Surgical Technique. and Product Catalog FOR NON-USA USE ONLY

Size: px
Start display at page:

Download "Surgical Technique. and Product Catalog FOR NON-USA USE ONLY"

Transcription

1 Surgical Technique and Product Catalog FOR NON-USA USE ONLY

2 INTRODUCTION Percutaneous vertebroplasty and, more recently, kyphoplasty, are Vertebral Body Augmentation (VBA) procedures used for the treatment of painful vertebral compression fractures caused by osteoporosis, tumors or trauma. Unlike other vertebroplasty or kyphoplasty systems that inject cement directly into the vertebral body, with the CONFIDENCE PERIMETER SPINAL CEMENT SYSTEM, the cement is injected into a mesh bag. This mesh bag provides additional control over placement of the cement. As the cement expands the mesh bag, there may also be the possibility of re-gaining some of the lost vertebral body height1, if the fracture is still mobile. The porous construction of the mesh allows the cement to extrude through the barrier and interdigitate with the surrounding cancellous bone. Additionally, it allows the bag to be removed once cement introduction is complete. The CONFIDENCE PERIMETER SPINAL CEMENT SYSTEM is only used in conjunction with the CONFIDENCE SPINAL CEMENT SYSTEM s high viscosity cement and hydraulic delivery system. For a complete list of warnings, precautions and contra-indications, please refer to the CONFIDENCE PERIMETER SPINAL CEMENT SYSTEM Packaging Insert (IFU).

3 CONTENTS SURGICAL TECHNIQUE Preparation 4 Bone Access Needle Placement 5 Biopsy Needle Placement (if required) 6 Bone Drill - Space Creation 7 Insertion of the Mesh Assembly 7 Cement Preparation 8 Cement Delivery 10 Retracting the Mesh Bag Assembly 13 Working Time and Temperature 13 CONFIDENCE PERIMETER SPINAL CEMENT SYSTEM Mesh Cutter Surgical Technique 14 PRODUCT CATALOG CONFIDENCE PERIMETER SPINAL CEMENT SYSTEM 18 CONFIDENCE SPINAL CEMENT SYSTEM 19 CONFIDENCE Needleless Kit 20 INDICATIONS FOR USE The CONFIDENCE PERIMETER SPINAL CEMENT SYSTEM is indicated for use in combination with the CONFIDENCE SPINAL CEMENT SYSTEM. The CONFIDENCE SPINAL CEMENT SYSTEM is intended for percutaneous delivery of CONFIDENCE High Viscosity Spinal Cement, which is indicated for fixation of pathological fractures of the vertebral body during vertebroplasty or kyphoplasty procedures. Painful vertebral compression fractures may result from osteoporosis, benign lesions (haemangioma), and malignant lesions (metastatic cancer, myeloma). The CONFIDENCE PERIMETER cement injection cannula has a specifically designed luer lock fitting. It is designed for use only with CONFIDENCE SPINAL CEMENT SYSTEM and the CONFIDENCE High Viscosity Spinal Cement. The device is not compatible with alternate materials. A single CONFIDENCE SPINAL CEMENT SYSTEM is restricted to be used with only one CONFIDENCE PERIMETER SPINAL CEMENT SYSTEM. When using a bilateral approach or doing multiple levels with the CONFIDENCE PERIMETER SPINAL CEMENT SYSTEM, a new CONFIDENCE SPINAL CEMENT SYSTEM should be opened and used for each level/approach.

4 Vertebral Body Augmentation SURGICAL TECHNIQUE PATIENT POSITIONING AND PREPARATION The patient should be in a prone position on a radiolucent table, permitting AP and lateral imaging using uni or bi-planar x-ray and/or CT. Either local or general anaesthesia may be used. Prepare the operative site according to standard surgical techniques and hospital procedures. INSTRUMENT PREPARATION The bone access needle is packaged with a beveled tip stylet. If a diamond tip is preferred, remove the beveled stylet and replace with the diamond stylet. Ensure that the diamond stylet is adequately locked to the handle of the access needle. The mesh needle assembly should not be disassembled. Disassembly will lead to disengagement of the stylet from the mesh bag which may result in piercing of the mesh bag when reinserted. 4

5 Surgical Technique BONE ACCESS NEEDLE PLACEMENT Under fluoroscopy, insert the bone access needle (working cannula and stylet) down to the appropriate entry location to the vertebral body. The trajectory by which the needle is inserted varies depending on the vertebra to be treated. Most commonly a transpedicular or extrapedicular approach is performed. A uni-pedicular or bi-pedicular approach may be used at the operator s discretion. Using alternating A/P and lateral fluoroscopy, the needle can be guided into the vertebral body. The tip of the needle should commonly begin superior and lateral to the border of the pedicle on an A/P image. The needle is advanced, under lateral fluoroscopy, into the vertebral body. During needle advancement, great care should be taken to avoid breaching the medial border of the pedicle. This is accomplished by checking that when the tip of the needle on lateral fluoroscopy is flush with the posterior border of the vertebral body, that it is not medial to the medial border of the pedicle on the A/P image. Care should be taken to position the distal tip of the working cannula to allow space for the CONFIDENCE PERIMETER Mesh Bag (23mm beyond the distal tip). 5

6 Vertebral Body Augmentation The needle holder, in combination with the long-armed mallet, may be used to position and advance the needle under fluoroscopy, while keeping the physician s hands outside the x-ray field. In case of beveled tip needles, the arrows on the handle of the needle point to the direction of the bevel. After passing the vertebral posterior wall using the CONFIDENCE PERIMETER bone access needle, hold the working cannula firmly in place and remove the stylet. BIOPSY NEEDLE PLACEMENT (IF REQUIRED) Under fluoroscopy and/or CT guidance, insert the CONFIDENCE biopsy needle through the bone access needle for biopsy collection. According to the user s preference in biopsy technique, the stylet may be removed either before insertion of the biopsy needle, or after the vertebral posterior wall has been passed. 6

7 Surgical Technique BONE DRILL - SPACE CREATION If the predicted location of the mesh bag (indicated by the location of the drill tip) is unsatisfactory, re-position the working cannula and drill prior to insertion of the mesh assembly. Under fluoroscopy and/or CT guidance, fully insert the bone drill through the working cannula, using a clockwise, rotating motion. This will create a working space for the mesh bag. When the bone drill is fully inserted through the working cannula, the distal tip of the drill represents the location of the distal tip of the mesh bag. Attention should be paid not to breach the anterior wall of the vertebral body while advancing the bone drill. Once completed, remove the drill. INSERTION OF THE MESH ASSEMBLY To prevent advancement of the working cannula, firmly hold the cannula in place while inserting the mesh assembly. Insert the mesh assembly into the working cannula using a uni-directional motion. Withdrawal and re-insertion may cause the stylet s tip to disengage from the tip of the mesh bag and potentially pierce the bag. Under fluoroscopy and/or CT guidance, insert the mesh assembly through the working cannula until it is firmly attached to the handle of the working cannula. While firmly holding the working cannula s handle, remove the mesh assembly stylet using a counter-clockwise rotation. The CONFIDENCE PERIMETER cement injection cannula has a specifically designed luer lock fitting. It is designed for use only with CONFIDENCE SPINAL CEMENT SYSTEM and the CONFIDENCE High Viscosity Spinal Cement. The device is not compatible with alternate materials. Mesh bag tip Once removed, do not reinsert stylet as it may result in piercing of the mesh bag. 7

