Fast, Low Pressure Analysis of Food and Beverage Additives Using a Superficially Porous Agilent Poroshell 120 EC-C18 Column

Similar documents
Fast Analysis of Environmental Phenols with Agilent Poroshell 120 EC-C18 Columns

EPA 8330A Analysis of Explosives Using Agilent Poroshell 120 EC-CN and EC-C18 Columns

Determination of Asarinin in Xixin (Asari Radix Et Rhizoma)

Application Note. Authors. Abstract. Introduction. Pharmaceutical

Agilent 1290 Infinity Binary LC System with ISET Emulation of a Waters Alliance 2695 LC Applying Concave, Convex, and Linear Gradients

Agilent 1290 Infinity Quaternary LC Support of Columns with 2.1 to 4.6 mm ID to 1200 bar

Fast and High-Resolution Reversed-Phase Separation of Synthetic Oligonucleotides

Performance Characteristics of the Agilent 1220 Infinity Gradient LC system

Rapid UHPLC Analysis of Reduced Monoclonal Antibodies using an Agilent ZORBAX Rapid Resolution High Definition (RRHD) 300SB-C8 Column

Reversed-phase Separation of Intact Monoclonal Antibodies Using Agilent ZORBAX Rapid Resolution High Definition 300SB-C8 1.

Automated Scouting of Stationary and Mobile Phases Using the Agilent 1290 Infinity II Method Development Solution

Application Note. Biopharma. Authors. Abstract. James Martosella, Phu Duong Agilent Technologies, Inc Centreville Rd Wilmington, DE 19808

Emulation of the Agilent 1100 Series LC Through Waters Empower Software Analysis of an Analgesic Mixture

Method Translation in Liquid Chromatography

Technical Overview. Author. Abstract. A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany

Performance characteristics of the 1260 Infinity Quaternary LC system

Fast Method Development Using the Agilent 1290 Infinity Quaternary LC System with Column Selection Valve

Charge Heterogeneity Analysis of Rituximab Innovator and Biosimilar mabs

Agilent 1290 Infinity LC with ISET under Waters Empower control Emulation of Agilent 1100 Series LC

Disulfide Linkage Analysis of IgG1 using an Agilent 1260 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column

Application Note. Environmental. Agilent 1290 Infinity Binary LC. with ISET and fine-tuning. Agilent 1290 Infinity Binary LC.

High-throughput and Sensitive Size Exclusion Chromatography (SEC) of Biologics Using Agilent AdvanceBio SEC Columns

Method Transfer from an Agilent 1200 Series LC to an Agilent 1260 Infinity II LC

Comparing gradient transfer of isocratic hold and delay volume addition using the Agilent 1290 Infinity LC with ISET

Application Note. Author. Abstract. Pharmaceuticals. A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany

Quality-by-Design-Based Method Development Using an Agilent 1290 Infinity II LC

Agilent 1290 Infinity Binary LC System with ISET - Emulation of the Waters Alliance 2695 LC system analyzing b-blockers

Transferring a Method from a Conventional 3.5 µm or 5 µm to an Agilent Poroshell 120 Column to Improve Analytical Speed and Resolution

Technical Overview. Author. Abstract. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

EPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube

ISET. Agilent 1290 Infinity Binary LC with ISET, emulation of the Waters Alliance 2695 LC system analyzing aromatic acids. Application Note.

