Pulsed Flame Photometric Detector (PFPD) for CP-3800

Size: px
Start display at page:

Download "Pulsed Flame Photometric Detector (PFPD) for CP-3800"

Transcription

1 Pulsed Flame Photometric Detector (PFPD) for CP-3800 Operator s Manual Varian Analytical Instruments 2700 Mitchell Drive Walnut Creek, CA /US Varian, Inc Printed in U.S.A :REV 3

2 Table of Contents SECTION 1 INTRODUCTION OVERVIEW DESCRIPTION OF THE PFPD BASIC COMPONENTS OF THE PFPD FEATURES OF THE PFPD ASSEMBLY FEATURES OF THE PFPD ELECTROMETER BOARD THEORY OF OPERATION OF THE PFPD STAGES OF FLAME PULSATION STAGES OF PULSED-FLAME OPERATION HOW DOES THE PFPD ACHIEVE ITS GREATER DETECTIVITY AND SELECTIVITY? FLOW PATH OF THE PFPD REFERENCES SPECIFICATIONS OF THE PFPD NITROGEN OR OTHER ELEMENT OPERATION SECTION 2 SETTING UP THE PFPD OVERVIEW INSPECTING THE PFPD CONTENTS OF THE PFPD ACCESSORY KIT CONTENTS OF THE PFPD DIFFUSION SOURCE KIT SECTION 3 OPERATION OVERVIEW INSTALLING A CHROMATOGRAPHIC COLUMN COLUMN INSTALLATION PREPARE THE COLUMN END PFPD ASSEMBLY DISASSEMBLE THE PFPD INSERT THE COLUMN REASSEMBLE THE PFPD ADJUSTING THE GAS FLOW RATES SPLIT VALVE PFPD GAS FLOW DIAGRAM TO ADJUST THE GAS FLOW RATES CALIBRATING THE FLOW CONTROLLERS :3 i

3 Table of Contents 3.5 OPTIMIZING THE PFPD DETECTIVITY SELECTIVITY THROUGHPUT SUGGESTED FUEL GAS FLOW RATES SUGGESTED PFPD ADJUSTMENT SETTINGS OPTIMIZATION CRITERIA ESTABLISHING TICK-TOCK PHOTOMULTIPLIER TUBE (PMT) HIGH VOLTAGE ADJUSTING THE PHOTOMULTIPLIER TUBE (PMT) VOLTAGE OPTIMIZING THE PFPD REPEATED INJECTIONS OF STANDARD CONTINUOUS INFUSION OF STANDARD TESTING THE PFPD PERFORMANCE TEST THE PERFORMANCE OF THE PFPD PFPD TEST CHROMATOGRAM: SULFUR PERFORMANCE PFPD TEST CHROMATOGRAM: PHOSPHORUS PERFORMANCE DETECTIVITY PFPD TEST CHROMATOGRAM: SULFUR DETECTIVITY PFPD TEST CHROMATOGRAM: PHOSPHORUS DETECTIVITY USING THE GATE DELAY AND GATE WIDTH PARAMETERS SETTING THE TRIGGER LEVEL SETTING GATE DELAY AND GATE WIDTH SET GATE WIDTH BUILDING A GC METHOD BUILDING A GC WORKSTATION METHOD QUANTITATION QUANTITATION WITH THE PFPD PEAK AREA QUANTITATION PEAK HEIGHT QUANTITATION PEAK RESOLUTION WITH THE PFPD QUANTITATION IN THE SULFUR MODE SQUARE ROOT FUNCTION SECTION 4 ADVANCED OPERATION OVERVIEW THEORY OF OPERATION TABLE OF OPTIMUM FLOW RATES INSTALLATION OF THE NITROGEN MODE PARTS NITROGEN MODE OPERATION PHOSPHORUS MODE OPERATION SULFUR MODE OPERATION SPARE PARTS AND SUPPLIES ii :3

4 Table of Contents 4.8 ARSENIC SELECTIVE DETECTION EMISSION WAVELENGTHS EMISSION TIMING FLAME COMPOSITION FILTERS SELENIUM SELECTIVE DETECTION EMISSION WAVELENGTHS EMISSION TIMING FLAME COMPOSITION FILTERS TIN SELECTIVE DETECTION EMISSION WAVELENGTHS EMISSION TIMING FLAME COMPOSITION FILTERS MANGANESE SELECTIVE DETECTION EMISSION WAVELENGTHS EMISSION TIMING FLAME COMPOSITION FILTERS SECTION 5 MAINTAINING THE PFPD OVERVIEW REPLACING THE OPTICAL FILTER CHANGING THE COMBUSTOR SIZE REPLACING A DIRTY COMBUSTOR ORDERING PARTS AND SUPPLIES FOR THE PFPD SECTION 6 TROUBLESHOOTING OVERVIEW FLAME DOES NOT IGNITE BASELINE NOISE IS TOO HIGH BASELINE IS ERRATIC (TICK-TOCK) :3 iii

5 1. Section 1 Introduction 1.1 OVERVIEW Topic See Page... Description of the PFPD Theory of Operation of the PFPD Specifications of the PFPD Nitrogen or Other Element Operation DESCRIPTION OF THE PFPD The Pulsed Flame Photometric Detector (PFPD) is the newest member of the family of flame-based gas chromatographic detectors. The PFPD operates in a pulsed-flame rather than continuous-flame mode and offers a number of significant improvements over the conventional flame photometric detector: Increased detectivity Increased selectivity (relative to hydrocarbons) Decreased hydrogen and air consumption BASIC COMPONENTS OF THE PFPD PFPD assembly PFPD electrometer board PFPD pneumatics :3 1-1

6 Introduction FEATURES OF THE PFPD ASSEMBLY :3

7 Introduction Item Description Function 1 Ignitor coil assembly Provides the heat that ignites the flame. 2 Photomultiplier tube (PMT) Converts the photons emitted in the flame into a current for measurement. 3 Ignitor cap Holds the ignitor coil and minimizes reflected light from the ignitor coil. 4 Detector tower Supports the optics, quartz combustor tube, and ignitor cap. 5 Quartz combustor A transparent inert surface that contains the column effluent for combustion and the transmission of emitted light. 6 Detector base Holds the detector tower and secures the chromatographic column to the detector. 7 Light pipe Carries light from the combustion region toward the optical filter and photomultiplier tube. 8 Combustor support Holds and positions the quartz combustor within the detector tower. 9 Combustor support seal Seals the space between the combustor support and detector base. 10 Hydrogen-air center flow inlet Hydrogen-air mixture enters here and flows up through the center of the quartz combustor. 11 Hydrogen-air wall flow inlet Hydrogen-air mixture enters here and flows along the outside wall of the quartz combustor and into the ignitor chamber. 12 Air 2 inlet Air 2 enters here and mixes with wall flow. 13 Crunch washer, upper Seals the space between the ignitor cap and detector tower. 14 Crunch washer, lower Seals the space between the detector tower and detector base. 15 Allen screws, 9/64" Secures the ignitor cap to the detector tower. 16 Allen screws, 8/32 x 1 1/2" Secures the detector tower to the detector base. 18 Phillips-head screws, 8/32 x 3/8" Secures the photomultiplier tube assembly to the support bracket :3 1-3

8 Introduction FEATURES OF THE PFPD ELECTROMETER BOARD Description J1: Electrometer input signal cable BNC connector J2: Photomultiplier tube high voltage cable TNC connector Function Photomultiplier signal output is directed to the electrometer board through this BNC connector. High voltage source on the electrometer board is directed to the photomultiplier tube through this TNC connector. J3: Ignitor cable Molex connector Current to the ignitor coil is directed through this Molex connector. SIGNAL (TP1): Test point TRIGGER (TP2): Test point GATE (TP3): Test point J4: Top Signal Trigger Gate Ground With an oscilloscope, used to set the Gate Width and Gate Delay. Used to monitor the gate trigger (threshold). Used to monitor when the integration period (Gate Width) begins. These are the same signals provided at TP1, 2 and 3, organized to provide easy connection to a cable. 1.3 THEORY OF OPERATION OF THE PFPD By limiting the flow rates of air and hydrogen to approximately one tenth those typically used in an FPD, the PFPD achieves substantial improvements in both detectivity and selectivity. At these low flow rates, a continuous flame cannot exist. Instead, the combustor and ignitor chambers fill with the combustible mixture of air and hydrogen. When this mixture enters the ignitor, the mixture ignites and a flame front propagates back through the combustor, terminating when the flame reaches the combustor support. The detector then refills with the combustible mixture and the process repeats :3

9 Introduction STAGES OF FLAME PULSATION The flame in the PFPD pulses because the flow rates of hydrogen and air will not sustain continuous combustion. Flame pulsation involves four stages: Fill: Air and hydrogen mix and enter the combustion chamber at two points. Part of the combustible gas stream mixes with the column effluent and moves upward through the interior of the quartz combustor tube. A second part of the gas stream sweeps along the outside of the quartz combustor tube and into the ignitor chamber. Ignite: The ignitor chamber contains a continuously heated ignitor coil. When the combustible gas mixture reaches the heated coil, the mixture ignites. Propagate: The flame front propagates downward into the combustion chamber. At the bottom of the combustion chamber, the flame extinguishes. During this propagation phase, the sample molecules in the flame are broken down into simpler molecules or atoms. Emit: During and after flame propagation, the sample atoms of interest undergo further reaction to form electronically excited species, from which light is emitted. The flame background emission is complete within less than 0.3 millisecond after propagation, whereas phosphorus and sulfur molecular species emit over a much longer time. This difference in emission times yields the enhanced detectivity and selectivity of the PFPD :3 1-5

10 Introduction STAGES OF PULSED-FLAME OPERATION HOW DOES THE PFPD ACHIEVE ITS GREATER DETECTIVITY AND SELECTIVITY? The enhanced detectivity and selectivity of the PFPD are a consequence of two main factors: Separation (in time) of the specific element emission from the background emission Concentration (in time) of the emitted light Background Emission in a Pulsed-flame Like the conventional FPD, the PFPD utilizes an air:hydrogen flame. Hydrogenrich flames favor a variety of gas phase chemical reactions, some of which yield molecular products that emit light (e.g., chemiluminescent reactions). Important excited combustion products include CH*, C2*, and OH*. The light from such combustion products is referred to as the background emission. Hydrogen-rich air:hydrogen flames typically yield chemiluminescence hydrocarbon background emission in the blue region with a maximum near 436 nm :3

11 Introduction Time-dependence of Emission Profiles An important advantage of pulsed-flame operation is the time-dependence of the resulting emission profiles. Flame pulsation enables the molecular emissions of sulfur- and phosphorus-containing compounds to be separated or resolved in time from the flame background emission itself. This resolution is possible because the emission time of the flame background is much shorter than the emission time of the excited molecular products of compounds of analytical interest. As an example, CH* and OH* emission in a hydrogen-rich flame occurs within the time it takes the flame to propagate through the combustor, typically 2 to 4 msec. In contrast, S2* emission reaches its maximum 5-6 msec after the CH*, and OH* emission has ceased. This separation in time enables the sulfur emission to be monitored when the flame background has dropped to a negligible level. The result is a great improvement in the detectivity and selectivity of sulfur and phosphorus detection. Emission Profiles The following figure shows the hydrocarbon and sulfur emission profiles as a function of time (Cheskis, S.; Atar, E.; Amirav, A. Anal. Chem. 1993, 65, 539) :3 1-7

12 Introduction Concentration (in time) of the Emitted Light In the PFPD, detectivity is greater because all the emission occurs during and immediately following the flame pulse. This increases the brightness of the analytical signal relative to the dark current and other electronic noise. For example, at a typical pulse rate of 2.5 Hz, all the sample eluting from the column during 400 msec is burned in a single pulse, emitting over perhaps 10 msec. This results in a signal which is not only approximately 40 times as bright as that from a continuous flame source, but also is resolved in time from the interference of the flame background. In addition to the general improvement in detectivity and selectivity as a result of flame pulsation, a further increase in the relative emissions of elements of interest is achieved through the appropriate choice of gas composition. As an example. the formation of S 2 * is favored in cooler hydrogen-rich air:hydrogen flames. Hydrogen content has less effect on phosphorus emission. Separation of the Element Emission from Background Emission The flame pulsation in the PFPD yields time as an added dimension to flame photometric detection. Both selectivity and detectivity are improved through the ability to: Resolve emissions of sulfur or phosphorus from the flame emission background Resolve emissions of sulfur and phosphorus from each other As mentioned, the emission times in an air:hydrogen flame of such excited molecular entities as S2* or HPO* are generally much longer than the emission times of species responsible for the background emission (e.g., CH*, C2*, and OH*). This delay allows the sulfur or phosphorus emission to be integrated after the flame background has dropped to a negligible level. The result is a lower overall noise level and thereby greater detectivity. The electrometer Gate Delay and Gate Width are used to select the time period of the emission that is to be integrated. In this way, only a small segment of the background noise is integrated during any pulsation. For example, if the time between pulses is 400 msec and the Gate Width is set to 20 msec, 95% of the total background noise is discarded, which further contributes to an improvement in detectivity :3

13 Introduction Effect of Gate Control The following figure illustrates the effect of Gate Delay on selectivity and detectivity. Adjustment of the Gas Composition The flow rates of air and hydrogen affect the detectivity and selectivity of the PFPD. The total flow rates of air and hydrogen set the operating frequency (rate of pulsation) of the PFPD. Typical operating frequency for the PFPD is near 2-4 pulses/second. The relative proportions of air and hydrogen influence the selectivity by favoring the formation of an emitting species or suppressing the formation of an unwanted interfering species. For example, when detecting phosphorus pesticides in plant tissues, the use of a slightly more air rich flame than usual will suppress formation of the interfering S 2 * molecule, improving P/S selectivity :3 1-9

