The Determination of Alloying Elements in High-Carbon Steel using Radially Viewed Inductively Coupled Plasma Optical Emission Spectrometry (ICP-OES)

Similar documents
Analysis of Stainless Steel by Dual View Inductively Coupled Plasma Spectrometry

Analysis of Stainless Steel by Dual View Inductively Coupled Plasma Spectrometry

Analysis of Mineral and Heavy Metal Content in Beverages Using the Teledyne Leeman Labs Prodigy Plus ICP-OES

Table I: MCL and MRL Concentrations for Contaminants Monitored Under the Safe Drinking Water Act National Primary Drinking Water Regulations

Analysis of steel and its alloys using the GB/T standard and an Agilent 5100 ICP-OES in dual view mode

Ultra-fast ICP-OES determinations of base metals in geochemical samples using next generation sample introduction technology

Analysis of 14 Trace Elements in Stainless Steel

Sprint analysis of lubricating oils using the Thermo Scientific icap 7600 ICP-OES

Determination of 22 Elements Following US EPA Guidelines with a New Megapixel CCD ICP-OES

DETERMINATION OF ELEMENTS IN DRINKING WATER AS PER BUREAU OF INDIAN STANDARDS 10500, & USING THE AGILENT 5100 ICP-OES

Composition Analysis of Animal Feed by HR ICP-OES

Tender Document. National Centre of Excellence in Analytical. Chemistry, University of Sindh, Jamshoro. Inductive Coupled Plasma (ICP -OES) ISSUED BY:

High matrix tolerance of the Thermo Scientific icap 7000 Plus Series ICP-OES with the radial Ceramic D-Torch and the sheath gas adaptor

US EPA Method using the Thermo Scientific icap 7600 ICP-OES Duo

Benefits of running organic matrices using the Agilent 5100 ICP-OES fast, robust, high performance analysis

Analysis of Heavy Metals in Iron-based Fertilizers by HR ICP-OES

Increased Sample Throughput for ICP-OES Applied to U.S. EPA Method 200.7

Analysis of wear metals and contaminants in engine oils using the 4100 MP-AES

ABSTRACT. This study investigates the use of inductively coupled plasma-optical emission spectrometry

EXPAND YOUR RANGE EXTEND YOUR RESOURCES. Avio 500 ICP Optical Emission Spectrometer

Analysis of Toxic Trace Elements in Coffee Bean Products by HR ICP-OES

Determination of Elemental Impurities in Graphite-based Anodes using the Agilent 5110 ICP-OES

Accurate. Robust. Reliable. AGILENT 710 SERIES ICP-OES

Accurate. Robust. Reliable. AGILENT 710 SERIES ICP-OES

Functionality. Normal Operation

Analysis of Samples Related to Nuclear Submarine Operation

Radial, Axial or Dual View ICP: Which Do You Choose? Manny Almeida Teledyne Leeman Labs, Inc. Hudson, NH

Multi-Element Analysis of Trace Metals in Animal Feed using ICP-OES

EU water analysis using the Thermo Scientific icap 7400 ICP-OES Duo

Agilent 5100 ICP-OES DUAL VIEW ICP-OES MINUS THE WAIT

Analysis of Trace Metals in Surface and Bottled Water with the Optima 7300 DV ICP-OES

PROCUREMENT PROCEDURE OF CPRI (NON WORKS) Dt of Revision : Issue Dt. :

Analysis of inorganic impurities in lubricating oils by ICP-MS

Atomic Spectroscopy Solutions for Environmental Applications

Increase productivity for environmental sample analysis using the SVS 2+ Switching Valve System for Agilent 5100 SVDV ICP-OES

ICP-OES. Varian Vista-MPX CCD Simultaneous ICP-OES

Trace Elements in Petrochemical Samples: Achieving Excellent Analytical Results

Direct Analysis of Photoresist by ICP-MS. Featuring the Agilent Technologies 7500s ICP-MS

DETERMINATION OF SULPHUR CONTENT OF UN-ALLOYED AND LOW-ALLOYED STEEL BY ICP-AES SPECTROMETRY USING WET CHEMICAL SAMPLE PREPARATION

Analysis of Environmental Samples with the Agilent 730-ES Following US EPA Guidelines

Analysis of Trace Elements in Seawater Using the Thermo Scientific icap 7000 Series ICP-OES Duo

US EPA Method using the Thermo Scientific icap 7600 ICP-OES Duo

Eco Plasma Innovative Inductively Coupled Plasma Technology

Analytically Monitoring the Effect of Fracturing Activity

Application Note PlasmaQuant PQ 9000 Elite. Analysis of Rare Earth Elements in Granite and Sandstone by HR ICP-OES. Challenge.