8 Vertebral Body Augmentation CEMENT PREPARATION Pour the entire contents of the cement powder into the mixing bowl. Pour the entire contents of the liquid monomer vial over the cement powder in the mixing bowl. Prior to cement mixing, ensure that the mesh assembly is in its final position. Important: CONFIDENCE PERIMETER working time = cement mixing, transfer, injection and mesh bag retraction. Powder should always be poured into the mixing bowl first. Care should be taken not to open the liquid monomer vial over the mixing bowl due to risk of glass splinters. Important: Timings provided begin at start of mixing. WORKING TIME AND TEMPERATURE Theatre Temperature ( C) End of Working Time (Cement mixing, transfer, injection and mesh bag retraction) 18 7 minutes 14 minutes 22 6 minutes 12 minutes Start of Setting Time (Working cannula removal) 25 5 minutes 10 minutes Prior to introducing the mixer handle into the mixing bowl, verify that the wiper blade is attached to the inside of the mixer handle. If it is not, attach the wiper blade to the mixer handle by snapping it into place. Keep in mind that the teeth should point upwards. Important: Closely monitor the working time of the procedure (see chart above). While holding the mixing bowl vertically on the table, push down on the outer ring of the mixer handle and screw this ring onto the mixing bowl in a clockwise direction. wiper blade 8

9 Surgical Technique Mix the cement by turning the handle 3 turns in one direction, then 3 turns in the opposite direction. Repeat for a total of seconds. Unscrew the ring on the mixer handle in a counter-clockwise direction and remove the handle from the mixing bowl. The cement and wiper blade will remain in the mixing bowl. Push and screw the cement reservoir adaptor onto the mixing bowl with a clockwise rotation until firmly seated in the bottom of the bowl. This process will force the cement into the cement reservoir. Ensure that the cement is fully distributed to the distal tip of the reservoir. Unscrew the cement reservoir from the cement reservoir adaptor with a counter-clockwise rotation. 9

10 Vertebral Body Augmentation CEMENT DELIVERY Ensure the piston inside the cap is properly attached to the cap before threading onto the reservoir. The reservoir cap must be fully seated against the flange on the cement reservoir. Screw the cement reservoir cap securely onto the cement reservoir. Care should be taken to align the cap carefully with the threaded reservoir to ensure proper thread engagement. It is important that the cap is fully tightened before pressurizing or delivering cement. A small amount of cement may extrude from the tip and should be removed before assembly to the needle. Connect the hydraulic pump s flexible tube to the cement reservoir cap by pushing and snapping the quick-connect assembly onto the cement reservoir cap. cap flange If the quick connect will not attach to the reservoir cap, rotate the handle of the pump counter-clockwise to relieve some pressure. This will allow the connector to attach to the cap. 10

11 Surgical Technique Care must be taken to avoid overtightening the cement reservoir to the mesh assembly (white luer). Screw the reservoir tip onto the luer fitting of the mesh assembly cement injection cannula (white in color). Keep hold of the working cannula to prevent advancing it further into the vertebral body. Prepare the system for delivery by turning the pump handle a few turns, until a small amount of cement extrudes into the mesh bag. This should be done under fluoroscopic guidance. While monitoring the mesh bag under fluoroscopy, slowly rotate the pump handle clockwise half a turn per second until cement begins to fill the mesh bag. The hydraulic pump has a safety release valve that restricts the build-up of excess pressure in the reservoir. The scale on the cement reservoir is intended to approximate the amount of cement introduced. Fluoroscopy should be used to confirm the correct amount of cement introduced into the vertebral body. Note that the volume of cement in the injection cannula is approximately 0.3ml. Each half turn (180 degrees) of the handle will inject approximately 0.15 ml/cc of cement. As the cement is introduced, the CONFIDENCE PERIMETER mesh bag will expand within the vertebral body. It takes approximately 2 ml/cc of cement to fully expand the bag. 11

12 Vertebral Body Augmentation Cement can continue to be introduced and will extrude through the pores of the mesh bag and interdigitate with the surrounding bone structure. Upon completion, or at any point during the procedure, cement flow can be stopped by rapidly rotating the handle counter-clockwise until pump handle spins freely. When cement introduction is complete, rotate the pump handle counter-clockwise until pump handle spins freely to depressurize the cement reservoir. While firmly holding the injection cannula cap (white in color) in one hand, disconnect the cement reservoir from the mesh assembly. Firmly hold the working cannula s handle (blue in color) and remove the injection cannula (white in color) from the mesh assembly, using a counter-clockwise rotation. 12

13 Surgical Technique Guide RETRACTING THE MESH BAG ASSEMBLY Ensure removal of the mesh bag from the vertebral body prior to the end of the cement working time. (see chart below). Flip up the winding handle on the mesh assembly (grey in color). Rotate the winding handle 3-4 full 360 turns until the mesh bag is completely retracted into the working cannula. Confirm that this step is complete using fluoroscopy. Prior to the start of the cement setting time (see chart below), remove the working cannula/ mesh assembly from the vertebral body by rotating and pulling the cannula. This will allow easier removal of the cannula. The needle introduction site can be closed with a single suture, a steri-strip, or in any other manner preferred by the operator. The CONFIDENCE PERIMETER SPINAL CEMENT SYSTEM must be stored at temperatures between 5 C (41 F) to 25 C (77 F). WORKING TIME AND TEMPERATURE Theatre Temperature ( C) End of Working Time (Cement mixing, transfer, injection and mesh bag retraction) Start of Setting Time (Working cannula removal) 18 7 minutes 14 minutes 22 6 minutes 12 minutes 25 5 minutes 10 minutes Please note that mesh bag is intended to be retrieved and working times in Time/Temperature Table should be followed carefully to retrieve the mesh bag during the required retraction time period. If the physician encounters a situation where they are unable to retrieve the mesh bag for any reason, or if it is in the best interest of the patient to keep the mesh bag in place, the following steps should be taken: 1. Pull out the winding handle from the mesh assembly. 2. Remove the working cannula. 3. Open skin incision to accept PERIMETER Mesh Cutter, if necessary. 4. Clamp onto the tail of the mesh bag and run the cutting tool (e.g. surgical scissors) down the length of the mesh bag. 5. Cut the mesh bag as close to the pedicle as possible and close the incision. 13

14 Vertebral Body Augmentation PERIMETER MESH CUTTER SURGICAL TECHNIQUE Please note that mesh bag is intended to be retrieved and working times in Time/Temperature Table should be followed carefully to retrieve the mesh bag during the required retraction time period. If the physician encounters a situation where they are unable to retrieve the mesh bag for any reason, or if it is in the best interest of the patient to keep the mesh bag in place, the following steps should be taken: 1. Pull out the winding handle from the mesh assembly. 2. Remove the working cannula from the patient. 14

15 Surgical Technique Guide 3. Open skin incision as needed to accept the PERIMETER Mesh Cutter. 4. Insert end of mesh bag into the slot of PERIMETER Mesh Cutter tip. 15

16 Vertebral Body Augmentation 5. Hold on to end of mesh bag while advancing PERIMETER Mesh Cutter along mesh bag until the Mesh Cutter contacts vertebral body. Use flouro to verify that PERIMETER Mesh Cutter is at the pedicle. 6. Cut the mesh bag as close to the pedicle as possible. 16

17 Surgical Technique Guide If desired, distal tip of working cannula stylet can be used to pack any protruding mesh into pedicle. 7. Remove PERIMETER Mesh Cutter and end of mesh bag from patient. 8. Close the incision. 17

18 Vertebral Body Augmentation PRODUCT CATALOG CONFIDENCE PERIMETER KIT / CATALOG NO Components included in kit: Bone Access Needle/Working Cannula, Beveled Tip, 11G, 100mm Additional Stylet, Diamond tip, 11G, 100mm Bone Drill Mesh Assembly Component Descriptions Bone Access Needle/Working Cannula Consists of a working cannula and a beveled tip stylet for percutaneous bone access. The needle also acts as a cannula through which the bone drill and PERIMETER Mesh Assembly are introduced into the vertebra. Stylet - Diamond Tip The beveled tip stylet in the introducer needle can be removed and replaced with the diamond tip stylet, if preferred. Bone Drill Inserted through the working cannula to create a working space for the introduction of the CONFIDENCE PERIMETER Mesh Assembly within the vertebral body. CONFIDENCE PERIMETER Mesh Assembly Consists of the mesh bag, an injection needle assembly and a blunt tip stylet. The mesh bag is made from a porous fabric and is equipped with a winding handle at its proximal end. It protrudes 23mm beyond the working cannula. Cement is introduced into the mesh bag through a small diameter injection cannula. PERIMETER MESH CUTTER / CATALOG NO MESH CUTTER CASE / CATALOG NO Advance mesh cutter along the mesh bag until the mesh cutter contacts vertebral body. Then cut the mesh bag as close to the pedicle as possible. 18

