1 Laboratory Page 1 of 17 Laboratory December 6, by A2LA All rights reserved. No part of this document may be reproduced in any form or by any means without the prior written permission of A2LA.

2 Laboratory Page 2 of 17 Table of Contents 1.0 SCOPE REFERENCES TERMS AND DEFINITIONS SPECIFIC CRITERIA... 5 DOCUMENT REVISION HISTORY... 15

3 Laboratory Page 3 of SCOPE This document describes the requirements for threat agent testing organizations seeking A2LA accreditation. For the purposes of this document threat agent testing refers to testing performed on submitted or collected chemical materials and/or biological organisms or products intended for harm and where the result of that testing could be used in criminal or civil litigation. Participation in this program is voluntary. Organizations seeking accreditation for threat agent testing must also meet A2LA policy and requirements documents: P102a Policy on Reference Material Traceability for Life Sciences Testing Laboratories.doc P103b Annex-Policy on Estimating Measurement Uncertainty for Life Sciences Testing Labs.doc P113 A2LA Policy on Measurement Traceability for Life Sciences Testing Laboratories R101 General Requirements: Accreditation of ISO/IEC laboratories R102 Conditions of Accreditation R103 General Requirements: Proficiency Testing for ISO/IEC Laboratories R103a Annex: Proficiency Testing for ISOIEC Laboratories R105 Requirements When Making Reference to A2LA Accredited Status When testing is performed outside the organization s permanent facility, R104 General Requirements: Accreditation of Field Testing and Field Calibration Laboratories applies. GENERAL CRITERIA: The general criteria for accreditation are contained in ISO/IEC 17025:2005, General requirements for the competence of testing and calibration laboratories, as referenced in Part A of R101 General Requirements: Accreditation of ISO/IEC Laboratories. PROFICIENCY TESTING: Please refer to R103 General Requirements: Proficiency Testing for ISO/IEC Laboratories for the proficiency testing requirements applicable to this program. 2.0 REFERENCES Food and Drug Administration ORA-LAB ORA Laboratory Procedure Volume II - Methods, Method Verification and Validation. Integrated Consortium of Laboratory Networks (ICLN) Guidelines for Comparison of Validation Levels Between Networks. 3.0 TERMS AND DEFINITIONS The following definitions were drawn from the ICLN s Guidelines for Comparison of Validation Levels Between Networks. Accuracy: A measure of the degree of conformity of a value generated by a specific procedure to the assumed or accepted true value, and includes precision and bias.

4 Laboratory Page 4 of 17 Certified Reference Material (CRM): reference material, accompanied by a certificate, one or more of whose property values are certified by a procedure which establishes metrological traceability to an accurate realization of the unit in which the property values are expressed, and for which each certified value is accompanied by an uncertainty at a stated level of confidence (slightly modified from VIM04). NOTE: The term "Standard Reference Material" (SRM) is the name of a certified reference material (CRM), which is the trademark name of a certified reference material that has been certified and is distributed by the National Institute of Standards and Technology (NIST). Limit of Quantification (LOQ): Lowest amount or concentration of analyte that can be quantitatively determined with an acceptable level of uncertainty, also referred to as the limit of determination. Linearity: Defines the ability of the method to obtain test results proportional to the concentration. Method Detection Limit (MDL): Lowest amount or concentration of analyte that a specific method can statistically differentiate from analyte-free sample matrix. This is dependent on sensitivity, instrumental noise, blank variability, sample matrix variability, and dilution factor. Minimum Detectable Concentration (MDC): An estimate of the minimum true concentration of analyte that must be present in a sample to ensure a specified high probability (usually 95%) that the measured response will exceed the detection threshold (i.e., critical value), leading one to conclude correctly that the analyte is present. Minimum Quantifiable Concentration (MQC): The smallest concentration of analyte whose presence in a laboratory sample ensures the relative standard deviation of the measurement does not exceed a specified value, usually 10 percent. Precision: Degree of agreement of measurements under specified conditions. The precision is described by statistical methods such as a standard deviation or confidence limit. See also Random Error. Repeatability expresses the precision under the same operating conditions over a short period of time. Intermediate precision expresses within-laboratory variations, such as different days, different analysts, and different equipment. Reproducibility expresses the precision between laboratories. Reference Material: A material or substance, one or more of whose property values are sufficiently homogenous and well established to be used for the calibration of an apparatus, the assessment of a measurement method, or for assigning values to materials. Repeatability: The closeness of the agreement between the results of successive measurements of the same measurand carried out under the same conditions of measurement. Sensitivity: The lowest concentration that can be distinguished from background noise or the smallest amount of a substance or organism that can accurately be measured by a method or test system is the analytical sensitivity. However, sensitivity is commonly defined as the slope of the calibration curve at a level near the LOQ. For assays that will be used to test human clinical specimens, the method's analytical sensitivity is distinct from the method's clinical diagnostic sensitivity. Clinical diagnostic sensitivity is the percentage of persons who have a given condition who

5 Laboratory Page 5 of 17 are identified by the method as positive for the condition (high analytical sensitivity does not guarantee acceptable diagnostic sensitivity). Specificity: Analytical specificity is the ability of a method to measure one particular analyte in the presence of components which may be expected to be present. For methods that will be used to test human clinical specimens, the method's analytical specificity is distinct from the method's clinical diagnostic specificity. Clinical diagnostic specificity is the percentage of persons who do not have a given condition who are identified by the method as negative for the condition. Trueness: The degree of agreement of the expected value from a measurement with the true value or accepted reference value. This is related to systematic error (bias). 4.0 SPECIFIC CRITERIA Specific criteria are used to augment the general criteria applicable to a certain field of testing, testing technology, or specific test. Along with the requirements of ISO/IEC 17025:2005, the specific requirements for the Threat Agent Testing laboratory program are outlined below. 4.0 Management Requirements 4.13 Control of Records : Records of original observations, derived data and sufficient information to establish an audit trail are required in order to enable the test to be repeated under conditions as close as possible to the original. These records include, but may not be limited to: Incubator temperature (start/end or at least twice daily); Oven temperature (start/end or at least twice daily); Time incubation started and ended (if no timer set); Extraction time started and ended (if no timer set); Initials of analyst involved in each step of the preparation and testing process; Lot numbers of reagents used (commercial and prepared); Lot numbers of media used (commercial and prepared); Reagent & media preparation records (including lot numbers of all components); Lot numbers of materials critical to the process (ex. extraction filters, petri dishes, etc.); Lot numbers and Certificates of Analysis of Certified Reference Materials (CRMs) and Reference Materials (RMs) (also see );

6 Laboratory Page 6 of 17 Identification of equipment used that is critical to the process (ex. pipettors, ovens, incubators, HPLC instruments, balances, etc.); Purchasing records, including POs and vendor evaluation/selection records; and QC of materials received critical to the test process. 5.0 Technical Requirements 5.3 Accommodation and Environmental Conditions 5.3.2: The laboratory shall monitor, control and record environmental conditions as required by the relevant specifications, methods and procedures or where they influence the quality of the results. This includes the following: Air exposure plates - daily (when plating is performed); Environmental swabs (contamination control swabs) of the test area to ensure cross-contamination has not occurred - daily (when testing performed) these swabs may be tested with daily samples, but are required to be tested if suspect target is identified; Restricted access to the test area when testing is in process, restricted access of unauthorized personnel, procedures to handle breach in access; and Evaluation of previous use of the area before new testing activities commence. 5.4 Test Methods and Method Validation The requirements for method validation are drawn from the minimal guidelines for the ICLN to validate methods which shall be used by the respective laboratory : Prior to developing the proposed methods: : The intended use of the method shall be defined; : The intended use should be aligned with current capabilities; : A written protocol which includes all of the following elements shall be prepared: Intended use and criteria for use; Assay principle and safety precautions; Acceptable sample types, sample collection method, preparation, preservation, storage and transportation conditions; Description of reagents supplied or directions for preparation and QC;

