Local dose reference levels for abdominal interventional radiology procedures

Size: px
Start display at page:

Download "Local dose reference levels for abdominal interventional radiology procedures"

Transcription

1 Local dose reference levels for abdominal interventional radiology procedures Poster No.: C-2053 Congress: ECR 2014 Type: Scientific Exhibit Authors: L. Hadid-Beurrier 1, M.-J. Waryn 2, Y. Ajavon 1, O. SEROR 1, N. Keywords: DOI: SELLIER 1 ; 1 Bondy/FR, 2 Bobigny/FR Radioprotection / Radiation dose, Interventional vascular, Abdomen, Fluoroscopy, Dosimetry, Embolisation, Shunts, Dosimetric comparison /ecr2014/C-2053 Any information contained in this pdf file is automatically generated from digital material submitted to EPOS by third parties in the form of scientific presentations. References to any names, marks, products, or services of third parties or hypertext links to thirdparty sites or information are provided solely as a convenience to you and do not in any way constitute or imply ECR's endorsement, sponsorship or recommendation of the third party, information, product or service. ECR is not responsible for the content of these pages and does not make any representations regarding the content or accuracy of material in this file. As per copyright regulations, any unauthorised use of the material or parts thereof as well as commercial reproduction or multiple distribution by any traditional or electronically based reproduction/publication method ist strictly prohibited. You agree to defend, indemnify, and hold ECR harmless from and against any and all claims, damages, costs, and expenses, including attorneys' fees, arising from or related to your use of these pages. Please note: Links to movies, ppt slideshows and any other multimedia files are not available in the pdf version of presentations. Page 1 of 15

2 Aims and objectives The frequency of interventional radiology (IR) procedures has seen significant growth over the past decades. This is justified by the undisputed clinical benefits of the IR practice, the increasing clinical skills of the operators and the development of new techniques with improved X-ray systems, digital detectors, new post processing, threedimensional imaging etc. However, IR can involve high doses to the patients. Specifically, the dose to the patients' skin can, in some cases, exceed the threshold of tissue reactions [1-3]. As a consequence, optimization of radiation protection during IR procedures is necessary. The concept of reference levels (RL) is the main tool for the optimization of patient dose. It represents an investigational level used to identify unusually high radiation doses for common medical imaging procedures. The International Commission on Radiological Protection (ICRP) introduced this term for the first time in 90s for only diagnostic common procedures and standard size patients [4-6]. Establishing these levels on a national stage for common interventional radiology examinations is a complex task. In fact, the definition of 'common' examinations can be difficult because of the variations in patients' anatomy and pathology. Nevertheless, many international bodies focused on this topic and published guidances for establishing RL for interventional procedures [7, 8]. However, no similar national reference levels have been developed infrance. In order to optimize patients' radiation protection, it is recommended to establish and use local reference levels for the most frequent and exposing procedures [9]. Jean Verdier Hospital is one of the reference hospitals in treatment of hepato-digestive diseases in France. Moreover, at present, there are still very few references in the literature about patient dosimetry in abdominal IR procedures [10, 11]. In December 2012, the radiology department of Jean Verdier Hospital has acquired a new IR interventional system with flat-panel digital detector in addition to a dose management solution, DoseWatch. Consequently, the radiology department has implemented a quality assurance (QA) program including many processes like the training of the radiology staff in patient radiation management, standardization of all the interventional procedures labels and patient dose tracking. In order to complete the process of QA, the purpose of the present study is to suggest initial values for local reference levels (RL) for abdominal fluoroscopically guided procedures in Jean Verdier Hospital in France. Page 2 of 15

3 Methods and materials Figure 1 illustrates the method of setting up local reference levels for abdominal IR procedures. The present study focused on four of the most irradiating abdominal procedures in the interventional radiology department: hepatic embolizations (HE), gastro-intestinal embolizations (GE), Transjugular Intrahepatic Porto-Systemic Shunt (TIPSS) and Percutaneous Transhepatic Cholangiography (PTC). Using the dose management solution, data on the fluoroscopy time, Kerma Area Product (KAP) and reference air kerma (Ka,r) were collected over a one year period. From these records, a set of interventional RL was deduced using the 75 th percentile method which is principally used for the establishment of diagnostic reference levels. This method prescribes the use of a value corresponding to the 75% of the distribution established by a dosimetric survey. The RL values found were compared to the European and American data published in the literature. Page 3 of 15

4 Images for this section: Fig. 1: Establishment of local reference levels for abdominal interventional procedures in Jean Verdier Hospital. Page 4 of 15

5 Results 99 patients were included in this study: 32 TIPSS, 24 PTC, 11 GE and 32 HE. KAP, Ka,r and fluoroscopy time distributions are given in figure 2 for TIPSS, in figure 3 for PTC, in figure 4 for GE and in figure 5 for HE procedures. The 75th percentiles are also shown on each histogram. The suggested KAP RL deduced from the 75 th percentile of the distributions are 350 Gy.cm 2, 41 Gy.cm 2,120 Gy.cm 2, and 280 Gy.cm 2 for TIPSS, PTC, GE and HE respectively. The estimated Ka,r are 1100 mgy, 190 mgy, 430 mgy and 660 mgy and the RL for the fluoroscopy time are 54 min, 12 min, 32 min and 18 min, respectively. In tables 1 to 3, a comparison of the proposed reference levels was done with the data published in the United States of America and Europe [10-12]. Our data are generally within the range of other reported values. For TIPSS procedures, tables 1 to 3 show that proposed KAP RL are similar to the Swiss RL and 20% lower than the US RL. However, we can notice that the RL for Jean Verdier Hospital in terms of fluoroscopy time is 35% higher than that of the Swiss national RL. This is explained by the TIPSS procedure that is applied in our centre. In fact, unlike the procedure used generally, apart from the TIPSS placement, physicians embolize systematically the gastric varicoce veins. The latter step can be very complex and thus increases the fluoroscopy time without increasing the doses. Indeed, the Ka,r RL of this study are 50% lower than the US RL. This can be explained by the fact that the mean weight of the American population is higher than the European population (table 4). Furthermore, it is important to note that the US reference levels were corrected for variations in patients' weight. In Jean Verdier Hospital, the range of weight for the TIPSS patients undergoing interventional procedures is kg. The mean weight of the 99 adult subjects of this study is 71,3 kg which is conform to the standard 70 kg. Concerning PTC procedures, our values are slightly higher compared to the published ones. While no published data were found for the Ka,r, the comparison with the Swiss RL shows 20% differences for the KAP and 37% for the fluoroscopy time. For HE procedures, RL of this study for the three dose indicators are lower than the other values. Indeed, slight variations are found between the proposed KAP RL and the Swiss and Spanish values (6,7% and 3,1% respectively). For the fluoroscopy time, our Page 5 of 15

