INTERNATIONAL STANDARDS ON FLUID PREPARATION FOR CONVECTIVE THERAPIES. Richard A. Ward, Ph.D.

Similar documents
International Water Standards

Evolution of AAMI/ANSI Standards: Harmonizing with ISO. V. Taaffe 2018

photo: Ameriwater.com Dialysis Water Treatment Equipment Monitoring December 2017 Developed by the BCPRA Hemodialysis Committee

PROVINCIAL STANDARDS & GUIDELINES

Appendix: Water purification methods

Hot Topics in Drug Product Process Validation: A Reviewer s Perspective

Water and Dialysate Safety: Your Role in Maintaining Quality. Glenda Payne, MS, RN, CNN

DIALYSIS WATER TREATMENT DIALYSIS PROCESS WATER SOLUTIONS THE PREMIER PROVIDER OF WATER TREATMENT SOLUTIONS TO THE DIALYSIS INDUSTRY

Compounding Pharmacies and Water

Dialysis Water PREVIEW and Dialysate COPY Recommendations:

Annex A2. Guidance on Process Validation Scheme for Aseptically Processed Products

USP Perspective on Pharmaceutical Waters. Antonio Hernandez-Cardoso, M.Sc. Senior Scientific Liaison September 7, 2017

HaemoDialysis: New Technology. Peter G Kerr Monash Medical Centre and Monash University Clayton, Vic, Australia

The new AK 96 machine

Dialysis Water, The Essential Basics February 27, 2003

Microbiological Quality: Understanding Drinking Water Quality and Management

Water for Instrument Processing

Customer Application Brief. Bioprocess, Biologicals, & Pharmaceutical. The Use Of Cartridge Filtration in the Production Of Pharmaceutical Grade Water

SOUTH AFRICAN NATIONAL STANDARD

Dialysis water purification

Overview of a sterility assurance program for PET drugs

UltraPure Water Application Culture of Nerve Cells

GUIDE TO INSPECTIONS OF HIGH PURITY WATER SYSTEMS

Sterile Compounding elearning Course Curriculum 28 lessons with 33 hours of CE. Fundamentals of Sterile Compounding (8 lessons/8 hours CE)

Tests to Support Sterility Claim. Imtiaz Ahmed

Quantitative Microbiological Monitoring of Hemodialysis Fluids: Evaluation of Methods and Demonstration of Lack of

Membrane Filtration Technology: Meeting Today s Water Treatment Challenges

An AAMI Webinar Series:

Aseptic Process Validation

Water Use in the Food Industry

CMS, FDA and AAMI Requirements for Bicarbonate Systems Made Simple. Compiled January 2012 by AmeriWater Incorporated

Microbiological Cleaning Method Validation

MDSAP Purchasing Process

Filtration in Pharmaceutical Water Systems

Purified Water Solutions for Dialysis WATER TECHNOLOGIES

arium mini Ultrapure Water System

Disclaimer. Healthcare facilities are required to have written policies approved by their governing body concerning the services that they provide.

AFS 8D / 16D Degassed Water Purification Systems

arium mini Ultrapure Water System

Protect MORE than just your drinking water. Purify water for your entire home.

SA 450/60 Material Data Sheet

ISO How to translate into action Sterile container systems

Christina Bradley Endoscope Decontamination A Webber Training Teleclass. Christina Bradley

Environmental Monitoring of Aseptic Processing Areas - 1

a solution for reducing the

Requirements of validation of packaging systems and sterility assurance

Validation of Sterilizing Grade Filters

Purified Water Systems

Revaclear. For best in class high flux performance. Smarter dialyzer technology

ISO INTERNATIONAL STANDARD

Flexible Endoscopes: The A,B,C s of Monitoring Manual Cleaning Efficacy!

GUIDE TO INSPECTIONS OF STERILE DRUG SUBSTANCE MANUFACTURERS

REUSABLE MEDICAL DEVICE VALIDATION

Guidelines for Selection and Use of Disinfectants

Water Based Critical Utilities. RO, WFI, and Steam

Dentistry Test methods for dental unit waterline biofilm treatment

USP Chapter 823 USP 32 (old) vs. USP 35 (new)

Ultra pure water, 18,2 M cm. TKA GenPure. For Life Sciences, Analyses and Laboratory applications WATER PURIFICATION SYSTEMS

Advantage System Elix The Best in Pure Water

THE CLEAR ADVANTAGE. INNOVATIVE FILTERS FOR process and potable water

Whitepaper. High temperature HEPA filtration. Dr.-Ing. Marc Schmidt, Dr.-Ing. Lothar Gail, Hugo Hemel MSc. Preview

RiOs Essential 5, 8, 16, 24 Water Purification Systems

Questions and Answers on allogenic stem cell-based products for veterinary use: specific questions on sterility

STERILIZATION VALIDATION IN THIS SECTION REPROCESSING VALIDATIONS FOR REUSABLE MEDICAL DEVICES. Radiation Sterilization Validation

FREQUENTLY ASKED QUESTIONS FOR THE ADVANTAGE

AK 95 S The benefits of experience? A perfect fit

On-line Preparation of Solutions for Dialysis: Practical Aspects Versus Safety and Regulations