Application Note. Author. Abstract. Small Molecule Pharmaceuticals. Sonja Krieger Agilent Technologies, Inc. Waldbronn, Germany

Method Transfer from an Agilent 1260 Infinity LC to an Agilent 1260 Infinity II LC

Application Note. Author. Abstract. Small Molecule Pharmaceuticals & Generics. A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany

Reducing Cycle Time for Charge Variant Analysis of Monoclonal Antibodies

Application Note. Authors. Abstract. Biopharmaceuticals

Size Exclusion Chromatography of Biosimilar and Innovator Insulin Using the Agilent AdvanceBio SEC column

Optimizing Productivity and Reliability for Monocyclic Aromatic Hydrocarbon Purity Analysis According to ASTM D7504 on the Agilent 8890 GC System

Faster Separations Using Agilent Weak Cation Exchange Columns

Analysis of Intact and C-terminal Digested IgG1 on an Agilent Bio MAb 5 µm Column

Separate and Quantify Rituximab Aggregates and Fragments with High-Resolution SEC

Application Note. Author. Abstract. Pharmaceuticals. Detlef Wilhelm ANATOX GmbH & Co. KG. Fuerstenwalde, Germany mau

High-resolution Analysis of Charge Heterogeneity in Monoclonal Antibodies Using ph-gradient Cation Exchange Chromatography

Peptide Mapping: A Quality by Design (QbD) Approach

Fast and High Resolution Analysis of Intact and Reduced Therapeutic Monoclonal Antibodies (mabs)

Quantification of genotoxic "Impurity D" in Atenolol by LC/ESI/MS/MS with Agilent 1200 Series RRLC and 6410B Triple Quadrupole LC/MS

ADVANCE ACCURACY AND PRODUCTIVITY FOR FASTER ANALYSIS

High-Throughput LC/MS Purification of Pharmaceutical Impurities

Application Note. Author. Abstract. Biopharmaceuticals. Verified for Agilent 1260 Infinity II LC Bio-inert System. Sonja Schneider

Analytical Instrument Qualification According to USP <1058>: Requirements and Examples for the Agilent 1290 Infinity LC System

Characterize mab Charge Variants by Cation-Exchange Chromatography

Clone Selection Using the Agilent 1290 Infinity Online 2D-LC/MS Solution

Preparative Purification of Corticosteroids by HPLC; Scalability and Loadability Using Agilent Prep C18 HPLC Columns Application

Scale-up with the Agilent SD-1 Purification System analytical and preparative runs on a single system

Automated QbD-Based Method Development and Validation of Oxidative Degraded Atorvastatin

ADVANCE ACCURACY AND PRODUCTIVITY FOR FASTER ANALYSIS

Automated alternating column regeneration on the Agilent 1290 Infinity LC

Seamless Method Transfer from an Agilent 1260 Infinity Bio-inert LC to an Agilent 1260 Infinity II Bio-inert LC

Analysis of Nucleosides Using an Agilent Infinity II High Speed UHPLC with the 6130 Single Quadrupole Mass Selective Detector

Developing Purification Strategies for the Agilent 1260 Infinity II Preparative LC/MSD System

Fast and Efficient Peptide Mapping of a Monoclonal Antibody (mab): UHPLC Performance with Superficially Porous Particles

Analysis of Atrazine in Drinking Water at the ppb Level Using New Agilent Reversed Phase LC Columns. Application. Author. Abstract.

Agilent AdvanceBio SEC Columns for Aggregate Analysis: Instrument Compatibility

New Advances in UHPLC -Resolution, Speed & Sensitivity

Analysis of TCM injections using the Agilent 1290 Infinity LC system

Analysis of amoxicillin and five impurities on the Agilent 1220 Infinity LC System

Separation of Recombinant Human Erythropoietin (repo) Using Agilent Bio SEC-3

Protein Separation with ph Gradients Using Composite Buffer Systems Calculated by the Agilent Buffer Advisor Software

The Agilent 1260 Infinity BioInert Quaternary Pump. Scope of a low-pressure mixing UHPLC pump with Bio-Inert Capabilities

Characterize Fab and Fc Fragments by Cation-Exchange Chromatography

Fraction Analysis of Cysteine Linked Antibody-Drug Conjugates Using Hydrophobic Interaction. chromatography. Agilent 1260 Infinity II Bio-Inert System

Application Note. Abstract. Authors. Introduction. Environmental

A Guide To Speeding Up Your Separation

Ensure high performance chromatography by using particle free mobile phases and sample solutions

