1 Polymer Block Cleaning Kit For Applied Biosystems 3730/3730xl DNA Analyzers Protocol April 29, :17 am, Title.fm

2 Copyright 2003, Applied Biosystems. All rights reserved. For Research Use Only. Not for use in diagnostic procedures. NOTICE TO PURCHASER: REFER TO THE APPLIED BIOSYSTEMS 3730/3730XL DNA ANALYZER USER GUIDE AND THE REAGENT PACKAGE INSERTS FOR LIMITED LABEL LICENSE OR DISCLAIMER INFORMATION. Information in this document is subject to change without notice. Applied Biosystems assumes no responsibility for any errors that may appear in this document. This document is believed to be complete and accurate at the time of publication. In no event shall Applied Biosystems be liable for incidental, special, multiple, or consequential damages in connection with or arising from the use of this document. TRADEMARKS: ABI PRISM and its design and Applied Biosystems are registered trademarks of Applera Corporation or its subsidiaries in the U.S. and certain other countries. AB (Design) and Applera are trademarks of Applera Corporation or its subsidiaries in the U.S. and certain other countries. All other trademarks are the sole property of their respective owners. Part Number Rev. B 04/2003 April 29, :17 am, Title.fm

3 Contents About This Kit Safety Polymer Block Care Cleaning the Upper Polymer Block Cleaning the Lower Polymer Block Reinstalling the Upper and Lower Polymer Blocks Obtaining Technical Support Polymer Block Cleaning Kit for 3730/3730xl DNA Analyzers April 29, :17 am, PolyBlockCleanTOC.fm iii

4 Polymer Block Cleaning Kit for 3730/3730xl DNA Analyzers April 29, :17 am, PolyBlockCleanTOC.fm iv

5 About This Kit About This Kit Purpose When to Use The Polymer Block Cleaning Kit (P/N ) allows you to clean the upper and lower polymer block assemblies of Applied Biosystems 3730/3730xl DNA Analyzers. Clean the upper and lower polymer block assemblies when you: Install a syringe onto the polymer block Remove or replace a capillary array Perform a shut down Have polymer in the syringe or blocks at room temperature for more than 1 week Kit Contents The Polymer Block Cleaning Kit contains the following items: Item Part Number Syringe, 20-mL, silicone-free Syringe adaptor, 6-mm, 2 each, black Polymer Block Cleaning Kit for 3730/3730xl DNA Analyzers 1

6 Safety Warnings POP-7 Polymer Heed the following warning while preforming the procedures in this protocol. CHEMICAL HAZARD. POP-7 polymer causes eye, skin, and respiratory tract irritation. Read the MSDS, and follow the handling instructions. Wear appropriate protective eyewear, clothing, and gloves. Polymer Block Care Avoiding Damage to the Blocks The upper and lower polymer blocks can be irreversibly damaged if: Polymer dries in the polymer channels of the upper or lower polymer block The inside of the channel surface is scratched The block is exposed to organic solvent The block is exposed to temperatures greater than 70 C IMPORTANT! Use clean, powder-free, silicone-free latex gloves whenever you handle the polymer block or any item in the polymer path to minimize background fluorescence. 2 Polymer Block Cleaning Kit for 3730/3730xl DNA Analyzers

7 Cleaning the Upper Polymer Block Cleaning the Upper Polymer Block Installed Polymer Blocks The polymer blocks as they appear when correctly installed on the instrument are shown in Fig µL glass syringe Upper polymer block SST check valve Polymer cap tube assembly Polymer bottle Buffer valve pin Polymer block interconnect tubing Lower polymer block Anode Buffer jar GR2213 Figure 1 Installed polymer blocks Upper Polymer Block The upper polymer block with the plastic cleaning syringe and syringe adaptor attached is shown in Fig 2. GR2208 Upper polymer block SST check valve 6-mm syringe luer adaptor 20-mL silicone-free plastic syringe Interconnecting polymer PEEK tubing Figure 2 Upper polymer block cleaning setup Polymer Block Cleaning Kit for 3730/3730xl DNA Analyzers 3

8 Removing the Upper Block Cleaning the Upper Polymer Block To remove the upper polymer block: 1. Remove the 500-µL glass syringe from the upper polymer block. 2. Remove the capillary array from the upper polymer block. IMPORTANT! Do not remove the stainless steel (SST) check valve from the bottom of the upper polymer block. 3. Disconnect the polymer tubing from the lower polymer block. 4. Remove the polymer cap tubing from the SST check valve (finger tight). To clean the upper polymer block: 1. Fill the 20-mL silicone-free plastic syringe (P/N ) with warm deionized water. IMPORTANT! Do not use water above 70 C. 2. Fit the 6-mm luer plastic syringe adaptor (P/N ) onto the plastic 20-mL syringe. IMPORTANT! Do not use the 500-µL glass syringe to force air or water through the channels. This action damages the syringe plunger and causes the syringe to leak. 3. Thread the 6-mm luer plastic syringe adaptor into the stainless steel check valve. IMPORTANT! To clean out the check valve, attach the luer fitting as shown in figure 2. Water must flow freely only from the plastic syringe to the upper polymer block. If the plastic syringe is able to aspirate water, the check valve is clogged and requires more rinsing. 4. Force at least 3 plastic syringe volumes of warm deionized water (70 C or below) through each channel in turn by sealing their openings with your fingers. 5. Force deionized water through the polymer block tubing. 4 Polymer Block Cleaning Kit for 3730/3730xl DNA Analyzers

9 Cleaning the Upper Polymer Block 6. Inspect the channels for dried polymer, visible as white residue. Wash partially obstructed channels with warm deionized water until the dried polymer is removed. IMPORTANT! It may take a while for the warm water to clear the obstruction. Do not use any pointed or sharp objects to clear the channel, even if the channel is completely obstructed with dried polymer. 7. After washing, rinse the upper polymer block and all the fittings thoroughly with deionized water. 8. After rinsing, follow the procedures described below: If you... Are placing the upper polymer block on the instrument for immediate use directly after cleaning Are not placing the upper polymer block on the instrument for immediate use directly after cleaning Then... Shake the water out of the channels and install the upper polymer block on the instrument. During the bubble purge wizard procedure, any remaining deionized water is flushed out with fresh polymer. Leave the upper polymer block in a beaker of deionized water until you are ready to use it. Note: Make sure that the check valve is completely submerged. IMPORTANT! Do not use canned compressed air to dry polymer blocks. Do not use the 500-µL glass syringe to force air or water through the channels. This action damages the syringe plunger and causes the syringe to leak. Polymer Block Cleaning Kit for 3730/3730xl DNA Analyzers 5