19 Product Catalog CONFIDENCE PLUS KIT / CATALOG NO Components included in kit: CONFIDENCE High Viscosity Spinal Cement (11ml) Hydraulic Pump Cement Reservoir & Mixer Mallet & Needle Holder 2 Introducer Needles, Diamond Tip, 13G, 100mm Biopsy Needle, 15G, 230mm CONFIDENCE High Viscosity Spinal Cement (11ml) Component Descriptions CONFIDENCE High Viscosity Spinal Cement Consists of a ampoule of liquid monomer and a packet of powder polymer. Mixing the 2 components for seconds results in a high viscosity, radiopaque cement. This high viscosity texture allows for an 8 10 minutes at 68F (20C), and enables a controlled introduction of cement. Hydraulic Pump The hydraulic pump enables a smooth and effortless introduction of highly viscous cement into the vertebra. It is provided pre-filled with sterile water. Its flexible extension tube enables the physician to keep his/her hands outside of the direct fluoroscopy exposure field. Cement Mixer Consisting of a mixing bowl and mixer handle, this unique, closed mixer enables easy mixing of the cement. Cement Reservoir & Cement Reservoir Adaptor The cement reservoir is an 11 ml/cc transparent barrel with volumetric reference markings. The cement reservoir adaptor enables the transfer of the cement into the cement reservoir following mixing. Long Handled Radiolucent Mallet A long handled, 150-gram disposable mallet to aid in needle introduction. The long radiolucent handle allows the physician to tap the needle during fluoroscopy, while keeping his/her hands outside of the direct fluoroscopy field. Needle Holder Radiolucent needle holder, allowing for manipulation of needle placement while keeping the hands of the physician outside the direct fluoroscopy field. Introducer Needle, Diamond Tip, 13G, 100mm Consists of a needle and a stylet for bone access. The needle also acts as a cannula through which the cement is introduced into the vertebra. Biopsy Needle, 15G, 230mm The biopsy needle consists of a needle and stylet intended for use through an 11 gauge or 13 gauge introducer needle. 19

20 Vertebral Body Augmentation CONFIDENCE NEEDLELESS KIT / CATALOG NO Components included in kit: CONFIDENCE High Viscosity Spinal Cement (11ml) Hydraulic Pump Cement Reservoir & Mixer Mallet & Needle Holder Component Descriptions CONFIDENCE High Viscosity Spinal Cement Consists of a ampoule of liquid monomer and a packet of powder polymer. Mixing the 2 components for seconds results in a high viscosity, radiopaque cement. This high viscosity texture allows for an 8 10 minutes at 68F (20C), and enables a controlled introduction of cement. Hydraulic Pump The hydraulic pump enables a smooth and effortless introduction of highly viscous cement into the vertebra. It is provided pre-filled with sterile water. Its flexible extension tube enables the physician to keep his/her hands outside of the direct fluoroscopy exposure field. Cement Mixer Consisting of a mixing bowl and mixer handle, this unique, closed mixer enables easy mixing of the cement. Cement Reservoir & Cement Reservoir Adaptor The cement reservoir is an 11 ml/cc transparent barrel with volumetric reference markings. The cement reservoir adaptor enables the transfer of the cement into the cement reservoir following mixing. Long Handled Radiolucent Mallet A long handled, 150-gram disposable mallet to aid in needle introduction. The long radiolucent handle allows the physician to tap the needle during fluoroscopy, while keeping his/her hands outside of the direct fluoroscopy field. Needle Holder Radiolucent needle holder, allowing for manipulation of needle placement while keeping the hands of the physician outside the direct fluoroscopy field. 20

21 Product Catalog CONFIDENCE NEEDLELESS KIT / CATALOG NO Components included in kit: CONFIDENCE High Viscosity Spinal Cement (7cc) Hydraulic Pump Cement Reservoir & Mixer Component Descriptions CONFIDENCE High Viscosity Spinal Cement Consists of a ampoule of liquid monomer and a packet of powder polymer. Mixing the 2 components for seconds results in a high viscosity, radiopaque cement. This high viscosity texture allows for an 8 10 minutes at 68F (20C), and enables a controlled introduction of cement. Hydraulic Pump The hydraulic pump enables a smooth and effortless introduction of highly viscous cement into the vertebra. It is provided pre-filled with sterile water. Its flexible extension tube enables the physician to keep his/her hands outside of the direct fluoroscopy exposure field. Cement Mixer Consisting of a mixing bowl and mixer handle, this unique, closed mixer enables easy mixing of the cement. Cement Reservoir & Cement Reservoir Adaptor The cement reservoir is an 11 ml/cc transparent barrel with volumetric reference markings. The cement reservoir adaptor enables the transfer of the cement into the cement reservoir following mixing. 21

22 Vertebral Body Augmentation NOTES 22

23 Surgical Technique Guide 23

24 Reference: 1. An Ex Vivo Biomechanical Evaluation of the CONFIDENCE PERIMETER SPINAL CEMENT SYSTEM used in the Treatment of Compression Fractures. White Paper, DePuy Spine TM, (Re-order code ) INDICATIONS The CONFIDENCE PERIMETER SPINAL CEMENT SYSTEM is indicated for use in combination with the CONFIDENCE SPINAL CEMENT SYSTEM. The CONFIDENCE SPINAL CEMENT SYSTEM is intended for percutaneous delivery of CONFIDENCE High Viscosity Spinal Cement, which is indicated for fixation of pathological fractures of the vertebral body during vertebroplasty or kyphoplasty procedures. Painful vertebral compression fractures may result from osteoporosis, benign lesions (hemangioma), and malignant lesions (metastatic cancer, myeloma). CONTRAINDICATIONS The use of CONFIDENCE PERIMETER SPINAL CEMENT SYSTEM is contraindicated in patients presenting with any of the following conditions: Use of CONFIDENCE High Viscosity Spinal Cement for prophylaxis (such as in metastatic or osteoporotic patients with no evidence of acute vertebral fracture). Coagulation disorders, or severe cardiopulmonary disease. Haemorrhagic diasthesis. Non-pathological, acute, traumatic fractures of the vertebra. Patient clearly improving on medical therapy. Spinal stenosis (>20% by retropulsed fragments). Compromise of the vertebral body or walls of the pedicles. Compromise or instability of vertebral fractures due to posterior involvement. Anatomical damage of the vertebra that prevents safe access of the needle to the vertebral body. Vertebral body collapse to less than 1/3 (33%) original height. Vertebral plana (collapse >90%). Active or incompletely treated infection. Coagulopathy or inability to reverse anti-coagulant therapy (both during and approximately 24 hours post-procedure). Severe pulmonary insufficiency. Allergic reaction to any of the components of the CONFIDENCE High Viscosity Spinal Cement. Limited Warranty and Disclaimer: DePuy Spine products are sold with a limited warranty to the original purchaser against defects in workmanship and materials. Any other express or implied warranties, including warranties of merchantability or fitness, are hereby disclaimed. DePuy Spine EMEA is a trading division of DePuy International Ltd. DePuy International Ltd. DePuy France SAS DePuy Orthopädie GmbH Registered office: St. Anthony s Road, Leeds LS11 8DT, England Registered in England No Not all products are currently available in all markets. DePuy International Ltd. St. Anthony s Road Leeds LS11 8DT England Tel: +44 (0) Fax: +44 (0) FOR NON-USA USE ONLY DePuy Spine, Inc All rights reserved. ADDB/UM 09/10 30%

Bone Marrow Concentration. For customized cellular concentrations of platelet-rich plasma from bone marrow aspirate

Bone Marrow Concentration. For customized cellular concentrations of platelet-rich plasma from bone marrow aspirate ARTHREX TM SYSTEM Bone Marrow Concentration For customized cellular concentrations of platelet-rich plasma from bone marrow aspirate Arthrex Angel System Technology is what sets Angel apart from the competition.

