Monitoring of radiation dose using dose tracking software in a neurovascular intervention room

Size: px
Start display at page:

Download "Monitoring of radiation dose using dose tracking software in a neurovascular intervention room"

Transcription

1 Monitoring of radiation dose using dose tracking software in a neurovascular intervention room Poster No.: ESI-0017 Congress: EuroSafe Imaging 2017 Type: EuroSafe Imaging Authors: H. Acton 1, K. James 1, F. Moloney 1, N. Fanning 2, G. Wyse 3, M. M. Maher 1, O. O'Connor 1 ; 1 Cork/IE, 2 University Hospital Cork/IE, 3 Cork University Hospital Cork/IE Keywords: DOI: Embolism / Thrombosis, Arteriovenous malformations, Aneurysms, Embolisation, Diagnostic procedure, Arterial access, Fluoroscopy, Radioprotection / Radiation dose, Interventional vascular, CNS, Action 6 - Education and training in medical radiation protection, Action 3 - Optimisation, diagnostic reference levels, image quality, Action 2 - Clinical audit /esi2017/ESI-0017 Any information contained in this pdf file is automatically generated from digital material submitted to EPOS by third parties in the form of scientific presentations. References to any names, marks, products, or services of third parties or hypertext links to thirdparty sites or information are provided solely as a convenience to you and do not in any way constitute or imply ECR's endorsement, sponsorship or recommendation of the third party, information, product or service. ECR is not responsible for the content of these pages and does not make any representations regarding the content or accuracy of material in this file. As per copyright regulations, any unauthorised use of the material or parts thereof as well as commercial reproduction or multiple distribution by any traditional or electronically based reproduction/publication method ist strictly prohibited. You agree to defend, indemnify, and hold ECR harmless from and against any and all claims, damages, costs, and expenses, including attorneys' fees, arising from or related to your use of these pages. Page 1 of 12

2 Please note: Links to movies, ppt slideshows and any other multimedia files are not available in the pdf version of presentations. Page 2 of 12

3 Background/Introduction Neurointerventional procedures have the potential for greater radiation exposure compared with other radiological procedures. While the benefits of the procedures outweigh the risks, patient radiation exposure can cause adverse effects such as radiation burn, cataracts and malignancy and should be kept as low as possible. Article 56(1) of the European Union Directive 2013/59/Euratom requires that all member states ensure that radiation doses for interventional procedures are kept as low as reasonably achieveable. Article 56(2) states that all member states must ensure the establishment, and use of diagnostic reference levels for radiodiagnostic examinations, and where appropriate, interventional radiology procedures[1]. The introduction of dose tracking software in Ireland is a welcome addition to radiology departments to aid in patient dose monitoring. Dose tracking software allows faster analysis of patient dose, decreased risk of input error, and alert to increased dose. The use of dose monitoring software will aid in analysis of patient dose levels with the end goal of producing accurate national dose reference levels for individual types of neurointerventional procedures (which is currently lacking). Combining this data with national hospitals performing such procedures will allow for the formation of national dose reference guidelines for these procedures which can be used national as guidelines and as a reference both within the European Union and internationally. Aim The aim for this study was to analyse neurointerventional radiation dose levels using dose-tracking software to find determinants of increased radiation dose Objectives Review radiation dose for neurointerventional procedures performed over a one-year period Analyse aneurysm coiling procedure variables to determine dose technique relationships Page 3 of 12

4 Description of activity and work performed Methods A retrospective cohort study was performed at Cork University Hospital, in Cork, Ireland. Radiation dose was measured using Dose Area Product (DAP) measured in mgy.cm 2. Data collection was carried out on the Picture Archiving and Communication System (Impax 6.5.3; Agfa Healthcare, Morstel, Belgium) and dose tracking software (DoseWatch, GE, Milwaukee, WI). Cases were excluded if dose level data was missing. Additional variable data on aneurysm coiling procedures was provided by a prospectively gathered patient procedure database continuously maintained by a consultant neurointerventional radiologist. Statistical analysis was performed using IBM SPSS software version 23. Two hundred and sixty-four patients who underwent neurointerventional procedures at Cork University Hospital between November 2014 and November 2015 were selected. thirty-two cases were excluded due to insufficient data. twenty-two of these cases were diagnostic; The remaining ten were therapeutic. Four neurointerventional procedures were studied: (1) four vessel angiogram, (2) aneurysm coiling, (3) arteriovenous malformation embolization, and (4) stroke thrombolysis/thrombectomy. One hundred and nine aneurysm coiling procedures were selected for the second objective, including cases selected for objective 1. the remaining cases were collected from the end period of objective 1 (November 2015) to September Results: Objective 1 Procedure type Of the two-hundred and sixty-four procedures selected, one hundred and eighty nine were four-vessel angiograms, fifty-nine were aneurysm coiling procedures, ten were stroke thrombolysis/thrombectomy, and six were arteriovenous malformation embolization. Patient demographic Sixty-one percent of patients were female and thirty-nine percent were male. Procedures distinguished by sex differed substantially. The number of aneurysm coiling procedures Page 4 of 12

5 performed on women was 3.7 times the number performed on men. Overall, women were more likely to undergo both diagnostic and therapeutic procedures. Patient age was categorized into groups of equal distribution and numbered for data analysis. Patients aged (group 4) had the highest number of therapeutic procedures performed. Patients aged (groups 4 and 5) shared the highest number of overall procedures performed. Radiation dose The median radiation DAP for diagnostic angiography (four-vessel angiogram) was 73.7 Gy.cm 2 (DAP range ). The remaining procedures were therapeutic. The median DAP for aneurysm coiling was Gy.cm 2 (DAP range ). The median DAP for arteriovenous malformation embolization was Gy.cm 2 (DAP range ), the median DAP for stroke thrombolysis/thrombectomy was 87 Gy.cm 2 (DAP range ). The Health Services Executive (Irish health service) have published dose reference levels (75th percentile DAP) for other interventional procedures. Other interventional procedures generally have lower DAP. For example, the DRL for a barium enema is 47 Gy.Cm 2, however, the DRL for a transjugular intrahepatic portosystemic shunt procedure (TIPS), a more complicated procedure, and more comparable to a challenging embolisation is referenced at Gy.cm 2 [2][3]. A number of outliers were found for both aneurysm coiling and four-vessel angiogram groups. The patient case files were reviewed on PACS to determine possible causes for the increased doses. Five of the six four-vessel angiogram outlier cases had arteriovenous malformations which required addiation imaging projections and more contrast studies than normal. Three of the five-aneurysm coiling procedure outliers were posterior circulation aneurysm with/without anterior circulation aneurysms. The two remaining cases were anterior circulation aneurysms developed complications, such as coil herniation and vasospasm. Results: Objective 2 Aneurysm coiling variables were analysed to determined variables that had a significant effect on radiation dose to the patient. Overall, nine variables were found to have a significant effect ( Table 3 on page 7 ). These variables were then analysed further to determine what vairable had the single biggest effect on radiation dose. Page 5 of 12

