1 Copyright JCPDS-International Centre for Diffraction Data 2014 ISSN ISO Simplified Borate Fusion / WDXRF Analytical Method for Iron Ore Including Total Iron Analysis: Part 2 Mathieu Bouchard, Alex Milliard, and Sebastien Rivard Corporation Scientifique Claisse, Québec, QC, Canada S. Ness Intertek Genalysis, Perth, Australia ABSTRACT A very efficient analytical method using an automated fusion machine as sample preparation tool and a Wavelength Dispersive X-ray Fluorescence (WDXRF) spectrometer for the determination of all the elements of interest for the iron ore industry has been optimized from the ISO standard method. This updated method allows for the simplification of both laboratory and spectrometry processes and so, in comparison with the original ISO method, becomes less restrictive in practice. This method was used to prepare a large variety of iron ores and exploration samples from all over the world. Results of the prevailing XRF application based on pure oxide standards as described in the ISO standard method are compared to the results of a unique XRF calibration application based on Certified Reference Materials (CRMs) for iron ores and iron ores exploration materials. The universal sample preparation fusion method for iron ores and exploration samples developed during the first phase of this project was used to select and evaluate a set of CRMs. Selected throughout the world from well recognized sources, the chosen set of CRMs allows a wide coverage for all the elements of interest of the iron ore industry, excluding tin oxide. This fusion method allows a matrix match for materials from different origins. A critical evaluation of precision and accuracy has been performed against the ISO standard method. Reference materials not included in the calibration (Control samples) was also investigated for accuracy evaluation. Furthermore, comparisons will be made between the data collected from this fusion method implemented in combination with a Bruker S4 Explorer WDXRF Spectrometer, and the analytical requirements of ISO. All deviations from the prevailing standard method parameters (calibration, standards, flux, Fluxer, etc.) will also be pointed out and discussed.
2 This document was presented at the Denver X-ray Conference (DXC) on Applications of X-ray Analysis. Sponsored by the International Centre for Diffraction Data (ICDD). This document is provided by ICDD in cooperation with the authors and presenters of the DXC for the express purpose of educating the scientific community. All copyrights for the document are retained by ICDD. Usage is restricted for the purposes of education and scientific research. DXC Website ICDD Website -
3 Copyright JCPDS-International Centre for Diffraction Data 2014 ISSN INTRODUCTION Whether it is dealing with high-grade hematite, low-grade hematite or magnetite, assays performed on exploration sample, concentrate, internal standard and/or reference material, will always be of paramount financial importance when it comes to the survival of a mining company. In this regard, the existence of a small bias associated with the total iron analysis process could result in millions of dollars in revenue losses for mining companies given the fact that their products are sold in millions of tons. The current International Standard Method, which remains widely accepted among peers, is the one entitled: Iron ores -- Determination of Various Elements by X-ray Fluorescence Spectrometry -- Part 1: Comprehensive Procedure (ISO :2003). However, there are a number of unveiled and corroborated limitations to the predominant version of this standard; it lacks adaptability when coping with recent advances in the fields of sample preparation by fusion and Wavelength Dispersive X-ray Fluorescence (WDXRF) spectrometry. The preparation of standards made of pure oxides is a complex and time consuming task and the calibration ranges do not cover for the exploration samples. The industry needs a single calibration for both the iron ore products and the exploration samples and forced us to rethink the calibration methodology that is described in the standard method. Good Certified Reference Materials (CRMs) from this industry are readily available and allow the evaluation of a simplified calibration strategy to the usual mixes of pure oxides. Using CRMs from worldwide sources that cover both the iron ores and the exploration samples as required by the industry, makes the matrix covered by this application much more complex. It also makes using pressed powder much more challenging due to the difficulties of matrix matching the calibration standards and the samples from the ores and exploration samples. Taking this information into account, using the borate fusion preparation allows for a more accurate analysis and requires less calibration curves because the technique removes particle size and mineralogy effects [1, 2]. For these reasons, and also to facilitate the lab work needed for the calibration, it is desirable to use a single fusion method, for the preparation of all the iron ore types and most of the common exploration sample types, combined with WDXRF, to comply with the ISO analytical performance requirements. To reach these objectives, a robust analytical method using an automated fusion machine as sample preparation tool and a WDXRF spectrometer has been optimized from the methodology described in the ISO standard method for the quantification of all elements of interest in the iron ore industry. This single method was used to prepare fused disks from more than 150 types of materials (CRMs and various samples) covering a very vast range of compositions. A set of CRMs from more than 10 suppliers of different origins were selected as calibration standards and allowed a matrix match for worldwide origin iron ores. The performance evaluation was completed using the new calibration approach and according to the instructions provided by the International Organization for Standardization (ISO), through the standard method for analysis of iron ores by X-ray fluorescence: ISO [3]. To better evaluate the accuracy of the CRM based calibration, a pure oxide calibration was prepared according to the ISO guidelines and served as a reference.
4 Copyright JCPDS-International Centre for Diffraction Data 2014 ISSN EXPERIMENTAL -Apparatus and instrumental conditions A Claisse M4 TM propane fired automatic Fluxer was used to generate all calibration standard fusion glass disks, but sample preparation for the precision evaluation were run using both the M4 TM and the TheOx Fluxers; The M4 TM auto-regulating gas system fluxer and the TheOx electric fluxer have both been designed with pre-set fusion programs allowing for the most repeatable and reproducible fusion conditions as well as the capability to retain the volatile elements.. A Fisher Scientific drying oven was used for moisture determinations. This method was used for all the samples and consisted in drying at 105 C in a clean ceramic/porcelain crucible for 120 minutes. A Fisher Scientific Isotemp programmable muffle furnace was used for the Loss On Ignition (LOI) determinations. The LOI method used for all the samples was an ignition at 1000 C in a clean ceramic/porcelain crucible for 60 minutes. This method was completed after the moisture determination, using the same crucible and dried sample. Table 1. Spectrometer Operation Parameters Element / Line kv ma Crystal Collimator Detector a Peak ( 2 ) Time (s) Bkd 1 b ( 2 ) Bkd 2 b ( 2 ) Time bkd (s) Fe K LiF 220 Fine SC Si K PET Fine FC Ca K LiF 200 Coarse FC Mn K LiF 200 Coarse SC Al K PET Coarse FC Ti K LiF 200 Coarse FC Mg K XS-55 Coarse FC P K Ge Coarse FC S K Ge Coarse FC K K LiF 200 Coarse FC Sn L LiF 200 Fine FC V K LiF 200 Coarse FC Cr K LiF 200 Coarse SC Co K LiF 200 Coarse SC Ni K LiF 200 Coarse SC Cu K LiF 200 Coarse SC Zn K LiF 200 Coarse SC As K LiF 200 Coarse SC Pb L LiF 200 Coarse SC Ba L LiF 200 Coarse FC Na K XS-55 Coarse FC Sr K LiF 200 Fine SC Zr K LiF 200 Coarse SC a FC = gas flow proportional counter; SC = Scintillation Counter b Bkd 1 and Bkd 2 = position value for the needed background when used
