Product Guide for LudgerSep TM N2 High Resolution Amide HPLC Columns for Glycan Analysis

Similar documents
Product Guide for LudgerClean TM D1. Glycan Cleanup Cartridges

Product Guide for LudgerClean TM S. Glycan Cleanup Cartridges

Product Guide for LudgerClean TM CEX Cartridges

ION EXCHANGE KIT FOR MAB SEPARATIONS

HPLC Columns. HILICpak VN-50 2D, 4D MANUAL. Shodex HPLC Columns Europe, Middle East, Africa, Russia

Agilent MetaCarb 87H Organic Acids Column H + Form User Manual Part No. A5214

for Acclaim Carbamate Column

Protein-Pak Hi Res HIC Column and HIC Protein Standard

Agilent MetaCarb Ca PLUS Carbohydrate Column Ca ++ Form User Manual Part No. A5205

the power of discovery USER GUIDE FOR ORGANIC ACIDS ANALYSIS COLUMNS

Analysis of Atrazine in Drinking Water at the ppb Level Using New Agilent Reversed Phase LC Columns. Application. Author. Abstract.

PuriTox Aflatoxin Product Code: P25

Chem 321 Lecture 23 - Liquid Chromatography 11/19/13

HPLC Columns. Protein LW-803 MANUAL. Shodex HPLC Columns Europe, Middle East, Africa, Russia

Econo-Pac High Capacity Ion Exchange Cartridges, 5 ml Instruction Manual. Catalog Numbers , ,

Application Note. Author. Abstract. Pharmaceuticals. Detlef Wilhelm ANATOX GmbH & Co. KG. Fuerstenwalde, Germany mau

10. Validated Normal Phase HPLC Method for the Determination. Fulvestrant is primarily used in the treatment of hormone receptor

TSK-GEL BioAssist Series Ion Exchange Columns

AFLARHONE WIDE Product Code: P116/25, P116/100

Performance characteristics of the 1260 Infinity Quaternary LC system

R R Innovation Way P/N SECKIT-7830 Newark, DE 19711, USA Tel: Fax: Website: Published in November 2013

Fast Preparative Column Liquid Chromatography (PCLC)

BioTrap 500 C18/C8 INSTRUCTION MANUAL. for direct injection of up to 500 µl plasma in HPLC. Illustration of a BioTrap 500 particle

Scale-up with the Agilent SD-1 Purification System analytical and preparative runs on a single system

A Rapid Method for Trace Analysis of Organophosphorus Pesticides in Drinking Water

DNA Synthesis With IRDye 700 Phosphoramidite. Part #

HPLC Columns HIC PH-814 MANUAL. Shodex HPLC Columns Europe, Middle East, Africa, Russia. For technical support please use contact details shown below:

Determination of Adenovirus pviii (31K) Concentration for Estimation of the Empty Capsid Concentration of the Adenovirus Reference Material (ARM)

ENVIRONMENTAL ANALYSIS. Solutions for Your Analytical Business Markets and Applications Programs. Solution Note

Improving Retention Time Precision and Chromatography of Early Eluting Peptides with Acetonitrile/Water Blends as Solvent B

Thermo Scientific. MAbPac HIC-10. Product Manual. P/N: June Part of Thermo Fisher Scientific

PuriTox AflaZON Product Code: TC-M160

Columns for Biomolecules BioLC Column Lines

The residue then is redissolved in HPLC solvent, e.g. in 400 µ l in example calculation see below, and a aliquot is finally injected into the system.

INTERNATIONAL PHARMACOPOEIA MONOGRAPH ON ARTEMETHER AND LUMEFANTRINE CAPSULES REVISED DRAFT FOR DISCUSSION

Agilent Multiple Affinity Removal Spin Cartridges for the Depletion of High-Abundant Proteins from Human Proteomic Samples.

Instructions for Capillary Electrophoresis Peptide Analysis Kit

Reversed-phase Separation of Intact Monoclonal Antibodies Using Agilent ZORBAX Rapid Resolution High Definition 300SB-C8 1.