More information

RapidPort EZ Access Port Kit DIRECTIONS FOR USE (DFU)

RapidPort EZ Access Port Kit DIRECTIONS FOR USE (DFU) RapidPort EZ Access Port Kit DIRECTIONS FOR USE (DFU) RapidPort EZ Access Port Kit Ref. No. C-2304 Standard RapidPort EZ Access Port Kit Ref. No. C-2306 Large RapidPort EZ Access Port Kit INTRODUCTION

More information

NORIAN DRILLABLE TECHNICAL MONOGRAPH. Fiber reinforced calcium phosphate phate bone void filler

NORIAN DRILLABLE TECHNICAL MONOGRAPH. Fiber reinforced calcium phosphate phate bone void filler NORIAN DRILLABLE Fiber reinforced calcium phosphate phate bone void filler TECHNICAL MONOGRAPH INDICATIONS & CONTRAINDICATIONS Drillable Drillable is intended for bony voids or defects of the extremities

More information

LAP-BAND System Access Port II Kit

LAP-BAND System Access Port II Kit LAP-BAND System Access Port II Kit DIRECTIONS FOR USE (DFU) Access Port Needle Access Port Priming Needle Access Port II End Plug Band Priming Needle Stainless Steel Connector Tubing LAP-BAND System Access

More information

RapidPort EZ Port Applier DIRECTIONS FOR USE (DFU)

RapidPort EZ Port Applier DIRECTIONS FOR USE (DFU) RapidPort EZ Port Applier DIRECTIONS FOR USE (DFU) RapidPort EZ Port Applier Ref. No. C-20390 RapidPort EZ Port Applier INTRODUCTION The RapidPort EZ Port Applier is an optional accessory for the LAP-BAND

More information

Product Name. Here. FLUORAway. Operator s Manual. Histology Innovation for a NEW Generation. Catalog # FA-S. A Division of General Data Healthcare

Product Name. Here. FLUORAway. Operator s Manual. Histology Innovation for a NEW Generation. Catalog # FA-S. A Division of General Data Healthcare A Division of General Data Healthcare Histology Innovation for a NEW Generation Pub No.: OM FA-S June 2013, Version 1.3 FA-S Operator s Manual Product Name FLUORAway Here Catalog # FA-S TABLE OF CONTENTS

More information

EasyClip Xpress Sterile staple solution. Operative technique

EasyClip Xpress Sterile staple solution. Operative technique EasyClip Xpress Sterile staple solution Operative technique Sterile staple solution Table of contents Indications 3 Operative technique 4 This publication sets forth detailed recommended procedures for

More information

PHP8 & PHP30 Series Hydraulic Pushers

PHP8 & PHP30 Series Hydraulic Pushers SPX Corporation 5885 th Street Rockford, IL 609-3699 USA Internet Address: http://www.powerteam.com Tech. Services: (800) 477-8326 Fax: (800) 765-8326 Order Entry: (800) 54-48 Fax: (800) 288-703 PHP8 PHP8H

More information

NeuroPort Array PN 4382, 4383, 6248, and 6249 Instructions for Use

NeuroPort Array PN 4382, 4383, 6248, and 6249 Instructions for Use 630 Komas Drive Suite 200 Salt Lake City UT 84108 USA P +1 801.582.5533 F +1 801.582.1509 www.blackrockmicro.com NeuroPort Array PN 4382, 4383, 6248, and 6249 Instructions for Use Revision 2.00 / LB-0612

More information

CAUTION: U.S. Federal law restricts this device to sale by or on the order of a licensed physician.

CAUTION: U.S. Federal law restricts this device to sale by or on the order of a licensed physician. TM CAUTION: U.S. Federal law restricts this device to sale by or on the order of a licensed physician. TABLE OF CONTENTS Section Port Styles 4 Description 5 Indications 5 Contraindications 5-6 Information

More information

Instructions for Use Reprocessed Ethicon ENDOPATH XCEL TM Universal Trocar Stability Sleeve. Reprocessed Device for Single Use.

Instructions for Use Reprocessed Ethicon ENDOPATH XCEL TM Universal Trocar Stability Sleeve. Reprocessed Device for Single Use. Reprocessed by Instructions for Use Reprocessed Ethicon ENDOPATH XCEL TM Universal Trocar Stability Sleeve Reprocessed Device for Single Use Caution: Federal (U.S.A.) law restricts this device to sale

More information

Instructions for Use Reprocessed Ethicon ENDOPATH XCEL TM Dilating Tip Trocar with OPTIVIEW Technology

Instructions for Use Reprocessed Ethicon ENDOPATH XCEL TM Dilating Tip Trocar with OPTIVIEW Technology Reprocessed by Instructions for Use Reprocessed Ethicon ENDOPATH XCEL TM Dilating Tip Trocar with OPTIVIEW Technology Reprocessed Device for Single Use Caution: Federal (U.S.A.) law restricts this device

More information

Fiberglass Dispenser Sump

Fiberglass Dispenser Sump Fiberglass Dispenser Sump Large Mouth LMM-XXXX-X models Single Wall and Double Wall Installation Instructions Franklin Fueling Systems 3760 Marsh Rd. Madison, WI 53718 USA Tel: +1 608 838 8786 800 225

More information

TREE I.V. Micro-Infusion System

TREE I.V. Micro-Infusion System TREE I.V. Micro-Infusion System Training Manual Revision Date: 07/24/15 Subject Headings Table of Contents Pages Introduction: TREE I.V. Micro-infusion 3 Parts of the TREE I.V. Kit 4 3 Easy Steps for the

More information

Trabecular Metal Revision Shell. Surgical Technique IMAGE TO COME. The Best Thing Next to Bone

Trabecular Metal Revision Shell. Surgical Technique IMAGE TO COME. The Best Thing Next to Bone Trabecular Metal Revision Shell Surgical Technique IMAGE TO COME The Best Thing Next to Bone Trabecular Metal Revision Shell Surgical Technique Initial and Long-Term Stability Fully-interconnected trabecular

More information

An improved mounting device for attaching intracranial probes in large animal models

An improved mounting device for attaching intracranial probes in large animal models Dunster Intensive Care Medicine Experimental (2015) 3:10 DOI 10.1186/s40635-015-0047-0 METHODOLOGY Open Access An improved mounting device for attaching intracranial probes in large animal models Kimble

More information

Innovations. For Life. EXAFLOW EXTERNAL CSF DRAINAGE PRODUCT CATALOGUE

Innovations. For Life. EXAFLOW EXTERNAL CSF DRAINAGE PRODUCT CATALOGUE Innovations. For Life. EXAFLOW EXTERNAL CSF DRAINAGE PRODUCT CATALOGUE Table of contents Ordering and Warranty Information... 1 External Drainage Systems... 2. External Drainage Set (EDS)... 2. External

More information

VALUE WITHOUT COMPARISON DIAGNOSTIC PROCEDURE TRAYS. Quality Without Compromise VALUE WITHOUT COMPARISON

VALUE WITHOUT COMPARISON DIAGNOSTIC PROCEDURE TRAYS. Quality Without Compromise VALUE WITHOUT COMPARISON Quality Without Compromise Value Without Comparison VALUE WITHOUT COMPARISON Quality Without Compromise QUALITY WITHOUT COMPROMISE VALUE WITHOUT COMPARISON VALUE WITHOUT COMPARISON DIAGNOSTIC PROCEDURE

More information

CRANIOS REINFORCED BONE VOID FILLER

CRANIOS REINFORCED BONE VOID FILLER CRANIOS REINFORCED BONE VOID FILLER The tougher bone cement TECHNICAL MONOGRAPH INTRODUCTION CRANIOS REINFORCED Bone Void Filler contains reinforcing fibers that impart an increase in toughness* to reduce

More information

HydroSet. Injectable HA Bone Substitute

HydroSet. Injectable HA Bone Substitute HydroSet Injectable HA Bone Substitute The Latest Advancement in Bone Substitute Technology Fast Setting Injectable or Manual Implantation Osteoconductive Excellent Wet-field Characteristics Isothermic

More information

Guidelines for QualiTru Line Sampling

Guidelines for QualiTru Line Sampling Guidelines for QualiTru Line Sampling QUALITRU GUIDELINES FOR LINE SAMPLING 1. Purpose: Following these Standard Operating Procedures, farms and processing plants may use the QualiTru aseptic sampling

More information

Description. Used materials. Warnings and precautions

Description. Used materials. Warnings and precautions USER MANUAL Description The GPS II System separates the blood components, present in blood, on the basis of density, through application of the GPS -Thermo International Equipment Company (IEC)-Centrifuge.

