Reference Value Assessment for DAP Measured with a Transmission Ion Chamber in Gastrointestinal Tract Examinations

Size: px
Start display at page:

Download "Reference Value Assessment for DAP Measured with a Transmission Ion Chamber in Gastrointestinal Tract Examinations"

Transcription

1 Reference Value Assessment for DAP Measured with a Transmission Ion Chamber in Gastrointestinal Tract Examinations M. Río, S. Pérez, A. Herreros, A. Coll, A. Ruiz, J.Baró and J. Fernández Asesoría en Control y Protección Radiológica, ACPRO, S.L. Rafael Batlle, 24, Barcelona, Spain. josep.baro@acpro.es Abstract. In diagnostic radiological departments in the Health Service of Catalonia, Institut Català de la Salut, ICS, two different digital fluoroscopy table systems, the General Electric Prestige II and the Philips Optimus 80 Telediagnost, are being introduced. These digital devices are provided with a transmission ion chamber, a Wellhöffer Kermax Plus and a RTI Doseguard 100, respectively, in order to determine Dose Area Product, DAP. The need to establish dose reference values for complete examinations has led us, as a Radiation Protection Service, to undertake a study to evaluate partial and total Dose Area Products from typical complete medical examinations. These medical examinations are typically barium meals or barium enemas carried out in the gastrointestinal tract. The results obtained for DAP values, expressed in Gy cm 2, can be represented by a function of parameters that has an influence on the results: body mass index BMI from the patient being examined, fluoroscopic time, number of (digital and/or conventional), type of examination and working practices by doctors and technicians. Consequently, a relationship is estimated between the most significant parameters which enabled us to establish DAP reference values for barium meals and barium enemas. For barium meals the 3rd quartile observed for DAP values goes from 14 to 36 Gy cm 2 and in the case of barium enemas DAP values between 38 and 68 Gy cm 2 are found. For both types of examinations a contribution of 66% from total DAP is due to the fluoroscopic mode and the remainder, for barium meals, comes from digital. This contribution is the same for digital and conventional in the case of barium enemas. 1. Introduction The establishment and revision of reference values for dosimetric indicators has become important for radiodiagnostic and nuclear medicine departments due to the increasing streamlining of the equipment and methodology used. Nowadays, in the radiodiagnostic field in Catalonia there is a gradual replacement of conventional fluoroscopy systems for digital systems and the unavoidable need arises to determine the associated dose reference values of diagnostic examinations carried out with this equipment. Hence, ACPRO, S.L., as a Radiation Protection Service for primary medical centers in the Catalan Health Service Network, Institut Català de la Salut (ICS), has carried out a detailed dosimetric study of barium meals and barium enemas which are used for complex examinations in the gastroinstestinal tract:. The aim of this study is to determine reference values for these diagnostic examinations. This study is based on registered data (dose area product, fluoroscopic time, ) obtained from 145 barium meal and 74 barium enema examinations by integrating partial dose received at each anatomical stage of the examination. These medical examinations have been carried out at 4 different primary medical centers belonging to the Catalan Health Service in the Province of Barcelona. Two of the centers use table digital fluoroscopy systems with the PHILIPS brand, model Optimus 80 Telediagnost, provided with a transmission ion chamber by RTI Doseguard (Equipment A) and the other two centers, work with the same system made by General Electric, model Prestige II, equipped with a Wellhöffer Kermax Plus transmission ion chamber (Equipment B). The parameter dose area product (DAP) has been used as a reference value, due to the Real Decreto 1976/1999[1], which states that fluoroscopy systems must have a device that provides a DAP value throughout all of the examination. 1

2 In this study the obtained reference values for both types of tests are shown and the parameters on which they depend are analysed. Furthermore, the level of effective dose received by the patient, compared with the reference DAP values, is indicated. 2. Methodology For each of the complete examinations in the gastrointestinal tract: partial and total DAP values, partial and total time elapsed by the use of fluoroscopy mode, the number of conventional and digital, data of each patient (height, weight and age), the reason for prescription of the diagnostic examination, as well as details of the system operator, were registered. Additionally, DAP values associated with each conventional radiograph and each fluoroscopic time interval were monitored at each center. Furthermore, characteristics such as voltage (kv), intensity (ma), load (mas) and time (ms) were recorded for each partial DAP value generator. These values have been used to estimate effective dose using the EffDose 1 program. 3. Study of DAP values obtained The distribution of total DAP values obtained for the four primary medical centers considered in this study is shown in FIG. 1, for barium enema examinations, and in FIG. 2, for barium meal examinations. FIG. 1. Distribution of DAP values obtained for 74 barium enema examinations. Solid bars stand for centers with fluoroscopic equipment A and the dotted bars for equipment B. 1 A program using NRPB-SR262[2] for the Estimation of Effective Dose in Diagnostic Radiology from Entrance Surface Dose and Dose-Area Product Measurements. The Program could also be used to estimate the Dose or Dose-Area Product on the basis of Tube Potential, Total Filtration, mas, Single or Three Phase and Source Skin Distance base on ICRP 34[3] data. 2

3 FIG. 2. Distribution of DAP values obtained for 145 barium meal examinations. Solid bars stand for centers with fluoroscopic Equipment A and dotted bars for Equipment B. DAP mean values obtained for each center provide reference values for medical examinations of this type and are presented in Table I expressed in terms of the third quartile. Table I. Results for DAP values for barium enema and barium meal examinations at 4 different centers. Dose Area Product (Gy cm 2 ) Mean(N) (2) Minimum 1st 3 rd Median quartile quartile Maximum Center 1/A (1) 38 (13) Center 2/B 32 (24) Barium Center 3/A 51 (18) Enema Center 4/B 29 (19) Total 37 (74) Center 1/A 11(41) Center 2/B 29(34) Barium Center 3/A 19(27) Meal Center 4/B 13(43) Total 17(145) (1) The capital letter A and B is used for differentiating the two models of fluoroscopic equipment analysed. (2) N is the number of examinations studied. Based on these results, no correlation between DAP values and type of fluroscopic table system can be found, since for both brands similar results are observed. Analysis of the data does not reveal a significant correlation between total DAP and body mass index (BMI) values for the patients, which reflects their physical constitution. The distribution of this index, 3

