MODEL 3340 LASER AEROSOL SPECTROMETER

Similar documents
Steven L. Alderman, Michael S. Parsons, and Charles A. Waggoner. Institute for Clean Energy Technology, Mississippi State University

MEASUREMENT OF BRAKE WEAR EMISSIONS BY THE ENGINE EXHAUST PARTICLE SIZER (EEPS) SPECTROMETER AND OPTICAL PARTICLE SIZER (OPS)

To: PMP Working Group From: Chris Parkin

Joint Committee Issue Paper

Development of Test Stand and Test Plan for Qualification of Sintered Fiber Metal Media Filters

Product data sheet Palas Condensation particle counter UF-CPC 100

Validation Guide USTR 2114 (2) Validation Guide for Pall Emflon PFR Filter Cartridges

Wierzbicka, Aneta; Gudmundsson, Anders; Pagels, Joakim; Dahl, Andreas; Löndahl, Jakob; Bohgard, Mats

THE WORLD LEADER IN CLEAN AIR SOLUTIONS. Pharmaceutical Clean Air Solutions PARTICULATE AND GASEOUS FILTRATION

THE LEADER IN PERFORMANCE INDOOR AIR QUALITY MEASUREMENTS

WM2014 Conference, March 2 6, 2014, Phoenix, Arizona, USA

ASEPTIC BIOPHARMACEUTICAL MANUFACTURING: FAQs

DEKATI EJECTOR DILUTER IN EXHAUST MEASUREMENTS V2.11

Dekati. ELPI Electrical Low Pressure Impactor. Real-time particle size distribution Wide particle size range Wide range of applications

CHAPTER 6 : GAS SAMPLING SYSTEMS

PAO Compatible eptfe Technology HEPA Filters for Cleanroom Pharmaceutical Applications

Release characteristics of single-wall carbon nanotubes during manufacturing and handling

A LOW PRESSURE DROP SINTERED METAL FILTER FOR ULTRA-HIGH PURITY GAS SYSTEMS

Line Filters Solid, simple, smart. Advanced reliability in compressed air. TECHNOLOGY YOU CAN TRUST

Inter-laboratory tests of the methodology for filtration efficiency tests in different filter media against nanoparticles

Performance Evaluation of a Novel Sampling and Measurement System for Exhaust Particle Characterization

Counting the Way to Cleaner Air

METHOD 201A - DETERMINATION OF PM EMISSIONS. 1.1 Applicability. This method applies to the in-stack measurement of

Special reprint from Gefahrstoffe Reinhaltung der Luft, Ausgabe 78 (2018) Nr. 3, Seiten Sonderdrucke aus Fachzeitschriften

Chambers and Instrumentation CHAMBERS. STORA: 22 m 3 chamber

0.01 Micron Membrane Filters

Particle precipitation in medium- and large-scale biomass combustion plants

ABSTRACT INTRODUCTION

An Automated System for At-Line Process Analysis of Biopharmaceutical Fermentation Reactions

Compressed Air Filters 1/4 and 1/2 Line Size Filters

289 Series Spring-Loaded Relief Valves

Microfil GP. Range. Validation Guide. World Class Filtration Solutions

The Leader in Performance Indoor Air Quality Measurements

Recent Trends for Filter Development for Diesel Particulate Aftertreatment

F-Series 1 Elements Performance Validation

LABORATORY CONTROL SOLUTIONS FROM TSI

Determination of the filtration performance of air filter MV-F8-03 according to EN 779:2002

Particle Loading Time and Humidity Effects on the Efficiency of an N95 Filtering Facepiece Respirator Model under Constant and Inhalation Cyclic Flows

A Sample Extraction Diluter for Ultrafine Aerosol Sampling

ASI MODEL DM 100 SERIES DILUTING MODULES

289 Series Spring-Loaded Relief Valves

Evaluation of Emissions from HEPA Filters as a Function of Challenge Conditions

Line filters LINE FILTERS

VERIFICATION STATEMENT

When Quality Matters. Microbial Air Samplers: Selection Criteria in Biopharma Manufacturing. Celebrating Over 50 Years In Business!

IN-PLACE HEPA FILTER PENETRATION TEST*

2008 Beltwide Cotton Conferences, Nashville, Tennessee, January 8-11, 2008

A. Submit manufacturer s data, qualifying documents, and shop drawings on the proposed water treatment system before starting any work.