Instruction of the Clean-up Process Using B-TeZ IAC DonZon Clean-up of Commodity Extracts of Food and Feed Samples containing DON and ZON by Simultane

THIOPHANATE-METHYL in air 5606

Maximizing Sample Recovery on the ACCQPrep HP125

DIDANOSINE ORAL POWDER Final text for addition to The International Pharmacopoeia

A Simple Rapid and Sensitive Method Development for Quantification of Quetiapine Fumarate in Bulk and Dosage Forms Using RP-HPLC

SAFETY AND QUALITY CONTROL

Ultrahydrogel Columns

BÜCHI Labortechnik AG

Preparative HPLC is still the

Abstract. 1. Introduction

Chromatography column for therapeutic protein analysis

Retrovirus Purification Miniprep Manual

Hamilton PRP Polymeric Columns

Experiment 11. DNA Composition by HPLC 1,2

DACLATASVIR TABLETS (DACLATASVIRI COMPRESSI) Proposal for The International Pharmacopoeia. (May 2018)

Version 2.2 CONTENTS. Introduction 3. Frequently Asked Questions 4. Kit Contents 5. Materials and Equipment Needed 5. Standard Protocol 6

Inertsil ODS-3 2 µm. Base Silica Physical Properties and Chemical Modification. Base Silica: High Purity Silica Gel % Surface Area: 450 m 2 /g

A method to pack Agilent Prep C18 Media on Load & Lock Column

Optimizing Purified Sample Recovery on Reverse-Phase HPLC Systems

Size Exclusion BioHPLC columns Ion Exchange BioHPLC columns

Bio-Scale Mini Affi-Prep Protein A Cartridges, 1 and 5 ml. Instruction Manual. Catalog # ,

Application Note. Abstract. Authors. Introduction. Environmental

Separation of Recombinant Human Erythropoietin (repo) Using Agilent Bio SEC-3

Column for High Performance,High-Binding Capacity Ion Exchange Chromatography:TSKgel SuperQ-5PW and Its Applications

Application Note. Author. Abstract. Biopharmaceuticals. Verified for Agilent 1260 Infinity II LC Bio-inert System. Sonja Schneider

G Sep Size Exclusion Columns

Disulfide Linkage Analysis of IgG1 using an Agilent 1260 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column

Empore. Solid Phase Extraction Cartridges. Technical Information. Product Description. Product Characteristics. Sorbents Available

Performance Characteristics of the Agilent 1220 Infinity Gradient LC system

Simple Techniques for Improving the Isolation of Synthetic Peptides Jo-Ann Jablonski Principal Scientist Waters Corporation

CX-1 ph Gradient Buffer

Sulfhydryl Immobilization Kit for Proteins

Introduction Ron Majors is a Senior Scientist at Agilent. Bill Champion is a chemist in Agilent s HPLC Columns tech support group.

Online LC Sample Preparation with BioTrap 500 C18

Instruction of the Clean-up Process Using B-TeZ IAC Afla-Ochra 3ml Clean-up of Commodity Extracts of Food and Feed Samples containing Aflatoxins and O

ProteoEnrich CAT-X SEC Kit Merck: Lichrospher Composition. Silica Particle size. 25 µm particles with 6 nm pore size Ligand SO 3

Fast and High-Resolution Reversed-Phase Separation of Synthetic Oligonucleotides

mdi Filters for Sample Preparation

Emulation of the Agilent 1100 Series LC Through Waters Empower Software Analysis of an Analgesic Mixture

Developing Quantitative UPLC Assays with UV

ultra why micro-lc? product note introducing the ExpressLC -Ultra system: UHPLC with all the advantages of micro-lc

2 Liquid chromatography of biomolecules

EASI-EXTRACT VITAMIN B12 (LGE) Product Code: P88

Techniques for Reducing the Effects of Sample Solvent on UHPLC Analyses

CLEAN SCREEN FASt FILTER AND SHOOT UCT FORENSICS. CLEAN SCREEN FASt SPE COLUMNS

High Throughput Sub-4 Minute Separation of Antibodies using Size Exclusion Chromatography

EVOLUTE ABN FOR EXTRACTION OF DRUGS FROM BIOLOGICAL FLUIDS

ph gradient analysis of IgG1 therapeutic monoclonal antibodies using a 5 µm WCX column

Erstellt mit einer Testversion von PDF Annotator - FLUIDIC & OPTICAL PRODUCTS AND INFORMATION

