A novel calibration strategy for residual solvents analysis by static headspace gas chromatography

Size: px
Start display at page:

Download "A novel calibration strategy for residual solvents analysis by static headspace gas chromatography"

Transcription

1 A novel calibration strategy for residual solvents analysis by static headspace gas chromatography Keith Freebairn Strategic Technologies Stevenage, UK 1

2 Introduction We have substantial experience over recent years of using a calibration strategy that involves the use of historically-derived stored response factors rather than the labour-intensive reference to calibration standards This strategy offers improved performance provided that it is adopted as part of a very rigorous Quality by Design approach 2

3 Old ways. Samples analysed by project teams using a range of hand-crafted solvent methods Standards prepared on the day Analysts - Wide range of skills and knowledge Relatively poor understanding of methodology - and sources of variability 3

4 Old ways the result Poor understanding of method performance and variability Poor method robustness Higher costs 4

5 New ways A standardised approach Fixed method Fixed infrastructure instrument, equipment ways of working etc. Reduced variability No standard preparation System suitability assessment Automation Instrument assessment of quality 5

6 A cultural shift.. A cultural shift is currently taking place in the Pharma industry This concerns the adoption of Quality by Design principles For analytical purposes this is all about understanding the performance and variability associated with analytical methods 6

7 QbD and analysis - Standardisation of approach and platform technology Standardisation is a key enabler in a QbD world 6 years developing analytical systems that provide standardised solutions for our key analytical tasks GSK has developed a QbD approach to the lifecycle of analytical methods* *Application of the principles of Quality By Design to Analytical Methods..Phil Borman, PhRMA API Workshop, Colorado, 26 th Sept

8 Standardisation and platform technology - analysis Why? Enhanced quality of output Greater efficiency Ease of use Lower method diversity Greater opportunities to exert control Enables continuous improvement Reduced method variability 8

9 A solvents analyser.mk1 Based on Agilent 6890 GC CTC combipal headspace analyser DB-624 column Samples dissolved in DMSO 11 min chromatographic run 9

10 A solvents analyzer.mk1 A conventional approach but automated Sample headspace taken by heated syringe directly into the chromatograph Instrument calibrated at the start of each day using five injections of a mixed solvent standard (containing 22 solvents). A further standard is run after every 8 sample injections a rolling response factor is calculated throughout the day and applied to the samples under analysis 10

11 A solvents analyzer.mk1 Launched in May 2002 Dedicated instrument always ready walk-up format Standardised method Custom user interface 22 common process solvents On-board system suitability and quality checks 11

12 12

13 Mk1 analyser The same excellent separation every day of the the year always available high performance - guaranteed Minutes pa Methanol Ethanol Acetone Propan2ol Acetonitrile Dichloromethane MEK Butan1ol Dioxan DMF Ethylacetate MIBK Toluene nhexane Tetrahydrofuran Isopropylacetate Pyridine Diethylether TBME Isopropylether Cyclohexane Heptane DMSO

14 Mk1 analyser - metrics Analysed ca 10,000 samples on 4 instruments over 2.5 years Total number of mixed standards :- 12,000 23,000 injections in total 14

15 Standardisation - Automated integration Solvents analyser integrated over 7000 chromatograms between July 02 and July 03 Not one was re-integrated - illustrates the power of standardisation and platform technology

16 Box Plot displaying variability of Retention Times Acetone Acetonitrile ichloromethane Diethyl ether Ethanol Ethyl Acetate Hexane Iso-propyl ether MEK Methanol Propan-2-ol TBME Retention Time GC solvents analyser Retention time reliability Variation typically ±0.05min for Acetone over 2.5 years, 3 sites

17 Mk1 solvents analyser Typical trend data for acetone response over 2.5 years 4 instruments Acetone Response / 07/ / 01/ / 07/ / 01/ / 07/ / 01/ 2005 Date 17

18 Acetone area response Variation about ±16% St Dev St Dev Acetone Peak area trend plot over a 2.5 year period for standards single site data with ±3Standard Deviation limits superimposed 18 11/07/ /01/ /07/ /01/ /07/ /01/2005 Date

19 Measurements systems analysis (MSA) Set up designed experiment to look at variability associated with the following aspects Injector Precision Concentrated standard preparation Working standard preparation Analyst Site The study concluded that the main source of error is concerned with preparing the standards 19

20 Acetone Peak Area Data from MSA Graphical representation of variability associated with preparation of standards Significant difference in instrument response at some sites cprep preparer (code) analyst Site (code) Raw Data 20 Substantial variability in the preparation of standards at two sites

21 MSA study conclusions Historical Acetone area response data (one site) with estimated variability of standard preparations from MSA Study The majority of the variability seen in the historical response dataset is accounted for by the preparation of standards 21

22 A counter-intuitive conclusion Analyser performance could be significantly improved if an average response factor is used rather than a response factor obtained on a given day 22

23 a Mk 2 analyser Use instrument-specific response factors generated over a period of 2 years to give an average response factor for that instrument Calibrate electronically using the historically generated mean response factor Check response using reference mixture containing only three solvent components that are representative of the total (system suitability) 23

24 Mk 2 solvents analyser launched September 2005 System suitability test for three component test mixture (PCS) Check areas (+/-20%) also indirect check on retention behaviour Check peak tail <1.5 Check DMSO >800 checks injection has been made this indirectly checks the ramp is correct Check blank If there are any issues with system suitability instrument will stop 24

25 Design requirements Fundamental requirements for calibration:- Consistently identify target peak Consistently integrate standards and sample peaks without interference Consistently establish target response factor Sample response to be within calibrated linear range 25

26 Typical PCS Test Solutions 12 overlays, one site, Jan 06 to Nov DMSO Response pa Acetone Tetrahydrofuran Pyridine Minutes Retention Time Time

27 Typical Blank Solutions 12 overlays, Stevenage, Jan 06 to Nov DMSO Response pa Dimethyl sulphide Pyridine Minutes Retention Time Time