Fully Comprehensive Premium Range

THANKS TO T.C. SOLI, Ph.D. T. C. Soli, President, Consulting & Training Services since 2004 after 25 years in big pharma

Microbiological Consideration for Non-Sterile Pharmaceutical

ISO INTERNATIONAL STANDARD

P URE PRO DRINKING WATER SYSTEM USER'S MANUAL REVERSE OSMOSIS SYSTEM. Type of product. Date of purchase. Address

Instantaneous Microbial Detection for Pharmaceutical Waters. IMD-W Instantaneous Microbial Detection System

ISO INTERNATIONAL STANDARD

Universidad Centroamericana José Simeón Cañas Dept. of Process Engineering and Environmental Sciences

High Quality Pipette Tips for Consistent and Reliable Results

4400M Digital Reverse Osmosis System

Mar Cor Purification. Carbon Block Filtration System. Operation & Maintenance Manual

California Leafy Greens Handlers Marketing Agreement (CA LGMA) Water Provisions

WHO GOOD MANUFACTURING PRACTICES: WATER FOR PHARMACEUTICAL USE PROPOSAL FOR REVISION DRAFT FOR COMMENTS

AAMI Quality Systems White Paper: Comparison of 21 CFR Part 820 to ISO 13485:2016 1

AK 95 S Dialysis machine. Part of Gambro s complete HD system

Best Practices for Reprocessing Endoscopes

MicroFunnel Filter Funnels

Review Validation of aseptic processes for pharmaceuticals

Validation Study on How to Avoid Microbial Contamination during Pharmaceutical Production

Ultrafiltration Technical Manual

Save your money and the environment!!!! BIONEST onsite waste water treatment solutions

Aseptic Processing Current Issues & Trends

TUV SUD BABT PRODUCTION QUALITY CERTIFICATION SCHEME

EU and FDA GMP Regulations: Overview and Comparison

for IND and RDRC Regulated PET Compounding

ISO INTERNATIONAL STANDARD

ISO/TC 198. Secretariat: ANSI

Sterilization & Disinfection

PHCT 401 Aseptic Processes & Techniques

Direct Water purification system Pure & ultrapure water directly from tap water

Company Profile. So Pure Enterprises FZC

Transcription:

INTERNATIONAL STANDARDS ON FLUID PREPARATION FOR CONVECTIVE THERAPIES Richard A. Ward, Ph.D.

THE EVOLUTION OF CONVECTIVE THERAPIES Convective therapies enhance large molecule removal by using ultrafiltration rates exceeding those required to achieve dry weight; volume balance is maintained by infusing substitution solution directly into the blood line. In the original practice of convective therapies, substitution solution was provided using pre-packaged bags or bottles of fluid sterilized using conventional terminal sterilization methods. The logistics of using pre-packaged fluids restricted the volume of substitution solution that could be used for a single treatment. This limitation was removed by the development of systems that prepared substitution solution on-line by sequential filtration of dialysate through bacteria- and endotoxin-retentive filters. The fluid produced by these systems should comply with relevant quality standards.

INTERNATIONAL STANDARDS FOR HEMODIALYSIS AND RELATED THERAPIES ISO 13958:2009, Concentrates for haemodialysis and related therapies ISO 13959:2009, Water for haemodialysis and related therapies ISO 11663:2009, Quality of dialysis fluid for haemodialysis and related therapies ISO 26722:2009, Water treatment equipment for haemodialysis applications and related therapies ISO 23500:2011, Guidance for the preparation and quality management of fluids for haemodialysis and related therapies

QUALITY STANDARD FOR SUBSTITUTION SOLUTION According to ISO 11663:2009, Clause 4.1.4, Microbiological requirements for on-line prepared substitution fluid: This fluid shall be sterile and non-pyrogenic. Sterile: Free from viable microorganisms with a sterility assurance level (SAL) of 10-6. One in a million chance of contamination. Non-pyrogenic: Having less than 0.03 EU/mL [of endotoxin]. 5 EU/kg/h is considered the minimum dose required to produce a fever. For a 70 kg person receiving 6 L/h of substitution solution, an endotoxin concentration of 0.03 EU/mL corresponds to a dose of 2.6 EU/kg/h.

COMPLIANCE WITH THE QUALITY STANDARD Compliance of online produced fluid with the requirements of this International Standard [ISO 11663:2009] cannot be demonstrated with traditional test procedures. For this reason, compliance with this International Standard shall be ensured by proper operation of a validated system, verified according to the manufacturer's instructions at the time of installation, and confirmed by the user with a regular monitoring and maintenance schedule. ISO 23500:2011 provides guidance to users on how to routinely achieve compliance with the quality requirements of ISO 11663:2009.

