LightScanner Hi-Res Melting Comparison of Six Master Mixes for Scanning and Small Amplicon and LunaProbes Genotyping

Similar documents
LightScanner Hi-Res Melting Comparison of Six Master Mixes for Scanning and Small Amplicon and LunaProbes Genotyping

LightScanner. Systems. Get There Faster. Achieve High Throughput Mutation Discovery and Genotyping.

HRM Powered by Precision Melt Analysis Software

Chapter 2. Genotyping DNA Variants with High-Resolution Melting Analysis. Rolf H.A.M. Vossen. Abstract. 1 Introduction

High-Resolution Melting Interactive Workshop AACC, July 25 th, 2006 Assay Description

Applications and Uses. (adapted from Roche RealTime PCR Application Manual)


Detection of FactorV Leiden (R506Q)

High-Resolution Melting analysis as a tool for rapid and sensitive detection of genotypes in cattle population

High-throughput BRCA1 mutation scanning using HR-MCA MCA. Evaluation project :

Rapid Cycle PCR, Real Time Analysis, and Hi-Res Melting

LightCycler 480 qpcr Tools. Meeting the Challenge of Your Research

For outstanding performance in real-time PCR

amplification High Resolution Melt Parameter Considerations for Optimal Data Resolution tech note 6009

Thermal cycler amplification robustness: a comparison of several models

Precipio, Inc. Instructions for Use. PIK3CA Exon 9 Mutation Analysis using ICE COLD-PCR for Detection with High Resolution Melting

3color RT HS-PCR Mix SYBR Ready-to-use mix for real-time Hot Start PCR with SYBR Green. Dedicated for white reaction tubes and plates.

Tamiflu resistance detection kit

Applied Biosystems Real-Time PCR Rapid Assay Development Guidelines

Detection of CFTR(F508) MyBioSource.com. Real-time PCR mutation detection/allelic discrimination kit. For general laboratory and research use only

Perform an HRM Methylation Study

QPCR-S100 (100 rxns) Store -20. Functional analysis RealHelix qpcr Kit was evaluated by real-time PCR using the 10-fold serial-diluted human

User Bulletin. Veriti 96-Well Thermal Cycler AmpFlSTR Kit Validation. Overview

PowerPlex. Y System Validation

2-3. Real-time PCR conditions using Roche LightCycler Nano Real-time PCR conditions using Bio-Rad MiniOpticon


KAPA Library Quantification Kits For Applied Biosystems SOLiD platform

THUNDERBIRD SYBR qpcr Mix

EGFR T790M mutation mutation detection by quantitative allele specific amplification (quasa) EGFR (T790M)


Supplementary Methods

High Resolution Melting Protocol

BRAF V600E mutation mutation detection by quantitative allele specific amplification (quasa) BRAF (V600E)

Cat. # RR430S RR430A. For Research Use. SYBR Fast qpcr Mix. Product Manual. v201610da

SYBR Green Realtime PCR Master Mix -Plus-

Detection of Prothrombin (G20210A)

JAK2 V617F mutation mutation detection by quantitative allele specific amplification (quasa) JAK2(V617F)

RealHelix TM qrt-pcr Kit [Intercalator type]

#K0262 For 1000 reactions of 25 µl Lot Exp.. Store at -20 C in the dark. V

Table of contents. I. Description...2. II. Principle...2. III. Kit Components...3. IV. Storage...3. V. Features...4. VI. Precautions for Operation...

Rotor-Gene Q Pure Detection

User Manual. NGS Library qpcr Quantification Kit (Illumina compatible)

M. tuberculosis_mpb64/is61 10 genesig Standard Kit

Factors affecting PCR

SYBR Green Realtime PCR Master Mix

Novallele Genotyping Assay User Guide

Positive control kit for enzymatic mismatch cleavage and agarose gel visualization (version 2.4)

User Manual. Chromofy dsdna binding dye Version 1.0 April 2009 For use in quantitative real-time PCR

Idaho Technology Inc. Kit No. HRLS-ASY-0001

Successful gene expression studies using validated qpcr assays. Jan Hellemans, CEO Biogazelle webinar October 28 th, 2015

