Critical Quality Attribute Assessment by Peptide Mapping Using LC/MS with Superficially Porous Columns

Size: px
Start display at page:

Download "Critical Quality Attribute Assessment by Peptide Mapping Using LC/MS with Superficially Porous Columns"

Transcription

1 Critical Quality Attribute Assessment by Peptide Mapping Using LC/MS with Superficially Porous Columns

2 Analytical groups are tasked with method development, sample analysis, and method transfer DNA Vector Establishment of genetically engineered cells (e.g. CHO) that produce desired product Biomarker ID & validation Samples Human gene HT Clone mab Generation Platform Screening & Cell Culture Scale-up & Production e.g. transgenic mice Selection Samples Samples Samples QA/QC Methods Analytical Groups Develop analytical methods; analyze samples from other functions, methods transfer Centrifuge Samples Samples Market Manufacturing Clinical Pre-clinical Chromatography Column #3 Chromatography Column #2 Chromatography Column # 2

3 Critical Quality Attributes & Testing Methods Not just used for monoclonal antibodies same techniques are used for any potentially therapeutic protein RP Disulfide Shuffling SEC Aggregation Pyro- Glutamate IEC LC/MS Intact mass Fragment mass Identification PTM ID & locations Glycan ID Glycosylation (G, G, G2) HILIC Fragmentation (Hinge) Truncation (Lys,, 2) IEC RP Deamidation /Oxidation IEC 3

4 Peptide Mapping Workflow Solutions to Accelerate Biomolecule Characterization A(3-6) B(67-73) B(33-337) B(278-29) B ( ) Data Acquisition Peptide mapping chromatogram Peptide ID by MS/MS Sequence Matching Determine Post Translational Modifications Chromatographic Separation Reporting Sample Preparation Complete Agilent Workflow Solution Delivery of Results 4

5 Sample Preparation for Peptide Mapping A time consuming, labor-intensive process to do manually, with multiple steps that can be optimized. Sample Preparation Depletion of matrix proteins, enrichment of target protein, dialysis or desalting Reduce disulfide bonds, and block to prevent re-formation Denaturation, Reduction, & Alkylation Digestion Trypsin is the most common digestion enzyme, but others may be called for depending on the protein and experiment. Sample concentration, enrichment for specific PTMs, or desalting Enrichment & Cleanup 5

6 AssayMAP Bravo Sample Prep Automation Peptide mapping Detect 6

7 AssayMAP Sample Prep Portfolio Antibody Purification Immunoaffinity Purification Fractionation Protein Digestion Peptide Clean Up Protein Clean Up Phosphopeptide Enrichment N-glycan Sample Prep 7

8 General Considerations for Peptide Mapping Column Selection Instrument capabilities and requirements - UV vs MS detection or both - Pressure capabilities Mobile phase requirements - High ph required for sample - TFA vs Formic acid Sample - Hydrophilic vs hydrophobic peptides - Larger polypeptides present Column dimensions - Generally prefer longer columns, especially for more complicated maps - 5, 5, and 25 mm lengths available - Use 2. mm i.d. for MS-sensitivity - 3. and 4.6 mm i.d. also available - Smaller pore sizes ideal, usually -5 Å 8

9 Detectors for Peptide Mapping UV Often used in routine process measurements, QA/QC labs No peptide mass or structural information; relies on reproducibility of RT, peak area, and peak area ratios Traditionally uses TFA modifier MS Often used in discovery/early development stage, R&D labs MW and structural information (MS/MS) to assist with co-elution challenges, verify sequence coverage, and to ID unknown peaks when they appear Generally prefer formic acid modifier to avoid ion suppression from TFA 9

10 Mobile Phases for Peptide Mapping Ion pairing agents used for RP LC Increase retention of small, poorly retained peptides Block exposed silanol sites on silica-based columns Acid helps solubilize and ionize peptide and protein samples TFA works well on traditional silicabased columns, but formic acid does not Formic acid produces broad, asymmetric peaks Poor resolution and poor peak capacity Need column solution for formic acid modifier AdvanceBio Peptide Plus Charged surface chemistry solves the problems associated with formic acid

11 Agilent Bio-LC Column Portfolio Agilent Bio-LC Columns Affinity Reversed Phase HILIC Size Exclusion Ion Exchange Bio-Monolith Protein A AdvanceBio Peptide Plus AdvanceBio Glycan Mapping AdvanceBio SEC Bio mab Bio-Monolith Protein G Multiple Affinity Removal System AdvanceBio Peptide Mapping AdvanceBio RP mab AdvanceBio Oligonucleotides AdvanceBio Desalting-RP ZORBAX RRHD 3-HILIC Bio SEC-3 Bio SEC-5 ProSEC 3S ZORBAX GF-25 Bio IEX (SAX, SCX, WAX, WCX) PL SAX PL SCX Bio-Monolith (QA, DEAE, SO 3 ) PLRP-S ZORBAX GF-45 ZORBAX RRHD 3Å,.8 μm AdvanceBio Amino Acid Analysis ZORBAX Amino Acid Analysis ZORBAX 3SB Poroshell 3

12 Agilent Bio-LC Column Portfolio Agilent Bio-LC Reverse Phase Columns for Protein Analysis Intact Proteins (& Fragments) Peptides Amino Acids PLRP-S, 5 µm, Å AdvanceBio Peptide Plus AdvanceBio Amino Acid Analysis AdvanceBio RP mab AdvanceBio Peptide Mapping ZORBAX Amino Acid Analysis ZORBAX RRHD.8 µm, 3 Å ZORBAX Eclipse Plus C8 ZORBAX 3SB ZORBAX 3SB* Poroshell 3 AdvanceBio Desalting-RP *unusually large pore size for peptide mapping, but some people really like it nonetheless 2

13 AdvanceBio Peptide Plus Poroshell particle UHPLC performance at lower pressure Unique charge hybrid/c8 bonded phase Improved peak shape and peak capacity for peptide analysis with formic acid and MS detection To provide AdvanceBio Peptide Plus that gives benefits in accuracy and reproducibility of results for characterization of mab identity, PTMs 3

14 Columns Individually Tested for Efficiency AdvanceBio Peptide Plus Column 4

15 Column Lots QA Tested Specifically for Peptide Separations Peptides Standard (p/n G ) AdvanceBio Peptide Plus Column 5

16 LC/MS Instrumentation 29 Infinity II LC system 6545XT AdvanceBio LC/Q-TOF BioConfirm B.8 for intact protein and peptide mapping data analysis 6

17 6545XT Features for Peptide Mapping Sensitive peptide detection featuring Agilent Jet Stream Quick-start peptide mapping method Access low intensity peptides/ptms with the new IterativeMS/MS mode Sub-ppm mass accuracy with 5k resolution from improved beam optics Peptide fragmentation performance verification at install Ease of maintenance with vent-free capillary removal

18 Method Conditions Parameter 29 Infinity II LC Column AdvanceBio Peptide Plus 2. 5mm, 2.7µm, 2Å Mobile phase A:.%FA B:.% FA in ACN Time (min) %B 3 3 Gradient Post time min Inj vol µl Sample thermostat 5 C Post time 5 min Column temp 55 C Flow rate.5 ml/min Sample trastuzumab digest (ug/ul) All results shown are with formic acid mobile phase Parameter 6545XT AdvanceBio LC/Q-TOF Ion mode Positive ion mode, dual AJS ESI (profile) Drying gas temp 325 C Drying gas flow 3 L/min Sheath gas temp 275 C Sheath gas flow 2 L/min Nebulizer Capillary voltage Fragmentor voltage Skimmer voltage Oct RF Vpp Acquisition parameters 35 psi 4 V 75 V 65 V 75 V Data were acquired at 2 GHz Extended Dynamic Range MS mass range 7 m/z MS/MS mass range 5 7 MS scan rate (spectra/second): 8 MS/MS scan rate (spectra/second): 3 Ramped collision energy Charge state Slope Offset and > Data analysis BioConfirm software B.8.8 8

