CMS, FDA and AAMI Requirements for Bicarbonate Systems Made Simple. Compiled January 2012 by AmeriWater Incorporated

Similar documents
DIALYSIS WATER TREATMENT DIALYSIS PROCESS WATER SOLUTIONS THE PREMIER PROVIDER OF WATER TREATMENT SOLUTIONS TO THE DIALYSIS INDUSTRY

Dialysis Water PREVIEW and Dialysate COPY Recommendations:

Lesson #9: Water and Dialysate

Water treatment system and water quality Standard, Configuration and Practical Considerations

Resolutions to Comments on AAMI/CDV-1 RD61, Concentrates for hemodialysis

Condition: Water & Dialysate Quality. Glenda Payne, RN, MS, CNN

LABORATORY & DEIONIZED WATER SYSTEMS DESIGN AND CONSTRUCTION STANDARD

Evolution of AAMI/ANSI Standards: Harmonizing with ISO. V. Taaffe 2018

Water for Dialysis Part 1: Components

Disclaimer. Healthcare facilities are required to have written policies approved by their governing body concerning the services that they provide.

AAMI: What is it and how does it impact the biomedical Technician?

photo: Ameriwater.com Dialysis Water Treatment Equipment Monitoring December 2017 Developed by the BCPRA Hemodialysis Committee

Tank Modifications & Accessories

Water treatment in dialysis technology

Water and Dialysate Safety: Your Role in Maintaining Quality. Glenda Payne, MS, RN, CNN

FREQUENTLY ASKED QUESTIONS FOR THE ADVANTAGE

Guidance for the Content of Premarket Notifications for Hemodialysis> Delivery Systems

Engineering Excellence

International Water Standards

Presented to the IAPMO Standards Review Committee on December 10, 2018

Dialysis water purification

APPENDIX C TECHNICAL SPECIFICATIONS HIGH DENSITY POLYETHYLENE (HDPE) PIPE AND FITTINGS

Draft proposals for Test methods for close-coupled solar water heating systems - Reliability and safety

ANSI/AAMI ST79 A2:2009 Key Changes in the 2009 Amendment

recycle water reuse it save up to from showers & baths for flushing toilets 30% of water use

UFR+L FILTER REGULATOR + LUBRICATOR

200 / 300 Eazy Retro Upgrade Kit

Line filters LINE FILTERS

SMF WATER SYSTEMS. (PIC/S July.2004) By Anand Kuptamatee Thai FDA Conference September 19-21, 2005

SECTION COOLING TOWER

Energy Transfer in a Hydraulic System. Lecture (1) Hydraulic and Pneumatic Circuits 4th year Mech. power Engineering Dept.

Regulation. Code of Federal Regulations in the United States. Chapter 1 Title 21 CFR is the title for pharmaceuticals.

Line Filters Solid, simple, smart. Advanced reliability in compressed air. TECHNOLOGY YOU CAN TRUST

VEGETABLE MILKS. Via Don Battistoni, Jesi (AN) Italy Phone Fax

Environmental SAVE ON WATER BILLS MONEY. Rainwater Harvesting Systems for Domestic Applications. Sustainable, Reliable, Affordable

BIO-WASTE SYSTEMS. Thermal & Chemical Effluent Kill Systems. Biotech & Pharmaceutical. Batch Process Systems. Continuous Process Systems

INTERNATIONAL STANDARDS ON FLUID PREPARATION FOR CONVECTIVE THERAPIES. Richard A. Ward, Ph.D.

Assessing the Condition & Location of Your Rain Water Collection System (Cistern)

A. This Section includes the following sump pumps and accessories, inside the building, for building storm drainage systems:

hebrochemie Premium coolants and metalworking fluids from hebro. Product group overview. Stimuli for people and the environment

hebrochemie GmbH Premium coolants from hebro. Product group overview. Stimuli for people and the environment A member of the Chemetall Group

CARR ViaFuge Technology

REVERSE OSMOSIS WATER TREATMENT

Drinking systems. for sows, piglets and finishing pigs

SECTION WELDED STEEL WATER STORAGE TANKS

HRO-50 Water Purifier

Technical Information ANTICORA. Concentrate. High-Performance Cooling Brine for Systems made of Stainless Steel or Non-Alloyed Steel

Welcome! How do I get a CE Certificate?

Monsanto Bulk Pesticide Repackaging Program 2011

SAF Filters. The automatic self-cleaning filter - suitable for more applications than ever.

SUMP / EFFLUENT PUMPS ARE SINGLE PHASE WITH PIGGYBACK FLOAT SWITCH. Max. Flow 44. Max. Head 20. Voltage 115

Captor Containment DOUBLE WALL TANK SYSTEMS.

Captor Containment. DOUBLE WALL TANK SYsTEMS.

