Scope and application: For water and wastewater. Turbid samples are filtered during the analysis.

Similar documents
Color, True and Apparent, LR

Scope and application: For water, wastewater, seawater, brine solutions, produced waters and hydraulic fracturing waters.

USEPA 1 Periodate Oxidation Method 2 Method to 20.0 mg/l Mn (HR) Powder Pillows

USEPA 1,2 Bicinchoninate Method 3 Method 8506 (CuVer 1) and Method 8026 (CuVer 2) 0.04 to 5.00 mg/l Cu Powder Pillows or AccuVac Ampuls

Hardness, Calcium/Magnesium

1-(2-Pyridylazo)-2-Naphthol PAN Method 1 Method to mg/l Mn (LR) Powder Pillows

Calcium and Magnesium; Chlorophosphonazo Rapid Liquid Method Method to 1000 µg/l Ca and Mg as CaCO 3 (ULR) Pour-Thru Cell

Phosphorus, Reactive (Orthophosphate)

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856

Instrument Adapters Light shield DR 6000, DR 5000 DR 3900 LZV849 DR 3800, DR 2800 LZV646 DR or (A)

Sulfate. USEPA 1 SulfaVer 4 Method 2 Method to 70 mg/l SO 4. Powder Pillows. Test preparation. Before starting.

Nitrogen, Simplified TKN (s-tkn )

Phosphorus, Reactive (Orthophosphate)

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

Method 8017 (0.7 to 80.0 µg/l) Powder Pillows. Scope and Application: For water and wastewater; digestion is required to determine total cadmium.

Dimethylphenol Method Method LR (0.23 to mg/l NO 3 N or 1.00 to mg/l NO 3 ) TNTplus 835

Oxygen Demand, Chemical

Direct ISE Method Method to 4.00 mg/l NO 3 N Nitrate ISE

Sulfate DOC SulfaVer 4 Method 1. Method (0 70, and mg/l) Powder Pillows. Test procedure

Table 163 Instrument-specific information. Instrument Pour-thru Kit Cell orientation Adapter. DR 6000 LQV Arrow faces right

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

Oxygen Demand, Chemical

Direct ISE Method Method to 10.0 mg/l F Fluoride ISE

Oxygen Demand, Chemical

DR/4000 PROCEDURE. IRON, Ferrous. Using Powder Pillows

Method to 500 µg/l CH 2 O Powder Pillows

DR/4000 PROCEDURE. OXYGEN DEMAND, Chemical. Colorimetric Measurement

Method (0.1 to 8.0 mg/l Cu) TNTplus 860

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

Phosphorus, Reactive (Orthophosphate)

Nitrogen, Simplified TKN (s-tkn ) DOC

DR/4000 PROCEDURE. OXYGEN DEMAND, Chemical. Colorimetric Measurement, 0 to 40 mg/l COD

PHOSPHORUS, REACTIVE (0.00 to 5.00 mg/l PO 4

Oxygen Demand, Chemical

USEPA Membrane Filtration Method Method m-tec. Scope and Application: For potable water, nonpotable water, recreation water and wastewater.

STP (off-line) Water Technologies & Solutions analytical procedure. colorimetric method DCA reagent 1 (5/5/1) 2 to 15 mg/l as STP

Coliforms, Total and E. coli

3. Fill a 50-mL. 2. Press: 1 ENTER The display will show mg/l, Al and the ZERO icon.

Heterotrophic Bacteria

TPTZ Method * (Powder Pillows or AccuVac Ampuls) Using Powder Pillows

1,5-Diphenylcarbohydrazide Method* (Powder Pillows or AccuVac Ampuls) USEPA accepted for wastewater analyses** Using Powder Pillows

Amperometric Titrator assembly. Digital Titrator. Beaker, low-form, 250-mL. Stir bar, octagonal, Teflon-coated, 50.8 x 7.9 mm

FerroVer Method (Powder Pillows or AccuVac Ampuls) USEPA approved for reporting wastewater analysis (digestion is required; see Section 2 * )

3. Add 0.4 ml of. 7. Use a TenSette

Oxygen Demand, Biochemical

3. Add 0.4 ml of. 7. Use a TenSette

DR/4000 PROCEDURE OXYGEN SCAVENGERS. 4. Fill a second sample cell (the blank) with 25 ml of deionized water.

