Performance of the Micro Photon Devices PDM 50CT SPAD detector with PicoQuant TCSPC systems

Similar documents
UV MicroTime 200. Crossing the Border towards Deep UV Time-resolved Microscopy of Native Fluorophores

Wavelength LDH - P / D - _ / C / F / FA / TA - N - XXX - _ / B / M / L / XL. Narrow linewidth (on request) Tappered amplified

MicroTime 200 STED. Super-resolution add-on for the confocal time-resolved microscopy platform

Super Resolution Microscopy - Breaking the Diffraction Limit Radiological Research Accelerator Facility

Real-Time Multiplex Kinase Phosphorylation Sensors in Living Cells

Supplementary Table 1. Components of an FCS setup (1PE and 2PE)

Supporting On-line Materials

Toggling Between Blue and Red Emitting Fluorescent Silver Nanoclusters

Department of Chemistry, Center for Photochemical Sciences, Bowling Green State University,

Confocal Microscopy & Imaging Technology. Yan Wu

The new LSM 700 from Carl Zeiss

DeltaPro. DeltaFlex FORENSICS PARTICLE CHARACTERIZATION ELEMENTAL ANALYSIS FLUORESCENCE GRATINGS & OEM SPECTROMETERS OPTICAL COMPONENTS RAMAN

DeltaPro. DeltaFlex FLUORESCENCE FORENSICS PARTICLE CHARACTERIZATION ELEMENTAL ANALYSIS FLUORESCENCE GRATINGS & OEM SPECTROMETERS OPTICAL COMPONENTS

Experts in Femtosecond Laser Technology. DermaInspect. Non-invasive multiphoton tomography of human skin

Nayar Prize I Quarterly Progress Report (Quarters 2&3) August, 2016

STED microscopy with single light source. TeodoraŞcheul

Supporting Information

Using a Scientifica Multiphoton SliceScope for Fluorescence Lifetime Imaging

Analysis of inhomogeneous samples and trace element detection in alloys using QUANTAX Micro-XRF on SEM

Leica TCS SMD Series. Single Molecule Detection Platform for Obtaining Meaningful, Reliable Results: FCS, FCCS, FLIM, FRET, FLCS, Gated FCS

SBDUV-APD Solar-Blind Deep UV APDs

The Sweet PIE (Pulsed Interleaved Excitation)

Fluorescence Nanoscopy 高甫仁 ) Institute of Biophotonics, National Yang Ming University. Outline

Macromolecular environments influence proteins

Confocal Microscopy of Electronic Devices. James Saczuk. Consumer Optical Electronics EE594 02/22/2000

F* techniques: FRAP, FLIP, FRET, FLIM,

Microscopy from Carl Zeiss

Supplementary Figures

Workshop advanced light microscopy

single-molecule fluorescence spectroscopy

Femtosecond micromachining in polymers

Monitoring and Optimizing the Lipopolysaccharides-plasmid DNA interaction by FLIM-FRET

Supporting Information for Matching nanoantenna field confinement to FRET distances enhances Förster energy transfer rates

Picosecond Transient Absorption Spectroscopy System. picotas

Special Techniques 1. Mark Scott FILM Facility

Supporting Information to Carbon Nanodots Towards a Comprehensive Understanding of their Photoluminescence

How to Build the World s Fastest Spectrofluorometer. MF 2 Multi-Frequency Fluorometer

Multiplexed 3D FRET imaging in deep tissue of live embryos Ming Zhao, Xiaoyang Wan, Yu Li, Weibin Zhou and Leilei Peng

In-situ laser-induced contamination monitoring using long-distance microscopy

FLIM Fluorescence Lifetime IMaging

SUPPLEMENTARY INFORMATION

Chapter 4. Fluorescence Studies of Hydroxy Indole Derivatives

Supporting information

Fluorescence & UV- Visible Workshop

Measurement of surface concentration of fluorophores using. fluorescence fluctuation spectroscopy

1. INTRODUCTION 2. EXPERIMENTAL 3. REFERENCES

Supporting Information for. Electrical control of Förster energy transfer.

Skills and excellence formation on basis of Laboratory of Plasma Physics & Atomic Spectroscopy Institute of Spectroscopy (ISAN) of Russian Academy of

Confocal Microscopes. Evolution of Imaging

Fs- Using Ultrafast Lasers to Add New Functionality to Glass

Phosphor Screens. Applications. Features. Taper and glass substrates with phosphor layer.

