PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION

Size: px
Start display at page:

Download "PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION"

Transcription

1 Hand Motion Shaker MPA-3 Operation Manual PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION Hagavish st. Israel Tel: , Fax: mrc@mrclab.com MRC.12.16

2 Contents I. SUMMARY II.TECHNICAL SPECIFICATIONS III. STRUCTURE IV. OPERATION V. NOTICES VI. MAINTENANCE VII. SUPPLEMENT

3 Warning: The operator shall be clear that the protection function of the instrument may be weakened if the instrument is not used in the method required by the manufacturer. Warning: Please handle with care all kinds of solution used in accordance with laboratory safety instructions and consult relevant safety documents of corresponding substances. Make sure to wear lab coat, goggles and rubber gloves whenever you make a test. Please be careful not to be burnt while handling hot reagent. Warning: Risk of electric shock. Only professionals can open the instrument cover and panel.

4 I. SUMMARY According to the definition in physical chemistry, the melting point of a substance means the temperature at which it changes from solid state into liquid state. In the field of organic chemistry, melting point test is the basic method to identify the nature character of a substance, and is one of the important means to measure purity. Model MPA-3 Automatic Melting Point Apparatus adopts excellent ergonomic design, color LCD touching screen, photoelectric inspection and other advanced technology and can test simultaneously three different or identical samples. It has the functions of automatic initial temperature and final temperature display, automatic record of melting process curve and etc. Linear correction platinum resistor is employed for temperature system of the apparatus as its detection component and it is characterized by accuracy, high efficiency and convenience, and so on. Model MPA-3 Automatic Melting Point Apparatus is widely used in chemical industry and medical research and is one of the necessary instruments for drug, flavor, dye and other organic crystalline substance production

5 II. TECHNICAL SPECIFICATIONS Melting point range Initial temperature" setting rate 7 min( ) 7.5 min( ) Temperature resolution 0.1 Repeatability 0.2 at heating rate of 0.2 /min Temperature accuracy ±0.4 (<200 ),±0.7 (<300 ) Temperature gradients /min(default 0.1 /min, 50 classes) Standard capillary size I.D. 1.0mm, O.D. 1.4mm Capacity per batch 3pc./batch (3 different samples) Control & Display model LCD color touch screen Power supply 220VAC±10% 50Hz Power 300W Net weight 17kg Dimensions 523mmX332mmX216mm - 2 -

6 III. STRUCTURE 1. Components Model MPA-3 is a melting point apparatus used for melting point measurement. The apparatus consists of setting, heating and testing systems. Structure of the instrument refers to the following figure (Figure 1). Figure 1 1. Indicator 2. Color LCD touching screen 3. Heating furnace 4. Heating furnace 5. Heating furnace 6. Power adapter 7. Power socket (with fuse in) 8. Power switch 9. Power output line 10. USB data interface 11. Cooling fan 12. Power input interface 13. Serial data interface 14. Furnace protective cover - 3 -

7 2. Functions 2.1 Welcome Interface When the instrument starts, a welcome screen as shown in Figure 2 Figure Function Interface After the instrument is started, it will automatically enter Function Interface, as shown in Figure 3. The function interface includes: Test, Data, Setting, Calibration and Help. Figure 3-4 -

8 2.3 Test Interface In Function Interface, click [Test] to enter sample-test select interface as shown in Figure 4. Users can select desired procedures to test the same samples or different samples. Figure Measurement of Same Samples Click [Same Sample] at sample test select interface and then click [OK] button to enter parameter setting interface for same sample test, as shown in Figure 5. Corresponding heating furnace can be selected and controlled by click the [Switch] under the numbers. If is displayed under the number, it indicates that the furnace is going to be put into work; and mark x indicates that the furnace will not be used. Click [Cancel] button to return to the previous interface. Figure 5 shows that No.2 Furnace will not be used this time

9 Figure 5 Click [Sample Name] in the test parameter setting interface to enter sample name setting interface as shown in Figure 6 for sample name setting. For example, if the sample is named A1, just input A1 (at most six letters or numbers can be used for sample name) with the screen keyboard. Click [OK] to complete name setting, or [Del] to delete the sample name, or [Cancel] to return to the previous interface. Figure 6-6 -

10 Click [Preheating Temperature] in the [Test Parameter] interface to enter preheating temperature setting interface, as shown in Figure 7 to set preheating temperature (preheating temperature setting range 50 ~400 ) for the sample. If preheating temperature is 76.0, click the keyboard to enter 76.0 and then click [Enter] button and [OK] button to complete the setting. Figure 7 Click [Heating Rate] item in the test parameter interface to enter heating rate setting interface, as shown in Figure 8 to set heating rate (settable heating rate is 0.1 /min~ 5.0 /min). For example, if desired heating rate is 1.0 /min, click the keyboard to enter 1.0 and then click [Enter] button and [OK] button to complete the setting

11 Figure 8 Figure 9 After completion of test parameter settings, system will return to the main test interface as shown in Figure 9. Click [OK] button to enter an interface as shown in Figure 10 and heating furnace begins pre-heating process

12 Figure 10 Note: In the figure shown above, TT represents current temperature of the heating furnace; TP means set pre-heating temperature; RH indicates set heating rate, TA indicates initial melting value of sample and TC is the final melting value of sample. When the temperature in the heating furnace reaches preset heating temperature and becomes stable as shown in Figure 11, put the test sample in. Click [Heating] item to start sample test. The test program can be checked via the melting curve with temperature as its abscissa and transmittance as its ordinate, as shown in the figure. Click [Return] to terminate the test and return to setting interface. Figure

13 When the sample test process is completed, initial melting value and final melting value of each sample are displayed in the interface. If you are going to save the data, click [Save] item in the interface to save the data into data storage table. After the data are saved, it will automatically drop to the preheating temperature. When the temperature in the heating furnace is stable at preheating temperature, the apparatus will start the next test of the same samples. Click [Stop] to stop the test at any time and return to the preheating temperature automatically. When the temperature in the heating furnace is stable at preheating temperature, the apparatus will start the next test of the same samples Measurement for Different Samples Click [Different Sample] at sample test select interface and then click [OK] button to enter parameter setting interface for different sample test, as shown in Figure 12. Referring to the setting methods as parameter settings for the same sample, users can control working status, sample name, preheating temperature and heating rate of each heating furnace. Figure 12 And then click [OK] button to preheat the heating furnace and enter an interface as shown in Figure 13. Click [Cancel] button to return to the previous interface. When the real-time temperature of heating furnace reaches preheating temperature, click [Heating] and [Stop] buttons for separate control of test processes of the three samples. Sample measuring

14 process and data storage process refer to those for [Same Sample]. Figure Data Inquiry In function interface, click [Data] button to enter data inquiry interface, as shown in Figure 14. Click [Page Up] and [Page Down] to check results of tested samples. Click [Return] button to return to the function interface. Figure Setting Interface In function interface, click [Setting] button to enter setting interface as shown in Figure

