ENUMERATION OF LYMPHOCYTES SUBSETS (IMMUNOPHENOTYPING-IPT) (CD3, CD4, CD8, CD19 & CD16/56)
|
|
- Christopher Hardy
- 6 years ago
- Views:
Transcription
1 Hospital Universiti Sains Malaysia for ENUMERATION OF LYMPHOCYTES SUBSETS (IMMUNOPHENOTYPING-IPT) (CD3, CD4, CD8, CD19 & CD16/56) Prepared by: Checked by: Approved by: En. Jamaruddin Mat Asan Date: Dr. Noor Suryani Mohd Ashari Date: PM Dr. Che Maraina Che Hussin Date: Amendment / Review No. Date Page Description Checked by Approved by / 7 Reviewed document / 7 Amendment 1 : Add procedure
2 Page 2 / 7 Enumeration of Lymphocytes Subsets ( Immunophenotyping IPT) (CD3, CD4, CD8, CD19 & CD16/56) CONTENTS: 1. Technique used 2. Principle of the test method 3. Clinical significance 4. Preparation of specimen 5. Equipment requirement 6. Reagent requirement 7. Test procedure 8. Report and interpretation 9. Internal quality control 10. Reference
3 Page 3 / 7 1. Technique used Immunophenotyping using direct immunoflourescence (read with flowcytometer) 2. Principle When whole blood is added to the reagent, the flourochrome labeled antibodies in the reagent bind specifically to leukocyte surface antigens. The stained samples are treated with lysing solution to lyse erythrocytes. During acquisition, the cells travel past the laser beam and scatter the laser light. The stained cells fluoresce. These scatter and fluorescence signals, detected by the instrument, provide information about the cell s size, internal complexity and relative fluorescence intensity 3. Clinical significance 1. Immunodeficiency cases 2. Human Immunodeficiency Virus cases 4. Specimen ml whole blood in EDTA container 2. Paediatric samples minimum 1 ml 5. Equipment and materials requirement 1. Flow cytometer machine 2. Micropipette (20µl, 50µl, 450µl) 3. Disposable 12x75mm polystyrene test tubes 4. Haematology Counter 5. Vortex
4 Page 4 / 7 6. Reagent requirement BD Multitest IMK kit containing - Fluorochrome- conjugated monoclonal antibody - 10X FACS Lysing Solution * Dilute FACS lysing solution 1:10 with distilled water 7. Procedure Staining direct immunoflourescence of whole blood using lyse 1. For each patient sample, label two 12x75mm tubes with number 1 and 2 2. Pipette 20uL of BD Multitest CD3/CD8/CD45/CD4 reagent into the bottom of each tube labeled 1 3. Pipette 20uL of BD Multitest CD3/CD16+CD56/CD45/CD19 reagent into the bottom of each tube labeled 2 4. Pipette 50uL of well-mixed, anticoagulated whole blood into the bottom of each tube 5. Cap the tubes and vortex gently to mix. Incubate for 15 minutes in the dark at room temperature (20 0 C C) 6. Add 450 ul of 1X BD Multitest lysing solution to each tube 7. Cap the tubes and vortex gently to mix. Incubate for 15 minutes in the dark at room temperature (20 0 C C) Samples are now ready to be analyzed on the flow cytometer
5 Page 5 / 7 Amendment 1 Haematology Counter 1. Turn on the printer, CPU and main unit 2. Log on user name [husm], password [husm] and press [OK] 3. Click [F6] and followed by [F9] to register sample 4. Insert sample number (lab number), test required (CBC + DIFF), patient ID, patient name and click [OK] 5. Click [F2], insert sample number, patient ID and click [OK] 6. Mix the sample blood thoroughly by inversion several times 7. Insert the probe to the bottom of the tube and press start button 8. The green LED flashes during sample aspiration, then buzzer beeps 2 times to indicate end of aspiration. Remove the tube 9. Click [F7] to display the result 10. Go to file and select print to print the results 11. Click [F4] and double click the [shutdown]. Click [execute] 12. After the shutdown sequence is completed, turn off the power of main unit and computer
