SCICLONE SURESELECT XT NGS WORKSTATION USER S GUIDE: AGILENT TARGET ENRICHMENT ON THE

Size: px
Start display at page:

Download "SCICLONE SURESELECT XT NGS WORKSTATION USER S GUIDE: AGILENT TARGET ENRICHMENT ON THE"

Transcription

1 USER S GUIDE: AGILENT SURESELECT XT TARGET ENRICHMENT ON THE SCICLONE NGS WORKSTATION Agilent SureSelect XT Target Enrichment System for Illumina Paired-End Sequencing Libraries with Multiplexing (Protocol v1.2 May 2011) Sciclone NGS Workstation with Maestro 6.0 Sciclone NGSx Workstation with Maestro 6.2

2 Table of Contents Introduction 3 Maestro SureSelect XT Overview 3-6 Required Materials and Reagents 7-8 Reagents 7 General Laboratory Equipment and Supplies 7 Sciclone NGS Accessories 7 Consumables 8 Running the Maestro SureSelect XT Workflow 8 Sample Preparation 8 SureSelectXT Initial SPRI Cleanup Application (Day 1) 8 Quality Control of Initial SPRI Samples (Optional) 10 SureSelectXT Library Prep Application (Day 2) 10 Quality Control and Quantification of Amplified Libraries 13 SureSelectXT Pre-Capture Normalization Application (Day 3) 14 SureSelectXT Hyb Setup Application (Day 3) 15 SureSelectXT Target Selection Application (Day 4) 17 Quality Control and Quantification of Captured Libraries 20 Appendix A: Step-by-Step Guide to the SureSelectXT Initial SPRI Cleanup Application 21 Appendix B: Step-by-Step Guide to the SureSelectXT Library Prep Application 21 Appendix C: Step-by-Step Guide to the SureSelectXT Normalization Application 22 Appendix D: Step-by-Step Guide to the SureSelectXT Hyb Setup Application 22 Appendix E: Step-by-Step Guide to the SureSelectXT Target Selection Application 22 2

3 Introduction Preparation of DNA samples for cluster generation and sequencing on the Illumina platform requires a series of manipulations to efficiently ligate appropriate indexed adapters onto DNA fragments to produce paired-end libraries. Selection of specific regions of the genome for sequencing requires additional steps including hybridization with capture probes, isolation of captured sequences, and enrichment of captured libraries. Automating the process has the advantage of avoiding sample tracking errors and reducing sample-tosample variability while dramatically increasing throughput. The Maestro-based SureSelect XT Workflow from PerkinElmer provides a pre-programmed solution for the SureSelect XT Target Enrichment System on the PerkinElmer Sciclone NGS Workstation. Overview of the Maestro Workflow The Maestro Workflow for SureSelect XT sample preparation is a validated process for library preparation from fragmented DNA samples that follows the steps outlined in Figure 1. Samples are processed in 96-well PCR plates, and the number of samples to process (1 to 12 columns of 8 samples each) is selected at the start of each run. Pre-set tip-tracking utilities written into the Maestro applications guide the instrument to pick up appropriate numbers of tips, and refill/replace tip boxes as needed. INHECO temperature blocks installed on the deck of the Sciclone NGS Workstation allow for appropriate 4 C or heated storage of reagents and controlled incubation temperatures for reactions and wash steps. Reaction mixes are pre-arrayed prior to addition to sample to ensure equal incubation times across the sample plate. Easy-to-follow user interfaces guide the reagent and deck setup process and prompt the user for any necessary interventions. Figure 1. Overview of the Maestro Workflow for the SureSelect XT protocol. 3

4 The Maestro Workflow for the SureSelect XT DNA Sample Preparation Consists of 5 Independent Maestro Applications Run Time (Including Run Time Set-up Setup Thermocycler (including Maestro Application Time thermocycler Maestro Application Time steps) Steps) 1 SureSelectXT Initial SPRI Cleanup SureSelectXT Initial SPRI Cleanup 10 min 10 min 1 hour 1 hour 2 SureSelectXT Library Prep SureSelectXT Library Prep 1 hour 1 hour 6 hours 6 hours 3 SureSelectXT PreCapture Normalization 10 min SureSelectXT PreCapture Normalization 10 min up to 1.5 hours up to 1.5 hours SureSelectXT 4 SureSelectXT Hyb Hyb Setup Setup min min 15 min 15 min SureSelectXT 5 SureSelectXT Target Target Selection Selection min min 7 hours 7 hours In the SureSelectXT Initial SPRI Cleanup, 130 µl sheared DNA samples are split into two aliquots for the AMPure XP bead cleanup to accommodate the necessary sample/bead volumes in standard Hard-Shell PCR plates. After elution in 15 µl each, the split samples are pooled back together for a total sample volume of 30 µl. The SureSelectXT Initial SPRI Cleanup application processes a maximum of 48 samples (6 columns); it is implemented twice if more than 48 samples are being processed. In the SureSelectXT Library Prep procedure, the volume of the End Repair reaction is reduced from the Agilent - recommended volume of 100 µl. Instead, the End Repair reaction is set up in a total volume of 50 µl (30 µl sample plus 20 µl End-Repair Mix) to allow for post End-Repair SPRI Cleanup (with 90 µl AMPure XP beads) in a Bio- Rad 96-well PCR plate. A single aliquot of AMPure XP beads is used per sample for SPRI Cleanups following the End-Repair, A-Tailing, and Ligation steps. The DNA in the sample is driven on and off the beads via changes in PEG and NaCl concentrations. This strategy increases yield by limiting the number of times samples are transferred to new wells/plates. For the library prep, use of Elution Buffer (Qiagen EB), rather than water, is recommended for all pre-pcr elution steps to ensure proper buffering during resuspension. When the post-ligation SPRI Cleanup is complete, the samples are split into two aliquots: 15 µl is added directly to PCR master mix for enrichment and further processing and 15 µl is transferred to a clean PCR plate and stored at -20 C for future use. A Post-PCR SPRI Cleanup step is integrated into the SureSelectXT Library Prep application. This step requires fresh AMPure XP beads and elution in water rather than EB. The full library preparation (End-Repair through Post-PCR SPRI Cleanup) can be completed for up to 96 samples in about 6 hours. The quality, size distribution, and concentration of library in each sample must be checked prior to continuing to the next steps. In the SureSelectXT PreCapture Normalization procedure, samples are diluted with water to a concentration of 18.9 ng/µl in 26.4 µl (total 500 ng DNA). This is undertaken one sample at a time and requires up to 1.5 hours for a 96 sample plate. The Maestro application simplifies sample tracking by importing concentration data directly from a Microsoft Excel spreadsheet and automatically calculating the proper sample and water volumes for each well. The excess volume in the normalized sample is reduced appropriately by evaporation during the high temperature denaturing step of hybridization setup. For SureSelectXT Hyb Setup, hybridization buffer and capture library are mixed by the user at room temperature, then arrayed into a 96-well plate and heated to 65 C on the Sciclone. Samples are mixed with blocking solution, denatured on the deck at 95 C, then transferred to the 65 C plate containing the capture library. The user is required to promptly seal and transfer the plate to a nearby thermocycler for the overnight incubation. In the SureSelectXT Target Selection application, wash steps are carried out in 96-well PCR plates with 150 µl wash buffer per well, rather than the Agilent recommended 200 and 500 µl per sample. Additional wash cycles are used to compensate for reduced wash volumes. At the completion of the wash steps, samples (with streptavidin beads) are resuspended in 50 µl 10 mm Tris ph 8.0. The SureSelect XT Elution Buffer and SureSelect XT Neutralization Buffer are not used. Half the resuspended beads/sample (25 µl) is transferred to a plate containing PCR Master Mix for PCR enrichment followed by a post-pcr SPRI Cleanup. The remaining 25 µl resuspended beads/sample (25 µl) is stored at -20 C for future use. The SureSelectXT Target Selection application, including the PCR and Post-PCR SPRI Cleanup steps, can process up to 96 libraries in 7 hours. 4

5 Flowcharts for Individual Steps of the Maestro Workflow for the SureSelect XT Protocol Figure 2. SureSelectXT Initial SPRI Cleanup Application (Day 1). Figure 3. SureSelectXT Library Prep Application (Day 2). 5

6 Flowcharts for Individual Steps of the Maestro Workflow of the SureSelect XT Workflow Figure 4. SureSelectXT Pre-Capture Normalization Application (Day 3). Figure 5. SureSelectXT Hyb Setup Application (Daay 3). Figure 6. SureSelectXT Target Selection Application (Day 4). 6

7 Required Materials and Reagents Reagents Reagent Sciclone NGS Workstation Accessories Accessory Vendor and Part No. Agilent SureSelect XT Human All Exon Kit or Agilent SureSelectXT Custom Target Enrichment System Kit Agilent Herculase II Fusion DNA Polymerase* Agilent Dynabeads MyOne Streptavidin T1 Nuclease-Free Water Ampure XP Beads (200 rxn) (400 rxn) Invitrogen (2 ml) (10 ml) (100 ml) PerkinElmer Part No. Agencourt 96-ring Magnet (2) CLS Spacer Assembly for Agencourt 96-ring Magnet CLS (qty 4) and CLS (1) INHECO Deepwell plate adapter (Corning 1 ml) CLS INHECO Deepwell plate adapter (12-row Pyramid) CLS INHECO 384-well plate adapter CLS INHECO 96-well adapters (3) CLS INHECO 96-well adapter/shaker CLS Various Beckman Coulter A63880 (5 ml) A63881 (60 ml) A63882 (450 ml) 100% EtOH Sigma E % PEG 8000 Sigma M NaCl Sigma S mm Tris ph 8.0 Various Elution Buffer (EB) Qiagen General Laboratory Equipment and Supplies Equipment Microfuge Vortexer Supplier 200 µl Multichannel pipettor and appropriate barrier tips Various 1000 µl Multichannel pipettor and appropriate barrier tips Various Covaris S2 or E210 System and appropriate tubes Microplate centrifuge Thermocycler** Various Various Various Various Bio-Rad Laboratories (MJ Research) DNA Engine PTC-200, or equivalent MicroAmp Clear Adhesive Plate Seals Applied Biosystems LabChip GX or Agilent Bioanalyzer with appropriate chips and reagents ** The standard sample plates used in this application are fully skirted Bio-Rad Hard-Shell 96-PCR plates. Please check if your thermocycler is compatible with this plate type. If it is not, please contact your PerkinElmer Field Application Scientist to discuss modifications to the application to support semi-skirted PCR plates. Various 7