15 Figure 15 Click [Clock] button in the setting interface to enter time-date setting interface as shown in Figure 16. Move cursor with left and right narrows and change numbers with + and buttons. Click [OK] button to save the settings. Click [Cancel] to abandon the setting data. Click [Return] to return to the setting interface. Figure 16 Click [Communication] in the setting interface to enter communication setting interface, as shown in Figure 17. Select in the interface the port to connect with PC and then click [Return] to return to the main setting interface

16 Figure 17 Click [System Parameter] button in the setting interface to enter system parameter setting interface, as shown in Figure 18. Figure 18 Click the number of heating furnace, you can set parameters of the selected one separately, as shown in Figure 19. Click [Actual Tem] as indicated in the interface, system will display real-time temperature of current heating furnace. Click [Final Melt Tem] button, you can set final melting temperature value with the help of the numeric keys in the interface. For example, for sample naphthalene, you can input 85 as its final melting temperature, and then

17 put naphthalene into the furnace and the heating furnace begins to heat the sample. Click [Transmittance] to display the transmittance value of the tested sample in the controlled heating furnace. When the transmittance value is stable, check if the sample is completely melted at the temperature. If the sample is melted, you can put sample in again till it reaches the maximum transmittance value. Click [Save Transmittance] to save current transmittance of the sample in the furnace chamber at final melting point. Click [Final Melt Per]. Use the numeric keys under it to set the instrument and determine transmittance percent at final melting point. Click [Initial Melt Per]. Use the numeric keys under it to set the instrument and determine transmittance percent at initial melting point. Figure 19 If the set parameters are to be saved, click [Save Setting] to complete the storage. If you want to resume factory settings, click [Restore Factory Settings] and enter an interface as shown in Figure 20. Click [OK] to resume factory parameter settings. Click [Cancel] to return to the previous interface

18 Figure Calibration Interface In function interface, click [Calibration] button to enter calibration interface, as shown in Figure 21. Apparatus can be calibrated with this interface. Naphthalene, adipate and anthraquinone are used respectively as standard substances for melting point calibration. Set working status of heating furnace, melting point and preheating temperature of standard samples in this interface. Parameter setting method is the same as the one used for identical sample measurement. Figure

19 After setup of all the parameters in the interface as shown in Figure 22, click [OK] button to enter an interface as shown in Figure 23. The apparatus begins to preheat the furnace. When the heating furnace reaches the preheating temperature, put the standard sample in and click [Heating] button and the instrument begins melting point test on the standard sample. Figure 22 Figure 23 When the melting point of the standard sample is measured, click [Save] button to complete the first calibration. Take out the sample and click [Stop] button and wait the furnace to get to the preheating temperature and perform

20 the second calibration of the sample. Calibrate three times on each sample. The standard sample is defaulted by the instrument. After calibration, the system displays OK. Take out the sample and start calibration on another type of standard sample with the same test procedures. 2.7 Help Interface In function interface, click [Help] button to enter help interface as shown in Figure 24. If you need to learn more about the use of the instrument, click [Operation Instruction] to enter instruction interface as Figure 25 and Figure 26 show. Click [Return] to return to function interface. Figure 24 Figure

21 Figure 26 IV. OPERATION 1. Capillary Sample Preparation Put test sample into agate mortar, pulverize and dry it. Take an 800mm long, dry and clean glass tube and stand it vertically on a magnetic plate or a glass plate. Put the capillary with test sample into the glass tube and make it fall freely. Repeat the free fall eight times till sample powder (about 4~5mm tall) gathers tightly at the bottom. 2. Melting Point Apparatus Test 2.1 Turn on the power switch and warm up the apparatus about 20 minutes. 2.2 Enter function interface and select Test. Set proper preheating temperature and heating rate. 2.3 Preheat the furnace after confirmation. When it reaches preheating temperature and is stable, put sample in and test. 2.4 When the test is completed, initial melting temperature and final melting temperature are displayed. Record or save test data and take out tested sample. Click [Stop] and wait till the real-time temperature gets to preheating temperature. Repeat testing sample of the same group

22 2.5 Click Return to return and enter parameter setting interface after test to re-set parameters, and then test the next group of samples. 3. Apparatus Calibration Test 3.1 Start the instrument and warm up 20 minutes. 3.2 Select [Calibration] to enter calibration interface. 3.3 Select the heating furnace to be calibrated and input melting point of standard sample, preheating temperature and heating rate. 3.4 Click [OK] and begin preheating. When it reaches preheating temperature and is stable, put sample in and click [Heating] for melting point measurement. 3.5 After the test, the instrument will display initial and final melting temperatures. Click [Save] button to complete the first melting point calibration. Take out the sample and click [Stop] button and wait the furnace to get to the preheating temperature and perform the second calibration of the sample. Calibrate three times on each sample. The standard sample is defaulted by the instrument. After calibration, the system displays OK. Take out the sample and start calibration on another type of standard sample with the same test procedures. V. NOTICES 1. Test sample shall be baked as required, pulverized in dry and clean mortar. Tap the capillary with free falling method till sample is filled in tightly. Height of sample in the capillary shall not be less than 3mm. Height of sample within the same batch number shall be consistent to ensure accordant test results. 2. Turn on the instrument and then set preheating temperature. When it gets close to the preheating temperature, put sample in and operate as instructions. 3. Starting temperature set shall not exceed operation range of the apparatus, or the instrument may be damaged

23 4. Starting temperature of some samples has great effect on melting point test results, please operate according to instructions. We suggest capillary shall be inserted 5 minutes ahead. If linear heating rate is selected as 1.0 /min, starting temperature shall be 3-5 lower than melting point; if heating rate is 3.0 /min, starting temperature shall be 9-15 lower than melting point. Usually optimum test conditions are determined according to specific test conditions 5. Different linear heating rates lead to different test results. Proper instructions and standards shall be worked out. Usually the higher the heating rate, the test result will be bigger than the actual value. Measured melting point value at each heating rate can be uniformed by test-modified values. For the sample with unknown melting point, quick heating or higher heating rate is employed first; when it reaches initial melting point range, fine measurement is used. 6. When there is a reference sample, test the reference sample first. Select certain starting temperature and heating rate for comparative measurement as required. Initial and final melting point readings of reference sample can be used as examination evidence. 7. It is better to fill test sample into several capillaries once a time to avoid accidental error caused by capillaries and filling in of test sample. Test them respectively and delete the maximum and minimum values. Take the average value of the remaining readings as test results. 8. Capillary inserted in the sample chamber must be clean and standard, or dirt will deposit and it is easy to damage the capillary, making test impossible. 9. Handle with care the instrument in use. Keep it away from acid and alkaline solution. Prevent it from use in strong light. Please do not touch sample chamber during heating process in case of being burnt. 10. Take care while putting in or taking out capillaries to prevent them from being broken. If a capillary is broken or damaged in the sample chamber, please turn off power supply, cool the instrument, and then take it out. Employ a capillary pulverizer to mix the broken capillary. 11. If the instrument may not be used for a long time, please disconnect power cable