I. Preconstruction Meeting and Site Preparation Date: Inspector: A. MDE Certified Installer Name

AIR TREATMENT BREATHING AIR PURIFIERS. DBS Series

SOPHISTICATED DESIGN REQUIREMENTS IN PLANT ENGINEERING FOR AN XLPE EXTRUDER FEEDING PLANT AS AN EXAMPLE

S.M.A.R.T.Green Test Pack for Prevacuum Sterilizers Technology Overview

RAINWATER HARVESTING SYSTEMS FOR DOMESTIC APPLICATIONS

Rainwater Harvesting Systems for Domestic Applications

Architect s Project Number of 5 Decentralized HVAC Equipment District Master Specs 2013 Edition

Advanced water purifier system operation manual and setup guide

A Green Solutions Company TM ENGINEERED PRODUCTS

PROVEN IN THE FIGHT AGAINST

2.2 Elevators may not include any proprietary controls or systems.

Orion. Reliable and proven purified water technology with intelligent innovation in the age of sustainability WATER TECHNOLOGIES

AIR TREATMENT BREATHING AIR PURIFIERS. DBS Series

REVERSE OSMOSIS SYSTEM USER'S MANUAL. Excellent Product, Excellent Water! My PurePro X6

STEAM TURBINE-GENERATOR & AUXILLIARY SYSTEMS Presentation by: RANA NASIR ALI General Manager, Power Plants Projects, at PITCO November 02, 2017

Filters & Elements. People. Passion. Performance

A c t i v a t e d C a r b o n F i l t e r Installation Operation Maintenance Manual

THE FRIENDLY TO KEEP OUR WORLD CLEANS

Sea Water Reverse Osmosis Container. 1. Description:

SECTION CHEMICAL FEED SYSTEM

Pharmastill multi effect Stills

SEPTIC TANK CONSTRUCTION GUIDELINES

Fencing, locks on access manholes, and other necessary precautions shall be provided to prevent trespassing, vandalism, and sabotage.

Charles P. Cavedoni PE Mechanical Engineer PO Box 252 Hakalau, HI (808) ph/fax

FAQ about OIL SIGHT GLASSES.

Assessment Schedule - TS.5

When Safety Comes First. Emergency Safety Shower and Eye Wash Catalogue

Filters & Elements. People. Passion. Performance

SECTION GROUND STORAGE TANK

SECTION CONCRETE ABOVE GROUND FUEL STORAGE

Enviro Voraxial Technology

Mission Statement. Profile and History. R.O. UltraTec s Advantages

RADIAL FLOW SETTLER 60 CONE (RFS-60)

HVAC WATER TREATMENT 1.01 RELATED CORNELL DESIGN AND CONSTRUCTION STANDARDS. D. Section Plumbing (Potable) Water Treatment

flow rates filtration degrees water for cleaning minimum operating pressure micron less than 1% of the total flow

Government Center Water Treatment Plant Kamphaeng Phet Province, Thailand

Rainwater Reuse WHAT TYPE OF SYSTEM? Click to edit Master text styles

Chapter 5 Water, Plumbing, and Waste

Water Service Inspection Checklist

Specco Rainwater Harvesting Products

SUGGESTED SPECIFICATIONS Pump Frames through 64132

Reverse Osmosis Purifiers

F u e l T a n k M a i n t e n a n c e & N e w I n s p e c t i o n R e q u i r e m e n t s

Line filters LINE FILTERS

SUPPLEMENTAL MECHANICAL CORRECTION SHEET FOR SEWAGE EJECTORS AND SUMP PUMPS 2011 LAPC

Water Based Critical Utilities. RO, WFI, and Steam

Proposed LNG Pump Requirements for Regulation 110

Transcription:

CMS, FDA and AAMI Requirements for Bicarbonate Systems Made Simple Compiled January 2012 by AmeriWater Incorporated

Background on the Standards FDA, 21 CFR Sec. 876.5820(a) states that hemodialysis systems and accessories are Class II medical devices used for the treatment of patients with renal failure or toxemic conditions. Bicarbonate concentrate mixing and distribution systems (Bicarb Systems) are included in this category and must be cleared for market through the FDA 510K process. In addition to FDA requirements, ANSI/AAMI RD52 and ANSI/AAMI/ISO 23500:2011 and ANSI/AAMI/ISO 11663:2009 (a voluntary industry standard) defines several requirements for bicarb systems. The ANSI/AAMI standards for hemodialysis have been adopted by the Centers for Medicare and Medicaid Services (CMS) as their conditions for coverage. The interpretive guidance for 42 CFR Part 494 establishes that when the words should or recommend are included in the adopted AAMI standards language, the referenced item must be put into use in the dialysis facility. The latest CMS provisions, defined in 42 CFR Part 494, became effective October 14, 2008. These provisions establish the conditions for coverage of service under Medicare and are the basis for survey activities for the purpose of determining whether an ESRD facility s services may be covered. There are several requirements for bicarbonate concentrate mixing and distribution systems within these provisions. This document summarizes the CMS requirements that will require close attention to ensure that Bicarbonate concentrate mixing and distribution systems are in compliance with the CMS Conditions for Coverage and FDA regulations. Materials Compatibility The components used in bicarbonate concentrate mixing and distribution systems (including mixing and storage tanks, pumps, valves, and piping) must be fabricated from materials that do not interact chemically or physically with the concentrate so as to affect its purity. Furthermore, the materials must be compatible with the chemicals or processes used to disinfect the equipment. Some examples of acceptable materials include plastic and appropriate stainless steel. It is strictly prohibited to use materials known to cause toxicity in hemodialysis, such as copper, brass, galvanized material and aluminum. Design Requirements Bicarbonate concentrate mixing and distribution systems require a purified water source, a suitable drain, and a ground fault protected electrical outlet. The mixing tank must be designed to drain completely, for example, a tank with a sloping bottom and a drain at the lowest point. The tank must also have a tight fitting lid, and must be designed to allow all internal surfaces to be disinfected and rinsed. Because of the highly corrosive nature of the concentrate solutions, the system must be designed and maintained to prevent corrosion. The mixing tank must also include a system to prevent over-mixing, such as a timer integrated into the mixing system for automatic cut-off.