6 Flow cytometer FACSCanto II Page 6 / 7 1. Turn on the main power of the flow cytometer, computer and printer 2. Insert password BDIS 3. Click [BDFACSCanto] on the desktop to open the BD FACSCanto software. 4. Insert password husm. After opening the software, fluidics startup will be automatically processed. (If not, go to cytometer bar and select fluidics startup) 5. Wait for the processing finish, then click [OK] 6. Insert the [PATIENT NAME],[PATIENT ID(lab no)], [CASE NUMBER (test quantity)], [PANEL (choose 4 colour TBNK)] and [LYMPHS (lymphocyte count)]. 7. Click [RUN] from main menu. Click [YES] for the message would you like to save the changes you ve made to the current work list 8. Use the of the test done for the [FILE NAME] box 9. Then click [SAVE] 10. When the message show you may load the tube CD3/CD4/CD45/CD8, insert the 1 st tube at the Sample Injection Tube (SIT), then click [OK] 11. Remove the tube from SIT after the message shows you may now unload the tube 12. Insert the 2 nd tube when the message shows you may load the tube CD3/CD16+56/CD45/CD Remove the tube from SIT after the message shows you may now unload the tube 14. Double click to the bar [STATUS] to see the result on the screen 15. Click [FILE] and [PRINT] to print out the result 16. Click [FILE], then [EXIT] 17. Choose [RUN FLUIDICS SHUTDOWN AND EXIT] for the messages please select how you want to exit 18. Then click [OK] 19. Switch off the computer and main power of flow cytometer
7 Page 7 / 7 8. Internal Quality Control 1. Percentage of CD3 difference must be less then 5% 2. Percentage of T-Sum must be less then 10% 3. Percentage of Lymph sum count (T Cells, B Cells, NK Cells) must be % 9. Reporting and interpretations of result Normal values lymphocyte populations as a function of age These values are derived from studies on the Caucasian population. Therefore they can only be used as a rough guide for the local population Absolute count expressed in 10 3 cells mm -3 Data are expressed as median together with percentile P25 and P75 values. Previous studies have shown that the distributions of many lymphocyte subsets are asymmetric 10. Reference 1. Pamphlet of the BD Multitest IMK kit 2. Flow cytometer manual 3. Hematology counter manual
PERFECT-COUNT MICROSPHERES
PERFECT-COUNT MICROSPHERES Perfect-Count Microspheres-Product code PCB-100 for 100 tests Introduction In recent years, the determination of absolute cell counts has been shown to be relevant in different
More informationApplication Information Bulletin: DuraClone IM Tubes Compensation setup for high content DuraClone reagents
Application Information Bulletin: DuraClone IM Tubes Compensation setup for high content DuraClone reagents Compensation setup for high content DuraClone reagents INTRODUCTION High content flow cytometry
More informationIdentification of red and white blood cells from whole blood samples using the Agilent 2100 bioanalyzer. Application Note
Identification of red and white blood cells from whole blood samples using the Agilent 2100 bioanalyzer Application Note Sylvie Veriac Valérie Perrone Madeleine Avon Abstract Agilent Equipment: 2100 bioanalyzer
More informationBD Stabilizing Fixative
BD Stabilizing Fixative 04/2015 23-7725-05 IVD 3X concentrate Catalog No. 339860 BD, BD Logo and all other trademarks are property of Becton, Dickinson and Company. 2015 BD Becton, Dickinson and Company
More informationACTG Laboratory Technologist Committee Revised Version 2.0 ACTG Lab Man CD38 Quantitation 07 May 2004
1. Principle Flow Cytometric quantitation of CD38 expression on CD8+ T lymphocytes The purpose of this assay is to quantitate the surface expression of the CD38 molecule on CD8 positive T lymphocytes.
More informationImmunophenotyping of Peripheral Blood and Bone Marrow Cells by Flow Cytometry *Akanni EO and # Palini A.
Immunophenotyping of Peripheral Blood and Bone Marrow Cells by Flow Cytometry *Akanni EO and # Palini A. * Department of Haematology & Blood Transfusion,College of Health Science, Ladoke Akintola University
More informationFlow Cytometry Immune Activation SOP
Flow Cytometry Immune Activation SOP Purpose This SOP standardizes the procedure for measuring immune activation of T cells using flow cytometry in ACTG Immunology Laboratories. Materials 1. 12x75mm flow
More informationXS-1000i New Sysmex 5-part diff haematology analyser with fluorescence technology
XS-1000i New Sysmex 5-part diff haematology analyser Fig. 1: sysmex xs-1000i Since the introduction of fluorescence technology for leukocyte differentiation by sysmex in the year 1999, thousands of laboratories
More informationAQUIOS CL LOAD & GO AND MAKE FLOW CYTOMETRY SIMPLICITY ITSELF
LOAD & GO AND MAKE FLOW CYTOMETRY SIMPLICITY ITSELF FOCUS ON YOUR PATIENT. NOT ON SAMPLE PREPARATION Sample preparation and data management usually are the major bottlenecks in routine flow cytometry.