19 Effect of Flow Rate on Peak Capacity AdvanceBio Peptide Plus Column x ml/min Increasing the flow rate from. ml/min to.5 ml/min leads to an increase of twice the peak capacity x 8.2 ml/min.5 x 8.3 ml/min.5 x 8.4 ml/min.5 x ml/min.5 x ml/min.5 TIC Counts vs. Acquisition Time (min) ug trastuzumab digest 9

20 Effect of Gradient Length on Peak Capacity AdvanceBio Peptide Plus Column x x 8 min Increasing the gradient time from min to 6 min leads to an increase 2.5 times peak capacity, After 3min, no significant increase in peak capacity 2 min.5 x min.4 x min.4 x min.4 x min.4 TIC Counts vs. Acquisition Time (min) ug trastuzumab digest 2

21 Effect of Gradient Slope on Peak Capacity AdvanceBio Peptide Plus Column x8.5-4% in 3min Slope.3%/min Higher efficiency sharper peaks Increased sensitivity Peak capacity = 42 Sequence coverage = 96.69% x % in 6min Slope.6%/min Improved resolution Peak capacity = 59 Sequence coverage = 94.59%.5 TIC Acquisition Time (min) ug trastuzumab digest 2

22 Effect of Temperature on Peak Capacity AdvanceBio Peptide Plus Column x6 Higher temperature produces narrower peaks, improving resolution Change in temperature changes selectivity ~2% more peaks resolved by increasing the column temperature to 55 C 2 45 o C Pc = 54 x o C Pc = 6 LC/MS Chromatogram Co unts vs. Acquisition Time (min) ug trastuzumab digest 22

23 Back Pressure Characterization AdvanceBio Peptide Plus Column 2. x 5 mm 6 bar pressure maximum Temperature *pressure reading observed (max) Elevated temp produces narrower peak bands and lower pressure 23

24 Good Peak Shape & Retention Time Stability with High Mass Load AdvanceBio Peptide Plus Competitor Sample: 3-peptide mixture in.%fa Bradykinin Angiotensin II Venin substrate Competitor 2 Competitor 3 Key for transferring methods between LC/MS and LC/UV All 2. x 5 mm columns 24

25 Highly Reproducible Separations Co unts x Hydrophilic Hydrophobic AdvanceBio Peptide Plus Column Acquisition Time (min) n = %RSD Peak no. RT FWHM Area Height Consistent chromatography RSD values demonstrate the robustness of the method and precision of the LC system Especially crucial for LC/UV peptide mapping 25

26 Lifetime Testing Retention time % Column: AdvanceBio Peptide Plus Sample: peptide standard (59-583) Injection: 3 µl Flow rate:.5 ml/min Mobile phase: A.%FA B.% FA in ACN R t dropped < % Plates count % The column has subjected to injections and lost no more than % of the initial plate and retention value at the end of the test % Plate count Plates dropped < % Good chemical stability under formic acid conditions 26

27 Column Options for Alternate Selectivity x AdvanceBio Peptide Mapping Peptide Standard (p/n ) x AdvanceBio Peptide Plus FWHM Better C ounts v s. Acquisition Time (min).4.2 AdvBio Peptide Mapping AdvBio Peptide Plus TIC. Bradykinin frag Bradykinin 3. Angiotensin II (human) 4. Neurotensin 5. Angiotensin I (human) 6. Renin substrate porcine 7. [Ace-F-3,-2 H-] Angiotensinogen (-4) 8. Ser/Thr Protein Phosphatase (5-3) 9. [F4] Ser/Thr Protein Phosphatase (5-3). Melittin (honey bee venom) Peptide peak 27

28 mab Peptide Mapping x8 5 LC/MS of mab tryptic digest on 2. x 5mm AdvanceBio Peptide Plus Column Hydrophilic Hydrophobic 4 3 Counts Acquisition Time (min) Hydrophilic peptide retention Narrow Peaks with baseline resolution Hydrophobic peptide elution Reduced and fast analysis time Critical and desired peptide mapping components to achieve fast, selective, and highly efficient peptide separations across a wide dynamic range. 28

29 Tryptic Peptides of Trastuzumab 29

30 LC/MS Peptide Map AdvanceBio Peptide Plus Column sequence coverage 99.25% mass accuracy of 5 ppm 3

31 MassHunter BioConfirm for Peptide Mapping Sample Prep Separation Detection Data Processing & Report

32 Counts Separation of Oxidized Peptides x DLTMISR peptide Unmodified peptide x x DLTMISR Unmodified peptide b -H2O L,I y b y b y T R y 2-NH y y b 4-H2O Counts v s. Mass-to-Charge (m/z) DLTMISR Oxidized peptide AdvanceBio Peptide Plus Column MS/MS spectra y Oxidized peptide L,I y y b y b -H2O Acquisition Time (min) y -NH b -H2O EIC Relative % of Oxidation =.69% Counts vs. Mass-to-Charge (m/ z) Increase in mass (6Da) 32

33 Separation of Deamidated Peptides NTAYLQMNSLR Deamidated peptides NTAYLQMNSLR peptide Relative % of Deamidation =.2% 5.9% EIC AdvanceBio Peptide Plus Column ASQDVNTAVAWYQQKPGK Non-deamidated peptide Relative % deamidation = 9.% Deamidated peptide ASQDVNTAVAWYQQKPGK peptide Relative % of Deamidation = 9.% 33 non-deamidated peptide NTAYLQMNSLR NTAYLQMNSLR 5.9% NTAYLQMNSLR.23% NTAYLQMNSLR 5.5% NTAYLQMNSLR.97% NTAYLQMNSLR.24% NTAYLQMNSLR 2.5% NTAYLQMNSLR.2%

34 Critical Quality Attributes & Testing Methods Not just used for monoclonal antibodies same techniques are used for any potentially therapeutic protein RP Disulfide Shuffling SEC Aggregation Pyro- Glutamate IEC LC/MS Intact mass Fragment mass Identification PTM ID & locations Glycan ID Glycosylation (G, G, G2) HILIC Fragmentation (Hinge) RP Deamidation /Oxidation IEC With a Multi-Attribute Method, peptide mapping can monitor many of these CQAs simultaneously. Truncation (Lys,, 2) IEC 34

35 Multi-Attribute Method (MAM) 35

36 Separation of Deamidated Peptides VVSVLTVLHQDWLnGK or VVSVLTVLHqDWLNGK m/z , doubly charged peptide Traditional silica columns don t resolve deamidation as well as charged surface columns. Traditional Silica Column AdvanceBio Peptide Plus Determination of deamidated peptides is an important goal of peptide mapping. Chromatographic resolution is important. MS/MS confirms which peak corresponds to which deamidation site. AdvanceBio Peptide Plus enhances analysis because it has higher selectivity for deamidated forms of peptide over the normal form.

