1 Installation of Vacuklav -B The new Europe-autoclave Vacuklav -B uses fractionated pre-vacuum (air removal by sub atmospheric pulsing before sterilization). A high vacuum is achieved using a water-ring pump, which is both very efficient and extremely robust in continual operation. These instructions contain important information about the installation of the Vacuklav -B. Please read them carefully. General features Rear view Front view 0 Side view Figure Safety valve Serial printer port (RS ) Sterile filter Main fuses - x A /" cooling water inlet 0 Power switch /" cooling water outlet Vacuum pump safety switch Feed-water intake, Swivel screw-fit for pipe x Sliding door lock Power lead -. metre Door (hinges on left) Adjustable front feet Display and operating panel Page of
2 Installation requirements Installation of Vacuklav -B Before setting up the Vacuklav -B you should make sure that the following requirements are met: Space requirements In order to ensure adequate ventilation, there should be at least cm clearance to the sides and above the autoclave (for dimensions see Figure ), with cm clearance at the rear for connections. Sufficient room must also be available either for the MELAG reservoir container for distilled or demineralised water (0 cm x cm x cm), for a MELAdem or MELAdem 0 water demineraliser, or e.g. for the original container provided by a supplier.. Horizontal surface: A plane, horizontal surface is required which is sufficiently stable to support the autoclave (unloaded weight kg).. Power supply The power supply should be 0 V, 0 Hz, and 00 VA, with a A fuse protection, optionally with a 0 ma circuit breaker. The power cable supplied is. m long. If the MELAprint printer is also used, then an additional power socket will be required.. Water supply Connecting the Vacuklav -B autoclave to the water supply is much the same as fitting up a domestic washing machine. Care should be taken to ensure that the installation is in accordance with the relevant standards on drinking water supply systems (e.g. EN ). Whereas in domestic settings there are few problems, the threat of water contamination in doctors' or dentists' surgeries is considerable, and requirements should be adhered to strictly. [Part of EN on the protection of the drinking water supply system specifies a safety combination of nonreturn valve and pipe air-release valve ).] Many buildings already have such fittings. Additional it is practical to use a tap with integrated non-return valve and pipe air-release valve. You should ensure that the Vacuklav -B meets industrial standards, we recommend one of the following options as shown in Figure : Option I : Option II : The installation of a water tap with integrated security combination (MELAG Art. No.: 0) on an available or newly installed water pipe (rated width DN, /") The installation of an additional stop cock with integrated security combination (MELAG Art. No.: 0) directly at a corner valve of an existing water supply (e.g. for a sink or wash basin) with 0 mm pipe For a flow rate of litre/min the flow pressure should be >. bar. The cooling water pipe supplied with the Vacuklav -B is metres long. You should consider installing a leak detector with water-stop valve (Figure -Figure MELAG Art.-No.: 00). ) The combination of non-return and pipe-air release valves is sufficient for the general use of the autoclave (sterilisation of noninfectious material, prior disinfection). For the use of the autoclave as a "waste steriliser" for the sterilisation of microbiologically contaminated materials, EN requires a funnel between the drinking water supply and the cooling water inlet. Page of
3 Installation of Vacuklav -B Rear view Vacuklav -B Option I Option II Water tap with non-return and pipe-air release valves MELAG-Art-No: 0 Additional stop cock with nonreturn valve and pipe-air release valve (for use with existing corner valve) MELAG-Art-No.: 0 0 mm cold water supply (to mixing battery) Cooling water supply pipe (DN ) existing corner valve Figure. Cooling water outlet An outlet is required for the cooling water - either a wastewater outlet in the wall (rating DN 0) or a connection to an existing trap (e.g. under a sink). Examples for the connections are shown on pages and. In order to reduce noise caused by outflowing water, MELAG supplies a special quiet double-chamber trap which can replace the existing one. MELAG also supplies a metre pipe to connect the autoclave to the wastewater outlet, which should be at least 0 cm lower than the Vacuklav. The connecting pipe should be free of twists and kinks, and have a constant gradient, without any points where water can collect. The supply with feed water (distilled or demineralised) If an autoclave has a closed-loop water system, impurities in the water regularly lead to damage. Therefore, the Vacuklav -B has a one-way system. It either draws on water from a reservoir (which staff refill at regular intervals, e.g. with distilled water from the MELAdest ), or, even easier, it draws water directly from the ME- LAdem or MELAdem 0 water demineraliser. The feed-water (for steam production) must at least comply with the requirements for purified water in accordance with VDE Code of Practice 00.. Indirect supply (see page, Figure ) The Vacuklav -B draws distilled or demineralised water as required from a MELAG container (or other container) along a m pipe. The container should not be more than. m below the autoclave. Of course, it can also be placed alongside the Vacuklav -B.. Direct supply (see page, Figure ) Connection of the ion exchanger MELAdem 0 with spray gun MELAjet The simplest option is when the MELAdem 0 is connected directly to the feed water inlet of the Vacuklav -B. The MELAjet is used to clean off instruments with purified water before sterilisation.. Direct supply (see page, Figure ) Connection of reverse osmosis system MELAdem. The MELAdem reverse osmosis demineraliser can be connected directly to the feed-water inlet of the Vacuklav. Before using other water-treatment systems, you should consult with your MELAG-dealer to ensure that the water quality is adequate. Page of