More information2. SUMMARY AND EXPLANATION
English Stem-Trol Control Cells 1-10 1. INTENDED USE 2 2. SUMMARY AND EXPLANATION 2 3. PRINCIPLE OF TEST 2 4. REAGENT CONTENTS 2 5. STATEMENT OF WARNINGS 2 6. STORAGE CONDITIONS AND STABILITY 3 6.1 Evidence
More informationHigh-throughput automation with the Attune NxT Autosampler: consistent results across all wells and across plates
APPLICATION NOTE Attune NxT Flow Cytometer with Autosampler High-throughput automation with the Attune NxT Autosampler: consistent results across all wells and across plates Introduction The emerging field
More informationDepartment of Laboratories St. Louis, MO ISSUE DATE: June 2007 REVISION DATE: May 2012 REVIEWED DATE: May 2018 PRINCIPLE:
Page 1 of 7 PROCEDURE: LEAD ESA LEADCARE II / HEALTHY KIDS EXPRESS ISSUE DATE: June 2007 REVISION DATE: May 2012 REVIEWED DATE: May 2018 PRINCIPLE: The LEADCARE II System relies on electrochemistry and
More informationInhibiScreen Kinase Inhibitor Assay
InhibiScreen Kinase Inhibitor Assay Flow Cytometric assessment of efficacy of PI3Kγ, PI3Kδ, BTK and SYK pathway inhibitors using the measurement of basophil activation in human EDTA and Heparin Whole Blood.
More informationProtocol for FACS analysis of HeLa cell transfectants
Protocol for FACS analysis of HeLa cell transfectants You can refer to: Marks et al., 1995, J. Cell Biol. 131: 351-369; Voorhees et al., 1995, EMBO J. 14: 4961-4975; or Marks et al., 1996, J. Cell Biol.
More informationFlow Cytometry SOP: 14 color flow for immune activation, senescence, and exhaustion
Flow Cytometry SOP: 14 color flow for immune activation, senescence, and exhaustion Purpose This SOP standardizes the procedure for measuring immune cells using flow cytometry in ACTG Immunology Laboratories.
More informationFlow Cytometry SOP: Monocytes from Frozen Cells
Flow Cytometry SOP: Monocytes from Frozen Cells Purpose This SOP standardizes the procedure for measuring immune cells using flow cytometry in ACTG Immunology Laboratories. Materials 1. 12x75mm flow tubes
More informationRef: CYT-MM-MRD. For Research Use Only. Not for use in diagnostic procedures
Ref: CYT-MM-MRD For Research Use Only. Not for use in diagnostic procedures MM MRD VIALS AND COMPENSATION TUBES ARE LYOPHILIZED PRODUCTS. READ CAREFULLY THE FOLLOWING INSTRUCTIONS FOR RECONSTITUTION: The
More informationPrinciples of Immunophenotyping
Principles of Immunophenotyping % of Cell types? Immune activation? Changes based on health state? Moving from a heterogeneous population of blood cells to identifying the presence and proportion of different
More informationTECHNICAL BULLETIN. QUANTUM SIMPLY CELLULAR KIT Product Numbers QSC-20 AND QSC-100 Storage Temperature 2-8 C. Do Not Freeze
QUANTUM SIMPLY CELLULAR KIT Product Numbers QSC-20 AND QSC-100 Storage Temperature 2-8 C. Do Not Freeze TECHNICAL BULLETIN Product Description The Quantum Simply Cellular Kit provides a convenient method
More informationKit Contents FL-10 FL-12 FL-07 FL-09. Important: Before You Begin. New for this Mailing. Reporting Code Changes. Critical Reporting Information
FL-B 2014 Kit Instructions CAP 2014 Flow Cytometry Survey FL FL1 FL2 Table of Contents Kit Contents Kit Contents... 1 Important: Before You Begin... 1 Detailed Testing Instructions... 2 FL FL1 FL2 FL-07
More informationBD FACSCalibur. The flow cytometer for your routine cell analysis needs
BD FACSCalibur The flow cytometer for your routine cell analysis needs A system with a rich application basis and a modular approach that continues to meet evolving needs of cell analysis worldwide. The
More information18/08/2011. Principles of Flow Cytometry (Practised in a Clinical Laboratory) Cytometer Components. Noel Williams Immunobiology Division of Immunology
Optical Measurement Principles Principles of Flow Cytometry (Practised in a Clinical Laboratory) Noel Williams Immunobiology Division of Immunology Cytometer Components Reagents Cytometer Setup Cytometer
More informationPURPOSE: To delineate the subsets of human lymphocytes based on the expression profiles of different phenotypic markers by FACS analysis
LABORATORY PROCEDURE: IMMUNOPHENOTYPING: Lymphocyte Staining for FACS Analysis Date: April 29 2014 Authors: Jennifer Hossler PURPOSE: To delineate the subsets of human lymphocytes based on the expression
More informationApplication Note. Assay Portability on the BD FACSVerse System. Summary. Maria Jaimes, Yibing Wang, Catherine McIntyre, and Dev Mittar
September Assay Portability on the BD FACSVerse System Maria Jaimes, Yibing Wang, Catherine McIntyre, and Dev Mittar Contents Summary Introduction 3 Objective 4 Methods 6 Results Discussion Conclusions