More information

Marrow Cellution. Autologous Bone Marrow Aspiration & Cancellous Bone Graft Harvesting. aspire-medical.eu

Marrow Cellution. Autologous Bone Marrow Aspiration & Cancellous Bone Graft Harvesting. aspire-medical.eu Marrow Cellution Autologous Bone Marrow Aspiration & Cancellous Bone Graft Harvesting aspire-medical.eu Aspire Medical Innovation Einsteinstr. 167 D-81677 München Traditional Trocar Design & Technique

More information

Instructions for Use Reprocessed Balloon Inflation Devices

Instructions for Use Reprocessed Balloon Inflation Devices Reprocessed by Instructions for Use Reprocessed Balloon Inflation Devices Reprocessed Device for Single Use Caution: Federal (U.S.A.) law restricts this device to sale by or on the order of a physician.

More information

CoAxial Spray Kit And Applicator Tips. Simplifying the Application of Platelet and Plasma Concentrate

CoAxial Spray Kit And Applicator Tips. Simplifying the Application of Platelet and Plasma Concentrate CoAxial Spray Kit And Applicator Tips Simplifying the Application of Platelet and Plasma Concentrate Intended for international use only. Indications for use of products and /or therapies contained herein

More information

Technical Bulletin No. MK3176. PREPARED: July, 2010

Technical Bulletin No. MK3176. PREPARED: July, 2010 Geoprobe DT45 Dual Tube Sampling System Standard Operating Procedure Technical Bulletin No. MK3176 PREPARED: July, 2010 A. B. C. D. E. A. Assembled outer casing with sample sheath and liner driven to collect

More information

ASAHI Masters PARKWAY HF KIT

ASAHI Masters PARKWAY HF KIT ASAHI Masters PARKWAY HF KIT Guide Wire Pre-loaded Microcatheter SYMBOLS 1 English 2 AMK-DT202 Ver.1.30 / 11TS023 SYMBOLS Legal manufacturer GC Minimum guiding catheter I.D. Do not use if package is damaged

More information

Percutaneous Spinal Injections problem presentation for HW1

Percutaneous Spinal Injections problem presentation for HW1 Percutaneous Spinal Injections problem presentation for HW1 Gabor Fichtinger, PhD Director of Engineering, Associate Research Professor of Computer Science and Radiology NSF-Funded Funded Engineering Research

More information

Model No. BWTDWFK-TRIFLO.

Model No. BWTDWFK-TRIFLO. Model No. BWTDWFK-TRIFLO www.bwt-uk.co.uk Before you start This tap is suitable for use at water supply pressures of 2 bar (min.) to 5 bar (max.). This product must be connected to the water supply in

More information

Life Sustaining Bone Marrow Stem Cells & Cancellous Bone Graft

Life Sustaining Bone Marrow Stem Cells & Cancellous Bone Graft Life Sustaining Bone Marrow Stem Cells & Cancellous Bone Graft Bone Marrow Cells (BMC) reside deep inside bone cavities in the most protected part of the body and are redundant throughout the organism.

More information

Artist Proof: Clotalyst. Autologous Serum Collection System

Artist Proof: Clotalyst. Autologous Serum Collection System Artist Proof: 01-15-10 Clotalyst Autologous Serum Collection System Clotalyst Autologous Serum The kit is designed for the preparation of autologous serum that is to be mixed with PRP and autograft or

More information

EPK- EasyTest Patch Test Kit Manual

EPK- EasyTest Patch Test Kit Manual EPK- EasyTest Patch Test Kit Manual L-2730 The Patch Test Kit from Schroeder provides the necessary tools to determine levels of both water and solid particulate contamination present in a particular fluid

More information

Advanced Life Support Basic Life Support Trainer

Advanced Life Support Basic Life Support Trainer Advanced Life Support Basic Life Support Trainer Model numbers 101-080 and 101-085 Phone: (845)-679-2475 (800)-431-4310 Fax: (845)-679-8996 www.simulaids.com Intubation Head Installation: Connect color

More information

RePlay Hemostasis Clip

RePlay Hemostasis Clip RePlay Hemostasis Clip Instructions for use. Read carefully prior to use. Caution: Federal (U.S.A.) Law restricts this device to sale by or on the order of a physician. Diversatek Healthcare Corporate

More information

BOLT MAINTENANCE GUIDE

BOLT MAINTENANCE GUIDE BOLT MAINTENANCE GUIDE Begin by disassembling your Bolt unit by unscrewing the two halves counterclockwise. 1 Important note: When you take the two halves apart it is important to pay attention to the

More information

ODYSSEY FLOOR MODEL. Owner s Manual. O-6400-F shown rev 118

ODYSSEY FLOOR MODEL. Owner s Manual. O-6400-F shown rev 118 Owner s Manual ODYSSEY FLOOR MODEL O-6400-F shown 67-1455 rev 118 3730 E. Southern Avenue, Phoenix, AZ 85040 USA 800-778-8779 Workhorseproducts.com 1 Table of Contents I. Introduction & Safety Information.

More information

PURE WATER. Installation, Operation and Maintenance Manual. Under Counter Water Filtration System Model PWDWHCUC1. Table of Contents !

PURE WATER. Installation, Operation and Maintenance Manual. Under Counter Water Filtration System Model PWDWHCUC1. Table of Contents ! Installation, Operation and Maintenance Manual Under Counter Water Filtration System Model PWDWHCUC1 PURE WATER! WARNING Please read carefully before proceeding with installation. Your failure to follow

More information

Merit Medical s Innovative Family. of Vascular Access Products. A Great Introduction

Merit Medical s Innovative Family. of Vascular Access Products. A Great Introduction Merit Medical s Innovative Family of Vascular Access Products A Great Introduction Sheath Introducers The Prelude family of sheath introducers is a comprehensive product line offering excellent transitions,

More information

Platelet Concentrate in Total Knees BIOLOGICS. This brochure is for International use only. It is not for distribution in the United States.

Platelet Concentrate in Total Knees BIOLOGICS. This brochure is for International use only. It is not for distribution in the United States. Platelet Concentrate in Total Knees BIOLOGICS This brochure is for International use only. It is not for distribution in the United States. Platelet Concentrate in Total Knees Total Knee Arthroplasty Total

More information

Advanced FEA Stress Analysis of Medical Instruments and Devices using Femap, NX Nastran and LS-DYNA

Advanced FEA Stress Analysis of Medical Instruments and Devices using Femap, NX Nastran and LS-DYNA Advanced FEA Stress Analysis of Medical Instruments and Devices Title: Medical device (aspiration tip) for bone marrow extraction for endoscopic surgery Keywords: Endoscopic surgical anvil, powder metallurgical

More information

Mounting Screw. Channel Filler. Light Angle Adjustable Screw. LED Board Vertical stud and blocking Locking Nose. Driver.

