testo 350 M/XL, flue gas measuring system TECHRENTAL NOTE: Pages selected from Testo catalogue Options not included with TR assets are removed or "greyed-out" mith Ltd. Cont. switch-off function PC analysis Data communication via PC, barcode pen printer Prints readings on site 4 user defined function buttons 0 11.2 Burner 5/1 324 23.6 C % H2 522 292 193 % Ptop Zoom Gas dp = 0 Milltown Location: Burner 5 002 02/06 Robust housing with impact protection Print Large memory testo 350 is a flexible, portable measuring system for a broad range of combustion applications. This advanced analyser can be used to adjust all types of industrial burners measure concentrations in raw and clean gas over a longer time period check the atmospheres of all types of process furnaces maintain stationary motors such as modular cogeneration stations check gas pressures and gas velocities in flue gas and/or fresh air ducts. V Testo data bus User defined probe socket Connection for flue gas probe + additional temperature probe Robust universal plug for flue gas and differential pressure connection differential pressure/velocity Peltier gas preparation unit testo 350 M 0563 0351 testo 350 XL 0563 0350 Particle filter differential pressure probe The testo 350 M basic version includes control unit, analysis box and a flue gas probe. O2,, NO (optional, NO2, O2 (optional or 2 are measured via the infrared mesurement module (optional(max.4 measuring modules. Temperature and differential pressure, as well as the usual calculated variables 2 etc. are also measured. The detachable control unit can also be used as a separate analyser for temperature, velocity, differential pressure, relative humidity etc. The readings can be printed on the integrated printer. A complete Peltier gas preparation unit for controlled condensate removal is also built into the analysis box. The more advanced testo 350 XL includes, in addition to the M version, NO and NO2 parameters, an event trigger socket, a fresh air valve as well as the option of upgrading an additional 2 measuring modules. The, H2, O2 or 2 measuring modules are available in the XL version. Additional advantages at a glance Battery operation, also with gas preparation unit (up to 2-3 h with data memory function, also without control unit Easy and fast operation via touchscreen (optional Measuring range extension (optional for uitable flue gas probe available for every application Extremely high accuracies for low NOx and thanks to special gas sensors Easy handling due to low weight (approx. 4.5 kg and compact size Can withstand tough conditions thanks to N m
272 testo 350 M/XL, Portable flue gas analyser testo 350 M/XL C NO/N The portable concept An analyser must be easy to transport and robust to make it suitable for portable applications in industrial systems. A common desire is to have a remote display at the burner controls. For your advantage, the control unit in testo 350 M/XL is detachable and can be operated at up to 50 m away from the analysis box. Gas conditioning for industrial applications is needed whether you are carrying out a spot check or long-term s lasting for several hours. Testo s integrated gas conditioning system achieves accurate s of NOx or O2 and the analyser is also protected from corrosive condensate from the flue gas. Fast, easy inexpensive maintenance is assured! The user can exchange consumables such as measuring cells on site. Testo has a worldwide spare parts delivery service., %RH ΔP, Pa The control unit can operate the entire system and read out data or be used as a stand alone hand-held instrument. It alone can measure differential pressure, temperature, humidity, velocity etc. Additional benefits: PC interface (R232 to read out data Backlit graphic display User-friendly, menu guided operation with user defined function buttons data memory (1 MB Printout of current or saved data Touchscreen operation (option for quick operation and input Magnets on the rear for positioning on location Robust housing with impact protection Power supply via analysis box, replaceable battery or mains unit, C (IAQ,, NO X, O,... ΔP, Pa The analysis box is the heart of the analyser. The following is inside the housing: Respective gas sensors and differential pressure ample gas pump Peltier gas preparation unit (with hose pump for disposing of condensate switch/off (rinse function to protect the sensor Rechargeable NiMH battery (without memory effect Built-in mains unit (110/230 V, 50 to 60Hz Data memory (1MB Options such as fresh air valve (for longterm s with testo 350 M, XL included Can be used for portable operation in the following 3 positions: Vertical position (e.g. lying in case Edgewise (as data logger in connection with wall holder Hanging down from carrying handle