4 BMI, calculated as the quotient between the weight and the squared height of the patients, reproduces the typical distribution of the population that lives in the province of Barcelona. However, as is shown in Table II, for the centers where the mean values of total DAP are greater, it is seen that the procedures or the methodology followed by the operators that perform the examinations mean greater fluoroscopic times and an increase in the number of. Table II. Methodology of the different centers during the performance of barium enema and barium meal examinations expressed in terms of mean values and standard deviations of the different kinds of radiological projections that are used. Digital Conventional time (s) Center 1/A (1) 11.7 ± ± ± 71.7 Barium Enema Center 2/B 12.9 ± ± ± Center 3/A 11.7 ± ± ± 71.4 Center 4/B 8.9 ± ± ± Center 1/A 14.3 ± ± ± 34.8 Barium Meal Center 2/B 22.1 ± ± ± Center 3/A 11.1 ± ± ± 94.6 Center 4/B 9.8 ± ± ± 58.9 (1) The capital letters A and B are used for differentiating the two models of fluoroscopic equipment analysed. To determine the influence of methodology in the DAP results during performance of the examinations, the results corresponding to one center have been analysed by grouping the data depending on the operator that has carried out the examinations. Table III. Mean values and standard deviation of different parameters that characterize the barium enema examinations for two operators at the same center, Center 3/A. Number of examinationss Total DAP (Gy cm 2 ) Number of digital time (s) DAP (Gy cm 2 ) Operator ± ± ± ± 11 Operator ± ± ± ± 13 As is shown in Table III, the methodology followed by the operator represents an important factor, because it conditions the fluoroscopic time used and hence, the greater contribution to the DAP, which constitutes almost % of the total value for both barium enemas or barium meals. 4. Estimation of effective dose The recommendations of the ICRP (1990) [4] establish that effective dose is the best for comparing the risks associated with x-ray examinations. However, its routine calculation from radiological parameters is not possible at a radiological facility, as specific information is required about dosage to each individual organ of the body. Hence, the dose area product is closely related to effective dose, since it takes into account the x-ray beam which affects the irradiated organs and also allows measurement of the total dose involved in 4

5 complete examinations, where different projections are used. This can be a useful parameter to assess patient effective dose. In the present study, the partial DAP values and the different projections applicable to the registered examinations have been used to estimate patient effective dose in radiography and fluoroscopic gap by means of the EffDose program, and hence, for the whole examination. These data have enabled conversion coefficients to be established to provide patient effective dose from DAP, which could be included in the patient s medical record. 5. Conclusions Since 75% of DAP values are located below the 3 rd quartile, we can establish from the results obtained in this work a DAP reference value of 46 Gy cm 2 for barium meal examinations and 23 Gy cm 2 for barium enemas. These proposed reference doses should be seen as a practical aid to increase awareness of the significance of observed levels of patient dose and hence to the promotion of optimisation of radiation protection in medical radiology. From the large number of parameters that can affect this dosimetric indicator, the operator that carries out the examination and its protocol is by far the most important factor, since he or she conditions the fluoroscopy time and the number of during the medical examination. However, the BMI parameter and the fluoroscopic system type do not display any significant correlation with DAP values, contrary to what was expected before this study was started. The detailed study of the projections done in each kind of examination makes their simulation possible by means of the EffDose program, as well as providing an estimation for the patient effective dose for all of the x-ray examination. A conversion factor for relating measured dose-area product to effective dose was estimated from these results. As a preliminary value, we have found a conversion factor of 0.28 msv/gy cm 2 for barium meals and 0.26 msv/gy cm 2 for barium enemas. On the other hand, it is observed from the data that in almost 90 % of the studied cases no pathologies are observed after the x-ray examination and that the mean age value for patients who undergo diagnostic examinations of this type is 60, although there are also cases of complete examinations in young patients. Therefore, we would like to point out the importance of carrying out an accurate medical diagnosis prior to prescribing radiological examinations of this type. REFERENCES [1] Real Decreto 1976/1999, de 23 de diciembre, por el que se establecen los criterios de calidad en radiodiagnóstico. [2] NRPB-262 [3] International Commission on Radiological Protection, Protection of the Patient in Diagnostic Radiology. Publication 34. Annals of the ICRP, Vol. 9/2. Pergamon Press (1983). [4] ICRP-60 5

Patient Dose in Single Radiodiagnostic Examinations in Cantabria (Spain) over a 10-year Period

Patient Dose in Single Radiodiagnostic Examinations in Cantabria (Spain) over a 10-year Period Patient Dose in Single Radiodiagnostic Examinations in Cantabria (Spain) over a 10-year Period M. Alonso 1, M.J. Castañeda 1, N. Díaz-Caneja 2, I. Gutiérrez 2, G. López 2 1 Servicio de Radiofísica y Protección

More information

Patient Dose Research in Interventional Radiology Suites in Rio de Janeiro, Brazil

Patient Dose Research in Interventional Radiology Suites in Rio de Janeiro, Brazil Patient Dose Research in Interventional Radiology Suites in Rio de Janeiro, Brazil Canevaro, L.V. (Presenting); Oliveira da Silva, M.W.; Dias Rodrigues, B.B. Serviço de Física Médica, Instituto de Radioproteção

More information

The role of Diagnostic Reference Levels in Optimization

The role of Diagnostic Reference Levels in Optimization The role of Diagnostic Reference Levels in Optimization Eliseo Vano Radiology Department. COMPLUTENSE UNIVERSITY MADRID Medical Physics Service. SAN CARLOS UNIVERSITY HOSPITAL MADRID - SPAIN Thursday 19th

More information

DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY: A STRATEGY FOR INTRODUCING DIAGNOSTIC REFERENCE LEVELS TAKING INTO ACCOUNT THE NATIONAL PRACTICE

DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY: A STRATEGY FOR INTRODUCING DIAGNOSTIC REFERENCE LEVELS TAKING INTO ACCOUNT THE NATIONAL PRACTICE DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY: A STRATEGY FOR INTRODUCING DIAGNOSTIC REFERENCE LEVELS TAKING INTO ACCOUNT THE NATIONAL PRACTICE Ph. R. Trueb #, A. Aroua *, H. Jung #, Th. Theiler #, J.-F. Valley

More information

Implementation of a Quality System for Medical Physics Patient Radiation Protection Management based on ISO 9001:2000

Implementation of a Quality System for Medical Physics Patient Radiation Protection Management based on ISO 9001:2000 Implementation of a Quality System for Medical Physics Patient Radiation Protection Management based on ISO 9001:2000 Sáez C, Gimeno V, Seoane A, Hermida M Servei de Física i Protecció Radiològica, Hospital

More information

A dosimetric trial for the clinical follow-up of potential skin injuries on patients undergoing interventional cardiology procedures