Mounting Screw. Channel Filler. Light Angle Adjustable Screw. LED Board Vertical stud and blocking Locking Nose. Driver. The Knife Edge Cove is a modular lighting system consisting of two parts: 1- An aluminum fixture housing extrusion 2- Drop-in lighting modules The fixture will provide a continuous knife edge architectural

More information

CEMEX Mixing. Cemex System Cemex Automatic Mixer Cemex Hand Mix

CEMEX Mixing. Cemex System Cemex Automatic Mixer Cemex Hand Mix CEMEX Mixing Techniques Cemex System Cemex Automatic Mixer Cemex Hand Mix Safety, Simplicity and Reliability in Bone Cement Cemex System Cemex System features a unique self-contained delivery system that

More information

Needle Loading. Seed Sterilization. Shielding. Seed Handling. SeedVac Seed Slider. Seed Sterilization and Sorting Tray Seed Sterilization Pill Box

Needle Loading. Seed Sterilization. Shielding. Seed Handling. SeedVac Seed Slider. Seed Sterilization and Sorting Tray Seed Sterilization Pill Box Prepare Prostate Brachytherapy treatments fast and with minimal exposure. Standard Imaging can provide you with complete kits that easily provide everything needed to start a prostate implant program.

More information

ASAHI Masters PARKWAY SOFT. ASAHI Masters PARKWAY HF

ASAHI Masters PARKWAY SOFT. ASAHI Masters PARKWAY HF ASAHI Masters PARKWAY SOFT Microcatheter ASAHI Masters PARKWAY HF Microcatheter SYMBOLS 1 English 2 AMK-DT207 Ver.1.40 / 11TS028 SYMBOLS Legal manufacturer Consult instructions for use Do not use if package

More information

High Temperature Aging Cell Instruction Manual

High Temperature Aging Cell Instruction Manual High Temperature Aging Cell Instruction Manual Part No. 760140001EA Rev. D TABLE OF CONTENTS Section Page Figure 1 General Information... 1 2 Safety Considerations... 3 3 Un-pressurized Testing Procedure...

More information

Trabecular Metal Acetabular Revision System Cup-Cage Construct

Trabecular Metal Acetabular Revision System Cup-Cage Construct Trabecular Metal Acetabular Revision System Cup-Cage Construct Surgical Technique The Best Thing Next to Bone Trabecular Metal Acetabular Revision System Cup-Cage Construct Trabecular Metal Acetabular

More information

VASCULAR ACCESS PRELUDE SHEATH INTRODUCERS. Merit Medical s Innovative Family of Vascular Access Products A GREAT INTRODUCTION

VASCULAR ACCESS PRELUDE SHEATH INTRODUCERS. Merit Medical s Innovative Family of Vascular Access Products A GREAT INTRODUCTION VASCULAR ACCESS PRELUDE SHEATH INTRODUCERS Merit Medical s Innovative Family of Vascular Access Products A GREAT INTRODUCTION THE PRELUDE FAMILY OF SHEATH INTRODUCERS A comprehensive product line offering

More information

2S and 3S Under-Counter UltraFiltration Drinking Water System Installation and Service Manual

2S and 3S Under-Counter UltraFiltration Drinking Water System Installation and Service Manual 2S and 3S Under-Counter UltraFiltration Drinking Water System Installation and Service Manual Table of Contents Section I. Section II. Section III. Section IV. Section V. Section VI. Introduction Specifications

More information

510(k) Summary. 510(k) Number: K Summary Prepared: August 31, 2017

510(k) Summary. 510(k) Number: K Summary Prepared: August 31, 2017 510(k) Summary 510(k) Number: K162613 Summary Prepared: August 31, 2017 Submitter: Dr. Fred Ma Repro Med System, Inc., D/B/A RMS Medical Products 24 Carpenter Road, Chester, NY 10918 Tel: 845-469-2042

More information

New consumer units. Design and functionality in plain sight.

New consumer units. Design and functionality in plain sight. New consumer units. Design and functionality in plain sight. CONSUMER UNITS A wide range of stylish solutions for any type of masonry or plasterboard wall. Consumer unit with white blind cover Consumer

More information

Drone Overview. Quick Start Guide. BACK 10 - Drone Status Indicator FRONT. 1 - Drone Overview. 4 - Disclaimers, Warnings & Guidelines

Drone Overview. Quick Start Guide. BACK 10 - Drone Status Indicator FRONT. 1 - Drone Overview. 4 - Disclaimers, Warnings & Guidelines 1 Drone Overview FRONT Quick Start Guide Drone Body ROTORS 1 - Drone Overview Orange LED Indicator 2 - Transmitter Overview Landing Skids Rotor Nut 4 - Disclaimers, Warnings & Guidelines Rotor 6 - Charging

More information

Circular Expansion Cement Curing Kit Instruction Manual

Circular Expansion Cement Curing Kit Instruction Manual Circular Expansion Cement Curing Kit Instruction Manual Manual No. 101443617, Revision C Instrument No. 205814 2015 Fann Instrument Company 2015 Fann Instrument Company Houston, Texas, USA All rights reserved.

More information

Installation Guide. Rain Screen Panel System

Installation Guide. Rain Screen Panel System Installation Guide Rain Screen Panel System Main Office 6860 South Abbott Road Orchard Park, New York 14127 tel 716.649.7490 toll free 1.888.214.3655 www.bostonvalley.com General Inquiries info@bostonvalley.com

More information

CapSure Permanent Fixation System

CapSure Permanent Fixation System CapSure Permanent Fixation System Permanent Fixation Redefined Advancing the Fixation Experience Recipient of 2015 SLS Innovations of the Year recognition. SOFT TISSUE REPAIR Right Procedure. Right Product.

More information

168 West Main Street, #922, New Market, Md Revision 5.2 6/6/2017

168 West Main Street, #922, New Market, Md Revision 5.2 6/6/2017 QUICK START GUIDE 168 West Main Street, #922, New Market, Md. 21774 TABLE OF CONTENTS Introduction... 2 General Culture Guidelines... 3 Reservoir Bottle and Cap Assembly Sterilization... 4 Pre-culture...

More information

Revision 5.1 4/17/17

Revision 5.1 4/17/17 QUICK START GUIDE 905 West 7th Street # 334, Frederick, MD 21701 TABLE OF CONTENTS Introduction... 2 General Culture Guidelines... 3 Reservoir Bottle and Cap Assembly Sterilization... 4 Pre-culture...

More information

RapidBrace Installation Guide

RapidBrace Installation Guide RapidBrace Installation Guide RapidBrace cast-in Brace Anchor System enables structural bracing in early strength concrete. RapidBrace is cast-into the concrete floor and is ready to load in concrete with

More information

MIIG INJECTABLE Graft The following languages are included in this packet:

MIIG INJECTABLE Graft The following languages are included in this packet: MIIG INJECTABLE Graft 128801-10 The following languages are included in this packet: English (en) Deutsch (de) Nederlands (nl) Français (fr) Español (es) Italiano (it) Português (pt) 中文 - Chinese (sch)

More information

Models FWB, LWB, MWB and HWB Brakes AIR CHAMP PRODUCTS. User Manual. (i) MTY (81)

Models FWB, LWB, MWB and HWB Brakes AIR CHAMP PRODUCTS. User Manual. (i) MTY (81) DIST. AUTORIZADO AIR CHAMP PRODUCTS User Manual Models FWB, LWB, MWB and HWB Brakes (i) DIST. AUTORIZADO In accordance with Nexen s established policy of constant product improvement, the specifications

More information

INSTRUCTIONS AIR JACKING BEAM. Installation, operation maintenance instructions & spare parts list for Molnar Air Jacking Beam JB11 / JB11A

INSTRUCTIONS AIR JACKING BEAM. Installation, operation maintenance instructions & spare parts list for Molnar Air Jacking Beam JB11 / JB11A Installation, operation maintenance instructions & spare parts list for Molnar Air Jacking Beam JB11 / JB11A AIR JACKING BEAM INSTRUCTIONS The Air Jacking Beam is for use with Molnar Four Post Hoists.