6 values are 10% less than the Swiss RL and 26% less than the Spanish RL. The biggest differences are noticed for the Ka,r RL where our values are 55% lower than the US ones. This is probably owing to the difference in patients' weight between American and European states. The largest discrepancies are found for the GE procedures. The differences between our values and the US RL amount to about 74% for the KAP and 79% for the Ka,r. These variations can also be explained by the high divergence in GE patients' weight between the two countries. Indeed, a difference of 33% is noticed between the average weights (Table 4). Page 6 of 15

7 Images for this section: Fig. 2: Distribution of the KAP (Gy.cm2), Ka,r (mgy) and the fluoroscopy time (s) for the TIPSS procedure Page 7 of 15

8 Fig. 3: Distribution of the KAP (Gy.cm2), Ka,r (mgy) and the fluoroscopy time (s) for the PTC procedure Page 8 of 15

9 Fig. 4: Distribution of the KAP (Gy.cm2), Ka,r (mgy) and the fluoroscopy time (s) for the GE procedure Page 9 of 15

10 Fig. 5: Distribution of the KAP (Gy.cm2), Ka,r (mgy) and the fluoroscopy time (s) for the HE procedure Page 10 of 15

11 Table 1: Comparison of KAP RL (Gy.cm2) of the present study with published data Table 2: Comparison of Ka,r (mgy) RL of the present study with published data Table 3: Comparison of fluoroscopy time RL (s) of the present study with published data Page 11 of 15

12 Table 4: Weight distribution for adult patients of Jean Verdier Hospital Page 12 of 15

13 Conclusion The establishment of reference levels in interventional radiology requires a close and continuous cooperation between the physician, the medical physicist and the radiologic technologist. The local provisional RL values established at the Jean Verdier Hospital are comparable to those proposed by other international studies. Using the dose monitoring software, alerts will be sent to the physician and the medical physicist when dose levels exceed the defined local reference levels for each exam. In this case, investigation is initiated and data are analysed to propose corrective actions if needed. These RL should be updated regularly in order to optimize radiation protection of patients in interventional radiology. The revision of these values and the follow up of the tendency in the next years is part of the quality assurance process implemented in the radiology department of Jean Verdier Hospital. In the future, the percutaneous tumour ablations using 3D images will be investigated using the same method. Page 13 of 15

14 References 1. Koenig TR, Wolff D, Mettler FA, Wagner LK (2001) Skin injuries from fluoroscopically guided procedures: Part 1. Characteristics of radiation injury. AJR Am J Roentgenol 177(1): Koenig TR, Mettler FA, Wagner LK (2001) Skin injuries from fluoroscopically guided procedures: Part 2. Review of 73 cases and recommendations for minimizing dose delivered to patient. AJR Am J Roentgenol 177(1): Batler S, Hopewell JW, Miller DL, et al. (2010) Fluoroscopy guided interventional procedures: A review of radiation effects on patients' skin and hair. Radiology Wall BF, Shrimpton PC (1998). The historical development of reference doses in diagnostic radiology. Radiat. Prot Dosimetry; 80: Ann ICRP (2001) Diagnostic reference levels in medical imaging: review and additional advice. 31: Ann ICRP (1991) 1990 Recommendations of the International Commission on Radiological Protection. 1991; 21: Batler S, Miller DL, Vano E, et al (2008). A pilot study exploring the possibility of establishing guidance levels in x-ray directed interventional procedures. Med Phys; 35: Aroua A, Rickli H, Stauffer JC, et al (2007). How to set up and apply reference levels in fluoroscopy at a national level. Eur Radiol; 17: GTRI/GPMED (2010) Rapport du groupe de travail sur la radiologie interventionnelle : Recommandations pour l'amélioration de la radioprotection en radiologie interventionnelle 10. Vano E,R Sanchez, JM Fernandez et al. (2009) Patient Dose Reference levels for interventional radiology : A national approach. Cardiovac Intervent Radiol 32:19-24 Page 14 of 15

15 11. Miller DL, Deukwoo K, Bonavia GH (2009) Reference levels for patient radiation doses in interventional radiology: Proposed initial values for US practice. Radiology Radiation Protection Division, Federal Office of Public Health, Diagnostic Reference Levels in Cardiology and Interventional Radiology (2008)Switzerland Notice R Page 15 of 15

Dose at an interventional reference point: Comparison between displayed values of X-ray units and actual measured values

Dose at an interventional reference point: Comparison between displayed values of X-ray units and actual measured values Dose at an interventional reference point: Comparison between displayed values of X-ray units and actual measured values Poster No.: C-1844 Congress: ECR 2012 Type: Authors: Keywords: DOI: Educational

More information

Patient doses in lower limb angiography in a recently installed flat detector x-ray system

Patient doses in lower limb angiography in a recently installed flat detector x-ray system Patient doses in lower limb angiography in a recently installed flat detector x-ray system Poster No.: C-0939 Congress: ECR 2011 Type: Authors: Keywords: DOI: Scientific Paper V. Tsapaki, D. Papavdis,

More information

Evaluation of organ and effective doses for CAG based on in-phantom dosimetry

Evaluation of organ and effective doses for CAG based on in-phantom dosimetry Evaluation of organ and effective doses for CAG based on in-phantom dosimetry Poster No.: C-0619 Congress: ECR 2013 Type: Scientific Exhibit Authors: Y. Matsunaga, A. Kawaguchi, S. Suzuki, Y. Takikawa,

More information

Monitoring of radiation dose using dose tracking software in a neurovascular intervention room

Monitoring of radiation dose using dose tracking software in a neurovascular intervention room Monitoring of radiation dose using dose tracking software in a neurovascular intervention room Poster No.: ESI-0017 Congress: EuroSafe Imaging 2017 Type: EuroSafe Imaging Authors: H. Acton 1, K. James

More information

Abdominal CT with Single-Energy Metal Artifact Reduction (SEMAR): Initial Experiences

Abdominal CT with Single-Energy Metal Artifact Reduction (SEMAR): Initial Experiences Abdominal CT with Single-Energy Metal Artifact Reduction (SEMAR): Initial Experiences Poster No.: C-0674 Congress: ECR 2014 Type: Scientific Exhibit Authors: K. Sofue 1, T. Yoshikawa 1, N. Negi 1, Y. Ohno