A dosimetric trial for the clinical follow-up of potential skin injuries on patients undergoing interventional cardiology procedures A dosimetric trial for the clinical follow-up of potential skin injuries on patients undergoing interventional cardiology procedures Vano E (1,2), Aviles P (1,2), Prieto C (2), Fernandez JM (1,2), Guibelalde

More information

Radiation Protection Dosimetry (2008), Vol. 129, No. 4, pp Advance Access publication 30 October 2007

Radiation Protection Dosimetry (2008), Vol. 129, No. 4, pp Advance Access publication 30 October 2007 Radiation Protection Dosimetry (2008), Vol. 129, No. 4, pp. 419 425 Advance Access publication 30 October 2007 doi:10.1093/rpd/ncm456 IMAGE QUALITY EVALUATION AND PATIENT DOSE ASSESSMENT OF MEDICAL FLUOROSCOPIC

More information

Study of the Spatial Distribution of Scattered Radiation Dose Around a Surgical C-arm. Assessment of Radiation Exposure to Staff.

Study of the Spatial Distribution of Scattered Radiation Dose Around a Surgical C-arm. Assessment of Radiation Exposure to Staff. Study of the Spatial Distribution of Scattered Radiation Dose Around a Surgical C-arm. Assessment of Radiation Exposure to Staff. D. Jurado, R. Pallerol, N. Jornet, J. Martin-Viera, P. Carrasco, T. Eudaldo,

More information

1. Executive Summary

1. Executive Summary 1. Executive Summary 1.1 General The fluoroscope is defined as an instrument used chiefly in industry and in the practice of medicine for observing the internal structure of objects (such as the living

More information

1. Executive Summary

1. Executive Summary 1. Executive Summary Diagnostic reference levels (DRLs), which are a form of investigation levels (ICRP, 1996), represent an important tool to optimize image quality and the radiation dose delivered to

More information

PATIENT DOSIMETRY IN INTERVENTIONAL RADIOLOGY

PATIENT DOSIMETRY IN INTERVENTIONAL RADIOLOGY 2009 International Nuclear Atlantic Conference - INAC 2009 Rio de Janeiro,RJ, Brazil, September27 to October 2, 2009 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-03-8 PATIENT DOSIMETRY

More information

Patient doses in interventional cardiology procedures.

Patient doses in interventional cardiology procedures. Patient doses in interventional cardiology procedures. J. Domienik *a, S.Papierz a, J. Jankowski a, J.Z. Peruga b a Nofer Institute of Occupational Medicine in Lodz, Radiation Protection Department, Lodz,

More information

REFERENCES LEVELS OF PATIENT RADIATION DOSE IN INTERVENTIONAL RADIOLOGY: A SPANISH CONTRIBUTION

REFERENCES LEVELS OF PATIENT RADIATION DOSE IN INTERVENTIONAL RADIOLOGY: A SPANISH CONTRIBUTION REFERENCES LEVELS OF PATIENT RADIATION DOSE IN INTERVENTIONAL RADIOLOGY: A SPANISH CONTRIBUTION F. Carrera 1,3, R. Ruiz-Cruces 3, E. Vano 2, J. Hernandez-Armas 4, M. Fernandez-Bordes 5, C. Ojeda 3 1 Radiological

More information

Quality ID #145: Radiology: Exposure Dose or Time Reported for Procedures Using Fluoroscopy National Quality Strategy Domain: Patient Safety

Quality ID #145: Radiology: Exposure Dose or Time Reported for Procedures Using Fluoroscopy National Quality Strategy Domain: Patient Safety Quality ID #145: Radiology: Exposure Dose or Time Reported for Procedures Using Fluoroscopy National Quality Strategy Domain: Patient Safety 2018 OPTIONS FOR INDIVIDUAL MEASURES: REGISTRY ONLY MEASURE

More information

Gafchromic XR-RV2 radiochromic films dosimetry in interventional radiology

Gafchromic XR-RV2 radiochromic films dosimetry in interventional radiology Gafchromic XR-RV2 radiochromic films dosimetry in interventional radiology A. Neocleous *1, E. Yakoumakis 1, G. Gialousis 2, A. Dimitriadis 1, N. Yakoumakis 3 and E. Georgiou 1 1 Medical Physics Department,

More information

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May Dose Reference Levels

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May Dose Reference Levels 2033-7 Joint ICTP/ Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dose Reference Levels Peter Homolka EFOMP Training Course on Medical Physics in Diagnostic

More information

Radiation Protection Dosimetry (2008), Vol. 132, No. 1, pp Advance Access publication 25 September 2008

Radiation Protection Dosimetry (2008), Vol. 132, No. 1, pp Advance Access publication 25 September 2008 Radiation Protection Dosimetry (2008), Vol. 132, No. 1, pp. 18 24 Advance Access publication 25 September 2008 doi:10.1093/rpd/ncn247 CORRELATION OF PATIENT MAXIMUM SKIN DOSES IN CARDIAC PROCEDURES WITH

More information

Justification for Training

Justification for Training This educational program is being offered to provide training to medical practitioners who perform or supervise the performance of diagnostic procedures utilizing fluoroscopic x- ray units. There are many

More information

New Joint Commission Requirements for Fluoroscopy Services

New Joint Commission Requirements for Fluoroscopy Services New Joint Commission Requirements for Fluoroscopy Services Melissa C. Martin, M.S., FAAPM, FACR Past President AAPM Florida Chapter Fall Meeting Orlando, FL September 14, 2018 New Requirements Effective

More information

Radiation Protection in Fluoroscopy

Radiation Protection in Fluoroscopy Radiation Protection in Fluoroscopy Introduction Physicians performing fluoroscopic guided procedures should be aware of the potential for serious radiation-induced skin injury to the patient. Occasionally

More information

Equipment Quality Control for Fluoroscopy February 22, Imaging Physics CancerCare Manitoba

Equipment Quality Control for Fluoroscopy February 22, Imaging Physics CancerCare Manitoba Equipment Quality Control for Fluoroscopy February 22, 2018 Imaging Physics CancerCare Manitoba Purpose An equipment quality control (QC) program establishes baseline performance levels, tracks system

More information

Radiation protection for theatre staff (refer guidelines) DISTANCE, TIME, SHIELDING principle should be applied

Radiation protection for theatre staff (refer guidelines) DISTANCE, TIME, SHIELDING principle should be applied Radiation protection for theatre staff (refer guidelines) DISTANCE, TIME, SHIELDING principle should be applied Monitoring (Refer guidelines) All full time theatre personnel must be monitored with dosimeters.

