1 High Five for venous blood collection for immunoassays Preanalytical errors are said to be the reason for up to 62 % of all errors in laboratory medicine [1]. Here are some tips on how to avoid the most frequent errors. 1. Carraro P et al. Errors in a stat laboratory: Types and frequencies 10 years later. Clin Chem 2007; 53,7:

2 This presentation tells you about the venous blood collection process for immunoassays Immunoassays are used to detect biomarkers that are generally only present in small amounts in the blood; therefore correct venous blood collection is pivotal for reliable, accurate test results. Radiometer provides immunoassays for cardiac markers, infection, hemostasis and pregnancy on the AQT90 FLEX analyzer platform. 2

3 Quality results begin with a quality specimen Agenda Why the preanalytical phase is important High Five for safe venous blood collection Blood collection tubes 3

4 The preanalytical phase of venous blood collection Preanalytical errors are said to be the reason for up to 62 % of all errors in laboratory medicine [1]. Error rate The errors that can occur during the collection and handling of blood specimens are potentially numerous, and the complications to the patient potentially harmful. Standards for venipuncture can reduce or alleviate many of these errors [2] Preanalytical phase 62 % CLSI Analytical phase 15 % Postanalytical phase 23 % CLSI: Vision: To be the leader in clinical and laboratory standards to improve the quality of medical care. [3] 1. Carraro P et al. Errors in a stat laboratory: Types and frequencies 10 years later. Clin Chem 2007; 53,7: CLSI. Procedures for the Collection of Diagnostic Blood Specimens by Venipuncture; Approved Standard Sixth Edition. CLSI document GP41-A6. Wayne, PA: Clinical and Laboratory Standards Institute;

5 High Five for safe venous blood sampling: Path of workflow: This workflow is based on CLSI international guidelines [1,2] 1. Patient preparation patient assessment and correct data registration to maximize patient safety 2. Blood collection device greater sample integrity and operator safety 3. Sample collection maximum safety for patient and operator 4. Sample handling key steps for greater sample integrity 5. Sample transport and storage minimum time to patient results 1. CLSI. Procedures for the Collection of Diagnostic Blood Specimens by Venipuncture; Approved Standard Sixth Edition. CLSI document GP41-A6. Wayne, PA: Clinical and Laboratory Standards Institute; CLSI. Tubes and Additives for Venous and Capillary Blood Specimen Collection; Approved Standard - Sixth Edition. CLSI document GP39-A6. Wayne, PA: Clinical and Laboratory Standards Institute;

6 1. Patient preparation A 6 6

7 1. Patient preparation Patient assessment and correct data registration to maximize patient safety Accurate patient identification is fundamental for patient safety Use at least 2 patient identifiers [1] Proper specimen labeling practices are critical components of effective and accurate patient identification [2] Proper sample labeling ensures the right result for the right patient Attach patient ID label to the sample tubes before you leave the patient Carefully select a sampling site with optimal access and blood flow The preferred venipuncture site is the area of either arm that is in front of and below the bend of the elbow with a number of superficial large veins [3] When the preferred veins are not acceptable or available, veins on the back of the hand are also acceptable [3] Tips! Establish a dedicated procedure for identifying patient and sample. 1. Joint Commission: National Patient Safety Goals Effective January 1, 2014 (Hospital Accreditation Program, Goal 1). 2. Kahn S. Specimen mislabeling: A significant and costly cause of potentially serious medical errors. Apr CLSI. Procedures for the Collection of Diagnostic Blood Specimens by Venipuncture; Approved Standard Sixth Edition. CLSI document GP41-A6. Wayne, PA: Clinical and Laboratory Standards Institute;

8 1. Patient preparation Errors can be caused by: Correct data registration to maximize patient safety Lack of patient identification and/or sample labeling Transcription errors due to manual data entry Lack of a dedicated procedure for identifying patient and samples Errors can lead to: Non-compliance Misdiagnosis Incorrect treatment Resampling Lost billing opportunities 8 8

9 2. Blood collection device 9 9

10 2. Blood collection device Greater sample integrity and operator safety Dedicated collection tube ensures greater sample integrity Select a self-filling vacuum tube prefilled with an anticoagulant If pre-evacuation by the end user is required, evacuate the tubes immediately before use Choose an anticoagulant that is compatible with the specific assay(s) to be used All blood collection equipment (tube, adaptor, needle) should come from the same manufacturer to ensure compatibility Anticoagulants and assay compatibility for the AQT90 FLEX analyzer EDTA TnI, TnT, CKMB, Myo, NT-proBNP, D-dimer, βhcg, CRP Lithium heparin TnI, TnT, CKMB, Myo, NT-proBNP, D-dimer, βhcg, CRP Citrate 3.2 % D-dimer Do NOT use tubes that contain a gel Do NOT use serum tubes 1. Ball J et al. Needlestick injury in Results from a survey of RCN members. Royal College of Nursing Sharps safety. RCN guidance to support implementation of the EU Directive 2010/32/EU on the prevention of sharps injuries in the health care sector. Royal College of Nursing

11 2. Blood collection device Greater sample integrity and patient safety with the AQT90 FLEX analyzer Do NOT use gel tubes Blood collection tube gels are thixotropic polymers. A thixotropic gel is semi-solid under static conditions it becomes fluid when stirred or shaken and returns to the semi-solid state upon standing. Risk of wrong results Whole-blood samples in gel tubes If a tube with a gel is placed on the AQT90 FLEX analyzer and mixed by the inlet wheel, we risk that part of the gel is mixed with the blood when the sample is aspirated Plasma samples from gel tubes If the aspirated sample contains gel droplets, then the actual volume of sample will be too low and the result will of course be affected Highly increased imprecision has been seen when plasma from gel tubes were used [1] 1. Unpublished data performed on the AQT90 FLEX

