High level of agreement between CP3000 and five other coagulation analyzers for INR in a multicenter study
|
|
- Erik Burke
- 6 years ago
- Views:
Transcription
1 DIAGNOSTICS Gabriella Lakos, MD, PhD. Sr. Associate Medical Director Medical & Scientific Affairs Abbott Diagnostics Larry Smith, PhD. Medical & Scientific Affairs Manager Abbott Diagnostics High level of agreement between CP000 and five other coagulation analyzers for INR in a multicenter study July 07 Sarah Just South West Area Pathology Service Liverpool Hospital Sydney, Australia Stephen So Shing Chau Clinical & Histopathology Laboratory St. Teresa s Hospital Kowloon, Hong Kong Ivana Malikova General University Hospital (Všeobecná Fakultní Nemocnice v Praze) Prague, Czech Republic Volker Drescher Medizinisches Labor Westsachsen Fachlabor Zwickau, Germany 5 Peter Reichardt Medizinisches Labor Westsachsen Fachlabor Zwickau, Germany Neda Razavi Marketing Manager Abbott Laboratories, Santa Clara, CA USA Abbott Diagnostics, Abbott Park, Illinois 600. Authors contributed equally to this document. Page of 7
2 Objectives In this multicenter study, the performance of the Sekisui CP000 Coagulation System (with Coagpia PT-N reagent) was assessed for prothrombin time (PT) and International Normalized Ratio (INR) in comparison with that of the ACL TOP-00 (Instrumentation Laboratories), Cobas t (Roche), Siemens BCS (Siemens), Sysmex CA-7000 (Siemens) and STAGO STA-R Evolution (Diagnostica STAGO) in routine laboratory settings. Methods A total of 95 samples (up to 0 patient samples per comparative instrument) were assayed by the CP000. Results were compared using regression statistics, Pearson correlation and qualitative agreement methods. Results The overall tested range spanned from 9.9 to 77. seconds. PT results obtained with the CP000 analyzer showed excellent correlation with those obtained with each comparative system: the Pearson correlation coefficient ranged from 0.96 (ACL TOP-00 and Cobas) to 0.99 (Siemens BCS) (Table ). Correlation coefficients for INR results ranged from 0.95 (Cobas) to.0 (Siemens BCS and STA-R Evolution) (Table ). Passing and Bablok regression analysis resulted in slopes of.0 (Siemens BCS) to. (Cobas) for INR. The percentage of samples in the INR range of.0 to.0 with < 0.5 difference in pairwise comparison (CLSI H57-A criteria) ranged from 50% (in comparison with ACL TOP-00) to 00% (in comparison with Siemens BCS). The mean difference in the therapeutic range (.0-.0 INR) ranged from (STA-R Evolution) to 0. (Roche t). Formal throughput testing with routine samples containing 7% specimens with prolonged PT resulted in 55 PT/hour performances. Based on 5 individual tests, average time to first (PT) result was minutes and seconds, and for a panel of PT, activated partial thromboplastin time (APTT) and fibrinogen concentration was 5 minutes and 7 seconds. Conclusion INR results obtained with the CP000 analyzer showed high level of agreement with those generated with the Siemens BCS, Sysmex CA-7000, STAGO STA-R Evolution, ACL TOP-00 and Cobas t instruments. Page of 7
3 Objectives This multicenter study compared prothrombin time (PT) and International Normalized Ratio (INR) results obtained with the CP000 Coagulation System (Sekisui Medical Co. Ltd, Tokyo) and Coagpia PT-N (Sekisui) reagent to those obtained with five different coagulation platforms: ACL TOP-00 (Instrumentation Laboratories), Cobas t (Roche), BCS (Siemens), CA-7000 (Sysmex/Siemens), and STA-R Evolution (Diagnostica STAGO). In addition, the study assessed the throughput and turnaround time of STAT results on CP000. Methods A total of 99 samples (up to 0 samples per comparative instrument) were assayed in four clinical laboratories with the CP000 and the comparative instruments. The following calculations were performed on these data: ) Passing-Bablok regression analysis ) Pearson correlation ) Predicted bias calculation at medical decision levels ) Percent of samples in the INR range of.0-.0 with <0.5 difference according to CLSI H57-A guideline 5) Mean difference between INR values in the INR range of.0-.0 Throughput was assessed for PT with normal and abnormal samples, and time to first result was measured for PT alone processed as a STAT sample, and for a panel of PT, activated partial thromboplastin time (APTT), and fibrinogen assays. Results Pearson correlation coefficients were 0.95 for PT and INR (Table and Figure A-E) in all comparisons. The slope and intercept of the regression line suggested the strongest concordance of CP000 results with those obtained with Siemens BCS and Sysmex CA-7000 (both with Thromborel S reagent), and the least amount of concordance with those obtained with Cobas t and STA-R Evolution. Slope and intercept values, however, significantly improved in these two comparisons when INR was considered instead of PT (Table ). Page of 7