37 Choosing a Column Chemistry For LC/MS work, AdvanceBio Peptide Plus with formic acid is the first choice. Best peak shape with formic acid, high peak capacity, high mass load tolerance, excellent PTM resolution (especially deamidation), peak shape and recovery of hydrophobic peptides, unique selectivity Exception : Sample is especially rich in small, hydrophilic peptides. Poor retention of small, hydrophilic peptides is a feature common to charged surface columns. In this case, try AdvanceBio Peptide Mapping or Eclipse Plus C8 first. Exception 2: Different selectivity is needed to separate a key pair. Charged surface chemistry AdvanceBio Peptide Plus vs Regular silica chemistry AdvanceBio Peptide Mapping Eclipse Plus RRHD ZORBAX 3 37

38 Choosing a Column Chemistry Sample Detection First Choice of column Alternative column Peptides LC/MS (FA) AdvanceBio Peptide Plus AdvanceBio Peptide Mapping LC/UV AdvanceBio Peptide Mapping AdvanceBio Peptide Plus ZORABX RRHD 3SB-C8 or Diphenyl For maximum flexibility to transfer methods across instrument platforms, use AdvanceBio Peptide Plus with formic acid. 38

39 Method Development Tips & Reminders Bonded Phase: C8 is routinely used. Try AdvanceBio Peptide Plus first; for alternate selectivity or highly hydrophilic samples try AdvanceBio Peptide Mapping or a ZORBAX column. Column dimension: Good start with 2.mm 5mm, for higher resolution use longer column Gradient: Acetonitrile : Water with.% formic acid, other organic solvent substitutions can be used for different selectivity Starting gradient time: -4% in 3min (.3%/min) Temperature: Higher column temperature can dramatically improve both resolution and recovery (6 C max) Desalt peptide mixture prior to injection Sample handling may cause artifacts (peptide modifications) 39

40 Troubleshooting Reasons for < % Sequence Coverage Question to ask: Looking at the protein sequence, what s missing? Very small and/or very hydrophilic peptides? If using AdvanceBio Peptide Plus, try AdvanceBio Peptide Mapping for better retention and/or resolution. Very large and/or very hydrophobic peptides? Can you go to a higher percent organic mobile phase? Are peptides sticking to sample tubes or otherwise lost during sample preparation? Or on the LC? If using AdvanceBio Peptide Mapping, try AdvanceBio Peptide Plus - May even need to consider a C8 column rather than C8 Very large, or very small peptides of any hydrophobicity? Consider a different digestion enzyme 4

41 More Information How-to Guide Application Notes: LC/MS Analysis of Peptide Mapping with Formic Acid Ion-Pairing Agent ( EN) Enhancing the Quality of Peptide Mapping Separation for the Analysis of Post- Translational Modifications ( EN) 4

42 Standards & Reagents Peptide Standard (p/n ) Co unts x Blank A cquisition Time (min) 7 Trypsin Digest Methylated BSA Standard (p/n G99-85) 9 8. Bradykinin frag Bradykinin 3. Angiotensin II (human) 4. Neurotensin 5. Angiotensin I (human) 6. Renin substrate porcine 7. [Ace-F-3,-2 H-] Angiotensinogen (-4) 8. Ser/Thr Protein Phosphatase (5-3) 9. [F4] Ser/Thr Protein Phosphatase (5-3). Melittin (honey bee venom) HSA peptides Standard (p/n G ) Trypsin. AAFTE[CAM][CAM]QAADK 2. YLYEIAR 3. LVNEVTEFAK 4. KVPQVSTPTLVEVSR 5. RP[CAM]FSALEVDETYVPK 6. AVMDDFAAFVEK 7. HPYFYAPELLFFAK 42

43 Summary High efficiency, high resolution separation of peptides using AdvanceBio Peptide Plus column MS-compatible AdvanceBio Peptide Plus column delivers high peak capacity with sharp and narrow peaks using MS-friendly formic acid modifier mobile phase conditions. AdvanceBio Peptide Plus and AdvanceBio Peptide Mapping are superficially porous columns that yield low back pressures & STM-like performance, allowing peptide mapping separations to be run on both HPLC and UHPLC systems Well-resolved PTM modified peptide peaks enabled reliable mab peptide maps and development of Multi-Attribute Methods High quality separations in ~3 minutes with AdvanceBio Peptide Plus make peptide mapping more time-effective Multiple column options for every customer s unique needs 43

CQA Assessment by Peptide Mapping Using LC/MS with an AdvanceBio Peptide Plus Column

CQA Assessment by Peptide Mapping Using LC/MS with an AdvanceBio Peptide Plus Column CQA Assessment by Peptide Mapping Using LC/MS with an AdvanceBio Peptide Plus Column Anne Blackwell, Ph.D. BioColumns Product Support Scientist June 7, 27 ASMS Analytical groups are tasked with method

More information

Reproducible LC/MS Peptide Separation Using Agilent AdvanceBio Peptide Plus Columns

Reproducible LC/MS Peptide Separation Using Agilent AdvanceBio Peptide Plus Columns Reproducible LC/MS Peptide Separation Using Agilent AdvanceBio Peptide Plus Columns Application Note Biotherapeutics and Biosimilars Author Suresh Babu C.V. Agilent Technologies Inc. Introduction Peptide

More information

PLRP-S Polymeric Reversed-Phase Column for LC/MS Separation of mabs and ADC

PLRP-S Polymeric Reversed-Phase Column for LC/MS Separation of mabs and ADC PLRP-S Polymeric Reversed-Phase Column for LC/MS Separation of mabs and ADC Analysis of Intact and Fragmented mabs and ADC Application Note Biotherapeutics and Biologics Author Suresh Babu C.V. Agilent

More information

Development of Analysis Methods for Therapeutic Monoclonal Antibodies Using Innovative Superficially Porous Particle Biocolumns

Development of Analysis Methods for Therapeutic Monoclonal Antibodies Using Innovative Superficially Porous Particle Biocolumns Development of Analysis Methods for Therapeutic Monoclonal Antibodies Using Innovative Superficially Porous Particle Biocolumns Anne Blackwell Bio Columns Product Support Scientist Suresh Babu Senior Application

More information

BioHPLC columns. Tim Rice Biocolumn Technical Specialist

BioHPLC columns. Tim Rice Biocolumn Technical Specialist BioHPLC columns Tim Rice Biocolumn Technical Specialist AU Typical Application Areas Size Exclusion: Aggregation Analysis Ion Exchange: Charge Isoform Analysis 0.035 Monomer 0.030 0.025 0.020 0.015 Dimer

More information

4/4/2013. BioHPLC columns. Paul Dinsmoor Biocolumn Technical Specialist. April 23-25, Size Exclusion BioHPLC Columns

4/4/2013. BioHPLC columns. Paul Dinsmoor Biocolumn Technical Specialist. April 23-25, Size Exclusion BioHPLC Columns BioHPLC columns Paul Dinsmoor Biocolumn Technical Specialist April 23-25, 2013 Size Exclusion BioHPLC Columns 1 NEW Size Exclusion Columns 5 m Particle 100Å, 150Å, 300Å, 500Å, 1000Å, 2000Å pore sizes High

More information

Precise Characterization of Intact Monoclonal Antibodies by the Agilent 6545XT AdvanceBio LC/Q-TOF

Precise Characterization of Intact Monoclonal Antibodies by the Agilent 6545XT AdvanceBio LC/Q-TOF Precise Characterization of Intact Monoclonal Antibodies by the Agilent 6545XT AdvanceBio LC/Q-TOF Application Note Author David L. Wong Agilent Technologies, Inc. Santa Clara, CA, USA Introduction Monoclonal

More information

Put the PRO in Protein Characterization

Put the PRO in Protein Characterization Put the PRO in Protein Characterization Aaron Boice LC/Q-TOF Product Manager 5-June-2017 Agenda The Right Tools for the Job Tackling Intact Proteins Break it Down Peptides and PTMs Sweet Success Released

More information

Fast and Efficient Peptide Mapping of a Monoclonal Antibody (mab): UHPLC Performance with Superficially Porous Particles

Fast and Efficient Peptide Mapping of a Monoclonal Antibody (mab): UHPLC Performance with Superficially Porous Particles Fast and Efficient Peptide Mapping of a Monoclonal Antibody (mab): UHPLC Performance with Superficially Porous Particles Application Note Biotherapeutics and Biosimilars Authors James Martosella, Alex

More information

AdvanceBio Peptide Mapping

AdvanceBio Peptide Mapping AdvanceBio Peptide Mapping An HPLC Column Technology for Faster Protein Biocharacterizations Tim Rice BioColumn Technical Specialist 1 What Is Peptide Mapping? The chemical or enzymatic treatment of a

More information

Profiling Glycosylation of Monoclonal Antibodies at Three Levels Using the Agilent 6545XT AdvanceBio LC/Q TOF

Profiling Glycosylation of Monoclonal Antibodies at Three Levels Using the Agilent 6545XT AdvanceBio LC/Q TOF Application Note Biotherapeutics and Biosimilars Profiling Glycosylation of Monoclonal Antibodies at Three Levels Using the Agilent 64XT AdvanceBio LC/Q TOF Author David L. Wong Agilent Technologies, Inc.