Optima 8x00 Series. ICP Optical Emission Spectrometers

Characterization of Surface Metals on Silicon Wafers by SME-ICP-MS. Featuring the Agilent Technologies 7500s ICP-MS

OPTIMA 4000 DV Series of ICP-OES Systems

EU Water Analysis Using the Thermo Scientific icap 7400 ICP-OES Duo

Water Analysis Using ICP-OES with an Ultrasonic Nebulizer

icap 7000 Plus Series ICP-OES Streamlined performance and ultra-low ICP-OES detection limits for routine laboratories

AMAZINGLY CAPABLE REMARKABLY AFFORDABLE. Avio 200 ICP Optical Emission Spectrometer. For research use only. Not for use in diagnostic procedures.

High performance radio frequency generator technology for the Thermo Scientific icap 7000 Plus Series ICP-OES

Wavelength Dispersive XRF Spectrometer

Automating EPA 6020 Compliant Analysis with the Agilent 7900 ICP-MS and ESI prepfast Autodilution System

The Determination of Trace Elements in Waters using the US EPA Method and the Thermo Scientific icap 6500 ICP-OES (Duo)

Avio 200 ICP-OES Flat Plate

Analysis of elemental impurities in drug products using the Thermo Scientific icap 7400 ICP-OES Duo

ElvaX ProSpector in Exploration & Mining

VHG Labs. Standards and Supplies for Spectrochemical Analysis QUALITY. TRACEABILITY. DEPENDABILITY. WORLDWIDE

ICP-AES Determination of 15 Kind of impurity elements in the Vanadium-aluminum alloy

The image part with relationship ID rid4 was not found in the file. Welcome

ESI SC-FAST Sample Introduction Solutions for PerkinElmer ICP-OES and ICP-MS Systems

ANALYTICAL PERFORMANCE OF A HANDHELD EDXRF SPECTROMETER WITH MINIATURE X-RAY TUBE EXCITATION

Thermo Scientific icap 6000 Series ICP Emission Spectrometer

Analysis of Cast Iron Using Shimadzu PDA-7000

Kunikazu Ide and Yoshisuke Nakamura. 1. Introduction

The Determination of Trace Elements in Stainless Steel by Forked Platform GFAAS

Optima 8x00 Series. ICP Optical Emission Spectrometers

VHG Labs/LGC Standards Wear Metals Performance Testing Program (PTP)

The Agilent Atomic Spectroscopy portfolio for Environmental applications AA, MP-AES, ICP-OES, ICP-MS & ICP-QQQ

Multi-element determination in pharmaceutical preparations using the Thermo Scientific icap Q ICP-MS

ICP Expert software. Technical Overview. Introduction

Inductively coupled plasma (ICP) spectroscopy for online measurements of trace metal emissions at the 250 kw PACT furnace

Agilent Atomic Spectroscopy Solutions for Energy & Chemicals AA, MP-AES, ICP-OES, ICP-MS.

Common Challenges facing Laboratories doing Elemental Analysis Today

Agilent 5100 ICP-OES DUAL VIEW ICP-OES MINUS THE WAIT

Amplifying ICPMS Productivity Using Discrete Sampling Technology

FOUNDRY-MASTER Pro. First class laboratory optical emission spectrometer for complete professional metal analysisoes

Thermo Scientific icap 7000 Series ICP-OES Accessories Guide

US EPA Method using the XSERIES 2 ICP-MS

Certificate of Analysis BS 110B

Ultima Expert Ultimate ICP-OES Spectrometer Best Performance & Ease-of-Use

Ultima Expert Ultimate ICP-OES Spectrometer Best Performance & Ease-of-Use

Thermo Scientific icap 7000 Plus Series ICP-OES. Gain more power. experience more performance

EDXRF APPLICATION NOTE IRON IN ALUMINUM # 1237

Assaying and Analysis Choices in Practice

METHOD 3040A DISSOLUTION PROCEDURE FOR OILS, GREASES, OR WAXES

USABILITY BY DESIGN Every component of the Agilent 7900 ICP-MS is engineered for performance and reliability

