PHARMA TEST Dissolution Mechanical Calibration
|
|
- Job Jones
- 6 years ago
- Views:
Transcription
1 Pharma Test Dissolution Automatisation PHARMA TEST Dissolution Mechanical Calibration Dissolution Testers, Pumps and Spectrophotometers 24-Nov-11
2 Requirements EP, USP, FDA, ASTM Dissolution 2
3 24-Nov-11 Pharma Test Dissolution Automatisation
4 Requirements EP, USP, FDA, ASTM Dissolution EP Requirements Temperature 37 +/- 0,5 C - Distance between the Apparatus and the vessel bottom 25+/-2mm - Vessel Height: mm, Vessel diameter mm - Stirrer Speed +/-4% - Stainless Steel 316 or equivalent should be used for the parts in contact with the media - Shaft is positioned max 2mm from the axis 4
5 Requirements EP, USP, FDA, ASTM Dissolution -EP Requirements Dissolution Medium Volume Accuracy +/-1% - Remove Thermometer or valididate system with thermometer - Sampling position is min 1cm away from the vessel wall, half way between Apparatus and liquid level. - Where multiple sampling times are specified, replace the aliquots withdrawn for analysis with equal volumes of fresh dissolution medium at 37 c or if not necessary (not under -1% volume lost), the volume loss must be corrected in the calculation 5
6 Requirements EP, USP, FDA, ASTM Dissolution EP Requirements If bubbles have an influence on the dissolution test, gases must be removed prior to testing - Samples are to be withdrawn only at the stated times, within a tolerance of +/- 2 per cent. - Delayed-release solid dosage forms Method A Place 750ml 0,1M HCl in the vessel. After 2 h withdraw an aliquot and add 250ml of 0,2M Trisodiumphosphate Dodecahydrate Buffer (37 +/- 0,5 C!!) 6
7 Requirements EP, USP, FDA, ASTM Dissolution EP Requirements Quality in design, any irregularities such as vibration are to be avoided. - The performance of the dissolution test equippment may be monitored by testing a reference product which is sensitve to hydrodynamic conditions (!!!USP Prednisone!!!) or Salicylic Acid. - During testing, critical inspection and observation are required. - Conventional Release dosage forms: min 75% after 45min - Prolonged-reease dosage forms: 3 Testpoints, when 20-30per cent are dissolved, when around 50% are dissolved and endpoint (more than 80%) - Delayed release dosage forms: 1h and 2h in 0,1N HCl, after buffer change when the tablet is dissolved completely. 75% min. 7
8 Requirements EP, USP, FDA, ASTM Dissolution Additional USP requirements - Performance Verification Test (PVT), Apparatus 1 and 2 with Prednisone - Periodicity is 6 Month for mechanical calibration and PVT Prednisone test - Tests need to be repeated if the instrument is translocated or repaired. - Test starts if the tablet rests at the bottom of the vessel > Staggared - Media Should be deaerated for the Prednisone Test. Nitrogen sparging is not a suitable dearation procedure and therefor should not be used. - The volume should be controlled over a more accurate weight measurement as the weight is temperature-independent attention!! Which weight? 500g? Correct density at 37 C must be choosen >> DDS4 8
9 Requirements EP, USP, FDA, ASTM Dissolution 9
10 Requirements EP, USP, FDA, ASTM Spectrophotometer EP
11 Requirements EP, USP, FDA, ASTM Spectrophotometer EP
12 Requirements EP, USP, FDA, ASTM Spectrophotometer Control of wavelength 12
13 Requirements EP, USP, FDA, ASTM Spectrophotometer 13
14 Requirements EP, USP, FDA, ASTM Spectrophotometer Control of absorbance 14
15 Requirements EP, USP, FDA, ASTM Spectrophotometer 15
16 Requirements EP, USP, FDA, ASTM Spectrophotometer Straylight, resolution, Cells When using 2mm, 5mm or 10 mm kuvettes? If the concentration and the Absorbance of the solution is to high for a measurement with a spectrophotometer it s possible to dilute the sample or to take a smaller pathlength of the küvette. Law of Lambert Beer: A = e x b x c Where A is absorbance e is the extinction koefficient (the molar absorbtivity with units of L mol -1 cm -1 ) b is the path length of the sample - that is, the path length of the cuvette in which the sample is contained. c is the concentration of the compound in solution, expressed in mol L-1 Most of the spectrophotometers have an Absorbance measurement limit of ~1,2. If you take a look on the Equation the only possibility to change the Absorbance is to change b (the pathlength) or c (concentration by dilution) because e is a constant. Accurcy of the pathlength is +/- 0,005 cm! 16
17 PHARMA TEST Vibration and Deaeration Studies 17
18 Vibration Studies PHARMA TEST Vibration Studies - USP/FDA states that Vibrations have a significant influence on Dissolution Tests - No Tolerances given yet by USP/FDA and EP. - FDA is currently performing vibration studies to implement tolerances. - Only Recommendation is given by AAPS. Vibrations should be under 0.2 mil ~ 0.5 µm displacement - For the vibration study Prednisone Tablet are used as they are known to be extremely sensitive to vibrations. AAPS (American Association of Pharmaceutical Scientists) 18
19 Vibration Studies PHARMA TEST Test Performance Materials USP Prednisone Tablets (10 mg) Lot P0E203 [Promochem, Germany]: Instruments - PTW S III C [PHARMA TEST, Germany] and PT-DT 8 [PHARMA TEST, Germany] - Media Deaeration System PT-DDS-4 [PHARMA TEST, Germany] - Vibration Meter [Model VM 110: Monitran, UK] - Spectrophotometer: Lambda 2 [Perkin Elmer, Germany] - 50W Amplifier and Tone generator (to generate vibrations) - Circulation pump (to generate vibrations) 19
20 Vibration Studies PHARMA TEST 20
21 Vibration Studies PHARMA TEST Dissolution Tests Process Based on USP requirements: the tests were performed using: - USP Apparatus 2 50 rpm ml of preheated and deaerated water at 37 C nm Results: (at time of publication): - The range of dissolved Prednisone: 37 70% for Lot P0E203 Types of Tests: Without vibration and with deaerated media the procedures are the same as mentioned above. With vibration an amplifier (with tone 50Hz) producing displacements approx. 0.4, 0.5, 0.6, 0.8 µm displacement. A circulation pump (50Hz) producing a displacement of 1.1 µm 21