Analytical Instrument Qualification According to USP <1058>: Requirements and Examples for the Agilent 1290 Infinity LC System

On-Line. User s Guide SPE CARTRIDGES. for Rapid Cleanup and Extraction of Analytes

Contents. Page 1. M13 Isolation Kit. EZgene TM

Supporting Information. Identification of N-(2-Phenoxyethyl)imidazo[1,2-a]pyridine- 3-carboxamides as New Anti-tuberculosis Agents

Separate and Quantify Rituximab Aggregates and Fragments with High-Resolution SEC

MERmaid Kit preps preps Filters/Catch Tubes Filters/Catch Tubes Filters/Catch Tubes

Developing Robust and Efficient IEX Methods for Charge Variant Analysis of Biotherapeutics Using ACQUITY UPLC H-Class System and Auto Blend Plus

Improving Resolution and Column Loading Systematically in Preparative Liquid Chromatography for Isolating a Minor Component from Peppermint Extract

EasyPrep TM Adeno-Associated. Virus Purification Mini Kit Manual. (All Serotype) Catalog#: AAV11-00, AAV11-01, AAV11-02

Superdex 200 HR 10/30 INSTRUCTIONS

Hichrom Limited 1 The Markham Centre, Station Road, Theale, Reading, Berkshire, RG7 4PE, UK Tel: +44 (0) Fax: +44 (0)

Transcription:

Product Guide for LudgerSep TM N2 High Resolution Amide HPLC Columns for Glycan Analysis (Ludger Product Codes: LS-N2-4.6x150 and LS-N2-2.0x150) Ludger Document # LS-N2-Guide-v4.0 Ludger Ltd Culham Science Centre Oxford OX14 3EB United Kingdom Tel: +44 1865 408 554 Fax: +44 870 163 4620 Email: info@ludger.com www.ludger.com

Contents Page Contents... 2 Specifications for LudgerSep N2 Columns... 3 HPLC System Requirements... 5 Installation of the Column... 5 Preconditioning of the Newly Installed Column...6 Column Cleaning and Storage... 6 Light Contamination...6 Cleaning Medium Contamination...6 Cleaning Heavy Contamination...6 Long Term Storage...7 Sample Preparation... 7 Fluorescent Labeling...7 Filtering Samples...8 Sialylated Glycans...8 Operating Schedule and System Suitability Test for 4.6 x 150 mm Column... 8 Glycan Analysis with the 4.6 x 150 mm LS-N2 Amide HPLC Column... 11 Solvents... 11 Gradient # LSN2-10m-Start-35% : N-Glycan Analysis Startup... 11 Gradient # LSN2-30m-Nlink-35% : N-Glycan Analysis Gradient... 11 Gradient # LSN2-10m-Start-20% : O-Glycan and N-Glycan Analysis Startup... 12 Gradient # LS-N2-60m-ONlink-20% : O-Glycan and N-Glycan Analysis Gradient... 12 Operating Schedule and System Suitability Test for two 2.0 x 150 mm Columns... 13 Glycan Analysis with two 2 x 150 mm LS-N2 Amide HPLC Columns... 15 Solvents... 15 Gradient # LSN2x2-5m-Start-30% : N-Glycan Analysis Startup... 15 Gradient # LSN2x2-30m-Nlink-30% : N-Glycan Analysis Gradient... 15 Warranties and Liabilities... 16 Document Revision Number... 16 Ludger Limited, 2010 Page 2

