Monitoring THMs and TOC in High-Purity Process Water Systems

Size: px
Start display at page:

Download "Monitoring THMs and TOC in High-Purity Process Water Systems"

Transcription

1 World Leader in Total Organic Carbon Analysis Model A-1000/A-1000XP/Access 643 Total Organic Carbon Analysis Systems Monitoring THMs and TOC in High-Purity Process Water Systems Introduction Trihalomethanes (THMs) are potential byproducts of the drinking water disinfection process and are formed when naturally occurring organic materials found in water react with chlorine, the most common drinking water disinfectant. These byproducts have been implicated to be carcinogenic 1 and are therefore regulated by the EPA through the Disinfection Byproducts Rule. Current EPA regulations allow no more than 80 ppb of total THMs in drinking water supplies. The amount of THMs allowed in drinking water supplies is important because of their impact on the source water or feedwater for process water purification systems in the semiconductor and pharmaceutical industries. By law the pharmaceutical industry is required to prepare water for injection and purified water from water complying with US EPA National Primary Drinking Water Regulations. 2 The semi-conductor industry also uses water that complies with these same drinking water regulations. Controlling THM Formation in Drinking Water The concentration of THMs in drinking water is controlled by monitoring the reduction in TOC between raw water and treated water. The Disinfection Byproducts Rule specifies a percentage removal of TOC by enhanced coagulation. Removal efficiencies from 20-50% are required based on the value of TOC and alkalinity concentration of the raw untreated water. Monitoring TOC in Drinking Water and High- Purity Water Systems TOC is monitored in drinking water facilities as outlined in the D/DBP Rule. 3 A common method of analysis is Standard Method 5310C. 4 Semiconductor manufacturing facilities continuously monitor and control TOC levels in their high-purity water systems due to the detrimental effects that TOC has on product yields. Pharmaceutical manufacturers must perform TOC testing according to USP Method <643> 2 and EP Method Semiconductor and pharmaceutical facilities most commonly use conductometric based TOC measurement technologies for low level TOC analysis. In all cases, TOC measurements are used as a process control attribute to monitor the performance of unit operations comprising the water s purification and distribution system. An increase in TOC levels can indicate the possible presence of THMs in high-purity water systems. THM Detection Methods and Sensitivity Gas Chromatography (GC) is the most commonly used method for analyzing total THMs and individual compound levels in water samples. The sensitivity of traditional GC methods is approximately 0.5 ppb. A new, high-sensitivity method combining solid phase microextraction (SPME) and GC-MS has been developed with detection as low as 0.04 ppb. 6 This new technique has been developed for determining the THM contamination in high-purity water systems. Key Words: Disinfection Byproducts Rule Trihalomethanes (THMs) Total Organic Carbon High-Purity Water Oxidation Efficiency

2 Examples of Removal Efficiency of THMs by High-Purity Process Water Systems Table 1 shows the results of two water system designs tested for THM removal as measured by SPME/GCMS. System 1 is a small laboratory system. The DI make-up loop includes a 5 µm filter, followed by an activated carbon unit, RO unit, EDI, storage tank, a mixed bed DI column and a 0.2 µm final filter. The polishing system contains activated carbon, mixed bed DI and UV. System 2 is a medium sized ultrapure water system. The DI make-up loop consists of multimedia, prefilter, RO, mixed bed DI and UV. The DI polishing loop on this system includes UV, mixed bed DI and a final filter. Sampling Point System 1 (total THM ppb) System 2 (total THM ppb) Incoming city water DI Make-up 16 not detected DI Polish System 1 (no UV) DI Polish System 2 (with UV) DI Polish Table 1: SPME-GCMS determination of THMs in high-purity water systems The data show that System 1 had more than the maximum allowed level of THMs (80 ppb) for its incoming city water. In the DI make-up loop the THM concentration was reduced to 16 ppb and consisted totally of chloroform. 6 Final polishing with UV reduced the chloroform levels to 0.06 ppb. This result demonstrates that THMs can be easily oxidized by UV and that the use of UV in the purification process dramatically reduces THM levels. System 2 had much lower THMs levels in the incoming city water and THMs were not detected after treatment. However, a small amount of THMs, namely chloroform, was found in the polished DI water. It was speculated that another source of chloroform existed in the polishing loop. These results show that THMs can be effectively removed by standard high-purity water purification processes, but can also reappear if the purification systems are not properly maintained. Ultrapure Water TOC and THM Contaminant Guidelines for Semiconductor Facilities Balazs Analytical Laboratory has published Ultrapure Water Monitoring Guidelines 7 for the year 2000 and beyond that have demonstrated acceptable product yields for semiconductor companies. Listed below are the maximum acceptable levels of TOC and total THMs compared to the line width of a device. Modern water purification operations such as those discussed in Table 1, can remove THMs down to or below these maximum acceptable levels. Test 16M 0.5 µm 64M 0.35 µm 256M 0.25 µm 1G 0.18 µm 4G 0.13 µm Resistivity TOC (ppb) On-line 2 2 < 2 < 2 < 1 THMs (ppb) < 2 < 2 < 2 < 2 < 1 Table 2: Balazs ultrapure water TOC and THM contaminant guidelines

3 Conductivity Based TOC Measurement Systems For low level TOC measurements, semiconductor and pharmaceutical water systems utilize conductometric based detection systems. Some organizations 7 have reported erroneous and falsely high TOC results in the presence of THMs from Anatel analyzers, without considering some very important experimental factors. The primary issues to be addressed are just how accurate are these TOC analyzers at such low concentration levels of both TOC and THMs, and the impact of THMs on their measurement accuracy. Two common conductometric based TOC analyzer designs are A) stop-flow, UV/TiO 2 oxidation, with differential conductivity measurement and B) continuous flow, UV oxidation, CO 2 membrane separation with differential conductivity measurement. Advantages and disadvantages of each system are as follows: Analyzer A (A-1000/A-1000XP/Access 643) Advantages Disadvantages On-line, no sample contamination, trending Limited resistivity range Limits of detection (parts per trillion levels) No Reagents or gases Analyzer B (Sievers Model 800, Model 400, Ultrapure PPT) Advantages On-line, no sample contamination, trending Limits of detection Disadvantages Limited resistivity range Cannot prove complete oxidation No oxidation catalyst in low range Incomplete CO 2 transfer through membrane A dosing experiment was conducted to evaluate the accuracy and oxidation efficiency of an Anatel A-1000XP and Sievers Model 800 analyzer. Background levels of TOC in an ultrapure water loop were determined with each instrument to generate a baseline and then a known TOC concentration was injected into the loop. A schematic of the experimental setup is shown in Figure 1. Sucrose and isopropyl alcohol were tested at 1 ppb, 5 ppb, 25 ppb, 50 ppb, 75 ppb and 100 ppb as carbon. Sucrose was selected to demonstrate oxidation efficiencies with a compound not typically found in water systems, and IPA was chosen because it can be a contaminant at points-of use (POU) where TOC measurements are performed. Figure 1: Dosing loop experimental setup