22 Vibration Studies PHARMA TEST Result and Discussion Vibration measurement Absence of vibration exposure PTW S III C suspected to have higher internal vibration value than PT-DT 8. Some possible causes are: - High vibration level on its paddle motor and drive belts (8 years old instrument) - Older design with the water bath placed directly on work bench resting on a rubber mat - Higher influence of environmental vibration sources Note: The vibration levels seen are still within proposed guideline levels (AAPS). 22
23 Vibration Studies PHARMA TEST PT DT 8. Lower vibration level, some causes are: - lower vibration level on its paddle motor and drive belts (new instrument, different bearings) - Latest design with the water bath placed onto Vibration Absorbers (Vibroban ). - Vibrations caused by the Circulation Pump are minimized.. 23
24 Vibration Studies PHARMA TEST 24
25 Vibration Studies PHARMA TEST PTWS III Testresults without forced Vibration Standard solution gave an absorbance of 0.87AU (represents 100% dissolved) Vibration level = 0.32 µm (measured on the instrument) 25
26 Vibration Studies PHARMA TEST PT DT8 Testresults without forced Vibration Vibration level = 0.21 µm (measured on the instrument) 26
27 Vibration Studies PHARMA TEST 0.32µm displacement, PTWSIII PT DT8 0.21µm displacement Differences between the two sets of results are not significant!! It is about 0.3% 27
28 % Dissolution Vibration Studies PHARMA TEST % Dissolution for PTWS 3C and PTDT 8 (Without Vibration) % dissolution test without vibration 103,00 70,00 37,00 4, Vessel PTW S III C PT-DT8 28
29 Vibration Studies PHARMA TEST Results recieved with forced Vibrations (0,5 µm, Amplifier) Prednisone test (PT DT 8), vibration level = 0.5µm reduced to 0.22µm by vibration absorbers on the instrument Prednisone test (PTW S III C) vibration level = 0.5 µm µm (both on the bench and at instrument) 29
30 % Dissolution Vibration Studies PHARMA TEST Graphic of % Dissolution between Table (With Vibrations) Prednisone Test with 0.5 µ m of Displacement (50Hz) 103,00 70,00 37,00 4, Vessel PT DT 8 PTW S III C % Differences between two results = 30.3% 30
31 Vibration Studies PHARMA TEST Vibration level = 0.6 µm displacement PTW S III C Vibration level = 0.8 µm displacement PTW S III C 31
32 Vibration Studies PHARMA TEST Vibration level = 1,1 µm PTW S III (with pump) 32
33 % Dissolution Vibration Studies PHARMA TEST 103,00 % Dissolution of Prednisone Test with Vibration Exposure 70,00 37,00 4, Vessel PTW S III C (0.8 µm) PT-DT 8 (0.22 µm) PTW S III C (1.1 µm) PTW S III C ( µm) PTW S III C ( µm) These results for the newer design PT DT8 are lower than the other sets regardless of the vibration levels 33
34 Vibration Studies PHARMA TEST Higher vibrational level Higher % Dissolution Results!! The highest increase is between 0.32 and 0.5 µm displacement. 34
35 % Dissolution Vibration Studies PHARMA TEST 103,00 Prednisone Test with and without Vibration 70,00 37,00 4, PTW S III C (without vibr = 0.32µm) Vessel PT-DT 8 (without vibr = 0.21µm) PTW S III C (vibr = 0.8 µm) PT-DT 8 (vibr = 0.22 µm) PTW S III C (vibr = 1.1 µm) PTW S III C (vibr µm) PTW S III C (vibr µm) 35
36 Vibration Studies FDA 36
37 Vibration Studies FDA 37
38 Vibration Studies FDA 38
39 Vibration Studies FDA 39
40 Deaeration Studies PHARMA TEST Prednisone Test, Paddle 40
41 Deaeration Studies PHARMA TEST >50% increase between non deaerated and deaerated media *) This value is out of range: Note that this study was carried out when the Prednisone LOT POE203 ranges were : 37-70% for Apparatus 2 (paddles) 41
42 % Dissolution Deaeration Studies PHARMA TEST 103 % Dissolution of Prednisone Test with Deaeration and Non- Deaeration Vessel 42
43 Deaeration Studies USP Nithyanandan et al. (USP), Dissolution Technologies 2006,
44 High Precision Vessels ID no. Upper Diameter Lower Diameter Diameter Difference W W W W W W W Mean Standard deviation ID no. Upper Diameter Lower Diameter Diameter Difference Mean Standard deviation
45 High Precision Vessels Dissolution Tester DT7 Serial Nr Vessel Old Vessels 4/28/2011 New Vessels 5/2/2011 Abs Conc. Abs Conc Std Dev Average LOT Nr. P1l300 P1l300 Test Result Pass Pass %CV Upper 4.7% 5.0% GeoMean GeoMean Tolerance Min Max %Coefficient of Variation Max
46 Deaeration Studies USP 46
47 Deaeration Studies USP 47
48 Operation Qualification (OQ) and Mechanical Calibration 48
49 Mechanical Calibration Standard Measurements and Requirements - Basket Dimensions (Certificate) - Paddle Dimensions (Certificate) - Vessel Dimensions (Certificate) - Rotation Speed (Tachometer) - Shaft Wobble (Watch) - Basket Wobble (Watch) - Paddle and Basket Depth (Certified 25mm balls) - Vessel Temperature (Thermometer) - Maintenance 49
50 Mechanical Calibration Shaft Wobble The Shaft Wobble is tested 2cm above the Apparatus Basket Wobble The Basket Wobble is tested at the bottom of the basket. The Tolerance is +/- 1mm. Test is performed with a Wobble Gauge. 50
51 Mechanical Calibration Paddle and Basket Depth The Depth is checked with certified 25mm balls which are placed at the bottom of the vessel. Rotation Speed 2 different speeds are checked with a certified Tachometer. Vessel Temperature A certified Thermometer is used to check the temperature in the vessel. The temperature Tolerance given is 37 +/- 0,5 C at a set temperature of 37 C. New USP requirement. All Temperatures must be in a range of 0,4 C 51
52 Mechanical Calibration Vessel Centering Should be measured at 2 different points of the shaft Tolerances (+/- 1mm) from the axes. 52
53 Mechanical Calibration Shaft Verticality Measure at 2 different points. Tolerances are 90 +/- 0,5 ) Measured with a digital level Vessel Verticality Also measured with a digital level. Tolerances are 90 +/-0,5 53
54 Mechanical Calibration Degrees out of Verticality = Arctan (X mm/ Y mm) = Angle (must be smaller than 1 ) 54
55 Mechanical Calibration Vibrations Vibrations should be not significantly higher than 0,2mils (0,5microns) displacement according to AAPS. Official Tolerances will be published soon. We PHARMA TEST recommend values under 0,4 microns displacement 55
56 Mechanical Calibration Vibration Absorbers 56
57 Mechanical Calibration Deaeration 40% of dissolved gases should be removed. Amount of Oxygen should be Less than 6ppm. PHARMA TEST DDS4 Media Preparator 24l, preheated, dosed and deaerated media 57
58 Mechanical Calibration 58