7 Laboratory Page 7 of 17 Description of quality controls to be used; Detailed instructions on how to perform the method; Detailed instructions on how to interpret and report the result; Any limitations of the method that are known or suspected; A summary of the performance characteristics of the method; and A statement on the target uncertainty of measurements for each analyte (analytical goal for accuracy) in order for the values to be fit for their intended purpose : Methods shall be validated whenever any of the following occur: : Development of a proposed new official method; : Expansion of the scope of an existing method to include additional analytes or new matrices; : Modification of a method s range beyond validated levels; : Modification of a method that may alter its performance specifications. All but the most trivial of changes shall be evaluated for effects on method performance. This includes: Changes to the fundamental science of an existing method and Equivalence issues such as substitutions of reagents/apparatus, or changes to some instrumental parameters since it is difficult to predict the results of any change : Performance specifications required to validate a method include the following: : Performance specifications that should be determined to validate a method will vary depending on the intended use, the type of method being validated, and the degree to which it has previously been validated. All methods submitted shall have all the proper controls and the parameters for calibrating and operating the method instrumentation included in the written procedure : Typical validation characteristics which shall be considered are the following: : Characteristics of quantitative methods: Method uncertainty at range of use; Minimum quantifiable concentration; Applicable analyte concentration range;

8 Laboratory Page 8 of 17 Accuracy (trueness); Precision (repeatability); Analytical specificity; Linearity; and Ruggedness/robustness : Characteristic for qualitative methods: Evaluate for reliable identification of an analyte at or below some target level; Sensitivity; Specificity; Limit of detection (See Detection limit, MDL, MDC); and Ruggedness/robustness : The following validation tools shall be used to demonstrate the ability to meet method specifications of performance: Blanks: Use of various types of blanks enables assessment of how much of the result is attributable to the analyte in relation to other causes; Reference materials and certified reference materials with the typical interferences expected. Use of known materials can be incorporated to assess the accuracy of the method, as well as for obtaining information on interferences; Fortified (spiked) materials and solutions: Recovery determinations can be estimated from fortification or spiking with a known amount of analyte. (Note: Understand that spiked recovery may not be truly representative of recovery from naturally incurred analytes); Repeatability: Replicate analyses provide a means of checking for changes in precision in an analytical process which could adversely affect the results; Statistics: Statistical techniques are employed to evaluate accuracy, trueness and precision, linear range, limits of detection and quantification, and measurement uncertainty. NOTE 1: Please see ORA-LAB in section 2.0 for reference : General Validation Protocol : The following provides practices that shall be used to determine method performance characteristics: Quantitative measurements, e.g. determine limit of quantification (LOQ), linear response.

9 Laboratory Page 9 of 17 Minimally need LOD; Prepare and analyze spiked blanks, matrix samples of known concentration utilizing one to three different concentration levels: low, medium, high based on the intended use of the method. These samples are carried through the complete sample preparation procedure, extraction and analytical steps of a particular method. Matrix effects shall be assessed with these samples. Accuracy or bias and precision are calculated from these results; data will also evaluate robustness of the method resulting from changes in the sample matrix. (Note: Proper certified reference materials and reference standards are used when available.); Assure that adequate sample replicates are performed and that results from replicate measurements of each analyte are compared; Analyze blanks (reagent and matrix) and compare these results to the reported limit of detection; and Evaluate interferences: spectral, physical, chemical or memory by analyzing samples containing various suspected interferences in the presence of the measurand : Validation of Methods (Original, New or Modified) : The following practices shall include, but not be limited to, matrix extensions and platform changes: In cases where the sample preparation and/or the extraction procedure/analytical method is modified from the existing test procedure and protocol, the new method shall demonstrate that the modifications do not adversely affect the precision and accuracy or bias of the data obtained; In order to implement the modified method, the standard or existing method is first performed. The modified method is then verified against the original method validation protocol; For original or new methods the authors shall pick a validation level that is suitable for their situation; Statistical methods are employed to verify performance between the original validated and new method sample means and to determine the degree of accuracy. (Note, for example: The t-test assesses whether the means of two groups are statistically different. The t-test is to be less than or equal to the t-critical value. The F-test is used to determine the significance of difference between two sample variances. The F value is to be less than or equal to the F-critical value). Note: See Tables for details : Estimation of Uncertainty of Measurement : Measurement Uncertainty is required to be estimated for tests with a qualitative reported result, if the result is based on quantitative endpoint, such as presence of a compound above a threshold value. 5.6 Measurement Traceability

10 Laboratory Page 10 of : In the event that CRMs are not available from an Asia Pacific Laboratory Accreditation Cooperation (APLAC) recognized, ISO/IEC accredited source, the laboratory shall provide records of material evaluation and traceability consistent with A2LA Policy P102a, Policy on Reference Material Traceability for Life Sciences Testing Laboratories. This includes the following records: : When Reference Materials are used as process controls: Unique identifier or lot number of the material (required); Description including (as appropriate) Origin, (e.g. inorganic, human or animal, vegetable, or microbial); Who and where manufactured (may be generated in-house or provided by supplier); Other pertinent information; and Records of use and process control results (required) : When Reference Materials are used to validate methods: Unique identifier or lot number (required); Origin (if applicable) (e.g. inorganic, human or animal, vegetable, or microbial); Who and where manufactured (may be generated in-house or provided by supplier); Issuing authority (if applicable) (e.g. WHO, BCR, IRMM, USP, ATCC); Records of characterization (if applicable); Molecular form(s) of or surrogate for the analyte (e.g. steric isomer for an amino acid, or glycerol for glycerol ester); Matrix (e.g. buffered bovine albumin solution); State(s) of aggregation (gas, liquid, solid); Phase(s) (solution, suspension, lyophilized); Records of use (required); and Records of Materials validation activities including an appropriate validation plan, acceptance criteria, and results (as appropriate) : When Reference Materials used are obtained from an authoritative source or prepared by a published standard procedure (to demonstrate operational traceability): Unique identifier or lot number (required);

11 Laboratory Page 11 of 17 Origin (if applicable) (e.g. inorganic, human or animal, vegetable, or microbial); Who and where manufactured (may be generated in-house or provided by supplier); Issuing authority (e.g. WHO, BCR, IRMM, USP, ATCC); Records of characterization (if applicable); Molecular form(s) of, or surrogate for the analyte (e.g. steric isomer for an amino acid, or glycerol for glycerol ester); Method used to establish value (AL); Matrix (e.g. buffered bovine albumin solution, analyte concentration in water, or soil, analyte concentration in air); State(s) of aggregation (gas, liquid, solid); Phase(s) (solution, suspension, lyophilized); and Records of use (required). 5.9 Assuring the Quality of Testing and Calibration Results 5.9.1: When commercial proficiency testing programs are available, laboratories shall participate in these programs annually for all targets, unless the programs are not relevant to the testing performed. Often, proficiency testing is not available for specific threat agents tested. In the event an external proficiency testing program is not available, the laboratory shall demonstrate competency annually, using blind spiked samples (spiked with target or surrogate). Prior to achieving initial accreditation, the laboratory shall have successfully participated in one of the two requirements listed above for each method on their desired Scope.

