboratory Automation, Palm Springs, CA, January Nakayama N, Amano M, Katsumi A, Kaneko T, Kawabata S, Takefuji M, et al: Rho-kinase and myosin II are required for cell type and environment specific migration. Genes Cells 2005;10: Address correspondence to: Wendy Gough, M.S. Lilly Corporate Center Drop Code 0703, Indianapolis, IN gough_wendy@lilly.com Journal of Biomolecular Screening 16(2);

Measuring Wound Healing and Cell Migration using Celigo Imaging Cytometer

Measuring Wound Healing and Cell Migration using Celigo Imaging Cytometer Measuring Wound Healing and Cell Migration using Celigo Imaging Cytometer Nexcelom Bioscience LLC. 360 Merrimack Street, Building 9 Lawrence, MA 01843 T: 978.327.5340 F: 978.327.5341 E: info@nexcelom.com

More information

Introduction. Figure 1. Oris Cell Migration Assay Principle

Introduction. Figure 1. Oris Cell Migration Assay Principle Optimizing Performance of the Membrane-free, Oris Cell Migration Assay for High Throughput Screening using the BioTek Synergy HT Multi-Mode Microplate Reader Keren I. Hulkower, Renee L. Herber, and Scott

More information

Investigation of Cell Migration using a High Density Cell Exclusion Assay and Automated Microplate Imager

Investigation of Cell Migration using a High Density Cell Exclusion Assay and Automated Microplate Imager A p p l i c a t i o n N o t e Investigation of Cell Migration using a High Density Cell Exclusion Assay and Automated Microplate Imager Peter J. Brescia and Peter Banks, Applications Department, BioTek

More information

Introduction to Assay Development

Introduction to Assay Development Introduction to Assay Development A poorly designed assay can derail a drug discovery program before it gets off the ground. While traditional bench-top assays are suitable for basic research and target

More information

Oris TM Pro 384 Cell Migration Assay Tissue Culture Treated

Oris TM Pro 384 Cell Migration Assay Tissue Culture Treated Oris TM Pro 384 Cell Migration Assay Tissue Culture Treated Product No.: PRO384CMA5 384-well, 2-D Assay for Investigating Cell Migration of Adherent Cell Lines PROTOCOL & INSTRUCTIONS Table of Contents

More information

Oris TM Pro 384 Cell Migration Assay Collagen I Coated

Oris TM Pro 384 Cell Migration Assay Collagen I Coated Oris TM Pro 384 Cell Migration Assay Collagen I Coated Product No.: PRO384CMACC1 & PRO384CMACC5 384-well, 2-D Assay for Investigating Cell Migration of Adherent Cell Lines on Collagen I PROTOCOL & INSTRUCTIONS

More information

In Vitro Angiogenesis Assay Kit

In Vitro Angiogenesis Assay Kit In Vitro Angiogenesis Assay Kit Catalog Number KA1323 100 assays Version: 02 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Background... 3 Principle of the Assay...

More information

Membrane Potential Assays Using the ValiScreen Human Kv1.3 Voltage-Gated K + Channel Cell Line on the EnVision Multilabel Plate Reader

Membrane Potential Assays Using the ValiScreen Human Kv1.3 Voltage-Gated K + Channel Cell Line on the EnVision Multilabel Plate Reader TECHNICAL BRIEF ValiScreen Stable Recombinant Ion Channel Cell Line and the EnVision Multilabel Plate Reader Membrane Potential Assays Using the ValiScreen Human Kv1.3 Voltage-Gated K + Channel Cell Line

More information

Apoptosis And Anti-tumor Effect Induced By Mtor Inhibitor And Autophagy Inhibitor In Human Osteosarcoma Cells

Apoptosis And Anti-tumor Effect Induced By Mtor Inhibitor And Autophagy Inhibitor In Human Osteosarcoma Cells Apoptosis And Anti-tumor Effect Induced By Mtor Inhibitor And Autophagy Inhibitor In Human Osteosarcoma Cells Ryosuke Horie. Kagawa University of medecine, Kita-gun, Japan. Disclosures: R. Horie: None.

More information

Oris TM Pro Cell Migration Assay Collagen I Coated

Oris TM Pro Cell Migration Assay Collagen I Coated Oris TM Pro Cell Migration Assay Product No.: PROCMACC1 & PROCMACC5 96-well, 2-D Assay for Investigating Cell Migration of Adherent Cell Lines on Collagen I PROTOCOL & INSTRUCTIONS Table of Contents I.