12 3. Sample collection 12 12

13 3. Sample collection Maximum safety for patient Be sure to follow local procedures and guidelines for sample collection Stop infusion A sample from a limb receiving infusion: Always stop infusion for a period of at least 2 minutes before you start to collect a sample [1] Using a tourniquet Tourniquet application should not exceed 1 minute [1] Indwelling lines sample collection procedure Removing sufficient flush solution prevents sample contamination. Note: Without complete, thorough, and documented training, phlebotomists should not collect blood from indwelling lines [1]. Mix with anticoagulant Immediately after the sample is collected, gently invert the tube according to recommendations to mix the anticoagulant with the blood. 1. CLSI. Procedures for the Collection of Diagnostic Blood Specimens by Venipuncture; Approved Standard Sixth Edition. CLSI document GP41-A6. Wayne, PA: Clinical and Laboratory Standards Institute;

14 3. Sample collection Maximum safety for patient and operator Sample contamination can be caused by: Operator safety can be compromised by: Diluting the sample with flush solution if an insufficient amount of flush solution has been removed This can lead to: Contaminating the sample with flush solution will alter the sample result so that it no longer represents the patient status Lack of dedicated procedure for operator safety This can lead to: Needle stick injury Operator concern over own safety Infection by blood-borne pathogens 14 14

15 4. Sample handling 15 15

16 4. Sample handling Key steps for greater sample integrity Proper mixing of the sample immediately after collection for a clot-free sample Mix to dissolve the anticoagulant to prevent clots to form Invert EDTA and lithium-heparin tubes at least 5-10 times unless otherwise specified by the tube manufacturer [1] Invert citrate tubes 3-4 times [1] Note: An inversion is one complete turn of the wrist, back and forth. 1. CLSI. Procedures for the Handling and Processing of Blood Specimens for Common Laboratory Tests; Approved Guideline Fourth Edition. CLSI document GP44-A4. Wayne, PA: Clinical and Laboratory Standards Institute;

17 4. Sample handling Key steps for greater sample integrity Microclots can be formed if the sample is not mixed properly immediately after its collection. Clots in the sample can lead to: Wrong results Analyzer downtime due to blocked sample fluid pathway 17 17

18 5. Sample transport and storage 18 18

19 5. Sample transport and storage Minimum time to patient results with the AQT90 FLEX analyzer Analyze sample immediately for optimal sample integrity and short TAT If samples cannot be analyzed immediately, separate plasma immediately after sample collection Store plasma at 2-8 C [1] Analyze plasma within 24 hours of collection time NOTE: Analyze TnI whole-blood samples within 2 hours after sample collection Analyze other whole-blood samples within 3 hours after sample collection The ability to run tests directly on whole-blood samples in capped tubes minimizes sample handling time and thereby the time to patient results. Tips! Note time of sample collection. Stay within the specified time range. 1. Sample stability claims validated by internal testing. 19

20 5. Sample transport and storage Minimum time to patient results Prolonged storage time can lead to: Prolonged storage time can alter the sample result so that it no longer matches the patient s disease state Incorrect results Incorrect storage temperature can lead to: Incorrect storage temperature can alter the sample result so that it no longer matches the patient s disease state Incorrect results 20 20

21 High Five for safe venous blood sampling: Path of workflow: 1. Patient preparation patient assessment and correct data registration to maximize patient safety 2. Blood collection device greater sample integrity and operator safety 3. Sample collection maximum safety for patient and operator 4. Sample handling key steps for greater sample integrity 5. Sample transport and storage minimum time to patient results 1. CLSI. Procedures for the Collection of Diagnostic Blood Specimens by Venipuncture; Approved Standard Sixth Edition. CLSI document GP41-A6. Wayne, PA: Clinical and Laboratory Standards Institute; CLSI. Tubes and Additives for Venous and Capillary Blood Specimen Collection; Approved Standard - Sixth Edition. CLSI document GP39-A6. Wayne, PA: Clinical and Laboratory Standards Institute;

22 Blood collection tubes for the AQT90 FLEX analyzer 22 22

23 Venous blood collection tubes Tube manufacturers providing tubes supported by the AQT90 FLEX analyzer Becton Dickinson (BD) Greiner International Terumo Sarstedt All the manufacturers have tubes using the American/ international color codes for caps. NOTE: In addition, Sarstedt also has tubes using the European color codes for caps. There are minor deviations from the international color codes for citrate tubes in the Japanese market. Sample tubes with one of three anticoagulants are used: EDTA Lithium heparin Citrate 3.2 % 23

24 Venous blood collection tubes Venous blood collection tubes supported by the AQT90 FLEX analyzer Tube manufacturers specifications to sample tubes Becton Dickinson Greiner Int. Terumo Sarstedt Tube size: 13 mm 75 mm Cap types: Conventional Hemoguard Anticoagulants: EDTA Li-heparin Citrate 3.2 % Draw volume: >1.8* ml Tube size: 13 mm 75 mm Cap types: Safety cap Anticoagulants: EDTA Li-heparin Citrate 3.2 % Draw volume: >2 ml Tube size: 13 mm 75 mm Cap type: Venosafe Anticoagulants: EDTA Li-heparin Citrate 3.2 % Draw volume: >1.8* ml Tube size: 13 mm 65 mm 11 mm 66 mm Cap type: Monovette Anticoagulants: EDTA Li-heparin Citrate 3.2 % Draw volume: >2.6 ml *1.8 ml for citrate tube containing 0.2 ml of citrate so that final volume is 2.0 ml 24

25 Venous blood collection tubes The AQT90 FLEX analyzer helps prevent the user from choosing the wrong tube for a specific test After installation of an AQT90 FLEX analyzer, the user sets up the analyzer to fit local preferences. NOTE: The tube type is defined as tube cap color and a name that lets users easily and correctly identify the tube Tube setup is a vital part of analysis setup. A tube type is defined for each type of sample tube approved for use on the analyzer, and used locally. For each tube type it is defined which tests can be done with the tube. A test cannot be run on a tube if the test is not defined for that tube type 25

Reducing Pre-analytical Errors

Reducing Pre-analytical Errors Reducing Pre-analytical Errors Christopher R. McCudden, Ph.D., FACB, FCACB, DABCC University of Ottawa The Ottawa Hospital Eastern Ontario Regional Laboratory Association Ontario, Canada Objectives List