8 Consumables PerkinElmer No. Used Consumable Description Part No. Vendor and Part No. per Run PCR Plates 96-well PCR Plate, Bio-Rad Hard-Shell, Full Skirt CLS Bio-Rad HSP Boxed Tips Pipette Tip 150 µl, Art, Box, Sterile Racks CLS PerkinElmer 58 (96 samples) Deepwell Plates Deepwell-96 POS, Square 2.0 ml Well, Polypro, CLS Seahorse Bioscience Seahorse Bioscience Reservoir- 12-Column Pyramid Bottom, 290 ml CLS Seahorse Bioscience Deepwell Seahorse Bioscience Lids 946 Lid Universal, Robotic Friendly, Polystyrene CLS Seahorse Bioscience well Plates Microplate 384-well, Round Bottom, CLS Corning, Inc Polypropylene (Pkg. 10) Deepwell Plates Microplate-96 well, 1 ml Well Volume, Polypropylene CLS Corning, Inc (1 ml) Corning Running the Maestro SureSelect XT Workflow Please read and familiarize yourself with all steps described in this section prior to beginning the run. A 4-day process is suggested for running the full set of Maestro SureSelect XT applications, but the process may be modified according to the laboratory schedule. Approximate times for setting up and running each application are indicated in the chart on Page 4. Be sure to plan properly for the 24 hour hybridization step to allow enough time to complete the subsequent Target Selection application during the following working day. SureSelectXT Initial SPRI Cleanup application (Day 1) This application serves to concentrate the fragmented DNA in order to reduce the volume of sample presented for the library prep application. Because each sample is split into two aliquots, the maximum sample number for each run of the SureSelectXT Initial SPRI Cleanup application is 48 samples. All subsequent applications in the workflow can process up to 96 samples. Sample preparation Genomic DNA should be fragmented to an average size of bp on a Covaris S2 or E210 instrument as described in the Agilent SureSelect XT Target Enrichment System for Illumina Paired-Ended Sequencing Library with Multiplexing Protocol v1.2 May Samples should be presented for the Maestro SureSelectXT Initial SPRI Cleanup run as up to 3 µg sheared genomic DNA in 130 µl TE in a Bio-Rad Hard-Shell 96-well PCR plate. Note: AMPure XP beads should be warmed at room temperature for about 30 min before use. They may be taken out of 4 C storage before beginning. If necessary, boot up the system by first starting the Sciclone NGS Workstation and the INHECO units, then starting the PC controller. 1. Modify the SS Initial SPRI worksheet in the SureSelectXT Workbook to specify the number of samples to run The workbook must be located in the filepath: C:\Program Data\CaliperLS\Maestro\Workbooks and must have the name SureSelectXT Workbook.xls. If the file is moved or the name is changed, Maestro will not be able to find the information necessary to begin the run. 8

9 The sample number must be set by indicating the number of columns to run in the worksheet titled SS Initial SPRI. The Maestro application only processes full columns of 8 samples each. 3. Prepare sample plate Up to 48 samples may be processed (as sets of 8 samples per column) in the SureSelectXT Initial SPRI application. Transfer 130 µl sheared DNA sample (containing up to 3 µg) to a Bio-Rad 96-well PCR plate. Fill columns 1-6 in order from left to right. Inspect the plate to ensure that air has not been trapped in the wells. If necessary, spin briefly to bring liquid to the bottom of the wells. If >48 samples are to be processed, prepare a second plate containing up to 48 additional samples. After processing the first set of 48 samples, all steps of the SureSelectXT Initial SPRI application must be repeated with the second plate. 4. Prepare reagents for the run Figure 7. SS Initial SPRI worksheet. After modifying the entry for the number of columns to process, the spreadsheet will update the appropriate volumes/ wells to fill on the reagent plates. Save the modified spreadsheet with its original name in its original file path. Print the SS Initial SPRI spreadsheet to use as a guide while setting up reagents. 2. Start the Maestro SureSelectXT Initial SPRI Cleanup run Launch the Maestro software and open the SureSelectXT Initial SPRI Cleanup Application. Ensure that the INHECO units have the correct adapters for the run: Position A3 Position A4 Position D2 Position D4 96-well Plate Adapter 96-well Plate Adapter 96-well Plate Adapter 96-well Plate Shaker Adapter Start the run by selecting the play button. If running in Edit mode, be sure to start the Main Method. Verify that the INHECO units for position A4 is set to 4 C and is cooling. Verify that the INHECO units for positions A3, D2, and D4 are set to 22 C. Note: when the run is started, the instrument will complete all initialization steps for the hardware and the specific application. The run will automatically pause and prompt the user to set up the deck and confirm proper setup prior to beginning the SPRI Cleanup steps. Ensure AMPure XP beads are at room temperature and thoroughly resuspended in storage solution. Using a multichannel pipettor, aliquot 90 µl AMPure XP beads per well into a Bio-Rad Hard-Shell PCR plate as shown in the SS Initial SPRI spreadsheet. Inspect the plate to ensure that air has not been trapped in the wells. If necessary, spin gently to bring liquid to the bottom of the wells. Store the bead plate at room temperature. Note: Two columns of AMPure XP beads will be used for each column of samples to be processed in the run. Make 50 ml fresh 80% EtOH solution by diluting 40 ml 100% Ethanol with 10 ml nuclease-free molecular biology grade water. Pour into a Seahorse Deepwell reservoir, cover with lid, and store at room temperature. Pour 20 ml Qiagen Elution Buffer (EB), or 10 mm Tris ph 8.0 into a Seahorse Deepwell reservoir and store at room temperature. If running a limited number of samples, a 12-column Seahorse Deepwell reservoir may be used instead. For every column to be run, fill 2 columns of the reservoir with 3 ml EB. 5. Set up the Sciclone NGS Workstation deck Confirm that the Maestro Sciclone NGS Workstation software has correctly read the workbook and is set to run the correct number of columns. Step through the pictures, placing the indicated consumables/prepared plates in the indicated locations. Take care to note whether or not a lid is needed at each location. Place new tip boxes in the indicated locations. 9

10 6. Run the SureSelectXT Initial SPRI Cleanup Steps Confirm that the deck setup matches the final picture in the setup window. Selecting Finished will prompt the application to begin the run. The Maestro SureSelectXT Initial SPRI application will automatically proceed through the steps indicated in the flowchart on Page 5 and Appendix A. While the application is running, the green light at the top of the instrument will blink. If there is a problem with the run, the light will change to yellow and an alarm will sound to indicate that user intervention is necessary. The application may be set up so that an is sent if an error occurs or when additional tip boxes are needed on the deck. After the final elution of sample from beads, the split samples will be re-pooled into single wells with a total volume of 30 µl. If properly sheared and size selected with the Initial SPRI application, the DNA samples will contain a majority of fragments in the bp (±10%) range. Note that the Initial SPRI Step will select against fragments smaller than 100 bp, so it is likely that the profile of the smear will shift as shown below. 7. Application complete When the Initial SPRI is complete, the application will pause and show a message indicating that the sample plate should be removed from the deck. The samples may be stored at -20 C, checked for quality/concentration, or passed immediately into Library Preparation steps. If processing > 48 samples through the SureSelect XT procedure, repeat steps 1-7 with the second plate of samples. Quality control of Initial SPRI samples (optional) For high throughput projects, a LabChip GX Automated Electrophoresis System is recommended for assessing sample quality and concentration. Samples, diluted into water in a 96-well PCR plate, may be analyzed sequentially by the LabChip GX System in a single run at a rate of about 1 sample per minute. To set up the LabChip GX System, follow the instructions in the LabChip GX System user guides, making sure to allow 30 minutes for the chip and reagents to equilibrate to room temperature prior to preparing the chip. DNA 1K assay: dilute sample 1:10 in a Bio-Rad 96-well PCR plate. DNA High Sensitivity assay: dilute sample 1:25 in a Bio-Rad 96-well PCR plate. Ensure the diluted sample is well mixed by pipetting up and down at least 3 times. Centrifuge the plate at 1000 xg for 1 min to eliminate bubbles and pellet any beads prior to beginning the LabChip GX System run. For lower throughput projects, the Agilent Bioanalyzer can be used in place of the LabChip GX System. Use the Agilent Bioanalyzer DNA 1000 kit for sheared DNA QC. Allow 30 minutes for reagents to equilibrate to room temperature, and prepare the chip according to the Agilent protocol. Figure 8. Representative data tracings showing LabChipGX analysis of 1:25 dilutions of a sheared total human DNA sample before (blue) and after (red) the Initial SPRI Cleanup Step. SureSelectXT Library Prep Application (Day 2) Please read and familiarize yourself with all steps described in this section prior to beginning the run. The Library Preparation application includes a thermocycler step and a post-pcr SPRI Cleanup. It is recommended to run the entire application prior to storing samples. If time allows, quantification of the library may be carried out on the same day as the library prep. Sample preparation Genomic DNA should be fragmented to an average size of bp on a Covaris S2 or E210 instrument as described in the Agilent SureSelect XT Target Enrichment System for Illumina Paired-Ended Sequencing libraries with multiplexing v1.2 May Samples should be purified and concentrated with the Maestro SureSelectXT Initial SPRI Cleanup application (or manual cleanup according to the Agilent protocol) and presented in 30 µl EB Buffer or 10 mm Tris ph 8.0 in a Bio-Rad 96-well PCR plate for the Maestro SureSelectXT Library Prep run. Note: AMPure XP beads should be warmed at room temperature for about 30 min before use. They may be taken out of 4 C storage before beginning. Do not thaw SureSelect XT reagents until steps 1 and 2 have been completed. If necessary, boot up the system by first starting the Sciclone NGS Workstation and the INHECO units, then starting the PC controller. 10

11 1. Modify the SS XT Lib Prep spreadsheet in the SureSelectXT Workbook to specify the number of samples to run The workbook must be located in the filepath: C:\ProgramData\CaliperLS\Maestro\Workbooks and must have the name SureSelectXT Workbook. If the file is moved or the name is changed, Maestro will not be able to find the information necessary to begin the run. The sample number must be set by indicating the number of columns to run in the worksheet titled SS XT Lib Prep. The Maestro application only processes full columns of 8 samples each. Figure 9. SS XT Lib Prep worksheet. After modifying entries for the number of columns to process, the spreadsheet will update the recipes for the reagent mixes and the appropriate volumes for the reagent plates. Save the modified spreadsheet with its original name in its original file path. Print the SS XT Lib Prep spreadsheet to use as a guide while setting up reagents. 2. Start the Maestro SureSelectXT Library Prep run Launch the Maestro software and open the SureSelectXT Library Prep Application. Ensure that the INHECO units have the correct adapters for the run: Position A3 Position A4 Position D2 Position D4 96-well PCR Plate Adapter 384-well Plate Adapter 96-well PCR Plate Adapter 96-well PCR Plate Shaker Adapter Start the run by selecting the play button. If running in Edit mode, be sure to start the Main Method. Verify that the INHECO units for positions A3 and A4 are set to 4 C and are cooling. Verify that the INHECO units for positions D2 is set to 22 C and D4 is set to 20 C. Note: when the run is started, the instrument will complete all initialization steps for the hardware and the specific application. The run will automatically pause and prompt the user to set up the deck and confirm proper setup prior to beginning the library preparation steps. Starting the application prior to thawing and diluting reagents ensures that the cold blocks are pre-chilled and ready for on-deck reagent storage. 3. Thaw the SureSelect XT library preparation reagents and place on ice Using the SS XT Lib Prep spreadsheet as a guide, thaw the reagents from the SureSelect XT Library Preparation Kit necessary to make the End Repair Mix, A-Tailing Mix, and Adapter Ligation Mix. Do not thaw the reagents for the PCR Master Mix at this time. Note: care should be taken to avoid freeze-thaw cycles with all SureSelect XT reagents. Reagents should be aliquoted appropriately if planning to use a kit for more than 2 independent runs. 4. Prepare the PEG, Elution Buffer, AMPure XP bead plates, and 80% EtOH reservoir Make fresh 20% PEG/2.5 M NaCl solution according to the recipe in the SS XT Library Prep spreadsheet. Using a multichannel pipettor, aliquot 190 µl 20% PEG/2.5 M NaCl solution per well into a Bio-Rad 96-well PCR plate for each column of samples to be run. Cover the plate with a lid and store at room temperature. The Elution Buffer (EB Buffer from Qiagen ), should be placed in a Seahorse Deepwell 12-column reservoir. Starting with column 1, add 3 ml buffer for each column of samples to be run. For higher throughput runs, an open Seahorse Deepwell reservoir with 50 ml Elution Buffer may be used instead. Note: water is not recommended for elution during the library preparation steps, as the lack of buffering can interfere with proper performance. Water is only used as elution buffer in the Post-PCR SPRI Cleanup just prior to sample normalization and hyb setup. Maestro will prompt the user to exchange the Elution Buffer for water prior to starting the Post-PCR SPRI Cleanup step. 11

