Let s take a look several devices and methods for obtaining measurements of water quality in the field.

Size: px
Start display at page:

Download "Let s take a look several devices and methods for obtaining measurements of water quality in the field."

Transcription

1 EHST 2361 Environmental Sampling and Analysis Calibration and Use of Dissolved Oxygen and ph/conductivity Field Meters and Colorimeters in Water Quality Investigations Introduction We are taught that samples are taken and delivered to the analytical laboratory for analysis. In reality, some analytes including several indicators of water quality can be measured in the field. Of these, a few must be measured in the field because it is either easier to do so or the process of sampling and delivery to the laboratory alters the analyte in an unpredictable way. Four water quality parameters typically measured in the field are dissolved oxygen concentration (DO) in mg O 2/l (also called ppm O 2) ph in standard units or s.u. specific conductance (conductivity) in µs/cm or µmhos/cm temperature in F or C To aid the sampling technician, the instrument industry has developed a variety of rugged field meters that provide accurate, defensible results. We rely on these meters and must know how to calibrate and maintain them in order to produce data that meet the needs of our study. Over the last decade devices that, due to their size, power requirements, or complexity, were once only available in the laboratory are now available for use in the field. Once such device is the colorimeter, a portable type of spectrophotometer. Water quality parameters commonly measured in the field using a colorimeter are turbidity in TU or total suspended solids (TSS) in mg/l phosphate or orthophosphate in mg/l ammonia in mg/l nitrate in mg/l Let s take a look several devices and methods for obtaining measurements of water quality in the field. The HACH HQ40d Portable Meter and LDO Dissolved Oxygen/Temperature/% Saturation Probe The DO meter replaces the Winkler Method, a messy wet method (titration) that requires the use of caustics, special glassware, pipets, and burets. This HACH digital meter offers significant advantages over the older, analog meters like the YSI 54A. The primary advantage over the older technology of the YSA 54A is the LDO probe is calibrated much less frequently, as few as one time per year which translates to less preparatory work before use. Another advantage is the meter also accepts ph and conductivity IntelliCAL probes creating a single device that can collect multiple water quality parameters, saving you the effort of carrying several devices to each field site. The LDO probe requires much less maintenance than the YSI DO probe. However, the LDO requires the annual purchase of a probe sensor cap with matching calibration IntelliCAL ibutton at a cost of approximately $150. Annual maintenance cost for an analog meter and probe is literally pennies. Principle of operation: The HACH LDO sensor is coated with a luminescent material. Blue light from an LED is transmitted to the sensor surface. The blue light excites the luminescent material. As the material relaxes, it emits red light. The time from when the blue light was sent and the red light is emitted is measured. The more oxygen that is present the shorter the time it takes for the red light to be emitted. This time is measured and correlated to the oxygen concentration. Between the flashes of blue light a red LED is flashed on the sensor and used as an internal reference. (From the HACH website) Calibration While calibration of the HACH HQ40d Portable Meter and LDO probe need only occur annually, you will perform a calibration as part of today s lab. For most field meters, calibration is essential prior to use. Typical calibration is the process of showing the meter one or more solutions containing a known amount of the analyte (in this case dissolved O 2). This is most often accomplished using specially-prepared standards that are purchased from a reliable source. The

2 probe is placed into the standard and the reading on the meter adjusted to read the value of the standard. The meter then stores the calibration values electronically and is assumed to perform accurately for the period specified by the manufacturer at the end of which another calibration is required. Calibration is performed in the field at the beginning of the day and checked or recalibrated periodically throughout the day. THE DO METER IS CALIBRATED A LITTLE DIFFERENTLY. The particulars of calibrating the DO meter are as follows. The amount of O 2 that can dissolve in a liquid like surface water is a function of field temperature and atmospheric pressure. Therefore both must be known to properly calibrate the meter, that is, to tell the meter how much O 2 to expect to see in the standard. The temperature probe built into the meter is factory calibrated and assumed accurate between periodic factory servicing. This probe will provide the temperature value we need for calibration. The meter also measures the atmospheric pressure needed to complete the calibration. The HACH LDO does not need a standard solution for calibration. Instead the probe is fitted with a small calibration bottle. Once placed on the probe, the calibration bottle creates a water-saturated air environment around the probe that simulates a water sample holding as much O 2 as possible at a given temperature and pressure. After stabilization, the temperature and pressure values are used to compute the standard DO saturation value used to calibrate the meter. This calibration method is called saturated air calibration. Other methods of calibration do exist. Procedure for Calibrating the HACH HQ40d and LDO: 1. LDO Calibration is performed manually using water-saturated air. 2. Locate the calibration bottle and add a small amount (about 1 cm) of water to the bottle, cap, and shake vigorously for several minutes. 3. Press the BLUE/LEFT key under Calibrate. 4. Remove the bottle cap, carefully blot the probe dry, and insert the LDO into the calibration bottle BEING CAREFUL NOT TO TOUCH THE PROBE CAP. 5. Check to make sure there is no water adhering to the probe cap. 6. Press the GREEN/RIGHT key under Read. 7. When the reading is stable the standard value will be highlighted on the screen and the calibrated reading will appear on the screen. 8. Press the UP key under Done. 9. Press the GREEN/RIGHT key under Store to accept the calibration and return to the measurement mode. 10. When calibration is successful, the display will show OK in the upper left hand corner. Additional Considerations: 1. The DO saturation is inversely proportional to temperature. Cold water holds the most O Supersaturation of water can occur when a cold water sample is warmed without agitation or if pure oxygen is bubbled into water. Supersaturation can be detrimental to aquatic life. 3. Low DO is detrimental to aquatic life. Signs of stress are likely to be seen at values below 4 mg O 2/liter. 4. Vigorous agitation while sampling or analyzing water can artificially introduce O 2 and result in higher than actual readings 5. Some gases or dissolved substances may interfere with the workings of the probe. Check the information accompanying the specific unit and probe you plan to use to understand potential interferences. Continued

3 HACH HQ40d Portable Meter and LDO probe Specifications: Dissolved Oxygen Range mg/l 1-200% saturation Dissolved Oxygen Accuracy ±0.1 mg/l for mg/l ±0.2 mg/l for greater than 8.0 mg/l % Saturation 1.0% Temperature Range 0-50 C Temperature Resolution 0.1 C Temperature Accuracy ±0.3 C Analyzing DO in Water: 1. The probe may be placed directly into surface water or collected sample. 2. Place the probe in the sample, no need to swirl. Press the GREEN/RIGHT key under Read. The display will show Stabilizing and a progress bar will fill from 0 to 100% as the probe stabilizes in the sample. When the result has stabilized, the lock icon will appear and the result will be stored automatically in the data log. 3. Record the DO in mg O 2/liter. Record the temperature. 4. When not in use, protect the probe from damage. There are no special storage requirements for the HACH HQ40d and LDO. Laboratory Activity - Learning the Capabilities of the DO Meter: 1. HACH HQ40d and LDO as instructed. 2. Draw a tap water sample into a 400 ml beaker. Allow the sample to equilibrate for a few minutes. 3. Determine the DO and temperature for the tap water. Determine the DO three more times waiting about one minute between each determination. List your results in the table below. HACH HQ40d & LDO Time DO mg O 2/l Temperature 0 min min min min Average Range 4. Compute the average and the range. If precision is defined as the closeness of repeated measurements of the same quantity, how precise is the meter? 5. Check the accuracy of the meter by immersing the probe into the zeroing solution. Does the meter read zero? 6. Rinse the probe in clean water. Measure the DO and temperature of the pond water samples. Record your results below. Try not to vigorously agitate any sample. Next page for data table