10 Cleaning the Lower Polymer Block Lower Polymer Block The lower polymer block with the syringe and syringe adaptor attached is shown in Fig. 3. Buffer valve pin 20-mL silicone-free syringe 6-mm syringe luer adaptor (black) Lower polymer block GR2191 Figure 3 Lower polymer block cleaning setup Removing the Lower Block To remove the lower polymer block: 1. Make sure that the buffer valve is open (in the up position). 2. Remove the interconnect PEEK tube and fitting from the lower polymer block. 3. Remove the lower polymer block from the instrument. 6 Polymer Block Cleaning Kit for 3730/3730xl DNA Analyzers

11 Cleaning the Lower Polymer Block Cleaning the Lower Block To clean the lower polymer block: 1. Rinse all the fittings with warm deionized water. Soak any fittings that are covered with polymer. IMPORTANT! Do not use water above 70 C to rinse the fittings or the polymer block. 2. With the lower polymer block in warm water (70 C or below), move the buffer valve in and out to ensure any encrusted polymer is removed from the guide channel. IMPORTANT! Do not remove any of the components from the lower polymer block. 3. Fill the 20-mL silicone-free plastic syringe (P/N ) with warm water (70 C or below). IMPORTANT! Do not use the 500-µL glass syringe to force air or water through the channels. This action damages the syringe plunger and causes the syringe to leak. 4. Fit the 6-mm plastic syringe adaptor (P/N ) onto the 20-mL silicone-free plastic syringe. 5. Thread the 6-mm plastic syringe adaptor into the polymer block where the polymer fitting was originally located. 6. Force several plastic syringe volumes of warm water (70 C or below) through the channel. 7. Visually inspect the channels for dried polymer, which appears as white residue. Wash partially obstructed channels with warm water (70 C or below) until the dried polymer is removed. IMPORTANT! It may take a while for the warm water to clear the obstruction. Do not use any pointed or sharp objects to clear the channel, even if the channel is completely obstructed with dried polymer. 8. Rinse the lower polymer block and all the fittings thoroughly using deionized water. IMPORTANT! Do not leave the lower polymer block soaking in warm deionized water because the O-ring will crack over time. 9. Shake excess deionized water from the lower polymer block, then store it until needed. Polymer Block Cleaning Kit for 3730/3730xl DNA Analyzers 7

12 Reinstalling the Upper and Lower Polymer Blocks Reinstalling the Polymer Blocks To reinstall the polymer blocks on the instrument: 1. Connect the tubing between the two blocks before attaching the blocks to the instrument: a. Insert a ferrule, at one of the tubing, into the upper polymer block and then rotate until finger tight. Do not over tighten. Note: The end of the interconnect PEEK tube must press against the end of the hole. This ensures that no bubbles are trapped during Bubble Purge wizard procedure. Also, make sure the interconnect PEEK tubing is not crimped. b. Insert a ferrule, at the other end of the tubing, into the lower polymer block and then rotate until finger tight. Do not over tighten. IMPORTANT! An air gap between the interconnect PEEK tubing and the stop may cause arcing. IMPORTANT! To ensure that you correctly attach the upper polymer block to the instrument, make sure that the check valve on the bottom of the block faces down (as shown below). Check valve at bottom of the upper block, facing down toward the lower block. Figure 4 Check valve on upper block 8 Polymer Block Cleaning Kit for 3730/3730xl DNA Analyzers

13 Reinstalling the Upper and Lower Polymer Blocks 2. Simultaneously push the upper polymer block and the lower polymer block onto their respective guide pins (as shown below). Then, simultaneously push both blocks to a point about halfway onto the guide pins (toward the instrument wall). Upper block guide pins IMPORTANT! Be sure not to kink the interconnect PEEK tube (tube between the upper and lower polymer block). A kink causes a restriction in the tube which can result in softening during electrophoresis requiring the tube to be replaced. Lower block guide pins Figure 5 Reinstalling the polymer blocks I Polymer Block Cleaning Kit for 3730/3730xl DNA Analyzers 9

14 3. Finish by pushing each block individually until they are flush against the instrument wall (as shown below). Push one......and then the other Yoke IMPORTANT! Be sure to align the yoke and the pin before the final push to the wall of the instrument. Any misalignment may cause pin damage Pin Figure 6 Reinstalling the polymer blocks II µL glass syringe Upper polymer block SST check valve Polymer cap tube assembly Polymer bottle Buffer valve pin Polymer block interconnect tubing Lower polymer block Anode Buffer jar GR Install a clean drip tray. 5. Install the capillary array using the Install Capillary Array wizard in the 3730/3730xl Data Collection software. 6. Install remaining components on the block assembly. 10 Polymer Block Cleaning Kit for 3730/3730xl DNA Analyzers

15 Obtaining Technical Support Obtaining Technical Support Service and Support A services and support page is available on the Applied Biosystems Web site: At the services and support page, you can: Search through frequently asked questions (FAQs) Submit a question directly to Technical Support Order Applied Biosystems user documents, MSDSs, certificates of analysis, and other related documents Download PDF documents Obtain information about customer training Download software updates and patches In addition, the services and support page provides worldwide telephone and fax numbers to contact Applied Biosystems Technical Support and Sales facilities. Polymer Block Cleaning Kit for 3730/3730xl DNA Analyzers 11

16 Intentionally blank 12 Polymer Block Cleaning Kit for 3730/3730xl DNA Analyzers

17 Intentionally blank Polymer Block Cleaning Kit for 3730/3730xl DNA Analyzers 13

18 Intentionally blank 14 Polymer Block Cleaning Kit for 3730/3730xl DNA Analyzers

19

20 Headquarters 850 Lincoln Centre Drive Foster City, CA USA Phone: Toll Free (In North America): Fax: Worldwide Sales and Support Applied Biosystems vast distribution and service network, composed of highly trained support and applications personnel, reaches 150 countries on six continents. For sales office locations and technical support, please call our local office or refer to our Web site at Applera Corporation is committed to providing the world s leading technology and information for life scientists. Applera Corporation consists of the Applied Biosystems and Celera Genomics businesses. Printed in USA, 04/2003 Part Number Rev. B

SUBJECT: System Software Patch Provides Support for Fast SYBR Green Master Mix and RNA-to-C T

SUBJECT: System Software Patch Provides Support for Fast SYBR Green Master Mix and RNA-to-C T User Bulletin Applied Biosystems StepOne and StepOnePlus Real-Time PCR Systems March 2008 SUBJECT: System Software Patch Provides Support for Fast SYBR Green Master Mix and RNA-to-C T In This User Bulletin

