ABB Bomem Inc. 585 Charest Blvd. East, Suite 300 Québec, Québec G1K 9H4, Canada Phone (418) , Fax (418)

Size: px
Start display at page:

Download "ABB Bomem Inc. 585 Charest Blvd. East, Suite 300 Québec, Québec G1K 9H4, Canada Phone (418) , Fax (418)"

Transcription

1 585 Charest Blvd. East, Suite 300 Québec, Québec G1K 9H4, Canada Phone (418) , Fax (418) PROPRIETARY The information contained herein is proprietary to and should be considered a trade secret of It shall not be reproduced in whole or in part without the written authorization of

2 Test Report Document No: BOM Date: March 04, 2009 i Table of Contents 1. INTRODUCTION Scope of document Purpose of document Applicable Documents Acronyms GENERALITIES Test flow QA INSPECTION FUNCTIONAL INSPECTION SYSTEM LEVEL VERIFICATIONS AND SYSTEM TESTS Test Summary Tests Nominal FOV mapping Secondary FOV mapping Spectral range Spectral resolution Spectral stability Sampling rate Noise Pressure test SUMMARY...33 APPENDIX A APPENDIX B QA INSPECTION DOCUMENT...34 ILS CHARACTERIIZATION...38 Do not copy or distribute.

3 Test Report Document No: BOM Date: March 04, 2009 ii List of Figures Figure 1: Test flow chart... 7 Figure 2: FOV mapping of MCT for x axis before rework activities Figure 3: FOV mapping of MCT for x axis after rework activities Figure 4: FOV mapping of MCT for y axis before rework activities Figure 5: FOV mapping of MCT for y axis after rework activities Figure 6: FOV mapping of InSb for x axis before rework activities Figure 7: FOV mapping of InSb for x axis after rework activities Figure 8: FOV mapping of InSb for y axis before rework activities Figure 9: FOV mapping of InSb for y axis after rework activities Figure 10: FOV mapping of MCT for x axis before rework activities Figure 11: FOV mapping of MCT for y axis before rework activities (at 6.5 degrees above nominal FOV).. 18 Figure 12: FOV mapping of InSb for x axis before rework activities Figure 13: FOV mapping of InSb for y axis before rework activities (at 6.5 degrees above nominal FOV) Figure 14: Angle between nominal and secondary FOV Figure 15: MCT spectral range (raised cosine apodization) Figure 16: InSb spectral range (raised cosine apodization) Figure 17: ILS fitting of cm -1 before rework activities Figure 18: FWHM of cm -1 water vapour line before rework activities Figure 19: ILS fitting of cm -1 after rework activities Figure 20: FWHM of cm -1 water vapour line after rework activities Figure 21: ILS fitting of 1550 nm DFB laser diode after rework activities Figure 22: Spectral stability estimated from 1918 cm-1 water vapor line over 2 hours of operation Figure 23: Sampling rate of instrument at 1 cm -1 of resolution Figure 24: NESR of MCT between 950 to 1050 cm Figure 25: NESR of InSb between 2650 to 2750 cm Figure 26: Apparent temperature of MCT between 950 to 1050 cm -1 when observing a 50 o C scene Figure 27: NEDT of MCT between 950 to 1050 cm Figure 28: Apparent temperature of InSb between 2650 to 2750 cm -1 when observing a 50 o C scene Figure 29: NEDT of InSb between 2650 to 2750 cm Figure 30: Initial QA inspection (1/2) Figure 31: Initial QA inspection (2/2) Figure 32: Final QA inspection Do not copy or distribute. List of Tables Table 1: Functional inspections... 9 Table 2: Test Summary Table 3: FWHM of nominal FOV Table 4: FWHM of secondary FOV Table 5: Spectral limit of MCT and InSb channels Table 6: Fitting parameters of cm -1 before rework activities Table 7: Fitting parameters of cm -1 after rework activites Table 8: Fitting parameters of 1550 nm DFB laser diode Table 9: RMS NESR Table 10: RMS NEDT... 32

4 Test Report Document No: BOM Date: March 04, 2009 iii Table 11: Internal pressure of instrument Do not copy or distribute.

5 Test Report Document No: BOM Date: March 04, 2009 iv Document Change Record Issue Rev. Date Change Description 1 - March 04, 2009 First release of document Distribution List Recipient 1- Document Version ABB project team Met Office Do not copy or distribute.

6 Test Report Document No: BOM Date: March 04, INTRODUCTION 1.1 SCOPE OF DOCUMENT The scope of this document is to cover system level verifications and tests of instrument that follow January 2009 delivery under RMA number Results of previous deliveries (1995 & 2005) are also presented to compare instrument performances. 1.2 PURPOSE OF DOCUMENT The purpose of this document is to present system level verifications and test results of instrument that follow January 2009 delivery. System level verifications have one main goal (see Figure 1): - Perform initial characterization (before rework) of instrument to validate and quantify issues reported by the customer. System tests have one main goal (see Figure 1): - Perform final characterization (after rework) of instrument to validate system performances and to compare initial (before rework) and final (after rework) results. 1.3 APPLICABLE DOCUMENTS AD 1 AD 2 BOM Verification & Test Plan Flow Chart.vsd BOM Verification & Test Plan.doc 1.4 ACRONYMS DFB FOV FWHM ILS InSb MCT NEDT NESR PPM PSI Airborne Research Interferometer Evaluation System Distributed FeedBack Field Of View Full Width at Half Maximum Instrument Line Shape Indium Antimony Mercury Cadmium Telluride Noise Equivalent Delta Temperature Noise Equivalent Spectral Radiance Parts Per Million Pounds per Square Inche Do not copy or distribute.

7 Test Report Document No: BOM Date: March 04, QA RMA RMS ZnSe Quality Assurance Return Material Authorization Root Mean Square Zinc Selenide Do not copy or distribute.

8 Test Report Document No: BOM Date: March 04, GENERALITIES 2.1 TEST FLOW Figure 1 shows the test flow of instrument. There are six test activities and two integration activities. Initial QA inspection Functional inspection System level verifications Functional inspection System tests Rework activities Final QA inspection Packing & delivery Test activities Integration activities Figure 1: Test flow chart Do not copy or distribute.

9 Test Report Document No: BOM Date: March 04, QA INSPECTION QA inspections have been done before (initial inspection) and after (final inspection) rework activities of instrument. Inspection reports are attached in Appendix A. Do not copy or distribute.

10 Test Report Document No: BOM Date: March 04, FUNCTIONAL INSPECTION Table 1 presents functional inspections of instrument. Note that functional inspections have been performed before and after rework activities. Table 1: Functional inspections ID Description Goal Passed / Failed (before rework activities) Passed / Failed (after rework activities) 1 Connection / disconnection to / from instrument Verify integrity of cable connectors by connecting / disconnecting to / from instrument. Passed / Failed 1 Passed / Failed 2 Resolution setting Verify resolution selection using software (click on each resolution and verify the number of points). Passed / Failed Passed / Failed 3 Gain setting for both channels Verify gain selection using software (click on each gain and verify each spectrum signal amplitude). Passed / Failed Passed / Failed 4 Pointing mirror position setting Verify that mirror control is working properly using software (click on mirror position feature and select different positions). Passed / Failed Passed / Failed 5 Interferogram data acquisition and real time display Verify that interferogram is saving properly and that real time display is working properly. Passed / Failed Passed / Failed 6 Housekeeping/status data acquisition and real time display Verify that housekeeping status is saving properly and that real time display is working properly. Passed / Failed Passed / Failed 1 See initial inspection report on Figure 30. Do not copy or distribute.