6 Overall, aneurysm location (posterior circulation vs. anterior circulation) was found to have the single biggest effect on radiation dose (p=.004, see Table 4 on page 8 ). Images for this section: Table 1: Summary of DAP measurement for neurointerventional procedures - Cork/IE Page 6 of 12

7 Fig. 3: Box-plot chart depicting DAP for neurointerventional procedures - Cork/IE Table 2: DAP percentiles for each neurointerventional procedure - Cork/IE Page 7 of 12

8 Table 3: Aneurysm coiling variables affecting DAP - Cork/IE Table 4: Aneurysm location - Cork/IE Page 8 of 12

9 Conclusion and Recommendations Exposure of the patient to high levels of radiation is a serious risk when performing neurointerventional procedures. It is the duty of the neurointerventional radiologist and team to perform the procedure to the best of their ability whilst also minimizing radiation exposure for the patient. The results of this study show that a patient undergoing a neurointerventional procedure will be subjected to variable radiation exposures depending on the abnormality and/or vascular location. These results demonstrate which patient sub groups are at increased risk for higher exposure, which has prompted review of techniques to help optimise radiation dose to the patient. This data is also useful for the purposes of patient consent where radiation dose should be discussed if a high radiation dose is anticipated. In addition the present data have implications for staff exposures especially in the context of tighter limits on operator eye doses proposed in Euratom Personal/Organisational information Holly Acton, Final Year Medical Student, University College Cork, Ireland Karl James, Radiology Specialist Registrar, Cork University Hospital, Ireland Fiachra Moloney, Radiology Fellow (MSK), Ontario, Canada Noel Fanning, Consultant Neuroradiologist, Cork University Hospital, Ireland Gerald Wyse, Consultant Neuroradiologist, Cork University Hospital, Ireland Michael Maher, Consultant Radiologist and Professor of Radiology UCC/Cork and Mercy University Hospitals, Ireland Owen O'Connor, Consultant Radiologist and Senior Lecturer UCC/Cork and Mercy University Hospitals, Ireland Page 9 of 12

10 Images for this section: Fig. 1 Cork University Hospital Page 10 of 12

11 Fig. 2 University College Cork Page 11 of 12

12 References [1] Council Directive 2013/59/Euratom of 5 December 2013 laying down basic safety standards for protection against the dangers arising from exposure to ionizing radiation and repealing Directives 89/618/Euratom, 90/641/Euratom, 92/29/Euratom, 97/43 Euratom and 2003/122/Euratom [2] Medical Council. Diagnostic Reference Levels: Position Paper. September safepatientcare/medexpradiatonunit/diagnostic_reference_levels.pdf [3]HSE. Template for developing a Patient Radiation Protection Manual. August medexpradiatonunit/rp%20manual% pdf Page 12 of 12

Local dose reference levels for abdominal interventional radiology procedures

Local dose reference levels for abdominal interventional radiology procedures Local dose reference levels for abdominal interventional radiology procedures Poster No.: C-2053 Congress: ECR 2014 Type: Scientific Exhibit Authors: L. Hadid-Beurrier 1, M.-J. Waryn 2, Y. Ajavon 1, O.

More information

Efficacy of a new post processing workflow for CTA head and neck

Efficacy of a new post processing workflow for CTA head and neck Efficacy of a new post processing workflow for CTA head and neck Poster No.: C-1760 Congress: ECR 2013 Type: Authors: Keywords: DOI: Scientific Exhibit N. Ardley 1, K. Buchan 2, K. K. Lau 3 ; 1 Clayton,

More information

Robot-assisted navigation system for CT-guided percutaneous lung lesions procedures: our experience in Hong Kong

Robot-assisted navigation system for CT-guided percutaneous lung lesions procedures: our experience in Hong Kong Robot-assisted navigation system for CT-guided percutaneous lung lesions procedures: our experience in Hong Kong Poster No.: B-0714 Congress: ECR 2015 Type: Scientific Paper Authors: C. M. Chu, S. Yu;

More information

Ultrasound guided biopsy using magnetic resonance volume navigation for the histological approach of additional breast lesions: a new technique

Ultrasound guided biopsy using magnetic resonance volume navigation for the histological approach of additional breast lesions: a new technique Ultrasound guided biopsy using magnetic resonance volume navigation for the histological approach of additional breast lesions: a new technique Poster No.: C-1100 Congress: ECR 2011 Type: Authors: Scientific

More information

Abdominal CT with Single-Energy Metal Artifact Reduction (SEMAR): Initial Experiences

Abdominal CT with Single-Energy Metal Artifact Reduction (SEMAR): Initial Experiences Abdominal CT with Single-Energy Metal Artifact Reduction (SEMAR): Initial Experiences Poster No.: C-0674 Congress: ECR 2014 Type: Scientific Exhibit Authors: K. Sofue 1, T. Yoshikawa 1, N. Negi 1, Y. Ohno

More information

Understanding Imaging of Gastric and Intestinal Surgery

Understanding Imaging of Gastric and Intestinal Surgery Understanding Imaging of Gastric and Intestinal Surgery Poster No.: C-2605 Congress: ECR 2013 Type: Educational Exhibit Authors: R. E. Ochoa Albíztegui, S. E. Vazquez ; Mexico, D.F./MX, 1 2 1 2 Mexico,

More information

Time-resolved 4D TWIST MR angiography of congenital heart and vessel anomalies: image quality and added diagnostic value compared to 3D MRA

Time-resolved 4D TWIST MR angiography of congenital heart and vessel anomalies: image quality and added diagnostic value compared to 3D MRA Time-resolved 4D TWIST MR angiography of congenital heart and vessel anomalies: image quality and added diagnostic value compared to 3D MRA Poster No.: C-1663 Congress: ECR 2012 Type: Scientific Paper

More information

Implementation of a multi country CT Dose Management system

Implementation of a multi country CT Dose Management system Implementation of a multi country CT Dose Management system Poster No.: ESI-0015 Congress: EuroSafe Imaging 2016 Type: EuroSafe Imaging Authors: K. Katsari 1, R. Illing 2, A. Papachristodoulou 3, C. A.