4 HACH HQ40d & LDO Pond Water Sample DO mg O 2/l Temperature Cold Room temperature Warm Assuming the DO in each sample started at the same concentration, what effect does temperature have on DO? We will use your results and Microsoft s Excel spreadsheet program to demonstrate the precision of the DO meter and the effect of temperature on dissolved oxygen. Please be ready to present your data. The Hanna HI Portable ph/ec/tds/temperature Meter and Probe The ph of Water The ph of a liquid is defined as the negative log of the hydrogen ion concentration or log [H + ]. In pure water, H + and the hydroxyl anion (OH - ) exist in equilibrium with water molecules (H 2O). The equilibrium equation is H + + OH - H 2O. Theoretically, pure water has a ph of 7, otherwise known as neutral, and the concentration of each ion is the equation is essentially equal. The addition of acidic or alkaline (basic) materials to pure water changes the ph because the concentration of H + in the solution is affected by the addition. Bases add OH - thereby increasing the concentration of hydroxyl anion and decreasing (relatively) the H + concentration. Basic solutions have ph values ranging from 7 to 14. Acids add H + and decrease (relatively) the OH - concentration. Acidic solutions have ph values ranging from 7 to 0. Beyond the underlying chemistry of acids and bases, ph plays an important role in understanding the hazard properties of materials and the environmental effects of hazardous materials releases. Under RCRA, a waste can be considered corrosive if its ph is less than 2.5 and greater than 12.5 causing it to become subject to RCRA hazardous waste regulations. Ecologically speaking, animals and plants are only found in habitats where the ph is moderate. As a general rule, organisms are not found living in waters or soils with ph values less than 4 and more than 11. Most organisms on the planet are not adapted to ph extremes. Releases of corrosive materials will cause damage to the environment if they are not neutralized or otherwise cleaned up. Areas of distressed vegetation, fish kills, bird kills, and injured or debilitated wildlife may be signs of releases. Neutralization involves chemical treatment of acids or bases resulting in more neutral (ph = 7) materials. For example, HCl could be neutralized with the addition of NaOH: HCl + NaOH H 2O + NaCl a very violent reaction. A better neutralizing agent is sodium bicarbonate (NaHCO 3), HCl + NaHCO3 H 2O + NaCl + CO 2, a weaker base that reacts less violently. There are several different types of ph meters available to the technician. Most have digital readout and data logging capabilities; older models may use analog readout (needle and scale). Meters may be benchtop devices, portable devices, or field devices. The HI is an example of a rugged field meter that produces reliable, defensible results. Benchtop and portable meters are not as rugged as field meters. Field meters are almost always weather (or water) resistant (able to withstand rain and splashes), but rarely water proof (able to withstand immersion in water; well, maybe a brief dunking). Meters need a probe or electrode to sense the ph of the sample. The electronics of the meter then interprets the signal from the electrode and provides a display of the ph. The meter also stores calibration information. Electrodes need to

5 be calibrated frequently. For example, daily, twice-daily, and before each use, may be recommended by meter manufacturers or EPA. The meter is calibrated using ph standards, also called standard solutions or ph buffers. A two-point calibration uses two standards: ph 4 and 7 or ph 7 and 10. Which standards you choose depends on the expected ph of the samples you will be analyzing. Try to select standards that bracket the expected ph of the samples you will be analyzing. If acidic samples are expected choose ph 4 and 7, for basic samples choose ph 7 and 10. A few ph meters use a three-point calibration. Once calibrated, meter function should be periodically recalibrated or checked against a single standard (like ph 7) throughout the day. Procedure for Calibrating the ph function of the HI : 1. Inspect the meter for any signs of damage. Check the probe connector and adjust to finger tight. Examine the probe tip. It should be free from any build up or cracks. 2. If necessary, connect the probe (or electrode) to the meter, remove the probe s protective cap, rinse it with distilled water, and place the probe end into distilled water or ph 4 buffer. 3. Turn on the meter by pressing ON. 4. The meter displays the remaining battery percentage every time it is turned on. Low batteries will give erroneous readings, replace if necessary. 5. Allow the probe and meter a short time to stabilize. 6. Of the three standards available, choose two standards that will bracket the sample ph. Unless contaminated in some way, surface water samples will be between ph 6 and 9. Check the expiration date of the standards use only the freshest solutions within their expiration dates. 7. Follow the manufacturer s instructions and perform a two-point calibration (attached). Note the calibration date, time, standards used, and any useful comments about meter function or maintenance performed in the meter s calibration and maintenance logbook or field notebook. 8. The meter is now ready to be used. Protect the probe when not in use by placing it into ph 4 buffer or its protective cap (filled with ph 4 buffer). Leave the meter on or in stand by mode while traveling between sampling sites or recalibrate in the field. Analyzing the ph of Water: 1. The probe may be placed directly into surface water or a collected sample. The stability indicator will appear in the LCD screen. 2. The probe is gently swirled in the sample until the ph stabilizes. If the water is flowing, there is no need for swirling. The stability indicator will disappear. 3. Record the ph in standard units. Record the temperature. 4. Rinse the probe with distilled water between samples. 5. When not in use, rinse the probe in distilled water and place it into ph 4 buffer or its protective cap. Additional Considerations: 1. Since periodic calibration and checking is needed to ensure accuracy, ph buffers, distilled water wash bottles, and lint-less wipes (like Kimwipes) will be required in the field. 2. The ph probe tip is typically made of thin glass and is therefore susceptible to breakage. 3. Our most common use of the ph function is water and wastewater analysis but the probe can also be used with other liquids. Check the manufacturer s specifications for other uses before measuring the ph of liquids other than water. 4. Vigorous agitation of the sample while measuring may alter its ph. 5. Samples with ph outside of the calibration range will not be read accurately. The error is proportional to the distance away from the calibration range. 6. ph buffers have a limited shelf life and are somewhat susceptible to dilution by samples, other buffers, and wash waters. Obtain fresh buffer after a few uses.

6 Electrical Conductance, Specific Conductance, or Conductivity of Water Surface waters and groundwaters are accurately described as water solutions. In reality, pure water does not exist in nature. Even precipitation, i.e., rain, quickly dissolves atmospheric gases, vapors, and particulates. These enter and leave solution with water according to their equilibrium constants, the presence of other solutes and temperature. Some alter the electrical conductance of water. Pure water is a poor conductor of electricity. However, water solutions improve conductance with increasing solute concentration. Conductivity is therefore a measure of the amount of solute in water and reported as µs/cm (microsiemens per centimeter), µmho/cm, or as total dissolved solids (TDS) in ppm. As a general rule, pollutant contamination increases conductivity. It is not unusual to see industrial effluents with conductivities greater than 1,000 or 5,000 µs/cm. For comparison, pure water has conductivity at or near 0 µs/cm, rainwater is typically less than 40 µs/cm, and surface waters in Ohio are usually less than 500 µs/cm. Conductivity is not only another water quality parameter but useful as a tool for delineating effluent mixing zones or purging wells (as discussed in a previous laboratory). High water conductivity values are physiologically challenging to freshwater aquatic organisms. Some species do not possess the ability to adapt to high conductivity and will leave the area or be killed as a result. In this way and others, conductivity can be an indicator of water quality. Like DO and ph meters, conductivity meters are calibrated. The HI uses a one-point calibration. Like DO, the amount of dissolved substances water can hold is dependent on temperature and pressure, with pressure being less important than it is in DO. The conductivity standard 1413 µs/cm is often used for meter calibration. I assume that this value is a compromise between typical values expected in the field and the stability of the chemicals used to make the standard. Note that with all calibrations, samples with values outside the calibration range or far from the calibration point are subject to error. Therefore, other conductivity standards do exist. Procedure for Calibrating the Conductivity Function of the HI : 1. Inspect the meter for any signs of damage. Check the probe connector and adjust to finger tight. Examine the probe tip. It should be free from any build up or cracks. 2. If necessary, connect the probe (or electrode) to the meter, remove the probe s protective cap, rinse it with distilled water, and place the probe end into distilled water or ph 4 buffer. 3. Turn on the meter by pressing ON. 4. The meter displays the remaining battery percentage every time it is turned on. Low batteries will give erroneous readings, replace if necessary. 5. Allow the probe and meter time to stabilize. 6. Check the expiration date of the standard use only the freshest solution. 7. Follow the manufacturer s instructions and perform a one-point calibration (attached). Note the calibration date, time, standard used, and any useful comments about meter function or maintenance performed in the meter s calibration and maintenance logbook or field notebook. 8. The meter is now ready to be used. Protect the probe when not in use by placing it into ph 4 buffer or its protective cap. Leave the meter on or in stand by mode while traveling between sampling sites or recalibrate in the field. Analyzing the Conductivity of Water: 1. The probe may be placed directly into surface water or collected sample. Gently tap probe to ensure no bubbles are present in the electrode. Note: the conductivity electrode is the slit in the plastic housing at the very tip of the combination probe. 2. The probe is held in the sample until the conductivity stabilizes. The stability indicator will appear in the LCD screen 3. Record the conductivity in µs/cm and the temperature ( C). 4. Rinse the probe with distilled water between samples. 5. When not in use, rinse the probe in distilled water and place it into ph 4 buffer or its protective cap.