24 V. MAINTENANCE 1. Instrument shall be used in dry and ventilated rooms. Prevent it from water and humidity. The instrument uses three-core power plug. The grounding terminal shall be grounded to the earth and can not be replaced by a neutral line. 2. Capillary used by the instrument must be supplied by the Company. No manually drawn capillary can be used instead, in case it is too tight and may break. Fine capillaries shall be selected for test purpose. 3. If a capillary is broken in the sample chamber, please first turn off power supply, and cool the furnace, then pull out the tube seat. The tube seat shall be pulled out like this: one hand holds the heating furnace, and the other hand lifts the tube seat. Please do not just lift the tube seat. Improper force on the seat may damage the components of the instrument. Then use a capillary pulverizer to smash the broken capillary in the tube seat. Empty the shatter from the seat and then insert the tube seat again into the furnace. The notch on the tube seat shall be aligned with the flange on the furnace. Check the entire instrument after that and contact the service department of the company for repair if necessary. 4. If there is no display on the screen when the apparatus is turned on, or cooling fan does not work, or heating furnace does not heat, please shut off power supply and check if the fuse on the power socket of power adapter is burnt, and replace it in time. 5. Accuracy test: Accuracy test is performed according to rules stipulated by GBW13238 Melting Point Certified Reference Materials (1989, No.335) approved by State Bureau of Technical Supervision. The apparatus uses three of the reference materials: naphthalene (final melting point ), adipate (final melting point ) and anthraquinone (final melting point ), as examination standard (final melting point of sample subject to test report). 1.0 /min heating rate is selected, and starting temperature is set 5 higher than final melting point value. Repeat the test for five times as required. Delete the maximum and minimum values and the remaining three

25 values are used as test results. Average the three values and the deviation value shall be within stipulated range. 6. Instrument Calibration: Enter calibration interface and use reference material naphthalene, adipate and anthraquinone to calibrate the apparatus one by one. After test of the three substances, turn off the instrument and complete calibration. VII. SUPPLEMENT Symptoms Causes Troubleshooting Panel does not work when instrument is started. Function of each operation unit fails. Melting point tested is not accurate. Poor repeatability. Power is not supplied. Program-controlled heating system is damaged. Position of platinum resistor changes. There is dirt in the sample chamber. Inconsistent sample filling. Sizes of capillaries are different. Low sample purity. Re-plug power plug or turn on the main switch. Return to the factory for repair. Reset and fix the platinum resistor. Clean the sample chamber frequently. Make sample filling strictly as per instructions. Use high-purity sample and make sure the instrument is normal

64C Amplification Block. Instruction Manual

64C Amplification Block. Instruction Manual 64C Amplification Block Instruction Manual Caution: All rights reserved. Quidel Corporation reserves the right to modify this manual at any time without notice. Any part of the manual shall not be duplicated,

More information

Operational Manual. Spectrophotometer Model: SP-830 PLUS. Metertech Inc. Version 1.07

Operational Manual. Spectrophotometer Model: SP-830 PLUS. Metertech Inc. Version 1.07 Operational Manual Spectrophotometer Model: SP-830 PLUS Metertech Inc. Version 1.07 Metertech Inc. provides this publication as is without warranty of any kind, either express or implied, including, but

More information

Rebar Locator with Scanning Trolley TC110. Instruction Manual

Rebar Locator with Scanning Trolley TC110. Instruction Manual Rebar Locator with Scanning Trolley TC110 Instruction Manual CHAPTER I OVERVIEW... 1 1.1 BRIEF INTRODUCTION... 1 1.2 MAIN FUNCTIONS... 1 1.3 MAIN FEATURES... 2 CHAPTER II INSTRUMENT COMPOSITION AND MAINTENANCE

More information

Warning: This device complies with part 15 of the FCC Rules. Operation is subject to

Warning: This device complies with part 15 of the FCC Rules. Operation is subject to CR2 www.carnation-inc.com Warning: This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2)

More information

Operation Manual INCUBATOR2 Dual Dry Block Incubator

Operation Manual INCUBATOR2 Dual Dry Block Incubator Operation Manual INCUBATOR2 Dual Dry Block Incubator Hygiena USA info@hygiena.com 805-388-8007 / 888-494-4362 Hygiena International enquiries@hygiena.net +44 0 1923 818821 Foreword Thank you for purchasing

More information

Coefficient of Friction Tester

Coefficient of Friction Tester 1.Ambient Request & Examination 1.1 Lab condition, Temperature 23 2 C, Humidity 50 10%RH. Difference of temperature and humidity will result in obvious influence on test results. Especially in high-humidity

More information

Operation Manual. Dual Temperature Digital Dry Block Incubator. Catalog No. INCUBATOR Rev C. April 2016

Operation Manual. Dual Temperature Digital Dry Block Incubator. Catalog No. INCUBATOR Rev C. April 2016 Operation Manual Dual Temperature Digital Dry Block Incubator Catalog No. INCUBATOR2 Rev C. April 2016-1 - Thank you for your purchase. This manual contains operation information for the Dual Temperature

More information

User s Manual. KSV NIMA Dip Coaters

User s Manual. KSV NIMA Dip Coaters User s Manual KSV NIMA Dip Coaters DipCoater Revision 1.3 Table of Contents 1 INTRODUCTION -------------------------------------------------------------------------------------------- 1 2 KSV NIMA INTERFACE

More information

Bioneer Corporation 8-11, Munpyeongseo-ro, Daedeok-gu, Daejeon 34302, Republic of Korea Tel: Fax:

Bioneer Corporation 8-11, Munpyeongseo-ro, Daedeok-gu, Daejeon 34302, Republic of Korea Tel: Fax: Bioneer Corporation 8-11, Munpyeongseo-ro, Daedeok-gu, Daejeon 34302, Republic of Korea Tel: +82-42-930-8777 Fax: +82-42-930-8688 Email: sales@bioneer.co.kr Safety Warnings and Precautions ExiPrep TM Plus

More information

HOBO Plug Load Logger (UX ) Manual

HOBO Plug Load Logger (UX ) Manual HOBO Plug Load Logger (UX120-018) Manual The HOBO Plug Load logger is designed to monitor energy consumption of AC-powered plug in loads. This compact device can be used as a power meter with its built-in

More information

POWER FACTOR REGULATOR. Computer-12e-xx INSTRUCTION MANUAL ( M / 02A ) (c) CIRCUTOR S.A.