Applied Biosystems 7500 Fast, 7500 and 7300 Real-Time PCR Systems

SYBR Premix Ex Taq II (Tli RNaseH Plus), Bulk

Phosphate buffered saline (PBS) for washing the cells TE buffer (nuclease-free) ph 7.5 for resuspending the SingleShot RNA control template

mirna QPCR Master Mix

SYBR Advantage qpcr Premix User Manual

Detection of Biological Threat Agents by Real-Time PCR - Comparison of Assay

M. tuberculosis_mpb64/is611. genesig Advanced Kit. 150 tests. Primerdesign Ltd. For general laboratory and research use only

GreenMasterMix (2X) b i o s c i e n c e. G E N A X X O N b i o s c i e n c e. High ROX (500nM)

PCR SYSTEMS. a new era in high-productivity qpcr. Applied Biosystems ViiA 7 Real-Time PCR System

Mycobacterium Tuberculosis

Cat. # RR391A. For Research Use. Probe qpcr Mix. Product Manual. v201610da

Application Note Detecting low copy numbers. Introduction. Methods A08-005B

Rotor-Gene Q Pure Detection. High Resolution Melting. From Assay to Analysis. Nan Fang, Ph.D. Senior Scientist R&D QIAGEN. Sample & Assay Technologies

SeCore. GSSP Kits. Instructions for Use. 1 Research Use Only

Single-Nucleotide Polymorphisms (snps) are the

BioBank cdna kit. Instructions for the use of BioBank control cdna in real-time PCR

A faster, smaller, better way to qpcr

DNA Genotyping from Human FFPE Samples Reliable and Reproducible

mirna QPCR Master Mix

Soya Round up Ready (RR)

TB Green Premix Ex Taq (Tli RNaseH Plus)

Boost Your Real-Time Results

Escherichia coli_o104: H4

TB Green Premix Ex Taq II (Tli RNaseH Plus)

CaMV promoter & NOS terminator detection

Sus scrofa Pig genesig Advanced Speciation Kit

Roche Molecular Biochemicals Technical Note No. LC 9/2000

SYBR Premix DimerEraser (Perfect Real Time)

Platinum II Taq Hot-Start DNA Polymerase for high-throughput PCR

Exceed the limit. SensiFAST Real-Time PCR Family. Superior Sensitive Quantification

Sus scrofa Pig genesig Speciation Kit

Report of Analyzing Short Tandem Repeats for Parentage Testing

Phosphate buffered saline (PBS) for washing the cells TE buffer (nuclease-free) ph 7.5 for resuspending the SingleShot RNA control template

Reference gene detection assay. Instructions for detection and quantification of a reference gene using SYBR Green detection chemistry

QuantiNova proven compatibility with various real-time cyclers in gene expression analysis

Exceed the limit. SensiFAST Real-Time PCR Family. Superior Sensitive Quantification

Closed-Tube SNP Genotyping Without Labeled Probes A Comparison Between Unlabeled Probe and Amplicon Melting

Human Y chromosome detection assay. Detection of the human Y chromosome by real-time PCR

Amplicon Library Preparation Method Manual. GS FLX Titanium Series October 2009

Lithos qpcr Master Mix - 2mM

Have your qpcr reagents evolved?

Library Quantification Kit User Manual

Fully Automated Library Quantification for Illumina Sequencing on the NGS STAR

Supplemental Materials. DNA preparation. Dehalogenimonas lykanthroporepellens strain BL-DC-9 T (=ATCC

SYBR Advantage qpcr Premix. User Manual

hpsc Genetic Analysis Kit

Name Order # 1234 Company Received 03/15/16 Reported 03/16/16. Sample

One Step SYBR PrimeScript RT-PCR Kit II (Perfect Real Time)

PrimeTime Pre-designed qpcr Assays

Real-Time PCR Principles and Applications

Transcription:

LightScanner Hi-Res Melting Comparison of Six Master Mixes for Scanning and Small Amplicon and LunaProbes Genotyping Introduction Commercial master mixes are convenient and cost-effective solutions for producing reliable scanning and genotyping results. was the first to offer Hi-Res Melting instruments, analysis software, and chemistry solutions. This study performed side-by-side comparisons of six commercial master mixes indicated for Hi-Res Melting: LightScanner Master Mix (Figures 1, 3) and the LightScanner High- Sensitivity Master Mix (Figure 2) (with calibrators) with LCGreen Plus dye (, LLC) LightCycler 480 High Resolution Melting Master with ResoLight dye (sold by the Roche Group; LightCycler is a registered trademark of the Roche Group) SsoFast EvaGreen Supermix with EvaGreen dye (sold by Bio-Rad Laboratories; EvaGreen is a registered trademark of Biotium, Inc.) Fast EvaGreen qpcr Master Mix for qpcr and high-resolution melting with EvaGreen dye (sold by Biotium, Inc.; EvaGreen is a registered trademark of Biotium, Inc.) EXPRESS SYBR GreenER qpcr SuperMix Universal (sold by Invitrogen; SYBR is a registered trademark of Molecular Probes Inc.; GreenER is a trademark of Life Technologies, Inc.) MeltDoctor HRM Master Mix (sold by Applied Biosystems, Inc.) A panel of human genomic DNA samples with known genotypes in the LIPC (large amplicon scanning), human tyrosine hydroxylase (small amplicon genotyping), and BX647987 (LunaProbes genotyping) genes were used in this evaluation. Methods Whole blood was drawn from 30 human donors, and DNA was isolated using standard extraction kits and procedures. Samples were diluted in TE 1X buffer, ph 8.0, to a final concentration of 15 ng/µl. Genotypes for all comparisons were identified using Hi-Res Melting technology and sequence-verified. All master mixes were used in accordance with manufacturer recommendations. We optimized PCR protocols for each master mix to generate the cleanest amplicons for analysis on the LightScanner system. Standard agarose gels were run to verify annealing temperature optimization. PCR duplicates (Figure 1) or triplicates (Figures 2 and 3) were run for each DNA sample. All PCRs for master mix comparisons were performed on calibrated 96-well plate thermal block cyclers, followed by Hi-Res Melting on the LightScanner. Data analysis was performed using standard software supplied with the LightScanner. High Sensitivity master mix (with calibrators) was used in place of the LightScanner master mix in the small amplicon genotyping study. It should be noted that this is the only commercial master mix that contains a finely tuned internal calibration system designed to work with High-Res Melting calibration software. Because of this, peak alignment is highest in Figure 2 with this master mix. Using this master mix ensured the highest resolution available despite the inevitable yet subtle sample-to-sample temperature, volume, and buffer variation.

Results Figure 1. Scanning sensitivity variation across the six master mixes. Gene scanning results for the LIPC amplicon (302 bp, 52% GC) are shown in Figure 1. LightScanner, LightCycler 480, and MeltDoctor mix correctly identified all seven genotype groups. In contrast, SsoFast EvaGreen, Fast EvaGreen, and EXPRESS SYBR GreenER master mixes were only able to differentiate five groups. This was due to the homozygous variant samples being incorrectly grouped with the wild-type profiles and two different heterozygous variants being grouped together. This decreased sensitivity is probably caused from reduced or absent heteroduplex content at approximately 87.5 ºC. This second melting domain is clearly observed with the LightScanner Master Mix and is partially observed with the LightCycler 480 mix.

Figure 2. Human tyrosine hydroxylase locus small amplicon genotyping results. The human tyrosine hydroxylase locus was assayed using a set of primers immediately flanking a common short tandem repeat used in forensics (HUMTHO1) yielding a variable length fragment ranging from 67 82 bp. Results are shown in Figure 2 above. Sample genotypes were established by fragment analysis performed independently by the Los Angeles Police Department s forensics laboratory. Regular LightScanner Master Mix did not perform as well (data not shown) as the LightScanner High Sensitivity Master Mix. The regular LightScanner Master Mix is designed for scanning not small amplicon genotyping, which requires temperature calibrators for the highest Tm separation. The High Sensitivity Master Mix correctly identified all eight genotype combinations. The MeltDoctor, SsoFast EvaGreen, Fast EvaGreen, and EXPRESS SYBR GreenER master mixes identified six of eight groups correctly. Several errors were found with the LightCycler 480 mix. In two instances, two genotypes were combined into a single group and also resulted in three unique genotypes being clustered into only two groups. The LCGreen dye in the High Sensitivity Master Mix creates maximal heteroduplex content as can be seen in Figure 2 (note green and orange group profile differences). All other dyes have less heteroduplex content, partly explaining the loss of sensitivity in this small amplicon genotyping application.