Rubber and plastics gloves for food services Limits for extractable substances

Fields of Application / Industry:

Analysis of Copper and Copper Base Alloys, Using Shimadzu PDA-7000

TECHNIQUES OF INDUCTIVELY COUPLED PLASMA OPTICAL EMISSION SPECTROMETER S. KARTHIKEYAN SSA CENTRAL POLLUTION CONTROL BOARD SOUTH ZONE OFFICE BANGALORE

Collaborative Efforts for Creating a Robust Trace Metals Analytical Procedure for Flue Gas Desulfurization Wastewaters by ICP-MS

Skoog, Holler and Nieman, Principles of Instrumental Analysis, 5th edition, Saunders College Publishing, Fort Worth, TX 1998, Ch 33.

SUB-Programs - Calibration range Fe Base for "PMI-MASTER Pro" Spark - mode Fe 000

Ultra-Trace Analysis of Beryllium in Water and Industrial Hygiene Samples by ICP-MS Application

SECTIOn 4. Claritas PPT Multi-Element Standards for ICP-MS

Standard Solutions (Traceable to NIST)

Transcription:

Application Note - AN1504 The Determination of Alloying Elements in High-Carbon Steel using Radially Viewed Inductively Coupled Plasma Optical Emission Spectrometry (ICP-OES) Introduction Page 1 Carbon steels comprise more than 85% of the steel produced in the United States. Classified by the amount of carbon they contain, variations in carbon content (shown in Table I) significantly affect the mechanical properties of the final steel product. Carbon is the most cost-effective alloying material for iron. Higher carbon content leads to increasing hardness and strength, while lower carbon content produces steels that are easier to bend and form. Other alloying elements can also be used to alter the steel s properties including Manganese and Vanadium. Consequently, accurate analysis of the steel for alloying elements is a requisite quality control measure. This application note will demonstrate the Teledyne Leeman Lab s Prodigy7 High-Dispersion ICP s ability to determine alloying elements in high-carbon steel. Table I Carbon Steel Classification Type Carbon Content Typical Uses Low < 0.30% Auto Body Panels, Wire Products Medium 0.30-0.60% Axles, Gears, Crankshafts High 0.60-1.00% Springs, Dies, High-Strength Wires Ultra-High 1.25-2.0% Metal Cutting Tools, Truck Springs Instrument The Prodigy7 is a compact bench-top simultaneous ICP-OES system featuring a 500 mm focal length Echelle optical system coupled with a mega-pixel large format CMOS detector. At 28 x 28 mm, the active area of the L-CMOS is significantly larger than any other solid-state detector currently used for ICP-OES. This combination allows the Prodigy7 to achieve higher optical resolution than other solid-state detector-based ICP systems. The detector also provides continuous wavelength coverage from 165 to 1100 nm, permitting measurement over the entire ICP spectrum in a single reading without sacrificing wavelength range or resolution. This detector design is inherently anti-blooming and is capable of random access, non-destructive readout that results in a dynamic range of more than 6 orders of magnitude. The Prodigy7 also uses a 40.68 MHz rugged, water-cooled, free-running RF Generator, allowing it to handle the most difficult sample matrices, as well as common organic solvents. The sample introduction system consists of a four-channel peristaltic pump, HF resistant Ryton Scott spray chamber, demountable quartz torch with an alumina injector and a Hildebrand Grid nebulizer.