5 Copyright JCPDS-International Centre for Diffraction Data 2014 ISSN A Bruker-AXS S4 Explorer sequential WDXRF spectrometer with a rhodium endwindow X-ray tube was used for data generation. A 34 mm collimator mask and vacuum were used for all the measurements. Spectrometer analytical conditions, peak-line, background measurements, background position, pulse-height, counting time and others were selected and optimized following the guidelines of ISO and by wavelength step-scanning of selected standard disks. The spectrometer set-up and performance evaluation guide also included in the ISO were used to verify proper spectrometer operation. The optimal spectrometer analytical conditions for the measurement of all elements are listed in Table 1. -Sample preparation method The optimization of the sample preparation was performed on both the iron ore and the exploration samples using a majority of parameters described in ISO , still some parameters had to be modified to achieve such a wide calibration. The platinum ware used for this project was made from 95% platinum and 5% gold. This alloy is accepted by the ISO standard and for decades, it has been the most commonly used material for this purpose. As described in the ISO test method, it was proven that our optimized fusion method can be used to prepare glass disks with a diameter ranging between 32 to 40 mm. Nevertheless, all glass disks produced for this application had a 40 mm diameter. The sample-to-flux ratio was as described in the ISO standard and kept at 1: Sample can be fused on an original basis or on a dry basis. The "Catch Weight" correction used for weighing was applied as described in the standard method. The standard method recommends using one of the three following fluxes: Pure sodium tetraborate, pure lithium tetraborate (LiT) or a mix of 35% lithium tetraborate with 65% of lithium metaborate (LiM). The flux used for this study was 50% lithium tetraborate with 50% of lithium metaborate which made the preparation of exploration samples easier without decreasing the success rate of the iron ore preparation. The oxidizer recommended by the standard method is Sodium Nitrate (NaNO 3 ), but after evaluating three different oxidizers (NaNO 3, LiNO 3 and NH 4 NO 3 ), ammonium nitrate (NH 4 NO 3 ) was selected for this project since it allowed analyzing Sodium (Na) on top of obtaining very precise results. This substitution was made because Na is now considered as an element of interest in the iron ore related materials. The ammonium iodine can be used as a Non-Wetting Agent (NWA) when needed, as stated in the standard method. However, the methodology described in this paper does not recommend using NWA. It has also been determined that using a VortexMixer TM to mix the sample and flux homogeneously prior to the fusion is more advantageous than using a glass rod, as it minimizes the possibility of material loss. The VortexMixer TM speed was controlled to avoid any loss of material because variance from the ratio of flux to sample weight causes error in the results [4]. The temperature range of the fusion process was kept between 1000 and 1050 C. It is well known that over the critical temperature of 1050 C, flux [5] and other compounds like SO 3 [2] begin to volatilize without consistency and can change the sample-to-flux ratio. Finally, the fusion process had pre-programmed steps with fixed times in order to obtain the highest level of precision and accuracy. The study presented in the following paper proves that a high level of precision is attainable for this sample preparation using both gas powered and electric powered fully automatic fusion instruments.
6 Copyright JCPDS-International Centre for Diffraction Data 2014 ISSN Global sample preparation method A 40 mm diameter, 1 mm thick shallow mold 1 was used to eliminate the curvature effect, which can occur after multiple heating cycles. Pure grade pre-fused flux * composition of 50.0% lithium tetraborate, 50.0% lithium metaborate was selected to increase the homogeneity and make stable glass disks. The maximum fusion temperature used for this fusion is 1050 C for both gas and electric Fluxers. The cooling process was achieved with forced air for 5 minutes. -Step by step procedure It is important to understand that this method can fuse "as received" basis samples or dried basis samples. If the disks are made using dried samples, the LOI determination must be established in parallel to the fusion process for spectrometry correction. If the "as received" basis samples are preferred over the dry basis samples, a weigh factor will have to be calculated from the moisture determination in parallel to establishing the LOI determination for spectrometry correction. For a 40 mm glass preparation, weigh 0.66 g of sample in a clean and dry Pt/Au crucible. Report the value of the weighed sample portion with ± g tolerance before any further calculations. Tare the analytical scale. Add 0.40 g of NH 4 NO 3 with a tolerance of ± 0.01 g on top of the sample portion. Tare the scale again. Then add 6.80 g of Claisse LiT/LiM: 50.0/50.0, Pure Grade Flux on top of the sample/oxidizer portion. Report the value of the weighed flux portion with ± g tolerance before any further calculations. Use a VortexMixer TM to mix the sample/oxidizer with the flux. The VortexMixer TM speed must be controlled to avoid any material loss, because variance from the ratio of flux to sample weight causes error in the results [4]. Place the crucible with its preparation on the selected fusion equipment. Utilize the appropriate optimize fusion program for the selected fusion equipment. Analyze the glass preparation as soon as possible and/or protect the obtained glass disk in a safety container on a vacuumed desiccator. -Robustness of the fusion method More than 160 different samples from 12 renowned mining companies and reference materials from more than 10 recognized reference material manufacturers were fused using the global project. This analytical method for iron ore products and exploration samples revealed good efficiency to prepare homogenous and stable lithium borate glass disks with all of the materials; it had limitations only for the preparation of iron ore related materials that contained relatively high level of Copper (Cu). When the content of this element was higher than 500 to 700 ppm, the glass disk had a tendency to stick to the mold and often lead to disk cracking. For this reason, it was determined that high Copper 1 Available at Corporation Scientifique Claisse
7 Copyright JCPDS-International Centre for Diffraction Data 2014 ISSN iron ores needed a different sample preparation fusion methodology that includes a NWA to avoid having the disk to stick to the mold. -Preparation for calibration, selection of control samples Pure Oxide Calibration: For this part of the work, we referred directly to the ISO standard method [3]. Only a few modifications were brought to the described procedure. First, we added a binary standard made of 30% Al 2 O 3 and 70% of Fe 2 O 3 to cover a wider range of exploration sample. We then prepared the synthetic calibration standard to add low concentration sodium points on the calibration curve. Finally, two binary overlap calibration standards were modified from 10% Cr 2 O 3 : 90% SiO 2 and 10% ZnO: 90% SiO 2 to 2% Cr 2 O 3 : 98% SiO 2 and 2% ZnO: 98% SiO 2, because they were impossible to fuse with the optimized automated fusion sample preparation. The only way to prepare these two calibration standards in their original composition was by increasing the fusion time and/or by increasing the temperature of the method. It was preferred to modify the calibration standard compositions instead of modifying the sample preparation method in a way which would of decreased the productivity and/or the precision of the sample preparation method. Certified Reference Materials Calibration: As previously mentioned, more than 80 CRM glass preparations were produced and evaluated only to select the best set of standards for the calibration of the borate fusion and XRF analytical application for iron ores and the related exploration samples. Out of all the evaluated CRMs, 16 were selected as iron ore calibration standards and 12 were selected as exploration calibration standards. Table 2 shows the compound concentrations for the two separate CRM sets as well as for the global application. The concentrations of the compounds of interest are presented in element form or oxide form depending on the market requirements. Table 2 also presents the composition of the selected control samples. The selected CRMs originated from the following list of producers: - Bureau of Analysed Samples Ltd (BAS) - CCRMP, CANMET Mining and Mineral Sciences Laboratories - China National Analysis Center for Iron and Steel - Dillinger Hütte Laboratory - European Committee for Iron and Steel Standardization (ECISS) - European Coal and Steel Community (ECSC) - Geological Institute for Chemical Minerals - Geostats Pty Ltd - Institut de Recherche de la Sidérurgie (IRSID) - Institute for Geology Ore Deposits, Petrography, Mineralogy and Geochemistry (IGEM) - National Institute of Standards & Technology (NIST) - Swedish Institute for Metals Research
8 Copyright JCPDS-International Centre for Diffraction Data 2014 ISSN Table 2. CRM sets compound concentration and control samples composition Iron Ore Exploration Global Control a Control a Control a Control a Compound Min. Max. Min. Max. Min. Max Fe (%) SiO 2 (%) Al 2 O 3 (%) TiO 2 (%) Mn (%) CaO (%) P (%) S (%) MgO (%) K 2 O (%) Na 2 O (%) V (%) N/A N/A Cr (%) N/A Co (%) N/A Ni (%) N/A Cu (%) N/A Zn (%) N/A As (%) N/A N/A N/A Sr (%) N/A N/A Zr (%) N/A N/A Ba (%) N/A N/A Pb (%) N/A N/A N/A a Control sample: One or more certified RM, not used in the calibration and having a composition within the calibration range for each element to be analyzed. When using only one validation certified RM, select a sample in the middle of the concentration ranges. When several validation CRMs are used, select samples covering high and low values.