More information

Understanding Imaging of Gastric and Intestinal Surgery

Understanding Imaging of Gastric and Intestinal Surgery Understanding Imaging of Gastric and Intestinal Surgery Poster No.: C-2605 Congress: ECR 2013 Type: Educational Exhibit Authors: R. E. Ochoa Albíztegui, S. E. Vazquez ; Mexico, D.F./MX, 1 2 1 2 Mexico,

More information

Time-resolved 4D TWIST MR angiography of congenital heart and vessel anomalies: image quality and added diagnostic value compared to 3D MRA

Time-resolved 4D TWIST MR angiography of congenital heart and vessel anomalies: image quality and added diagnostic value compared to 3D MRA Time-resolved 4D TWIST MR angiography of congenital heart and vessel anomalies: image quality and added diagnostic value compared to 3D MRA Poster No.: C-1663 Congress: ECR 2012 Type: Scientific Paper

More information

Ultrasound guided biopsy using magnetic resonance volume navigation for the histological approach of additional breast lesions: a new technique

Ultrasound guided biopsy using magnetic resonance volume navigation for the histological approach of additional breast lesions: a new technique Ultrasound guided biopsy using magnetic resonance volume navigation for the histological approach of additional breast lesions: a new technique Poster No.: C-1100 Congress: ECR 2011 Type: Authors: Scientific

More information

An optimal protocol for full-spine radiography

An optimal protocol for full-spine radiography An optimal protocol for full-spine radiography Poster No.: C-0649 Congress: ECR 2016 Type: Scientific Exhibit Authors: C. Ernst, N. Buls, A. Laumen, E. ASSENOVA, S. Assouti, G. 1 1 1 1 4 1 2 2 3 3 Van

More information

Do we need MRI quality assurance: experience from a multiunit imaging center with 14 MRI systems

Do we need MRI quality assurance: experience from a multiunit imaging center with 14 MRI systems Do we need MRI quality assurance: experience from a multiunit imaging center with 14 MRI systems Poster No.: B-0321 Congress: ECR 2015 Type: Scientific Paper Authors: L. J. Kuusela, M. Timonen, T. Mäkelä,

More information

Ultrasound probes: risk of cross infection and methods to reduce it

Ultrasound probes: risk of cross infection and methods to reduce it Ultrasound probes: risk of cross infection and methods to reduce it Poster No.: C-0932 Congress: ECR 2013 Type: Authors: Keywords: DOI: Educational Exhibit N. Ramesh, N. Mukiza; Portlaoise/IE Management,

More information

Robot-assisted navigation system for CT-guided percutaneous lung lesions procedures: our experience in Hong Kong

Robot-assisted navigation system for CT-guided percutaneous lung lesions procedures: our experience in Hong Kong Robot-assisted navigation system for CT-guided percutaneous lung lesions procedures: our experience in Hong Kong Poster No.: B-0714 Congress: ECR 2015 Type: Scientific Paper Authors: C. M. Chu, S. Yu;

More information

Interventional Fluoroscopy

Interventional Fluoroscopy NATIONAL CANCER INSTITUTE THE SOCIETY OF INTERVENTIONAL RADIOLOGY Interventional Fluoroscopy Reducing Radiation Risks for Patients and Staff CONTENTS Introduction 1 Increasing use and complexity of interventional

More information

What Have We Done To Deserve This? Key Performance Indicators in Radiology.

What Have We Done To Deserve This? Key Performance Indicators in Radiology. What Have We Done To Deserve This? Key Performance Indicators in Radiology. Poster No.: C-1505 Congress: ECR 2017 Type: Authors: Educational Exhibit S. Vujnovic 1, S. Vujmilovic 2, D. A. Stojanov 3 ; 1

More information

This prospective study was approved by our institutional review board.

This prospective study was approved by our institutional review board. Digital subtraction angiography during TIPS creation or revision: data on radiation exposure and image quality obtained using a standard and a low-dose acquisition protocol in a flat-panel detector-based

More information

Efficacy of a new post processing workflow for CTA head and neck

Efficacy of a new post processing workflow for CTA head and neck Efficacy of a new post processing workflow for CTA head and neck Poster No.: C-1760 Congress: ECR 2013 Type: Authors: Keywords: DOI: Scientific Exhibit N. Ardley 1, K. Buchan 2, K. K. Lau 3 ; 1 Clayton,

More information

Adverse reactions to CT contrast media: incidence, severity and risk factors

Adverse reactions to CT contrast media: incidence, severity and risk factors Adverse reactions to CT contrast media: incidence, severity and risk factors Poster No.: C-0827 Congress: ECR 2014 Type: Authors: Keywords: DOI: Scientific Exhibit S. Lee, K. Bae, K. N. Jeon; Jinju/KR

More information

Automatic multi-slice CT injectors: hygiene and efficiency of disposable, prefilled and multi-dosing roller pump systems in clinical routine

Automatic multi-slice CT injectors: hygiene and efficiency of disposable, prefilled and multi-dosing roller pump systems in clinical routine Automatic multi-slice CT injectors: hygiene and efficiency of disposable, prefilled and multi-dosing roller pump systems in clinical routine Poster No.: C-0305 Congress: ECR 2011 Type: Authors: Keywords:

More information

Purpose. Methods and Materials

Purpose. Methods and Materials MR cholangiopancreatography: Comparison between 3D fast recovery fast spin echo and 2D single shot fast spin echo sequences in the evaluation of choledocholithiasis. Poster No.: C-1193 Congress: ECR 2012

More information

Dark is the night (or how fatigue increases errors in abdominal CT reports).

Dark is the night (or how fatigue increases errors in abdominal CT reports). Dark is the night (or how fatigue increases errors in abdominal CT reports). Poster No.: C-2214 Congress: ECR 2012 Type: Authors: Keywords: DOI: Scientific Exhibit C. Serrano Burgos, A. Malet, F. Novell

More information

Comparison of MR sequences for in vivo olfactory bulb volumetry: reproducibility and repeatability of volumetric measurements using 3 Tesla MRI

Comparison of MR sequences for in vivo olfactory bulb volumetry: reproducibility and repeatability of volumetric measurements using 3 Tesla MRI Comparison of MR sequences for in vivo olfactory bulb volumetry: reproducibility and repeatability of volumetric measurements using 3 Tesla MRI Poster No.: C-0504 Congress: ECR 2012 Type: Scientific Paper

More information

Digital breast tomosynthesis-guided biopsy: preliminary experience

Digital breast tomosynthesis-guided biopsy: preliminary experience Digital breast tomosynthesis-guided biopsy: preliminary experience Poster No.: C-2207 Congress: ECR 2016 Type: Authors: Keywords: DOI: Scientific Exhibit F. Pipan, E. Cimino, E. Zanelli, A. Dal Col, R.