The Bernoulli Filter

DOWNSTREAM DEFENDER TESTING USING MAINE DEPARTMENT OF ENVIRONMENTAL PROTECTION PROTOCOL UTILIZING OK-110 FEED SAND

Dr. Peter Gäng. FilT Eq Filtration Testing Equipment & Services GmbH. Amthausstr. 14, D Karlsruhe, Germany

Understanding specifications and performance of indoor air quality test instruments

For Safe and Delicious Water

Clean air solutions CAMSAFE 2

Protected zone ventilation reduces personal exposure to indoor pollution

EU Water Analysis Using the Thermo Scientific icap 7400 ICP-OES Duo

Compact flue gas analyser: testo 300 M-I

TSI Model 3063 Thermal Mass Flowmeter

Biogas analysers. Product overview. MRU over 30 years of innovative gas analysis! Biogas, landfill gas, biomethane, coal mine gas, CHP exhaust gas

URSA AIR ALU ALU P5858

Parker Balston Validation Studies

PERFORMANCE QUALIFICATION PROTOCOL HVAC SYSTEM

Sampling method for particle measurements of vehicle exhaust

Fpollutant-free zero air for analytical

Effect of the Relative Humidity on Formaldehyde Emission from Tackifier- Coated HVAC Filters

- 1 - Agenda Items. atmospheric dust the invisible cesspool. 1 micrometer = 1/25,400 of an inch. Fiber Size. Lab v Real life conditions

Fisher Cavitrol III One, Two, and Three Stage Trims

A pin-on-disc study of the rate of airborne wear particle emissions from railway braking materials

TECHNICAL DATA SHEET

PURELAB CHORUS SYSTEMS

CyFlow Cube series Appealing from every angle

Particle Measurement Programme

Sample Conditioning Instruments. Jonna Kannosto Technical Marketing Specialist, PhD

Technical Overview. Author. Abstract. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Sprint analysis of lubricating oils using the Thermo Scientific icap 7600 ICP-OES

Preliminary Field Evaluation of EPA Method CTM-039 (PM2.5 Stack Sampling Method) Michael D. Buser and Derek P. Whitelock

Fully Extractive Vs. Dilution Extractive CEMS

On-Line. User s Guide SPE CARTRIDGES. for Rapid Cleanup and Extraction of Analytes

A class of its own JUMAG STEAM BOILER COMPACT INFORMATION STEAM BOILER WATER SUPPLY MODULES BLOW DOWN VESSEL COMPACT STEAM SYSTEM.

NetfilTM. World-class Solutuions in. Proven Technology of High Temp HEPA Filters for Sterilizing Zone. Clean Room Equipment & Filters

Five compelling reasons to replace sieve analysis with laser diffraction

Testing of Carbon Filters in Australian

AIR FILTERS FOR GAS TURBINES. Clean air solutions for turbo machinery. Camfil Clean Air Solutions 13

Filtration in Preparation of Cell Culture Media and Buffers

A Hands-On Guide to Ultrafiltration/ Diafiltration Optimization using Pellicon Cassettes

Proven Flow Measurement Solutions for the Oil and Gas Industry

TWIN LASER PARTICLE COUNTER

This document is a preview generated by EVS

Leadership in Filtration. ScandMist Industrial air cleaners

Aerosol Management in Amine Scrubbing Steven Fulk & Dr. Gary Rochelle

Study on magnetic separation device for scale removal from feed-water in thermal power plant

SAP Solution Manager Focused Insights Setup for ST-OST SP4. AGS Solution Manager SAP Labs France

Skoog, Holler and Nieman, Principles of Instrumental Analysis, 5th edition, Saunders College Publishing, Fort Worth, TX 1998, Ch 33.

ON-LINE OIL TRACE MONITOR

RiOs Essential 5, 8, 16, 24 Water Purification Systems

Application Note. Yeast Analysis with a life - dead staining kit (Yeast control - viability)

BX-895 REAL TIME MODULE OPERATION MANUAL

UniGas Hand-held combustion gas analyzer with 4 sensors

Filter Clogging by Bimodal Aerosol

Transcription:

MODEL 3340 LASER AEROSOL SPECTROMETER Application ote 3340-001 Use of the TSI LAS Model 3340 for HEPA Filter Testing Hans-Georg Horn TSI Incorporated Basic Setup for Filter Testing Measuring filter efficiency always follows the same scheme: A test aerosol is fed into a test duct upstream of the filter under test. Particle concentration measurements are then taken upstream and downstream of this filter. Either one particle measuring system [e.g., Scanning Mobility Particle Sizer (SMPS ) spectrometer, Laser Aerosol Spectrometer, Aerodynamic Particle Sizer (APS ) spectrometer] or two parallel instruments of the same kind may be used for the measurement. In case only one instrument is used, up- and downstream measurements can be made either without or with the filter in place, respectively, or a sample flow switching valve can be used to connect the single instrument to an upstream and downstream sampling port. For most filters, the particle concentration on the upstream side must be quite high to allow reasonably short measurement time on the downstream (clean) side. Therefore, the upstream sample must be diluted before it enters the particle instrument for measurement. In addition to the particle measurement, the measurement of pressure, temperature, relative humidity, filter flow rate and filter resistance (pressure drop) are common for a complete filter test. For most filter tests, industry standards or regulations apply which define the test aerosol (size range and particle material), the type of particle instrument to be used [Airborne Particle Counter (APC), Aerodynamic Particle Sizer (APS ) spectrometer, Condensation Particle Counter, Condensation Particle Counter plus Differential Mobility Analyzer, Scanning Mobility Particle Sizer (SMPS ) spectrometer, photometer, etc.), the size classes in which the filter penetration or efficiency shall be reported, and the test procedure itself. Figure 1 shows the schematic setup of a filter test. TSI, TSI logo, Scanning Mobility Particle Sizer (SMPS ), and Aerodynamic Particle Sizer (APS ) are trademarks of TSI Incorporated.

P, T, R.H., Flow Rate Delta-P Inlet air filter Filter under test Aerosol Generator Dilution Up/Downstream- Switching if one counter/ sizer Is used Particle Counter or Sizer Optional 2nd Counter or Sizer Figure 1: Schematic Filter Test Setup Methods for HEPA & ULPA Filter Testing Available from TSI Table 1: Overview Over Test Methods Applicable for HEPA and ULPA Filter Media And Filters Method Description Advantage Disadvantage Monodisperse test aerosol (from DMA), measurement with one or two CPC(s) e.g., AFT 3160 Monodisperse aerosol from DMA counted by CPC upstream and downstream Polydisperse test aerosol and sampling through DMA to CPC Polydisperse test aerosol and LAS 3340 Aerosol generator (8130 oil generator or 9306A) feeds into test duct, CPC samples known size through DMA Aerosol generator (8130 oil generator or 9306A) feeds into test duct, LAS measures upstream and downstream size distribution Size range 20 nm to 800 nm applicable for electret filters and glass fiber filters. Low upstream dilution 100:1 can be verified/calibrated in situ. Software available (CERTITEST Model 3160) Size range 20 nm to 800 nm applicable for electret filters and glass fiber filters. Low upstream dilution 100:1 can be verified/calibrated in situ. Software available (CERTITEST 3160 special) Works with high flow rates up to 300 m³/h. Above this, more than one generator (parallel) necessary. Simultaneous measurement of all sizes reduces necessary measurement time. Method works for low flow (filter media) and high flow applications. One measurement per size (time consuming) Only applicable for low flow rates <100 L/min (e.g., filter media and cartridges) Low downstream concentration after DMA requires long sampling time. Filter loading effects can bias measurement. Separate upstream and downstream measurement (also time consuming) or two DMAs and CPCs (expensive) necessary. High upstream dilution 1000:1 to 10000:1 cannot be verified/calibrated in situ. Size range (>90 nm) limits this method to glass fiber HEPA/ULPA filters All methods listed in Table 1 can be realized using TSI equipment. The size range from 0.09 µm to 7.5 µm makes the Laser Aerosol Spectrometer LAS 3340 a very interesting instrument for air filter testing. Especially for glass fiber HEPA and ULPA filters (most penetrating particle size MPPS between 100 nm and 200 nm), but also for HVAC filters, the instrument offers advantages. Simultaneous measurement of all particle sizes can reduce the necessary measurement time. -2-