12 Laboratory Page 12 of 17 Table 1. Validation Recommendations for Chemistry Methods Originating Laboratory Study Level One: Urgent Usage/ Matrix Extension Level Two: Single Lab Validation Level Three: Independent Lab Validation Level Four: Collaborative Study a,b # of matrices 1 or more as the situation At least as the 3-5 as the 3-5 as the 3-5 as the # of sources situation situation situation situation # participating laboratories to 10 (qualitative) # minimum analyte level 1 spike level (spike at LOD) and 1 matrix blank 2 spike levels (spike at LOD and mid level) and 1 matrix blank 2 spike levels (spike at LOD and mid level) and 1 matrix blank 3 spike levels and 1 matrix blank Replicates per matrix tested at each level 2 2 (qualitative) 2 4 (qualitative) 2 6 (qualitative) 2 6 (qualitative) a Pure and Applied Chemistry, 67, No. 2, 1995,

13 Laboratory Page 13 of 17 b AOAC International, AOAC Peer-Verified Methods Program Manual on Policies and Procedures, AOAC international 1998.

14 Laboratory Page 14 of 17 Table 2. Validation Recommendations for Microbiological Methods Originating Laboratory Study Level One: Urgent usage Level Two: Single lab Validation Level Three: Independent Lab Validation Level Four: Collaborative Study i,j # of strains of target organism 1 to 5b 30b 30b 50b (inclusivity) a # of strains of non-target organism (exclusivity) 1 to 5 c 30 c 30 c 30 c # of matrices 1 or more d 1 or more d 1 or more d Up to 20 matrices d # analyte levels/matrix Set level based on intended use and uninoculated control One inoculated level e, one at 1 log higher, and uninoculated control One inoculated level e, one at 1 log higher, and uninoculated control One inoculated level e, one at 1 log higher (optional), and uninoculated control Replicates per matrix f 2 or more 6 or more 6 or more at least 10 Aging of inoculated samples prior to testing No No Yes g Yes g

15 Laboratory Page 15 of 17 Originating Laboratory Study Level One: Urgent usage Level Two: Single lab Validation Level Three: Independent Lab Validation Level Four: Collaborative Study i,j In 1 matrix at +1 In 1 matrix at +1 In 1 matrix at +1 Addition of competitor strain h Normal background flora log >analyte at fractional positive d analyte log >analyte at fractional positive d analyte log >analyte at fractional positive d analyte level level level Comparison to recognized method No Yes, if available Yes, if available Yes, if available a For bacteriological methods at 10 3 CFU/ ml (g) following the method protocol. b Select agent organisms may have limited strain availability for inclusion and exclusion studies. An attempt should be made to obtain necessary strains when possible. c For bacteriological methods at 10 3 CFU/mL for non-target organisms grown in a non-selective rich medium d Depends on applicability of method e Must be adjusted to achieve fractional positive results (one or both methods give 40-90% positive results) f Except for Level 1, must run six replicates per matrix, with as many different matrix sources as possible, up to six. Additional replicates may be required based on the specific method requirements for bias, precision, uncertainty, etc. g Period of aging depends on material being tested h An appropriate competitor is one that gives similar reactions in enrichment and detection systems i Pure and Applied Chemistry, 67, No. 2, 1995, j AOAC International, AOAC Peer-Verified Methods Program Manual on Policies and Procedures, AOAC international 1998.

16 Laboratory Page 16 of 17 Table 3. Validation Recommendations for Radiological Methods Originating Laboratory Study Level One: Urgent Usage Matrix Extension Level Two: Single Lab Validation Level Three: Independent Lab Validation Level Four: Collaborative Study a,b 1 or more as 1 or more as 1 or more as 1 or more as # of matrices the situation the situation the situation the situation 3-5 as the 3-5 as the 3-5 as the # of sources situation situation situation At least 5 # participating laboratories to 10 (qualitative) # minimum analyte levels/ matrix 1 spike level and 1 matrix blank 2 spike levels and 1 matrix blank 2 spike levels and 1 matrix blank 3 spike levels and 1 matrix blank Replicates per matrix tested at each level 2 2 (qualitative) 2 6 (qualitative) 2 6 (qualitative) 2 6 (qualitative) a Pure and Applied Chemistry, 67, No. 2, 1995, b AOAC International, AOAC Peer-Verified Methods Program Manual on Policies and Procedures, AOAC international 1998.

17 Laboratory Page 17 of 17 Document Revision History Date Description 4/1/2016 Initial version released. 12/6/2017 Section 4.0: Addition of C231 clauses. See C231 revision history from August 23, 2017 for further details. Formatting changes (throughout document) by A2LA All rights reserved. No part of this document may be reproduced in any form or by any means without the prior written permission of A2LA.

A2LA. P113a Policy on Reference Material Traceability for Life Sciences Testing Laboratories

A2LA. P113a Policy on Reference Material Traceability for Life Sciences Testing Laboratories Page 1 of 6 Introduction The Life Sciences Advisory Committee realizes that the traceability of reference materials (RMs) per the requirements of ISO/IEC 17025:2005 (Section 5.6.3) or ISO/IEC 17025:2017(Section

More information

CRITERIA FOR VALIDATION OF METHODS USED BY CHEMICAL LABORATORIES AND RELATED INDUSTRIES

CRITERIA FOR VALIDATION OF METHODS USED BY CHEMICAL LABORATORIES AND RELATED INDUSTRIES Document No: SADCAS TR 17 Issue No: 1 CRITERIA FOR VALIDATION OF METHODS USED BY CHEMICAL LABORATORIES AND RELATED INDUSTRIES Prepared by: SADCAS Advisory Committee TLAP Approved by: SADCAS CEO Approval

More information

TECHNICAL GUIDANCE FOR THE VALIDATION OF METHODS USED BY CHEMICAL LABORATORIES IN THE FOOD, WATER AND RELATED INDUSTRIES

TECHNICAL GUIDANCE FOR THE VALIDATION OF METHODS USED BY CHEMICAL LABORATORIES IN THE FOOD, WATER AND RELATED INDUSTRIES TECHNICAL GUIDANCE FOR THE VALIDATION OF METHODS USED BY CHEMICAL LABORATORIES IN THE FOOD, WATER AND RELATED INDUSTRIES Approved By: Chief Executive Officer: Ron Josias Senior Manager: Christinah Leballo

More information

Guide to Fulfillment of Validation and Verification of Examination Requirements

Guide to Fulfillment of Validation and Verification of Examination Requirements DIAGNOSTIC ACCREDITATION PROGRAM College of Physicians and Surgeons of British Columbia 300 669 Howe Street Telephone: 604-733-7758 ext. 2635 Vancouver BC V6C 0B4 Toll Free: 1-800-461-3008 (in BC) www.cpsbc.ca

More information

RAJESH NAIR DIRECTOR CALF-NDDB, ANAND, GUJARAT

RAJESH NAIR DIRECTOR CALF-NDDB, ANAND, GUJARAT RAJESH NAIR DIRECTOR CALF-NDDB, ANAND, GUJARAT 1 Coverage Need of AQC in food testing laboratories Analytical quality system of food testing laboratories Relationship between Quality System, Quality Assurance