POWER FACTOR REGULATOR. Computer-12e-xx INSTRUCTION MANUAL ( M / 02A ) (c) CIRCUTOR S.A. POWER FACTOR REGULATOR Computer-12e-xx INSTRUCTION MANUAL ( M 981 606 / 02A ) (c) CIRCUTOR S.A. -------- POWER FACTOR REGULATOR COMPUTER- 12e --------- Page 2 POWER FACTOR REGULATOR COMPUTER- 12e 1.- POWER

More information

Lab 3: Calibration of a Melting Point Apparatus

Lab 3: Calibration of a Melting Point Apparatus Written by Danielle M. Solano & Jesse Bergkamp Department of Chemistry & Biochemistry California State University, Bakersfield Objectives By the end of this laboratory, you should have developed the skills

More information

Furnaces for Research and Laboratory. Testing Drying Ashing Tempering Preheating Sintering Firing Annealing

Furnaces for Research and Laboratory. Testing Drying Ashing Tempering Preheating Sintering Firing Annealing Furnaces for Research and Laboratory Testing Drying Ashing Tempering Preheating Sintering Firing Annealing Muffle Furnaces KL 03/11 KL 15/12 T max 1100 C and 1200 C Universal muffle furnaces for many laboratory

More information

BEFORE USING THE COUNTING SCALE

BEFORE USING THE COUNTING SCALE π H-1114, H-1115 H-1116, H-1117 INDUSTRIAL COUNTING SCALES 1-800-295-5510 uline.com WEIGH T UNIT WEIGH T lb/1000 Pc s u 1-800-295-5510 7 8 9 4 5 6 1 2 3 QTY UNIT PST UNIT PST ZERO TOTAL 0. CE TARE TOTAL

More information

PINLESS LCD MOISTURE METER WITH TRICOLOR BAR GRAPH USER S MANUAL

PINLESS LCD MOISTURE METER WITH TRICOLOR BAR GRAPH USER S MANUAL PINLESS LCD MOISTURE METER WITH TRICOLOR BAR GRAPH USER S MANUAL MMD7NP Please read this manual carefully and thoroughly before using this product. TABLE OF CONTENTS Introduction.................................

More information

OXYGEN UV+ ABSOLUTE RANGE

OXYGEN UV+ ABSOLUTE RANGE PURAPOOL OXYGEN UV+ ABSOLUTE RANGE OWNER S MANUAL OXYGEN UV MODE FACTORY SETTING 1. PARTS IDENTIFICATION LIST 1. 5. 2. 4. 3. 6. Figure 1.1 6. 1. Oxygen UV+ Power pack with timer option (x1) 2. Cell (x1)

More information

Table of contents. 1. Introduction

Table of contents. 1. Introduction 2 Year Warranty Table of contents 1. Introduction 1-1. Transportation 1-2. Preliminary steps 1-3. Initial setup 1-4. Important safety instructions 1-5. Maintenance, service and faults 2. Operation 2.1

More information

Operating instructions and installation information. METTLER TOLEDO MultiRange ID7-Form-XP application software

Operating instructions and installation information. METTLER TOLEDO MultiRange ID7-Form-XP application software T Operating instructions and installation information METTLER TOLEDO MultiRange ID7-Form-XP application software ID7-Form-XP Contents Contents Page 1 Introduction and assembly... 2 1.1 Introduction...

More information

Operation Manual for

Operation Manual for Operation Manual for PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 mrc@mrclab.com MRC.VER.01-8.13 Contents I. Summary Summary...1...1

More information

Best Selection for Your Business. Electronic Pricing Scale. User s Manual. (Model: TP-31)

Best Selection for Your Business. Electronic Pricing Scale. User s Manual. (Model: TP-31) Best Selection for Your Business Electronic Pricing Scale User s Manual (Model: TP-31) CATALOG 1. Foreword... 2 1.1. Introductions... 2 1.2. Main functions and features... 2 1.3. Specification... 2 1.4.

More information

OXYGEN UV+ ABSOLUTE RANGE

OXYGEN UV+ ABSOLUTE RANGE PURAPOOL OXYGEN UV+ ABSOLUTE RANGE OWNER S MANUAL OXYGEN UV MODE FACTORY SETTING 1. PARTS IDENTIFICATION LIST 1. 5. 2. 4. 3. 6. Figure 1.1 6. 1.! Oxygen UV+ Power pack with timer option (x1) 2.! Cell (x1)

More information

iloader LTS On-Board Scale for Front Loader (User Manual)

iloader LTS On-Board Scale for Front Loader (User Manual) 1 iloader LTS On-Board Scale for Front Loader (User Manual) Istanbul, 2018 2 Contents 1. User Interface Summary...3 2. Installation and Wiring...5 3. Calibration (Commissioning)...8 4. Material and Customer

More information

#201 Migun-Techno-World, 533 Yongsan-Dong, Yuseong-Gu, Daejeon-City, KOREA Tel: , 1688 Fax:

#201 Migun-Techno-World, 533 Yongsan-Dong, Yuseong-Gu, Daejeon-City, KOREA Tel: , 1688 Fax: #201 Migun-Techno-World, 533 Yongsan-Dong, Yuseong-Gu, Daejeon-City, KOREA 305-500 Tel: 82-42-861-1687, 1688 Fax: 82-42-861-1689 http://www.coxem.com E-mail: infor@coxem.com Table of contents Before using

More information

Normalization, Dilution and Transfer Methods for the Maxprep TM Liquid Handler Technical Manual

Normalization, Dilution and Transfer Methods for the Maxprep TM Liquid Handler Technical Manual Normalization, Dilution and Transfer Methods for the Maxprep TM Liquid Handler Technical Manual Instructions for use of Products AS9100, AS9101, AS9200, AS9201 TM530 11/17 Table of Contents 1 Description...2

More information

Manual Rotary Heat Sealers. Type: F108TX. List of content : Introduction 2. General description 3. Application 4. Safety precautions 5

Manual Rotary Heat Sealers. Type: F108TX. List of content : Introduction 2. General description 3. Application 4. Safety precautions 5 Manual Rotary Heat Sealers Type: F108TX List of content : Page: Introduction 2 General description 3 Application 4 Safety precautions 5 Transport and storage 6 Installation 7 First Set-up 8 Connection

More information

FIXED-WINDOW POWER DEMAND CONTROLLER BY ENERGY PULSES CVM-R8D-CPP INSTRUCTION MANUAL

FIXED-WINDOW POWER DEMAND CONTROLLER BY ENERGY PULSES CVM-R8D-CPP INSTRUCTION MANUAL FIXED-WINDOW POWER DEMAND CONTROLLER BY ENERGY PULSES CVM-R8D-CPP INSTRUCTION MANUAL ( M 981 316 / 99B ) (c) CIRCUTOR S.A. ------ Power demand controller by energy pulses - CVM-R8D-CPP ---------- Page

More information

ALTRONIC RESEARCH, INC.