More information

The workload of UK radiologists: how does it compare to that of international radiologists?

The workload of UK radiologists: how does it compare to that of international radiologists? The workload of UK radiologists: how does it compare to that of international radiologists? Poster No.: C-1386 Congress: ECR 2013 Type: Authors: Keywords: DOI: Scientific Exhibit W. Hedges 1, S. Khan 2

More information

Quality ID #145: Radiology: Exposure Dose or Time Reported for Procedures Using Fluoroscopy National Quality Strategy Domain: Patient Safety

Quality ID #145: Radiology: Exposure Dose or Time Reported for Procedures Using Fluoroscopy National Quality Strategy Domain: Patient Safety Quality ID #145: Radiology: Exposure Dose or Time Reported for Procedures Using Fluoroscopy National Quality Strategy Domain: Patient Safety 2018 OPTIONS FOR INDIVIDUAL MEASURES: REGISTRY ONLY MEASURE

More information

Assessment of commercially available deformable image registration (DIR) software for CT-MRI registration in breast radiotherapy patients

Assessment of commercially available deformable image registration (DIR) software for CT-MRI registration in breast radiotherapy patients Assessment of commercially available deformable image registration (DIR) software for CT-MRI registration in breast radiotherapy patients Poster No.: R-0019 Congress: 2014 CSM Type: Scientific Exhibit

More information

Strategies for the reduction of exposure of medical staff in the CT operating room

Strategies for the reduction of exposure of medical staff in the CT operating room Strategies for the reduction of exposure of medical staff in the CT operating room Poster No.: C-3025 Congress: ECR 2010 Type: Educational Exhibit Topic: Radiographers Authors: M. Körner, U. Linsenmaier,

More information

Poster No.: R Page 1 of 12

Poster No.: R Page 1 of 12 Abdominal endovascular aortic repair (EVAR): Comparison of image fusion (IF3D) angiography versus conventional (2D) angiography with respect to intravenous contrast dose, radiation dose and total fluoroscopy

More information

Diagnostic Performance of Strain- and Shear-Wave- Elastography in an Elasticity Phantom: A Comparative Study

Diagnostic Performance of Strain- and Shear-Wave- Elastography in an Elasticity Phantom: A Comparative Study Diagnostic Performance of Strain- and Shear-Wave- Elastography in an Elasticity Phantom: A Comparative Study Poster No.: C-2135 Congress: ECR 2014 Type: Scientific Exhibit Authors: J. F. Carlsen 1, M.

More information

Implementation of a multi country CT Dose Management system

Implementation of a multi country CT Dose Management system Implementation of a multi country CT Dose Management system Poster No.: ESI-0015 Congress: EuroSafe Imaging 2016 Type: EuroSafe Imaging Authors: K. Katsari 1, R. Illing 2, A. Papachristodoulou 3, C. A.

More information

The role of Diagnostic Reference Levels in Optimization

The role of Diagnostic Reference Levels in Optimization The role of Diagnostic Reference Levels in Optimization Eliseo Vano Radiology Department. COMPLUTENSE UNIVERSITY MADRID Medical Physics Service. SAN CARLOS UNIVERSITY HOSPITAL MADRID - SPAIN Thursday 19th

More information

DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY: A STRATEGY FOR INTRODUCING DIAGNOSTIC REFERENCE LEVELS TAKING INTO ACCOUNT THE NATIONAL PRACTICE

DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY: A STRATEGY FOR INTRODUCING DIAGNOSTIC REFERENCE LEVELS TAKING INTO ACCOUNT THE NATIONAL PRACTICE DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY: A STRATEGY FOR INTRODUCING DIAGNOSTIC REFERENCE LEVELS TAKING INTO ACCOUNT THE NATIONAL PRACTICE Ph. R. Trueb #, A. Aroua *, H. Jung #, Th. Theiler #, J.-F. Valley

More information

Aims and objectives. Page 2 of 13

Aims and objectives. Page 2 of 13 Review of individual radiation exposure and the impact of real-time radiation information on an Interventional Radiology (IR) team using real-time individual dosimeters for the insertion of central venous

More information

Installation of a regional CT dose registry with a dose management software

Installation of a regional CT dose registry with a dose management software Installation of a regional CT dose registry wi a dose management software Poster No.: ESI-0036 Congress: EuroSafe Imaging 2016 Type: EuroSafe Imaging Auors: S. Schindera, A. Euler, A. Parakh ; Riehen/CH,

More information

Ultrasound Bi-dimensional Navigation Technology for Breast application

Ultrasound Bi-dimensional Navigation Technology for Breast application Ultrasound Bi-dimensional Navigation Technology for Breast application Poster No.: C-1824 Congress: ECR 2016 Type: Scientific Exhibit Authors: A. Morresi, V. Girardi, S. de Beni, M. battaglia, L. Forzoni,

More information

Optimization of anode-filter combinations in digital mammography: An experimental study

Optimization of anode-filter combinations in digital mammography: An experimental study Optimization of anode-filter combinations in digital mammography: An experimental study Poster No.: C-2915 Congress: ECR 2010 Type: Scientific Exhibit Topic: Physics in Radiology Authors: A. Peruzzo Cornetto

More information

A dosimetric trial for the clinical follow-up of potential skin injuries on patients undergoing interventional cardiology procedures

A dosimetric trial for the clinical follow-up of potential skin injuries on patients undergoing interventional cardiology procedures A dosimetric trial for the clinical follow-up of potential skin injuries on patients undergoing interventional cardiology procedures Vano E (1,2), Aviles P (1,2), Prieto C (2), Fernandez JM (1,2), Guibelalde

More information

US-guided procedures in the management of muscular trauma

US-guided procedures in the management of muscular trauma US-guided procedures in the management of muscular trauma Poster No.: P-0132 Congress: ESSR 2015 Type: Educational Poster Authors: B. Ruiz, J. L. del Cura, F. Diez Renovales, I. Corta, J. Cardenal Urdampilleta,

More information

Low radiation dose subtraction CT angiography (CTA) for diagnosis of peripheral arterial occlusive disease

Low radiation dose subtraction CT angiography (CTA) for diagnosis of peripheral arterial occlusive disease Low radiation dose subtraction CT angiography (CTA) for diagnosis of peripheral arterial occlusive disease Poster No.: C-1978 Congress: ECR 2013 Type: Authors: Scientific Exhibit M. Suzuki 1, R. Tanaka

More information

White Paper. Dose Tracking System. Patient Skin Dose Estimates in Real Time

White Paper. Dose Tracking System. Patient Skin Dose Estimates in Real Time White Paper Dose Tracking System Patient Skin Dose Estimates in Real Time INTRODUCTION In an era of increased utilization of ionizing radiation for medical diagnostic and interventional procedures and

More information

In recent years, the benefits and complexity of interventional

In recent years, the benefits and complexity of interventional PATIENT SAFETY Patient Radiation Dose Management in the Follow-Up of Potential Skin Injuries in Neuroradiology E. Vano, J.M. Fernandez, R.M. Sanchez, D. Martinez, L. Lopez Ibor, A. Gil, and C. Serna-Candel

More information

2018/7/31. Primary Objective

2018/7/31. Primary Objective Andrew Kuhls-Gilcrist, PhD, DABR July 31, 2018 Primary Objective The purpose of this session is to allow vendors to present their solutions and describe how their product can empower the clinician and

More information

Patient doses in interventional cardiology procedures.