4 INR - CP000 INR - CP000 Table. Correlation results according to comparative instrument. Reagents are shown in parenthesis. Comparative system n Range tested Passing-Bablok regression Pearson s r ACL TOP-00 (RecombiPlastin G) Roche Cobas t (Coasys thromboplastin) Siemens BCS (Thromborel S) Sysmex CA-7000 (Thromborel S) STAGO STA-R Evolution (Neoplastine CI Plus) Slope Intercept PT (sec) INR PT (sec) INR PT (sec) INR PT (sec) INR PT (sec) INR INR - ACL TOP INR - Roche Cobas t Page of 7
5 INR - CP000 INR - CP000 INR - CP INR - BCS INR - CA INR - STAGO STA-R-Evolution Figure. INR correlation across platforms. Predicted bias at medical decision levels was the smallest in comparison with Siemens BCS, STA- R Evolution and Sysmex CA-7000 (Table ). Table. Predicted bias Predicted bias according to comparator system INR medical decision levels ACL TOP-00 Roche Cobas t Siemens BCS Sysmex CA-7000 STAGO STA-R Evolution Page 5 of 7
6 The average difference between INR values in the INR range of.0-.0 was <0.5 in each comparison (Table ). CLSI guideline H57-A considers two PT methods equivalent if at least 85% of the samples in the INR range of.0-.0 exhibit < 0.5 difference. This criterion was fulfilled in comparison with Siemens BCS and STAGO STA-R Evolution, and it was not fulfilled in comparison with Cobas t. The number of samples in the INR range of.0-.0 was and 6 for ACL TOP-00 and Sysmex CA-7000, respectively, so the results are inconclusive (Table ). Table. Average and pair-wise differences Comparative system CP000 vs comparative instrument n samples in INR range of.0-.0 % samples in INR range of.0-.0 with < 0.5 difference Average difference in INR range of ACL TOP % 0.7 Roche Cobas t 8.6% 0. Siemens BCS % 0.0 Sysmex CA % 0.5 STAGO STA-R Evolution % Formal throughput testing with routine samples containing 7% specimens with prolonged PT resulted in 55 PT/hour performance. Based on 5 individual tests, average time to first PT result was minutes and seconds, and for a panel of PT, APTT and fibrinogen concentration was 5 minutes and 7 seconds. Conclusion Results obtained from these studies demonstrate that INR values generated on the CP000 with Coagpia PT-N reagent show strong concordance with those obtained with five other coagulation analyzers. INR equivalency was successfully demonstrated in comparison with Siemens BCS and STAGO STA-R Evolution. Throughput was in line with the manufacturer s claim and time to first results was rapid. Page 6 of 7
7 REFERENCES Sara Just, Larry Smith, Gabriella Lakos. CP000 vs STA-R Evolution. Study ID: C-6-06; Haematology Department, Sydney South West Pathology Service, New South Wales Health Pathology, Liverpool Hospital, Sydney, Australia. Stephen So Shing Chau, Larry Smith, Gabriella Lakos. CP000 vs. ACL TOP-00. Study ID: C- 6-0; Clinical & Histopathology Laboratories, St. Teresa s Hospital, Kowloon, Hong Kong. Ivana Malikova, Larry Smith, Gabriella Lakos. CP000 vs BCS and CA Study ID: C-6-0; General University Hospital (Všeobecná Fakultní Nemocnice v Praze), Prague, Czech Republic. & 5 Volker Drescher, Peter Reichardt, Larry Smith, Gabriella Lakos. CP000 vs t. Study ID: C-6-0; Medizinschces Labor Westsachsen Fachlabor, Zwickau, Germany Presented at XXXth International Symposium on Technical Innovations in Laboratory Hematology, Honolulu, HI, May -6, 07, Abstract Number. ISLH 07, Honolulu, HA. CP000 is manufactured by Sekisui Medical Co., Ltd. Distributed by Abbott Laboratories. CP000 is a mark of Sekisui Medical Co., Ltd. Coagpia is a registered trademark of Sekisui Medical Co., Ltd. All other trademarks are the property of their respective owners ADD Abbott Laboratories. Page 7 of 7
HEMOCHRON. Whole Blood Coagulation Systems
HEMOCHRON Whole Blood Coagulation Systems Activated Partial Thromboplastin Time (APTT) Cuvette Correlation Protocol for HEMOCHRON Microcoagulation Instruments MSIG: 49 10/06 Dear Medical Professional:
More informationWith Stago, discover an outstanding Routine range
In Haemostasis, There s routine... and then there s Routine With Stago, discover an outstanding Routine range An optimal Routine range for guaranteed satisfaction 1 Comprehensive range Stago s extensive
More informationManagement & Administration
Laboratory Evaluation of the STA Satellite Benchtop Coagulation Instrument Ha T Su, MT(ASCP) CM, 1 Katherine L Whelchel, MT(ASCP)SH, 2 Mary Jo Sonnek, MT(ASCP), 1 and Gwyn A. Cutsforth, PhD 2 * ABSTRACT
More informationEffect of Direct Thrombin Inhibitors, Bivalirudin, Lepirudin, and Argatroban, on Prothrombin Time and INR Values
Coagulation and Transfusion Medicine / DIRECT THROMBIN INHIBITOR EFFECT ON INR Effect of Direct Thrombin Inhibitors, Bivalirudin, Lepirudin, and Argatroban, on Prothrombin Time and INR Values Robert C.