ISO 23500:2011 Guidance for the preparation and quality management of fluids for haemodialysis and related therapies 1. Scope 2. Normative references 3. Terms and definitions 4. Summary of quality requirements of ISO 13958, ISO 13959 and ISO 11663 5. Critical aspects of system design 6. Validation of system performance 7. Quality management 8. Strategies for microbiological control 9. Environment 10. Personnel

PREPARATION OF SUBSTITUTION SOLUTION BY SEQUENTIAL FILTRATION Substitution solution is produced by sequential filtration through bacteria- and endotoxin-retentive filters capable of reducing the level of bacteria and endotoxin by > 10 6 10 7 and 10 3 10 4, respectively The manufacturer s validation of these systems is applicable only when they are operated under specified conditions. In particular, the incoming fluid quality must comply with the maximum contaminant levels set by the manufacturer of the system. For example: Dialysis water contains < 100 CFU/mL and < 0.25 EU/mL; or, Standard dialysis fluid contains < 1000 CFU/mL and < 1 EU/mL.

PREPARATION OF SUBSTITUTION SOLUTION BY SEQUENTIAL FILTRATION The challenge for a dialysis facility is to establish a Quality Management System that ensures the manufacturer s requirements for the incoming fluid are routinely met.

BLENDING VALVE REVERSE OSMOSIS DEPTH FILTER CARBON SOFTENER 5µm FILTER MUNICIPAL WATER DIALYSIS MACHINE DIALYZER DIALYSIS STATIONS STORAGE TANK DISTRIBUTION LOOP CONCENTRATE The goal of the Quality Management System is to maintain an intact hygenic chain between the product water side of the reverse osmosis module and the point at which dialysis fluid enters the dialyzer or becomes substitution solution.

QUALITY MANAGEMENT SYSTEM The Quality Management System should be designed to prevent bacterial infiltration and limit biofilm formation. Critical aspects of the Quality Management System include: Equipment design (ISO 23500:2011, Annex B); Installation and operational verification of equipment (ISO 23500:2011 clause 6); Maintenance of equipment (ISO 23500:2011 clause 8); and, Monitoring of system performance (ISO 23500:2011 clauses 7, 8 and Annexes C and D).

SYSTEM DESIGN All fluid handing systems should be designed to allow frequent disinfection. frequent disinfection. The preferred methods are hot water pasteurization and ozone since neither requires extensive rinsing to eliminate residuals. All fluid handling systems should be constructed from materials that allow the use of hot water or ozone.

SYSTEM OPERATION The operation of all fluid preparation systems (dialysis water, concentrate, dialysis fluid) should be subject to formal documentation that: Clearly and concisely describes the systems and their operational status; Provides detailed procedures for ongoing maintenance and operational verification of the systems performance; Defines responsibility for the systems; and, Establishes training requirements for all personnel involved in any aspect of fluid preparation. This documentation should be consistent with the manufacturer s instructions for use of the system used to prepare the substitution solution.

INSTALLATION AND OPERATIONAL VERIFICATION AND QUALIFICATION Installation should be followed by: Formal verification that the installed system matches the pre-approved plans. Verification that system performance meets the functional specifications of the system, including the operation of safety systems and compliance with applicable standards for all contaminants. A period of normal operation during which there is frequent data acquisition to demonstrate consistent performance.

SYSTEM MAINTENANCE The single most important aspect of maintenance is regular disinfection to suppress the formation of mature biofilm. Once established, biofilms are extremely difficult to eliminate. Therefore, disinfection must be proactive designed to suppress biofilm formation, not eliminate biofilm after it has formed. Hot water pasteurization or ozone are preferred. They are effective against biofilm and allow frequent disinfection because they leave no chemical residuals (hot water) or residuals with a very short half-life (ozone).

ESTABLISHING MAINTENANCE AND MONITORING SCHEDULES Data obtained during the initial period of performance qualification should be used to establish initial disinfection schedules, monitoring plans, and action levels. CAUTION: Cultures and endotoxin tests may not accurately reflect the bacterial burden in a newly installed system because biofilm takes time to form and mature.

MONITORING The purpose of monitoring is to demonstrate the adequacy of maintenance procedures and to ensure that the inputs to the system used to prepare substitution solution meet the requirements of that system s manufacturer. Chemical contaminants: Rejection by reverse osmosis and conductivity of product water (continuous) Total chlorine (at least daily when present as monochloramine) Chemical analysis (at least annually) Microbial contaminants: Cultures Endotoxin testing

MONITORING FOR MICROBIAL CONTAMINANTS Available culturing methods are not sufficiently sensitive to demonstrate sterility by testing the final substitution solution. Sterility of the substitution solution is assured by using a validated system and testing the inputs to that system to show that they comply with the system manufacturer s specifications. Validated system of bacteria- and endotoxin-retentive filters TEST HERE Substitution Solution Blood Line Inputs (dialysis water, dialysis fluid)

SUMMARY Equipment to perform on-line convective therapies is widely available and has been shown to perform safely and effectively in large numbers of patients in multiple centers and over long periods. Equipment for on-line convective therapies incorporates many safety systems; however, the user is responsible for certain residual risks. One important user responsibility is maintaining the quality of the fluids supplying the on-line equipment within the limits for which the equipment performance has been validated. This responsibility requires the user to establish a Quality Management System that encompasses all aspects of fluid preparation, including system design, performance validation, on-going performance verification, and staff training. ISO 23500:2011 provides a road map to guide users in establishing their quality management system.