7 Additional Considerations: 1. Salinity of seawater is related to conductivity but typically measured using a different device and reported using a different unit (ppt or o / oo). 2. Conductivity standards have a limited shelf life and are very susceptible to dilution by samples and wash waters. Dry the probe with a lint free wiper before placing into the standard. 3. Some older conductivity meters cannot be calibrated. Their function can only be checked against standards and the results recorded. These must be calibrated at the factory. Samples will be present in the laboratory so you can try out the meters on the real thing. These meters will be taken into the field on our surface water and sediment sampling trip. Remember the calibration procedures so you can repeat them in the field. Laboratory Activity - Learning the Capabilities of the ph and Conductivity Meter: If there is time, repeat the DO experiment above using the ph and conductivity function of the HI Record your results below. Remember to wait for the stability indicator before taking your reading. Precision Time ph (su) Temperature Conductivity µs/cm 0 min min min min Average ** Range Effect of Temperature Pond Water Sample ph (su) Temperature Conductivity µs/cm Cold Room temperature Warm Can you draw any conclusions about the precision of the ph/conductivity probe? The effect of temperature? We will use your results and Microsoft s Excel spreadsheet program to demonstrate the precision of the ph and Conductivity meter and the effect of temperature on ph and conductivity. **Average the ph values and place the result here. IMPORTANT NOTE FOR THE PREPARATION OF SUMMARY STATISTICS AND REPORTS USING ph DATA: ph values must be converted back to the H + concentration, averaged using a typical method, then converted back to log [H + ] for reporting. Averaging the ph meter results directly, without transformation, results in inaccurate mean and related statistics like standard deviation of the mean.

8 The HACH Company DR/850 Portable Datalogging Colorimeter A colorimeter depends on the reaction of various added reagents and dissolved analytes in samples to affect the absorbance of light passing through a water sample. The absorbance of a particular wavelength of light by the prepared sample is proportional to the concentration of the analyte dissolved in the sample. There are many procedures available for measuring analytes using a colorimeter, some are approved for use by regulatory agencies like EPA and some are not. Unapproved methods should not be used for regulatory compliance work. A sample of surface water is prepared according to a prescribed method. Typical preparations include measuring a volume of sample (10 ml is a common sample volume), adding one or more reagents, mixing, and allowing time for reactions to occur. The sample is either prepared in the colorimeter s supplied glass sample cell or measured into a glass sample cell after preparation in other glassware. The outer walls of the sample cell are then cleaned to remove fingerprints, water, and any other debris that might affect the transmittance of the beam of light and placed into the colorimeter s sample cell compartment. Colorimeters are capable of generating a beam of light of known wavelength. The wavelength chosen is adjusted, as needed, for different analytical methods. Once placed in the sample cell compartment, the beam of light passes through prepared sample in a glass sample cell. In most preparations, the presence of the analyte of interest, now reacted with the reagents added during the preparation step, causes an increase the absorbance of the beam of light by the sample, essentially the same thing as a decrease in the transmittance of the light through the sample. The decrease in transmittance or the increase in absorbance is directly proportional to the presence of the analyte in the sample. The analyte is now quantified! When EPA-approved methods are chosen for preparing the samples, the colorimeter becomes a very sensitive and powerful tool for quickly (usually less than 30 minutes) quantifying a number of water quality parameters and contaminants in water. With the addition of quality control measures which are part of a typical quality assurance program, data generated by the colorimeter are defensible and on par with laboratory data from approved laboratories. The following basic procedures for measuring turbidity, orthophosphate, and ammonia in surface waters using the HACH DR/850 are taken from the Euclid Creek Volunteer Watershed Monitoring Program: Directions for Performing Physical/Chemical Tests written by Dr. Mike Nichols of John Carroll University in Cleveland. They have been modified slightly for use in this laboratory exercise. Turbidity of Surface Water Measurement Using Colorimeter 1. Turn on the colorimeter by pressing the Exit Φ Button. 2. Remove the cover from colorimeter. 3. Obtain the two sample cells marked DI water and TURB/PO 4 sample; the DI water will already have distilled water in it. 4. Wipe the outside of the DI water cell with a lint-less paper towel to remove any fingerprints and water droplets. 5. Place the DI water cell into the hole (sample cell compartment) in the meter and place the cover on top of the sample to block out all of the light. 6. Press PRGM 9 4 ENTER. 7. The meter should show susld at the bottom of the screen. 8. Press ZERO. 9. The meter should read 0 in a couple of seconds. 10. Remove the DI water cell from the meter. 11. Fill the TURB/PO 4 sample cell to the 10 ml mark with stream water, using a plastic pipet if necessary. 12. Wipe the outside of the cell with a lint-less paper towel. 13. Place the TURB/PO 4 sample cell in meter and place cover on top. 14. Press READ. 15. Record the value on the data sheet. 16. Keep the TURB/PO 4 sample cell in the colorimeter. 17. Proceed immediately to the reactive phosphate procedure.

9 Reactive or Mono Phosphate in Surface Water 1. The TURB/PO 4 sample cell should still be in the colorimeter; if its not place it in the colorimeter. 2. Turn the colorimeter on (Exit Φ Button), if it is not still on. 3. Press PRGM 7 9 ENTER; the colorimeter should read P, P 2O 5, or PO 4 at the bottom of the screen (these are the units of measure). 4. Place the cover over the cell. 5. Press ZERO. 6. Remove the cover and TURB/PO 4 sample cell. 7. Remove the TURB/PO 4 sample cell cap and add the contents of 1 foil PhosVer 3 powder packet. 8. Recap the cell and shake vigorously for 30 seconds. Note: not all of the powder will dissolve. 9. Press TIMER and ENTER. 10. The reaction will take 2 minutes and the solution will turn blue if phosphate is present. 11. After the 2 minute period, wipe the outside of the cell with a lint-less paper towel. 12. Place the TURB/PO 4 sample cell in the meter, place the cover on top. 13. Press READ. 14. Record the value and the units of the measurement (listed as either P, P 2O 5, or PO 4 at the bottom of the screen) on the data sheet. 15. Should the reading be underrange - record the value as underrange, not zero. 16. If the reading is overrange - record it as overrange on the data sheet; this sample will need further analysis in the laboratory. Ammonia in Surface Water 1. Remove the cover of the colorimeter. 2. Turn the colorimeter on by pressing the Exit Φ Button. 3. Press PRGM 6 4 ENTER; the colorimeter should show NH 4, NH 3-N, or NH 3 on the bottom of the screen (these are the units of measure). 4. Use two clean sample cells, they are marked NH 4 blank and NH 4 sample. 5. Fill the NH 4 blank cell to the 10 ml mark with distilled water from the squirt bottle. 6. Fill the NH 4 sample cell to the 10 ml mark with stream water, using a plastic pipet if necessary. 7. Add the contents of 1 ammonium salicylate powder packet to each cell. 8. Cap each vial and shake each for 30 seconds. 9. Press TIMER and ENTER. 10. The first reaction will take 3 minutes, all of the solid will dissolve and the solutions will turn yellow. 11. After the 3 minute period has elapsed, add the contents of 1 ammonium cyanurate powder packet to each cell and cap and shake vigorously for 30 seconds. All of the powder will dissolve. 12. Press ENTER. 13. The second reaction will take 15 minutes and the NH 4 sample solution will turn green if ammonia is present. 14. After the 15 minute time period has passed, wipe the outside of both cells with a lint-less paper towel. 15. Place the NH 4 blank cell into the meter, place the cover over it and press ZERO; the colorimeter will now read Remove the NH 4 blank cell and place the NH 4 sample cell into the meter, replace the cover and press READ. 17. Record the value and units of measure (NH 4, NH 3-N, or NH 3) on the data sheet. 18. Should the reading be underrange - record the value as underrange, not zero. 19. If the reading is overrange - record it as overrange on the data sheet; this sample will need further analysis in the laboratory. After You Have Completed All of the Spectrometric Tests 1. Make sure that all of the solutions in the sample cells (except the DI water cell) have been emptied into a waste bottle or down the drain with water. 2. Each cell has been rinsed with 10 ml of distilled water from the squirt bottle. Rinse each cell at least three times. 3. The data sheets are complete. 4. All trash has been placed into a waste plastic bag. 5. The colorimeter is off. 6. All equipment and materials have been placed back into their storage containers.