9 Copyright JCPDS-International Centre for Diffraction Data 2014 ISSN RESULTS AND DISCUSSION -Calibration For the calibration of the WDXRF instrument and further evaluation of precision and accuracy of this borate fusion and XRF analytical method, selected CRMs were prepared in duplicates using an M4 TM gas Fluxer to verify the precision of the sample preparation over the wide range of composition that was covered. -Precision and accuracy For the precision evaluation, 12 glass disks were produced using both the M4 TM gas Fluxer and the TheOx electric Fluxer. All fusion positions of both instruments were used to produce the complete set of glass disks of the precision evaluation. The sample selected for the precision evaluation was a high iron magnetite known to be relatively difficult to prepare using fusion. The accuracy evaluation was conducted using four control samples which were CRMs that were not included in the calibration. Table 3. ISO precision test results Compound Concentration ISO d Limit XRF M4 TM TheOx (%) Fe SiO Al 2 O TiO Mn CaO P S <LLD N/A N/A N/A N/A MgO K 2 O Na 2 O N/A V Cr Co Ni Cu Zn As <LLD N/A N/A N/A N/A Sr N/A 0, Zr N/A 0, Ba Pb
10 Copyright JCPDS-International Centre for Diffraction Data 2014 ISSN It is important to note that the ISO limits for precision are not fixed. The ISO limits are variable as a function of the concentration of the element in the sample analyzed. The ISO precision test was applied as described in the method [3]. The standard deviation limit calculated for all the elements is shown in Table 3. The precision results of the spectrometer are presented as well as the precision of the M4 TM gas Fluxer and the TheOx electric Fluxer. These results were compared to the ISO precision limits. The values obtained for all the compounds met the specified limits. The ISO method refers to a trueness test for the accuracy evaluation. This test is complex even for experienced analysts. For this reason the accuracy results are presented in the following tables as the maximum deviation between certified values and results. Accuracy validation was examined on the 2 different calibration curves: a first calibration based on pure oxides fused into calibration glass disks acting as the reference methodology as described in the ISO test method and a second calibration based on CRMs fused into glass disks. This allowed evaluating the performance of the CRM based calibration against the reference methodology. The maximum deviation was calculated using two different preparations for the four control samples. Table 4. Control sample 1 accuracy evaluation JK 42 Certified Max. Dev. Max. Dev. Compound Values (%) (%) (%) Pure oxides CRMs Fe SiO Al 2 O TiO Mn CaO P S MgO K 2 O Na 2 O V Cr Co Ni Cu Zn As N/A N/A N/A Sr N/A N/A N/A Zr N/A N/A N/A Ba N/A N/A N/A Pb N/A N/A N/A LOI N/A N/A N/A Table 5. Control sample 2 accuracy evaluation GIOP-37 Certified Max. Dev. Max. Dev. Compound Values (%) (%) (%) Pure oxides CRMs Fe SiO Al 2 O TiO Mn CaO P S MgO K 2 O Na 2 O V Cr <LLD <LLD Co Ni Cu Zn As Sr N/A Zr N/A Ba Pb LOI
11 Copyright JCPDS-International Centre for Diffraction Data 2014 ISSN Table 6. Control sample 3 accuracy evaluation NIST SRM 692 Certified Max. Dev. Max. Dev. Compound Values (%) (%) (%) Pure oxides CRMs Fe SiO Al 2 O TiO Mn CaO P S MgO K 2 O Na 2 O V N/A N/A N/A Cr N/A N/A N/A Co N/A N/A N/A Ni N/A N/A N/A Cu N/A N/A N/A Zn N/A N/A N/A As N/A N/A N/A Sr N/A N/A N/A Zr N/A N/A N/A Ba N/A N/A N/A Pb N/A N/A N/A LOI N/A N/A N/A Table 7. Control sample 4 accuracy evaluation FER-2 Certified Max. Dev. Max. Dev. Compound Values (%) (%) (%) Pure oxides CRMs Fe SiO Al 2 O TiO Mn CaO P S MgO K 2 O Na 2 O V N/A N/A N/A Cr Co <LLD <LLD Ni Cu Zn As N/A N/A N/A Sr N/A Zr N/A Ba Pb N/A N/A N/A LOI N/A N/A N/A The accuracy results presented in the four previous tables demonstrate that both calibration strategies allow for similar accuracy levels for all the evaluated elements and over the different types of iron ore related material. The great benefit of using a CRM based calibration is the simplified calibration work as the CRMs are fused into glass disks in parallel of two simple operations: drying and ignition. The pure oxide calibration strategy as described in the ISO test method necessitates many steps in order to prepare all the mixes of the different oxide powders to produce the various calibration points (drying of all powders, ignition of some powders, weighing of all powders to produce all the mixes in the exact ratio, etc.). It is time consuming as well as expensive to buy all the oxide powders in analytical grade. It also opens the door to many manipulation errors that can bring bias in the analysis.