28 Acetone Total number area of response std points = 4600 over for 4 yrs single 4 mths instrument over 4½ years Peak Area Nov-01 Mar-03 Aug-04 Dec-05 Apr-07 Date Acetone Area WUS Acetone Area EOVI Upper limit (+20%) Lower limit (-20%) Expected area Total number of Std points = 4600 over ~4½ years

29 Experimental uncertainty associated with Mk2 analyser factors contributing to overall variance Estimated variance Note: Time effect/s not studied Injector variability -Dispensing Standard -Crimping of Vial -Working Standard Preparer -Working Standard Preparations -Site -Instrument -Gas Supply & Flow Small amount of this variability will still be associated with preparing samples ~1% Eliminated by using historic data to set response factors Eliminated by using Sitespecific response factors 20 0 Acetone Peak Area -Stock Std Solution Preparer Eliminated by using historical data to set response factors 29

30 Benefits of using a Stored Response Factor ±16% ±15% Acetone Peak Area Response Historical Data ±3 St Dev Variability associated with preparing standards from MSA ±3 St Dev ±3% Variability associated with error * from MSA ± 3 St Dev *Injector variability Dispensing Standard Crimping of Vial 105 Historical Data + Variability MSA + Standards Variability MSA + error Variability 30

31 Accuracy reference to external benchmarks Results for beverage analysis by Mk2 solvents analyser plotted against label ethanol content 16 Label Ethanol Content %v/v English beer #1 English beer #2 Dutch lager Californian white wine ±0.1%v/v (±2%) Ethanol content by Mk2 solvents analyser %v/v 31

32 Summary Stored response factors offer improved method performance A platform approach offers excellent stability, reliability and opportunities to both understand and control method variability 32

33 Acknowledgements - Phil Borman - Colin Donaghy - Marion Chatfield

Chem-Lab High Purity Solvents

Chem-Lab High Purity Solvents Chem-Lab High Purity Solvents Chem-lab is pleased to offer a large selection of High Purity Solvents. Our High Purity Solvents are filtered through 0.2 µm filters and bottled under nitrogen. Our screw

More information

Quantitative determination of residual 2-(2-chloroethoxy) ethanol (CEE) in quetiapine fumarate by gas chromatogaraphy

Quantitative determination of residual 2-(2-chloroethoxy) ethanol (CEE) in quetiapine fumarate by gas chromatogaraphy Advances in Bioscience and Biotechnology, 2010, 1, 367-371 doi:10.4236/abb.2010.15049 Published Online December 2010 (http://www.scirp.org/journal/abb/). Quantitative determination of residual 2-(2-chloroethoxy)

More information

Analysis of USP <467> Residual Solvents using the Agilent 7697A Headspace Sampler with the Agilent 7890B Gas Chromatograph

Analysis of USP <467> Residual Solvents using the Agilent 7697A Headspace Sampler with the Agilent 7890B Gas Chromatograph Analysis of USP Residual Solvents using the Agilent 797A Headspace Sampler with the Agilent 79B Gas Chromatograph Application Note Pharmaceuticals Authors Bart Tienpont, Frank David, and Pat Sandra

More information

ZLD solution for Pharmaceutical Industry

ZLD solution for Pharmaceutical Industry ZLD solution for Pharmaceutical Industry Hemant Wagh Associate Vice President Water & WW solutions Group Major Concerns... Where is the space for process plant??? Can I generate/save energy??? Can I use

More information

Fusion Analytical Method Validation

Fusion Analytical Method Validation Fusion QbD Software System Fusion Analytical Method Validation The Only Software That Has It All! 100% aligned with FDA/ICH Quality by Design (QbD) guidances! Can be used for LC and Non-LC methods (e.g.

More information

Residual Solvents in medical cannabis analysis using the SRI 8610C FID GC June 2014

Residual Solvents in medical cannabis analysis using the SRI 8610C FID GC June 2014 June 2014 The SRI Medical Herb Potency 8610C GC is shown at right. This GC can also be used to test for residual solvents (i.e. butane, acetone, gasoline residue, etc.) in medical cannabis. These solvents

More information

PPLICATION NOTE DETERMINATION OF PAH

PPLICATION NOTE DETERMINATION OF PAH DETERMIN OF PAH via GPC with On-line EVAporation and HPLC-FLD Meas urement APPLIKS www.lctech.dee As of: Mai 2015, Version: 1.11 Determination of PAH via GPC with on-line EVAporation and HPLC-FLD-measurement

More information

Fast Detailed Hydrocarbon Analysis by Modified ASTM Method D6730

Fast Detailed Hydrocarbon Analysis by Modified ASTM Method D6730 application Note Gas Chromatography Author Terry Osenbach PerkinElmer, Inc. Shelton, CT 06484 USA Fast Detailed Hydrocarbon Analysis by Modified ASTM Method D6730 Introduction Detailed hydrocarbon analysis

More information

Quality-by-Design-Based Method Development Using an Agilent 1290 Infinity II LC

Quality-by-Design-Based Method Development Using an Agilent 1290 Infinity II LC Quality-by-Design-Based Method Development Using an Agilent 129 Infinity II LC An Efficient Method Development Workflow Combined with ISET-mediated Method Transfer Under Waters Empower 3 CDS Control Application

More information

Formulas: See Table 1 MW: Table 1 CAS: Table 1 RTECS: Table 1. METHOD: 2555, Issue 1 EVALUATION: PARTIAL Issue 1: 15 March 2003

Formulas: See Table 1 MW: Table 1 CAS: Table 1 RTECS: Table 1. METHOD: 2555, Issue 1 EVALUATION: PARTIAL Issue 1: 15 March 2003 KETONES I 2555 Formulas: See Table 1 MW: Table 1 CAS: Table 1 RTECS: Table 1 METHOD: 2555, Issue 1 EVALUATION: PARTIAL Issue 1: 15 March 2003 OSHA : See Table 2 NIOSH: See Table 2 ACGIH: See Table 2 PROPERTIES:

More information

Appendix F: Hazardous Gases and Vapors

Appendix F: Hazardous Gases and Vapors Appendix F: Hazardous Gases and Vapors The following pages contain excerpts from the National Fire Protection Association (NFPA) publication NFPA 497M Classification of Gases, Vapors, and Dusts for Electrical

More information

Determination of organochlorine pesticide residues in eggs by gel-permeation chromatography (GPC) cleanup and GC-ECD

Determination of organochlorine pesticide residues in eggs by gel-permeation chromatography (GPC) cleanup and GC-ECD Determination of organochlorine pesticide residues in eggs by gel-permeation chromatography (GPC) cleanup and GC-ECD 1. Experimental Section LabTech, Inc. 1.1 Instruments and reagents AutoClean automatic

More information

Protocol for Quantitative Determination of Residual Solvents in Cannabis Concentrates Prepared by: Amanda Rigdon, May 23 rd, 2016

Protocol for Quantitative Determination of Residual Solvents in Cannabis Concentrates Prepared by: Amanda Rigdon, May 23 rd, 2016 Protocol for Quantitative Determination of Residual Solvents in Cannabis Concentrates Prepared by: Amanda Rigdon, May 23 rd, 2016 1.0 Method background: This method employs full evaporation technique headspace

More information

HYDROCARBONS, BP C 1500

HYDROCARBONS, BP C 1500 HYDROCARBONS, BP 36 216 C 1500 FORMULA: Table 1 MW: Table 1 CAS: Table 1 RTECS: Table 1 METHOD: 1500, Issue 3 EVALUATION: PARTIAL Issue 1: 15 August 1990 Issue 3: 15 March 2003 OSHA : Table 2 NIOSH: Table

More information

SOLVENT PROPERTIES Typical Physical Properties of More than 120 Commonly Used Organic Solvents

SOLVENT PROPERTIES Typical Physical Properties of More than 120 Commonly Used Organic Solvents SOLVENT PROPERTIES Typical Physical Properties of More than 20 Commonly Used Organic Solvents TECH DOC Nexeo Solutions is a leading distributor of solvents manufactured by the top suppliers across North

More information

PAL SPME Fibers Optimized for Automation

PAL SPME Fibers Optimized for Automation PAL SPME Fibers Optimized for Automation PAL SPME Fiber Since its introduction by Pawliszyn et al. (ref. 1) Solid Phase Micro Extraction (SPME) has seen a tremendous development. SPME is a very effective

More information

QUANLYNX APPLICATION MANAGER

QUANLYNX APPLICATION MANAGER QUANLYNX APPLICATION MANAGER OVERVIEW INTRODUCTION QuanLynx, an Application Manager included with Waters MassLynx Software, is designed for quantitative analysis. Simply acquiring mass spectrometry data

More information

Volatile Organic Compounds in Solids PBM

Volatile Organic Compounds in Solids PBM Organics Revision Date: Aug 15, 2014 Volatile Organic Compounds in Solids PBM Parameter Analytical Method Introduction Method Summary Parameter Applicability, MDLs Volatile Organic Compounds (VOCs) in

More information

CHROMATOGRAPHY/MASS SPECTROMETRY

CHROMATOGRAPHY/MASS SPECTROMETRY DETERMINATION OF 1,4-DIOXANE IN DRINKING WATER BY SOLID PHASE EXTRACTION (SPE) AND GAS CHROMATOGRAPHY/MASS SPECTROMETRY (GC/MS) WITH SELECTED ION MONITORING (SIM)* EPA Method 522 Part Number: EU52112M6

More information

Peptide Mapping: A Quality by Design (QbD) Approach

Peptide Mapping: A Quality by Design (QbD) Approach Peptide Mapping: A Quality by Design (QbD) Approach Application Note Bio-Pharmaceutical Authors Sreelakshmy Menon and Suresh babu C.V. Agilent Technologies, Inc. Richard Verseput S-Matrix Corporation Abstract

More information

Headspace Samplers. HS-20 Series C146-E209D

Headspace Samplers. HS-20 Series C146-E209D Headspace Samplers HS-20 Series C146-E209D HS-20 Series of Headspace Samplers A Revolutionary System Aimed at Performance and Ease of Use The HS-20 Series is the optimal solution for volatile component

More information

Agilent 7820A GC System THE EASY CHOICE FOR ROUTINE ANALYSES

Agilent 7820A GC System THE EASY CHOICE FOR ROUTINE ANALYSES Agilent 7820A GC System THE EASY CHOICE FOR ROUTINE ANALYSES THE AGILENT 7820A GC SYSTEM FORWARD-THINKING INNOVATION, EVERYDAY RELIABILITY AND VALUE Is your lab looking for better, more reproducible results

More information

APPLICATION NOTE. UOP : Analysis of Trace CO and CO 2 in bulk H 2 and Light Gaseous Hydrocarbons by GC

APPLICATION NOTE. UOP : Analysis of Trace CO and CO 2 in bulk H 2 and Light Gaseous Hydrocarbons by GC UOP 63-13: Analysis of Trace CO and CO 2 in bulk H 2 and Light Gaseous Hydrocarbons by GC Fast Analysis in

More information

Size Exclusion Chromatography of Biosimilar and Innovator Insulin Using the Agilent AdvanceBio SEC column

Size Exclusion Chromatography of Biosimilar and Innovator Insulin Using the Agilent AdvanceBio SEC column Size Exclusion Chromatography of Biosimilar and Innovator Insulin Using the Agilent AdvanceBio SEC column Application Note Bio-Pharmaceutical Authors M. Sundaram Palaniswamy and Andrew Coffey Agilent Technologies,

More information

Low-level Detection of Ethanol, 1,4-Dioxane, and Other Oxygenates Using the Eclipse Purge-and-Trap Sample Concentrator

Low-level Detection of Ethanol, 1,4-Dioxane, and Other Oxygenates Using the Eclipse Purge-and-Trap Sample Concentrator Application Note 25370206 Low-level Detection of,, and Other Oxygenates Using the Eclipse Purge-and-Trap Sample Concentrator Keywords Eclipse Fuel Oxygenates Methanol Model 4552 Purge and Trap Sample Concentrator