More information

Raad voor Accreditatie (Dutch Accreditation Council RvA) Explanatory document on microbiology

Raad voor Accreditatie (Dutch Accreditation Council RvA) Explanatory document on microbiology Raad voor Accreditatie (Dutch Accreditation Council RvA) Explanatory document on microbiology Document code: A RvA-Explanatory note describes the policy and/or the procedures of the RvA concerning a specific

More information

MODULE 2A GENERAL MANAGEMENT SYSTEM REQUIREMENTS

MODULE 2A GENERAL MANAGEMENT SYSTEM REQUIREMENTS MODULE 2A GENERAL MANAGEMENT SYSTEM REQUIREMENTS 2A.1 SCOPE (See ISO/IEC 17025:2005, Section 1) Laboratories shall meet all requirements of the ISO/IEC 17025:2005 International Standard and other AIHA-LAP,

More information

á1225ñ VALIDATION OF COMPENDIAL PROCEDURES

á1225ñ VALIDATION OF COMPENDIAL PROCEDURES 1640 á1224ñ Transfer of Analytical Procedures / General Information USP 39 THE ANALYTICAL PROCEDURE The procedure should be written with sufficient detail and explicit instructions, so that a trained analyst

More information

Validation of Analytical Methods used for the Characterization, Physicochemical and Functional Analysis and of Biopharmaceuticals.

Validation of Analytical Methods used for the Characterization, Physicochemical and Functional Analysis and of Biopharmaceuticals. Validation of Analytical Methods used for the Characterization, Physicochemical and Functional Analysis and of Biopharmaceuticals. 1 Analytical Method Validation: 1..1 Philosophy: Method validation is

More information

TNI s NATIONAL ENVIRONMENTAL LABORATORY ACCREDITATION PROGRAM (NELAP) Changing to the 2009 Standard

TNI s NATIONAL ENVIRONMENTAL LABORATORY ACCREDITATION PROGRAM (NELAP) Changing to the 2009 Standard TNI s NATIONAL ENVIRONMENTAL LABORATORY ACCREDITATION PROGRAM (NELAP) Changing to the 2009 Standard THE NEW NELAP STANDARDS Four Volumes = Four Standards Volume 1: Requirements for Laboratories 7 Modules

More information

GLP/SC/01 Basic statistical tools for analytical chemistry (2 days)

GLP/SC/01 Basic statistical tools for analytical chemistry (2 days) GLP Consulting http://consultglp.com Courses on offer (See outlines below) GLP/SC/01 Basic statistical tools for analytical chemistry (2 days) GLP/SC/02 The GUM bottom-up evaluation technique of measurement

More information

A2LA. G123 - Guidance for Cannabis Laboratories Preparing for Accreditation

A2LA. G123 - Guidance for Cannabis Laboratories Preparing for Accreditation Page 1 of 6 Purpose: This document is intended to act as a guide to new cannabis testing laboratories to help determine the steps to take to prepare for accreditation, what is required as part of an application

More information

APPENDIX H TRACEABILITY OF MEASUREMENT

APPENDIX H TRACEABILITY OF MEASUREMENT APPENDIX H TRACEABILITY OF MEASUREMENT Table of Contents 1. SCOPE... 2 2. REFERENCES... 2 3. TERMS AND DEFINITIONS... 2 4. BACKGROUND... 6 5. TRACEABILITY OF MEASUREMENT POLICY... 7 7. GUIDANCE ON IMPLEMENTING

More information

Examples of reference materials

Examples of reference materials Examples of reference materials JCTLM Working Group for Traceability: Education and Promotion (WG-TEP) Area 2: Mini-presentations to explain scientific concepts What are reference materials? Reference

More information

American Association for Laboratory Accreditation

American Association for Laboratory Accreditation ISO 15189 Page 1 of 17 Accredited clinical laboratories are required to meet the following additional requirements as contained in P905 - A2LA Metrological Traceability Policy for ISO 15189 Laboratory

More information

GUIDANCE NOTES ON ANALYTICAL METHOD VALIDATION

GUIDANCE NOTES ON ANALYTICAL METHOD VALIDATION ON ANALYTICAL METHOD VALIDATION HSA September 2004 Reproduction prohibited for commercial purposes. Reproduction for internal use is authorised, provided the source is acknowledged. MQA Dir: DISK1\GUIDE-MQA-012A-004.doc

More information

Specific Accreditation Criteria Materials ISO/IEC Annex. Cement testing

Specific Accreditation Criteria Materials ISO/IEC Annex. Cement testing Specific Accreditation Criteria Materials ISO/IEC 17025 Annex Cement testing January 2018 Copyright National Association of Testing Authorities, Australia 2013 This publication is protected by copyright

More information

Guidance Document. A Guide for the Validation and Approval of New Marine Biotoxin Test Methods. 10 April 2017

Guidance Document. A Guide for the Validation and Approval of New Marine Biotoxin Test Methods. 10 April 2017 Guidance Document A Guide for the Validation and Approval of New Marine Biotoxin Test Methods A guidance document issued by the Ministry for Primary Industries Title Guidance Document: Validation and Approval

More information

ISO/IEC 17025:2017 Working Document Instruction Page

ISO/IEC 17025:2017 Working Document Instruction Page Instruction Page NOTES: 1. This working document is intended as a checklist for the assessor when conducting Testing and Calibration Laboratory and Sampling Organization Assessments according to ISO/IEC

More information

The Importance of Validated Rapid Methods for International Trade

The Importance of Validated Rapid Methods for International Trade The Importance of Validated Rapid Methods for International Trade 3M 2009. All Rights Reserved. Robert Koeritzer Sr. Technical Manager 3M Food Safety Past-President, AOAC International Agenda Method validation

More information

PALCAN Guidance for the Validation of Test Methods

PALCAN Guidance for the Validation of Test Methods PALCAN Guidance for the Validation of Test Methods CAN-P-1629 PALCAN GUIDANCE FOR THE VALIDATION OF TEST METHODS POLITIQUE DU PALCAN CONCERNANT LA VALIDATION DES MÉTHODES D ESSAIS CAN-P-1629 Copyright

More information

How to use the TNI standards to meet essential QC elements in the MUR

How to use the TNI standards to meet essential QC elements in the MUR How to use the TNI standards to meet essential QC elements in the MUR August 7, 2012 Susan Wyatt Minnesota Department of Health EPA Methods Update Rule Finalized on May 18, 2012 (Proposed 9/23/2010) Effective

More information

APLAC GUIDANCE ON REFERENCE MATERIAL USE AND PRODUCTION

APLAC GUIDANCE ON REFERENCE MATERIAL USE AND PRODUCTION APLAC GUIDANCE ON REFERENCE MATERIAL USE AND PRODUCTION Issue 6 Issue Date: 2018/02 Page 1 of 18 PURPOSE Reference Materials (RMs) are used in all stages of the measurement process, for method validation,

More information

A24 CALA Checklist for Microbiology Revision 4.1 February 16, 2018

A24 CALA Checklist for Microbiology Revision 4.1 February 16, 2018 A24 CALA Checklist for Microbiology Revision 4.1 February 16, 2018 Laboratory Name: Appendix Name: Appendix Number: Assessor: Date: Rev. 4.1 CALA CHECKLIST FOR MICROBIOLOGY Item 01 SAMPLING AND SAMPLE

More information

TECHNICAL REQUIREMENTS (CLAUSE 5)