Approved for NPDES (Editorial Revision 1978) Silica, Dissolved (Colorimetric)

Professional Boiler Water Test Kit

PaSEC Colorimeter Instructions DR850

Engineered Systems & Designs TURBIDITY METER

3. Fill a 50-mL. 2. Press: 1 ENTER The display will show mg/l, Al and the ZERO icon.

BONDERITE NT-1 Phosphate-free, Regulated Heavy Metal-Free Pre-treatment

Free, Total and Double SO2 in Wine

1,10 Phenanthroline Method * (Powder Pillows or AccuVac Ampuls) Using Powder Pillows

Hach Method Determination of Turbidty by Laser Nephelometry*

PavCoTing. Chromium appearance and corrosion resistance Ease of plating bath control Wide current density range

Technical Process Bulletin

NACA Laboratory Equipment and Supplies

Technical Process Bulletin

SPECTROPHOTOMETRIC DETERMINATION OF IRON

STATE OF OHIO DEPARTMENT OF TRANSPORTATION SUPPLEMENT 1122 DETERMINING SULFATE CONTENT IN SOILS. July 17, 2015

Technical Process Bulletin

SECTION D.3 CHLOROPHYLL AND PHEOPHYTIN

Cu (s) Cu 2+ (aq) Cu(OH) 2 (s) CuO (s) Cu 2+ (aq) Cu (s)

Sampling Day Activities

Protocol. Envirochek HV Sampling Capsule. 1.0 Sample Collection Materials

Technical Process Bulletin

Standard Operating Procedure for GLNPO Turbidity: Nephelometric Method

Standard Methods for the Examination of Water and Wastewater

SOP-C-109 Determination of Filterable Residue

SurTec 650 chromital TCP

30th International Chemistry Olympiad

Rev Experiment 10

DAIRY WATERS. (Coliform Group and Escherichia coli) [E. coli verification required only on source water] IMS #24

OXIDATION-REDUCTION TITRATIONS-Permanganometry

Technical Process Bulletin

2. 47 mm grid marked, white sterile 0.45 micron membranes (Millipore or equivalent) 4. Vacuum pump capable of inches of vacuum

Suspended Metals Digestion Procedure

:00 am 6:00 pm Mon-Thu (MT) 6:00 am 5:00 pm Fri (MT)

METHOD DETERMINATION OF TURBIDITY BY NEPHELOMETRY. Edited by James W. O'Dell Inorganic Chemistry Branch Chemistry Research Division

The determination of copper in brass

Fines fraction of paper stock by wet screening

LABORATORY 3 SOIL ANALYSIS

EVENT: Perform the following steps of a BOD analysis using the YSI ProOBOD Optical BOD probe.

Ti-Pure TITANIUM DIOXIDE. CONCENTRATED ACID SOLUBILITY OF TiO 2 METHOD: T WP

An Oxidation-Reduction Titration: The Reaction of Fe 2+ and Ce 4+

Methods and Materials Manual Of Ion Chromatography Using DX-100 IC. Collecting Samples. LDPE bottle with cap (pre-rinsed with DI water)

CHM Gravimetric Chloride Experiment (r7) 1/5

MyBioSource.com. Human Vitamin K2 ELISA USER INSTRUCTION

Human Myostatin, ELISA Kit (MSTN)

scd134/ox40 (Human) ELISA

Turbidity by Nephelometric Determination

MyBioSource.com. Human Osteoprotegerin ELISA USER INSTRUCTION

SUMMARY OF STANDARDS FOR WATER ANALYSIS ON COURSE WAT-E2120 SPRING 2017

Total Dissolved Solids Dried at 180ºC. Total Volatile Dissolved Solids at 550 C

MyBioSource.com. Rat Hydroxyproline ELISA USER INSTRUCTION

Technical Process Bulletin

3.0 DETECTION LIMIT 3.1. See Procedure ALQAP Procedure for the Determination of Uncertainty

VISCOSITY, INHERENT (One Point)