More informationNovoCyte Flow Cytometer
NovoCyte Flow Cytometer The Flow Cytometer for Everyone 2 Experience the NovoCyte Advantage Focus on advancing your research. Let the flow cytometer do the rest. NovoCyte Flow Cytometer High Performance
More informationTITLE: LIVE/DEAD VIABILITY FOR CLINICAL SAMPLES
Paul K. Wallace, Ph.D. Director Roswell Park Cancer Institute Elm & Carlton Streets Voice:(716) 845-8471 Buffalo, NY 14263 Fax:(716) 845-8806 FILE NAME FL-SRP-2090.00 Live/Dead Viability for Clinical Samples
More informationPRINCIPLES OF THE PROCEDURE
CD19 For In Vitro Diagnostic use Fluorochrome Reference Clon Isotype Format µl/test FITC CYT-19F6 HIB19 IgG1 1 ml/200 test 5 PerCP-Cyanine5.5 CYT-19C8 HIB19 IgG1 150ul/50 test 3 PE-Cyanine5 CYT-19C3 HIB19
More informationBD Multicolor CompBeads
4/2015 23-9955-01 IVD BD Multicolor CompBeads 100 compensation setups Catalog No. 644204 BD, BD Logo and all other trademarks are property of Becton, Dickinson and Company. 2015 BD Becton, Dickinson and
More informationSupplementary File 3: DNA and RNA isolation
Supplementary File 3: DNA and RNA isolation Q-CROC-02 Biopsy protocol For the purposes of this protocol, four needle core biopsies (NCBs) of lymph node tissue are isolated from each patient using a 16G
More informationNavios EX FLOW CYTOMETER POWERFUL, DEPENDABLE CLINICAL FLOW CYTOMETRY
Navios EX FLOW CYTOMETER POWERFUL, DEPENDABLE CLINICAL FLOW CYTOMETRY BECAUSE EVERY EVENT MATTERS The Navios EX flow cytometer offers a solution for advanced cytometry applications optimized for the clinical
More informationLambda-APC-C750 Ref: CYT-LAC750
Lambda-APC-C750 Ref: CYT-LAC750 For In Vitro Diagnostic use INTENDED USE Lambda-APC-C750 is a polyclonal antibody (pab) labelled with the tandem allophycocyanine-c750 (APC- C750) and designed for use as
More informationRPCI 001 v.003 In vitro Intracellular Cytokine Staining With and Without Stimulation
Immune Tolerance Network RPCI 001 v.003 Author: Paul Wallace and Earl Timm, Director, RPCI Laboratory of Flow Cytometry Approved by: Paul Wallace, Director, RPCI Laboratory of Flow Cytometry 1.0 Title
More informationdetermine optimum instrument settings for their own instruments and establish their own daily values.
PC7 (770/488) SETUP KIT 6607121 PN 4299504-C FLOW CYTOMETER ALIGNMENT VERIFICATION FLUOROSPHERES FLOW CYTOMETER DETECTOR STANDARDIZATION FLUOROSPHERES INTENDED USE For Research Use Only. Not for use in
More informationEstablishing sirna assays in primary human peripheral blood lymphocytes
page 1 of 7 Establishing sirna assays in primary human peripheral blood lymphocytes This protocol is designed to establish sirna assays in primary human peripheral blood lymphocytes using the Nucleofector
More informationSee Brilliant Results
See Brilliant Results without all the reviews. * In Development Product not available in the U.S. First Pass Efficiency. Getting it right the first time. GOALS: More reportable WBC and WBC differential
More informationFLOW CYTOMETRY. CyAn ADP. Analyzer
FLOW CYTOMETRY CyAn ADP Analyzer Experience the Power of the CyAn ADP and its optimal performance The Power of Detection The Power of Speed The Power of Ease The CyAn ADP Analyzer is the next step in Advanced
More informationBD IntraSure Kit IVD. Cell fixation and permeabilization reagents. 50 tests per kit Catalog No
03/2015 23-8992-01 IVD BD IntraSure Kit Cell fixation and permeabilization reagents 50 tests per kit Catalog No. 641778 BD, BD Logo and all other trademarks are property of Becton, Dickinson and Company.
More informationCollection Manager Intraoperative Specimen Labeling Project
Collection Manager Intraoperative Specimen Labeling Project Specimen mislabeling is a persistent problem across many clinical settings at UCSF, particularly in the fast-paced and often stressful environment
More informationTitration of Fluorochrome-Conjugated Antibodies for Labeling Cell Surface Markers on Live Cells
Titration of Fluorochrome-Conjugated Antibodies for Labeling Cell Surface Markers on Live Cells Ruud Hulspas 1 UNIT 6.29 1 Cytonome/ST, Boston, Massachusetts ABSTRACT Nonspecific antibody binding is best
More informationSet-up, programming and analysis MagNA Pure 96 (Roche)
Set-up, programming and analysis MagNA Pure 96 (Roche) A. Start-up Before preparing the run, start the instrument and the control unit v 1. Starting the MagNA Pure 96 Software To start and log on to the
More informationINTENDED PURPOSE. TECHNICAL SUMMARY.