More information

Oris TM Pro Cell Migration Assay Tissue Culture Treated

Oris TM Pro Cell Migration Assay Tissue Culture Treated Oris TM Pro Cell Migration Assay Tissue Culture Treated Product No.: PROCMA1 & PROCMA5 96-well, 2-D Assay for Investigating Cell Migration of Adherent Cell Lines PROTOCOL & INSTRUCTIONS Table of Contents

More information

An Automated, Cell-based Platform for the Rapid Detection of Novel Androgen Receptor Modulators

An Automated, Cell-based Platform for the Rapid Detection of Novel Androgen Receptor Modulators A p p l i c a t i o n N o t e An Automated, Cell-based Platform for the Rapid Detection of Novel Androgen Receptor Modulators Brad Larson and Peter Banks, BioTek Instruments, Inc., Winooski, VT Bruce Sherf,

More information

CytoSelect 24-Well Wound Healing Assay

CytoSelect 24-Well Wound Healing Assay Product Manual CytoSelect 24-Well Wound Healing Assay Catalog Number CBA-120 CBA-120-5 24 assays 5 x 24 assays FOR RESEARCH USE ONLY Not for use in diagnostic procedures Introduction Wounded tissue initiates

More information

EarlyTox Cell Integrity Kit

EarlyTox Cell Integrity Kit EarlyTox Cell Integrity Kit The EarlyTox Cell Integrity Kit from Molecular Devices is an optimized set of reagents that simplifies the measurement of live and dead cells in a single well. The assay uses

More information

For detecting mitochondrial membrane potential changes in cells using our proprietary fluorescence probe.

For detecting mitochondrial membrane potential changes in cells using our proprietary fluorescence probe. ab112133 JC-10 Mitochondrial Membrane Potential Assay Kit Flow Cytometry Instructions for Use For detecting mitochondrial membrane potential changes in cells using our proprietary fluorescence probe. This

More information

ab JC-10 Mitochondrial Membrane Potential Assay Kit Flow Cytometry

ab JC-10 Mitochondrial Membrane Potential Assay Kit Flow Cytometry ab112133 JC-10 Mitochondrial Membrane Potential Assay Kit Flow Instructions for Use For detecting mitochondrial membrane potential changes in cells using our proprietary fluorescence probe. This product

More information

ICSB Workshop: Drug Response Measurement and Analysis Part 2: Best practices for experimental design, execution, and analysis

ICSB Workshop: Drug Response Measurement and Analysis Part 2: Best practices for experimental design, execution, and analysis ICSB Workshop: Drug Response Measurement and Analysis Part 2: Best practices for experimental design, execution, and analysis Caitlin Mills Kartik Subramanian Drug-response experiments are becoming increasingly

More information

NEW INSIGHTS. NEW DISCOVERIES. Real-time automated measurements of cell health, movement and function inside your incubator.

NEW INSIGHTS. NEW DISCOVERIES. Real-time automated measurements of cell health, movement and function inside your incubator. THE NEXT GENERATION HAS ARRIVED IncuCyte S3 Live-Cell Analysis System Real-time automated measurements of cell health, movement and function inside your incubator. NEW INSIGHTS. NEW DISCOVERIES. See what

More information

Cell Cycle Phase Determination Kit

Cell Cycle Phase Determination Kit Cell Cycle Phase Determination Kit Catalog Number KA1301 100 assays Version: 04 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Background... 3 General

More information

Radius 24-Well Cell Migration Assay (Fibronectin Coated)

Radius 24-Well Cell Migration Assay (Fibronectin Coated) Product Manual Radius 24-Well Cell Migration Assay (Fibronectin Coated) Catalog Number CBA-125-FN 24 assays FOR RESEARCH USE ONLY Not for use in diagnostic procedures Introduction Cell migration is a highly

More information

Radius 24-Well Cell Migration Assay (Laminin Coated)

Radius 24-Well Cell Migration Assay (Laminin Coated) Product Manual Radius 24-Well Cell Migration Assay (Laminin Coated) Catalog Number CBA-125-LN 24 assays FOR RESEARCH USE ONLY Not for use in diagnostic procedures Introduction Cell migration is a highly

More information

Radius 384-Well Cell Migration Assay

Radius 384-Well Cell Migration Assay Product Manual Radius 384-Well Cell Migration Assay Catalog Number CBA-127 CBA-127-5 384 assays 5 x 384 assays FOR RESEARCH USE ONLY Not for use in diagnostic procedures Introduction Cell migration is

More information

FLIPR Calcium 6 Assay Explorer Kit (R8190) FLIPR Calcium 6-QF Assay Explorer Kit (R8192)

FLIPR Calcium 6 Assay Explorer Kit (R8190) FLIPR Calcium 6-QF Assay Explorer Kit (R8192) FLIPR Calcium 6 Assay Explorer Kit (R8190) FLIPR Calcium 6-QF Assay Explorer Kit (R8192) About the FLIPR Calcium 6 and 6-QF Assay Kits The FLIPR Calcium 6 Assay Kits from Molecular Devices, LLC (hereafter

More information

Brightfield and Fluorescence Imaging using 3D PrimeSurface Ultra-Low Attachment Microplates

Brightfield and Fluorescence Imaging using 3D PrimeSurface Ultra-Low Attachment Microplates A p p l i c a t i o n N o t e Brightfield and Fluorescence Imaging using 3D PrimeSurface Ultra-Low Attachment Microplates Brad Larson, BioTek Instruments, Inc., Winooski, VT USA Anju Dang, S-BIO, Hudson,

More information

Automated, Kinetic Imaging of Cell Migration and Invasion Assays using Corning FluoroBlok Permeable Supports

Automated, Kinetic Imaging of Cell Migration and Invasion Assays using Corning FluoroBlok Permeable Supports A p p l i c a t i o n N o t e Automated, Kinetic Imaging of Cell Migration and Invasion Assays using Corning FluoroBlok Permeable Supports Brad Larson, Principal Scientist, Applications Department, BioTek