More information

Applied Biosystems 7900HT Fast Real-Time PCR System. Performing Fast Gene Quantitation with 384-Well Plates DRAFT

Applied Biosystems 7900HT Fast Real-Time PCR System. Performing Fast Gene Quantitation with 384-Well Plates DRAFT User Bulletin Applied Biosystems 7900HT Fast Real-Time PCR System October 3, 2005 SUBJECT: Performing Fast Gene Quantitation with 384-Well Plates In This User Bulletin System Overview System Requirements

More information

Applied Biosystems 7500 Fast Real-Time PCR System

Applied Biosystems 7500 Fast Real-Time PCR System User Bulletin Applied Biosystems 7500 Fast Real-Time PCR System July 2006 SUBJECT: Using Expert Mode on the Applied Biosystems 7500 Fast Real-Time PCR System with the Applied Biosystems 7500 Fast System

More information

Protocol. Path-ID Multiplex One-Step RT-PCR Kit Protocol TaqMan probe-based multiplex one-step real-time RT-PCR detection of RNA targets

Protocol. Path-ID Multiplex One-Step RT-PCR Kit Protocol TaqMan probe-based multiplex one-step real-time RT-PCR detection of RNA targets Protocol TaqMan probe-based multiplex one-step real-time RT-PCR detection of RNA targets For Research Use Only. Not for use in diagnostic procedures. Information in this document is subject to change without

More information

Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v3.0

Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v3.0 User Bulletin Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v3.0 September 9, 2010 SUBJECT: Protocols for Processing AmpFlSTR PCR Amplification Kit PCR Products In This

More information

PrepFiler Express and PrepFiler Express BTA Forensic DNA Extraction Kits

PrepFiler Express and PrepFiler Express BTA Forensic DNA Extraction Kits QUICK REFERENCE CARD PrepFiler Express and PrepFiler Express BTA Forensic DNA Extraction Kits Note: For safety and biohazard guidelines, refer to the Safety section in the PrepFiler Express and PrepFiler

More information

Perform an HRM Methylation Study

Perform an HRM Methylation Study Quick Reference Card Perform an HRM Methylation Study This quick reference card provides brief procedures for performing an HRM methylation study using MeltDoctor HRM Master Mix and High Resolution Melting

More information

Volume of Each Kit Reagent (ml) XTerminator Solution. 2-mL mL 1, mL 2,

Volume of Each Kit Reagent (ml) XTerminator Solution. 2-mL mL 1, mL 2, BigDye XTerminator Purification Kit Quick Reference Card For safety and biohazard guidelines, refer to the Safety section in the Applied Biosystems BigDye XTerminator Purification Kit Protocol (PN 4374408).

More information

Validation Using SDS Software Version on the Applied Biosystems 7500 Real-Time PCR System and the ABI PRISM 7000 Sequence Detection System

Validation Using SDS Software Version on the Applied Biosystems 7500 Real-Time PCR System and the ABI PRISM 7000 Sequence Detection System User Bulletin Quantifiler Kits April 2006 SUBJECT: Validation Using SDS Software Version 1.2.3 on the Applied Biosystems 7500 Real-Time PCR System and the ABI PRISM 7000 Sequence Detection System Note:

More information

User Bulletin. Applied Biosystems 3400 DNA Synthesizer. New Features in 3400 DNA Synthesizer Firmware v In This User Bulletin

User Bulletin. Applied Biosystems 3400 DNA Synthesizer. New Features in 3400 DNA Synthesizer Firmware v In This User Bulletin User Bulletin Applied Biosystems 3400 DNA Synthesizer April 2007 SUBJECT: New Features in 3400 DNA Synthesizer Firmware v1.4.1 In This User Bulletin This user bulletin covers: Overview...........................................................2

More information

Multiplex RT for TaqMan Array Human MicroRNA Panel

Multiplex RT for TaqMan Array Human MicroRNA Panel f Multiplex RT for TaqMan Low Density Array Multiplex RT for TaqMan Array Human MicroRNA Panel Quick Reference Card For safety and biohazard guidelines, refer to the Safety section in the Multiplex RT

More information

ABI PRISM 3100 Genetic Analyzer. ABI PRISM cm Capillary Array for High Throughput Microsatellite and SNP Genotyping

ABI PRISM 3100 Genetic Analyzer. ABI PRISM cm Capillary Array for High Throughput Microsatellite and SNP Genotyping User Bulletin ABI PRISM 3100 Genetic Analyzer November 12, 2001 SUBJECT: ABI PRISM 3100 22-cm Capillary Array for High Throughput Microsatellite and SNP Genotyping About This Bulletin High throughput Microsatellite

More information

PrepSEQ Nucleic Acid Extraction Kit

PrepSEQ Nucleic Acid Extraction Kit Quick Reference Card PrepSEQ Nucleic Acid Extraction Kit Note: For safety and biohazard guidelines, refer to the Safety appendix in the PrepSEQ Nucleic Acid Extraction Kit Protocol (PN 4400739). For all

More information

Supported Protocol: Getting Started with the QuantStudio 3D Digital PCR System

Supported Protocol: Getting Started with the QuantStudio 3D Digital PCR System USER BULLETIN Supported Protocol: Getting Started with the QuantStudio 3D Digital PCR System Publication Number MAN0009821 Revision A.0 Overview... 1 Reuired materials... 2 Workflow... 3 Protocol... 3

More information

RNAscope 2.5 LS Assay- RED Combined with Immunohistochemistry (IHC)

RNAscope 2.5 LS Assay- RED Combined with Immunohistochemistry (IHC) TECHNICAL NOTE RNAscope 2.5 LS Assay- RED Combined with Immunohistochemistry (IHC) Introduction This Technical Note provides guidelines for performing automated in situ hybridization (ISH) using an RNAscope

More information

TaqMan PreAmp Master Mix Kit. Protocol

TaqMan PreAmp Master Mix Kit. Protocol TaqMan PreAmp Master Mix Kit Protocol Copyright 2007, 2010 Applied Biosystems. All rights reserved. For Research Use Only. Not for use in diagnostic procedures. Information in this document is subject

More information

RNAscope 2.5 LS Assay- BROWN Combined with Immunohistochemistry (IHC)

RNAscope 2.5 LS Assay- BROWN Combined with Immunohistochemistry (IHC) TECHNICAL NOTE RNAscope 2.5 LS Assay- BROWN Combined with Immunohistochemistry (IHC) Introduction This Technical Note provides guidelines for performing automated in situ hybridization (ISH) using an RNAscope

More information

Axiom TM 2.0 Automated Target Prep Protocol

Axiom TM 2.0 Automated Target Prep Protocol Quick Reference Card Axiom TM 2.0 Automated Target Prep Protocol Stage 1. DNA Amplification Introduction Running the Axiom 2.0 Assay requires the following sets of steps: 1. Genomic DNA Prep, described

More information

User Bulletin. ABI PRISM 7000 Sequence Detection System DRAFT. SUBJECT: Software Upgrade from Version 1.0 to 1.1. Version 1.