More information

Dose at an interventional reference point: Comparison between displayed values of X-ray units and actual measured values

Dose at an interventional reference point: Comparison between displayed values of X-ray units and actual measured values Dose at an interventional reference point: Comparison between displayed values of X-ray units and actual measured values Poster No.: C-1844 Congress: ECR 2012 Type: Authors: Keywords: DOI: Educational

More information

Evaluation of organ and effective doses for CAG based on in-phantom dosimetry

Evaluation of organ and effective doses for CAG based on in-phantom dosimetry Evaluation of organ and effective doses for CAG based on in-phantom dosimetry Poster No.: C-0619 Congress: ECR 2013 Type: Scientific Exhibit Authors: Y. Matsunaga, A. Kawaguchi, S. Suzuki, Y. Takikawa,

More information

The workload of UK radiologists: how does it compare to that of international radiologists?

The workload of UK radiologists: how does it compare to that of international radiologists? The workload of UK radiologists: how does it compare to that of international radiologists? Poster No.: C-1386 Congress: ECR 2013 Type: Authors: Keywords: DOI: Scientific Exhibit W. Hedges 1, S. Khan 2

More information

Adverse reactions to CT contrast media: incidence, severity and risk factors

Adverse reactions to CT contrast media: incidence, severity and risk factors Adverse reactions to CT contrast media: incidence, severity and risk factors Poster No.: C-0827 Congress: ECR 2014 Type: Authors: Keywords: DOI: Scientific Exhibit S. Lee, K. Bae, K. N. Jeon; Jinju/KR

More information

Do we need MRI quality assurance: experience from a multiunit imaging center with 14 MRI systems

Do we need MRI quality assurance: experience from a multiunit imaging center with 14 MRI systems Do we need MRI quality assurance: experience from a multiunit imaging center with 14 MRI systems Poster No.: B-0321 Congress: ECR 2015 Type: Scientific Paper Authors: L. J. Kuusela, M. Timonen, T. Mäkelä,

More information

Comparison of MR sequences for in vivo olfactory bulb volumetry: reproducibility and repeatability of volumetric measurements using 3 Tesla MRI

Comparison of MR sequences for in vivo olfactory bulb volumetry: reproducibility and repeatability of volumetric measurements using 3 Tesla MRI Comparison of MR sequences for in vivo olfactory bulb volumetry: reproducibility and repeatability of volumetric measurements using 3 Tesla MRI Poster No.: C-0504 Congress: ECR 2012 Type: Scientific Paper

More information

Purpose. Methods and Materials

Purpose. Methods and Materials MR cholangiopancreatography: Comparison between 3D fast recovery fast spin echo and 2D single shot fast spin echo sequences in the evaluation of choledocholithiasis. Poster No.: C-1193 Congress: ECR 2012

More information

Poster No.: R Page 1 of 12

Poster No.: R Page 1 of 12 Abdominal endovascular aortic repair (EVAR): Comparison of image fusion (IF3D) angiography versus conventional (2D) angiography with respect to intravenous contrast dose, radiation dose and total fluoroscopy

More information

This prospective study was approved by our institutional review board.

This prospective study was approved by our institutional review board. Digital subtraction angiography during TIPS creation or revision: data on radiation exposure and image quality obtained using a standard and a low-dose acquisition protocol in a flat-panel detector-based

More information

An optimal protocol for full-spine radiography

An optimal protocol for full-spine radiography An optimal protocol for full-spine radiography Poster No.: C-0649 Congress: ECR 2016 Type: Scientific Exhibit Authors: C. Ernst, N. Buls, A. Laumen, E. ASSENOVA, S. Assouti, G. 1 1 1 1 4 1 2 2 3 3 Van

More information

Ultrasound probes: risk of cross infection and methods to reduce it

Ultrasound probes: risk of cross infection and methods to reduce it Ultrasound probes: risk of cross infection and methods to reduce it Poster No.: C-0932 Congress: ECR 2013 Type: Authors: Keywords: DOI: Educational Exhibit N. Ramesh, N. Mukiza; Portlaoise/IE Management,

More information

Installation of a regional CT dose registry with a dose management software

Installation of a regional CT dose registry with a dose management software Installation of a regional CT dose registry wi a dose management software Poster No.: ESI-0036 Congress: EuroSafe Imaging 2016 Type: EuroSafe Imaging Auors: S. Schindera, A. Euler, A. Parakh ; Riehen/CH,

More information

Dark is the night (or how fatigue increases errors in abdominal CT reports).

Dark is the night (or how fatigue increases errors in abdominal CT reports). Dark is the night (or how fatigue increases errors in abdominal CT reports). Poster No.: C-2214 Congress: ECR 2012 Type: Authors: Keywords: DOI: Scientific Exhibit C. Serrano Burgos, A. Malet, F. Novell

More information

Assessment of commercially available deformable image registration (DIR) software for CT-MRI registration in breast radiotherapy patients

Assessment of commercially available deformable image registration (DIR) software for CT-MRI registration in breast radiotherapy patients Assessment of commercially available deformable image registration (DIR) software for CT-MRI registration in breast radiotherapy patients Poster No.: R-0019 Congress: 2014 CSM Type: Scientific Exhibit

More information

What Have We Done To Deserve This? Key Performance Indicators in Radiology.