More information

GENERAL SPECIMEN COLLECTION AND HANDLING HANDBOOK

GENERAL SPECIMEN COLLECTION AND HANDLING HANDBOOK GENERAL SPECIMEN COLLECTION AND HANDLING HANDBOOK Building 27, Ibn Sina Building, Block B, Ground floor, Dubai Healthcare City, U.A.E. ABOUT US After working 20 years in USA, Dow Healthcare Management

More information

Blood Specimen Collections

Blood Specimen Collections Laboratory testing and results play an essential role in caring for our patients. More than 70% of diagnoses, treatment regimens and monitoring assessments are based off of laboratory results. The validity

More information

Preanalytical Variables and the Impact on Quality Test Results. Learning Outcomes. Introduction. Sheryl Whitlock, MA,MT(ASCP)BB

Preanalytical Variables and the Impact on Quality Test Results. Learning Outcomes. Introduction. Sheryl Whitlock, MA,MT(ASCP)BB Preanalytical Variables and the Impact on Quality Test Results By Sheryl Whitlock, MA,MT(ASCP)BB Learning Outcomes Differentiate the phases of testing: preanalytical, analytical, postanalytical. List and

More information

Improving Quality and Turnaround Time of Clinical Chemistry Specimens Using Plasma

Improving Quality and Turnaround Time of Clinical Chemistry Specimens Using Plasma Improving Quality and Turnaround Time of Clinical Chemistry Specimens Using Plasma April 20, 2017 Sol Green PhD FACB Director of Medical Affair BD Life Science Preanalytical Systems Clinical Professor

More information

Auditing The Preanalytical Phase: Lessons from 40+ Audits

Auditing The Preanalytical Phase: Lessons from 40+ Audits Auditing The Preanalytical Phase: Lessons from 40+ Audits Stephen Church Oxford, UK Dr. Kathrin Schlüter BD Diagnostics, Preanalytical Systems Laboratory Analysis The Preanalytical Phase Order Test Collect

More information

Special Specimen Collection Procedures for Coagulation Testing

Special Specimen Collection Procedures for Coagulation Testing Special Specimen Collection Procedures for Coagulation Testing The accuracy of hemostasis testing depends upon the quality of the specimen submitted. Hemostasis specimens must be properly collected, labeled,

More information

Serum vs Plasma. Which specimen should you use? American Society for Clinical Laboratory Science Michigan Annual Meeting

Serum vs Plasma. Which specimen should you use? American Society for Clinical Laboratory Science Michigan Annual Meeting Serum vs Plasma Which specimen should you use? American Society for Clinical Laboratory Science Michigan Annual Meeting Sol Green PhD FACB March 31 2016 Introduction This presentation is intended to provide

More information

SAMPLE. Validation and Verification of Tubes for Venous and Capillary Blood Specimen Collection; Approved Guideline

SAMPLE. Validation and Verification of Tubes for Venous and Capillary Blood Specimen Collection; Approved Guideline December 2010 Validation and Verification of Tubes for Venous and Capillary Blood Specimen Collection; Approved Guideline This document provides guidance for conducting validation and verification testing

More information

Title of Manual: Specimen Collection Document Number: GPA.SPC.48.0

Title of Manual: Specimen Collection Document Number: GPA.SPC.48.0 Page 1 of 5 I. PURPOSE Proper specimen handling assures the results reported reflects the concentration in which the analyte exists in the patient as precisely as technologically possible. Once outside

More information

Best Practices in Sample Handling: Part 1 Responses to Participant Questions

Best Practices in Sample Handling: Part 1 Responses to Participant Questions Best Practices in Sample Handling: Part 1 Responses to Participant Questions 1. Can Sodium Fluoride tubes be stored in the freezer for a lactic acid test? This has not been tested by Greiner Bio-One. Please

More information

Pre-Analytic Issues in Laboratory Medicine

Pre-Analytic Issues in Laboratory Medicine Pre-Analytic Issues in Laboratory Medicine Daniel J. Fink, MD, MPH Director, Core Laboratory New York Presbyterian hospital Columbia University Medical Center October 3, 2006 Learning Objectives Pre-Analytic

More information

Introduction. Opinion Paper

Introduction. Opinion Paper Clin Chem Lab Med 2015; 53(3): 371 376 Opinion Paper Ana-Maria Simundic*, Michael P. Cornes, Kjell Grankvist, Giuseppe Lippi, Mads Nybo, Ferruccio Ceriotti, Elvar Theodorsson and Mauro Panteghini on behalf

More information

SAMPLE H21-A5. January 2008

SAMPLE H21-A5. January 2008 January 2008 Collection, Transport, and Processing of Blood Specimens for Testing Plasma-Based Coagulation Assays and Molecular Hemostasis Assays; Approved Guideline Fifth Edition This document provides

More information

Blood collection systems

Blood collection systems Blood collection systems As individual as your patients The complete solution for all requirements SARSTEDT International Table of contents Your partner in medicine and science worldwide SARSTEDT Group

More information

Discussion. Objectives:

Discussion. Objectives: EXERCISE 13: SPECIMEN PROCESSING Skills: 15 points Objectives: 1. Explain how a centrifuge works. 2. Name five types of centrifuges and tell what each is used for. 3. Safely and correctly demonstrate the

More information

i-stat iccnet CHSA Edition 1.2 Page 1 of 7 i-stat Rosy Tirimacco / Paul Simpson

i-stat iccnet CHSA Edition 1.2 Page 1 of 7 i-stat Rosy Tirimacco / Paul Simpson Page 1 of 7 i-stat Edition No: 1.2 Author: Rosy Tirimacco / Paul Simpson Date: January, 2012 Operative Date: January, 2012 Review Date: January, 2013 Location: Country Health SA Integrated Cardiovascular

More information

An Analysis of Iceberg of Errors in Clinical Laboratory

An Analysis of Iceberg of Errors in Clinical Laboratory Original article: An Analysis of Iceberg of Errors in Clinical Laboratory Dr.Nandini T Name of the institution: Dept. of Biochemistry, Bhagavan Mahaveer Jain Hospital, India Corresponding author : Dr Nandini