User Bulletin. ABI PRISM 7000 Sequence Detection System DRAFT. SUBJECT: Software Upgrade from Version 1.0 to 1.1. Version 1. User Bulletin ABI PRISM 7000 Sequence Detection System 9/15/03 SUBJECT: Software Upgrade from Version 1.0 to 1.1 The ABI PRISM 7000 Sequence Detection System version 1.1 software is a free upgrade. Also

More information

PATH Protein Microarray Slide (Product Numbers and ) Sandwich Immunoassay Protocol

PATH Protein Microarray Slide (Product Numbers and ) Sandwich Immunoassay Protocol L002 Rev 092805 1 of 6 Table of Contents 1.0 Introduction 1 2.0 Contents...... 1 3.0 Materials Required, But Not Included....... 1 4.0 Storage and Stability......... 2 5.0 Safety and Handling... 2 6.0

More information

AccuSEQ Real-Time PCR Detection Software

AccuSEQ Real-Time PCR Detection Software Quick Reference Card AccuSEQ Real-Time PCR Detection Software Mycoplasma SEQ Experiments Note: For safety and biohazard guidelines, refer to the Safety section in the AccuSEQ Real-Time PCR Detection Software

More information

Prepare a Barcoded Fragment Library with the SOLiD Fragment Library Barcoding Kit 1 96

Prepare a Barcoded Fragment Library with the SOLiD Fragment Library Barcoding Kit 1 96 QUICK REFERENCE CARD Prepare a Barcoded Fragment Library with the SOLiD Fragment Library Barcoding Kit 1 96 Note: For safety and biohazard guidelines, refer to the Safety section in the Applied Biosystems

More information

Water-Water Titration for the MicroCal ITC200

Water-Water Titration for the MicroCal ITC200 Water-Water Titration for the MicroCal ITC200 Goal: Water-Water titrations are a good way to check for possible problems, especially the cleanliness of the ITC and your skills in cell filling. If a water-water

More information

TaqMan Sample-to-SNP Kit

TaqMan Sample-to-SNP Kit Quick Reference Card TaqMan Sample-to-SNP Kit Note: For safety and biohazard guidelines, refer to the Safety appendix in the Applied Biosystems TaqMan Sample-to-SNP Kit Protocol (PN 4402136). For all chemicals

More information

Calcium-EDTA Tutorial for the MicroCal ITC200

Calcium-EDTA Tutorial for the MicroCal ITC200 Calcium-EDTA Tutorial for the MicroCal ITC200 Getting Started 1. The computer and the ITC200 are normally left on. If they have been switched off for any reason makes sure that you boot them in the order

More information

BigDye Direct Cycle Sequencing Kit

BigDye Direct Cycle Sequencing Kit BigDye Direct Cycle Sequencing Kit QUICK REFERENCE CARD Note: For safety and biohazard guidelines, refer to the Safety section in the BigDye Direct Cycle Sequencing Kit Protocol (PN 4458040). For every

More information

Standard and Low-Flow 100 PLUS Powder Feed Pumps

Standard and Low-Flow 100 PLUS Powder Feed Pumps Instruction Sheet P/N 036G Standard and Low-Flow 00 PLUS Powder Feed Pumps Introduction Standard and low-flow 00 PLUS powder feed pumps are used to pump organic and metallic powder coatings to powder spray

More information

SCIEX Anion Analysis Kit For P/ACE MDQ and P/ACE MDQ plus Capillary Electrophoresis Systems. Instruction Guide

SCIEX Anion Analysis Kit For P/ACE MDQ and P/ACE MDQ plus Capillary Electrophoresis Systems. Instruction Guide For P/ACE MDQ and P/ACE MDQ plus Capillary Electrophoresis Systems A49108AE May 2015 AB Sciex Pte. Ltd and its affiliates disclaims all warranties with respect to this document, expressed or implied, including

More information

VetMAX -Plus Multiplex One-Step RT-PCR Kit TaqMan probe-based multiplex one-step real-time RT-PCR amplification of RNA targets.

VetMAX -Plus Multiplex One-Step RT-PCR Kit TaqMan probe-based multiplex one-step real-time RT-PCR amplification of RNA targets. VetMAX -Plus Multiplex One-Step RT-PCR Kit TaqMan probe-based multiplex one-step real-time RT-PCR amplification of RNA targets Protocol Copyright 2010, Life Technologies Corporation. All rights reserved.

More information

Operating Instructions

Operating Instructions POROS Affinity Depletion Cartridges Operating Instructions 1 Product Description POROS Affinity Depletion Cartridges and columns are available in two affinity chemistries: POROS Anti-HSA Cartridges Packed

More information

Electroelution. Teachers Handbook. (Cat. # BE 602) think proteins! think G-Biosciences

Electroelution. Teachers Handbook. (Cat. # BE 602) think proteins! think G-Biosciences PR097 G-Biosciences 1-800-628-7730 1-314-991-6034 technical@gbiosciences.com A Geno Technology, Inc. (USA) brand name Electroelution Teachers Handbook (Cat. # BE 602) think proteins! think G-Biosciences

More information

Relative Fluorescent Quantitation on Capillary Electrophoresis Systems:

Relative Fluorescent Quantitation on Capillary Electrophoresis Systems: Application Note Relative Fluorescent Quantitation on the 3130 Series Systems Relative Fluorescent Quantitation on Capillary Electrophoresis Systems: Screening for Loss of Heterozygosity in Tumor Samples

More information

Saliva DNA Collection, Preservation and Isolation Kit 50 Individual Devices

Saliva DNA Collection, Preservation and Isolation Kit 50 Individual Devices 3430 Schmon Parkway Thorold, ON, Canada L2V 4Y6 Phone: 866-667-4362 (905) 227-8848 Fax: (905) 227-1061 Email: techsupport@norgenbiotek.com Saliva DNA Collection, Preservation and Isolation Kit 50 Individual

More information

Arcturus CapSure LCM MicroCaps

Arcturus CapSure LCM MicroCaps Arcturus CapSure LCM MicroCaps USER BULLETIN Procedures for isolating DNA and RNA from laser capture microdissections Catalog Numbers A3053, A3054, A3055 Publication Number MAN004607 Revision B.0 For Research