More information

RAPIDSORB Injectable Polymer System (IPS) Surgical Technique

RAPIDSORB Injectable Polymer System (IPS) Surgical Technique RAPIDSORB Injectable Polymer System (IPS) Surgical Technique TABLE OF CONTENTS INTRODUCTION RAPIDSORB Injectable Polymer System (IPS) 2 Indications and Contraindications 4 SURGICAL TECHNIQUE Preparation

More information

Material List by Procedure. The following lists are provided to ensure that the correct supplies are in place prior to the procedure.

Material List by Procedure. The following lists are provided to ensure that the correct supplies are in place prior to the procedure. Material List by Procedure Purpose: Policy: The following lists are provided to ensure that the correct supplies are in place prior to the procedure. Changes to material lists must be approved by the medical

More information

UNITS MUST BE INSTALLED AFTER PRESSURE TANK

UNITS MUST BE INSTALLED AFTER PRESSURE TANK Chlorine setting in P.P.M. for center tube and cap Hole #1 = 2 P.P.M. Hole #2 = 8 P.P.M. Hole #3 = 30 P.P.M. Hole #A = 3.5 P.P.M. Hole #B = 13 P.P.M. Hole #C = 40 P.P.M. All holes closed will deliver 60

More information

Uncompromised accuracy in neuroimaging and treatment

Uncompromised accuracy in neuroimaging and treatment Uncompromised accuracy in neuroimaging and treatment 1 Helping clinicians improve patients lives. 2 Innovation built on strong foundations The tools used by the surgeon must be adapted to the task and,

More information

Geoprobe SDT45 Sonic Dual Tube Sampling System

Geoprobe SDT45 Sonic Dual Tube Sampling System Geoprobe SDT45 Sonic Dual Tube Sampling System Standard Operating Procedure Technical Bulletin No. MK3189 PREPARED: January, 2012 A. B. C. D. E. A. Assembled outer casing with sample sheath and liner advanced

More information

VARIO XX28 Stainless Steel Time And Temperature Controlled Electric Model

VARIO XX28 Stainless Steel Time And Temperature Controlled Electric Model VARIO XX28 Stainless Steel Time And Temperature Controlled Electric Model Operators Manual Issue 5 For correct operation of this autoclave it is essential that the chamber be filled with water up to the

More information

Leksell. Vantage Stereotactic System. Advancing stereotactic neurosurgery

Leksell. Vantage Stereotactic System. Advancing stereotactic neurosurgery Leksell Vantage Stereotactic System Advancing stereotactic neurosurgery Innovation built on strong foundations Since the founding of our company by Lars Leksell more than 45 years ago, Elekta has delivered

More information

VARIO XX28 Stainless Steel Time And Temperature Controlled Electric Model

VARIO XX28 Stainless Steel Time And Temperature Controlled Electric Model DIXONS SURGICAL INSTRUMENTS LTD, 1 ROMAN COURT HURRICANE WAY, WICKFORD BUSINESS PARK WICKFORD, ESSEX, SS11 8YB, UK TEL: +44 (0)1268 764614 - FAX: +44 (0)1268 764615 EMAIL: info@dixons-uk.com - WEB: http://www.dixons-uk.com

More information

RAPIDSORB Injectable Polymer System (IPS) Surgical Technique

RAPIDSORB Injectable Polymer System (IPS) Surgical Technique RAPIDSORB Injectable Polymer System (IPS) Surgical Technique TABLE OF CONTENTS INTRODUCTION RAPIDSORB Injectable Polymer System (IPS) 2 Indications and Contraindications 4 SURGICAL TECHNIQUE Preparation

More information

POOL STEP INSTRUCTIONS (NE1087WH)

POOL STEP INSTRUCTIONS (NE1087WH) POOL STEP INSTRUCTIONS (NE1087WH) 6492 The sequence of the instructions may vary at points depending on the type of step system that you have purchased. If there are different instructions for your system,

More information

Owner s Manual Calcite Cartridge System

Owner s Manual Calcite Cartridge System Owner s Manual Calcite Cartridge System Page 2 Copyright 2011 Enviro Water Solutions, Inc. All rights reserved. All information contained herein is the property of Pelican Water Technologies. Pelican Water

More information

SAN TM PRODUCT FAMILY SAFE AUTO-NEEDLES TM. Better injectors. Better compliance. Better care.

SAN TM PRODUCT FAMILY SAFE AUTO-NEEDLES TM. Better injectors. Better compliance. Better care. SAFE AUTO-NEEDLES TM SAN TM PRODUCT FAMILY Better injectors. Better compliance. Better care. Addressing patient-centered challenges in injectable drug delivery Compatible with: Prefilled syringes (luer

More information

8: IV/SQ Maintenance

8: IV/SQ Maintenance Section 8: IV/SQ Maintenance Section Author(s): abalmforth & wbethurum Field Guide Section 8: IV/SQ Maintenance 1 Section 8: IV & SQ Maintenance 2 Section 8: IV/SW Maintenance Field Guide Section Contents

More information

ORGANISATION AFRICAINE DE LA PROPRIETE INTELLECTUELLE

ORGANISATION AFRICAINE DE LA PROPRIETE INTELLECTUELLE 19 ORGANISATION AFRICAINE DE LA PROPRIETE INTELLECTUELLE 11 N 17145 8 51 Inter. CI. A61B 17/34 FASCICULE DE BREVET D INVENTION 21 22 Numéro de dépôt : 1201400470 (PCT/IB13/000649) Date de dépôt : 10/04/2013

More information

Product Bulletin 2K-3.2. stamixco. mix it up..

Product Bulletin 2K-3.2. stamixco. mix it up.. stamixco Product Bulletin 2K-3.2 mix it up.. 2-Component Resin Mixing Technology Plastic Disposable & Metal Static Mixers for Mixing Viscous Materials 2 Introduction For more than 40 years, static mixers

More information

Water-Water Titration for the MicroCal ITC200

Water-Water Titration for the MicroCal ITC200 Water-Water Titration for the MicroCal ITC200 Goal: Water-Water titrations are a good way to check for possible problems, especially the cleanliness of the ITC and your skills in cell filling. If a water-water

More information

booster EX2 / EX3 Operating Manual Extrusion Welder Weldy AG Galileo-Strasse 10 CH-6056 Kaegiswil / Switzerland TECHSPANGROUP

booster EX2 / EX3 Operating Manual Extrusion Welder Weldy AG Galileo-Strasse 10 CH-6056 Kaegiswil / Switzerland TECHSPANGROUP booster EX2 / EX3 Extrusion Welder Operating Manual Weldy AG Galileo-Strasse 10 CH-6056 Kaegiswil / Switzerland TECHSPANGROUP Australia: Tel. 1-800 148 791 Fax. 1-800 148 799 www.weldy.com.au New Zealand:

More information

EJOT Chemical Anchoring Systems. High strength anchoring of threaded resin studs into masonry blocks, cracked and non-cracked concrete

EJOT Chemical Anchoring Systems. High strength anchoring of threaded resin studs into masonry blocks, cracked and non-cracked concrete EJOT Chemical Anchoring Systems High strength anchoring of threaded resin studs into masonry blocks, cracked and non-cracked concrete The Quality Connection www.ejot.co.uk EJOT Multifix VSF EJOT Multifix

More information

LITe LIF Less invasive posterior lumbar interbody fusion procedure

LITe LIF Less invasive posterior lumbar interbody fusion procedure LITe LIF Less invasive posterior lumbar interbody fusion procedure LITe LIF Access LITe Midline Retractor Fixation Xia CT Interbody Tritanium PL Cage Instruments LITe Decompression 2.0 Power CD3 and RemB

More information

3.2 SuperTie 50K Installation Guide

3.2 SuperTie 50K Installation Guide 3.2 SuperTie 50K Installation Guide FORM ERECTION 1. This G50K Gripper has been shipped partially disassembled. It is necessary for you only to attach the base plate. The Gripper Body, Base Plate, and