Mobile Phase Optimization in SEC Method Development

BioHPLC columns. Tim Rice Biocolumn Technical Specialist

METHOD TRANSLATION OF HJ FOR INTUVO

Practical Applications of Method Translation Using the Agilent Method Translation Tool. eseminar and Workshop

A Direct Column-Performance Comparison for Rapid Contract Laboratory Program (CLP) Pesticide Analysis. Application. Authors. Abstract.

mab Titer Analysis with the Agilent Bio-Monolith Protein A Column

deltadot

Controlling Agilent 1200 Series Rapid Resolution LC systems through Waters Empower chromatography data software. Technical Overview.

Performance characteristics of the Agilent 1100 Series capillary LC system using diode-array UV and MS for detection. Technical Note.

Technical Overview. Author. Abstract. A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany

ENVIRONMENTAL ANALYSIS. Solutions for Your Analytical Business Markets and Applications Programs. Solution Note

Agilent OpenLAB CDS MatchCompare for the Comparison of Peptide Mapping Samples

Improved Sensitivity with the Agilent 1290 Infinity Evaporative Light Scattering Detector (ELSD)

Contract Laboratory Program (CLP) Pesticide Analysis with 0.18 mm ID High-Efficiency GC Columns Utilizing Helium Carrier Gas. Application.

Sean M. McCarthy and Martin Gilar Waters Corporation, Milford, MA, U.S. INTRODUCTION EXPERIMENTAL RESULTS AND DISCUSSION

Achieving High Performance With Rapid Resolution HT (1.8um) Columns

Organohalide Pesticides in Water by GC/µECD with Agilent J&W DB-CLP1 and DB-CLP2

Development of Ultra-fast ph-gradient Ion Exchange Chromatography for the Separation of Monoclonal Antibody Charge Variants

Encompassing Method Development, Small Scale Purification, and Post-Purification Analysis with the Investigator SFC System

Agilent 1220 Infinity LC. Infinitely more affordable

Simplifying Methods Transfer: Novel Tools for Replicating Your Established Methods on an ACQUITY Arc System

Optimizing Protein Separations with Agilent Weak Cation-Exchange Columns

Analysis of Suspected Flavor and Fragrance Allergens in Cosmetics Using the 7890A GC and Capillary Column Backflush Application

Peptide Mapping. Hardware and Column Optimization

Analysis of USP <467> Residual Solvents using the Agilent 7697A Headspace Sampler with the Agilent 7890B Gas Chromatograph

Transcription:

Fast, Low Pressure Analysis of Food and Beverage Additives Using a Superficially Porous Agilent Poroshell 12 EC-C18 Column Application ote Food and Beverage Authors Anne E. Mack and William J. Long Agilent Technologies, Inc. 28 Centerville Road Wilgton, DE 1988 USA Abstract Eleven non-nutritive food and beverage additives are analyzed in less than three utes using an Agilent Poroshell 12 EC-C18, 3. mm 1 mm, 2. µm column with an ammonium acetate/acetonitrile gradient. The maximum pressure of the analysis is under 4 bar, making this method suitable for use with standard HPLC instruments. Selectivity and efficiency of this Agilent Poroshell 12 EC-C18 column is similar to that of Agilent ZRBAX Eclipse Plus C18, 3. mm 1 mm, 1.8 µm and 4. mm 2 mm, µm columns. While the 1.8-µm Agilent ZRBAX Eclipse Plus column can achieve the desired separation in the same time as the Agilent Poroshell 12, the maximum pressure exceeds 4 bar making a high pressure HPLC necessary. Conversely, the -µm Agilent ZRBAX Eclipse Plus column can perform this analysis at less than 4 bar, but requires substantially more time to accomplish the separation. This application note demonstrates the transfer of older methods using Agilent ZRBAX Eclipse Plus C18 columns to new Agilent Poroshell 12 EC-C18 columns on an HPLC with similar efficiency and selectivity, while saving substantial time.