12 Copyright JCPDS-International Centre for Diffraction Data 2014 ISSN CONCLUSIONS To match the actual needs and expectations of the iron ore industry regarding the elemental characterization of iron ore materials and exploration minerals, a new borate fusion and XRF application was evaluated through this study. Consisting in a single method of preparation by fusion, it allows fusing the various iron ore types and the exploration samples normally found in this industry in a lithium borate glass disk. A CRM based calibration proved to provide the same level of precision and accuracy for iron ore samples when compared to the traditional pure oxide calibration as recommended in the ISO international test method. In addition to providing high analytical performance in line with the ISO analytical targets, the CRM based calibration allows for extended calibration ranges that cover both the iron ores and the exploration samples and also makes the analysis of Sodium (Na) possible and easier. Finally, this alternative CRM based application significantly simplifies the preparation of the calibration standard fused disks which allows saving time and minimize the work needed to prepare the application. It was also proven that both the M4 TM gas Fluxer and the TheOx electric Fluxer offer the same versatility and ability to fuse the full range of samples all while offering similar analytical performances. References 1. ANZELMO, J. A., The Role of XRF, Inter-Element Corrections, and Sample Preparation Effects in the 100-Year Evolution of ASTM Standard Test Method C114, Journal of ASTM International, Vol. 6, No. 2, Paper ID JAI101730, available online at , pp SPANGENBERG, J. and FONTBOTÉ, L., X-Ray Fluorescence Analysis of Base Metal Sulphide and Iron-Manganese Oxide Ore Samples in Fused Glass Disk, X-Ray Spectrometry, Vol. 23, 1994, pp ISO :2003 (First edition, ), Iron ores Determination of various elements by X-ray fluorescence spectrometry Part 1: Comprehensive procedure (ISO :2003), 72 pp. 4. BÉRUBÉ, L., RIVARD, S., ANZELMO, J. A., XRF Fusion Precision with TheAnt, International Cement Review, March, 2008, 4 pp. 5. LOUBSER, M., STRYDOM, C., and POTGIETER, H., A Thermogravimetric Analysis Study of Volatilization of Flux Mixtures Used in XRF Sample Preparation, X-Ray Spectrom. 2004; 33: , Published online January 29 th 2004 in Wiley InterScience ( DOI: /xrs.700
Global Cement and Raw Materials Fusion/XRF Analytical Solution: Part 2
246 247 Global Cement and Raw Materials Fusion/XRF Analytical Solution: Part 2 Mathieu Bouchard, John Anzelmo, and Sebastien Rivard Corporation Scientifique Claisse, Quebec, Canada Alexander Seyfarth and
More informationGlobal Cement and Raw Materials Fusion/XRF Analytical Solution
263 Global Cement and Raw Materials Fusion/XRF Analytical Solution Mathieu Bouchard, John Anzelmo, and Sebastien Rivard Corporation Scientifique Claisse, Quebec, Canada Alexander Seyfarth and Larry Arias
More informationRAW MATERIALS XRF APPLICATION USING A
RAW MATERIALS XRF APPLICATION USING A UNIVERSAL BORATE FUSION METHODOLOGY Anne-Catherine Breton, Mathieu Bouchard, John Anzelmo and Sébastien Rivard Corporation Scientifique Claisse, Québec, Canada Corporation
More informationXRF DRIFT MONITORS DATA CALIBRATION MATERIAL
UNIQUE PRODUCTS FROM ONE SOURCE XRF DRIFT MONITORS DATA CALIBRATION MATERIAL KEY FEATURES Monitor Composition The monitors are manufactured as stable fortified glass discs that are used to correct for
More informationXRF ANALYSIS OF HIGH GAIN-ON-IGNITION SAMPLES BY FUSION METHOD USING FUNDAMENTAL PARAMETER METHOD
Copyright JCPDS-International Centre for Diffraction Data 2013 ISSN 1097-0002 177 RF ANALYSIS OF HIGH GAIN-ON-IGNITION SAMPLES BY FUSION METHOD USING FUNDAMENTAL PARAMETER METHOD Mitsuru atanabe 1, Hisashi
More informationISO INTERNATIONAL STANDARD. Iron ores Determination of various elements by X-ray fluorescence spectrometry Part 1: Comprehensive procedure
INTERNATIONAL STANDARD ISO 9516-1 First edition 2003-04-01 Iron ores Determination of various elements by X-ray fluorescence spectrometry Part 1: Comprehensive procedure Minerais de fer Dosage de divers
More informationXRF S ROLE IN THE PRODUCTION OF MAGNESIUM METAL BY THE MAGNETHERMIC METHOD
Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol.44 398 XRF S ROLE IN THE PRODUCTION OF MAGNESIUM METAL BY THE MAGNETHERMIC METHOD H. L. Baker Northwest Alloys,
More informationS2 RANGER LE: Analysis of Light Elements in Cement, Slags and Feldspar
S2 RANGER LE: Analysis of Light Elements in Cement, Slags and Feldspar 1 Welcome Today s topics: S2 RANGER LE o o o Overview XFlash LE detector Improved X-ray tube Analysis of Light Elements in Cement
More informationTHE CLAISSE-CALCULATOR OF SAMPLES ACIDITY APPLICATIONS
71 THE CLAISSE-CALCULATOR OF SAMPLES ACIDITY APPLICATIONS Fernand Claisse Fernand CLAISSE inc 2270, Léon Harmel, Suite 165 Québec, QC, G1N 4L2 ABSTRACT The CLAISSE-CALCULATOR is a new tool for XRF analysts
More informationX-RAY FLUORESCENCE SPECTROMETRY: DEVELOPMENT OF EFFECTIVE FUSION PROCEDURES FOR LIME PRODUCTS
76 X-RAY FLUORESCENCE SPECTROMETRY: DEVELOPMENT OF EFFECTIVE FUSION PROCEDURES FOR LIME PRODUCTS Jean-Philippe Gagnon Corporation Scientifique Claisse, 35 rue Franquet, Québec, QC, G1P 4P3, CANADA ABSTRACT
More informationSPECTRAL INTERFERENCE IN X-RAY FLUORESCENCE ANALYSIS OF COMMON MATERIALS
Copyright JCPDS - International Centre for Diffraction Data 2003, Advances in X-ray Analysis, Volume 46. 38 ISSN 097-0002 SPECTRAL INTERFERENCE IN X-RAY FLUORESCENCE ANALYSIS OF COMMON MATERIALS Frank
More informationElvaX ProSpector in Exploration & Mining
ElvaX ProSpector in Exploration & Mining Introduction ElvaX ProSpector is a fast, accurate and easy tool for different mining applications. It provides onsite analysis of ore samples with minimal sample
More informationA QUICK AND RELIABLE FUSION METHOD FOR SILICON AND FERROSILICON
Copyright (c)jcpds-international Centre for Diffraction Data 2002, Advances in X-ray Analysis, Volume 45. 415 A QUICK AND RELIABLE FUSION METHOD FOR SILICON AND FERROSILICON Jimmy Blanchette Corporation
More informationCreating Value in Mining: How XRF Delivers Accuracy and Precision in Daily Routine
Creating Value in Mining: How XRF Delivers Accuracy and Precision in Daily Routine Presented by: Kai Behrens & Arkady Buman Bruker AXS November 19, 2012 1 Webinar: Creating Value in Mining Today s Speakers
More informationLithium based Borate Fusion of Gypsum/High Sulfate Samples: A New Approach.