14 Introduction FLOW PATH OF THE PFPD The figure below shows the flow path of gases entering and exiting the PFPD. The air:hydrogen mixture enters the PFPD at two points: around the outer wall and through the center of the quartz combustor tube. The wall flow is used to refill the upper ignitor chamber of the detector after each pulse, while the combustor flow mixes with the column effluent and refills the combustor. For proper operation, the combustor should refill just slightly faster than the ignitor. The refill rates are balanced by use of a needle valve on the air:hydrogen mixture, and by the flow of air2, which bypasses the combustor and goes directly to the ignitor REFERENCES Refer to the following articles for additional information on the PFPD. Cheskis, S.; Atar, E.; Amirav, A. Anal. Chem. 1993, 65, 539. Amirav, A.; Jing, H. Anal. Chem. 1995, 67, :3

15 Introduction 1.4 SPECIFICATIONS OF THE PFPD Specification Sulfur Performance Phosphorus Performance Detectivity 1 pg S/sec 0.1 pg P/sec Selectivity 10 6 S/Carbon NOTE: The selectivity of Sulfur/Carbon is a function of the sulfur concentration at any time (from 10 5 near its detection limit to 10 6 at higher sulfur concentrations) P/Carbon Linear Dynamic Range Drift (in 20 minutes) 10 x peak-peak noise 10 x peak-peak noise 1.5 NITROGEN OR OTHER ELEMENT OPERATION The nitrogen mode enhancement to the Pulsed Flame Photometric Detector extends the range of wavelengths available for detection of element-specific emissions to 900 nm in the near infrared. This not only allows detection of the nitrogen emissions between 700 nm and 800 nm, but also permits the detection of other elements, such as tin and arsenic. Section 4 of this manual gives details of operation in the nitrogen mode, as well as some preliminary information on detection of additional elements. Nitrogen Mode Specifications Detectivity: 20 pg N/seconds Selectivity: >10 4 N/C Linear Dynamic Range: >10 2 Drift: <10X peak to peak noise in 20 minutes :3 1-11

16 2. Section 2 Setting Up the PFPD 2.1 OVERVIEW Topic See Page... Inspecting the PFPD Contents of the PFPD Accessory Kit Contents of the Diffusion Source Kit INSPECTING THE PFPD The PFPD is installed and tested at the factory before the GC is shipped. Carefully inspect the PFPD. If there is evidence of damage or if parts are missing from the PFPD or its accessory kit, contact: Customer Service Varian Analytical Instruments 2700 Mitchell Drive Walnut Creek, CA (925) Outside of the USA, notify the nearest International Sales Office listed at the beginning of this manual :3 2-1

17 Setting Up the PFPD 2.3 CONTENTS OF THE PFPD ACCESSORY KIT The PFPD accessory kit (P/N ) contains tools to disassemble the PFPD and a supply of parts used in its routine maintenance. Item Part Number Qty Description Quartz combustor (2 mm ID, Sulfur) Combustor support (2 mm ID, Sulfur) Teflon combustor extraction device (wide bore, 3 mm ID) Teflon combustor extraction device (narrow bore, 2 mm ID) Combustor support wrench Teflon tweezers Allen wrench, 9/64" (cleaned) Crunch washers (Aluminum) Combustor support seal (Aluminum) Filter assembly, Sulfur Teflon pipe thread tape, not shown Reducing union (flow check) Please note that Item 13, Sulfur Linearization Software, is not used with the 3800 PFPD :3

18 Setting Up the PFPD 2.4 CONTENTS OF THE PFPD DIFFUSION SOURCE KIT The PFPD diffusion source kit (P/N ) is supplied as an option to the PFPD. Item Part Number Qty Description Adapter seal Permeation tube, Sulfur Diffusion source holder Spring Injector adapter Teflon pipe thread tape, 1/2" (not shown) :3 2-3

19 3. Section 3 Operation 3.1 OVERVIEW Topic See Page... Installing a Chromatographic Column Adjusting the Gas Flow Rates Calibrating the Flow Controllers Optimizing the PFPD Testing the PFPD Performance Using the User Defined Mode Building a GC Method Building a GC Workstation Method Quantitating in the Sulfur Mode INSTALLING A CHROMATOGRAPHIC COLUMN The following procedure describes the installation (or reinstallation) of a gas chromatographic column in a Varian GC equipped with a PFPD. To ensure good chromatographic performance with the PFPD, the tip of the column must extend mm above the top of the combustor support COLUMN INSTALLATION Item Description Capillary Column Nut Capillary Column Ferrule Correct Column Cut Incorrect Column Cut :3 3-1

20 Operation Tools and Equipment Needed These tools are needed to complete the following procedure: Tungsten carbide pencil (P/N ) or a scoring tool Magnifying lens (P/N ) PREPARE THE COLUMN END Follow these steps to prepare the end of the chromatographic column for insertion into the PFPD. 1. Before cutting the end of the column, slide the capillary column nut (Item 1) over the fused silica column end (Item 2). 2. Install the appropriate Vespel or graphite ferrule (Item 3) over the column end. Make certain that the tapered edge of the ferrule is toward the inside of the column nut (or away from the column end). NOTE: Typically, graphite ferrules contain sulfur as a contaminant. If you use graphite ferrules, expect a higher background during preconditioning of the chromatographic column. The background may be particularly high when using a new graphite ferrule that has not be used and conditioned in the past. 3. Hold the column securely between the thumb and forefinger of one hand. Use the scoring tool to score the column once lightly about 1 to 2 cm from the end of the column. 4. With your thumbs and forefingers, grasp the column on each side of the score and bend the column carefully away from the score mark until it breaks at the score mark. 5. Examine the cut with the magnifying lens to decide whether the column was properly cut (Item 4) or improperly cut (Item 5). If you observe small splinters of silica or bits of the outer coating on the column end, repeat the above procedure and make a fresh cut. If you observe a clean cut, you may use either of two techniques to install the column. If you wish to visually confirm the position of the column end above the combustor support, go to Disassemble the PFPD. If you would rather not disassemble the detector, go to To Install a Column Without Disassembling the Detector :3

21 Operation PFPD ASSEMBLY Refer to paragraph (page 1-2) when disassembling the PFPD, inserting the column, and reassembling the PFPD DISASSEMBLE THE PFPD These tools are needed to complete the following procedure: Phillips-head screwdriver 9/64" Allen wrench Metal forceps Teflon combustor extraction device Teflon tweezers Lab tissues To Disassemble the PFPD 1. Turn the column oven OFF to stop the oven fan. 2. Open the column oven door and wait for the column to cool to room temperature (or to a temperature at which you can comfortably handle the column). 3. Turn OFF the PFPD electronics. 4. Remove the detector cover from the top of the GC. 5. Before using, thoroughly clean the 9/64" Allen wrench, 1/4" wrench, Teflon combustor extraction device, and Teflon tweezer with methanol or isopropanol. Set these tools aside on a clean lab tissue. NOTE: Dust and dirt inside the combustor can interfere with emission and thereby lower the sensitivity of the PFPD. Therefore, never touch the internal parts of the PFPD with bare hands. Always use a clean lab tissue, Teflon combustor extraction device, or Teflon tweezer to remove or replace detector parts. Place parts on a clean lab tissue as they are removed from the PFPD. 6. Use the 9/64" Allen wrench to remove the 2 Allen screws (Item 1) from the top of the ignitor cap (Item 2). Place the Allen screws on a clean lab tissue :3 3-3

22 Operation WARNING: BURN HAZARD The ignitor cap and Allen screws may be very hot. Wait until the ignitor cap and Allen screws have cooled or use metal forceps to remove the Allen screws from the ignitor cap. 7. Use the metal forceps to lift the ignitor cap from the detector tower (Item 3). Place the ignitor cap bottom up on a clean surface (e.g., a clean lab tissue). The top of the quartz combustor (Item 4) is visible in the center of the detector tower. WARNING: BURN HAZARD The ignitor cap may be very hot. Wait until the ignitor cap has cooled or use metal forceps to remove the ignitor cap from the detector body. 8. Gently insert the appropriate Teflon combustor extraction device into the opening in the top of the combustor. Wait 2-3 seconds for the Teflon combustor extraction device to seat, then lift the quartz combustor from the combustor support (Item 5). Place the quartz combustor on a clean surface (e.g., a clean lab tissue). 9. Use the metal forceps to remove the top aluminum crunch washer (Item 6) from the detector tower. Discard this used crunch washer INSERT THE COLUMN These tools are needed to complete the following procedure: Phillips-head screwdriver 5/16" wrench Capillary pick or combustor support seal extraction tool Ceramic scoring tool (P/N ) Magnifying lens (P/N ) 3 mm combustor extraction tool/column Depth Gauge (P/N ) :3

23 Operation To Insert the Column 1. Check that the capillary column nut and the ferrule (reversed) are installed on the column as described in the manual. Check that approximately 20 cm of the detector end of the column is free so that it can be inserted into the detector. 2. Insert the Column Depth Gauge into the hole from which the combustor was removed. Insert the column into the detector and push the column up until it begins to push up the depth gauge. Tighten the column nut gently until a small amount of force is required to move the column. Gently apply force to the depth gauge to push the column down until the depth gauge contacts the detector tower. See the following figure. Tighten the column nut to secure the column in this position. Depth Gauge Column Combustor Support :3 3-5

24 Operation REASSEMBLE THE PFPD Tools and equipment needed to complete the following procedure: Phillips-head screwdriver 9/64" Allen wrench Teflon combustor extraction device Metal forceps To Reassemble the PFPD 1. Use the clean metal forceps to place a new bottom aluminum crunch washer (Item 12) on the detector base. 2. Hold the PMT assembly (Item 11) with one hand and the detector tower with the other. Sight the top of the combustor support (Item 5) through the detector tower. Keep the combustor support centered as you lower the detector tower (Item 3) over the support. CAUTION Be very careful not to crush the end of the column. 3. Reinstall the 2 Allen screws (Item 7) that hold the detector tower to the detector base (Item 8). Tighten the screws alternately. 4. Use the Teflon combustor extraction device (of the correct taper) to insert the quartz combustor (Item 4). 5. Use the clean metal forceps to place a new top aluminum crunch washer (Item 6) on the detector tower. 6. Replace the ignitor cap (Item 2) on the detector tower. Use the 9/64" Allen wrench to reinstall in the top of the ignitor cap the 2 Allen screws (Item 1) that secure the ignitor cap to the detector tower. 7. Reinstall the detector cover on the top of the GC. 8. Turn ON the PFPD electronics and PFPD Oven Power. Be sure that the detector gases are turned on. The detector may not begin pulsing normally until the detector tower has begun to heat up somewhat. Allow the detector to come to temperature and equilibrate for at least 15 minutes. During this equilibration period the polyimide coating on the portion of the column exposed to the flame is burned off, leaving a clean fused silica surface :3

25 Operation 9. Did you install a quartz combustor with a different diameter? If Yes, go to Optimizing the PFPD for instructions on optimizing the performance of the PFPD with the quartz combustor just installed. If No, the PFPD is ready for operation. 3.3 ADJUSTING THE GAS FLOW RATES The following procedure describes the adjustment of the flow rates of the gases to the PFPD. If you frequently adjust the column flow rate or change the carrier gas, you may want to calibrate the air and hydrogen flow controllers. Once the flow controllers are calibrated, you can quickly select new flow rates and use the optimization tables that are provided. If the PFPD detector EFC option is installed on your 3800 GC, detector gas flows are set by pressing the Detector key, selecting the PFPD, then pressing the Adjustments softkey. Air1, hydrogen and Air2 are set on page 2 of the Adjustments. The Split Valve adjustment is described below SPLIT VALVE A split valve is used to finely adjust the flow rate through the combustor section of the PFPD combustor. Follow these directions when adjusting this split valve: 1. Disconnect the outlet line from the needle valve and replace with a flowmeter. 2. With Air 1 and Hydrogen turned on, close the needle valve by turning the knob fully clockwise. The flowmeter should indicate a flow of less than 1 ml/minute. If this is the case, disconnect the flowmeter and reattach the outlet line. 3. If the flow from the needle valve exceeds 1 ml/minute, remove the knob by loosening the set screw. Rotate the valve stem clockwise gently until the flow from the outlet drops below 1 ml/minute. Do not overtighten the valve stem, as the needle valve does not need to shut off completely to function properly, and overtightening can damage the delicate needle in the valve. After making the adjustment, reinstall the knob on the valve stem and secure with the set screw :3 3-7

26 Operation PFPD GAS FLOW DIAGRAM Refer to the following figure when adjusting the flow rates of the gases to the PFPD. Item Description 5 Split valve 1 Air-1 mass flow controller 6 Wall flow (H 2 + Air-1 + Air-2) 2 Air-2 mass flow controller 7 Center flow (H 2 + Air-1) 3 Center flow tee 8 Filter 4 Hydrogen mass flow controller 9 Ignitor cap TO ADJUST THE GAS FLOW RATES Tools and equipment needed to complete the following procedure: Varian Intelligent Digital Flow Meter (P/N ) or Soap flow meter (P/N ) Stopwatch Flow measuring adapter (1/8"-to-1/16" Swagelok reducing union) 1/4" flat blade screwdriver :3

27 Operation NOTE: For best stability of flows, flow controllers should have pressure applied to them for at least 24 hours before adjustments are made. A moderate flow of gas through the flow controller during this conditioning period is also helpful in stabilizing the flow controller. 1. Turn the ignitor OFF. 2. Locate the Air-1 mass flow controller valve (Item 3). Note the current dial setting. Turn Air-1 OFF. 3. Locate the Air-2 mass flow controller valve (Item 1). Note the current dial setting. Turn Air-2 OFF. 4. Locate the Hydrogen mass flow controller valve (Item 4). Note the current dial setting. Turn Hydrogen OFF. 5. Install the flow measuring adapter (Swagelok 1/8" to 1/16" reducing union). Use Teflon thread tape to make the connection to the ignitor cap gas tight. Measure and record the column flow rate, then turn the column flow OFF. NOTE: A Varian Intelligent Digital Flowmeter (P/N ) is highly recommended for this procedure. IMPORTANT: Be sure the ignitor seal is completely leak tight before attempting to measure flow rates. This seal can loosen due to temperature cycling during normal operation. This does not affect detector performance, but can lead to serious errors in setting flow rates. Tighten securely before beginning to measure flows. 6. Adjust the Air-1 mass flow controller valve until flow rate is near 15 ml/min. Write down and save the dial setting for the Air-1 mass flow controller valve. Turn the Air-1 flow controller Off. 7. Adjust the Air-2 mass flow controller valve until flow rate is near 15 ml/min. Write down and save the dial setting for the Air-2 mass flow controller valve. Turn the Air-2 flow controller Off :3 3-9