Patient doses in interventional cardiology procedures. Patient doses in interventional cardiology procedures. J. Domienik *a, S.Papierz a, J. Jankowski a, J.Z. Peruga b a Nofer Institute of Occupational Medicine in Lodz, Radiation Protection Department, Lodz,

More information

Transcatheter arterial chemoembolization of the extrahepatic collateral arteries in patients with hepatocellular carcinoma

Transcatheter arterial chemoembolization of the extrahepatic collateral arteries in patients with hepatocellular carcinoma Transcatheter arterial chemoembolization of the extrahepatic collateral arteries in patients with hepatocellular carcinoma Poster No.: C-1945 Congress: ECR 2010 Type: Topic: Scientific Exhibit Interventional

More information

Establishing a patient safety program in Interventional Radiology. A. Kyle Jones, Ph.D., DABR Assistant Professor MD Anderson Cancer Center

Establishing a patient safety program in Interventional Radiology. A. Kyle Jones, Ph.D., DABR Assistant Professor MD Anderson Cancer Center Establishing a patient safety program in Interventional Radiology A. Kyle Jones, Ph.D., DABR Assistant Professor MD Anderson Cancer Center Disclosure I am co owner of Fluoroscopic Safety, LLC, a company

More information

Modified WHO Surgical safety check-lists for MRI-related safety in hybrid operating room

Modified WHO Surgical safety check-lists for MRI-related safety in hybrid operating room Modified WHO Surgical safety check-lists for MRI-related safety in hybrid operating room Poster No.: C-0014 Congress: ECR 2016 Type: Authors: Keywords: DOI: Educational Exhibit M. Cherkashin 1, N. Berezina

More information

Ultrasound diagnosis of pneumothorax after Fine Needle Aspiration of the lung

Ultrasound diagnosis of pneumothorax after Fine Needle Aspiration of the lung Ultrasound diagnosis of pneumothorax after Fine Needle Aspiration of the lung Poster No.: C-0727 Congress: ECR 2013 Type: Scientific Exhibit Authors: A. Lasarte Izcue, J. M. Navasa Melado, R. Pellón, G.

More information

Patient Dose Research in Interventional Radiology Suites in Rio de Janeiro, Brazil

Patient Dose Research in Interventional Radiology Suites in Rio de Janeiro, Brazil Patient Dose Research in Interventional Radiology Suites in Rio de Janeiro, Brazil Canevaro, L.V. (Presenting); Oliveira da Silva, M.W.; Dias Rodrigues, B.B. Serviço de Física Médica, Instituto de Radioproteção

More information

Radiation Protection Dosimetry (2008), Vol. 132, No. 1, pp Advance Access publication 25 September 2008

Radiation Protection Dosimetry (2008), Vol. 132, No. 1, pp Advance Access publication 25 September 2008 Radiation Protection Dosimetry (2008), Vol. 132, No. 1, pp. 18 24 Advance Access publication 25 September 2008 doi:10.1093/rpd/ncn247 CORRELATION OF PATIENT MAXIMUM SKIN DOSES IN CARDIAC PROCEDURES WITH

More information

1. Executive Summary

1. Executive Summary 1. Executive Summary Diagnostic reference levels (DRLs), which are a form of investigation levels (ICRP, 1996), represent an important tool to optimize image quality and the radiation dose delivered to

More information

Importance of a Patient Dosimetry and Clinical Follow-up Program in the Detection of Radiodermatitis After Long Percutaneous Coronary Interventions

Importance of a Patient Dosimetry and Clinical Follow-up Program in the Detection of Radiodermatitis After Long Percutaneous Coronary Interventions Cardiovasc Intervent Radiol (2013) 36:330 337 DOI 10.1007/s00270-012-0397-x CLINICAL INVESTIGATION ARTERIAL INTERVENTIONS Importance of a Patient Dosimetry and Clinical Follow-up Program in the Detection

More information

program in Interventional Radiology Assistant Professor U.T. M. D. Anderson Cancer Center

program in Interventional Radiology Assistant Professor U.T. M. D. Anderson Cancer Center Establishing a patient safety program in Interventional Radiology A. Kyle Jones, Ph.D. Assistant Professor U.T. M. D. Anderson Cancer Center Or Complying with JC standards JC Sentinel Event Defined as:

More information

Error rates in breast imaging reports generated with automatic speech recognition technology versus traditional transcriptionist-generated reports

Error rates in breast imaging reports generated with automatic speech recognition technology versus traditional transcriptionist-generated reports Error rates in breast imaging reports generated with automatic speech recognition technology versus traditional transcriptionist-generated reports Poster No.: B-866 Congress: ECR 2011 Type: Scientific

More information

New Joint Commission Requirements for Fluoroscopy Services

New Joint Commission Requirements for Fluoroscopy Services New Joint Commission Requirements for Fluoroscopy Services Melissa C. Martin, M.S., FAAPM, FACR Past President AAPM Florida Chapter Fall Meeting Orlando, FL September 14, 2018 New Requirements Effective

More information

FLUOROSCOPIC PATIENT & OPERATOR RISK REDUCTION

FLUOROSCOPIC PATIENT & OPERATOR RISK REDUCTION FLUOROSCOPIC PATIENT & OPERATOR RISK REDUCTION Ray Dielman RSO ST ANTHONY S HOSPITAL James Menge CHP ARSO ST ANTHONY S HOSPITAL BAYCARE HEALTH SYSTEM 1 FLUROSCOPY EDUCATIONAL OBJECTIVES I. Understand the