4 Fitted installation: Installation of Vacuklav -B If you wish to fit the Vacuklav -B under a working surface we can supply special information on request. Installation material: The following installation material is supplied: Cooling water supply hose (m) Cooling water outlet hose (m) Double-chamber trap Storage container with tube. metre long and suction filter (if installation as Figure ) Supply hose. metre long and connector (if installation as Figure or Figure ) The following articles can be ordered: Wall outlet with Y-junction MELAG Art.-No.: 0 Y-junction with non-return valve for use with an existing trap MELAG Art.-No.: 00 / water tap with security combination MELAG Art.-No.: 0 Additional stop cock with non-return valve and pipe-air release valve (to be attached to an existing corner valve) MELAG Art.-No.: 0 Leak detector with stop valve and sensor MELAG-Art.-No.: 00 Filter for MELAdem MELAG-Art.-No.: 0 Safety combination for EN MELAG-Art.-No.: 0 Installation examples. Example : Feed water from a storage container and outlet via a wall-mounted trap 0 Figure Vacuklav -B rear view Safety combination for EN Cooling water supply pipe Water outlet pipe DN Leak detector with stop valve and sensor (MELAG Art. No.: 00 - optional) Water tap - belongs to the house installation Feed-water supply line with filter MELAG container for distilled or demineralised water - 0x0x0 mm (depth x width x height) Y-junction with non-return valve (MELAG Art.-No.: 00) 0 Double-chamber trap (MELAG Art.-No.: - in the supply included) Wall outlet (NW 0 - belongs to the house installation) Power connection Feed-water inlet Cooling water outlet Cooling water inlet Page of
5 Installation of Vacuklav -B. Example : Feed water from the MELAdem 0 and outflow via a double-chamber trap Figure 0 MELAjet (MELAG Art.-No.: 000) optional MELAdem 0 Cooling water supply pipe Cooling water inlet and supply piece for MELAdem 0 (MELAG Art.-No. - in the supply included) Cooling water outlet Water outlet pipe DN Double-chamber trap (MELAG Art.-No.: - in the supply included) Demineralised water feed line Filter for MELAdem 0 Water tap - belongs to the house installation Leak detector with stop valve and sensor (MELAG Art. No.: 00 - optional) Demineralised water feed line - in the supply included Vacuklav -B rear view Safety combination for EN. Example : Feed water from the MELAdem and outlet via a double-chamber trap 0 Figure Vacuklav -B rear view Safety combination for EN /" Y-junction (MELAG-Art.-No.: 00 - in the supply included)cooling water inlet Cooling water supply pipe Cooling water inlet Water outlet pipe DN Power connection Additional demineralised water outlet Double-chamber trap (MELAG Art.-No.: - in the supply included) 0 Y-junction with non-return valve (MELAG Art.-No.: 00 - in the supply included) Waste outflow of MELAdem Reverse osmosis system MELAdem Demineralised water feed line Filter for MELAdem Water tap - belongs to the house installation Leak detector with stop valve and sensor (MELAG Art. No.: 00 - optional) Feed water intake IA_00-00_GB_B.doc Rev.: 0/0 Page of
The Premium-Plus-Class Now even better. Up to 9 kg of instruments!
The Premium-Plus-Class Now even better Up to 9 kg of instruments! The success of a family-led company - through innovation and quality As an owner-run and operated family concern founded in 1951, we have
More informationPROCLASS A plus in quality: Vacuklav +
PROCLASS A plus in quality: Vacuklav + THE SUCCESS OF A FAMILY-RUN COMPANY THROUGH INNOVATION AND QUALITY As an owner-run and operated family concern founded in 1951, we have a long history of successful
More informationLaboklav. Type 55 to 195 Liter. Technical Datasheet. Dampfsterilisatoren Steam sterilizers
Laboklav Dampfsterilisatoren Steam sterilizers Type 55 to 195 Liter Technical Datasheet Edition 07/2008 SHP Steriltechnik AG Gustav-Ricker-Str. 62 D-39120 Magdeburg Tel: +49 (0) 391 626988-0 Fax: +49 (0)
More informationRay s top tips: Things to consider when buying an Autoclave:
136 / AUTOCLAVES Different types of Autoclaves Class B autoclaves must meet the most stringent requirements. They work with a fractionated pre-vacuum and their performance can be compared to that of large
More informationOperating Manual. MELAquick 12+
Operating Manual MELAquick 12+ Software version 4.12 Please read this Operating Manual before you start operation of the autoclave. The instructions include important safety information. Make sure to keep
More informationOperations Manual. For Research Use Only. PH: EM: WEB:
Operations Manual For Research Use Only PH: 908-769-5555 EM: info@benchmarkscientific.com WEB: www.benchmarkscientific.com Thank you for choosing the BioClave Benchtop Sterilizer. Prior to operating this
More informationREVERSE OSMOSIS SYSTEM USER'S MANUAL. Excellent Product, Excellent Water! My PurePro X6
REVERSE OSMOSIS SYSTEM USER'S MANUAL UV Plus! Excellent Product, Excellent Water! My PurePro X6 PUREPRO X6 Thank you very much for selecting Pure-Pro Water Corp. In order to bring the best use of your
More informationSTERILIZATION DEVICES
STERILIZATION DEVICES I. Central Sterilization Units II. Prismatic Steam Sterilizer III. Washing Steam Sterilizer IV. Laboratory Type Vertical Steam Sterilizer 2195 13485 14001 9001 About us Sümer A.Ş.