4 The A-1000XP routinely returned accurate (within 2% recovery) TOC values for all concentrations tested independent of the type of organic compound measured. The Sievers Model 800 consistently underreported (< 80% recovery at 25 ppb and below; <90% recovery at 50 ppb to 100 ppb) TOC values again independent of the type of organic compound. Representative results from a 25 ppb dose of sucrose and IPA are shown in Figures 2 and 3 respectively. Data collected from this dosing study indicate that the A-1000XP, due to its stop-flow method of analysis coupled with complete oxidation using dynamic endpoint detection, provides accurate and repeatable TOC values. The Sievers Model 800 because of its flow through method of analysis, reports lower than expected TOC values, presumably due to the instrument s reliance on an analytically inaccurate single-point calibration curve and/or partial oxidation of the TOC in the water. Figure 2: 25 ppb TOC Dose of Sucrose Figure 3: 25 ppb TOC Dose of IPA

5 Analyzing THMs with On-line Conductivity based TOC Analyzers As was stated previously, The Environmental Protection Agency (EPA) has set total THM concentration limits in all drinking waters at a maximum of 80 ppb. Chloroform is the most common THM. If it is assumed that all the THM present is chloroform, then 80 ppb of chloroform would be the limit that could be present in drinking water. Chloroform is 10% carbon: CHCl 3 = 121 g/mol 1 C = 12 g/mol (12gmol - /121 gmol - )100 = 10% 80 ppb of chloroform is 8 ppb as carbon. Table 1 shows the effectiveness of ultrapure water systems in removing THMs. Therefore, realistic levels of chloroform at the POU where on-line conductivity based TOC analyzers are operating should be close to 0 ppb or sub ppb levels. Using the same dosing setup as for the accuracy studies, a 9.24 ppb solution of chloroform was injected into the dosing loop and TOC measurements were taken with the A-1000XP and the Sievers Model 800. A 9.24 ppb chloroform solution contains ppb as carbon. Table 3 shows the average TOC values and percent error obtained from each analyzer. The A-1000XP read approximately 3 times higher than expected and the Sievers Model 800 read 1.5 times lower than expected. The following equation illustrates the oxidation of chloroform: CHCl 3 + HOH 1CO 2 + 3HCl H + + HCO (H + + Cl - ) Each H + ion contributes ~ 350 µs/mmol, the Cl - contributes ~ 60 µs/mmol, and the HCO 3 - adds approximately 45 µs/mmol. If only 1 mole each of H + and HCO 3 - were produced during oxidation the resulting conductivity would be: H + + HCO 3 - = 350 µs/mmol + 45 µs/mmol = 395 µs/mmol Since 3 moles each of H + + Cl - are also produced during oxidation, the resulting solution total conductivity is: H + + HCO H + + 3Cl - = 350 µs/mmol + 45 µs/mmol + 3(350) µs/mmol + 3(60 µs/mmol) = 1625 µs/mmol The presence of 3 moles of HCl elevates the solution conductivity by 1230 µs/mmol, which is approximately 3 times more than if only 1 mole each of H + + HCO 3 - were present. The A1000XP, and related products, determine TOC concentration by measuring the difference between an initial and final conductivity reading after oxidation. The oxidation of chloroform, using the A-1000XP, produces an elevated TOC reading as much as 3 times the actual value due to the additional 1230 µs/mmol of conductivity that is generated by the HCl. TOC levels in excess of 3 times the expected amount are not possible due to the chemistry previously demonstrated. Chloroform has limited solubility in water and does not dissolve easily. An emulsion is formed when preparing a solution of chloroform. This emulsified solution requires sonication for extended periods to dissolve the chloroform. If the chloroform is analyzed as an emulsion, extremely high TOC readings will result because it is not completely dissolved in the water. Abnormally high percent recoveries in excess of 3000%, as previously published by Ionics Sievers in MICRO magazine 8 and elsewhere are totally inaccurate and physically impossible. These high percent recoveries are due to the incomplete solubility of chloroform in the test solutions. A-1000XP Sievers Model ppb 0.62 ppb % error = 291% % error = 33% Table 3: Dosing of ppb chloroform as carbon The 1.5 times lower than expected % recovery by the Sievers Model 800 is assumed to be a result of partial oxidation due to its flow-though technology and its single point calibration. Both analyzers have trouble measuring TOC in the presence of THMs. According to the Balazs ultrapure water THM contaminant guidelines, 7 less than 2 ppb THM should be present in ultrapure water systems. At this level, the TOC concentration would be 0.2 ppb or less. Realistically, the A-1000XP, and associated products, could read approximately 0.6 ppb TOC. According to the results presented above, the Sievers Model 800 could read as low as 0.07 ppb ppb is the published detection limit for the Sievers Model 800. For standard analytical techniques, the limit of quantitation (LOQ), where reliable and accurate data can be obtained, is defined as 3 times the method detection limit. For the Sievers Model 800, the LOQ would be 0.15 ppb. Therefore, the Sievers Model 800 can not accurately measure or report THMs at the levels that are required to be in ultrapure water because the measured value is less than their LOQ.