59 Mechanical Calibration - Prepared medium to reproducible standards - Pre-heated medium saves time for dissolution runs - Reproducible dispensed volumes (reduces any random errors - Programmable, up to 20 methods - GMP-stainless steel - Printer included (all necessary data such as Volumes, Date etc are mentioned - GMP-stainless steel - Volume up to 24 Liters - Gets you through the prednisone tests 59
60 Mechanical Calibration Preparing, deaerating, vacuum pump or ultrasonic Heating time Conventional Dissolution ~ 20 min. every dissolution run ~ 20 min. (from room temperature 25 C to 37 C) DDS4 ~ 40 min. for 4 dissolution runs 0 min. Appr. 40min/Dissolution Test Appr. 10min/Dissolution Test 60
61 Mechanical Calibration 61
62 Mechanical Calibration 62
63 Mechanical Calibration 63
64 Mechanical Calibration -No Salicylic Acid Test Required -LOT POE203 is not valid anymore -New LOT P1I300 -New Test Procedure with new Tolerances -Single Stage Test -Second Stage Test 64
65 Mechanical Calibration USP FDA Speed Test 50 and 100 rpm +/- 1rpm 50 and 100 rpm +/- 2rpm Wobble Test Shaft 1cm above the paddle < 1mm 2cm above Paddle < 1mm Wobble Test Basket Bottom Basket < 1mm Bottom Basket < 1mm 65
66 Mechanical Calibration USP FDA Temperature Test 37 +/- 0,5 C (0,4 C Range e.g C) 37 C +/- 0,5% Vessel Centricity Shaft Verticality In Vessel, max 2cm under the vessel surface. +/- 2mm At 2 Positions 90 +/ Positions (2mm and 80mm above the paddle) +/- 1mm At 2 Positions 90 +/
67 Mechanical Calibration USP FDA Hight Check 25mm +/- 2mm 25mm +/- 2mm Volume +/- 1% - Vessel Verticality At 2 Positions 90 +/- 0.5 At 2 Positions 90 +/- 1 67
68 Mechanical Calibration USP FDA Deaeration Vibrations.dissolved gases should be removed prior to testing USP 711 Table should have a high inertial mass to limit vibration Medium should be degassed less than 60% saturation of total dissolved gases...there should be no significant Vibrations in the dissolution apparatus or Medium 68
69 Mechanical Calibration USP Old Requirement USP New Requirement Prednisone 37-70% 27-38% 1 Stage CV%: 5, % 2 Stage CV%: 6,8 (2 Stage Test with 6 Tablets) Salicylic Acid Used Not used anymore 69
70 Mechanical Calibration 70
71 Mechanical Calibration 71
72 Harness Testers Introduction Purpose - USP <1217> What Is the Purpose Of A Tablet Hardness Test? According to USP Monograph <1217>: Tablets must be able to withstand the rigors of handling and transportation experienced in the manufacturing plant, in the drug distribution system, and in the field at the hands of the end users (patients/consumers)., the mechanical strength of tablets is of considerable importance and is routinely measured. Tablet strength serves both as a criterion by which to guide product development and as a quality control specification. 72
73 Hardness Testers Introduction Purpose - EP <2.9.8> What Is the Purpose Of A Tablet Hardness Test? According to EP Monograph <2.9.8>: This test is intended to determine, under defined conditions, the resistance to crushing of tablets, measured by the force needed to disrupt them by crushing. 73
74 Hardness Testers Technical Details: Typical Expression Take 10 samples, place them between the jaws, taking into account the shape, the break-line and any inscription. The jaws must have the height of the sample for full contact. Express the results as mean, minimum and maximum values of the forces, all expressed in Newton. Indicate the apparatus and orientation of the tablet. Accuracy ± 1N. All Hardness Testers have a stainless steel finish to meet GLP criteria in the production area. 74
75 Hardness Tester Adjustment PTB-CAL15 3 5, 10 and 15kg Calibration Weight Calibration handle with interchangeable weights to allow for 5, 10 and 15kg configurations 75
76 Hardness Testers Calibration PT MT 2: The Instrument The PT-MT2 can be used to qualify each PHARMA TEST tablet tester: PTB 311/411 PTBA 211 PTB 302 WHT ME 76
77 Tablet and Ampoule Testing Instrument Range: What can we offer? Instrument Range: What can we offer? PTB 302 PTB 311 E PTB 320 WHT 3 ME PT-MT2 PTBA 211 E 77
78 Tablet and Ampoule Testing Physical Parameters: What will be measured? Physical Parameters: What will be measured? Instrument Hardness Width Diameter Weight PTB 302 Yes No No No PTB 311 E Yes Yes Yes No PTB 320 * Yes Yes Yes Yes* WHT 3 ME + WHTSM Yes Yes Yes Yes PTBA 211E Yes (ampoules No No No only) * Analytical Balance (Mettler or Sartorius) to be added All systems are also available as 500 N versions! 78
79 Disintegration EP Physical Parameters: Typical Values for Tablets and Capsules Raising and lowering smoothly at a constant frequency between 28 and 32 cycles per minute through a distance of 50 to 60 mm. Volume is such that in the highest position the wire mesh is 15 mm at least below the surface of the liquid and in the lowest position the wire mesh is 25 mm at least above the bottom of the beaker and the glass tubes remain above the surface of the liquid. Temperature of the disintegration liquid should be maintained at 36 C to 39 C. 79
80 Disintegration Time Testing Instrument Range: What Can We Offer? Instrument Range: What Can We Offer? PTZ S DIST 3 PTZ AUTO PTZ AUTO EZ PTS 3 E SPT 6 80
81 Disintegration Time Testing PTZ S and DIST 3: The Instruments PTZ S and DIST 3: The Instruments Disintegration Basket Plastic Disks for manual use 81
82 Fully Automated Tablet Disintegration Time Testing PTZ AUTO 1EZ- 4EZ Instrument The Automated Detection Principle The disintegration basket MKT for automated detection is different from a manual basket. The bottom sieves are split into 2 segments. The disks have a metal ring at the bottom end. A signal transfer cable is attached. Like with the manual baskets on top of the tablet a disk is placed. If the Tablet is disintegrated the metal ring gives a contact to the 2 sieve segments. 82
83 Suppository Testing Instrument Range Disintegration of Suppositories and Pessaries Disintegration and Penetration Time Test of Suppositories and Pessaries SPT 6 Penetration Time PTS 3 E Disintegration Time 83
84 Suppository Testing Instrument Range Disintegration of Suppositories and Pessaries Disintegration and Penetration Time Test of Suppositories and Pessaries SPT 6 Penetration Time PTS 3 E Disintegration Time 84