Introduction Conventional 4-bar HPLC systems are found in many laboratories. With the introduction of UHPLC, there is a desire to move towards sub-2-µm particles. HPLC columns packed with sub-2-µm particles provide superior resolution and decreased analysis time over traditional -µm particles. However this also increases system back pressure. For analysts that wish to translate their current methods to a sub-2-µm method but cannot afford a high pressure LC, columns packed with superficially porous particles may be the answer. Columns like the Agilent Poroshell 12 column offer resolution and speed similar to columns packed with 1.8-µm particles, without generating high back pressure. The high efficiency of Agilent Poroshell 12 particles is similar to sub-2-µm totally porous particles. This is due to short mass transfer distance and substantially narrower particle size distribution. The larger 2.-µm Agilent Poroshell 12 particles generate very low back pressure, about 4% to % of the back pressure generated by sub-2 µm totally porous particles. This allows the columns to run faster flow rates without exceeding HPLC pressures. Agilent Poroshell 12 columns with 2-µm frits are more forgiving with dirty samples than 1.8-µm columns, providing a more seamless method transfer from traditional -µm columns [1-3]. This application note demonstrates the transfer of established methods using longer columns packed with -µm particles to a UHPLC-like method using superficially porous particles packed in shorter columns, while keeping the system back pressure below 4 bar. The benefits are a decrease in sample and mobile phase consumption, significant time savings, and the selectivity and efficiency of a sub-2-µm analysis on virtually any HPLC system. A group of 11 non-nutritive food and beverage additives is used to demonstrate this method translation. These compounds include preservatives, artificial sweeteners, an energy supplement and a flavoring agent. While these compounds are not harmful in appropriate amounts, they can cause sensitization and allergic reactions after excessive exposure [4]. Therefore, detection and quantification of these additive compounds are important. Experimental An Agilent 12 Rapid Resolution LC (RRLC) system was used for this work: G1312B Binary Pump SL with mobile phase A: 2 mm ammonium acetate (ph 4.8) and. Gradient was 14% B at t o, ramp to 2% B. Gradient times vary depending on column dimensions and flow rate, see Table 1. G13C Automatic Liquid Sampler (ALS) SL. Injection volume was 11.8 and 2. µl for the 4. mm 2 mm and 3. mm 1 mm columns respectively. G131B Thermostatted Column Compartment (TCC) SL with temperature set to 3 C. G131C Diode Array Detector (DAD) SL with the signal set to 23, 4 nm and reference set to 3, 1 nm, using a G131-24 micro flow cell (3-mm path, 2-µL volume). ChemStation version B.4.1 (491) was used to control the HPLC and process the data. Table 1. HPLC Method Parameters for Various Columns Column Flow rate Gradient time Stop time Post run time (ml/) () () () Agilent ZRBAX Eclipse Plus C18 1. 12. 13.1. 4. mm 2 mm,. µm Agilent ZRBAX Eclipse Plus C18.38 3. 3. 2. 3. mm 1 mm, 3. µm Agilent ZRBAX Eclipse Plus C18.81 2.1 2. 1.8 3. mm 1 mm, 1.8 µm Agilent Poroshell 12 EC-C18.81 2.1 2. 1.8 3. mm 1 mm, 2. µm 2