10 This page intentionally left blank

11

12

13 GENERAL INFORMATION SITE NAME OR NUMBER: STREAM NAME: INVESTIGATOR(S): DATE AND TIME STARTED: Surface Water Quality Monitoring Data Sheet WATERSHED NAME: WEATHER PAST 24 HOURS: HEAVY RAIN STEADY RAIN INTERMITTENT RAIN OVERCAST CLEAR/SUNNY WEATHER NOW: HEAVY RAIN STEADY RAIN INTERMITTENT RAIN OVERCAST CLEAR/SUNNY AIR TEMPERATURE: F OR C STREAM OBSERVATIONS: WATER COLUMN STREAM BOTTOM STREAM BANK FISH PRESENT? YES NO MONITORING DATA WATER TEMPERATURE: F OR C DISSOLVED O2: (MG/L) METHOD/METER USED: CALIBRATION/NOTES PH: (S.U.) CALIBRATION NOTES CONDUCTIVITY: (µs/cm OR µmhos/cm) CALIBRATION/NOTES METHOD/METER USED: STANDARDS USED EXPIRATION DATES METHOD/METER USED: STANDARD(S) USED EXPIRATION DATES TURBIDITY TURBIDITY TUBE: (CM OR IN) RESULTING TSS: (MG TSS/L, FROM TABLE) circle TURBIDITY METER: (MG TSS/L) METER USED: OTHER PARAMETERS ORTHOPHOSPHATE: (MG/L) AMMONIA: (MG/L) CALIBRATION/NOTES OTHER (LIST) STREAM FLOW: (FT 3 /SEC, CFS) METHOD/METER USED: NOTES

PaSEC Colorimeter Instructions DR850

PaSEC Colorimeter Instructions DR850 PaSEC Colorimeter Instructions DR850 OXYGEN, DISSOLVED, High Range (0 15.0 mg/l O 2 ) Dissolved Oxygen Test Colorimeter Instruction #4 Replacement: Using the AccuVac Snapper Tool Breaking the Dissolved

More information

Phosphorus, Reactive (Orthophosphate)

Phosphorus, Reactive (Orthophosphate) Phosphorus, Reactive (Orthophosphate) DOC316.53.01118 USEPA 1, 2 PhosVer 3 Method 3 Method 8048 0.06 to 5.00 mg/l PO 4 (0.02 to 1.60 mg/l P) Test N Tube Vials Scope and application: For drinking water,

More information

PHOSPHORUS, REACTIVE (0.00 to 5.00 mg/l PO 4

PHOSPHORUS, REACTIVE (0.00 to 5.00 mg/l PO 4 PHOSPHORUS, REACTIVE (0.00 to 5.00 mg/l PO 4 ) PhosVer 3 Method, Test N Tube Procedure USEPA accepted for reporting wastewater analysis * Method 8048 For water, wastewater, and seawater 1. Enter the stored

More information

Collecting a Surface Water Sample

Collecting a Surface Water Sample Collecting a Surface Water Sample CAUTION: Never carry or lift the pole above your head, as touching power lines could cause electrocution. Notes: Students (

More information

Total Dissolved Solids

Total Dissolved Solids Total Dissolved Solids LabQuest 12 INTRODUCTION Solids are found in streams in two forms, suspended and dissolved. Suspended solids include silt, stirred-up bottom sediment, decaying plant matter, or sewage-treatment

More information

Sampling Day Activities

Sampling Day Activities Sampling Day Activities Approved by: WWA Director Reviewed by: Volunteer Monitoring Coordinator On the day of sampling, the following activities are to be conducted at the sampling site in the order presented

More information

Biochemical Oxygen Demand

Biochemical Oxygen Demand Biochemical Oxygen Demand Computer 20 Oxygen available to aquatic organisms is found in the form of dissolved oxygen. Oxygen gas is dissolved in a stream through aeration, diffusion from the atmosphere,

More information

Phosphorus, Reactive (Orthophosphate)

Phosphorus, Reactive (Orthophosphate) Method 8048 DR/2400 * Adapted from Standard Methods for the Examination of Water and Wastewater ** Procedure is equivalent to USEPA method 365.2 and Standard Method 4500-P-E for wastewater. Phosphorus,

More information

Color, True and Apparent, LR

Color, True and Apparent, LR Color, True and Apparent, LR DOC316.53.01252 Platinum-Cobalt Standard Method 1, 2, 3 Method 8025 3 to 200 color units Scope and application: For water, wastewater and seawater; equivalent to NCASI method

More information

Direct ISE Method Method to 10.0 mg/l F Fluoride ISE

Direct ISE Method Method to 10.0 mg/l F Fluoride ISE Fluoride DOC316.53.01237 Direct ISE Method Method 8323 0.1 to 10.0 mg/l F Fluoride ISE Scope and application: For drinking water 1 and wastewater 2. 1 The equipment that is specified in this method meets

More information

Boiler Water No.2 Test Kit Instruction Manual

Boiler Water No.2 Test Kit Instruction Manual Boiler Water No.2 Test Kit Instruction Manual BOILER WATER No.2 Steam boilers require careful treatment in order to maintain efficiency, steam purity and to prevent scale and corrosion of pipework and

More information

Sulfate. USEPA 1 SulfaVer 4 Method 2 Method to 70 mg/l SO 4. Powder Pillows. Test preparation. Before starting.

Sulfate. USEPA 1 SulfaVer 4 Method 2 Method to 70 mg/l SO 4. Powder Pillows. Test preparation. Before starting. Sulfate DOC316.53.01473 USEPA 1 SulfaVer 4 Method 2 Method 8051 2 to 70 mg/l SO 4 2 Powder Pillows Scope and application: For water, wastewater and seawater. 1 USEPA accepted for reporting wastewater analyses.

More information

Dissolved Oxygen. LabQuest INTRODUCTION

Dissolved Oxygen. LabQuest INTRODUCTION Dissolved Oxygen LabQuest 5 INTRODUCTION Oxygen gas dissolved in water is vital to the existence of most aquatic organisms. Oxygen is a key component in cellular respiration for both aquatic and terrestrial

More information

3. Fill a 50-mL. 2. Press: 1 ENTER The display will show mg/l, Al and the ZERO icon.

3. Fill a 50-mL. 2. Press: 1 ENTER The display will show mg/l, Al and the ZERO icon. ALUMINUM (0 to 0.80 mg/l) Aluminon Method * Method 8012 For water and wastewater 1. Enter the stored program number for aluminum (Al). Press: PRGM The display will show: PRGM? Note: Adjust the ph of stored

More information

3. Add 0.4 ml of. 7. Use a TenSette

3. Add 0.4 ml of. 7. Use a TenSette Method 10173 ORGANIC CARBON, TOTAL, Mid Range (15 150 mg/l C) Direct Method * For wastewater and industrial waters HRS MIN SEC 1. Turn on the DRB 200 reactor. Heat to 103 105 C. Note: See DRB 200 user

More information

:00 am 6:00 pm Mon-Thu (MT) 6:00 am 5:00 pm Fri (MT)

:00 am 6:00 pm Mon-Thu (MT) 6:00 am 5:00 pm Fri (MT) 800.372.0122 303.431.8454 6:00 am 6:00 pm Mon-Thu (MT) 6:00 am 5:00 pm Fri (MT) info@eraqc.com www.eraqc.com www.facebook.com/eraqc www.youtube.com/eraqc 2013 Waters Corporation 1 Top Tips for Conducting