More information

Quantification of Host Cell Protein Impurities Using the Agilent 1290 Infinity II LC Coupled with the 6495B Triple Quadrupole LC/MS System

Quantification of Host Cell Protein Impurities Using the Agilent 1290 Infinity II LC Coupled with the 6495B Triple Quadrupole LC/MS System Application Note Biotherapeutics Quantification of Host Cell Protein Impurities Using the Agilent 9 Infinity II LC Coupled with the 6495B Triple Quadrupole LC/MS System Authors Linfeng Wu and Yanan Yang

More information

Analysis of Monoclonal Antibody (mab) Using Agilent 1290 Infinity LC System Coupled to Agilent 6530 Accurate-Mass Quadrupole Time-of-Flight (Q-TOF)

Analysis of Monoclonal Antibody (mab) Using Agilent 1290 Infinity LC System Coupled to Agilent 6530 Accurate-Mass Quadrupole Time-of-Flight (Q-TOF) Analysis of Monoclonal Antibody (mab) Using Agilent 9 Infinity LC System Coupled to Agilent Accurate-Mass Quadrupole Time-of-Flight (Q-TOF) Application Note Authors Ravindra Gudihal and Suresh Babu CV

More information

Application of Agilent AdvanceBio Desalting-RP Cartridges for LC/MS Analysis of mabs A One- and Two-dimensional LC/MS Study

Application of Agilent AdvanceBio Desalting-RP Cartridges for LC/MS Analysis of mabs A One- and Two-dimensional LC/MS Study Application of Agilent AdvanceBio Desalting-RP Cartridges for LC/MS Analysis of mabs A One- and Two-dimensional LC/MS Study Application note Biotherapeutics and Biologics Authors Suresh Babu C.V., Anne

More information

ADVANCE ACCURACY AND PRODUCTIVITY FOR FASTER ANALYSIS

ADVANCE ACCURACY AND PRODUCTIVITY FOR FASTER ANALYSIS Agilent AdvanceBio Columns ADVANCE ACCURACY AND PRODUCTIVITY FOR FASTER ANALYSIS with Agilent ZORBAX RRHD 3Å 1.8 µm columns ns Rapid resolution high definition columns for UHPLC protein and peptide separations

More information

Peptide Mapping of Glycoprotein Erythropoietin by HILIC LC/MS and RP-LC/MS

Peptide Mapping of Glycoprotein Erythropoietin by HILIC LC/MS and RP-LC/MS Peptide Mapping of Glycoprotein Erythropoietin by HILIC LC/MS and RP-LC/MS Application Note BioPharma Authors Alex Zhu, James Martosella, and Phu T Duong Agilent Technologies, Inc Introduction Peptide

More information

Biotherapeutic Method Development Guide

Biotherapeutic Method Development Guide Biotherapeutic Method Development Guide HALO BIOCLASS 1000 Å PROTEIN SELECTIVITY KIT Fused-Core Particle Technology BIOCLASS Strategy for Optimizing Protein Separations Using Reversed-phase Liquid Chromatography

More information

Biotherapeutic Method Development Guide

Biotherapeutic Method Development Guide Biotherapeutic Method Development Guide HALO BIOCLASS 1000 Å PROTEIN SELECTIVITY KIT Fused-Core Particle Technology BIOCLASS Strategy for Optimizing Protein Separations Using Reversed-phase Liquid Chromatography

More information

Host Cell Protein Analysis Using Agilent AssayMAP Bravo and 6545XT AdvanceBio LC/Q-TOF

Host Cell Protein Analysis Using Agilent AssayMAP Bravo and 6545XT AdvanceBio LC/Q-TOF Application Note Biotherapeutics and Biosimilars Host Cell Protein Analysis Using Agilent AssayMAP Bravo and 6545XT AdvanceBio LC/Q-TOF Authors Linfeng Wu, Shuai Wu and Te-Wei Chu Agilent Technologies,

More information

Ensure your Success with Agilent s Biopharma Workflows

Ensure your Success with Agilent s Biopharma Workflows Ensure your Success with Agilent s Biopharma Workflows Steve Madden Software Product Manager Agilent Technologies, Inc. June 5, 2018 Agenda Agilent BioPharma Workflow Platform for LC/MS Intact Proteins

More information

ADVANCE ACCURACY AND PRODUCTIVITY FOR FASTER ANALYSIS

ADVANCE ACCURACY AND PRODUCTIVITY FOR FASTER ANALYSIS Agilent AdvanceBio Columns ADVANCE ACCURACY AND PRODUCTIVITY FOR FASTER ANALYSIS with Agilent ZORBAX RRHD 3Å 1.8 µm columns ns Rapid resolution high definition columns for UHPLC protein and peptide separations

More information

Fast and High-Resolution Reversed-Phase Separation of Synthetic Oligonucleotides

Fast and High-Resolution Reversed-Phase Separation of Synthetic Oligonucleotides Fast and High-Resolution Reversed-Phase Separation of Synthetic Oligonucleotides High-pH-stable, superficially porous particle columns for LC/UV and LC/MS Application Note Biologics and Biosimilars Authors

More information

Clone Selection Using the Agilent 1290 Infinity Online 2D-LC/MS Solution

Clone Selection Using the Agilent 1290 Infinity Online 2D-LC/MS Solution Clone Selection Using the Agilent 9 Infinity Online D-LC/MS Solution Application Note Biopharmaceuticals Authors Abstract Ravindra Gudihal and This Application Note describes the Agilent 9 Infinity online

More information

Separation of Native Monoclonal Antibodies and Identification of Charge Variants:

Separation of Native Monoclonal Antibodies and Identification of Charge Variants: Separation of Native Monoclonal Antibodies and Identification of Charge Variants: Teamwork of the Agilent 31 OFFGEL Fractionator, Agilent 21 Bioanalyzer and Agilent LC/MS Systems Application Note Biosimilar

More information

Agilent 1290 Infinity II 2D-LC Solution Biopharmaceutical Polymer Analysis. WCBP Jan 2017 Washington, DC

Agilent 1290 Infinity II 2D-LC Solution Biopharmaceutical Polymer Analysis. WCBP Jan 2017 Washington, DC Agilent 1290 Infinity II 2D-LC Solution Biopharmaceutical Polymer Analysis WCBP Jan 2017 Washington, DC 1 Overview Resolving power and how to measure it Why two-dimensional LC? Setup of a 2D-LC System

More information

Agilent Technologies April 20,

Agilent Technologies April 20, Agilent Technologies April 20, 2015 1 Improving HPLC Characterization of Biomolecules Agilent Solutions to Separation Challenges Paul Dinsmoor Technical Specialist, Bio- Columns Agilent Technologies April

More information

A Comprehensive Approach for Monoclonal Antibody N-linked Glycan Analysis from Sample Preparation to Data Analysis

A Comprehensive Approach for Monoclonal Antibody N-linked Glycan Analysis from Sample Preparation to Data Analysis Application ote Biotherapeutics and Biosimilars A Comprehensive Approach for Monoclonal Antibody -linked Glycan Analysis from Sample Preparation to Data Analysis Authors David L. Wong, scar Potter, Jordy