What Have We Done To Deserve This? Key Performance Indicators in Radiology. What Have We Done To Deserve This? Key Performance Indicators in Radiology. Poster No.: C-1505 Congress: ECR 2017 Type: Authors: Educational Exhibit S. Vujnovic 1, S. Vujmilovic 2, D. A. Stojanov 3 ; 1

More information

Automatic multi-slice CT injectors: hygiene and efficiency of disposable, prefilled and multi-dosing roller pump systems in clinical routine

Automatic multi-slice CT injectors: hygiene and efficiency of disposable, prefilled and multi-dosing roller pump systems in clinical routine Automatic multi-slice CT injectors: hygiene and efficiency of disposable, prefilled and multi-dosing roller pump systems in clinical routine Poster No.: C-0305 Congress: ECR 2011 Type: Authors: Keywords:

More information

Digital breast tomosynthesis-guided biopsy: preliminary experience

Digital breast tomosynthesis-guided biopsy: preliminary experience Digital breast tomosynthesis-guided biopsy: preliminary experience Poster No.: C-2207 Congress: ECR 2016 Type: Authors: Keywords: DOI: Scientific Exhibit F. Pipan, E. Cimino, E. Zanelli, A. Dal Col, R.

More information

Patient doses in lower limb angiography in a recently installed flat detector x-ray system

Patient doses in lower limb angiography in a recently installed flat detector x-ray system Patient doses in lower limb angiography in a recently installed flat detector x-ray system Poster No.: C-0939 Congress: ECR 2011 Type: Authors: Keywords: DOI: Scientific Paper V. Tsapaki, D. Papavdis,

More information

Modified WHO Surgical safety check-lists for MRI-related safety in hybrid operating room

Modified WHO Surgical safety check-lists for MRI-related safety in hybrid operating room Modified WHO Surgical safety check-lists for MRI-related safety in hybrid operating room Poster No.: C-0014 Congress: ECR 2016 Type: Authors: Keywords: DOI: Educational Exhibit M. Cherkashin 1, N. Berezina

More information

Ultrasound Bi-dimensional Navigation Technology for Breast application

Ultrasound Bi-dimensional Navigation Technology for Breast application Ultrasound Bi-dimensional Navigation Technology for Breast application Poster No.: C-1824 Congress: ECR 2016 Type: Scientific Exhibit Authors: A. Morresi, V. Girardi, S. de Beni, M. battaglia, L. Forzoni,

More information

US-guided procedures in the management of muscular trauma

US-guided procedures in the management of muscular trauma US-guided procedures in the management of muscular trauma Poster No.: P-0132 Congress: ESSR 2015 Type: Educational Poster Authors: B. Ruiz, J. L. del Cura, F. Diez Renovales, I. Corta, J. Cardenal Urdampilleta,

More information

Low radiation dose subtraction CT angiography (CTA) for diagnosis of peripheral arterial occlusive disease

Low radiation dose subtraction CT angiography (CTA) for diagnosis of peripheral arterial occlusive disease Low radiation dose subtraction CT angiography (CTA) for diagnosis of peripheral arterial occlusive disease Poster No.: C-1978 Congress: ECR 2013 Type: Authors: Scientific Exhibit M. Suzuki 1, R. Tanaka

More information

Diagnostic Performance of Strain- and Shear-Wave- Elastography in an Elasticity Phantom: A Comparative Study

Diagnostic Performance of Strain- and Shear-Wave- Elastography in an Elasticity Phantom: A Comparative Study Diagnostic Performance of Strain- and Shear-Wave- Elastography in an Elasticity Phantom: A Comparative Study Poster No.: C-2135 Congress: ECR 2014 Type: Scientific Exhibit Authors: J. F. Carlsen 1, M.

More information

Strategies for the reduction of exposure of medical staff in the CT operating room

Strategies for the reduction of exposure of medical staff in the CT operating room Strategies for the reduction of exposure of medical staff in the CT operating room Poster No.: C-3025 Congress: ECR 2010 Type: Educational Exhibit Topic: Radiographers Authors: M. Körner, U. Linsenmaier,

More information

Quality ID #145: Radiology: Exposure Dose or Time Reported for Procedures Using Fluoroscopy National Quality Strategy Domain: Patient Safety

Quality ID #145: Radiology: Exposure Dose or Time Reported for Procedures Using Fluoroscopy National Quality Strategy Domain: Patient Safety Quality ID #145: Radiology: Exposure Dose or Time Reported for Procedures Using Fluoroscopy National Quality Strategy Domain: Patient Safety 2018 OPTIONS FOR INDIVIDUAL MEASURES: REGISTRY ONLY MEASURE

More information

Aims and objectives. Page 2 of 13

Aims and objectives. Page 2 of 13 Review of individual radiation exposure and the impact of real-time radiation information on an Interventional Radiology (IR) team using real-time individual dosimeters for the insertion of central venous

More information

Error rates in breast imaging reports generated with automatic speech recognition technology versus traditional transcriptionist-generated reports

Error rates in breast imaging reports generated with automatic speech recognition technology versus traditional transcriptionist-generated reports Error rates in breast imaging reports generated with automatic speech recognition technology versus traditional transcriptionist-generated reports Poster No.: B-866 Congress: ECR 2011 Type: Scientific

More information

Ultrasound diagnosis of pneumothorax after Fine Needle Aspiration of the lung

Ultrasound diagnosis of pneumothorax after Fine Needle Aspiration of the lung Ultrasound diagnosis of pneumothorax after Fine Needle Aspiration of the lung Poster No.: C-0727 Congress: ECR 2013 Type: Scientific Exhibit Authors: A. Lasarte Izcue, J. M. Navasa Melado, R. Pellón, G.