More information

A Comparison of AlphaLISA and TR-FRET Homogeneous Immunoassays in Serum-Containing Samples

A Comparison of AlphaLISA and TR-FRET Homogeneous Immunoassays in Serum-Containing Samples application Note A Comparison of and Homogeneous Immunoassays in Serum-Containing Samples Authors Anuradha Prasad, PhD, Catherine Lautenschlager, PhD, Stephen Hurt, PhD, David Titus, PhD and Stéphane Parent,

More information

Cell Health and Viability Assays Real-time automated measurements of cell health and viability inside your incubator

Cell Health and Viability Assays Real-time automated measurements of cell health and viability inside your incubator INCUCYTE LIVE-CELL ANALYSIS SYSTEM Cell Health and Viability Assays Real-time automated measurements of cell health and viability inside your incubator See what your cells are doing and when they do it

More information

Z -LYTE KINASE ASSAY KIT TYR 3 PEPTIDE PROTOCOL

Z -LYTE KINASE ASSAY KIT TYR 3 PEPTIDE PROTOCOL Z -LYTE KINASE ASSAY KIT TYR 3 PEPTIDE PROTOCOL Cat. no. PV3192 O-062193-r1 US 0405 TABLE OF CONTENTS 1.0 INTRODUCTION... 2 2.0 ASSAY THEORY... 2 3.0 Z9-LYTE KINASE ASSAY KIT TYR 3 PEPTIDE MATERIALS PROVIDED...

More information

NEW INSIGHTS. NEW DISCOVERIES. Real-time automated measurements of cell health, movement and function inside your incubator.

NEW INSIGHTS. NEW DISCOVERIES. Real-time automated measurements of cell health, movement and function inside your incubator. THE NEXT GENERATION HAS ARRIVED IncuCyte S3 Live-Cell Analysis System Real-time automated measurements of cell health, movement and function inside your incubator. NEW INSIGHTS. NEW DISCOVERIES. See what

More information

For identifying inhibitors and activators of mitochondrial biogenesis in adherent cultured cells.

For identifying inhibitors and activators of mitochondrial biogenesis in adherent cultured cells. ab110216 MitoBiogenesis TM In-Cell ELISA Kit (IR) Instructions for Use For identifying inhibitors and activators of mitochondrial biogenesis in adherent cultured cells. This product is for research use

More information

System. Dynamic Monitoring of Receptor Tyrosine Kinase Activation in Living Cells. Application Note No. 4 / March

System. Dynamic Monitoring of Receptor Tyrosine Kinase Activation in Living Cells. Application Note No. 4 / March System Application Note No. 4 / March 2008 Dynamic Monitoring of Receptor Tyrosine Kinase Activation in Living Cells www.roche-applied-science.com Dynamic Monitoring of Receptor Tyrosine Kinase Activation

More information

Performance of cell viability and cytotoxicity assays on the IN Cell Analyzer 3000

Performance of cell viability and cytotoxicity assays on the IN Cell Analyzer 3000 GE Healthcare Application Note 28-4070-51 AA IN Cell Analyzer 3000 Performance of cell viability and cytotoxicity assays on the IN Cell Analyzer 3000 Key words: cell-based assay viability cytotoxicity

More information

CloneSelect Imager. Objective, quantitative assessment of cell growth. Genetix Now part of Molecular Devices.

CloneSelect Imager. Objective, quantitative assessment of cell growth. Genetix Now part of Molecular Devices. CloneSelect Imager Objective, quantitative assessment of cell growth www.moleculardevices.com / genetix Genetix Now part of Molecular Devices Assess cell confluence objectively and quantitatively Rapid

More information

High-throughput automation with the Attune NxT Autosampler: consistent results across all wells and across plates

High-throughput automation with the Attune NxT Autosampler: consistent results across all wells and across plates APPLICATION NOTE Attune NxT Flow Cytometer with Autosampler High-throughput automation with the Attune NxT Autosampler: consistent results across all wells and across plates Introduction The emerging field

More information

MitoBiogenesis In-Cell ELISA Kit (Colorimetric)

MitoBiogenesis In-Cell ELISA Kit (Colorimetric) PROTOCOL MitoBiogenesis In-Cell ELISA Kit (Colorimetric) DESCRIPTION 1850 Millrace Drive, Suite 3A Eugene, Oregon 97403 MS643 Rev.2 For identifying inhibitors and activators of mitochondrial biogenesis

More information

Development and Utilization of Bromodomain and Extra Terminal Domain (BET) Assays Using AlphaLISA Technology

Development and Utilization of Bromodomain and Extra Terminal Domain (BET) Assays Using AlphaLISA Technology APPLICATION NOTE AlphaLISA Technology Authors: Hong Cao Daniel Cardillo Jen Carlstrom Dawn Nida PerkinElmer, Inc. Hopkinton, MA Development and Utilization of Bromodomain and Extra Terminal Domain (BET)

More information

Z -LYTE fluorescent kinase assay technology. Z -LYTE Kinase Assay Platform

Z -LYTE fluorescent kinase assay technology. Z -LYTE Kinase Assay Platform Z -LYTE fluorescent kinase assay technology Z -LYTE Kinase Assay Platform Z -LYTE Kinase Assay Platform Non-radioactive assay format for screening a diverse collection of over 185 kinase targets Assay