More information

INTERNATIONAL PHARMACOPOEIA MONOGRAPH ON ARTEMETHER AND LUMEFANTRINE CAPSULES REVISED DRAFT FOR DISCUSSION

INTERNATIONAL PHARMACOPOEIA MONOGRAPH ON ARTEMETHER AND LUMEFANTRINE CAPSULES REVISED DRAFT FOR DISCUSSION October 2007 RESTRICTED ` INTERNATIONAL PHARMACOPOEIA MONOGRAPH ON ARTEMETHER AND LUMEFANTRINE CAPSULES REVISED DRAFT FOR DISCUSSION World Health Organization 2007 All rights reserved. This draft is intended

More information

Qualification of High-Performance Liquid Chromatography Systems

Qualification of High-Performance Liquid Chromatography Systems Qualification of High-Performance Liquid Chromatography Systems L. Huber Hewlett- Packard GmbH, Waldbronn December 1998 Manuscript of an article published in BioPharm, Vol 11 Number 11, November 1998,

More information

Certified Reference Material - Certificate of Analysis

Certified Reference Material - Certificate of Analysis Certified Reference Material - Certificate of Analysis S(-)-Nicotine, Primary Standard N-008 Page 1 of 7 Catalog Number: N-008 H Lot: Expiration: October 2019 N Description: S(-)-Nicotine in Methanol.

More information

Determining the Exempt Solvent Content of Coatings Using Gas Chromatography

Determining the Exempt Solvent Content of Coatings Using Gas Chromatography Determining the Exempt Solvent Content of Coatings Using Gas Chromatography by Philip Patterson and William Lum ARL-TN-194 December 2002 Approved for public release, distribution is unlimited. NOTICES

More information

Chem 321 Lecture 23 - Liquid Chromatography 11/19/13

Chem 321 Lecture 23 - Liquid Chromatography 11/19/13 Chem 321 Lecture 23 - Liquid Chromatography 11/19/13 Student Learning Objectives High Performance Liquid Chromatography With the advent of relatively inexpensive and reliable pumps, the development of

More information

Agilent 8355 Sulfur Chemiluminescence Detector REIMAGINED. REDESIGNED.

Agilent 8355 Sulfur Chemiluminescence Detector REIMAGINED. REDESIGNED. Agilent 8355 Sulfur Chemiluminescence Detector REIMAGINED. REDESIGNED. AGILENT 8355 SULFUR CHEMILUMINESCENCE DETECTOR (SCD) THE ONLY INTEGRATED SYSTEM FOR LOW-LEVEL SULFUR ANALYSIS Why did Agilent reimagine

More information

Gas Chromatography Assignment Chem 543/443

Gas Chromatography Assignment Chem 543/443 Gas Chromatography Assignment Chem 543/443 1. Introduction Capillary gas chromatography (GC) is one of the most popular analytical techniques used in today s research. Its popularity is mainly due to efficient

More information

Developing Robust and Efficient IEX Methods for Charge Variant Analysis of Biotherapeutics Using ACQUITY UPLC H-Class System and Auto Blend Plus

Developing Robust and Efficient IEX Methods for Charge Variant Analysis of Biotherapeutics Using ACQUITY UPLC H-Class System and Auto Blend Plus Developing Robust and Efficient IEX Methods for Charge Variant Analysis of Biotherapeutics Using ACQUITY UPLC H-Class System and Auto Blend Plus Robert Birdsall, Thomas Wheat, and Weibin Chen Waters Corporation,

More information

ISSN India; g,secunderabad. Abstractt. a flow rate. of 1ml/min. di hydrogen. which acts. and chronic. including minimize (5) Figure

ISSN India; g,secunderabad. Abstractt. a flow rate. of 1ml/min. di hydrogen. which acts. and chronic. including minimize (5) Figure B.Lakshmi et al SPJTS,2013,1(1),015-023 RP-HPLC METHOD FOR THE QUANTIFICATION OF ROFLUMILAST IN FORMULATIONS ISSN 2321-4597 B..Lakshmi 1, Prof. T.V.Reddy 2 1.Kallam Haranadha Reddy Institute of Technology,NH-5,

More information

STANDARD OPERATING PROCEDURES

STANDARD OPERATING PROCEDURES PAGE: 1 of 5 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING, AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information

SCION GC Series. The Gas Chromatographers Choice for Separations

SCION GC Series. The Gas Chromatographers Choice for Separations SCION GC Series The Gas Chromatographers Choice for Separations Innovation in Gas Chromatography Scion Instrument s tradition of innovation and product reliability have combined to create the next generation

More information

Research Paper Development of Stability Indicating Reverse Phase HPLC Method for Aripiprazole from Solid Dosage form

Research Paper Development of Stability Indicating Reverse Phase HPLC Method for Aripiprazole from Solid Dosage form International Journal of Pharmaceutical Sciences and Nanotechnology 572 International Journal of Pharmaceutical Sciences and Nanotechnology Volume 2 Issue 2 July - September 2009 Volume 2 Issue 2 July

More information

Separation and Quantitation of a Mixture of Alcohols by Gas Chromatography

Separation and Quantitation of a Mixture of Alcohols by Gas Chromatography Separation and Quantitation of a Mixture of Alcohols by Gas Chromatography 1. Purpose This procedure will determine the percentage of two alcohols in an unknown aqueous mixture. 2. Background Gas chromatography

More information

Metal-Organic Frameworks for Thin-Layer Chromatographic Applications

Metal-Organic Frameworks for Thin-Layer Chromatographic Applications Supporting Information for: Metal-Organic Frameworks for Thin-Layer Chromatographic Applications Claudia Schenk, Christel Kutzscher, Franziska Drache, Stella Helten, Irena Senkovska and Stefan Kaskel*

More information

Fast Preparative Column Liquid Chromatography (PCLC)

Fast Preparative Column Liquid Chromatography (PCLC) Fast Preparative Column Liquid Chromatography (PCLC) Application Note 224 Joan Stevens, Ph.D. and Gary Scharrer Introduction Preparative HPLC is recognized as a prime method for obtaining pure compounds