More information

PowerPlex 5-Dye Matrix Standards, 3100/3130

PowerPlex 5-Dye Matrix Standards, 3100/3130 TECHNICAL BULLETIN PowerPlex 5-Dye Matrix Standards, 3100/3130 Instructions for Use of Product DG4700 Note: These matrix fragments are compatible with the ABI PRISM 3100 and 3100-Avant and Applied Biosystems

More information

LIFluor EnhanCE. for use with dsdna 1000 and dsdna 20,000 Kits

LIFluor EnhanCE. for use with dsdna 1000 and dsdna 20,000 Kits LIFluor EnhanCE for use with dsdna 1000 and dsdna 20,000 Kits For the Separation of Double Stranded DNA by P/ACE Fluorescent Capillary Electrophoresis 725824 AB September 2006 Beckman Coulter, Inc. 4300

More information

Continuously Regenerated Trap (CR-TC 500, Capillary)

Continuously Regenerated Trap (CR-TC 500, Capillary) Thermo Scientific Continuously Regenerated Trap (CR-TC 500, Capillary) Column Product Manual P/N: November 2012 Part of Thermo Fisher Scientific Product Manual for Continuously Regenerated Trap Column

More information

Protocol for DNA and RNA Isolation from the Same FFPE Sample

Protocol for DNA and RNA Isolation from the Same FFPE Sample C19682AA December 2017 FormaPure Total: Protocol for DNA and RNA Isolation from the Same FFPE Sample Refer to www.beckmancoulter.com/ifu for updated protocols. For questions regarding this protocol, call

More information

User Bulletin. Veriti 96-Well Thermal Cycler AmpFlSTR Kit Validation. Overview

User Bulletin. Veriti 96-Well Thermal Cycler AmpFlSTR Kit Validation. Overview User Bulletin Veriti 96-Well Thermal Cycler AmpFlSTR Kit Validation June 2009 SUBJECT: AmpFlSTR PCR Amplification Kit validation on the Veriti 96-Well Thermal Cycler with 0.2 ml sample block format In

More information

BrightStar -Plus Membranes

BrightStar -Plus Membranes BrightStar -Plus Membranes Positively Charged Nylon Membrane Part Number AM10100, AM10102, AM10104 A. Product Description Part Number AM10100 AM10102 AM10104 Contents 5 membranes: 15 cm x 15 cm Small roll:

More information

Operating Instructions

Operating Instructions Sequazyme DNA Standards Kit Operating Instructions 1 Product Description The Sequazyme DNA Standards Kit allows you to do either or both of the following: Test the MassGenotyping Solution 1 (MGS1) System,

More information

Extended Protocol for RNA Isolation from FFPE Sample

Extended Protocol for RNA Isolation from FFPE Sample C19683AA December 2017 FormaPure RNA: Extended Protocol for RNA Isolation from FFPE Sample Refer to www.beckmancoulter.com/ifu for updated protocols. For questions regarding this protocol, call Technical

More information

MagMAX DNA Multi-Sample Ultra Kit

MagMAX DNA Multi-Sample Ultra Kit MagMAX DNA Multi-Sample Ultra Kit High-throughput isolation of PCR-ready DNA from buccal swabs Catalog Numbers A25597 and A25598 Pub. No. MAN0010293 Rev. D.0 WARNING! Read the Safety Data Sheets (SDSs)

More information

Instructions for Use Life Science Kits & Assays

Instructions for Use Life Science Kits & Assays Instructions for Use Life Science Kits & Assays Order No.: 845-KS-6010010 10 reactions 845-KS-6010050 50 reactions 845-KS-6010250 250 reactions Publication No.: HB_KS-6010_e_180329 This documentation describes

More information

Eureka Genotyping Assay

Eureka Genotyping Assay Quick Reference Card Eureka Genotyping Assay This quick reference card is intended for experienced users. For detailed instruction, please refer to the Eureka Genotyping Workflow User Guide (P/N 703400

More information

Axiom TM 2.0 for NIMBUS TM Automated Target Prep Protocol

Axiom TM 2.0 for NIMBUS TM Automated Target Prep Protocol Quick Reference Card Axiom TM 2.0 for NIMBUS TM Automated Target Prep Protocol Introduction Running the Axiom 2.0 Assay requires the following sets of steps: 1. Genomic DNA Prep, described in either the

More information

Calcium and Magnesium; Chlorophosphonazo Rapid Liquid Method Method to 1000 µg/l Ca and Mg as CaCO 3 (ULR) Pour-Thru Cell

Calcium and Magnesium; Chlorophosphonazo Rapid Liquid Method Method to 1000 µg/l Ca and Mg as CaCO 3 (ULR) Pour-Thru Cell Hardness, Total DOC316.53.01045 Calcium and Magnesium; Chlorophosphonazo Rapid Liquid Method Method 8374 4 to 1000 µg/l Ca and Mg as CaCO 3 (ULR) Pour-Thru Cell Scope and application: For boiler and ultrapure

More information

PTC1. Paint Test Cell Kit

PTC1. Paint Test Cell Kit PTC1 Paint Test Cell Kit Operator s Manual Copyright 2017 Gamry Instruments, Inc. Revision 2.1 September 15, 2017 900-00057 Limited Warranty Gamry Instruments, Inc. warrants to the original user of this

More information

Size-Select Kit for NGS Library Preparation Product # 53600

Size-Select Kit for NGS Library Preparation Product # 53600 3430 Schmon Parkway Thorold, ON, Canada L2V 4Y6 Phone: 866-667-4362 (905) 227-8848 Fax: (905) 227-1061 Email: techsupport@norgenbiotek.com Size-Select Kit for NGS Library Preparation Product # 53600 Product

More information

MagMAX DNA Multi-Sample Ultra 2.0 Kit

MagMAX DNA Multi-Sample Ultra 2.0 Kit USER GUIDE MagMAX DNA Multi-Sample Ultra 2.0 Kit High throughput isolation of DNA from saliva Catalog Number A36570 Pub. No. MAN0017324 Rev. C.0 WARNING! Read the Safety Data Sheets (SDSs) and follow the

More information

Tempus Blood RNA Tube and Tempus 12-Port RNA Isolation Kit. Protocol

Tempus Blood RNA Tube and Tempus 12-Port RNA Isolation Kit. Protocol Tempus Blood RNA Tube and Tempus 12-Port RNA Isolation Kit Protocol Copyright 2007, Applied Biosystems. All rights reserved. For Research Use Only. Not for use in diagnostic procedures. Information in

