Revision No. : 03 Page 1 of 14 1 Place where equipment to be supplied 2 Scope (Supply, Supply & Installation/Supply, Installation & Training) DMD, CPRI, Bengaluru Supply, Installation, commissioning and Training 3 ICP OES Coupled with Micro wave Digestion system. 1.Simultaneous ICP-OES with dual view mode (both axial and radial in a single run) suitable for elemental analysis in ppb level coupled with sample digestion system. 2.The equipment shall be suitable for multi elemental analysis of Paints, Chemicals, wear metals, petrochemicals, organics, inorganics, HFs, water, soil, fly ash, dust, air particulates as per National/International Standards
Revision No. : 03 Page 2 of 14 3.Capable to perform elemental analysis on both aqueous and non aqueous samples (matrices). 4.The supplier shall be responsible for supply of main equipment, Micro wave digestion system and for their integration, warranty,malfunction, and other complications, if any. 4 Detectors UV sensitive SC/CCD/CMOS for the wave length range 165-900nm or better. 5 Optical System 1.Instrument shall be equipped with high energy Echelle based Grating with a focal length of 300-500mm or better. 2.The system shall have dual view with axial and radial measured together in the same method and in a single run 3.Resolution - 0.009nm or better at 200nm
Revision No. : 03 Page 3 of 14 6 Wave length 1.Range - 165 to 900nm or better. 2.All elements must consists by at least 3 sensitive and 3 secondary lines to satisfy the requirement of wide range of matrices. 7 Gas system 1.Gas system shall have software control automatic mass flow controllers(mfc) for Plasma, Nebulizer and Auxilary gases to adjust the rate of flow automatically. 2.System shall have least Argon Consumption and shall be indicated. 8 Sample Introduction Kit Sample introduction kits includes, nebulizer, chamber, torch and tubes shall be suitable for both aqueous, non-aqueous, HFs, in-orgnics, Organics for all the matrices indicated in S.No 3.2.
Revision No. : 03 Page 4 of 14 9 Neubilizer & Pump Shall provide cyclonic spray chamber with high performance glass concentric neublizer and neublizer for organic samples and software controlled 3 to 5 channel peristaltic pump to minimize cross contamination with sprint valve/switching valve system for reduction in rinsing time and sample uptake time 10 Torch Alignment and Plasma view Shall have Axial/Vertical mounted torch and plasma, which includes quick release pre-aligned mounting block. 11 Plasma Source 1.Integral water cooled solid state 30-40 MHz RF power generator with a delivery output of 1500 watts or better having 10 watt or better increment. 2.Plasma ignition shall be computer-controlled, user determined and totally automated.
Revision No. : 03 Page 5 of 14 12 Micro wave Digestion system 1.Dual magnetron with delivery power capacity of 1600 watts or better. 2.System shall have contactless insitu temperature sensors for better accuracy to precisely measure the temperature of liquid inside all the vessels in a real time with an accuracy of ±1 0 C. 3. The system shall have provision for 8 or more slots to load the samples 8 or more 4.The system shall offer 8 or more high pressure HF Resistant vessels with 100 ml capacity. Vessels shall with stand up to 250 0 C or more with a working pressure of 100 bar or more. 8 or more
Revision No. : 03 Page 6 of 14 4. Dual IR temperature sensors (Accuracy ±1 0 C) with dual controller (for better accuracy) to be provided with a provision for uniform temperature sensing of all the vessels(regardless of the sample quantity in vessels). 5.Magnetron protection with safety interlocks to be provided. 6.Automatic adjustment of Microwave power out put to be provided. 7.Built in touch screen with method library for all types of matrices for micro wave digestive system 15and main unit.. 13 Software 1.Operating System with latest windows to be provided.
Revision No. : 03 Page 7 of 14 2. All instrument functions like Plasma ignition, gas flow, position viewing, safety interlocks, monitoring etc., to be controlled through software. 3. Built-in in touch screen, preset methods for selection of matrices type, wave length library ( with an indication for selecting sensitive or secondary lines, mode of analysis - Axial or Radial or Dual view, for faster and easier operation to be provided.) 4. Software should have provision for auto and manual peak selection and contain method library 5.Facility for online back ground corrections and provision for auto and manual spectral interface corrections. 6.In built software validation for main equipment and Micro wave digestive system.