More informationHarmonization of red blood cell distribution width (RDW): an attainable target?
Original Article Page 1 of 5 Harmonization of red blood cell distribution width (RDW): an attainable target? Giuseppe Lippi 1, Silvia Pipitone 2, Emmanuel J. Favaloro 3,4 1 Section of Clinical Biochemistry,
More informationcobas 8100 automated workflow series 3-D intelligence in lab automation
3-D intelligence in lab automation Personalized Lab Automation Maximizing Testing Efficiency and Medical Value Hospital network Commercial laboratories Independent hospitals At Roche, laboratory automated
More informationcobas 8100 automated workflow series 3-D intelligence in lab automation
3-D intelligence in lab automation Personalized Lab Automation Maximizing Testing Efficiency and Medical Value Hospital network At Roche, laboratory automated solutions deliver the quality and reliability
More informationNote: The taller 100 x 13 mm tube is not designed to run on the ACL TOP. Plasma after centrifugation of these tubes must be placed in sample cups.
Created Revised Reviewed Approved Lupus Insensitive aptt on the ACL TOP HEM 4.15.1 7/13/2012 4/11/2013 4/11/2013 4/11/2013 I. PRINCIPLE The activated partial thromboplastin time is a global screening procedure
More information9/7/17. Implementing the ARK Immunoassay for Therapeutic Drug Monitoring of High-Dose Methotrexate. Disclosure. Learning Objectives
Implementing the ARK Immunoassay for Therapeutic Drug Monitoring of High-Dose Methotrexate Kristine R. Crews, PharmD, FCCP, BCPS St. Jude Children s Research Hospital September 14, 2017 Disclosure I declare
More informationcobas p 512 pre-analytical system Adapting to today s needs. Flexible for tomorrow s demand.
cobas p 512 pre-analytical system Adapting to today s needs. Flexible for tomorrow s demand. Personalized Lab Automation Maximizing testing efficiency and medical value Hospital network At Roche, laboratory
More informationcobas p 512 pre-analytical system Adapting to today s needs. Flexible for tomorrow s demand.
cobas p 512 pre-analytical system Adapting to today s needs. Flexible for tomorrow s demand. Personalized Lab Automation Maximizing Testing Efficiency and Medical Value At Roche, laboratory automated solutions
More informationActivated Protein C Resistance vs Factor V Leiden assay: Which is the most cost effective?
Activated Protein C Resistance vs Factor V Leiden assay: Which is the most cost effective? Rajiv K. Pruthi, M.B.B.S Co-Director, Special Coagulation Laboratory & Director, Mayo Comprehensive Hemophilia
More informationMarcia L. Zucker, Ph.D. ZIVD LLC
Marcia L. Zucker, Ph.D. ZIVD LLC 1 Monitoring hemostasis Bleeding Clotting 2 Picture courtesy of Helena Laboratories 3 Extrinsic Pathway Monitor with ACT / aptt WARFARIN Monitor with PT Common Pathway
More informationcobas p 612 pre-analytical system Adapting to today s needs. Flexible for tomorrow s demand.
cobas p 612 pre-analytical system Adapting to today s needs. Flexible for tomorrow s demand. Personalized Lab Automation Maximizing Testing Efficiency and Medical Value At Roche, laboratory automated solutions
More informationSAMPLE. Determination of Coagulation Factor Activities Using the One-Stage Clotting Assay
2nd Edition H48 Determination of Coagulation Factor Activities Using the One-Stage Clotting Assay This guideline provides recommendations regarding the proper collection and handling of specimens, reagents,
More informationPlease note that the uses described in the following page(s) have not been approved or cleared by FDA, with respect to the described assay or test.
Please note that the uses described in the following page(s) have not been approved or cleared by FDA, with respect to the described assay or test. In the US, the product is intended For Research Use Only.