More information

PACKAGE INSERT. For use in the preparation and isolation of purified lymphocytes directly from whole blood. For In Vitro Diagnostic Use

PACKAGE INSERT. For use in the preparation and isolation of purified lymphocytes directly from whole blood. For In Vitro Diagnostic Use For use in the preparation and isolation of purified lymphocytes directly from whole blood PACKAGE INSERT For In Vitro Diagnostic Use PI-TT.610-UK-V6 Instructional Information Intended Use The T-Cell Xtend

More information

Ministry Saint Michael s Hospital Laboratory LABORATORY POLICY AND PROCEDURE

Ministry Saint Michael s Hospital Laboratory LABORATORY POLICY AND PROCEDURE Ministry Saint Michael s Hospital Laboratory LABORATORY POLICY AND PROCEDURE TITLE/SUBJECT: FILE NUMBER: ASSIGNED MANUAL: ORIGINATION DATE: 1/24/08 Blood Culture Collection MB013.05 Phlebotomy Manual and

More information

Pre-analytics in coagulation lab: why struggle for better results?

Pre-analytics in coagulation lab: why struggle for better results? Pre-analytics in coagulation lab: why struggle for better results? Valdas Banys, MD, PhD Vilnius University Faculty of Medicine Vilnius University Hospital Santariskiu Klinikos, Center of Laboratory Medicine

More information

Achieving the advantages of Plasma: How to prepare to convert to Plasma, the experience and perspective of a European hospital

Achieving the advantages of Plasma: How to prepare to convert to Plasma, the experience and perspective of a European hospital Preanalytical Education Series Achieving the advantages of Plasma: How to prepare to convert to Plasma, the experience and perspective of a European hospital Prof. Tim James Prof. Tim James, Laboratory

More information

Blood Collection Systems

Blood Collection Systems Blood Collection Systems As individual as your patients The complete solution for all requirements SARSTEDT International Your partner in medicine and science worldwide For over years 2 Table of contents

More information

Blood Collection Systems

Blood Collection Systems Blood Collection Systems As individual as your patients The complete solution for all requirements SARSTEDT International Table of contents Your partner in medicine and science worldwide SARSTEDT Group

More information

YOU Can Make a Difference!

YOU Can Make a Difference! YOU Can Make a Difference! Lucia M. Berte Laboratories Made Better! www.laboratoriesmadebetter.com 1 What is Quality? Quality is doing the right things and doing those things right. Philip Crosby 2 Do

More information

Preanalytical Variables in Blood Collection: Impact on Precision Medicine

Preanalytical Variables in Blood Collection: Impact on Precision Medicine Preanalytical Variables in Blood Collection: Impact on Precision Medicine Carolyn Compton, MD, PhD Chair, Scientific Advisory Committee, Indivumed GmbH Professor of Life Sciences, ASU Professor of Laboratory

More information

BIOSPECIMEN COLLECTION SOP Phenotype, Genotype, and Biomarkers (PGB) in ALS and Related Disorders

BIOSPECIMEN COLLECTION SOP Phenotype, Genotype, and Biomarkers (PGB) in ALS and Related Disorders The purpose of this Standard Operating Procedure (SOP) is to describe the instructions for collection, processing, and storage of collected specimens for the PGB Study. 1. Definitions 1. EDTA An EDTA tube

More information

Tests recently added to the NMS Labs test menu. New Tests are effective immediately.

Tests recently added to the NMS Labs test menu. New Tests are effective immediately. Immediate Action In our continuing effort to provide you with the highest quality toxicology laboratory services available, we have compiled important changes regarding a number of tests we perform. Listed

More information

SAMPLE. Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline Third Edition

SAMPLE. Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline Third Edition October 2010 Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline Third Edition This document contains guidelines for determining reference values and

More information

Platelet Function Testing Kits (catalogue numbers PSK-001, PSK-002, PSK-003)

Platelet Function Testing Kits (catalogue numbers PSK-001, PSK-002, PSK-003) Customer Information Sheet Platelet Function Testing Kits (catalogue numbers PSK-001, PSK-002, PSK-003) For remote testing of platelet function and the inhibitory effects of antiplatelet agents such as

More information

Human D-Dimer, D2D ELISA Kit

Human D-Dimer, D2D ELISA Kit Human D-Dimer, D2D ELISA Kit Catalog No: E0506h 96 Tests Operating instruction www.eiaab.cn FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS! PLEASE READ THROUGH ENTIRE PROCEDURE BEFORE

More information

Porcine Interleukin 1 Beta(IL-1β) ELISA Kit

Porcine Interleukin 1 Beta(IL-1β) ELISA Kit Porcine Interleukin 1 Beta(IL-1β) ELISA Kit Cat.No: DEIA-T6104 Lot. No. (See product label) Size 96T Intended Use The Porcine Interleukin 1 Beta(IL-1β) ELISA Kit is to be used for the vitro quantitative

More information

trucollect -plus - Stabilization and Transport Kit (10)

trucollect -plus - Stabilization and Transport Kit (10) trucollect -plus - Stabilization and Transport Kit (10) Whole blood specimen dry stabilization, transport, and storage For Research Use Only Not for use in diagnostic procedures PN 520254 Patents Granted

More information

Fibrinogen Universal Kit

Fibrinogen Universal Kit Order references Reagents CONT FITUR-B00 Universal kit 1 x 50 ml R1 + 1 x 5 ml R2 FITUR-00 Universal kit 2 x 50 ml R1 + 1 x 15 ml R2 FITUR-C00 Universal kit 2 x 17 ml R1 + 1 x 4 ml R2 Other necessary products

More information

WORTHAM LABORATORIES, INC. DRAFT Thrombin Reagent

WORTHAM LABORATORIES, INC. DRAFT Thrombin Reagent WORTHAM LABORATORIES, INC. DRAFT Thrombin Reagent Intended Use Wortham Laboratories Thrombin Reagent is intended for thrombin to convert fibrinogen in the quantitative determination of fibrinogen in plasma

More information

Date. Albumin measurements are used in the diagnosis and treatment of numerous diseases primarily involving the liver and/or kidneys.

