Food Contaminants and Residue Analysis
|
|
- Lauren Moore
- 5 years ago
- Views:
Transcription
1 Food Contaminants and Residue Analysis by Yolanda Picó Laboratori de Bromatologia i Toxicologia, Facultat de Farmàcia, Universitat de València, Burjassot, València, Spain. Publisher: Elsevier Science Ltd, 2008 Series: Comprehensive Analytical Chemistry Chapter 3 Guidelines on quality implementation for analytical methods Sergio Caroli Istituto Superiore di Sanità, Viale Regina Elena 299, Rome. caroli@iss.it Introduction The concept that no laboratory can survive in the long term without having a fit-forpurpose quality system in place is nowadays widely accepted by the international scientific community. Experimental information must be, in fact, supported by documented evidence to be valid, credible and comparable. To date, quality systems can be traced back basically to two distinct, yet complementary, approaches, i.e., those adopting the criteria for accreditation, mostly those worked out by the International Standardisation Organisation (ISO) and those based on compliance with the principles of Good Laboratory
2 Practice (GLP), in particular as developed by the Organisation for Economic Co-operation and Development (OECD) (1 3). The rationale behind the latter roots in the need for assessing the integrity of experimental studies, while the former aims at evaluating the competence of a laboratory to perform measurements. In other words, in a GLP system the validity of a completed study is challenged. In so doing, necessarily all the phases of the study are reconstructed and the laboratory performance and facilities are checked. On the other hand, in the case of a quality system based on accreditation, the target is to gain confidence in the ability of the laboratory to generate defendable experimental data. More in detail, a GLP system aims at providing the decision maker with reliable experimental information on new chemical substances so as to allow for a sound assessment of the benefit-to-risk ratio well before chemicals are produced and marketed. The cornerstone of this policy is the implementation of the OECD GLP principles whenever non-clinical safety studies on new chemicals are undertaken by test facilities. Such principles can be found in the Decision of the Council concerning the Mutual Acceptance of Data in the Assessment of Chemicals [C(81)30(Final)], the Council Decision- Recommendation on Compliance with Principles of GLP [C(89)87(Final)] and the Council Needless to say, the GLP quality system is compulsory resorted to whenever there is, e.g., a registration obligation for the commercialization of a new substance. Decision concerning the Adherence of Non-member Countries to the Council Acts related to the Mutual Acceptance of Data in the Assessment of Chemicals [C(97)114(Final)]. In turn, an accreditation-oriented quality system is governed by the ISO/IEC Standard and has to cover the administrative and technical issues specified in the Standard, including internal audits, job descriptions and responsibilities, procedures for equipment/instrument maintenance and calibration, document control, handling of reagents, chemicals and reference materials, sample delivery and storage, validation of test methods, traceability and uncertainty of the test results, training of personnel, client complaints and corrective and preventive actions. Several of these issues are also required for compliance with the OECD GLP principles either with a different emphasis or with additional requirements. Although the two systems have been designed to meet largely different needs, it is hoped that they can more and more support each other to minimize redundancy and to provide the end user with experimental information as trustable as possible The role of quality
3 General aspects Important decisions are often taken on the basis of experimental data. Hence, it is crucial that such data be comparable, reliable and valid. No laboratory can in fact be run without a fit-for-purpose quality system in place. Quality has been defined by ISO as The totality of features and characteristics of a product or service that bear on its ability to satisfy stated or implied needs. To date, quality systems are basically inspired either by the Good Laboratory Practice (GLP) principles or by the accreditation criteria. Laboratory work may be of two different types: i) the outcome of the investigation are exact figures, to which precision and reproducibility are expected to be attached; ii) the outcome of the investigation is, in a general sense, complex information which should be credible, reliable and comparable. In the former case, what matters more are the experimental measurements. In this context quality is assessed in terms of precision and reproducibility of the numerical data obtained. The ability of the laboratory to generate such data is thus of primary importance. Quality systems based on accreditation criteria are ideal in this respect. In the latter case, the focus is on the overall study as such. Third parties should be enabled to reconstruct the whole course of the study and to check its integrity so that confidence can be gained in the way the study results have been obtained. Under such circumstances, quality systems based on the GLP principles do apply. Which approach is to be preferred depends only on the scope and goals of the activities performed in the laboratory, although it should not be overlooked that accreditation is basically voluntary, whereas the GLP system is prescribed by law for those Test Facilities (TFs) undertaking non-clinical safety studies. There is still some confusion surrounding the terms of accreditation and certification. As this may well misleading, consensus has been reached on the following definitions: accreditation is a means used to identify competent testing laboratories, whereas certification is the official approval granted by a given authority Key aspects of a quality system based on the accreditation criteria As set forth by the IUPAC, The international scientific community recognizes that a laboratory must take appropriate measures to ensure that it is capable of providing