6 Conclusions The accuracy of measuring TOC in ultrapure water systems using on-line conductivity based TOC analyzers is dependent upon the analysis techniques used to measure the organic carbon content. Reliable and accurate measurements of ultrapure water that contains THMs at recommended guideline levels (<0.2 ppb) cannot be achieved with the Sievers Model 800. At these levels, the Sievers Model 800 will report TOC concentrations that are less than the instrument s limit of quantitation. The flow-through technology of the Sievers Model 800 allows only partial oxidation of the organics in the water. This partial oxidation, in combination with a single-point calibration, results in under-reported TOC values. The Anatel A-1000XP, and related products, report TOC concentrations that are approximately 3 times higher than the actual water THM levels. The measurements from the A-1000XP, and related products, are reliable and predictable based on water chemistry and the instrument s stop-flow, dynamic endpoint analysis technique. The A-1000XP, and related products, completely oxidize the organics in the water, but in the presence of THMs, in particular chloroform, the instrument over-reports TOC values. At best, a false positive result obtained with an Anatel A- 1000XP, or other related products, will alert personnel if there is an upset or undesirable contaminants in the water system. Depending on the magnitude of the upset, the Sievers Model 800 will not indicate that an upset has occurred. If THMs are present, the most accurate way of determining actual concentration levels is to use GC-MS. References 1. Environmental Protection Agency s Information Collection Rule and disinfection byproducts monitoring; Federal Register 61, USP 23 Fifth Supplement; November 15, PHARMEUROPA; vol. 11, No. 1. March United States Environmental Protection Agency s Office of Ground water and Drinking Water; Microbial and Disinfection Byproduct Rules: Stage 1 D/DBP Rule, EPA 815-F , December 1998 ( th edition of Standard Methods for the Examination of Water and Wastewater, American Public Health Association, 1015 Fifteenth Street NW, Washington D.C Wang, K, and Samantha Tan, Solid Phase Micro-extraction and GC-MS Analysis of Trihalomethanes in Ultrapure Water Systems ; ChemTrace Corp; Presented at WATERTECH 99 Conference, October 5-6, Ultra Pure Water Monitoring Guidelines 2000; Balazs Analytical Laboratory, Sunnyvale, CA; Franklin, K. et al. Verifying the Accuracy of Low-Level TOC Measurements ; MICRO magazine, vol. 17, No. 8. September

RELIABILITY OF TOC RESULTS IN ULTRAPURE WATER ANALYSIS

RELIABILITY OF TOC RESULTS IN ULTRAPURE WATER ANALYSIS Tech Notes: 030101 RELIABILITY OF TOC RESULTS IN ULTRAPURE WATER ANALYSIS The intent of this technical note is to discuss current methods of TOC (Total Organic Carbon) analysis, including their advantages

More information

UTILIZING AN ON-LINE TTHM ANALYZER TO AID IN COMPLIANCE WITH DBP REGULATIONS

UTILIZING AN ON-LINE TTHM ANALYZER TO AID IN COMPLIANCE WITH DBP REGULATIONS UTILIZING AN ON-LINE TTHM ANALYZER TO AID IN COMPLIANCE WITH DBP REGULATIONS SHARON FILLMANN OPERATIONS & WATER QUALITY MANAGER MAY 7, 2014 1 Disinfection By-Product Formation DBPs Naturally-Occurring

More information

UPW Recycle Strategy

UPW Recycle Strategy UPW Recycle Strategy Semiconductor Safety Association April 28, 2000 Texas Instruments Inc. Recycling Water in Dallas since 1994! TI Water Recycling Strategy Strategy: Match Water Qualities to Water Requirements;

More information

THM Analysis From Compliance to Optimization

THM Analysis From Compliance to Optimization THM Analysis From Compliance to Optimization Kazi Hassan Ph.D. Senior Engineer (Water) April 20, 2015 Disinfection Byproducts (DBP) Disinfection byproducts are formed when disinfectants used in water treatment

More information

USP Perspective on Pharmaceutical Waters. Antonio Hernandez-Cardoso, M.Sc. Senior Scientific Liaison September 7, 2017

USP Perspective on Pharmaceutical Waters. Antonio Hernandez-Cardoso, M.Sc. Senior Scientific Liaison September 7, 2017 USP Perspective on Pharmaceutical Waters Antonio Hernandez-Cardoso, M.Sc. Senior Scientific Liaison September 7, 2017 Why is Water Important? Raw material Solvent Ingredient Reagent Cleaning agent (hot

More information

PURELAB Chorus. Solutions for Type I Ultrapure Water

PURELAB Chorus. Solutions for Type I Ultrapure Water PURELAB Chorus Solutions for Type I Ultrapure Configure your solution Step 1: Choose your system Integrated Technology Typical Applications Select The Impurities You Want To Remove Advanced deionization

More information

Technical Challenges and Solutions to Pharmaceutical Water System Testing during Validation and Operation

Technical Challenges and Solutions to Pharmaceutical Water System Testing during Validation and Operation Technical Challenges and Solutions to Pharmaceutical Water System Testing during Validation and Operation Presented by Mark Suparat Tang, Ph.D. September 2005 Thai FDAj Presentation 08.04.05 Topics of

More information

PURELAB Chorus. Solutions for Type I Ultrapure Water

PURELAB Chorus. Solutions for Type I Ultrapure Water PURELAB Chorus Solutions for Type I Ultrapure Configure your solution Step 1: Choose your system Integrated Technology Typical Applications Select The Impurities You Want To Remove Advanced deionization

More information

Importance of pure water in modern ion chromatography

Importance of pure water in modern ion chromatography Importance of pure water in modern ion chromatography Author: Paul Whitehead, PhD, Chem, FRSC, Laboratory Manager, ELGA R&D Facility Ion chromatography (IC) is evolving to meet the analytical demands for

More information

Trihalomethane Formation Potential in Rainwater Harvested from Different Roofing Materials

Trihalomethane Formation Potential in Rainwater Harvested from Different Roofing Materials Trihalomethane Formation Potential in Rainwater Harvested from Different Roofing Materials Project Number: 570652 Principal Investigator: Carolina Mendez, M.S., Department of Civil, Architectural and Environmental

More information

Protegra CS RO Series. Meeting All Your Pure Water Demands. Water Technologies

Protegra CS RO Series. Meeting All Your Pure Water Demands. Water Technologies Protegra CS RO Series Meeting All Your Pure Water Demands Water Technologies Solenoid switch gauge Pump RO-Module Flow meter Pre-Filter control Solenoid Service Restrictor Protegra CS RO Series flowsheet

More information

PURELAB CHORUS SYSTEMS

PURELAB CHORUS SYSTEMS REACTION TANKS PURELAB CHORUS SYSTEMS SOLUTIONS FOR TYPE I ULTRAPURE WATER CONFIGURE YOUR SOLUTION Integrated Technology STEP 1: CHOOSE YOUR SYSTEM Typical Applications Select the Impurities You Want to

More information

water purification system

water purification system selecting a laboratory water purification system Water Purification Systems UK selecting a laboratory water purification system Clean water is a basic human requirement, purified water is a laboratory

More information

Water Purification System

Water Purification System Water Purification System Heal Force leads you to healthier life Heal Force R NW Series Ultrapure Water System NW series ultrapure water system can be connected directly to tap water source.the system

More information

Purite Water Purification Systems

Purite Water Purification Systems SELECT WATER PURIFICATION Water Purification The Select range of high performance water purification systems offers a unique combination of reliability, quality and functionality for all laboratory applications.