Date. Albumin measurements are used in the diagnosis and treatment of numerous diseases primarily involving the liver and/or kidneys. SYNCHRON System(s) Chemistry Information Sheet Copyright 2013 Beckman Coulter, Inc. Albumin REF 442765 For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by Date Reviewed by Date PRINCIPLE INTENDED USE

More information

Please consult the Product Package Insert before running the test

Please consult the Product Package Insert before running the test Please consult the Product Package Insert before running the test Sample Collection, Preparation, Storage and Volume Q. What type of blood sample should be used with the PIFA Heparin/PF4 Rapid Assay? A.

More information

GENERAL INFORMATION ORDERS AND RESULTS

GENERAL INFORMATION ORDERS AND RESULTS GENERAL INFORMATION ORDERS AND RESULTS ORDERING LABORATORY TESTS: When ordering lab tests care should be taken to write orders legibly and with as much detail as possible. Aliases and abbreviations should

More information

HEMOCHRON. Whole Blood Coagulation Systems

HEMOCHRON. Whole Blood Coagulation Systems HEMOCHRON Whole Blood Coagulation Systems Activated Partial Thromboplastin Time (APTT) Cuvette Correlation Protocol for HEMOCHRON Microcoagulation Instruments MSIG: 49 10/06 Dear Medical Professional:

More information

Stability of Plasma for Add-On PT and APTT Tests

Stability of Plasma for Add-On PT and APTT Tests COAGULATION AND TRANSFUSION MEDICINE Original Article Stability of Plasma for Add-On PT and APTT Tests DEMETRA NEOFOTISTOS, MT, MARIA OROPEZA, MT(ASCP), AND CHUNG-HSIN TS'AO, PhD We conducted studies to

More information

Mouse helicobacter pylori antibody (IgG) ELISA Kit

Mouse helicobacter pylori antibody (IgG) ELISA Kit Mouse helicobacter pylori antibody (IgG) ELISA Kit Catalog Number. CSB-E14132m For the qualitative determination of mouse helicobacter pylori antibody (IgG) concentrations in serum, plasma. This package

More information

Human anti-endomysial (EMA) antibody (IgG) ELISA Kit

Human anti-endomysial (EMA) antibody (IgG) ELISA Kit Human anti-endomysial (EMA) antibody (IgG) ELISA Kit Catalog Number. CSB-EQ027572HU For the qualitative determination of human anti-endomysial (EMA) antibody (IgG) concentrations in serum, plasma. This

More information

Human heparin-platelet factor 4 complex antibody (IgG) ELISA Kit

Human heparin-platelet factor 4 complex antibody (IgG) ELISA Kit Human heparin-platelet factor 4 complex antibody (IgG) ELISA Kit Catalog Number. CSB-E09027h For the qualitative determination of human heparin-platelet factor 4 complex antibody(igg) concentrations in

More information

Applicable To Employees of the Gundersen Palmer Lutheran Hospitals and Clinics laboratories.

Applicable To Employees of the Gundersen Palmer Lutheran Hospitals and Clinics laboratories. Subject ABO Forward Grouping - D Antigen Typing - Palmer Index Number Lab-8791 Section Laboratory Subsection Regional/Affiliates Category Departmental Contact Betty Tilleraas Last Revised 10/31/2017 References

More information

IMPROVING THE PREANALYTICAL PROCESS: THE FOCUS ON SPECIMEN QUALITY

IMPROVING THE PREANALYTICAL PROCESS: THE FOCUS ON SPECIMEN QUALITY JMB 2008; 27 (3) DOI: 10.2478/v10011-008-0013-4 UDK 577.1 : 61 ISSN 1452-8258 JMB 27: 343 347, 2008 Review article Pregledni ~lanak IMPROVING THE PREANALYTICAL PROCESS: THE FOCUS ON SPECIMEN QUALITY POBOLJ[ANJE

More information

SAMPLE. Validation, Verification, and Quality Assurance of Automated Hematology Analyzers; Approved Standard Second Edition

SAMPLE. Validation, Verification, and Quality Assurance of Automated Hematology Analyzers; Approved Standard Second Edition June 2010 Validation, Verification, and Quality Assurance of Automated Hematology Analyzers; Approved Standard Second Edition This document provides guidance for the validation, verification, calibration,

More information

Human anti-gliadin antibody (IgG) ELISA Kit

Human anti-gliadin antibody (IgG) ELISA Kit Human anti-gliadin antibody (IgG) ELISA Kit Catalog Number. CSB-E09554h For the qualitative determination of human anti-gliadin antibody (IgG) concentrations in serum, plasma. This package insert must

More information

PLATELET-ORIENTED INHIBITION IN NEW TIA

PLATELET-ORIENTED INHIBITION IN NEW TIA PLATELET-ORIENTED INHIBITION IN NEW TIA AND MINOR ISCHEMIC STROKE ANCILLARY BIOMARKERS STUDY BLOOD SPECIMEN PROCEDURE MANUAL US Sites DNA Only DNA and Plasma CONTENTS I. DNA ONLY SPECIMEN PROCEDURES...

More information

Asian Journal of Research in Biological and Pharmaceutical Sciences Journal home page:

Asian Journal of Research in Biological and Pharmaceutical Sciences Journal home page: Research Article ISSN: 2349 4492 Asian Journal of Research in Biological and Pharmaceutical Sciences Journal home page: www.ajrbps.com IDENTIFICATION OF THE PRE-ANALYTICAL ERRORS AMONG GOVERNMENT HOSPITALS

More information

Pre-analytical errors in Clinical Chemistry Laboratory of a Tertiary care Hospital in Western Uttar Pradesh, India

Pre-analytical errors in Clinical Chemistry Laboratory of a Tertiary care Hospital in Western Uttar Pradesh, India IJBBS: v.2 n.1, p. 1-7. Dec. 2014 1 Pre-analytical errors in Clinical Chemistry Laboratory of a Tertiary care Hospital in Western Uttar Pradesh, India Sangeeta Kapoor 1, Anjali Verma 2*, Indu Saxena 3