4 data of the required quality. Such measures include: i) internal quality control procedures; ii) participation in proficiency testing schemes; iii) validated methods of analysis; iv) accreditation to an international standard. (4). Accreditation-based quality systems are governed by the international Standard ISO/IEC (5). This standard exploits the extensive experience gained in implementing the ISO/IEC Guide 25 and EN norms and replaces them both. The ISO/IEC Standard sets forth the requirements a laboratory has to meet to be recognized as competent to carry out tests and/or calibrations, including sampling. The pillars of an accreditation system are listed in Table 1. Method validation is central to the accreditation process as reliability and comparability of data are crucial to perform experimental meaningful tests and to achieve credible results which can be profitably used by the client, i.e., the end-user. It should be noted that the overall validation process covers all of the pivotal phases of an experimental measurement and not only the mere quantification step, as illustrated in Fig. 1. In turn, method validation as such should at least cover the parameters given in Table Key aspects of a quality system based on the principles of good laboratory practice In the early 1960 s the US Food and Drug Administration (FDA) became aware that some studies on the safety of new chemicals performed by TFs for regulatory purposes were basically unreliable. Evidence was in fact provided of major adverse effects of such substances which had not been reported at the time when the authorization to production and commerce was granted. In the early 1970 s the US Congress undertook the reassessment of studies submitted by some TFs to Regulatory Authorities (RAs) and suspected to be fraudolent. Under such conditions thousands and thousands of safety studies on industrial chemicals, pesticides, herbicides, drugs, cosmetics, and food and feed additives were conducted for years (about % of all toxicological studies authorized in the USA in that period). As an example, an article published by The Washington Post in 1997 is shown in Fig. 2. Senator Edward Kennedy declared at the US Congress of January 20, 1976, that unreliable, undocumented and fraudolent research is the most frightening menace to the health and safety of people. That research be wrong because of technical problems or
5 because of the lack of competence or even due to criminal negligence is less important than the very fact that it is wrong The principles GLP were conceived to harmonize the conduct of non-clinical safety studies and to minimize the risk of fraud. Since the early years, this matter became a priority for the Organisation for Economic Co-operation and Development (OECD) which set up the GLP principles in order to promote and manage the mutual acceptance of nonclinical safety studies in the Member Countries. According to OECD, the principles of GLP are a quality system concerned with the organizational process and the conditions under which safety studies are planned, conducted, controlled, recorded, reported and archived. In practice, they form a body of reciprocally dependent documented items that make the falsification of a study more time-consuming and expensive than its actual correct performance. The three major acts of OECD in the field of GLP are as follows: i) Decision of the Council concerning the Mutual Acceptance of Data (MAD) in the Assessment of Chemicals [C(81)30(Final)]; ii) Council Decision-Recommendation on Compliance with Principles of Good Laboratory Practice; iii) [C(89)87(Final)] Council Decision concerning the Adherence of Non-member Countries to the Council Acts related to the Mutual Acceptance of Data in the Assessment of Chemicals [C(97)114(Final)]. As a part of the permanent activities of its Environment, Health and Safety Programme, the OECD also prepares and publishes Test Guidelines for Chemical Substances to be used when performing GLP studies and thus enhance their reliability. The Series on the GLP principles and compliance monitoring consists at present of 14 monographs, as detailed in Table 3, whereas Table 4 details the pillars of a GLP system (6, 7). These guides form the core of the legal provisions of the European Union in the field of GLP (8, 9) Conclusions The two quality systems have been conceived to meet quite different needs. In other words, the accreditation criteria are designed to manage activities in a laboratory where routine quantitative measurements (such as analytical determinations) are carried out, whereas the GLP principles are intended to guarantee the integrity of data generated in
6 non-clinical safety studies. Their respective fundamental characteristics are summarized in Table 5. From this standpoint, it is worth mentioning that, e.g., the GLP system prescribes that the Director of the TF, the person responsible of the Quality Assurance Unit, the Study Director and the Archivist be all independent of each other to fully guarantee the fair conduct of the study, while in the case of the accreditation system the first two functions can coincide and the third one does not exist. On the other hand, in the accreditation system, it is imperative to have a quality manual, which in turn is not formally requested in the GLP system, although in the latter the Standard Operating Procedures play basically the same role. Moreover, a study plan, mandatory in the GLP system, is not needed in the accreditation one, not to speak of the fact that management of complaints and participation in proficiency testing is mandatory in the latter, but not necessary in the former. As regards validation of methods, the GLP system requires that validated methods are in place, but does not impose that such methods are set up according to the GLP principles, any other fit-for-purpose quality system being acceptable to this end. All this provides clear evidence of the profound diversity in the approaches and goals of the two systems, although some common aspects are present. In this regard, in recent years, the OECD has established a dialogue group to verify where the two systems can actually interact, thus minimizing useless duplication of efforts. In conclusion, the selection of the quality system to be adopted should be carefully made on the basis of the prevailing activities carried out in the laboratory. Quality is inescapable, but it has a cost: a wrong decision can only lead to failure. References 1. Ph. Quevauviller, E. A. Maier. B. Griepink (Eds.), Quality Assurance for Environmental Analysis, Elsevier, 1995, pp. xx J. P. Seiler, Good Laboratory Practice. The Why and the How, Springer, 2001, pp. x