Specifications for LudgerSep N2 Columns Application Analysis and purification by HPLC of LudgerTag fluorophore and UV-chromophore labeled glycans. The 2mm diameter column is recommended for LC-MS applications. Description LudgerSep N2 HPLC columns contain particles with a polymeric amide coating optimized for high resolution chromatography of complex glycan mixtures. Particles 3 μm particle size with 80 angstrom pores and polymeric amide coating. Column Size Cat # Description Dimensions LS-N2-4.6x150 LudgerSep N2 HPLC Column 4.6 x 150 mm LS-N2-2.0x150 LudgerSep N2 HPLC Column 2.0 x 150 mm Flow Rates Typical flow rates = 0.4-1.0 ml/min for 4.6mm column = 0.15-0.22ml/min for 2.0mm column Column Pressure Maximum pressure = 2250 psi (150 kg/cm 2 ) for 4.6mm column = 2900 psi (195 kg/cm 2 ) for 2.0mm column ph Range 2.0-7.5 Temperature Typical operating temperature = 30 o C. Temperature range = 10-40 o C. Solvents Typical solvent systems for glycan analysis include gradients of acetonitrile and aqueous buffers containing ammonium formate, ph 4.4. The recommended buffer is 50 mm ammonium formate ph 4.4 which is available in 50 ml aliquots for making up 2 liters: (Ludger catalog # : LS-N-BUFFX40) Shipping Solvent 75% acetonitrile - 25% water Storage For long-term storage, the column should be washed with a gradient going from the operational buffer to 100% water then to the storage solvent of >50% acetonitrile in water (v/v). Column Protection Filter all solvents to 0.2 μm and degas using either helium sparging or vacuum degassing. Filter all samples using a 0.2 μm filter membrane before loading onto the column. Install an in-line filter between the sample injector and the column. Ludger Limited, 2010 Page 3

Amount of Sample The maximum amount of glycan sample that can be loaded on the column depends on the number and type of glycan components as well as the nature of any non-glycan material. The typical range for successful analytical runs is 10fmol - 1 nmol per sample peak and up to 200 nmol of total glycans. Suitable Samples Suitable samples include glycans labeled with the following LudgerTag labels : 2-AA (2-aminobenzoic acid), 2-AB (2-aminobenzamide), AA-Ac (3-(acetylamino)-6- aminoacridine Sample Filter samples to 0.2 μm then dry using a centrifugal evaporator. Preparation For the 4.6mm column inject the sample in up to 100 μl of the starting buffer (i.e. the solvent mixture used at the very start of the HPLC gradient). For the 2mm column inject the sample in up to 25 μl of the starting buffer (i.e. the solvent mixture used at the very start of the HPLC gradient) Sample Detection Either fluorescence, mass spectrometry or UV-absorbance depending on the dye used (see the appropriate LudgerTag instruction guide). Handling: Ensure that any glass, plasticware or solvents used are free of glycosidases and environmental carbohydrates. Use powder-free gloves for all sample handling procedures and avoid contamination with environmental carbohydrate. Safety: Please read the Material Safety Data Sheets (MSDS's) for all chemicals used. All processes involving labeling reagents should be performed using appropriate personal safety protection - eyeglasses, chemically resistant gloves (e.g. nitrile), etc. - and where appropriate in a laboratory fume cupboard For research use only. Not for human or drug use Ludger Limited, 2010 Page 4

HPLC System Requirements LudgerSep N2 columns can be used with an HPLC system capable of delivering accurate gradients at a flow rate of 0.3 to 1.0 ml/min for the 4.6 mm diameter column, or 50 to 300 μl/min for the 2mm diameter column. In general, systems which mix eluants at high pressure (after the pump head) have lower dead volumes and supply more accurate gradients that are appropriate at the flow rate needed for LudgerSep columns. For the 4.6mm column inject the sample in up to 100 μl of the starting buffer (i.e. the solvent mixture used at the very start of the HPLC gradient). For the 2mm column inject the sample in up to 25 μl of the starting buffer A fluorescence detector is required with the following detection wavelengths: Fluorescence Label λ ex λ em (nm) (nm) 2-AB [2-aminobenzamide] 330 420 2-AA [2-aminobenzoic acid] 330 420 AA-Ac [3-(acetylamino)-6-aminoacridine] 442 525 For optimal detection, use wide slit widths (e.g. 10 20 nm). Sub-picomole levels of 2-AB or 2-AA labelled glycans can be detected with good signal-to-noise (depending on the sensitivity of the detector used). To improve repeatability and intermediate precisions for glycan analyses use a column temperature controller. Good results can be obtained with a column temperature of 35 o C. Installation of the Column During column installation we recommend that : You should connect the LudgerSep N2 column to your HPLC system using standard 1/16 OD tubing and 10-32 (1/16 ) fittings in either stainless steel or PEEK (polyetheretherketone). Finger-tight PEEK fittings and tubing (0.17 mm / 0.007 ID) are recommended for ease of connection and to minimise damage to the column threads. Flow direction is indicated by an arrow on the column label. Keep the lengths of tubing between the injector to column and column to detector as short as possible to minimise dispersion effects. Install an in-line 0.2 μm filter with minimal dead volume between the injector and the head of the Ludger Limited, 2010 Page 5

