EXPERIMENT 5. Molecular Absorption Spectroscopy: Determination of Iron with 1,10-Phenanthroline

Similar documents
Rev Experiment 10

SPECTROPHOTOMETRIC DETERMINATION OF IRON

IRON (Colorimetric) 2. Muffle Furnace: Equipped with pyrometer and capable of operating at controlled temperatures up to 600 C

Experiment 18 Determination of Iron by Visible Spectrophotometry

Chem 2115 Experiment #9. Consumer Chemistry: Determining the Iron Content in Supplements

Measuring Manganese Concentration Using Spectrophotometry

Technical Procedure for Ultraviolet Spectroscopy

Determination of the Mass Percentage of Copper in a Penny

An Oxidation-Reduction Titration: The Reaction of Fe 2+ and Ce 4+

SIMULTANEOUS SPECTROPHOTOMETRIC DETERMINATION OF IRON (II) AND TOTAL IRON USING FLOW INJECTION ANALYSIS

Determination of Ferrous Iron in EASTMAN Color Films, KUL Bleach

Determination of Iron Content in Different Hemoglobin Samples from Some Patients by UV-Visible Spectrophotometer

DR/4000 PROCEDURE. IRON, Ferrous. Using Powder Pillows

Scope and application: For water, wastewater, seawater, brine solutions, produced waters and hydraulic fracturing waters.

Spectrophotometry of DNA and RNA

Effects of Gamma Radiation on Iron (III) Complex with Sodium Salicylate in Aqueous Solutions

DR/4000 PROCEDURE. OXYGEN DEMAND, Chemical. Colorimetric Measurement, 0 to 40 mg/l COD

Write your own lab Spectrophotometry/Colorimetry

Standard Methods for the Examination of Water and Wastewater

Polyvidone Polyvinylpyrrolidone H 2 C H C N

SECTION D.3 CHLOROPHYLL AND PHEOPHYTIN

Materials Required for Lab Report Resubmissions: Original graded report with all required components

Spectroscopy Solutions for Pharmaceuticals:

SUMMARY OF STANDARDS FOR WATER ANALYSIS ON COURSE WAT-E2120 SPRING 2017

1,10 Phenanthroline Method * (Powder Pillows or AccuVac Ampuls) Using Powder Pillows

1-(2-Pyridylazo)-2-Naphthol PAN Method 1 Method to mg/l Mn (LR) Powder Pillows

Determination of Aluminum by Fluorometry

CHM Gravimetric Chloride Experiment (r7) 1/5

Experiment 11. DNA Composition by HPLC 1,2

Application Notes for COD Analysis DETERMINATION OF CHEMICAL OXYGEN DEMAND (COD) IN WATER AND WASTE WATER.

DR/4000 PROCEDURE. OXYGEN DEMAND, Chemical. Colorimetric Measurement

The Importance of Water Absorbance/Transmittance on the Efficiency of Ultraviolet Disinfection Reactors

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

Method 8017 (0.7 to 80.0 µg/l) Powder Pillows. Scope and Application: For water and wastewater; digestion is required to determine total cadmium.

COPPER CYCLE EXPERIMENT 3

CHM111 Lab Redox Titration Grading Rubric

Approved for NPDES (Editorial Revision 1978) Silica, Dissolved (Colorimetric)

Measurement of Enzyme Kinetics by UV-Visible Spectroscopy

OBSERVATIONS OF AN OPERATOR IN THE LAB IMPROVING LABORATORY RESULTS. Shane Jordan-Hill. Merck Millipore

Final Report. Characterisation of Sample Report. Job No 2016/11/12-34 AS No. 1234A. Client Example Contact Sample. Signed. Example report.

Experiment 8. Determination of Iron in an Ore by Potentiometric Titration. Iron ores are often completely decomposed in hot concentrated HCl.