7 3. Ph. Quevauviller (Ed.), Quality Assurance in Environmental Monitoring. Sampling and Sample Pretreatment, 1995, VCH, pp. xv IUPAC Technical Report, Harmonized Guidelines for Single-Laboratory. Validation of Methods of Analysis. 5. ISO/IEC 17025, General Requirements for the Competence of Testing and Calibration Laboratories. 6. OECD Series on Principles of Good Laboratory Practice and Compliance Monitoring, Number 1, OECD Principles on Good Laboratory Practice (as revised in 1997), ENV/MC/CHEM (98) S. Caroli (Ed.), The New Principles of Good Laboratory Practice: Priorities, Problems, Perspectives, Ann. Ist. Super. Sanità, 38 (2002), pp Directive 2004/9/EC of the European Parliament and Council, 11 February 2004, Off. J. EU, L50 (2004), Directive 2004/10/EC of the European Parliament and Council, 11 February 2004, Off. J. EU, L50 (2004),
8 Table 1. Key aspects of a laboratory compliant with an accreditation system. Service to the client Policy for complaints Control of non-conformities Quality manual Management of records Internal audits Measurement traceability Motivation of personnel Laboratory setting Validation of methods Equipment Management reviews Test and calibration items Report of results
9 Table 2. Parameters to be ascertained to validate an analytical method. Applicability Selectivity Calibration and linearity Trueness Accuracy Precision Recovery Range Limit of detection Limit of quantification Sensitivity Ruggedness Robustness Fitness for purpose Matrix variation Measurement uncertainty
10 Table 3. The OECD series on the GLP principles and compliance monitoring. No. 1. OECD Principles of Good Laboratory Practice. No. 2. Revised Guides for Compliance Monitoring Procedures for Good Laboratory Practice (1995). No. 3. Revised Guidance for the Conduct of Laboratory Inspections and Study Audits (1995). No. 4. Quality Assurance and GLP (as revised in 1999). No. 5. Compliance of Laboratory Suppliers with GLP Principles (as revised in 1999). No. 6. The Application of the GLP Principles to Field Studies (as revised in 1999). No. 7. The Application of the GLP Principles to Short Term Studies (as revised in 1999). No. 8. The Role and Responsibilities of the Study Director in GLP Studies (as revised in 1999). No. 9. Guidance for the Preparation of GLP Inspection Reports (1995). No. 10. The Application of the Principles of GLP to Computerised Systems (1995). No. 11. The Role and Responsibilities of the Sponsor in the Application of the Principles of GLP (1999). No. 12. Requesting and Carrying Out Inspections and Study Audits in Another Country (2000). No. 13. The Application of the OECD Principles of GLP to the Organisation and Management of Multi-site Studies (2002). No. 14. The Application of the OECD GLP Principles to in vitro Studies (2004). No. 15. Establishment and Control of Archives That Operate in Compliance with the Principles of Good Laboratory Practice (2007)
11 Table 4. Key aspects of a Test Facility compliant with a GLP system. Director of the Test Facility Study Director Quality Assurance Unit Archivist Sponsor Test and reference items Study plan Final report Standard Operative Procedures Test Site (if applicable) Principal Investigator (if applicable)
12 Table 5. Key elements of the accreditation and GLP system. Accreditation quality system GLP quality system Overlapping aspects Management of complaints Master schedule Management Uncertainty of measurements Study director Motivation Proficiency testing Archivist Training Preventive actions Quality assurance unit Reference materials Service to the client Study plan Equipment and Sampling Test article maintenance Reports Method validation Chain of custody Quality control procedures Corrective action Audits Sample reception
13 Hardware validation Software validation/qualification Method validation System suitability Fig. 1. Major steps of the validation process. VALIDATION
14 Fig. 2. Article published by The Washington Post in 1974 on the falsification of experimental data
Overview of Good Food Laboratory Practices
Overview of Good Food Laboratory Practices Dr. Anne Bridges Technical Director, AACC International International Workshop & Training Program On Good Food Laboratory Practices Jointly Organized By Food
More informationA comparison of the complimentary and different issues in ISO/IEC and OECD GLP
Accred Qual Assur (2003) 8:208 212 DOI 10.1007/s00769-003-0589-9 GENERAL PAPER Tina Engelhard Etty Feller Zahava Nizri A comparison of the complimentary and different issues in ISO/IEC 17025 and OECD GLP
More informationGeneral Accreditation Criteria Equipment assurance, in-house calibration and equipment verification
General Accreditation Criteria Equipment assurance, in-house calibration and equipment verification January 2018 Copyright National Association of Testing Authorities, Australia 2009 This publication is
More informationGuidance on the Application. of ISO / IEC Accreditation International Association for Certifying Bodies
Accreditation International Association for Certifying Bodies Guidance on the Application of ISO / IEC 17020 Guidance on the Application of ISO/IEC 17020 Page 1 of 16 Introduction This guidance document
More informationPALCAN Guidance for the Validation of Test Methods
PALCAN Guidance for the Validation of Test Methods CAN-P-1629 PALCAN GUIDANCE FOR THE VALIDATION OF TEST METHODS POLITIQUE DU PALCAN CONCERNANT LA VALIDATION DES MÉTHODES D ESSAIS CAN-P-1629 Copyright
More informationLaboratory Quality Assurance Manager & Laboratory Assessor RULES & HANDBOOK
EOQ Personnel Registration Scheme Laboratory Quality Assurance Manager & RULES & HANDBOOK Prepared by: Dr. Eugenia Soboleva, Quality Austria In accordance with the working group on EOQ product development
More informationQuality System ACCREDITATION For the Clinical Laboratories
Quality System ACCREDITATION For the Clinical Laboratories A NEW PERSPECTIVE BY: Accreditation Commission For Conformity Assessment Bodies (ACCAB) 1 Objective What are Standards? Voluntary & Mandatory
More informationAnimal Products Notice
Animal Products Notice Specifications for Laboratories.. An animal products notice issued under the Animal Products Act 1999 TITLE Animal Products Notice: Specifications for Laboratories COMMENCEMENT This
More informationEA Guidance on Accreditation of Pesticide Residues Analysis in Food and Feed
Publication Reference EA-4/22 G: 2018 EA Guidance on Accreditation of Pesticide Residues Analysis in Food and Feed PURPOSE This paper provides guidance to develop harmonized accreditation of Pesticide
More information1 INTRODUCTION TO QUALITY MANAGEMENT
1 INTRODUCTION TO QUALITY MANAGEMENT This section will introduce you to the basic concepts and definitions in Quality Management applied to XRF. The scope and contents of the Standards from the ISO 9000
More informationISO 9001:2000 What does it mean in the supply chain?
