I-STAT POCT System PRINCIPLE:

Similar documents
i-stat iccnet CHSA Edition 1.2 Page 1 of 7 i-stat Rosy Tirimacco / Paul Simpson

PHSA Laboratories CW Site - Point of Care Title: CWPC_BGE_0125 istat1 Aqueous Quality Control Procedure

Department of Laboratories St. Louis, MO ISSUE DATE: June 2007 REVISION DATE: May 2012 REVIEWED DATE: May 2018 PRINCIPLE:

Type Description Date Version Performed By Notes

Design and implementation of a microchemistry analyzer

i-stat SYSTEM PRINCIPLE:

CARTRIDGE AND TEST INFORMATION

The following reagents and supplies are regarded as critical to performance and accuracy.

PERFUSION TECHNOLOFISTS OF GREATER CHICAGO, INC. POLICY AND PROCEDURE MEDTRONIC HEMOTEC ACTIVATED CLOTTING TIME (ACT) OPERATING PROCEDURE

Note: The taller 100 x 13 mm tube is not designed to run on the ACL TOP. Plasma after centrifugation of these tubes must be placed in sample cups.

PERFORM Operating Document. Use and Maintenance of Chemistry Analyzer Piccolo Xpress

Hemoglobin HemoCue Hb method

ITC Hemochron Signature Elite

CLIA Complexity: Waived for Finger-stick Whole Blood. Instructional video available online at quidel.com/sofia2lyme.

Introduction of PoCT. Synopsis of Talk. Quality Assurance. PoCT should not compromise standards of patient care and clinical decision-making

Title: Hemoglobin HemoCue Hb 301 method

In-Country shipment : How to safely ship human blood samples from suspected Ebola cases within a country by road, rail and sea

Department Of Pathology Point of Care Testing POC i-stat One - SHB Version #16

Implementation Guide. 3a. Skills Checklist. 3b. Competency Exam. 3c. Answer Key: Competency Exam. 3d. Competency Assessment Checklist

How to safely ship human blood samples from suspected Ebola cases within a country by road, rail and sea

Quick Reference Instructions

For certification, manufacturers must perform a sample comparison with a Secondary Reference Laboratory (SRL) following the NGSP protocol.

How to safely ship human blood samples from Lassa cases within a country by road, rail and sea

Na + K + Cl - TCO 2 Ca ++ Mg ++ Li + TCa Glu BUN Crea. Hct ph. Nova Electrolyte/Chemistry Analyzers Versatility for All Your Testing Requirements

Please consult the Product Package Insert before running the test

Contents: POCT Program Massachusetts General Hospital - Pathology Service 55 Fruit Street, Boston, MA 02114

Special Specimen Collection Procedures for Coagulation Testing

All data should be sent by from the laboratory and from the SRL by directly to the NGSP Network Coordinator (above address).

NGSP Level II Laboratory Information Packet

QUICK REFERENCE INSTRUCTIONS For use with Sofia and Sofia 2.

IQCP for the Akers Bio PIFA Heparin Platelet Factor 4 Rapid Assay

Copyright 2017 Aegis Sciences Corporation All Rights Reserved

QUICK REFERENCE INSTRUCTIONS For use with Sofia and Sofia 2. Rx only

Goat IgG ELISA Catalog #:

Hepatitis C virus Ab ELISA Kit

Proficiency Testing on the i-stat System

Date. Albumin measurements are used in the diagnosis and treatment of numerous diseases primarily involving the liver and/or kidneys.

UNIVERSITY OF TOLEDO HEALTH SCIENCE CAMPUS

HBeAg and HBeAg Ab ELISA Kit

Department Of Pathology Point of Care Testing POC Quick Vue In-Line Strep A Version#4

Discussion. Objectives:

HEMOCHRON. Whole Blood Coagulation Systems

QUICK REFERENCE GUIDE. Easy operation fast results. Closer to Care.

Human CRP Precoated ELISA kit

MyBioSource.com. Human Vitamin K2 ELISA USER INSTRUCTION

Blood Gas Analyzer-BGA-2100

MyBioSource.com. Human Osteoprotegerin ELISA USER INSTRUCTION

MyBioSource.com. Rat Hydroxyproline ELISA USER INSTRUCTION

Please consult the Product Package Insert before running the test

Procine CRP ELISA Kit

Date. Alkaline phosphatase measurements are used in the diagnosis and treatment of liver, bone, parathyroid, and intestinal diseases.