868 Lithium based Borate Fusion of Gypsum/High Sulfate Samples: A New Approach. Abstract Donald J. Broton Construction Technology Laboratories, Inc Skokie, Illinois Manufacturing portland cement requires
More informationPURE & PRODUCTIVE FLUXES
GOOD REASONS FOR PURE & PRODUCTIVE FLUXES KEY FEATURES Quantifiable Pure Products Lithium Borate Internal Standard Fully Integrated Additives Prior Preparation Ready for Fusion Open the flux vial, place
More informationWavelength-dispersive X-ray fluorescence spectrometer
New products Wavelength-dispersive X-ray fluorescence spectrometer 1. Introduction The newly-released sequential general-purpose wavelength-dispersive X-ray fluorescence spectrometer, ZSX PrimusIII is
More informationS8 TIGER Series 2. Lab Report XRF 144. Innovation with Integrity. Accurate Quality Control of Tool Steels XRF
Lab Report XRF 144 S8 TIGER Series 2 Accurate Quality Control of Tool Steels Tool steels are iron-based alloys that are particularly well-suited to be made into tools. Their suitability comes from their
More informationSIMPLIFIED MATERIALS ANALYSIS VIA XRF. Ravi Yellepeddi and Didier Bonvin Thermo Fisher Scientific (Ecublens) SARL, Ecublens/Switzerland
SIMPLIFIED MATERIALS ANALYSIS VIA XRF Ravi Yellepeddi and Didier Bonvin Thermo Fisher Scientific (Ecublens) SARL, Ecublens/Switzerland SIMPLIFIED MATERIALS ANALYSIS VIA XRF Ravi Yellepeddi and Didier Bonvin
More informationDEVELOPMENT AND APPLICATION OF A THULIUM BASED INTERNAL STANDARD METHOD FOR HIGH ACCURACY NICKEL ANALYSIS
15 DEVELOPMENT AND APPLICATION OF A THULIUM BASED INTERNAL STANDARD METHOD FOR HIGH ACCURACY NICKEL ANALYSIS Shane J. Reeves 1, David Brown 2 and Jorg Metz 3 1 BHP Billiton, Nickel West, Leinster Nickel
More informationOPTIMIZING THE BALANCE OF QUALITY AND TURNAROUND TIME FOR A PROCESS CONTROL XRF LABORATORY
104 OPTIMIZING THE BALANCE OF QUALITY AND TURNAROUND TIME FOR A PROCESS CONTROL XRF LABORATORY ABSTRACT Suzanne W. Bowe Kennecott Utah Copper, Magna, Utah Kennecott Utah Copper, outside Salt Lake City,
More informationX-ray Fluorescence Spectrometry X-ray Fluorescence Spectrometry (XRF) is a non-destructive, quantitative technique for determining chemical
X-ray Fluorescence Spectrometry X-ray Fluorescence Spectrometry (XRF) is a non-destructive, quantitative technique for determining chemical composition. When a primary x-ray collides with an atom in the
More informationProcess and quality control in the aluminum industry using ARL 9900 XRF-XRD integrated spectrometer
Lecture Transcript: LT O4/05 Process and quality control in the aluminum industry using ARL 9900 XRF-XRD integrated spectrometer Authors: R. Yellepeddi, A. Kohler and D. Bonvin Thermo Electron SA, Ecublens/Switzerland
More informationWavelength Dispersive XRF Spectrometer
ISO9001 ISO14001 for Wafers, Media Disks, and Large Samples Wavelength Dispersive XRF Spectrometer ADVANCED XRF For Large and Irregularly Shaped Samples Is there any solution to analyze a large sample
More informationSample preparation for X-ray fluorescence analysis V. Fusion bead method part 2: practical applications
Technical articles Sample preparation for X-ray fluorescence analysis V. Fusion bead method part 2: practical applications Mitsuru Watanabe* 1. Introduction The general preparation method of fusion bead,
More informationAPPROACHING A UNIVERSAL SAMPLE PREPARATION METHOD FOR XRF ANALYSIS OF POWDER MATERIALS
Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol.44 368 APPROACHING A UNIVERSAL SAMPLE PREPARATION METHOD FOR XRF ANALYSIS OF POWDER MATERIALS Abstract John
More informationXRF ANALYSIS OF AUTOMOTIVE CATALYSTS BY FLUX/FUSION
Copyright (c)jcpds-international Centre for Diffraction Data 2002, Advances in X-ray Analysis, Volume 45. 472 XRF ANALYSIS OF AUTOMOTIVE CATALYSTS BY FLUX/FUSION A. R. Drews Physical and Environmental
More informationKeeping ahead through Claisse expertise in sample preparation by fusion. The standard in sample preparation by fusion
Keeping ahead through Claisse expertise in sample preparation by fusion The standard in sample preparation by fusion how TO reach efficiency with Claisse Expertise? Claisse offers a global solution in
More informationX-Ray-Fluorescence Analysis Major Eiements in Silicate Minerals
LA-94-MS UC- Issued: September 982 LA 94-MS DE83 39G8 X-Ray-Fluorescence Analysis Major Eiements in Silicate Minerals Roland C. Hagan DISCLAIMER oi 'est> ni> ' * o' ^ *..i v. iproifui-r. is/ e jr imply
More information--> Buy True-PDF --> Auto-delivered in 0~10 minutes. GB/T Translated English of Chinese Standard: GB/T
Translated English of Chinese Standard: GB/T36164-2018 www.chinesestandard.net Buy True-PDF Auto-delivery. Sales@ChineseStandard.net NATIONAL STANDARD OF THE PEOPLE S REPUBLIC OF CHINA GB ICS 77.080.20
More informationAbstract No. 63. Applications of Borate Fusion for X-Ray Fluorescence Analysis
Abstract No. 63 Applications of Borate Fusion for X-Ray Fluorescence Analysis Author: Dr. Rainer Schramm, FLUXANA GmbH & Co. KG, Borschelstr.3, 47551 Bedburg-Hau, Germany Tel. +49 2821 99732-0, Fax +49
More informationDevelopment and Application of a Thulium Based Internal Standard Method for High Accuracy Nickel Analysis
Development and Application of a Thulium Based Internal Standard Method for High Accuracy Nickel Analysis Shane J. Reeves a, David Brown b and Jorg Metz c a BHP Billiton, Nickel West, Leinster Nickel Operation,
More informationElvaX Light for Petrochemical analysis
ElvaX Light for Petrochemical analysis Introduction ElvaX Light is a powerful and productive instrument for analysis of petroleum products, such as automotive fuels (gasoline, diesel), lubricating oil,
More informationAnalytical Techniques for Grade and Quality Control in Coal Mining. Dr Paul O Meara XRD Application Specialist
Analytical Techniques for Grade and Quality Control in Coal Mining Dr Paul O Meara XRD Application Specialist Topics Pulsed Fast & Thermal Neutron Activation (PFTNA) Cross-belt Elemental Analysis of Coal
More informationKeeping ahead through Claisse expertise in sample preparation by fusion. Get ready for high productivity!