More information

For the preparation of purified mononuclear immune cells directly from whole blood PACKAGE INSERT. For In Vitro Diagnostic Use PI-TS-IVD-UK-V3

For the preparation of purified mononuclear immune cells directly from whole blood PACKAGE INSERT. For In Vitro Diagnostic Use PI-TS-IVD-UK-V3 For the preparation of purified mononuclear immune cells directly from whole blood PACKAGE INSERT For In Vitro Diagnostic Use Intended Use The T-Cell Select kit is intended for the isolation of mononuclear

More information

NT-proBNP (Human) ELISA Kit

NT-proBNP (Human) ELISA Kit NT-proBNP (Human) ELISA Kit Catalog Number KA3099 96 assays Version: 21 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Background... 3 Principle of

More information

Medical Staff, House Staff, Patient Care Centers, Outpatient Clinics and UC Med Labs Clients

Medical Staff, House Staff, Patient Care Centers, Outpatient Clinics and UC Med Labs Clients TO: FROM: Medical Staff, House Staff, Patient Care Centers, Outpatient Clinics and UC Med Labs Clients Krzysztof Mikrut, B.S. MT (ASCP) Technical Director, Coagulation Laboratory Jonathan L. Miller, M.D.,

More information

SAMPLE. Risk Management Techniques to Identify and Control Laboratory Error Sources; Approved Guideline Second Edition

SAMPLE. Risk Management Techniques to Identify and Control Laboratory Error Sources; Approved Guideline Second Edition November 2009 Risk Management Techniques to Identify and Control Laboratory Error Sources; Approved Guideline Second Edition This guideline describes risk management techniques that will aid in identifying,

More information

SAMPLE. User Evaluation of Between-Reagent Lot Variation; Approved Guideline

SAMPLE. User Evaluation of Between-Reagent Lot Variation; Approved Guideline September 2013 User Evaluation of Between-Reagent Lot Variation; Approved Guideline This document provides guidance for laboratories on the evaluation of a new reagent lot, including a protocol using patient

More information

For the preparation of purified mononuclear immune cells directly from whole blood PACKAGE INSERT. For In Vitro Diagnostic Use PI-TS-IVD-UK-V4

For the preparation of purified mononuclear immune cells directly from whole blood PACKAGE INSERT. For In Vitro Diagnostic Use PI-TS-IVD-UK-V4 For the preparation of purified mononuclear immune cells directly from whole blood PACKAGE INSERT For In Vitro Diagnostic Use Intended Use The T-Cell Select kit is intended for the isolation of mononuclear

More information

Human Myostatin, ELISA Kit (MSTN)

Human Myostatin, ELISA Kit (MSTN) Human Myostatin, ELISA Kit (MSTN) 96 Tests Catalog Number: MBS733837 Store all reagents at 2-8 C Valid Period: six months For samples: Cell culture fluid & body fluid & tissue homogenate Serum or blood

More information

Human hepatitis E virus antibody (IgG) ELISA Kit

Human hepatitis E virus antibody (IgG) ELISA Kit Human hepatitis E virus antibody (IgG) ELISA Kit Catalog Number. NBP2-60048 For the qualitative determination of human hepatitis E virus antibody (IgG) concentrations in serum, plasma. This package insert

More information

Human Angiotensin 2 (Ang2) ELISA

Human Angiotensin 2 (Ang2) ELISA Human Angiotensin 2 (Ang2) ELISA For the quantitative determination of human Ang2 in serum, plasma, cell culture fluid and other biological fluids Cat. No. KT-52748 For Research Use Only. Not for use in

More information

Specimen Collection and Preparation

Specimen Collection and Preparation Specimen Collection and Preparation Laboratory test results are dependent on the quality of the specimen submitted. It is important that all specimens and request forms be properly labeled with the name

More information

Standard Operating Procedure

Standard Operating Procedure Subject Blood Culture Collection Index Number Lab-1325 Section Laboratory Subsection Specimen Collection and Processing Category Departmental Contact Regina Collins Last Revised 3/17/2017 References Required

More information

Clinical Biochemistry 3H03 Session 1

Clinical Biochemistry 3H03 Session 1 Clinical Biochemistry 3H03 Session 1 Dr J. Macri Associate Professor Department of Pathology and Molecular Medicine Clinical Biochemist Hamilton Regional Laboratory Medicine Program Your Most Important

More information

Fibrinogen Universal Kit

Fibrinogen Universal Kit Order references Reagents CONT FITUR-B00 Universal kit 1 x 50 ml R1 + 1 x 5 ml R2 FITUR-00 Universal kit 2 x 50 ml R1 + 1 x 15 ml R2 FITUR-C00 Universal kit 2 x 17 ml R1 + 1 x 4 ml R2 Other necessary products

More information

Rabbit C-reactive protein(crp) ELISA Kit

Rabbit C-reactive protein(crp) ELISA Kit Rabbit C-reactive protein(crp) ELISA Kit Catalog Number. CSB-E06847Rb For the quantitative determination of rabbit C-reactive protein(crp) concentrations in serum, plasma. This package insert must be read

More information

Human anti-toxoplasma (TOX) antibody (IgG) ELISA Kit

Human anti-toxoplasma (TOX) antibody (IgG) ELISA Kit Human anti-toxoplasma (TOX) antibody (IgG) ELISA Kit Catalog Number. CSB-E09532h For the qualitative determination of human anti-tox antibody (IgG) concentrations in serum, plasma. This package insert

More information

Hemoglobin HemoCue Hb method

Hemoglobin HemoCue Hb method TLM_PRO_027 and Version No. 001 Page 1 of 5 Department: Clinical Sciences Unit: Tropical Laboratory Medicine Hemoglobin HemoCue Hb 201 + method Project/study: Not applicable 1. Scope and application This

More information

Sysmex CS-2500 Hemostasis System Unrestricted Siemens Healthcare Diagnostics Inc All rights reserved.