With a well adapted aerosol generator, measurements can be applied for filter media (low flow rate), filter cartridges (medium flow rate) and filter panels (high flow rate). However, the requirements in filter testing may be challenging, which often makes optimization of test aerosol concentration, upstream dilution, used size resolution and downstream sampling time necessary. Figure 2 demonstrates how the parameters characterizing a filter test are linked to each other. Filter Test Dependencies Particle Measurement Aerosol Filter Instrument Filter Flow Rate umber Concentration Generator Flow Rate Maximum umber Concentratiton ecessary Upstream Dilution Upstream umber Concentration Efficieny or Penetration Particle Instrument Detector Flow Rate Downstream umber Concentration Size Resolution (# of size bins) Count rate (#/s) per Size Bin Allowed Measurement Uncertainty ecessary Downstream Sampling Time Figure 2 The most critical parameter in filter testing is the number of particles counted in each size bin on the downstream side of the filter. The more particles counted, the lower the measurement uncertainty, see section on counting statistics. For a given uncertainty requirement (equals total number of particles counted in a size bin), the downstream count rate for each size bin determines the necessary sampling time. If, for example, the downstream sampling time shall be shortened, this can be achieved in several ways: The aerosol generator output (#/s) can be increased, which then increases both the upstream and downstream number concentration and therefore the downstream count rate. It is important to check if the upstream dilution must be increased as well in order to keep the upstream particle concentration within the specification of the LAS. The detector flow rate of the LAS can be increased (if possible). The number of size bins can be reduced, thus making the size bins wider (result: more particles counted in each bin). This can either be done in the instrument configuration or by post processing (see measurement example). Last but not least, a higher measurement uncertainty could be accepted if this is allowed (reduce the necessary number of particles counted in the size bins). As already explained above, it is also very important to dilute the upstream particle concentration for the measurement: The LAS 3340 allows a maximum particle count rate of 3000 particles per second (with -3-

5% coincidence error). This corresponds to 1800 #/cm³ at maximum flow rate, 3600 #/cm³ at 50 cm³/min and 18000 #/cm³ at 10 cm³/min, respectively. Measurement Uncertainty and Counting Statistics Detecting particles in a single particle measuring instrument like the LAS 3340 is a statistical process, simply because the particles do not appear in the detector evenly distributed over time. If, over the measurement time period, a certain number of particles have been counted, this number is directly linked to uncertainty via Poisson statistics. The uncertainty is usually described by means of the 95% confidence interval. If, for example, 1000 particles have been accumulated in a size bin during the measurement, there is a 5 % uncertainty (or: there is 95 % confidence) that the true number concentration, which would be found by infinite counting, is between 938 and 1062. This range is called the 95% confidence interval. If the number of counted particles is larger than 100, the lower and upper boundary of the 95% confidence interval can be calculated using 95 % 1. 962 (1) In this case, the relative uncertainty U in the measurement is U 95% [%] 100% 95% 1.962 100% (2) This means that the statistical uncertainty for 1000 counted particles is ±6.2 %. Inverting equation 2 leads to the necessary particle counts i in a size bin for an allowed uncertainty U: 2 100% i ( U) 1.962 (3) U For example, 1537 particles must be counted to reach ±5% uncertainty or 384 particles must be counted to reach ±10% uncertainty, respectively. Filter Measurement Example This example shows results from the measurement of a HEPA filter; the following parameters characterize the measurement: Parameter Value Filter type HEPA, H14 according to E1822 Filter flow rate 300 m³/h (5000 l/min) Test aerosol Paraffin Oil, D50 200 nm, GSD 1.9 Upstream number concentration 2.4E6 P/cm³ Upstream dilution 1000:1 Particle instrument TSI LAS Model 3340 Flow rate for upstream measurement 39.8 cm³/min Flow rate for downstream measurement 99.2 cm³/min Selected size range 90 nm to 1000 nm Selected number of size bins 30 Selected sampling time 5 x 2 min on each filter side Results: With above parameters, five measurement runs on both the upstream and downstream side were made. To reduce the downstream measurement uncertainty, three adjacent size bins were combined into one new size bin when post-processing the data. The data shown below are based on two minute measurements. -4-