More information

TurboMass GC/MS Software

TurboMass GC/MS Software TurboMass GC/MS Software GAS CHROMATOGRAPHY / MASS SPECTROMETRY P R O D U C T N O T E Today s demanding laboratory requires software that is easy to learn yet offers sophisticated instrument control, robust

More information

Optimization of Volatile Petroleum Hydrocarbons Analysis by Automated Headspace Using Method Development Tools

Optimization of Volatile Petroleum Hydrocarbons Analysis by Automated Headspace Using Method Development Tools Optimization of Volatile Petroleum Hydrocarbons Analysis by Automated Headspace Using Method Development Tools Application Note Abstract The Commonwealth of Massachusetts Department of Environmental Protection

More information

Technical Design by AMCEC, USA. Solvent Recovery Solutions For Profitability with Compliance

Technical Design by AMCEC, USA. Solvent Recovery Solutions For Profitability with Compliance Technical Design by AMCEC, USA Solvent Recovery Solutions For Profitability with Compliance A Winning Combination Two well known names AMCEC, USA and Dalal Engineering, India have come together to offer

More information

Chrom-Card Data System. A simple, chemist-oriented data handling system...productive from the first day!

Chrom-Card Data System. A simple, chemist-oriented data handling system...productive from the first day! c h r o m a t o g r a p h y Chrom-Card Data System A simple, chemist-oriented data handling system......productive from the first day! Analyze Detect Measure Control Chrom-Card Data System A simple, chemist-oriented

More information

technical information

technical information CHEMICAL RESISTANCE TABLES Version: 19-08-10 ADD Resins and Chemicals (Pty) Ltd Distribute a comprehensive range of unsaturated polyester resins for chemical resistance applications. The resins recommended

More information

Results of Proficiency Test Fuel/Bio-ethanol December 2015

Results of Proficiency Test Fuel/Bio-ethanol December 2015 Results of Proficiency Test Fuel/Bio-ethanol December 215 Organised by: Authors: Correctors: Report: Institute for Interlaboratory Studies (iis) Spijkenisse, the Netherlands ing. R.J. Starink dr. R.G.

More information

I n d u P r o AB. Measuring range (ppm) Detector tube name. Gas or vapour to be measured. Tube Code No.

I n d u P r o AB. Measuring range (ppm) Detector tube name. Gas or vapour to be measured. Tube Code No. Detector tube name Tube Code No. Gas or vapour to be measured Measuring range (ppm) No. Of Pump Strokes Shelf Life (year) Note TLV TWA, C (ACGIH) (ppm) Acetaldehyde SAT 4M Acetaldehyde 0 300 C5 Acetic

More information

Certificate of Accreditation

Certificate of Accreditation PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of: (Hereinafter called the Organization) and hereby declares

More information

Total Petroleum Hydrocarbons By Gas Chromatography in Colombian Waters and Soils

Total Petroleum Hydrocarbons By Gas Chromatography in Colombian Waters and Soils American Journal of Environmental Science, 2012, 8 (4), 396-402 ISSN: 1553-345X 2012 Science Publication doi:10.3844/ajessp.2012.396.402 Published Online 8 (4) 2012 (http://www.thescipub.com/ajes.toc)

More information

On-line SPE-LC/MS/MS to Detect Organonitrogen and Triazine Pesticides at 10ng/L in Drinking Water

On-line SPE-LC/MS/MS to Detect Organonitrogen and Triazine Pesticides at 10ng/L in Drinking Water application note On-line SPE-LC/MS/MS of Pesticides On-line SPE-LC/MS/MS to Detect Organonitrogen and Triazine Pesticides at 10ng/L in Drinking Water API 3200 LC/MS/MS System Overview The feasibility of

More information

Agilent 7693A Series Automatic Liquid Sampler. Inject new performance into your gas chromotography

Agilent 7693A Series Automatic Liquid Sampler. Inject new performance into your gas chromotography Agilent 7693A Series Automatic Liquid Sampler Inject new performance into your gas chromotography Agilent 7693A Series Automatic Liquid Sampler A powerful combination of productivity and flexibility How

More information

Available Online through (or) IJPBS Volume 2 Issue 3 JULY-SEPT Research Article Pharmaceutical Sciences

Available Online through  (or)  IJPBS Volume 2 Issue 3 JULY-SEPT Research Article Pharmaceutical Sciences Page1 Research Article Pharmaceutical Sciences RP-HPLC DETERMINATION OF RELATED SUBSTANCES OF TAPENTADOL IN BULK AND PHARMACEUTICAL DOSAGE FORM EDIGA SASI KIRAN GOUD* 1, V.KRISHNA REDDY 2 * 1,2 Department

More information

Method Translation in Liquid Chromatography

Method Translation in Liquid Chromatography Method Translation in Liquid Chromatography Technical Overview Abstract Ronald E. Majors Agilent Technologies, Inc. 2850 Centerville Rd Wilmington, DE 19808 USA With the recent emphasis on high performance

More information

Agilent 7693A Series Automatic Liquid Sampler INJECT NEW PERFORMANCE INTO YOUR GAS CHROMATOGRAPHY

Agilent 7693A Series Automatic Liquid Sampler INJECT NEW PERFORMANCE INTO YOUR GAS CHROMATOGRAPHY Agilent 7693A Series Automatic Liquid Sampler INJECT NEW PERFORMANCE INTO YOUR GAS CHROMATOGRAPHY AGILENT 7693A SERIES AUTOMATIC LIQUID SAMPLER A powerful combination of productivity and flexibility Autosampler

More information

AppNote 2/2011 KEYWORDS ABSTRACT

AppNote 2/2011 KEYWORDS ABSTRACT AppNote 2/2011 Quantifi cation of Ethanol in Complex Oil Samples: A Comparison of Different Headspace Methods and an Automated Direct Injection Procedure Susanne Sperling, Eike Kleine-Benne Gerstel GmbH

More information

easy. HPLC has never been so Agilent 1120 Compact LC Our measure is your success.