12 Thoroughly resuspend AMPure XP beads (warmed to room temperature) by inverting/rotating the bottle. Using a multichannel pipettor, aliquot 95 µl AMPure XP beads per well into a Bio-Rad 96-well PCR plate for each column of samples to be run (see SS XT Library Prep spreadsheet). Store at room temperature. Note: a second plate of AMPure XP beads, with 90 µl beads per well, will be necessary for the Post-PCR SPRI Cleanup step at the end of Library Prep. AMPure XP beads for this plate may be left at room temperature at this time and aliquoted into a Bio-Rad 96-well PCR plate during the amplification step. Make 100 ml fresh 80% EtOH solution by diluting 80 ml 100% Ethanol with 20 ml nuclease-free molecular biology grade water. Pour into a Seahorse Deepwell reservoir, cover with a lid and store at room temperature. 5. Aliquot the adapter mixes Transfer the appropriate volumes of adapter mix into the Adapter Oligo Plate as indicated in the SS XT Library Prep spreadsheet. Cover the plate with a lid and store on ice or at 4 C. 6. Make the reaction mixes and aliquot into the Master Mix Plate The following mixes should be prepared at this step according to the recipes on the SS XT Library Prep spreadsheet: End-Repair Mix, A-Tailing Mix, Ligation Mix. Care should be taken to pipet accurately, as the reaction mixes have minimal overage volumes. Keep the reaction mixes on ice. Note: to avoid loss of polymerase activity, the PCR Mix is not prepared at this time. The application will prompt for placement of the PCR mix on the deck at the appropriate time during the run. Prepare the Master Mix plate as specified on the spreadsheet. Keep the plate on ice, pipette carefully into the bottom of the wells, and avoid trapping air or creating bubbles. Cover the plate with a lid and store on ice or at 4 C. 7. Set Up the Sciclone NGS Workstation Deck Inspect all prepared reagent plates to ensure that air has not been trapped in the wells. If necessary, spin the plates briefly to bring reagents to the bottom of the wells. Confirm that the Maestro software has correctly read the workbook and is set to run the correct number of columns. Step through the pictures, placing the indicated consumables/prepared plates in the indicated locations. Take care to note whether or not a lid is needed at each location. Place new tip boxes in the indicated locations. Note: when the SureSelectXT Library Prep application is started, the variables used for tip tracking are reset. The run must be started with new, full tip boxes in the indicated positions, as Maestro will not retain tip tracking information from the previous run. 8. Run the Library Preparation Steps Confirm that the deck setup matches the final picture in the setup window. Selecting Finished will prompt the application to begin the library preparation steps. While the application is running, the green light at the top of the instrument will blink. If there is a problem with the run, the light will change to yellow and an alarm will sound to indicate that user intervention is necessary. The application may be set up so that an is sent if an error occurs or when additional tip boxes are needed on the deck. The Maestro SureSelectXT Library Prep application will automatically proceed through End Repair, A-tailing, Ligation, and PCR setup steps as indicated in the flowchart on Page 5 and Appendix B. 9. Prepare the PCR Master Mix During the Post-Ligation SPRI Cleanup (Step 4), the PCR Master Mix should be prepared according to the recipe in the SS XT Lib Prep spreadsheet. Thaw the reagents, prepare the master mix, and store it on ice or at 4 C. At the start of Step 5 PCR Setup, the application will prompt the user to add the PCR Master Mix to the reagent plate at A3. The Sciclone NGS Workstation will then distribute the Master Mix from the plate at A3 to the appropriate wells of a clean plate for the PCR reactions. At the completion of the post-ligation SPRI Cleanup, the sample is eluted in 30 µl total volume. 15 µl is transferred to a clean plate for storage and 15 µl is transferred to a plate containing PCR mix. The stored sample may be used in the event that later steps in the SureSelect XT workflow need to be repeated. 10. Amplify Adapter-Ligated Library Note: the PCR reaction will be setup in a Bio-Rad Hard-Shell skirted 96-well PCR plate. Please ensure that your thermocycler is compatible with this plate type. If it is not, the application will need to be modified to transfer samples to a thermocycler-compatible plate for the amplification step and then to transfer amplified libraries back to a Bio-Rad Hard-Shell skirted 96-well PCR plate for post-pcr purification. When the PCR setup is complete, the application will pause and prompt the user to remove the plate containing the PCR reactions, seal, spin, and transfer it to a thermocycler for 12

13 library amplification. The plate containing excess sample is sealed and stored at -20 C. The SureSelectXT Library Prep application will remain paused until the user indicates that the plate containing the completed PCR reactions has been spun, unsealed, and returned to the deck of the Sciclone NGS Workstation. Agilent recommends the following program for amplification of libraries prepared from 3 µg input DNA. The number of cycles may need to be modified according to the amount of input DNA in the sample. Use a heated lid to prevent condensation. 98 C for 2 minutes 4 cycles of: 98 C for 30 seconds 65 C for 30 seconds 72 C for 1 minute 72 C for 10 minutes Hold at 4 C While the samples are amplifying, prepare the reagents for Post-PCR SPRI Cleanup. Ensure that AMPure XP beads are at room temperature and are well resuspended. Aliquot 90 µl per well to a clean Bio-Rad 96-well PCR plate (see SS XT Lib Prep spreadsheet). Prepare a new Deepwell Seahorse reservoir to replace the reservoir containing Elution Buffer. Either an open reservoir containing 50 ml of water or a 12-column reservoir containing 3 ml per column (one column for each column of samples in the run) may be used. 11. Purify amplified libraries Because amplification is minimal at this stage, Agilent does not find it necessary to recommend segregation of this SPRI Cleanup to a specified Post-PCR area. Therefore, the Post- PCR SPRI Cleanup step is integrated into the SureSelectXT Library Prep application. The application will prompt the user to place the new AMPure XP bead plate at location D4, the reservoir containing water at position D3, and the plate containing the amplified library at position B4. Once the user has indicated that the samples are back on the deck, the run will resume. Note: it is critical that Nuclease-Free water is used in place of Elution Buffer during this cleanup step. Samples will be evaporated prior to hybridization, and it is important not to concentrate any buffer/salts during the evaporation. 12. Application complete When the Post-PCR SPRI is complete and the library has been eluted in 30 µl water, the application will pause and show a message indicating that the sample plate should be removed from the deck. Click through the dialogs in Maestro to shut down the INHECO units and complete the run. The user may proceed directly to QC or store libraries at -20 C. Quality control and quantification of amplified libraries It is necessary to quantify and normalize amplified DNA libraries to 500 ng per sample prior to hybridization. For a high throughput project, a LabChip GX System is recommended for quality control steps. Either the DNA 1K Assay Kit or DNA High Sensitivity Assay Kit may be used. To set up the LabChip GX System, follow the instructions in the LabChip GX user guides, making sure to allow 30 minutes for the chip and reagents to equilibrate to room temperature. For DNA 1K assay, dilute the sample 1:10 in a Bio-Rad 96-well plate. For DNA High Sensitivity assay, dilute the sample 1:25 in a Bio-Rad 96-well plate. Mix the diluted sample by pipeting up and down at least 3 times. Spin the plate at 1000 xg for 1 minute to eliminate any bubbles and pellet any beads. Load the plate into the LabChip GX System and run the appropriate assay. For lower numbers of samples, an Agilent Bioanalyzer can be used in place of the LabChip GX System. Use the Agilent Bioanalyzer DNA 1000 kit for library quantitation. It is also possible to quantify the library using a fluorescence-based assay such as Pico Green or Qubit. Absorbance readings are not recommended for library quantification, as they tend to over-represent the concentration of DNA in the library. Quantification data, representing the concentration in ng/ul of each undiluted library sample, should be saved in an Microsoft Excel -compatible format for import into the SureSelectXT Pre-Capture Normalization worksheet. Expected results The total amount of DNA will vary depending on amount of input DNA, how the input DNA was quantified, and quality of the input DNA. Generally, yields of 1-2 µg of library with maximum concentration at bp are expected. 500 ng is required to proceed to hybridization steps. Figure 10. Representative data tracing from LabChipGX analysis of a 1:25 dilution of adapterligated library after 4 cycles of amplification and post-pcr SPRI cleanup. 13

14 1. Modify the Normalization spreadsheet in the SureSelectXT Workbook to specify the number of samples to run and the concentration of each sample. Figure 11. Gel image from LabChipGX analysis of 1:25 dilutions of 16 replicate libraries prepared from 1000 ng total human DNA. Libraries were amplified with 4 cycles of PCR. The workbook must be located in the filepath: C:\ ProgramData\CaliperLS\Maestro\Workbooks and must have the name SureSelectXT Workbook. If the file is moved or the name is changed, Maestro will not be able to find the information necessary to begin the run. a. Specify the number of samples to be normalized by setting the number of columns to run in cell D2 (the application only processes full columns of 8 samples each). b. Ensure the value for Total Volume (µl) entered in cell D6 is c. Fill in the Sample, Source Well and Conc. (ng/µl) columns with information specific for the samples to be run. d. Save the modified spreadsheet with its original name in its original file path. Figure 12. Total yield of amplified libraries as calculated from LabChipGX data shown in figure 10. SureSelectXT Pre-Capture Normalization Application (Day 3) Please read and familiarize yourself with all steps described in this section prior to beginning the run. Libraries are normalized one at a time (by transfer to a clean plate and dilution with water) to a concentration of 18.9 ng/µl in 26.4 µl for a total of 500 ng DNA per library. Evaporation during the denaturing step of hybridization setup will further concentrate the sample prior to the hybridization incubation. Normalization will take up to 90 minutes for 96 samples. Sample preparation Library samples prepared with Agilent SureSelect XT adapters and PCR amplification through 4-6 cycles with the SureSelect XT Primer (forward) and SureSelect XT Indexing Pre-Capture Primer (reverse) must be used. After post-pcr purification, the concentration of each amplified library must be determined. Analyzing libraries using a LabChip GX System or Agilent 2100 Bioanalyzer is recommended in order to visualize the size distribution of the library fragments and quantify the DNA in the appropriate size range. If necessary, boot up the system by first starting the Sciclone NGS Workstation and the INHECO units, then starting the PC controller. Figure 13. Normalization workbook. The Volume for 500 ng value will be calculated by Microsoft Excel based on the value for Conc. (ng/µl). The Source well location will also be the destination well location in the normalized sample plate. Columns with green headers provide additional information on the samples, and will be displayed by Maestro when validating the workbook. They may be deleted if desired. Do not delete any columns highlighted in yellow. 14