More information

Applied Biosystems TaqMan Array Microfluidic Cards

Applied Biosystems TaqMan Array Microfluidic Cards Performing Rapid Cycling Gene Quantitation on TaqMan Array Microfluidic Cards Applied Biosystems TaqMan Array Microfluidic Cards User Bulletin JULY 2010 SUBJECT: In this user bulletin Performing Rapid

More information

ULTRAhyb -Oligo Buffer

ULTRAhyb -Oligo Buffer ULTRAhyb -Oligo Buffer Hybridization Buffer for Oligonucleotides Part Number AM8663 A. Product Description ULTRAhyb -Oligo Hybridization Buffer has been optimized specifically for hybridization of oligonucleotides

More information

CAT. NO. BR BR BR SIZE 10 preps 50 preps 250 preps. Buffer LYSIS LC 12 ml 25 ml 120 ml. Buffer LYSIS LD 10 ml 25 ml 125 ml

CAT. NO. BR BR BR SIZE 10 preps 50 preps 250 preps. Buffer LYSIS LC 12 ml 25 ml 120 ml. Buffer LYSIS LD 10 ml 25 ml 125 ml Product Insert GenUP Plant DNA Kit LOT: See product label EXPIRY DATE: See product label ORDERING INFORMATION PRODUCT GenUP Plant DNA Kit CAT. NO. BR0700801 BR0700802 BR0700803 SIZE 10 preps 50 preps 250

More information

Fast, easy, reliable, ultra-pure transfection grade plasmid DNA Maxiprep using a microcentrifuge spin-column.

Fast, easy, reliable, ultra-pure transfection grade plasmid DNA Maxiprep using a microcentrifuge spin-column. INSTRUCTION MANUAL ZymoPURE II Plasmid Maxiprep Kit Catalog Nos. D4202 & D4203 (Patent Pending) Highlights Fast, easy, reliable, ultra-pure transfection grade plasmid DNA Maxiprep using a microcentrifuge

More information

Fast, easy, reliable, ultra-pure transfection grade plasmid DNA Maxiprep using a microcentrifuge spin-column.

Fast, easy, reliable, ultra-pure transfection grade plasmid DNA Maxiprep using a microcentrifuge spin-column. INSTRUCTION MANUAL ZymoPURE II Plasmid Maxiprep Kit Catalog Nos. D4202 & D4203 (Patent Pending) Highlights Fast, easy, reliable, ultra-pure transfection grade plasmid DNA Maxiprep using a microcentrifuge

More information

Mini SAL Salivary DNA Isolation Kit (Spin Column Method)

Mini SAL Salivary DNA Isolation Kit (Spin Column Method) Mini SAL Salivary DNA Isolation Kit (Spin Column Method) Catalog No. DNAX-504-30 Oasis Diagnostics Corporation - Pioneering Oral Fluid Diagnostic Solutions TABLE OF CONTENTS Storage, Stability & Intended

More information

Expression Array System

Expression Array System Integrated Science for Gene Expression Applied Biosystems Expression Array System Expression Array System SEE MORE GENES The most complete, most sensitive system for whole genome expression analysis. The

More information

Circular Expansion Cement Curing Kit Instruction Manual

Circular Expansion Cement Curing Kit Instruction Manual Circular Expansion Cement Curing Kit Instruction Manual Manual No. 101443617, Revision C Instrument No. 205814 2015 Fann Instrument Company 2015 Fann Instrument Company Houston, Texas, USA All rights reserved.

More information

ebioscience BrdU Kit for IHC/ICC Colorimetric Catalog Number: RUO: For Research Use Only. Not for use in diagnostic procedures.

ebioscience BrdU Kit for IHC/ICC Colorimetric Catalog Number: RUO: For Research Use Only. Not for use in diagnostic procedures. Page 1 of 1 ebioscience BrdU Kit for IHC/ICC Colorimetric Catalog Number: 8800-6599 RUO: For Research Use Only. Not for use in diagnostic procedures. Product Information Contents: ebioscience BrdU Kit

More information

TaKaRa MiniBEST Plasmid Purification Kit Ver.4.0

TaKaRa MiniBEST Plasmid Purification Kit Ver.4.0 Cat. # 9760 For Research Use TaKaRa MiniBEST Plasmid Purification Kit Ver.4.0 Product Manual Table of Contents I. Description... 3 II. Kit Components... 3 III. Shipping and Storage... 4 IV. Preparation

More information

Agilent Multiple Affinity Removal Spin Cartridges for the Depletion of High-Abundant Proteins from Human Proteomic Samples.

Agilent Multiple Affinity Removal Spin Cartridges for the Depletion of High-Abundant Proteins from Human Proteomic Samples. Agilent Multiple Affinity Removal Spin Cartridges for the Depletion of High-Abundant Proteins from Human Proteomic Samples Instructions Third edition July 2005 General Information Introduction The Agilent

More information

BigDye Terminator v1.1 Cycle Sequencing Kit

BigDye Terminator v1.1 Cycle Sequencing Kit BigDye Terminator v1.1 Cycle Sequencing Kit Protocol August 27, 2002 12:32 pm, 4337036A_v1.1Title.fm Copyright 2002, Applied Biosystems. All rights reserved. For Research Use Only. Not for use in diagnostic

More information

PowerPlex 4C Matrix Standard

PowerPlex 4C Matrix Standard TECHNICAL MANUAL PowerPlex 4C Matrix Standard Instruc ons for Use of Product DG4800 Printed 10/15 TMD048 PowerPlex 4C Matrix Standard All technical literature is available at: www.promega.com/protocols/

More information

AmpFlSTR Identifiler PCR Amplification Kit

AmpFlSTR Identifiler PCR Amplification Kit Product Bulletin Human Identification AmpFlSTR Identifiler PCR Amplification Kit Fifteen STR (short tandem repeat) loci and Amelogenin co-amplified in a single tube Incorporation of the same proven, reliable

More information

User Bulletin. ABI PRISM 310 Genetic Analyzer. A Fast Native Protocol for the Analysis of PCR Products. In This Bulletin

User Bulletin. ABI PRISM 310 Genetic Analyzer. A Fast Native Protocol for the Analysis of PCR Products. In This Bulletin User Bulletin June 23, 1999 (updated 12/2000) A Fast Native Protocol for the Analysis of PCR Products In This Bulletin This User Bulletin describes the Fast Native Protocol for the separation of DNA fragments

More information

Use of the MSI Analysis System with the Applied Biosystems 3500 Genetic Analyzer and POP-7 Polymer

Use of the MSI Analysis System with the Applied Biosystems 3500 Genetic Analyzer and POP-7 Polymer Use of the MSI Analysis System with the Applied Biosystems 3500 Genetic Analyzer and POP-7 Polymer MSI Analysis System Application Note Promega Corporation Notes: The MSI Analysis System is For Research