Mouse RF IgM ELISA Kit

REAGENTS AND MATERIALS SUPPLIED Control 2: 25-OH vitamin D in human serum with <0.1% Streptomycin Sulfate (2)

Section: Administration Date of Issue: 05/09/2011 Issued By: Administration Part: Pneumatic Tube System Revision #: Revision Date:

Instructions for Use: Duodenoscope Sampling Kit

Maxwell RSC Buffy Coat DNA Kit

BioSpecimen Exchange for Neurological Disorders (BioSEND) LBD Training Webinar

Human High sensitivity C Reactive Protein ELISA KIT

i-stat 1 System Manual

Monkey apolipoprotein A1 (Apo-A1) ELISA Kit. For the quantitative determination of monkey apolipoprotein A1 (Apo-A1) concentrations in serum, plasma.

For In Vitro Diagnostic Use

Mouse Noradrenaline (NA) ELISA Kit

M-TP PRINCIPLE REF ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

Goat immunoglobulin G (IgG) ELISA Kit

Guinea pig PGI2 ELISA Kit

Canine PHA ELISA Kit

Human LENK ELISA Kit

Human Angiotensin 2 (Ang2) ELISA

Unique patient demographic identifiers: Name, Date of Birth (DOB), medical record number (MRN), Social security number (SS#)

Chicken VEGF ELISA Kit

Human connective tissue growth factor (CTGF) ELISA Kit. MyBioSource.com. This package insert must be read in its entirety before using this product.

Rabbit C-reactive protein(crp) ELISA Kit

Guinea pig TIMP-1 ELISA Kit

Human CoxV-A16 ELISA Kit

Rat GLO ELISA Kit. For the quantitative in vitro determination of Rat Glyoxalase concentrations in

Guinea pig Hyp ELISA Kit

Laboratory Procedure Manual

Bovine IgG-Ab ELISA Kit

Mouse helicobacter pylori antibody (IgG) ELISA Kit

Human anti-endomysial (EMA) antibody (IgG) ELISA Kit

Human CTRP5 ELISA Kit

Rabbit immunoglobulin M(IgM) ELISA Kit

Human heparin-platelet factor 4 complex antibody (IgG) ELISA Kit

Human anti-gliadin antibody (IgG) ELISA Kit

Monkey hs-tnt ELISA Kit

Mouse immunoglobulin M(IgM) ELISA Kit

Human anti-cytomegalovirus(cmv) antibody (IgM) ELISA Kit

Guinea pig 25-OH-VD ELISA Kit

BioSpecimen Exchange for Neurological Disorders (BioSEND) LBD Training Webinar

Fish Estradiol(E2) ELISA Kit

Hamster follicle stimulating hormone (FSH) ELISA Kit

Human hepatitis E virus antibody (IgG) ELISA Kit

Human salmonella typhi (ST) antibody (IgM) ELISA kit

Human 24(S) Hydroxycholesterol ELISA KIT

Human placenta lactogen (HPL) ELISA Kit

Rat follicle-stimulating hormone (FSH) ELISA Kit

Mouse Oxytocin(OT) ELISA Kit

BioSpecimen Exchange for Neurological Disorders, BioSEND

Human nuclear matrix protein 22 (NMP-22) ELISA Kit. MyBioSource.com

QUICK USER GUIDE CAPILLARY BLOOD INR TESTING FOR USING THE THROMBI-STAT MC1 WB INSTRUMENT MANCHESTER CAPILLARY REAGENT INTRODUCTION AND

Transcription:

PRINCIPLE: I-STAT POCT System The i-stat system uses a single disposable cartridge. The cartridge, which has been filled with fresh blood, is placed into a hand-held analyzer for analysis. The results are printed and also transmitted to a computer where they can be stored and transferred to another system for charting.

PRINCIPLES OF MEASUREMENT: Sodium, Potassium, Chloride,Ionized Calcium are measured by ion-selective potentiometry. Concentrations are calculated from the measured potential through the Nernst Equation. Urea is first hydrolyzed to ammonium ions to be measured by an ion-selective electrode. Concentrations are calculated from the measured potential through the Nernst Equation. Glucose is measured amperometrically. Oxidation of glucose produces hydrogen peroxide, which is oxidized at an electrode to produce an electric current, which is proportional to the glucose concentration. Creatinine is hydrolyzed in several steps to hydrogen peroxide. Hydrogen Peroxide is then oxidized at an electrode to produce a current, which is proportional to the sample Creatinine concentration. Hematocrit is determined conductometrically. The measured conductivity, after correction for electrolyte concentration, is inversely related to the hematocrit