Keeping ahead through Claisse expertise in sample preparation by fusion Get ready for high productivity! how TO reach efficiency with Claisse Expertise? Claisse offers a global solution in sample preparation
More informationAnalysis of Cast Iron Using Shimadzu PDA-7000
Analysis of Cast Iron Using Shimadzu PDA-7000 C112-0510M The analysis of low and high alloy cast iron by optical emission spectrometry is presented. Cast iron alloys are classified by their mechanical
More informationSorting and Drying Code Price Unit. Sorting and Boxing of Samples, received as pulps SORTBOX 0.00 Sample
Sample Preparation Sorting and Drying Sorting and Boxing of Samples, received as pulps SORTBOX 0.00 Sample Sorting and Drying of Samples, prior to preparation SORTDRY 0.00 Sample Reporting of Sample Weight,
More informationHow to Increase Productivity in Mining with Automated Simultaneous XRF Analysis. Arkady Buman and Dr. Kai Behrens Bruker AXS Inc.
How to Increase Productivity in Mining with Automated Simultaneous XRF Analysis Arkady Buman and Dr. Kai Behrens Bruker AXS Inc. Today s Speakers Arkady Buman Product Manager XRF Madison, WI, USA Dr. Kai
More informationThe Application of X-Ray Fluorescence to Assess Proportions of Fresh Concrete
Civil, Construction and Environmental Engineering Civil, Construction and Environmental Engineering Conference Presentations and Proceedings 2012 The Application of X-Ray Fluorescence to Assess Proportions
More informationREVISIT OF FORENSIC ANALYSIS OF ARSENIC POISONING CASE 1998
Copyright JCPDS-International Centre for Diffraction Data 2014 ISSN 1097-0002 177 REVISIT OF FORENSIC ANALYSIS OF ARSENIC POISONING CASE 1998 Jun Kawai Department of Materials Science and Engineering,
More informationEDXRF APPLICATION NOTE IRON IN ALUMINUM # 1237
EDXRF APPLICATION NOTE IRON IN ALUMINUM # 1237 SCOPE The analysis of iron in aluminum extrussion billets is demonstrated. BACKGROUND Often, selection of a bauxite mine site is determined by the alumina,
More informationS2 PUMA for Analysis of Geological Materials. Innovation with Integrity. Precise and Reliable Grade Control of Geological Materials EDXRF
S2 PUMA for Analysis of Geological Materials Precise and Reliable Grade Control of Geological Materials Innovation with Integrity EDXRF Maximizing Your Process Yield and Improve Material Grades with S2
More informationEVALUATION OF FUNDAMENTAL PARAMETERS FOR QUANTITATIVE X-RAY FLUORESCENCE OF POLYOLEFINS. The Dow Chemical Company, Freeport, TX 77541
Copyright JCPDS-International Centre for Diffraction Data 2014 ISSN 1097-0002 116 EVALUATION OF FUNDAMENTAL PARAMETERS FOR QUANTITATIVE X-RAY FLUORESCENCE OF POLYOLEFINS Abstract: Authors: Burns, D.W.
More informationQuantitative Analysis of FeMo Alloys by X-Ray Fluorescence Spectrometry
American Journal of Analytical Chemistry, 2014, 5, 766-774 Published Online September 2014 in SciRes. http://www.scirp.org/journal/ajac http://dx.doi.org/10.4236/ajac.2014.512085 Quantitative Analysis
More informationThe preparation of fused samples in X-ray fluorescence analysis
MINERALOGICAL MAGAZINE, SEPTEMBER 1969, VOL. 37, NO. 287 The preparation of fused samples in X-ray fluorescence analysis P. R. HOOPER AND LYNNE ATK1NS Department of Geology, University College of Swansea,
More informationONLINE X-RAY ELEMENTAL ANALYSIS OF ORE AND OTHER BULKY MATERIALS
ONLINE X-RAY ELEMENTAL ANALYSIS OF ORE AND OTHER BULKY MATERIALS Dr. Albert Klein 1 Abstract Online elemental Analysis directly at the material stream or at a bypass is enabled by a modified X- ray fluorescence
More informationLoss On Ignition measurement for oxidizable iron ores with C / H analyzer
Progress in Analytical Chemistry & Materials characterisation in the Steel and Metals Industries 19-21/05/2015 2015 Loss On Ignition measurement for oxidizable iron ores with C / H analyzer L. DAHERON
More informationKeeping ahead through Claisse expertise in sample preparation by fusion. Let the rfusion system work for you!
Keeping ahead through Claisse expertise in sample preparation by fusion Let the rfusion system work for you! how TO reach efficiency with Claisse Expertise? Claisse offers a global solution in sample preparation
More informationAnalysis of Copper and Copper Base Alloys, Using Shimadzu PDA-7000
C112-0514M Analysis of Copper and Copper Base Alloys, Using Shimadzu PDA-7000 Copper alloys are designated by their chemical composition with characteristic properties such as high corrosion resistance,
More informationCenter for By-Products Utilization
Center for By-Products Utilization DRAFT Evaluation of Biron #5 Boiler Ash from Stora Enso North America By Tarun R. Naik and Rudolph N. Kraus REP-500 February 2003 Submitted to Mr. John Michalsen, Stora
More informationCertificate of Analysis
Certificate of Analysis First issued: June 2018 Version: June 2018 REE-3 Certified Reference Material for an Ore with Rare Earth Elements and Zirconium Table 1 REE-3 Certified Values note: Please refer
More informationON-STREAM XRF ANALYSIS OF HEAVY METALS AT PPM CONCENTRATIONS
Copyright JCPDS - International Centre for Diffraction Data 2004, Advances in X-ray Analysis, Volume 47. 130 ABSTRACT ON-STREAM XRF ANALYSIS OF HEAVY METALS AT PPM CONCENTRATIONS G Roach and J Tickner
More informationAssaying and Analysis Choices in Practice
Assaying and Analysis Choices in Practice Au assays at the GDL laboratories Dirk Hofmans Umicore UPMR Hoboken, Belgium March 2015 Agenda LBMA Good delivery rules for laboratories How to assay > 995 Gold?