P (%) E (%) The 95% confidence value for the penetration was then calculated by using the lower limit of the 95% confidence interval for the upstream counts and the upper limit of the 95% confidence interval for the downstream counts. UPSTREAM Counts in 2 min 11301 19072 29138 35005 35290 27801 Counts 95% min 11092.6 18801.7 28803.2 34637.9 34921.8 27473.8 Concentration (1/cm³) 1.42E+05 2.40E+05 3.66E+05 4.40E+05 4.43E+05 3.49E+05 Concentration min (1/cm³) 1.39E+05 2.36E+05 3.62E+05 4.35E+05 4.39E+05 3.45E+05 DOWSTREAM Counts in 2 min 339 703 857 473 127 18 Counts 95% max 374.9 755.0 914.8 515.2 149.5 26.8 Concentration (1/cm³) 1.71E+00 3.54E+00 4.32E+00 2.38E+00 6.42E-01 9.27E-02 Concentration max (1/cm³) 1.89E+00 3.80E+00 4.61E+00 2.60E+00 7.54E-01 1.35E-01 PEETRATIO & EFFICIECY 101.5 129.2 164.3 209.1 266.0 338.4 P (%) 1.20E-03 1.48E-03 1.18E-03 5.42E-04 1.45E-04 2.66E-05 P-95 (%) 1.36E-03 1.61E-03 1.27E-03 5.97E-04 1.72E-04 3.92E-05 E (%) 99.99880 99.99852 99.99882 99.99946 99.99986 99.99997 E-95 (%) 99.99864 99.99839 99.99873 99.99940 99.99983 99.99996 P (%) P-95 (%) E (%) E-95 (%) 1.00E-02 100.00000 99.99950 1.00E-03 99.99900 1.00E-04 99.99850 1.00E-05 99.99800 Alternatively, it would have been possible to add all five measurements, which then corresponds to a 10 minute measurement on each side of the filter. This further reduces the measurement uncertainty, as can be seen in the following data: UPSTREAM Counts in 10 min 56505 95362 145689 175023 176450 139003 Counts 95% min 56039.1 94756.7 144940.9 174203.0 175626.7 138272.3 Concentration (1/cm³) 1.42E+05 2.40E+05 3.66E+05 4.40E+05 4.43E+05 3.49E+05 Concentration min (1/cm³) 1.41E+05 2.38E+05 3.64E+05 4.38E+05 4.41E+05 3.47E+05 DOWSTREAM Counts in 10 min 1694 3515 4287 2363 637 92 Counts 95% max 1774.7 3631.2 4415.3 2458.3 686.5 110.8 Concentration (1/cm³) 1.71E+00 3.54E+00 4.32E+00 2.38E+00 6.42E-01 9.27E-02 Concentration max (1/cm³) 1.79E+00 3.66E+00 4.45E+00 2.48E+00 6.92E-01 1.12E-01 PEETRATIO & EFFICIECY 101.5 129.2 164.3 209.1 266.0 338.4 P (%) 1.20E-03 1.48E-03 1.18E-03 5.42E-04 1.45E-04 2.66E-05 P-95 (%) 1.27E-03 1.54E-03 1.22E-03 5.66E-04 1.57E-04 3.22E-05 E (%) 99.99880 99.99852 99.99882 99.99946 99.99986 99.99997 E-95 (%) 99.99873 99.99846 99.99878 99.99943 99.99984 99.99997-5-

P (%) E (%) P (%) P-95 (%) E (%) E-95 (%) 1.00E-02 100.00000 99.99950 1.00E-03 99.99900 1.00E-04 99.99850 1.00E-05 99.99800 Reducing Sampling Losses Particle losses in the sampling line from the instrument to the sampling ports on the test bench may bias the measurements. To overcome this problem, tubing of equal length (or better: the same tubing) should be used to connect to both the upstream and the downstream sampling port. While this will minimize the differences between upstream and downstream particle loss, the losses themselves might still be significant, which unnecessarily reduces especially the downstream particle counts and results in increased measurement uncertainty. The sampling losses can be reduced by several means: Keep sampling tubing as short as possible. Use either conductive flexible tubing or stainless steel tubing Increase the sampling flow rate, but keep the flow laminar (recommendation: 1000 < Re < 1500) The sampling flow rate of the LAS 3340 is relatively low, which makes it difficult to achieve low sampling losses. Therefore, we recommend the use of a simple adapter, which serves two purposes: It allows upstream sample dilution with a TSI diluter Model 3302A and, at the same time, increases the flow through the sampling tube to 3 L/min driven by an external sampling pump. The following pictures illustrate the sampling adapter and its use for filter testing. -6-

Schematic of the Sampling Adapter S P A C E R 3302A Diluter 3340 LAS Photos of the sampling adapter: Parts (upper and lower left) and installation between the LAS 3340 and a Diluter 3302A (upper right) To pump @ 3 L/min, controlled by critical orifices To LAS 3340: Isokinetic flow @ 0.1 L/min TSI Incorporated Visit our website www.tsi.com for more information. USA Tel: +1 800 874 2811 UK Tel: +44 149 4 459200 France Tel: +33 4 91 11 87 64 Germany Tel: +49 241 523030 India Tel: +91 80 67877200 China Tel: +86 10 8251 6588 Singapore Tel: +65 6595 6388 3340-001 Rev. B 2012 TSI Incorporated Printed in U.S.A.