11 Test Report Document No: BOM Date: March 04, ID Description Goal Passed / Failed (before rework activities) Passed / Failed (after rework activities) 7 Real time of raw or calibrated data (radiance, apparent intensity) Verify that raw or calibrated data are display properly in real time. Passed / Failed Passed / Failed 8 Data visualization of saved data (interferograms, raw or calibrated spectra, status) Verify that acquired interferograms can be transformed to spectrum and calibrate. Check also, if they can be display properly with their status. Passed / Failed Passed / Failed 9 Miscellaneous functionalities to ease data browsing and inspection Verify miscellaneous functionalities like ones find in Visualization and Curves Viewer tabs. Passed / Failed Passed / Failed 10 Shortcuts to ease manipulation of the software and instrument commanding (e.g. from the aircraft) Verify shortcuts. Passed / Failed Passed / Failed Do not copy or distribute.

12 Test Report Document No: BOM Date: March 04, SYSTEM LEVEL VERIFICATIONS AND SYSTEM TESTS 5.1 TEST SUMMARY Table 2 presents a summary of test results. Each test is described in section 5.2. Table 2: Test Summary Require ment ID 2 Section # Test Name Requirement Results Compliance Status R1a Nominal FOV mapping Total FOV shall be less than 3 degrees. Total FOV is less than 3 degrees for x and y axis for both channels. Compliant R1b Secondary FOV mapping Secondary FOV shall be inexistent. Secondary FOV is inexistent. Compliant R Spectral range Spectral range shall be < 500 cm -1 (lower limit) and > 2940 cm -1 (upper limit) at 1 % of relative amplitude for MCT and InSb respectively. MCT lower spectral limit at 1 % relative amplitude is 499 cm -1. InSb upper spectral limit at 1 % relative amplitude is 3160 cm -1. Compliant R Spectral resolution FWHM of cm -1 water vapor line shall be < 0.7 cm -1 (unapodized). FWHM of cm -1 water vapor line is 0.67 cm -1. Compliant R Spectral stability Spectral stability shall be 10 ppm over 2 hours of operation. Standard deviation of variation of cm -1 water vapour line position is 0.7 ppm. Compliant R Sampling rate Sampling rate: shall be 120 scans/min at 1 cm -1 of resolution. Sampling rate is 121 scans/min. Compliant 2 Requirement ID is defined in BOM Verification & Test Plan.doc Do not copy or distribute.

13 Test Report Document No: BOM Date: March 04, Require ment ID 2 Section # Test Name Requirement Results Compliance Status R Noise R Pressure test RMS NESR shall be < 3.5*10-7 W/(cm 2 sr cm -1 ) between 950 to 1050 cm -1 (MCT) and < 4.82*10-9 W/(cm 2 sr cm -1 ) between 2650 to 2750 cm -1 (InSb). RMS NEDT shall be < 0.2 K between 950 to 1050 cm -1 (MCT) and < 0.2 K between 2650 to 2750 cm -1 (InSb). Internal pressure of instrument shall remain constant (at 6 psi) over a period of 1 hour. MCT RMS NESR is 5.6*10-9 W/(cm 2 sr cm -1 ) between 950 to 1050 cm -1. InSb RMS NESR is 5.9*10-10 W/(cm 2 sr cm -1 ) between 2650 to 2750 cm -1. MCT RMS NEDT is 0.11 K. InSb RMS NEDT is 0.09 K. Pressure remains at 6 psi over one hour of operation. Compliant Compliant Do not copy or distribute.

14 Test Report Document No: BOM Date: March 04, TESTS Nominal FOV mapping Requirement: [R1a]: Total FOV shall be less than 3 degrees. Mapping of nominal FOV has been done for both channels (MCT and InSb) before and after rework activities. Figure 2 to Figure 9 show normalized response with respect to the angle for x axis and y axis. A red rectangular box has been drawn on each figure at 50 % of amplitude Relative amplitude Degree Figure 2: FOV mapping of MCT for x axis before rework activities Do not copy or distribute.

15 Test Report Document No: BOM Date: March 04, Relative amplitude Degree Figure 3: FOV mapping of MCT for x axis after rework activities Relative amplitude Degree Figure 4: FOV mapping of MCT for y axis before rework activities Do not copy or distribute.

16 Test Report Document No: BOM Date: March 04, Relative amplitude Degree Figure 5: FOV mapping of MCT for y axis after rework activities Relative amplitude Degree Figure 6: FOV mapping of InSb for x axis before rework activities Do not copy or distribute.

17 Test Report Document No: BOM Date: March 04, Relative amplitude Degree Figure 7: FOV mapping of InSb for x axis after rework activities Relative amplitude Degree Figure 8: FOV mapping of InSb for y axis before rework activities Do not copy or distribute.

18 Test Report Document No: BOM Date: March 04, Relative amplitude Degree Figure 9: FOV mapping of InSb for y axis after rework activities Table 3 indicates FWHM of nominal FOV for x and y axes of both channels. Table 3: FWHM of nominal FOV FWHM (before rework activities) FWHM (after rework activities) MCT x axis 38.0 mrad 39.6 mrad MCT y axis 38.2 mrad 40.5 mrad InSb x axis 37.8 mrad 38.4 mrad InSb y axis 40.0 mrad 41.7 mrad Secondary FOV mapping Requirement: [R1b]: Secondary FOV shall be inexistent. Mapping of secondary FOV has been done for both channels (MCT and InSb) before and after rework activities. Figure 10 to Figure 13 show normalized response with respect to the angle for x axis and y axis. A red rectangular box has been drawn on each figure at 50 % of amplitude. Do not copy or distribute.

19 Test Report Document No: BOM Date: March 04, Relative amplitude Degree Figure 10: FOV mapping of MCT for x axis before rework activities Relative amplitude Degree Figure 11: FOV mapping of MCT for y axis before rework activities (at 6.5 degrees above nominal FOV) Do not copy or distribute.

20 Test Report Document No: BOM Date: March 04, Relative amplitude Degree Figure 12: FOV mapping of InSb for x axis before rework activities Relative amplitude Degree Figure 13: FOV mapping of InSb for y axis before rework activities (at 6.5 degrees above nominal FOV) Do not copy or distribute.

21 Test Report Document No: BOM Date: March 04, Table 4 indicates FWHM of secondary FOV for x and y axis of both channels. Table 4: FWHM of secondary FOV FWHM (before rework activities) FWHM (after rework activities) MCT x axis 36.7 mrad Secondary FOV is non-existent MCT y axis 38.4 mrad Secondary FOV is non-existent InSb x axis 35.3 mrad Secondary FOV is non-existent InSb y axis 38.4 mrad Secondary FOV is non-existent Note: Data obtained during initial characterization indicates that the ratio between on-axis (direct beam) and offaxis (first reflected beam) FOV amplitudes is approximately 0.04 while theoretical value is 0.03 (calculated using a bare ZnSe substrate). Moreover, using a ZnSe entrance window with a wedge of 1.3 degrees, we simulated an angle of 6.4 degrees between nominal and secondary FOV as shown on Figure 14. ZnSe window (1.3 degrees of wedge) Secondary FOV 6.4 degrees Nominal FOV Figure 14: Angle between nominal and secondary FOV Do not copy or distribute.