28 Operation 8. Adjust the Hydrogen mass flow controller valve until flow rate is near 12 ml/min. NOTE: If you will be using hydrogen as your carrier gas, the total hydrogen flow to the detector should be approximately 12 ml/minute. The flow rate set in this step should thus be 12 ml/min minus the anticipated column flow rate. Write down and save the dial setting for the Hydrogen mass flow controller valve. 9. Restore the Air-1 and Air-2 flow controllers to the settings recorded above. Restore the carrier gas flow to the column. Remove the flow measurement adapter, remove any Teflon thread tape residue from the ignitor cap. NOTE: Leave the flow controllers ON at all times unless you must turn them off to re-plumb or reconfigure the gas chromatograph. If you turn the flow controllers off for an extended period then on again, they must equilibrate overnight before the PFPD can achieve optimum performance. 10. Turn the ignitor ON. 3.4 CALIBRATING THE FLOW CONTROLLERS NOTE: For best stability of flows, flow controllers should have pressure applied to them for at least 24 hours before adjustments are made. A moderate flow of gas through the flow controller during this conditioning period is also helpful in stabilizing the flow controller. Tools and equipment needed to complete the following procedure: Varian Intelligent Digital Flow Meter (P/N ) or Soap flow meter (P/N ) Stopwatch Flow measuring adapter (1/8"-to-1/16" Swagelok reducing union) 1/4" flat blade screwdriver To Calibrating the Flow Controllers 1. Turn off the column flow. Or, if you want flow to continue to the column, disconnect the column from the detector and use a column nut and 1/16" nohole ferrule (P/N ) to seal the detector. 2. Write down the current dial settings for Air-1, Air-2, and Hydrogen, then turn off each flow controller. 3. Install the flow measuring adapter (1/8" to 1/16" reducing union). Use Teflon thread tape to make a gas-tight seal :3

29 Operation NOTE: A Varian Intelligent Digital Flowmeter (P/N ) is highly recommended for this procedure. IMPORTANT: Be sure the ignitor seal is completely leak tight before attempting to measure flow rates. This seal can loosen due to temperature cycling during normal operation. This does not affect detector performance, but can lead to serious errors in setting flow rates. Tighten securely before beginning to measure flows. 4. Connect the flowmeter to the adaptor. Turn the Air-1 flow controller (clockwise) until the flowmeter indicates flow through the system. Slowly turn the Air-1 flow controller (counter-clockwise) until the flow just stops. Again, write down the dial setting. NOTE: It is recommended that the flow controller dial be set so that it reads zero at this setting. Remove the knob then loosen the two set screws holding the dial to the flow controller shaft. While holding the flow controller shaft so that it does not rotate, turn the dial to 000. Retighten the set screws and replace the knob. 5. Set the dial to 100, 200, etc., and allow the flow to equilibrate. Measure the flow and record. When you have completed all measurements, turn off flow. 6. Repeat steps 1-5 for the Air-2 and Hydrogen flow controllers. NOTE: Chart the flow rate versus dial setting so that you have a reference when setting flow rates as suggested in Optimizing the PFPD on Page If you removed the column prior to calibrating the flow controllers, reinstall it now. 3.5 OPTIMIZING THE PFPD The PFPD is optimized to maximize: Detectivity of Sulfur or Phosphorus Selectivity of Sulfur and Phosphorus to Carbon or each other Analytical throughput (shorten total analysis time) :3 3-11

30 Operation DETECTIVITY Detectivity is the minimum detectable quantity (MDQ) of Sulfur or Phosphorus (usually expressed as femtograms of Sulfur (or Phosphorus) per second. Detectivity is the preferred method of expressing the sensitivity of a detection method because it includes a consideration of the Signal-to-Noise ratio. Sensitivity alone is insufficient in that the sensitivity of the PFPD can be increased simply by increasing the voltage of the photomultiplier tube, but the noise increases as the signal increases, yielding little enhancement in the minimum amount detectable SELECTIVITY An important question when optimizing the selectivity of the PFPD is, "Selective relative to what?." In some situations, the Sulfur or Phosphorus response is best optimized relative to Carbon (the principal background emission in the airhydrogen flame). At other times, it may be more important to optimize the Phosphorus response relative to the Sulfur response, as when Phosphorus is analyzed in certain plant tissue extracts that are high in Sulfur content THROUGHPUT The additional selectivity of the PFPD can result in shorter analysis times and thereby greater sample throughput. As an example, by optimizing the sulfur response relative to hydrocarbons, the PFPD may detect fewer peaks thereby simplifying the chromatogram and increasing the chromatographic resolution. With greater resolution, the chromatography can be adjusted such that the peaks of interest elute faster thereby decreasing the analysis time. In addition, when hydrogen is used as the carrier gas, the PFPD can be operated at higher flow rates without greatly affecting the detectivity of the method. Higher carrier gas flow rates translates into shorter analysis times and greater sample throughput SUGGESTED FUEL GAS FLOW RATES The following table lists flow rates for several carrier gases that are often used in common analysis situations when operating the PFPD. Use these flow rates as starting points when optimizing the performance of the PFPD :3

31 Operation Carrier Gas Flow Rate (ml/min) H 2 (ml/min) Air-1 (ml/min) Air-2 (ml/min) Helium Hydrogen Nitrogen You May Not Need to Adjust Flow Rates The difference in optimum gas flow rates between sulfur and phosphorus is small. If you don t require the maximum sensitivity for an analysis, leave the flow rates as they are and change only the optical filter, gate delay, and integration time, when shifting between the sulfur and phosphorus detection modes SUGGESTED PFPD ADJUSTMENT SETTINGS The following adjustment settings are recommended for setup and routine operation of the PFPD. These adjustments are accessed by pressing the Adjustments softkey in the PFPD detector section of the 3800 method. Element Gate Delay Gate Width Trigger Sulfur Phosphorus Nitrogen OPTIMIZATION CRITERIA Refer to the table below when optimizing the PFPD :3 3-13

32 Operation To Maximize... Consider... Phosphorus response Install wide bore quartz combustor Use moderately hydrogen-rich flame Adjust split valve such that detector is just out of tick-tock Set detector temperature 300 C Sulfur response Install narrow bore quartz combustor. Phosphorus-to- Carbon selectivity Set hydrogen carrier gas flow rate to a value that is near, but just above, the minimum of the van Deemter curve for the column and maximum column temperature used. Adjust split valve such that detector is just out of tick-tock Use optimum hydrogen-rich flame Increase Air-2 to enhance ignition and flame propagation in the hydrogen-rich flame Set detector temperature 200 C NOTE: For some compounds, sulfur response decreases with increasing detector temperature. Set the detector temperature as low as possible while preserving chromatographic peak shape. Install wide bore quartz combustor Use hydrogen as carrier gas Use moderately hydrogen-rich flame Adjust split valve such that detector is just out of tick-tock, then turn counterclockwise an additional 1/2 turn. Detector temperature 300 C NOTE: The Gate Delay can be adjusted to increase or decrease selectivity. However, as the selectivity increases, the detectivity often decreases. Set the Gate Delay to balance these performance criteria at levels that are appropriate to the application. Also, the gate should be set to open (or trigger) as close to the main peak as possible rather than before or after. Setting the gate correctly prevents the hydrocarbon and phosphorus responses from overlapping as the flame propagation slows in hydrocarbon-rich flames :3

33 Operation To Maximize... Consider... Sulfur-to-Carbon selectivity Install wide bore quartz combustor Use hydrogen as carrier gas Use moderately hydrogen-rich flame Adjust split valve such that detector is just out of tick-tock, then adjust counterclockwise an additional 1/2 turn. Set detector temperature ~200 C Increase sample amount injected (inject larger volume or more concentrated form of sample) The trigger level should be set to trigger on the main flame peak. Setting the trigger level correctly prevents the hydrocarbon and sulfur responses from overlapping as the flame propagation slows in hydrocarbon-rich flames. The Sulfur response is proportional to the square of the Sulfur concentration in the flame whereas the Carbon response is directly proportional to the Carbon concentration. Therefore, when the amount of sample that is injected is doubled, the Sulfur-Carbon selectivity doubles (all other parameters being equal). Phosphorus-to- Sulfur selectivity Install wide bore quartz combustor Use hydrogen as carrier gas Use slightly hydrogen-rich flame. Increasing Air1 flow until phosphorus response decreases by a factor of two will improve P/S selectivity greatly. Adjust split valve such that detector is just out of tick-tock Set detector temperature ~300 C Inject the smallest amount of sample consistent with adequate phosphorus signal. Sulfur interference is reduced in direct proportion to the reduction in sample size. For example, a five fold reduction in sample size will result in a 25 fold reduction in sulfur interference response, but only a five fold reduction in phosphorus response, for a net enhancement in selectivity of a factor of :3 3-15

34 Operation To Maximize... Consider... Decrease Gate Delay and Gate Width as much as possible while maintaining an acceptable Carbon selectivity. NOTE: Because phosphorus emits sooner than Sulfur, a shorter Gate Delay and Gate Width improves the Phosphorus-Sulfur selectivity. Also, because air-rich flames favor Phosphorus emission over Sulfur emission, decrease the hydrogen-to-air ratio to increase the Phosphorus-Sulfur selectivity. Analytical throughput (decrease analysis time) Install wide bore quartz combustor. Use hydrogen as the carrier gas at flow rates up to 10 ml/min. Bracket only the region of the chromatogram that the PFPD can detect and for which the detector conditions have been optimized. NOTE: Because the van Deemter curve for hydrogen is flat at very high flow rates, there is little loss of efficiency as the flow rate is increased to shorten the analysis time. The wide bore quartz combustor permits the use of higher flow rates, but this may not be a good choice if maximum Sulfur response is also the objective of the application. Effect of PFPD temperature An initial detector temperature of 200 C is recommended when optimizing the PFPD, but your sample may require a higher detector temperature to adequately separate and detect higher boiling sample components. The figure below shows the effect of increasing the PFPD temperature by 100 C. The sample eluted from the column at a column temperature of 260 C. At a detector temperature of 200 C, the peak is broad with a slight shoulder. When the detector temperature was increased to 300 C, the peak sharpened and the shoulder disappeared :3

35 Operation To Maximize... Consider ESTABLISHING TICK-TOCK Tick-tock occurs when the combustible mixture ignites in the ignitor chamber but does not propagate into the combustor chamber because the combustor has not yet filled with a combustible gas mixture. Optimum performance is generally achieved when the detector is operating near, but not in, the tick-tock mode. Follow these steps to adjust the gas flow rates for optimum performance of the PFPD. This adjustment should be done with the column oven at the highest temperature which will be used in the analysis. This will assure that the detector will remain in the normal operating mode throughout the column temperature program. 1. To monitor the detector baseline, connect a strip chart recorder to the detector analog output. Turn off the AutoZero function (in the Detector section of the GC method). Set the Gate Delay to 4 msec and the Gate Width to 10 msec. Leave the trigger level at the default value, 200 mv. (This makes Tick-Tock easier to observe. Remember to switch back to the desired gate delay and width settings after setting the needle valve.) :3 3-17

36 Operation 2. Adjust the gas flow rates so that the PFPD is pulsing at about 2-4 Hz (pulses/second). 3. Adjust the Air-Hydrogen needle valve fully clockwise. This shuts off the flow of air-hydrogen to the combustor chamber while permitting the air-hydrogen mixture to pass around the outside of the combustor and enter the ignitor chamber. The pulse rate should increase. As the flame stops propagating into the combustor, the detector baseline drops and becomes smoother (less noisy). NOTE: At low column flow rates (particularly with the 3 mm combustor), there may be enough diffusion of fuel mixture into the combustor chamber that the baseline is not completely smooth (noiseless). 4. Slowly turn the Air-Hydrogen needle valve counter-clockwise. With each turn of the needle valve, more of the combustible mixture enters the combustor chamber. As the amount of combustible mixture entering the combustor chamber increases, the flame begins to propagate into the combustor chamber. First, the flame propagates into the combustor chamber on every third or fourth pulse and then, as the flow rate of the combustible mixture increases, on alternate pulses (tick-tock mode), and then finally on every pulse. As you increase the flow rate of combustible mixture that enters the combustor, the baseline should rise sharply and become very noisy as the detector enters tick-tock. As the detector leaves the tick-tock mode, the baseline will rise and then fall and become less noisy. The pulse rate will also decrease as you pass through tick-tock, since more of the combustible mixture is burned on each pulse as you adjust the needle valve for normal operation :3

37 Operation :3 3-19

38 Operation 5. Once the baseline reaches a steady value, you may adjust the needle valve an additional 1/2-turn counter-clockwise to ensure stable operation of the PFPD. If you changed the gate delay and width, reset them to the proper values for your analysis PHOTOMULTIPLIER TUBE (PMT) HIGH VOLTAGE The PFPD detector is shipped with the electrometer and detector set up for the mode specified or the phosphorus mode, if not specified. The PMT high voltage is set at 510 volts typically for phosphorus. For the sulfur mode, the recommended voltage is somewhat higher, usually 550 volts. For the nitrogen mode, volts is typical ADJUSTING THE PHOTOMULTIPLIER TUBE (PMT) VOLTAGE The following procedure describes how to adjust the PMT voltage when changing from one mode to the other, or to compensate for aging of the PMT over time. 1. If you have not already done so, adjust the detector gas flow rates for the mode of operation desired (Sulfur or Phosphorus). Use flow rate values that you determined previously or the recommended flow rates listed in the Gas Flow Rates table on Page When the detector is pulsing normally, press the Detector key on the GC control panel. Select Front, Middle, or Rear (the position in which the PFPD is installed). Press Enter, then Page Down until the PMT voltage entry field is displayed. 3. If the PMT voltage is: within volts of the nominal value for the desired mode of operation, you may not need to make any further adjustment. not within this range, enter a new PMT voltage while observing the PFPD output on a strip chart recorder or on the Workstation System Control display. 4. If you want to optimize the detector sensitivity with respect to PMT voltage: Set up the PFPD to operate in the desired mode. Slowly adjust the HV setting while watching the baseline noise (displayed on a strip chart recorder or other output device) :3