More informationAnalysis of Pharmaceuticals by ICP-MS and WDXRF
Analysis of Pharmaceuticals by ICP-MS and WDXRF Frank X. Weber, James M. Harrington, Andrea McWilliams, James Medlin, Keith E. Levine Trace Inorganics Department, RTI International, Research Triangle Park,
More informationReduction of the Sample Size in the Analysis of Rock by EDXRF
Copyright (C) JCPDS International Centre for Diffraction Data 1999 873 Reduction of the Sample Size in the Analysis of Rock by EDXRF Riidiger Harmel, Ulrike Otto, Olaf Haupt, Clemens Sch$er and Walter
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 VALE CANADA LTD. COPPER CLIFF ANALYTICAL SERVICES 18 Rink Street Copper Cliff, Ontario P0M 1N0 Canada Mr. Claude Serre Phone: 705 682 7501 CHEMICAL Valid To:
More informationCore Analysis with the Tracer
BRUKER ELEMENTAL Core Analysis with the Tracer Prepared by: Lee Drake, Senior Application Scientist January 23, 2014 BRUKER ELEMENTAL Key Points OBJECTIVE 3 EXPLORATION 3 DECISION MAKING 3 RECLAMATION
More informationCertificate of Analysis First issued: May 2008 Version: May 2008
Certificate of Analysis First issued: May 2008 Version: May 2008 DS-1 Certified Reference Material for a Gold Ore Table 1 DS-1 Certified Values Within-lab Betweenlabs 95% Confidence Interval of Mean Ag
More informationFor personal use only
Venture Upgrades DSO Resource Base and Delivers a 4mt Maiden Ore Reserve. ASX Announcement Thursday, 26/07/12 Ref: /VMS/606/VMS00287 Australian mineral exploration company, Venture Minerals Limited (ASX
More informationGlass beads by borate fusion
XRF SAMPLE PREPARATION Glass beads by borate fusion James P. Willis The Analytical X-ray Company XRF SAMPLE PREPARATION Glass beads by borate fusions James P. Willis About James Willis Prof. Emeritus James
More informationCERTIFICATE OF ANALYSIS
CERTIFIED REFERENCE MATERIAL Prepared by and Distributed by MINTEK P/Bag X3015 Randburg 2125 Republic of South Africa Email : asdenquries@mintek.co.za Telephone : + 27 11 7094046 Fax : + 27 11 7094006
More informationPlatinum Care Extending the life of your Labware
Platinum Care Extending the life of your Labware Introduction The harsh environments that platinum labware is exposed to can significantly reduce the lifetime of the product and the accuracy of the measurement
More informationCement analysis using FP program (Fundamental Parameters)
Application note issue #5 Cement Analysis August 2017 Cement analysis using FP program (Fundamental Parameters) Cement is a blend of several minerals. It is critical to control the elemental composition
More informationCERTIFIED REFERENCE MATERIALS (CRM)
CERTIFIED REFERENCE MATERIALS (CRM) Sufficiently Homogenized Indian Reference Materials certified by Round robin analysis for the use of validation of methods and calibration. CSIR-National Metallurgical
More informationNew Advances using Handheld XRF Technology for the Prevention of Flow-Accelerated Corrosion (FAC)
New Advances using Handheld XRF Technology for the Prevention of Flow-Accelerated Corrosion (FAC) Joey Cheverie, Vice President Elemental Controls Ltd www.elementalcontrols.com 1 Agenda Flow-accelerated
More informationCertificate of Analysis
Brammer Standard Company, Inc. Certificate of Analysis BS 625A Certified Reference Material for Inconel 625 - UNS Number N06625 1 Certified Estimate of Certified Estimate of 2 3 2 Value Uncertainty Value
More informationTable I: MCL and MRL Concentrations for Contaminants Monitored Under the Safe Drinking Water Act National Primary Drinking Water Regulations
Application Note - AN1303 Water Analysis Following U.S. EPA Method 200.7 Using the Teledyne Leeman Lab s Prodigy7 ICP-OES Under the Safe Drinking Water Act (SDWA) and the Clean Water Act (CWA), the USEPA
More informationCertificate of Analysis BS 110B
Brammer Standard Company, Inc. Certificate of Analysis BS 110B 1 Certified Reference Material for Grade 11000 Electronic Tough Pitch Copper (UNS Number C11000) Certified Estimate of Certified Estimate
More informationCETAS Iron sinter process control using XRD. Uwe König 1 & Nicholas Norberg 1 1 PANalytical B.V., Almelo, Netherlands
Abstract No. 44 Iron sinter process control using XRD Uwe König 1 & Nicholas Norberg 1 1 PANalytical B.V., Almelo, Netherlands The use of high speed detectors made X-ray diffraction (XRD) become an important
More informationamazing performance in WDXRF Cement Slags Petroleum Chemicals Oxides Metals Glasses Ceramics Alternate fuels Food
Thermo Scientific ARL OPTIM X Series WDXRF Spectrometer amazing performance in WDXRF Cement Slags Petroleum Chemicals Oxides Metals Glasses Ceramics Alternate fuels Food amazing performance in WDXRF 200
More informationThermo Scientific X-ray product range. For chemical and phase analysis of solids, liquids and powders
X-ray product range For chemical and phase analysis of solids, liquids and powders Metals, cement, mining, petrochemicals, environment, electronics, geology, glass, polymers, forensics, materials science,
More informationThermo Scientific X-ray product range. For chemical and phase analysis of solids, liquids and powders
X-ray product range For chemical and phase analysis of solids, liquids and powders Metals, cement, mining, petrochemicals, environment, electronics, geology, glass, polymers, forensics, materials science,
More informationZETIUM. Petro edition
ZETIUM Petro edition Elemental innovation Continuous development, improved customer experience Scientifically-sound, benefits-driven innovations implemented in the Petro edition of the Zetium spectrometer
More informationSTUDY OF ELEMENT FINGERPRINTING IN GOLD DORÉ BY GLOW DISCHARGE MASS SPECTROMETRY
STUDY OF ELEMENT FINGERPRINTING IN GOLD DORÉ BY GLOW DISCHARGE MASS SPECTROMETRY MICHAEL W. HINDS, Ph.D. 1 ELEMENT FINGERPRINTING In 1994, John Whatling* used laser ablation inductively coupled mass spectrometry
More informationKEEPING AHEAD THROUGH CLAISSE EXPERTISE IN SAMPLE PREPARATION BY FUSION. A class of its own
KEEPING AHEAD THROUGH CLAISSE EXPERTISE IN SAMPLE PREPARATION BY FUSION A class of its own HOW TO REACH EFFICIENCY WITH CLAISSE EXPERTISE? Claisse offers a global solution in sample preparation by fusion
More informationCERTIFICATE OF ANALYSIS
Certified values MONGOLIA CENTRAL GEOLOGICAL LABORATORY CERTIFICATE OF ANALYSIS CERTIFIED REFERENCE MATERIAL TRHB RARE EARTH ORE CGL 126 No. Oxides/ Elements Unit CV 1 ± А 2 N 3 1. SiO 2 % m/m 71.38 0.39
More informationAdvanced Materials Analysis with Micro-XRF for SEM
Advanced Materials Analysis with Micro-XRF for SEM Birgit Hansen, Application Scientist EDS and Micro-XRF for SEM Stephan Boehm, Product Manager Micro-XRF for SEM Bruker Nano GmbH, Berlin, Germany Innovation
More informationSchedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK
2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK Metals and Minerals Division 2 Perry Road Witham Essex CM8 3TU Contact: Mrs C Hargreaves Tel: +44 (0)1376 536800 Fax: +44 (0)1376 520819 E-Mail:
More informationOn site heavy metal testing and monitoring using monochromatic XRF methods
On site heavy metal testing and monitoring using monochromatic XRF methods X-Ray Fluorescence Analysis hn o hn i 8/5/2017 XOS Confidential File 2 Counts (100 seconds) Limitation of XRF High scattering
More informationCertificate of Analysis First issued: March 2010 Version: March 2010
Certificate of Analysis First issued: March 2010 Version: March 2010 RTS-3a Certified Reference Material for a Sulphide Ore Mill Tailings Element Units Mean Table 1 RTS-3a Certified Values Within-lab Betweenlabs
More informationEvaluation of Energy-Dispersive X-Ray Fluorescence for Mobile Analysis of Hazardous Metals in Transuranic Waste
Copyright(C)JCPDS-International Centre for Diffraction Data 2000, Advances in X-ray Analysis, Vol.42 146 Copyright(C)JCPDS-International Centre for Diffraction Data 2000, Advances in X-ray Analysis, Vol.42
More informationUsing your Vulcan in combination with the extractor hood will ensure that every bead is produced under the same constant conditions!