More information

Radiography Curriculum Analysis

Radiography Curriculum Analysis Program Number Program Name Date / /20 Radiography Curriculum Analysis DIRECTIONS: Determine the course(s) in which each of the following content area is covered and enter the course number(s) and/or title(s).

More information

FLUOROSCOPIC SAFETY AND PATIENT AND OPERATOR RADIATION DOSE REDUCTION: CREATING A FOCUSED FLUOROSCOPIC CURRICULUM FOR RADIOLOGY RESIDENTS

FLUOROSCOPIC SAFETY AND PATIENT AND OPERATOR RADIATION DOSE REDUCTION: CREATING A FOCUSED FLUOROSCOPIC CURRICULUM FOR RADIOLOGY RESIDENTS FLUOROSCOPIC SAFETY AND PATIENT AND OPERATOR RADIATION DOSE REDUCTION: CREATING A FOCUSED FLUOROSCOPIC CURRICULUM FOR RADIOLOGY RESIDENTS Rutgers Robert Wood Johnson Medical School, New Brunswick, NJ Abstract

More information

Patient doses from CT examinations in region of Prishtina, Kosovo

Patient doses from CT examinations in region of Prishtina, Kosovo Patient doses from CT examinations in region of Prishtina, Kosovo Sehad KADIRI, Gëzim HODOLLI, Kostandin DOLLANI Institute of Occupational Medicine, Radiation Protection Service, Obiliq, Kosovo Institute

More information

Original article Paediatrics Today 2016;12(1):75-80 DOI /p

Original article Paediatrics Today 2016;12(1):75-80 DOI /p Original article Paediatrics Today 2016;12(1):75-80 DOI 10.5457/p2005-114.138 THE EFFECT OF ANTI-SCATTER GRID ON RADIATION DOSE IN CHEST RADIOGRAPHY IN CHILDREN Igor ŠABIČ 1*, Damjana KLJUČEVŠEK 1, Martin

More information

ESTIMATION OF ORGAN DOSES OF PATIENT UNDERGOING HEPATIC CHEMOEMBOLIZATION PROCEDURES

ESTIMATION OF ORGAN DOSES OF PATIENT UNDERGOING HEPATIC CHEMOEMBOLIZATION PROCEDURES X Congreso Regional Latinoamericano IRPA de Protección y Seguridad Radiológica Radioprotección: Nuevos Desafíos para un Mundo en Evolución Buenos Aires, 12 al 17 de abril, 2015 SOCIEDAD ARGENTINA DE RADIOPROTECCIÓN

More information

Interventional Fluoroscopy

Interventional Fluoroscopy NATIONAL CANCER INSTITUTE THE SOCIETY OF INTERVENTIONAL RADIOLOGY Interventional Fluoroscopy Reducing Radiation Risks for Patients and Staff CONTENTS Introduction 1 Increasing use and complexity of interventional

More information

FLUOROSCOPIC PATIENT & OPERATOR RISK REDUCTION

FLUOROSCOPIC PATIENT & OPERATOR RISK REDUCTION FLUOROSCOPIC PATIENT & OPERATOR RISK REDUCTION Ray Dielman RSO ST ANTHONY S HOSPITAL James Menge CHP ARSO ST ANTHONY S HOSPITAL BAYCARE HEALTH SYSTEM 1 FLUROSCOPY EDUCATIONAL OBJECTIVES I. Understand the

More information

Radiology Equipment and Reports

Radiology Equipment and Reports Section 2 Radiology Equipment and Reports Contents Risk Assessment RPII Licence of Equipment Inventory of Radiation Equipment Criteria of Acceptability Setting and Reviewing Nominal Replacement Dates National

More information

White Paper. Dose Tracking System. Patient Skin Dose Estimates in Real Time

White Paper. Dose Tracking System. Patient Skin Dose Estimates in Real Time White Paper Dose Tracking System Patient Skin Dose Estimates in Real Time INTRODUCTION In an era of increased utilization of ionizing radiation for medical diagnostic and interventional procedures and

More information

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method Iran. J. Radiat. Res., 2004; 1(4): 187-194 Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method M.R. Ay 1, M. Shahriari 2, S. Sarkar 3, P.

More information

Patient Dosimetry and Quality Control in Diagnostic Radiology

Patient Dosimetry and Quality Control in Diagnostic Radiology بسم االله الرحمن الرحيم Sudan Academy of Sciences (SAS) Atomic Energy Council Patient Dosimetry and Quality Control in Diagnostic Radiology Ibrahim Idris Suliman B.Sc. (Physics), M.Sc. (Nuclear Physics)

More information

Minnesota Rules, Chapter 4732 X-ray Revision

Minnesota Rules, Chapter 4732 X-ray Revision Minnesota Rules, Chapter 4732 X-ray Revision FLUOROSCOPIC X-RAY SYSTEMS, 1.0 Subpart 17. Subpart 18. Subpart 19. Subpart 20. Subpart 21. Subpart 22. Subpart 23. Subpart 24. Subpart 25. Equipment performance

More information

Types of Radiation Units

Types of Radiation Units RADIATION UNITS Types of Radiation Units Exposure in air - amount of ionization produced by or radiation in air. Unit - (R) or / (c/kg) Absorbed Dose energy transfer of / (unit of energy) of any. Used

More information

Reducing Unnecessary Medical Imaging Exposure

Reducing Unnecessary Medical Imaging Exposure Reducing Unnecessary Medical Imaging Exposure CDR Sean M. Boyd, MPH, USPHS U.S. Food and Drug Administration Center for Devices and Radiological Health April 30, 2010 Objectives FDA s role in ensuring

More information

Glossary to The Practice Standards for Medical Imaging and Radiation Therapy

Glossary to The Practice Standards for Medical Imaging and Radiation Therapy Glossary to The Practice Standards for Medical Imaging and Radiation Therapy Accuracy Ability of the bone mineral densitometry system to measure the true value of an object. Act anything done, being done,

More information

Donna Newman B.A. RT ( R) CNMT Director of Professional Practice ISRRT

Donna Newman B.A. RT ( R) CNMT Director of Professional Practice ISRRT Donna Newman B.A. RT ( R) CNMT Director of Professional Practice ISRRT Responsibilities Worker (BSS 3.83) Workers shall: follow any applicable rules for protection use properly the monitoring devices and