ISO 9001:2000 What does it mean in the supply chain? What is ISO 9001:2000? ISO 9001:2000 is an international standard that gives requirements for an organization s Quality Management System ( QMS ). It
More informationTECHNICAL DIRECTIVE PART ONE: STANDARDIZATION AND RELATED ACTIVITIES GENERAL VOCABULARY
SAUDI ARABIAN STANDARDS ORGANIZATION (SASO) TECHNICAL DIRECTIVE PART ONE: STANDARDIZATION AND RELATED ACTIVITIES GENERAL VOCABULARY D7-24 5-2004 FOREWORD This Part of SASO s Technical Directives is Adopted
More informationOMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT
OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT PA/PH/OMCL (13) 113 2R Evaluation and Reporting of Results Core document Full document title and reference Evaluation and Reporting of Results
More informationThe importance of valid analytical measurements
The importance of valid analytical measurements The importance of valid analytical measurements Why we need reliable measurements Analytical measurements made in laboratories are always required to fulfil
More informationILAC P10 TRACEABILITY
ACTIVITY 1.3 TRAINING SESSION FOLLOWING THE FINDINGS FROM PREVIOUS ACTIVITIES: ILAC P10 TRACEABILITY Maria Pia Toni Slide 1 of XX REFERENCE ILAC P10:2013 ILAC Policy on the Traceability of Measurement
More informationDECISION TO ACCREDIT
DECISION TO ACCREDIT Determine the requirements of the regulator, legislation and customer s needs prior to deciding on the standard to implement. Process is based on the following: Top management s decision
More informationRules for accreditation of providers of proficiency testing/interlaboratory
Rules for accreditation of providers of proficiency testing/interlaboratory tests Scope: With this regulation the requirements for accreditations of providers of proficiency testing/interlaboratory tests/ring
More informationA2LA. R231 Specific Requirements: Threat Agent Testing Laboratory Accreditation Program. December 6, 2017
Laboratory Page 1 of 17 Laboratory December 6, 2017 2017 by A2LA All rights reserved. No part of this document may be reproduced in any form or by any means without the prior written permission of A2LA.
More informationTitle Procedure for setting up the accreditation of new conformity assessment schemes
Title Procedure for setting up the accreditation of new conformity assessment schemes Reference PG-13-01 Revision 1 Date 09-04-2018 NOTE: The present document represents the English version of document
More informationEstablishment of an Accredited Reference Measurement Laboratory
Establishment of an Accredited Reference Measurement Laboratory Francesco Dati, PhD IVD-Consulting Marburg / Germany E-mail: f.dati@t-online.de Quality of Analytical Systems Reference Measurement Systems
More informationBEGINNERS GUIDE TO ISO 9001 : 2000
INTRODUCTION ISO 9000 is a standard for Quality Management Systems. It is a series of documents that are designed to help organisations achieve better quality of product and service. This guide is split
More informationDutch Accreditation Council (RvA) Explanatory notes on EN-ISO/IEC 17025
Dutch Accreditation Council (RvA) Explanatory notes on EN-ISO/IEC 17025 Document code: RvA-T015-UK Version 6, 13-4-2018 Rv A explanatory notes describe the policy and/or the procedures of the Rv A concerning
More informationEA Procedure and Criteria for the Evaluation of Conformity Assessment Schemes by EA Accreditation Body Members
Publication Reference EA-1/22 A-AB: 2014 EA Procedure and Criteria for the Evaluation of Conformity Assessment Schemes by EA PURPOSE This document contains the procedure and criteria to be applied by EA
More informationGLP/SC/01 Basic statistical tools for analytical chemistry (2 days)
GLP Consulting http://consultglp.com Courses on offer (See outlines below) GLP/SC/01 Basic statistical tools for analytical chemistry (2 days) GLP/SC/02 The GUM bottom-up evaluation technique of measurement
More informationSpecific Accreditation Criteria
Specific Accreditation Criteria ISO/IEC 17025 Application Document Materials - Annex Characterisation of industrial materials - General July 2018 Copyright National Association of Testing Authorities,
More informationMODULE 2A GENERAL MANAGEMENT SYSTEM REQUIREMENTS
MODULE 2A GENERAL MANAGEMENT SYSTEM REQUIREMENTS 2A.1 SCOPE (See ISO/IEC 17025:2005, Section 1) Laboratories shall meet all requirements of the ISO/IEC 17025:2005 International Standard and other AIHA-LAP,
More informationREQUIREMENTS FOR ACCREDITATION OF INSPECTION BODIES
NATIONAL CENTRE FOR ACCREDITATION REQUIREMENTS FOR ACCREDITATION OF INSPECTION BODIES according to SM SR EN ISO/IEC 17020:2013 Code DR-OI-07 Edition 2 Page 1/15 Approved by Technical Certification Committee
More informationGood Laboratory Practice (GLP) Philip Cunningham St Vincent s s Hospital Sydney
Good Laboratory Practice (GLP) Philip Cunningham St Vincent s s Hospital Sydney Pharmaceutical clinical trials may require research or non-routine routine laboratory test results for study endpoints Why?
More informationEA Procedure and Criteria. For the Evaluation of Conformity. Assessment Schemes by EA. Accreditation Body Members
Schemes by EA Accreditation Body Members Publication Reference EA-1/22 A: 2016 EA Procedure and Criteria For the Evaluation of Conformity Assessment Schemes by EA Accreditation Body Members PURPOSE This
More informationGMP, GLP OR ISO 17025: HOW DO THESE APPLY TO OUTSOURCED ANALYTICAL TESTING?