Four Agilent columns were used in this work: Agilent Poroshell 12 EC-C18, 3. mm 1 mm, 2. µm (p/n 99-32) Agilent ZRBAX Eclipse Plus C18, 3. mm 1 mm, 1.8 µm (p/n 9994-32) Agilent ZRBAX Eclipse Plus C18, 3. mm 1 mm, 3. µm (p/n 9991-32) Agilent ZRBAX Eclipse Plus C18, 4. mm 2 mm, µm (p/n 9999-92) The compounds of interest are shown in Figure 1, with their respective structure, pka value and additive function. Compounds were dissolved in water at 1 mg/ml. Equal aliquots were combined to produce a mixed sample. Compounds were purchased from Sigma Aldrich (Bellefonte, PA). Additionally, acetonitrile and ammonium acetate were purchased from Sigma Aldrich. Water used was 18 M-Ω Milli-Q water (Bedford, MA). Results and Discussion Figure 2 shows the separation of 11 food and beverage additives on a traditional Agilent ZRBAX Eclipse Plus C18, 4. mm 2 mm, µm column in just over 13 utes. Figure 3 shows the same separation scaled to an Agilent ZRBAX Eclipse Plus C18, 3. mm 1 mm, 3. µm with a Food and Beverage Additives H H H H H H Ascorbic Acid pka = 4.2 S - K + Acesulfame K pka = 2. Sweetener H S Saccharin pka = 2. Sweetener H p-hydroxybenzoic Acid pka = 4. CH 3 H H 3 C CH 3 H H 2 H CH 3 Caffeine pka =. Stimulant Benzoic Acid pka = 4.2 Aspartame pka =. Sweetener H CH 3 H H H Sorbic Acid pka = 4.8 Quinine pka = 8. Bittering Agent Dehydroacetic Acid pka =.3 Sweetener Methylparaben pka = 8. Figure 1. Compounds of interest. 3

significantly shorter analysis time of 3. utes. The flow rate for this analysis was adjusted to compensate for the smaller internal diameter and for the smaller particle size. In this case, the 3.-µm column does not resolve two peak pairs: benzoic acid/aspartame and dehydroacetic acid/methylparaben. Figure 4 shows the same separation, scaled to an Agilent ZRBAX Eclipse Plus C18, 3. mm 1 mm, 1.8 µm column. Flow rate for the 1.8 µm column was adjusted to optimize the smaller sub-2-µm particles. All compounds were well resolved in just over 2. utes. This faster analysis can enhance laboratory productivity, and lower the mobile phase and solvent consumption. A smaller sample volume is required for the smaller 3. mm 1 mm column, thus yielding sample preservation. The benefits of this sub-2-µm separation, however are at the cost of higher back pressure (483 bar), resulting in the need for a high pressure LC system. Figure shows the same separation on a superficially porous Agilent Poroshell 12 EC-C18, 3. mm 1 mm, 2.-µm with the same shortened analysis time as the 1.8-µm column. This separation has the added benefit of lower back pressure (3 bar). The more than 1-bar difference in pressure is very significant, because it deteres if a high pressure HPLC system is needed or if a traditional system (4 bar maximum) is sufficient for this analysis. 2 2 1 1 P max = 12 bar 1 2 3 riginal 13.1 Minute Separation of 11 Food and Beverage Additive Compounds on a 2 mm, -µm Column 4 8 9 * 1 11 Gradient: 14% B at t o, ramp to 2% B in 12. Flow rate: 1. ml/ 1. Ascorbic Acid 2. Acesulfame K 3. Saccharin 4. p-hydroxybenzoic Acid. Caffeine. Benzoic Acid. Aspartame 8. Sorbic Acid 9. Quinine 1. Dehydroacetic Acid 11. Methylparaben * Quinine Impurity 2 4 8 1 12 Figure 2. Separation of 11 food and beverage additives on a Agilent ZRBAX Eclipse Plus C18, 4. mm 2 mm, -µm column using an ammonium acetate/acetonitrile gradient. 12 1 8 4 2 P max = 123 bar 1 2 3 Shorter 3.-µm Column Does ot Resolve Benzoic Acid/Aspartame and Dehydroacetic Acid/Methylparaben Peaks Poor resolution 8 9 Mobile phase: A: 2 mm Ammonium acetate, ph 4.8 B: Acetonitrile Gradient: 14% B at t o, ramp to 2% B in 3. Flow rate:.38 ml/ 1 4 11 * 1. Ascorbic Acid 2. Acesulfame K 3. Saccharin 4. p-hydroxybenzoic Acid. Caffeine. Benzoic Acid. Aspartame 8. Sorbic Acid 9. Quinine 1. Dehydroacetic Acid 11. Methylparaben * Quinine Impurity. 1 1. 2 2. 3 Figure 3. Separation of 11 food and beverage additives on a Agilent ZRBAX Eclipse Plus C18, 3. mm 1 mm, 3.-µm column using an ammonium acetate/acetonitrile gradient. 4