Method Page 2 Sample Preparation A high-carbon steel (0.87% C) reference material, SRM 364 (NIST), was used in this study. Approximately 1 gram of the material was placed in a Teflon beaker, covered with a minimum of deionized water (DIW) and placed on a hot plate. The samples were dissolved using 5 ml of nitric acid (HNO 3 ) and 1 ml hydrofluoric acid (HF) while gently heating. Once the dissolution was complete, the samples were diluted to 100 ml with DIW. Calibration Standards Calibration standards were made from single-element stock solutions (VHG Labs). Standards were matrix-matched to the Fe concentration of the sample by dissolving an electrolytic iron reference material (NIST SRM 365). The final acid concentration in the standards was 5% HNO 3 /1% HF. The concentrations of the calibration standards are listed in Table II. Table II Calibration Standards Concentrations, ppm Element STD1 (ppm) STD2 (ppm) STD3 (ppm) Mo 0 50 100 Mn, Cu, Ti 0 15 30 Ni, V, W, Co, Nb, Ta 0 10 20 Si, Cr, As, Zr 0 5 10 Sn, B, Pb, Sb 0 2.5 5 Instrument Operating Conditions The Prodigy7 operating parameters are listed in Table III. Table III Instrument Operating Parameters Instrument RF Power 1.20 kw Coolant Flow 14 L/min Auxiliary Flow 0.0 L/min Nebulizer Pressure 1.0 L/min Uptake Rate 25 rpm Sample Introduction Nebulizer Type Hildebrand Grid Spray Chamber HF Resistant Ryton Scott Torch Quartz Demountable Injector Alumina, 2.5 mm bore Uptake Tubing Aqueous, Black Tab 030 Drain Tubing Aqueous, Red Tab 045 Sample Internal Standard None Integration Time 30 seconds Contact Teledyne Leeman Labs for the appropriate sample introduction component part number. The analytical viewing zone was set by using a 10 ppm Mn standard. The optimum viewing position is automatically selected by the Prodigy7 s Salsa software.

Page 3 Wavelength Parameters The Prodigy7 typically uses a 27 pixel wide subarray, centered on the wavelength of interest, to collect data for each analyte. However, subarrays can be up to 57 pixels in width if needed. The analytical peaks and background correction points are defined in each subarray with pixel position and width values. In Table IV below, wavelength and background correction position data are listed. Where possible, two wavelengths were used for each element. All data in the subarrays is collected simultaneously. Additionally, all pixel data are saved, permitting recalculation of results at a later time. Table IV Wavelength Parameters Element Wavelength, nm Left Background Correction Right Background Correction Position Width Position Width Mn 257.610 8 2 27 2 259.372 2 2 27 2 Si 251.611 9 2 26 2 288.158 2 0 27 2 Cu 324.754 - - 27 2 327.396 6 2 - - Ni 231.604 2 2 28 2 221.648 2 2 - - Cr 267.716 2 0 28 2 205.552 - - 28 2 V 309.311 2 2 27 2 310.230 2 2 27 2 Mo 202.030 2 2 27 2 W 207.911 2 2 27 2 224.875 2 2 26 2 Co 228.615 2 2 27 2 345.351 2 0 27 2 Ti 337.280 2 2 27 2 334.941 2 2 27 2 As 189.042 - - 27 2 197.262 - - 27 2 Sn 189.991 - - 28 2 Nb 313.079 2 2 27 2 Ta 263.558 - - 21 2 226.230 2 2 27 2 B 182.641 - - 27 2 Pb 220.353 2 2 27 2 Zr 343.823 2 2 22 2 339.198 2 2 24 2 Sb 206.833 7 7 27 2

Page 4 Figure 1 illustrates the element parameters for the W 207.911 nm line as an example. The left and right background regions begin at pixel positions 2 and 27, respectively. The width of both positions is 2 pixels. The analytical region of interest, where the W peak is found, begins at pixel position 13 and has a width of 5 pixels. Figure 1 W 207.911 nm Wavelength Parameter Example Figure 2 illustrates a calibration curve showing typical precision and linearity for the concentration range used. Figure 2 Typical Calibration Curve Results After igniting the plasma and allowing a 15 minute warm-up period, the Prodigy7 was calibrated using the Teledyne CETAC ASX-520 autosampler. Once the calibration was complete, a 1 ppm QC Standard was analyzed with an acceptance criteria of ±10%. Upon successful completion of the QC Standard analysis, the reference sample was analyzed. After the sample analysis, the QC Standard was reanalyzed. The Prodigy7 s Salsa software allows the entire sequence to be run unattended.