More information

Transcatheter arterial chemoembolization of the extrahepatic collateral arteries in patients with hepatocellular carcinoma

Transcatheter arterial chemoembolization of the extrahepatic collateral arteries in patients with hepatocellular carcinoma Transcatheter arterial chemoembolization of the extrahepatic collateral arteries in patients with hepatocellular carcinoma Poster No.: C-1945 Congress: ECR 2010 Type: Topic: Scientific Exhibit Interventional

More information

How to perform multi-modality 3D roadmapping in neurovascular procedures

How to perform multi-modality 3D roadmapping in neurovascular procedures How to perform multi-modality 3D roadmapping in neurovascular procedures Poster No.: C-1183 Congress: ECR 2012 Type: Educational Exhibit Authors: D. Ruijters, R. Homan, P. Mielekamp, D. Babic; Best/NL

More information

Poster No.: C-1848 Congress: ECR Educational Exhibit. Authors:

Poster No.: C-1848 Congress: ECR Educational Exhibit. Authors: Understanding " the fusion imaging " - Methods of fusion with their advantages and disadvantages and study various fusion combination with their applications. Poster No.: C-1848 Congress: ECR 2017 Type:

More information

DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY: A STRATEGY FOR INTRODUCING DIAGNOSTIC REFERENCE LEVELS TAKING INTO ACCOUNT THE NATIONAL PRACTICE

DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY: A STRATEGY FOR INTRODUCING DIAGNOSTIC REFERENCE LEVELS TAKING INTO ACCOUNT THE NATIONAL PRACTICE DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY: A STRATEGY FOR INTRODUCING DIAGNOSTIC REFERENCE LEVELS TAKING INTO ACCOUNT THE NATIONAL PRACTICE Ph. R. Trueb #, A. Aroua *, H. Jung #, Th. Theiler #, J.-F. Valley

More information

Cost analysis of "CEUS first" approach in the assessment of focal liver lesions.

Cost analysis of CEUS first approach in the assessment of focal liver lesions. Cost analysis of "CEUS first" approach in the assessment of focal liver lesions. Poster No.: C-1152 Congress: ECR 2014 Type: Scientific Exhibit Authors: R. Dežman, D. Vidmar-Bracika, P. Gregoric, M. Gorenc,

More information

Optimization of anode-filter combinations in digital mammography: An experimental study

Optimization of anode-filter combinations in digital mammography: An experimental study Optimization of anode-filter combinations in digital mammography: An experimental study Poster No.: C-2915 Congress: ECR 2010 Type: Scientific Exhibit Topic: Physics in Radiology Authors: A. Peruzzo Cornetto

More information

Interventional Fluoroscopy

Interventional Fluoroscopy NATIONAL CANCER INSTITUTE THE SOCIETY OF INTERVENTIONAL RADIOLOGY Interventional Fluoroscopy Reducing Radiation Risks for Patients and Staff CONTENTS Introduction 1 Increasing use and complexity of interventional

More information

ESR Clinical Audit booklet Esperanto

ESR Clinical Audit booklet Esperanto ESR Clinical Audit booklet Esperanto This booklet was named after the most successful constructed language in the world, Esperanto, created by the Polish ophthalmologist Ludwik L. Zamenhof. In this booklet

More information

The role of Diagnostic Reference Levels in Optimization

The role of Diagnostic Reference Levels in Optimization The role of Diagnostic Reference Levels in Optimization Eliseo Vano Radiology Department. COMPLUTENSE UNIVERSITY MADRID Medical Physics Service. SAN CARLOS UNIVERSITY HOSPITAL MADRID - SPAIN Thursday 19th

More information

DENOMINATOR: All final reports for patients, regardless of age, undergoing a CT procedure

DENOMINATOR: All final reports for patients, regardless of age, undergoing a CT procedure Quality ID #363: Optimizing Patient Exposure to Ionizing Radiation: Search for Prior Computed Tomography (CT) Studies Through a Secure, Authorized, Media-Free, Shared Archive National Quality Strategy

More information

A post-processing system uses these differential data to generate:

A post-processing system uses these differential data to generate: Iterative reconstruction in single source dual-energy CT pulmonary angiography: Is sufficient to achieve a radiation dose as low as state-of-the-art single-energy CTPA? Poster No.: B-0300 Congress: ECR

More information

Comparison of quantitative methods for the assessment of interstitial lung disease in MDCT

Comparison of quantitative methods for the assessment of interstitial lung disease in MDCT Comparison of quantitative methods for the assessment of interstitial lung disease in MDCT Award: Certificate of Merit Poster No.: C-0140 Congress: ECR 2014 Type: Scientific Exhibit Authors: N. Papapanagiotou,

More information

Poster No.: C-1217 Congress: ECR Scientific Exhibit. Authors:

Poster No.: C-1217 Congress: ECR Scientific Exhibit. Authors: Is rotational cholangiogram a feasible investigation for obstructive jaundice? A comparison study with conventional percutaneous transhepatic cholangiogram. Poster No.: C-1217 Congress: ECR 2013 Type:

More information

Breast cancer: let's prepare breast reconstruction.

Breast cancer: let's prepare breast reconstruction. Breast cancer: let's prepare breast reconstruction. Poster No.: C-2273 Congress: ECR 2011 Type: Educational Exhibit Authors: F. Bidault, M. Sahnoun, C. Balleyguier, S. Canale, B. Boulet, 1 1 1 1 2 1 1

More information

ESR AUDIT TOOL. myesr.org

ESR AUDIT TOOL. myesr.org myesr.org Endorsed by the EFRS (European Federation of Radiographer Societies) in January 2016 2 INTRODUCTION As stated by the European Society of Radiology (ESR), Clinical Audit is defined as a tool designed

More information

DIRECT MEASUREMENT OF SKIN DOSE WITH RADIOCHROMIC FILMS IN FLUOROSCOPY GUIDED VESSEL REPAIR

DIRECT MEASUREMENT OF SKIN DOSE WITH RADIOCHROMIC FILMS IN FLUOROSCOPY GUIDED VESSEL REPAIR DIRECT MEASUREMENT OF SKIN DOSE WITH RADIOCHROMIC FILMS IN FLUOROSCOPY GUIDED VESSEL REPAIR J. KALEF-EZRA, EZRA, * 1 E. KATSAROU, 1 S. KARAVASILIS, 1 M. MATSAGKAS, 2 L.K. MICHALIS 3 1 Medical Physics Department,

More information

RADIATION EXPOSURE IN ENDOVASCULAR PROCEDURES - AN ASSESSMENT BASED ON EXAMINATIONS PERFORMED AT SAHLGRENSKA UNIVERSITY HOSPITAL

RADIATION EXPOSURE IN ENDOVASCULAR PROCEDURES - AN ASSESSMENT BASED ON EXAMINATIONS PERFORMED AT SAHLGRENSKA UNIVERSITY HOSPITAL SAHLGRENSKA ACADEMY DEPARTMENT OF RADIATION PHYSICS RADIATION EXPOSURE IN ENDOVASCULAR PROCEDURES - AN ASSESSMENT BASED ON EXAMINATIONS PERFORMED AT SAHLGRENSKA UNIVERSITY HOSPITAL M.Sc. thesis Roham D.