39 Operation When the noise approaches the values noted below, begin making injections of test sample and calculating signal to noise ratio for the peaks of interest. Increase the PMT voltage in increments of 20 to 50 volts until the signal to noise ratio no longer increases, or until the noise becomes unacceptably large. NOTE: For the sulfur mode, the noise should be about 1% of full scale (for Attenuation = 1, Range = 10). For the phosphorus mode, the noise should be about 1% of full scale (for Attenuation = 4, Range = 10). 5. Note the HV value displayed on the GC Detector Status display as above. NOTE: As the PMT ages, the voltage required to generate adequate sensitivity gradually increases. Replace the PMT when you can no longer achieve adequate sensitivity within the range of HV available OPTIMIZING THE PFPD The PFPD can be optimized in two ways: Repeated injections of standard Continuous infusion of standard Tools and Equipment Needed These tools are needed to complete the procedure for optimizing the PFPD using the method of repeated injections of standard: Flat tipped screwdriver Printer/plotter or stripchart recorder FPD test sample (P/N ) 10 µl syringe (P/N ) :3 3-21

40 Operation REPEATED INJECTIONS OF STANDARD Follow these steps to optimize the PFPD using the method of repeated injections of a standard. 1. Install the test column. 2. Set the Detector temperature to 200 C. Set injector temperature to 200 for isothermal injectors; 80 initial, 0 hold, program to 200 for Set column oven to 80 initial, hold 0.5 minutes, then program at 25 /minute to 200, and hold 2 minutes, or until the last peak of interest elutes.. 3. Allow the chromatographic system to equilibrate for 1 hour. 4. Turn the stripchart recorder or other display device ON. 5. Adjust the gas flow rates to those recommended for your carrier gas and flow rate in the Suggested Fuel Gas Flow Rates table on Page If necessary, adjust the Air-2 flow rate to give a flame pulse rate of 2-4 pulses/second. 7. Inject 1.0 µl of the FPD Test Sample split ~20:1 if using a 1079 split injection. Inject 1 µl of a 20:1 dilution of the test sample in iso-octane if using a 1079 splitless or on-column injection or a 1041/1061 injector. 8. Note the peak height of the sample peak eluting last, methyl parathion. When the GC is Ready again, increase the hydrogen flow by 20 digits on the counter, inject the sample again, and note whether the peak height increases. If so, continue to increase the hydrogen flow and reinject the sample until the peak height no longer increases. If the peak height does not increase by increasing the hydrogen flow, then decrease the flow by 20 digits on the counter and reinject the sample. Continue this procedure until the methylparathion peak is maximized. NOTE: Allow the system to equilibrate for 2-3 minutes between flow rate adjustments so that the air and hydrogen filters are completely flushed with the new gas composition. 9. When the peak height is maximized, tune the fill time of the ignitor volume to be just slightly greater than the fill time of the combustor volume. See Establishing Tick-Tock on Page :3

41 Operation Tools and Equipment Needed These tools are needed to complete the procedure for optimizing the PFPD using the method of continuous infusion of standard: Printer/plotter or stripchart recorder Sulfur diffusion source kit (P/N ) CONTINUOUS INFUSION OF STANDARD Follow these steps to optimize the PFPD response using the method of continuous infusion of a standard. 1. Install the Sulfur diffusion source. Disconnect the column at the injector end. Insert the sulfur permeation tube (P/N ) into the diffusion source holder (P/N ), wrap the threads of the holder with Teflon thread sealant tape and attach the injector adaptor. Tighten the cap just enough to make a gas-tight seal. Insert an aluminum adaptor seal (P/N ) in the end of the injector adaptor. Check that the crunch washer is properly seated at the bottom of the holder and not resting on the threads. Screw the injector adaptor-end of the diffusion source holder onto the injector outlet fitting. Tighten the holder just enough to make a gas-tight seal. NOTE: If the permeation tube has not been used recently, set the injector and column temperature to 50 C and pass carrier gas over the permeation tube for ~1/2 hour. This decreases the surface concentration of the permanent. 2. Install the test column. If the test column is installed, shorten the length of column extending above the injector nut to ~5 mm. Insert the column into the column-end of the diffusion source holder. Tighten the column nut as usual. The outlet of the holder uses the same fitting as the injector outlet except that the holder requires an insertion depth of only 1-2 cm :3 3-23

42 Operation 3. Set the Injector and Column temperatures to 60 C. NOTE: The concentration of standard that enters the detector is adjusted by increasing or decreasing the Injector and Column temperatures. However, 60 C should be near optimum for the optimization procedure. 4. Adjust the column flow rate as required (~2 ml/min for the 320µ test column). Allow the chromatographic system and permeation tube and diffusion source holder to equilibrate for 1 hour. 5. Adjust the gas flow rates to the recommended flow rates listed in the Gas Flow Rates table of Optimizing the PFPD on Page Set the detector range to 10. Set Autozero to OFF. Adjust the attenuation to keep the emission on scale. 6. If necessary, adjust the Air-1 and Hydrogen flow rates to give a flame pulse rate 2-4 pulses/second. Alternately increase or decrease Air-1 and Hydrogen so that the flow rate remains approximately the same as the gas composition is changed. Allow the system to equilibrate for 2-3 minutes between flow rate adjustments so that the air and hydrogen filters are completely flushed with the new gas composition. NOTE: When adjusting the flow rate, turn the Air and Hydrogen flow controller dials in increments of 10 or Increase the flow rate of Hydrogen by 10 dial units. Examine the emission on the stripchart recorder. Monitor the emission after each flow rate adjustment. If the sulfur emission increases, continue to increase the hydrogen flow rate until the response no longer increases. If the pulse rate increases beyond 4 Hz, reduce the Air-1 flow by 10 units and continue the procedure. Typically, noise is relatively independent of gas composition so that the relative gas composition can be adjusted with little or no effect on the noise. NOTE: Sulfur emission is maximized when the gas mixture is very rich in hydrogen. 8. Once the sulfur emission has been maximized, adjust the fill time of the ignitor volume so that it is just slightly less than the fill time of the combustor volume. See Establishing Tick-Tock on Page Remove the column from the diffusion device holder, remove the diffusion device holder from the injector outlet, and reinstall the test column or the appropriate analytical column in the injector :3

43 Operation 3.6 TESTING THE PFPD PERFORMANCE The following procedure describes how to test the performance of the PFPD after it has been optimized for the desired detectivity and selectivity. Tools and Equipment Needed These tools are needed to complete the following procedure: FPD test sample (P/N ) 10 µl syringe (P/N ) Test column, DB-1 320µ x 4M (P/N ) TEST THE PERFORMANCE OF THE PFPD Follow these steps to test the chromatographic performance of the PFPD. 1. Configure the GC as described below: Column: Test column (DB-1 320µ x 4M) Injector: 1079 (preferred) Stripchart recorder 2. Select the mode (sulfur or phosphorus): Turn the PFPD electrometer OFF Remove the GC top covers Install the correct optical filter for the element of interest. Set the Gate Delay and Gate Width for the element of interest. Install a 2mm quartz combustor and support for sulfur mode, or a 3mm combustor and support for phosphorus. Replace the GC top covers and turn the PFPD electrometer ON. 3. Set the GC method (or GC Workstation method) for the following bake-out conditions: Oven temperature: 240 C Detector temperature: 350 C Injector temperature: 250 C :3 3-25

44 Operation Disconnect the test column, cap off the detector, and operate the GC for at least 4 hours to bake out the column. Set the oven temperature to room temperature and wait until the oven and column are cool. Reconnect the test column to the detector. 4. Set the GC method (or GC Workstation method) for the following test conditions: Injector: 200 isothermal or 80 initial, 0 hold, program to 200 for 1079 modes Detector: Temperature 200, Range 10, Autozero ON Column Oven: 80 initial, 0.5 min. hold, 25 /minute to 200, hold 2 minutes, or until last peak of interest elutes. Set PMT voltage to 530V for phosphorus or 570V for sulfur. 5. Operate the GC for at least 1 hour to ensure that the PFPD and its electronics are thermally stable. Re-optimize the PFPD as recommended in Optimizing the PFPD, Page 3-11). 6. If you are using a split injection, inject 1 µl of FPD test mix with a split ratio of 20:1. If you are using a non-splitting injector, dilute the FPD Test Sample 1:20 in iso-octane, and inject 1.0 µl of the diluted FPD Test Sample. 7. The resulting chromatographic peaks should have similar retention times, heights, and resolution to those shown below. 8. Are the peak retention times substantially different? If Yes, adjust the column temperature and/or column flow rate and repeat steps 6-8. If No, replace the test column with the column to be used in your particular application, adjust the GC conditions to appropriate values, and proceed with the analysis :3

45 Operation PFPD TEST CHROMATOGRAM: SULFUR PERFORMANCE Refer to the figure below to check the chromatographic performance of the PFPD for sulfur. Compare the test chromatogram to the chromatogram shown below in terms of peak retention times, peak heights and peak resolution :3 3-27

46 Operation PFPD TEST CHROMATOGRAM: PHOSPHORUS PERFORMANCE Refer to the figure below to check the chromatographic performance of the PFPD for phosphorus. Compare the test chromatogram to the chromatogram shown below in terms of peak retention times, peak heights and peak resolution DETECTIVITY Refer to the instructions below to calculate the detectivity for sulfur and phosphorus with the PFPD: Determine the peak-to-peak noise (Np-p) on the baseline. Collect at least 5 minutes of baseline noise at an attenuation which will allow easy measurement of the noise. Inject a known amount of standard ( 100 x noise) and determine the peak height (H) for Sulfur or peak area (A) for Phosphorus. The instructions above will give a methyl parathion peak which contains 120 pg of both sulfur and phosphorus. Determine the width (W) of the peak at half-height (W ½ ) for phosphorus, or the width at 1/4 height for sulfur. Calculate detectivity (D) according to the formula below: Dphosphorus = [Wt x (2 x Np-p)]/[W ½ x H] Dsulfur = [Wt/W ¼ ] x SQRT [(2 x Np-p)/H] :3

47 Operation NOTE: Detectivity is expressed in units of weight/time. Wt is the weight of the specific element injected (e.g., weight of methyl parathion multiplied by 0.12 for either sulfur or phosphorus). W ¼ 1.5 x W ½. The term SQRT means to take the square root of the product in brackets. If you are using peak areas or peak heights that have been calculated by a data system, use the counts-to-mm conversion factor from the data system PFPD TEST CHROMATOGRAM: SULFUR DETECTIVITY Refer to the figure below to check the detectivity of the PFPD for sulfur. D s = [W s /(1.5 x W 1/2)] x SQRT(2 x N/H) where D s = Detectivity (pg/sec) for sulfur, W s = Weight (pg) of sulfur, W 1/2 = width (sec) at half height of sulfur peak, N = peak-to-peak baseline noise (mm), H = height (mm) of sulfur peak, SQRT = square root As an example, for the following test chromatogram: W s = 120 pg, W 1/2 = 2.2 sec, N = 1 mm (1 mm x attenuation of 1), H = 4992 mm (78 mm x attenuation of 64) then D s = [120 pg /1.5 x 2.2 sec] x SQRT(2 x 1 mm/4992 mm) = 0.73 pg/sec :3 3-29

48 Operation PFPD TEST CHROMATOGRAM: PHOSPHORUS DETECTIVITY Refer to the following figure to check the detectivity of the PFPD for phosphorus. D p = [(W p x 2 x N)/(H x W 1/2)] where D p = Detectivity (pg/sec) for phosphorus, W s = Weight (pg) of phosphorus, W 1/2 = width (sec) at half height of phosphorus peak, N = peak-to-peak baseline noise (mm), H = height (mm) of phosphorus peak As an example, for the following test chromatogram: W p = 120 pg, W 1/2 = 2.5 sec, N = 16 mm (4 mm x attenuation of 4), H = mm (67 mm x attenuation of 256) then D p = [(120 pg x 2 x 16 mm)/(17152 mm x 2.5 sec) = pg/sec 3.7 USING THE GATE DELAY AND GATE WIDTH PARAMETERS Through the Gate Delay and Gate Width parameters you have access to the control functions of the gated integrator located on the PFPD electrometer board. These control functions enable you to select the time period (or region) of the emission spectrum that is to be integrated, thereby improving the sensitivity and selectivity of the PFPD :3

49 Operation What Can Be Controlled Through the Adjustments? Three functions can be controlled through the Adjustments: Trigger Level: signal required to start (trigger) the delay and width timer Gate Delay: time in milliseconds after triggering before integration of the detector signal begins. Gate Width: time in milliseconds after the Gate Delay during which the detector signal is integrated. The appropriate values for the Gate Width and the Gate Delay are determined from the emission profile (time x emission intensity) of the element of interest. Appropriate values for a number of elements are listed in the following table. Preset and Range of the Control Functions Refer to the table below for a description of the preset values and ranges of the functions that can be adjusted. Function Preset Range Trigger Level 200 mv mV Gate Delay 4 msec msec Gate Width 10 msec msec For those users interested in gaining a more detailed understanding of the PFPD emission profile, the analog detector signal, the trigger state, and the integrator gate state are available at test points on the edge of the electrometer board, and on the pins of connector J4. When displayed on an oscilloscope, these signals allow detailed investigation of the emission time profiles of the flame background and of elements of interest, so that the detector parameters may be fine tuned for optimum selectivity and detectivity. For more information on the use of these signals, please contact your Varian representative SETTING THE TRIGGER LEVEL Tools and Equipment Needed These tools and equipment are needed to complete the following procedures: Printer/plotter or stripchart recorder FPD test sample (P/N ) 10 µl syringe (P/N ) :3 3-31