More information

Evaluation of organ and effective doses for CAG based on in-phantom dosimetry

Evaluation of organ and effective doses for CAG based on in-phantom dosimetry Evaluation of organ and effective doses for CAG based on in-phantom dosimetry Poster No.: C-0619 Congress: ECR 2013 Type: Scientific Exhibit Authors: Y. Matsunaga, A. Kawaguchi, S. Suzuki, Y. Takikawa,

More information

The Identification and Evaluation of Likely High-Dose Electrophysiology (EP) Procedures at a Large Teaching Hospital

The Identification and Evaluation of Likely High-Dose Electrophysiology (EP) Procedures at a Large Teaching Hospital The Identification and Evaluation of Likely High-Dose Electrophysiology (EP) Procedures at a Large Teaching Hospital Bradford Taylor, MS, RT(R) Associate RSO, UNC-Chapel Hill October 18, 2007 NCHPS Agenda

More information

1 Digital Imaging and Communications in Medicine 2 (DICOM) 3 Sup Retirement of Radiation Dose Module from Modality. 4 Performed Procedure Step

1 Digital Imaging and Communications in Medicine 2 (DICOM) 3 Sup Retirement of Radiation Dose Module from Modality. 4 Performed Procedure Step 1 Digital Imaging and Communications in Medicine 2 (DICOM) 3 Sup 201 - Retirement of Radiation Dose Module from Modality 4 Performed Procedure Step 5 DICOM Standards Committee - Working Group 8 - Structured

More information

Patient doses in lower limb angiography in a recently installed flat detector x-ray system

Patient doses in lower limb angiography in a recently installed flat detector x-ray system Patient doses in lower limb angiography in a recently installed flat detector x-ray system Poster No.: C-0939 Congress: ECR 2011 Type: Authors: Keywords: DOI: Scientific Paper V. Tsapaki, D. Papavdis,

More information

Function of this control

Function of this control Page 1 of 7 Introduction The goals of this training are to develop a better appreciation of equipment controls, scatter radiation in the interventional suite, the radiation output capabilities of the angiography

More information

Improving protocols and procedures for strengthened radiation protection in interventional procedures

Improving protocols and procedures for strengthened radiation protection in interventional procedures Improving protocols and procedures for strengthened radiation protection in interventional procedures R. Loose German Commission on Radiological Protection (SSK) German Roentgen Society (DRG) Institute

More information

What s the Issue? Radiation Dose in the OR. Exposure to OR Personnel. Effects of Proper Collimation. Image-guided procedures are increasingly complex

What s the Issue? Radiation Dose in the OR. Exposure to OR Personnel. Effects of Proper Collimation. Image-guided procedures are increasingly complex Radiation Dose in the OR Robert G. Gould, Sc.D. Professor and Vice Chair Department of Radiology and Biomedical Imaging University of California San Francisco What s the Issue? Image-guided procedures

More information

The following 2 points present a particular concern for us:

The following 2 points present a particular concern for us: COCIR Position Paper on the Proposal for a Directive laying down basic safety standards for protection against the dangers arising from exposure to ionizing radiation 1 COCIR is the leading voice of the

More information

Survey of Radiation Exposure for adult and pediatric patients undergoing CT procedures in Latin America

Survey of Radiation Exposure for adult and pediatric patients undergoing CT procedures in Latin America Survey of Radiation Exposure for adult and pediatric patients undergoing CT procedures in Latin America Simone Kodlulovich *1, Helen Khoury 2, Patricia Mora 3, Alfredo Ramirez 4, Fernando Leyton 5, Juan

More information

A Paradigm Shift in Patient Dose Monitoring

A Paradigm Shift in Patient Dose Monitoring VL White Paper A Paradigm Shift in Patient Dose Monitoring Dose Tracking System Andrew Kuhls-Gilcrist, Ph.D., D.A.B.R. Senior XRVL Clinical Science Manager Canon Medical Systems, Inc. Millions of fluoroscopically

More information

Introductory Presentation: EURAMED s Vision on Medical Radiation Protection (Research)

Introductory Presentation: EURAMED s Vision on Medical Radiation Protection (Research) Introductory Presentation: EURAMED s Vision on Medical Radiation Protection (Research) Christoph Hoeschen Otto-von-Guericke University Magdeburg on behalf of the executive board What is EURAMED?!!!Since

More information

Glossary to The Practice Standards for Medical Imaging and Radiation Therapy

Glossary to The Practice Standards for Medical Imaging and Radiation Therapy 0 0 Glossary to The Practice Standards for Medical Imaging and Radiation Therapy Accuracy Ability of the bone mineral densitometry system to measure the true value of an object. Act anything done, being

More information

Table of Contents. 1. Introduction. 2. Authorisation Process. 3. X-Ray Facility Design Considerations and Minimum Requirements. 3.

Table of Contents. 1. Introduction. 2. Authorisation Process. 3. X-Ray Facility Design Considerations and Minimum Requirements. 3. Table of Contents 1. Introduction 2. Authorisation Process 3. X-Ray Facility Design Considerations and Minimum Requirements 3.1 Location 3.2 Room Size 3.3 Walls 3.4 Viewing Window Requirements 3.5 Change

More information

In recent years, the benefits and complexity of interventional

In recent years, the benefits and complexity of interventional PATIENT SAFETY Patient Radiation Dose Management in the Follow-Up of Potential Skin Injuries in Neuroradiology E. Vano, J.M. Fernandez, R.M. Sanchez, D. Martinez, L. Lopez Ibor, A. Gil, and C. Serna-Candel

More information

Local dose reference levels for abdominal interventional radiology procedures

Local dose reference levels for abdominal interventional radiology procedures Local dose reference levels for abdominal interventional radiology procedures Poster No.: C-2053 Congress: ECR 2014 Type: Scientific Exhibit Authors: L. Hadid-Beurrier 1, M.-J. Waryn 2, Y. Ajavon 1, O.