15 2. Start the Maestro SureSelectXT Pre-Capture Normalization run Launch the Maestro software and open the SureSelectXT Pre-Capture Normalization Application. Ensure that the INHECO units have the correct adapters for the run: Position A3 Position A4 Position D2 Position D4 96-well PCR Plate Adapter 96-well PCR Plate Adapter 96-well PCR Plate Adapter 96-well PCR Plate Shaker Adapter Start the run by selecting the play button. If running in Edit mode, be sure to start the Main Method. Verify that the INHECO units for positions A3 and A4 are set to 4 C and are cooling. Verify that the INHECO units for positions D2 and D4 are set to 22 C. 3. Set Up the Sciclone NGS Workstation Deck Confirm that the software has correctly read the normalization worksheet, is set to run the correct number of samples, and is displaying data specific for the samples to be normalized. Step through the pictures, placing the indicated consumables/tip boxes/prepared plates in the indicated locations. Take care to note whether or not a lid is needed at each location. Note: the buffer reservoir for sample dilution must contain water, as no additional salts should be added to the library samples prior to the hybridization setup. Note: when the SureSelectXT Pre-Capture Normalization application is started, the variables used for tip tracking are reset. The run must be started with a new, full tip box at C3, as Maestro will not retain tip tracking information from the library prep run. An empty tip box must be placed at position C5, as shown in the setup pictures. 4. Run the SureSelectXT Pre-Capture Normalization application The head of the Sciclone NGS Workstation will transfer a single column of tips to the empty tip box at position C5, then load tips one at a time on mandrel A1. The tip will first aspirate the appropriate volume of water, then the appropriate volume of sample. Both water and sample will be dispensed to the clean plate at position A4, and the tip will be ejected. The application will continue through the total number of samples, taking about 1 minute to process each sample. 5. Application complete When prompted to do so, seal the original library plate (position A3) for long-term storage at -20 C. The normalized plate (position A4) may be sealed and stored at -20 C or may be used immediately for hybridization setup. Click through the dialogs in Maestro to shut down the INHECO units and complete the run. SureSelectXT Hyb Setup Application (Day 3) Please read and familiarize yourself with all steps described in this section prior to beginning the run. The minimum incubation time for SureSelect XT Hybridization is 24 hours. The subsequent washing, PCR amplification and cleanup steps require 6 hours to complete. If planning to run the target capture steps the day following hybridization setup, it is best to complete the SureSelectXT Hyb Setup Application no later than 11:00 am to allow sufficient time during normal working hours to complete the target capture. The SureSelectXT Hyb Setup Application involves adding block mix to the libraries and denaturing at > 95 C on the deck. Water is expected to evaporate from the samples during this incubation. The denatured and concentrated library/block mix is then added to a plate held at > 65 C containing capture probes in hybridization buffer. The user must immediately seal the plate and transfer it to a thermocycler for extended incubation at 65 C. Actual run time for the application is about 15 minutes. The user is advised to attend the entire run to ensure that no delays occur during the hybridization setup steps and subsequent transfer to thermocycler. Note: it is important that the thermocycler be located in immediate proximity to the Sciclone. The hybridization reactions must not be allowed to cool below 65 C while transferring from the Sciclone deck to the thermocycler. Prior to running SureSelectXT Hyb Setup Application for the first time, it is advised to check thermocycler incubation conditions with appropriate volumes of water to ensure that evaporation is not excessive. Please consult the Agilent SureSelect XT protocol for information on sealing plates and testing your specific incubation conditions for the 24 hr (or longer) hybridization. 15

16 Sample preparation Samples must be presented for hybridization setup in a Bio-Rad 96-well Hard-Shell PCR plate as 500 ng of sample in a total volume of 26.4 µl of water. If necessary, boot up the system by first starting the Sciclone NGS Workstation and the INHECO units, then starting the PC controller. 1. Modify the SS Hyb Setup spreadsheet in the SureSelectXT Workbook to specify the number of samples to run. The workbook must be located in the filepath: C:\Program Files\CaliperLS\Maestro\Workbooks and must have the name SureSelectXT Workbook. If the file is moved or the name is changed, Maestro will not be able to find the information necessary to begin the run. The sample number must be set by indicating the number of columns to run in the worksheet titled SS Hyb Setup. The Maestro application only processes full columns of 8 samples each. Start the run by selecting the play button. If running in Edit mode, be sure to start the Main Method. Verify that the INHECO unit for position A3 is set to 105 C, and D4 is set to 75 C. Verify that the INHECO units for positions D2 and A4 are set to 22 C. A timer will appear to prevent the user from continuing with the run before the INHECO units at A3 and D4 have equilibrated for a full 30 minutes. 3. Thaw the SureSelect XT hybridization reagents and place on ice The following reagents need to be thawed at this time: Indexing Block #1, Indexing Block #2, Indexing Block #3, Hyb#3, RNase Block, SureSelect Capture Library. Additional reagents: Hyb #1, Hyb #2, Hyb #4 are stored at room temperature. Note: the SureSelect XT Capture library is composed of RNA baits. To prevent degradation, be sure to handle the baits and all hybridization reagents under RNAse-free conditions. Upon first use of the capture library, aliquot the library appropriately to avoid multiple freeze-thaw cycles for subsequent runs. 4. Prepare the thermocycler The thermocycler must be located in immediate proximity to the Sciclone NGS Workstation. Turn on the thermocycler and set it to hold temperature indefinitely at 65 C with a heated lid set to > 75 C. 5. Prepare the hybridization reagents Figure 14. SS Hyb Setup worksheet. After modifying entries for the number of columns to process, the spreadsheet will update the recipes for the reagent mixes and the appropriate volumes for the reagent plates. Save the modified spreadsheet with its original name in its original file path. Print the SS Hyb Setup spreadsheet to use as a guide while setting up reagents. 2. Start the Maestro SureSelectXT Hyb Setup run Launch the Maestro software and open the SureSelectXT Hyb Setup Application. Ensure that the INHECO units have the correct adapters for the run: Position A3 Position A4 Position D2 Position D4 96-well PCR Plate Adapter 96-well PCR Plate Adapter 96-well PCR Plate Adapter 96-well PCR Plate Shaker Adapter Using the SS Hyb Setup spreadsheet as a guide, prepare the appropriate volumes of Block Mix and Hybridization Buffer in RNase-Free microtubes. Mix by inversion and spin briefly. If a precipitate forms in the hybridization buffer mix, warm the buffer to 65 C for 5 minutes then allow it to cool to room temperature. Store the reagent mixes at room temperature. In a RNase-Free 1.5 ml tube, dilute an aliquot of RNase block 1:3 with Nuclease-free water to reach the volume required for the Capture Probe Library mix. Mix by inversion or gentle vortexing and spin briefly. Keep the diluted RNase block at room temperature. In a separate RNase-Free 1.5 ml tube, mix the indicated volumes of diluted RNase block and capture library. Allow this tube with the Baits/RNase dilution to equilibrate to room temperature. Return the stock of capture library to -80 C storage. 16

17 Once the hybridization buffer and Baits/RNase mixture are at room temperature, add the indicated volume of hybridization buffer to the tube containing the Baits/RNase mixture. Mix gently by inversion and spin briefly. Keep at room temperature. Aliquot the block mix and the capture library in hyb buffer to the 96-well PCR plate as specified on the worksheet. Note: the worksheet only accounts for dead volume in the PCR plate, but, to conserve capture library, no additional volume is added for pipetting error. To avoid shortage of reagents, transfer 0.3 µl less than the indicated aliquot amount. 6. Set Up the Sciclone NGS Workstation Deck Confirm that the software has correctly read the workbook and is set to run the correct number of columns. Step through the pictures, placing the indicated consumables/ prepared plates in the indicated locations. Place new tip boxes in the indicated locations. Note: when the SureSelectXT Hyb Setup application is started, the variables used for tip tracking are reset. The run must be started with new, full tip boxes in the indicated positions, as Maestro will not retain tip tracking information from the previous run. 7. Run the SureSelectXT Hyb Setup application Note: the user is advised to attend the entire run to ensure that no delays occur during the hybridization setup steps and subsequent transfer to thermocycler. Block mix (5.6 µl) will be aliquoted from the plate at C4 into the normalized library plate at B4. After the blockers are aliquoted, the library plate at B4 will be moved to location A3 for denaturation. It is critical that the samples are denatured for exactly 5 minutes. To ensure that the incubation is completed correctly, it is recommend to manually run a timer for 5 minutes starting when the sample plate is moved to position A3. During the incubation, 20 µl of Capture Library/Hybridization Buffer mix will be arrayed to the appropriate number of columns in a clean PCR plate at D4. When the 5 minute incubation is complete, the sample plate at A3 will be moved back to B4 and the denatured libraries will be transferred from B4 to the Capture Library/Hybridization buffer mix at D4 (75 C). After mixing denatured library with capture library at D4, the head of the Sciclone NGS Workstation will move out of the way and the user will be prompted to seal the plate while on the deck at D4. For successful hybridization, the plate must be double-sealed IMMEDIATELY after mixing and BEFORE removing the plate from the 65 C INHECO block at D4. When completely sealed, quickly transfer the plate to the pre-heated thermocycler. Note the time. Samples should hybridize for at least 24 hours, but can incubate up to 72 hours if necessary. 8. Application complete Click through the dialogs in Maestro to shut down the INHECO units and complete the run. SureSelectXT Target Selection Application (Day 4) Please read and familiarize yourself with all steps described in this section prior to beginning the run. This application will take 6-7 hours and should be started 1 hour before the end of the hybridization incubation. The first step in the Target Selection application involves preparing the MyOne Streptavidin T1 beads for target capture. When the bead washing step is complete, the application will prompt the user to transfer the hybridized samples directly from the thermocycler to the deck of the Sciclone NGS Workstation. The samples will be immediately transferred onto the streptavidin beads. Sample preparation Samples must be hybridized for 24 hours in a 96-well PCR plate incubating at 65 C in a thermocycler. The samples will be transferred from the thermocycler to the deck of the Sciclone NGS Workstation when prompted by the application at the start of Step 2. If necessary, boot up the system by first starting the Sciclone NGS Workstation and the INHECO units, then starting the PC controller. 1. Modify the SS Target Selection spreadsheet in the SureSelectXT Workbook to specify the number of samples to run The workbook must be located in the filepath: C:\Program Data\CaliperLS\Maestro\Workbooks and must have the name SureSelectXT Workbook. If the file is moved or the name is changed, Maestro will not be able to find the information necessary to begin the run. 17

18 4. Prepare the streptavidin beads and resuspension buffer plate Thoroughly resuspend the MyOne Streptavidin T1 beads by rotating the bottle. Using the SS Target Selection spreadsheet as a guide, aliquot the specified volumes of streptavidin beads into Column 1 of a 96-well Seahorse 2 ml Deepwell plate. Figure 15. SS Target Selection worksheet. The sample number must be set by indicating the number of columns to run in the spreadsheet titled SS Target Selection. The Maestro application only processes full columns of 8 samples each. After modifying entries for the number of columns to process, the spreadsheet will update the recipes for the reagent mixes and the appropriate volumes for the reagent plates. Save the modified spreadsheet with its original name in its original file path. Print the SS Hyb Setup spreadsheet to use as a guide while setting up reagents. 2. Start the Maestro SureSelectXT Target Selection run Launch the Maestro software and open the SureSelectXT Target Selection Application. Start the run by selecting the play button. If running in Edit mode, be sure to start the Main Method. Ensure that the INHECO units have the correct adapters for the run: Position A3 Position A4 Position D2 Position D4 96-Deepwell Corning 3959 Adapter 12-Row Pyramid Reservoir Adapter 96-well PCR Plate Adapter 96-well PCR Plate Shaker Adapter Verify that the INHECO units for positions D2 and A3 are set to 65 C. Verify that the INHECO units for position D4 is set to 22 C and A4 is set to 4 C. 3. Prepare the Binding Buffer, Wash Buffer I, and Wash Buffer II plates Using the SS Target Selection spreadsheet as a guide, aliquot 775 µl of Binding Buffer and 350 µl of Wash Buffer I into the indicated columns of separate 96-well Seahorse 2 ml Deepwell plates. Aliquot 1000 µl of Wash Buffer II to the indicated columns of a 1 ml Costar 96 Deepwell plate. Cover the plates with lids, inspect for bubbles, spin if necessary, and store at room temperature. Aliquot 10 mm Tris ph 8.0 to Column 6 of the same 96-well Seahorse 2 ml Deepwell plate. Do not prepare the PCR Master Mix, PCR Master Mix plate, or Indexing Primers plate at this time. The PCR Master Mix and Indexing Primer dilutions should be made and aliquoted during Step 4 (Wash Buffer II) of the application. The application will prompt the user to place the PCR Master Mix plate and the Indexing Primers plate on the deck at the start of Step 5 (PCR Setup) Do not aliquot the AMPure XP beads to this plate at this time. Provided that the Post-PCR SPRI Cleanup will be performed on the same instrument, the application will prompt for the addition of AMPure XP beads to Column 4 of the Seahorse 2 ml Deepwell plate when the PCR setup has been completed. 5. Set Up The Sciclone NGS Workstation Deck Confirm that the software has correctly read the workbook and is set to run the correct number of columns. Step through the pictures, placing the indicated consumables/ prepared plates in the indicated locations. Take care to note whether or not a lid is needed at each location. Place new tip boxes in the indicated locations. Note: when the SureSelectXT Target Selection application is started, the variables used for tip tracking are reset. The run must be started with new, full tip boxes in the indicated positions, as Maestro will not retain tip tracking information from the previous run. 6. Run the Maestro SureSelectXT Target Selection application Confirm that the deck setup matches the final picture in the setup window. Selecting Finished will prompt the application to begin the target selection steps. While the application is running, the green light at the top of the instrument will blink. If there is a problem with the run, the light will change to yellow and an alarm will sound to indicate that user intervention is necessary. The application may be set up so that an is sent if an error occurs or when additional tip boxes are needed on the deck. 18