Page 5 If a QC Standard result is out of specification, the Prodigy7 will automatically recalibrate and rerun the QC Standard and any samples that were analyzed since the last successful QC Standard was analyzed. Results from the analysis are shown in Table III. All concentrations are shown in percentages. The values measured by the Prodigy7 are contained in the column labeled Found % while the certified values are in the column labeled Certified %. The certified value listed is not expected to deviate from the true value by more than ±1 in the last significant figure reported. For a subscript figure, the deviation is not expected to be more than ±5. The agreement between the measured and certified values is quite good. Table V NIST 364 Analysis Results, % Mn 257.610 r 0.25 Mn 259.372 r 0.25 Si 251.611 r 0.05 Si 288.158 r 0.05 Cu 324.754 r 0.24 Found % Certified % 0.25 0.06 5 0.24 9 Cu 327.396 r 0.24 Ni 231.604 r 0.14 Ni 221.648 r 0.13 0.14 4 Cr 267.716 r 0.06 Cr 205.552 r 0.06 0.06 3 V 309.311 r 0.11 V 310.230 r 0.11 0.10 5 Mo 202.030 r 0.50 0.49 W 207.911 r 0.09 W 224.875 r 0.10 0.1 Co 228.615 r 0.15 Co 345.351 r 0.15 0.15 Ti 337.280 r 0.25 Ti 334.941 r 0.25 0.24 As 189.042 r 0.04 As 197.262 r 0.05 0.05 2 Sn 189.991 r 0.008 0.008 Nb 313.079 r 0.15 0.15 7 Ta 263.558 r 0.10 Ta 226.230 r 0.10 0.11 B 182.641 r 0.011 0.0106 Pb 220.353 r 0.024 0.023 Zr 343.823 r 0.069 0.068 Zr 339.198 r 0.065 Sb 206.833 r 0.033 0.034

Discussion The determination of the alloying elements in the 364 high-carbon steel reference material was straightforward. Of all the elements determined, the most challenging was Lead (Pb). Page 6 During line selection, a number of Pb lines were examined. All were found to have significant interferences. The most promising line was the primary Pb line at 220.353 nm. A scan of a 10 ppm Pb standard (not matrix matched) and the 364 reference sample is displayed in Figure 3, where the interference is clearly seen. Figure 3 Interference on Pb 220.353 nm Line The Prodigy7 software has sufficient flexibility to provide solutions to deal with this type of interference. First, the analytical Region Of Interest (ROI) can be reduced in size thereby avoiding the contribution from the interfering peak. The advantage in this method is that the identity of the interfering peak does not necessary need to be determined. An example of this technique is shown in Figure 4. Without reducing the ROI interval, the Pb concentration was reported as 0.348%. The reduced size ROI reported the accurate value of 0.024%, as shown in Figure 4.

Figure 4 Example of Reduced Size Region Of Interest (ROI) Page 7 The second method is to perform an Interfering Element Correction (IEC). In order to perform this correction, the element responsible for the interfering peak must be determined. The Prodigy7 software also provides tools to easily identify unknown peaks. The first step in identifying an unknown peak is to take a Full Frame Image of the sample. The Full Frame image takes a complete spectrum from 165 to 1100 nm without any gaps in a single exposure. A Full Frame Image of the 364 reference sample, with analytical wavelengths superimposed is shown in Figure 5. Figure 5 NIST SRM 364 High-Carbon Steel Full Frame Image

The next step is to zoom into the wavelength in question and use the Identifier Button (Figure 6). Figure 6 Full Frame Image Zoomed to Pb 220.353 nm Subarray Page 8 By clicking on the Identifier Button and then inside the Pb subarray (as indicated by the red arrow in Figure 6), the wavelength line library will be searched for possible elements. The result of the search is displayed in Figure 7. Figure 7 Search Results The search results suggest the interfering peak is due to Niobium (Nb). A scan of an aqueous Nb standard added to the scans in Figure 4 confirms the identification (see Figure 8). Once the interfering element has been identified, the IEC factor can be easily determined by analyzing the Nb standard to obtain the correction ratio.

Figure 8 Confirmation of Nb as the Interfering Element Page 9 Conclusion The analysis of alloying elements in high-carbon steel has been carried out for 18 elements using a radially viewed Teledyne Leeman Lab s, Prodigy7 High-Dispersion ICP. Accurate results were obtained by carefully matrix matching the base iron concentration of the samples to the calibration standards. The HF sample introduction system performed without any clogging of the torch or nebulizer and did not require the use of an argon humidifier. The Hildebrand Grid nebulizer with its platinum screens is an ideal nebulizer for use with HF containing samples, as well as samples containing high levels of dissolved solids. The image stabilized plasma and the simultaneous data collection of both peak and background data combine to provide exceptionally precise and stable results.