More information

DIRECT MEASUREMENT OF SKIN DOSE WITH RADIOCHROMIC FILMS IN FLUOROSCOPY GUIDED VESSEL REPAIR

DIRECT MEASUREMENT OF SKIN DOSE WITH RADIOCHROMIC FILMS IN FLUOROSCOPY GUIDED VESSEL REPAIR DIRECT MEASUREMENT OF SKIN DOSE WITH RADIOCHROMIC FILMS IN FLUOROSCOPY GUIDED VESSEL REPAIR J. KALEF-EZRA, EZRA, * 1 E. KATSAROU, 1 S. KARAVASILIS, 1 M. MATSAGKAS, 2 L.K. MICHALIS 3 1 Medical Physics Department,

More information

Deviation Value for Conventional X-ray in Hospitals in South Sulawesi Province from 2014 to 2016

Deviation Value for Conventional X-ray in Hospitals in South Sulawesi Province from 2014 to 2016 Journal of Physics: Conference Series PAPER OPEN ACCESS Deviation Value for Conventional X-ray in Hospitals in South Sulawesi Province from 2014 to 2016 To cite this article: Ilham Bachtiar et al 2018

More information

IV. 1. Skin Dose Measurement for Patients Using Imaging Plates in Interventional Radiology Procedures

IV. 1. Skin Dose Measurement for Patients Using Imaging Plates in Interventional Radiology Procedures CYRIC Annual Report 2004 IV. 1. Skin Dose Measurement for Patients Using Imaging Plates in Interventional Radiology Procedures Ohuchi H., Satoh T. *, Eguchi Y. *, and Mori K. ** Graduate School of Pharmaceutical

More information

ISSN Radiation. Protection. European Guidelines on Diagnostic Reference Levels for Paediatric Imaging. Energy

ISSN Radiation. Protection. European Guidelines on Diagnostic Reference Levels for Paediatric Imaging. Energy ISSN 2315-2826 Radiation Protection N 185 European Guidelines on Diagnostic Reference Levels for Paediatric Imaging Energy This report was prepared under contract ENER/2013/NUCL/Sl2.671441. The information

More information

Act anything done, being done, or to be done; the process of doing. Synonymous with procedure and clinical services.

Act anything done, being done, or to be done; the process of doing. Synonymous with procedure and clinical services. Act anything done, being done, or to be done; the process of doing. Synonymous with procedure and clinical services. Action plan A program or method that explains the actions or steps to be taken. Advanced-practice

More information

Production of 131-I iodide capsules in Argentina

Production of 131-I iodide capsules in Argentina Production of 131-I iodide capsules in Argentina J.O. Nicolini (1), G. Baró (1), G. Casale (1), L. Silva (1), M. Nicolini (1), P. Apolinario (1), G. Oviedo (1), G. Orlandino (1), R. Caro (1), M.J. Salgueiro

More information

How to Optimize Radiation Protection During Complex Peripheral

How to Optimize Radiation Protection During Complex Peripheral How to Optimize Radiation Protection During Complex Peripheral G. Bartal MD, FCIRSE, FSIR Dept of Medical Imaging Meir MC, Kfar Saba, Affiliated to Sackler Medical School, Tel Aviv University, Tel Aviv

More information

Computed Tomography: Optimization of acquisition protocols & Justification of clinical referrals. Koos Geleijns, medical physicist

Computed Tomography: Optimization of acquisition protocols & Justification of clinical referrals. Koos Geleijns, medical physicist Computed Tomography: Optimization of acquisition protocols & Justification of clinical referrals Koos Geleijns, medical physicist CT delivers excellent 3D image quality CT delivers excellent 3D image quality

More information

Radiation Exposure During Endovascular Procedures: Methods to Protect Patients, Physicians and OR Personnel

Radiation Exposure During Endovascular Procedures: Methods to Protect Patients, Physicians and OR Personnel Radiation Exposure During Endovascular Procedures: Methods to Protect Patients, Physicians and OR Personnel Melissa L. Kirkwood MD Associate Professor Division of Vascular and Endovascular Surgery UT Southwestern

More information

Patient radiation dose and image quality optimization in CT. Optimiser la dose patient et la qualité image en scanner

Patient radiation dose and image quality optimization in CT. Optimiser la dose patient et la qualité image en scanner Conférence SPIMED - 8 juin 2012 IQ Patient safety Dose Reduction Patient radiation dose and image quality optimization in CT Optimiser la dose patient et la qualité image en scanner Philippe Coulon, PhD.

More information

Disclaimer and Notices.

Disclaimer and Notices. Fluoroscopy Dose Reduction and Clinical Management Fluoroscopy Trajectories (Curves), Global Dose Monitoring and Tracking of Patient Dose. Pei Jan Paul Lin, Ph.D. Division of Radiation Physics and Biology

More information

Optimizing Radiation Dose for Samsung GC85A

Optimizing Radiation Dose for Samsung GC85A White Paper Article # WP201602-DR-Dose/ Issue Date 23 Feb., 2016 Optimizing Radiation Dose for Samsung GC85A Hong Eo, M.D. / Hyesun Hwang, M.D. / Samsung Medical Center Introduction ALARA (As Low As Reasonably

More information

Robert Boyle Prof of Medical Physics Trinity College, g, Dublin IAEA, Vienna

Robert Boyle Prof of Medical Physics Trinity College, g, Dublin IAEA, Vienna The Blue Book Published, 1988, by predecessor of Radiological Protection Institute of Ireland (RPII) Prof fjim Ml Malone Robert Boyle Prof of Medical Physics Trinity College, g, Dublin IAEA, Vienna Problems

More information

Journal of Theoretical and Applied Information Technology.

Journal of Theoretical and Applied Information Technology. MONITORING AND QUALITY CONTROL TESTS OF NIGERIAN NATIONAL PETROLEUM CORPORATION (NNPC) DIAGONSTIC FACILITIES: PART OF QUALITY ASSURANCE PROGRAMME OF RADIOLOGY IN NIGERIA 1 OLUWAFISOYE, P.A, 2 OLOWOOKERE,

More information

IMAGE GUIDANCE DOSES IN RADIOTHERAPY. Scott Crowe

IMAGE GUIDANCE DOSES IN RADIOTHERAPY. Scott Crowe IMAGE GUIDANCE DOSES IN RADIOTHERAPY Scott Crowe IMAGING DOSE ALARA requires that imaging dose is managed and optimised Imaging dose presents an increased risk for paediatric patients Increased effective

More information

Radiation Protection in Veterinary Medicine

Radiation Protection in Veterinary Medicine Auslntiian Go\'t~ nllnl'nt Australian Radiation Protl'Ction and NudC!lf S!lfCl~' Agency Radiation in Veterinary Medicine RADIATION PROTECTIO N SERIES No. :17 Radiation The Radiation is published by the

More information

Dose at an interventional reference point: Comparison between displayed values of X-ray units and actual measured values

Dose at an interventional reference point: Comparison between displayed values of X-ray units and actual measured values Dose at an interventional reference point: Comparison between displayed values of X-ray units and actual measured values Poster No.: C-1844 Congress: ECR 2012 Type: Authors: Keywords: DOI: Educational