TECHNICAL REQUIREMENTS (CLAUSE 5) LECTURE 4 THE ELEMENTS OF ISO/IEC 17025 - TECHNICAL REQUIREMENTS (CLAUSE 5) Slide 1 ISO 17025 - RELATIONSHIP OF EACH REQUIREMENTS Management System Organization Quality system Continual Improvement Management

More information

Procedure for Equipment Calibration and Maintenance

Procedure for Equipment Calibration and Maintenance Procedure for Equipment Calibration and Maintenance 1.0 Purpose This procedure specifies the schedule and requirements for calibration, performance verification, and maintenance of State Crime Laboratory

More information

Implementing the New TNI Standard. April 7, 2011

Implementing the New TNI Standard. April 7, 2011 Implementing the New TNI Standard April 7, 2011 On July 1, 2011, the 2009 TNI standard, Volume 1: Management and Technical Requirements for Laboratories Performing Environmental Analysis, will become effective

More information

Verification of Method XXXXXX

Verification of Method XXXXXX Reviewed By: Date: Technical Supervisor Reviewed By: Date: General Supervisor Reviewed By: Date: Quality Assurance Coordinator Approved By: Date: Director Page 1 of 11 Table of Contents Purpose and Principle:...

More information

Specific Accreditation Criteria

Specific Accreditation Criteria Specific Accreditation Criteria ISO/IEC 17025 Application Document Materials - Annex Characterisation of industrial materials - General July 2018 Copyright National Association of Testing Authorities,

More information

METHOD VALIDATION TECHNIQUES PREPARED FOR ENAO ASSESSOR CALIBRATION COURSE OCTOBER/NOVEMBER 2012

METHOD VALIDATION TECHNIQUES PREPARED FOR ENAO ASSESSOR CALIBRATION COURSE OCTOBER/NOVEMBER 2012 METHOD VALIDATION PREPARED FOR ENAO ASSESSOR CALIBRATION COURSE TECHNIQUES OCTOBER/NOVEMBER 2012 Prepared by for ENAO Assessor Calibration B SCOPE Introduction House Rules Central Tendency Statistics Population

More information

Quality System Guidance

Quality System Guidance Page 1 of 16 TABLE OF CONTENTS 1.0 SCOPE... 2 2.0 REFERENCES... 2 3.0 TERMS AND DEFINITIONS... 2 4.0 BACKGROUND... 5 5.0 GUIDANCE ON IMPLEMENTATION... 5 5.1 Choosing Type A or Type B... 6 5.2 General considerations...

More information

Revision of 30 April 2013 draft, 4 November 2013

Revision of 30 April 2013 draft, 4 November 2013 GUIDANCE DOCUMENT FOR SINGLE LABORATORY VALIDATION OF QUANTITATIVE ANALYTICAL METHODS USED IN SUPPORT OF PRE- AND POST-REGISTRATION DATA REQUIREMENTS FOR PLANT PROTECTION AND BIOCIDAL PRODUCTS INTRODUCTION

More information

ISO and SANAS guidelines for use of reference materials and PT scheme participation. T & M Conference

ISO and SANAS guidelines for use of reference materials and PT scheme participation. T & M Conference ISO 17025 and SANAS guidelines for use of reference materials and PT scheme participation T & M Conference Presentation layout 1. Introduction ISO/IEC 17025 2. Requirements for use Reference Materials

More information

Quality System Guidance

Quality System Guidance Page 1 of 16 TABLE OF CONTENTS 1.0 SCOPE... 2 2.0 REFERENCES... 2 3.0 TERMS AND DEFINITIONS... 2 4.0 BACKGROUND... 5 5.0 GUIDANCE ON IMPLEMENTATION... 5 5.1 Choosing Type A or Type B... 6 5.2 General considerations...

More information

Standard Practices for Measurement Traceability in Forensic Toxicology

Standard Practices for Measurement Traceability in Forensic Toxicology ANSI/ASB Standard 017, First Edition 2018 Standard Practices for Measurement Traceability in Forensic Toxicology This document is copyrighted by the AAFS Standards Board, LLC. 2018 All rights are reserved.

More information

Guidelines for the assessment of the appropriateness of small interlaboratory comparisons within the process of laboratory accreditation

Guidelines for the assessment of the appropriateness of small interlaboratory comparisons within the process of laboratory accreditation Publication Reference EA-4/21 INF: 2018 Guidelines for the assessment of the appropriateness of small interlaboratory comparisons within the process of laboratory accreditation PURPOSE This paper provides

More information

Procedure for Biomedical Equipment Calibration and Maintenance

Procedure for Biomedical Equipment Calibration and Maintenance Procedure for Biomedical Equipment Calibration and Maintenance 1.0 Purpose This procedure specifies the schedule and requirements for calibration, performance verification, and maintenance of Biomedical

More information

Thank You to United States Pharmacopeia for Their Sponsorship and Assistance

Thank You to United States Pharmacopeia for Their Sponsorship and Assistance Use of Measurement Uncertainty in Testing a Drug Substance M.L. Jane Weitzel Quality Analysis Consultants mljweitzel@msn.com Eurachem/CITAC Workshop June 2011 Thank You to United States Pharmacopeia for

More information

EA Guidance on Accreditation of Pesticide Residues Analysis in Food and Feed

EA Guidance on Accreditation of Pesticide Residues Analysis in Food and Feed Publication Reference EA-4/22 G: 2018 EA Guidance on Accreditation of Pesticide Residues Analysis in Food and Feed PURPOSE This paper provides guidance to develop harmonized accreditation of Pesticide

More information

Laboratory Developed Tests. William Castellani, MD Inter-regional Commissioner Clinical Pathologist, Penn State Hershey Medical Center

Laboratory Developed Tests. William Castellani, MD Inter-regional Commissioner Clinical Pathologist, Penn State Hershey Medical Center Laboratory Developed Tests William Castellani, MD Inter-regional Commissioner Clinical Pathologist, Penn State Hershey Medical Center Objectives Compare and contrast new test validation requirements for

More information

American Association for Laboratory Accreditation

American Association for Laboratory Accreditation Page 1 of 14 P113 A2LA Policy on Measurement Traceability for Life (November 2012) 2012 by A2LA. All rights reserved. No part of this document may be reproduced in any form or by any means without the

More information

A016 Traceability of measurement results compared to national and international measurement standards

A016 Traceability of measurement results compared to national and international measurement standards 08.04.2016 Version 08 Page 1 de 9 A016 Traceability of measurement results compared to national and international measurement standards Modifications: p 2 South Lane Tower I 1, avenue du Swing L-4367 Belvaux

More information

CRITERIA FOR VALIDATION AND QUALITY ASSURANCE IN MICROBIOLOGICAL TESTING

CRITERIA FOR VALIDATION AND QUALITY ASSURANCE IN MICROBIOLOGICAL TESTING Document No: SADCAS TR 18 Issue No: 1 CRITERIA FOR VALIDATION AND QUALITY ASSURANCE IN MICROBIOLOGICAL TESTING Prepared by: SADCAS Advisory Committee TLAP Approved by: SADCAS CEO Approval Date: 2018-03-24

More information

Guidelines for the validation and verification of chemical test methods

Guidelines for the validation and verification of chemical test methods Technical Note 17 April 2009 Issued: August 2004 Amended and reissued: December 2006, April 2009 Guidelines for the validation and verification of chemical test methods Be Absolutely Assured Copyright

More information

ENVIRONMENTAL LABORATORY SECTOR VOLUME 1. Management and Technical Requirements for Laboratories Performing Environmental Analysis.