19 The Maestro SureSelectXT Target Selection application will automatically proceed through the streptavidin bead preparation. The user will be prompted to place the hybridized samples on the deck when the instrument has finished preparing the beads. Once the user has indicated that the samples have been added to the deck, the application will continue through the DNA Capture, Wash Buffer I and Wash Buffer II steps. At the completion of washing, the streptavidin beads will be resuspended in 10 mm Tris ph 8.0, and the user will be prompted to place the PCR Master Mix and Indexing Primers plates on the deck for PCR Setup. 7. PCR setup During Step 4 (Wash Buffer II), prepare the PCR Master Mix on ice according to the recipe on the SS Target Selection spreadsheet. Aliquot the indicated volume of PCR Master Mix to the indicated wells of the 96-well PCR Master Mix plate. Keep the plate on ice, pipette carefully into the bottom of the wells, and avoid trapping air or creating bubbles. Cover the plate with a lid and store on ice or at 4 C. Note: if more than 6 columns of samples are run, PCR Master Mix is added to column 7 of the plate, in addition to column 1. Dilute appropriate volumes of the desired set of SureSelect XT Indexing Primers 1:5 with nuclease-free water. Aliquot 5 µl per well of the appropriate diluted indexing primer to each well of the Indexing Primers Plate. One well should be filled for each sample to be run. Pipette carefully into the bottom of the wells, and avoid trapping air or creating bubbles. Cover the plate with a lid and store on ice or at 4 C. Note: the SureSelectXT Target Selection application is written for pre-arrayed indexing primers. The Indexing spreadsheet in the SureSelectXT Workbook may be used to record which indexes are to be used with each sample, but it is not used directly by the Maestro application. The user must create the Indexing Primers Plate with the appropriate indexing primer in each well. Inspect the PCR Master Mix Plate and the Indexing Primers Plate to ensure that air has not been trapped in the wells. If necessary, spin the plates briefly to bring reagents to the bottom of the wells. Place the plates in the specified locations on the deck of the Sciclone NGS Workstation when prompted to do so. Continue with the PCR Setup step. The Sciclone NGS Workstation will array the PCR Master Mix across the PCR Master Mix plate, add the Indexing Primers, and transfer 25 µl resuspended streptavidin beads/sample to the plate for PCR. The 25 µl resuspended streptavidin beads/sample remaining in the sample plate should be saved and stored at -20 C for future use. 8. PCR enrichment of the captured library When PCR Setup is complete, the application will prompt the user to seal the PCR reaction plate and run the appropriate program on the thermocycler. Agilent recommends the following program for enrichment of libraries prepared from 3 µg input DNA. The number of cycles should be adjusted according to the amount of input DNA in the sample and the capture library size (consult the Agilent SureSelect XT protocol for guidance). Use a heated lid to prevent condensation. 98 C for 2 minutes 10 to 16 cycles of: 98 C for 30 seconds 65 C for 30 seconds 72 C for 1 minute 72 C for 10 minutes Hold at 4 C 9. Post-PCR SPRI Cleanup If desired, the application may be modified by a PerkinElmer field application scientist to stop after PCR Setup. This allows Post-PCR SPRI Cleanup to be carried out on a separate instrument to avoid potential cross-contamination of pre-pcr samples with post-pcr samples. This section assumes that the Post-PCR SPRI Cleanup step will be performed on the Sciclone NGS Workstation as Step 6 of the SureSelectXT Target Selection application. Note: AMPure XP beads should be warmed at room temperature for about 30 min before use. They should be taken out of 4 C storage during the PCR Setup step and resuspended thoroughly prior to use. Make 50 ml fresh 80% EtOH solution by diluting 40 ml 100% Ethanol with 10 ml nuclease-free molecular biology grade water. Pour into a Seahorse Deepwell reservoir, cover with a lid and store at room temperature. Follow the prompts in the Maestro dialog boxes to reset the deck for the SPRI Cleanup step. Discard the Deepwell plates at A5, B4, and B5. Place a clean Deepwell plate for liquid waste at A5. Place the lidded Seahorse Deepwell reservoir with 80% EtOH at B5. Aliquot the appropriate volume of SPRI beads (refer to the SS Target Selection spreadsheet) to each well in column 4 of the Seahorse Deepwell plate at A4. When the PCR program is complete, unseal the sample plate, place it on the magnet at the indicated location on the deck, and continue with the run. The Sciclone NGS Workstation will broadcast the AMPure XP beads to a clean plate, and transfer the samples from the streptavidin beads to the AMPure XP beads. After the SPRI Cleanup, the samples will be resuspended in 30 µl 10 mm Tris ph

20 10. Application complete Seal the library plate and store at -20 C for up to 7 days, or proceed directly into library validation prior to storing. Click through the dialogs in Maestro to shut down the INHECO units and complete the run. Quality control and quantification of amplified libraries The LabChip GX System may be used to check the size distribution of fragments in the enriched captured library and estimate the concentration of fragments in the appropriate size range. Make a 1:20 dilution of library into molecular biology grade water in a Bio-Rad 96-well PCR plate. Mix well by pipetting up and down, and spin the plate to remove any bubbles. Run the samples on the LabChip GX System using a High Sensitivity DNA chip and kit according to the standard LabChip protocol. Additional/alternative validation and quantification, including qpcr quantification, should be done according to the user s standard practices. Figure 17. Gel image from LabChipGX analysis of 1:10 dilutions of 16 captured libraries. Capture was performed with 500 ng adapter-ligated library per sample (as determined per LabChipGX analysis) and captured libraries were amplified with 12 cycles of PCR. Expected Results The total amount of DNA in the final library will vary depending on the quality and quantity of the input DNA, the size of the capture library, and the number of postcapture amplification cycles. Generally, libraries will have concentrations of 3-4 pg/µl and will contain approximately 100 ng total DNA in 30 µl (as measured by smear analysis with the LabChip GX). CV values across replicate samples should be less than 25%. Figure 18. Total yield of captured amplified libraries as calculated from LabChipGX data shown in figure 17. Figure 16. Representative data tracing from LabChipGX analysis of a 1:10 dilution of captured library after 12 cycles of amplification and post-pcr SPRI cleanup. 20

TRUSEQ RNA ON THE SCICLONE LIBRARY PREPARATION NGS LIQUID HANDLER USER S GUIDE ILLUMINA

TRUSEQ RNA ON THE SCICLONE LIBRARY PREPARATION NGS LIQUID HANDLER USER S GUIDE ILLUMINA Compatible with: Illumina TruSeq RNA Sample Prep v2 kits Sciclone NGS Hardware with Maestro 6.0 Sciclone NGSx Hardware with Maestro 6.2 USER S GUIDE ILLUMINA TRUSEQ RNA LIBRARY PREPARATION ON THE SCICLONE

More information

SCICLONE NGS STRANDED RNA SAMPLE PREPARATION ON THE ILLUMINA TRUSEQ WORKSTATION. User Guide. Workflow: Maestro TruSeq Stranded RNA Rev1

SCICLONE NGS STRANDED RNA SAMPLE PREPARATION ON THE ILLUMINA TRUSEQ WORKSTATION. User Guide. Workflow: Maestro TruSeq Stranded RNA Rev1 ILLUMINA TRUSEQ STRANDED RNA SAMPLE PREPARATION ON THE SCICLONE NGS WORKSTATION User Guide Workflow: Maestro TruSeq Stranded RNA Rev1 Compatible with: Illumina TruSeq Stranded Total RNA Sample Prep kits

More information

Automation of the IDT xgen Lockdown Panels on the Sciclone G3 NGS Workstation

Automation of the IDT xgen Lockdown Panels on the Sciclone G3 NGS Workstation A PPLI C A TION NOT E NGS Library Preparation Sciclone G3 NGSx Workstation Automation of the IDT xgen Lockdown Panels on the Sciclone G3 NGS Workstation Introduction Integrated DNA Technologies (IDT) xgen

More information

Automation of xgen hybridization capture on the Sciclone G3 NGS Workstation

Automation of xgen hybridization capture on the Sciclone G3 NGS Workstation next generation sequencing protocol Automation of xgen hybridization capture on the Sciclone G3 NGS Workstation www.idtdna.com For Research Use Only Version 2 Revision history Document version Date released

More information

Methods S1. Minimal Starting Amount Sample Preparation Protocol (MSA-Cap)

Methods S1. Minimal Starting Amount Sample Preparation Protocol (MSA-Cap) 1 Methods S1 Minimal Starting Amount Sample Preparation Protocol (MSA-Cap) 1. Methods. 1.1. Fragmentation. Start from 50-60 ng DNA in 10-30 µl Elution buffer (EB) or TE buffer (10mM Tris, ph 7.5/ 1 mm

More information

Hybridization capture of DNA libraries using xgen Lockdown Probes and Reagents

Hybridization capture of DNA libraries using xgen Lockdown Probes and Reagents Hybridization capture of DNA libraries using xgen Lockdown Probes and Reagents For use with: llumina TruSeq adapter ligated libraries xgen Universal Blockers TS Mix (Catalog # 1075474, 1075475, 1075476)

More information

Complete protocol in 110 minutes Enzymatic fragmentation without sonication One-step fragmentation/tagging to save time

Complete protocol in 110 minutes Enzymatic fragmentation without sonication One-step fragmentation/tagging to save time Molecular Cloning Laboratories Manual Version 1.2 Product name: MCNext UT DNA Sample Prep Kit Cat #: MCUDS-4, MCUDS-24, MCUDS-96 Description: This protocol explains how to prepare up to 96 pooled indexed

More information

NEXTFLEX ChIP-Seq Kit (For Illumina Platforms) Catalog #NOVA (Kit contains 8 reactions) Bioo Scientific Corp V15.

NEXTFLEX ChIP-Seq Kit (For Illumina Platforms) Catalog #NOVA (Kit contains 8 reactions) Bioo Scientific Corp V15. NEXTFLEX ChIP-Seq Kit (For Illumina Platforms) Catalog #NOVA-5143-01 (Kit contains 8 reactions) Bioo Scientific Corp. 2015-2018 V15.07 This product is for research use only. Not for use in diagnostic procedures.