More information

5DGLDWLRQSURWHFWLRQ *XLGDQFHRQPHGLFDO H[SRVXUHVLQPHGLFDODQG ELRPHGLFDOUHVHDUFK (XURSHDQ&RPPLVVLRQ

5DGLDWLRQSURWHFWLRQ *XLGDQFHRQPHGLFDO H[SRVXUHVLQPHGLFDODQG ELRPHGLFDOUHVHDUFK (XURSHDQ&RPPLVVLRQ 5DGLDWLRQSURWHFWLRQ *XLGDQFHRQPHGLFDO H[SRVXUHVLQPHGLFDODQG ELRPHGLFDOUHVHDUFK (XURSHDQ&RPPLVVLRQ European Commission 5DGLDWLRQSURWHFWLRQ *8,'$1&(210(',&$/(;32685(6,10(',&$/$1'%,20(',&$/5(6($5&+ 1998 Directorate-General

More information

Evaluation of the shielding efficiency and space dose of the tungsten shielding system

Evaluation of the shielding efficiency and space dose of the tungsten shielding system Volume 118 No. 19 2018, 2267-2281 ISSN: 1311-8080 (printed version); ISSN: 1314-3395 (on-line version) url: http://www.ijpam.eu ijpam.eu Evaluation of the shielding efficiency and space dose of the tungsten

More information

2018/7/31. Primary Objective

2018/7/31. Primary Objective Andrew Kuhls-Gilcrist, PhD, DABR July 31, 2018 Primary Objective The purpose of this session is to allow vendors to present their solutions and describe how their product can empower the clinician and

More information

4p All 2017 JUN 27 AN 11:29. Medical reportable event for radiation-producing diagnostic or. interventional X-ray procedures -

4p All 2017 JUN 27 AN 11:29. Medical reportable event for radiation-producing diagnostic or. interventional X-ray procedures - TO: PA DEP Bureau of Radiation Protection interventional X-ray procedures - It is not clear what the intent of Radiation Dose to Skin of the Neck, Torso, Pelvis, Buttocks, or Arms, in an article published

More information

Optimizing the Use of Pulsed Fluoroscopy to Reduce Radiation Exposure to Children

Optimizing the Use of Pulsed Fluoroscopy to Reduce Radiation Exposure to Children Optimizing the Use of Pulsed Fluoroscopy to Reduce Radiation Exposure to Children Mervyn D. Cohen, MB, ChB, MD Radiologists desire to keep radiation dose as low as possible. Pulsed fluoroscopy provides

More information

Methods for measuring fluoroscopic skin dose

Methods for measuring fluoroscopic skin dose Pediatr Radiol (2006) 36 (Suppl 2): 136 140 DOI 10.1007/s00247-006-0193-3 ALARA Stephen Balter Methods for measuring fluoroscopic skin dose Published online: 22 July 2006 # Springer-Verlag 2006 Abstract

More information

DETERMINATION OF DOSES AND CANCER RISK TO PATIENTS UNDERGOING DIGITAL X-RAY EXAMINATIONS AT THE TAMALE TEACHING HOSPITAL

DETERMINATION OF DOSES AND CANCER RISK TO PATIENTS UNDERGOING DIGITAL X-RAY EXAMINATIONS AT THE TAMALE TEACHING HOSPITAL DETERMINATION OF DOSES AND CANCER RISK TO PATIENTS UNDERGOING DIGITAL X-RAY EXAMINATIONS AT THE TAMALE TEACHING HOSPITAL A THESIS SUBMITTED TO THE DEPARTMENT OF NUCLEAR SAFETY AND SECURITY OF THE SCHOOL

More information

The Role of Tomosynthesis Imaging at Our Institute and X-ray Dose Optimization

The Role of Tomosynthesis Imaging at Our Institute and X-ray Dose Optimization R/F The Role of Tomosynthesis Imaging at Our Institute and X-ray Dose Optimization Tetsuta Izumi R.T. Osaka International Cancer Institute, Diagnostic and Interventional Radiology Section 1 Osaka International

More information

Dose management at Ziehm Imaging for 2D imaging

Dose management at Ziehm Imaging for 2D imaging 01 Dose management at Ziehm Imaging for 2D imaging An educational paper on general dose principles and concepts as well as further insights on dose management for 2D imaging at Ziehm Imaging. International

More information

Ocean 2014 Your X-ray QA Partner

Ocean 2014 Your X-ray QA Partner Ocean 2014 Your X-ray QA Partner Ocean 2014 is the diagnostic software to use with your RTI instrument. It displays all your measurements and waveforms gathered on an easy-to-read screen. Ocean suits everybody

More information

ATOMSCOPE HF200A. Operating Instruction Manual. Ver. 3.1

ATOMSCOPE HF200A. Operating Instruction Manual. Ver. 3.1 ATOMSCOPE HF00A Operating Instruction Manual Ver.. This special purpose portable equipment is for diagnostic radiography only. It is not designed, nor does it meet, the standards for any other use. CONTENTS

More information

GE Healthcare. Senographe Crystal. The clear and simple choice in digital mammography

GE Healthcare. Senographe Crystal. The clear and simple choice in digital mammography GE Healthcare Senographe Crystal The clear and simple choice in digital mammography The Senographe Crystal mammography system makes it easy to transition to full-field digital mammography. Ease of use,

More information

ICRP 2006 RECOMMENDATIONS

ICRP 2006 RECOMMENDATIONS ICRP 2006 RECOMMENDATIONS Annie SUGIER GT CIPR - 7 décembre 2006 GT CIPR 7 décembre 2006 1 Why ICRP Recommendations? To take account of new biological and physical information and of trends in the setting

More information

Image-Quality Perception as a Function of Dose in Digital Radiography

Image-Quality Perception as a Function of Dose in Digital Radiography Medical Physics and Informatics Original Research Medical Physics and Informatics Original Research Thomas Lehnert 1 Nagy N. N. Naguib Huedayi Korkusuz Ralf W. Bauer J. Matthias Kerl Martin G. Mack Thomas

More information

Comprehensive Dose Management for Improved Patient Care.