22 5.2.3 Spectral range Test Report Document No: BOM Date: March 04, Requirement: [R2]: Spectral range shall be < 500 cm-1 (lower limit) and > 2940 cm-1 (upper limit) at 1 % of relative amplitude for MCT and InSb respectively. Spectral range has been measured for both channels (MCT and InSb). Figure 15 and Figure 16 show results of MCT and InSb respectively Relative amplitude Wavenumber (cm -1 ) Figure 15: MCT spectral range (raised cosine apodization) Do not copy or distribute.

23 Test Report Document No: BOM Date: March 04, Relative amplitude Wavenumber (cm-1) Figure 16: InSb spectral range (raised cosine apodization) Table 5 indicate spectral limit of MCT and InSb channels at 0 %, 1%, 5 % and 10 % of relative amplitude. Table 5: Spectral limit of MCT and InSb channels Description Measured at 0 % Measured at 1 % Measured at 5 % Measured at 10 % MCT lower spectral limit InSb upper spectral limit 455 cm cm cm cm cm cm cm cm Spectral resolution Requirement: [R3]: FWHM of cm-1 water vapor line shall be less than 0.7 cm -1. ILS has been characterized using cm -1 water vapor line and a 1550 nm DFB laser diode. Figure 17 to Figure 21 show theoretical ILS fit and measured ILS while Table 6 to Table 8 include parameters that have been used to fit theoretical ILS. Do not copy or distribute.

24 Test Report Document No: BOM Date: March 04, Residual [-] Fit [-] Theoritical ILS fit Measurement Wavenumber [cm -1 ] Figure 17: ILS fitting of cm -1 before rework activities Table 6: Fitting parameters of cm -1 before rework activities Fitting parameter Gas pressure (fix parameter) Gas temperature (fix parameter) Path length (fix parameter) H 2 O (fit parameter) Spectral stretch (fit parameter) FOV diameter (fit parameter) FOV off-axis diameter (fit parameter) Value kpa K 200 cm ppm 52.2 ppm 45.0 mrad 9.0 mrad According to Table 6, the effective FOV is equals to the design value of 45.0 mrad and the FOV is misaligned by 20 % of the FOV size. FWHM of cm -1 water vapour line is 0.72 cm -1 as shown on Figure 18. Do not copy or distribute.

25 Test Report Document No: BOM Date: March 04, Figure 18: FWHM of cm -1 water vapour line before rework activities Residual [-] Fit [-] Theoretical ILS fit Measurement Wavenumber [cm -1 ] Do not copy or distribute. Figure 19: ILS fitting of cm -1 after rework activities

26 Test Report Document No: BOM Date: March 04, Table 7: Fitting parameters of cm -1 after rework activites Fitting parameter Gas pressure (fix parameter) Gas temperature (fix parameter) Path length (fix parameter) H 2 O (fit parameter) Spectral stretch (fit parameter) FOV diameter (fit parameter) FOV off-axis diameter (fit parameter) Value kpa (20 PSI) K 200 cm ppm 11.5 ppm 45.0 mrad 1.9 mrad According to Table 7, the effective FOV is equals to the design value of 45.0 mrad, and the FOV is misaligned by 4 % of the FOV size. FWHM of cm -1 water vapour line is 0.67 cm -1 as shown on Figure 20. Figure 20: FWHM of cm -1 water vapour line after rework activities Do not copy or distribute.

27 Test Report Document No: BOM Date: March 04, Theoretical ILS fit Measurement a.u Wavenumber [cm -1 ] Figure 21: ILS fitting of 1550 nm DFB laser diode after rework activities Table 8: Fitting parameters of 1550 nm DFB laser diode Fitting parameter Value Laser wavenumber (fit parameter) cm -1 FOV diameter (fit parameter) FOV off-axis diameter (fit parameter) 45.4 mrad 1.9 mrad Shear (fit parameter) µm According to Table 8, the effective FOV is 0.9 % larger than the design value of 45.0 mrad, and the FOV is misaligned by 4 % of the FOV size. FWHM of laser line is 1.56 cm -1. See Appendix B regarding ILS characterization from Raphaël Desbiens Spectral stability Requirement: [R4]: Spectral stability shall be 10 ppm over a period of 2 hours. Spectral stability of instrument has been estimated from cm -1 water vapour line over two hours of operation. Estimated stability is 0.7 ppm rms over two hours of operation as shown on Figure 22. Do not copy or distribute.

28 Test Report Document No: BOM Date: March 04, Spectral shift (ppm) Scan number (@ 121 scan/min.) Figure 22: Spectral stability estimated from 1918 cm-1 water vapor line over 2 hours of operation Sampling rate Requirement: [R5]: Sampling rate shall be 120 scans/min at 1 cm-1 of resolution. Sampling rate is reported by software and monitored by firmware. Sampling rate is 121 scan/min as shown on Figure 23. Do not copy or distribute.

29 Test Report Document No: BOM Date: March 04, Figure 23: Sampling rate of instrument at 1 cm -1 of resolution Noise Requirement: [R6]: RMS NESR shall be < 3.5*10-7 W/(cm 2 sr cm -1 ) between 950 to 1050 cm -1 (MCT) and < 4.82*10-9 W/(cm 2 sr cm -1 ) between 2650 to 2750 cm -1 (InSb). RMS NEDT shall be < 0.2 K between 950 to 1050 cm -1 (MCT) and < 0.2 K between 2650 to 2750 cm -1 (InSb). NESR has been measured between 950 to 1050 cm-1 (MCT) and 2650 to 2750 cm-1 (InSb). Figure 24 and Figure 25 shown results while Table 9 indicates values. Do not copy or distribute.

30 Test Report Document No: BOM Date: March 04, E E-08 NESR (W/(cm 2 sr cm -1 ) 5.00E E E E Wavenumber (cm-1) Figure 24: NESR of MCT between 950 to 1050 cm E E E-09 NESR (W/cm 2 sr cm -1 ) 1.00E E E E E E Wavenumber (cm -1 ) Do not copy or distribute. Figure 25: NESR of InSb between 2650 to 2750 cm -1

31 Test Report Document No: BOM Date: March 04, Table 9: RMS NESR Spectral interval RMS NESR (W/(cm 2 sr cm -1 )) 950 to 1050 cm -1 (MCT) 5.6* to 2750 cm -1 (InSb) 5.9*10-10 Apparent temperature and NEDT have also been measured between 950 to 1050 cm-1 (MCT) and 2650 to 2750 cm-1 (InSb). Figure 26 to Figure 29 shown results while Table 10 indicates values of NEDT Apparent temperature (K) Wavenumber (cm -1 ) Figure 26: Apparent temperature of MCT between 950 to 1050 cm -1 when observing a 50 o C scene Do not copy or distribute.