12 1 8 4 2 _ 2 P max = 483 bar 1 14 2 High pressure 3 4 1.8-µm Column Resolves 11 Food and Beverage Additive Compounds in 2. Minutes, but at the Cost of High Pressure. 1 1. 2 2. 8 9 * 1 11 Gradient: 14% B at t o, ramp to 2% B in 2.1 Flow rate:.81 ml/ 1. Ascorbic Acid 2. Acesulfame K 3. Saccharin 4. p-hydroxybenzoic Acid. Caffeine. Benzoic Acid. Aspartame 8. Sorbic Acid 9. Quinine 1. Dehydroacetic Acid 11. Methylparaben * Quinine Impurity Figure 4. Separation of 11 food and beverage additives on a Agilent ZRBAX Eclipse Plus C18, 3. mm 1 mm, 1.8 µm column using an ammonium acetate/acetonitrile gradient. 1 14 12 1 8 4 2 _ 2 P max = 3 bar 1 2 3 4 Superficially Porous Agilent Poroshell 12 Column Resolves 11 Food and Beverage Additives in 2. Minutes and Under 4 bar 9 8. 1 1. 2 2. * 1 11 Gradient: 14% B at t o, ramp to 2% B in 2.1 Flow rate:.81 ml/ 1. Ascorbic Acid 2. Acesulfame K 3. Saccharin 4. p-hydroxybenzoic Acid. Caffeine. Benzoic Acid. Aspartame 8. Sorbic Acid 9. Quinine 1. Dehydroacetic Acid 11. Methylparaben * Quinine Impurity Figure. Separation of 11 food and beverage additives on a Agilent Poroshell 12 EC-C18, 3. mm 1 mm, 2. µm superficially porous column using an ammonium acetate/acetonitrile gradient.

Figure shows an overlay of the original -µm method compared to the new Agilent Poroshell 12 method. Analysis time is reduced from 13.1 to 2., with the post run time reduced from to 1.8. Solvent and mobile phase consumption are reduced by more than 8%. Resolution of the critical pair (dehydroacetic acid and methylparaben) improved from 1.9 to 3.1 on the Agilent Poroshell 12 method, compared to the longer -µm Agilent ZRBAX Eclipse Plus method. ote in Figure, that the last peak on Agilent Poroshell 12 elutes at approximately the same time as the first peak on the -µm Agilent ZRBAX Eclipse Plus. Four consumer product samples are successfully analyzed in less than 2. utes and under 4 bar (Figure ). The energy drink contains caffeine, benzoic acid and sorbic acid. The diet soda has saccharin, caffeine, benzoic acid and aspartame. Mouthwash includes saccharin and benzoic acid. The sugar-free chewing gum contains acesulfame k and aspartame. When high throughput is important and HPLC system limits allow, the flow rate can be increased with little loss in chromatographic quality, as shown in Figure 8. The flow rate on the Agilent Poroshell 12, 3. mm 1 mm column can be increased from.81 to 1.489 ml/ to further reduce run time by 4% in under bar. This achieves little loss in resolution of the critical pair and has imal effects on conditional peak capacity (n c ) []. The result is a baseline separation of 11 compounds in under bar back pressure with an analysis time of 1. utes. Fast, Low Pressure Analysis; All 11 Peaks on Agilent Poroshell 12 are Resolved by the Time the First Peak Elutes on the riginal -µm Agilent ZRBAX Eclipse Plus Method 2 1 1 -µm Agilent ZRBAX Eclipse Plus, P max = 12 bar Critical pair Gradient: 14% B at t o, ramp to 2% B in 12. Flow rate: 1. ml/ 2 4 8 1 12 14 12 1 8 4 2 Agilent Poroshell 12, P max = 3 bar Analysis reduced from 13.1 to 2. Post run reduced from to 1.8 Solvent and mobile phase consumption reduced by more than 8% Resolution of critical pair improved from 1.9 to 3.1 on Poroshell 12 Gradient: 14% B at t o, ramp to 2% B in 2.1 Flow rate:.81 ml/ 2 4 8 1 12 Figure. An overlay of the original Agilent Eclipse Plus -µm method and new Agilent Poroshell 12 method.