More informationPhoenix - Mixing System MIX08/20 Inline 20
Phoenix - Mixing System MIX08/20 Inline 20 Content. General ------------------------------------------------------------------------------------------------------ 3.0 Operation ----------------------------------------------------------------------------------------------
More informationCliniklav 25. For highest quality in surgical applications MELAG. Evidence Based Sterilization
Cliniklav 25 For highest quality in surgical applications MELAG Evidence Based Sterilization Cliniklav 25 The big one from MELAG a powerful piece of technology Built-in printer for documenting program
More informationPRO. Pre-Installation guide. Video. Complete Walkthrough. Scanomat Download Center. Scanomat Connection Diagram. Scanomat Architect Pack (2D/3D Files)
PRO Video Complete Walkthrough Pre-Installation guide Scanomat Download Center Scanomat Connection Diagram Scanomat Architect Pack (2D/3D Files) For TopBrewer Pro Version: 3.1 01-06-18 CONTENTS Introduction
More informationPRO. Pre-Installation guide. Video. Complete Walkthrough. Scanomat Download Center. Scanomat Connection Diagram. Scanomat Architect Pack (2D/3D Files)
PRO Video Complete Walkthrough Pre-Installation guide Scanomat Download Center Scanomat Connection Diagram Scanomat Architect Pack (2D/3D Files) For TopBrewer Pro Version: 3.0 01-09-16 CONTENTS Introduction
More informationLUBRICATION PUMP Z2 CENTRAL LUBRICATION
LUBRICATION PUMP APPLICATION Lubrication pumps are used as a source of pressure lubricant in the circuits of two-line central lubrication systems. Pumps can be used for lubrication of high number of lubricated
More informationUser Manual. Vacuklav 40-B Vacuklav 44-B. Autoclave. Dear doctor!
User Manual Autoclave Vacuklav 40-B Vacuklav 44-B Dear doctor! We thank you for your confidence demonstrated by the purchase of this autoclave. More than 55 years ago, MELAG began as a medium-sized family-run
More informationPre-installation Manual. Plate Processor 68/85 Thermal
Pre-installation Manual Plate Processor 68/85 Thermal Pre-installation Manual Plate Processor 68/85 Thermal T1965 Edition DC, October 2011 Part No 51280 68/85 PLATE PROCESSOR INTRODUCTION RESERVATIONS
More informationPartnership. HSG Pre & Post Vacuum - Medium Size Mobile Autoclave. Economic Line
Partnership HSG Pre & Post Vacuum - Medium Size Mobile Autoclave Economic Line Fast,Heavy Duty, for Medical Clinics & OR Economic Line The HSG Economic Line includes 85 liter and 160 liter single door
More informationAUTOCLAVE SERIES VERSION 18 AND 23 CLASS B CYCLES EXPERIENCE AND INNOVATION SINCE 1948
AUTOCLAVE SERIES VERSION 18 AND 23 CLASS B S EXPERIENCE AND INNOVATION SINCE 1948 USER-FRIENDLY INTUITIVE INTERFACE The 5'' high-definition colour touch screen display, and its intuitive graphics, make
More informationUser Manual. Vacuklav 23 B+ Vacuklav 31 B+ Autoclave. as of software version Dear doctor,
User Manual Autoclave Vacuklav 23 B+ Vacuklav 31 B+ as of software version 4.06 Dear doctor, We thank you for your confidence demonstrated by the purchase of this autoclave. Since 1951, MELAG began as
More informationRO102 P URE PRO DRINKING WATER SYSTEM USER'S MANUAL. Memo REVERSE OSMOSIS SYSTEM. Type of product. Date of purchase. Address RO102
Type of product RO102 Memo Date of purchase Name Address Tel REVERSE OSMOSIS SYSTEM PURE- REVERSE OSMOSIS SYSTEM 15 RO102 USER'S MANUAL 01 Introduction of RO102 02 What is reverse osmosis 03 Components
More informationAQUAPROF BASIC AQUAPROF TOP ACTIVE SWITCH