More informationUpon completion of the Clinical Hematology rotation, the MLS student will be able to:
Clinical Performance Objectives in Clinical Hematology Department of Medical and Research Technology University of Maryland School of Medicine Spring 2015 Upon completion of the Clinical Hematology rotation,
More informationStago Newsletter. Diagnostica Stago Pty Ltd 651 Doncaster Road Doncaster, Vic Phone:
Diagnostica Stago Pty Ltd 651 Doncaster Road Doncaster, Vic 3108 Phone: +61 3 9840 5555 E-mail: info@stago.com Stago Newsletter V O L U M E 1, I S S U E 2 A U G U S T 2 0 1 0 I N S I D E T H I S I S S
More informationCoagulation Laboratory: Methods, Standards & Cost Effective Testing Part 1
Coagulation Laboratory: Methods, Standards & Cost Effective Testing Part 1 Donna D. Castellone, MS, MT (ASCP) SH Clinical Project Manager Hematology and Hemostasis Siemens Healthcare Diagnostics Disclosures
More informationKlinisch Laboratorium Declerck Ardooie, Belgium
Klinisch Laboratorium Declerck Ardooie, Belgium Dominique Declerck, Lab Director Sofie Arens, Clinical Biologist The Klinisch Laboratorium Declerck is known for its ability to provide clinicians with consistently
More informationUrine Protein Electrophoresis and Immunoelectrophoresis Using Unconcentrated or Minimally Concentrated Urine Samples
Immunopathology / Electrophoresis of Unconcentrated Urine Samples Urine Protein Electrophoresis and Immunoelectrophoresis Using Unconcentrated or Minimally Concentrated Urine Samples Anja C. Roden, MD,
More informationOn-Site Specialty Analyzer
Specialty Analyzer On-Site Specialty Analyzer The VetScan VSpro is a state-of-the-art specialty analyzer that offers a growing test menu useful to veterinarians of any specialty. The PT/aPTT and fibrinogen
More informationHematology and Coagulation Controls
Bio-Rad Laboratories QUALITY CONTROL SOLUTIONS FOR HEMATOLOGY AND COAGULATION Hematology and Coagulation Controls Your Quality Goals Realized Comprehensive QC solutions for building an effective quality
More informationEffect of Higher Centrifugation Speed and Shortened Centrifugation Time on Prothrombin and Activated Thromboplastin Time
TECHNICAL REPORT Effect of Higher Centrifugation Speed and Shortened Centrifugation Time on Prothrombin and Activated Thromboplastin Time Azlin I, Hafiza A, Azma RZ, Aidifitrina RK, Hamidah NH Haematology
More informationInfluence of Residual Platelet Count on Routine Coagulation, Factor VIII, and Factor IX Testing in Postfreeze-Thaw Samples
834 Influence of Residual Platelet Count on Routine Coagulation, Factor VIII, and Factor IX Testing in Postfreeze-Thaw Samples Giuseppe Lippi, MD 1 Rossana Rossi, PhD 1 Luigi Ippolito, MD 1 Valentina Zobbi,
More informationSAMPLE H21-A5. January 2008
January 2008 Collection, Transport, and Processing of Blood Specimens for Testing Plasma-Based Coagulation Assays and Molecular Hemostasis Assays; Approved Guideline Fifth Edition This document provides
More informationComparison of Several Activated Partial Thromboplastin Time Methods
Comparison of Several Activated Partial Thromboplastin Time Methods ROBERT J. MOWN, M.D., AND DOROTHY WILLOUGHBY, B.S., M.T. (ASCP) Department of Pathology, Harbor General Hospital, UCLA School of Medicine,
More informationInternal Quality Control Practices in Coagulation Laboratories: recommendations based on a patterns-of-practice survey
International Journal of Laboratory Hematology ORIGINAL ARTICLE The Official journal of the International Society for Laboratory Hematology INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY Internal Quality
More informationSPRIworks Systems. Push button. Walk away. Fully automated library construction systems with built-in size selection and cleanup BR-15981B
SPRIworks Systems Fully automated library construction systems with built-in size selection and cleanup Push button. Walk away. BR-15981B SPRIworks Technology Fully automated fragment library construction
More informationResidual Serum Thrombin
Residual Serum Thrombin Activity Murray Weiner N RECENT YEARS a remarkable number of blood and tissue components have been found to play a role in the formation of thrombin, the enzyme ultimately responsible
More informationGlobal In-Vitro Diagnostic Market Report
Global In-Vitro Diagnostic Market Report ----------------------------------------------------- 2013 Executive Summary The process of performing a diagnostic test outside of a living organism in an artificial