32 Test Report Document No: BOM Date: March 04, NEDT (K) Wavenumber (cm -1 ) Figure 27: NEDT of MCT between 950 to 1050 cm Apparent temperature (K) Wavenumber (cm -1 ) Figure 28: Apparent temperature of InSb between 2650 to 2750 cm -1 when observing a 50 o C scene Do not copy or distribute.

33 Test Report Document No: BOM Date: March 04, NEDT (K) Wavenumber (cm-1) Figure 29: NEDT of InSb between 2650 to 2750 cm -1 Table 10: RMS NEDT Spectral interval RMS NEDT (K) 950 to 1050 cm -1 (MCT) to 2750 cm -1 (InSb) Pressure test Requirement: [R7]: Internal pressure of instrument shall remain constant (at 6 PSI) over a period of 1 hour. Table 11 indicates pressure value during two hours of operation. Elapsed time (hours) Table 11: Internal pressure of instrument Pressure psi psi psi Do not copy or distribute.

34 Test Report Document No: BOM Date: March 04, SUMMARY instrument was tested at ABB Bomem before shipment. Results of tests and compliance are summarized in Table 2. Do not copy or distribute.

35 Test Report Document No: BOM Date: March 04, APPENDIX A QA INSPECTION DOCUMENT Do not copy or distribute.

36 Test Report Document No: BOM Date: March 04, Figure 30: Initial QA inspection (1/2) Do not copy or distribute.

37 Test Report Document No: BOM Date: March 04, Figure 31: Initial QA inspection (2/2) Do not copy or distribute.

38 Test Report Document No: BOM Date: March 04, Figure 32: Final QA inspection Do not copy or distribute.

39 Test Report Document No: BOM Date: March 04, APPENDIX B ILS CHARACTERIIZATION Do not copy or distribute.

40 Test Report Document No: BOM Date: March 04, Do not copy or distribute.

41 Test Report Document No: BOM Date: March 04, Do not copy or distribute.

42 Test Report Document No: BOM Date: March 04, Do not copy or distribute.

43 Test Report Document No: BOM Date: March 04, Do not copy or distribute.

44 Test Report Document No: BOM Date: March 04, Do not copy or distribute.

45 Test Report Document No: BOM Date: March 04, Do not copy or distribute.

46 Test Report Document No: BOM Date: March 04, Do not copy or distribute.

47 Test Report Document No: BOM Date: March 04, End of document - Do not copy or distribute.

MB3600. Versatile FT-NIR analyzer designed for your industry. The most reliable FT-NIR specifically designed for QA / QC.

MB3600. Versatile FT-NIR analyzer designed for your industry. The most reliable FT-NIR specifically designed for QA / QC. MB3600 Versatile FT-NIR analyzer designed for your industry. The most reliable FT-NIR specifically designed for QA / QC. MB3600 FT-NIR Spectrometer for your industry Our industry needs versatility and

More information

MB3000-PH. FT-IR Spectrometer for the Pharmaceutical Industry Designed for QA/QC, Formulation, Research and Development Laboratory Applications.

MB3000-PH. FT-IR Spectrometer for the Pharmaceutical Industry Designed for QA/QC, Formulation, Research and Development Laboratory Applications. MB3000-PH FT-IR Spectrometer for the Pharmaceutical Industry Designed for QA/QC, Formulation, Research and Development Laboratory Applications. MB3000-PH FT-IR Spectrometer for the Pharmaceutical Industry

More information

Liquid Metal Cleanliness Analyzer for Continuous Monitoring

Liquid Metal Cleanliness Analyzer for Continuous Monitoring LiMCA CM Liquid Metal Cleanliness Analyzer for Continuous Monitoring n Direct, in-line and real-time inclusion measurement n Fully automated n Minimum maintenance n Integrated to the Process To maintain

More information

LiMCA CM. Liquid Metal Cleanliness Analyzer for Continuous Monitoring

LiMCA CM. Liquid Metal Cleanliness Analyzer for Continuous Monitoring LiMCA CM Liquid Metal Cleanliness Analyzer To maintain their edge, world-class aluminum plants must be able to monitor in real-time and with precision the quality of their liquid aluminum. They must be

More information

MB3000-PH FT-IR spectrometer for the pharmaceutical industry Designed for QA/QC, formulation, research and development laboratory applications

MB3000-PH FT-IR spectrometer for the pharmaceutical industry Designed for QA/QC, formulation, research and development laboratory applications Measurement & Analytics Measurement made easy MB3000-PH FT-IR spectrometer for the pharmaceutical industry Designed for QA/QC, formulation, research and development laboratory applications MB3000-PH FT-IR

More information

SUPPLIER QUALITY SYSTEM SURVEY

SUPPLIER QUALITY SYSTEM SURVEY SUPPLIER QUALITY SYSTEM SURVEY DATE (M/d/yy): 55 Dragon Court Woburn, MA 01801 Tel: (781) 933-7300 Fax: (781) 935-4529 GENERAL INFORMATION Supplier Name: Physical Address: City: State: Zip: Sr. Company

More information

SUPER-SEALED LIQUID FLOW CELLS OPERATION AND USER MANUAL

SUPER-SEALED LIQUID FLOW CELLS OPERATION AND USER MANUAL SUPER-SEALED LIQUID FLOW CELLS OPERATION AND USER MANUAL Optimized Design for Continuous Flow Applications Rugged, heavy duty design compatible with continuous use Large IR port diameter to accommodate

More information

Raman Spectroscopy Measurement System and Data Analysis for Characterization of Composite Overwrapped Pressure Vessels (COPVs)

Raman Spectroscopy Measurement System and Data Analysis for Characterization of Composite Overwrapped Pressure Vessels (COPVs) Raman Spectroscopy Measurement System and Data Analysis for Characterization of Composite Overwrapped Pressure Vessels (COPVs) Phillip A. Williams and Buzz Wincheski Nondestructive Evaluation Sciences

More information

Automated Data Analysis & Report Generation of Key Characteristics (KCs) Measurements

Automated Data Analysis & Report Generation of Key Characteristics (KCs) Measurements ABSTRACT The purpose of this paper is to discuss an automated analysis and report generation of key characteristics measurement data. Key characteristics (KCs) are those features of a part whose measurements

More information

Unrivaled. Sensitive. Flexible. Agilent Cary 600 Series FTIR

Unrivaled. Sensitive. Flexible. Agilent Cary 600 Series FTIR Unrivaled. Sensitive. Flexible. Agilent Cary 600 Series FTIR Agilent Cary 600 Series FTIR Agilent Technologies is your premier resource and partner for molecular spectroscopy. With the addition of the

More information

ISO Non-destructive testing of welds Ultrasonic testing Use of automated phased array technology

ISO Non-destructive testing of welds Ultrasonic testing Use of automated phased array technology INTERNATIONAL STANDARD ISO 13588 First edition 2012-10-01 Non-destructive testing of welds Ultrasonic testing Use of automated phased array technology Contrôle non destructif des assemblages soudés Contrôle

More information

Supplier Quality Requirements Document (SQRD)

Supplier Quality Requirements Document (SQRD) Supplier Quality Requirements Document (SQRD) D127135 Rev. AE Supplier Quality Requirements Document Page 2 of 8 1.0 SCOPE This document applies to all worldwide production suppliers of Coherent, Inc.