More information

BIOO LIFE SCIENCE PRODUCTS

BIOO LIFE SCIENCE PRODUCTS BIOO LIFE SCIENCE PRODUCTS FOR REFERENCE PURPOSES This manual is for Reference Purposes Only. DO NOT use this protocol to run your assays. Periodically, optimizations and revisions are made to the kit

More information

Genome Reagent Kits v2 Quick Reference Cards

Genome Reagent Kits v2 Quick Reference Cards Chromium Genome Reagent Kits v2 Quick Reference Cards FOR USE WITH Chromium Genome Library Kit & Gel Bead Kit v2, 16 rxns PN-120258 Chromium Genome Chip Kit v2, 48 rxns PN-120257 Chromium i7 Multiplex

More information

BIOO LIFE SCIENCE PRODUCTS

BIOO LIFE SCIENCE PRODUCTS BIOO LIFE SCIENCE PRODUCTS FOR REFERENCE PURPOSES This manual is for Reference Purposes Only. DO NOT use this protocol to run your assays. Periodically, optimizations and revisions are made to the kit

More information

BIOO LIFE SCIENCE PRODUCTS

BIOO LIFE SCIENCE PRODUCTS BIOO LIFE SCIENCE PRODUCTS FOR REFERENCE PURPOSES This manual is for Reference Purposes Only. DO NOT use this protocol to run your assays. Periodically, optimizations and revisions are made to the kit

More information

DRAFT. Apollo 324 System. Protocol. PrepX-32i DNA Library

DRAFT. Apollo 324 System. Protocol. PrepX-32i DNA Library Apollo 324 System PrepX-32i DNA Library Protocol Copyright 2012, IntegenX Inc. All rights reserved. Information in this document is subject to change without notice. IntegenX assumes no responsibility

More information

Supplemental File 1: Modified Nextera XT DNA Sample Preparation Guide (Illumina, USA, Part # rev. C, October 2012).

Supplemental File 1: Modified Nextera XT DNA Sample Preparation Guide (Illumina, USA, Part # rev. C, October 2012). Supplemental File 1: Modified Nextera XT DNA Sample Preparation Guide (Illumina, USA, Part # 15031942 rev. C, October 2012). Required Kit content: Box1 ATM Amplicon Tagment Mix TD Tagment DNA Buffer NPM

More information

sparq HiFi PCR Master Mix

sparq HiFi PCR Master Mix sparq HiFi PCR Master Mix Cat. No. 95192-050 Size: 50 reactions Store at -25 C to -15 C 95192-250 250 reactions Description The sparq HiFi PCR Master Mix is a high efficiency, high-fidelity, and low bias

More information

FOR REFERENCE PURPOSES

FOR REFERENCE PURPOSES FOR REFERENCE PURPOSES This manual is for Reference Purposes Only. DO NOT use this protocol to run your assays. Periodically, optimizations and revisions are made to the kit and protocol, so it is important

More information

Quick Reference Card. Axiom Automated Target Prep Protocol Stage 1. DNA Amplification. Introduction. STAGE 1: DNA Amplification

Quick Reference Card. Axiom Automated Target Prep Protocol Stage 1. DNA Amplification. Introduction. STAGE 1: DNA Amplification Quick Reference Card Axiom Automated Target Prep Protocol Stage 1. DNA Amplification Introduction Running the Axiom Assay requires the following sets of steps: 1. Genomic DNA Prep as described in Axiom

More information

For detailed instructions about the TruSeq Custom Amplicon library preparation methods, refer to your reference guide.

For detailed instructions about the TruSeq Custom Amplicon library preparation methods, refer to your reference guide. 1 TruSeq Custom Amplicon Script Welcome Navigation Objectives Welcome to the TruSeq Custom Amplicon course. Click next to begin. Take a moment to familiarize yourself with the navigation for this course.

More information

NEXTFLEX 16S V4 Amplicon-Seq Kit (For Illumina Platforms) Catalog #NOVA (Kit contains 8 reactions) Bioo Scientific Corp V18.

NEXTFLEX 16S V4 Amplicon-Seq Kit (For Illumina Platforms) Catalog #NOVA (Kit contains 8 reactions) Bioo Scientific Corp V18. NEXTFLEX 16S V4 Amplicon-Seq Kit 2.0-4 (For Illumina Platforms) Catalog #NOVA-4203-01 (Kit contains 8 reactions) Bioo Scientific Corp. 2018 V18.07 This product is for research use only. Not for use in

More information

NEXTFLEX Rapid Directional RNA-Seq Kit (For Illumina Platforms) Catalog #NOVA (Kit contains 8 reactions)

NEXTFLEX Rapid Directional RNA-Seq Kit (For Illumina Platforms) Catalog #NOVA (Kit contains 8 reactions) NEXTFLEX Rapid Directional RNA-Seq Kit (For Illumina Platforms) Catalog #NOVA-5138-07 (Kit contains 8 reactions) Bioo Scientific Corp. 2014-2018 V18.07 This product is for research use only. Not for use

More information

HyperCap, an automatable workflow on the Agilent Bravo B

HyperCap, an automatable workflow on the Agilent Bravo B Automation Note February 2018 HyperCap, an automatable workflow on the Agilent Bravo B 1. OVERVIEW As the demand for next-generation sequencing (NGS) grows, laboratories must adapt to manage increased

More information

Single Cell 3 Reagent Kits v2 Quick Reference Cards

Single Cell 3 Reagent Kits v2 Quick Reference Cards Chromium Single Cell 3 Reagent Kits v2 Quick Reference Cards FOR USE WITH Chromium Single Cell 3' Library & Gel Bead Kit v2, 16 rxns PN-120237 Chromium Single Cell 3' Library & Gel Bead Kit, 4 rxns PN-120267

More information

TruSight DNA Amplicon Sequencing Panel

TruSight DNA Amplicon Sequencing Panel FOR RESEARCH USE ONLY Date: Illumina Kit Description: New or less experienced users are strongly advised to follow the protocol in the latest version of the Library Prep Kit Guide (part # 15054779) before

More information

GENERAL INFORMATION...

GENERAL INFORMATION... BIOO LIFE SCIENCE PRODUCTS NEXTflex-96 TM DNA Barcodes (Illumina Compatible) Catalog #: 514106 BIOO Scientific Corp. 2012 V12.11 TABLE OF CONTENTS GENERAL INFORMATION... 1 Product Overview... 1 Contents,

More information

Fragment Library Preparation Using the AB Library Builder System

Fragment Library Preparation Using the AB Library Builder System Fragment Library Preparation Using the AB Library Builder System 5500 Series SOLiD Systems QUICK REFERENCE Note: For safety and biohazard guidelines, refer to the Safety section in the Fragment Library

More information

Amplicon Library Preparation Method Manual. GS FLX Titanium Series October 2009

Amplicon Library Preparation Method Manual. GS FLX Titanium Series October 2009 GS FLX Titanium Series 1. Workflow 3. Procedure The procedure to prepare Amplicon libraries is shown in Figure 1. It consists of a PCR amplification, performed using special Fusion Primers for the Genome

More information

SureSelect Target Enrichment System for Illumina Paired-End Sequencing Library

SureSelect Target Enrichment System for Illumina Paired-End Sequencing Library SureSelect Target Enrichment System for Illumina Paired-End Sequencing Library SureSelect Target Enrichment for Illumina Paired-End Multiplexed Sequencing Protocol Version 1.0, May 2010 SureSelect platform

More information

Single Cell 3 Reagent Kits v2 Quick Reference Cards

Single Cell 3 Reagent Kits v2 Quick Reference Cards Chromium Single Cell 3 Reagent Kits v2 Quick Reference Cards FOR USE WITH Chromium Single Cell 3' Library & Gel Bead Kit v2 PN-120237 Chromium Single Cell 3 Chip Kit v2 PN-120236 Chromium i7 Multiplex

More information

BIOO LIFE SCIENCE PRODUCTS

BIOO LIFE SCIENCE PRODUCTS BIOO LIFE SCIENCE PRODUCTS FOR REFERENCE PURPOSES This manual is for Reference Purposes Only. DO NOT use this protocol to run your assays. Periodically, optimizations and revisions are made to the kit

More information

Fragment Library Preparation

Fragment Library Preparation Fragment Library Preparation 5500 Series SOLiD Systems QUICK REFERENCE Note: For safety and biohazard guidelines, refer to the Safety section in the Fragment Library Preparation: 5500 Series SOLiD Systems

More information

SureSelect Target Enrichment System for Illumina Paired-End Sequencing Library

SureSelect Target Enrichment System for Illumina Paired-End Sequencing Library SureSelect Target Enrichment System for Illumina Paired-End Sequencing Library SureSelect Human All Exon and Human All Exon Plus Protocol Version 2.0.1, May 2010 SureSelect platform manufactured with Agilent

More information

Scaling Up NGS Sample Preparation: Automation and Lab Preparation. Alicia Burt Sr. Product Manager

Scaling Up NGS Sample Preparation: Automation and Lab Preparation. Alicia Burt Sr. Product Manager Scaling Up NGS Sample Preparation: Automation and Lab Preparation Alicia Burt Sr. Product Manager Alicia_burt@agilent.com Steve Lappin Application Scientist AGILENT AUTOMATED NGS SAMPLE PREPARATION Alicia

More information

NEXTflex Cystic Fibrosis Amplicon Panel. (For Illumina Platforms) Catalog # (Kit contains 8 reactions) Bioo Scientific Corp V17.

NEXTflex Cystic Fibrosis Amplicon Panel. (For Illumina Platforms) Catalog # (Kit contains 8 reactions) Bioo Scientific Corp V17. NEXTflex Cystic Fibrosis Amplicon Panel (For Illumina Platforms) Catalog #4231-01 (Kit contains 8 reactions) Bioo Scientific Corp. 2017 V17.02 This product is for research use only. Not for use in diagnostic

More information

COFACTOR GENOMICS Cofactor ImmunoPrism Kit Version 1.0 THE LEADERS IN RNA. Cofactor ImmunoPrism Kit Version 1.0

COFACTOR GENOMICS Cofactor ImmunoPrism Kit Version 1.0 THE LEADERS IN RNA. Cofactor ImmunoPrism Kit Version 1.0 THE LEADERS IN RNA Cofactor ImmunoPrism Kit Version 1.0 Table of Contents Introduction Procedure Protocol Notes Thermal Cycler Programs Optional Protocol Modifications Support Protocol Overview Kit Reagents

More information

NEBNext. Ultra II RNA Library Prep Kit for Illumina

NEBNext. Ultra II RNA Library Prep Kit for Illumina LIBRARY PREPARATION NEBNext Ultra II RNA Library Prep Kit for Illumina Instruction Manual NEB #E7770S/L, #E7775S/L 24/96 reactions Version 1.0 4/17 be INSPIRED drive DISCOVERY stay GENUINE This product

More information

NEBNext Direct Custom Ready Panels

NEBNext Direct Custom Ready Panels LIBRARY PREPARATION NEBNext Direct Custom Ready Panels Instruction Manual NEB #E6631S/L/X 8/24/96 reactions Version 1.0 4/18 be INSPIRED drive DISCOVERY stay GENUINE i This product is intended for research

More information

5X WGS Fragmentation Mix

5X WGS Fragmentation Mix 4.1. 5X WGS Fragmentation Mix Instructions for Use Product Number Y9410L and Y9410F Product Description The 5X WGS Fragmentation Mix is an enzyme mix to perform DNA fragmentation, end-repair and da-tailing

More information

Hashimshony, Wagner, Sher & Yanai. CEL-Seq: Single cell RNA-Seq by multiplexed linear amplification (Cell Reports).