Product: FITC Anti-Human CD2 Cat. Ref: 2F-100T Reagent provided: 100 test (20μl/test) Description: Mouse Monoclonal Anti-Human CD2 FITC is recommended for use in flow cytometry for identification of peripheral
More informationSimplexa EBV Quantitation Standards
Simplexa EBV Quantitation Standards REF MOL2410 Rev. C Simplexa EBV Quantitation Standards are intended to establish a standard curve with the Simplexa EBV assay on the 3M Integrated Cycler. For in vitro
More informationReagent Titration for the ICS Assay
Purpose This standard operating procedure (SOP) describes how to titrate new lots of fluorochromeconjugated antibody reagents and the cell viability marker for use in the intracellular cytokine staining
More informationLONDON HEALTH SCIENCES CENTRE (LHSC) WORKFLOW COMPARISON STUDY WITH THE AQUIOS CL FLOW CYTOMETER
CASE WHITE STUDY PAPER LONDON HEALTH SCIENCES CENTRE (LHSC) WORKFLOW COMPARISON STUDY WITH THE AQUIOS CL FLOW CYTOMETER Laboratory Profile London Health Sciences Centre (LHSC), Flow Cytometry Laboratory,
More informationCD34 Stem Cell Programme Sample All Participant Report. Date Issued - 06 March 2017 Closing Date - 24 March 2017 Machine Used - Facscanto II
Distribution - 16176 CD3 Stem Cell Programme - 3 All Participant Report Participant ID - XXXX Date Issued - 6 March 17 Closing Date - March 17 Machine Used - Facscanto II Trial Comments This programme
More informationApplication Note. Yeast Analysis with a life - dead staining kit (Yeast control - viability)
Application Note Yeast Analysis with a life - dead staining kit (Yeast control - viability) (CyFlow Cube 8 and CyFlow Cube 6 with CyView TM version 1.5 CyFlow Space) Introduction The Partec Yeast control
More informationPowERFUL. INTUITIvE. customizable. ThE NEw STAR of. BENchToP FLow cytometry
PowERFUL. INTUITIvE. customizable. ThE NEw STAR of BENchToP FLow cytometry ABoUT AcEA Established in 2002, ACEA Biosciences, Inc. develops cutting-edge cell analysis platforms for life science research.
More informationMaxPar Cytoplasmic/Secreted Antigen Staining Protocol
MaxPar Cytoplasmic/Secreted Antigen Staining Protocol CHEMICAL HAZARD: Before handling any chemicals, refer to the Material Safety Data Sheet (MSDS) provided by the manufacturer, and observe all relevant
More informationMinimum Information about a Flow Cytometry Experiment (MIFlowCyt) Annotation
Minimum Information about a Flow Cytometry Experiment (MIFlowCyt) Annotation 1. Experiment Overview 1.1 Purpose The purpose of these sets of experiments is to develop a methodology of identifying and quantifying
More informationBD IMag. Streptavidin Particles Plus - DM. Technical Data Sheet. Product Information
Technical Data Sheet Streptavidin Particles Plus - DM Product Information Material Number: Size: Storage Buffer: 557812 5 ml Aqueous buffered solution containing BSA and 0.09% sodium azide. Description
More informationNavios EX FLOW CYTOMETER POWERFUL, DEPENDABLE FLOW CYTOMETRY
Navios EX FLOW CYTOMETER POWERFUL, DEPENDABLE FLOW CYTOMETRY BECAUSE EVERY EVENT MATTERS The Navios EX flow cytometer offers a solution for advanced cytometry applications with optimized workflows for
More informationBD Stem Cell Enumeration Application Guide
BD Stem Cell Enumeration Application Guide for BD FACSCalibur Flow Cytometers IVD For In Vitro Diagnostic Use bdbiosciences.com 23-11206-01 Rev. 01 12/2011 Becton, Dickinson and Company BD Biosciences
More informationProtein and transcriptome quantitation using BD AbSeq Antibody-Oligonucleotide
Protein and transcriptome quantitation using BD AbSeq Antibody-Oligonucleotide technology and the 10X Genomics Chromium Single Cell Gene Expression Solution Jocelyn G. Olvera, Brigid S. Boland, John T.
