Mass Flow Controllers and Carbon Dioxide. - Their Effects on Quantitation

Similar documents
Calibration Standards for Measurement of Toxic Organics in Biogas and Biomethane Using Permeation Tubes

Optimum Parameters for Detection of Long- Lived Green House Gases in Ambient Air

Tap Water Analysis by Atomic Absorption and Ion Chromatography Michelle Youn Chem. 31: Inorganic Quantitative Analysis Section 6

Analytical Method of Limit Test for Hexachlorobenzene. in Picloram TC and Method Validation Data*

June 8, Submitted to: The Environmental Protection Agency Research Triangle Park, NC. Prepared by:

APPLICATION NOTE. UOP : Analysis of Trace CO and CO 2 in bulk H 2 and Light Gaseous Hydrocarbons by GC

PRIMER ON TEMPERATURE CONTROL AND MONITORING IN GAS CHROMATOGRAPHY

List of Acronyms and Abbreviations

TECHNICAL DATA SHEET

DFE. Dynamic Filter Efficiency. hyprofiltration.com/dfe

SCR Control Strategy and CEMS for an Ultra-Low NO x Emitter

Background Statement for Ratification Ballot of SEMI Draft Document #5765 Revision to SEMI C TEST METHOD FOR ppm AND ppb HUMIDITY STANDARDS

A Simple Rapid and Sensitive Method Development for Quantification of Quetiapine Fumarate in Bulk and Dosage Forms Using RP-HPLC

DFE... The Evolution in Filter Element Multi-Pass Development Testing Hy-Pro s Competitive Advantage

STANDARD OPERATING PROCEDURES

LD8000. DESIGN REPORT v2. Trace Nitrogen in Argon or Helium analyzer

10. Validated Normal Phase HPLC Method for the Determination. Fulvestrant is primarily used in the treatment of hormone receptor

Background Statement for SEMI Draft Document #5765 Revision to SEMI C TEST METHOD FOR ppm AND ppb HUMIDITY STANDARDS

Fusion Analytical Method Validation

VALIDATION OF ANALYTICAL PROCEDURES: METHODOLOGY *)

REPORT ON THE BENEFITS OF ICE- BASED THERMAL STORAGE FOR DISTRICT COOLING IN THE UAE MARKETS

Determination of the Ethylene Oxide Control Efficiency from Catalytic Oxidizers. A Guidance Document

SIMULTANEOUS SPECTROPHOTOMETRIC DETERMINATION OF IRON (II) AND TOTAL IRON USING FLOW INJECTION ANALYSIS

of Methane, Ethylene, Acetylene and Ethane in Water by Headspace-Gas Chromatography (HS-GC) with

Trace Moisture Contamination in Ultra-High Purity Phosphine: Techniques for Measurement and Control

CX-1 ph Gradient Buffer

CORESTA Recommended Method No. 57

Automating HPLC and GC Analytical Method Validation Fusion AE Software Program White Paper

DETERMINATION OF METHOD UNCERTAINTY A PRACTICAL EXAMPLE FOR THE ANALYSIS OF TRIHALOMETHANES IN DRINKING WATER USING PURGE&TRAP GC-MS

Final DRAFT API TECHNICAL REPORT. Carbon Content, Sampling, & Calculation

Verification of Method XXXXXX

Simplicity. Accuracy. Durability. Lotix. Automated High Temperature Combustion TOC Analyzer

Review of ConocoPhillips Alaska Inc. Continuous Emissions Monitoring Data from a 2014 Drill Rig Study at Kuparuk Drill Site 2N

477 kcmil, 3M Brand Composite Conductor Mechanical Properties, Volume 1 Tensile and Stress-Strain Tests

APPENDIX III SAMPLE LAB REPORT. Experiment 1. High Performance Liquid Chromatography. Alfred E. Neuman

VICH Topic GL2 (Validation: Methodology) GUIDELINE ON VALIDATION OF ANALYTICAL PROCEDURES: METHODOLOGY

Characterization of IgG monomers & their aggregates

Protocol for Quantitative Determination of Residual Solvents in Cannabis Concentrates Prepared by: Amanda Rigdon, May 23 rd, 2016

Simultaneous Oxygen, Nitrogen and Hydrogen Determination of Metals. Dennis Lawrenz, John Hawkins

Atmospheric Emissions from Stationary Combustion Turbines

Title: Standard Operating Procedure for Measurement of CO in Ambient Air by Gas Filter Correlation (GFC)

easurement of trace hydrofluorocarbons, perfluoroent air. Details are provided for the hardware reed with a high performance gas chromatograph and

a.k.a. David N. Shaw, P. E.

prepfast Autodilution Autocalibration Auto QC Dilution Elemental Scientific

Automating HPLC and GC Analytical Method Validation Fusion AE Software Program White Paper