More information

Measurement & Analytics Measurement made easy. Advanced spectroscopy software for quantitative and qualitative analysis Horizon MB FTIR

Measurement & Analytics Measurement made easy. Advanced spectroscopy software for quantitative and qualitative analysis Horizon MB FTIR Measurement & Analytics Measurement made easy Advanced spectroscopy software for quantitative and qualitative analysis Horizon MB FTIR Intuitive Spectroscopy Software for MB3000 and MB3600 Spectrometers

More information

DETECTION OF LASER ULTRASONIC SURFACE DISPLACEMENT BY WIDE APERTURE FIBER OPTIC AMPLIFIER M.L. Rizzi and F. Corbani CESI, Milano, Italy

DETECTION OF LASER ULTRASONIC SURFACE DISPLACEMENT BY WIDE APERTURE FIBER OPTIC AMPLIFIER M.L. Rizzi and F. Corbani CESI, Milano, Italy DETECTION OF LASER ULTRASONIC SURFACE DISPLACEMENT BY WIDE APERTURE FIBER OPTIC AMPLIFIER M.L. Rizzi and F. Corbani CESI, Milano, Italy Abstract: In the frame of the European Project INCA, CESI is in charge

More information

SPECIFICATION FOR NICKEL ALLOY FORGINGS

SPECIFICATION FOR NICKEL ALLOY FORGINGS SPECIFICATION FOR NICKEL ALLOY FORGINGS 01 (Identical with ASTM Specification B 564-00 except that certification has been made mandatory.) 1. Scope 1.1 This specification covers forgings of nickel alloy

More information

GOOD LABORATORY PRACTICE Agilent Cary 8454 UV-Visible Spectroscopy System

GOOD LABORATORY PRACTICE Agilent Cary 8454 UV-Visible Spectroscopy System GOOD LABORATORY PRACTICE Agilent Cary 8454 UV-Visible Spectroscopy System Introduction The objective of good laboratory practice (GLP) is to obtain accurate and precise results, using a measurement system

More information

Langham Creek Machine Works, Inc.

Langham Creek Machine Works, Inc. Langham Creek Machine Works, Inc. A Can Do Company Specializing in the Machining of Exotic Materials Quality Assurance Manual UNCONTROLLED COPY This manual is the property of Langham Creek Machine Works,

More information

The MIDAC Low Moisture in Corrosive Gases Process Analyzer for Continuous On-Line Applications

The MIDAC Low Moisture in Corrosive Gases Process Analyzer for Continuous On-Line Applications AP-156 The MIDAC Low Moisture in Corrosive Gases Process Analyzer for Continuous On-Line Applications Dr. Peter G. Zemek MIDAC Corporation 17911 Fitch Avenue Irvine, California USA 92614 Email: sales@midac.com

More information

Guide to Weak-Story Tool

Guide to Weak-Story Tool Appendix A Guide to Weak-Story Tool A.1 Introduction The Guidelines introduce topics that may be unfamiliar to some engineers. Though the calculations required in Chapters 4 and 5 are not conceptually

More information

FT-IR. varian, inc. 660/670/680-IR.

FT-IR. varian, inc. 660/670/680-IR. FT-IR varian, inc. 660/670/680-IR ft-ir spectroscopy SoLUTIONS www.varianinc.com VARIAN, INC. The World s Best FT-IR In a Class of Their Own The world leader in molecular spectroscopy innovation now introduces

More information

Laser Speckle Reducer LSR-3000 Series

Laser Speckle Reducer LSR-3000 Series Datasheet: LSR-3000 Series Laser Speckle Reducer LSR-3000 Series Speckle noise from a laser-based system is reduced by dynamically diffusing the laser beam. A diffuser is bonded to a thin elastic membrane,

More information

Agilent 8453 Spectrophotometer and UV-visible ChemStation for Dissolution Testing. Solution Bundles for automated Dissolution Testing

Agilent 8453 Spectrophotometer and UV-visible ChemStation for Dissolution Testing. Solution Bundles for automated Dissolution Testing Agilent 8453 Spectrophotometer and UV-visible ChemStation for Dissolution Testing Solution Bundles for automated Dissolution Testing General considerations UV benefits Simple, rugged and reliable equipment

More information

Supplier Quality Manual

Supplier Quality Manual Supplier Quality Manual DuraTech Industries Supplier Quality Manual Page 2 of 13 Table of Contents 1.0 Introduction 1.1. Overview 1.2. Scope 1.3. Customer Responsibilities 3 3 3 2.0 Supplier Quality Expectations

More information

The ratio of total pulse energy to maximum pulse power.

The ratio of total pulse energy to maximum pulse power. Guide Number Revision Effective Date Page C 04/13/2015 1 of 6 Subject Date of Issue Parent Command Media Optical Coating Laser Damage Resistance and Laser Damage Threshold Testing Plan 09/08/2014 P708

More information

Supplier Quality System Survey

Supplier Quality System Survey Supplier Division Address Date: Quality Contact: Reports to: Title: Title: Phone Fax e-mail Plant size (ft 2 ): Union: Contract expires: Total Employment: Direct Labor: Shifts: QA employees: This report

More information

Merrimack and Suwannee ACSS/TW/HS285 Conductor Stress-Strain Tests. Southwire Company

Merrimack and Suwannee ACSS/TW/HS285 Conductor Stress-Strain Tests. Southwire Company Merrimack and Suwannee ACSS/TW/HS285 Conductor Stress-Strain Tests Southwire Company NEETRAC Project Number: 05-204 November, 2005 A Research Center of the Georgia Institute of Technology Requested by:

More information

SUPPLIER QUALITY SYSTEM SURVEY

SUPPLIER QUALITY SYSTEM SURVEY SUPPLIER NAME DATE WEB ADDRESS ADDRESS CITY STATE ZIP PUBLICLY OR PRIVATELY HELD? TEL FAX CONTACT INFORMATION POSITION CONTACT NAME PHONE FAX E-MAIL PRESIDENT OPERATIONS MGR. QUALITY MGR. ENGINEERING MGR.