More informationInstitute of Science & Technology in Medicine Control of Substance Hazardous to Health COSHH assessment
Institute of Science & Technology in Medicine Control of Substance Hazardous to Health COSHH assessment Title of Experiment/Procedure: Cell counting and viability using FACS, counting beads and propidium
More informationFor the preparation of purified mononuclear immune cells directly from whole blood PACKAGE INSERT. For In Vitro Diagnostic Use PI-TS-IVD-UK-V3
For the preparation of purified mononuclear immune cells directly from whole blood PACKAGE INSERT For In Vitro Diagnostic Use Intended Use The T-Cell Select kit is intended for the isolation of mononuclear
More informationStrategies for Assessment of Immunotoxicology in Preclinical Drug Development
Strategies for Assessment of Immunotoxicology in Preclinical Drug Development Rebecca Brunette, PhD Scientist, Analytical Biology SNBL USA Preclinical Immunotoxicology The study of evaluating adverse effects
More informationAffymetrix Processing of ChIP or DamID DNA to GeneChip Drosophila Tiling 2.0R Array
http://www.flychip.org.uk/affy_chip.php Affymetrix Processing of ChIP or DamID DNA to GeneChip Drosophila Tiling 2.0R Array Outline For the following protocol 7.5 µg DNA containing dutp (see the corresponding
More informationAtherosclerosis Risk in Communities Carotid MRI Study Manual 6 Flow Cytometry Laboratory
Atherosclerosis Risk in Communities Carotid MRI Study Manual 6 Flow Cytometry Laboratory Prepared by the ARIC Carotid MRI Study Investigators For Copies, Please Contact ARIC Coordinating Center Department
More informationNo-wash, no-lyse detection of phagocytic cells via a phrodo BioParticles functional assay in human whole blood on the
APPLICATION NOTE Attune NxT Flow Cytometer No-wash, no-lyse detection of phagocytic cells via a phrodo BioParticles functional assay in human whole blood on the Attune NxT Flow Cytometer Introduction Analysis
More informationAntibody Titrations. Institut für HIV Forschung SOP #06-03 (Nov BS) Background. Reagents
Institut für HIV Forschung SOP #06-03 (Nov 2017 - BS) Antibody Titrations Reagents Reagent Vendor Catalogue # Stock Conc. FACS Tubes BD Falcon 352054 - Anti-CD28/CD49d Antibody BD Fastimmune 347690 - GolgiPlug
More informationFor the preparation of purified mononuclear immune cells directly from whole blood PACKAGE INSERT. For In Vitro Diagnostic Use PI-TS-IVD-UK-V4
For the preparation of purified mononuclear immune cells directly from whole blood PACKAGE INSERT For In Vitro Diagnostic Use Intended Use The T-Cell Select kit is intended for the isolation of mononuclear
More informationPACKAGE INSERT. For use in the preparation and isolation of purified lymphocytes directly from whole blood. For In Vitro Diagnostic Use
For use in the preparation and isolation of purified lymphocytes directly from whole blood PACKAGE INSERT For In Vitro Diagnostic Use PI-TT.610-UK-V6 Instructional Information Intended Use The T-Cell Xtend
More informationMaxpar Phospho-Protein Staining
PRD016 Version 3 PROTOCOL Maxpar Phospho-Protein Staining WARNING Before handling any chemicals, refer to the Safety Data Sheet (SDS) provided by the manufacturer, and observe all relevant precautions.
More informationThis kit is intended for Research Use Only. Not for use in diagnostic procedures.
This kit is intended for Research Use Only. Not for use in diagnostic procedures. Introduction The DRG Herpes Simplex Virus Type 1+2 IgG Enzyme Immunoassay Kit provides materials for determination of IgG-class
More informationCFSE Cell Division Assay Kit
CFSE Cell Division Assay Kit Item No. 10009853 www.caymanchem.com Customer Service 800.364.9897 Technical Support 888.526.5351 1180 E. Ellsworth Rd Ann Arbor, MI USA TABLE OF CONTENTS GENERAL INFORMATION
More informationQUICK REFERENCE INSTRUCTIONS For use with Sofia and Sofia 2.
QUICK REFERENCE INSTRUCTIONS For use with Sofia and Sofia 2. Study the Package Insert and User Manual thoroughly before using Quick Reference Instructions. This is not a complete Package Insert. CLIA Complexity:
More informationRuud Hulspas, Mike Keeney, Ben Hedley and Andrea Illingworth
Sponsored and reviewed by ICCS Quality and Standards Committee Title: Quality of Reagents Monoclonal Antibodies Written by: Date: March 6, 2018 Ruud Hulspas, Mike Keeney, Ben Hedley and Andrea Illingworth
More informationDetecting human circulating endothelial cells using the Attune Acoustic Focusing Cytometer
APPLICATION NOTE Attune Acoustic Focusing Cytometer Detecting human circulating endothelial cells using the Attune Acoustic Focusing Cytometer Circulating endothelial cells (CECs) are mature cells shed
More informationCell surface antibody staining for flow cytometry with optional amine reactive dye. Type SOP
1 of 5 1.0 PURPOSE 1.1 This procedure describes the process for cell surface staining with fluorescent conjugated antibody reagents and optional amine reactive dye for viability measurement 2.0 SCOPE 2.1
More informationTime Clock Time Clock
The Last option on the Times/Commissions menu is the Time Clock button. The Time Clock feature allows you to track employees hours. Employees clock in and out, and the system tracks their hours. Employees
More informationMedonic M-series M32. User s Manual
Medonic M-series M32 User s Manual Page ii Table of Contents TABLE OF CONTENTS Section 1. Introduction 5 Medonic M-series M32 Systems... 5 Contact Details... 5 Analyzer Overview... 6 Consumable Overview...