More information

ab JC1 - Mitochondrial Membrane Potential Assay Kit

ab JC1 - Mitochondrial Membrane Potential Assay Kit ab113850 JC1 - Mitochondrial Membrane Potential Assay Kit Instructions for Use For the measurement of mitochondrial membrane potential by fluorescence plate reader This product is for research use only

More information

CytoSelect 24-Well Anoikis Assay

CytoSelect 24-Well Anoikis Assay Product Manual CytoSelect 24-Well Anoikis Assay Catalog Number CBA-080 24 assays FOR RESEARCH USE ONLY Not for use in diagnostic procedures Introduction Adhesion to the extracellular matrix (ECM) is essential

More information

Label-free, real-time live-cell assays for spheroids: IncuCyte bright-field analysis

Label-free, real-time live-cell assays for spheroids: IncuCyte bright-field analysis Introduction APPLICATION NOTE IncuCyte Live-Cell Analysis System Label-free, real-time live-cell assays for spheroids: IncuCyte bright-field analysis Susana L. Alcantara, Miniver Oliver, Kalpana Patel,

More information

CytoSelect 24- Well Cell Contraction Assay Kit (Floating Matrix Model)

CytoSelect 24- Well Cell Contraction Assay Kit (Floating Matrix Model) Product Manual CytoSelect 24- Well Cell Contraction Assay Kit (Floating Matrix Model) Catalog Number CBA- 5020 24 assays FOR RESEARCH USE ONLY Not for use in diagnostic procedures Introduction Wound healing

More information

High Throughput Quantitation of Cytokine Biomarkers using LANCE Ultra TR-FRET Assays

High Throughput Quantitation of Cytokine Biomarkers using LANCE Ultra TR-FRET Assays APPLIATION NOTE LANE TR-FRET Authors: Jen arlstrom Stephen Hurt Roger osse PerkinElmer, Inc. Hopkinton, MA High Throughput Quantitation of ytokine iomarkers using LANE Ultra TR-FRET Assays Introduction

More information

CytoSelect 24-Well Cell Haptotaxis Assay (8 µm, Collagen I-Coated, Fluorometric Format)

CytoSelect 24-Well Cell Haptotaxis Assay (8 µm, Collagen I-Coated, Fluorometric Format) Product Manual CytoSelect 24-Well Cell Haptotaxis Assay (8 µm, Collagen I-Coated, Fluorometric Format) Catalog Number CBA-101-COL 12 assays FOR RESEARCH USE ONLY Not for use in diagnostic procedures Introduction

More information

Platypus Technologies, LLC 5520 Nobel Drive, Suite 100 Madison, WI Toll Free: (866) Phone: (608) Fax: (608)

Platypus Technologies, LLC 5520 Nobel Drive, Suite 100 Madison, WI Toll Free: (866) Phone: (608) Fax: (608) Universal Cell Migration Assembly Kit Product No.: CMAU101 & CMAU505 96-well Assay for Investigating Cell Migration, Cell Invasion and 2-D Closure of Adherent Cell Lines Protocol & Instructions Patent

More information

Comparison of LANCE Ultra TR-FRET to PerkinElmer s Classical LANCE TR-FRET Platform for Kinase Applications

Comparison of LANCE Ultra TR-FRET to PerkinElmer s Classical LANCE TR-FRET Platform for Kinase Applications Comparison of LANCE Ultra TR-FRET to PerkinElmer s Classical LANCE TR-FRET Platform for Kinase Applications LANCE ULTRA TR-FRET TECHNOLOGY A P P L I C A T I O N N O T E Introduction Protein kinases play

More information

Lab Module 7: Cell Adhesion

Lab Module 7: Cell Adhesion Lab Module 7: Cell Adhesion Tissues are made of cells and materials secreted by cells that occupy the spaces between the individual cells. This material outside of cells is called the Extracellular Matrix

More information

Use of Phase Contrast Imaging to Track Morphological Cellular Changes due to Apoptotic Activity

Use of Phase Contrast Imaging to Track Morphological Cellular Changes due to Apoptotic Activity A p p l i c a t i o n N o t e Use of Phase Contrast Imaging to Track Morphological Cellular Changes due to Apoptotic Activity Brad Larson and Peter Banks, Applications Department, BioTek Instruments, Inc.,

More information

Z -LYTE KINASE ASSAY KIT SER/THR 9 PEPTIDE PROTOCOL

Z -LYTE KINASE ASSAY KIT SER/THR 9 PEPTIDE PROTOCOL Z -LYTE KINASE ASSAY KIT SER/THR 9 PEPTIDE PROTOCOL Part # PV3324 Lit. # 762-038412 020304 TABLE OF CONTENTS 1.0 INTRODUCTION... 2 2.0 ASSAY THEORY... 2 3.0 Z -LYTE KINASE ASSAY KIT SER/THR 9 PEPTIDE MATERIALS

More information

Characterization of anti-inflammatory substances from Cyanobacterias

Characterization of anti-inflammatory substances from Cyanobacterias Studying the international way Characterization of anti-inflammatory substances from Cyanobacterias Dipl.Biol. Maren Pflüger, Ph.D. 22.06.2016 Cyanobacteria European Territorial Co-operation Czech Academy