85 Friability Testing Instrument Range: What can we offer? Instrument Range: What can we offer? PTF 10 E/ER PTF 20 E/ER PTF 30/60 ERA PTF 1DR PTF 3DR 85
86 Friability Testing EP Physical Parameters: Typical Values Tablets weighing up to 0.65 g take a sample of 20 tablets, for tablets weighing more than 0.65 g take 10 samples remove any loose dust (sieve no. 1000, soft brush, aid of air pressure). Rotate drum 100 times, at speed of 25 rotations per minute,remove tablets of the drum and any loose dust. Maximum loss of 1 per cent of the mass of the tablet is accepted. If test failed repeat twice and calculate the average of all the runs. For Tablets of a Diameter > 16 mm use a 10 angle to test. 86
87 PTG S4 EP The Instrument A4 PTG-ER2 Stirrer Funnel Release flap Powder cone h2 Powder receipt phi h1 ø100 Light barrier for the measurement of the cone height Powder overflow Collector Balance PTG S3 h1 = Pouring cone height h2 = 100mm phi = Cone angle zul. Abw. Oberfläche Maßstab Gewicht Verwendungsbereich ISO DIN 7151 IT 12 Gez. Datum ISO 1302 Ra Reihe 1 Name Werkstoff Benennung Operation diagramme 87
88 PTG S4 What will be measured 3. What will be measured? The automated powder analyzer PTG S4 is capable to test: powder flow time of a pre-defined mass powder cone volume powder cone density cone angle (angle of repose) powder flowability as per EP <2.9.16> amount of sample flow (g/in preset time) flow chart of sample (g/time) 88
89 PTG S4 USP and EP 2. USP and EP USP <1174>, EP <2.9.36> and ISO 4324 (12/83) Powder flow and Angle of repose Flow Properties and Corresponding Angles of Repose Flow property Angle of repose (degrees) Excellent Good Fair aid not needed Angle of repose Passable may hang up Poor must agitate, vibrate Very poor Very, very poor > 65 89
90 PTG S4 USP and EP USP and EP EP <2.9.16> Flowability 2. EP stainless steel funnel 1. Plastic Beaker 3. Pouring nozzle 10, 15, 25 mm 90
Automated Powder Analyzer - Type PTG-S _09.01E Page 1 of 5
The automated PTG-S4 powder testing system is used to measure the flow behaviour of granules and powders in compliance with the EP , USP Pharmacopoeia and ISO 4324 standards. This instrument
More informationPortfolio Overview Galenic Instruments
Galenic Instruments Portfolio Overview Galenic Instruments www.pharma-test.com Galenic Instruments For over 35 years, Pharma Test has been synonymous with the development and production of high-value test
More informationPTG-S4 Automated Powder Flow Analyzer
PTG-S4 Automated Powder Flow Analyzer The automated PTG-S4 powder testing system is used to measure the flow behavior of granules and powders in compliance with the EP , USP Pharmacopoeia
More informationTHE DISSOLUTION PROCEDURE: DEVELOPMENT AND VALIDATION
THE DISSOLUTION PROCEDURE: DEVELOPMENT AND VALIDATION Pharmacopeial Forum Vol. 31(5)(Sept.-Oct. 2005) By : Mr. Seubpong Kumpusiri Mrs. Patima Maneesatid 26 May 2006 THE DISSOLUTION PROCEDURE: DEVELOPMENT
More informationPractical Aspects of Dissolution Instrument Qualification a European Perspective
dx.doi.org/10.14227/dt180211p11 Practical Aspects of Dissolution Instrument Qualification a European Perspective e-mail: JKraemer@phast.com Johannes Kraemer* and Rolf Schwan PHAST, Kardinal-Wendel-Str.
More informationSuccessfully Implementing Mechanical Qualification for Dissolution Apparatus 1 and 2
Successfully Implementing Mechanical Qualification for Dissolution Apparatus 1 and 2 Ken Boda and Bryan Crist Agilent Technologies Dissolution Systems Agenda Mechanical Qualification Standards When You
More informationMagnetic Tablet for Dynamic Calibration and Periodical Qualification of Tablet hardness Testing Instruments
PT-MT3 Magnetic Tablet for Dynamic Calibration and Periodical Qualification of Tablet hardness Testing Instruments This instrument represents a system for calibrating and qualifying the hardness range
More informationMethod Development and Validation for Online UV-Dissolution Methods Using Fiber-Optic Technology
Technical Overview Method Development and Validation for Online UV-Dissolution Methods Using Fiber-Optic Technology Introduction Online fiber-optic and multicell UV-dissolution systems have become increasingly
More informationBEP2 Powder Flowability Tester Application Note
BEP2 Powder Flowability Tester Application Note The new harmonized chapters in the Pharmacopoeia (USP, Chapter / EP, Chapter 2.9.36) on powder flow list 4 well-defined methods for powder testing
More informationEnhanced Mechanical Calibration of Dissolution Test Equipment
dx.doi.org/10.14227/dt180211p25 Enhanced Mechanical Calibration of Dissolution Test Equipment Alger Salt 1, * and John Glennon 2 1 GlaxoSmithKline, Product Development Laboratory Automation and Platforms,
More informationA Simple and Economical Approach/Concept to Evaluate Quality of Pharmaceutical Products Based on an Improved Dissolution Testing Methodology
The Open Drug Delivery Journal, 2008, 2, 33-37 33 Open Access A Simple and Economical Approach/Concept to Evaluate Quality of Pharmaceutical Products Based on an Improved Dissolution Testing Methodology
More informationCE 7smart Flow-Through Dissolution System
CE 7smart Flow-Through Dissolution System In-vitro drug release testing in compliance with USP apparatus 4, EP, and JP Configurations for various methods and analytical requirements Specific flow-through
More informationAutomatic hardness testing devices
Automatic hardness testing devices The measure of hardness, applied to rubber and plastic, is a measure of stiffness obtained from an indentation test. An indentor is pressed into the sample under a given
More informationDeveloping and Validating Dissolution Procedures for Improved Product Quality
W H I T E P A P E R Developing and Validating Dissolution Procedures for Improved Product Quality By Michael Swartz, Ph. D., Director of Research and Development, and Mark Emanuele, Chemist Abstract In
More informationPerformance Testing of Novel Dosage Forms
RQA Ireland Regional Forum - Athlone, May 2016 Quality Considerations Pharma and Biopharma Performance Testing of Novel Dosage Forms Terry Way BPharm MAPS Dissolution Science Consultant Glasside Technologies
More informationThe widespread use of powders in the pharmaceutical industry has led to a proliferation of test methods for measuring powder flow and density.