More informationGRAM STAIN/CULTURE of CSF and OTHER BODY FLUIDS by HEMATOLOGY LAB (11:30pm 07:30am)
Created Rebecca Riemann Clin. Lab 5/22/2012 Tech 1 Revised Eric Spitzer Chief of 5/22/2012 Micro/Assoc. Dir Labs Reviewed Jay Bock M.D. 5/22/2012 Approved Eric Spitzer Chief of Micro/Assoc. Dir Labs mic
More informationUniversity of Illinois HR Operations Scanner Survey Respondent Overview and Instructions
Introducing the University of Illinois The HR Operations Scanner (Scanner) is an activity-based survey that is used to help understand the range of tasks performed by HR at the University of Illinois and
More informationSAMPLE PROCESSING AND PCR SETUP ON THE GS1
SAMPLE PROCESSING AND PCR SETUP ON THE GS1 For use with the EasyScreen Sample Processing Kit (SP003) and EasyScreen Detection Kit(s). PROCESSING STEPS 1. Turn on and set heat block to 95 C. 2. Add Reagent
More informationPCCS Growth Media, Cell Tagging, Cell Separation Final Assignment. Igneris Rosado-Erazo. Panama College of Cell Science
Running Head: Growth Media, Cell Tagging, Cell Separation PCCS Growth Media, Cell Tagging, Cell Separation Final Assignment Igneris Rosado-Erazo Panama College of Cell Science In partial fulfillment of
More informationDRG International, Inc., USA Fax: (908)
Please use only the valid version of the package insert provided with the kit. This kit is intended for Research Use Only. Not for use in diagnostic procedures. 1 INTENDED USE AND PRINCIPLE OF THE TEST
More informationQuality Control in Flow. Dr David Westerman Head of Haematopathology Peter MacCallum Cancer Centre
Quality Control in Flow Dr David Westerman Head of Haematopathology Peter MacCallum Cancer Centre Aims Quality Assurance Quality Control Literature In house competencies SHOT DATA 1996-2009 Ref: SHOT Annual
More informationCyfra 21-1 IRMA. Product information Information about other products is available at: Userś Manual DE52100
Product information Information about other products is available at: www.demeditec.com Userś Manual Cyfra 21-1 IRMA The CYFRA 21.1 IRMA system provides a direct in vitro quantitative determination of
More informationApplication Protocol: CD45 CK Immunostaining for patient blood
REPRODUCTION AND USE This document is protected by copyright and it cannot be used or shared without permission from Vortex Biosciences, Inc. Such permission is given on condition that Vortex Biosciences
More informationQUICK USER GUIDE CAPILLARY BLOOD INR TESTING FOR USING THE THROMBI-STAT MC1 WB INSTRUMENT MANCHESTER CAPILLARY REAGENT INTRODUCTION AND
QUICK USER GUIDE FOR CAPILLARY BLOOD INR TESTING USING THE THROMBI-STAT MC1 WB INSTRUMENT AND MANCHESTER CAPILLARY REAGENT INTRODUCTION The combination of the Hart Biologicals Thrombi-stat MC1WB coagulation
More informationFlow Cytometric Devices Draft Guidance for Industry and Food and Drug Administration Staff DRAFT GUIDANCE
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 36 Flow Cytometric Devices Draft Guidance for Industry and Food and Drug Administration Staff DRAFT GUIDANCE This guidance document
More informationE. coli Phagocytosis Assay Kit
E. coli Phagocytosis Assay Kit Item No. 601370 www.caymanchem.com Customer Service 800.364.9897 Technical Support 888.526.5351 1180 E. Ellsworth Rd Ann Arbor, MI USA TABLE OF CONTENTS GENERAL INFORMATION
More informationLab 2. Isolation of mononuclear cells from peripheral blood and separation into subpopulations.
Lab 2 Isolation of mononuclear cells from peripheral blood and separation into subpopulations. Supervisors: Ulrika Andreasson ulrika.andreasson@immun.lth.se tel: 222 9264 Niclas Olsson niclas.olsson@immun.lth.se
More informationCellular Immunology. Leukocyte Function
Clinical Laboratory s of Practice All laboratories shall comply with the applicable requirements contained in the Clinical Laboratory s of Practice- General Systems. In addition, the laboratory shall meet
More informationMagniSort Mouse CD3 Positive Selection Kit Catalog Number: RUO: For Research Use Only. Not for use in diagnostic procedures.