AQUAPROF BASIC AQUAPROF TOP ACTIVE SWITCH SYSTEM FOR THE EXPLOITATION OF RAINWATER AQUAPROF TOP ACTIVE SWITCH AQUAPROF BASIC / AQUAPROF TOP AQUAPROF TOP Applications The AQUAPROF unit is used for the management
More informationS800-Direct Flow USER'S MANUAL. Memo REVERSE OSMOSIS SYSTEM. Type of product. Date of purchase. Address
Memo Type of product S800-Direct Flow Date of purchase Name Address Tel REVERSE OSMOSIS SYSTEM 15 S800-Direct Flow USER'S MANUAL 01 Introduction of S800-Direct Flow 02 What is reverse osmosis 03 Components
More informationCustomer Responsibilities
Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation
More informationInstallation plan. Washer-extractor PW 6107 EL. en - GB M.-Nr / 01
Installation plan Washer-extractor PW 6107 EL en - GB 10.11 M.-Nr. 09 270 450 / 01 M.-Nr. 09 270 450 / 01 2 Technical Datasheet Washer-extractor: Heating: PW 6107 Electric (EL) Legend: Solid circle
More informationPrice list valid:
Vertical laboratory autoclaves Price list valid: 01.01. - 31.12 2019 CertoClav laboratory autoclave EL (Thermostat controlled) 8501312 12 l GS 125/140 C 2 fixed 125 C or 140 C 12 l Aluminum 24 x 26 cm
More informationDRINKING WATER CONDITIONERS WATER PURIFIERS U.V. STERILIZERS REVERSE OSMOSIS GELPUR FILTER 113 GELPUR FILTER SI 113 GELPUR SL
DRINKING WATER CONDITIONERS WATER PURIFIERS U.V. STERILIZERS REVERSE OSMOSIS GELPUR FILTER 113 GELPUR FILTER SI 113 GELPUR SL 250 114 GELPUR SL 250 FAST 115 GELPUR DF 116 GELPUR DF TOP 117 GELPUR CMJ 80
More informationPRODUCT DESCRIPTION PARAMETERS
PRODUCT SHEET PRODUCT DESCRIPTION PARAMETERS Steam boilers PB-P PB-PP PB-NP Moderated-pressure three-pass steam boilers combusting gaseous and liquid fuels in the saturated steam and superheated steam
More informationChungHo Dealer Technical Guide
ChungHo Dealer Technical Guide Recommendations Prior to Installation -Flush out the system -Check for leaks / hose connections -Test Board connections Installation Procedures -Basic POU Installation -Do
More informationThank you for Being our Customer!
USER MANUAL Hisha Thank you for Being our Customer! Dear Customer, We sincerely appreciate your purchase of KYK alkaline water Ionizer medical substance producer and we hope your health and happiness.
More informationUser Manual. MELAtronic 15 EN+ Autoclave. as of software version V4.07. Dear Dr.
User Manual Autoclave MELAtronic 15 EN+ as of software version V4.07 Dear Dr. We thank you for your confidence demonstrated by the purchase of this product. Since 1951, MELAG began as a medium-sized family-run
More informationMELAtronic 15 MELAtronic 17 MELAtronic 23
Operating Manual For the MELAG Autoclaves -Not for medical use / only for veterinary use- MELAtronic 15 MELAtronic 17 MELAtronic 23 Dear Doctor: Thank you very much for the trust that you have shown in
More informationAqualife Products (Australia) Pty Ltd Steam Sterilisor Water System
Aqualife Products (Australia) Pty Ltd Steam Sterilisor Water System AquaClave P5 REVERSE OSMOSIS WATER TREATMENT SYSTEM Attention USER S MANUAL 01 Introduction of AquaClave 02 What is reverse osmosis 03
More informationUser Manual. Euroklav 23 VS+ Euroklav 29 VS+ as of software version Autoclave. Dear doctor,
User Manual Autoclave Euroklav 23 VS+ Euroklav 29 VS+ as of software version 4.06 Dear doctor, We thank you for your confidence demonstrated by the purchase of this autoclave. More than 55 years ago, MELAG
More informationFast. Easy to use. Secure traceability.
Sterilization, Hygiene and Maintenance Fast. Easy to use. Secure traceability. NEW Lisa Sterilizer PERFORMANCE* Accelerated type B cycle thanks to Eco Dry technology EASY TO USE thanks to a new user interface
More informationModel No. BWTDWFK-TRIFLO.