Four Consumer Product Samples Successfully Analyzed in 2. Minutes each in under 4 bar Standard Ascorbic Acid Acesulfame K Saccharin p-hydroxybenzoic Acid Caffeine Benzoic Acid Aspartame Sorbic Acid Quinine Dehydroacetic Acid Methylparaben Energy drink Diet soda Gradient: 14% B at t o, ramp to 2% B in 2.1 Flow rate:.81 ml/ Mouthwash Sugar-free gum. 1 1. 2 2. Figure. Four consumer product samples analyzed on a Agilent Poroshell 12 EC-C18, 3. mm 1 mm, 2. µm superficially porous column. Flow Rate on Agilent Poroshell 12 Can Be Increased to Further Enhance Productivity with Minimal Impact on Resolution and Conditional Peak Capacity (n c ).81 ml/ t g = 2.1 3 bar n c =4.9 ml/ t g = 1.9 39 bar n c = 3 Rs = 3.1.3 ml/ t g = 1. 43 bar n c = 2 1.2 ml/ t g = 1.3 1.489 ml/ t g = 1.1 8 bar n c = 8 13 bar n c = Rs = 2.8 where t R,n and t R,1 are the retention times of the last and the first eluting peaks and W is the average 4σ peak width. 1 1. 2 2. Figure 8. verlay showing effects on chromatographic quality with increased flow rate on Agilent Poroshell 12.

Conclusion HPLC columns packed with superficially porous particles offer many advantages over columns packed with conventional, fully porous particles. The superficially porous 2.-µm Agilent Poroshell 12 EC-C18 column offers similar efficiency and selectivity to the 1.8-µm Agilent ZRBAX Eclipse Plus C18 column, without the high back pressure. While larger -µm particles packed in longer columns can yield similar efficiency without high back pressure, they result in a significantly longer analysis time. Due to the similar selectivity between Poroshell 12 EC-C18 and Eclipse Plus C18, methods can easily be transferred from older Eclipse Plus C18 columns to new Poroshell 12 EC-C18 columns. This achieves lower back pressure for older 4 bar HPLC systems, shorter run times, and solvent savings. References 1. W. Long, and A. Mack, Fast Analysis of Sulfa Drugs using the Agilent 11 Series LC with Agilent Poroshell 12 EC-C18 columns, Agilent Technologies publication 99-2E, 21. 2. A. Gratzfeld-Hüsgen, E. aegele, Maximizing efficiency using Agilent Poroshell 12 columns, Agilent Technologies publication 99-2E, 21. 3. V. R. Meyer, Practical High Performance Liquid Chromatography, Fourth Ed., p 34, Wiley, 24. 4. A. Brooks, and W. Long, Selective Analysis of on- utritive Food Additives Using Agilent ZRBAX Eclipse Plus 18, Eclipse Plus Phenyl-Hexyl, Eclipse XDB-Phenyl, and StableBond SB-Phenyl Columns. Agilent Technologies publication 989-991E, 28.. J. Henderson, and J. Berry, UHPLC of Polyphenols in Red Wine, Agilent Technologies publication 99-9E, 21. For More Information For more information on our products and services, visit our Web site at www.agilent.com/chem. www.agilent.com/chem Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc., 21 Printed in the USA July 8, 21 99-82E