More informationThe China In Vitro Diagnostics (IVD) Market has been analyzed from 4 View Points:
China In Vitro Diagnostics (IVD) Market - Overview China In-Vitro Diagnostics (IVD) market is among the fastest growing in the world. It is anticipated that China IVD market will more than triple by 2020
More informationClinical Chemistry Approach to Evaluation of Commutability
Clinical Chemistry Approach to Evaluation of Commutability Hubert W. Vesper, Ph.D. Director, Clinical Standardization Programs Division of Laboratory Sciences Centers for Disease Control and Prevention,
More informationMonitoring and improvement of intralaboratory turnaround time at the university hospital Campus Bio-Medico in Rome
SCIENTIFIC PAPERS CONTRIBUTI SCIENTIFICI Monitoring and improvement of intralaboratory turnaround time at the university hospital Campus Bio-Medico in Rome Silvia Angeletti 1, Marina De Cesaris 1, Siriana
More informationINNOVANCE DTI Assay Delivering the Reliability Your Lab Needs to Confidently Test for Dabigatran
Eva-Maria Landmann Marketing Manager Hemostasis INNOVANCE DTI Assay Delivering the Reliability Your Lab Needs to Confidently Test for Dabigatran INNOVANCE DTI Assay* Introduction Clinical Information Dabigatran
More informationExpert workflow management cobas p 512 pre-analytical system Professional management of laboratory processes
Expert workflow management cobas p 512 pre-analytical system Professional management of laboratory processes COBAS, COBAS INTEGRA, COBAS E, COBAS P, MODULAR PRE-ANALYTICS EVO, ELECSYS and LIFE NEEDS ANSWERS
More informationUnited States Lateral Flow Assay Market Report 2017
Published on Market Research Reports Inc. (https://www.marketresearchreports.com) Home > United States Lateral Flow Assay Market Report 2017 United States Lateral Flow Assay Market Report 2017 Publication
More informationHematotoxin producing snakes in Thailand พญ.ม ชฌ มา รอดเช อ พญ.อ สร ย หาญอ ท ยร ศม
Hematotoxin producing snakes in Thailand พญ.ม ชฌ มา รอดเช อ พญ.อ สร ย หาญอ ท ยร ศม Hematotoxin producing snakes Pit vipers (Crotalinae) Malayan pit viper (Calloselasma rhodostoma) Green pit viper (Trimeresurus
More informationOriginal Article. A Novel Automated Slide-Based Technology for Visualization, Counting, and Characterization of the Formed Elements of Blood
Original Article A Novel Automated Slide-Based Technology for Visualization, Counting, and Characterization of the Formed Elements of Blood A Proof of Concept Study James W. Winkelman, MD; Milenko J. Tanasijevic,
More informationLonger acting/extended Half Life Products: Laboratory Monitoring
Longer acting/extended Half Life Products: Laboratory Monitoring Steve Kitchen Lead Clinical scientist Sheffield Haemophilia Centre & UK NEQAS Blood Coagulation Declaration of Interests -1 Speaker/ consultancy
More informationSEED Coagulation. Sysmex Educational Enhancement and Development April 2014
SEED Coagulation Sysmex Educational Enhancement and Development April 2014 The Thrombin Time Test and Reptilase Test what is their role in coagulation testing? Baseline screening tests of coagulation The
More informationCompleting the performance picture.
Completing the performance picture. AU5800 Clinical Chemistry Systems Blood Banking Centrifugation Chemistry Flow Cytometry Hematology Hemostasis Immunoassay Information Systems Lab Automation Molecular
More informationBio-Rad Laboratories BIOPLEX 2200 SYSTEM. BioPlex 2200 ANA Screen with MDSS. The first and only fully-automated, multiplexed ANA Screen
Bio-Rad Laboratories BIOPLEX 2200 SYSTEM BioPlex 2200 ANA Screen with MDSS The first and only fully-automated, multiplexed ANA Screen Bio-Rad Laboratories BIOPLEX 2200 SYSTEM Like no other The BioPlex
More informationcobas p 312 pre-analytical system Compact automation for maximum efficiency
cobas p 312 pre-analytical system Compact automation for maximum efficiency Personalized Lab Automation Maximizing testing efficiency and medical value At Roche, laboratory automated solutions deliver
More informationSirtex Medical Limited
Sirtex Medical Limited Market Update 2011 Half Year Results Gilman Wong CEO Darren Smith CFO 9-10 March 2011 SIR-Spheres is a registered trademark of Sirtex SIR-Spheres Pty Ltd. Understanding the Sirtex
More informationApplication Note. Authors. Abstract
Automated, High Precision Tryptic Digestion and SISCAPA-MS Quantification of Human Plasma Proteins Using the Agilent Bravo Automated Liquid Handling Platform Application Note Authors Morteza Razavi, N.