ALTRONIC RESEARCH, INC. ALTRONIC RESEARCH, INC. P.O. BOX 249 YELLVILLE, ARKANSAS 72687-0249 U.S.A. CALORIMETRY MODEL 3500 TABLE OF CONTENTS CALORIMETRY MODEL 3500 SECTION PAGE Warranty... 3 Introduction... 4 I. Description and

More information

Ceramic Engineering 122 DTA and Glass Crystallization

Ceramic Engineering 122 DTA and Glass Crystallization CER122 (1) DTA/Crystallization Ceramic Engineering 122 DTA and Glass Crystallization OBJECTIVE The objectives of this lab are to teach the students how to use differential thermal analysis (DTA) to analyze

More information

Contents. 1. About this edition of user instructions

Contents. 1. About this edition of user instructions 2 Contents 1. About this edition of user instructions... 3 2. Safety precautions... 4 3. General information... 5 4. Getting started... 6 5. Operation... 7 6. Calibration... 9 7. Specifications... 11 8.

More information

Model BF 56 Standard-Incubators with forced convection

Model BF 56 Standard-Incubators with forced convection Model BF 56 Standard-Incubators with forced convection The BINDER incubator of the BF Avantgarde.Line series with forced convection is suitable for all gentle incubation applications, particularly under

More information

User Manual. XK315A1 water proof series Weighing Indicators. Applications: Electronic platform scale, weighing scale, etc.

User Manual. XK315A1 water proof series Weighing Indicators. Applications: Electronic platform scale, weighing scale, etc. User Manual XK315A1 water proof series Weighing Indicators Applications: Electronic platform scale, weighing scale, etc. 1 Introductions 1.1 Safety precautions WARNING! Don t use XK315A1 series indicators

More information

Gelovit 200 en English

Gelovit 200 en English Gelovit 200 en English Translation of the original instructions BA 85 981 / 02 334.00001... en Gelovit 200 This unit documentation is part of the unit and must be enclosed when selling or passing on the

More information

USER S MANUAL release 1.0

USER S MANUAL release 1.0 SLIGHT two RGB-W IP66 halogen colour changer USER S MANUAL release 1.0 This manual must be considered an integral part of the projector. BEFORE CONNECTING AND USING THE PROJECTOR, IT IS IMPORTANT TO READ

More information

Anyvib Vibration Test Chambers

Anyvib Vibration Test Chambers Anyvib Vibration Test Chambers WHY ANYVIB? Because anyvib is the easiest way to combine any kind of shaker with an environmental test chamber. anyvib is a new family of climatic chambers that can be interfaced

More information

AP5K-C Precision AC Double-Pulse Spot Welding Machine User s Manual Shenzhen Will-Best Electronics Co., Ltd

AP5K-C Precision AC Double-Pulse Spot Welding Machine User s Manual Shenzhen Will-Best Electronics Co., Ltd AP5K-C Precision AC Double-Pulse Spot Welding Machine User s Manual Shenzhen Will-Best Electronics Co., Ltd 1 Content 1. Introduction...3 1.1 Functions...3 1.2 Units of AP5K-C...4 2. The Initial Installation

More information

USER MANUAL. ExStik CL200A. Waterproof Total Residual Chlorine Tester. Patent Pending. Additional User Manual Translations available at

USER MANUAL. ExStik CL200A. Waterproof Total Residual Chlorine Tester. Patent Pending. Additional User Manual Translations available at USER MANUAL ExStik CL200A Waterproof Total Residual Chlorine Tester Patent Pending Additional User Manual Translations available at www.extech.com Introduction Congratulations on your purchase of the ExStik

More information

Models TJ30K TJ15K TJ6K TJ3K. Resolution Ratio 1g 0.5g 0.2g 0.1g Calibration graduation 10d 10d 10d 10d

Models TJ30K TJ15K TJ6K TJ3K. Resolution Ratio 1g 0.5g 0.2g 0.1g Calibration graduation 10d 10d 10d 10d Operating Instruction Manual of TJ Series Electronic counting balance ( for following models: TJ3K,TJ6K,TJ15K,TJ30K,TJ3KA,TJ6KA,TJ15KA, TJ30KA) 1. Overview TJ series electronic counting weighing apparatus

More information

Xxxxx Xxxxxx OM-CP-TEMP101A. Temperature Data Logger with 10 year battery life INSTRUCTION SHEET MQS5026/0117

Xxxxx Xxxxxx OM-CP-TEMP101A. Temperature Data Logger with 10 year battery life INSTRUCTION SHEET MQS5026/0117 Xxxxx Xxxxxx Temperature Data Logger with 10 year battery life INSTRUCTION SHEET MQS5026/0117 Shop online at omega.com SM e-mail: info@omega.com For latest product manuals: www.omegamanual.info Product

More information

OM-CP-QUADPROCESS 4-Channel Low Level DC Current Data Logger. OM-CP-OCTPROCESS 8-Channel Low Level DC Current Data Logger INSTRUCTION SHEET

OM-CP-QUADPROCESS 4-Channel Low Level DC Current Data Logger. OM-CP-OCTPROCESS 8-Channel Low Level DC Current Data Logger INSTRUCTION SHEET Xxxxx OM-CP-QUADPROCESS Xxxxxx 4-Channel Low Level DC Current Data Logger OM-CP-OCTPROCESS 8-Channel Low Level DC Current Data Logger INSTRUCTION SHEET MQS5087/1216 Shop online at omega.com SM e-mail:

More information

Installation & Calibration

Installation & Calibration Installation & Calibration Type to enter text ED4-SR SkidWeigh Plus System Lift Truck Onboard Weighing System - Active Session Recordings (Recycling Industry Application) Version: V1.02 General Installation

More information

Model KMF 115 Constant climate chambers with expanded temperature / humidity range

Model KMF 115 Constant climate chambers with expanded temperature / humidity range Model KMF 115 Constant climate chambers with expanded temperature / humidity range The BINDER KMF ensures absolutely constant test conditions throughout the testing area. A big advantage of this constant

More information

Melting Point M-565 Technical data sheet

Melting Point M-565 Technical data sheet Melting Point M-565 Technical data sheet The video camera equipped Melting Point M-565 for the automatic determination of melting and boiling points is designed for true, professional quality control.

More information

Manual. AGRETO Hydraulic Scale AGRETO electronics GmbH

Manual. AGRETO Hydraulic Scale AGRETO electronics GmbH Manual AGRETO Hydraulic Scale 22.04.2014 AGRETO electronics GmbH Content 1 Introduction... 3 2 Scope of Delivery... 3 3 Intended Use... 3 4 Security... 4 4.1 Safety Instructions for the Buyer... 4 4.2

More information

Instruction Manual HI Portable ph/ec/tds/ C Meters

Instruction Manual HI Portable ph/ec/tds/ C Meters Instruction Manual HI 9811-5 Portable ph/ec/tds/ C Meters www.hannainst.com 1 Dear Customer, Thank you for choosing a Hanna product. Please read this instruction manual carefully before using the meter.