GMP, GLP OR ISO 17025: HOW DO THESE APPLY TO OUTSOURCED ANALYTICAL TESTING? Joe Grappin, MS, EAG Laboratories 2018 MD&M West Tech Theater 1 OVERVIEW Background: What Led to the Presentation History of
More informationEU food safety legislation and official controls
EU food safety legislation and official controls A Seminar on the EU Official Control Regulation, Food and Nutrition Labelling Bangkok, 8 June 2018 Dr. Tsviatko Alexandrov (DVM, PhD) Bulgarian Food Safety
More informationACCREDITATION A TOOL TO SUPPORT REGULATORS
ACCREDITATION A TOOL TO SUPPORT REGULATORS ACCREDITATION 5 CONFORMITY ASSESSMENT 9 ASSURANCE OF COMPETENCE 13 EA, EUROPEAN CO-OPERATION FOR ACCREDITATION 17 INTERNATIONAL AGREEMENTS 19 PUBLIC AND PRIVATE
More informationCurrent quality status of CABETs environmental laboratory and necessary steps to meet GLP requirements
Current quality status of CABETs environmental laboratory and necessary steps to meet GLP requirements June 18 22, 2007 Paul de Bie (Alterra) Jiang Hui (ICAMA) Workshop Report WP1 1 Beijing / Wageningen,
More informationAPLAC GUIDANCE ON REFERENCE MATERIAL USE AND PRODUCTION
APLAC GUIDANCE ON REFERENCE MATERIAL USE AND PRODUCTION Issue 6 Issue Date: 2018/02 Page 1 of 18 PURPOSE Reference Materials (RMs) are used in all stages of the measurement process, for method validation,
More informationACCREDITATION A TOOL TO SUPPORT REGULATORS
ACCREDITATION A TOOL TO SUPPORT REGULATORS ACCREDITATION CONFORMITY ASSESSMENT ASSURANCE OF COMPETENCE EA, EUROPEAN CO-OPERATION FOR ACCREDITATION INTERNATIONAL AGREEMENTS PUBLIC AND PRIVATE BENEFITS
More informationACCREDITATION A TOOL TO SUPPORT REGULATORS
ACCREDITATION A TOOL TO SUPPORT REGULATORS ACCREDITATION CONFORMITY ASSESSMENT ASSURANCE OF COMPETENCE EA, EUROPEAN CO-OPERATION FOR ACCREDITATION INTERNATIONAL AGREEMENTS PUBLIC AND PRIVATE BENEFITS
More informationEnsuring Food Safety. Through Accredited Third-Party Conformity Assessment. An ANSI-ASQ National Accreditation Board White Paper
Ensuring Food Safety Through Accredited Third-Party Conformity Assessment An ANSI-ASQ National Accreditation Board White Paper Executive Summary According to a February 2007 Government Accountability Office
More informationA2LA. R104 - General Requirements - ISO-IEC Accreditation of Field Testing and Field Calibration Laboratories
and Field Calibration Page 1 of 15 R104 - General Requirements - ISO/IEC 17025:2017 Accreditation of Field Testing and Field Calibration 2018 2018 by A2LA All rights reserved. No part of this document
More information1201 Maryland Avenue SW, Suite 900, Washington, DC ,
1201 Maryland Avenue SW, Suite 900, Washington, DC 20024 202-962-9200, www.bio.org February 22, 2011 Dockets Management Branch (HFA-305) Food and Drug Administration 5600 Fishers Lane, Rm. 1061 Rockville,
More informationMeasuring Instruments Directive 2014/32/EU Assessment of Notified Bodies in Charge of Type Examination Presumption of Conformity based on EN 17065
WELMEC 8.5, 2017 Measuring Instruments Directive 2014/32/EU Assessment of Notified Bodies in Charge of Type Examination Presumption of Conformity based on EN 17065 For information: This guide is available
More informationOECD Series on Principles of GLP and Compliance Monitoring Number 7 (Revised)
Unclassified ENV/JM/MONO(99)23 ENV/JM/MONO(99)23 Or. Eng. Unclassified Organisation de Coopération et de Développement Economiques OLIS : 14-Sep-1999 Organisation for Economic Co-operation and Development
More informationISCC 204 AUDIT REQUIREMENTS AND RISK MANAGEMENT. Version 3.0
ISCC 204 AUDIT REQUIREMENTS AND RISK MANAGEMENT Version 3.0 II Copyright notice 2016 ISCC System GmbH This ISCC document is protected by copyright. It is freely available from the ISCC website or upon
More informationGeneral Accreditation Guidance. ISO/IEC 17025:2017 Gap analysis. April 2018
General Accreditation Guidance Gap analysis April 2018 Copyright National Association of Testing Authorities, Australia 2018 This publication is protected by copyright under the Commonwealth of Australia
More informationA016 Traceability of measurement results compared to national and international measurement standards
08.04.2016 Version 08 Page 1 de 9 A016 Traceability of measurement results compared to national and international measurement standards Modifications: p 2 South Lane Tower I 1, avenue du Swing L-4367 Belvaux
More informationGeneral requirements for the competence of testing and calibration laboratories. In this presentation:
General requirements for the competence of testing and calibration laboratories ISO/IEC 17025:2017 In this presentation: Explanation New requirement Interpretation / Examples / Questions Agenda Welcome
More informationGuidance for Swedac s assessors on the assessment of laboratories based on the requirements set out in SS-EN ISO/IEC 17025:2005
SWEDAC DOC 02:5 Version 10 2016-05-16 Guidance for Swedac s assessors on the assessment of laboratories based on the requirements set out in SS-EN ISO/IEC 17025:2005 Swedac, Swedish Board for Accreditation
More informationGuidelines for the assessment of the appropriateness of small interlaboratory comparisons within the process of laboratory accreditation
Publication Reference EA-4/21 INF: 2018 Guidelines for the assessment of the appropriateness of small interlaboratory comparisons within the process of laboratory accreditation PURPOSE This paper provides
More informationPRINCIPLES FOR THE USE OF SAMPLING AND TESTING IN INTERNATIONAL FOOD TRADE CAC/GL Adopted in Revision: 2015.