Tap Water Analysis by Atomic Absorption and Ion Chromatography Michelle Youn Chem. 31: Inorganic Quantitative Analysis Section 6

LABORATORY 10 SOIL FERTILITY. Objectives

Experiment #8. Redox Titration

Gravimetric Analysis: Determination of % Sulfur in Fertilizer

Differences in Free Energy of Solvation in a Two-Layer System

INTRODUCTION TO ELECTROCHEMISTRY: CURRENT, VOLTAGE, & BATTERIES. Introduction. Electrochemistry Revised 4/28/14

Skoog, Holler and Nieman, Principles of Instrumental Analysis, 5th edition, Saunders College Publishing, Fort Worth, TX 1998, Ch 33.

A facile and fast method for quantitating lignin in lignocellulosic biomass using acidic

Enzymes and Enzymatic Reactions Adapted with permission from the original author, Charles Hoyt

Applications of Oxidation/Reduction Titrations. Lecture 6

Demonstration of osmosis

Seamless Method Transfer between the Agilent Cary 8454 UV- Visible Spectrophotometer and the 8453 UV-Visible Spectrophotometer

Optical spectroscopy of sodium silicate glasses prepared with nano- and micro-sized iron oxide particles

Orono Spectral Solutions, Inc.

USEPA 1 Periodate Oxidation Method 2 Method to 20.0 mg/l Mn (HR) Powder Pillows

LAD B3 (pg! 1 of 6! ) Analysis by Redox Titration Name Per

PREPARATION & ANALYSIS OF AN IRON COORDINATION COMPOUND PART A: PREPARATION OF AN IRON COORDINATION COMPOUND

Dimethylphenol Method Method LR (0.23 to mg/l NO 3 N or 1.00 to mg/l NO 3 ) TNTplus 835

Clinical Chemistry (CHE221) Professor Hicks Lecture 9

Experiment 13: Determination of Molecular Weight by Freezing Point Depression

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856

Spectrophotometric Estimation of Cefprozil Sulphate in Pharmaceutical Formulations

Determination of sodium, calcium, copper, iron and manganese in pulp and paper by atomic absorption spectroscopy (Five-year review of T 266 om-11)

Standard Test Methods for Rubber from Natural Sources Chemical Analysis 1

OXIDATION-REDUCTION TITRATIONS-Permanganometry

T 619 cm-84 TENTATIVE STANDARD 1933 OFFICIAL STANDARD 1935 CORRECTED 1944 CORRECTED 1953 CLASSICAL METHOD TAPPI. Analysis of salt cake

EXTRA CREDIT - EXPERIMENT G ELECTROCHEMISTRY ACTIVITY OF METALS

DIDANOSINE ORAL POWDER Final text for addition to The International Pharmacopoeia

Color, True and Apparent, LR

Calcium and Magnesium; Chlorophosphonazo Rapid Liquid Method Method to 1000 µg/l Ca and Mg as CaCO 3 (ULR) Pour-Thru Cell

Experiment 1: DNA Isolation from Plant and Animal Tissue and Gel Electrophoresis. Amber Broadbooks, Adrienne Martinez, Quenton Sayles

Analysis of dissolved nitrite (NO 2- )

(aq) + 5e - Mn 2+ (aq) + 4H 2

Spectrophotometric Determination of Phosphates in Water

Lab #7: Redox Titration Lab Exercise Chemistry II 10 points Partner: USE BLUE/BLACK INK!!!! Date: Hour:

Comparative methods applied for the determination of total iron from beer samples

METHOD #: Approved for NPDES and SDWA (Issued 1974) Mercury (Automated Cold Vapor Technique) ANALYTE: CAS # Hg Mercury

Instrument Adapters Light shield DR 6000, DR 5000 DR 3900 LZV849 DR 3800, DR 2800 LZV646 DR or (A)

Analytical Method of Limit Test for Hexachlorobenzene. in Picloram TC and Method Validation Data*