More information

By Jon S. Kauffman PhD., Lancaster Laboratories

By Jon S. Kauffman PhD., Lancaster Laboratories Robust analytical TOC method validation is essential to the success of any online TOC system, particularly systems that release pharmaceutical-grade water in real time. Meeting USP 643 or EP 2.2.44 specifications

More information

Bucking the Industry Trend. Simple Change to Water System Saves New Jersey Manufacturer $23,000 Annually

Bucking the Industry Trend. Simple Change to Water System Saves New Jersey Manufacturer $23,000 Annually Simple Change to Water System Saves New Jersey Manufacturer by Kevin Preising Sales Engineer . Summary: Over the past two decades, the trend in the water treatment industry has been towards more and more

More information

Controlling Trihalomethanes and the Costs Associated with their Minimization using Automated On-Line Monitoring with the THM-RR system

Controlling Trihalomethanes and the Costs Associated with their Minimization using Automated On-Line Monitoring with the THM-RR system Controlling Trihalomethanes and the Costs Associated with their Minimization using Automated On-Line Monitoring with the THM-RR system Gary L. Emmert Professor and Chair Department of Chemistry The THMs

More information

Sensitive Determination of Hexavalent. Brian De Borba, Lipika Basumallick, Jeffrey Rohrer

Sensitive Determination of Hexavalent. Brian De Borba, Lipika Basumallick, Jeffrey Rohrer Sensitive Determination of Hexavalent Chromium in Drinking Water Brian De Borba, Lipika Basumallick, Jeffrey Rohrer Outline Why do we need a sensitive method for hexavalent chromium analysis? U.S. EPA

More information

The Right Analytical Method for the Right Application: TOC Analysis for Cleaning Validation

The Right Analytical Method for the Right Application: TOC Analysis for Cleaning Validation GE Water & Process Technologies Analytical Instruments Paper The Right Analytical Method for the Right Application: TOC Analysis for Cleaning Validation Advances in technologies and mounting manufacturing

More information

Approaches and Benefits to Online Water Quality Monitoring

Approaches and Benefits to Online Water Quality Monitoring Approaches and Benefits to Online Water Quality Monitoring Matt Umberg, U.S. EPA National Environmental Monitoring Conference Orange County, CA; August 11, 2016 Outline Topic 1: What is Online Water Quality

More information

0UltraClean. Ion Exchange Resins for Ultra-pure Water in the Electronics and Semiconductor Industry

0UltraClean. Ion Exchange Resins for Ultra-pure Water in the Electronics and Semiconductor Industry Ion Exchange Resins for Ultra-pure Water in the Electronics and Semiconductor Industry Lenntech info@lenntech.com www.lenntech.com Tel. +31-15-261.09.00 Fax. +31-15-261.62.89 0UltraClean Introduction Since

More information

Thermo Scientific HyClone WFI Quality Water. Ultra pure to meet today s rapid manufacturing advances

Thermo Scientific HyClone WFI Quality Water. Ultra pure to meet today s rapid manufacturing advances Thermo Scientific HyClone WFI Quality Water Ultra pure to meet today s rapid manufacturing advances What sets Thermo Scientific HyClone WFI Quality Water apart? Our cgmp ISO 9001 manufacturing facility,

More information

UNDERAPPRECIATED RISKS AND DANGERS OF ORGANIC / TOC CONTAMINATION IN POWER PLANT CYCLE CHEMISTRY

UNDERAPPRECIATED RISKS AND DANGERS OF ORGANIC / TOC CONTAMINATION IN POWER PLANT CYCLE CHEMISTRY UNDERAPPRECIATED RISKS AND DANGERS OF ORGANIC / TOC CONTAMINATION IN POWER PLANT CYCLE CHEMISTRY Mr. Kirk Buecher Director Of Power Solutions - Mettler-Toledo Thornton, USA WATER CHEMISTRY MANAGEMENT This

More information

Drinking Water Treatment Optimization Tools

Drinking Water Treatment Optimization Tools Drinking Water Treatment Optimization Tools Introduction Treating water to meet today s potable drinking water regulations requires a delicate balance between seemingly conflicting or competing regulations,

More information

Electro Deionization: EDI Systems.

Electro Deionization: EDI Systems. Electro Deionization: EDI Systems. Electro Pure EDI, Inc.: High technology water tm www.cswaters.co.kr Introduction: EDI Electro Deionization 1. Where does EDI belong in a Pure Water System? 2. How does

More information

News. Power Generation 11. ph Control Reduces Acid Consumption in Neutralization Process up to 90 % Perspectives in Pure Water Analytics THORNTON

News. Power Generation 11. ph Control Reduces Acid Consumption in Neutralization Process up to 90 % Perspectives in Pure Water Analytics THORNTON Power Generation 11 Perspectives in Pure Water Analytics News THORNTON Leading Pure Water Analytics ph Control Reduces Acid Consumption in Neutralization Process up to 90 % THORNTON new instrumentation

More information

A Method to Determine Biodegradable Dissolved Organic Nitrogen in Water ND 58108, USA

A Method to Determine Biodegradable Dissolved Organic Nitrogen in Water ND 58108, USA A Method to Determine Biodegradable Dissolved Organic Nitrogen in Water Tanush Wadhawan a, Halis Simsek a, Murthy Kasi a, Kristofer Knutson b, John McEvoy c, Birgit Pruess c and Eakalak Khan a* a Department

More information

Disinfection Byproducts in the Drinking Water What Does it Mean and What is Being Done

Disinfection Byproducts in the Drinking Water What Does it Mean and What is Being Done Disinfection Byproducts in the Drinking Water What Does it Mean and What is Being Done Existing Water System What Are Disinfection Byproducts (DBPs)? DBPs are formed when disinfectants used in water treatment

More information

adrona Crystal E series Crystal E Ultrapure Crystal E HPLC Crystal E Bio Crystal E Pure Water purification systems

adrona Crystal E series Crystal E Ultrapure Crystal E HPLC Crystal E Bio Crystal E Pure Water purification systems series adrona Water purification systems HPLC Bio Pure Features The highest water quality Perfect water quality guaranteed by two sensors directly at the Grade I dispense point and at the entrance of the

More information

L A B W A T E R ULTRA-PURE WATER. PURELAB Ultra. Ultra-Pure Water Polishing System. The Laboratory Water Specialists

L A B W A T E R ULTRA-PURE WATER. PURELAB Ultra. Ultra-Pure Water Polishing System. The Laboratory Water Specialists ULTRA-PURE WATER L A B W A T E R PURELAB Ultra Ultra-Pure Water Polishing System The Laboratory Water Specialists PURELABultra The PURELAB Ultra The new PURELAB Ultra is the state of the art ultra-pure

More information

Department of Civil Engineering, IIT Delhi (India) CVL722 (Sem1, ) (Dr. Arun Kumar;

Department of Civil Engineering, IIT Delhi (India) CVL722 (Sem1, ) (Dr. Arun Kumar; Department of Civil Engineering, IIT Delhi (India) CVL722 (Sem1, 2015-16) (Dr. Arun Kumar; Email: arunku@civil.iitd.ac.in) Set 1 Do not submit. Please mention your assumptions explicitly. Q1. Say a raw

More information

Engineering & Equipment Division

Engineering & Equipment Division Since their development as practical unit operations in the late 1950 s and early 1960 s, reverse osmosis (RO) and ultra filtration (UF) have been continually expanding the scope of their applications.