A cost-effective fluorescence detection system for pulsed laser analysis

OCTOPLUS QPLEX FLUORESCENCE IMAGER. for fast & powerful fast 2D Gel image acquisition

Electronics from the Printing Press

Raman Spectroscopy Measurement System and Data Analysis for Characterization of Composite Overwrapped Pressure Vessels (COPVs)

Building An Ultrafast Photon-Induced Near-field Transmission Electron Microscope

Carbon Black At-line Characterization. Using a Portable Raman Spectrometer

Quantum dots CdSe/ZnS core-shell with maximum emissions of approximately 520 and 600

Bringing Raman Spectroscopy to the Field

Supporting Information

Aluminum / Copper oscillation welding with a 500 W direct diode laser

Quantum Dot applications in Fluorescence Imaging for Calibration and Molecular Imaging

Strong electromagnetic confinement near dielectric microspheres to enhance single-molecule fluorescence

Fabrication of Micro and Nano Structures in Glass using Ultrafast Lasers

Reproducible copper welding

Laser Processing and Characterisation of 3D Diamond Detectors

Advantages of Photodiode Array

Rice/TCU REU on Computational Neuroscience. Fundamentals of Molecular Imaging

Surface characterization of stainless HP-40 steel using laser induced μ- breakdown spectroscopy (μ -LIBS)

Lab 5: Optical trapping and single molecule fluorescence

Supplemental Information. Ca 2+ and Myosin Cycle States Work as Allosteric Effectors of Troponin. Activation

Contact Details. Dr Alexander Galkin. Office: MBC Room 186. Tel: (028) Frequency and wavelength.

Components for Highly Advanced time-resolved fluorescence Microscopy based on Nonlinear Glass fibres

Shaped Femtosecond Laser Pulse Spectroscopy for Nuclear Forensics. Phyllis Ko ANS Student Conference 2011

Time resolved fluorescence spectroscopy: An old technique to monitor protein higher order structure changes

The DCS-120 Confocal Scanning FLIM System

Introduction. (b) (a)

Fabrication of micro/nano structures in glass by lasers

Confocal Microscopy Analyzes Cells

Sub-micron scale patterning of fluorescent. silver nanoclusters using low-power laser

Single Molecules and Single Gold Nanoparticles: Detection and Spectroscopy

PoS(PhotoDet 2012)055

Lambda Square Mapping and FLIM

Intracellular Delivery Using Nanosecond-Laser Excitation of Large-Area Plasmonic Substrates

Supplementary Information

Simultaneous single molecule atomic force and fluorescence lifetime imaging

Picosecond Laser Patterning of ITO Thin Films

Rapid Microtooling with laser based methods

In situ semi-quantitative assessment of single cell viability by resonance

Cell Nucleus-Targeting Zwitterionic Carbon Dots

Supporting Information. Two-Photon Luminescence of Single Colloidal Gold NanoRods: Revealing the Origin of Plasmon Relaxation in Small Nanocrystals

average diameter = 3 nm, from PlasmaChem) was mixed in NLCs to produce QDembedded

Example SimphoSOFT simulation of Thulium-Ion-Doped Pulsed Amplification

Supporting Information. Quantum Wells

Dynamics of Crystallization and Melting under Pressure

Sapphire. Biomolecular Imager THE NEXT GENERATION OF LASER-BASED IMAGING

LASERS, LEDS AND OTHER LIGHTING SOURCES FOR LIFE SCIENCE APPLICATIONS. Wallace Latimer Coherent

Supplementary Note 1: Estimation of the number of the spectroscopic units inside the single Pdots

Scintillating Optical Fibers

Micron-Resolution Photocurrent of CdTe Solar Cells Using Multiple Wavelengths

Transcription:

Technical Note Performance of the Micro Photon Devices PDM 5CT SPAD detector with PicoQuant TCSPC systems Rolf Krahl, Andreas Bülter, Felix Koberling, PicoQuant GmbH These measurements were performed to investigate the ability to use the recently developed thin film CMOS APDs both for ultrafast single photon counting as well as ultra sensitive fluorescence detection in a single molecule spectroscopy system. Detector The PDM 5CT module from Micro Photon Devices (MPD) is an actively quenched Single Photon Avalanche Diode (SPAD), specifically designed for photon counting applications. It is based on CMOS technology and was developed using the long term expertise of Prof. Sergio Cova in Milano together with Microgate Engineering [1,2]. The SPADs of the PDM series are available with three different detector diameters (2, 5 and 1 µm) and can be supplied with a TE cooler to reduce the number of dark counts. We selected for this test a pre-produced cooled detector module with an active diameter of 5 µm compromising between dark counts and the demands of the single molecule system. The manufacturer specifications of the PDM 5CT are given to be: Active area: 5 µm Timing resolution (FWHM): Photon Detection Efficiency: Dark counts: <5 ps 25% @ 4 nm, 47% @ 532 nm, 34% @ 65 nm typ. 75 counts per second Afterpulsing probability:.5% Dead time: Output pulse shape: 7 ns TTL and NIM for improved timing resolution quantum efficiency [%] 6 5 4 3 2 1 4 5 6 7 8 9 1 Fig. 1: Quantum efficiency as a function of wavelength as specified by the manufacturer. Test Setup wavelength [nm] The PDM 5CT was tested in two different setups: In order to test the SPAD for potential applications in a single molecule detection system, we integrated the SPAD in the confocal fluorescence microscope MicroTime 2 [3] whereas to investigate the timing behaviour, we used the following setup: Fig. 2: Experimental setup for the timing analysis of the PDM 5 CT (for details see text). Light pulses from different laser heads of the LDH- PicoQuant 25 TechNote MPD SPAD Page 1

series, driven by the PDL 8-B generator [4] were directly focussed on the active area of the SPAD. The output power of the laser pulses was regulated by the PDL 8-B and if necessary also attenuated by suited ND-optical filters. The NIM-output signal from the SPAD was used because of its faster timing response and directly connected to the PicoHarp 3 [5], which was used for the time-correlated single photon measurements. The reference signal was directly taken from the PDL 8-B. To characterize the SPAD different laser heads of the LDH-series with wavelengths between 375 and 67 nm were used. The UV and blue wavelengths are included into the test because they are of high importance in all fields of cell biology and other bioanalytical applications. Furthermore, as the specified timing resolution is close to the value of a typical MCP-PMT, the measurement was repeated under the same conditions but with a R389U-5 MCP-PMT from Hamamatsu [4] as detector along with the laser head at 67 nm, which shows an exceptionally short pulse width of 32 ps. Instrument Response Function (IRF) The Instrument Response Function for the setup in fig. 2 was measured with five different wavelengths between 375 and 67 nm. Some result of this measurement are shown in fig. 3 and all measured IRF widths along with the respective laser pulse widths are summarized in table 1. counts 6 5 4 3 2 1,4,6,8 1, 1,2 Fig. 3: Wavelength dependence of the IRF. 45 nm 47 nm 67 nm The results clearly indicate that the IRF width depends on the wavelength. In general, all curves show a fast rise time in the order of 1 ps or below. The trailing edges on the other hand are significantly different and the formation of shoulders can clearly be seen for wavelengths at 47 nm and below. These shoulders are the main reason for the larger IRF width at these wavelengths. Wavelength [nm] IRF, FWHM [ps] 375 14 45 45 154 55 47 71 56 635 83 66 67 49 32 Laser pulse, Table 1: Measured IRF widths at different laser wavelengths. However, the IRF width is still significantly smaller as compared to photomultiplier tubes (typ. 2 ps), which are typically used in this spectral region and even much smaller as compared to currently used SPADs of the SPCM-AQR series from Perkin- Elmer (typ. 35-5 ps). For wavelengths above 47 nm, the measured IRF width is within the specifications of the PDM 5CT, if one takes the pulse width of the excitation laser into account. Comparison with MCP-PMT A comparison of the IRF with similar measurements using a R389U-5 MCP-PMT from Hamamatsu [6] is shown in fig. 4 and summarized in table 2. As can be seen, the rising edges of both detectors are nearly identical, whereas the trailing edges of the SPAD are longer, which leads to the slightly larger values of the IRF width. Wavelength [nm] IRF, MPD, IRF, MCP-PMT, 635 83 73 67 49 48 Table 2: Measured IRF width with the PDM SPAD and a MCP- PMT. counts 7 6 5 4 3 2 1,4,8 1,2 1,6 2, 2,4 PDM @ 63 nm MCP-PMT @ 63 nm PDM @ 67 nm MCP-PMT @ 67 nm Fig. 4: Comparison between IRFs of a MCP-PMT and the PDM SPAD. PicoQuant 25 TechNote MPD SPAD Page 2