PRINCIPLES FOR THE USE OF SAMPLING AND TESTING IN INTERNATIONAL FOOD TRADE CAC/GL 83-2013 Adopted in 2013. Revision: 2015. CAC/GL 83-2013 2 1. INTRODUCTION Sampling and testing are, among others, procedures
More informationIL AC-G22:2004. Use of Proficiency Testing as a Tool for Accreditation in Testing
IL AC-G22:2004 Use of Proficiency Testing as a Tool for Accreditation in Testing Copyright ILAC 2004 ILAC encourages the authorised reproduction of its publications, or parts thereof, by organisations
More informationMark Citriglia Manager of Analytical Services Northeast Ohio Regional Sewer District
Mark Citriglia Manager of Analytical Services Northeast Ohio Regional Sewer District Licensure Non-voluntary process by which a government agency regulates and grants permission for an individual to engage
More informationINTERNATIONAL FEDERATION OF INSPECTION AGENCIES CODE OF PRACTICE CONSUMER PRODUCT TESTING RELEASE st May 2018
INTERNATIONAL FEDERATION OF INSPECTION AGENCIES CODE OF PRACTICE CONSUMER PRODUCT TESTING RELEASE 1.0 1 st May 2018 Consumer Product Testing Code of Practice - Release 1.0-1st May 2018 Page 1 of 8 EXECUTIVE
More informationCOMMISSION RECOMMENDATION. of XXX. on the audits and assessments performed by notified bodies in the field of medical devices
EUROPEAN COMMISSION Brussels, XXX [ ](2013) XXX draft COMMISSION RECOMMENDATION of XXX on the audits and assessments performed by notified bodies in the field of medical devices (Text with EEA relevance)
More informationSRI RESEARCH/RATING QUALITY STANDARD
April 2008, CSRR-QS 2.1, page 1/15 Association for Independent Corporate Sustainability & Responsibility Research SRI RESEARCH/RATING QUALITY STANDARD CSRR-QS 2.1 Description: CSRR-QS 2.1 is a quality
More informationKey Elements of Quality Assurance in the Good Laboratory Practice (GLP) Environment
Key Elements of Quality Assurance in the Good Laboratory Practice (GLP) Environment The principles of Good Laboratory Practice (GLP) define a set of rules and criteria for a quality system concerned with
More informationQuality & Safety GLP. 44 q&more 02/14
Quality & Safety GLP 44 q&more 02/14 More than keeping a lab notebook GLP regulations and their new role in the approval of medicines for advanced therapies (ATMPs) Dr Katja Schellenberg, Translation Centre
More informationHandbook ISO/IEC 17025:2017
Handbook ISO/IEC 17025:2017 Page 1 of 32 General considerations This new standard ISO/IEC 17025 includes some noteworthy changes related to its structure and scope that should be mentioned before we go
More informationSpecific Accreditation Criteria Materials ISO/IEC Annex. Cement testing
Specific Accreditation Criteria Materials ISO/IEC 17025 Annex Cement testing January 2018 Copyright National Association of Testing Authorities, Australia 2013 This publication is protected by copyright
More informationc) Have personnel been appointed to supervise the production operations across all shifts in order to ensure the product quality?
Factory s Quality Assurance Ability 1 Responsibilities and Resources 1.1 Responsibilities a) Have the responsibilities and interrelation of various personnel involved in quality activities been defined?
More informationAmerican Association for Laboratory Accreditation
ISO 15189 Page 1 of 17 Accredited clinical laboratories are required to meet the following additional requirements as contained in P905 - A2LA Metrological Traceability Policy for ISO 15189 Laboratory
More informationSam Chappell, CIML Honorary Member, USA
22 OPPORTUNITIES AND FUTURE TRENDS IN LEGAL METROLOGY CONTROL OF MEASURING INSTRUMENTS Sam Chappell, CIML Honorary Member, USA Introduction Currently, legal metrology control generally includes type evaluation
More informationGood Clinical Practice (GCP): how does a laboratory that supports a clinical trial ensure compliance with these Regulations
Good Clinical Practice (GCP): how does a laboratory that supports a clinical trial ensure compliance with these Regulations 9th edition of Synevo Clinical Research Symposium. 17 th November 2017 Bucharest.
More informationPROTOCOLS: WHY BOTHER?
PROTOCOLS: WHY BOTHER? UC CEIN Student & Postdoc Leadership Workshop September 28, 2013 Have you ever used a protocol? When are protocols useful? Purpose of Protocols Protocols are the driving force behind
More informationWELMEC European cooperation in legal metrology
WELMEC 8.6 Issue 1 WELMEC European cooperation in legal metrology Measuring Instruments Directive 2004/22/EC Presumption of Conformity of the Quality System of Manufacturers with Module D or H 1 when EN
More informationSupport for the National Accreditation Centre MOLDAC to successfully undergo the EA peer evaluation process. Twinning Project MD14/ENPI/TR/20
Activity1.5 ME Work in newareas November 23-27 2015 Anna A. Sampò Slide 1of 36 Wednesday November 23 Slide 2of 36 Medical device A medical device is "an instrument, apparatus, implement, machine, implant,
More informationQuality Manual. Index
This quality policy manual is for the use of, its Clients, Vendors, and the appropriate Regulatory Agencies. Unauthorized duplication or transfer to others is prohibited. Index Authorization... 3 Introduction...