Model No. BWTDWFK-TRIFLO www.bwt-uk.co.uk Before you start This tap is suitable for use at water supply pressures of 2 bar (min.) to 5 bar (max.). This product must be connected to the water supply in
More informationSTERIVAPHP. Original without compromises. protecting human health. new line steam sterilizers
STERIVAPHP new line steam sterilizers Original without compromises protecting human health Tradition, quality, innovation Long-term experiences of BMT Medical Technology s.r.o., advanced patented solutions
More informationStack Parking Surface Mounted Stackers
1 / 6 Stack Parking Surface Mounted Stackers swiss-park V-2 Comfort The V-2 is designed to park cars of any size thanks to its vertical adjustability. It will accommodate cars as tall as the garage ceiling
More informationINTRODUCTION. History RAYPA was founded by Mr. Ramón Espinar nd generation take over New makeover
INTRODUCTION INTRODUCTION History 1970 RAYPA was founded by Mr. Ramón Espinar 2008 2 nd generation take over 2016 New makeover 2018 30 employees Turn over: 3.6 M 85% Sales on Export Sales in more than
More informationBasic design of laboratory bioreactor
Basic design of laboratory bioreactor In terms of the construction, the following variants of the laboratory bioreactor can be made: 1. Glass bioreactor (without a jacket) with an upper stainless steel
More informationAutomatic Tabletop Autoclaves
Automatic Tabletop Autoclaves Fully Automatic Autoclaves Robust & Reliable The Fully Automatic Tabletop Autoclave is designed for the needs of modern clinics where autoclaves are expected to operate non-stop
More informationPRIMARY GRADE WATER L A B W A T E R. PURELAB TM Prima. Primary Grade Water Purification Systems. The Laboratory Water Specialists
PRIMARY GRADE WATER L A B W A T E R PURELAB TM Prima Primary Grade Water Purification Systems The Laboratory Water Specialists PURELAB Prima The simple and dependable choice for primary grade water The
More informationPCB Water Cleaning System
PCB Water Cleaning System User Manual for OKO-2000uP 2004. November ~ INDEX 1. Safety Notes () ---------------------------------------------------------------------------------------------------- 3 2.
More informationSampling. Stationary Portable. Sampling. Sewage Treatment. Municipal Sewerage Systems. Water Protection Control
Sampling Stationary Portable Sampling is an important factor for the compliance with required regulations and efficient operation. It forms the basis for continuous monitoring objectives and criteria,
More informationSYSTEM 600 BILL CHANGER INSTALLATION INSTRUCTIONS Congratulations...
www.standardchange.com 1-800-968-6955 Technical Phone Support is from 8:00AM to 7:30PM E.S.T., Monday-Friday Walk-in Service is from 8:00AM to 4:30PM E.S.T., Monday-Friday Parts Department is from 8:00AM
More informationDemineralisation. PUROTAP expert. Installation Function Operation Service. Demineralised heating water.
EN Demineralised heating water Demineralisation PUROTAP expert Installation Function Operation Service Demineralised heating water makes heating systems safe in accordance with VDI 205 and SWKI BT 102-01
More informationUSER'S MANUAL. Memo PURE-PRO REVERSE OSMOSIS SYSTEM ET-PURE. Type of product. Date of purchase. Address
Type of product ET-PURE Memo Date of purchase Name Address Tel REVERSE OSMOSIS SYSTEM 15 ET-PURE ET-PURE USER'S MANUAL 01 Introduction of ET-PURE 02 What is reverse osmosis 03 ET-PURE layout and components
More informationVARIO XX28 Stainless Steel Time And Temperature Controlled Electric Model
VARIO XX28 Stainless Steel Time And Temperature Controlled Electric Model Operators Manual Issue 5 For correct operation of this autoclave it is essential that the chamber be filled with water up to the
More informationINSTALLATION & OPERATING INSTRUCTIONS
MEDIA FILTER MODELS: Turbidex 21" Media Filters 21 MAGNUM FILTER AUTOMATIC TIMER VALVE TURBIDEX HYPER FILTRATION MEDIA TurbidexTM is more than our name it's the product as well. TurbidexTM granules consist
More informationBTT : 1 autoclave bin
3' dia x 3' long (NB#195) 3' ft. diameter, 3' ft. length, 75 PSI pressure, 500 Deg. F temperature Contact: Pietro Wooldridge Phone: 606-677-2616 BTT : 1 autoclave bin 1.1.0 BONDTECH AUTOCLAVE/STERILIZER
More informationTechnical Description
Technical Description Autoclave Euroklav 23 VS+ Euroklav 29 VS+ as of software version 4.06 Dear doctor, We thank you for your confidence demonstrated by the purchase of this autoclave. More than 55 years
More informationVARIO XX28 Stainless Steel Time And Temperature Controlled Electric Model
DIXONS SURGICAL INSTRUMENTS LTD, 1 ROMAN COURT HURRICANE WAY, WICKFORD BUSINESS PARK WICKFORD, ESSEX, SS11 8YB, UK TEL: +44 (0)1268 764614 - FAX: +44 (0)1268 764615 EMAIL: info@dixons-uk.com - WEB: http://www.dixons-uk.com
More informationSPIRIT. Water purifier. Complete User Manual
SPIRIT Water purifier Complete User Manual Document No 822901692 Revision BB1522 Bluewater 2015 Contents INTRODUCTION...03 About this manual...03 Limited warranty...03 Environment information...03 Disposal...03
More informationWARNINGS AND PRECAUTIONS
WARNINGS AND PRECAUTIONS If you have questions about the unit you are repairing, please do not hesitate to contact your local SciCan representative for information. Also, the STATIM is heavy. Exercise
More informationModel KMF 115 Constant climate chambers with expanded temperature / humidity range
Model KMF 115 Constant climate chambers with expanded temperature / humidity range The BINDER KMF ensures absolutely constant test conditions throughout the testing area. A big advantage of this constant
More informationENDOSCOPIC REPROCESSING ETD DOUBLE Advancing the Efficiency of Reprocessing.