Instrument Response Function as a function of the count rate All measurements shown above were done with a constant count rate of approx. 1 khz. Real biological samples in scanning microscopy, however, generate a strongly varying count rate. The currently used SPADs from Perkin-Elmer (SPCM-AQR series) show under these conditions a broadening as well as a temporal shift of the IRF. In order to test the corresponding behaviour of the PDM 5CT, the measurements were repeated with different count rates using the 67 nm laser for the excitation. The results are shown in fig. 5. autocorrelation,4,3,2,1, norm. intensity [a.u.] 1,,8,6,4,2 count rate: 12.5 khz max. @ 192 ps, FWHM 52 ps count rate: 375 khz max. @ 184 ps, FWHM 48 ps count rate: 15 khz max. @ 184 ps, FWHM 52 ps,,,2,4,6,8 Fig. 5: Influence of the count rate on the IRF position and width. As can be seen the time shift of the maximum IRF channel is extremely low (less than 1 ps) and also the full width at half maximum (FWHM) of the IRF is almost constant. Afterpulsing For correlation measurements a detector with almost no afterpulsing is needed. Unfortunately, the SPCM-AQR SPADs from Perkin Elmer show a significant afterpulsing probability which makes it very difficult to extract correct particle numbers and triplet state dynamics from an Fluorescence Correlation Spectroscopy (FCS) measurement. Therefore, many users currently prefer to do a cross correlation measurement which is free of this artifact but requires a second detection channel and therefore a substantial financial investment. The afterpulsing measurements were done with a cw-laser running at 635 nm and the TimeHarp 2 [7] in the Time-Tagged Time Resolved (TTTR) mode. The count rate in these experiments was 1 khz. The resulting autocorrelation curve in fig. 6 shows only very little afterpulsing. In detail the afterpulsing ceases at about 1 µs, which means that typical diffusion times can smoothly be measured with a single detector. 1-3 1-2 1-1 1 1 1 lag time [ms] Fig. 6: Autocorrelation function of the PDM 5CT SPAD using uncorrelated light. Applications in a Single Molecule Detection system The biggest advantage of a SPAD compared with a MCP-PMT based system is the high quantum efficiency, which is especially important in the field of Single Molecule Detection (SMD). However, SPADs usually have a relatively small active area onto which the emission light needs to be focussed, which puts some demands on the optical setup. The currently used SPCM-AQR SPADs from Perkin Elmer have an active area of 18 µm, whereas the PDM 5CT has an active area of 5 µm. In a microscope system, like the MicroTime 2 confocal microscope [3], this can be problematic as beampath alignment and confocal pinhole exchange might lead to small beam displacements which in turn can lead to only partial illumination of the SPAD active area. The currently used SPADs from Perkin Elmer, for example, show a broadening of the IRF for out of center focussing. In order to evaluate the influence of the focus adjustment on the active area of the PDM 5 CT, the light focus was moved diagonally across the active area of the SPAD for different pinhole sizes. The results of these measurements are shown in fig. 7. It can clearly be seen, that even at a large pinhole size of 2 µm a displacement of approx. 2 µm in total does not change the count rate at all for the small confocal pinhole size of 5 µm the area of constant count rate even extends to 35 µm. As such relatively large displacements are usually not encountered in a well aligned and stable system, a SPAD with an active area of 5 µm can be very well used in the MicroTime 2 and, in fact, using the sophisticated optics of the MicroTime 2 it was no problem at all to integrate the PDM 5CT into the setup. PicoQuant 25 TechNote MPD SPAD Page 3