SPECIMEN REQUIREMENTS: Note: All blood specimens are potentially infectious. Observe standard precautions at all times when handling athlete or race participant specimens. Chem 8 cartridge : (measures NA, K, CL, CO2, BUN, GLU, Creatinine, Ionized Calcium, Hct calculated Anion Gap and HB) Fresh whole blood collected in a green top (lithium heparin) collection tube Test Immediately -Fresh whole-blood collected into a plastic syringe or blood that has not been put into a green top tube. TEST WITHIN 30 MINUTES OF COLLECTION IN GREEN TOP TUBE Remix thoroughly before testing

REAGENTS AND SUPPLIES i-stat System 1. Cartridges stored at 2-8 o C (36-46 o F) Store main supply refrigerated until expiration date. DO NOT FREEZE. Cartridge stability at room temperature is noted on back of each cartridge A single cartridge must be at room temperature for 5 minutes prior to use. A box of cartridges must be at room temp for an hour before use. Date room temperature cartridges with new room temp expiration date. Do not return room temperature cartridges to refrigerator. 2. Analyzer 3. Lithium 9V batteries 4. Printer and paper (if used in area, not required)

Blood Collection and Transfer Equipment 1. Gloves 2. Venipuncture collection equipment 3. Green top heparin tubes 4. Transfer device 5. 1cc plain syringe

Liquid Controls 1. I-STAT Chem 6 or Chem 8 Controls depending or cartridges used may vary from event to event. 2. Liquid controls (supplied by Abbott) should be run on each new shipment of cartridges.

SAFETY PRECAUTIONS Always wear gloves when performing a test and change them if they become contaminated. Treat all specimens as capable of transmitting disease. Never access clean supplies with contaminated gloves. Put extra supplies on a paper towel before applying gloves. When test is complete: remove gloves and perform hand hygiene Clean the i-stat with a germicidal if the i-stat is visibly contaminated with blood. Do not immerse or spray analyzer

CALIBRATION: Calibration is automatically performed as part of the test cycle on each cartridge. Operator intervention is not necessary.

QUALITY CONTROL: #1 Upon receipt of a new Shipment of Cartridges (new lot number) testing of a sample cartridge using a liquid test sample (available from Abbott Laboratories) is performed. All control test results must be within Manufacturer s acceptance range. Document results on New Cartridge Shipment Log.

QUALITY CONTROL: #2 Acceptance Criteria for shipment of cartridges Temperature of cartridge shipping does not exceed 8 o C. Only window A/1 of temperature Monitor on the package is colored. Windows B/2, C/3 and D/3 are colorless. If windows B,C, and D are blue. This is an indication that shipping container temperature has been compromised and the cartridges should not be used. Results of any of the liquid controls are outside of manufacturer s acceptance range the cartridges should not be used. Document results on New Cartridge Shipment Log along with corrective actions taken. Quarantine lot of cartridges and notify Abbott Customer Service for advice on return or removal.

An electronic stimulator is provided with the i-stat machine, and is used to test the internal electronics. This is performed each time the i-stat is taken out for use. In many cases, this procedure will be completed before you receive the i-stat in the field. We advise that you check with your Medical Director to ensure this testing has been completed.

ISTAT TESTING PROCEDURE: #1 Athlete or Race Participant Testing Press ON/OFF to turn analyzer on. Select 2- I-STAT Cartridge. Dispense the sample until it reaches the fill mark (blue mark) on the cartridge. Fold the SNAP COVER over the sample well until it snaps into place. Press on the round tab, NOT overfill the sample well or a pre-burst error will occur and a new cartridge will need to be used. * Make sure not to touch the platinum mental on the end of the cartridge. This will contaminate the cartridge and render the test inaccurate.

ISTAT TESTING PROCEDURE: #2 Athlete or Race Participant Testing Carefully insert cartridge into port of I-STAT analyzer. Do not attempt to remove cartridge while Cartridge Locked remains on the screen. Note, for best results keep the i-stat on a level surface and do not move the i-stat while it is running. When screen displays Operator ID, enter your assigned ID # (volunteer number or equivalent) and ENTER. Repeat step. At Patient ID (athlete bib number or equivalent)enter assigned Patient ID # and ENTER. Repeat step. Results will appear when tests are complete. Remove cartridge after Cartridge Locked message disappears. Dispose in a Sharps container. The analyzer is ready for the next test immediately.

Sampling Tips 1. Tilt cartridge if sample does not flow to fill mark. 2. Cartridge may not seal properly if sample well is overfilled. Do not draw back excess with syringe. Use a gauze pad over snap closure when covering well. This will absorb the excess but not affect the cell/plasma ratio. 3. Do not use cartridge on which blood or any other liquid has spilled as the internal contact pads in analyzer may become contaminated. 4. Be sure sample is well mixed incorrect results may occur with poorly mixed specimen. 5. Falsely high potassium results can occur from hemolyzed specimens. 6. If results do not match patient condition, repeat the testing with a new cartridge.