More information

Characterize mab Charge Variants by Cation-Exchange Chromatography

Characterize mab Charge Variants by Cation-Exchange Chromatography Characterize mab Charge Variants by Cation-Exchange Chromatography Application Note Biologics and Biosimilars Authors Isabel Vandenheede, Emmie Dumont, Pat Sandra, and Koen Sandra Research Institute for

More information

Proteins. Patrick Boyce Biopharmaceutical Marketing Manager Waters Corporation 1

Proteins. Patrick Boyce Biopharmaceutical Marketing Manager Waters Corporation 1 Routine Characterization of mabs and Other Proteins Patrick Boyce Biopharmaceutical Marketing Manager Europe and India 2011 Waters Corporation 1 Agenda Why? What scientific challenges? Technology Example

More information

Disulfide Linkage Analysis of IgG1 using an Agilent 1260 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column

Disulfide Linkage Analysis of IgG1 using an Agilent 1260 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column Disulfide Linkage Analysis of IgG1 using an Agilent 126 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column Application Note Biotherapeutics & Biosimilars Author M. Sundaram

More information

Application Note. Author. Abstract. Biotherapeutics and Biologics. Sonja Schneider Agilent Technologies, Inc. Waldbronn, Germany

Application Note. Author. Abstract. Biotherapeutics and Biologics. Sonja Schneider Agilent Technologies, Inc. Waldbronn, Germany D-LC/MS Characterization of Charge Variants Using Ion Exchange and Reversed-Phase Chromatography Multiple Heart-Cutting D-LC Analysis of Innovator versus Biosimilar Monoclonal Antibodies Application Note

More information

LC/MS Based Quantitation of Intact Proteins for Bioanalytical Applications

LC/MS Based Quantitation of Intact Proteins for Bioanalytical Applications LC/MS Based Quantitation of Intact Proteins for Bioanalytical Applications Alex Zhu, Ph.D. Agilent Technologies Wilmington, DE ASMS,2017-06 1 Outline Introduction on intact protein quantitation Agilent

More information

Characterize Fab and Fc Fragments by Cation-Exchange Chromatography

Characterize Fab and Fc Fragments by Cation-Exchange Chromatography Characterize Fab and Fc Fragments by Cation-Exchange Chromatography Application Note Biologics and Biosimilars Authors Isabel Vandenheede, Emmie Dumont, Pat Sandra, and Koen Sandra Research Institute for

More information

Peptide Mapping. Hardware and Column Optimization

Peptide Mapping. Hardware and Column Optimization Peptide Mapping Hardware and Column Optimization 1 Peptide Map: definition Analysis of PEPTIDES that are generated from the digestion or fragmentation of a protein or mixture of PROTEINS, by ELECTROPHORESIS;

More information

mabs and ADCs analysis by RP

mabs and ADCs analysis by RP mabs and ADCs analysis by RP Shanhua Lin, Ph.D. The world leader in serving science Protein and mab Separation by HPLC Size difference? YES Size Exclusion Chromatography (SEC) MAbPac SEC-1 NO NO Charge

More information

Peptide Mapping: A Quality by Design (QbD) Approach

Peptide Mapping: A Quality by Design (QbD) Approach Peptide Mapping: A Quality by Design (QbD) Approach Application Note Bio-Pharmaceutical Authors Sreelakshmy Menon and Suresh babu C.V. Agilent Technologies, Inc. Richard Verseput S-Matrix Corporation Abstract

More information

Size Exclusion BioHPLC columns Ion Exchange BioHPLC columns

Size Exclusion BioHPLC columns Ion Exchange BioHPLC columns Confidently separate and characterize bio-molecules with Agilent BioHPLC columns Size Exclusion BioHPLC columns Ion Exchange BioHPLC columns "It's a struggle to isolate and identify charge variants of

More information

Chromatographic Workflows for Biopharmaceutical Characterization

Chromatographic Workflows for Biopharmaceutical Characterization Chromatographic Workflows for Biopharmaceutical Characterization Dr. Ken Cook European Bioseparation Sales Support Expert Thermo Fisher Scientific, Hemel Hempstead/Germany The world leader in serving science

More information

Cell Clone Selection Using the Agilent Bio-Monolith Protein A Column and LC/MS

Cell Clone Selection Using the Agilent Bio-Monolith Protein A Column and LC/MS Cell Clone Selection Using the Agilent Bio-Monolith Protein A Column and LC/MS Application Note Biopharmaceuticals and Biosimilars Authors Emmie Dumont, Isabel Vandenheede, Pat Sandra, Koen Sandra Research

More information

Application Note. Authors. Abstract. Ravindra Gudihal Agilent Technologies India Pvt. Ltd. Bangalore, India

Application Note. Authors. Abstract. Ravindra Gudihal Agilent Technologies India Pvt. Ltd. Bangalore, India Quantitation of Oxidation Sites on a Monoclonal Antibody Using an Agilent 1260 Infinity HPLC-Chip/MS System Coupled to an Accurate-Mass 6520 Q-TOF LC/MS Application Note Authors Ravindra Gudihal Agilent

More information

Enabling routine characterization of proteins. Agilent MassHunter BioConfirm software

Enabling routine characterization of proteins. Agilent MassHunter BioConfirm software Enabling routine characterization of proteins Agilent MassHunter BioConfirm software Bioconfirm a key component of your major biopharma workflows Intact protein analysis BioConfirm will rapidly deconvolute

More information

Rapidly Characterize Antibody- Drug Conjugates and Derive Drug-to-Antibody Ratios Using LC/MS

Rapidly Characterize Antibody- Drug Conjugates and Derive Drug-to-Antibody Ratios Using LC/MS Rapidly Characterize Antibody- Drug Conjugates and Derive Drug-to-Antibody Ratios Using LC/MS Ning Tang, Ph.D. Agilent Technologies Santa Clara, California, USA Heterogeneity in ADCs Complexity T-DM1 (Genentech)

More information

Application Note. Authors. Abstract. Introduction. Pharmaceutical

Application Note. Authors. Abstract. Introduction. Pharmaceutical Scalability of Agilent s Across HPLC and UHPLC Instruments Application Note Pharmaceutical Authors Anne E. Mack, William J. Long Agilent Technologies, Inc. 2850 Centerville Road, Wilmington, DE 19808 USA

More information

Charged Surface Hybrid C18 for High Resolution LC and LC/MS Peptide Separations

Charged Surface Hybrid C18 for High Resolution LC and LC/MS Peptide Separations Charged Surface Hybrid C18 for High Resolution LC and LC/MS Peptide Separations Higher Quality Peptide Separations Using Hybrid Particle-Based Reversed-Phase Columns and CSH Technology 2013 Waters Corporation

More information

Strategies for the Separation and Characterization of Protein Biopharmaceuticals

Strategies for the Separation and Characterization of Protein Biopharmaceuticals Strategies for the Separation and Characterization of Protein Biopharmaceuticals Koen Sandra Webinar in association with SelectScience and Agilent Technologies January 28, 2015 Protein biopharmaceuticals

More information

Agilent Prep LC Columns for Small Molecules and Biomolecules MAINTAIN RAPID, RELIABLE SEPARATIONS AS YOU SCALE-UP

Agilent Prep LC Columns for Small Molecules and Biomolecules MAINTAIN RAPID, RELIABLE SEPARATIONS AS YOU SCALE-UP Agilent Prep LC Columns for Small Molecules and Biomolecules MAINTAIN RAPID, RELIABLE SEPARATIONS AS YOU SCALE-UP AGILENT PREP COLUMNS FOR HPLC FLEXIBLE, COST-EFFECTIVE OPTIONS FOR SCALING AND PREPARATIVE