More informationDry Hematology Analyzer Fast, Simple CBC Analysis Designed for the Point of Care
QBC STAR TM Dry Hematology Analyzer Fast, Simple CBC Analysis Designed for the Point of Care Reach for the STAR The QBC STAR hematology analyzer expands and advances the unique capabilities of previous
More informationHistology Pattern Recognition Software in Investigative Pathology
Histology Pattern Recognition Software in Investigative Pathology J. Webster, DVM, PhD, DACVP Laboratory of Cancer Biology and Genetics National Cancer Institute, Bethesda, MD Pathology Visions 2011 November
More informationEvaluation of the Automated Immature Granulocyte Count (IG) on Sysmex XE-2100 Automated Haematology Analyser vs. Visual Microscopy (NCCLS H20-A)
Evaluation of the Automated Immature Granulocyte Count (IG) on Sysmex XE-21 Automated Haematology Analyser vs. Visual Microscopy (NCCLS H2-A) Th WEILAND *1, H KALKMAN *2, and H HEIHN *1 *1 Central Laboratory,
More informationThe Simple Solution for CBC Testing
QBC STAR Centrifugal Hematology Analyzer The Simple Solution for CBC Testing Diagnostics Innovative Solutions for a Healthier World UNRIVALED SIMPLICITY The QBC STAR Centrifugal Hematology System is the
More informationInstrument specifications Technology Software Host interfaces Printer interfaces Data Station Physical dimensions Weight Certifications
The evolution of PCR Introducing the COBAS TaqMan 48 Analyzer At the forefront of PCR evolution Since Roche acquired the rights to PCR in 1991, its mission has been to fully realize the potential of this
More informationProcedure. Prothrombin Time with INR (PT/INR) Hemochron Signature+ Hemochron Signature Elite. Page 1 of 19. UAMS pt-inr-clsi-procedure
Procedure Prothrombin Time with INR (PT/INR) Hemochron Signature+ Hemochron Signature Elite Page 1 of 19 Purpose The Hemochron Jr. Prothrombin Time (PT) Test is a quantitative assay intended for in vitro
More informationComparison of Six Automated Treponema-Specific Antibody Assays
JCM Accepted Manuscript Posted Online 11 November 2015 J. Clin. Microbiol. doi:10.1128/jcm.02593-15 Copyright 2015, American Society for Microbiology. All Rights Reserved. 1 Comparison of Six Automated
More informationPhysician Orders - Adult
Physician Orders - Adult attach patient label here Title: Direct Thrombin Inhibitor (DTI) Protocol Orders Height: cm Weight: kg Allergies: [ ] No known allergies [ ]Medication allergy(s): [ ] Latex allergy
More informationMarcia L. Zucker, Ph.D. ZIVD LLC
Marcia L. Zucker, Ph.D. ZIVD LLC 1 Explain why ACTs from different systems are not the same Develop a plan for switching from one ACT system to another Describe why ACT and aptt are not interchangeable
More informationProduct Catalog # Description List Price (JPY) Primer Assays (desalted)
Assays and Controls Primer Assay Pricing Primer Assays (desalted) 10025636 Primer assay desalted, 200 reactions 16,000 (Wet-lab validated human, mouse, and rat) 10025637 Primer assay desalted, 1,000 reactions
More informationValidating, Verifying, and Evaluating Your Test Methods: It s NOT a Regulatory Exercise!
Validating, Verifying, and Evaluating Your Test Methods: It s NOT a Regulatory Exercise! Pat Garrett, Ph.D., DABCC Renee Howell, Ph.D., MT(ASCP) SeraCare Life Sciences, Inc. AACC Annual Meeting July 29,
More informationBasic coagulation applications and case studies
Basic coagulation applications and case studies Jing Jin Clinical laboratory Scientist (MLS, ASCP) - Coagulation/Hematology Stanford University Hospital and Clinics 1 Agenda Overview about 3 major phases
More informationBioPlex 2200 System. Like no other. The first and only fully-automated, random access, multiplex testing platform. Bio-Rad Laboratories
Bio-Rad Laboratories BIOPLEX 2200 SYSTEM BioPlex 2200 System Like no other. The first and only fully-automated, random access, multiplex testing platform Bio-Rad Laboratories BIOPLEX 2200 SYSTEM Fully
More informationExpert workflow management cobas p 612 pre-analytical system. Professional management of laboratory processes
Expert workflow management cobas p 612 pre-analytical system Professional management of laboratory processes Expert workflow management Concept Professional management of processes in diagnostic laboratories
More informationK. Doubleday S. Kumnick. Clot Signature Curves and the ACLAdvance TM
COAGULATION 12 K. Doubleday S. Kumnick Clot Signature Curves and the ACLAdvance TM K. Doubleday S. Kumnick Clot Signature Curves and the ACL Advance TM Index 3 What is a Clot Signature Curve Pag. 5 How
More informationTable of Contents. Proficiency Testing (PT) Manual. 4 How to Interpret Your Evaluation General Customer Information...
2017Pr o ci encytest i ngmanual Per f or manceyoucanmeasur e Accur acyyoucant r ust Table of Contents Proficiency Testing (PT) Manual 1 General Customer Information... 3 10 Order Confirmation...3 Regulatory
More informationDade Behring. Analyst Meeting July 18, Page 1
Dade Behring Analyst Meeting July 18, 2007 Page 1 Agenda Welcome / Introductions Overview Products and Solutions Update U.S. / Chemistry Spain / Microbiology Germany / Hemostasis Summary Jim Reid-Anderson
More informationR.Li, C.Swaelens, F.Vandermijnsbrugge, B.Cantinieaux BSTH Laboratory of haematology, Porte de Hal,
Institut J. Bordet Normal with Actin-FS avoids intrinsic pathway factors assays in the presence of an isolated prolongation of (Platelin-LS) without hemorrhagic history R.Li, C.Swaelens, F.Vandermijnsbrugge,
More informationBio-Rad Laboratories BIOPLEX 2200 SYSTEM. BioPlex 2200 System. Workflow: Redefined
BioPlex 2200 System Workflow: Redefined Workflow: Redefined Leveraging the power of multiplex testing, the BioPlex 2200 system brings flexible workflow solutions to your laboratory. Disease Focused Multiplex
More informationComparison of 13 Commercially Available Cardiac Troponin Assays in a Multicenter North American Study
ARTICLES SC Comparison of 13 Commercially Available Cardiac Troponin Assays in a Multicenter North American Study Robert H. Christenson, 1 * Ellis Jacobs, 2,3 Denise Uettwiller-Geiger, 4 Mathew P. Estey,
More informationLaboratory Assessment of Novel Oral Anticoagulants: Method Suitability and Variability between Coagulation Laboratories
Clinical Chemistry 59:5 807 814 (2013) Hemostasis and Thrombosis Laboratory Assessment of Novel Oral Anticoagulants: Method Suitability and Variability between Coagulation Laboratories Tuukka A. Helin,
More informationClotting tests. June 2009 V Diagnostica Stago - All right reserved
7 Clotting tests June 2009 V2 Reagents Technologies Clotting tests Chromogenic assays Enzyme Immuno Assays (ELISA) Latex Immuno Assays (LIA) Latex slide agglutination Flow cytometry 2 Clotting tests Samples
More information2015 Catalogue Coagulation
2015 Catalogue Coagulation 1 CONTENTS Reagents TriniCLOT TM Routine Reagents PT 4 aptt 4 Fibrinogen 5 Thrombin Time 5 Factor Deficient Plasmas 5 Solutions 5 TriniLIA TM D-Dimer TriniLIA TM D-Dimer 6 TriniCLOT
More informationBench top / Desktop. English HumaClot Pro GUI. Open at launch, but system will be closed mandatorily in the future.