ENVIRONMENTAL LABORATORY SECTOR VOLUME 1. Management and Technical Requirements for Laboratories Performing Environmental Analysis. EL-V1-2009 ENVIRONMENTAL LABORATORY SECTOR VOLUME 1 Management and Technical Requirements for Laboratories Performing Environmental Analysis TNI Standard P.O. Box 2439 Weatherford, TX 76086 817-598-1624

More information

ISO 13528:2015 Statistical methods for use in proficiency testing by interlaboratory comparison

ISO 13528:2015 Statistical methods for use in proficiency testing by interlaboratory comparison ISO 13528:2015 Statistical methods for use in proficiency testing by interlaboratory comparison ema training workshop August 8-9, 2016 Mexico City Class Schedule Monday, 8 August Types of PT of interest

More information

Exposing Myths, Legends and Tales About Reference Materials. Presented by: Matt Sica - ANAB Steve Arpie Absolute Joe Konschnik - RESTEK

Exposing Myths, Legends and Tales About Reference Materials. Presented by: Matt Sica - ANAB Steve Arpie Absolute Joe Konschnik - RESTEK Exposing Myths, Legends and Tales About Reference Materials Presented by: Matt Sica - ANAB Steve Arpie Absolute Joe Konschnik - RESTEK Overview Reference Materials Accreditation Reference Materials Production

More information

Application of ISO/IEC 17025

Application of ISO/IEC 17025 Application of ISO/IEC 17025 A Testing Perspective Colleen M. Cotter, Accreditation Manager Outline An overview of CALA Application of ISO/IEC 17025 Specific requirements for Testing Laboratories (5.6.2.2)

More information

GENERAL GUIDELINES FOR THE ACCREDITATION OF FOOD AND ENVIRONMENTAL MICROBIOLOGICAL LABORATORIES

GENERAL GUIDELINES FOR THE ACCREDITATION OF FOOD AND ENVIRONMENTAL MICROBIOLOGICAL LABORATORIES GENERAL GUIDELINES FOR THE ACCREDITATION OF FOOD AND ENVIRONMENTAL MICROBIOLOGICAL LABORATORIES ESYD G-MICROBIOL/01/02/20-10-2016 1/12 ESYD G-MICROBIOL Issue: 01 Revision: 02 Issue Date: 20-06-2007 Revision

More information

Pesticides: Method Validation and Quality Control

Pesticides: Method Validation and Quality Control The guidance in 10684/2009 is intended for laboratories control or in the monitoring of pesticide residues in food involved in official and feed in the European Union. The document describes the method

More information

Establishment of an Accredited Reference Measurement Laboratory

Establishment of an Accredited Reference Measurement Laboratory Establishment of an Accredited Reference Measurement Laboratory Francesco Dati, PhD IVD-Consulting Marburg / Germany E-mail: f.dati@t-online.de Quality of Analytical Systems Reference Measurement Systems

More information

INTERNATIONAL STANDARD

INTERNATIONAL STANDARD INTERNATIONAL STANDARD FDIS ISO/IEC 17025 General requirements for the competence of testing and calibration laboratories Exigences générales concernant la compétence des laboratoires d'étalonnages et

More information

INTERPRETATION AND GUIDANCE ON THE ESTIMATION OF UNCERTAINTY OF MEASUREMENT IN TESTING

INTERPRETATION AND GUIDANCE ON THE ESTIMATION OF UNCERTAINTY OF MEASUREMENT IN TESTING KAN-G-20 INTERPRETATION AND GUIDANCE ON THE ESTIMATION OF UNCERTAINTY OF MEASUREMENT IN TESTING Issue number : 1 April 2016 Komite Akreditasi Nasional National Accreditation Body of Indonesia Gedung I

More information

Policy on the Traceability of Measurement Results

Policy on the Traceability of Measurement Results Page 1 of 14 Reviewed by: Getnet Tsigemelak Position: Deputy Director General Signature: Approved by: Araya Fesseha Position: Director General Signature: Contents pages 1. Purpose and scope... 2 2. References...

More information

DESIGN AND OPERATION OF AN INTERLABORATORY COMPARISON SCHEME

DESIGN AND OPERATION OF AN INTERLABORATORY COMPARISON SCHEME DESIGN AND OPERATION OF AN INTERLABORATORY COMPARISON SCHEME R.M. VOICULESCU, M. C. OLTEANU, V.M. NISTOR Institute for Nuclear Research Pitesti, Romania, ramona.voiculescu@nuclear.ro ABSTRACT The competence

More information

Measurement Uncertainty Guide. ISO Accreditation Program

Measurement Uncertainty Guide. ISO Accreditation Program Measurement Uncertainty Guide ISO 15189 Accreditation Program Background Why This is Necessary The ISO 15189:2012 standard contains enhanced expectations regarding measurement uncertainty (MU) in clause

More information

CLSI C60: Assay Validation & Post-Validation Monitoring

CLSI C60: Assay Validation & Post-Validation Monitoring CLSI C60: Assay Validation & Post-Validation Monitoring Ross J. Molinaro, MT(ASCP), PhD, DABCC, FACB Medical Director Core Laboratory, Emory University Hospital Midtown Emory Clinical Translational Research

More information

A61-01 CALA Traceability Policy Revision 3.2 September 16, 2016

A61-01 CALA Traceability Policy Revision 3.2 September 16, 2016 A61-01 CALA Traceability Policy Revision 3.2 September 16, 2016 PAGE II TABLE OF CONTENTS TABLE OF CONTENTS... 1 CALA TRACEABILITY POLICY... 2 1.0 SCOPE... 2 2.0 BACKGROUND... 2 3.0 DEFINITIONS... 3 4.0

More information

validation report for TaqMan Roundup Ready Quantification Kit

validation report for TaqMan Roundup Ready Quantification Kit validation report for TaqMan Roundup Ready Quantification Kit Life Technologies offers a breadth of products DNA RNA protein cell culture instruments For Research Use Only. Not intended for any animal

More information

Enhancing trueness and accuracy of procedures for food, feed and environmental samples

Enhancing trueness and accuracy of procedures for food, feed and environmental samples Enhancing trueness and accuracy of procedures for food, feed and environmental samples EQALM, Dublin 20 October Hilde Skår Norli Chair of NordVal International Norwegian Veterinary Institute Contents Rational

More information

Agreed by VICH Steering Committee January Adoption by CVMP 12 February Date for coming into effect January 2016

Agreed by VICH Steering Committee January Adoption by CVMP 12 February Date for coming into effect January 2016 12 February 2015 EMA/CVMP/VICH/463202/2009 Committee for Medicinal Products for Veterinary Use (CVMP) VICH topic GL49: Studies to evaluate the metabolism and residues kinetics of veterinary drugs in human

More information

A69 Checklist for Bulk Asbestos Revision 1.2 May 5, 2015

A69 Checklist for Bulk Asbestos Revision 1.2 May 5, 2015 A69 Checklist for Bulk Asbestos Revision 1.2 May 5, 2015 Laboratory Name: Appendix Name: Appendix Number: Assessor: Date: 01 DOCUMENT CONTROL 01 4.3 : is there a documented method? : Verify that the current

More information

Traceability Exercise Documenting Vendor Evaluations and Calibration Certificate Review