More information

3. Add 0.4 ml of. 7. Use a TenSette

3. Add 0.4 ml of. 7. Use a TenSette Method 10129 ORGANIC CARBON, TOTAL, Low Range (0.0 20.0 mg/l C) Direct Method * For water, drinking water, and wastewater HRS MIN SEC 1. Turn on the DRB 200 reactor. Heat to 103-105 C. Note: See DRB 200

More information

TPTZ Method * (Powder Pillows or AccuVac Ampuls) Using Powder Pillows

TPTZ Method * (Powder Pillows or AccuVac Ampuls) Using Powder Pillows IRON, TOTAL (0 to 1.80 mg/l) TPTZ Method * (Powder Pillows or AccuVac Ampuls) Using Powder Pillows Method 8112 For water, wastewater, and seawater 1. Enter the stored program number for iron (Fe)- powder

More information

Manual No Revision MODEL GLI HAND-HELD/PORTABLE TOTAL/FREE CHLORINE METER

Manual No Revision MODEL GLI HAND-HELD/PORTABLE TOTAL/FREE CHLORINE METER Manual No. 93711 Revision 0-401 MODEL GLI-93711 HAND-HELD/PORTABLE TOTAL/FREE CHLORINE METER Please read this instruction manual carefully before using the instrument. This manual will provide you with

More information

Direct ISE Method Method to 4.00 mg/l NO 3 N Nitrate ISE

Direct ISE Method Method to 4.00 mg/l NO 3 N Nitrate ISE Nitrate DOC316.53.01239 Direct ISE Method Method 8359 0.04 to 4.00 mg/l NO 3 N Nitrate ISE Scope and application: For drinking water. Test preparation Instrument-specific information This procedure is

More information

Rev Experiment 10

Rev Experiment 10 Experiment 10 SPECTROPHOTOMETRIC DETERMINATION OF IRON IN DRINKING WATER 2 lab periods Reading: 1) Chapter 17, pg 393-403, Quantitative Chemical Analysis, 8 h Edition, Daniel C. Harris (7 th Edition: Chapter

More information

OPERATING INSTRUCTIONS. Portable Turbidity Meter

OPERATING INSTRUCTIONS. Portable Turbidity Meter OPERATING INSTRUCTIONS Portable Turbidity Meter ELE International Chartmoor Road, Chartwell Business Park, Leighton Buzzard Bedfordshire, LU7 4WG England Sales: Civil +44 (0) 870 777 7706 Sales: Env. +44

More information

Measuring Dissolved Oxyg Instructions for the La Motte Field-Test Kit

Measuring Dissolved Oxyg Instructions for the La Motte Field-Test Kit Then, immediately add 8 drops of alkaline potassium iodide azide to the water sample Recap the bottle and shake to mix. Immediately add 8 drops of manganous sulfate to the water Record the temperature

More information

Collecting a Surface Water Sample

Collecting a Surface Water Sample FIELD PROCEDURE Collecting a Surface Water Sample CAUTION: Never carry or lift the pole above your head, as touching power lines could cause electrocution. Notes: Students (

More information

Refractometer Calibration, Use and Maintenance Stephen Vasquez and Shannon Mueller

Refractometer Calibration, Use and Maintenance Stephen Vasquez and Shannon Mueller University of California Cooperative Extension Fresno County Refractometer Calibration, Use and Maintenance Stephen Vasquez and Shannon Mueller Harvest season will soon be upon us in the San Joaquin Valley,

More information

Evaluation copy. Total Dissolved Solids. Computer INTRODUCTION

Evaluation copy. Total Dissolved Solids. Computer INTRODUCTION Total Dissolved Solids Computer 12 INTRODUCTION Solids are found in streams in two forms, suspended and dissolved. Suspended solids include silt, stirred-up bottom sediment, decaying plant matter, or sewage-treatment

More information

Instruction Manual HI Portable ph/ec/tds/ C Meters

Instruction Manual HI Portable ph/ec/tds/ C Meters Instruction Manual HI 9811-5 Portable ph/ec/tds/ C Meters www.hannainst.com 1 Dear Customer, Thank you for choosing a Hanna product. Please read this instruction manual carefully before using the meter.

More information

Groundwater Monitoring on a Budget. June Mooney, Water Quality Specialist, Greater Augusta Utility District

Groundwater Monitoring on a Budget. June Mooney, Water Quality Specialist, Greater Augusta Utility District Groundwater Monitoring on a Budget June Mooney, Water Quality Specialist, Greater Augusta Utility District "Common Interests, Shared Solutions" September 18, 2012 Why We Monitor Groundwater Rule Wellhead

More information

HI EPA Compliant Turbidity Meter HI EPA Compliant Turbidity and Chlorine Meter

HI EPA Compliant Turbidity Meter HI EPA Compliant Turbidity and Chlorine Meter HI 98703 EPA Compliant Turbidity Meter HI 93414 EPA Compliant Turbidity and Chlorine Meter SPECIFICATIONS Turbidity (HI 93414 & HI 98703) Range Range Selection ACCESSORIES HI 710005 Voltage adapter HI

More information

EPA Compliant Turbidity & Chlorine Meter

EPA Compliant Turbidity & Chlorine Meter EPA Compliant Turbidity & Chlorine Meter New Turbidity Technology from HANNA HANNA s new HI 98703 and HI 93414 utilize an EPA compliant tungsten light source and state-of-the-art optical system that allow

More information

DR/4000 PROCEDURE. OXYGEN DEMAND, Chemical. Colorimetric Measurement, 0 to 40 mg/l COD

DR/4000 PROCEDURE. OXYGEN DEMAND, Chemical. Colorimetric Measurement, 0 to 40 mg/l COD Method 8000 DR/4000 PROCEDURE Reactor Digestion Method* (0 to 40.0** mg/l COD) Scope and Application: For water, wastewater and seawater; digestion is required. The estimated detection limit for program

More information

Scope and application: For water, wastewater, seawater, brine solutions, produced waters and hydraulic fracturing waters.

Scope and application: For water, wastewater, seawater, brine solutions, produced waters and hydraulic fracturing waters. Iron, Ferrous DOC316.53.01049 1,10-Phenanthroline Method 1 Method 8146 0.02 to 3.00 mg/l Fe 2+ Powder Pillows or AccuVac Ampuls Scope and application: For water, wastewater, seawater, brine solutions,

More information

1,5-Diphenylcarbohydrazide Method* (Powder Pillows or AccuVac Ampuls) USEPA accepted for wastewater analyses** Using Powder Pillows

1,5-Diphenylcarbohydrazide Method* (Powder Pillows or AccuVac Ampuls) USEPA accepted for wastewater analyses** Using Powder Pillows Method 8023 CHROMIUM, HEXAVALENT (0 to 0.60 mg/l Cr 6+ ) For water and wastewater 1,5-Diphenylcarbohydrazide Method* (Powder Pillows or AccuVac Ampuls) USEPA accepted for wastewater analyses** Using Powder

More information

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document By the Authority Vested By Part 5 of the United States Code 552(a) and Part 1 of the Code of Regulations 51 the attached document has

More information

Evaluation copy. Chloride and Salinity. Computer INTRODUCTION

Evaluation copy. Chloride and Salinity. Computer INTRODUCTION Chloride and Computer 15 INTRODUCTION Chloride Chloride, in the form of the Cl ion, is one of the major inorganic anions, or negative ions, in saltwater and freshwater. It originates from the dissociation

More information

STEPS TO DMR-QA SUCCESS. BY: Marcy Bolek

STEPS TO DMR-QA SUCCESS. BY: Marcy Bolek STEPS TO DMR-QA SUCCESS BY: Marcy Bolek marcy@alloway.com Presentation Overview Purpose of DMR-QA Preparation DMR-QA Ethics Analysis Ammonia BOD Residual Chlorine ph TSS Quality Control Reporting What

More information

Approved for NPDES (Editorial Revision 1978) Silica, Dissolved (Colorimetric)