More informationJournal of Chemical and Pharmaceutical Research
Available online www.jocpr.com Journal of Chemical and Pharmaceutical Research ISSN No: 0975-7384 CODEN(USA): JCPRC5 J. Chem. Pharm. Res., 2011, 3(6):88-98 Conduct of Inspections for Pharmaceutical Manufactures
More informationTerms and conditions for laboratories appointed to undertake the testing of compost under the Compost Certification Scheme aligned to PAS 100 and the
Terms and conditions for laboratories appointed to undertake the testing of compost under the Compost Certification Scheme aligned to PAS 100 and the Compost Quality Protocol Renewable Energy Assurance
More informationGreen Product Mark Certification Scheme
Green Product Mark Certification Scheme www.tuv.com FOREWORD Product environmental labels are claims regarding the environmental impact of a product. They provide information about the product s overall
More informationCO-ORDINATION OF NOTIFIED BODIES PPE Regulation 2016/425 RECOMMENDATION FOR USE
CO-ORDINATION OF NOTIFIED BODIES PPE Regulation 2016/425 PPE-R/00.018 Version 1 RECOMMENDATION FOR USE Number of pages: 5 Approval stage : Approved on : Origin : Horizontal Committee, C2D Ad hoc group
More informationOMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT
OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT PA/PH/OMCL (05) 49 DEF CORR UNCERTAINTY OF MEASUREMENT - PART 1 Full document title and reference Document type Legislative basis Uncertainty
More informationPolicy on Measurement Traceability
Policy on Measurement Traceability Date of Issue: January 1, 2006 Voluntary EMC Laboratory Accreditation Center, Inc. (VLAC) This document is copyrighted under the Copyright Act. Unauthorized reproduction,
More informationAPLAC GUIDELINES FOR FOOD TESTING LABORATORIES
APLAC GUIDELINES FOR FOOD TESTING LABORATORIES Issue No. 2 Issue Date: 01/08 Page 1 of 13 PURPOSE This document provides guidance to APLAC members and their accredited laboratories on the operation of
More informationStandards and accreditation. Tools for delivering better regulation
Standards and accreditation Tools for delivering better regulation Introduction Standards and accreditation are marketbased tools that can be used by Government policy makers to deliver better regulation.
More informationEA-7/04 Legal Compliance as a part of accredited ISO 14001: 2004 certification
Publication Reference EA-7/04 Legal Compliance as a part of Accredited ISO 14001: 2004 certification PURPOSE The text of this document has been produced by a working group in the European co-operation
More informationThe Importance of Validated Rapid Methods for International Trade
The Importance of Validated Rapid Methods for International Trade 3M 2009. All Rights Reserved. Robert Koeritzer Sr. Technical Manager 3M Food Safety Past-President, AOAC International Agenda Method validation
More informationThe Third-Party Process: Waste of Resources or Added Value?
The Third-Party Process: Waste of Resources or Added Value? Discussing accredited certification at a cocktail party isn t likely to make you the center of attention. But depending on where you work, you
More informationINTRODUCTION TO QUALITY ASSURANCE
INTRODUCTION TO QUALITY ASSURANCE Basic Quality Terminology No more definitions...please! Page 1 Quality is the degree to which a specific product conforms to a design or specification H.L. Gilmore Product
More informationPALCAN Policy on Use of Proficiency Testing as a Tool for Accreditation in Testing (ILAC-G22:2004)
PALCAN Policy on Use of Proficiency Testing as a Tool for Accreditation in Testing (ILAC-G22:2004) CAN-P-1624 PALCAN POLICY ON USE OF PROFICIENCY TESTING AS A TOOL FOR ACCREDITATION IN TESTING (ILAC-G22:2004)
More informationQSS 0 Products and Services without Bespoke Contracts.