More information

Investigation Of Tissue Elasticity Measurements Using Shear Wave Ultrasound Elastography

Investigation Of Tissue Elasticity Measurements Using Shear Wave Ultrasound Elastography Investigation Of Tissue Elasticity Measurements Using Shear Wave Ultrasound Elastography Poster No.: C-1915 Congress: ECR 2014 Type: Scientific Exhibit Authors: H. E. Ting, C. H. Yeong, B. J. J. Abdullah,

More information

TRUFILL DCS ORBIT Detachable Coil System

TRUFILL DCS ORBIT Detachable Coil System ORBIT Conforming to Your Complex Needs Excellent Conformability and Concentric Filling for Outstanding Packing Density ORBIT Full range of Mini Complex and new Tight Distal Loop Technology coils Our Complex

More information

By completing this questionnaire, you are indicating your consent to participate in this research.

By completing this questionnaire, you are indicating your consent to participate in this research. Introduction "Is your Imaging EuroSafe?" comprises a series of monthly surveys on CT DRLs for different indications, which are being carried out by the EuroSafe Imaging campaign, coordinated by the European

More information

A failure mode and effects analysis (FMEA) to assess a protocol for MRI scanning in patients with ICDs and pacemakers

A failure mode and effects analysis (FMEA) to assess a protocol for MRI scanning in patients with ICDs and pacemakers A failure mode and effects analysis (FMEA) to assess a protocol for MRI scanning in patients with ICDs and pacemakers Poster No.: C-0969 Congress: ECR 2018 Type: Scientific Exhibit Authors: P. Gilligan,

More information

Clinical Applications. ImagingRite. Interventional Radiology

Clinical Applications. ImagingRite. Interventional Radiology Clinical Applications ImagingRite Interventional Radiology ImagingRite, a comprehensive suite of imaging tools offered with Infinix -i angiographic systems, was designed to assist clinicians in optimizing

More information

Assessing Intra-procedural Arterial Complications Following Lobar Radioembolization with Yttrium-90 Microsphere

Assessing Intra-procedural Arterial Complications Following Lobar Radioembolization with Yttrium-90 Microsphere Assessing Intra-procedural Arterial Complications Following Lobar Radioembolization with Yttrium-90 Microsphere Abdulrahman Masrani, MD ; Bradford J Chong, BS; Osman Ahmed, MD; Sreekumar Madassery, MD;

More information

How Tracking Radiologic Procedures and Dose Helps: Experience From Finland

How Tracking Radiologic Procedures and Dose Helps: Experience From Finland Medical Physics and Informatics Original Research Seuri et al. Tracking Radiologic Procedures and Dose Medical Physics and Informatics Original Research FOCUS ON: Raija Seuri 1 Madan M. Rehani 2 Mika Kortesniemi

More information

5DGLDWLRQSURWHFWLRQ *XLGDQFHRQPHGLFDO H[SRVXUHVLQPHGLFDODQG ELRPHGLFDOUHVHDUFK (XURSHDQ&RPPLVVLRQ

5DGLDWLRQSURWHFWLRQ *XLGDQFHRQPHGLFDO H[SRVXUHVLQPHGLFDODQG ELRPHGLFDOUHVHDUFK (XURSHDQ&RPPLVVLRQ 5DGLDWLRQSURWHFWLRQ *XLGDQFHRQPHGLFDO H[SRVXUHVLQPHGLFDODQG ELRPHGLFDOUHVHDUFK (XURSHDQ&RPPLVVLRQ European Commission 5DGLDWLRQSURWHFWLRQ *8,'$1&(210(',&$/(;32685(6,10(',&$/$1'%,20(',&$/5(6($5&+ 1998 Directorate-General

More information

Diagnosis for Open Wounds as a Result of Cancer Resection

Diagnosis for Open Wounds as a Result of Cancer Resection Diagnosis for Open Wounds as a Result of Cancer Resection December 15, 2016 What diagnosis code do we use when we are reconstructing a defect after the Moh s surgeon, or someone else removed the cancer?

More information

University of Cape Town

University of Cape Town Radiation dose optimization in interventional radiology and cardiology using diagnostic reference levels Hendrik Johannes de Vos DVSHEN3 Thesis presented in fulfilment of the requirements for the degree

More information

Radiology Equipment and Reports

Radiology Equipment and Reports Section 2 Radiology Equipment and Reports Contents Risk Assessment RPII Licence of Equipment Inventory of Radiation Equipment Criteria of Acceptability Setting and Reviewing Nominal Replacement Dates National

More information

Endovascular Simulators Reduce Patient and Staff Exposure

Endovascular Simulators Reduce Patient and Staff Exposure Endovascular Simulators Reduce Patient and Staff Exposure Gabriel Bartal MD, FCIRSE, FSIR Dept of Diagnostic and Interventional Radiology, Meir MC, Kfar-Saba, Affiliated to Tel-Aviv University, Tel-Aviv,

More information

1. Executive Summary

1. Executive Summary 1. Executive Summary Diagnostic reference levels (DRLs), which are a form of investigation levels (ICRP, 1996), represent an important tool to optimize image quality and the radiation dose delivered to

More information

#Patients develop allergic or nephrotoxic reactions to iodinated media # carbon dioxide (CO2) gas angiography can be used.