More information

Laboratory Water Quality Affects Protein Separation by 2D Gel Electrophoresis

Laboratory Water Quality Affects Protein Separation by 2D Gel Electrophoresis Laboratory Water Quality Affects Protein Separation by 2D Gel Electrophoresis 2D gel electrophoresis remains a dominant technique in proteomics. Obtaining high quality gels requires careful and tedious

More information

Investigations into Method Improvements for Hexavalent Chromium. Presented by Aaron Rose Thermo Fisher/Dionex Products For 2012 Watercon

Investigations into Method Improvements for Hexavalent Chromium. Presented by Aaron Rose Thermo Fisher/Dionex Products For 2012 Watercon Investigations into Method Improvements for Hexavalent Chromium Presented by Aaron Rose Thermo Fisher/Dionex Products For 2012 Watercon Recent EPA Approvals and Announcements December 22, 2010 Statement

More information

Blank Optimization Using Ultrapure Water Suitable for Trace Ion Analysis.

Blank Optimization Using Ultrapure Water Suitable for Trace Ion Analysis. Blank Optimization Using Ultrapure Suitable for Trace Ion Analysis. Daniel Darbouret and Ichiro Kano Millipore Corporation Laboratory Division Research and Development BP 307 78054 St Quentin-en-Yvelines

More information

Field experience by water treatment facilities of Chlorine Dioxide & Chlorine Compound Disinfectant, replacing the traditional chlorine treatment

Field experience by water treatment facilities of Chlorine Dioxide & Chlorine Compound Disinfectant, replacing the traditional chlorine treatment JSC The Urals Research chemical Institute with an Experimental plant (JSC UNICHIM & EP ) Field experience by water treatment facilities of Chlorine Dioxide & Chlorine Compound Disinfectant, replacing the

More information

City of Paso Robles Uses Online Monitor to Detect Low-Level of THMs in Treated Wastewater

City of Paso Robles Uses Online Monitor to Detect Low-Level of THMs in Treated Wastewater City of Paso Robles Uses Online Monitor to Detect Low-Level of THMs in Treated Wastewater The City of Paso Robles, California used an online water quality instrument to characterize and monitor low-levels

More information

PRODUCTION OF PW/HPW/WFI AND USP/EP REQUIREMENTS. Erez Eilam F&U Area Lead Novartis Singapore Bio production Operations

PRODUCTION OF PW/HPW/WFI AND USP/EP REQUIREMENTS. Erez Eilam F&U Area Lead Novartis Singapore Bio production Operations PRODUCTION OF PW/HPW/WFI AND USP/EP REQUIREMENTS Erez Eilam F&U Area Lead Novartis Singapore Bio production Operations Agenda Background USP/EP Standard Steps in production Available technologies Key aspects

More information

Standard Specification for Reagent Water 1

Standard Specification for Reagent Water 1 Designation: D 1193 06 Federal Test Method Standard No. 7916 Standard Specification for Reagent Water 1 This standard is issued under the fixed designation D 1193; the number immediately following the

More information

Using a Recirculating Aeration System to Remove Bromoform and Other THMs. Russell Porter, P.E. May 9, 2014

Using a Recirculating Aeration System to Remove Bromoform and Other THMs. Russell Porter, P.E. May 9, 2014 Using a Recirculating Aeration System to Remove Bromoform and Other THMs Russell Porter, P.E. May 9, 2014 Presentation Outline Plant background & DBP history Description of pilot efforts Carbon Aeration

More information

Technology Note. Monitoring TOC in ultra-pure laboratory water. Control of impurity levels. Figure 1 Control of impurities

Technology Note. Monitoring TOC in ultra-pure laboratory water. Control of impurity levels. Figure 1 Control of impurities Monitoring TOC in ultra-pure laboratory water The serious consequences of organic contamination of ultra-pure water have resulted in the widespread of use of TOC monitoring in addition to resistivity as

More information

Understanding Reverse Osmosis Mark Rowzee Mark Rowzee

Understanding Reverse Osmosis Mark Rowzee Mark Rowzee Understanding Reverse Osmosis Mark Rowzee Mark Rowzee RO produces some of the highest-purity water and is a workhorse technology for drinking water applications. Many people have heard of reverse osmosis

More information

A SIMPLE METHOD OF REMOVING DISINFECTION BYPRODUCT PRECURSORS IN WATER

A SIMPLE METHOD OF REMOVING DISINFECTION BYPRODUCT PRECURSORS IN WATER A SIMPLE METHOD OF REMOVING DISINFECTION BYPRODUCT PRECURSORS IN WATER Tony Robles Town of Kirkland Lake WTP roblest@ntl.sympatico.ca Overall Responsible Operator 705-642-5626 Operator-In-Charge Abstract

More information

5.B Generation of pharmaceutical water Author: Michael Gronwald Co-Author: Dr. Ralph Gomez / Up06

5.B Generation of pharmaceutical water Author: Michael Gronwald Co-Author: Dr. Ralph Gomez / Up06 Generation of pharmaceutical water Generation of pharmaceutical water Author: Michael Gronwald Co-Author: Dr. Ralph Gomez / Up06 Here you will find answers to the following questions: What are the different

More information

Disinfection and Disinfection Byproducts (DBP)

Disinfection and Disinfection Byproducts (DBP) Disinfection and Disinfection (DBP) 1 Regulatory Requirements for Disinfection By-Product Management & Disinfection By-Product Monitoring for Water Treatment Plants 2 DPB Rule Requirements Rule/Requirement

More information

SOURCE WATER MONITORING. Taseen Karim SE PA Regional Sales Manager Greg Leininger- NE Divisional Sales Manager

SOURCE WATER MONITORING. Taseen Karim SE PA Regional Sales Manager Greg Leininger- NE Divisional Sales Manager SOURCE WATER MONITORING Taseen Karim SE PA Regional Sales Manager Greg Leininger- NE Divisional Sales Manager SOURCE WATER MONITORING IMPORTANCE Nice to know something is changing before it gets into your

More information

INDUSTRIAL WATER ENGINEERS (M) SDN. BHD.