50 Operation The trigger level is preset to 200 mv. You may need to set the trigger level lower if you are using a narrow band interference filter which does not transmit much of the flame background. You may need to set the trigger level higher than 200 mv for optimum operation with some elements which have no delayed emission, or very short emission delay times. To set the trigger level, follow the steps below: 1. Press the Detector key and select the PFPD you wish to adjust. Turn OFF initial Autozero in the active method. 2. Press the Adjustments softkey to access the integrator parameters. Set Gate delay to 3.5 ms. This allows the tail of the prompt flame background emission to be integrated as a test signal. This figure shows a typical flame background emission profile. Trigger Level OK NOISY OK 1. Set the trigger level to the minimum and observe the noise on the detector output using a strip chart recorder or the Star Workstation status display. Gradually increase the trigger level. The signal level should decrease and the noise should also decrease. Select a trigger level which gives minimum noise. This will usually be triggering on the preliminary emission pulse :3

51 Operation 2. If you will be detecting an element such as nitrogen, which emits very promptly after the main flame propagation, you may get better results by continuing to increase the trigger level until triggering is occurring on the main background emission pulse itself. As you increase the trigger level, the noise will increase substantially at one point. This is due to the trigger occurring sometimes on the apex of the preliminary emission pulse, and sometimes on the main pulse. Increasing the trigger level 10 to 20% above this point should provide very stable triggering for detection of prompt emitting elements. This setting is also preferred when the effects of large amounts of hydrocarbons on emission timing must be minimized. The following figure illustrates the signal observed during such an adjustment SETTING GATE DELAY AND GATE WIDTH The following table contains gate delay and gate width settings which are recommended for routine detection of common elements. Try these settings first when setting up your PFPD to detect these elements. Element Gate Delay Gate Width Sulfur 6 20 Phosphorus 4 10 Nitrogen :3 3-33

52 Operation Set the Gate Delay by Repeated Injections of Standard The appropriate Gate Delay can be determined by injecting a standard and adjusting the Gate Delay until the required selectivity is achieved. To set the Gate Delay by the method of repeated injections of standard, follow the steps below: 1. Install the test column. 2. Set the Detector temperature to 200 C. Set injector temperature to 200 for isothermal injectors; 80 initial, 0 hold, program to 200 for 1079 mode. Set column oven to 80 initial, hold 0.5 minutes, then program at 25 /minute to 200, and hold 2 minutes, or until the last peak of interest elutes. 3. Allow the chromatographic system to equilibrate for 1 hour. 4. Turn the stripchart recorder ON. 5. Adjust the gas flow rates to those recommended for your carrier gas and flow rate in the Suggested Fuel Gas Flow Rates table in Optimizing the PFPD (Page 3-12). The Air-2 flow rate has little effect on the detectivity of the PFPD and can be set to 10 ml/min. 6. If necessary, adjust the Air-1 and Hydrogen flow rates to give a flame pulse rate 2-4 pulses/second. 7. Inject 1.0 µl of the FPD Test Sample split ~20:1 if using a split injector. Inject 1 µl of a 20:1 dilution of the test sample in iso-octane if using a non-splitting injector. Or, inject another sample of interest. 8. If the desired selectivity has not been achieved, increase the Gate Delay by 0.5 msec and inject the standard to check the effect of this change. Continue to increase the Gate Delay until the desired selectivity is achieved. NOTE: In the case of elements which emit very near the flame background peak, selectivity may be improved by opening the combustor flow needle valve somewhat beyond the position determined in the procedure Establishing Tick Tock. An increase of 1/4 to 1/2 turn counterclockwise is usually sufficient :3

53 Operation SET GATE WIDTH Tools and Equipment Needed These tools and equipment are needed to complete the following procedures: Printer/plotter or stripchart recorder FPD test sample (P/N ) or other standard 10 µl syringe (P/N ) 1. Set Gate Width to minimum 2. Inject an appropriate standard, calculate the detectivity for this Gate Width. 3. See: Testing the Performance of the PFPD (Page 3-25) for instructions on calculating the detectivity of sulfur or phosphorus with the PFPD. 4. Increase Gate Width by 1 msec, inject standard, calculate detectivity. 5. Continue to increase Gate Width as long as detectivity continues to improve. 3.8 BUILDING A GC METHOD The procedure for building a GC method involves 4 tasks: Turn the PFPD electrometer on Set the detector temperature Set the range of the PFPD electrometer Set the gated integrator parameters To Build a GC Method 1. Press the Detector key and select the PFPD you wish to use. 2. Turn the detector oven ON and set the desired detector temperature. 3. Turn the detector electronics ON. 4. Select the Range you wish to use and turn ON initial Autozero if desired. You may also time program Range and Autozero functions if your analysis requires it :3 3-35

54 Operation 5. Press the Adjustments softkey to access the Gated Integrator parameters. Enter the Gate Delay, Gate Width, and Trigger Level parameters as determined in the preceding procedures, or from the table provided. 6. If you wish to use the analog output from your detector, select one of the Analog Output ports, enter the position of the PFPD detector you are using (Front, Middle, or Rear) and enter an attenuation value. If the analog output is going to a data system or integrator, an attenuation of 1 is strongly recommended. If the output is going to a strip chart recorder, note that you may time program the attenuation of the analog output port. 3.9 BUILDING A GC WORKSTATION METHOD The procedure for building a PFPD Detector section as part of a GC method involves 4 tasks: Turn on the PFPD electrometer Set the detector temperature Set the range of the PFPD electrometer Set the Gated Integrator parameters To Build a GC Workstation Method 1. Open the method to which you wish to add the PFPD section. Select the 3800 section, and select the Detector page. Select the position of the PFPD you wish to add (Front, Middle or Rear). 2. Set the Electronics and Detector Oven to ON. 3. Set the Detector Temperature to an appropriate value for your analysis. 4. Set the Range and Autozero as required. Note that these parameters are time programmable. 5. Set the Trigger level, Gate Delay, and Gate Width as required. 6. If you wish to use the Analog Output in addition to the digital signal used by the Star Workstation, select the Analog Output page, select the output channel you wish to use, and select the PFPD position and attenuation :3

55 Operation 3.10 QUANTITATION QUANTITATION WITH THE PFPD When using the PFPD, it is important to understand how its pulsing flame can affect the accuracy and precision of a quantitative analysis. As described previously, the combustor and ignitor chambers of the PFPD fill with a combustible mixture of air and hydrogen. The combustor chamber fills with the combustible mixture at about the same time as the ignitor chamber. When the gas mixture enters the ignitor, the mixture ignites and the flame propagates back through the combustor and then extinguishes. The combustor then begins to fill with the gas mixture and the process repeats. The rate of this repetitive process can vary, but for optimum performance of the PFPD, you must adjust the gas flow rates in order to strike a balance between the fill time of the combustor chamber and the fill time of the ignitor chamber. It is important that the combustor chamber fill slightly before the ignitor chamber in order to ensure that the flame propagates completely back to the combustor chamber. However, the combustor chamber should not fill too fast (relative to the ignitor chamber) or some of the sample exits the detector before it is combusted resulting in poorer detectivity. Because the PFPD pulses, the emission derived from the PFPD appears as discrete signals separated in time. This requires electronics that are quite different from the conventional non-pulsed FPD in which the emission signal increases then decreases without interruption. That is, the PFPD electronics are not sampling discrete points on the eluting peak, as in the case of a conventional FPD. Rather, the PFPD stores the emission output derived during each pulse and generates a peak (or waveform) that may or may not be identical to the actual eluting peak shape, but which contains all the signal information from the chromatographic peak. Given these unique properties of the PFPD, there are differences between quantitation based on the height or area of the chromatographic peaks PEAK AREA QUANTITATION In general, quantitation based on peak areas is preferred because it is least affected by the pulsing nature of the PFPD and the way the PFPD electronics handles the resulting emission signal. Peak areas are best because once the fill rates of the PFPD are appropriately adjusted, any sample that enters the detector contributes to the resulting signal (i.e., all the sample that enters the detector between pulses is combusted), no matter how narrow the chromatographic peak :3 3-37

56 Operation PEAK HEIGHT QUANTITATION Quantitation based on peak height, on the other hand, is less precise and depends, somewhat, on the widths and resolution of the eluting peaks. Peak height is determined by the pulse which contains the maximum concentration of the eluting peak, and to some extent is dependent on how well centered the concentration maximum of the eluting peak is relative to a pulse. The quantitative precision of the PFPD is not affected much by very narrow or very wide peaks. For very narrow peaks, almost all of the peak is combusted in one pulse and the resulting peak height is thereby equivalent to the total emission. For very broad peaks, the pulsed output follows the chromatographic peak shape. There may be some intermediate width peaks for which peak height precision is worse, but these cases can be addressed by changing the pulse repetition rate appropriately. For these intermediate width peaks, the peak is eluting during 2 or 3 pulses and the position of these pulses relative to the peak position can cause peak height to vary. Peak area is not greatly affected PEAK RESOLUTION WITH THE PFPD Higher repetition rates (up to 10 Hz) are necessary when you must resolve two closely spaced peaks. Higher repetition rates give the best possible peak shape definition (at a small cost in detectivity). However, peak resolution is not usually a problem at lower repetition rates since the number of possible sulfur or phosphorus containing peaks is fairly small at the beginning of a chromatogram where peak resolution is likely to be a problem QUANTITATION IN THE SULFUR MODE Since the electronically excited S 2 molecule is the emitting species in the sulfur mode, the signal generated by the PFPD is proportional to the square of the sulfur concentration in the combustor. Because it takes two sulfur atoms to form each excited S 2 molecule, detectivity is improved by increasing the concentration of sulfur in the combustor by confining the emission to a smaller volume (i.e., 2 mm combustor) :3

57 Operation SQUARE ROOT FUNCTION The PFPD electrometer (through a square root function) converts the quadratic response (sulfur emission varies as the square of sulfur concentration) to a linear response (sulfur signal varies as a direct function of the sulfur concentration) by taking the square root of the response. The square root function can be used for the determination of total sulfur in a hydrocarbon matrix, for example, where individual resolved peaks may not be present, and a total area of sulfur response is needed. Normal/Square Root setting The square root mode is selected by the NORMAL/Square Root field. When Square Root is selected, the output from the PFPD is the square root of the sulfur emission. When NORMAL is selected, the output of the PFPD is not converted or adjusted. Use this NORMALAL setting when operating the PFPD in the phosphorus mode, and when the sulfur signal is to be processed by a data system using the square root of height. Detectivity of the NORMAL Versus Square Root Modes In the Square Root mode, the amplifier signal is applied to the input of a square root circuit which produces an output equal to 10 times the square root of V in /10. This output is linear with sulfur concentration but by the nature of the square root function is noisy. Detectivity in the hardware Square Root mode is often 2 to 4 times worse than in the Normal mode under the same detector conditions. Using the Star Workstation and the Square Root of Height Mode The square root of height mode of the Star Workstation is most useful for those cases where quantitation of reasonably resolved peaks is required. The peak height reported is the square root of the net peak height, that is, after correction for the position of the baseline. Detectivity is not affected by use of this square root of height. Do not use the square root of height mode when the method is in the SQUARE ROOT mode, since this would result in taking the fourth root of the original signal. The Star Workstation square root of height mode can be used in conjunction with the Multilevel Calibration function of the Star Workstation :3 3-39

58 Operation Multilevel Calibration for Sulfur Analysis Finally, multilevel calibration in the Star Workstation allows you to construct response curves which can be used to obtain quantitative results, even in the presence of interferences which alter the normal quadratic response. The Star Workstation Multilevel Calibration function can be used, with either quadratic or linearized data, for the quantitation of discrete sulfur containing peaks. This function allows for individual calibration of peaks at up to 9 levels of each component and provides accurate quantitation even under conditions where the sulfur response may not be quadratic because of interferences or matrix effects :3

59 4. Section 4 Advanced Operation 4.1 OVERVIEW This section describes the operation of the PFPD in the Nitrogen mode, including: Theory of operation, describing the basis for nitrogen detection and including suggested gas flow rates. Installation of the nitrogen mode parts - for users who are adding the nitrogen mode capability to an existing PFPD. Includes an exploded view of the parts. Nitrogen mode operation - provides some suggestions for enhancing or modifying nitrogen mode performance. Phosphorus mode/sulfur mode operation - describes some changes which may be required to detect phosphorus or sulfur using the nitrogen mode PMT. Spare Parts and Supplies Additional Elements - describes preliminary operating conditions for the detection of the following elements: Arsenic, Selenium, Tin, and Manganese. 4.2 THEORY OF OPERATION Nitrogen mode is based upon combustion of nitrogen containing compounds to form NO. NO then reacts with hydrogen atoms remaining in the combustor after the flame is extinguished to form HNO*, an electronically excited molecule which emits light in the nm range of the near infrared. Because these emission wavelengths are beyond the range of the photomultiplier tube used in the standard PFPD, the Nitrogen mode requires a new red-sensitive photomultiplier and PMT holder hardware, along with an appropriate dark red colored glass filter. Because N mode is based on complete combustion of the sample molecule to form HNO*, the PFPD response is much less dependent on the chemical nature of the sample than is the TSD response, which depends on the presence of a C- N bond in the molecule. Thus even inorganic nitrogen compounds such as ammonia, hydrazine, and nitrogen oxides are detectable by the N mode PFPD, whereas these compounds give no response on the TSD. The only nitrogen compound known to give no useful response on the PFPD is molecular nitrogen, :3 4-1