Transcription:

Color, ADMI DOC316.53.01122 ADMI Weighted Ordinate Method 1 Method 10048 5 to 300 Pt-Co units 2 Scope and application: For water and wastewater. Turbid samples are filtered during the analysis. 1 Adapted from Allen, et al., 1972. Determination of color of water and wastewater by means of ADMI Color Values. Proc. 28th Ind. Waste Conf., Purdue Univ., Eng. Ext. Ser. no. 142:661. 2 DR 5000: 5 to 250 Pt-Co units Test preparation Instrument-specific information Table 1 shows all of the instruments that have the program for this test. The table also shows the sample cell options and orientation requirements. To use the table, select an instrument, then read across to find the applicable information for this test. Table 1 Instrument-specific information Instrument Sample cell orientation 1-inch sample cell 10-mm sample cell 50-mm sample cell DR 6000 DR 5000 DR 3900 Before starting The fill line is to the right (1-inch cell) The clear side is to the right (10-mm, 50-mm cells) The fill line is toward the user (1-inch cell) The clear side is toward the user (10-mm, 50- mm cells) 2495402 2095100 2629250 Review the Safety Data Sheets (MSDS/SDS) for the chemicals that are used. Use the recommended personal protective equipment. Dispose of reacted solutions according to local, state and federal regulations. Refer to the Safety Data Sheets for disposal information for unused reagents. Refer to the environmental, health and safety staff for your facility and/or local regulatory agencies for further disposal information. Items to collect Description Quantity Beaker, 250 ml polypropylene 2 Cylinder, graduated, 100 ml polypropylene 1 Filtration apparatus: funnel/filter holder, 0.45 µm filter, flasks (2) and aspirator 1 ph meter with electrode 1 Sodium hydroxide or sulfuric acid solution, 10 N and 0.100 N Water, deionized Sample cells (For information about sample cells, adapters or light shields, refer to Instrumentspecific information on page 1.) varies varies 2 Refer to Consumables and replacement items on page 4 for order information. 1

Sample collection Test procedure Collect samples in clean glass or plastic bottles with tight-fitting caps. Completely fill the bottle and immediately tighten the cap. Prevent agitation of the sample and exposure to air. Analyze the samples as soon as possible for best results. If immediate analysis is not possible, keep the samples at or below 6 C (43 F) for a maximum of 24 hours. Let the sample temperature increase to room temperature before analysis. Start 1. Start program 97 Color ADMI 1 inch, 96 Color ADMI 10 mm or 98 Color ADMI 50 mm. For information about sample cells, adapters or light shields, refer to Instrumentspecific information on page 1. 2. Pour 100 ml of sample into each of two beakers. 3. Adjust the ph of the sample in one of the beakers to ph 7.6 with a minimum volume of sodium hydroxide or sulfuric acid solution. Use a 10 N solution to make large ph adjustments and a 0.1 N solution to make small ph adjustments. Note: Sodium hydroxide increases the ph value and sulfuric acid decreases the ph value. 4. If the sample is turbid, set up the filtration apparatus (0.45 micron membrane filter, filter holder/funnel, filter flask, and aspirator). If the sample is not turbid, go to step 10. 5. Pour approximately 50 ml of the original sample through the filter to rinse the filter. Discard the rinse water. 6. Pour the remaining 50 ml of original sample through the filter. Write "original" on the flask. 7. Install a clean filter flask for the ph-adjusted sample. 8. Pour approximately 50 ml of the ph-adjusted sample through the filter. Discard the rinse water. 2 Color, ADMI Method (300 Pt-Co units)