More information

Accelerate mab Characterization Using Automated Sample Prep

Accelerate mab Characterization Using Automated Sample Prep Accelerate mab Characterization Using Automated Sample Prep David Knorr, Ph.D. Automation Solutions Ning Tang, Ph.D. LC/MS 15 February 2012 Page 1 Protein Sample Processing Workflows Glycan Profiling Biological

More information

Columns for Biomolecules BioLC Column Lines

Columns for Biomolecules BioLC Column Lines Columns for Biomolecules BioLC Column Lines Monoclonal Antibodies Glycans GlycanPac Nucleic Acids DNAPac Protein A Accucore Amide-HILIC DNAPac PA1 SEC-1 GlycanPac AXH-1 DNAPac PA2 SCX-1 GlycanPac AXR-1

More information

N-Glycan Profiling Analysis of a Monoclonal Antibody Using UHPLC/FLD/Q-TOF

N-Glycan Profiling Analysis of a Monoclonal Antibody Using UHPLC/FLD/Q-TOF N-Glycan Profiling Analysis of a Monoclonal Antibody Using UHPLC/FLD/Q-TOF Application Note Authors Xianming Liu, Wei Zhang, Yi Du, Sheng Yin, Hong Que, and Weichang Zhou WuXi AppTec iopharmaceuticals

More information

Maximizing Chromatographic Resolution of Peptide Maps using UPLC with Tandem Columns

Maximizing Chromatographic Resolution of Peptide Maps using UPLC with Tandem Columns Maximizing Chromatographic Resolution of Peptide Maps using UPLC with Tandem Columns Hongwei Xie, Martin Gilar, and Jeff Mazzeo Waters Corporation, Milford, MA U.S. APPLICATION BENEFITS The ACQUITY UPLC

More information

Agilent BioHPLC Columns PROTEIN IDENTIFICATION AND IMPURITY PROFILING USING REVERSED-PHASE HPLC/UHPLC

Agilent BioHPLC Columns PROTEIN IDENTIFICATION AND IMPURITY PROFILING USING REVERSED-PHASE HPLC/UHPLC Agilent BioHPLC Columns PROTEIN IDENTIFICATION AND IMPURITY PROFILING USING REVERSED-PHASE HPLC/UHPLC REVERSED-PHASE HPLC/UHPLC Agilent can boost your accuracy and productivity Reversed-phase is used to

More information

Reverse-Phase Separation of Proteins, Peptide and Other Biomolecules. Bill Long Applications Engineer October 8, 2008

Reverse-Phase Separation of Proteins, Peptide and Other Biomolecules. Bill Long Applications Engineer October 8, 2008 Reverse-Phase Separation of Proteins, Peptide and Other Biomolecules Bill Long Applications Engineer October 8, 28 Where is Reversed Phase HPLC Useful in Bio Separations? Proteomics New Bio Pharmaceutical

More information

Analysis of Nucleosides Using an Agilent Infinity II High Speed UHPLC with the 6130 Single Quadrupole Mass Selective Detector

Analysis of Nucleosides Using an Agilent Infinity II High Speed UHPLC with the 6130 Single Quadrupole Mass Selective Detector Analysis of Nucleosides Using an Agilent Infinity II High Speed UHPLC with the 6130 Single Quadrupole Mass Selective Detector Application Note Clinical Research Author Patrick Cronan Agilent Technologies,

More information

Highly Confident Peptide Mapping of Protein Digests Using Agilent LC/Q TOFs

Highly Confident Peptide Mapping of Protein Digests Using Agilent LC/Q TOFs Technical Overview Highly Confident Peptide Mapping of Protein Digests Using Agilent LC/Q TOFs Authors Stephen Madden, Crystal Cody, and Jungkap Park Agilent Technologies, Inc. Santa Clara, California,

More information

Bravo with AssayMAP. Automated Sample Preparation

Bravo with AssayMAP. Automated Sample Preparation Bravo with AssayMAP Automated Sample Preparation Why Automate? At some point, you cannot do it well without automation! Maintain data quality with increased sample capacity. Consistency scales better with

More information

Characterization of Glycosylation in the Fc Region of Therapeutic Recombinant Monoclonal Antibody

Characterization of Glycosylation in the Fc Region of Therapeutic Recombinant Monoclonal Antibody Characterization of Glycosylation in the Fc Region of Therapeutic Recombinant Monoclonal Antibody Application Note BioPharma Authors James Martosella, Phu Duong, and Alex Zhu Agilent Technologies, Inc.

More information

MAbPac RP Column. High-performance reverse phase chromatography column for monoclonal antibody analysis

MAbPac RP Column. High-performance reverse phase chromatography column for monoclonal antibody analysis CHROMATOGRAPHY MAbPac RP Column High-performance reverse phase chromatography column for monoclonal antibody analysis Product Specifications The Thermo Scientific MAbPac RP is a reverse phase (RP) liquid

More information

Solutions for the Core and Protein Laboratory: MassHunter Walkup

Solutions for the Core and Protein Laboratory: MassHunter Walkup Solutions for the Core and Protein Laboratory: MassHunter Walkup Jade C. Byrd MS Software Product Manager Page 1 Common Challenges in Characterizing Biomolecules Molecular size (insulin is 5 kda, mabs

More information

Application Note. Author. Introduction. Ravindra Gudihal Agilent Technologies India Pvt Ltd

Application Note. Author. Introduction. Ravindra Gudihal Agilent Technologies India Pvt Ltd Peptide Mapping of Innovator and iosimilar Monoclonal Antibod Using an Agilent 9 Infinit UHPLC Coupled to an Agilent 655 ifunnel Q-TOF LC/MS Sstem Application Note Author Ravindra Gudihal Agilent Technologies

More information

Fast and High Resolution Analysis of Intact and Reduced Therapeutic Monoclonal Antibodies (mabs)

Fast and High Resolution Analysis of Intact and Reduced Therapeutic Monoclonal Antibodies (mabs) Fast and High Resolution nalysis of Intact and Reduced Therapeutic Monoclonal ntibodies (mbs) The gilent io-inert L and dvanceio RP-mb olumns pplication Note io-pharmaceutical uthor M. Sundaram Palaniswamy

More information

NISTmAb characterization with a high-performance RP chromatography column

NISTmAb characterization with a high-performance RP chromatography column APPLICATION NOTE 21848 NISTmAb characterization with a high-performance RP chromatography column Author Xin Zhang Thermo Fisher Scientific, Sunnyvale, CA, USA Keywords MAbPac RP column, inter-column reproducibility,

More information

An Integrated Workflow for Intact and Subunits of Monoclonal Antibody Accurate Mass Measurements

An Integrated Workflow for Intact and Subunits of Monoclonal Antibody Accurate Mass Measurements An Integrated Workflow for Intact and Subunits of Monoclonal Antibody Accurate Mass Measurements Application Note Biotherapeutics and Biosimilars Authors Shuai Wu and Maryann Shen Agilent Technologies,

More information

Rapid UHPLC Analysis of Reduced Monoclonal Antibodies using an Agilent ZORBAX Rapid Resolution High Definition (RRHD) 300SB-C8 Column

Rapid UHPLC Analysis of Reduced Monoclonal Antibodies using an Agilent ZORBAX Rapid Resolution High Definition (RRHD) 300SB-C8 Column Rapid UHP Analysis of Reduced Monoclonal Antibodies using an Agilent ZORBAX Rapid Resolution High Definition (RRHD) 3SB-C8 Column Application Note BioPharma Authors James Martosella, Phu Duong, Susanne

More information

Reversed-phase Separation of Intact Monoclonal Antibodies Using Agilent ZORBAX Rapid Resolution High Definition 300SB-C8 1.