11157 ENDOSCOPIC REPROCESSING ETD DOUBLE Advancing the Efficiency of Reprocessing. ETD DOUBLE: INTRODUCING THE FIRST OLYMPUS PASS- THROUGH WASHER-DISINFECTOR With the ETD Double, Olympus Has Advanced Its
More informationAqua-Recycling-Control 80B Plus
Aqua-Recycling-Control 80B Plus Item No G17200, G17205 Example picture, Aqua-Recycling-Control including Aqua-Control (is not included in scope of supply) Typical applications - Decentralized waste water
More informationMRI Chiller Installation Guidelines
MRI Chiller Installation Guidelines *These Items are the responsibility of the installing contractor to complete prior to the startup of the MRI Chiller.* **Please refer to DTS chiller manual for further
More informationUser Manual. Vacuklav 41 B+ Vacuklav 43 B+ Autoclave. as of software version Dear Dr.
User Manual Autoclave Vacuklav 41 B+ Vacuklav 43 B+ as of software version 3.055 Dear Dr. We should like to extend our thanks for the expression of trust in our company which you have displayed through
More informationPRIMARY GRADE WATER. PURELAB Prima. Primary Grade Water Purification Systems. The Laboratory Water Specialists
PRIMARY GRADE WATER PURELAB Prima Primary Grade Water Purification Systems The Laboratory Water Specialists PURELAB Prima The simple and dependable choice for primary grade water The PURELAB Prima range
More informationHRO-50 Water Purifier
HRO-50 Water Purifier Product Installation & Operation Manual Honeywell Environmental & Combustion Controls (Tianjin) Co., Ltd. Content Important Notes P2 Product Functions P3-P5 Installation Instructions
More informationLow-profile, 11U high cabinets hold networking equipment, servers, and phone equipment.
RM140A-R3 RM140A-M-R3 RM145A-R3 RM145A-M-R3 Small Office/Home Office (SOHO) Cabinets Low-profile, 11U high cabinets hold networking equipment, servers, and phone equipment. Fit under most desktops in the
More informationEN STERILINE 9.5. Class B and S sterilisation systems in compliance with international standards
AUTOMATIC STERILIZATION SYSTEMS EN 13060 STERILINE 9.5 Class B and S sterilisation systems in compliance with international standards EN 13060 STERILINE 9.5 B CLASS The class B sterilization in compliance
More informationSTERIVAP. Excellent Price High Quality Engineering. protecting human health. Large steam sterilizer for health care
STERIVAP Large steam sterilizer for health care Excellent Price High Quality Engineering protecting human health MMM Group leading service supplier for health care Since 1954 the MMM Group has been active
More informationMASTER. Water Conditioning Corp. OPERATION MANUAL. PURO PRO 1200 SC Series Modular Reverse Osmosis System. January 2012 Version
MASTER Water Conditioning Corp. www.masterwater.com OPERATION MANUAL PURO PRO 1200 SC Series Modular Reverse Osmosis System January 2012 Version Table of Contents Model Number... 1 Shipping Description
More informationFor Safe and Delicious Water
RESIDUAL CHLORINE ANALYZER 400G NON-REAGENT FREE CHLORINE ANALYZER FC400G 400G RESIDUAL CHLORINE ANALYZER FC400G NON-REAGENT FREE CHLORINE ANALYZER Bulletin 12F01A01-01E www.yokogawa.com/an/ For Safe and
More informationCompact. Pre-Installation guide. Video. Complete Walkthrough. Scanomat Download Center. Scanomat Connection Diagram
Compact Video Complete Walkthrough Pre-Installation guide Prepare for your TopBrewer. A comprehensive step-by-step installation guide is available through your dealer, authorized installer or our website,
More informationERS-106UV USER'S MANUAL. Memo PURE-PRO REVERSE OSMOSIS SYSTEM. Type of product. Date of purchase. Address
Type of product Memo Date of purchase Name Address Tel REVERSE OSMOSIS SYSTEM 15 ERS-106UV USER'S MANUAL 01 Introduction of ERS-106UV 02 What is reverse osmosis 03 Components & Selected filters 04 Cartridge
More informationSCO Series - Air Jacketed Units. SHEL LAB CO2 Incubators
SCO Series - Air Jacketed Units SHEL LAB CO2 Incubators Ideal for Clinical, Analytical, and General Laboratory Applications SCO air jacketed CO2 incubators provides an optimized culturing environment with
More informationGerber M Series Site Requirements
Gerber M Series Site Requirements Preparing for the arrival of your new machine. P rior to the arrival of your Gerber M Series Cutting System there are certain things you can prepare to facilitate a quick