More information

A Paradigm Shift in Patient Dose Monitoring

A Paradigm Shift in Patient Dose Monitoring VL White Paper A Paradigm Shift in Patient Dose Monitoring Dose Tracking System Andrew Kuhls-Gilcrist, Ph.D., D.A.B.R. Senior XRVL Clinical Science Manager Canon Medical Systems, Inc. Millions of fluoroscopically

More information

ESTIMATION OF ORGAN DOSES OF PATIENT UNDERGOING HEPATIC CHEMOEMBOLIZATION PROCEDURES

ESTIMATION OF ORGAN DOSES OF PATIENT UNDERGOING HEPATIC CHEMOEMBOLIZATION PROCEDURES X Congreso Regional Latinoamericano IRPA de Protección y Seguridad Radiológica Radioprotección: Nuevos Desafíos para un Mundo en Evolución Buenos Aires, 12 al 17 de abril, 2015 SOCIEDAD ARGENTINA DE RADIOPROTECCIÓN

More information

REFERENCES LEVELS OF PATIENT RADIATION DOSE IN INTERVENTIONAL RADIOLOGY: A SPANISH CONTRIBUTION

REFERENCES LEVELS OF PATIENT RADIATION DOSE IN INTERVENTIONAL RADIOLOGY: A SPANISH CONTRIBUTION REFERENCES LEVELS OF PATIENT RADIATION DOSE IN INTERVENTIONAL RADIOLOGY: A SPANISH CONTRIBUTION F. Carrera 1,3, R. Ruiz-Cruces 3, E. Vano 2, J. Hernandez-Armas 4, M. Fernandez-Bordes 5, C. Ojeda 3 1 Radiological

More information

Poster No.: C-1848 Congress: ECR Educational Exhibit. Authors:

Poster No.: C-1848 Congress: ECR Educational Exhibit. Authors: Understanding " the fusion imaging " - Methods of fusion with their advantages and disadvantages and study various fusion combination with their applications. Poster No.: C-1848 Congress: ECR 2017 Type:

More information

Cost analysis of "CEUS first" approach in the assessment of focal liver lesions.

Cost analysis of CEUS first approach in the assessment of focal liver lesions. Cost analysis of "CEUS first" approach in the assessment of focal liver lesions. Poster No.: C-1152 Congress: ECR 2014 Type: Scientific Exhibit Authors: R. Dežman, D. Vidmar-Bracika, P. Gregoric, M. Gorenc,

More information

7/30/2018. Paid Speaker, Bayer. Radiology Cardiology Interventional Endoscopy Pain Medicine Speech Language Pathology Surgery List is longer than this

7/30/2018. Paid Speaker, Bayer. Radiology Cardiology Interventional Endoscopy Pain Medicine Speech Language Pathology Surgery List is longer than this http://www.wpclipart.com (Public Domain) 7/30/2018 William F. Sensakovic, PhD, DABR, MRSC Medical Physicist Florida Hospital Co-Founder, Telerad Physics Teaching, LLC Assoc. Prof. University of Central

More information

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May Dose Reference Levels

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May Dose Reference Levels 2033-7 Joint ICTP/ Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dose Reference Levels Peter Homolka EFOMP Training Course on Medical Physics in Diagnostic

More information

Experience With Patient Dosimetry and Quality Control Online for Diagnostic and Interventional Radiology Using DICOM Services

Experience With Patient Dosimetry and Quality Control Online for Diagnostic and Interventional Radiology Using DICOM Services Medical Physics and Informatics Review Vano et al. Online Quality Control for Diagnostic and Interventional Radiology Medical Physics and Informatics Review FOCUS ON: Eliseo Vano 1,2 Jose I. Ten 2,3 Jose

More information

Poster No.: C-1217 Congress: ECR Scientific Exhibit. Authors:

Poster No.: C-1217 Congress: ECR Scientific Exhibit. Authors: Is rotational cholangiogram a feasible investigation for obstructive jaundice? A comparison study with conventional percutaneous transhepatic cholangiogram. Poster No.: C-1217 Congress: ECR 2013 Type:

More information

Methods for measuring fluoroscopic skin dose

Methods for measuring fluoroscopic skin dose Pediatr Radiol (2006) 36 (Suppl 2): 136 140 DOI 10.1007/s00247-006-0193-3 ALARA Stephen Balter Methods for measuring fluoroscopic skin dose Published online: 22 July 2006 # Springer-Verlag 2006 Abstract

More information

Ultrasound (US)-guided percutaneous core biopsy of pancreatic tumor:tips and tricks to ensure safe and effective biopsy

Ultrasound (US)-guided percutaneous core biopsy of pancreatic tumor:tips and tricks to ensure safe and effective biopsy Ultrasound (US)-guided percutaneous core biopsy of pancreatic tumor:tips and tricks to ensure safe and effective biopsy Poster No.: C-2368 Congress: ECR 2014 Type: Scientific Exhibit Authors: J. W. Kim,

More information

1. Executive Summary

1. Executive Summary 1. Executive Summary 1.1 General The fluoroscope is defined as an instrument used chiefly in industry and in the practice of medicine for observing the internal structure of objects (such as the living

More information

Minnesota Rules, Chapter 4732 X-ray Revision

Minnesota Rules, Chapter 4732 X-ray Revision Minnesota Rules, Chapter 4732 X-ray Revision FLUOROSCOPIC X-RAY SYSTEMS, 1.0 Subpart 17. Subpart 18. Subpart 19. Subpart 20. Subpart 21. Subpart 22. Subpart 23. Subpart 24. Subpart 25. Equipment performance

More information

Patient skin doses in complex cardiac procedures

Patient skin doses in complex cardiac procedures S1.P16 Patient skin doses in complex cardiac procedures R. Borisova* 1, J. Vassileva 2, V. Gelev 3, A. Doganov 3 1 Tokuda Hospital Sofia, Sofia, Bulgaria 2 National Centre of Radiobiology and Radiation

More information

Investigation Of Tissue Elasticity Measurements Using Shear Wave Ultrasound Elastography

Investigation Of Tissue Elasticity Measurements Using Shear Wave Ultrasound Elastography Investigation Of Tissue Elasticity Measurements Using Shear Wave Ultrasound Elastography Poster No.: C-1915 Congress: ECR 2014 Type: Scientific Exhibit Authors: H. E. Ting, C. H. Yeong, B. J. J. Abdullah,

More information

Radiological Protection of Patient and Operator in Interventional Radiology

Radiological Protection of Patient and Operator in Interventional Radiology Radiological Protection of Patient and Operator in Interventional Radiology JMAJ 44(11): 484 489, 2001 Hironobu NAKAMURA Professor and Chairman, Department of Radiology, Osaka University Graduate School