easy. HPLC has never been so Agilent 1120 Compact LC Our measure is your success. easy. HPLC has never been so Agilent 1120 Compact LC Our measure is your success. Introducing the Agilent 1120 Compact LC Simply a better value. Is your lab looking for better, more reproducible results

More information

SCOTT Method TO-14A/15/17

SCOTT Method TO-14A/15/17 SCOTT Method TO-14A/15/17 Calibration Standards Benefits and Features SCOTT technical expertise in engineering Method 14A/15/17, HAPS and TIC calibration standards, sulfur mixes and vapor intrusion products

More information

SCION GC Series. The Gas Chromatographers Choice for Separations

SCION GC Series. The Gas Chromatographers Choice for Separations SCION GC Series The Gas Chromatographers Choice for Separations Innovation in Gas Chromatography Bruker s Scion Instrument s long tradition long of tradition innovation of innovation and product and reliability

More information

PC0 2 Carbon Dioxide Polishers for the Beverage Industry.

PC0 2 Carbon Dioxide Polishers for the Beverage Industry. PC0 2 Carbon Dioxide Polishers for the Beverage Industry www.domnickhunter.com 1 PC0 2 Carbon Dioxide Polishers for the Beverage Industry Why choose a domnick hunter PCO 2 system? The PCO 2 system is designed

More information

Agilent Ultimate Plus Fused Silica Tubing

Agilent Ultimate Plus Fused Silica Tubing Agilent Ultimate Plus Fused Silica Tubing Technical Overview Introduction Deactivated fused silica tubing is widely used for guard columns, transfer lines, or long retention gaps in GC and GC/MS applications

More information

Agilent 7693A Series Automatic Liquid Sampler INJECT NEW PERFORMANCE INTO YOUR GAS CHROMATOGRAPHY

Agilent 7693A Series Automatic Liquid Sampler INJECT NEW PERFORMANCE INTO YOUR GAS CHROMATOGRAPHY Agilent 7693A Series Automatic Liquid Sampler INJECT NEW PERFORMANCE INTO YOUR GAS CHROMATOGRAPHY AGILENT 7693A SERIES AUTOMATIC LIQUID SAMPLER A powerful combination of productivity and flexibility It

More information

Oxygen Sensor Probes. Options for a Wide Range of Sensing Environments. Sensors

Oxygen Sensor Probes. Options for a Wide Range of Sensing Environments. Sensors Oxygen Sensor Probes Oxygen sensor probes typically are available in a variety of designs and with each of our standard coating formulations (FOXY, FOSPOR and HIOXY). Custom probes and accessories are

More information

EPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube

EPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube EPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube Application Note Environmental Authors Don Noot, Ralph Hindle Vogon Laboratory Services Cochrane, AB

More information

Certificate of Analysis

Certificate of Analysis Certificate of Analysis ISO GUIDE 34 AAB Cert# AR-1470 ISO/IEC 17025 AAB Cert# AT-1467 FLUCOAZOLE CERTIFIED REFERECE MATERIAL F OH F CERTIFIED PURITY: 99.6%, U crm = ±0.2% k = 2 (Mass Balance/as is basis)

More information

Protein-Pak Hi Res HIC Column and HIC Protein Standard

Protein-Pak Hi Res HIC Column and HIC Protein Standard Protein-Pak Hi Res HIC Column and HIC Protein Standard CONTENTS I. INTRODUCTION II. a. Mobile Phase b. Flow Direction CONNECTING COLUMN TO LC SYSTEM I. INTRODUCTION This offering contains non-porous, polymethacrylate-based

More information

Application Note. Abstract. Authors. Introduction. Environmental

Application Note. Abstract. Authors. Introduction. Environmental EPA Method 538: Determination of Selected Organic Contaminants in Drinking Water by Direct Aqueous Injection with the Agilent 646 Triple Quadrupole LC/MS System Application Note Environmental Authors Imma

More information

Low mg/m

Low mg/m 1 (7) CWA-9.2.3 Nerve Blister Library specified agents GA, GB, GD, GF VX HD L Low mg/m 3 0.1 0.1 2 2 Library unspecified agents/cases Nerve precursors Vesicant precursor - - s in hazardous concentration

More information

R R Innovation Way P/N SECKIT-7830 Newark, DE 19711, USA Tel: Fax: Website: Published in November 2013

R R Innovation Way P/N SECKIT-7830 Newark, DE 19711, USA Tel: Fax: Website:  Published in November 2013 5-100 Innovation Way Newark, DE 19711, USA Tel:302-3661101 Fax:302-3661151 Website: www.sepax-tech.com Published in November 2013 P/N SECKIT-7830 These Phases are developed based on innovative surface

More information

The Analysis of 4-Vinyl-Cyclo-Hexene (VCH) in Butadiene with the DVLS LGI Injector

The Analysis of 4-Vinyl-Cyclo-Hexene (VCH) in Butadiene with the DVLS LGI Injector The Analysis of 4-Vinyl-Cyclo-Hexene (VCH) in Butadiene with the DVLS LGI Injector Introduction In synthetic rubber manufacturing, 4-Vinylcyclohexene (VCH) can be found as a byproduct from the chemical

More information

The Importance of High Speed Data Acquisition in the Design and Implementation of GC/MS Applications

The Importance of High Speed Data Acquisition in the Design and Implementation of GC/MS Applications TECHNICAL BRIEF Gas Chromatography/ Mass Spectrometry The Importance of High Speed Data Acquisition in the Design and Implementation of GC/MS Applications C 14 scans Introduction As gas chromatography

More information

Controlling Agilent 1200 Series Rapid Resolution LC systems through Waters Empower chromatography data software. Technical Overview.