INDUSTRIAL WATER ENGINEERS (M) SDN. BHD. INDUSTRIAL WATER ENGINEERS (M) SDN. BHD. The most cost-effective Water & Wastewater Treatment System: Design and Build Locally State-Of-Art Air/Biogas Scrubbing System and Biogas Solutions With the most

More information

IMEC. March PW & HPW Generation

IMEC. March PW & HPW Generation IMEC March 2014 PW & HPW Generation CONTENT Introduction In 10 Seconds.. Our Business Approach PHARMA : Purified Water Generation Case Study MICRO ELECTRONIC : High Purified Water Generation Case Study

More information

CTB3365x Introduction to Water Treatment

CTB3365x Introduction to Water Treatment CTB3365x Introduction to Water Treatment D5a Direct Surface water treatment Luuk Rietveld Although surface water is mostly contaminated, we need this source for drinking water production. Welcome! Today

More information

10. Validated Normal Phase HPLC Method for the Determination. Fulvestrant is primarily used in the treatment of hormone receptor

10. Validated Normal Phase HPLC Method for the Determination. Fulvestrant is primarily used in the treatment of hormone receptor 229 10. Validated Normal Phase HPLC Method for the Determination of Fulvestrant in Pharmaceutical Dosage Forms 10.1 Introduction Fulvestrant is primarily used in the treatment of hormone receptor positive

More information

Microbial Detection. Continuous At-Line Microbial Monitoring For Pharmaceutical Waters

Microbial Detection. Continuous At-Line Microbial Monitoring For Pharmaceutical Waters Microbial Detection 7000RMS Microbial Detection Analyzer Continuous, real-time analysis Results in seconds No sample preparation No reagents required Process/batch sampling modes Continuous At-Line Microbial

More information

Town of Salem, New Hampshire Water Treatment Plant Disinfection Update

Town of Salem, New Hampshire Water Treatment Plant Disinfection Update Town of Salem, New Hampshire Water Treatment Plant Disinfection Update Beginning in July of 2007, the Town of Salem will be changing the disinfectant it maintains in its distribution system from chlorine

More information

Water Production for Pharmaceutical Manufacturing: Looking Beyond Compliance at the Entire Manufacturing Process

Water Production for Pharmaceutical Manufacturing: Looking Beyond Compliance at the Entire Manufacturing Process Water Production for Pharmaceutical Manufacturing: Looking Beyond Compliance at the Entire Manufacturing Process Grantek is a Systems Integrator focused on automation with extensive expertise and partnerships

More information

The evolution of high purity water production

The evolution of high purity water production FEDI Fractional Electrodeionization from EDI to FEDI The evolution of high purity water production EDI THE CONVENTIOL PROCESS Reject Electrodeionization Cell DEIONIZED WATER Reject Electrodeionization

More information

performance of the ultrapure make-up system at Turkey Point Nuclear Power Plant

performance of the ultrapure make-up system at Turkey Point Nuclear Power Plant Water Technologies & Solutions technical paper performance of the ultrapure make-up system at Turkey Point Nuclear Power Plant Authors: Stephen P. Ellis, Laboratory Manager, Ionics and Richard Steinke,

More information

OVIVO. Online THM Analyzer

OVIVO. Online THM Analyzer OVIVO Online THM Analyzer Ovivo Online THM Analyzer OVIVO has developed a specific online process analyzer for trihalomethanes detection in potable, process and high purity water: OVIVO THM1501 Measured

More information

FEDI. Fractional Electrodeionization. The evolution of high purity water production

FEDI. Fractional Electrodeionization. The evolution of high purity water production FEDI Fractional Electrodeionization from EDI to FEDI The evolution of high purity water production EDI THE CONVENTIONAL PROCESS Reject Electrodeionization Cell DEIONIZED WATER Reject The Electrodeionization

More information

Balancing Microbial Control and Stage 2 DBP Rule Compliance (at Ocean City, Maryland)

Balancing Microbial Control and Stage 2 DBP Rule Compliance (at Ocean City, Maryland) Balancing Microbial Control and Stage 2 DBP Rule Compliance (at Ocean City, Maryland) May 4, 2012 Presented by: Earl Swartzendruber, Jr., PE Whitman, Requardt and Associates, LLP Acknowledgements Town

More information

HPLC Sample Preparation

HPLC Sample Preparation HPLC Sample Rely on Sartorius solutions and benefit from optimized workflows for clean results mau 400 350 300 250 200 150 100 50 HPLC-Analyse 0 min 50 0,0 1,0 2,0 3,0 4,0 5,0 6,0 7,0 8,0 9,0 10,0 11,0

More information

NJWA Presents DISINFECTION BYPRODUCTS: TREATMENT OPTIONS & CHALLENGES. Stage 2 D/DBPR. Trihalomethanes DISINFECTION BYPRODUCTS.

NJWA Presents DISINFECTION BYPRODUCTS: TREATMENT OPTIONS & CHALLENGES. Stage 2 D/DBPR. Trihalomethanes DISINFECTION BYPRODUCTS. NJWA Presents DISINFECTION BYPRODUCTS: TREATMENT OPTIONS & CHALLENGES Donald C. Kucher Chlorinator & Instrument Service, Inc. Hampton, NJ 908-735-2303 email cl2fixd@aol.com www.cl2fixd.com Stage 2 D/DBPR

More information

Mitigation of Bromate during Ozonation

Mitigation of Bromate during Ozonation Mitigation of Bromate during Ozonation John C. Crittenden, Ph.D., P.E., N.A.E. Daisuke Minakata Arizona State University Zaid K. Chowdhury A Malcolm Pirnie Perspective 1 Outline Application of ozone to

More information

produce washing, horticulture, carcass washing, brewing and beverage production.

produce washing, horticulture, carcass washing, brewing and beverage production. CHLORINE DIOXIDE (ClO2) The Gold Standard of Complete Water Disinfection Solution FLORIDA ENVIRO SOLUTIONS LLC., The Authorized Representative & Distributor of Two Components Solid Forms of CHLORINE DIOXIDE

More information

Water supplied by Marafiq does not meet the process requirements.