Sysmex CS-2500 Hemostasis System Unrestricted Siemens Healthcare Diagnostics Inc All rights reserved. Excel with Confidence Sysmex CS-2500 Hemostasis System Unrestricted Siemens Healthcare Diagnostics Inc. 2015 All rights reserved. General Lab Challenges Test Volume Financial Performance Accuracy Reliability

More information

Human papillomavirus type 16 (HPV16) antibody (IgM) ELISA

Human papillomavirus type 16 (HPV16) antibody (IgM) ELISA Human papillomavirus type 16 (HPV16) antibody (IgM) ELISA For the qualitative determination of human papillomavirus type 16 (HPV16) antibody (IgM) in serum and plasma Cat. No. KU-442 For Research Use Only.

More information

Preanalytical note's for Blood coagulation tests

Preanalytical note's for Blood coagulation tests Preanalytical note's for Blood coagulation tests importance The use of modern laboratory instrumentation with high levels of test reliability and appropriate quality assurance measures will lead to very

More information

Human anti-hepatitis C virus (HCV) antibody ELISA Kit

Human anti-hepatitis C virus (HCV) antibody ELISA Kit Human anti-hepatitis C virus (HCV) antibody ELISA Kit Catalog Number. CSB-E09525h For the qualitative determination of human anti-hepatitis C virus (HCV) antibody concentrations in serum, plasma. This

More information

Human anti-cytomegalovirus(cmv) antibody (IgM) ELISA Kit

Human anti-cytomegalovirus(cmv) antibody (IgM) ELISA Kit Human anti-cytomegalovirus(cmv) antibody (IgM) ELISA Kit Catalog Number. CSB-E09546h For the qualitative determination of human anti-cytomegalovirus(cmv) antibody (IgM) concentrations in serum, plasma.

More information

Human Coxsackie virus (Cox V) antibody (IgG) ELISA Kit

Human Coxsackie virus (Cox V) antibody (IgG) ELISA Kit Human Coxsackie virus (Cox V) antibody (IgG) ELISA Kit Catalog Number. CSB-E04998h For the qualitative determination of human Coxsackie virus (Cox V) antibody (IgG) concentrations in serum, plasma. This

More information

For In Vitro Diagnostic Use

For In Vitro Diagnostic Use SYNCHRON System(s) Chemistry Information Sheet Lactate Dehydrogenase REF 442660 For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Date Reviewed by: Date PRINCIPLE INTENDED USE reagent, when used in

More information

ab FAP Human ELISA Kit

ab FAP Human ELISA Kit ab193701 FAP Human ELISA Kit Instructions for Use For the quantitative measurement of human Fibroblast Activation Protein, alpha (FAP) in serum, plasma, and cell culture supernatants. This product is for

More information

Department of Laboratories St. Louis, MO ISSUE DATE: June 2007 REVISION DATE: May 2012 REVIEWED DATE: May 2018 PRINCIPLE:

Department of Laboratories St. Louis, MO ISSUE DATE: June 2007 REVISION DATE: May 2012 REVIEWED DATE: May 2018 PRINCIPLE: Page 1 of 7 PROCEDURE: LEAD ESA LEADCARE II / HEALTHY KIDS EXPRESS ISSUE DATE: June 2007 REVISION DATE: May 2012 REVIEWED DATE: May 2018 PRINCIPLE: The LEADCARE II System relies on electrochemistry and

More information

Bovine prolactin/luteotropic hormone (PRL/LTH) ELISA Kit

Bovine prolactin/luteotropic hormone (PRL/LTH) ELISA Kit Bovine prolactin/luteotropic hormone (PRL/LTH) ELISA Kit Catalog Number. MBS703224 For the quantitative determination of bovine prolactin/luteotropic hormone (PRL/LTH) concentrations in serum, plasma.

More information

PAXgene Blood DNA PAXgene Blood DNA Tube (IVD) for clinical use PAXgene Blood DNA System (RUO) for research use. Explore more at

PAXgene Blood DNA PAXgene Blood DNA Tube (IVD) for clinical use PAXgene Blood DNA System (RUO) for research use. Explore more at PAXgene Blood DNA PAXgene Blood DNA Tube (IVD) for clinical use PAXgene Blood DNA System (RUO) for research use Explore more at www.preanalytix.com Situation The composition, amount, quality and integrity

More information

Human salmonella typhi (ST) antibody (IgM) ELISA kit

Human salmonella typhi (ST) antibody (IgM) ELISA kit Human salmonella typhi (ST) antibody (IgM) ELISA kit Catalog Number. CSB-E15070h For the qualitative determination of human salmonella typhi (ST) antibody (IgM) concentrations in serum, plasma. This package

More information

Mouse myeloperoxidase-antineutrophil cytoplasmic antibody IgG (MPO-ANCA IgG) ELISA Kit

Mouse myeloperoxidase-antineutrophil cytoplasmic antibody IgG (MPO-ANCA IgG) ELISA Kit Mouse myeloperoxidase-antineutrophil cytoplasmic antibody IgG (MPO-ANCA IgG) ELISA Kit Catalog No. CSB-E08676m (96 tests) This immunoassay kit allows for the in vitro semi-quantitative determination of

More information

Deer insulin-like growth factors 1 (IGF-1) ELISA Kit. This package insert must be read in its entirety before using this product.

Deer insulin-like growth factors 1 (IGF-1) ELISA Kit. This package insert must be read in its entirety before using this product. Deer insulin-like growth factors 1 (IGF-1) ELISA Kit Catalog Number. For the quantitative determination of deer insulin-like growth factors 1 (IGF-1) concentrations in serum, plasma, tissue homogenates.