More information

RPBS-1 PORTABLE SCALE

RPBS-1 PORTABLE SCALE PORTABLE SCALE User Manual Software version: 3.4 Date of issue: January 2006 This manual is published by ROTEM Computerized Controllers Ltd. All rights to this publication are reserved. No part of this

More information

Model BF 260 Incubators Avantgarde.Line with forced convection

Model BF 260 Incubators Avantgarde.Line with forced convection Model BF 260 Incubators Avantgarde.Line with forced convection The BINDER incubator of the BF Avantgarde.Line series with forced convection is suitable for all gentle incubation applications, particularly

More information

Manual Supplement. This supplement contains information necessary to ensure the accuracy of the above manual.

Manual Supplement. This supplement contains information necessary to ensure the accuracy of the above manual. Manual Supplement Manual Title: 4180, 4181 Users Supplement Issue: 4 Part Number: Web-Only Issue Date: 4/18 Print Date: March 2013 Page Count: 5 Revision/Date: This supplement contains information necessary

More information

Daker DK kva. Parallel installation quick start. Part. LE08927AA-09/15-01 GF

Daker DK kva. Parallel installation quick start. Part. LE08927AA-09/15-01 GF Daker DK 4.5-10 kva Parallel installation quick start Part. LE08927AA-09/15-01 GF Daker DK 4.5-10 kva UK ENGLISH 3 2 Daker DK 4.5-10 kva Index Introduction 2 1 Important Safety Instructions 2 2 Parallel

More information

Moving money forward UV MG

Moving money forward UV MG Moving money forward T This manual contains important information on safety measures and operational features, please read it carefully before operating your currency counter and keep it for future reference.

More information

INSTRUCTION MANUAL Wardrobe lift 717

INSTRUCTION MANUAL Wardrobe lift 717 English Coatlift 717 Doc. No: M717 Edition: 1 Date: 2009-04-02 INSTRUCTION MANUAL Wardrobe lift 717 Granberg Interior AB Box 6112 SE-600 06 Norrköping SWEDEN For more information, see: www.granberg.se

More information

The Biology CRO. Protein Staining System. For electric staining of proteins in mini polyacrylamide gels. Version 1.0 April 3, 2010.

The Biology CRO. Protein Staining System. For electric staining of proteins in mini polyacrylamide gels. Version 1.0 April 3, 2010. The Biology CRO Protein Staining System For electric staining of proteins in mini polyacrylamide gels Version 1.0 April 3, 2010 User Manual Table of Contents Product Contents 1 Product Specifications 3

More information

LED ROTATION BEAM LIGHT

LED ROTATION BEAM LIGHT LED ROTATION BEAM LIGHT MJ-1018(White) INSTRUCTION MANUAL Thank you for choosing our LED rotation beam light. For the sake of your safety, Please read and follow these instructions carefully and keep this

More information

PD6770 VANTAGEVIEW LOOP-POWERED PROCESS METER

PD6770 VANTAGEVIEW LOOP-POWERED PROCESS METER PD6770 VANTAGEVIEW LOOP-POWERED PROCESS METER 4-20 ma input 1 V drop (4 V with Backlight) 3½ Digits LCD, 1" High Loop-Powered Backlight Option HART Protocol Transparent Plastic NEMA 4X, IP65 Enclosure

More information

PHS-3D-01 Lab ph Meter. User s Guide. Shanghai San-Xin Instrumentation, Inc.

PHS-3D-01 Lab ph Meter. User s Guide. Shanghai San-Xin Instrumentation, Inc. PHS-3D-01 Lab ph Meter User s Guide Shanghai San-Xin Instrumentation, Inc. Table of Contents 1. Brief Introduction----------------------------------------------------------------------- 2. Technical Specifications

More information

Exceed IRELAND Electronic Postage Computing Scale

Exceed IRELAND Electronic Postage Computing Scale Exceed IRELAND Electronic Postage Computing Scale OPERATING MANUAL CONTENTS INTRODUCTION Specifications Preparing the scale for use Operating the Exceed postal scale KEYBOARD FUNCTIONS Basic services Special

More information

U28. Name of Parts 1. Drone. Blade Guard. Battery Cover. Drone Body Blade Motor. Body Light. Camera. Landing Gear. Blade Guard Installation Diagram

U28. Name of Parts 1. Drone. Blade Guard. Battery Cover. Drone Body Blade Motor. Body Light. Camera. Landing Gear. Blade Guard Installation Diagram Name of Parts 1. Drone Blade Guard Drone Body Blade Motor Body Light Camera Landing Gear Blade Guard Installation Diagram U8 TF Card Installation Diagram . Transmitter Power Switch Throttle/ Up and Down

More information

Automatic Key Cutting Machine - CONDOR XC-007 CONDOR. Automatic Key Cutting Machine XC-007. User Manual V1.2

Automatic Key Cutting Machine - CONDOR XC-007 CONDOR. Automatic Key Cutting Machine XC-007. User Manual V1.2 1 CONDOR Automatic Key Cutting Machine XC-007 User Manual V1.2 2 Table of Contents 1) Safety Precautions and Warnings --------------------------------------------3 2) General Information -------------------------------------------------------4

More information

Welding & Fabrication Tools Mini VaR Tool Kit

Welding & Fabrication Tools Mini VaR Tool Kit With Volta Tools You Can Never Go Wrong! Fast and simple belt installation. Unique and versatile design - compact, rugged and easy-to-use. Designed for both shop and field use. Light-weight construction.

More information

Operations Manual. For Research Use Only. PH: EM: WEB:

Operations Manual. For Research Use Only. PH: EM: WEB: Operations Manual For Research Use Only PH: 908-769-5555 EM: info@benchmarkscientific.com WEB: www.benchmarkscientific.com Thank you for choosing the BioClave Benchtop Sterilizer. Prior to operating this

More information

Model S-2200 Single Pocket Currency Discriminator Operating Manual

Model S-2200 Single Pocket Currency Discriminator Operating Manual Model S-2200 Single Pocket Currency Discriminator Operating Manual SEMACON MODEL S-2200 SINGLE POCKET CURRENCY DISCRIMINATOR This Operating Manual provides general guidelines for the operation of your

More information

Operating Manual. Professional Weighing Equipment. TB Series intelligent weighing machine. Touch Color Screen HIGH RESOLUTION BALANCE

Operating Manual. Professional Weighing Equipment. TB Series intelligent weighing machine. Touch Color Screen HIGH RESOLUTION BALANCE Professional Weighing Equipment Operating Manual TB Series intelligent weighing machine Touch Color Screen HIGH RESOLUTION BALANCE 1 Introduction What you should know about these Operating Instructions:

More information

Operate Manual. DA-1600i Stud Welding Machine. DA-2000i Stud Welding Machine. DA-2500i Stud Welding Machine

Operate Manual. DA-1600i Stud Welding Machine. DA-2000i Stud Welding Machine. DA-2500i Stud Welding Machine DA-1600i Stud Welding Machine DA-2000i Stud Welding Machine DA-2500i Stud Welding Machine Operate Manual KENT STUD WELDING CO., LTD 1 Content 1. Preface------------------------------------------------------------------------------

More information

POWER PINNER BENCH HAND WELDER 7250 OPERATOR S MANUAL

POWER PINNER BENCH HAND WELDER 7250 OPERATOR S MANUAL POWER PINNER BENCH HAND WELDER 7250 OPERATOR S MANUAL Copyright: November 20, 2018 Revised: Serial No. Gripnail Corporation An Employee Owned Company 97 Dexter Road East Providence, Rhode Island 02914-2045

More information

State of Nevada Department of Transportation Materials Division METHOD OF TEST FOR DETERMINING BITUMEN RATIO IN HOT MIX ASPHALT BY THE IGNITION METHOD

State of Nevada Department of Transportation Materials Division METHOD OF TEST FOR DETERMINING BITUMEN RATIO IN HOT MIX ASPHALT BY THE IGNITION METHOD State of Nevada Department of Transportation Materials Division METHOD OF TEST FOR DETERMINING BITUMEN RATIO IN HOT MIX ASPHALT BY THE IGNITION METHOD SCOPE This test method covers the determination of

More information

! CAUTION: Read and follow these instructions.