Page 1 of 2 MagniSort Mouse CD3 Positive Selection Kit RUO: For Research Use Only. Not for use in diagnostic procedures. Mouse splenocytes were unsorted (left) or sorted with the MagniSort Mouse CD3 Positive
More informationStandard Operation Procedure (SOP) for Biobanking Sampling Procedure Manual Use
Standard Operation Procedure (SOP) for Biobanking Sampling Procedure Manual Use 1. Oragene TM DNA Purification Protocol for use with Self-Collection kit of human sliva samples, OG-500 1.1. Equipment and
More informationThe NIAID Flow Cytometry Advisory Committee; the Guidelines Subcommittee
January, 1999 From: To: Subject: The NIAID Flow Cytometry Advisory Committee; the Guidelines Subcommittee NIAID DAIDS Flow Cytometry Laboratories Comparison study information for labs wishing to switch
More informationPhagocytosis Assay Kit (IgG PE)
Phagocytosis Assay Kit (IgG PE) Item No. 600540 www.caymanchem.com Customer Service 800.364.9897 Technical Support 888.526.5351 1180 E. Ellsworth Rd Ann Arbor, MI USA TABLE OF CONTENTS GENERAL INFORMATION
More informationCD45 (2D1) IVD. Table 1 Bottling concentrations. Monoclonal mouse anti-human reagent for identification of cells expressing the CD45 antigen
04/2015 23-97-06 IVD CD45 (2D1) Monoclonal mouse anti-human reagent for identification of cells expressing the CD45 antigen Form Catalog No. FITC 345808 PerCP 345809 PerCP-Cy5.5 332784 APC 340910 APC-Cy7
More informationFeatures & Benefits Five-part differential on cat, dog, horse, cow, alpaca and llama; three-part differential on the nine other species
Telephone: Facsimile: e-mail: + 27 (0)11 791 0025 + 27 (0)86 551 0832 johan@anjotech.co.za VetScan HM5 Hematology System Advanced Hematology Five-Part Differential The VetScan HM5 is a fully automated
More informationHuman hepatitis E virus antibody (IgG) ELISA Kit
Human hepatitis E virus antibody (IgG) ELISA Kit Catalog Number. NBP2-60048 For the qualitative determination of human hepatitis E virus antibody (IgG) concentrations in serum, plasma. This package insert
More informationEach question may have MULTIPLE correct answers. Select all that are correct.
Knowledge Assessment Flow Cytometry Workshop, Part 1 April 20, 2015 Each question may have MULTIPLE correct answers. Select all that are correct. 1. Tandem dyes are a. highly stable fluorophores after
More informationHuman soluble TREM-2 ELISA Kit
Human soluble TREM-2 ELISA Kit Cat.No: DEIA-SY005 Lot. No. (See product label) Size 96T Intended use For the quantitative measurement of human soluble TREM-2 concentrations in serum, plasma. General Description
More informationHuman Beta 2-Microglobulin ELISA Kit
Human Beta 2-Microglobulin ELISA Kit CATALOG NO: IRKTAH1125 LOT NO: SAMPLE INTENDED USE The Beta 2-Microglobulin test kit is a highly sensitive two-site enzyme linked immunoassay (ELISA) for measuring
More informationFluorochrome: Fluorescein isothiocyanate (Molecular Probes) INTENDED PURPOSE.
Product: FITC Anti-Human CD14 Cat. Ref: 14F-100T Reagent provided: 100 test (20μl / test) Description: Monoclonal Mouse Anti-Human CD14 FITC is recommended for use in flow cytometry for identification
More informationB2M (Human) ELISA Kit
B2M (Human) ELISA Kit Catalog Number KA0222 96 assays Version: 02 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Background... 3 Principle of the Assay...
More informationHuman free Fatty Acid (FFA) ELISA kit
Human free Fatty Acid (FFA) ELISA kit Cat.No: DEIA3480 Lot. No. (See product label) Size 96T Intended Use This CD Free Fatty acid ELISA kit is a 1.5 hour solid-phase ELISA designed for the quantitative
More informationTECHNICAL BULLETIN. QUANTUM FLUORESCENCE KITS FOR MESF UNITS OF FITC Product Numbers QMF-2 AND QMF-10 Storage Temperature 2-8 C.
QUANTUM FLUORESCENCE KITS FOR MESF UNITS OF FITC Product Numbers QMF-2 AND QMF-10 Storage Temperature 2-8 C. Do Not Freeze TECHNICAL BULLETIN Product Description Quantum Fluorescence Kits for MESF Units
More information