#Patients develop allergic or nephrotoxic reactions to iodinated media # carbon dioxide (CO2) gas angiography can be used. 320-row multidetector CT angiography for hepatocellular carcinoma using CO2 gas instead of iodinated contrast agents: Experiment and preliminary clinical study Poster No.: C-1123 Congress: ECR 2011 Type:

More information

Introducing a new take on efficient workflow: Deeply integrated clinical applications

Introducing a new take on efficient workflow: Deeply integrated clinical applications Clinical Applications Introducing a new take on efficient workflow: Deeply integrated clinical applications The EXPERIENCE of TRUE CLINICAL APPLICATIONS While developments in image acquisition technology

More information

On the Importance of Computation in Clinical Radiology

On the Importance of Computation in Clinical Radiology On the Importance of Computation in Clinical Radiology Manuel Arreola, Ph.D., DABR Director of Radiological Physics Radiology Department University of Florida/Shands Healthcare My Disclaimer I ve received

More information

HERCA: Heads of the European Radiological Protection Competent Authorities. Multi-stakeholder Meeting on the optimised use of CT scanners

HERCA: Heads of the European Radiological Protection Competent Authorities. Multi-stakeholder Meeting on the optimised use of CT scanners HERCA: Heads of the European Radiological Protection Competent Authorities Multi-stakeholder Meeting on the optimised use of CT scanners i. The self commitments approach with COCIR regarding CT scanners

More information

Setting The setting was secondary care. The economic study was carried out in the Netherlands.

Setting The setting was secondary care. The economic study was carried out in the Netherlands. Contrast-enhanced MR angiography and digital subtraction angiography in living renal donors: diagnostic agreement, impact on decision making, and costs Kock M C, Ijzermans J N, Visser K, Hussain S M, Weimar

More information

Various Iterative Reconstruction Techniques in Chest CT: Impact for Interstitial Lung Disease under Low-dose Condition

Various Iterative Reconstruction Techniques in Chest CT: Impact for Interstitial Lung Disease under Low-dose Condition Various Iterative Reconstruction Techniques in Chest CT: Impact for Interstitial Lung Disease under Low-dose Condition Poster No.: C-1803 Congress: ECR 2018 Type: Scientific Exhibit Authors: M. Kim, M.

More information

Protecting Patients 3/5/2014. Educating and Engaging your Staff in Radiation Safety

Protecting Patients 3/5/2014. Educating and Engaging your Staff in Radiation Safety Educating and Engaging your Staff in Radiation Safety Robert Vincent MD;CM FACC Professor of Pediatrics Emory University School of Medicine Ralf Holzer MD Associate Professor of Pediatrics Ohio State University

More information

Head Fixation Using a Vac-Lok Cushion during Neuroendovascular Therapy

Head Fixation Using a Vac-Lok Cushion during Neuroendovascular Therapy Journal of Neuroendovascular Therapy 2016; 10: 116 120 Online June 10 2016 DOI: 10.5797/jnet.oa.2016-0021 Head Fixation Using a Vac-Lok Cushion during Neuroendovascular Therapy Daisuke Uesaka, 1 Akinori

More information

Comparison of planar and SPECT perfusion lung scintigraphy in the diagnosis of pulmonary embolism

Comparison of planar and SPECT perfusion lung scintigraphy in the diagnosis of pulmonary embolism Comparison of planar and SPECT perfusion lung scintigraphy in the diagnosis of pulmonary embolism Poster No.: C-1798 Congress: ECR 2015 Type: Scientific Exhibit Authors: E. Gera, A. Samuel, F. Zarb; Msida/MT

More information

How to Optimize Radiation Protection During Complex Peripheral

How to Optimize Radiation Protection During Complex Peripheral How to Optimize Radiation Protection During Complex Peripheral G. Bartal MD, FCIRSE, FSIR Dept of Medical Imaging Meir MC, Kfar Saba, Affiliated to Sackler Medical School, Tel Aviv University, Tel Aviv

More information

PATIENT RADIATION DOSE RANGES FOR PROCEDURES IN UNIVERSITAS HOSPITAL VASCULAR LABORATORIES

PATIENT RADIATION DOSE RANGES FOR PROCEDURES IN UNIVERSITAS HOSPITAL VASCULAR LABORATORIES PATIENT RADIATION DOSE RANGES FOR PROCEDURES IN UNIVERSITAS HOSPITAL VASCULAR LABORATORIES Henra Muller N Dip Radiography (Diagnostic) B Tech Radiography (Diagnostic) Dissertation submitted in fulfilment

More information

DATA SHARING AGREEMENT

DATA SHARING AGREEMENT DATA SHARING AGREEMENT Effective Date: This Data Sharing Agreement ( Agreement ) is effective as of the Effective Date between EMD Serono, One Technology Place, Rockland, MA 02370 and Name ( Institution

More information

RADIATION DOSE MONITOR. The DACS for Building a Low-Dose Culture

RADIATION DOSE MONITOR. The DACS for Building a Low-Dose Culture RADIATION DOSE MONITOR The DACS for Building a Low-Dose Culture Radiation Dose Monitor / WHAT IS DACS? / DACS (Dose Archiving and Communication System) is to radiation dose reports what PACS is to DICOM

More information

RADIATION DOSE MONITOR

RADIATION DOSE MONITOR RADIATION DOSE MONITOR The DACS for Building a Low-Dose Culture Radiation Dose Monitor / WHAT IS DACS? / DACS (Dose Archiving and Communication System) is to radiation dose reports what PACS is to DICOM

More information

Clinical Applications of the Latest Ultra-High-Resolution CT and Iterative Reconstruction Algorithm: Things Radiologist and Radiographer Need to Know

Clinical Applications of the Latest Ultra-High-Resolution CT and Iterative Reconstruction Algorithm: Things Radiologist and Radiographer Need to Know Clinical Applications of the Latest Ultra-High-Resolution CT and Iterative Reconstruction Algorithm: Things Radiologist and Radiographer Need to Know Poster No.: C-1477 Congress: ECR 2018 Type: Educational

More information

GENDER PAY GAP REPORT MARCH 2018

GENDER PAY GAP REPORT MARCH 2018 GENDER PAY GAP REPORT MARCH 2018 CEO S MESSAGE Welcome to the first Electrocomponents plc Gender Pay Gap report. We re committed to creating an inclusive and diverse place to work where everyone can deliver

More information

Reducing Fluoroscopic Doses in Routine Musculoskeletal Joint Related Procedures

Reducing Fluoroscopic Doses in Routine Musculoskeletal Joint Related Procedures Reducing Fluoroscopic Doses in Routine Musculoskeletal Joint Related Procedures Mital Patel MD, Pranav Chitkara MD, Vikram Kinni MD, Rati Patel MD Naveen Subhas MD, Ryan Fisher PhD Disclosure No relevant

More information

Measuring Radiologist workload

Measuring Radiologist workload Measuring Radiologist workload Position of Faculty of Radiologists, RCSI Measurements used elsewhere and in the past Historically, the number of studies reported by individual radiologists was used as

More information

BBC Statutory Gender Pay Report 2017

BBC Statutory Gender Pay Report 2017 BBC Statutory Gender Pay Report 2017 BBC Statutory Gender Pay Report 2017 2 Statutory Gender Pay Report Contents 3 Introduction 4 Why equal pay and the gender pay gap are not the same 5 Gender at the BBC

More information

The Consistent Quality of Connected Radiology

The Consistent Quality of Connected Radiology The Consistent Quality of Connected Radiology Radiology is Evolving New compliance requirements, heightened radiation awareness and increasing technical complexity present new challenges to radiology professionals.