RESULT REPORTING Results are available for viewing on i-stat screen or can be printed using the option i-stat printer LINEARITY / REPORTABLE RANGES OF i-stat (provided by i-stat) Measured: SODIUM POTASSIUM CHLORIDE BUN GLUCOSE Calcium, Ionized CREATININE TCO2 HEMATOCRIT 100-180 mmol/l 2.0-9.0 mmol/l 65-140 mmol/l 3-140 mg/dl 20-700 mg/dl 1.0 10.0 mg/dl 0.2-20.0 mg/dl 5 50 mmol/l 10 75 PCV% Calculated: HEMOGLOBIN 3.4-25.5 g/dl Anion Gap -10 - +99

Suppressed Results There are three conditions under which the i-stat System will not display results: Results outside the System s reportable ranges are flagged with a < or >, indicating that the result is below the lower limit or above the upper limit of the reportable range. The <> flag indicates that the results for this test were dependent on the result of a flagged as either > or <. Action: Specimen would need to be sent to a laboratory for analysis.

Suppressed Results Cartridge results were not reportable based on internal QC rejection criteria are flagged with ***. Action: Analyze the specimen again using a fresh sample from the well mixed blood tube and another cartridge. The results that are not suppressed may be reported in the usual manner. If the result is suppressed a second time, send the sample to a lab for analysis.

Suppressed Results A quality Check message will be reported instead of results if the analyzer detects a problem with the sample, calibrant solution, sensors, or mechanical or electrical functions of the analyzer during the test cycle. Action: Take the action displayed with the message that identifies the problem.

Printing the i-stat results Align the IR windows of the analyzer and the printer. To print displayed results, push the print key.

CALCULATIONS: The i-stat analyzer contains a microprocessor that performs all calculations required for reporting results. LIMITATIONS: Results outside the systems reportable ranges are flagged with a "< or >"or <>, send to the lab for analysis. Results which are unreportable *** based on internal rejection criteria should be retested again with another cartridge. Results will be suppressed if a test cycle has a problem with sample, calibrant solution, sensors, mechanical or electrical functions of the analyzer. Refer to i-stat ERROR CODE LIST to troubleshoot problem. Specimens should be rejected for analysis; When evidence of clotting is present. If specimen is collected with an anticoagulant other than heparin.

RECALLING RESULTS: Press On/Off key Press 1 to continue if prompted Press Menu key to display Administration Menu Select 2 Data Review Select 1 Patient Enter patient ID Bottom of screen will indicate number of pages of results. Scroll back and forth using the 1 and 2 keys. Press Menu key to access another patient or On/Off key to end.

MAINTENANCE: Disinfect analyzer and remove any residual dried blood using a germicidal wipe, or by using a cloth moistened with disinfectant. 1. DO NOT spray disinfectants onto the analyzer. 2. DO NOT use alcohol wipes. That will cause deterioration of the screen and outer covering of analyzer. 3. Avoid wiping plastic screen with disinfectant unless screen is visibly contaminated

QUALITY CONTROL: Daily Procedures Verify the operation of each i-stat analyzer with an electronic simulator. The Internal Electronic is set to automatically be activated every 8-hours if i-stat is in use. INTERNAL ELECTRONIC SIMULATOR Internal Electronic Simulator will be automatically initiated by the insertion of a cartridge every 8-hours as needed. Results are transmitted to the lab for review.

QUALITY CONTROL: If result is PASS Pass will not display on screen and analyzer will perform testing on inserted cartridge If result is FAIL - analyzer will display FAIL and you will have to remove test cartridge and reinsert to initiate a test cycle. If FAIL is displayed a second time, do not use analyzer and contact the POCT Coordinator. The integrity of each new shipment of cartridges is verified with assayed controls. Verify that the i-stat cartridges at room temperature have not been out of the refrigerator more than cartridge packaging specifies and that room temp expiration date is documented on cartridge box or individual cartridge.

CALIBRATION For cartridges, calibration is automatically performed as part of the test cycle on each cartridge type except for coagulation tests (ACT, PT/INR). Operator intervention is not necessary.

REFERENCE RANGES Sodium 138-146 mmol/l Potassium 3.5-4.9 mmol/l Chloride 98-109 mmol/l Total C02 24-29 mmol/l BUN 7-25 mg/dl Glucose 65-100 mg/dl Creatinine 0.8-1.3 mg/l Ionized calcium 4.5 5.3 mg/dl Hematocrit 38-51 % Hemoglobin (calc) 12.0 17.0 g/dl Anion Gap 10-20 ***Note: Ranges listed above are for adult population