More information

Dose Reduction during EVAR in Hybrid Room. Multicentric Study

Dose Reduction during EVAR in Hybrid Room. Multicentric Study Dose Reduction during EVAR in Hybrid Room Multicentric Study Golden rules to reduce radiation dose in the hybrid room to a minimum Stéphan Haulon, MD, Best Practice from Lille Aortic Center Aotic Centre,

More information

Diagnostic reference levels and complexity indices in interventional radiology: a national programme

Diagnostic reference levels and complexity indices in interventional radiology: a national programme DOI 10.1007/s00330-016-4334-2 INTERVENTIONAL Diagnostic reference levels and complexity indices in interventional radiology: a national programme R. Ruiz-Cruces 1 & E. Vano 2 & F. Carrera-Magariño 3 &

More information

SEMINAR 1. Radiological Protection Of Patients. Radiation Protection in Paediatric Cardiology. Dra. Patricia Miranda González

SEMINAR 1. Radiological Protection Of Patients. Radiation Protection in Paediatric Cardiology. Dra. Patricia Miranda González SEMINAR 1 Radiological Protection Of Patients Radiation Protection in Paediatric Cardiology Dra. Patricia Miranda González CARDIOVASCULAR DEPARTMENT, CALVO MACKENNA S HOSPITAL SANTIAGO CHILE Pediatric

More information

Protecting Patients 3/5/2014. Educating and Engaging your Staff in Radiation Safety

Protecting Patients 3/5/2014. Educating and Engaging your Staff in Radiation Safety Educating and Engaging your Staff in Radiation Safety Robert Vincent MD;CM FACC Professor of Pediatrics Emory University School of Medicine Ralf Holzer MD Associate Professor of Pediatrics Ohio State University

More information

The Radiation Dose in Interventional Radiology Study: Knowledge Brings Responsibility

The Radiation Dose in Interventional Radiology Study: Knowledge Brings Responsibility Commentary The Radiation Dose in Interventional Radiology Study: Knowledge Brings Responsibility M. Victoria Marx,MD J Vasc Interv Radiol 2003; 14:947 951 Abbreviations: CD cumulative dose, FDA Food and

More information

Gafchromic XR-RV2 radiochromic films dosimetry in interventional radiology

Gafchromic XR-RV2 radiochromic films dosimetry in interventional radiology Gafchromic XR-RV2 radiochromic films dosimetry in interventional radiology A. Neocleous *1, E. Yakoumakis 1, G. Gialousis 2, A. Dimitriadis 1, N. Yakoumakis 3 and E. Georgiou 1 1 Medical Physics Department,

More information

RECONSTRUCTION OF ORGAN DOSES IN CHILDREN UNDERGOING LIFE-SAVING FLUOROSCOPY-GUIDED CARDIAC SURGERY

RECONSTRUCTION OF ORGAN DOSES IN CHILDREN UNDERGOING LIFE-SAVING FLUOROSCOPY-GUIDED CARDIAC SURGERY RECONSTRUCTION OF ORGAN DOSES IN CHILDREN UNDERGOING LIFE-SAVING FLUOROSCOPY-GUIDED CARDIAC SURGERY November 13 th, 2015 Emily Marshall, B.S. J Crayton Pruitt Family Department of Biomedical Engineering

More information

Reducing Unnecessary Medical Imaging Exposure

Reducing Unnecessary Medical Imaging Exposure Reducing Unnecessary Medical Imaging Exposure CDR Sean M. Boyd, MPH, USPHS U.S. Food and Drug Administration Center for Devices and Radiological Health April 30, 2010 Objectives FDA s role in ensuring

More information

Justification for Training

Justification for Training This educational program is being offered to provide training to medical practitioners who perform or supervise the performance of diagnostic procedures utilizing fluoroscopic x- ray units. There are many

More information

Radiation Dose Index Monitoring (RDIM)

Radiation Dose Index Monitoring (RDIM) Radiation Dose Index Monitoring (RDIM) Renée (Dickinson) Butler, M.S. Texas Health Resources Presbyterian Hospital Dallas AAPM Medical Physics Practice Guideline (MPPG) 6.a. Selection of a Radiation Dose

More information

IV. 1. Skin Dose Measurement for Patients Using Imaging Plates in Interventional Radiology Procedures

IV. 1. Skin Dose Measurement for Patients Using Imaging Plates in Interventional Radiology Procedures CYRIC Annual Report 2004 IV. 1. Skin Dose Measurement for Patients Using Imaging Plates in Interventional Radiology Procedures Ohuchi H., Satoh T. *, Eguchi Y. *, and Mori K. ** Graduate School of Pharmaceutical

More information

Radiation Dose of Interventional Radiology System Using a Flat-Panel Detector

Radiation Dose of Interventional Radiology System Using a Flat-Panel Detector Vascular and Interventional Radiology Original Research Chida et al. Radiation Dose in Interventional Radiology Vascular and Interventional Radiology Original Research Koichi Chida 1 Yohei Inaba 2 Haruo

More information

University of Cape Town

University of Cape Town Radiation dose optimization in interventional radiology and cardiology using diagnostic reference levels Hendrik Johannes de Vos DVSHEN3 Thesis presented in fulfilment of the requirements for the degree

More information

Breast cancer: let's prepare breast reconstruction.

Breast cancer: let's prepare breast reconstruction. Breast cancer: let's prepare breast reconstruction. Poster No.: C-2273 Congress: ECR 2011 Type: Educational Exhibit Authors: F. Bidault, M. Sahnoun, C. Balleyguier, S. Canale, B. Boulet, 1 1 1 1 2 1 1

More information

#Patients develop allergic or nephrotoxic reactions to iodinated media # carbon dioxide (CO2) gas angiography can be used.