Approved for NPDES (Editorial Revision 1978) Silica, Dissolved (Colorimetric) METHOD #: 370.1 TITLE: Approved for NPDES (Editorial Revision 1978) Silica, Dissolved (Colorimetric) ANALYTE: Silica, SiO 2 INSTRUMENTATION: Spectrophotometer STORET No. Dissolved 00955 1.0 Scope and Application

More information

Instruction Manual. HI HI Portable ph/ec/tds/ C Meters

Instruction Manual. HI HI Portable ph/ec/tds/ C Meters Instruction Manual HI 9812-0 HI 9812-5 Portable ph/ec/tds/ C Meters w w w. h a n n a i n s t. c o m Dear Customer, Thank you for choosing a Hanna product. Please read this instruction manual carefully

More information

3. Fill a 50-mL. 2. Press: 1 ENTER The display will show mg/l, Al and the ZERO icon.

3. Fill a 50-mL. 2. Press: 1 ENTER The display will show mg/l, Al and the ZERO icon. ALUMINUM (0 to 0.80 mg/l) Aluminon Method * Method 8012 For water and wastewater 1. Enter the stored program number for aluminum (Al). Press: PRGM The display will show: PRGM? Note: Adjust the ph of stored

More information

Phosphorus, Reactive (Orthophosphate)

Phosphorus, Reactive (Orthophosphate) Phosphorous, Reactive (Orthophosphate),TNTplus 846, 10214 Phosphorus, Reactive (Orthophosphate) DOC316.53.01120 Molybdovanadate Method Method 10214 5.0 to 90.0 mg/l PO 3 4 or (1.6 to 30 mg/l PO 4 P) TNTplus

More information

Water Analysis Kit for Home Brewers

Water Analysis Kit for Home Brewers Brew your best batch EVERY time! TM Water Analysis Kit for Home Brewers Code 7188-01 PLUS Scan for direct access to LaMotte Brewlab web page. For further tips and information, go to instructions/tips tab.

More information

An Oxidation-Reduction Titration: The Reaction of Fe 2+ and Ce 4+

An Oxidation-Reduction Titration: The Reaction of Fe 2+ and Ce 4+ An Oxidation-Reduction Titration: The Reaction of Fe 2+ and Ce 4+ LAB ADV COMP 8 From Advanced Chemistry with Vernier, Vernier Software & Technology, 2004 INTRODUCTION A titration, as you recall, is a

More information

DR/4000 PROCEDURE. OXYGEN DEMAND, Chemical. Colorimetric Measurement

DR/4000 PROCEDURE. OXYGEN DEMAND, Chemical. Colorimetric Measurement Method 8000 DR/4000 PROCEDURE Reactor Digestion Method* (0 to 150.0 mg/l COD) Scope and Application: For water, wastewater and seawater; digestion is required; USEPA Approved** for wastewater analyses.

More information

Oxygen Demand, Chemical

Oxygen Demand, Chemical Oxygen Demand, Chemical DOC316.53.01101 Manganese III Reactor Digestion Method (with chloride removal) Method 10067 30 to 1000 mg/l COD Mn Test N Tube Vials Scope and application: For water and wastewater

More information

Methods and Materials Manual Of Ion Chromatography Using DX-100 IC. Collecting Samples. LDPE bottle with cap (pre-rinsed with DI water)

Methods and Materials Manual Of Ion Chromatography Using DX-100 IC. Collecting Samples. LDPE bottle with cap (pre-rinsed with DI water) Methods and Materials Manual Of Ion Chromatography Using DX-100 IC Collecting Samples Field Material: LDPE bottle with cap (pre-rinsed with DI water) permanent marker thermometer field notebook 1. Record

More information

A natural impurity in water. The test will measure the amount of free oxygen "gas" dissolved in your water sample in ppm (mg/l).

A natural impurity in water. The test will measure the amount of free oxygen gas dissolved in your water sample in ppm (mg/l). What is? A natural impurity in water. The test will measure the amount of free oxygen "gas" dissolved in your water sample in ppm (mg/l). Aquatic animals such as fish and zooplankton, can not use the oxygen

More information

QAM-I-113 Operation and Calibration of the Dissolved Oxygen Meter

QAM-I-113 Operation and Calibration of the Dissolved Oxygen Meter 1.0 Applicability and Purpose i. This procedure applies to the operation and calibration of the YSI Model 5100 dissolved oxygen meter or equivalent and associated probe. By performing the calibration procedure

More information

HACH TU 5 SERIES TURBIDIMETERS SOP: COMPARISON LAB AND PROCESS. The next standard in the evolution of turbidity

HACH TU 5 SERIES TURBIDIMETERS SOP: COMPARISON LAB AND PROCESS. The next standard in the evolution of turbidity HACH TU 5 SERIES TURBIDIMETERS SOP: COMPARISON LAB AND PROCESS The next standard in the evolution of turbidity COMPARE TU5 LAB AND PROCESS GUIDE & TROUBLESHOOT To guarantee that lab and process readings

More information

Typical Experiments. How it Works

Typical Experiments. How it Works ph Sensor DT016A The ph sensor can be connected to the Nova5000, MultiLogPRO or TriLink data loggers. The ph sensor is capable of measuring the entire range of 0-14 ph and is used for various experiments

More information

HYDROLOGY INVESTIGATION Electrical Conductivity. What is Conductivity?

HYDROLOGY INVESTIGATION Electrical Conductivity. What is Conductivity? What is Conductivity? Conductivity of a water sample is a measure of its ability to carry an electric current. The more impurities (total dissolved solids) in water, the greater its electrical conductivity.

More information

EXPERIMENT 5. Molecular Absorption Spectroscopy: Determination of Iron with 1,10-Phenanthroline

EXPERIMENT 5. Molecular Absorption Spectroscopy: Determination of Iron with 1,10-Phenanthroline EXPERIMENT 5 Molecular Absorption Spectroscopy: Determination of Iron with 1,10-Phenanthroline UNKNOWN Submit a clean, labeled 100-mL volumetric flask to the instructor so that your unknown iron solution

More information

Water Quality Consists of these Eight Important Components:

Water Quality Consists of these Eight Important Components: Water Quality Consists of these Eight Important Components: ph The H in ph stands for Hydrogen and the p stands for potential or the power of that Hydrogen, the concentration of hydrogen ions in a solution.

More information

STP (off-line) Water Technologies & Solutions analytical procedure. colorimetric method DCA reagent 1 (5/5/1) 2 to 15 mg/l as STP

STP (off-line) Water Technologies & Solutions analytical procedure. colorimetric method DCA reagent 1 (5/5/1) 2 to 15 mg/l as STP Water Technologies & Solutions analytical procedure AP901 STP (off-line) colorimetric method DCA reagent 1 (5/5/1) 2 to 15 mg/l as STP scope and application: for water analysis. Test preparation instrument

More information

Oxygen Demand, Chemical

Oxygen Demand, Chemical , TNTplus 820, 10211 DOC316.53.01103 USEPA Reactor Digestion Method Method 10211 ULR (1 60 mg/l COD) TNTplus 820 Scope and Application: For wastewater, process water, surface water, and cooling water;

More information

Hach Method Luminescence Measurement of Dissolved Oxygen (LDO ) in Water and Wastewater

Hach Method Luminescence Measurement of Dissolved Oxygen (LDO ) in Water and Wastewater Hach Method 10360 Luminescence Measurement of Dissolved Oxygen (LDO ) in Water and Wastewater LED Photo Diode LED Hach Method 10360 Luminescence Measurement of Dissolved Oxygen in Water and Wastewater

More information

Oyster ph - Conductivity - TDS - Temperature Meters

Oyster ph - Conductivity - TDS - Temperature Meters User's Guide Oyster ph - Conductivity - TDS - Temperature Meters Models 34135-A 341350 341350-P 34145A 341450 341450-P INTRODUCTION Congratulations on your purchase of the Extech's Oyster Series meter