Oxygen Demand, Chemical

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

Method (0.1 to 8.0 mg/l Cu) TNTplus 860

The Synthesis of Copper Metal

National Food Safety Standard

SOP_Chem_001 Chemical Analysis_ Chemical Oxygen Demand. Standard Operation Procedure Chemical Oxygen Demand Closed Reflux, Titrimetric Method

The following are the completed but unbalanced equations. Each equation is numbered to match each step of the cycle:

Human CRP Precoated ELISA kit

GOOD LABORATORY PRACTICE Agilent Cary 8454 UV-Visible Spectroscopy System

Background The document will be revised for its five year review and updated per the committee agreed upon changes.

APPLICATION NOTE. Application Note. Summary/Abstract:

Oxygen Demand, Chemical

Method 108C - Determination of Arsenic Content in Ore Samples From Nonferrous Smelters (Molybdenum Blue Photometric Procedure)

Oxygen Demand, Chemical

ITS Sequencing in Millepora. 10/09 Subcloning DNA Fragments into pbluescript Preparation of pbluescript Vector

GB (GB/T )

EXPERIMENT 5 Chemistry 110 COMPOSITION OF A MIXTURE

30th International Chemistry Olympiad

Water Analysis Kit for Home Brewers

Lab 1: Tools for Biologists

Transcription:

EXPERIMENT 5 Molecular Absorption Spectroscopy: Determination of Iron with 1,10-Phenanthroline UNKNOWN Submit a clean, labeled 100-mL volumetric flask to the instructor so that your unknown iron solution can be issued. Your name, section number, and locker number should be written legibly on this flask. The flask does not need to be dry on the inside, but needs to have been rinsed with distilled water after it has been washed. The flask must be turned in at least 1 lab period before you plan to do the experiment so that the Teaching Assistants will have time to prepare the unknown. Each student will have his or her own unknown to analyze even if you are working in pairs. As a backup precaution, bring an MS-DOS formatted floppy disk, or other data-storage medium, so that you can copy the files obtained during the experiment. BACKGROUND 1,10-phenanthroline (C 12 H 8 N 2, ortho-phenanthroline or o-phen) is a tricyclic nitrogen heterocyclic compound that reacts with metals such as iron, nickel, ruthenium, and silver to form strongly colored complexes. This property provides an excellent and sensitive method for determining these metal ions in aqueous solution. For example, o-phen reacts with ferrous ion to produce a deeply colored red complex: Fe 2+ + 3 o-phen Fe(o-Phen) 3 2+ The molar absorptivity (ε) of the ferrous complex, [(C 12 H 8 N 2 ) 3 Fe] 2+, is 11,100 L/mol-cm at the wavelength of maximum absorbance intensity, λ max = 508 nm. This large value indicates the complex absorbs very strongly. The intensity of the color is independent of ph in the range 2 to 9. The complex is very stable and the color intensity does not change appreciably long periods of time. Beer s law is obeyed, over about 1.5-2 orders of magnitude of iron concentration. Beer s Law is a very simple relationship: A = εbc where A is the absorbance of a substance at a specified wavelength λ, in units of nm for light in the ultraviolet and visible regions of the electromagnetic spectrum; b is the length of the light path through the sample, usually in cm; ε is the molar absorptivity of the absorbing species at λ, when the concentration is in M; and c is the concentration in molar units, M. CHE 226 Analytical Chemistry Laboratory 30 Iron Absorption