! CAUTION: Read and follow these instructions. OmniLog SAFETY AND INSTALLATION INSTRUCTIONS! CAUTION: Read and follow these instructions. Avoid Contact with Hazardous Live Parts WARNING: Risk of electrical shock. Refer servicing to qualified personnel.

More information

Automatic Isokinetic Sampler ST5

Automatic Isokinetic Sampler ST5 Automatic Isokinetic Sampler ST5 Automatic Isokinetic Sampler ST5 Description The ST5 is the next generation isokinetic sampler, developed to be efficient, rugged, accurate and user friendly. The perfect

More information

1. About this edition of the operating manual

1. About this edition of the operating manual Contents 1. About this edition of the operating manual... 2 2. Safety precautions... 3 3. General information... 4 4. Getting started... 5 5. Operation... 5 6. Calibration... 6 7. Specification... 8 8.

More information

TB100 Portable Turbidity Meter Instruction Manual

TB100 Portable Turbidity Meter Instruction Manual TB100 Portable Turbidity Meter Instruction Manual BANTE INSTRUMENTS CO., LTD TB100 Portable Turbidity Meter 1 Introduction Thank you for selecting the TB100 portable turbidity meter. This manual provides

More information

Determining Melting Temperature. Sample

Determining Melting Temperature. Sample Determining Melting Temperature Experiment 1 The melting temperature of a compound is the temperature at which it changes from a solid to a liquid. This is a physical property often used to help identify

More information

INSTRUCTION MANUAL WARNING: BEFORE USING THE MACHINE READ THE INSTRUCTION MANUAL CAREFULLY!

INSTRUCTION MANUAL WARNING: BEFORE USING THE MACHINE READ THE INSTRUCTION MANUAL CAREFULLY! INSTRUCTION MANUAL WARNING: BEFORE USING THE MACHINE READ THE INSTRUCTION MANUAL CAREFULLY! SECTION 1 : PRODUCT AND COMPANY IDENTIFICATION PRODUCT NAME : SYNONYMS : PRODUCT CODES : Weldability ARC140i

More information

INSTRUCTION MANUAL Wardrobe lift 718

INSTRUCTION MANUAL Wardrobe lift 718 Wardrobe lift - English Doc. No: M718 Edition: 2 Date: 2006-04-20 INSTRUCTION MANUAL Wardrobe lift 718 Granberg Interior AB Box 6112 SE-600 06 Norrköping SWEDEN Tel: +46 (0)11-19 77 50 Fax: +46 (0)11-12

More information

Melting Point Determination Application Note #1

Melting Point Determination Application Note #1 Application Note #1 Introduction A few basic guidelines must be carefully followed to avoid errors during melting point determinations with OptiMelt. The way in which the sample is prepared and the instrument

More information

DJ Switch 10F V2 ORDERCODE 50365

DJ Switch 10F V2 ORDERCODE 50365 DJ Switch 10F V2 ORDERCODE 5035 Congratulations! You have bought a great, innovative product from Showtec. The Showtec DJ Switch brings excitement to any venue. Whether you want simple plug-&-play action

More information

Whitening Lamp. Instructions for use. flaesh.com

Whitening Lamp. Instructions for use. flaesh.com Whitening Lamp Instructions for use flaesh.com Whitening System Latest LED technology Guided treatment function Determination of whitening shades achieved Cosmetic and medical tooth whitening 2flaesh.com

More information

LABORATORY INVESTIGATION

LABORATORY INVESTIGATION LABORATORY INVESTIGATION Seed Respiration Pea seeds, each, contain an embryo plant, a supply of nutrients and a protective covering called the seed coat. Assuming the temperature is high enough, germination

More information

Operating Instructions

Operating Instructions 028028 Operating Instructions Monarch 928 Stacker 23456 78901 Fishing Tackle $40.00 23456 78901 Fishing Tackle $40.00 23456 78901 Fishing Tackle $40.00 23456 78901 Fishing Tackle $40.00 TC0928OI Rev. AJ

More information

AA Portable Power Corp MSK 320B2 User Manual

AA Portable Power Corp    MSK 320B2 User Manual MSK 320B2 User Manual 1 Table of content 1 Overview:... 3 2 Precaution... 3 3 Key Technical Specification:... 4 4 Machine structure: (Fig.1)... 4 5 Operation instruction:... 5 6 General Troubleshooting...

More information

NOTES TABLE OF CONTENTS

NOTES TABLE OF CONTENTS NOTES TABLE OF CONTENTS 1.0 Measuring procedure.. 2 2.0 Design of the device 4 3.0 Calibration curves 5 4.0 Activation of the super user function.. 6 5.0 Changing batteries.. 7 6.0 Determination of the

More information

PURCHASE OF BATTERY TESTER ON SUPLY BASIS.

PURCHASE OF BATTERY TESTER ON SUPLY BASIS. Scope of Work 1. Introduction PURCHASE OF BATTERY TESTER ON SUPLY BASIS. PTCL intends to purchase battery testers on suply basis for preventive maintenance of batteries installed in its network. It is

More information

Engineering Guide For Zircoa s Foundry Crucibles

Engineering Guide For Zircoa s Foundry Crucibles Engineering Guide For Zircoa s Foundry Crucibles 1. Introduction Thank you for your interest in Zircoa crucibles. The information contained in this Engineering Guide has been compiled over a period of

More information

Coating Thickness Meter Data Sheet / Instructions

Coating Thickness Meter Data Sheet / Instructions Paint Test Equipment Coating Thickness Meter Data Sheet / Instructions Coating Thickness Meter Information The Paint Test Equipment Coating Thickness Meter easily measures all coatings on metallic substrates

More information

Coating Thickness Meter Coating Thickness

Coating Thickness Meter Coating Thickness Paint Test Equipment Coating Thickness Meter Coating Thickness Coating Thickness Meter Information The Paint Test Equipment Coating Thickness Meter easily measures all coatings on metallic Substrates using