LudgerSep N2 column to prevent damage to the column by particles. Where two 150 x 2mm columns are joined together use as short a length of tubing as possible. Before analysing any samples, condition your newly installed column as follows Preconditioning of the Newly Installed Column Prepare your newly installed column for glycan analysis as follows: Precondition the column by running two complete LSN2-30m-Nlink-35% gradient cycles (see below) without any sample injected. Check the column performance by running a fluorescently labeled glucose homopolymer (GHP) as a system suitability standard using a LSN2-30m-Nlink-35% gradient for the 4.6 mm column or a LSN2x2-60m-Nlink-30% gradient for two 2mm columns (see section on suitability suitability below). Column Cleaning and Storage After heavy use, your LudgerSep N2 column may become contaminated with strongly adsorbed sample constituents that will lead to a loss in column performance. Light Contamination 1 Clean the column following the protocol in the Column Preconditioning section above. 2 Equilibrate in the starting buffer for the LSN2-30m-Nlink-35% (4.6 mm column) or LSN2x2-60m-Nlink- 30% (2 mm columns) gradient 3 Do a blank run without injecting a sample to check the baseline. Cleaning Medium Contamination 1 Run through one LSN2-30m-Nlink-35% (4.6 mm column) or LSN2x2-60m-Nlink-30% (2mm columns) gradient. 2 Wash with 50% (v/v) 50 mm ammonium formate ph 4.4 (cat. No. LS-N-BUFFX40) 50% (v/v) acetonitrile at 0.1 ml/min overnight 3 Run through one LSN2-30m-Nlink-35% (4.6 mm column) or LSN2x2-60m-Nlink-30% (2 mm columns) gradient 4 Run a system suitability check Cleaning Heavy Contamination Wash the column with the following: Ludger Limited, 2010 Page 6

1 Water [4 h or more at 0.1 ml/min] - to remove very polar solutes from the bonded phase 2 45% acetonitrile (aq) [4 h or more at 0.1 ml/min] - to desorb hydrophobic compounds 3 0.1% triethylamine in 80% acetonitrile [4 h or more at 0.1 ml/min] - to remove desorbed basic compounds 4 50 mm ammonium formate ph 4.4 / acetonitrile (1:1 v/v) [4 h or more at 0.1 ml/min] - to remove ionic compounds 5 Run through one LSN2-30m-Nlink-35% gradient (4.6 mm column) or LSN2x2-60m-Nlink-30% (2 mm columns) 6 Run a system suitability check Long Term Storage For long-term storage, the column should be washed with a gradient going from the operational buffer to 100% water then to the storage solvent of >50% acetonitrile in water (v/v). Sample Preparation LudgerSep N2 columns should be used for analysis of purified glycans (fluorescently labeled or unlabeled). Samples must be free of particulates. The sample should be dissolved in water before addition of acetonitrile to make up to the same composition as the start of the gradient (e.g. dissolve sample in 35 μl water, then add 65 μl acetonitrile before injection to run with the N-glycan analysis gradient method: LSN2-30m-Nlink-35%). Fluorescent Labeling Fluorescent glycans can be prepared by derivatizing pure glycans using reductive amination with a LudgerTag label and purification on a LudgerClean cartridge. The following are commonly used labeling and purification systems for biopharmaceutical glycosylation analysis: Labeling System 2-AB (LudgerTag kit # LT-KAB-A2) 2-AA (LudgerTag kit # LT-KAA-A2) AA-Ac (LudgerTag kit # LT-KAAAC-A2) Post-Labeling Purification System LudgerClean S-Cartridge (Cat # LC-S-A6) LudgerClean S-Cartridge (Cat # LC-S-A6) LudgerClean D1-Cartridge (Cat # LC-D1-A6) Ludger Limited, 2010 Page 7