METHOD 30B DETERMINATION OF TOTAL VAPOR PHASE MERCURY EMISSIONS FROM COAL-FIRED COMBUSTION SOURCES USING CARBON SORBENT TRAPS

Minimizing Carbon Dioxide Contamination in Anion Reagent-Free Ion Chromatography

Fusion Analytical Method Validation

OVERVIEW OF THE IPCC GUIDELINES

A Statistical Comparison Of Accelerated Concrete Testing Methods

Automating EPA 6020 Compliant Analysis with the Agilent 7900 ICP-MS and ESI prepfast Autodilution System

METER SELECTION FOR VARIOUS LOAD REQUIREMENTS. Edgar B. Bowles, Jr. Southwest Research Institute 6220 Culebra Road San Antonio, TX USA

Paraffin Deposition Progress Report April June 2003

PAMS: A Study of Performance in Low and High Humidity Environments

Chem 321 Lecture 23 - Liquid Chromatography 11/19/13

Universal Solution for Monoclonal Antibody Quantification in Biological Fluids Using Trap-Elute MicroLC-MS Method

Lab 6 Measurement of Ozone

Guidance for Portable Electrochemical Analyzer Testing Used for Compliance Monitoring

Zinc Testing and On Site Calibration for Water Testing Detection Range: 1 15ppm

EXPERIMENT 5. Molecular Absorption Spectroscopy: Determination of Iron with 1,10-Phenanthroline

The MIDAC Low Moisture in Corrosive Gases Process Analyzer for Continuous On-Line Applications

Petroleum Fuels. ogen process analyzers 6000 SERIES

A Single Analyzer Solution for Refinery Flare Emissions Monitoring (40 CFR 60 Subpart Ja) Using Mass Spectrometry

Cadmium Testing and On Site Calibration for Water Testing Detection Range: ppm

Validation of Analytical Methods used for the Characterization, Physicochemical and Functional Analysis and of Biopharmaceuticals.

Measurement of part per billion H2-NMHC-CH4-N2-CO2-CO for semiconductor gases

Feature Article. Improving the accuracy of Fuel Consumption Measurement in CVS system. Tatsuki KUMAGAI. Introduction

OSA100 XRF On-stream Slurry Analyzer

Skoog, Holler and Nieman, Principles of Instrumental Analysis, 5th edition, Saunders College Publishing, Fort Worth, TX 1998, Ch 33.

Landfill Bioreactor Protocol May 2008 SPECIFIED GAS EMITTERS REGULATION. Withdrawn. MAY 2008 Version 1. Page 1

Landfill Bioreactor Protocol May 2008 SPECIFIED GAS EMITTERS REGULATION. MAY 2008 Version 1. Page 1

PAHs in Surface Water by PDA and Fluorescence Detection

Uranium Testing and Site Calibration for Water Testing Detection Range: 2 60ppb

Developing a Fluorinated Greenhouse Gas Stack Testing Method using FTIR

Diode laser analyser LS4000 Ammonia slip measurement for DeNOx process

CO Sensor Response Study NO & NOx Analyzer Validation. June 6, 2018

THE LEAD PROFILE AND OTHER NON-PARAMETRIC TOOLS TO EVALUATE SURVEY SERIES AS LEADING INDICATORS

// XPLORER-NS. Full range Total Nitrogen, Total Sulfur instrument for the modern Laboratory.

Automatic Determination of Greenhouse Gases by GC

Application Note. Hydrogen Purity Analysis by FTIR PROBLEM BACKGROUND

Method X: Determination of Metal Concentrations in Emissions from Stationary Sources (Instrumental Analyzer Procedure)

Laboratory Validation of AT575 Nitrous Oxide Vapor Sampler

Discharge Measurements using Radiotracer Techniques

Urinalysis and Body Fluids CRg. Laboratory Regulation. Laboratory Regulation for Quality Assessment. Unit 1 B. Quality Assessment

ADVANCES IN MICRO GAS CHROMATOGRAPHY

SAN JOAQUIN VALLEY UNIFIED AIR POLLUTION CONTROL DISTRICT COMPLIANCE DEPARTMENT COM 2025

A70 Checklist for Asbestos Fibre Counting Revision 1.11 March 21, 2018

Field Tests of Chloroform Collection/Analysis Methods

ABB Measurement & Analytics Monitoring cement plant stack emissions using FTIR

Inquiry CFA #1. Student Name: Date: Teacher Name: Jasmine Straughter. Score:

AUTOMATIC PEROXYACETYL NITRATE (PAN) ANALYZER

For Condensing Economizers ENBRIDGE MAY Topics

Heated FID OVF Portable THC/TVOC Analyzer

Pipe Flow Hydraulics

Performance of Single Point Monitor in Measuring Ammonia and Hydrogen Sulfide Gases

Greenhouse effect. Measuring temperature inside and outside a greenhouse

The Rapid Development of. Injection Control. 1. Introduction. 2. Hybrid optimization using an RSM

Comparison Continuous Emission Monitoring Techniques

Odyssey Conductivity & Temperature Logger.