More information

Battery Maintenance - + Section 6-2 HOW IT WORKS

Battery Maintenance - + Section 6-2 HOW IT WORKS Section 6-2 Battery Maintenance One of the important components of your unit is the storage battery. We depend on it from the moment of unit start up to shutdown, yet, it is surprising how many operators

More information

Tindo Karra. Installation and Specifications Manual. Introduction

Tindo Karra. Installation and Specifications Manual. Introduction Tindo Karra Installation and Specifications Manual Introduction Thank you for choosing a Tindo Solar Module. Background to our name TINDO Tindo is a word from the Kaurna warra Aboriginal language which

More information

TARGAS-1. Portable Photosynthesis System. Photosynthesis Soil Respiration Canopy Assimilation Environmental Monitoring

TARGAS-1. Portable Photosynthesis System. Photosynthesis Soil Respiration Canopy Assimilation Environmental Monitoring TARGAS-1 Portable Photosynthesis System Photosynthesis Soil Respiration Canopy Assimilation Environmental Monitoring TARGAS-1 Main Console CO 2 & H 2 O Gas Analysis The TARGAS-1 console is compact, lightweight

More information

Olivia. Dedicated Analytical Solutions

Olivia. Dedicated Analytical Solutions Olivia Olivia is a robust, simple-to-use and accurate olive analyser that allows small and medium olive processors to make informed decisions about the fat and moisture content of their raw materials.

More information

ecosphere water purifier 3in1 USER MANUAL

ecosphere water purifier 3in1 USER MANUAL ecosphere water purifier 3in1 USER MANUAL TABLE OF CONTENTS System Components 3 Function Overview 4 How to Set up Your Purifier 6 3in1 Cartridge Set Up 7 Initial Run 8 Use 8 Purified Water 8 Tap Water

More information

booster EX2 / EX3 Operating Manual Extrusion Welder Weldy AG Galileo-Strasse 10 CH-6056 Kaegiswil / Switzerland TECHSPANGROUP

booster EX2 / EX3 Operating Manual Extrusion Welder Weldy AG Galileo-Strasse 10 CH-6056 Kaegiswil / Switzerland TECHSPANGROUP booster EX2 / EX3 Extrusion Welder Operating Manual Weldy AG Galileo-Strasse 10 CH-6056 Kaegiswil / Switzerland TECHSPANGROUP Australia: Tel. 1-800 148 791 Fax. 1-800 148 799 www.weldy.com.au New Zealand:

More information

ZX70 & ZX80 SERIES PRECISION DECADE RESISTANCE BOX

ZX70 & ZX80 SERIES PRECISION DECADE RESISTANCE BOX ZX70 & ZX80 SERIES PRECISION DECADE RESISTANCE BOX OPERATING INSTRUCTIONS SUBJECT TO CHANGE WITHOUT NOTICE This manual superseded all previous versions please keep for future reference TINSLEY, A DIVISION

More information

Model BF 720 Incubators Avantgarde.Line with forced convection

Model BF 720 Incubators Avantgarde.Line with forced convection Model BF 720 Incubators Avantgarde.Line with forced convection The BINDER incubator of the BF Avantgarde.Line series with forced convection is suitable for all gentle incubation applications, particularly

More information

Collator Series 300/400 Operator Manual

Collator Series 300/400 Operator Manual Collator Series 300/400 Operator Manual January 2003 Part No. 90907 Table of contents 1. Introduction page 1.1 Operational safety 1-1 2. Installation 2.1 Installation 2-1 3. Getting to know the collator

More information

D-Series Warming Cabinet Unpacking and Setup

D-Series Warming Cabinet Unpacking and Setup Unpacking Instructions Receiving Requirements The customer is responsible for making sure the loading dock at their facility can accommodate a shipping carton approximately 70 inches (1.778 m) long and

More information

User s Guide. Contents. Advanced Biometric Access Control System. Fingerprint & Digital Lock. Fingerprint & Digital Lock

User s Guide. Contents. Advanced Biometric Access Control System. Fingerprint & Digital Lock. Fingerprint & Digital Lock Contents User s Guide Advanced Biometric Access Control System 1. Features ---------------------------------------------- 3 page 2. Specifications ----------------------------------------- 4 page 3. How

More information

Magner /35-3 Series Currency Counter. Operator's Manual

Magner /35-3 Series Currency Counter. Operator's Manual Magner 35-2003/35-3 Series Currency Counter Operator's Manual Magner 35-2003 / 35-3 Series Introduction. The MAGNER 35-2003 / 35-3 is the most advanced Currency Counter available today. MAGNER's Design

More information

Technical Information. to the. Constant Torque Gieseler Plastometer PF-22

Technical Information. to the. Constant Torque Gieseler Plastometer PF-22 DASFOS Czr, s.r.o. Technology-innovation center Bozkova 45 14, 702 00, Ostrava 2-Privoz, Czech Republic Phone: + 420 59 6612092 Fax: + 420 59 6612094 E-mail: dasfos@dasfos.com Web:http://www.dasfos.com

More information

AUTOMATIC CAPACITOR BANK FOR LOW VOLTAGE NETWORKS INSTRUCTION MANUAL. MIBATBT_05i

AUTOMATIC CAPACITOR BANK FOR LOW VOLTAGE NETWORKS INSTRUCTION MANUAL. MIBATBT_05i AUTOMATIC CAPACITOR BANK FOR LOW VOLTAGE NETWORKS INSTRUCTION MANUAL MIBATBT_05i COMMISSIONING AND OPERATION MANUAL FOR LOW VOLTAGE AUTOMATIC CAPACITOR BANKS 1. DESCRIPTION The automatic capacitor banks

More information

Model FD 115 Drying and heating chambers Avantgarde.Line with forced convection

Model FD 115 Drying and heating chambers Avantgarde.Line with forced convection Model FD 115 Drying and heating chambers Avantgarde.Line with forced convection A BINDER FD series Avantgarde.Line heating oven is always used when fast drying and sterilization is required. Thanks to

More information

Version Software User Guide

Version Software User Guide 06 Version 0.0. Software User Guide Long Range LLC. 6 Tannery Street Franklin, NH 05 Table of Contents Introduction Features... Included... Installation 5 Migrating Your Personal Access File... 7 Set Range

More information

PFBH room 341 Steris Autoclave Operation

PFBH room 341 Steris Autoclave Operation University of California San Diego Dept. of Bioengineering Standard Operating Procedure SOP000026 V0.5 PFBH room 341 Steris Autoclave Operation Issuing Office: Dept. of Bioengineering, JSOE Responsible

More information

VARIO XX28 Stainless Steel Time And Temperature Controlled Electric Model

VARIO XX28 Stainless Steel Time And Temperature Controlled Electric Model VARIO XX28 Stainless Steel Time And Temperature Controlled Electric Model Operators Manual Issue 5 For correct operation of this autoclave it is essential that the chamber be filled with water up to the

More information