Filtering Samples Remove particulates from samples by filtering through a spin filter or syringe filter with 0.2 μm pore size membrane. Sialylated Glycans Sialylated glycans can become desialylated if exposed to acidic conditions and elevated temperatures. Avoid desialylation with such samples by a. minimizing exposure to acid (if possible, keep the ph between 5-8), and b. minimizing exposure to temperatures greater than 25 o C. Operating Schedule and System Suitability Test for 4.6 x 150 mm Column Two running methods are listed below: the 30 min run starting at 35% A is suitable for analysis of N- glycans; the 60 min run starting at 20% A is suitable for analysis of both O-glycans and N-glycans When starting the analysis bring the flow rate up slowly by running the startup method: LSN2-10m- Start-35% for N-glycan analysis or LSN2-10m-Start-20% for O-glycan and N-glycan analysis Follow this by running a complete gradient with no sample to condition the column with the running method: LSN2-30m-Nlink-35% : N-Glycan Analysis Gradient or LS-N2-60m-ONlink-20% (see below for gradient methods) Run one of the following system suitability standards using the LSN2-30m-Nlink-35% or LS-N2-60m- ONlink-20% gradient: Product # CAB-GHP-30 CAA-GHP-30 Description 2-AB GHP (2-AB labelled glucose homopolymer ladder) 2-AA GHP (2-AA labelled glucose homopolymer ladder) Ludger Limited, 2010 Page 8

Figure 1: Typical HPLC chromatogram from 2-AB GHP run with method LSN2-30m-Nlink-35% on a 4.6 x 150 mm LudgerSepN2 column. Sample injected in 35%aqueous, 65% acetonitrile. Figure 2: Typical HPLC chromatogram from 2-AB GHP run with method LSN2-60m-Nlink-20% on a 4.6 x 150 mm LudgerSepN2 column. Sample injected in 20%aqueous, 80% acetonitrile. The GHP system suitability profile should be similar to that in figure 1 using the gradient LSN2-30m- Nlink-35%, or figure 2 if using the gradient LSN2-60m-Nlink-20%,. If all 12 peaks do not elute in this time, then this is likely to be due to a high dead volume in the HPLC system. Increase the run time to take this into account. (e.g. gradient of 35-47% A over 24 min) The peaks should be symmetrical, if there is tailing or the peaks are too broad then try cleaning the column (see Section: Column Cleaning and Storage) or change to a new column. Continue running fluorescent labelled GHP suitability standards approximately every 10 sample runs Ludger Limited, 2010 Page 9

The GHP ladder can also be used to calibrate against day-to-day and system-to-system changes by expressing the elution time of peaks as glucose units (GU). The GU value is calculated by fitting a cubic spline distribution curve to the GHP ladder (GU 1-12 in figure 1), this curve is then used to allocate GU values from retention times. These GU values are very reproducible for neutral glycans (+/- 0.03), with more variation in sialylated glycans (+/- 0.3) and can be compared to database values [http://glycobase.ucd.ie/cgi-bin/public/glycobase.cgi Ludger Limited, 2010 Page 10

Glycan Analysis with the 4.6 x 150 mm LS-N2 Amide HPLC Column Solvents The glycan analysis gradients in this guide are based on the following solvents: Solvent A : 50 mm ammonium formate ph 4.4, (50 ml of Ludger product # LS-N-BUFFX40 diluted to 2 litres with water) Solvent B : Acetonitrile Gradient # LSN2-10m-Start-35% : N-Glycan Analysis Startup Use as a start up method for gradient for analysis of N-glycans which starts at 35% A. Time % A %B 4.6mm column (min) Flow Rate (ml/min) 0 35 65 0.0 2 35 65 1.0 10 35 65 1.0 Gradient # LSN2-30m-Nlink-35% : N-Glycan Analysis Gradient Use as a gradient for analysis of N-glycans with GU between 4 and 12. Time % A %B 4.6mm column (min) Flow Rate (ml/min) 0 35 65 1.0 22 46 54 1.0 22.5 100 0 1.0 24.5 100 0 1.0 26 35 65 1.0 30 35 65 1.0 Ludger Limited, 2010 Page 11

Gradient # LSN2-10m-Start-20% : O-Glycan and N-Glycan Analysis Startup Use as a start up method for gradient for analysis of O-glycans and N-glycans which starts at 20% A. Time % A %B 4.6mm column (min) Flow Rate (ml/min) 0 20 80 0.0 2 20 80 1.0 10 20 80 1.0 Gradient # LS-N2-60m-ONlink-20% : O-Glycan and N-Glycan Analysis Gradient Use as a gradient for analysis of O-glycans and N-glycans with GU between 1 and 12. Time % A %B 4.6mm column (min) Flow Rate (ml/min) 0 20 80 1.0 52 46 54 1.0 52.5 100 0 1.0 54.5 100 0 1.0 54 20 80 1.0 60 20 80 1.0 Ludger Limited, 2010 Page 12