More informationSOPHISTICATED DESIGN REQUIREMENTS IN PLANT ENGINEERING FOR AN XLPE EXTRUDER FEEDING PLANT AS AN EXAMPLE
SOPHISTICATED DESIGN REQUIREMENTS IN PLANT ENGINEERING FOR AN XLPE EXTRUDER FEEDING PLANT AS AN EXAMPLE Dipl.-Ing. Reza Izadpanah, Senior Manager Sales, Zeppelin Systems GmbH CONTENTS 1. Material handling
More informationUser Manual. Euroklav 23 VS+ Euroklav 29 VS+ as of software version Autoclave. Dear doctor,
User Manual Autoclave Euroklav 23 VS+ Euroklav 29 VS+ as of software version 4.06 Dear doctor, We thank you for your confidence demonstrated by the purchase of this autoclave. Since 1951, MELAG began as
More informationLARGE CAPACITY PHARMACEUTICAL AUTOCLAVES
LARGE CAPACITY PHARMACEUTICAL AUTOCLAVES www.eryigit.com.tr Our Vision Our vision is to make Eryiğit an example among its class on its way to become the first medical device manufacturer that comes to
More informationPreparation of Reagent Water
General 07_02_09.003 Dates Active: March 15 th, 2018 - Present Preparation of Reagent Water 1 INTRODUCTION Preparation of reagent water is one of the most important processes in the laboratory. Purified
More informationMRI Chiller Installation Guidelines
MRI Chiller Installation Guidelines *These Items are the responsibility of the installing contractor to complete prior to the startup of the MRI Chiller.* **Please refer to DTS chiller manual for further
More informationDraft proposals for Test methods for close-coupled solar water heating systems - Reliability and safety
IEA/SHC Task 57, Subtask B Draft proposals for new test procedures B4: Final Draft Draft proposals for Test methods for close-coupled solar water heating systems - Reliability and safety HE Zinian Beijing
More informationSterilization & Infection Control Products
Sterilization & Infection Control Products Mechanical Autoclaves Tuttnauer semi-automatic (M) autoclave is designed to complement any healthcare and laboratory facility. Time tested reliability, durability
More informationBoss Water Systems Australia Reverse Osmosis System Model 021-4P-GM
Boss Water Systems Australia Reverse Osmosis System Model 021-4P-GM Congratulations on your purchase of the most advanced water purifier system available! Installation and Service Guide short web version,
More informationRO400 USER'S MANUAL. Memo PURE-PRO REVERSE OSMOSIS SYSTEM RO400. Type of product. Date of purchase. Address
Type of product RO400 Memo Date of purchase Name Address Tel REVERSE OSMOSIS SYSTEM 19 RO400 USER'S MANUAL 01 Introduction of RO400 02 What is Reverse Osmosis 03 Components& Selections 04 Cartridge filters
More informationEnewall Render Silo User Guide
At Enewall, we have always placed a considerable emphasis on customer care. It is our aim to make things as easy as possible for our customers by offering innovative products and technology backed up by
More informationRD01 WHAT S IN THE BOX INSTALLATION MANUAL
INSTALLATION MANUAL RD0 WHAT S IN THE BOX Rain Director control panel with mode indicators, programming buttons and control valves and 0 V AC to V wall adapter. Control panel measures 80mm W x 70mm H x
More informationAdvanced water purifier system operation manual and setup guide
Advanced water purifier system operation manual and setup guide What is Reverse Osmosis? Reverse osmosis, also known as hyper filtration, is the finest filtration known. This process will allow the removal
More informationRC400G FC400G RESIDUAL CHLORINE ANALYZER RC400G NON-REAGENT FREE CHLORINE ANALYZER FC400G RESIDUAL CHLORINE ANALYZER
RESIDUAL CHLORINE ANALYZER 400G NON-REAGENT FREE CHLORINE ANALYZER FC400G 400G RESIDUAL CHLORINE ANALYZER FC400G NON-REAGENT FREE CHLORINE ANALYZER Bulletin 12F01A01-01E www.yokogawa.com/an/ For Safe and
More informationSMF WATER SYSTEMS. (PIC/S July.2004) By Anand Kuptamatee Thai FDA Conference September 19-21, 2005
SMF WATER SYSTEMS (PIC/S July.2004) By Anand Kuptamatee Thai FDA Conference September 19-21, 2005 DESCRIPTION OF WATER SYSTEMS CONTENT: Water Purification Process System capacity Construction materials
More informationWater treatment in dialysis technology
Fig.: Twin pass dialysis water treatment, supports full thermal disinfection Water treatment in dialysis technology in partnership with Herco Dialysis water treatment it s a matter of trust At a glance