9. Pour the remaining 50 ml of ph-adjusted sample through the filter. Write "ph-adjusted" on the flask. 10. Prepare the blank: Fill a sample cell with deionized water. 11. Prepare the sample: Fill a second sample cell with the ph-adjusted sample. Discard the remaining sample in the flask. 12. Clean the blank sample cell. Zero 13. Insert the blank into the cell holder. 14. Push ZERO. The display shows the scan status, then 0 ADMI Value. 15. Clean the prepared sample cell. 16. Insert the prepared sample into the cell holder. Read 17. Push READ. Results show in ADMI Value. 18. Do steps 11 to 17 again for the original sample. For USEPA reporting, report both results. Interferences Accuracy check Turbidity interferes directly. The filtration steps in the test procedure remove turbidity. Standard solution method Use the standard solution method to validate the test procedure, the reagents and the instrument. Items to collect: Color standard solution, 500 platinum-cobalt units 100-mL volumetric flask, Class A 20-mL volumetric pipet, Class A and pipet filler Color, ADMI Method (300 Pt-Co units) 3

Deionized water 1. Prepare a 100 platinum-cobalt units standard solution as follows: a. Use a pipet to add 20.00 ml of a 500 platinum-cobalt units standard solution into the volumetric flask. b. Dilute to the mark with deionized water. Mix well. 2. Use the test procedure to measure the concentration of the prepared standard solution. 3. Compare the expected result to the actual result. Note: The factory calibration can be adjusted slightly with the standard adjust option so that the instrument shows the expected value of the standard solution. The adjusted calibration is then used for all test results. This adjustment can increase the test accuracy when there are small variations in the reagents or instruments. Summary of method The three properties of color are hue, chroma and value. Hue is the color name, such as blue, red, green or yellow. Chroma is the color intensity (e.g., bright or dull). Value is how light or dark the color is. This method measures the color value independently of the hue and the chroma. The sample transmittance is measured from 400 to 700 nm. The transmittance values are calculated to give a set of numbers for the color that is seen by typical humans. Further calculations give a single number for the color value. This number can be found on a scale used by the American Dye Manufacturers Institute (ADMI) for color value. The ADMI uses the platinum-cobalt standard from the American Public Health Association (APHA) as the standard for color value. Although this standard is yellow, the ADMI method is applicable to all hues. Consumables and replacement items Required reagents Description Quantity/test Unit Item no. Sodium Hydroxide Solution, 10 N varies 500 ml 2545049 Sodium Hydroxide Standard Solution, 0.100 N varies 1000 ml 19153 Sulfuric Acid, 10 N varies 1000 ml 93153 Sulfuric Acid Standard Solution, 0.100 N varies 100 ml MDB 20232H Water, deionized varies 4 L 27256 Required apparatus Description Quantity/test Unit Item no. Beaker, polypropylene, 250 ml 2 each 108046 Cylinder, graduated, polypropylene, 100 ml 1 each 108142 ph meter with electrode 1 each HQ11d Recommended standards Description Unit Item no. Color standard solution, 500 platinum-cobalt units 1 L 141453 4 Color, ADMI Method (300 Pt-Co units)

Optional reagents and apparatus Description Unit Item no. Filter pump, aspirator each 213100 Filter holder, magnetic each 1352900 Filter membrane, 0.45-micron, 47-mm 100/pkg 1353000 Flask, filtering, 500 ml each 54649 Flask, volumetric, Class A, 100-mL each 1457442 Pipet, volumetric Class A, 20 ml each 1451520 Pipet filler, safety bulb each 1465100 Stopper, rubber, one-hole, No. 7 6/pkg 211907 Tubing, rubber, 5/16-in. inside diameter 3.66 m (12 ft) 56019 Color, ADMI Method (300 Pt-Co units) 5

FOR TECHNICAL ASSISTANCE, PRICE INFORMATION AND ORDERING: In the U.S.A. Call toll-free 800-227-4224 Outside the U.S.A. Contact the HACH office or distributor serving you. On the Worldwide Web www.hach.com; E-mail techhelp@hach.com HACH COMPANY WORLD HEADQUARTERS Telephone: (970) 669-3050 FAX: (970) 669-2932 Hach Company/Hach Lange GmbH, 2007 2015, 2017. All rights reserved. 08/2017, Edition 9