60 Advanced Operation N 2. Compounds such as nitroglycerin and other nitrate esters, nitroaromatics, and nitroso compounds give a somewhat enhanced response when compared to more reduced nitrogen compounds, but the enhancement amounts to only ~20%, so for most purposes the N mode PFPD can be considered to give a uniform response to all nitrogen compounds, as it does for sulfur and phosphorus compounds. Uniformity of response may be enhanced by increasing the oxygen flow into the combustor, at the cost of some reduction in sensitivity. Oxidation of nitrogen requires a hotter flame than either phosphorus or sulfur. This is reflected in the following table of recommended flame gas flow rates. In general, the nitrogen mode requires a more air rich flame, and suffers a larger reduction in sensitivity as carrier flow rate increases. Use of the 3 mm combustor provides best detectivity TABLE OF OPTIMUM FLOW RATES Optimum Flow Rate, ml/minute Carrier Gas Flow Rate, ml/minute Air 1 Hydrogen Air 2 Hydrogen Helium Nitrogen INSTALLATION OF THE NITROGEN MODE PARTS If you are upgrading a Phosphorus/Sulfur PFPD to add the Nitrogen mode, you will need to replace the photomultiplier tube housing and filter currently installed on your PFPD with the PMT assembly and filter supplied with your upgrade kit. Refer to the figure below for identification of the components of the N mode assembly. The steps involved are: 1. Turn off the PFPD. Alternatively, turn off GC power. 2. Disconnect the high voltage and signal cables from the base of the P/S PMT housing :3

61 Advanced Operation 3. Remove the PMT housing by removing the two thumb screws securing the housing to the light pipe flange. Store the P/S PMT housing so as to protect it from mechanical damage and bright light. 4. Insert the Nitrogen mode filter in the recess in front of the PMT in the N mode housing. Be sure the silicone O-ring is in its place in the groove on the face of the housing flange. Using the two screws supplied with the upgrade kit, attach the N mode PMT housing to the light pipe flange. Be sure that the O- ring makes a light-tight seal. 5. Reconnect the signal and high voltage cables to the connectors on the base of the N mode PMT housing. Restore power to the PFPD. 6. Press the Detector key, select the PFPD, and press the Adjustments soft key. Set a Gate Delay of 3.5 msec, a Gate width of 8 msec, and a PMT Voltage of 600 volts. Item Description 1 Filter Mount 2 O-ring 3 O-ring, PMT Mounting 4 PMT Housing 5 Set Screws 6 O-ring 7 PMT 8 Socket 9 Retainer 10 Connectors with Mounting Plate :3 4-3

TRACE 1300 and TRACE 1310 Gas Chromatographs Dual Detector Microfluidics Installation Guide

TRACE 1300 and TRACE 1310 Gas Chromatographs Dual Detector Microfluidics Installation Guide TRACE 1300 and TRACE 1310 Gas Chromatographs Dual Detector Microfluidics Installation Guide P/N 31709711 Revision A October 2013 2013 Thermo Fisher Scientific Inc. All rights reserved. TSQ 8000, ISQ, TRACE

More information

Using the Pulsed Flame Photometric Detector (PFPD) for Low-level Analysis of Organophosphorus Pesticides

Using the Pulsed Flame Photometric Detector (PFPD) for Low-level Analysis of Organophosphorus Pesticides Application Note 25390206 Using the Pulsed Flame Photometric Detector (PFPD) for Low-level Analysis of Organophosphorus Pesticides Keywords Model 5380 Organophosphorus Pesticide PFPD PFPDView Phosphorus

More information

TRACE 1300 and TRACE 1310 Gas Chromatographs. NoVent Microfluidics. Installation Guide Revision A March 2013

TRACE 1300 and TRACE 1310 Gas Chromatographs. NoVent Microfluidics. Installation Guide Revision A March 2013 TRACE 1300 and TRACE 1310 Gas Chromatographs NoVent Microfluidics Installation Guide 120807-0001 Revision A March 2013 2013 Thermo Fisher Scientific Inc. All rights reserved. TSQ 8000, ISQ, TRACE 1300,

More information

DETECTORS Dry Electrolytic Conductivity Detector - DELCD. Overview

DETECTORS Dry Electrolytic Conductivity Detector - DELCD. Overview Dry Electrolytic Conductivity Detector - Overview on an 8610C GC The Dry Electrolytic Conductivity detector, or, is selective to chlorinated and brominated molecules. It differs from the traditional wet

More information

Content Page OPTIC Liners OPTIC 2 OPTIC 3 OPTIC-4 LINEX PHASER

Content Page OPTIC Liners OPTIC 2  OPTIC 3 OPTIC-4 LINEX PHASER Content Page OPTIC Liners 3 OPTIC 2 Consumables 5 OPTIC 3 Consumables 7 Spare Parts 9 OPTIC-4 Consumables 12 Spare Parts 13 OPTIC accessories 15 LINEX Consumables, spare parts 16 PHASER Consumables 17

More information

Agilent Consumables Improve Performance with Shimadzu GC/MS

Agilent Consumables Improve Performance with Shimadzu GC/MS Agilent Consumables Improve Performance with Shimadzu GC/MS Application Note Environmental & Food Safety Author Ken Lynam Agilent Technologies, Inc. Abstract Leak-free performance over 150 heat cycles

More information

GC Troubleshooting. Frank Michel, Katherine K. Stenerson T HCE

GC Troubleshooting. Frank Michel, Katherine K. Stenerson T HCE GC Troubleshooting Frank Michel, Katherine K. Stenerson T404077 HCE GC Troubleshooting: Topics Basic Troubleshooting Strategy Preventing Problems Identifying Common Problems Recommended Reading Discussion

More information

Capillary GC Troubleshooting: a Practical Approach

Capillary GC Troubleshooting: a Practical Approach Capillary GC Troubleshooting: a Practical Approach Katherine Stenerson Gas Separations R&D Supelco, Supelco Park, Bellefonte, PA 16823 1999 Sigma-Aldrich Co. Outline Basic Troubleshooting Strategy Preventing

More information

Video Notes. LC Troubleshooting Series Peak Broadening. Why Peak Broadening Occurs

Video Notes. LC Troubleshooting Series Peak Broadening. Why Peak Broadening Occurs Video Notes LC Troubleshooting Series Peak Broadening Mike Woodman is an Application Scientist who specializes in LC instrument performance and works in the Agilent Midwest Center of Excellence. Bill Long

More information

Agilent InfinityLab Preparative Flow Cells for MWD & DAD

Agilent InfinityLab Preparative Flow Cells for MWD & DAD Agilent InfinityLab Preparative Flow Cells for MWD & DAD Technical Note This technical note describes the preparative flow cells for the Agilent 1260 Infinity II DAD WR (G7115A) and Agilent 1260 Infinity

More information

When optimization is discussed throughout this talk, it s in terms of performing more samples, in less time, while minimizing investment in capital

When optimization is discussed throughout this talk, it s in terms of performing more samples, in less time, while minimizing investment in capital 1 2 When optimization is discussed throughout this talk, it s in terms of performing more samples, in less time, while minimizing investment in capital instruments. 3 Efficiency is a measure of peak width.

More information

SCION GC Series. The Gas Chromatographers Choice for Separations

SCION GC Series. The Gas Chromatographers Choice for Separations SCION GC Series The Gas Chromatographers Choice for Separations Innovation in Gas Chromatography Scion Instrument s tradition of innovation and product reliability have combined to create the next generation

More information

SCION GC Series. The Gas Chromatographers Choice for Separations

SCION GC Series. The Gas Chromatographers Choice for Separations SCION GC Series The Gas Chromatographers Choice for Separations Innovation in Gas Chromatography Scion Instrument s tradition of innovation and product reliability have combined to create the next generation

More information

Prefractionator for Reliable Analysis of the Light Ends of Crude Oil and other Petroleum Fractions

Prefractionator for Reliable Analysis of the Light Ends of Crude Oil and other Petroleum Fractions Prefractionator for Reliable Analysis of the Light Ends of Crude Oil and other Petroleum Fractions Application Note Hydrocarbon Processing Author Roger L Firor Agilent Technologies, Inc. 2850 Centerville

More information

TurboMatrix Headspace and Headspace Trap Samplers

TurboMatrix Headspace and Headspace Trap Samplers S P E C I F I C A T I O N S Gas Chromatography TurboMatrix Headspace and Headspace Trap Samplers For added flexibility to meet your laboratory s changing requirements, five different PerkinElmer TurboMatrix

More information

A Rapid Method for Trace Analysis of Organophosphorus Pesticides in Drinking Water

A Rapid Method for Trace Analysis of Organophosphorus Pesticides in Drinking Water A Rapid Method for Trace Analysis of Organophosphorus Pesticides in Drinking Water Application Note Environmental Authors Min Cai and Yun Zou Agilent Technologies Co. Ltd, 412 Ying Lun Road Waigaoqiao

More information

Contract Laboratory Program (CLP) Pesticide Analysis with 0.18 mm ID High-Efficiency GC Columns Utilizing Helium Carrier Gas. Application.

Contract Laboratory Program (CLP) Pesticide Analysis with 0.18 mm ID High-Efficiency GC Columns Utilizing Helium Carrier Gas. Application. Contract Laboratory Program (CLP) Pesticide Analysis with 0.8 mm ID High-Efficiency GC Columns Utilizing Helium Carrier Gas Application Environmental Authors Ken Lynam and William Long Agilent Technologies,

More information

Varian Supplies/Parts. GC Accessories. Septum Nut. Merlin Microseal Kit

Varian Supplies/Parts. GC Accessories. Septum Nut. Merlin Microseal Kit Septum Nut Septum nut provides consistent septum tightness, fewer bent needles, and easy leak detection. The nut contains a needle guide, ensuring that the needle consistently penetrates the septum in

More information

Industrial flue gas probes. Instruction manual

Industrial flue gas probes. Instruction manual Industrial flue gas probes Instruction manual 2 1 Contents 1 Contents 1 Contents... 3 2 Safety and the environment... 4 2.1. About this document... 4 2.2. Ensure safety... 4 2.3. Protecting the environment...

More information

Gilian BDX-II Air Sampling Pump Setup and Use

Gilian BDX-II Air Sampling Pump Setup and Use Gilian BDX-II Air Sampling Pump Gilian BDX-II Air Sampling Pump Setup and Use Gilian BDX-II Designed for ease-of-use in abatement applications Lead abatement Asbestos abatement Robust design Light weight

More information

System Overview. Valve oven. Restrictor. Valve. Sample in and out

System Overview. Valve oven. Restrictor. Valve. Sample in and out System Overview The GC system is designed to provide a continous reading of total hydrocarbons in a gas stream, while periodically performing a chromatographic separation of the sample to determine the

More information

Shur-Shot X-Proof Hydrogen Fluoride Alarm Operations Manual

Shur-Shot X-Proof Hydrogen Fluoride Alarm Operations Manual Shur-Shot X-Proof Hydrogen Fluoride Alarm Operations Manual P/N 1000006053 Rev E $7%$QDO\WLFV// //& 733 Dairy Rd. Parkton, Md. 21120 www.atbanalytics.com (410) 733-6365 Table of Contents Chapter 1: Getting

More information

PerkinElmer Supplies/Parts. GC Accessories. Autosampler Syringes

PerkinElmer Supplies/Parts. GC Accessories. Autosampler Syringes Autosampler Syringes Volume, Code, Gauge, Length, Point Style QTY OEM No. PerkinElmer No. Part No. Hamilton Syringes 5μL, ASN, 23s, 70mm, #3 1 88035 N6101390 24-88035 5μL, ASN, 26s, 70mm, #3 1 88040 N6101380

More information

Gas Chromatography Assignment Chem 543/443

Gas Chromatography Assignment Chem 543/443 Gas Chromatography Assignment Chem 543/443 1. Introduction Capillary gas chromatography (GC) is one of the most popular analytical techniques used in today s research. Its popularity is mainly due to efficient

More information

KOBRA CELL Art. No. K01

KOBRA CELL Art. No. K01 R-Biopharm Rhône Ltd KOBRA CELL Art. No. K01 Instruction Manual R-Biopharm for the safety of your analysis. R-Biopharm Rhone Ltd 2 Contents 1. Introduction to the KOBRA CELL... 4 1.1 Intended Use... 4

More information

VWR Zero Air Generator Models , Installation, Operation, and Maintenance Manual. Bulletin TI-ZAG3500-L1466

VWR Zero Air Generator Models , Installation, Operation, and Maintenance Manual. Bulletin TI-ZAG3500-L1466 VWR Zero Air Generator Models 26000-020, 26000-022 Bulletin TI-ZAG3500-L1466 Installation, Operation, and Maintenance Manual Bulletin TI-ZAG3500-L1466 VWR Model 26000 Zero Air Generators Warranty United

More information

( ) IONPAC ATC-HC ANION TRAP COLUMN

( ) IONPAC ATC-HC ANION TRAP COLUMN PRODUCT MANUAL IONPAC ATC-3 2-mm ANION TRAP COLUMN (4 x 35 mm, P/N 059661) IONPAC ATC-3 4-mm ANION TRAP COLUMN (9 x 24 mm, P/N 059660) IONPAC ATC-HC ANION TRAP COLUMN (9 x 75 mm, P/N 059604) QUICKSTART

More information

SCIEX Anion Analysis Kit For P/ACE MDQ and P/ACE MDQ plus Capillary Electrophoresis Systems. Instruction Guide

SCIEX Anion Analysis Kit For P/ACE MDQ and P/ACE MDQ plus Capillary Electrophoresis Systems. Instruction Guide For P/ACE MDQ and P/ACE MDQ plus Capillary Electrophoresis Systems A49108AE May 2015 AB Sciex Pte. Ltd and its affiliates disclaims all warranties with respect to this document, expressed or implied, including

More information

Model: LoLander Series Truck Scale Model: LL FT Model: LL FT Installation, Set-Up and Operation Manual

Model: LoLander Series Truck Scale Model: LL FT Model: LL FT Installation, Set-Up and Operation Manual SCALE WORKS, INC. Model: LoLander Series Truck Scale Model: LL8035-10FT Model: LL8035-11FT Installation, Set-Up and Operation Manual Cambridge Scale Works, Inc. P.O. Box 670 Honey Brook, PA. 19344 (800)

More information

Transferring a Method from a Conventional 3.5 µm or 5 µm to an Agilent Poroshell 120 Column to Improve Analytical Speed and Resolution

Transferring a Method from a Conventional 3.5 µm or 5 µm to an Agilent Poroshell 120 Column to Improve Analytical Speed and Resolution Video Notes Transferring a Method from a Conventional 3.5 µm or 5 µm to an Agilent Poroshell 120 Column to Improve Analytical Speed and Resolution Ron Majors is Senior Scientist for Agilent s Columns and

More information

Analysis of Organotin Compounds in Biological and Environmental Samples by Gas Chromatography and Pulsed Flame Photometric Detection (GC-PFPD)

Analysis of Organotin Compounds in Biological and Environmental Samples by Gas Chromatography and Pulsed Flame Photometric Detection (GC-PFPD) Introduction Organotin compounds have a great many applications which include stabilizers, catalysts, biocides, and pesticides. One of the more significant uses was that of biocides employed as anti-fouling

More information

SYSTEM 600 GENERAL MAINTENANCE AND ADJUSTMENT GUIDE

SYSTEM 600 GENERAL MAINTENANCE AND ADJUSTMENT GUIDE SYSTEM 600 GENERAL MAINTENANCE AND ADJUSTMENT GUIDE If you follow a routine maintenance schedule your system 600 will give you years of reliable service with a minimum of down time. This guide outlines

More information

A Direct Column-Performance Comparison for Rapid Contract Laboratory Program (CLP) Pesticide Analysis. Application. Authors. Abstract.