Water supplied by Marafiq does not meet the process requirements. WATER TREATMENT In Industries Water is used for: a. Drinking b. Cleaning c. Cooling d. Producing Steam e. Process requirement Why we need to treat water? For human consumption a. Water to be purified (Make

More information

Additional guidance on: TNsG on Data Requirements, Part A, Chapter 2, Point 4. Analytical Methods for Detection and Identification.

Additional guidance on: TNsG on Data Requirements, Part A, Chapter 2, Point 4. Analytical Methods for Detection and Identification. Additional guidance on: TNsG on Data Requirements, Part A, Chapter 2, Point 4 Analytical Methods for Detection and Identification and Part B, Chapter 2, Point 4 Methods of Identification and Analysis These

More information

using EDI to meet the needs of pure water production

using EDI to meet the needs of pure water production Water Technologies & Solutions technical paper using EDI to meet the needs of pure water production Authors: Ted Prato and Christopher Gallagher, Ionics This paper was originally presented at the 61st

More information

Elix Gulfstream Clinical Water Purification System

Elix Gulfstream Clinical Water Purification System Elix Gulfstream Clinical Water Purification System All-in-one water purification system for clinical analyzers for analyzers with pure water needs of up to 2000 L daily Merck Millipore is a division of

More information

High Purity Water Needs for the PV Industry Part 1: Development of Cost Effective Specification

High Purity Water Needs for the PV Industry Part 1: Development of Cost Effective Specification High Purity Water Needs for the PV Industry Part 1: Development of Cost Effective Specification John Morgan Sarah Schoen, Andreas Neuber, Marty Burkhart, John Morgan, and Slava Libman SEMI PV Task Force

More information

ENVIRONMENTAL CHEMICAL ANALYSIS III

ENVIRONMENTAL CHEMICAL ANALYSIS III ENVIRONMENTAL CHEMICAL ANALYSIS III Chromatographic Methods Gas Chromatography (GC) Gas chromatography is a separation process in which a gaseous solute is carried through the column by a gaseous mobile

More information

Circulation model for water circulation and purification in a water Cerenkov detector *

Circulation model for water circulation and purification in a water Cerenkov detector * CPC(HEP & NP), 2009, 33(7): 567 571 Chinese Physics C Vol. 33, No. 7, Jul., 2009 Circulation model for water circulation and purification in a water Cerenkov detector * LU Hao-Qi( ) 1) YANG Chang-Gen(

More information

Local Water Contaminants

Local Water Contaminants Local Water Contaminants CWQA 2017 Annual General Meeting Waterloo, Ontario By: Mike Weatherill Purolite Canadian Technical Sales Specialist Presentation Overview Some Water Chemistry Basics Local Contaminant

More information

CITY OF HOUSTON, TEXAS: A CASE STUDY

CITY OF HOUSTON, TEXAS: A CASE STUDY CITY OF HOUSTON, TEXAS: A CASE STUDY Astro International Corporation sold Astro Model 1950s to the City of Houston prior to the sale of Astro by the current Star management. After Astro was sold to Zellweger

More information

Thermo Scientific Barnstead LabTower EDI water purification system

Thermo Scientific Barnstead LabTower EDI water purification system Thermo Scientific Barnstead LabTower EDI water purification system The Barnstead LabTower EDI is an integrated system combining purification with EDI technology and storage in one, as well as producing

More information

TREATMENT OF MTBE CONTAMINATED WATER USING UV/CHLORINE ADVANCED OXIDATION PROCESS

TREATMENT OF MTBE CONTAMINATED WATER USING UV/CHLORINE ADVANCED OXIDATION PROCESS TREATMENT OF MTBE CONTAMINATED WATER USING UV/CHLORINE ADVANCED OXIDATION PROCESS By A. Kedir, B. Tawabini, A. Shaibani, A. Bukhari Feb.19 th, 2015 Water Arabia 2015 Le Meridien Khobar, Saudi Arabia 1

More information

Water purification solutions

Water purification solutions Water purification solutions Razmišljate li o kvaliteti vode u Vašem laboratoriju? Osijek, 06.10.2011. Prehrambeno-tehnološki fakultet u Osijeku Jasenka Ivanković, Merck d.o.o. Zagreb Agenda - Water purification

More information

CLEANING VALIDATION FOOD & BEVERAGE WASTEWATER LABORATORY PHARMACEUTICAL & MEDICAL DEVICE

CLEANING VALIDATION FOOD & BEVERAGE WASTEWATER LABORATORY PHARMACEUTICAL & MEDICAL DEVICE CLEANING VALIDATION FOOD & BEVERAGE WASTEWATER LABORATORY PHARMACEUTICAL & MEDICAL DEVICE Company Overview International Products Corporation, incorporated in 1923, manufactures specialty cleaners and

More information

Online Organics Monitoring for Rapid Process Control of Drinking Water and Wastewater Treatment Operations

Online Organics Monitoring for Rapid Process Control of Drinking Water and Wastewater Treatment Operations Online Organics Monitoring for Rapid Process Control of Drinking Water and Wastewater Treatment Operations Amanda Scott Suez Water Technologies & Solutions 2018 ISA Water/Wastewater & Automatic Controls

More information

Use of Spiral Wound UF in RO Pretreatment

Use of Spiral Wound UF in RO Pretreatment Use of Spiral Wound UF in RO Pretreatment Authors: Harry Dalaly, Asia Pacific Sales Manager, Hydranautics and Yeoh Cheik How, Facilities Senior Engineer, Hewlett-Packard Singapore The Hewlett-Packard Singapore

More information

TKA LabTower EDI. Two in one! Pure and Ultrapure water WATER PURIFICATION SYSTEMS

TKA LabTower EDI. Two in one! Pure and Ultrapure water WATER PURIFICATION SYSTEMS TKA LabTower EDI Two in one! Pure and Ultrapure water WATER PURIFICATION SYSTEMS The integrated system. Purification and tank in one. High purity water ASTM II 15 10 MΩ cm. For daily needs of 100 to 500

More information

Analysis of Trace Organic Contaminants in Water by Direct Injection Using Agilent 6490 LC/MS/MS with Pos/Neg Switching

Analysis of Trace Organic Contaminants in Water by Direct Injection Using Agilent 6490 LC/MS/MS with Pos/Neg Switching Analysis of Trace Organic Contaminants in Water by Direct Injection Using Agilent 6490 LC/MS/MS with Pos/Neg Switching Application Note Environmental Authors Tarun Anumol, Shimin Wu, Sylvain Merel, and

More information

Production of High Purity Functional Water at Point-of-Use for Advanced Mask Cleaning Processes