Reversed-phase Separation of Intact Monoclonal Antibodies Using Agilent ZORBAX Rapid Resolution High Definition 300SB-C8 1. Reversed-phase Separation of Intact Monoclonal Antibodies Using Agilent ZORBAX Rapid Resolution High Definition 3SB-C8 1.8 µm Column Application Note Biopharmaceuticals Authors James Martosella and Phu

More information

Application Note. Authors. Abstract. Automated Protein Sample Preparation for LC/MS

Application Note. Authors. Abstract. Automated Protein Sample Preparation for LC/MS Automation of Sample Preparation for Accurate and Scalable Quantification and Characterization of Biotherapeutic Proteins Using the Agilent AssayMAP Bravo Platform Application Note Automated Protein Sample

More information

Application Note. Biopharma. Authors. Abstract. James Martosella, Phu Duong Agilent Technologies, Inc Centreville Rd Wilmington, DE 19808

Application Note. Biopharma. Authors. Abstract. James Martosella, Phu Duong Agilent Technologies, Inc Centreville Rd Wilmington, DE 19808 Reversed-Phase Optimization for Ultra Fast Profiling of Intact and Reduced Monoclonal Antibodies using Agilent ZORBAX Rapid Resolution High Definition 3SB-C3 Column Application Note Biopharma Authors James

More information

Intact & Subunit Purity

Intact & Subunit Purity Agilent Biocolumns Intact & Subunit Purity Application Compendium Contents Background 2 Getting Started 3 Featured Application Notes 5 Fast and High Resolution Analysis of Intact and Reduced Therapeutic

More information

Developing Quantitative UPLC Assays with UV

Developing Quantitative UPLC Assays with UV Developing Quantitative UPLC Assays with UV Detection for Antibodies & Other Proteins Steve Taylor 2011 Waters Corporation 1 Outline UPLC technology for RP protein separations Method development parameters

More information

Thermo Scientific MAbPac HIC Columns. Novel Hydrophobic Interaction HPLC Columns. Designed for Monoclonal Antibody Analysis

Thermo Scientific MAbPac HIC Columns. Novel Hydrophobic Interaction HPLC Columns. Designed for Monoclonal Antibody Analysis Thermo Scientific MAbPac HIC Columns Novel Hydrophobic Interaction HPLC Columns Designed for Monoclonal Antibody Analysis Introduction The Thermo Scientific MAbPac HIC column family is designed for separations

More information

Monoclonal Antibody Analysis on a Reversed-Phase C4 Polymer Monolith Column

Monoclonal Antibody Analysis on a Reversed-Phase C4 Polymer Monolith Column Monoclonal Antibody Analysis on a Reversed-Phase C4 Polymer Monolith Column Shane Bechler 1, Ken Cook 2, and Kelly Flook 1 1 Thermo Fisher Scientific, Sunnyvale, CA, USA; 2 Thermo Fisher Scientific, Runcorn,

More information

Chromatography column for therapeutic protein analysis

Chromatography column for therapeutic protein analysis PRODUCT SPECIFICATIONS ProPac Elite WCX Column Chromatography column for therapeutic protein analysis Benefits Superior resolution power for proteins, monoclonal antibodies, and associated charge variants

More information

Agilent 6545XT AdvanceBio LC/Q-TOF System DESIGNED FOR BIOPHARMA

Agilent 6545XT AdvanceBio LC/Q-TOF System DESIGNED FOR BIOPHARMA gilent 6545XT dvancebio LC/Q-TOF System DESIGNED FO BIOPHM ONE INSTUMENT, MULTIPLE WOKFLOWS Characterizing proteins requires multiple approaches to access multiple levels of information; you can examine

More information

Improved Peptide Maps Using Core-Shell Media: Explaining Better Biopharmaceutical Applications With Kinetex C18 Columns

Improved Peptide Maps Using Core-Shell Media: Explaining Better Biopharmaceutical Applications With Kinetex C18 Columns Improved Peptide Maps Using Core-Shell Media: Explaining Better Biopharmaceutical Applications With Kinetex C18 Columns Michael McGinley, Jeff Layne, and Vita Knudson Phenomenex, Inc., 411 Madrid Ave.,Torrance,

More information

Separation of Recombinant Human Erythropoietin (repo) Using Agilent Bio SEC-3

Separation of Recombinant Human Erythropoietin (repo) Using Agilent Bio SEC-3 Separation of Recombinant Human Erythropoietin (repo) Using Agilent Bio SEC-3 Application Note BioPharma Authors Phu T Duong and James Martosella Agilent Technologies, Inc. 285 Centerville Rd, Wilmington,

More information

Separate and Quantify Rituximab Aggregates and Fragments with High-Resolution SEC

Separate and Quantify Rituximab Aggregates and Fragments with High-Resolution SEC Separate and Quantify Rituximab Aggregates and Fragments with High-Resolution SEC The Agilent 126 Infinity Bio-Inert Quaternary LC System and the AdvanceBio SEC 3Å, 2.7 µm Column Application Note Biologics

More information

Cell Culture Optimization Using an Agilent Bio-Monolith Protein A Column and LC/MS

Cell Culture Optimization Using an Agilent Bio-Monolith Protein A Column and LC/MS Cell Culture Optimization Using an Agilent Bio-Monolith Protein A Column and LC/MS Application Note Biopharmaceuticals and Biosimilars Authors Emmie Dumont, Isabel Vandenheede, Pat Sandra, Koen Sandra

More information

Understanding the Effects of Common Mobile Phase Additives on the Performance of Size Exclusion Chromatography

Understanding the Effects of Common Mobile Phase Additives on the Performance of Size Exclusion Chromatography Understanding the Effects of Common Mobile Phase Additives on the Performance of Size Exclusion Chromatography 1 Presentation Importance of aggregate analysis in BioPharma manufacturing Analytical techniques

More information

Puzzled About LCMS? Sample. Sensitivity in Mass Spec Analysis. Prep. Adapting LC-UV. Optimizing and Maintaining Your Mass Spec LCMS

Puzzled About LCMS? Sample. Sensitivity in Mass Spec Analysis. Prep. Adapting LC-UV. Optimizing and Maintaining Your Mass Spec LCMS Puzzled About LCMS? Sensitivity in Mass Spec Analysis Sample Prep Adapting LC-UV Optimizing and Maintaining Your Mass Spec To LCMS Paul Altiero Alex Ucci Applications Phone Support Chemistry and Supplies

More information

Analysis of Monoclonal Antibody N-glycans by Fluorescence Detection and Robust Mass Selective Detection Using the Agilent LC/MSD XT

Analysis of Monoclonal Antibody N-glycans by Fluorescence Detection and Robust Mass Selective Detection Using the Agilent LC/MSD XT Analysis of Monoclonal Antibody N-glycans by Fluorescence Detection and Robust Mass Selective Detection Using the Agilent LC/MSD XT Application Note Biotherapeutics & Biologics Authors Oscar Potter, Gregory

More information

Sample Preparation Automation and QQQ Workflows for Peptide Quantitation in DMPK

Sample Preparation Automation and QQQ Workflows for Peptide Quantitation in DMPK Sample Preparation Automation and QQQ Workflows for Peptide Quantitation in DMPK Alex Zhu & Yanan Yang LCMS Applications Scientist Agilent Technologies Why is Agilent Workflow a Good Choice for DMPK? AssayMAP

More information

Advanced QA/QC characterization MS in QC : Multi Attribute Method

Advanced QA/QC characterization MS in QC : Multi Attribute Method Advanced QA/QC characterization MS in QC : Multi Attribute Method Global BioPharma Summit The world leader in serving science A Complex Problem: Drug Safety and Quality Safety Is the product safe to use?