More information

477 kcmil, 3M Brand Composite Conductor Mechanical Properties, Volume 1 Tensile and Stress-Strain Tests

477 kcmil, 3M Brand Composite Conductor Mechanical Properties, Volume 1 Tensile and Stress-Strain Tests 477 kcmil, 3M Brand Composite Conductor Mechanical Properties, Volume 1 Tensile and Stress-Strain Tests Minnesota Mining and Manufacturing (3M) Company Purchase Order 0000227040 NEETRAC Project Number:

More information

CRYSTALAS. UV Optics System for Excimer Laser Based Crystallization of Thin Silicon Films

CRYSTALAS. UV Optics System for Excimer Laser Based Crystallization of Thin Silicon Films L A S E R S Y S T E M G M B H CRYSTALAS UV Optics System for Excimer Laser Based Crystallization of Thin Silicon Films CRYSTALAS The New Optical Crystallization System CRYSTALAS is an excimer laser-based

More information

Spatter-Free Stable Conduction and Keyhole Welding of Copper with 275 Watt Blue Laser

Spatter-Free Stable Conduction and Keyhole Welding of Copper with 275 Watt Blue Laser Spatter-Free Stable Conduction and Keyhole Welding of Copper with 275 Watt Blue Laser Abstract Laser welding of highly reflective materials such as copper has been problematic for infrared lasers due to

More information

Multiplexed 3D FRET imaging in deep tissue of live embryos Ming Zhao, Xiaoyang Wan, Yu Li, Weibin Zhou and Leilei Peng

Multiplexed 3D FRET imaging in deep tissue of live embryos Ming Zhao, Xiaoyang Wan, Yu Li, Weibin Zhou and Leilei Peng Scientific Reports Multiplexed 3D FRET imaging in deep tissue of live embryos Ming Zhao, Xiaoyang Wan, Yu Li, Weibin Zhou and Leilei Peng 1 Supplementary figures and notes Supplementary Figure S1 Volumetric

More information

QUALITY MANUAL REVISION CONTROL LOG. Number Rev Rev Date SS Description of Change By Chk App

QUALITY MANUAL REVISION CONTROL LOG. Number Rev Rev Date SS Description of Change By Chk App Superecedes: QPOL-1001.24 Section: REVISION CONTRO LOG REVISION CONTROL LOG Number Rev Rev Date SS Description of Change By Chk App QPOL-0001 18 02-JAN-12 A Review & Update Quality Manual JD CN JD QPOL-0001

More information

Vendor Qualification Survey

Vendor Qualification Survey 1200 West 96 th St Minneapolis, MN 55431 Ph: 952-888-7900 Fax: 952-888-2719 Vendor Qualification Survey Vendor Information Company Name: Date: Address: City: Phone Number: email address: Product or Service

More information

Real-time tablet API analysis: a comparison of a palm-size NIR spectrometer to HPLC method

Real-time tablet API analysis: a comparison of a palm-size NIR spectrometer to HPLC method Real-time tablet API analysis: a comparison of a palm-size NIR spectrometer to HPLC method Presented by: Chris Pederson, Product Applications Engineer, JDS Uniphase Corp. Co-Authors: Nada O Brien, JDS

More information

EPICOR, INCORPORATED QUALITY ASSURANCE MANUAL

EPICOR, INCORPORATED QUALITY ASSURANCE MANUAL EPICOR, INCORPORATED QUALITY ASSURANCE MANUAL Revision: 6 Date 05/18/09 EPICOR, INCORPORATED 1414 E. Linden Avenue P.O. Box 1608 Linden, NJ. 07036-0006 Tel. 1-908-925-0800 Fax 1-908-925-7795 Table of Contents:

More information

NEW Spectrophotometer Packages from Beckman Coulter THE COMPLETE SPECTRO SOLUTION

NEW Spectrophotometer Packages from Beckman Coulter THE COMPLETE SPECTRO SOLUTION FL-8650C NEW Spectrophotometer Packages from Beckman Coulter THE COMPLETE SPECTRO SOLUTION UP TO$6,168 IN SAVINGS! Look inside for details! c o m ORDER NOW FROM VWR INTERNATIONAL! 1-800-932-5000 or www.vwr.com

More information

Supplier Quality Requirements for Control and Use of Digital Product Definition SQR-004 REVISION DATE: 22 NOVEMBER 2010

Supplier Quality Requirements for Control and Use of Digital Product Definition SQR-004 REVISION DATE: 22 NOVEMBER 2010 Supplier Quality Requirements for Control and SQR-004 REVISION DATE: 22 NOVEMBER 2010 Approved David Copeland, Manager Supply Chain Quality Suppliers may view this document via the Internet at: http://www.voughtaircraft.com/suppliers/

More information

A New Continuous Wave 2500W Semiconductor Laser Vertical Stack

A New Continuous Wave 2500W Semiconductor Laser Vertical Stack A New Continuous Wave 2500W Semiconductor Laser Vertical Stack Xiaoning Li 1,2, Chenhui Peng 1, Yanxin Zhang 1, Jingwei Wang 1, Lingling Xiong 1, Pu Zhang 1, Xingsheng Liu 1,3 1 State Key Laboratory of

More information

ASA-100 QUALITY MANUAL

ASA-100 QUALITY MANUAL Your Company Name ASA-100 QUALITY MANUAL Origination Date: XXXX Document Identifier: Date: Document Revision: QMS-00 Quality Manual Latest Revision Date Abstract: This quality manual describes (your Company's)

More information

Automating AFP Tuning Using a Laser Sensor

Automating AFP Tuning Using a Laser Sensor 11ATC-0340 Automating AFP Tuning Using a Laser Sensor Josh Cemenska Copyright 2011 SAE International ABSTRACT Electroimpact Automatic Fiber Placement (AFP) machines lay-up composite parts by accurately

More information

Standard Test Method for Measuring Compressive Properties of Thermal Insulations 1

Standard Test Method for Measuring Compressive Properties of Thermal Insulations 1 Designation: C 165 95 Standard Test Method for Measuring Compressive Properties of Thermal Insulations 1 This standard is issued under the fixed designation C 165; the number immediately following the

More information

Technical Standard Order

Technical Standard Order Department of Transportation Federal Aviation Administration Aircraft Certification Service Washington, D.C. TSO-C164a Effective Date: 10/02/15 Technical Standard Order Subject: Night Vision Goggles 1.

More information

Manual Analytical Data Adjustment. Environmental Health Laboratories A division of MAS Technology Corporation

Manual Analytical Data Adjustment. Environmental Health Laboratories A division of MAS Technology Corporation Corporate Policy for Quality Assurance Manual Analytical Data Adjustment Environmental Health Laboratories A division of MAS Technology Corporation 110 S. Hill Street South Bend, IN 46617 Phone: (219)

More information

MD Power Supply -48Vdc +Bias Tee

MD Power Supply -48Vdc +Bias Tee 215988-001MD Power Supply -48Vdc +Bias Tee Operation and Maintenance Manual Mitec Telecom Inc. Designers and manufacturers of telecom and wireless products 9000 Trans Canada, Pointe-Claire, Quebec, Canada

More information

INSULATED JOINTS. Monolithic Joints for Electrically Isolating Pipelines

INSULATED JOINTS. Monolithic Joints for Electrically Isolating Pipelines INSULATED JOINTS Monolithic Joints for Electrically Isolating Pipelines INSULATED JOINTS TUBE TURNS INSULATED JOINTS are a pre-fabricated, non-separable, union used to isolate specific sections of pipelines

More information

ICP Expert software. Technical Overview. Introduction

ICP Expert software. Technical Overview. Introduction ICP Expert software Technical Overview Introduction The Agilent 5110 ICP-OES provides fast sample analysis, using less gas, without compromising on performance for tough samples. It has been designed for

More information

Non-destructive testing of welds Ultrasonic testing Techniques, testing levels, and assessment

Non-destructive testing of welds Ultrasonic testing Techniques, testing levels, and assessment Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO 17640 Third edition 2017-10 Non-destructive testing of welds Ultrasonic testing Techniques, testing levels, and assessment Essais non destructifs

More information

Production Part Approval Process (PPAP)

Production Part Approval Process (PPAP) Document ID: QN-M04 Page 1 of 13 Aerojet Rocketdyne Quality Note M04 Date: 03/16/2016 Table of Contents 1. PURPOSE... 2 2. APPLICATION... 2 2.1. Resubmissions... 2 2.2. Applicable Documents... 3 3. CONVENTIONS...