Controlling Agilent 1200 Series Rapid Resolution LC systems through Waters Empower chromatography data software. Technical Overview. Controlling Agilent 1200 Series Rapid Resolution LC systems through Waters Empower chromatography data software Technical Overview Abstract The Waters Empower 2 chromatography data software with ICS 1.05

More information

2. Crystallization. A. Background

2. Crystallization. A. Background 2. Crystallization A. Background Crystallization is one of several available techniques available to purify organic compounds. Unlike other techniques, however, crystallization is specific to the purification

More information

ANALYSIS OF PESTICIDE RESIDUES IN DRINKING WATER AS PER BUREAU OF INDIAN STANDARDS USING THE AGILENT 7000 GC/MS/MS WITH PESTICIDES ANALYZER

ANALYSIS OF PESTICIDE RESIDUES IN DRINKING WATER AS PER BUREAU OF INDIAN STANDARDS USING THE AGILENT 7000 GC/MS/MS WITH PESTICIDES ANALYZER ENVIRONMENTAL ANALYSIS ANALYSIS OF PESTICIDE RESIDUES IN DRINKING WATER AS PER BUREAU OF INDIAN STANDARDS USING THE AGILENT 7000 GC/MS/MS WITH PESTICIDES ANALYZER Solutions for Your Analytical Business

More information

International Journal of Pharma Research & Review, Feb 2014; 3(2):11-16 ISSN:

International Journal of Pharma Research & Review, Feb 2014; 3(2):11-16 ISSN: Research Article Determination of Methyl Methanesulfonate, Ethyl Methanesulfonate and Isopropyl Methanesulfonate Impurities in Lopinavir API by GC/MS/MS using Electron Ionization Veenaeesh P, *Manikumar

More information

Overview of preparative HPLC. Analytical Technologies Limited

Overview of preparative HPLC. Analytical Technologies Limited Overview of preparative HPLC Analytical Technologies Limited 1.Review of liquid phase chromatography origin develop prospect Mobile phase Tewett 1903 Stationa ry phase origin Separation result Liquid-stationary

More information

Results of Proficiency Test Fuel/Bio-ethanol November 2013

Results of Proficiency Test Fuel/Bio-ethanol November 2013 Results of Proficiency Test Fuel/Bio-ethanol November 213 Organised by: Authors: Correctors: Report: Institute for Interlaboratory Studies (iis) Spijkenisse, the Netherlands ing. L. Dijkstra dr. R.G. Visser

More information

Technical Overview. Author. Abstract. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Technical Overview. Author. Abstract. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany New Features of the Agilent Method Scouting Wizard for Automated Method Development of Complex Samples Analysis of Large Data Sets by Method Scouting Reports and Automated Adjustment of Flow Rates and

More information

AppNote 5/2006 ABSTRACT

AppNote 5/2006 ABSTRACT AppNote 5/26 Stir Bar Sorptive Extraction Based on Restricted Access Material for the Direct Extraction of Drugs and Metabolites in Biological Fluids Wayne M. Mullett and Elizabeth Kwong Merck Frosst,

More information

Quick Manual to the Waters UPLC System

Quick Manual to the Waters UPLC System 1 Quick Manual to the Waters UPLC System 1. Application of Chiral HPLC Columns Warning! The HPLC columns are destroyed if the pressure overrides 2000 psi. The highest pressure applied for routine analysis

More information

Analysis of Suspected Flavor and Fragrance Allergens in Cosmetics Using the 7890A GC and Capillary Column Backflush Application

Analysis of Suspected Flavor and Fragrance Allergens in Cosmetics Using the 7890A GC and Capillary Column Backflush Application Analysis of Suspected Flavor and Fragrance Allergens in Cosmetics Using the 7890A GC and Capillary Column Backflush Application Food Authors Frank David Research Institute for Chromatography Pres. Kennedypark

More information

HPLC METHODOLOGY MANUAL

HPLC METHODOLOGY MANUAL HPLC METHODOLOGY MANUAL DISTRIBUTED PHARMACEUTICAL ANALYSIS LABORATORY (DPAL) Revision Date 2016-08-04 Professor Marya Lieberman Department of Chemistry and Biochemistry University of Notre Dame Notre

More information

Aligning Process Redesign and Change Management with Project Management (System Implementation Projects)

Aligning Process Redesign and Change Management with Project Management (System Implementation Projects) Aligning Redesign and Change Management with Project Management (System Implementation Projects) Kim Brant-Lucich Director of Redesign St. Joseph Health System Information Services HIMSS 2009: Project

More information

Fundamentals and Techniques of Preparative HPLC. Parto Zist Behboud Tel: 42108

Fundamentals and Techniques of Preparative HPLC. Parto Zist Behboud  Tel: 42108 Fundamentals and Techniques of Preparative HPLC Parto Zist Behboud www.partoz.com Tel: 42108 Today s Overview 1 Basic know-how for up-scaling 2 Practical examples of up-scaling 3 Recycle preparation 4

More information

Supplemental material:

Supplemental material: Supplemental material: REGIOSPECIFIC SYNTHESIS OF DIMETHYLPHENANTHRENES Thomas Böhme, Marianne Lorentzen and Kåre B. Jørgensen* Faculty of Science and Technology, University of Stavanger, N-4036 Stavanger,

More information

Physical Stability of a Silica- Based Size Exclusion Column for Antibody Analysis

Physical Stability of a Silica- Based Size Exclusion Column for Antibody Analysis Physical Stability of a Silica- Based Size Exclusion Column for Antibody Analysis Atis Chakrabarti* and Roy Eksteen + Tosoh Bioscience LLC, King of Prussia, PA 19406 *Corresponding Author. + Current address:

More information

Elute LC series. Innovation with Integrity. UHPLC Precision for MS applications LC-MS

Elute LC series. Innovation with Integrity. UHPLC Precision for MS applications LC-MS Elute LC series UHPLC Precision for MS applications Innovation with Integrity LC-MS Delivering Precision and Speed for Mass Spectrometry Designed for reproducibility in retention time and peak shape, Elute

More information

VICH Topic GL2 (Validation: Methodology) GUIDELINE ON VALIDATION OF ANALYTICAL PROCEDURES: METHODOLOGY