Revision No. : 03 Page 8 of 14 14 Accessories to be supplied 1.Multi element standards for Aqueous and organic medium of 1000ppm along with plain matrix (for dilution to different concentrations) with a combination of Pb, Hg, Cd, Sb, S, Si, Ag, Cu, Fe, Ca, Ba, Al, Cd, Mg, Mn, Hg, As, Na, P, Ni, Zn, etc., tracable to NIST with suitable matrix for ICP- OES with maximum shelf-life provided with certificate of analysis. 2.Suitable noise free air compressor (if required) 3.Suitable noise free chiller for rapid cooling (if required) each 4.Sampling tubes (Additional) 10 Mtrs 5.Torch (Additional) 6.Injector - (Additional)
Revision No. : 03 Page 9 of 14 7.Suitable PC with preloaded software and printer(configuration and model to be specified) 8.Suitable Hydride Generation System. 15 Indegenous Accessories to be supplied 1.15kVA UPS with 30 minutes backup 2.Filled Argon (4 No's, Purity - 99.999%) and Nitrogen gas cylinders (2 No's, Purity-99.999%) with a capacity of 6.3 cubic meters provided with 2 stage regulator ( each, SS) with purification panels and manifold Argon- 4No's N2-2No's 3.SS Fume hood both for ICP and Micro wave digestive system) 16 Good Laboratory Practice 1. User right restriction and administrative log in.
Revision No. : 03 Page 10 of 14 2.Good laboratory practice method for Documentation, Validation and Service/ calibration reminder facility including instrument serial number & batch number 3.LIMS enabled system 17 Demonstration The vendor shall demonstrate the total set up of the Model quoted for functional suitability at CPRI, Benglauru during the technical evaluation, if required 18 Power supply Power supply requirement: 230V AC, single phase 19 Operating condition The equipment should effectively function from ambient temperature to 35 C & RH between 10-80 % non-codensing
Revision No. : 03 Page 11 of 14 20 Warranty Warranty for the ICP-OES, Micro digestive system and Vessels one year from the date of successful installation & commissioning of equipment including for micro wave digester Vessels. 21 Instruction Manual 2 No's. of Instruction manual for both main instrument, and Sample digestive system should be provided with each equipment 2 No's 22 Training Training for 2 Officers on operation & maintenance of equipment "at CPRI, Bangalore" 23 Pre installation requirement Pre-installation requirements have to be mentioned along with the offer.
Revision No. : 03 Page 12 of 14 24 Compliance of specification Compliance of the specification shall be indicated unambiguously in the column provided above and shall be supported by documentary evidence such as product catalogues (Original)/ application notes. 25 Deviations Deviations if any from the specifications given above which provides improvement in the functioning of the above item/equipment shall be clearly brought out with their advantages. 26 Additional features Any additional features other than the above may be brought out clearly. 27 Customer list A list of customers (with details like name of contact person, address, phone No.) to whom the similar system was supplied shall be furnished along with performance certificate
Revision No. : 03 Page 13 of 14 28 Presentation The vendor shall make detailed presentation of the system offered, at CPRI Bangalore, during tender evaluation, if required. 29 Pre dispatch Inspection Required 30 Calibration Certificates Instrument and other accessories shall be supplied with Calibration certificate 31 Supply of spares The vendor shall commit for supply of spares for at least 7 years as and when required. 32 Safety 1.Shall have all safety provisions for complete system as per National/International standards 2.Shock proof & safety measurements in accordance to relevant national/international standards
Revision No. : 03 Page 14 of 14 3. Shall have saftey features to reduce the volaties in to the environment. 33 Company Status Orginal manufacturing company shall have ISO:9001 or Equivalent certification. PN: 1) Bidder shall indicate complete details/information how the CPRI GTP are complied with against each and every specification parameters & mere statement of 'complied' do not suffice the requirement. 2)A detailed technical/catalogue/literature/phamplet and any other details shall be submitted in hard copy in a sealed cover superscribing enquiry number and due date so as to reach the following address within the due date and time. 3) The Indian representative of the foreign/overseas firms shall submit the Letter of Authorization issued by their principals. 4)The bidder shall submit the quotation / offer in the above prescribed format of Section IA, IB and II only. Informations any other forms shall shall be rejected. Address : Joint Director (Purchase), Purchase Authrity, Central Power Research Institute, Sadashiva Nagar P.O., P.Box.No.8066, Bengaluru-560 80. Telefax: 080-2207 2010., email:purchase@cpri.in, web : www.cpri.in