Powders The widespread use of powders in the pharmaceutical industry has led to a proliferation of test methods for measuring powder flow and density. The harmonised chapters in the Pharmacopoeias on Powder
More informationMISSION STATEMENT ABOUT US DISTEK S MISSION DISTEK S VISION AND VALUES
ABOUT US MISSION STATEMENT DISTEK S MISSION Distek, Inc. is a leading manufacturer of pharmaceutical laboratory instruments, specializing in dissolution testing products, as well as an experienced provider
More information29-Apr Outline. Automation in dissolution testing. Mumbai , May 4 Samir Haddouchi
Automation in dissolution testing Mumbai - 2013, May 4 Samir Haddouchi samir.haddouchi@sps-pharma.com 1 Outline - Introduction about automation - Automation for dissolution testing - Validation of automated
More informationDissolution Apparatus Types
Dissolution Apparatus Types Thank you for reading. As you may know, people have look numerous times for their favorite books like this, but end up in infectious downloads. Rather than enjoying a good book
More informationAGILENT TECHNOLOGIES PRACTICAL SOLUTIONS NEWSLETTER
AGILENT TECHNOLOGIES PRACTICAL SOLUTIONS NEWSLETTER Volume 12, Issue 2 In This Issue: Enhanced Mechanical Qualification 1 Dissolution Qualification Timeline / Schedule 3 MQ Survey 4 MQ Guidance Summary
More informationAgilent 8453 Spectrophotometer and UV-visible ChemStation for Dissolution Testing. Solution Bundles for automated Dissolution Testing
Agilent 8453 Spectrophotometer and UV-visible ChemStation for Dissolution Testing Solution Bundles for automated Dissolution Testing General considerations UV benefits Simple, rugged and reliable equipment
More informationMulti Purpose Attachment
oating > Mixing > Grinding > Stirring > Kneading > Granulating > Filling > Homogenising > > Stirring > Kneading > Granulating > Filling > Homogenising > Pelletizing > Sieving > Sha > Dosing > Coating >
More informationInternational Journal of Research in Pharmaceutical and Nano Sciences Journal homepage:
Review Article CODEN: IJRPJK ISSN: 2319 9563 International Journal of Research in Pharmaceutical and Nano Sciences Journal homepage: www.ijrpns.com REVIEW ARTICLE ON INPROCESS PROBLEMS AND EVALUATION TESTS
More informationTroubleshooting Differences Between Labs. Ken Boda Dissolution Applications Engineer
Troubleshooting Differences Between Labs Ken Boda Dissolution Applications Engineer Agenda Dissolution Run Failure Investigation Differences in data between sites Potential causes of site to site differences
More informationAT 70smart Fully Automated Dissolution System
AT 70smart Fully Automated Dissolution System Fully automated from media preparation to data reporting Batch up to 40 lots of Apparatus 1 or 2 methods including the use of sinkers Powerful DoE screening
More informationPharmaceutical Solutions
Pharmaceutical Solutions HI-SPEED Checkweighing SAFELINE Metal Detection & X-ray Inspection Product Inspection For The Pharmaceutical Industry Industry Challenges We understand that product quality and
More informationComparative analysis of eight brands of sulfadoxinepyrimethamine
Tropical Journal of Pharmaceutical Research, June 2003; 2 (1): 161-167 Pharmacotherapy Group, Faculty of Pharmacy, University of Benin, Benin City, Nigeria. All rights reserved. Available online at http://www.tjpr.freehosting.net
More informationFleet Bioprocessing Ltd. Manufacturing Batch Record Title: FORMULATION OF AN ANTIBODY-BIOTIN CONJUGATE Lot Number(s):
Manufacturing Batch All unshaded boxes in this document should be completed. All operators involved in this formulation must add their name, signature and identifying initials to the Table below. NAME
More informationTransferring Dissolution Methods. Kenneth Boda Applications Engineer
Transferring Dissolution Methods Kenneth Boda Applications Engineer Agenda Prerequisites of a Method Transfer Transferring a Method to Automation Transferring a Method to Another Lab Determining Appropriate
More informationOur design fits everywhere
Fully Automated System with Basket Station BS 60 Our design fits everywhere ISO 9001 Up to 15 unattended paddle dissolution tests (USP Apparatus 2) with the AT 70smart Paddle Station Up to 10 unattended
More informationINTRODUCTION TO INDUSTRIAL PHARMACY LAB 1
INTRODUCTION TO INDUSTRIAL PHARMACY LAB 1 LAB INSTRUCTORS: LECT. ANAS T ARIK NAFEA ASSIST. LECT. ZAINAB HASSAN MAHDI Definition of Industrial Pharmacy The conversion of raw materials into certain dosage
More informationKinematic Viscosity Viscometer
Kinematic Viscosity Viscometer www.labtron.com BASING VIEW, BASINGSTOKE, HAMPSHIRE RG21 4RG, UK, PHONE: +44 208 4329420 KINEMATIC VISCOSITY TESTER LKVT-A11 It is used to determine automatically the kinematic
More informationApplication Note. Author. Abstract. Pharmaceuticals. Detlef Wilhelm ANATOX GmbH & Co. KG. Fuerstenwalde, Germany mau
Development, validation, and comparison of an HPLC method to analyze paracetamol and related impurities according to the European Pharmacopoeia (EP) and USP using the Agilent 1120 Compact LC and the Agilent
More informationRecent FDA Guidance For Industry; BCS Class 1 and 3 August 2015
Recent FDA Guidance For Industry; BCS Class 1 and 3 August 2015 Bryan Crist Scientific Affairs Manager, Agilent Technologies, Dissolution Systems Dissolution Exchange WebEx Bryan.crist@agilent.com August,
More informationAGILENT VACUUM-FORMED (VF) MOLDED DISSOLUTION VESSELS
AGILENT TECHNOLOGIES PRACTICAL SOLUTIONS NEWSLETTER VOLUME 15 ISSUE 3 PAGE 1 Agilent Vacuum-Formed (VF) Molded Dissolution Vessels PAGE 5 Learn. Solve. Discuss. PAGE 6 Gelatin Cross-linking and Dissolution
More informationDissolution Market. Vinod Mehta
Dissolution Market Vinod Mehta Objective Dissolution - What is it? Market and Product Situation Carl Zeiss Offering System Packages Competition What is Dissolution Solid dosage form assays in the pharmaceutical
More informationTURBIDITY TB 300 IR TB 210 IR TB 250 WL. General Catalogue April 2015
TURBIDITY 128 TB 300 IR TB 210 IR TB 250 WL www.lovibond.com General Catalogue April 2015 Turbidity measurement The term "turbidity is used to describe the cloudy or milky appearance of liquid or solid