VICH Topic GL2 (Validation: Methodology) GUIDELINE ON VALIDATION OF ANALYTICAL PROCEDURES: METHODOLOGY The European Agency for the Evaluation of Medicinal Products Veterinary Medicines Evaluation Unit CVMP/VICH/591/98-FINAL London, 10 December 1998 VICH Topic GL2 (Validation: Methodology) Step 7 Consensus

More information

N. Tamilselvi *, Dona Sara Kurian. Department of Pharmaceutical Analysis. KMCH college of pharmacy Coimbatore

N. Tamilselvi *, Dona Sara Kurian. Department of Pharmaceutical Analysis. KMCH college of pharmacy Coimbatore BIOANALYTICAL METHOD DEVELOPMENT AND VALIDATION OF PIRFENIDONE BY RPHPLC METHOD AND ITS APPLICATION TO THE DETERMINATION OF DRUG FOOD INTERACTION STUDY IN WISTER RATS ABSTRACT N. Tamilselvi *, Dona Sara

More information

EGB Relative Value Report. May152013

EGB Relative Value Report. May152013 EGB May152013 Jul-14 Jul-15 Sep-16 Sep-17 Mar-19 Jul-20 Sep-21 Apr-22 Nov-22 Oct-23 Mar-26 Jul-27 May-34 Mar-37 Jun-44 Jan-62 Austria, 15-May-13 45 bp 35 bp 25 bp 15 bp 5 bp -5 bp -15 bp -25 bp -35 bp

More information

Extreme Agile Implementation and Creating a Value Delivery Office

Extreme Agile Implementation and Creating a Value Delivery Office Extreme Agile Implementation and Creating a Value Delivery Office Anup Deshpande Sr. Leader of Agile Portfolio Management 1 COPYRIGHT 2017 Anup Deshpande. ALL RIGHTS RESERVED For Informational Purposes

More information

Monitoring for Microcystins in Raw Water Supply Reservoirs Using the Agilent 6410 Triple Quadrupole LC/MS

Monitoring for Microcystins in Raw Water Supply Reservoirs Using the Agilent 6410 Triple Quadrupole LC/MS Monitoring for Microcystins in Raw Water Supply Reservoirs Using the Agilent 641 Triple Quadrupole LC/MS Application Note Environmental Author Toni Hall Wessex Water Bath, UK Abstract A method for the

More information

VOC- AND SVOC-EMISSIONS FROM ADHESIVES, FLOOR COVERINGS AND COMPLETE FLOOR STRUCTURES

VOC- AND SVOC-EMISSIONS FROM ADHESIVES, FLOOR COVERINGS AND COMPLETE FLOOR STRUCTURES VOC- AND SVOC-EMISSIONS FROM ADHESIVES, FLOOR COVERINGS AND COMPLETE FLOOR STRUCTURES O Wilke *, O Jann and Brödner Federal Institute for Materials Research and Testing, 12200 Berlin, Germany ABSTRACT

More information

5 Star London Hotels - Example Report

5 Star London Hotels - Example Report 5 Star London Hotels - Example Report January 2018 CONTENTS Your Benchmark Report Website Traffic Conversion Rates Ecommerce Performance AdWords Spend Your Traffic Index Your Conversion Rate Index Your

More information

Adcock Ingram and Sub 2 Micron Column Technology

Adcock Ingram and Sub 2 Micron Column Technology Adcock Ingram and Sub 2 Micron Column Technology Heritage Quality Integrity Adcock Ingram unbundled from Tiger Brands Limited and independently listed on the JSE on 25 August 2008. Limited investment in

More information

Spectroscopy Solutions for Pharmaceuticals:

Spectroscopy Solutions for Pharmaceuticals: Spectroscopy Solutions for Pharmaceuticals: Confidence in Compliance to USP Using the Agilent Cary 60 UV-Vis Spectrophotometer White Paper Introduction UV-Vis spectroscopy is one of the most commonly

More information

Implementing USP Guidelines with Analytical Techniques: A Primer YOUR PRESCRIPTION FOR USP COMPLIANCE

Implementing USP Guidelines with Analytical Techniques: A Primer YOUR PRESCRIPTION FOR USP COMPLIANCE Implementing USP Guidelines with Analytical Techniques: A Primer YOUR PRESCRIPTION FOR USP COMPLIANCE CONTENTS An overview The United States Pharmacopeia (USP) 3 Implementation of USP and on

More information

Dynamic Reallocation of Portfolio Funds

Dynamic Reallocation of Portfolio Funds Complete Perspective. Smart Decisions. #StrategicPMO Dynamic Reallocation of Portfolio Funds Ben Chamberlain Chief Product & Marketing Officer Ben.Chamberlain@umt360.com Agenda What s wrong with traditional

More information

A Sub-picogram Quantification Method for Desmopressin in Plasma using the SCIEX Triple Quad 6500 System

A Sub-picogram Quantification Method for Desmopressin in Plasma using the SCIEX Triple Quad 6500 System A Sub-picogram Quantification Method for Desmopressin in Plasma using the SCIEX Triple Quad 6500 System A high-throughput method for detecting ultra-low levels (0.5 pg/ml) of a therapeutic peptide in human

More information

Electric Forward Market Report

Electric Forward Market Report Mar-01 Mar-02 Jun-02 Sep-02 Dec-02 Mar-03 Jun-03 Sep-03 Dec-03 Mar-04 Jun-04 Sep-04 Dec-04 Mar-05 May-05 Aug-05 Nov-05 Feb-06 Jun-06 Sep-06 Dec-06 Mar-07 Jun-07 Sep-07 Dec-07 Apr-08 Jun-08 Sep-08 Dec-08

More information

NATURALLY BETTER CHROMATOGRAPHY SOLUTIONS. Clarus Gas Chromatographs Clarus 680/580/480

NATURALLY BETTER CHROMATOGRAPHY SOLUTIONS. Clarus Gas Chromatographs Clarus 680/580/480 Clarus Gas Chromatographs Clarus 680/580/480 NATURALLY BETTER CHROMATOGRAPHY SOLUTIONS CLARUS DELIVERS POWER AND PERFORMANCE AT EVERY LEVEL The Clarus GC Family Performance you can rely on from the leader

More information