A Direct Column-Performance Comparison for Rapid Contract Laboratory Program (CLP) Pesticide Analysis. Application. Authors. Abstract. A Direct Column-Performance Comparison for Rapid Contract Laboratory Program (CLP) Pesticide Analysis Application Environmental Authors Ken Lynam and John W. Henderson, Jr. Agilent Technologies, Inc. 2850

More information

WARNINGS AND PRECAUTIONS

WARNINGS AND PRECAUTIONS WARNINGS AND PRECAUTIONS If you have questions about the unit you are repairing, please do not hesitate to contact your local SciCan representative for information. Also, the STATIM is heavy. Exercise

More information

Chem 321 Lecture 23 - Liquid Chromatography 11/19/13

Chem 321 Lecture 23 - Liquid Chromatography 11/19/13 Chem 321 Lecture 23 - Liquid Chromatography 11/19/13 Student Learning Objectives High Performance Liquid Chromatography With the advent of relatively inexpensive and reliable pumps, the development of

More information

GAS CHROMATOGRAPHY CLARUS 400/480 GC

GAS CHROMATOGRAPHY CLARUS 400/480 GC GAS CHROMATOGRAPHY CLARUS 400/480 GC Customer Hardware and Service Guide Clarus 400/480 GC Customer Hardware and Service Guide Release History Part Number Release Publication Date 09936811 C January 2010

More information

Skoog, Holler and Nieman, Principles of Instrumental Analysis, 5th edition, Saunders College Publishing, Fort Worth, TX 1998, Ch 33.

Skoog, Holler and Nieman, Principles of Instrumental Analysis, 5th edition, Saunders College Publishing, Fort Worth, TX 1998, Ch 33. CHEM 3281 Experiment Ten Determination of Phosphate by Flow Injection Analysis Objective: The aim of the experiment is to investigate the experimental variables of FIA for a model system and then to use

More information

AP5K-C Precision AC Double-Pulse Spot Welding Machine User s Manual Shenzhen Will-Best Electronics Co., Ltd

AP5K-C Precision AC Double-Pulse Spot Welding Machine User s Manual Shenzhen Will-Best Electronics Co., Ltd AP5K-C Precision AC Double-Pulse Spot Welding Machine User s Manual Shenzhen Will-Best Electronics Co., Ltd 1 Content 1. Introduction...3 1.1 Functions...3 1.2 Units of AP5K-C...4 2. The Initial Installation

More information

Operating the SRI RGD ( Reduction Gas Detector )

Operating the SRI RGD ( Reduction Gas Detector ) The SRI RGD detector is mounted on the right side of the column oven on a 8610C or 310 gas chromatograph. The RGD detector can also be mounted on a Model 110 stand-alone chassis ( not shown ). RGD detector

More information

CX-1 ph Gradient Buffer

CX-1 ph Gradient Buffer User Manual CX-1 ph Gradient Buffer 065534 Revision 03 May 2016 For Research Use Only. Not for use in diagnostic procedures. Product Manual for CX-1 ph Gradient Buffer A (ph 5.6) (125 ml P/N: 083273) (250

More information

High Temperature Aging Cell Instruction Manual

High Temperature Aging Cell Instruction Manual High Temperature Aging Cell Instruction Manual Part No. 760140001EA Rev. D TABLE OF CONTENTS Section Page Figure 1 General Information... 1 2 Safety Considerations... 3 3 Un-pressurized Testing Procedure...

More information

INSTRUCTIONS FOR THE ASSEMBLY AND INSTALLATION OF THE VORTOX LIQUID SAMPLER

INSTRUCTIONS FOR THE ASSEMBLY AND INSTALLATION OF THE VORTOX LIQUID SAMPLER INSTRUCTIONS FOR THE ASSEMBLY AND INSTALLATION OF THE VORTOX LIQUID SAMPLER I. SAMPLER SUSPENDED BY CABLE The Vortox Liquid Sampler is a precision instrument manufactured of stainless steel, easy to install,

More information

5000TOC Sensor Standard Operating Procedure

5000TOC Sensor Standard Operating Procedure Part No. 84460 5000TOC Sensor Standard Operating Procedure for System Suitability Testing to meet USP 643 and EP 2.2.44 This document contains proprietary information, which is protected by copyright.

More information

ASI MODEL DM 100 SERIES DILUTING MODULES

ASI MODEL DM 100 SERIES DILUTING MODULES ASI MODEL DM 100 SERIES DILUTING MODULES The ASI Model DM 100 Series Diluting Modules have been developed to meet the requirements of some of our more specialized applications, where our standard in situ

More information

Boss Water Systems Australia Reverse Osmosis System Model 021-4P-GM

Boss Water Systems Australia Reverse Osmosis System Model 021-4P-GM Boss Water Systems Australia Reverse Osmosis System Model 021-4P-GM Congratulations on your purchase of the most advanced water purifier system available! Installation and Service Guide short web version,

More information

SCION GC Series. The Gas Chromatographers Choice for Separations

SCION GC Series. The Gas Chromatographers Choice for Separations SCION GC Series The Gas Chromatographers Choice for Separations Innovation in Gas Chromatography Bruker s Scion Instrument s long tradition long of tradition innovation of innovation and product and reliability

More information

Gas Chromatograph. GC-8A Series C184-E016B

Gas Chromatograph. GC-8A Series C184-E016B Gas Chromatograph GC-8A Series C184-E016B GC-8A Series Dedicated Single Detector Gas Chromatograph GC that meets increasing analytical requirements GC that can analyze instantaneously GC that is tough

More information

Continuously Regenerated Trap (CR-TC 500, Capillary)

Continuously Regenerated Trap (CR-TC 500, Capillary) Thermo Scientific Continuously Regenerated Trap (CR-TC 500, Capillary) Column Product Manual P/N: November 2012 Part of Thermo Fisher Scientific Product Manual for Continuously Regenerated Trap Column

More information

Agilent Max-Light Cartridge Cell Information for G4212A and G4212B DAD

Agilent Max-Light Cartridge Cell Information for G4212A and G4212B DAD Agilent Max-Light Cartridge Cell Information for G4212A and G4212B DAD Information about the Max-Light Cartridge Cells 2 Specifications 2 Description of the dispersion volume V( ) of 2 Recommendations

More information

TABLE OF CONTENTS. Vacuum Cleaner. Model No. MC-CG Bagless Canister Red North America

TABLE OF CONTENTS. Vacuum Cleaner. Model No. MC-CG Bagless Canister Red North America Order No.: MAC1412002CE Vacuum Cleaner Model No. MC-CG902-02 Bagless Canister Red North America TABLE OF CONTENTS PAGE 1 Specifications ----------------------------------------------------- 2 2 Disassembly

More information

Sampling Techniques Handbook. Agilent Automatic Liquid Samplers

Sampling Techniques Handbook. Agilent Automatic Liquid Samplers Sampling Techniques Handbook Agilent Automatic Liquid Samplers Agilent Technologies 1997-2001 Target is a registered trademark and DP is a trademark of National Scientific Company. All Rights Reserved.

More information

EASY-Spray Columns. Guidance for column set up and installation Tips to maximize column lifetime

EASY-Spray Columns. Guidance for column set up and installation Tips to maximize column lifetime EASY-Spray Columns Guidance for column set up and installation Tips to maximize column lifetime EASY-Spray Column Tips and Tricks This document provides guidance for Thermo Scientific EASY-Spray column

More information

Mounting Screw. Channel Filler. Light Angle Adjustable Screw. LED Board Vertical stud and blocking Locking Nose. Driver.

Mounting Screw. Channel Filler. Light Angle Adjustable Screw. LED Board Vertical stud and blocking Locking Nose. Driver. The Knife Edge Cove is a modular lighting system consisting of two parts: 1- An aluminum fixture housing extrusion 2- Drop-in lighting modules The fixture will provide a continuous knife edge architectural

More information

The Retention Gap Effect

The Retention Gap Effect GC CONNECTIONS The Retention Gap Effect John V. Hinshaw, Serveron Corp., Hillsboro, Oregon, USA. In this month s GC Connections, the author examines the effects on retention times and peak shapes when

More information

2.1 This procedure provides steps and conditions for the determination of Vapor Space Organic Compounds in ground waters, and industrial waste waters.

2.1 This procedure provides steps and conditions for the determination of Vapor Space Organic Compounds in ground waters, and industrial waste waters. 1. TITLE: MSD Analytical Method for Vapor Space Organics 2. PURPOSE: 2.1 This procedure provides steps and conditions for the determination of Vapor Space Organic Compounds in ground waters, and industrial

More information

Operating & Maintenance Manual JEC JCP Series. Centrifugal Pumps

Operating & Maintenance Manual JEC JCP Series. Centrifugal Pumps Operating & Maintenance Manual JEC JCP Series Centrifugal Pumps JEC LTD. 408, Cheongyo-ri, Bibong-myun, Hwaseong-si, Gyeonggi-do, South Korea Tel: +82 31 355 0316 Fax: +82 31 355 0319 www.jecpump.com E-mail:

More information

Perkin Elmer HGA-600/ HGA-700 Graphite Furnaces

Perkin Elmer HGA-600/ HGA-700 Graphite Furnaces www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Perkin Elmer HGA-600/ HGA-700 Graphite Furnaces The HGA-600 and HGA-700 are

More information

SAFETY AND QUALITY CONTROL

SAFETY AND QUALITY CONTROL Note: this protocol is optimized for the Breeze software and for purifying peptides with a high concentration >30 mg/ml To purify peptides with high performance liquid chromatography. Samples can be vortexed

More information

Introduction Ron Majors is a Senior Scientist at Agilent. Bill Champion is a chemist in Agilent s HPLC Columns tech support group.

Introduction Ron Majors is a Senior Scientist at Agilent. Bill Champion is a chemist in Agilent s HPLC Columns tech support group. Video Notes LC Troubleshooting Series Ghost Peaks Introduction Ron Majors is a Senior Scientist at Agilent. Bill Champion is a chemist in Agilent s HPLC Columns tech support group. Ghost peaks can come

More information

Agilent Ultimate Plus Fused Silica Tubing

Agilent Ultimate Plus Fused Silica Tubing Agilent Ultimate Plus Fused Silica Tubing Technical Overview Introduction Deactivated fused silica tubing is widely used for guard columns, transfer lines, or long retention gaps in GC and GC/MS applications

More information

USER S OPERATING AND INSTRUCTION MANUAL

USER S OPERATING AND INSTRUCTION MANUAL Grand Rapids, Michigan, U.S.A. 49504-5298 USER S OPERATING AND INSTRUCTION MANUAL MODEL 738 MINI CHIP SLICER 0738S20000-CV INDEX Section Description Document No. Page No. SAFETY INSTRUCTIONS --------------------------------------

More information

Organohalide Pesticides in Water by GC/µECD with Agilent J&W DB-CLP1 and DB-CLP2

Organohalide Pesticides in Water by GC/µECD with Agilent J&W DB-CLP1 and DB-CLP2 Organohalide Pesticides in Water by GC/µECD with Agilent J&W DB-CLP and DB-CLP Application Note Environmental Authors Doris Smith and Ken Lynam Agilent Technologies, Inc. Abstract Organohalide pesticides

More information

Agilent MetaCarb 87H Organic Acids Column H + Form User Manual Part No. A5214

Agilent MetaCarb 87H Organic Acids Column H + Form User Manual Part No. A5214 Agilent MetaCarb 87H Organic Acids Column H + Form User Manual Part No. A5214 Warning: The Agilent MetaCarb 87H organic acids column is packed with a polymeric material that requires special care. Introduction

More information

EXIRA EXPOSED PIPE SHOWER INSTALLATION

EXIRA EXPOSED PIPE SHOWER INSTALLATION SKU(s): 923583 EXIRA EXPOSED PIPE SHOWER INSTALLATION BEFORE YOU BEGIN We recommend consulting a professional if you are unfamiliar with installing plumbing fixtures. Signature Hardware accepts no liability

More information

Mar Cor Purification. Carbon Block Filtration System. Operation & Maintenance Manual

Mar Cor Purification. Carbon Block Filtration System. Operation & Maintenance Manual Mar Cor Purification Carbon Block Filtration System Operation & Maintenance Manual 3027337 Rev. B 8Jan13 3027337 Rev. B 8Jan13 Operation & Maintenance Manual Table of Contents Page CHAPTER ONE: GENERAL

More information

Section 8 Technician Level Maintenance

Section 8 Technician Level Maintenance Section 8 Technician Level Maintenance Page 8-1 Table of Contents 1. Warning: Maintenance Hazards 1.3 Scope of this Section 2. Gasifier 2.1 Overhaul Gasifier 2.2 Overhaul Hopper 2.3 Replace Gasifier 3.