More information

Advanced Test Equipment Rentals ATEC (2832)

Advanced Test Equipment Rentals ATEC (2832) Established 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) Agilent 11896A and 8169A Polarization Controllers Product Overview Automatic polarization state adjustments for polarizationdependent

More information

REPORT: Study of a Sample of Raw Iron Ore Pellets. with MAYA Elemental Laser Analyzer in a Stream Mode

REPORT: Study of a Sample of Raw Iron Ore Pellets. with MAYA Elemental Laser Analyzer in a Stream Mode REPORT: Study of a Sample of Raw Iron Ore Pellets with MAYA Elemental Laser Analyzer in a Stream Mode Laser Distance Spectrometry. Report on the analysis of samples of unprocessed iron ore pellets using

More information

BRAVO. Innovation with Integrity. The next generation handheld Raman spectrometer. Raman

BRAVO. Innovation with Integrity. The next generation handheld Raman spectrometer. Raman BRAVO The next generation handheld Raman spectrometer Innovation with Integrity Raman BRAVO Raises Standards. BRAVO sets new standards in performance, safety and ease of use of handheld Raman analyzers

More information

REVISION HISTORY. ANY COMPANY ACP: Rev. A. SUBJECT: Gemstone Check In Process REVISION DESCRIPTION OF CHANGE DATE.

REVISION HISTORY. ANY COMPANY ACP: Rev. A. SUBJECT: Gemstone Check In Process REVISION DESCRIPTION OF CHANGE DATE. Page 1 of 9 REVISION HISTORY REVISION DESCRIPTION OF CHANGE DATE A Initial Release Supersedes AB22A633 Preparation and Checkout of Gemstone Runs, AB22A635 Standard Operating Procedures, section 8.1 Gemstone

More information

TurboMass GC/MS Software

TurboMass GC/MS Software TurboMass GC/MS Software GAS CHROMATOGRAPHY / MASS SPECTROMETRY P R O D U C T N O T E Today s demanding laboratory requires software that is easy to learn yet offers sophisticated instrument control, robust

More information

Keywords: leaf fluorescence; canopy fluorescence; fluorescence in-filling, O 2 -A band

Keywords: leaf fluorescence; canopy fluorescence; fluorescence in-filling, O 2 -A band Chlorophyll Fluorescence Detection with a High-Spectral Resolution Spectrometer through in-filling of the O 2 -A band as function of Water Stress in Olive Trees P.J. Zarco-Tejada 1, O. Pérez-Priego 1,

More information

Preliminary Investigation of Near Infrared Spectroscopy for Green Sand Component Identification

Preliminary Investigation of Near Infrared Spectroscopy for Green Sand Component Identification Preliminary Investigation of Near Infrared Spectroscopy for Green Sand Component Identification Dr. Scott R. Giese University of Northern Iowa Research Objective Green sand molding characterization for

More information

PCA ELECTRONICS, INC. Quality Manual

PCA ELECTRONICS, INC. Quality Manual PCA ELECTRONICS, INC. Conforms to ISO 9001:2015 (c) 2018 PCA Electronics, Inc.; all rights reserved. This document may contain proprietary information and may only be released to third parties with approval

More information

SUPPLEMENTARY INFORMATION

SUPPLEMENTARY INFORMATION Measuring subwavelength spatial coherence with plasmonic interferometry Drew Morrill, Dongfang Li, and Domenico Pacifici School of Engineering, Brown University, Providence, RI 02912, United States List

More information

SUPPLIER QUALITY ASSESSMENT

SUPPLIER QUALITY ASSESSMENT Supplier Organization Name: Supplier Number: Street Address: Date of This Audit: City, State, Zip Code: Date of Last Audit: Country: of Employees: Main Phone Number: of Buildings/Size: Fax Number: Principal

More information

PRODUCT SPECIFICATION

PRODUCT SPECIFICATION 1 of 23 B 1.0 GENERAL Scope This specification covers the insulation piercing Quickie Backplane Connector System designed for Backplane-to-flat cable (round conductor) interconnection in low power applications.

More information

Application of DO-254 Level A (Appendix B) Design Assurance Objectives of. Elemental Analysis. Mixed Signal (Analog/Digital) Discrete Circuitry

Application of DO-254 Level A (Appendix B) Design Assurance Objectives of. Elemental Analysis. Mixed Signal (Analog/Digital) Discrete Circuitry Application of DO-254 Level A (Appendix B) Design Assurance Objectives of Elemental Analysis To Mixed Signal (Analog/Digital) Discrete Circuitry By Dave Duncan Purple Seal Inc. THE INFORMATION CONTAINED

More information

LiMCA III Liquid Metal Cleanliness Analyzer

LiMCA III Liquid Metal Cleanliness Analyzer A B B M E A S U R E M E N T & A N A LY T I C S LiMCA III Liquid Metal Cleanliness Analyzer 2 LIMCA III LIQUID METAL CLEANLINESS ANALYZER The LiMCA III is a versatile aluminium cleanliness analyzer that

More information

Supplier Nonconformance Notification and Waiver Procedure

Supplier Nonconformance Notification and Waiver Procedure DOC NUMER: 1 of 11 Supplier Nonconformance Notification and Waiver Procedure Created y Checked y Checked y Approvals Name Title Date Signature Tamara Crismundo Jesus Villegas Denis eaulieu Sr. Quality

More information

SUPPLIER QUALITY REQUIREMENTS

SUPPLIER QUALITY REQUIREMENTS SUPPLIER QUALITY REQUIREMENTS 25 TH February 2010 CONTENTS 1. Introduction 2. Scope 3. Approval Validity 4. Monitoring of Suppliers Performance 5. Quality Clauses 6. Goods Inwards 7. Selection of Quality

More information

Demo Script. Quality Assurance Classification: Internal and for Partners. SAP Business ByDesign Reference Systems. Version:

Demo Script. Quality Assurance Classification: Internal and for Partners. SAP Business ByDesign Reference Systems. Version: ) Demo Script Classification: Internal and for Partners Version: 20180131 SAP Business ByDesign Reference Systems Disclaimer This demo script is based on the ByD Reference System for the Almika model company.