Traceability Exercise Documenting Vendor Evaluations and Calibration Certificate Review Traceability Exercise Documenting Vendor Evaluations and Calibration Certificate Review Douglas Berg Testing Technical Program Mgr Perry Johnson Laboratory Accreditation PJLA NIST Traceability What is

More information

This paper describes the interpretation and guidelines set

This paper describes the interpretation and guidelines set 1070 ROBERTSON & CHAN:JOURNAL OF AOAC INTERNATIONAL VOL. 86, NO. 5, 2003 SPECIAL GUEST EDITOR SECTION APLAC Interpretation and Guidance on the Estimation of Uncertainty of Measurement in Testing MAX ROBERTSON

More information

NATA Policy on Calibration and Measurement Traceability

NATA Policy on Calibration and Measurement Traceability NATA Policy on Calibration and Measurement Traceability Paul McMullen Sector Manager, Calibration This Session 1. VIM Definitions Calibration and Metrological Traceability 2. Previous Policy on Metrological

More information

Roundtable Discussion Method Validation & Verification

Roundtable Discussion Method Validation & Verification Roundtable Discussion Method Validation & Verification 2 May 2016 DeAnn Benesh 3M 2015. All Rights Reserved WHAT is Validation? Reference Codex CAC/GL 74 1 ISO 16140-1 2 USDA FSIS 3 US FDA 4 Validation

More information

Overview of Good Food Laboratory Practices

Overview of Good Food Laboratory Practices Overview of Good Food Laboratory Practices Dr. Anne Bridges Technical Director, AACC International International Workshop & Training Program On Good Food Laboratory Practices Jointly Organized By Food

More information

Traceability & Uncertainty. Accreditation Experience. David Ducroq. Weqas. Unit 6, Parc Tŷ Glas. Llanishen. Cardiff.

Traceability & Uncertainty. Accreditation Experience. David Ducroq. Weqas. Unit 6, Parc Tŷ Glas. Llanishen. Cardiff. Traceability & Uncertainty Accreditation Experience David Ducroq Weqas Unit 6, Parc Tŷ Glas Llanishen Cardiff www.weqas.com UKAS Perspective - Comments from lead assessor Traceability: Where available,

More information

ILAC-P10:01/2013 -Calibration-

ILAC-P10:01/2013 -Calibration- ACTIVITY 1.4 TRAININGS OF MOLDAC PERSONNEL INVOLVED IN ACCREDITATION PROCESS INCLUDING THE ISSUE REGARDING THE REQUIREMENTS OF THE NEW VERSIONS OF THE EA, IAF, ILAC DOCUMENTS ILAC-P10:01/2013 -Calibration-

More information

Proficiency Testing. Bryanne Shaw Biology Section Manager

Proficiency Testing. Bryanne Shaw Biology Section Manager Proficiency Testing Bryanne Shaw Biology Section Manager Proficiency Testing The evaluation of participant performance against pre-established criteria by means of interlaboratory comparisons. ISO 17025

More information

Presented By: Matt Sica

Presented By: Matt Sica FQS Accreditation for forensic testing agencies Presented By: Matt Sica ANSI-ASQ National Accreditation Board Implies an unbroken path to a higher level of accuracy or authority. Characterized by a number

More information

VALIDATION OF ANALYTICAL PROCEDURES: METHODOLOGY *)

VALIDATION OF ANALYTICAL PROCEDURES: METHODOLOGY *) VALIDATION OF ANALYTICAL PROCEDURES: METHODOLOGY *) Guideline Title Validation of Analytical Procedures: Methodology Legislative basis Directive 75/318/EEC as amended Date of first adoption December 1996

More information

General Accreditation Criteria Equipment assurance, in-house calibration and equipment verification

General Accreditation Criteria Equipment assurance, in-house calibration and equipment verification General Accreditation Criteria Equipment assurance, in-house calibration and equipment verification January 2018 Copyright National Association of Testing Authorities, Australia 2009 This publication is

More information

FQS Accreditation for forensic testing agencies NEMC ANSI-ASQ National Accreditation Board

FQS Accreditation for forensic testing agencies NEMC ANSI-ASQ National Accreditation Board FQS Accreditation for forensic testing agencies ISO Guide d 34 for f Environmentall Labs b NEMC 2012 ANSI-ASQ National Accreditation Board ISO/IEC 17025 Calibration and Testing Labs ISO/IEC 17020 Inspection

More information

TECHNICAL AND PRODUCTION REQUIREMENTS of ISO Guide 34:2009

TECHNICAL AND PRODUCTION REQUIREMENTS of ISO Guide 34:2009 SANAS Accr. No/s. TECHNICAL AND PRODUCTION REQUIREMENTS of ISO Guide 34:2009 Date/s of evaluation Assessor/s & Observers Facility Area / field of Representative operation This report covers the following:

More information

VICH Topic GL49. at step 4 GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS USED IN RESIDUE DEPLETION STUDIES

VICH Topic GL49. at step 4 GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS USED IN RESIDUE DEPLETION STUDIES European Medicines Agency Veterinary Medicines and Inspections London, 14 December 2009 Doc. Ref. EMEA/CVMP/VICH/463202/2009-CONSULTATION VICH Topic GL49 at step 4 GUIDELINES FOR THE VALIDATION OF ANALYTICAL

More information

Available online at ScienceDirect. Procedia Engineering 132 (2015 )

Available online at  ScienceDirect. Procedia Engineering 132 (2015 ) Available online at www.sciencedirect.com ScienceDirect Procedia Engineering 132 (2015 ) 811 815 The Manufacturing Engineering Society International Conference, MESIC 2015 Metrological Regulations for

More information

APLAC GUIDELINES FOR FOOD TESTING LABORATORIES

APLAC GUIDELINES FOR FOOD TESTING LABORATORIES APLAC GUIDELINES FOR FOOD TESTING LABORATORIES Issue No. 2 Issue Date: 01/08 Page 1 of 13 PURPOSE This document provides guidance to APLAC members and their accredited laboratories on the operation of

More information

Guidance document Regulation (EC) 882/2004 Microbiological sampling and testing of foodstuffs. Enne de Boer. II. Analysis

Guidance document Regulation (EC) 882/2004 Microbiological sampling and testing of foodstuffs. Enne de Boer. II. Analysis TAIEX workshop Ankara, 4-5 March 2013 Guidance document Regulation (EC) 882/2004 Microbiological sampling and testing of foodstuffs II. Analysis Enne de Boer Contents Sample testing at laboratories Requirements

More information

2010 PNCWA Conference Bend, OR

2010 PNCWA Conference Bend, OR 2010 PNCWA Conference Bend, OR Built in 1964,1971,1988, 1999, 2007, 2010 EPA Proposed Method Update Rule New Approved Methods and new versions of Old Approved Methods Revised Table of Holding Times and

More information

Laboratory Quality Assurance Manager & Laboratory Assessor RULES & HANDBOOK

Laboratory Quality Assurance Manager & Laboratory Assessor RULES & HANDBOOK EOQ Personnel Registration Scheme Laboratory Quality Assurance Manager & RULES & HANDBOOK Prepared by: Dr. Eugenia Soboleva, Quality Austria In accordance with the working group on EOQ product development

More information

Edward F. Askew PhD Askew Scientific Consulting

Edward F. Askew PhD Askew Scientific Consulting Edward F. Askew PhD Askew Scientific Consulting DO YOU KNOW WHAT IT IS?? (Are the Auditors Listening??) Askew Scientific Consulting LLC 2 Requires specific test procedure: Protocol for Review and Validation