More information

DRONIUM ZERO DRONE WITH LIVE STREAMING CAMERA

DRONIUM ZERO DRONE WITH LIVE STREAMING CAMERA DRONIUM ZERO DRONE WITH LIVE STREAMING CAMERA THANK YOU. Thank you for your purchase of Protocol s Dronium Zero With Live Streaming Camera. You are about to experience the best of what remote control

More information

Volatile Organic Carbon Contaminated Site Assessment

Volatile Organic Carbon Contaminated Site Assessment 92 Volatile Organic Carbon Contaminated Site Assessment Introduction Roughly 75 percent of the major cities in the U.S. depend, at least in part, on groundwater for their water supply. Various estimates

More information

Method Translation in Liquid Chromatography

Method Translation in Liquid Chromatography Method Translation in Liquid Chromatography Technical Overview Abstract Ronald E. Majors Agilent Technologies, Inc. 2850 Centerville Rd Wilmington, DE 19808 USA With the recent emphasis on high performance

More information

Apparatus for Vacuum Heating of TEM Samples

Apparatus for Vacuum Heating of TEM Samples Apparatus for Vacuum Heating of TEM Samples Adam Callens Capstone Project, Physics 492R May 1, 2007 Advisor: Richard Vanfleet Abstract The purpose of this capstone project was to create a sample holder

More information

Operational Manual. Spectrophotometer Model: SP-830 PLUS. Metertech Inc. Version 1.07

Operational Manual. Spectrophotometer Model: SP-830 PLUS. Metertech Inc. Version 1.07 Operational Manual Spectrophotometer Model: SP-830 PLUS Metertech Inc. Version 1.07 Metertech Inc. provides this publication as is without warranty of any kind, either express or implied, including, but

More information

Model: 770 Light Vehicle Scale Size: 8 x30 x5 Size: 10 x30 x5 Capacity: 40K CLC: 20K Capacity: 60K CLC:30K Installation, Set-Up and Operation Manual

Model: 770 Light Vehicle Scale Size: 8 x30 x5 Size: 10 x30 x5 Capacity: 40K CLC: 20K Capacity: 60K CLC:30K Installation, Set-Up and Operation Manual SCALE WORKS, INC. Model: 770 Light Vehicle Scale Size: 8 x30 x5 Size: 10 x30 x5 Capacity: 40K CLC: 20K Capacity: 60K CLC:30K Installation, Set-Up and Operation Manual Cambridge Scale Works, Inc. P.O. Box

More information

In-Ground LED Light Installation Guide

In-Ground LED Light Installation Guide TM PROFESSIONAL GRADE LED LIGHTING www.aspectled.com Copyright 2012-2016. aspectled, Inc. 1 OVERVIEW In-ground high power LEDs can be mounted in-ground or in-wall. These are available in Ultra Bright and

More information

C.R. LAURENCE CO., INC. Intelli-Track SPS. Phase One - Installation Instructions

C.R. LAURENCE CO., INC. Intelli-Track SPS. Phase One - Installation Instructions Tools and Supplies: C.R. LAURENCE CO., INC. Intelli-Track SPS Phase One - Installation Instructions Ratchet Wrench with 4" Extension 9/16" Open-End Wrench (two required for suspension mounting) 9/16" Hex

More information

Direct Injection of Fish Oil for the GC-ECD Analysis of PCBs: Results Using a Deans Switch With Backflushing. Application. Author.

Direct Injection of Fish Oil for the GC-ECD Analysis of PCBs: Results Using a Deans Switch With Backflushing. Application. Author. Direct Injection of Fish Oil for the GC-ECD Analysis of PCBs: Results Using a Deans Switch With Backflushing Application Environmental and Pharmaceutical Author Philip L. Wylie Agilent Technologies, Inc.

More information

Conversion Station Model 711

Conversion Station Model 711 Conversion Station Model 711 Automatic Tag Dispenser Guide 3M Library Systems 3M Center, Building 225-4N-14 St. Paul, Minnesota 55144-1000 www.3m.com/library Copyright 2000-2004, 3M IPC. All rights reserved.

More information

ABB MEASUREMENT & ANALYTICS DATA SHEET. PGC5007 Total sulfur analyzer

ABB MEASUREMENT & ANALYTICS DATA SHEET. PGC5007 Total sulfur analyzer ABB MEASUREMENT & ANALYTICS DATA SHEET PGC5007 Total sulfur analyzer 2 P G C 5 0 0 7 TOTA L S U L F U R A N A LY Z E R D S/ P G C 5 0 0 7- E N R E V. A Measurement made easy Online ASTM method D7041-04

More information

FORSYTH COUNTY OFFICE OF ENVIRONMENTAL ASSISTANCE AND PROTECTION

FORSYTH COUNTY OFFICE OF ENVIRONMENTAL ASSISTANCE AND PROTECTION Page 1 of 9 FORSYTH COUNTY OFFICE OF ENVIRONMENTAL ASSISTANCE AND PROTECTION STANDARD OPERATING PROCEDURE (SOP) Zero Air Generators Signature Page Section 13 Page 2 of 9 By the signatures below, the Forsyth

More information

Markes International Thermal Desorption G8125A, G8126A, G8127A, G8128A, G8131A, G8132A, G8136A, G8133A, MKI-MicroC Site Preparation Checklist

Markes International Thermal Desorption G8125A, G8126A, G8127A, G8128A, G8131A, G8132A, G8136A, G8133A, MKI-MicroC Site Preparation Checklist Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation

More information

Heated FID OVF Portable THC/TVOC Analyzer

Heated FID OVF Portable THC/TVOC Analyzer Heated FID OVF-3000 Portable THC/TVOC Analyzer The OVF-3000 is light weight and portable, competitively priced compact and fully heated FID (HFID) based total hydrocarbon analyzer for high accuracy, sensitivity

More information

Performance characteristics of the 1260 Infinity Quaternary LC system

Performance characteristics of the 1260 Infinity Quaternary LC system Performance characteristics of the 1260 Infinity Quaternary LC system The new standard in HPLC Technical Overview Introduction The Agilent 1260 Infinity LC system consists of modular units that operate

More information

FINAL REPORT. Participants in the reported tasks 1. Ravi Subramanian 2. Manoranjan Misra 3. Kent Hoekman

FINAL REPORT. Participants in the reported tasks 1. Ravi Subramanian 2. Manoranjan Misra 3. Kent Hoekman FINAL REPORT Subtask 2.2.1: Solar thermal hydrogen production Duration of the Project: October 01 2009 - Dec 31, 2010 Date of Final Report Submission: January 24, 2011 Primary Lead contact Vaidyanathan

More information

PURPOSE: To separate the pigments of spinach by Column Chromatography. To analyze Column Chromatography fractions by Thin Layer Chromatography.

PURPOSE: To separate the pigments of spinach by Column Chromatography. To analyze Column Chromatography fractions by Thin Layer Chromatography. LAB 2: CHROMATOGRAPHY PART 2: Column Chromatography PURPOSE: To separate the pigments of spinach by Column Chromatography. To analyze Column Chromatography fractions by Thin Layer Chromatography. SAFETY

More information

Autosampler Supplies for Agilent GC and HPLC Systems with CTC Autosamplers

Autosampler Supplies for Agilent GC and HPLC Systems with CTC Autosamplers Autosampler Supplies for Agilent GC and HPLC Systems with CTC Autosamplers Proven reliability that complements Agilent instruments the most reliable in the industry. A wide selection to suit your high-throughput

More information

Gas Filtration Catalogue

Gas Filtration Catalogue Gas Filtration Catalogue 2 SHIMADZU GAS FILTRATION CATALOGUE ABOUT SUPER CLEAN GAS FILTERS Cartridge systems make changing gas filters quick and easy. A base plate allows cartridges to be exchanged without

More information

// XPLORER-NS. Full range Total Nitrogen, Total Sulfur instrument for the modern Laboratory.

// XPLORER-NS. Full range Total Nitrogen, Total Sulfur instrument for the modern Laboratory. Configuration: XPLORER with ARCHIE and GLS* // XPLORER-NS Full range Total Nitrogen, Total Sulfur instrument for the modern Laboratory. TE Instruments has developed the XPLORER-NS, a Total Nitrogen and

More information

! CAUTION: Read and follow these instructions.

! CAUTION: Read and follow these instructions. OmniLog SAFETY AND INSTALLATION INSTRUCTIONS! CAUTION: Read and follow these instructions. Avoid Contact with Hazardous Live Parts WARNING: Risk of electrical shock. Refer servicing to qualified personnel.

More information

Ghost Peaks in Gas Chromatography Part 2: The Injection Port

Ghost Peaks in Gas Chromatography Part 2: The Injection Port Cd The Chrom Doctor Ghost Peaks in Gas Chromatography Part 2: The Injection Port Jaap de Zeeuw, Restek Corporation, Middelburg, The Netherlands. In the last issue we discussed possible sources for ghost

More information

Engineered Systems & Designs TURBIDITY METER

Engineered Systems & Designs TURBIDITY METER Engineered Systems & Designs TURBIDITY METER Operator s Manual for Model 800 Turbidimeter Model 800 with Optional Battery Pack and Carrying Case INDEX Specifications for Turbidity.................. 1 Introduction..............

More information

METHOD 3 - GAS ANALYSIS FOR THE DETERMINATION OF DRY MOLECULAR WEIGHT. NOTE: This method does not include all of the

METHOD 3 - GAS ANALYSIS FOR THE DETERMINATION OF DRY MOLECULAR WEIGHT. NOTE: This method does not include all of the 312 METHOD 3 - GAS ANALYSIS FOR THE DETERMINATION OF DRY MOLECULAR WEIGHT NOTE: This method does not include all of the specifications (e.g., equipment and supplies) and procedures (e.g., sampling) essential

More information

METHOD 17 - DETERMINATION OF PARTICULATE MATTER EMISSIONS FROM STATIONARY SOURCES. NOTE: This method does not include all of the

METHOD 17 - DETERMINATION OF PARTICULATE MATTER EMISSIONS FROM STATIONARY SOURCES. NOTE: This method does not include all of the 1035 METHOD 17 - DETERMINATION OF PARTICULATE MATTER EMISSIONS FROM STATIONARY SOURCES NOTE: This method does not include all of the specifications (e.g., equipment and supplies) and procedures (e.g.,

More information

Ultrahydrogel Columns

Ultrahydrogel Columns CONTENTS I. INTRODUCTION II. INSTALLATION a. Solvents b. Column c. Organic solvents d. Aqueous salt and buffer solution e. Flow rate f. Solvent and sample preparation g. Guard columns h. Column installation

More information

ChungHo Dealer Technical Guide

ChungHo Dealer Technical Guide ChungHo Dealer Technical Guide Recommendations Prior to Installation -Flush out the system -Check for leaks / hose connections -Test Board connections Installation Procedures -Basic POU Installation -Do

More information

ODYSSEY FLOOR MODEL. Owner s Manual. O-6400-F shown rev 118

ODYSSEY FLOOR MODEL. Owner s Manual. O-6400-F shown rev 118 Owner s Manual ODYSSEY FLOOR MODEL O-6400-F shown 67-1455 rev 118 3730 E. Southern Avenue, Phoenix, AZ 85040 USA 800-778-8779 Workhorseproducts.com 1 Table of Contents I. Introduction & Safety Information.

More information

// XPLORER-NS. Full range Total Nitrogen, Total Sulfur instrument for the modern Laboratory.

// XPLORER-NS. Full range Total Nitrogen, Total Sulfur instrument for the modern Laboratory. // XPLORER-NS Full range Total Nitrogen, Total Sulfur instrument for the modern Laboratory. TE Instruments has developed the Xplorer-NS, a Total Nitrogen and Total Sulfur combustion analyzer, offering

More information

Hydrocarbon Processing Industry Gas Chromatographs. Standard GC Analyzer Solutions

Hydrocarbon Processing Industry Gas Chromatographs. Standard GC Analyzer Solutions Hydrocarbon Processing Industry Gas Chromatographs Standard GC Analyzer Solutions Scion Instrument s GC Analyzer specialists have the knowledge and experience to provide pre-configured gas chromatographic

More information

Agilent G2615A Bar Code Reader

Agilent G2615A Bar Code Reader Agilent G2615A Bar Code Reader User Information Agilent Technologies Notices Agilent Technologies, Inc. 2002 No part of this manual may be reproduced in any form or by any means (including electronic storage

More information

Operating Instructions

Operating Instructions Operating Instructions Series B37000 Hose Reels BB37118 L CB37118 L EB37118 L12D BB37122 L CB37122 L EB37122 L12D BB37128 L CB37128 L EB37128 L12D Dimensional Data C A B D Part # A B C D BB/CB37118 L 31

More information

Superiorex ODS http://hplc.shiseido.co.jp/e/column/html/sp_index.htm Page 1 of 2 List of Sales Representatives Contact Shiseido Technical Materials Catalog List HPLC Columns HPLC Instruments Large-Diameter

More information

Product data sheet DUMATHERM N Pro. for the automatic serial determination of NITROGEN

Product data sheet DUMATHERM N Pro. for the automatic serial determination of NITROGEN for the automatic serial determination of NITROGEN DUMATHERM N Pro Automated system for fast nitrogen determination acc. to Dumas Technical Data DUMATHERM N Pro Data DUMATHERM N Pro Sample size 0,5 mg

More information