More information

Operating Instructions

Operating Instructions Operating Instructions Series B37000 Hose Reels BB37118 L CB37118 L EB37118 L12D BB37122 L CB37122 L EB37122 L12D BB37128 L CB37128 L EB37128 L12D Dimensional Data C A B D Part # A B C D BB/CB37118 L 31

More information

PTC1. Paint Test Cell Kit

PTC1. Paint Test Cell Kit PTC1 Paint Test Cell Kit Operator s Manual Copyright 2017 Gamry Instruments, Inc. Revision 2.1 September 15, 2017 900-00057 Limited Warranty Gamry Instruments, Inc. warrants to the original user of this

More information

Catalog of products Liquid Management

Catalog of products Liquid Management Catalog of products Liquid Management February 2018 Contents Company Profile 7-13 Extension Lines 14-23 Infusion Systems 24-27 Transfusion Systems 28-48 Spikes & Filters 49-51 Components & Parts Company

More information

SAFETECH 3.2K ST v1.0. User s manual. Opale-Parachutes.com. Please read this manual before first use.

SAFETECH 3.2K ST v1.0. User s manual. Opale-Parachutes.com. Please read this manual before first use. 2016 v1.0 User s manual SAFETECH 3.2K ST60 Please read this manual before first use. Thank you for choosing Opale-Parachutes. This user s guide includes all the information you need to assemble and use

More information

Installing the Sno-free Panel Eave System for New Roofs

Installing the Sno-free Panel Eave System for New Roofs Installing the Sno-free Panel Eave System for New Roofs SFP CLEAT SFP BASE SFP EXTRUSION SFP TOP COVER Step # 1 Placing the Cleat Install Ice and Water Shield as per code to the edge of sheeting. Install

More information

PVC Project Assembly Guide a FORMUFIT field manual

PVC Project Assembly Guide a FORMUFIT field manual PVC Project Assembly Guide a FORMUFIT field manual www.formufit.com 1 PVC Project Assembly Guide 1 Welcome. This guide is designed to be friendly, knowledgable and to provide options for different people

More information

Agilent G2615A Bar Code Reader

Agilent G2615A Bar Code Reader Agilent G2615A Bar Code Reader User Information Agilent Technologies Notices Agilent Technologies, Inc. 2002 No part of this manual may be reproduced in any form or by any means (including electronic storage

More information

First Experiences with the Ziehm Vision FD Mobile C-Arm with Flat-Panel Detector

First Experiences with the Ziehm Vision FD Mobile C-Arm with Flat-Panel Detector 01 White Paper No. 02/2009 First Experiences with the Ziehm Vision FD Mobile C-Arm with Flat-Panel Detector Leiden University Medical Center (LUMC) in the Netherlands is the first hospital in the world

More information

Physician Training Manual. LASE Endoscopic Discectomy

Physician Training Manual. LASE Endoscopic Discectomy Physician Training Manual LASE Endoscopic Discectomy 900248.001.06 2 DISCLAIMER This manual contains information about the LASE procedure. Clarus Medical, LLC recommends, that before performing a LASE

More information

User Manual. Electrochemical Test Cell ECC-Opto-Std. Release October EL-CELL GmbH

User Manual. Electrochemical Test Cell ECC-Opto-Std. Release October EL-CELL GmbH User Manual Release 2.61 Electrochemical Test Cell ECC-Opto-Std 24. October 2016 2016 EL-CELL GmbH The information in this manual has been carefully checked and believed to be accurate; however, no responsibility

More information

Everything in its place. OWNER S MANUAL. TRINITY ECOSTORAGE 6-TIER NSF 48 x 18 x 72 WIRE SHELVING RACK w/wheels CHROME Model # TBFC-0907

Everything in its place. OWNER S MANUAL. TRINITY ECOSTORAGE 6-TIER NSF 48 x 18 x 72 WIRE SHELVING RACK w/wheels CHROME Model # TBFC-0907 Everything in its place. OWNER S MANUAL TRINITY ECOSTORAGE 6-TIER NSF 48 x 18 x 72 WIRE SHELVING RACK w/wheels CHROME Model # TBFC-0907 PARTS LIST Your TRINITY EcoStorage 6-Tier NSF Wire Shelving Rack

More information

3034 Owen Drive Antioch, TN USA Fax: Maybachstraße 10. Revised 01/17

3034 Owen Drive Antioch, TN USA Fax: Maybachstraße 10. Revised 01/17 CLN VersaPak Modular Instrument Tray Symmetry Surgical Inc. 3034 Owen Drive Antioch, TN 37013 USA 1-800-251-3000 Fax: 1-615-964-5566 www.symmetrysurgical.com EC REP Symmetry Surgical GmbH 78532 Tuttlingen,

More information

LINEAR Lead model SC2108 Lead Extension model SC3108. Spinal Cord Stimulation System

LINEAR Lead model SC2108 Lead Extension model SC3108. Spinal Cord Stimulation System LINEAR Lead model SC2108 Lead Extension model SC3108 Spinal Cord Stimulation System Physician Lead Manual 9055095-001 Rev C Physician Lead Manual CAUTION. Investigational device. Limited by federal law

More information

ST XX28 Stainless Steel Thermostatically Controlled Electric Model

ST XX28 Stainless Steel Thermostatically Controlled Electric Model DIXONS SURGICAL INSTRUMENTS LTD, 1 ROMAN COURT HURRICANE WAY, WICKFORD BUSINESS PARK WICKFORD, ESSEX, SS11 8YB, UK TEL: +44 (0)1268 764614 - FAX: +44 (0)1268 764615 EMAIL: info@dixons-uk.com - WEB: http://www.dixons-uk.com

More information

Repackaging the Router

Repackaging the Router APPENDIXA Revised: March 16, 29, Introduction This appendix describes two separate but related router repackaging tasks: How to remove the router from the equipment rack and repackage the router for shipment

More information

ESSENTIAL FOR SURVIVAL

ESSENTIAL FOR SURVIVAL ESSENTIAL FOR SURVIVAL LIFESAVER LIBERTY USER MANUAL SLI00 YOUR LIFESAVER LIBERTY BOTTLE PRIMING Priming is an essential step that must be performed before you start to use the bottle to drink from. Correct

More information

Experiment 4: Heat Engine Cycle

Experiment 4: Heat Engine Cycle TD-8572A Experiment 4: Heat Engine Cycle Experiment 4: Heat Engine Cycle Equipment Needed Heat Engine/Gas Law Apparatus PASCO Interface and Data Collection Software PASCO Dual Pressure Sensor PASCO Rotary

More information

LSA-PLUS Series 2. Connection and Disconnection Modules and Accessories. 1 st Edition

LSA-PLUS Series 2. Connection and Disconnection Modules and Accessories. 1 st Edition LSA-PLUS Series 2 Connection and Disconnection Modules and Accessories 1 st Edition LSA-PLUS Series 2 Connection and Disconnection Modules and Accessories A technically and commercially superior quick

More information

Proweld Equipment Owner & Maintenance Manual FIELD 12 TRENCH (Widos 4900)

Proweld Equipment Owner & Maintenance Manual FIELD 12 TRENCH (Widos 4900) Proweld Equipment Owner & Maintenance Manual FIELD 12 TRENCH (Widos 4900) 35 Green Street, PO Box 653, Malden, MA 02148 Tel: (781) 321-5409 - Fax (781) 321-4421 - Toll Free: (800) 343-3618 www.asahi-america.com

More information

Premium Water Filtration Systems

Premium Water Filtration Systems SETUP AND MAINTENANCE GUIDE Premium Water Filtration Systems RO SYSTEM 150 GPD Product # 738300 & Product # 738306 RO SYSTEM 300 GPD Product # 738305 & Product # 738308 WHAT S IN THE BOX? Premium Water

More information