More informationInternational Journal of PharmTech Research CODEN (USA): IJPRIF, ISSN: Vol.7, No.1, pp ,
International Journal of PharmTech Research CODEN (USA): IJPRIF, ISSN: 0974-4304 Vol.7, No.1, pp 139-147, 2014-2015 Pharmaceutical Prospective Process Validation Of Chloroquiune Tablets Vilas S Jadhav*,
More informationWho we are? Dissolution testing of modified release forms. Dissolution testing of immediate/ modified release forms
Dissolution testing of modified release forms Mumbai - 2013, May 3 Samir Haddouchi samir.haddouchi@sps-pharma.com 1 Dissolution testing of immediate/ modified release forms Mumbai - 2013, May 3 Samir Haddouchi
More informationBAUSCH + STRÖBEL. Technical Specification SP100/200 SP3060. Filling machine for bottles Page 1 of 6
BAUSCH + STRÖBEL Technical SP100/200 Filling machine for bottles 2012-06-04 Page 1 of 6 Contents 1. General 3 1.1 Application 3 2. Machine design 3 2.1 Basic construction 3 2.2 Container infeed 3 2.3 Dosing
More informationRev Experiment 10
Experiment 10 SPECTROPHOTOMETRIC DETERMINATION OF IRON IN DRINKING WATER 2 lab periods Reading: 1) Chapter 17, pg 393-403, Quantitative Chemical Analysis, 8 h Edition, Daniel C. Harris (7 th Edition: Chapter
More informationTE 65 MULTIPLEX SAND/WHEEL ABRASION TESTER
TE 65 MULTIPLEX SAND/WHEEL ABRASION TESTER Background Abrasive wear of industrial materials handling plant (e.g. transfer chutes, conveyors and hoppers) is a direct cause of repair, replacement and downtime
More informationAMI INTEGRITY TEST SYSTEM. Innovative solutions for the pharmaceutical industry
AMI INTEGRITY TEST SYSTEM Innovative solutions for the pharmaceutical industry AMI INTEGRITY TEST SYSTEM Innovative solutions for the pharmaceutical industry Our know-how Integrity challenges An innovative
More informationCONSTRUCTION SPECIFICATION FOR CRACK REPAIR - CONCRETE
ONTARIO PROVINCIAL STANDARD SPECIFICATION METRIC OPSS 932 MARCH 1998 CONSTRUCTION SPECIFICATION FOR CRACK REPAIR - CONCRETE 932.01 SCOPE 932.02 REFERENCES 932.03 DEFINITIONS TABLE OF CONTENTS 932.04 SUBMISSION
More informationSpectroscopy Solutions for Pharmaceuticals:
Spectroscopy Solutions for Pharmaceuticals: Confidence in Compliance to USP Using the Agilent Cary 60 UV-Vis Spectrophotometer White Paper Introduction UV-Vis spectroscopy is one of the most commonly
More informationINSPECTION EXPERTISE FROM A TO Z
Inspection Systems INSPECTION EXPERTISE Efficient detection of foreign objects in food and other products is critical to consumer safety and brand survival. Heat and Control offers a complete line of
More informationEnvironmental simulation chamber for cyclical temperature
MK series (E2.1) Environmental simulation chambers Environmental simulation chamber for cyclical temperature The BINDER environmental simulation chamber of the MK series is suitable for heat or cold testing
More informationThe procurement, qualification and calibration of lab instruments: An Overview
The procurement, qualification and calibration of lab instruments: An Overview The reliability of analytical data generated from chemical and physical analyses is critically dependent on three factors:
More informationIJPRD, 2013; Vol 5(02): April-2013 ( ) International Standard Serial Number
IJPRD, 2013; Vol 5(02): April-2013 (106 112) International Standard Serial Number 0974 9446 --------------------------------------------------------------------------------------------------------------------------------------------------
More informationInvestigation of Aqueous Ethylcellulose Dispersion in Extended Release Metformin Inert Matrices
Surelease Application Data Aqueous Ethylcellulose Dispersion Investigation of Aqueous Ethylcellulose Dispersion in Extended Release Metformin Inert Matrices ABSTRACT SUMMARY: In the present study, Surelease,
More informationAGILENT TECHNOLOGIES PRACTICAL SOLUTIONS NEWSLETTER
AGILENT TECHNOLOGIES PRACTICAL SOLUTIONS NEWSLETTER VOLUME 16 ISSUE 4 PAGE 1 The 280-DS Mechanical Qualification System Not Just an Agilent Thing PAGE 3 You Asked, We Listened: New Inline Filtration for
More informationFor simple and accurate mechanical calibration.
For simple and accurate mechanical calibration. Regulatory Specifications and Recommendations Changes since 2007 in regulatory specifications now also allow enhanced mechanical calibration of dissolution
More informationROTARY CONE SAMPLE DIVIDER VIBRATORY FEEDERS ULTRASONIC CLEANERS FRITSCH DIVIDING FEEDING CLEANING IDEAL FOR REPRESENTATIVE SAMPLE PREPARATION
FRITSCH DIVIDING FEEDING CLEANING IDEAL FOR REPRESENTATIVE SAMPLE PREPARATION THE FOUNDATION FOR EVERY PRECISE ANALYSIS! ROTARY CONE SAMPLE DIVIDER VIBRATORY FEEDERS ULTRASONIC CLEANERS FRITSCH Rotary
More informationMethods for determination of mercury in LP gas
(JLPGA-S-07) Methods for determination of mercury in LP gas 1. Scope This standard specifies the methods for determination of gaseous mercury in vaporized liquid liquefied petroleum (LP) gas. Remarks 1
More informationThe entire document shall be read and understood before proceeding with a test.
Packaged-Products Over 150 lb (68 kg) 1B2 0 0 1 ISTA 1 Series Non-Simulation Integrity Performance Test Procedure ISTA, Your Alliance in Transport Packaging, is the world leader in Performance Tests for
More informationQUALITY CONTROL DEPARTMENT EQUIPMENT LIST
1. Precisa Analytical balance XB220A Precisa FTIR room PRL/QC/001 (XB220A) 2. Centrifuge machine -- Multitech Instrumentation Room PRL/QC/002 3. Disintegration test machine 1 -- Multitech Instrumentation
More informationGENERAL EQUIPMENT. 3 Electronic analytical balance (320gr capacity - 0,001gr readability)
1. Laboratory Equipment Every laboratory equipment has its own calibration certificate and is operatedmaintained according to its manual guide. The equipment of the laboratory is listed below. The laboratory
More informationHardness Testers Surface Testers
Hardness Testers Surface Testers 49 3814 Analogue Bench Hardness Tester The 3814 Hardness Tester provides reliable Rockwell Hardness values on all types of metal and alloys, hard or soft, and in many shapes.