More information

Application Note. BD PureCoat ECM Mimetic Cultureware Collagen I Peptide: Novel Synthetic, Animal-free Surface for Culture of Human Keratinocytes

Application Note. BD PureCoat ECM Mimetic Cultureware Collagen I Peptide: Novel Synthetic, Animal-free Surface for Culture of Human Keratinocytes Page 1 BD PureCoat ECM Mimetic Cultureware Collagen I Peptide: Novel Synthetic, Animal-free Surface for Culture of Human Keratinocytes Kerry Thompson, Jeff Partridge, Elizabeth Abraham, Paula Flaherty,

More information

A Comparison of PerkinElmer s LANCE camp Assay Performance to that of State of the Art Competitive Technologies

A Comparison of PerkinElmer s LANCE camp Assay Performance to that of State of the Art Competitive Technologies A Comparison of PerkinElmer s LANCE camp Assay Performance to that of State of the Art Competitive Technologies GPCR ASSAYS A P P L I C A T I O N N O T E Introduction G-protein coupled receptors (GPCRs)

More information

FAR RED FP. GDP FP Assay. Technical Manual. v071814

FAR RED FP. GDP FP Assay. Technical Manual. v071814 FAR RED FP GDP FP Assay Technical Manual v071814 Transcreener GDP FP Assay Instructions for Part Numbers 3009-1K and 3009-10K 1.0 Introduction p.2 2.0 Assay Components p.3 3.0 Protocol p.4 Set up GDP Detection

More information

ab In vitro Angiogenesis Assay Kit

ab In vitro Angiogenesis Assay Kit ab204726 In vitro Angiogenesis Assay Kit Instructions for Use For simple qualitative detection of angiogenesis. This product is for research use only and is not intended for diagnostic use. Version 1 Last

More information

SensoLyte Homogeneous AFC Caspase-8 Assay Kit *Fluorimetric*

SensoLyte Homogeneous AFC Caspase-8 Assay Kit *Fluorimetric* SensoLyte Homogeneous AFC Caspase-8 Assay Kit *Fluorimetric* Catalog # 72088-100 Kit Size 100 Assays (96-well plate) Optimized Performance: This kit is optimized to detect caspase-8 activity. Enhanced

More information

The Product and Experiment Guide

The Product and Experiment Guide The Product and Experiment Guide Providing Solutions for Your Research p. 3 p. 10 p. 4 CELL CULTURE & MICROSCOPY IMMUNO- FLUORESCENCE LIVE CELL IMAGING p. 8 p. 7 p. 11 ANGIOGENESIS ASSAYS CHEMOTAXIS ASSAYS

More information

CFSE Cell Division Assay Kit

CFSE Cell Division Assay Kit CFSE Cell Division Assay Kit Catalog Number KA1302 100 assays Version: 02 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Background... 3 General Information...

More information

Application Note. Introduction. Robbie Narang, Zhaoping Liu, Kim Luu IntelliCyt Corporation

Application Note. Introduction. Robbie Narang, Zhaoping Liu, Kim Luu IntelliCyt Corporation pplication Note High Throughput Flow ssays using the MultiCyt Cell Proliferation Reagent Dye Panel: Multiplexing with Membrane Integrity, Cell Cycle, and Immunophenotyping Endpoints Robbie Narang, Zhaoping

More information

Modeling Cardiomyocyte Differentiation:

Modeling Cardiomyocyte Differentiation: icell Cardiac Progenitor Cells Prototype Application Protocol Wnt- and Activin/TGFβ-inhibitor Induction with Flow Cytometry Analysis Introduction The ability of cardiac progenitor cells to proliferate

More information

ISCT Telegraft Column: Mesenchymal Stromal Cell (MSC) Product Characterization and Potency Assay Development

ISCT Telegraft Column: Mesenchymal Stromal Cell (MSC) Product Characterization and Potency Assay Development ISCT Telegraft Column: Mesenchymal Stromal Cell (MSC) Product Characterization and Potency Assay Development University of Wisconsin-Madison, Production Assistance for Cellular Therapies (PACT) Over the

More information

Monitoring cell health with alamarblue and PrestoBlue reagents using the Varioskan LUX Multimode Microplate Reader

Monitoring cell health with alamarblue and PrestoBlue reagents using the Varioskan LUX Multimode Microplate Reader APPLICATION NOTE Varioskan LUX Multimode Microplate Reader Monitoring cell health with alamarlue and Prestolue s using the Varioskan LUX Multimode Microplate Reader Abstract In this application note, the

More information

xcelligence Immunotherapy Kit - B Cell Killing Assay ASSAY MANUAL

xcelligence Immunotherapy Kit - B Cell Killing Assay ASSAY MANUAL xcelligence Immunotherapy Kit - B Cell Killing Assay ASSAY MANUAL Table of Contents I. List of Kit Components 2 II. Storage Conditions 3 III. Additional Materials & Reagents Required 3 IV. Introduction

More information

Add Live, Dead and Total Dyes. Overnight Variable 30 min. 10 min 5 min. 30 min

Add Live, Dead and Total Dyes. Overnight Variable 30 min. 10 min 5 min. 30 min Cell Viability Analysis using Calcein AM, Propidium Iodide and Hoechst Application Description Celigo Application Plate Type Major Steps Cell viability analysis using Calcein AM, Propidium Iodide and Hoechst