More information

DR/4000 PROCEDURE. IRON, Ferrous. Using Powder Pillows

DR/4000 PROCEDURE. IRON, Ferrous. Using Powder Pillows DR/4000 PROCEDURE Method 8146 Powder Pillows or AccuVac Ampuls 1,10 Phenanthroline Method* (0 to 3.000 mg/l) Scope and Application: For water, wastewater and seawater. The estimated detection limit for

More information

Human CRP Precoated ELISA kit

Human CRP Precoated ELISA kit Absorption 450 nm CL76153K-48/CL76153K-96 Human CRP Precoated ELISA kit Product Description: 2.4 The Human CRP ELISA kit is to be used for the in-vitro quantitative determination of CRP in human serum,

More information

USEPA 1,2 Bicinchoninate Method 3 Method 8506 (CuVer 1) and Method 8026 (CuVer 2) 0.04 to 5.00 mg/l Cu Powder Pillows or AccuVac Ampuls

USEPA 1,2 Bicinchoninate Method 3 Method 8506 (CuVer 1) and Method 8026 (CuVer 2) 0.04 to 5.00 mg/l Cu Powder Pillows or AccuVac Ampuls Copper DOC316.53.01039 USEPA 1,2 Bicinchoninate Method 3 Method 8506 (CuVer 1) and Method 8026 (CuVer 2) 0.04 to 5.00 mg/l Cu Powder Pillows or AccuVac Ampuls Scope and application: For water, wastewater

More information

seven mm screw top bottles Vernier computer interface shallow pan Vernier Dissolved Oxygen Probe scissors Logger Pro

seven mm screw top bottles Vernier computer interface shallow pan Vernier Dissolved Oxygen Probe scissors Logger Pro Primary Productivity AP Bio Lab 12B Name: Date: Oxygen is vital to life. In the atmosphere, oxygen comprises over 20% of the available gases. In aquatic ecosystems, however, oxygen is scarce. To be useful

More information

Turbidity Meter Instruction Manual

Turbidity Meter Instruction Manual Turbidity Meter 860040 Instruction Manual TABLE OF CONTENTS INTRODUCTION...2 FEATURES...3 MATERIALS SUPPLIED...3 FRONT PANEL DESCRIPTION...4 SET UP...5 CALIBRATION...6 MEASUREMENT PROCEDURES...8 ZERO ADJUSTMENT...9

More information

Parts Per Million Analysis (PPM)

Parts Per Million Analysis (PPM) Parts Per Million Analysis (PPM) Randy Sharpe, H2 Sciences Inc. What Is "PPM"? PPM is an acronym for "Parts Per Million", a unit of measure that describes how much of one substance (called the "solute")

More information

USEPA 1 Periodate Oxidation Method 2 Method to 20.0 mg/l Mn (HR) Powder Pillows

USEPA 1 Periodate Oxidation Method 2 Method to 20.0 mg/l Mn (HR) Powder Pillows Manganese DOC316.53.01058 USEPA 1 Periodate Oxidation Method 2 Method 8034 0.1 to 20.0 mg/l Mn (HR) Powder Pillows Scope and application: For soluble manganese in water and wastewater. 1 USEPA Approved

More information

Instruction Manual HI Portable Microprocessor Turbidity Meter

Instruction Manual HI Portable Microprocessor Turbidity Meter Instruction Manual HI 93703 Portable Microprocessor Turbidity Meter 1 PRELIMINARY EXAMINATION Dear Customer, Thank you for choosing a HANNA instruments product. Please read this instruction manual carefully

More information

z /zr/r/ lnitials: tv4'wl z"z{-t L 3 /, /tt Operation and Calibration of the Dissolved Oxygen Meter Revision 5 Approval: Laboratory Man ager/lqao/rso

z /zr/r/ lnitials: tv4'wl zz{-t L 3 /, /tt Operation and Calibration of the Dissolved Oxygen Meter Revision 5 Approval: Laboratory Man ager/lqao/rso QAM.I-1 13 Operation and Calibration of the Dissolved Oxygen Meter Revision 5 Approval: Laboratory Man ager/lqao/rso Date z /zr/r/ Date z"z{-t L Effective date: Renewal date: * 1r i l: 3 /, /tt lnitials:

More information

Dimethylphenol Method Method LR (0.23 to mg/l NO 3 N or 1.00 to mg/l NO 3 ) TNTplus 835

Dimethylphenol Method Method LR (0.23 to mg/l NO 3 N or 1.00 to mg/l NO 3 ) TNTplus 835 LR, 10206 DOC316.53.01070 Dimethylphenol Method Method 10206 LR (0.23 to 13.50 mg/l NO 3 N or 1.00 to 60.00 mg/l NO 3 ) TNTplus 835 Scope and Application: For wastewater, drinking water, surface water

More information

Water Quality Lab Packet Environmental Science

Water Quality Lab Packet Environmental Science Name Water Quality Lab Packet Environmental Science Instructions: 1. READ. Before measuring anything, you need to read the information provided at each station. 2. PREDICT. Using a scale of 1-10, 1 being

More information

Demonstration of Relationship between CO2 and ph Instructions

Demonstration of Relationship between CO2 and ph Instructions Demonstration of Relationship between CO2 and ph Instructions Note - If you are interested in running the demonstration in class, follow the Leading the Demonstration in Class instructions below to run

More information

Nitrogen, Simplified TKN (s-tkn )

Nitrogen, Simplified TKN (s-tkn ) Nitrogen, Simplified TKN (s-tkn ) DOC316.53.01258 s-tkn Method Method 10242 1 0 to 16 mg/l TKN TNTplus 880 Scope and application: For water and wastewater. Digestion is required. 1 USEPA approved for water

More information

HYDRA Ammonium & HYDRA Nitrate Analyzers

HYDRA Ammonium & HYDRA Nitrate Analyzers HYDRA Ammonium & HYDRA Nitrate Analyzers Presented by: Steve Rupert Sr. Product Manager December, 2009 ELECTRO-CHEMICAL DEVICES HYDRA Overview The HYDRA Analyzers are a Family of Nitrogen Analyzers C22

More information

Vernier Software & Technology

Vernier Software & Technology Vernier Software & Technology ORDER CODE U.S. EDUCATOR PRICE Product Page Download Manual PDF The Conductivity Probe can be used to measure either solution conductivity or total ion concentration of aqueous

More information

Module 2, Add on Lesson Turbidity Sensor. Teacher. 90 minutes

Module 2, Add on Lesson Turbidity Sensor. Teacher. 90 minutes Module 2, Add on Lesson Turbidity Sensor Teacher 90 minutes Purpose Construct a sensor to measure the turbidity of water Graph data and reason about curves and linear relationships Calibrate the turbidity

More information

FIELD PROCEDURE: TURBIDITY

FIELD PROCEDURE: TURBIDITY FIELD PROCEDURE: TURBIDITY EQUIPMENT NEEDED: Note: Don t leave this equipment out in freezing temperatures unless you re using it. If air temperatures are just below freezing, keep the turbidity standards

More information

Oxygen. Oxygen is one of the fundamental resources required by life forms on Earth. Aquatic ecosystems have a wide assortment of life forms.

Oxygen. Oxygen is one of the fundamental resources required by life forms on Earth. Aquatic ecosystems have a wide assortment of life forms. Oxygen Oxygen is one of the fundamental resources required by life forms on Earth. Aquatic ecosystems have a wide assortment of life forms. Oxygen is also required for some natural chemical decays. What

More information

LabQuest INTRODUCTION

LabQuest INTRODUCTION LabQuest 2 INTRODUCTION Water contains both hydrogen ions, H +, and hydroxide ions, OH. The relative concentrations of these two ions determine the value. 1 Water with a of 7 has equal concentrations of

More information

Tracer TM Total Chlorine PockeTester Code 1740

Tracer TM Total Chlorine PockeTester Code 1740 Tracer TM Total Chlorine PockeTester Code 1740 Water Testing Leader Since 1919! WARNING! This set contains chemicals that may be harmful if misued. Read cautions on individual containers carefully. Not

More information

Dissolved oxygen. What is it and how does it affect my farm?