More information

In recent years, the benefits and complexity of interventional

In recent years, the benefits and complexity of interventional PATIENT SAFETY Patient Radiation Dose Management in the Follow-Up of Potential Skin Injuries in Neuroradiology E. Vano, J.M. Fernandez, R.M. Sanchez, D. Martinez, L. Lopez Ibor, A. Gil, and C. Serna-Candel

More information

REFERENCES LEVELS OF PATIENT RADIATION DOSE IN INTERVENTIONAL RADIOLOGY: A SPANISH CONTRIBUTION

REFERENCES LEVELS OF PATIENT RADIATION DOSE IN INTERVENTIONAL RADIOLOGY: A SPANISH CONTRIBUTION REFERENCES LEVELS OF PATIENT RADIATION DOSE IN INTERVENTIONAL RADIOLOGY: A SPANISH CONTRIBUTION F. Carrera 1,3, R. Ruiz-Cruces 3, E. Vano 2, J. Hernandez-Armas 4, M. Fernandez-Bordes 5, C. Ojeda 3 1 Radiological

More information

Application for Training Centre. BSET Fellowships

Application for Training Centre. BSET Fellowships Application for Training Centre BSET Fellowships Institution: St George s Vascular Institute, London, UK Correspondence: Professor Matt Thompson St George s Vascular Institute 4 th Floor St James Wing

More information

Reducing Unnecessary Medical Imaging Exposure

Reducing Unnecessary Medical Imaging Exposure Reducing Unnecessary Medical Imaging Exposure CDR Sean M. Boyd, MPH, USPHS U.S. Food and Drug Administration Center for Devices and Radiological Health April 30, 2010 Objectives FDA s role in ensuring

More information

Integrating the Healthcare Enterprise (IHE) Integration Statement

Integrating the Healthcare Enterprise (IHE) Integration Statement GE Healthcare Technical Publications Document Number: DOC1439176 Revision: 1 DoseWatch Version 1.3 Integrating the Healthcare Enterprise (IHE) Integration Statement Copyright 2013 by General Electric Company

More information

Dose Reduction during EVAR in Hybrid Room. Multicentric Study

Dose Reduction during EVAR in Hybrid Room. Multicentric Study Dose Reduction during EVAR in Hybrid Room Multicentric Study Golden rules to reduce radiation dose in the hybrid room to a minimum Stéphan Haulon, MD, Best Practice from Lille Aortic Center Aotic Centre,

More information

PRESS RELEASE. At Arab Health 2015, Agfa HealthCare showcases how its solutions meet the evolving needs of the Future of Radiology

PRESS RELEASE. At Arab Health 2015, Agfa HealthCare showcases how its solutions meet the evolving needs of the Future of Radiology PRESS RELEASE Agfa Press Office Septestraat 27 B 2640 Mortsel Belgium Marc De Fré Global Marcom Director Agfa HealthCare T +32 (0)3 444 73 19 marc.defre@agfa.com Johan Jacobs Corporate Press Relations

More information

NHSLA Risk Management Standards for NHS Trusts Providing Community Services 2011/12

NHSLA Risk Management Standards for NHS Trusts Providing Community Services 2011/12 NHSLA Risk Management Standards for NHS Trusts Providing Community Services 2011/12 Milton Keynes Primary Care Trust Provider of Community and Mental Health Services Level 1 May 2011 Contents Page 1: Executive

More information

Quantitative biomarker for clinical decision making

Quantitative biomarker for clinical decision making Deep imaging Quantitative biomarker for clinical decision making Joerg Aumueller English, October 2017 54% of healthcare leaders see an expanding role of in medical decision support.* * The future of healthcare

More information

ISSN Radiation. Protection. European Guidelines on Diagnostic Reference Levels for Paediatric Imaging. Energy

ISSN Radiation. Protection. European Guidelines on Diagnostic Reference Levels for Paediatric Imaging. Energy ISSN 2315-2826 Radiation Protection N 185 European Guidelines on Diagnostic Reference Levels for Paediatric Imaging Energy This report was prepared under contract ENER/2013/NUCL/Sl2.671441. The information

More information

LINC James F. McKinsey, M.D.

LINC James F. McKinsey, M.D. Two-Year Evaluation of Fenestrated and Parallel Branch Endografts for the Treatment of Juxtrarenal, Suprarenal and Thoracoabdominal Aneurysms at a Single Institution LINC 2018 James F. McKinsey, M.D. The

More information

Ultrasound (US)-guided percutaneous core biopsy of pancreatic tumor:tips and tricks to ensure safe and effective biopsy

Ultrasound (US)-guided percutaneous core biopsy of pancreatic tumor:tips and tricks to ensure safe and effective biopsy Ultrasound (US)-guided percutaneous core biopsy of pancreatic tumor:tips and tricks to ensure safe and effective biopsy Poster No.: C-2368 Congress: ECR 2014 Type: Scientific Exhibit Authors: J. W. Kim,

More information

New Joint Commission Requirements for Fluoroscopy Services

New Joint Commission Requirements for Fluoroscopy Services New Joint Commission Requirements for Fluoroscopy Services Melissa C. Martin, M.S., FAAPM, FACR Past President AAPM Florida Chapter Fall Meeting Orlando, FL September 14, 2018 New Requirements Effective

More information

Clinical Study Synopsis

Clinical Study Synopsis Clinical Study Synopsis This file is posted on the Bayer HealthCare Clinical Trials Registry and Results website or on the website www.clinicalstudyresults.org hosted by the Pharmaceutical Research and

More information