More information

TRACE 1300 and TRACE 1310 Gas Chromatographs Dual Detector Microfluidics Installation Guide

TRACE 1300 and TRACE 1310 Gas Chromatographs Dual Detector Microfluidics Installation Guide TRACE 1300 and TRACE 1310 Gas Chromatographs Dual Detector Microfluidics Installation Guide P/N 31709711 Revision A October 2013 2013 Thermo Fisher Scientific Inc. All rights reserved. TSQ 8000, ISQ, TRACE

More information

Procedure for Use of the 3130XL Genetic Analyzer

Procedure for Use of the 3130XL Genetic Analyzer Procedure for Use of the 3130XL Genetic Analyzer 1.0 Purpose- This procedure specifies the steps for daily maintenance, preparing and running samples on the 3130XL ABI Genetic Analyzer. 2.0 Scope- This

More information

AmpFlSTR SEfiler. User s Manual. PCR Amplification Kit. DRAFT March 22, :04 pm, SEfiler_TitlePage.fm

AmpFlSTR SEfiler. User s Manual. PCR Amplification Kit. DRAFT March 22, :04 pm, SEfiler_TitlePage.fm AmpFlSTR SEfiler PCR Amplification Kit User s Manual March 22, 2012 2:04 pm, SEfiler_TitlePage.fm 2012 Life Technologies Corporation. All rights reserved. Printed in the U.S.A. TRADEMARKS: AB (Design),

More information

Protocol for Creating Custom RT and Preamplification Pools using TaqMan MicroRNA Assays

Protocol for Creating Custom RT and Preamplification Pools using TaqMan MicroRNA Assays USER BULLETIN Protocol for Creating Custom RT and Preamplification Pools using TaqMan MicroRNA Assays Publication Part Number 4465407 Revision Date January 2013 (Rev. C) SUBJECT: In this user bulletin

More information

Tempus Blood RNA Tube and Tempus Spin RNA Isolation Kit. Protocol

Tempus Blood RNA Tube and Tempus Spin RNA Isolation Kit. Protocol Tempus Blood RNA Tube and Tempus Spin RNA Isolation Kit Protocol Copyright 2008, Applied Biosystems. All rights reserved. For Research Use Only. Not for use in diagnostic procedures. Information in this

More information

Table 163 Instrument-specific information. Instrument Pour-thru Kit Cell orientation Adapter. DR 6000 LQV Arrow faces right

Table 163 Instrument-specific information. Instrument Pour-thru Kit Cell orientation Adapter. DR 6000 LQV Arrow faces right , 8374 Calcium and Magnesium; Chlorophosphonazo Rapid Liquid Method ULR (4 to 1000 µg/l Ca and Mg as CaCO 3 ) Scope and Application: For boiler and ultra pure water. DOC316.53.01045 Method 8374 Pour-Thru

More information

PowerPlex 5C Matrix Standards, 310

PowerPlex 5C Matrix Standards, 310 TECHNICAL MANUAL PowerPlex 5C Matrix Standards, 310 Instruc ons for Use of Product DG5640 Printed 10/15 TMD050 PowerPlex 5C Matrix Standards, 310 All technical literature is available at: www.promega.com/protocols/

More information

Color, True and Apparent, LR

Color, True and Apparent, LR Color, True and Apparent, LR DOC316.53.01252 Platinum-Cobalt Standard Method 1, 2, 3 Method 8025 3 to 200 color units Scope and application: For water, wastewater and seawater; equivalent to NCASI method

More information

foodproof Sample Preparation Kit II

foodproof Sample Preparation Kit II For food testing purposes FOR IN VITRO USE ONLY foodproof Sample Preparation Kit II Version 1, October 2007 For isolation of bacterial DNA from enrichment cultures of food samples prior to PCR analysis

More information

HighPrep FFPE DNA 96. Genomic DNA isolation from FFPE (Formalin-Fix, Paraffin-Embedded) Tissue Magnetic beads based chemistry High-throughput capable

HighPrep FFPE DNA 96. Genomic DNA isolation from FFPE (Formalin-Fix, Paraffin-Embedded) Tissue Magnetic beads based chemistry High-throughput capable MAGBIO ACCELERATING genomic research HighPrep FFPE DNA 96 Catalog Nos. HPFF-D96, HPFF-D96X4 Manual Revision v2.03 Genomic DNA isolation from FFPE (Formalin-Fix, Paraffin-Embedded) Tissue Magnetic beads

More information

PowerPlex Matrix Standards, 3100/3130

PowerPlex Matrix Standards, 3100/3130 TECHNICAL BULLETIN PowerPlex Matrix Standards, 3100/3130 Instruc ons for use of Product DG4650 Note: The PowerPlex Matrix Standards, 3100/3130, can be used to perform spectral calibra on on the Applied

More information

Tali Viability Kit Dead Cell Green

Tali Viability Kit Dead Cell Green Tali Viability Kit Dead Cell Green *for use with Tali Assays: Green, Green + Red* Catalog no. A10787 Table 1 Contents and storage Material Amount Concentration Storage* Stability Tali Dead Cell Green (SYTOX

More information

Life science unlimited

Life science unlimited Life science unlimited Manual innuprep DNA Micro Kit Publication No.: HB_KS-1010_e_080605 This documentation describes the state at the time of publishing. It needs not necessarily agree with future versions.

More information

BigDye Terminator v3.1 Cycle Sequencing Kit

BigDye Terminator v3.1 Cycle Sequencing Kit BigDye Terminator v3.1 Cycle Sequencing Kit Protocol August 27, 2002 12:32 pm, 4337035A_v3.1Title.fm Copyright 2002, Applied Biosystems. All rights reserved. For Research Use Only. Not for use in diagnostic

More information

CENTRUM CUP STACKER OPERATIONAL INSTRUCTIONS

CENTRUM CUP STACKER OPERATIONAL INSTRUCTIONS TM CENTRUM CUP STACKER OPERATIONAL INSTRUCTIONS Ensure that all personnel read the points listed within this leaflet and that a copy is given to staff involved with the installation and maintenance of

More information

Model No. BWTDWFK-TRIFLO.

Model No. BWTDWFK-TRIFLO. Model No. BWTDWFK-TRIFLO www.bwt-uk.co.uk Before you start This tap is suitable for use at water supply pressures of 2 bar (min.) to 5 bar (max.). This product must be connected to the water supply in

More information

MagniSort Mouse CD3 Positive Selection Kit Catalog Number: RUO: For Research Use Only. Not for use in diagnostic procedures.