More information

Data Sheet. Hippo Pathway TEAD Reporter MCF7 Cell Line Catalog #: 60618

Data Sheet. Hippo Pathway TEAD Reporter MCF7 Cell Line Catalog #: 60618 Data Sheet Hippo Pathway TEAD Reporter MCF7 Cell Line Catalog #: 6618 Background The Hippo pathway regulates cell proliferation and cell death. It is activated by high cell density and cell stress to stop

More information

Evaluating Microplate Detection Instruments

Evaluating Microplate Detection Instruments AppNote Evaluating Microplate Detection Instruments Introduction Important features to consider when evaluating a microplate reader are the following: sensitivity, dynamic range, throughput, integration

More information

Kinetic measurement of cytotoxicity using CellTox Green Cytotoxicity Assay on the IncuCyte TM FLR or ZOOM

Kinetic measurement of cytotoxicity using CellTox Green Cytotoxicity Assay on the IncuCyte TM FLR or ZOOM Kinetic measurement of cytotoxicity using CellTox Green Cytotoxicity Assay on the IncuCyte TM FLR or ZOOM The CellPlayer TM cytotoxicity assay described in this protocol utilizes CellTox Green Dye (Promega,

More information

LI-COR, Odyssey, Aerius, IRDye and In-Cell Western are registered trademarks or trademarks of LI-COR Biosciences Inc.

LI-COR, Odyssey, Aerius, IRDye and In-Cell Western are registered trademarks or trademarks of LI-COR Biosciences Inc. PROTOCOL PARP-1 (cleaved) In-Cell ELISA Kit (IR) 185 Millrace Drive, Suite 3A Eugene, Oregon 9743 MSA43 Rev. DESCRIPTION An In-Cell ELISA kit for measuring cleaved PARP-1 (89kDa fragment) in human adherent

More information

Assay Name: Antibody-Dependent Drug Uptake Assay

Assay Name: Antibody-Dependent Drug Uptake Assay Assay Name: Antibody-Dependent Drug Uptake Assay Assay ID: Celigo_02_0019 Table of Contents Experiment: Antibody-Dependent Drug Uptake Assay... 2 Celigo Setup...2 Assay Protocol and Plate Setup...3 Results...5

More information

CytoSelect 24-Well Cell Migration Assay (5 µm, Fluorometric Format)

CytoSelect 24-Well Cell Migration Assay (5 µm, Fluorometric Format) Revised Protocol Product Manual CytoSelect 24-Well Cell Migration Assay (5 µm, Fluorometric Format) Catalog Number CBA-102 12 assays FOR RESEARCH USE ONLY Not for use in diagnostic procedures Introduction

More information

JPMA 06, Dec., Case Studies of Established Conditions in Analytical Procedures. Biopharmaceutical Committee Technology Working Committee

JPMA 06, Dec., Case Studies of Established Conditions in Analytical Procedures. Biopharmaceutical Committee Technology Working Committee CMC Strategy Forum Japan 2016 Case Studies of Established Conditions in Analytical Procedures Biopharmaceutical Committee Technology Working Committee JPMA 06, Dec., 2016 Akira Nonoyama Santen Pharmaceutical

More information

ab Propidium Iodide Flow Cytometry Kit for Cell Cycle Analysis

ab Propidium Iodide Flow Cytometry Kit for Cell Cycle Analysis ab139418 Propidium Iodide Flow Cytometry Kit for Cell Cycle Analysis Instructions for Use To determine cell cycle status in tissue culture cell lines by measuring DNA content using a flow cytometer. This

More information

ab Propidium Iodide Flow Cytometry Kit for Cell Cycle Analysis

ab Propidium Iodide Flow Cytometry Kit for Cell Cycle Analysis ab139418 Propidium Iodide Flow Cytometry Kit for Cell Cycle Analysis Instructions for Use To determine cell cycle status in tissue culture cell lines by measuring DNA content using a flow cytometer. This

More information

Quantification of Cell Migration and Invasion Using the IncuCyte Chemotaxis Assay

Quantification of Cell Migration and Invasion Using the IncuCyte Chemotaxis Assay APPLICATION NOTE IncuCyte ZOOM Live-Cell Imaging System Quantification of Cell Migration and Invasion Using the IncuCyte Chemotaxis Assay Lindy O Clair, Meagan Roddy, Maria Tikhonenko, Clare Syzbut, Nicola

More information

Automated Optimization of Cell Transfection

Automated Optimization of Cell Transfection Automated Optimization of Cell Transfection Abstract Cell transfection is an essential technique for interrogating cellular pathways. Determining the optimal conditions for introducing nucleic acids into

More information

xcelligence Immunotherapy Kit - Liquid Tumor Killing Assay (anti-cd71) ASSAY MANUAL

xcelligence Immunotherapy Kit - Liquid Tumor Killing Assay (anti-cd71) ASSAY MANUAL xcelligence Immunotherapy Kit - Liquid Tumor Killing Assay (anti-cd71) ASSAY MANUAL Table of Contents I. List of Kit Components 2 II. Storage Conditions 3 III. Additional Materials & Reagents Required