The critical point of Beer s Law for use in quantitative analysis is that, theoretically, the absorbance of a solution is linear with the concentration of the absorbing substance, if the wavelength, the pathlength, and other experimental conditions are kept constant. There are a number of factors, however, that limit the concentration range over which Beer s Law is valid for a particular analysis. Perhaps the most critical is concentration. There are invariably deviations from Beer s law at high enough concentrations. To determine the total iron in the sample, it must be completely in the ferrous state, and Fe 2+ can readily be air-oxidized to the ferric state, Fe 3+. o-phen will form a colored complex with Fe 3+, but its spectrum is different from that of the ferrous complex and the color is not as intense. Thus, one could not determine the total iron present by making measurements at only one wavelength. Hence, a mild reducing agent is added before the color is developed in order to provide a measure of the total Fe present in solution. Hydroxylamine, as its hydrochloride salt, can be used. The reaction is 2 Fe 3+ + 2 NH 2 OH HCl + 2 OH - 2 Fe 2+ + N 2 + 4 H 2 O + H + + Cl - This is an equilibrium reaction, but even a mild reducing agent will drive it >99.99% to the right. The ph is adjusted to a value between 6 and 9 by addition of an ammonia or sodium acetate buffer. INSTRUMENTATION Varian Cary 50 UV-Vis Spectrophotometer. This is a computer-controlled double-beam grating spectrophotometer with a constant 20- nm bandpass. A high-intensity Xe flash lamp is used as the source for both UV and visible light, which permits taking 80 data per second. PREPARATION OF SOLUTIONS Stock Iron Standard Solution, 10 ppm Primary standard solid ferrous ammonium sulfate hexahydrate, (NH 4 ) 2 (SO 4 ) 2 6H 2O, 392.13 g/mol, is available on the side shelves for preparation of the standard iron solution. 1. Tap a small amount of the solid ferrous ammonium sulfate onto a sheet of glassine weighing paper that has been folded in the middle. Zero your balance. Accurately weigh about 0.07 g of pure dry ferrous ammonium sulfate (to +0.1 mg) onto a folded sheet of glassine paper or into a small, plastic weighing boat. 2. Transfer the ferrous ammonium sulfate quantitatively into a 1-L volumetric flask, carefully squirting down the weighing boat and the neck of the flask to ensure a quantitative transfer. Add about 100-200 ml of distilled water. Dissolve the solid completely before diluting to volume. This is critical. CHE 226 Analytical Chemistry Laboratory 31 Iron Absorption

3. Pipet 2.5 ml of concentrated sulfuric acid into the flask, rinse the neck of the flask down, and mix carefully with swirling. [Be very careful when using concentrated H 2 SO 4 ; it is quite caustic.] Dilute the solution to the mark. Calculate the iron concentration of the solution in µg of iron per ml (ppm) and in molar (M) units. Because this solution is used to calibrate absorbances and prepare a calibration curve, it must be prepared very carefully and accurately. The results of the entire experiment rest on preparing this solution accurately. The iron solution must be prepared daily, so there is no point in saving the solution to re-use it if you end up needing to re-do the experiment. You will need to prepare another standard solution. Iron Standard Calibration Solutions 1. Into each of five 100-mL volumetric flasks, pipet 1, 5, 10, 20, and 35 ml of the standard iron solution, respectively. Use a combination of 1-, 5-, 10-, and 25-mL volumetric pipets. The 1- and 5-mL pipets are located in the drawer marked for the experiment. 2. Pour about 50 ml of distilled water into a 6 th flask to serve as the blank (i.e. zero iron concentration). 3. Obtain the unknown sample from the Teaching assistants and treat it in the same manner as the standards, as indicated below. 4. Line all seven 100-mL volumetric flasks in this order: The blank, those with 1-35 ml of iron stock standard solution added, and your unknown sample. To each flask (including the distilled water blank and the unknowns), pipet in order a. 1 ml of the hydroxylamine solution, b. 10 ml of the 1,10-phenanthroline solution, and c. 8 ml of the sodium acetate solution. Note that the blank solution must have all the reagents in it except for any ferrous ammonium sulfate. 5. Swirl each flask to mix the contents, then carefully dilute each solution to the 100-mL mark and mix thoroughly. 6. Allow the solutions to stand for 10 minutes to fully develop the color. Mix well again. Fill each of seven clean, dry plastic cuvettes about two-thirds full with each of the seven solutions, keeping them in the same order. (If the insides of the cuvettes are wet or spotted, rinse them out twice with the appropriate solution first.) Stock Reagent Solutions The sodium acetate buffer (1.2 M), the 1,10-phenanthroline solution (1 g/l), and the hydroxylamine hydrochloride solution (100 g/l) are prepared by the Teaching Assistants and should be available for your use. CHE 226 Analytical Chemistry Laboratory 32 Iron Absorption