Hashimshony, Wagner, Sher & Yanai. CEL-Seq: Single cell RNA-Seq by multiplexed linear amplification (Cell Reports). CEL-Seq Protocol Hashimshony, Wagner, Sher & Yanai. CEL-Seq: Single cell RNA-Seq by multiplexed linear amplification. 2012 (Cell Reports). Reagents: LoBind tubes 0.5 ml Eppendorf 022431005 Ultra pure RNase

More information

Axiom TM 2.0 Automated Target Prep Protocol

Axiom TM 2.0 Automated Target Prep Protocol Quick Reference Card Axiom TM 2.0 Automated Target Prep Protocol Stage 1. DNA Amplification Introduction Running the Axiom 2.0 Assay requires the following sets of steps: 1. Genomic DNA Prep, described

More information

HALOPLEX PCR TARGET ENRICHMENT & LIBRARY PREPARATION PROTOCOL. Version 1.0.3, April 2011 For research use only

HALOPLEX PCR TARGET ENRICHMENT & LIBRARY PREPARATION PROTOCOL. Version 1.0.3, April 2011 For research use only HALOPLEX PCR TARGET ENRICHMENT & LIBRARY PREPARATION PROTOCOL Version 1.0.3, April 2011 For research use only HALOPLEX PCR TARGET ENRICHMENT & LIBRARY PREPARATION PROTOCOL Halo Genomics AB, 2011 No part

More information

3.1 RNA Fragmentation, Priming and First Strand cdna Synthesis. 3.1A RNA Fragmentation and Priming Starting from Intact or Partially Degraded RNA:

3.1 RNA Fragmentation, Priming and First Strand cdna Synthesis. 3.1A RNA Fragmentation and Priming Starting from Intact or Partially Degraded RNA: CHAPTER 3 Please refer to revision history for a summary of protocol updates Symbols SAFE STOP This is a point where you can safely stop the protocol and store the samples prior to proceeding to the next

More information

Apollo 324 System. PrepX PGM 200 DNA Library Protocol. User Guide

Apollo 324 System. PrepX PGM 200 DNA Library Protocol. User Guide Apollo 324 System PrepX PGM 200 DNA Library Protocol User Guide Copyright 2012, IntegenX Inc. All rights reserved. LIMITATIONS For research use only. Not intended for any human or animal diagnostic or

More information

GENERAL INFORMATION...

GENERAL INFORMATION... BIOO LIFE SCIENCE PRODUCTS NEXTflex TM DNA Barcodes - 6 (Illumina Compatible) Catalog #: 514101 (48 reactions) BIOO Scientific Corp. 2012 V12.11 TABLE OF CONTENTS GENERAL INFORMATION... 1 Product Overview...

More information

SureSelect Target Enrichment System for Illumina Paired-End Sequencing Library

SureSelect Target Enrichment System for Illumina Paired-End Sequencing Library SureSelect Target Enrichment System for Illumina Paired-End Sequencing Library SureSelect Target Enrichment for Illumina Paired-End Multiplexed Sequencing Protocol Version 1.1.1, November 2010 SureSelect

More information

Twist Human Core Exome EF Multiplex Complete Kit, 16 Samples PN

Twist Human Core Exome EF Multiplex Complete Kit, 16 Samples PN Twist Human Core Exome Enrichment Kit Twist Human Core Exome EF Multiplex Complete Kit, 16 Samples PN 100252 Reagents for preparing 16 exome-enriched libraries ready for sequencing from human genomic DNA

More information

SensationPlus FFPE Amplification and 3 IVT 1 Labeling Protocol

SensationPlus FFPE Amplification and 3 IVT 1 Labeling Protocol SensationPlus FFPE Amplification and 3 IVT 1 Labeling Protocol Protocol performed using the SensationPlus FFPE Amplification and 3 IVT Labeling Kit (902039 12 rxn) Dilution of Poly-A Controls Requires

More information

BIOO LIFE SCIENCE PRODUCTS. NEXTflex TM 16S V4 Amplicon-Seq Kit 4 (Illumina Compatible) BIOO Scientific Corp V13.01

BIOO LIFE SCIENCE PRODUCTS. NEXTflex TM 16S V4 Amplicon-Seq Kit 4 (Illumina Compatible) BIOO Scientific Corp V13.01 BIOO LIFE SCIENCE PRODUCTS NEXTflex TM 16S V4 Amplicon-Seq Kit 4 (Illumina Compatible) Catalog #: 4201-01 (16 reactions) BIOO Scientific Corp. 2013 V13.01 TABLE OF CONTENTS GENERAL INFORMATION... 1 Product

More information

SureSelect Target Enrichment System for Roche 454 GS FLX and GS Junior Sequencing Platforms

SureSelect Target Enrichment System for Roche 454 GS FLX and GS Junior Sequencing Platforms SureSelect Target Enrichment System for Roche 454 GS FLX and GS Junior Sequencing Platforms Protocol Version 1.4, February 2012 SureSelect platform manufactured with Agilent SurePrint Technology Research

More information

TruSeq ChIP Sample Preparation

TruSeq ChIP Sample Preparation FOR RESEARCH USE ONLY Date: Illumina Kit Description: NOTE Unless familiar with the protocol in the latest version of the TruSeq ChIP Sample Preparation Guide (part # 15023092), new or less experienced

More information

SensationPlus FFPE Amplification and WT 1 Labeling Protocol

SensationPlus FFPE Amplification and WT 1 Labeling Protocol SensationPlus FFPE Amplification and WT 1 Labeling Protocol Protocol performed using the SensationPlus FFPE Amplification and WT Labeling Kit (902042 12 rxn) Dilution of Poly-A Controls Requires the Affymetrix

More information

Apollo 324 System. PrepX SPIA RNA-Seq Library. Protocol

Apollo 324 System. PrepX SPIA RNA-Seq Library. Protocol Apollo 324 System PrepX SPIA RNA-Seq Library Protocol Copyright 2012, IntegenX Inc. All rights reserved. Information in this document is subject to change without notice. IntegenX assumes no responsibility

More information

Updated August well High-multiplexing Tagmentation and Amplification Robyn Tanny August Company Kit Catalog Number.

Updated August well High-multiplexing Tagmentation and Amplification Robyn Tanny August Company Kit Catalog Number. 96-well High-multiplexing Tagmentation and Amplification Robyn Tanny August 2015 This protocol uses the following purchased reagents: Company Kit Catalog Number Illumina Nextera DNA Sample Preparation

More information

TailorMix Stranded mrna Sample Preparation Kit

TailorMix Stranded mrna Sample Preparation Kit TailorMix Stranded mrna Sample Preparation Kit Catalog Numbers: TM200-A and TM200-B Introduction The TailorMix Stranded mrna Sample Preparation Kit from SeqMatic is a comprehensive solution for generating

More information

NEBNext Single Cell/Low Input RNA Library Prep Kit for Illumina

NEBNext Single Cell/Low Input RNA Library Prep Kit for Illumina LIBRARY PREPARATION NEBNext Single Cell/Low Input RNA Library Prep Kit for Illumina Instruction Manual NEB #E6420S/L 24/96 reactions Version 1.0 4/18 be INSPIRED drive DISCOVERY stay GENUINE i This product

More information

SureSelect RNA Capture Enrichment System for SOLiD Multiplexed Sequencing

SureSelect RNA Capture Enrichment System for SOLiD Multiplexed Sequencing SureSelect RNA Capture Enrichment System for SOLiD Multiplexed Sequencing Protocol Version 1.2, February 2012 SureSelect platform manufactured with Agilent SurePrint Technology Research Use Only. Not for

More information

Globin Block Modules for QuantSeq Instruction Manual

Globin Block Modules for QuantSeq Instruction Manual Globin Block Modules for QuantSeq Instruction Manual Catalog Numbers: 070 (RS-Globin Block, Homo sapiens, 96 rxn) 071 (RS-Globin Block, Sus scrofa, 96 rxn) 015 (QuantSeq 3 mrna-seq Library Prep Kit for

More information

T7-Based RNA Amplification Protocol (in progress)

T7-Based RNA Amplification Protocol (in progress) T7-Based RNA Amplification Protocol (in progress) Jacqueline Ann Lopez (modifications) Amy Cash & Justen Andrews INTRODUCTION T7 RNA Amplification, a technique originally developed in the laboratory of

More information

SeqCap RNA Enrichment System User s Guide Version 1.0

SeqCap RNA Enrichment System User s Guide Version 1.0 SeqCap RNA Enrichment System User s Guide Version 1.0 For life science research only. Not for use in diagnostic procedures. Copyright 2014 Roche NimbleGen, Inc. All Rights Reserved. Editions Version 1.0,

More information

JetSeq Flex DNA Library Preparation Kit. Product Manual

JetSeq Flex DNA Library Preparation Kit. Product Manual JetSeq Flex DNA Library Preparation Kit Product Manual 2 Product Manual bioline.com/jetseq JetSeq Flex DNA Library Preparation Kit JetSeq Flex DNA Library Preparation Kit TABLE OF CONTENTS 1 Kit contents

More information

Ren Lab ENCODE in situ HiC Protocol for Tissue

Ren Lab ENCODE in situ HiC Protocol for Tissue Ren Lab ENCODE in situ HiC Protocol for Tissue Pulverization, Crosslinking of Tissue Note: Ensure the samples are kept frozen on dry ice throughout pulverization. 1. Pour liquid nitrogen into a mortar

More information

User-Demonstrated Protocol: BD Single-Cell Multiplexing Kit Human

User-Demonstrated Protocol: BD Single-Cell Multiplexing Kit Human User-Demonstrated Protocol: BD Single-Cell Multiplexing Kit For use with the 10x Chromium Single Cell 3 Reagent Kit v2 01/2018 Doc ID: 179682 Rev. 1.0 Contents Disclaimer on page 3 Overview on page 3 Workflow

More information

Biotool DNA library prep kit V2 for Illumina

Biotool DNA library prep kit V2 for Illumina Biotool DNA library prep kit V2 for Illumina Description Biotool DNA library prep kit V2 for Illumina is developed specially for the Illumina high-throughput sequencing platform, and generates sequencing-ready

More information

Ion AmpliSeq Library Kit 2.0

Ion AmpliSeq Library Kit 2.0 QUICK REFERENCE Ion AmpliSeq Library Kit 2.0 DNA Library Preparation with 1- or 2-Pool Panels Using qpcr Quantification Catalog Numbers 4475345, 4480441, 4480442, 4479790, A31133, A31136, A29751 Pub. No.

More information

FFPE DNA Extraction Protocol

FFPE DNA Extraction Protocol FFPE DNA Extraction Protocol Introduction The number of archival formalin-fixed paraffin embedded (FFPE) samples is in the millions, providing an invaluable repository of information for genetic analysis.