More information

Recent years have witnessed an expansion in the disciplines encompassing drug

Recent years have witnessed an expansion in the disciplines encompassing drug Course overview Recent years have witnessed an expansion in the disciplines encompassing drug discovery outside the pharmaceutical industry. This is most notable with a significant number of Universities

More information

Morphology and ATP-based Cytotoxicity Profiling of Cancer Cell Lines using the EnSight Multimode Plate Reader with Well Imaging Technology

Morphology and ATP-based Cytotoxicity Profiling of Cancer Cell Lines using the EnSight Multimode Plate Reader with Well Imaging Technology APPLICATION NOTE Multimode Detection Authors: Philip Gribbon, PhD Fraunhofer Institute for Molecular Biology and Applied Ecology IME, Hamburg, Germany Morphology and ATP-based Cytotoxicity Profiling of

More information

In vitro Human Umbilical Vein Endothelial Cells (HUVEC) Tube-formation Assay. Josephine MY Ko and Maria Li Lung *

In vitro Human Umbilical Vein Endothelial Cells (HUVEC) Tube-formation Assay. Josephine MY Ko and Maria Li Lung * In vitro Human Umbilical Vein Endothelial Cells (HUVEC) Tube-formation Assay Josephine MY Ko and Maria Li Lung * Clinical Oncology Department, The Univerisity of Hong Kong, Hong Kong, Hong Kong SAR *For

More information

PolarScreen Estrogen Receptor-β Competitor Assay, Green. Table of Contents

PolarScreen Estrogen Receptor-β Competitor Assay, Green. Table of Contents PolarScreen Estrogen Receptor-β Competitor Assay, Green Catalog nos. P2615 and P2700 Storage: See page 2 Protocol part no. L0712 Revision date: 10 December 2009 Table of Contents Page 1. Introduction...

More information

WST-8 Cell Proliferation Assay Kit. Item No

WST-8 Cell Proliferation Assay Kit. Item No WST-8 Cell Proliferation Assay Kit Item No. 10010199 TABLE OF CONTENTS GENERAL INFORMATION 3 Materials Supplied 4 Precautions 4 If You Have Problems 4 Storage and Stability 4 Materials Needed but Not Supplied

More information

Autophagy/Cytotoxicity Dual Staining Kit

Autophagy/Cytotoxicity Dual Staining Kit Autophagy/Cytotoxicity Dual Staining Kit Catalog Number KA1299 96 assays Version: 05 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Background... 3 Principle of the Assay...

More information

CytoSelect 96-Well In Vitro Tumor Sensitivity Assay (Soft Agar Colony Formation)

CytoSelect 96-Well In Vitro Tumor Sensitivity Assay (Soft Agar Colony Formation) Product Manual CytoSelect 96-Well In Vitro Tumor Sensitivity Assay (Soft Agar Colony Formation) Catalog Number CBA-150 CBA-150-5 96 assays 5 x 96 assays FOR RESEARCH USE ONLY Not for use in diagnostic

More information

EdU Cell Proliferation Kit. User Manual

EdU Cell Proliferation Kit. User Manual EdU Cell Proliferation Kit User Manual Ordering information: (for detailed kit content see Table 2) EdU Click Kits: Product number EdU Used fluorescent dye BCK-EdU488 BCK-EdU555 BCK-EdU594 BCK-EdU647 5

More information

Quantification of Cell Migration and Invasion Using the IncuCyte Chemotaxis Assay

Quantification of Cell Migration and Invasion Using the IncuCyte Chemotaxis Assay Quantification of Cell Migration and Invasion Using the IncuCyte Chemotaxis Assay Lindy O Clair, Meagan Roddy, Maria Tikhonenko, Clare Syzbut, Nicola Bevan, Kirk Schroeder and Daniel M. Appledorn Essen

More information

CytoSelect 24-Well Cell Migration Assay (12 µm, Colorimetric Format)

CytoSelect 24-Well Cell Migration Assay (12 µm, Colorimetric Format) Product Manual CytoSelect 24-Well Cell Migration Assay (12 µm, Colorimetric Format) Catalog Number CBA-107 12 assays FOR RESEARCH USE ONLY Not for use in diagnostic procedures Introduction Cell migration

More information

MSD MULTI-SPOT Assay System

MSD MULTI-SPOT Assay System MSD MULTI-SPOT Assay System Human MMP 2-Plex Ultra-Sensitive Kit 1-Plate Kit 5-Plate Kit 25-Plate Kit K15033C-1 K15033C-2 K15033C-4 17335-v5-2012Mar 1 MSD Biomarker Assays Human MMP 2-Plex Ultra-Sensitive

More information

Cell Contraction Assay

Cell Contraction Assay Product Manual Cell Contraction Assay Catalog Number CBA-201 24 assays FOR RESEARCH USE ONLY Not for use in diagnostic procedures Introduction Wound healing comprises of three processes: epithelialization,

More information

Monitoring Cell Cycle Progression in Cancer Cells

Monitoring Cell Cycle Progression in Cancer Cells A p p l i c a t i o n N o t e Monitoring Cell Cycle Progression in Cancer Cells Using Nuclear Staining to Assess Cellular DNA Content Paul Held, Ph.D., Laboratory Manager, Applications Department, BioTek