High and low level alarms are not required on the mixing tank, but may be incorporated to help prevent overfilling or to signal low tank conditions. Translucent tanks are recommended to allow users to see the liquid level. Sight tubes should not be used because of the potential for bacteria, algae, and fungi growth. Distribution tanks must be equipped with a conical or bowl shaped bottom, tight fitting lids, a spray mechanism, high and low level alarms and 0.2 micron hydrophobic vent filters. There are two common configurations for distribution. Those are gravity feed and pressurized systems. Gravity feed systems require an elevated tank and pressurized systems use a pump and motor and do not require an elevated tank. Bicarbonate concentrate delivery piping must be color-coded blue at the point of use, and all joints must be sealed to prevent leakage of concentrate. Monitoring & Maintenance Requirements Bicarbonate concentrate mixing and distribution systems must be monitored according to the manufacturer s recommendations. The mixing system must be completely emptied, cleaned and disinfected at least weekly. Bicarb distribution systems must be disinfected at least weekly. The distribution systems must be purged of bicarbonate concentrate before disinfection. Furthermore, the distribution system must be cleared of bicarbonate solution at some point during the treatment day and rinsed clear. This is generally done at the end of the treatment day. Bacteria and endotoxin testing must be performed weekly until a pattern of consistent compliance can be demonstrated (four consecutive compliant results). The frequency of monitoring may then be reduced, but must be performed at least monthly. If elevated levels of bacteria or endotoxin are found, the system must be evaluated, appropriate action taken (including disinfection), and the frequency of monitoring must be increased until a pattern of consistent compliance can be demonstrated. References: Association for the Advancement of Medical Instrumentation. Concentrates for Hemodialysis and Related Therapies (ANSI / AAMI / ISO 13958:2009). Association for the Advancement of Medical Instrumentation. Dialysate for Hemodialysis (ANSI / AAMI RD52:2004). Association for the Advancement of Medical Instrumentation. Guidance for the Preparation and Quality Management of Fluids for Hemodialysis and Related Therapies (ANSI / AAMI / ISO 23500:2011). Association for the Advancement of Medical Instrumentation. Water for Hemodialysis and Related Therapies (ANSI / AAMI / ISO 13959:2009).

Association for the Advancement of Medical Instrumentation. Water Treatment Equipment for Hemodialysis Applications (ANSI / AAMI RD62:2006). Association for the Advancement of Medical Instrumentation. Water Treatment Equipment for Hemodialysis Applications and Related Therapies (ANSI / AAMI / ISO 26722:2009). Centers for Medicare and Medicaid Services (CMS), HHS. US Department of Health and Human Services. Conditions for coverage for end stage renal disease facilities (42 CFR, Part 494) Final Rule 2008. US Food and Drug Administration (FDA). US Department of Health and Human Services. Hemodialysis system and accessories (21 CFR 876.5820) April 1, 2009.

Compliance Checklist This checklist has been compiled by AmeriWater to help you quickly determine if your Bicarbonate system meets the latest standards. Yes No Bicarbonate Concentrate System Materials compatible with concentrates & disinfectants. (No copper, brass, galvanized material or aluminum) Bicarb system has purified water source, suitable drain, and GFI receptacle. Translucent tanks (no sight tubes). Yes No Mix System Designed to drain completely (cone bottom tank with drain at lowest point) Tank has a tight fitting lid (no opening for paddle type mixer). All internal surfaces can be disinfected & rinsed (spray ball). Includes a system to prevent over-mixing (such as a mix timer). Yes No Distribution System Tank with a conical or bowl shaped bottom (drain at lowest point). Tank has a tight fitting lid. All internal surfaces can be disinfected & rinsed (spray ball). Tank includes high and low level alarms. Includes 0.2 micron hydrophobic vent filters. Bicarbonate delivery piping color-coded blue at the point of use. Yes No Monitoring & Maintenance Mixing system completely emptied and disinfected at least weekly. Distribution system purged of bicarb solution & disinfected at least weekly. Distribution system cleared of bicarb solution & rinsed clear daily. Bacteria and endotoxin testing completed at least monthly.