Corn Syrup Analysis E-1a- 1. ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography)

Transcription:

Mass Flow Controllers and Carbon Dioxide - Their Effects on Quantitation by Randall Bramston-Cook Lotus Consulting 5781 Campo Walk Long Beach, California 90803 310/569-0128 Copyright 2003 Lotus Flower, Inc.

Concentration measurements in gas chromatography are relative calculations. The size of the eluting peak for an unknown is normally directly proportional to the corresponding size of a standard peak collected under identical conditions. Response Factor = Area of Standard [Standard] This determination is very reliant on knowing the injection volume. [Unknown] = Area of Unknown X Injection Volume of Sample Response Factor Injection Volume of Standard Standard procedures in chromatography mandate a verification of system linearity to ensure that the relationship given above applies to a range of concentrations. One process is to inject standards over a range of concentrations to establish that the response factor remains constant over a concentration range. In gas analysis, this process to generate accurate standards of various concentrations can be very tedious and subject to errors in the dilution process, or it can be very expensive with an inventory of multiple-level standard cylinders required. Mass flow controllers can be employed in an analytical gas system to set a sample flow rate into the injection process, typically into cold trapping systems. By timing the sample loading into the trap with a given flow rate, the injection volume is determined. For example, a flow rate of 50 ml/min for 3 minutes yields an injection volume of 150 ml. Loading Time Volume Loaded Effective Concentration 12.00 minutes 600 ml 8.00 ppb 6.00 minutes 300 ml 4.00 ppb 3.00 minutes 150 ml 2.00 ppb 2.00 minutes 100 ml 1.33 ppb 1.00 minutes 50 ml 0.67 ppb 0.50 minutes 25 ml 0.33 ppb A multipoint calibration is readily generated from one standard by simply varying the timing for the standard loading. The table above illustrates a typical application of a multipoint calibration using a single standard with a concentration of 2 ppb, based on a loading volume of 150 ml. And Figure 1 demonstrates typical results generated with the various sampling times of a single standard.

Multipoint Analysis for Trichloromonofluoromethane R = 0.9995, Max. Linear = 8.00 ppb Area Counts 450000 400000 350000 300000 250000 200000 150000 100000 50000 0 0 1 2 3 4 5 6 7 8 9 Conc. (ppb) Figure 1. A nice advantage of Mass Flow Controllers is that they can be used to generate multiple level calibrations, all based on a single standard. As the sample flow to the cold trap remains constant, the sampling time is varied to yield the different concentrations. The graph above illustrates the linearity from 0.33 ppb to 8.00 ppb by setting the MFC to 50 ml/min and varying the sampling time from 0.5 minutes to 12 minutes. All is good when the sample bulk gas matches the standard matrix. Mass flow controllers are calibrated by the manufacturer to a preset bulk gas, typically air. 1 But if the composition of the bulk gas deviates from the calibrated basis, systematic errors are introduced. 2 Errors generated from various concentrations of carbon dioxide in air are given in Figure 2. Adjustments are possible, but the new matrix composition must be known and correction factors applied mathematically, albeit subject to induced calculation errors. 3 The error can be significant if the carbon dioxide level exceeds 5% of the matrix (with balance air). 1 Air and nitrogen have identical responses with mass flow controllers - Operator s Manual, Sierra Series 830/840/860 Side-Trak and Auto-Trak Mass Flow Meters and Controllers, Sierra Instruments, Monterey, CA, 1994, A-6. 2 Ibid., A-2. 3 Important note about K-Factors (corrections to MFC calibrations): Please note that if you have a transducer calibrated for a gas such as methane and wish to use the K-factors to measure a gas such as air, the inaccuracy of the measurement can range from ±5 to 10%. The use of K- factors is, at best, only a rough approximation and should not used in applications that require a better than ±5 to 10% accuracy. Ibid., A-5.

0 Systematic Error Factor (%) -10-20 -30 0 20 40 60 80 100 Percent CO 2 in Sample Figure 2. Alterations in the overall coefficient of specific heat of the sample matrix, such as the inclusion of carbon dioxide, can greatly effect a change in the volume that is loaded into a trap with a mass flow controller. The estimated systematic error when the mass flow controller is installed upstream from the trap is plotted for various carbon dioxide levels. For example, a 50% carbon dioxide sample can generate an approximate error of 15% in volume due to the change in sample conductivity. When trapping gas samples into a cold trap warm enough to allow carbon dioxide to pass on through, the magnitude of the flow rate error is indicated above. If the trap is colder, normally below -78 o C, carbon dioxide is frozen out, held in the trap and taken out of the flow stream. The resulting flow into the mass flow controller is then less than anticipated, as illustrated below and in Figures 3 and 4. As a result, the indicated flow for the mass flow controller is not accurate, with the magnitude dependent on the concentration of carbon dioxide in the sample. 200 ml/min Cold 100 ml/min Trap 50% CO 2 Trapped 50% CO 2 /50% air 50% air Mass Flow Controller