Comprehensive Dose Management for Improved Patient Care. Comprehensive Dose Management for Improved Patient Care. Set the standard for dose management Toshiba is committed to continuously advancing dose reduction technologies and incorporating these into its

More information

Image-Quality Perception as a Function of Dose in Digital Radiography

Image-Quality Perception as a Function of Dose in Digital Radiography Medical Physics and Informatics Original Research Medical Physics and Informatics Original Research PROOF Copyrighted Material For Review Only Not for Distribution Page Copy Image-Quality Perception as

More information

NHS Imaging and Radiodiagnostic activity

NHS Imaging and Radiodiagnostic activity NHS Imaging and Radiodiagnostic activity NHS Imaging and Radiodiagnostic activity 2013/14 Release Version number: 1 First published: 6 th August 2014 Prepared by: NHS England Analytical Services (Operations)

More information

Contents Preface 1. Introduction and Recommendations

Contents Preface 1. Introduction and Recommendations Contents Preface............................................. iii 1. Introduction and Recommendations................. 1 1.1 Purpose and Scope............................... 1 1.2 Quantities and Units.............................

More information

There are three general methods for reducing one's personal radiation exposure. The application of each when working around fluoroscopy units will be

There are three general methods for reducing one's personal radiation exposure. The application of each when working around fluoroscopy units will be There are three general methods for reducing one's personal radiation exposure. The application of each when working around fluoroscopy units will be discussed in more detail. The ALARA concept was introduced

More information

Endovascular Simulators Reduce Patient and Staff Exposure

Endovascular Simulators Reduce Patient and Staff Exposure Endovascular Simulators Reduce Patient and Staff Exposure Gabriel Bartal MD, FCIRSE, FSIR Dept of Diagnostic and Interventional Radiology, Meir MC, Kfar-Saba, Affiliated to Tel-Aviv University, Tel-Aviv,

More information

CALIFORNIA LAW AND RESPONSE + FIVE QUICK LESSONS ON CT DOSE OPTIMIZATION

CALIFORNIA LAW AND RESPONSE + FIVE QUICK LESSONS ON CT DOSE OPTIMIZATION CALIFORNIA LAW AND RESPONSE + FIVE QUICK LESSONS ON CT DOSE OPTIMIZATION John M. Boone, PhD Professor and Vice Chair (Research) of Radiology Professor of Biomedical Engineering Department of Radiology

More information

Accurate Assessment of Patient Effective Radiation Dose and Associated Detriment Risk From Radiofrequency Catheter Ablation Procedures

Accurate Assessment of Patient Effective Radiation Dose and Associated Detriment Risk From Radiofrequency Catheter Ablation Procedures Accurate Assessment of Patient Effective Radiation Dose and Associated Detriment Risk From Radiofrequency Catheter Ablation Procedures Kostas Perisinakis, PhD; John Damilakis, PhD; Nicholas Theocharopoulos,

More information

Experience With Patient Dosimetry and Quality Control Online for Diagnostic and Interventional Radiology Using DICOM Services

Experience With Patient Dosimetry and Quality Control Online for Diagnostic and Interventional Radiology Using DICOM Services Medical Physics and Informatics Review Vano et al. Online Quality Control for Diagnostic and Interventional Radiology Medical Physics and Informatics Review FOCUS ON: Eliseo Vano 1,2 Jose I. Ten 2,3 Jose

More information

Quality Estimation of Filtration of Diagnostic X-Ray at Federal Medical Centre and Bishop Murray Hospital Makurdi using Half-Value Layer (HVL)

Quality Estimation of Filtration of Diagnostic X-Ray at Federal Medical Centre and Bishop Murray Hospital Makurdi using Half-Value Layer (HVL) Quality Estimation of Filtration of Diagnostic X-Ray at Federal Medical Centre and Bishop Murray Hospital Makurdi using Half-Value Layer (HVL) N.B. Akaagerger 1 F.O.Ujah 2 T.C. Akpa 3 1.Department of Physics,

More information

ACR AAPM TECHNICAL STANDARD FOR DIAGNOSTIC MEDICAL PHYSICS PERFORMANCE MONITORING OF FLUOROSCOPIC EQUIPMENT

ACR AAPM TECHNICAL STANDARD FOR DIAGNOSTIC MEDICAL PHYSICS PERFORMANCE MONITORING OF FLUOROSCOPIC EQUIPMENT The American College of Radiology, with more than 30,000 members, is the principal organization of radiologists, radiation oncologists, and clinical medical physicists in the United States. The College

More information

FLORIDA HOSPITAL DIAGNOSTIC RADIOLOGY RESIDENCY PROGRAM PHYSICS GOALS AND OBJECTIVES

FLORIDA HOSPITAL DIAGNOSTIC RADIOLOGY RESIDENCY PROGRAM PHYSICS GOALS AND OBJECTIVES FLORIDA HOSPITAL DIAGNOSTIC RADIOLOGY RESIDENCY PROGRAM PHYSICS GOALS AND OBJECTIVES Goals and objectives are based on recommendations and requirements from the AAPM, RSNA, NRC, FL DOH, and ACGME Module

More information

MANAGING PATIENT DOSE IN FLUOROSCOPIC PROCEDURES

MANAGING PATIENT DOSE IN FLUOROSCOPIC PROCEDURES MANAGING PATIENT DOSE IN FLUOROSCOPIC PROCEDURES Keith J. Strauss, MSc MSc,, FAAPM Director, Radiology Physics & Engineering Children s Hospital Harvard Medical School DILEMAS A. Fluoroscopically Guided

More information

Radiological Protection of Patient and Operator in Interventional Radiology

Radiological Protection of Patient and Operator in Interventional Radiology Radiological Protection of Patient and Operator in Interventional Radiology JMAJ 44(11): 484 489, 2001 Hironobu NAKAMURA Professor and Chairman, Department of Radiology, Osaka University Graduate School

More information

Low-dose and High-resolution Cardiovascular Imaging with Revolution* CT

Low-dose and High-resolution Cardiovascular Imaging with Revolution* CT GE Healthcare Case study Low-dose and High-resolution Cardiovascular Imaging with Revolution* CT Jean-Louis Sablayrolles, M.D. Laurent Macron, M.D. Jacques Feignoux, M.D. Centre Cardiologique du Nord,

More information

مباحث :IAEA Training Material on Radiation Protection in Cardiology. List of Topics & Objectives١.doc

مباحث :IAEA Training Material on Radiation Protection in Cardiology. List of Topics & Objectives١.doc - 2-3 دوره آموزش حفاظت در برابر پرتوهاي يونيزان در كارديولوژي پاييز 1394 سرفصل مطالب: 1 -فصل هاي 9 8 7 6 5 4 3 2 1 و 13 از كتاب 1 حفاظت در برابر اشعه (از 1 تا 13 بجز فصل هاي 11 10 و (12 فصل هاي 3 2 1 از

More information