276 Gas sampling probes C ampling probes have to endure extreme conditions when measuring flue gases for example: High temperatures Corrosive condensate Dust Mechanical loads. The selection of the right probe is critical for accurate and consistent s. Because the sampling locations are often different, it s beneficial to have a standard probe designed for a wide variety of applications. In addition to the standard sampling probes, Testo also offers probe systems for specific industrial applications. tandard sampling probes NO/N The standard sampling probe is available in lengths of 335 mm and 700 mm and for different temperature ranges. The outer pipe with a sintered filter is used for dusty flue gases. The hose has a standard length of 2.2 m (5 m optional. 335/700 mm long Ø 8 mm T max 500/1000 Material: probe pipe T max +500 : tainless steel 1.4361 Material: probe pipe T max +1000 : tainless steel 1.4841 Option: Outer pipe with filter for dusty flue gases Hose length: tandard 2.2 m / 5 m optional 335/700 mm long Ø 8 mm Pore size: 3 μm / Material of probe pipe: tainless steel 1.4841 Ø 10 mm T max 800 tandard probe: 335 mm long tandard probe: 700 mm long Basic sampling probe: 335 mm long, T max +500 0600 7451 Basic sampling probe: 700 mm long, T max +500 0600 7452 Options Options Higher flue gas temperature/dusty flue gases: Higher flue gas temperature/dusty flue gases: Outer pipe with filter for dusty flue gases, Tmax +800 0440 7435 or Outer pipe for high flue gas temperature, Tmax +1000 0440 7437 Hose: Hose length 5 m (not for O2 s 0440 7443 pecial hose for O2 s 1, 2.2 m long 0440 7442 pecial hose for O2 s 1, 5 m long 0440 7445 1 Use outer pipe with filter for dusty flue gases. Outer pipe with filter for dusty flue gases, Tmax +800 0440 7436 or Outer pipe for high flue gas temperature, Tmax +1000 0440 7438 Hose: Hose length: 5 m (not for O2 s 0440 7444 pecial hose for O2 s 1, 2.2 m long 0440 7442 pecial hose for O2 s 1, 5 m long 0440 7446
278 Measurement system and practical accessories Logger, measures and saves (max. 250,000 readings 0577 4540 Options for NO2, O2,..., can also be upgraded on site at a later time 2 module, 0 to 50 vol.%, infrared principle, absolute pressure, 2 absorption filter with refill pack 0440 0417 Protective cover for analysis box 0554 0199 Wall holder for analysis box, incl. heat protection plate 0554 0203 C NO/N Touchscreen with pen 0440 0559 Ordering data: Measurement system and accessories testo 350, control unit displays data and controls the measuring system, incl. built-in printer, pressure 80/200, 1 user defined probe socket, programmable s and memory space for 250,000 readings, connection for Testo data bus, incl. terminal plug Touch screen with pen (available only with original order For easy input of text and values pare thermal paper for printer (6 rolls Barcode pen to read in locations Quick and accurate allocation of reading to site Barcode labels, self-adhesive (1200 off Location marked with barcode, printed using software Adhesive pockets (50 off for printout, paper barcode labels... IRDA interface, from control unit to PC For direct online transfer of readings to PC Analog output box ( out Testo rechargeable battery pack NiMH for control unit, logger Mains unit 230 V/ 8 V/ 1 A, for instrument (European plug 0554 0845 0440 0559 0554 0569 0554 0460 0554 0411 0554 0116 0440 0560 0515 0097 0554 1084 Transport case 0516 0351 ystem case (aluminium, incl. drawer 0516 0352 Additional box for system case 0516 0352, can be snapped on 0516 0353 Ordering data: Measurement system and accessories testo 350 XL analysis box and equipment testo 350 XL analysis box, equipped with O2, (with switch-off and rinse function, NO, NO2, differential pressure, 2 temperature probe sockets, gas preparation, Testo data bus connection, automatic fresh air rinse with valve, integrated rechargeable battery, data memory, can be upgraded to max. 6 modules (With H2,, O2 low module, 0 to 500, highly accurate, instead of standard module, built into analysis box 2 module, 0 to 50 vol.%, infrared principle, absolute pressure, 2 absorption filter with refill pack NOlow module, 0 to 300, highly accurate, instead of standard NO module, built into analysis box O2 module, built into analysis box module (nonburned hydrocarbons, built into analysis box H2 module, built into analysis box Measuring range extension for module (dilution, built into analysis box, selectable dilution factors: 0, 2, 5, 10, 20, 40 Event trigger socket, for starting and stopping externally, built into analysis box 0563 0350 0440 3925 0440 0417 0440 3934 0440 3927 0440 3929 0440 3930 0440 0555 0440 3932 testo 350 M analysis box and equipment testo 350 M analysis box, with O2, (with switch-off and rinse functions, gas preparation, diff. pressure meas., 2 temperature probe sockets, can be upgraded to max. 4 modules (with NO/NO2/O2, Testo data bus connection, built-in rech. batt., data memory low module, 0 to 500, highly accurate, instead of standard module, built into analysis box 0563 0351 0440 3925 2 module, 0 to 50 vol.