PROCEDURE Determining the Maximum Absorbance of the Complex on the Cary 50 1. On the Cary WinUV window, select SCAN. 2. Click SET UP to specify method parameters. a. Set the appropriate mode for the scan in the X direction (nanometers). b. Set the wavelength range for the scan by entering 600 nm and 300 nm in the START and STOP fields. c. In the Y field select ABS for absorbance. d. Enter 0 and 1.0 for the Y min and Y max fields. 3. Under SET UP select OPTIONS. a. Set the beam mode - Select DUAL. b. Select MEDIUM for scan control. 4. Select REPORTS. a. Enter your name in the NAME field. b. Enter the sample name in the COMMENTS field. c. Select MAX PEAK to report the maxima for individual peaks. 5. Insert the blank sample and click ZERO. 6. Remove the blank, select one of the iron solutions, and click SCAN. 7. When the scan is completed, print 1 copy of the screen per student. Your lab report should include a printout of the spectrum obtained. 8. As a backup precaution, copy the spectrum file onto a floppy disk or other storage medium: Click FILE, select SAVE AS, and save on drive a: with an identifying filename. Determining the Absorbances of the Test Solutions 1. After scanning the spectrum, select the wavelength of maximum absorption (λ max )to use for the determination of iron with 1,10-phenanthroline. 2. To measure individual standards and your unknown sample, select SIMPLE READ in the Cary WinUV menu. 3. Click SET UP and select READ at WAVELENGTH and enter the wavelength of maximum absorption. CHE 226 Analytical Chemistry Laboratory 33 Iron Absorption

4. In the Y mode select ABS and click OK. 5. Insert the blank sample and click ZERO. 6. Measure the absorbance of each of the standard solutions and the unknowns by clicking READ. 7. Plot absorbance vs. the concentration of the standards. 8. Note whether Beer s law is obeyed. From the absorbance of the unknown solution, calculate the concentration of iron in the unknown in units of µg/ml and in M units. LABORATORY REPORT In addition to other items commonly required in a laboratory report, calculate the concentration of your unknown both (1) by reading the concentration directly off the calibration curve and (2) by using the average of the molar absorptivities (actually εb) calculated from each of the five standard solutions. If the plot is linear the values should be the same, and the slope of the absorbance vs. concentration (in M) plot should be equal to εb according to Beer s Law: A = εbc. Assuming the pathlength of the cuvette used is 1.00 cm, calculate the molar absorptivity, ε, of the ferrous o-phen complex and compare with the literature value of 11,100 L/mol-cm. Comment on the reasons for any difference. HAZARDOUS WASTE DISPOSAL Empty ALL the solutions and left over reagents into the proper Hazardous Waste Bottle for this experiment. This includes the blank, unknown, and standard solutions; the remaining stock iron standard solution; small amounts of concentrated sulfuric acid, hydroxylamine hydrochloride, and ortho-phenanthroline solutions that may be left over; and any solid ferrous ammonium sulfate that may be left over from your experiment. If you are unsure of the proper container, ASK. TEXT REFERENCE D. A. Skoog, D. M. West, F. J. Holler, and S. R. Crouch, Analytical Chemistry: An Introduction, 7th ed., Chapters 21 and 22, pp. 547-592. Revised September 30, 2005 Copyright by The Department of Chemistry,, 2005 CHE 226 Analytical Chemistry Laboratory 34 Iron Absorption