Operating Schedule and System Suitability Test for two 2.0 x 150 mm Columns Two 2.0 x 150 mm columns can be joined together in series to obtain double the column length. This setup produces similar resolution to that obtained using one 4.6 x 150 mm column, with the advantage of cutting solvent use to a third of that required for the larger diameter column. A single column can be used but the resolution is not as high. The method for N-glycan analysis is listed below. When starting the analysis bring the flow rate up slowly by running the startup method: LSN2x2-5m- Start-30% Follow this by running a complete gradient with no sample to condition the column with the running method: LSN2x2-60m-Nlink-30% Run one of the following system suitability standards using the LSN2x2-60m-Nlink-30% Product # CAB-GHP-30 CAA-GHP-30 Description 2-AB GHP (2-AB labelled glucose homopolymer ladder) 2-AA GHP (2-AA labelled glucose homopolymer ladder) GU 1 2 Fluorescence 3 4 5 6 7 8 9 10 15 17 0 10 20 30 40 50 60 Minutes Figure 3: Typical HPLC chromatogram from 2-AB GHP run with method LSN2-30m-Nlink-35% on two 2 x 150 mm LudgerSepN2 columns in series. Sample injected in 30%aqueous, 70% acetonitrile. Ludger Limited, 2010 Page 13

The GHP system suitability profile should be similar to that in figure 3 using the gradient LSN2x2-60m-Nlink-30%. If enough peaks do not elute in this time, then this is likely to be due to a high dead volume in the HPLC system. Increase the run time to take this into account. The peaks should be symmetrical, if there is tailing or the peaks are too broad then try cleaning the column (see Section: Column Cleaning and Storage) or change to a new column. Continue running fluorescent labelled GHP suitability standards approximately every 10 sample runs The GHP ladder can also be used to calibrate against day-to-day and system-to-system changes by expressing the elution time of peaks as glucose units (GU). The GU value is calculated by fitting a cubic spline distribution curve to the GHP ladder (GU 1-12 in figure 1), this curve is then used to allocate GU values from retention times. These GU values are very reproducible for neutral glycans (+/- 0.03), with more variation in sialylated glycans (+/- 0.3) and can be compared to database values [http://glycobase.ucd.ie/cgi-bin/public/glycobase.cgi Ludger Limited, 2010 Page 14

Glycan Analysis with two 2 x 150 mm LS-N2 Amide HPLC Columns Solvents The glycan analysis gradients in this guide are based on the following solvents: Solvent A : 50 mm ammonium formate ph 4.4, (50 ml of Ludger product # LS-N-BUFFX40 diluted to 2 litres with water) Solvent B : Acetonitrile Gradient # LSN2x2-5m-Start-30% : N-Glycan Analysis Startup Use as a start up method for gradient for analysis of N-glycans which starts at 30% A. Time % A %B 2 x 2mm column (min) Flow Rate (ml/min) 0 30 70 0.0 2 30 70 0.18 5 30 70 0.18 Gradient # LSN2x2-30m-Nlink-30% : N-Glycan Analysis Gradient Use as a gradient for analysis of N-glycans with GU between 4 and 12. Time % A %B 2 x 2mm column (min) Flow Rate (ml/min) 0 30 70 0.18 46 53 47 0.18 46.5 100 0 0.18 49.5 100 0 0.18 51 30 70 0.18 60 30 70 0.18 Ludger Limited, 2010 Page 15

Warranties and Liabilities Ludger warrants that the above product conforms to the attached analytical documents. Should the product fail for reasons other than through misuse Ludger will, at its option, replace free of charge or refund the purchase price. This warranty is exclusive and Ludger makes no other warrants, expressed or implied, including any implied conditions or warranties of merchantability or fitness for any particular purpose. Ludger shall not be liable for any incidental, consequential or contingent damages. This product is intended for in vitro research only. Document Revision Number Document # LS-N2-Guide-v4.0 Ludger Limited, 2010 Page 16