More information1. To test the Breaking Load of thread / yarn. thread / yarn. thread / yarn
TET FACILITIE AT QAE (G) NEW DELHI TORE DTE General tores and Textile Lab: Name of The dal Officer and Contact QAE(G) Anand Name of the dal Officer- Col Manish aran, QAO E-mail IDsqaegsdelhidgqa@nic.in
More informationHighway Engineering Materials - CE 400 Spring 2009 Aggregate Labs Draft Due February 4, 2009 Final Report Due February 11, 2009
Highway Engineering Materials - CE 400 Spring 2009 Aggregate Labs Draft Due February 4, 2009 Final Report Due February 11, 2009 January 21 st and January 28 th, 2009 T1) ASTM D4791-95: Flat and Elongated
More informationSECTION D.3 CHLOROPHYLL AND PHEOPHYTIN
SECTION D.3 CHLOROPHYLL AND PHEOPHYTIN 1.1 Scope and Application 1.1.1 This spectrophotometric method is used in the determination of chlorophyll a, b, and c, and pheophytin a. This method can be used
More information35 STATION DOUBLE ROTORY COMPRESSION MACHINE
Page 1 of 16 User Requirement Specifications 35 STATION DOUBLE ROTORY Equipment ID: T-35COMP 01 Page 2 of 16 Table of Contents 1.0 APPROVAL SIGNATURES 2.0 OVERVIEW 3.0 PROCESS DESCRIPTION 4.0 PRODUCTIVITY
More informationCHL Laboratory refrigerators
Application storing water and sewage samples, piezometer leachate storing ASA, GC or HPLC calibration standards storing reagents chemical storage general storage door lock (option) fan LCD display adjustable
More informationInnovative Solutions in Material Test Equipment
Innovative Solutions in Material Test Equipment Physical Test Solutions (PTS) offers a complete line of hardness testers for Rockwell testing applications. Reliability and repeatability is a trait in all
More informationThe use of surrogates for dissolution testing for Immediate Release (IR) formulations, when is it feasible?
The use of surrogates for dissolution testing for Immediate Release (IR) formulations, when is it feasible? Limin Zhang (Bristol-Myers Squibb Company) Andre Hermans (Merck & Co., Inc.) 2017 M-CERSI Workshop
More informationMODEL HB-3000 BRINELL HARDNESS TESTER. Feature & Use. Specifications
MODEL HB-3000 BRINELL HARDNESS TESTER. Mechanical reversing switch. It s steady and reliable for testing of curved surface. Precision conforms to the Standards of GB/T 231.2, ISO 6506-2 and ASTM E10 It
More informationKollidon SR: A polyvinyl acetate based excipient for DCsustained-release
Kollidon SR: A polyvinyl acetate based excipient for DCsustained-release oral dosage forms by Dr. Bernhard Fussnegger BASF Aktiengesellschaft, Ludwigshafen Strategic Marketing Pharma Excipients Introduction
More informationTHE PROCESS VALIDATION OF TABLET CONTAINING IRBESARTAN 300MG AND HYDROCHLOROTHIAZIDE 12.5mg Zamir Hussain, Baqir Shyum Naqvi & Muhammad Iqbal Nasiri
THE PROCESS VALIDATION OF TABLET CONTAINING IRBESARTAN 300MG AND HYDROCHLOROTHIAZIDE 12.5mg Zamir Hussain, Baqir Shyum Naqvi & Muhammad Iqbal Nasiri Faculty of Pharmacy, Department of Pharmaceutics, Hamdard
More informationMACRO PWD POWDER MACRODOSING MACHINE
MACRO PWD POWDER MACRODOSING MACHINE MACRO PWD POWDER MACRODOSING MACHINE MINIMUM & MAXIMUM BOTTLE INLET/ OUTLET LOADING SENSORS. BOTTLE CONFIRMATION PRESENCE AT MACHINE INFEED. NO BOTTLE NO FILL CONTROL
More informationFines content. SCAN-CM 66:05 Accepted Mechanical and chemical pulps
Accepted 2005 Mechanical and chemical pulps Fines content 0 Introduction This SCAN-test Method has been prepared to make it possible to determine the fines content of mechanical and chemical pulps. This
More informationLABORATORY GRINDING MILLS
LABORATORY GRINDING MILLS Union Process Research And Laboratory Attritors Produce Fine And Homogeneous Dispersions Quickly And Repeatedly Under Scientific Conditions. History From a revolutionary idea
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 QUALITY CONTROL SERVICES, INC. 2340 SE 11 th Avenue Portland, OR 97214 George Cahill Phone: 503 236 2712 CALIBRATION Valid To: September
More informationEquipment Needs and Maintenance, What is Required in the Wet Chemistry Laboratory for Manual Method Testing
Equipment Needs and Maintenance, What is Required in the Wet Chemistry Laboratory for Manual Method Testing Debra Waller NJDEP-Office of Quality Assurance debra.waller@dep.nj.gov Evaluation of Suppliers
More informationUniversity of Sulaimani College of Pharmacy Dept. of Pharmaceutics 5 th stage Second Semester
University of Sulaimani College of Pharmacy Dept. of Pharmaceutics 5 th stage Second Semester 2/26/2018 Industrial Pharmacy II, Dr. rer. nat. Rebaz Ali 1 Outlines Powder flow Introduction Factor affecting
More information10. Validated Normal Phase HPLC Method for the Determination. Fulvestrant is primarily used in the treatment of hormone receptor
229 10. Validated Normal Phase HPLC Method for the Determination of Fulvestrant in Pharmaceutical Dosage Forms 10.1 Introduction Fulvestrant is primarily used in the treatment of hormone receptor positive
More informationDesign of Controlled Release Non-erodible Polymeric Matrix Tablet Using Microwave Oven-assisted Sintering Technique
Pharmaceutics Design of Controlled Release Non-erodible Polymeric Matrix Tablet Using Microwave Oven-assisted Sintering Technique Patel DM Patel BK Patel A Patel CN Department of Pharmaceutics and Pharmaceutical
More informationColor, True and Apparent, LR
Color, True and Apparent, LR DOC316.53.01252 Platinum-Cobalt Standard Method 1, 2, 3 Method 8025 3 to 200 color units Scope and application: For water, wastewater and seawater; equivalent to NCASI method
More informationCalibration, Qualification and Preventive Maintenance Services Overview
, Qualification and Preventive Maintenance Services Overview PVSR Corporation, recently certified as an ISO 9001 organization is emerging as an innovator providing the Pharmaceutical, Medical Devices and
More informationDissolution Testing and Acceptance Criteria for Immediate-Release Solid Oral Dosage Form Drug Products Containing High Solubility Drug Substances
Dissolution Testing and Acceptance Criteria for Immediate-Release Solid Oral Dosage Form Drug Products Containing High Solubility Drug Substances Guidance for Industry U.S. Department of Health and Human