More information

NxSeq Long Mate Pair Library Kit

NxSeq Long Mate Pair Library Kit FOR RESEARCH USE ONLY. NOT FOR HUMAN OR DIAGNOSTIC USE. Lucigen Corporation 2905 Parmenter St, Middleton, WI 53562 USA Toll Free: (888) 575-9695 (608) 831-9011 FAX: (608) 831-9012 lucigen@lucigen.com www.lucigen.com

More information

KAPA Library Preparation Kits

KAPA Library Preparation Kits Technical Data Sheet KAPA Library Preparation Kits Illumina series Product Description The KAPA Library Preparation Kit provides all of the enzymes and reaction buffers required for constructing libraries

More information

ab High Sensitivity DNA Library Preparation Kit (For Illumina )

ab High Sensitivity DNA Library Preparation Kit (For Illumina ) ab185905 High Sensitivity DNA Library Preparation Kit (For Illumina ) Instructions for Use For the preparation of a DNA library using sub-nanogram amounts of DNA input for next generation sequencing applications

More information

ab High Sensitivity DNA Library Preparation Kit (For Illumina )

ab High Sensitivity DNA Library Preparation Kit (For Illumina ) ab185905 High Sensitivity DNA Library Preparation Kit (For Illumina ) Instructions for Use For the preparation of a DNA library using sub-nanogram amounts of DNA input for next generation sequencing applications

More information

Axiom TM 2.0 for NIMBUS TM Automated Target Prep Protocol

Axiom TM 2.0 for NIMBUS TM Automated Target Prep Protocol Quick Reference Card Axiom TM 2.0 for NIMBUS TM Automated Target Prep Protocol Introduction Running the Axiom 2.0 Assay requires the following sets of steps: 1. Genomic DNA Prep, described in either the

More information

Ion AmpliSeq Library Preparation on the Janus NGS Express

Ion AmpliSeq Library Preparation on the Janus NGS Express Ion AmpliSeq Library Preparation on the Janus NGS Express Application Guide Compatible with: Ion AmpliSeq Library Kit 2.0 Ion AmpliSeq Panels and Primer Pools Ion Express Barcode Adapters 1-96 Kits WinPrep

More information

Agilent SureSelect Automated Library Prep and Capture System

Agilent SureSelect Automated Library Prep and Capture System Agilent SureSelect Automated Library Prep and Capture System SureSelect XT Automated Target Enrichment for illumina Paired-End Multiplexed Sequencing Protocol Version C, July 2011 SureSelect platform manufactured

More information

Attention all MYbaits users: The following MYbaits protocol has been replaced with a newer version, which can be accessed from:

Attention all MYbaits users: The following MYbaits protocol has been replaced with a newer version, which can be accessed from: MYcroarray Attention all MYbaits users: The following MYbaits protocol has been replaced with a newer version, which can be accessed from: http://www.mycroarray.com/mybaits/manual s.html The following

More information

Procedure & Checklist - Preparing Asymmetric SMRTbell Templates

Procedure & Checklist - Preparing Asymmetric SMRTbell Templates Procedure & Checklist - Preparing Asymmetric SMRTbell Templates Before You Begin In this procedure, PCR products are generated using two rounds of amplification. The first round uses target specific primers

More information

ATAC-seq Protocol Kaestner Lab

ATAC-seq Protocol Kaestner Lab ATAC-seq Protocol Kaestner Lab Reagents 1X PBS Nuclease-free H 2O NP-40 10% (Sigma/Roche, catalog # 11332473001), store at 4 o C Tween-20 10% (Sigma/Roche, catalog # 11332465001), store at 4 o C Digitonin

More information

Table 1.1. Reagents Preparation Reagent Temp Out of Module* Treatment Store before using in Master Mix

Table 1.1. Reagents Preparation Reagent Temp Out of Module* Treatment Store before using in Master Mix Quick Reference Card Axiom Manual Target Prep Protocol Stage 1: DNA Amplification Intro to Manual Target Preparation Running the Axiom Assay requires the following sets of steps: 1. Genomic DNA Prep, described

More information

AMPURE PCR PURIFICATION PAGE 1 OF 7

AMPURE PCR PURIFICATION PAGE 1 OF 7 PCR PURIFICATION PAGE 1 OF 7 Please refer to http://www.agencourt.com/technical/reagent_information/ for updated protocols. AMPure is a registered trademark of Agencourt Bioscience and is for laboratory

More information

Illumina TruSeq RNA Access Library Prep Kit Automated on the Biomek FX P Dual-Hybrid Liquid Handler

Illumina TruSeq RNA Access Library Prep Kit Automated on the Biomek FX P Dual-Hybrid Liquid Handler Illumina TruSeq RNA Access Library Prep Kit Automated on the Biomek FX P Dual-Hybrid Liquid Handler Introduction David Horvath, M.S., Senior Applications Scientist, Beckman Coulter, Inc.; Tim Hill, Scientist,

More information

Procedure & Checklist - Preparing SMRTbell Libraries using PacBio Barcoded Universal Primers for Multiplex SMRT Sequencing

Procedure & Checklist - Preparing SMRTbell Libraries using PacBio Barcoded Universal Primers for Multiplex SMRT Sequencing Procedure & Checklist - Preparing SMRTbell Libraries using PacBio Barcoded Universal Primers for Multiplex SMRT Sequencing Before You Begin This document describes methods for generating barcoded PCR products

More information

Library Prep Using the With-Bead Method

Library Prep Using the With-Bead Method Target enrichment library preparation Library Prep Using the With-Bead Method Introduction This protocol is based on the method of Meyer and Kircher (2010) with modifications to accommodate the capturing

More information

sparq DNA Frag & Library Prep Kit

sparq DNA Frag & Library Prep Kit sparq DNA Frag & Library Prep Kit Cat. No. 95194-024 Size: 24 reactions Store at -25 C to -15 C 95194-096 96 reactions Description The sparq DNA Frag & Library Prep Kit provides reagents essential for

More information

Cell Hashing Protocol

Cell Hashing Protocol Cell Hashing Protocol For experiments involving Cell Hashing, use cost per cell calculator to plan experiments, determine number of hashes, number of cells to load, expected doublet rates (detected and

More information

ThruPLEX -FD Prep Kit Instruction Manual. Single Tube Library Preparation for Illumina NGS Platforms

ThruPLEX -FD Prep Kit Instruction Manual. Single Tube Library Preparation for Illumina NGS Platforms ThruPLEX -FD Prep Kit Instruction Manual Single Tube Library Preparation for Illumina NGS Platforms Contents Product Description... 2 Kit Contents... 2 Shipping and Storage... 2 Getting Started... 3 Input

More information

FFPE DNA Extraction kit

FFPE DNA Extraction kit FFPE DNA Extraction kit Instruction Manual FFPE DNA Extraction kit Cat. No. C20000030, Format 50 rxns Version 1 / 14.08.13 Content Introduction 4 Overview of FFPE DNA extraction workflow 4 Kit contents

More information

NEBNext FFPE DNA Repair Mix

NEBNext FFPE DNA Repair Mix LIBRARY PREPARATION NEBNext FFPE DNA Repair Mix Instruction Manual NEB #M6630S/L 24/96 reactions Version 4.0 4/18 be INSPIRED drive DISCOVERY stay GENUINE This product is intended for research purposes

More information

Purpose: To sequence 16S microbial DNA isolated from animal fecal matter and perform compositional analysis of bacterial communities.

Purpose: To sequence 16S microbial DNA isolated from animal fecal matter and perform compositional analysis of bacterial communities. Purpose: To sequence 16S microbial DNA isolated from animal fecal matter and perform compositional analysis of bacterial communities. Reference: This protocol was adapted from: A custom and streamlined

More information

Cleanup. Total Time 2.5 hr

Cleanup. Total Time 2.5 hr sparq DNA Library Prep Kit Cat. No. 95191-024 Size: 24 reactions Store at -25 C to -15 C 95191-096 96 reactions Description The sparq DNA Library Prep Kit provides components for the rapid construction

More information

Ion TrueMate Library Preparation

Ion TrueMate Library Preparation USER GUIDE Ion TrueMate Library Preparation for use with: Ion TrueMate Library Kit Ion TrueMate Plus Library Kit Catalog Numbers A25614 and A25656 Publication Number MAN0010280 Revision B.0 For Research

More information

Premium WGBS Kit. Whole Genome Bisulfite Sequencing. Cat. No. C (8 rxns) Version 1 I 07.15

Premium WGBS Kit. Whole Genome Bisulfite Sequencing. Cat. No. C (8 rxns) Version 1 I 07.15 Premium WGBS Kit Whole Genome Bisulfite Sequencing Cat. No. C02030034 (8 rxns) Version 1 I 07.15 Contacts diagenode headquarters Diagenode s.a. BELGIUM EUROPE LIEGE SCIENCE PARK Rue Bois Saint-Jean, 3

More information

PharmacoScan Assay 96-Array Format Manual Protocol Pub. No Rev. 1

PharmacoScan Assay 96-Array Format Manual Protocol Pub. No Rev. 1 QUICK REFERENCE PharmacoScan Assay 96-Array Format Manual Protocol Pub. No. 703461 Rev. 1 Introduction and Stage 1A: Multiplex PCR and 1B: DNA Amplification Introduction to PharmacoScan Assay 96-Array

More information

Library Loading Bead Kit (EXP-LLB001) NEBNext FFPE Repair Mix (M6630) Magnetic rack. NEBNext End repair / da-tailing Module (E7546)

Library Loading Bead Kit (EXP-LLB001) NEBNext FFPE Repair Mix (M6630) Magnetic rack. NEBNext End repair / da-tailing Module (E7546) Before start checklist Materials Consumables Equipment Low Input by PCR Barcoding Kit (SQK- LWB001) Agencourt AMPure XP beads Hula mixer (gentle rotator mixer) Library Loading Bead Kit (EXP-LLB001) NEBNext

More information

Automating Genomics Applications with the Agilent Bravo Workstation

Automating Genomics Applications with the Agilent Bravo Workstation Automating Genomics Applications with the Agilent Bravo Workstation Stephen Lappin Field Applications Specialist Agilent Automation Solution Page 0 AAS Tools for sample preparation Today we will present:

More information

E.Z.N.A. Size Select-IT Kit. D preps D preps D preps

E.Z.N.A. Size Select-IT Kit. D preps D preps D preps E.Z.N.A. Size Select-IT Kit D6488-00 10 preps D6488-01 50 preps D6488-02 200 preps September 2012 E.Z.N.A. Size Select-IT Table of Contents Introduction...2 Illustrated Protocol...3 Size Selection Guide...4

More information

SureSelect Target Enrichment System

SureSelect Target Enrichment System SureSelect Target Enrichment System Illumina Single-End Sequencing Platform Library Prep Protocol Version 2.0, December 2009 SureSelect platform manufactured with Agilent SurePrint Technology Research

More information

mrna-seq Library Prep Kit V2 on the PerkinElmer Sciclone/Zephyr NGS Workstations

mrna-seq Library Prep Kit V2 on the PerkinElmer Sciclone/Zephyr NGS Workstations Making sense of HT RNA Sequencing mrna-seq Library Prep Kit V2 on the PerkinElmer Sciclone/Zephyr NGS Workstations User Guide Catalog Numbers: 001.24 (SENSE mrna-seq Library Prep Kit V2 for Illumina, including

More information

Multiplexed Strand-specific RNA-Seq Library Preparation for Illumina Sequencing Platforms

Multiplexed Strand-specific RNA-Seq Library Preparation for Illumina Sequencing Platforms Multiplexed Strand-specific RNA-Seq Library Preparation for Illumina Sequencing Platforms Important Things to know before you start: This protocol generates strand-specific reads, but may lead to slightly

More information

PIPETMAX : Automation of the Illumina Nextera XT DNA Library Preparation Kit

PIPETMAX : Automation of the Illumina Nextera XT DNA Library Preparation Kit PIPETMAX : Automation of the Illumina Nextera XT DNA Library Preparation Kit SETUP GUIDE FOR APPLICATION NOTE AN1011 CONTENTS OVERVIEW SETUP NEXTERA XT SCRIPTS OVERVIEW FAQ SCRIPTS TAGMENTATION AMPLIFICATION

More information

Functional Genomics Research Stream. Research Meeting: November 8, 2011 cdna Library Construction for RNA-Seq

Functional Genomics Research Stream. Research Meeting: November 8, 2011 cdna Library Construction for RNA-Seq Functional Genomics Research Stream Research Meeting: November 8, 2011 cdna Library Construction for RNA-Seq Lab Issues Don t leave out boxes of water + ethidium bromide Minimize tip boxes in phenol waste

More information