#Patients develop allergic or nephrotoxic reactions to iodinated media # carbon dioxide (CO2) gas angiography can be used. 320-row multidetector CT angiography for hepatocellular carcinoma using CO2 gas instead of iodinated contrast agents: Experiment and preliminary clinical study Poster No.: C-1123 Congress: ECR 2011 Type:

More information

A failure mode and effects analysis (FMEA) to assess a protocol for MRI scanning in patients with ICDs and pacemakers

A failure mode and effects analysis (FMEA) to assess a protocol for MRI scanning in patients with ICDs and pacemakers A failure mode and effects analysis (FMEA) to assess a protocol for MRI scanning in patients with ICDs and pacemakers Poster No.: C-0969 Congress: ECR 2018 Type: Scientific Exhibit Authors: P. Gilligan,

More information

NEMA XR Standard Attributes on CT Equipment Related to Dose Optimization and Management

NEMA XR Standard Attributes on CT Equipment Related to Dose Optimization and Management NEMA XR 29-2013 Standard Attributes on CT Equipment Related to Dose Optimization and Management Published by: National Electrical Manufacturers Association 1300 North 17th Street Rosslyn, Virginia 22209

More information

Comparison of quantitative methods for the assessment of interstitial lung disease in MDCT

Comparison of quantitative methods for the assessment of interstitial lung disease in MDCT Comparison of quantitative methods for the assessment of interstitial lung disease in MDCT Award: Certificate of Merit Poster No.: C-0140 Congress: ECR 2014 Type: Scientific Exhibit Authors: N. Papapanagiotou,

More information

PATIENT DOSIMETRY IN INTERVENTIONAL RADIOLOGY

PATIENT DOSIMETRY IN INTERVENTIONAL RADIOLOGY 2009 International Nuclear Atlantic Conference - INAC 2009 Rio de Janeiro,RJ, Brazil, September27 to October 2, 2009 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-03-8 PATIENT DOSIMETRY

More information

DIRECT MEASUREMENT OF SKIN DOSE WITH RADIOCHROMIC FILMS IN FLUOROSCOPY GUIDED VESSEL REPAIR

DIRECT MEASUREMENT OF SKIN DOSE WITH RADIOCHROMIC FILMS IN FLUOROSCOPY GUIDED VESSEL REPAIR DIRECT MEASUREMENT OF SKIN DOSE WITH RADIOCHROMIC FILMS IN FLUOROSCOPY GUIDED VESSEL REPAIR J. KALEF-EZRA, EZRA, * 1 E. KATSAROU, 1 S. KARAVASILIS, 1 M. MATSAGKAS, 2 L.K. MICHALIS 3 1 Medical Physics Department,

More information

RADIATION EXPOSURE IN ENDOVASCULAR PROCEDURES - AN ASSESSMENT BASED ON EXAMINATIONS PERFORMED AT SAHLGRENSKA UNIVERSITY HOSPITAL

RADIATION EXPOSURE IN ENDOVASCULAR PROCEDURES - AN ASSESSMENT BASED ON EXAMINATIONS PERFORMED AT SAHLGRENSKA UNIVERSITY HOSPITAL SAHLGRENSKA ACADEMY DEPARTMENT OF RADIATION PHYSICS RADIATION EXPOSURE IN ENDOVASCULAR PROCEDURES - AN ASSESSMENT BASED ON EXAMINATIONS PERFORMED AT SAHLGRENSKA UNIVERSITY HOSPITAL M.Sc. thesis Roham D.

More information

TRIR project. The determination of Trigger Levels for patient doses in Interventional Procedures. End report

TRIR project. The determination of Trigger Levels for patient doses in Interventional Procedures. End report The determination of Trigger Levels for patient doses in Interventional Procedures TRIR project End report by Lara Struelens (SCK CEN, Mol) Klaus Bacher (Universiteit Gent) Fréderic Bleeser (CH Jolimont-Lobbes)

More information

PATIENT RADIATION DOSE RANGES FOR PROCEDURES IN UNIVERSITAS HOSPITAL VASCULAR LABORATORIES

PATIENT RADIATION DOSE RANGES FOR PROCEDURES IN UNIVERSITAS HOSPITAL VASCULAR LABORATORIES PATIENT RADIATION DOSE RANGES FOR PROCEDURES IN UNIVERSITAS HOSPITAL VASCULAR LABORATORIES Henra Muller N Dip Radiography (Diagnostic) B Tech Radiography (Diagnostic) Dissertation submitted in fulfilment

More information

Improving protocols and procedures for strengthened radiation protection in interventional procedures

Improving protocols and procedures for strengthened radiation protection in interventional procedures Improving protocols and procedures for strengthened radiation protection in interventional procedures R. Loose German Commission on Radiological Protection (SSK) German Roentgen Society (DRG) Institute

More information

How to perform multi-modality 3D roadmapping in neurovascular procedures

How to perform multi-modality 3D roadmapping in neurovascular procedures How to perform multi-modality 3D roadmapping in neurovascular procedures Poster No.: C-1183 Congress: ECR 2012 Type: Educational Exhibit Authors: D. Ruijters, R. Homan, P. Mielekamp, D. Babic; Best/NL

More information

There are three general methods for reducing one's personal radiation exposure. The application of each when working around fluoroscopy units will be

There are three general methods for reducing one's personal radiation exposure. The application of each when working around fluoroscopy units will be There are three general methods for reducing one's personal radiation exposure. The application of each when working around fluoroscopy units will be discussed in more detail. The ALARA concept was introduced

More information

Protecting patients by training physicians in fluoroscopic radiation management

Protecting patients by training physicians in fluoroscopic radiation management JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 1, NUMBER 1, WINTER 2000 Protecting patients by training physicians in fluoroscopic radiation management Benjamin R. Archer a) Department of Radiology,

More information

5DGLDWLRQSURWHFWLRQ *XLGDQFHRQPHGLFDO H[SRVXUHVLQPHGLFDODQG ELRPHGLFDOUHVHDUFK (XURSHDQ&RPPLVVLRQ

5DGLDWLRQSURWHFWLRQ *XLGDQFHRQPHGLFDO H[SRVXUHVLQPHGLFDODQG ELRPHGLFDOUHVHDUFK (XURSHDQ&RPPLVVLRQ 5DGLDWLRQSURWHFWLRQ *XLGDQFHRQPHGLFDO H[SRVXUHVLQPHGLFDODQG ELRPHGLFDOUHVHDUFK (XURSHDQ&RPPLVVLRQ European Commission 5DGLDWLRQSURWHFWLRQ *8,'$1&(210(',&$/(;32685(6,10(',&$/$1'%,20(',&$/5(6($5&+ 1998 Directorate-General

More information

Radiography Curriculum Analysis

Radiography Curriculum Analysis Program Number Program Name Date / /20 Radiography Curriculum Analysis DIRECTIONS: Determine the course(s) in which each of the following content area is covered and enter the course number(s) and/or title(s).

More information

Clinical Applications of the Latest Ultra-High-Resolution CT and Iterative Reconstruction Algorithm: Things Radiologist and Radiographer Need to Know

Clinical Applications of the Latest Ultra-High-Resolution CT and Iterative Reconstruction Algorithm: Things Radiologist and Radiographer Need to Know Clinical Applications of the Latest Ultra-High-Resolution CT and Iterative Reconstruction Algorithm: Things Radiologist and Radiographer Need to Know Poster No.: C-1477 Congress: ECR 2018 Type: Educational

More information