2000 1800 Systematic Error Factor (%) 1600 1400 1200 1000 800 600 400 200 0 0 20 40 60 80 100 Percent CO 2 in Sample Figure 3. When the mass flow controller is mounted downstream from the trap and the trap is cold enough to freeze out carbon dioxide (< -78 o C), the carbon dioxide component in the sample is snared out of the sample stream and the controller only meters the remaining gas. Possible systematic errors are charted for various carbon dioxide levels with a mass flow controller downstream from the trap. For example, if a sample possesses 50% carbon dioxide and if the trap is cold enough to trap out this carbon dioxide, the mass flow controller effectively sees only half of the sample and the resulting volume error measured by the mass flow controller will be +100%.

100 Systematic Error Factor (%) 80 60 40 20 0 0 10 20 30 40 50 Percent CO 2 in Sample Figure 4. While the magnitude of the error is catastrophic at high concentrations of carbon dioxide, when the mass flow controller is located downstream of the trap and carbon dioxide is frozen out of the stream (from Figure3), the error is still significant even at modest concentrations. For example, at a CO2 concentration of 20%, the flow passing into the trap is 25% higher than the flow measured by the controller.

If the sample matrix is ambient air, no corrections are required since the mass flow controller will indeed measure the actual loading volume properly. 4 With samples from combustion sources, such as vehicle exhaust or stack emissions, both possessing sizeable and variable levels of carbon dioxide, corrections for these errors become difficult to generate. An accurate assessment of the carbon dioxide concentration must be determined and then mathematical manipulations, outlined above and listed in Table II, are applied to yield the corrected injection volume and the effect on the results. Even then, an accumulation of errors can severely impact the accuracy of the final corrected results. These errors are not canceled out from the calibration sequence. The standard matrix is frequently either nitrogen or air; unlikely is the standard to possess the same carbon diode level as exhaust or stack emission samples. Besides, the carbon dioxide level can vary from sample to sample. Table I illustrates the possible systematic errors realized with a 50% carbon dioxide matrix. Concentration of Sample in Air Same Sample with 50% CO2 Relative Error Same Sample with 50% CO2 and MFC after Trap Relative Error 2.00 ppb 1.70 ppb -15% 4.00 ppb +100% Table I. The magnitude of possible errors is demonstrated for a 2.00 ppb sample in a matrix with 50% CO 2. The position of the cold trap relative to the mass flow controller alters the effect. If the sample matrix matches the standard matrix and it corresponds to the calibration of the mass flow controller, a proper aliquot of sample is correctly loaded, with the cold trap located either prior or after to the mass flow controller. However, when the sample matrix is different, then serious errors can be introduced, resulting in invalid results. These errors are not detected with calibration sequences, as the errors are only realized with sample runs with differing major components. The classical approach to eliminate any bias in results from the variation in matrix composition is to apportion the sample volume through a fixed volume sample loop. Now the sample loading is always measured volumetrically and is quite independent of the gas matrix. The only mandate is the precautions necessary to ensure that the sample loop is at constant temperature and the injection occurs when the loop has a consistent pressure for both standard and samples, due to the impact of the Gas Laws. 5 4 Levels of CO2 in ambient air are much too low (< 400 ppm) to make a significant contribution (< 0.04%) to the flow measurement errors discussed here. 5 From the Boyle-Mariotte Gas Law, volume in a loop has a direct relationship with pressure. Doubling of the sample pressure in the loop yields a twofold increase in its effective volume to the trap. Similarly, the Charles-Gay-Lussac Gas Law defines the relationship between volume and temperature. A five (5) degree Celsius increase in the loop temperature can generate nearly a 2% decrease in the effective loop volume.

Error with MFC prior to Trap Percent of CO 2 In Sample Error with MFC after Trap 0 0 % 0-3 % 10 % + 11 % - 6 % 20 % + 25 % - 9 % 30 % + 43 % - 12 % 40 % + 67 % - 15 % 50 % + 100 % - 17 % 60 % + 150 % - 20 % 70 % + 233 % - 22 % 80 % + 400 % - 24 % 90 % + 900 % - 26 % 100 % + Table II. The magnitude of systematic errors, in sample loading volume with a mass flow controller realized with various levels of carbon dioxide in the matrix, can be significant for even modest levels of CO 2. Lotus Consulting 310/569-0128 Fax 714/898-7461 email ebramstoncook@msn.com 5781 Campo Walk Long Beach, California 90803