More information

CytoSelect 48-Well Cell Adhesion Assay (ECM Array, Fluorometric Format)

CytoSelect 48-Well Cell Adhesion Assay (ECM Array, Fluorometric Format) Product Manual CytoSelect 48-Well Cell Adhesion Assay (ECM Array, Fluorometric Format) Catalog Number CBA-071 CBA-071-5 48 assays 5 x 48 assays FOR RESEARCH USE ONLY Not for use in diagnostic procedures

More information

OriGene GFC-Arrays for High-throughput Overexpression Screening of Human Gene Phenotypes

OriGene GFC-Arrays for High-throughput Overexpression Screening of Human Gene Phenotypes OriGene GFC-Arrays for High-throughput Overexpression Screening of Human Gene Phenotypes High-throughput Gene Function Validation Tool Introduction sirna screening libraries enable scientists to identify

More information

Focus Application. Cell Migration. Featured Study: Inhibition of Cell Migration by Gene Silencing. xcelligence System Real-Time Cell Analyzer

Focus Application. Cell Migration. Featured Study: Inhibition of Cell Migration by Gene Silencing. xcelligence System Real-Time Cell Analyzer xcelligence System Real-Time Cell Analyzer Focus Application Cell Migration Featured Study: Inhibition of Cell Migration by Gene Silencing Markus Greiner and Richard Zimmermann Department of Medical Biochemistry

More information

Rhod-4 No Wash Calcium Assay Kit

Rhod-4 No Wash Calcium Assay Kit ab112157 Rhod-4 No Wash Calcium Assay Kit Instructions for Use For detecting calcium in cells without washing by using our proprietary fluorescence probe This product is for research use only and is not

More information

Fast and Easy Calcium Flux Assays on the VICTOR Nivo Multimode Plate Reader

Fast and Easy Calcium Flux Assays on the VICTOR Nivo Multimode Plate Reader APPLICATION NOTE Multimode Detection Authors: Adriana-Maria Stanciu Maria Kuzikov Fraunhofer IME Screening Port Hamburg, Germany Fast and Easy Calcium Flux Assays on the VICTOR Nivo Multimode Plate Reader

More information

ATP Assay Kit. Catalog Number KA assays Version: 04. Intended for research use only.

ATP Assay Kit. Catalog Number KA assays Version: 04. Intended for research use only. ATP Assay Kit Catalog Number KA1661 100 assays Version: 04 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Principle of the Assay... 3 General Information...

More information

SensoLyte 490 MMP-9 Assay Kit *Fluorimetric*

SensoLyte 490 MMP-9 Assay Kit *Fluorimetric* SensoLyte 490 MMP-9 Assay Kit *Fluorimetric* Revision#1.2 Last updated: February 2017 Catalog # Kit Size AS-71134 500 assays (96-well) or 1250 assays (384-well) Convenient Format: All essential assay components

More information

The BioFlux 200 System Using Well Plate Microfluidics for Live Cell Assays Product Overview and Tutorial

The BioFlux 200 System Using Well Plate Microfluidics for Live Cell Assays Product Overview and Tutorial The BioFlux 200 System Using Well Plate Microfluidics for Live Cell Assays Product Overview and Tutorial Introduction to the BioFlux System Enables live-cell assays with precisely-controlled shear flow

More information

PARP-1 (cleaved) Human In-Cell ELISA Kit (IR)

PARP-1 (cleaved) Human In-Cell ELISA Kit (IR) ab110215 PARP-1 (cleaved) Human In-Cell ELISA Kit (IR) Instructions for Use For the quantitative measurement of Human PARP-1 (cleaved) concentrations in cultured adherent and suspension cells. This product

More information

Bafilomycins Produced by an Endophytic Actinomycete Streptomyces sp. YIM56209

Bafilomycins Produced by an Endophytic Actinomycete Streptomyces sp. YIM56209 Bafilomycins Produced by an Endophytic Actinomycete Streptomyces sp. YIM56209 Zhiguo Yu, 1,2 Li-Xing Zhao, 3 Cheng-Lin Jiang, 3 Yanwen Duan, 2 Lily Wong, 4 Kristopher C. Carver, 4 Linda A. Schuler, 4 and

More information

CytoSelect 96-Well Cell Migration Assay (8 µm, Fluorometric Format)

CytoSelect 96-Well Cell Migration Assay (8 µm, Fluorometric Format) Product Manual CytoSelect 96-Well Cell Migration Assay (8 µm, Fluorometric Format) Catalog Number CBA-106 96 assays FOR RESEARCH USE ONLY Not for use in diagnostic procedures Introduction Cell migration

More information

Application Note 493

Application Note 493 Corning PureCoat ECM Mimetic Cultureware Collagen I Peptide: Novel Synthetic, Animal-free Surface for Culture of Human Keratinocytes Kerry Thompson, Jeff Partridge, Elizabeth Abraham, Paula Flaherty, Susan

More information

In Vitro Angiogenesis Assay Kit. Item No

In Vitro Angiogenesis Assay Kit. Item No In Vitro Angiogenesis Assay Kit Item No. 10009964 TABLE OF CONTENTS GENERAL INFORMATION 3 Materials Supplied 3 Precautions 4 If You Have Problems 4 Storage and Stability 4 Materials Needed but Not Supplied

More information