The fusion technology is used to prepare samples to be analyzed with an X-ray fluorescence machine. It is also used to improve precision over pressed powder technology. Using your in combination with the
More informationENERGY-DISPERSIVE X-RAY FLUORESCENCE ANALYSIS OF MONO- AND POLYCRYSTALS OF SELENIDE SPINELS BY FUNDAMENTAL PARAMETER METHOD
322 ENERGY-DISPERSIVE X-RAY FLUORESCENCE ANALYSIS OF MONO- AND POLYCRYSTALS OF SELENIDE SPINELS BY FUNDAMENTAL PARAMETER METHOD ABSTRACT Rafa Sitko, Beata Zawisza, Ewa Malicka Institute of Chemistry, Silesian
More informationRIX3100. Fully Automated Sequential X-ray Spectrometer System. Product Information
The Rigaku Journal Vol. 14/ number 1/ 1997 Product Information Fully Automated Sequential X-ray Spectrometer System RIX3100 1. Introduction Wavelength dispersive X-ray fluorescence (WDXRF) analysis methods
More informationAPPLICATION OF A PORTABLE TXRF SPECTROMETER TO DETERMINE TRACE AMOUNTS OF TOXIC ELEMENTS
18 APPLICATION OF A PORTABLE TXRF SPECTROMETER TO DETERMINE TRACE AMOUNTS OF TOXIC ELEMENTS Shinsuke Kunimura 1 and Jun Kawai 2 1 Materials Fabrication Laboratory, RIKEN (The Institute of Physical and
More informationVULCAN - Fusion Technology
- Fusion Technology 6MA Tools: Fusion machine Flux Flux mixtures for specific applications Platinum ware Applications Sample storage Fusion technology is used to prepare samples to be analyzed with an
More informationThe Determination of Alloying Elements in High-Carbon Steel using Radially Viewed Inductively Coupled Plasma Optical Emission Spectrometry (ICP-OES)
Application Note - AN1504 The Determination of Alloying Elements in High-Carbon Steel using Radially Viewed Inductively Coupled Plasma Optical Emission Spectrometry (ICP-OES) Introduction Page 1 Carbon
More informationFor personal use only
Manganese Intercepted at Grenfell Project Bligh Resources Limited ACN 130 964 162 ASX: BGH ASX Release 24 December 2012 Suite 804 Level 8, 84 Pitt St Sydney New South Wales 2000 Tel: +61 2 9233 4677 Fax:
More informationComparison of Off-peak vs. MAN vs. Nth Point background measurements John Donovan,
Comparison of vs. vs. background measurements John Donovan, 09-03-2009 Comparison of methods The following is an attempt to compare three methods (, and ) for background determination using both a homogeneous
More informationX-ray diffraction (XRD) for grade control of iron ores
XRD APPLICATION NOTE X-ray diffraction (XRD) for grade control of iron ores Beneficiation techniques for iron ore are receiving worldwide attention due to the development of new deposits of lower grade
More informationPreconcentration of Environmental Waters by Agar for XRF Analysis
96 Preconcentration of Environmental Waters by Agar for XRF Analysis Kazuhiko NAKANO 1, 2), Kenta OKUBO 1), Kouichi TSUJI 1) 1 ) Department of Applied Chemistry, Graduate School of Engineering, Osaka City
More informationAn Evaluation of Portable X-Ray Fluorescence for Artifact Sourcing in the Field: Can a Handheld Device Differentiate Anatolian Obsidian Sources?
An Evaluation of Portable X-Ray Fluorescence for Artifact Sourcing in the Field: Can a Handheld Device Differentiate Anatolian Obsidian Sources? Ellery Frahm Departments of Anthropology and Geology & Geophysics
More informationCertificate of Analysis
Certificate of Analysis First issued: September 2010 Version: September 2010 GTS-2a Certified Reference Material for a Gold Ore Mill Tailings Table 1 GTS-2a Certified Values Within-lab Between-labs 95%
More information5th International Conference on Sustainable Solid Waste Management, Athens, June M. Contreras, M.J. Gázquez, J.P.
5th International Conference on Sustainable Solid Waste Management, Athens, 21 24 June 2017 M. Contreras, M.J. Gázquez, J.P. Bolívar Group of Radiations Physics and Environment (FRYMA) University of Huelva,
More informationThis section describes the chemicals, apparatus and experimental procedure used.
7 EXPERIMENTAL 7.1 Introduction This section describes the chemicals, apparatus and experimental procedure used. 7.2 Apparatus and Reagents 7.2.1 Reagents Table 7.1: Reagents used Name Purity Supplier
More informationFAYALITE SLAG MODIFIED STAINLESS STEEL AOD SLAG
FAYALITE SLAG MODIFIED STAINLESS STEEL AOD SLAG Shuigen HUANG, Muxing GUO, Peter Tom JONES, Bart BLANPAIN Department of Metallurgy and Materials Engineering, KU Leuven, 3001 Heverlee, Belgium shuigen.huang@mtm.kuleuven.be,
More informationAPPLICATIONS OF X-RAY STRESS MEASUREMENT FOR INTERFACE AREA OF Ni 3 Al SYSTEM INTERMETALLIC COMPOUND COATING
Copyright JCPDS - International Centre for Diffraction Data 2004, Advances in X-ray Analysis, Volume 47. 385 APPLICATIONS OF X-RAY STRESS MEASUREMENT FOR INTERFACE AREA OF Ni 3 Al SYSTEM INTERMETALLIC
More informationTHE POSSIBILITY OF USING HANDHELD XRF IN CEMENT APPLICATIONS
THE POSSIBILITY OF USING HANDHELD XRF IN CEMENT APPLICATIONS Dr. Michelle Cameron, Bruker Elemental, Kennewick, WA Denver X-ray Conference 2011 ABSTRACT Handheld XRF may become a useful and cost-saving
More informationCertificate of Analysis First issued: March 2008 Version: March 2008
Certificate of Analysis First issued: March 2008 Version: March 2008 CPB-2 Certified Reference Material for Lead Concentrate Table 1 CPB-2 Certified Values Within-lab Betweenlabs 95% Confidence Interval
More informationSecond Edition February 2003
This book is in constant evolution; updated topics are added as research progresses. The reader is strongly invited to fax us for more information or to contribute his own findings or tricks; any valuable
More information