More information

Agilent AdvanceBio SEC Columns for Aggregate Analysis: Instrument Compatibility

Agilent AdvanceBio SEC Columns for Aggregate Analysis: Instrument Compatibility Agilent AdvanceBio SEC Columns for Aggregate Analysis: Instrument Compatibility Technical Overview Introduction Agilent AdvanceBio SEC columns are a new family of size exclusion chromatography (SEC) columns

More information

Fast Agilent HPLC for Large Biomolecules

Fast Agilent HPLC for Large Biomolecules Fast Agilent HPLC for Large Biomolecules Technical Overview Introduction Agilent media for the analysis of large biomolecules is available in an array of pore sizes to maximize selectivity and capacity

More information

The clearly better choice

The clearly better choice Agilent Accurate-Mass 6500 Series Q-TOF and 600 Series TOF LC/MS Systems The clearly better choice in TOF and Q-TOF LC/MS Our measure is your success. products applications soft ware services The unmatched

More information

Confident Pesticides Analysis with the Agilent LC/Triple Quadrupole and TOF/QTOF Solutions

Confident Pesticides Analysis with the Agilent LC/Triple Quadrupole and TOF/QTOF Solutions Confident Pesticides Analysis with the Agilent LC/Triple Quadrupole and TOF/QTOF Solutions Screening and Confirmation Chemical Analysis Solution Unit November, 2009 Page 1 Agilent s New 6540 Ultra High

More information

Robustness of the Agilent Ultivo Triple Quadrupole LC/MS for Routine Analysis in Food Safety

Robustness of the Agilent Ultivo Triple Quadrupole LC/MS for Routine Analysis in Food Safety Technical Overview Robustness of the Agilent Ultivo Triple Quadrupole LC/MS for Routine Analysis in Food Safety Authors Mark Sartain, Theresa Sosienski, and Dan Hui Dorothy Yang Agilent Technologies, Inc.

More information

Sean M. McCarthy and Martin Gilar Waters Corporation, Milford, MA, U.S. INTRODUCTION EXPERIMENTAL RESULTS AND DISCUSSION

Sean M. McCarthy and Martin Gilar Waters Corporation, Milford, MA, U.S. INTRODUCTION EXPERIMENTAL RESULTS AND DISCUSSION UPLC Separation of DNA Duplexes Sean M. McCarthy and Martin Gilar Waters Corporation, Milford, MA, U.S. INTRODUCTION Over the past 2 years there has been a considerable amount of effort focused on the

More information

Application Note. Authors. Abstract. Ravindra Gudihal Agilent Technologies India Pvt. Ltd. Bangalore India

Application Note. Authors. Abstract. Ravindra Gudihal Agilent Technologies India Pvt. Ltd. Bangalore India Characterization of bacteriophage derived anti-staphylococcal protein (P28) from production to purification using Agilent HPLC-Chip Q-TOF LC/MS system Application Note Authors Ravindra Gudihal Agilent

More information

The Agilent 1260 Infinity BioInert Quaternary Pump. Scope of a low-pressure mixing UHPLC pump with Bio-Inert Capabilities

The Agilent 1260 Infinity BioInert Quaternary Pump. Scope of a low-pressure mixing UHPLC pump with Bio-Inert Capabilities The Agilent 1260 Infinity BioInert Quaternary Pump Scope of a low-pressure mixing UHPLC pump with Bio-Inert Capabilities Patrick Cronan Applications Scientist Agilent Technologies Boston, MA 1 Comparison

More information

Agilent AssayMAP Bravo Platform AUTOMATED PROTEIN AND PEPTIDE SAMPLE PREPARATION FOR MASS SPEC ANALYSIS

Agilent AssayMAP Bravo Platform AUTOMATED PROTEIN AND PEPTIDE SAMPLE PREPARATION FOR MASS SPEC ANALYSIS Agilent AssayMAP Bravo Platform AUTOMATED PROTEIN AND PEPTIDE SAMPLE PREPARATION FOR MASS SPEC ANALYSIS GET FAST, ACCURATE, REPRODUCIBLE RESULTS The Agilent AssayMAP platform is an easy to use yet flexible

More information

High-resolution Analysis of Charge Heterogeneity in Monoclonal Antibodies Using ph-gradient Cation Exchange Chromatography

High-resolution Analysis of Charge Heterogeneity in Monoclonal Antibodies Using ph-gradient Cation Exchange Chromatography High-resolution Analysis of Charge Heterogeneity in Monoclonal Antibodies Using ph-gradient Cation Exchange Chromatography Agilent 1260 Infinity Bio-inert Quaternary LC System with Agilent Bio Columns

More information

Comparability Analysis of Protein Therapeutics by Bottom-Up LC-MS with Stable Isotope-Tagged Reference Standards

Comparability Analysis of Protein Therapeutics by Bottom-Up LC-MS with Stable Isotope-Tagged Reference Standards Comparability Analysis of Protein Therapeutics by Bottom-Up LC-MS with Stable Isotope-Tagged Reference Standards 16 September 2011 Abbott Bioresearch Center, Worcester, MA USA Manuilov, A. V., C. H. Radziejewski

More information

Jet Steam Proteomics with Automated Sample Prep for Superior Peptide Quantitation

Jet Steam Proteomics with Automated Sample Prep for Superior Peptide Quantitation Jet Steam Proteomics with Automated Sample Prep for Superior Peptide Quantitation Christine Miller Jason Russell June 2, 205 Overview Automated sample preparation Jet Stream Proteomics Peptide quantitation

More information

mab and ADC Analysis Shanhua Lin, Ph.D. The world leader in serving science

mab and ADC Analysis Shanhua Lin, Ph.D. The world leader in serving science mab and ADC Analysis Shanhua Lin, Ph.D. The world leader in serving science 2 Structure of IgG and Typical Forms of Heterogeneity Protein and mab Separation by HPLC Size difference? YES Size Exclusion

More information

Identifying Monoclonal Antibody Mutation Sites Using the Agilent 1290 Infinity II 2D-LC Solution with Q-TOF LC/MS

Identifying Monoclonal Antibody Mutation Sites Using the Agilent 1290 Infinity II 2D-LC Solution with Q-TOF LC/MS Identifying Monoclonal Antibody Mutation Sites Using the Agilent 190 Infinity II D-LC Solution with Q-TOF LC/MS Application Note Biopharmaceuticals and Biosimilars Authors Koen Sandra, Gerd Vanhoenacker,

More information

A Highly Accurate Mass Profiling Approach to Protein Biomarker Discovery Using HPLC-Chip/ MS-Enabled ESI-TOF MS

A Highly Accurate Mass Profiling Approach to Protein Biomarker Discovery Using HPLC-Chip/ MS-Enabled ESI-TOF MS Application Note PROTEOMICS METABOLOMICS GENOMICS INFORMATICS GLYILEVALCYSGLUGLNALASERLEUASPARG CYSVALLYSPROLYSPHETYRTHRLEUHISLYS A Highly Accurate Mass Profiling Approach to Protein Biomarker Discovery

More information

Discovery BIO Wide Pore

Discovery BIO Wide Pore Discovery BIO Wide Pore Solutions to Protein and Peptide Separation Challenges T403118 2 Agenda: What is Discovery BIO Wide Pore Physical characteristics Why we developed it and for whom Performance demonstrations

More information

2012 Waters Corporation 1

2012 Waters Corporation 1 UPLC User meeeting April 2012 Principles and Practices for SEC, IEX for Intact Protein Analysis by UPLC anders_feldthus@waters.com 2012 Waters Corporation 1 Agenda Ion-Exchange Chromatography Theory and

More information

Analysis and Purification of Polypeptides by Reversed-Phase HPLC

Analysis and Purification of Polypeptides by Reversed-Phase HPLC Analysis and Purification of Polypeptides by Reversed-Phase HPLC Reversed-phase HPLC is a valuable tool for the analysis and purification of proteins and peptides. It is effective in separating peptide

More information