%, infrared 0440 0417 principle, absolute pressure, 2 absorption filter with refill pack NO measuring module, 0 to 3000, built into analysis box 0440 3935 NOlow module, 0 to 300, highly accurate, built into analysis box 0440 3928 NO2 measuring module, built into analysis box 0440 3926 O2 module, built into analysis box 0440 3927 Fresh air valve for long-term, built into analysis box 0440 0557 Measuring range extension for module (dilution, built into analysis box, selectable dilution factors: 0, 2, 5, 10, 20, 40 0440 0555 Accessories for analysis boxes Wall holder for analysis box incl. heat protection plate, can be locked Protective cover for analysis box (can also be used with wall holder Carrying belt set for analysis box and control unit Transport case for analyser, probes and accessories ystem case (aluminium, with drawer for accessories, for transport and protection during Additional box for system case 0516 0352, can be snapped on IO calibration certificate/flue gas Calculation of fuel-specific factors to accurately display calculated variables in deviating fuels (calculation for one fuel pare particle filter, pack of 20 Hose set to convey flue gas from analysis box, 5 m long Refill pack of filter pellets for 2 absorption filter 0554 0203 0554 0199 0554 0434 0516 0351 0516 0352 0516 0353 0520 0003 0991 0030 0554 3381 0554 0451 0554 0369
Technical data Technical data/testo 350 M analysis box, testo 350 XL Temperature (H2 compensated low meas. (H2 compensated C NO meas. (option for testo 350 M NOlow NO2 measuring module (option for testo 350 M O2 Meas. range -40 to +1200 0 to +25 0 to +10000 0 to +500 0 to C max Vol. % C 0 to +3000 NO 0 to +300 NO 0 to +500 N 0 to +5000 ±0.5% of mv (+100 to +1200 ±0.5 (-40 to +99.9 ±0.8% of f.v. +25 ±5% of mv (+200 to +2000 ±10% of mv (+2001 to +10000 ±10 +199 ±5% of mv (+40 to +500 ±2 +39.9 Calculated from to +1999.9 NO ±10% of mv (+2000 to +3000 NO ±5 NO +99 NO ±5% of mv (+40 to +300 NO ±2 NO +39.9 NO to +500 N ±5 N +99.9 N to +2000 ±10% of mv (+2001 to +5000 ±5 +99 Resolution 0.1 (-40 to +1200 0.01 (0 to +25 1 +10000 0.1 +500 0.01 C 1 NO +3000 NO 0.1 NO +300 NO 0.1 N +500 N 1 +5000 Reaction time 20 s 40 s 40 s 20 s 40 s V Reaction type t 95 t 95 Efficiency Flue gas loss Differential pressure 1 Differential pressure 2 Velocity C (IR Meas. range 0 to +120 % -20 to +99.9 % -200 to +200-40 to +40 0 to +40 0 to 50 C ±1.5% of mv (-50 to -200 ±1.5% of mv (+50 to +200 ±0.5 (-49.9 to +49.9 ±1.5% of mv (-40 to -3 ±1.5% of mv (+3 to +40 ±0.03 (-2.99 to +2.99 ±0.3 C + 1% of mv 25 C ±0.5 C + 1.5% of mv (>25 to 50 C N Resolution 0.1 % +120 % 0.1 % (-20 to +99.9 % 0.1 (-200 to +200 0.01 (-40 to +40 0.1 +40 0.01 C 25 C 0.1 C (>25 C Reaction time <10 s Reaction type Dimensions 395 x 275 x 95 mm Weight 3200 g torage temp. -20 to +50 Oper. temp. -5 to +45 Material/Housing AB Additional technical data: Memory: 250 000 readings Power supply: Via integrated mains unit (90 V to 260 V, 47 to 63 Hz or exchangeable rechargeable batteries Electrical power required: 0.5 A (110 V AC, 0.3 A (230 V AC Dew point calculation: 0 to 99 td Maximum positive pressure of flue gas: 50 (500 mm water column Maximum negative pressure: 200 (2000 mm water column Pump flow: 0.8 with flow monitoring Max. dust load: 20 g/m³ dust in flue gas Max. humidity load: +70 Dew point temperature at analyser box inlet Measuring range extension (dilution for : dilution factors 0, 2, 5, 10, 20, 40 Dilution gas: Fresh air or N2 : Reading plus max. 2% Trigger input testo 350XL: Voltage: 5 to 12 volt (rising or falling edge Pulse width > 1 second Load: 5 V/max. 5, 12 V/max. 40 Warranty: Analysers: 2 years (excluding working parts e.g. measuring cells... /NO/NO2/O2/H2//2 sensor: 1 year O2 measuring cell: 1.5 years Additional technical data for testo 350 XL analysis box only Meas. range Additional technical data: Trigger input: 5 to 12 V (rising or falling edge H2 0 to +300 ±5% of mv (+40 to +300 ±2 +39.9 Resolution 0.1 +300 Reaction time 35 s Reaction type Technical data for module Parameter Measurement range 100 to 40,000 Resolution Min. 02 requirement in flue gas Methane less than 400 (100 to 4000 less than 10 % of m.v. (greater than 4000 10 2% + (2 x methane reading Reaction time t90 less than 40 s Response factor * 1 Propane 100 to 21,000 less than 400 (100 to 4000 less than 10 % of m.v. (greater than 4000 10 2% + (5 x propane reading less than 40 s 1.5 Butane 100 to 18,000 less than 400 (100 to 4000 less than 10 % of m.v. (greater than 4000 10 2% + (6.5 x butane reading less than 40 s 2 ¹ Lower explosion limit must be adhered to. ² The module is adjusted to methane in the factory. It can be adjusted to another gas by the user. m