MagniSort Mouse CD3 Positive Selection Kit Catalog Number: RUO: For Research Use Only. Not for use in diagnostic procedures. Page 1 of 2 MagniSort Mouse CD3 Positive Selection Kit RUO: For Research Use Only. Not for use in diagnostic procedures. Mouse splenocytes were unsorted (left) or sorted with the MagniSort Mouse CD3 Positive

More information

Optima Vector Plank. Installation Instructions. Ceiling

Optima Vector Plank. Installation Instructions. Ceiling Ceiling Systems Between us, ideas become reality Optima Vector Plank Installation Instructions 1. GENERAL 1.1. Product Description The Vector products referenced in these instructions are made from fiberglass.

More information

Urethane Booster Pump Seal Kit (Zenith or Nordson -Type)

Urethane Booster Pump Seal Kit (Zenith or Nordson -Type) Instruction Sheet P/N Urethane Booster Pump Seal Kit (Zenith or Nordson -Type) 1. Introduction Use this instruction sheet to install a urethane booster seal kit. There are three kits available depending

More information

SQ-1 ROOF TRUSS WEG-IT TABLE

SQ-1 ROOF TRUSS WEG-IT TABLE WEG-IT Steel Top Table SQ1 Service SQ-1 ROOF TRUSS WEG-IT TABLE Operators Manual FOREWORD This manual explains the proper maintenance of Square 1 Design Roof Truss WEG-IT Table as well as periodic inspection

More information

Human DNA Alu Amplification by Polymerase Chain Reaction (PCR)* Laboratory Procedure

Human DNA Alu Amplification by Polymerase Chain Reaction (PCR)* Laboratory Procedure Human DNA Alu Amplification by Polymerase Chain Reaction (PCR)* Laboratory Procedure *Polymerase Chain Reaction is covered by patents owned by Hoffmann-La Roche, Inc. This experiment was adapted from Laboratory

More information

Antibody Array User s Guide

Antibody Array User s Guide Antibody Microarray User s Guide Rev. 9.1 Page 1 TABLE OF CONTENTS Introduction... 2 How it works... 3 Experimental considerations... 4 Components... 4 Additional Material Required... 6 Reagent Preparation...

More information

Thermo Scientific GeneJET Whole Blood RNA Purification Mini Kit #K0761

Thermo Scientific GeneJET Whole Blood RNA Purification Mini Kit #K0761 PRODUCT INFORMATION Thermo Scientific GeneJET Whole Blood RNA Purification Mini Kit #K0761 www.thermoscientific.com/onebio #K0761 Lot Exp. CERTIFICATE OF ANALYSIS Thermo Scientific GeneJET Whole Blood

More information

HighPrep FFPE RNA 96. RNA isolation from FFPE (Formalin-Fix, Paraffin-Embeded) Tissue Magnetic beads based chemistry High-throughput capable PROTOCOL

HighPrep FFPE RNA 96. RNA isolation from FFPE (Formalin-Fix, Paraffin-Embeded) Tissue Magnetic beads based chemistry High-throughput capable PROTOCOL MAGBIO ACCELERATING genomic research HighPrep FFPE RNA 96 Catalog Nos. HPFF-R96, HPFF-R96X4 Manual Revision v3.00 RNA isolation from FFPE (Formalin-Fix, Paraffin-Embeded) Tissue Magnetic beads based chemistry

More information

MagniSort Mouse B cell Enrichment Kit Catalog Number: RUO: For Research Use Only. Not for use in diagnostic procedures.

MagniSort Mouse B cell Enrichment Kit Catalog Number: RUO: For Research Use Only. Not for use in diagnostic procedures. Page 1 of 2 MagniSort Mouse B cell Enrichment Kit RUO: For Research Use Only. Not for use in diagnostic procedures. Mouse splenocytes were unsorted (left) or sorted with the MagniSort Mouse B cell Enrichment

More information

PrepSEQ Residual DNA Sample Preparation Kit

PrepSEQ Residual DNA Sample Preparation Kit QUICK REFERENCE PrepSEQ Residual DNA Sample Preparation Kit Catalog Numbers 4413686 (kit), 4415414 (kit with protocol and quick reference card), 4413713 (combo kit ), and 4415413 (combo kit with protocol

More information

RapidPort EZ Access Port Kit DIRECTIONS FOR USE (DFU)

RapidPort EZ Access Port Kit DIRECTIONS FOR USE (DFU) RapidPort EZ Access Port Kit DIRECTIONS FOR USE (DFU) RapidPort EZ Access Port Kit Ref. No. C-2304 Standard RapidPort EZ Access Port Kit Ref. No. C-2306 Large RapidPort EZ Access Port Kit INTRODUCTION

More information

HighPrep Blood & Tissue DNA Kit

HighPrep Blood & Tissue DNA Kit MAGBIO ACCELERATING genomic research HighPrep Blood & Tissue DNA Kit Manual Revision v1.1 Catalog Nos. HPBTS-D16, HPBTS-D96, HPBTS-D96X4 Genomic DNA isolation from 20-250 μl of blood, lysate of tissues,

More information

INSTALLATION MANUAL MOSER STAINLESS STEEL KITCHEN FAUCET MODEL MANCHESTER

INSTALLATION MANUAL MOSER STAINLESS STEEL KITCHEN FAUCET MODEL MANCHESTER INSTALLATION MANUAL MOSER STAINLESS STEEL KITCHEN FAUCET MODEL MANCHESTER Thank you for your purchase of a MoSER faucet. With proper care and maintenance, your new faucet will bring you years of trouble

More information

QIAGEN Supplementary Protocol

QIAGEN Supplementary Protocol QIAGEN Supplementary Protocol PCR and data analysis using the cador T. equigenitalis PCR Kit on Rotor-Gene Q instruments This protocol is designed for using the cador T. equigenitalis PCR Kit for detection

More information

PowerPlex 5-Dye Matrix Standards, 310

PowerPlex 5-Dye Matrix Standards, 310 TECHNICAL BULLETIN PowerPlex 5-Dye Matrix Standards, 310 Instructions for Use of Product DG4600 Revised 3/16 TBD023 PowerPlex 5-Dye Matrix Standards, 310 All technical literature is available at: www.promega.com/protocols/

More information

edna PROTOCOL SAMPLE COLLECTION Caren Goldberg and Katherine Strickler, Washington State University Revised January 2017

edna PROTOCOL SAMPLE COLLECTION Caren Goldberg and Katherine Strickler, Washington State University Revised January 2017 edna PROTOCOL SAMPLE COLLECTION Caren Goldberg and Katherine Strickler, Washington State University Revised January 2017 MATERIALS 1. Cellulose nitrate disposable filter funnels or other field-tested,

More information