More informationTechnical Procedure for Ultraviolet Spectroscopy
Technical Procedure for Ultraviolet Spectroscopy 1.0 Purpose - This procedure specifies the required elements for the performance check and use of the ultraviolet spectrophotometers. 2.0 Scope - This procedure
More informationGOOD LABORATORY PRACTICE Agilent Cary 8454 UV-Visible Spectroscopy System
GOOD LABORATORY PRACTICE Agilent Cary 8454 UV-Visible Spectroscopy System Introduction The objective of good laboratory practice (GLP) is to obtain accurate and precise results, using a measurement system
More informationDrilling Technologies
Drilling Technologies WTC-11-2550 September 2011 API 13C (ISO 13501) Sieve and Conductance Testing Global Wire de Colombia S. A. S. Intertek Westport Technology Center 6700 Portwest Drive Houston, Texas
More informationArtisan Technology Group is your source for quality new and certified-used/pre-owned equipment
Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment FAST SHIPPING AND DELIVERY TENS OF THOUSANDS OF IN-STOCK ITEMS EQUIPMENT DEMOS HUNDREDS OF MANUFACTURERS SUPPORTED
More informationENGLISH UNIVERSAL BRINELL UNIVERSAL BRINELL UNIVERSAL BRINELL UNIVERSAL BRINELL UNIVERSAL BRINELL 75 YEARS HARDNESS TESTING
ENGLISH UNIVERSAL BRINELL UNIVERSAL BRINELL UNIVERSAL BRINELL UNIVERSAL BRINELL UNIVERSAL BRINELL WOLPERT GROUP WOLPERT Group manufactures a comprehensive range of hardness testers for over 75 years. Our
More informationInspection and Approval of Asphalt Field Laboratories
Standard Method of Test for July 6, 2010 Inspection and Approval of Asphalt Field Laboratories SCDOT Designation: SC-M-404 (5/10) 1. SCOPE 1.1 This method covers the process for inspection and approval
More informationExotherm Dynamic Crystallization of Fat and Fat based products. A Complete Solution. Bench top instrument made of :
Exotherm 2000 Dynamic Crystallization of Fat and Fat based products A Complete Solution Bench top instrument made of : - Exotherm 2000, the main part where the product temperature and sheare are precisely
More informationMulti-function VAC measuring instrument
Data sheet testo 480 Multi-function VAC measuring instrument testo 480 Cutting-edge technology for professionals Measurement of all VAC-relevant parameters: flow velocity, temperature, humidity, pressure,
More informationUSP s Perspective on Drug Product Performance Test
USP s Perspective on Drug Product Performance Test Course Overview 1. The concept of in vitro dissolution Definition and application 2. Compendial dissolution/ drug release testing 3. Method development
More informationECO INSTRUMENT (NS M)
Computerized AC Servo Control Universal Material Testing Machine Safety Protection Frame Rubber Type 180 degree grip or 90 degree peel fixture Bending Test Compression Test Specimen Gripping Fixture Product
More informationMASS PER CUBIC FOOT (METER), YIELD, AND AIR CONTENT (GRAVIMETRIC) OF FRESHLY MIXED CONCRETE (Kansas Test Method KT-20)
5.9.20 MASS PER CUBIC FOOT (METER), YIELD, AND AIR CONTENT (GRAVIMETRIC) OF FRESHLY MIXED CONCRETE (Kansas Test Method ) 1. SCOPE This method of test covers the procedure for determining the mass per cubic
More informationPresented By. NicolScales.com Contact Us
Presented By Dallas Ft Worth ustin Houston NicolScales.com 800.225.8181 Contact Us Nicol Scales & Measurement is an ISO ccredited Calibration Company that has provided calibration, repair and sales of
More informationAvailable online at ScienceDirect. Procedia Engineering 132 (2015 )
Available online at www.sciencedirect.com ScienceDirect Procedia Engineering 132 (2015 ) 811 815 The Manufacturing Engineering Society International Conference, MESIC 2015 Metrological Regulations for
More informationNIR Checkmaster Near-infrared spectroscopy On-line analysis of active ingredients during tablet production
NIR Checkmaster Near-infrared spectroscopy On-line analysis of active ingredients during tablet production Slash release times with NIR Innovative features Fully automatic assay of tablet weight, hardness,
More information* Current monograph name valid until May 1, 2017 ** New monograph name valid as of May 1, 2012, mandatory as of May 1, 2017
Technical Information EUDRAGIT L 30 D-55 Specification and Test Methods Ph. Eur. Methacrylic Acid - Ethyl Acrylate Copolymer (1:1) Dispersion 30 Per Cent USP/NF Methacrylic Acid Copolymer Dispersion -
More informationAssessment of free floating dust and abrasion particles of treated seeds as a parameter of the quality of treated seeds
PHYSICAL METHOD Assessment of free floating dust and abrasion particles of treated seeds as a parameter of the quality of treated seeds HEUBACH TEST Authors: ESA STAT Dust Working Group PHYSICAL METHOD
More informationUnderstanding the Characteristics and Establishing Acceptance Criteria for Analytical Methods Validation
Understanding the Characteristics and Establishing Acceptance Criteria for Analytical Methods Validation Ying Verdi IVT LAB WEEK EUROPE June 2017 Partners in Health Since 1919 Regulatory View of Method
More informationOBSERVATIONS OF AN OPERATOR IN THE LAB IMPROVING LABORATORY RESULTS. Shane Jordan-Hill. Merck Millipore
OBSERVATIONS OF AN OPERATOR IN THE LAB IMPROVING LABORATORY RESULTS Paper Presented by: Shane Jordan-Hill Author: Shane Jordan-Hill, Business Manager Environmental Analysis, Merck Millipore 6 th Annual
More informationCells and Calibration Standards
TECHNICAL BULLETIN 913 Cells and Calibration Standards h earch A Res na IE IO RT IF N al tic ly 9001:2008 CE YEAR Rud ol p lytical Ana Se r 20 h rc Guarantee ce vi dolph Res ea Ru Cells, NIST Traceable
More informationDraft proposals for Test methods for close-coupled solar water heating systems - Reliability and safety
IEA/SHC Task 57, Subtask B Draft proposals for new test procedures B4: Final Draft Draft proposals for Test methods for close-coupled solar water heating systems - Reliability and safety HE Zinian Beijing
More informationfor IND and RDRC Regulated PET Compounding
Overview of USP Chapter for IND and RDRC Regulated PET Compounding Distributed Manufacturing of PET Radiopharmaceuticals for Multi-Center Clinical Trials SNM, Annual Meeting Toronto, Ontario, Canada
More information