Specifications & more HumaClot Pro > > System Specifications > > Scope of Supply > > Obligatory Items > > Optional Items System Overview Cat.No. 15800 Analyzer type Fully automated coagulation analyzer
More informationActivated Partial Thromboplastin Time and Anti-Xa Measurements in Heparin Monitoring. Biochemical Basis for Discordance
Coagulation and Transfusion Medicine / vs Level Discrepancies ctivated Partial Thromboplastin Time and Measurements in Heparin Monitoring iochemical asis for Discordance Clifford M. Takemoto, MD, 1 Michael.
More informationCoagulation in perspective: Blood management. Objectives
Coagulation in perspective: Blood management Julie Wegner, PhD jawrbl@gmail.com Objectives To gain a basic understanding of the following: 1. Coagulation components and processes Why patients bleed. 2.
More informationvalidation report for TaqMan Roundup Ready Quantification Kit
validation report for TaqMan Roundup Ready Quantification Kit Life Technologies offers a breadth of products DNA RNA protein cell culture instruments For Research Use Only. Not intended for any animal
More information***IMPORTANT INSTRUCTIONS PLEASE READ***
January 29, 2013 ***IMPORTAT ISTRUCTIOS PLEASE READ*** TO: FROM: SUBJECT: Laboratory Director Erasmus Schneider, Ph.D. Director, Diagnostic Oncology Section, Clinical Laboratory Evaluation Program OCOLOGY
More informationFirst Annual Biomarker Symposium Quest Diagnostics Clinical Trials
First Annual Biomarker Symposium Quest Diagnostics Clinical Trials Terry Robins, Ph.D. Director Biomarker R&D and Scientific Affairs Quest Diagnostics Clinical Trials Key Considerations: Biomarker Development
More informationNavigating Change How a Major Academic Medical Center Laboratory Switched Hematology Platforms, and No One Noticed
Navigating Change How a Major Academic Medical Center Laboratory Switched Hematology Platforms, and No One Noticed Sysmex Educational Symposium Santa Barbara, CA November 12, 2015 Melanie Browning Core
More informationTechnologies for Coagulation Instruments
C E U P D A T E I N S T R U M E N T A T I O N I V John A. Koepke, MD Technologies for Coagulation Instruments Both coagulation and fibrinolysis have traditionally been measured by functional assays. For
More informationThe Corporate Representative's Presentation
The Corporate Representative's Presentation IFCC General Conference, Madrid, 2016-03-19 Rolf Hinzmann, MD, PhD Corporate Representative at the IFCC Executive Board Contact: rolf.hinzmann@roche.com Agenda
More informationTrue Walkaway Automated Urinalysis, Chemistry to Microscopy The CLINITEK AUWi System Answers for life.