Figure 3. Sample mixing to demonstrate LunaProbes sensitivity. LunaProbes genotyping results for the BX647987 locus are presented in Figure 3 above. A common G:C single nucleotide polymorphism (SNP) (rs1869458) was assayed in the human BX647987 gene using a LunaProbes assay to evaluate each product s ability to detect minimal amounts of a given allele. In this case, the probe was designed as a perfect match to the G allele while creating a single base pair mismatch to the C allele. A homozygous G and homozygous C sample were identified, quantified, and mixed to create fractions of both alleles down to 5% in a 95% background of the other allele. A natural heterozygous sample was also run for comparison and validation of the mixed samples. Using the LightScanner and LightCycler 480 master mixes, the 50%C:50%G mixed sample (orange) performs similarly to the true heterozygous sample (olive green) relative to allele peak heights, verifying the mixed samples were accurately diluted. Table 1. Detection Sensitivity of rs1869458 LunaProbes for all high-resolution melting master mix products. Supplier BioFire Defense Roche Bio-Rad Biotium Invitrogen ABI High-resolution Melting Master Mix LightScanner Master Mix LightCycler 480 High Resolution Melting Master SsoFast EvaGreen Supermix Fast EvaGreen qpcr Master Mix for qpcr and HRM EXPRESS SYBR GreenER qpcr SuperMix Universal MeltDoctor HRM Master Mix Dye LCGreen Plus C allele Sensitivity G allele Sensitivity Primer asymmetry 5% 5% 5:1 ResoLight 5% 5% 10:1 Evagreen 25% 10% 5:1 Evagreen 25% 10% 5:1 SYBR GreenER 25% 25% 10:1 SYTO 9 10% 50% 10:1

Conclusions These experiments show that although several master mix products are marketed for high-resolution melting applications, the LightScanner Master Mix and High Sensitivity Master Mix with LCGreen Plus dye provide the best sensitivity and support the broadest range of applications for high-resolution melting. Experiments for each master mix product were independently optimized to produce high-quality PCR amplicons. LightScanner Master Mix with LCGreen Plus dye yielded the best overall results. This is presumably due to its ability to saturate PCR products and yield maximal heteroduplex detection relative to other master mixes. The other commercial master mix products tended to identify lower heteroduplex percentages in the melting curves resulting in reduced melting domain differentiation. Ultimately, this trend leads to a reduction in mutation scanning sensitivity. Small amplicon genotyping results clearly show increased heteroduplex content as well as tighter peak alignment. This enhancement over the other master mixes results from the unique calibration system included in the High Sensitivity Master Mix and associated LightScanner instrument and software. In addition, greater primer asymmetry was required for adequate LunaProbes signal in several master mixes (Table 1). A higher primer asymmetry makes PCR more difficult and reduces efficiency. For optimal high-resolution melting, a master mix and dye that provide greater LunaProbe signal at a given primer asymmetry is preferred, all else being equal. Patents: /LegalNotices LightScanner, LCGreen, LunaProbes,, BioFire Diagnostics and the BioFire logo are trademarks of BioFire Diagnostics, LLC and/or BioFire Defense, LLC and are registered trademarks in the United States. LS32 is a trademark of and/or BioFire Diagnostics. All other names of products and brands appearing in this manual are trademarks or registered trademarks of their respective owners. is leading the way for Hi-Res Melting chemistry, instruments, software analysis methods and customer support. For additional information on this technology and other products, please visit us at or call 1-801-262-3592. MRKT-PRT-0190-02, LLC formally known as Idaho Technology, Inc.