More informationUltraGas. Technical Data Output range 15 to 2000 kw. Condensing Gas Boiler /02-07/15 Subject to modifications
UltraGas Condensing Gas Boiler Technical Data Output range 15 to 2000 kw - 07/15 Subject to modifications UltraGas UltraGas Condensiing 15-2000D Gas Boiler Contents page Description 3 UltraGas 15-100 Technical
More informationR442A (RS-50) Features and uses of the R442A (RS-50) Lubricants TECHNICAL DATA SHEET. Barcelona Girona Madrid Zaragoza
SHEET R442A (RS-50) Features and uses of the R442A (RS-50) R442A (RS-50) is a non-flammable HFC blend with an ODP = 0 and low global warming potential (GWP), developed to comply with the new dictates laid
More informationMASTER. Water Conditioning Corp. OPERATION MANUAL. PURO PRO 1200 SCP Series Modular Reverse Osmosis System. January 2012 Version
MASTER Water Conditioning Corp. www.masterwater.com OPERATION MANUAL PURO PRO 1200 SCP Series Modular Reverse Osmosis System January 2012 Version Table of Contents Model Number... 1 Shipping Description
More informationRS-106M-P USER'S MANUAL. Memo PURE-PRO REVERSE OSMOSIS SYSTEM. Type of product. Date of purchase. Address
Type of product Memo Date of purchase Name Address Tel REVERSE OSMOSIS SYSTEM 15 RS-106M-P USER'S MANUAL 01 Introduction of RS-106M-P 02 What is reverse osmosis 03 Components & Selected filters 04 Cartridge
More informationChapter 2. TC 2020 R FMC compact installation conditions. Drawing number: EN Part number:
Chapter 2 TC 2020 R FMC compact installation conditions Drawing number: 90951-5-20EN Part number: 0347572 B435EN02.DOC TC 2020 R FMC compact installation conditions - Version 2 2-1 Table of Contents 1.
More informationVentum Series. The new Generation Wood Gasification Boilers. Europe s Fastest Selling Wood Gasification Boiler.
Ventum Series The new Generation Wood Gasification Boilers Europe s Fastest Selling Wood Gasification Boiler. The New Generation Ventum Series leaps forward in gasification technology making them the fastest
More informationSTERIVAP. Excellent Price Economic use of utilities. protecting human health. Steam sterilizer for health care
STERIVAP Steam sterilizer for health care Excellent Price Economic use of utilities protecting human health MMM Group leading supplier of health care systems Since 195 the MMM Group has been active world-wide
More informationSteam Sterilizer. SA Series. for hospitals, CSSD and medical centers
Steam Sterilizer SA Series for hospitals, CSSD and medical centers Driven by customer needs Colussi is a leading infection control solution provider, supplying the healthcare, laboratory research and pharma
More informationHDG Turbotec. HDG Log wood boilers. Comfortable and safe. 340-litre fuel chamber. Log wood sizes up to 1-metre long
HDG Log wood boilers HDG Turbotec Fuels: Log wood up to 1 metre logs Rough chips Capacity: 50 kw and 60 kw Comfortable and safe 340-litre fuel chamber Log wood sizes up to 1-metre long Highly effective
More informationSump Pump System OEM INSTALLATION MANUAL
Sump Pump System OEM INSTALLATION MANUAL TABLE OF CONTENTS System and Safety Information 2 Preparation 2 Resources Required 2 Installation 3 Plumbing 3 Electrical Connections 4 Operation 5 System and Safety
More informationINSTALLATION MANUAL RD01
INSTALLATION MANUAL RD0 WHAT S IN THE BOX Smart header tank for roof space, including level sensors, float switch and overflow tower. The 9 litre tank measures 665 L x 90 W x 50 H mm. Flow reducing valves
More informationAP5K-C Precision AC Double-Pulse Spot Welding Machine User s Manual Shenzhen Will-Best Electronics Co., Ltd
AP5K-C Precision AC Double-Pulse Spot Welding Machine User s Manual Shenzhen Will-Best Electronics Co., Ltd 1 Content 1. Introduction...3 1.1 Functions...3 1.2 Units of AP5K-C...4 2. The Initial Installation
More information! CAUTION: Read and follow these instructions.
OmniLog SAFETY AND INSTALLATION INSTRUCTIONS! CAUTION: Read and follow these instructions. Avoid Contact with Hazardous Live Parts WARNING: Risk of electrical shock. Refer servicing to qualified personnel.
More informationVaVie Water Purifier
VaVie Water Purifier Installation instructions Thank you for purchasing VaVie water purifier. Please read the instructions carefully before using. Please keep this booklet for later reference. Model: WA75-A
More information