More information

Verification of Compendial Methods

Verification of Compendial Methods Verification of Compendial Methods ภญ.ดร ดร.ส ภาณ ดวงธ รปร ชา ส าน กยาและว ตถ เสพต ด กรมว ทยาศาสตร การแพทย 7 ม ถ นายน 2556 1 Scopes : Overview Asean Analytical Validation Guidelines Verification of Compendial

More information

Introducing the Concept of Uncertainty of Measurement in Testing in Association with the Application of the Standard ISO/IEC 17025

Introducing the Concept of Uncertainty of Measurement in Testing in Association with the Application of the Standard ISO/IEC 17025 Introducing the Concept of Uncertainty of Measurement in Testing in Association with the Application of the Standard ISO/IEC 17025 ILAC-G17:2002 Copyright ILAC 2002 ILAC encourages the authorised reproduction

More information

MAURITAS R3:2017 Issue No. 7:Revision 1 Date: November 2017 MAURITAS. Traceability of measurement. Mauritius Accreditation Service. Copyright MAURITAS

MAURITAS R3:2017 Issue No. 7:Revision 1 Date: November 2017 MAURITAS. Traceability of measurement. Mauritius Accreditation Service. Copyright MAURITAS Issue No. 7:Revision 1 MAURITAS R3 Traceability of measurement Mauritius Accreditation Service Copyright MAURITAS CONTENTS FOREWORD... 2 ABOUT MAURITAS PUBLICATIONS... 2 1. PURPOSE... 3 2. SCOPE AND RESPONSIBILITIES...

More information

CRITICAL ASPECT ANALYTICAL TEST REVIEW

CRITICAL ASPECT ANALYTICAL TEST REVIEW CRITICAL ASPECT ANALYTICAL TEST REVIEW Jakarta 14 December 2017 Speaker: HERU PURNOMO, ST QC WORK FLOW Start Sample Received (In-Process & Finished Good) Testing Review Lab Result Yes No Non Conformance

More information

A02 Assessment Rating Guide Revision 2.9 August 21, 2016

A02 Assessment Rating Guide Revision 2.9 August 21, 2016 Revision 2.9 August 21, 2016 Laboratory Name: Assessment Date: (Lead) Assessor: Signature: ! TABLE OF CONTENTS 1.0! INTRODUCTION... 1 2.0! ASSESSOR NOTES ON THE USE OF CALA A02 ASSESSMENT RATING GUIDE...

More information

Erin Patton, MS Senior Product Specialist Charles River Labs, Microbial Solutions

Erin Patton, MS Senior Product Specialist Charles River Labs, Microbial Solutions Validation of a Rapid Microbial Method Approach to Validation Case Studies: Equivalence Verification of an Alternate Assay for Microbial Limits Screening and Sterility Testing of Pharmaceutical Products

More information

IL AC-G22:2004. Use of Proficiency Testing as a Tool for Accreditation in Testing

IL AC-G22:2004. Use of Proficiency Testing as a Tool for Accreditation in Testing IL AC-G22:2004 Use of Proficiency Testing as a Tool for Accreditation in Testing Copyright ILAC 2004 ILAC encourages the authorised reproduction of its publications, or parts thereof, by organisations

More information

Understanding the Characteristics and Establishing Acceptance Criteria for Analytical Methods Validation

Understanding the Characteristics and Establishing Acceptance Criteria for Analytical Methods Validation Understanding the Characteristics and Establishing Acceptance Criteria for Analytical Methods Validation Ying Verdi IVT LAB WEEK EUROPE June 2017 Partners in Health Since 1919 Regulatory View of Method

More information

Control Charts and Trend Analysis for ISO 17025

Control Charts and Trend Analysis for ISO 17025 2015 Webinar Series Control Charts and Trend Analysis for ISO 17025 5/27/2015 Speakers Cynthia Mangione, Microbiology QAO, New York State Department of Agriculture and Markets, Food Laboratory Division,

More information

Ready, Set, Test! AACC Conference Mass Spectrometry in the Clinical Lab: Best Practice and Current Applications September 17-18, 2013 St.

Ready, Set, Test! AACC Conference Mass Spectrometry in the Clinical Lab: Best Practice and Current Applications September 17-18, 2013 St. Ready, Set, Test! Ross Molinaro, PhD, MLS(ASCP) CM, DABCC, FACB Medical Director, Clinical Laboratories Emory University Hospital Midtown Emory Clinical Translational Research Laboratory AACC Conference

More information

WAIT! Ready, Set, Test! Financial Disclosure. Research/Educational grants/consulting/salary support

WAIT! Ready, Set, Test! Financial Disclosure. Research/Educational grants/consulting/salary support Ready, Set, Test! Ross Molinaro, PhD, MLS(ASCP) CM, DABCC, FACB Medical Director, Clinical Laboratories Emory University Hospital Midtown Emory Clinical Translational Research Laboratory AACC Conference

More information

INSTRUCTIONS FOR USE

INSTRUCTIONS FOR USE INSTRUCTIONS FOR USE Epower Microorganisms INTENDED USE Epower microorganisms are lyophilized, quantitative microorganism preparations to be used in industrial laboratories for quality control purposes.

More information

ISO/TS TECHNICAL SPECIFICATION. Water quality Guidance on analytical quality control for chemical and physicochemical water analysis

ISO/TS TECHNICAL SPECIFICATION. Water quality Guidance on analytical quality control for chemical and physicochemical water analysis TECHNICAL SPECIFICATION ISO/TS 13530 First edition 2009-03-15 Water quality Guidance on analytical quality control for chemical and physicochemical water analysis Qualité de l'eau Lignes directrices pour

More information

From purpose based interpretation to appropriate application of Calibration and Interlaboratory comparison results

From purpose based interpretation to appropriate application of Calibration and Interlaboratory comparison results From purpose based interpretation to appropriate application of Calibration and Interlaboratory comparison results Mulugeta Derebew mulugetaderebew@yahoo.com Test & Measurement 2017, Conference and Workshop,

More information

CALA Disputes

CALA Disputes Modify procedure so that management has formulated training goals, and that there are policies and procedures for identifying training needs and provision of training, with such training evaluated for

More information

The new ISO/IEC 17025:2017

The new ISO/IEC 17025:2017 The new ISO/IEC 17025:2017 By Dr. George Anastasopoulos Director, Conformity assessment, IAS Email: ganastasopoulos@iasonline.org Introduction - Background information ISO/IEC 17025 was first issued in

More information

Event-specific Method for the Quantification of Soybean CV127 Using Real-time PCR. Validation Report

Event-specific Method for the Quantification of Soybean CV127 Using Real-time PCR. Validation Report Event-specific Method for the Quantification of Soybean CV127 Using Real-time PCR Validation Report 20 September 2011 Joint Research Centre Institute for Health and Consumer Protection Molecular Biology

More information

R205: SPECIFIC REQUIREMENTS: CALIBRATION LABORATORY ACCREDITATION PROGRAM. December 2011

R205: SPECIFIC REQUIREMENTS: CALIBRATION LABORATORY ACCREDITATION PROGRAM. December 2011 Page 1 of 12 R205: SPECIFIC REQUIREMENTS: CALIBRATION LABORATORY ACCREDITATION PROGRAM December 2011 2011 by A2LA All rights reserved. No part of this document may be reproduced in any form or by any means

More information