counts 12 5 µ m pinhole size 2 µ m pinhole size 1 8 6 4 Fluorescence bursts from single Atto655 molecules which are diffusing statistically through the confocal investigation volume can be clearly seen. The typical count rate is about a factor of two lower compared with the SPCM-AQR SPADS, which is consistent with the theoretical data that shows about 3% quantum efficiency for the PDM 5CT compared to about 62% of the SPCM-AQR SPADs at this wavelength. However, the signal rate is still high enough to calculate the fluorescence correlation curve, as shown in fig. 9. 7 2 6-4 -2 2 4 Fig. 7: Signal loss due to partial illumination of the active area. Two different confocal pinholes were imaged and moved across the SPAD active area with a nominal sixfold image reduction between pinhole and active area. In order to evaluate the possibility of using the SPAD in single molecule applications, Fluorescence Correlation Spectroscopy (FCS) measurements of a very diluted solution (1x1-1 M) of freely diffusing Atto655 molecules in water were performed. The excitation wavelength for these measurements was 64 nm and a 1x 1.3 NA objective and a 5 µm pinhole was used. The emission wavelength of Atto655 is in the range of 68 nm. All measurements were done in the TTTR mode of the TimeHarp 2 board, our "workhorse" TCSPC solution for SMD. In principle, the PicoHarp 3 can also be used for this type of measurements. The fluorescence time trace of these measurements is shown in fig. 8. intensity [counts/ms] 35 3 25 2 15 1 5 displacement [µ m],,1,2,3,4,5,6 time [s] Fig. 8: Fluorescence time trace of single Atto655 molecules diffusing through the confocal detection volume. normalized autocorrelation 5 4 3 2 1 1E-3,1,1 1 1 1 Fig. 9: FCS curve calculated from the MCS trace in fig. 8 for a 1 s measurement. Summary lag time [ms] The MPD PDM 5CT SPAD has a very good timing response, which is comparable to MCP-PMTs. One striking feature compared to the currently used SPADs of the SPCM-AQR Series from Perkin Elmer is the independence of the temporal position and the full width at half maximum of the Instrument Response Function (IRF) from the signal count rate, which is of high importance for single molecule applications and Fluorescence Lifetime Imaging (FLIM). The PDM 5CT is by far more sensitive in the blue spectral region compared to the SPCM-AQR SPADs, whereas the spectral sensitivity in the red is lower, but still high enough to allow for single molecule measurements. The SPAD has been found to be very robust even exposure to daylight during operation did not do any damage. The integration of these SPADs into the MicroTime 2 is not problematic at all, especially as the position of the focus on the active area is relatively uncritical compared to the SPCM- AQR SPADs. First measurements showed, that the PDM 5CT can very well be used for single molecule measurements. Besides single molecule applications, the SPADs can also be used to replace expensive MCP-PMTs for fast timing applications, e.g. for laser pulse characterisation. The active area of the SPAD is, however, too small for ultra sensitive detection of large volume type samples (e.g. cuvettes) in spectrometer applications. PicoQuant 25 TechNote MPD SPAD Page 4

Acknowledgement We would like to thank Micro Photon Devices for giving us the opportunity to test the PDM 5 CT. Further reading [1] Bertone N., Biasi R., Dion B., Laser Focus World, p. 93, April 25 [2] Cova S., Ghioni M., Lotito A., Rech I., Zappa F., Journal of Modern Optics, Vol. 51, p. 1267-1288 (24) [3] Wahl, M., Koberling, F., Patting, M., Rahn, H., Erdmann, R., Curr. Pharm. Biotech., Vol. 5, p. 299-38 (24) and http://www.picoquant.com/products/microtime2/microtime2.htm [4] Erdmann R., Wahl M., Lauritsen K., Enderlein J., Legendre, Jr. B. L., Williams D. C., Soper S. A., Proceedings of SPIE, Vol.2985, p.138-144 (1997) and http://www.picoquant.com/products/pdl8b/pdl8b.htm [5] see http://www.picoquant.com/products/picoharp3/picoharp3.htm [6] see http://www.hamamatsu.com [7] Wahl M., Erdmann R., Lauritsen K., Rahn H.-J., Proceedings of SPIE, Vol.3259, p.173-178 (1998) and http://www.picoquant.com/products/timeharp2/timeharp2.htm Copyright of this document belongs to PicoQuant GmbH. No parts of it may be reproduced, translated or transferred to third parties without written permission of PicoQuant GmbH. PicoQuant GmbH Unternehmen für optoelektronische Forschung und Entwicklung Rudower Chaussee 29 (IGZ), 12489 Berlin, Germany Telephone: +49 / ()3 / 6392 656 Fax: +49 / ()3 / 6392 6561 e-mail: photonics@pq.fta-berlin.de WWW: http://www.picoquant.com PicoQuant 25 TechNote MPD SPAD Page 5