Dissolved oxygen. What is it and how does it affect my farm? Dissolved oxygen What is it and how does it affect my farm? Oxygen is crucial in any farming system, as it is the very basis of sustaining life. Dissolved oxygen, or D.O. as it is commonly called, is the

More information

Oxygen Demand, Chemical

Oxygen Demand, Chemical , TNTplus 823, 10212 DOC316.53.01104 USEPA Reactor Digestion Method Method 10212 UHR (250 15,000 mg/l COD) TNTplus 823 Scope and Application: For wastewater and process waters; digestion is required Test

More information

USER MANUAL. ExStik CL200A. Waterproof Total Residual Chlorine Tester. Patent Pending. Additional User Manual Translations available at

USER MANUAL. ExStik CL200A. Waterproof Total Residual Chlorine Tester. Patent Pending. Additional User Manual Translations available at USER MANUAL ExStik CL200A Waterproof Total Residual Chlorine Tester Patent Pending Additional User Manual Translations available at www.extech.com Introduction Congratulations on your purchase of the ExStik

More information

Operating Manual for Digital ph-meter GPHR 1400

Operating Manual for Digital ph-meter GPHR 1400 Operating Manual for Digital ph-meter GPHR 1400 Specification Measuring range: Resolution: ph-electrode: conductivity Input resistance: Accuracy (unit only): Working temperature: Nominal temperature: Power

More information

Silage DNA extraction procedure

Silage DNA extraction procedure Silage DNA extraction procedure Using the Fast DNA SPIN Kit for Soil by MP Biomedicals, LLC This Procedure is accompanied by a Video demonstration that can be downloaded at: http://animalscience.ucdavis.edu/faculty/robinson/videos.html

More information

EVENT: Perform the following steps of a BOD analysis using the YSI ProOBOD Optical BOD probe.

EVENT: Perform the following steps of a BOD analysis using the YSI ProOBOD Optical BOD probe. Operations Challenge 2014 Laboratory Event EVENT: Perform the following steps of a BOD analysis using the YSI ProOBOD Optical BOD probe. General NOTES: 1. Team Captain tells the Lead judge they are ready

More information

LAD B3 (pg! 1 of 6! ) Analysis by Redox Titration Name Per

LAD B3 (pg! 1 of 6! ) Analysis by Redox Titration Name Per LAD B3 (pg! 1 of 6! ) Name Per Introduction As you know, one common type of reaction in chemistry is oxidation-reduction. It involves the transfer of electrons from one species to another. Atoms undergo

More information

To whom it may concern, February 21, 2011

To whom it may concern, February 21, 2011 To whom it may concern, February 21, 2011 The following is an evaluation of the GEMS grant that I received to start my own C-MORE lab kits, specifically ocean acidification, plankton, marine debris and

More information

Introduction The Van London Co. Cadmium Ion Electrode is used to measure Cadmium ions in aqueous solutions.

Introduction The Van London Co. Cadmium Ion Electrode is used to measure Cadmium ions in aqueous solutions. GENERAL INSTRUCTIONS Introduction The Van London Co. Cadmium Ion Electrode is used to measure Cadmium ions in aqueous solutions. Required Equipment 1. An ion meter 2. Cadmium Ion Sensing Electrode 3. Polishing

More information

TB100 Portable Turbidity Meter Instruction Manual

TB100 Portable Turbidity Meter Instruction Manual TB100 Portable Turbidity Meter Instruction Manual BANTE INSTRUMENTS CO., LTD TB100 Portable Turbidity Meter 1 Introduction Thank you for selecting the TB100 portable turbidity meter. This manual provides

More information

Thermo Scientific Orion Dissolved Oxygen Probes

Thermo Scientific Orion Dissolved Oxygen Probes Thermo Scientific Orion Dissolved Oxygen Background Dissolved oxygen is often referred to as DO. It is a measure of the amount of oxygen in a solution. While some gases react chemically with water to form

More information

Salinity Sensor PS Salinity Sensor Probe. Salinity Sensor Probe, 10 X, Conductivity/Temperature PASPORT Extension Cable

Salinity Sensor PS Salinity Sensor Probe. Salinity Sensor Probe, 10 X, Conductivity/Temperature PASPORT Extension Cable Instruction Sheet 012-10546A *012-10546* Salinity Sensor PS-2195 Six pin mini- DIN connector Salinity Sensor Temperature Input port Conductivity Input port Salinity Sensor Probe Temperature Input plug

More information

CBF Water Quality Interactive Map

CBF Water Quality Interactive Map CBF Water Quality Interactive Map Student and adult groups that take part Often, they measure the water chemistry to evaluate the be doing it when you come out with us! By compiling these points on a map

More information

User's Guide. ExStik CL200A. Waterproof Total Residual Chlorine Tester Patent Pending

User's Guide. ExStik CL200A. Waterproof Total Residual Chlorine Tester Patent Pending User's Guide ExStik CL200A Waterproof Total Residual Chlorine Tester Patent Pending Introduction Congratulations on your purchase of the ExStik CL200 chlorine tester. The CL200 is a revolutionary first

More information

Method 8017 (0.7 to 80.0 µg/l) Powder Pillows. Scope and Application: For water and wastewater; digestion is required to determine total cadmium.

Method 8017 (0.7 to 80.0 µg/l) Powder Pillows. Scope and Application: For water and wastewater; digestion is required to determine total cadmium. , 8017 Dithizone Method 1 DOC316.53.01012 Method 8017 (0.7 to 80.0 µg/l) Powder Pillows Scope and Application: For water and wastewater; digestion is required to determine total cadmium. 1 Adapted from

More information

TN100 PortableTurbidity Meter. Instruction Manual.

TN100 PortableTurbidity Meter. Instruction Manual. TN100 PortableTurbidity Meter Instruction Manual. Table of Contents 1. Overview -------------------------------------------------------------------------------------- 1 2. Technical Specifications --------------------------------------------------------------

More information

INSTRUCTION MANUAL. ba76126e01 03/2013. photoflex Series TEST INSTRUCTIONS AND SPECIFICATIONS, PROG. V 2.02 Y

INSTRUCTION MANUAL. ba76126e01 03/2013. photoflex Series TEST INSTRUCTIONS AND SPECIFICATIONS, PROG. V 2.02 Y INSTRUCTION MANUAL photoflex Series TEST INSTRUCTIONS AND SPECIFICATIONS, PROG. V 2.02 Y photoflex Series For the most recent version of the manual, please visit www.ysi.com. Contact Copyright YSI 1725

More information

With Discovery Incubator.

With Discovery Incubator. TEST KIT POTABLE WATER With Discovery Incubator. FEATURES & BENEFITS 8 dip slide capacity Compact and convenient design Weighs only 1.2 kg Temperature range: ambient +5 to 45 C Accurate temperature control

More information

Chemical Oxygen Demand

Chemical Oxygen Demand HI83314 Multiparameter Photometer with COD for Wastewater with Digital ph Electrode Input HI83314 photometer measures 10 different key wastewater quality parameters using 20 different methods that allow

More information

OPERATIONS CHALLENGE LABORATORY PROCEDURE 2019

OPERATIONS CHALLENGE LABORATORY PROCEDURE 2019 OPERATIONS CHALLENGE LABORATORY PROCEDURE 2019 Version 1.21.19 Goal Analyzing and determining total suspended solids, conductivity / TDS and performing solids mass balance across a treatment system. Introduction

More information

Multiparameter. portable. ph/orp/ise, EC/TDS/ Resistivity/Salinity/Seawater σ, Turbidity, DO, Temperature and Atmospheric Pressure

Multiparameter. portable. ph/orp/ise, EC/TDS/ Resistivity/Salinity/Seawater σ, Turbidity, DO, Temperature and Atmospheric Pressure HI9829 GPS Meters ph/orp/ise, EC/TDS/ Resistivity/Salinity/Seawater σ, Turbidity, DO, Temperature and Atmospheric Pressure Logging Logging from probe or meter Fast Tracker Tag Identification System Sensor

More information

Measuring ORP : Tips, Cautions and Limitations

Measuring ORP : Tips, Cautions and Limitations Measuring ORP : Tips, Cautions and Limitations Introduction and Basic Theory As described in Standard Methods for the Examination of Water and Wastewater (Section 2580 B.), ORP is a potentiometric measurement

More information