QSS 0 Products and Services without Bespoke Contracts. Use: Typically Sub 50k processes without a bespoke contract. Amendment History Version Date Status V3 June 2018 Updated for 2018 deployment Contents
More informationCOMPLIANCE BY DESIGN FOR PHARMACEUTICAL QUALITY CONTROL LABORATORIES INSIGHT FROM FDA WARNING LETTERS
COMPLIANCE BY DESIGN FOR PHARMACEUTICAL QUALITY CONTROL LABORATORIES INSIGHT FROM FDA WARNING LETTERS Primer CONTENTS INTRODUCTION...3 QUALITY AND COMPLIANCE IN QUALITY CONTROL LABORATORIES...5 Compliance
More informationPART 58 GOOD LABORATORY PRACTICE FOR NONCLINICAL LABORATORY STUDIES
Part 58-explanation of changes Scope ( 58.1) We propose to expand the scope of FDA-regulated nonclinical laboratory studies to specifically include toxicity studies. For purposes of this proposal, toxicity
More informationPART 58 GOOD LABORATORY PRACTICE FOR NONCLINICAL LABORATORY STUDIES
Part 58-explanation of changes Scope ( 58.1) We propose to expand the scope of FDA-regulated nonclinical laboratory studies to specifically include toxicity studies. For purposes of this proposal, toxicity
More information3410N Assurance engagements relating to sustainability reports
3410N Assurance engagements relating to sustainability reports Royal NIVRA 3410N ASSURANCE ENGAGEMENTS RELATING TO SUSTAINABILITY REPORTS Introduction Scope of this Standard ( T1 and T2) 1. This Standard
More informationWELMEC Application of module D Quality assurance Under directive 2014/32/EU (MID) or directive 2014/31/EU (NAWID)
WELMEC 8.4 2017 Application of module D Quality assurance Under directive 2014/32/EU (MID) or directive 2014/31/EU (NAWID) WELMEC is a cooperation between the legal metrology authorities of the Member
More informationGuide to Fulfillment of Validation and Verification of Examination Requirements
DIAGNOSTIC ACCREDITATION PROGRAM College of Physicians and Surgeons of British Columbia 300 669 Howe Street Telephone: 604-733-7758 ext. 2635 Vancouver BC V6C 0B4 Toll Free: 1-800-461-3008 (in BC) www.cpsbc.ca
More informationAuthorized By: Issue No.: 01 Date: 01/09/2010 Doc. No.: ACCAB-QM-4.0 BOD Revision: 00 Date: 01/09/2010 Effective Date: 01/09/2010 Page 1 of 8
Revision: 00 Date: 01/09/2010 Effective Date: 01/09/2010 Page 1 of 8 4.1 Purpose: 4.1.1 To define the meanings of terminologies and abbreviations used in this manual. 4.2 Scope: 4.2.1 Applicable to quality
More informationOMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT
OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT PA/PH/OMCL (05) 48 DEF CORR SCOPE OF ACCREDITATION Full document title and reference Document type Legislative basis Scope of Accreditation
More informationWe are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists. International authors and editors
We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists 3,900 116,000 120M Open access books available International authors and editors Downloads Our
More informationWe are IntechOpen, the first native scientific publisher of Open Access books. International authors and editors. Our authors are among the TOP 1%
We are IntechOpen, the first native scientific publisher of Open Access books 3,350 108,000 1.7 M Open access books available International authors and editors Downloads Our authors are among the 151 Countries
More informationGeneral Requirements for the Competence of Testing and Calibration Laboratories
ISO 17025:2005 POSITION PAPER General Requirements for the Competence of Testing and Calibration Laboratories EXECUTIVE SUMMARY This position paper provides a high-level overview of how LabVantage supports
More informationCodex Committee on Methods of Analysis and Sampling (31 st Session) Budapest, Hungary (8-12 March 2010)
CRD 13 Codex Committee on Methods of Analysis and Sampling (31 st Session) Budapest, Hungary (8-12 March 2010) European Union comments on the Proposed Draft Guidelines on Criteria for Methods for Detection,
More informationISO/IEC INTERNATIONAL STANDARD. Conformity assessment General requirements for accreditation bodies accrediting conformity assessment bodies
INTERNATIONAL STANDARD ISO/IEC 17011 First edition 2004-09-01 Conformity assessment General requirements for accreditation bodies accrediting conformity assessment bodies Évaluation de la conformité Exigences
More informationOECD and EPA GLP Differences
OECD and EPA GLP Differences Annette Leslie RQAP-GLP Du Pont Pioneer Ankeny Iowa I conduct studies in the US, so Why do I care about oecd regulations? Disclaimer: I cannot cover all the differences in
More informationPACIFIC RIM CONSULTING COMMENTS ON FDA PROPOSAL TO REVISE GLPS
PACIFIC RIM CONSULTING COMMENTS ON FDA PROPOSAL TO REVISE GLPS 1. GLP Quality System While many of the requirements of the existing regulation are consistent with a GLP quality system, FDA believes that
More informationá1225ñ VALIDATION OF COMPENDIAL PROCEDURES
1640 á1224ñ Transfer of Analytical Procedures / General Information USP 39 THE ANALYTICAL PROCEDURE The procedure should be written with sufficient detail and explicit instructions, so that a trained analyst
More informationImplementing the New TNI Standard. April 7, 2011
Implementing the New TNI Standard April 7, 2011 On July 1, 2011, the 2009 TNI standard, Volume 1: Management and Technical Requirements for Laboratories Performing Environmental Analysis, will become effective
More informationLIMS FOR TESTING LABORATORIES AN OVERVIEW OF FEATURES AND COMPLIANCE FOR TESTING LABORATORIES
WHITE PAPER LIMS FOR TESTING LABORATORIES AN OVERVIEW OF FEATURES AND COMPLIANCE FOR TESTING LABORATORIES 427 N Tatnall St #42818 Wilmington DE 19801-2230 USA A laboratory that performs any type of test
More informationMETHOD VALIDATION: AN ESSENTIAL COMPONENT OF THE MEASUREMENT PROCESS. Presented by : Jerry Parr, The NELAC Institute
METHOD VALIDATION: AN ESSENTIAL COMPONENT OF THE MEASUREMENT PROCESS Presented by : Jerry Parr, The NELAC Institute Source Material The NELAC Institute, Volume 1, Module 2, Quality Systems General Requirements,
More informationWhile the recognition
Designing the Perfect Change Control System Change control systems today are expected to be designed in a way that provides a system to not only document and approve changes, but also to anticipate change
More informationTHE QUALITY INSTITUTIONS An enabling framework for international trade. Tom Rotherham International Institute for Sustainable Development (IISD)
THE QUALITY INSTITUTIONS An enabling framework for international trade Tom Rotherham International Institute for Sustainable Development (IISD) THE QUALITY INSTITUTIONS An enabling framework for international
More information