www.siemens.com/diagnostics True Walkaway Automated Urinalysis, Chemistry to Microscopy The CLINITEK AUWi System Answers for life. Can I achieve true unattended urinalysis testing in my laboratory? With
More informationIntegrated Radiology & Pathology Diagnostic Services: The Time is Now at UCLA
Integrated Radiology & Pathology Diagnostic Services: The Time is Now at UCLA Scott W Binder MD Pritzker Professor of Clinical Pathology and Dermatology Senior Vice Chair Director, Pathology Clinical Services
More informationVerifying New Reagent Lot Performance
Verifying New Reagent Lot Performance Julianne Addison Application Consultant, Siemens RSC Answers for life. Objectives Describe issues relating to changing reagent lots Discuss the reasons QC material
More informationNewsletter from All Eights
Guide to Newsletter from All Eights Focus Thrombophilia in Asian Countries: Myth or Reality? First Local Paper APCR test in the investigation of thrombophilic state Application System validation & correlation
More informationFlexible, robust solutions from BSI. An In Vitro Diagnostic Notified Body. Expertise and experience. IVD regulatory solutions
Flexible, robust solutions from BSI An In Vitro Diagnostic Notified Body Expertise and experience IVD regulatory solutions Updated May 2017 Your guide to the In Vitro Diagnostic Directive The purpose of
More informationCase No COMP/M SIEMENS / BAYER DIAGNOSTICS. REGULATION (EC) No 139/2004 MERGER PROCEDURE. Article 6(1)(b) NON-OPPOSITION Date: 31/10/2006
EN Case No COMP/M.4321 - SIEMENS / BAYER DIAGNOSTICS Only the English text is available and authentic. REGULATION (EC) No 139/2004 MERGER PROCEDURE Article 6(1)(b) NON-OPPOSITION Date: 31/10/2006 In electronic
More informationFibrinogen Universal Kit
Order references Reagents CONT FITUR-B00 Universal kit 1 x 50 ml R1 + 1 x 5 ml R2 FITUR-00 Universal kit 2 x 50 ml R1 + 1 x 15 ml R2 FITUR-C00 Universal kit 2 x 17 ml R1 + 1 x 4 ml R2 Other necessary products
More informationChanges in coagulation and tissue plasminogen activator after the treatment of cerebral infarction with lumbrokinase
Clinical Hemorheology and Microcirculation 23 (2000) 213 218 213 IOS Press Changes in coagulation and tissue plasminogen activator after the treatment of cerebral infarction with lumbrokinase Lirong Jin
More informationQC and Changing Lots
QC and Changing Lots Back to the Basics Karen Hoffman MT(ASCP) Senior Technical Integration Specialist Laboratory Solutions Services Sysmex America, Inc. Objectives On completion of this training you should
More informationFibrinogen Kaiserslautern III: A New Case of Congenital Dysfibrinogenemia with A 16 Arg j Cys Substitution
Original Paper Haemostasis 2001;31:12 17 Received: November 24, 1999 Accepted in revised form: December 8, 2000 Fibrinogen Kaiserslautern III: A New Case of Congenital Dysfibrinogenemia with A 16 Arg j
More informationPerformance of the new automated Abbott RealTime MTB assay for rapid detection of Mycobacterium tuberculosis complex in respiratory specimens
DOI 10.1007/s10096-015-2419-5 ARTICLE Performance of the new automated Abbott RealTime MTB assay for rapid detection of Mycobacterium tuberculosis complex in respiratory specimens J. H. K. Chen 1 & K.
More informationPersonal history of excessive mucocutaneous bleeding. Family history of excessive bleeding. Laboratory tests of hemostasis consistent with VWD
Diagnostic Approach Personal history of excessive mucocutaneous bleeding Family history of excessive bleeding Laboratory tests of hemostasis consistent with VWD Nose Bleeds Skin Bruising Gum Bleeding Menorrhagia
More informationCONCURRENT 10: THE COMPLETE BLOOD COUNT AND BEYOND: QUALITY ISSUES AND REFERENCE METHODS
CONCURRENT 10: THE COMPLETE BLOOD COUNT AND BEYOND: QUALITY ISSUES AND REFERENCE METHODS Saturday May 14 th 2016 1:30 3:00 PM (13.30 15.00hrs) International Society for Laboratory Hematology (ISLH) CONCURRENT
More informationIntersection of LEAN and the LIS. Megan Schmidt. Product Manager, Sunquest Information Systems, Inc. Patrick O Sullivan
Intersection of LEAN and the LIS Megan Schmidt Product Manager, Sunquest Information Systems, Inc. Patrick O Sullivan Laboratory Director, Florida Hospital Orlando Agenda Manufacturing Mindset LEAN Principles
More informationInnovations with "Best of Breed" Solutions:
Innovations with "Best of Breed" Solutions: Valuable Lessons Learned by PinnacleHealth Laboratories Laboratory Executive War College May 14, 2008 Presented by: Judy Judith R. Darr, MS MT (ASCP) SH DLM
More informationSerum vs Plasma. Which specimen should you use? American Society for Clinical Laboratory Science Michigan Annual Meeting
Serum vs Plasma Which specimen should you use? American Society for Clinical Laboratory Science Michigan Annual Meeting Sol Green PhD FACB March 31 2016 Introduction This presentation is intended to provide
More informationSpecial Coagulation - APC Resistance. DiaPharma Group, Inc. Customer Service: Technical Support:
Special Coagulation - APC Resistance DiaPharma Group, Inc. Customer Service: 1.800.526.5224 Technical Support:1.800.447.3846 www.diapharma.com 1 Review of Hemostasis Overview Pathways of coagulation, anticoagulation,
More informationcobas p 501/701 post-analytical units Automated storage and retrieval for maximum efficiency and safety
cobas p 501/701 post-analytical units Automated storage and retrieval for maximum efficiency and safety Personalized Lab Automation Maximizing Testing Efficiency and Medical Value At Roche, laboratory
More information