Production of High Purity Functional Water at Point-of-Use for Advanced Mask Cleaning Processes purify protect transport Production of High Purity Functional Water at Point-of-Use for Advanced Mask Cleaning Processes 2010 International Workshop on EUV Lithography June 22-24, 2010 Annie Xia, Sr. Applications

More information

The U.S Environmental Protection

The U.S Environmental Protection FWRJ Distribution System Water Quality Models Support Treatment Process Decisions Brandon Bryant, Michael Harber, and Robert Dehler The U.S Environmental Protection Agency (EPA) has estimated that over

More information

REACTION TANKS CENTRA CENTRALIZED LABORATORY WATER SYSTEMS

REACTION TANKS CENTRA CENTRALIZED LABORATORY WATER SYSTEMS REACTION TANKS CENTRA CENTRALIZED LABORATORY WATER SYSTEMS Total confidence in your Pure Water system. The CENTRA product family includes a Centralized Recirculating Distribution System and Centralized

More information

THORNTON 5000TOC Sensor and 700MAX demonstrate online measurement capability

THORNTON 5000TOC Sensor and 700MAX demonstrate online measurement capability 5 THORNTON Leading Pure Water Analytics THORNTON 5000TOC Sensor and 700MAX demonstrate online measurement capability A US-based wafer tool manufacturer to the Semiconductor industry produces ultrapure

More information

Wastewater Treatment Technology and Applications in Industrial Facilities

Wastewater Treatment Technology and Applications in Industrial Facilities Wastewater Treatment Technology and Applications in Industrial Facilities Treatment that industrial facilities give wastewater before discharging it to the local wastewater treatment facility is referred

More information

Water purification consumable information sheet

Water purification consumable information sheet Water purification consumable information sheet Protect your investment and maintain your Thermo Scientific water purification system for optimal performance. Use this guide to identify replacement consumables

More information

NEW TECHNOLOGIES & BREAKTHROUGHS TO OLD DEIONIZATION APPLICATIONS

NEW TECHNOLOGIES & BREAKTHROUGHS TO OLD DEIONIZATION APPLICATIONS NEW TECHNOLOGIES & BREAKTHROUGHS TO OLD DEIONIZATION APPLICATIONS By: Jim Wark, jwark@culligantx.net STS AIChE March 2018 Monthly Dinner Meeting Workshop Thursday, March 1, 2018, 5:30pm-6:30pm CST The

More information

Ultrapure water: LC-MS suitability tests

Ultrapure water: LC-MS suitability tests Ultrapure water: LC-MS suitability tests Application details Contaminations such as ions and organic traces present in ultrapure water affect LC- MS analyses and decrease perfomance of the analytical instumentation.

More information

Drinking Water Disinfection A History and Improved Monitoring Techniques. Randy Turner Technical Director, Chemist Swan Analytical USA

Drinking Water Disinfection A History and Improved Monitoring Techniques. Randy Turner Technical Director, Chemist Swan Analytical USA Drinking Water Disinfection A History and Improved Monitoring Techniques By Randy Turner Technical Director, Chemist Swan Analytical USA History In many countries we take for granted the availability and

More information

Unit Treatment Processes in Water and Wastewater Engineering

Unit Treatment Processes in Water and Wastewater Engineering Unit Treatment Processes in Water and Wastewater Engineering T J Casey AQUAVARRA RESEARCH LIMITED 22A Brookfield Avenue Blackrock Co. Dublin. October 2006 Author s Note Water and wastewater treatment technology

More information

Importance of TOC measurement in WFI in light of European Pharmacopoeia change

Importance of TOC measurement in WFI in light of European Pharmacopoeia change WITE PAPER Importance of TOC measurement in WFI in light of European Pharmacopoeia change Tony arrison, Applications Specialist Beckman Coulter, Inc. Abstract The new revision of the European Pharmacopoeia

More information

Chlorine Dioxide One Year Later. Shazelle Terry Treatment & WQ Department Manager

Chlorine Dioxide One Year Later. Shazelle Terry Treatment & WQ Department Manager Chlorine Dioxide One Year Later Shazelle Terry Treatment & WQ Department Manager Presentation Agenda Background JVWTP Chlorine Dioxide History JVWTP Full Scale Implementation Results Summary 2 Chlorinated

More information

Pure and ultrapure laboratory water from a single system

Pure and ultrapure laboratory water from a single system Pure and ultrapure laboratory water from a single system Your laboratory water specialists ELGA has been a trusted name in water purification for over 50 years, pioneering innovative technologies and award

More information

Disinfectant dosing of blended drinking waters

Disinfectant dosing of blended drinking waters 18 th World IMACS / MODSIM Congress, Cairns, Australia 13-17 July 2009 http://mssanz.org.au/modsim09 Disinfectant dosing of blended drinking waters van Leeuwen, J. 1,3, D. Cook 2, C. Chow 1,2,3 and M.

More information

YOUR PARTNER FOR TURNKEY SOLUTIONS IN PHARMA & BIOTECH INDUSTRY AN ISO 9001:2008 CERTIFIED COMPANY

YOUR PARTNER FOR TURNKEY SOLUTIONS IN PHARMA & BIOTECH INDUSTRY AN ISO 9001:2008 CERTIFIED COMPANY CHEMION PURIFIED WATER GENERATION, STORAGE AND DISTRIBUTION SYSTEM YOUR PARTNER FOR TURNKEY SOLUTIONS IN PHARMA & BIOTECH INDUSTRY AN ISO 9001:2008 CERTIFIED COMPANY CHEMION ENGINEERING THE COMPANY The

More information

Milli-Q Reference Water Purification System. The reference for ultrapure water systems

Milli-Q Reference Water Purification System. The reference for ultrapure water systems Milli-Q Reference Water Purification System The reference for ultrapure water systems Ultrapure water by Merck Millipore Easily prepare solutions with the low flow function to precisely adjust the meniscus

More information

ARSENIC IN DRINKING WATER

ARSENIC IN DRINKING WATER ARSENIC IN DRINKING WATER Treatment Technologies: Removal Background In water, the most common valence states of arsenic are As(V), or arsenate, which is more prevalent in aerobic surface waters and As(III),

More information

[PPT PDF] Pharmaceutical Water System Design Validation - Microbial Testing of Water is discussed in detail in this article.

[PPT PDF] Pharmaceutical Water System Design Validation - Microbial Testing of Water is discussed in detail in this article. [PPT PDF] Pharmaceutical Water System Design Validation - Microbial Testing of Water is discussed in detail in this article. Pharmaceutical Water System: Classical Culture Approach for microbial testing

More information