More information

Rabbit immunoglobulin M(IgM) ELISA Kit

Rabbit immunoglobulin M(IgM) ELISA Kit Rabbit immunoglobulin M(IgM) ELISA Kit Catalog Number. CSB-E06950Rb For the quantitative determination of rabbit immunoglobulin M (IgM) concentrations in serum, plasma. This package insert must be read

More information

Human nuclear matrix protein 22 (NMP-22) ELISA Kit. MyBioSource.com

Human nuclear matrix protein 22 (NMP-22) ELISA Kit. MyBioSource.com Human nuclear matrix protein 22 (NMP-22) ELISA Kit Catalog Number. For the quantitative determination of human nuclear matrix protein 22 (NMP-22) concentrations in serum, plasma and urine. This package

More information

SUBJECT: INTRAVENOUS SOLUTIONS AND ADMIXTURE I.V.SOLUTIONS

SUBJECT: INTRAVENOUS SOLUTIONS AND ADMIXTURE I.V.SOLUTIONS COALINGA STATE HOSPITAL NURSING POLICY AND PROCEDURE MANUAL SECTION - Medications POLICY NUMBER: 519 Effective Date: August 31, 2006 SUBJECT: INTRAVENOUS SOLUTIONS AND ADMIXTURE I.V.SOLUTIONS 1. DEFINITION:

More information

SAFETY DATA SHEET. Radiometer MSDS M ; 942-/903/906/909/915/918/930/936/940; 944-/212/214/216/220/222/234/235/236/237/258..

SAFETY DATA SHEET. Radiometer MSDS M ; 942-/903/906/909/915/918/930/936/940; 944-/212/214/216/220/222/234/235/236/237/258.. 1/7 SECTION 1: IDENTIFICATION OF THE SUBSTANCE/MIXTURE AND OF THE COMPANY/UNDERTAKING 1.1. Product identifier Container size: 1.2. Relevant identified uses of the substance or mixture and uses advised

More information

Mouse immunoglobulin M(IgM) ELISA Kit

Mouse immunoglobulin M(IgM) ELISA Kit Mouse immunoglobulin M(IgM) ELISA Kit Catalog Number. CSB-E07977m For the quantitative determination of mouse immunoglobulin M (IgM) concentrations in serum, plasma. This package insert must be read in

More information

LabNotes. Blood Testing: A Newsletter from BD Vacutainer Systems, Preanalytical Solutions

LabNotes. Blood Testing: A Newsletter from BD Vacutainer Systems, Preanalytical Solutions LabNotes Volume 11 No.1, Summer 2002 A Newsletter from BD Vacutainer Systems, Preanalytical Solutions Blood Testing: Choosing the Right Specimen IN THIS ISSUE FEATURE: Blood Testing: Choosing the Right

More information

Human Selenoprotein P1, Plasma (SEPP1) ELISA

Human Selenoprotein P1, Plasma (SEPP1) ELISA Human Selenoprotein P1, Plasma (SEPP1) ELISA For the quantitative determination of human SEPP1 in serum, plasma, cell culture fluid and other biological fluids Cat. No. KT-50806 For Research Use Only.

More information

ab IL10 (Interleukin-10) Canine ELISA Kit

ab IL10 (Interleukin-10) Canine ELISA Kit ab193685 IL10 (Interleukin-10) Canine ELISA Kit Instructions for Use For the quantitative measurement canine IL10 in serum, plasma and cell culture supernatants. This product is for research use only and

More information

Human high sensitivity C-Reactive protein (hs-crp) ELISA Kit

Human high sensitivity C-Reactive protein (hs-crp) ELISA Kit Human high sensitivity C-Reactive protein (hs-crp) ELISA Kit Catalog No. CSB-E08617h (96T) This immunoassay kit allows for the in vitro quantitative determination of human hs-crp concentrations in serum,

More information

Mario Plebani University-Hospital of Padova, Italy

Mario Plebani University-Hospital of Padova, Italy Mario Plebani University-Hospital of Padova, Italy 70-80% of clinical decisions require laboratory medicine information (sources: UK Department of Health, Italian Health Minister)

More information

Human salmonella typhi antibody(st-ab) ELISA Kit

Human salmonella typhi antibody(st-ab) ELISA Kit Human salmonella typhi antibody(st-ab) ELISA Kit Catalog Number. CSB-E13576h For the qualitative determination of human salmonella typhi antibody (St-Ab) concentrations in serum, plasma. This package insert

More information

Human Collagen Type III (COL3) ELISA

Human Collagen Type III (COL3) ELISA Human Collagen Type III (COL3) ELISA For the quantitative determination of human COL3 in serum, plasma, cell culture fluid and other biological fluids Cat. No. KT-61018 For Research Use Only. Not for use

More information

cobas p 312 pre-analytical system Compact automation for maximum efficiency

cobas p 312 pre-analytical system Compact automation for maximum efficiency cobas p 312 pre-analytical system Compact automation for maximum efficiency Personalized Lab Automation Maximizing testing efficiency and medical value At Roche, laboratory automated solutions deliver

More information

Rat α-melanocyte stimulating hormone (α-msh) ELISA Kit

Rat α-melanocyte stimulating hormone (α-msh) ELISA Kit Rat α-melanocyte stimulating hormone (α-msh) ELISA Kit For the quantitative determination of rat α-melanocyte stimulating hormone (α-msh) concentrations in serum, plasma, tissue homogenates. This package

More information

Monkey apolipoprotein A1 (Apo-A1) ELISA Kit. For the quantitative determination of monkey apolipoprotein A1 (Apo-A1) concentrations in serum, plasma.

Monkey apolipoprotein A1 (Apo-A1) ELISA Kit. For the quantitative determination of monkey apolipoprotein A1 (Apo-A1) concentrations in serum, plasma. Monkey apolipoprotein A1 (Apo-A1) ELISA Kit Catalog Number. For the quantitative determination of monkey apolipoprotein A1 (Apo-A1) concentrations in serum, plasma. This package insert must be read in

More information

Sheep oxytocin (OT) ELISA Kit

Sheep oxytocin (OT) ELISA Kit Sheep oxytocin (OT) ELISA Kit Catalog Number. CSB-E13236Sh For the quantitative determination of sheep oxytocin (OT) concentrations in serum, plasma. This package insert must be read in its entirety before

More information

Mouse Coagulation Factor X (F10) ELISA

Mouse Coagulation Factor X (F10) ELISA Mouse Coagulation Factor X (F10) ELISA For the quantitative determination of mouse F10 in biological samples. Cat. No. KT-52133 For Research Use Only. Not for use in diagnostic procedures. Page 1 of 5

More information