More information

Request for Proposal of GEO-KOMPSAT-2 EEE Part Procurement HEATER (FLEXIBLE)

Request for Proposal of GEO-KOMPSAT-2 EEE Part Procurement HEATER (FLEXIBLE) Korea Aerospace Research Institute GEO-KOMPSAT Program Office 45 Eoeun-dong Yuseong-gu Daejeon 305-333 Republic of Korea Request for Proposal of GEO-KOMPSAT-2 EEE Part Procurement HEATER (FLEXIBLE) Date:

More information

Global Greenhouse Gas Observation by Satellite

Global Greenhouse Gas Observation by Satellite Global Greenhouse Gas Observation by Satellite Greenhouse gases Observing SATellite Figure 1. Overview of GOSAT ( JAXA) The Greenhouse Gases Observing Satellite (GOSAT) will be the world s first satellite

More information

ABSTRACT INTRODUCTION

ABSTRACT INTRODUCTION 1 Developing a System-Level Monitoring Instrument for UV-Curing Systems Dr. Peyman H. Dehkordi, James L. Riner, and Douglas E. Ingram Applied Computing Technology, Inc. 1010D Commerce Park Drive Oak Ridge,

More information

Bridgelux IB Series. Product Data Sheet DS340

Bridgelux IB Series. Product Data Sheet DS340 Bridgelux IB Series Product Data Sheet DS340 Introduction IB Series The Bridgelux IB Series TM is a simple plug and play solution that enables highly uniform glare free lighting. The unique 240 degree

More information

Quality Assurance for particle beam therapy

Quality Assurance for particle beam therapy Quality Assurance for particle beam therapy PTCOG Educational Workshop, Essen 2013 Heidelberg Ion Beam Therapy Center at the University Hospital German Cancer Research Center, Heidelberg, Germany Outline

More information

NEMA/MITA Standards. Good Refurbishment Practices for Medical Imaging Equipment

NEMA/MITA Standards. Good Refurbishment Practices for Medical Imaging Equipment NEMA/MITA Standards Good Refurbishment Practices for Medical Imaging Equipment Jeroen van Nistelrooij, R.S Director of Marketing Rick Cardoso, R.S Sr. Field Marketing Manager July, 2016 Purpose of NEMA/MITA

More information

Qualification of High-Performance Liquid Chromatography Systems

Qualification of High-Performance Liquid Chromatography Systems Qualification of High-Performance Liquid Chromatography Systems L. Huber Hewlett- Packard GmbH, Waldbronn December 1998 Manuscript of an article published in BioPharm, Vol 11 Number 11, November 1998,

More information

Thermo Scientific Nicolet is20 FTIR Spectrometer

Thermo Scientific Nicolet is20 FTIR Spectrometer Thermo Scientific Nicolet is20 FTIR Spectrometer The perfect fit for busy analytical laboratories Step into a new dimension of reliable answers Trusted by thousands of users around the world, the Thermo

More information

Evaluation of Ultrasonic Attenuation and Estimation of Ultrasonic Grain Noise in Copper

Evaluation of Ultrasonic Attenuation and Estimation of Ultrasonic Grain Noise in Copper Evaluation of Ultrasonic Attenuation and Estimation of Ultrasonic Grain Noise in Copper T. Stepinski and P. Wu Uppsala University, Signals & Systems, SE 751 0, Uppsala, Sweden Abstract. This paper presents

More information

Environmental Testing

Environmental Testing Nebraska Center for Excellence in Electronics 4740 Discovery Drive Lincoln, NE 68521-5376 Phone: 402-323-NCEE (6233) Fax: 402-323-6238 Environmental Testing Company: Crystal Group 850 Kacena Road Hiawatha,

More information

Optical Gateway Interface (OGI) Connector System Qualification Test Result Summary Telcordia GR 1435 CORE

Optical Gateway Interface (OGI) Connector System Qualification Test Result Summary Telcordia GR 1435 CORE Optical Gateway Interface (OGI) Connector System Qualification Test Result Summary Telcordia GR 1435 CORE (Generic Requirements for Multi-Fiber Optical Connectors) Technical Report June 2003 78-8133-4065-6-A

More information

PS850110E ACCEPTANCE TESTING OF INCOMING AEROSPACE FASTENERS

PS850110E ACCEPTANCE TESTING OF INCOMING AEROSPACE FASTENERS SECURITY CLASS: UNCLASSIFIED COML. CLASS: UNCLASSIFIED DOCUMENT IDENTIFIER SUBJECT: MATERIAL ENGINEERING & TECHNOLOGY DEVELOPMENT ENGINEERING DIVISION / ENGINEERING & DEVELOPMENT GROUP PREPARED BY: A.

More information

ISO INTERNATIONAL STANDARD. Non-destructive testing of welds Ultrasonic testing Techniques, testing levels, and assessment

ISO INTERNATIONAL STANDARD. Non-destructive testing of welds Ultrasonic testing Techniques, testing levels, and assessment INTERNATIONAL STANDARD ISO 17640 Second edition 2010-12-15 Non-destructive testing of welds Ultrasonic testing Techniques, testing levels, and assessment Contrôle non destructif des assemblages soudés

More information

Developments in Ultrasonic Phased Array Inspection I

Developments in Ultrasonic Phased Array Inspection I 6th International Conference on NDE in Relation to Structural Integrity for Nuclear and Pressurized Components October 2007, Budapest, Hungary For more papers of this publication click: www.ndt.net/search/docs.php3?mainsource=70

More information

Counterfeit Control Plan for Electrical, Electronic and Electromechanical Parts

Counterfeit Control Plan for Electrical, Electronic and Electromechanical Parts Your Logo Counterfeit Control Plan for Electrical, Electronic and Electromechanical Parts Origination Date: XXXX Document Identifier: Date: Project: Document Status: Counterfeit Control Plan for Electrical,

More information

Investigation of high voltage induced damage of GaN grown on Si substrate by terahertz nano-imaging and spectroscopy 1.

Investigation of high voltage induced damage of GaN grown on Si substrate by terahertz nano-imaging and spectroscopy 1. Investigation of high voltage induced damage of GaN grown on Si substrate by terahertz nano-imaging and spectroscopy 1 Introduction May 18, 2017 Anis Rahman, PhD Applied Research & Photonics, Inc. ( ARP

More information

Electro-Miniatures Corp. 68 West Commercial Ave. Moonachie New Jersey Phone: Fax:

Electro-Miniatures Corp. 68 West Commercial Ave. Moonachie New Jersey Phone: Fax: Electro-Miniatures Corp. 68 West Commercial Ave. Moonachie New Jersey 07074 Phone: 201-460-0510 Fax: 201-935-8153 1. PURPOSE GENERAL SUPPLIER REQUIREMENTS CLAUSE This general requirement describes the

More information

MMOG/LE Evaluation. Data Collection Process Improvement

MMOG/LE Evaluation. Data Collection Process Improvement MMOG/LE Evaluation Data Collection Process Improvement 1. STRATEGY AND IMPROVEMENT 1.2 Objectives 1.2.2 Key Performance Indicators (KPIs) shall cover objectives for all areas of the materials planning

More information

Supplier Quality Requirements Q Clauses

Supplier Quality Requirements Q Clauses Supplier Quality Requirements Q Clauses QA-0100032-Rev D DCR# 1603139 4/8/16 The following supplier quality requirements (Q Clauses), when specifically referenced in the Purchase Order by number, form

More information

Global Water globalw.com. Global Water. Instrumentation, Inc.

Global Water globalw.com. Global Water. Instrumentation, Inc. Global Water Instrumentation, Inc. 151 Graham Road P.O. Box 9010 College Station, TX 77842-9010 Telephone : 800-876-1172 International : (979) 690-5560, Fax : (979) 690-0440 e-mail : globalw@globalw.com

More information

HOMOSIL, HERASIL 1, 2 and 3

HOMOSIL, HERASIL 1, 2 and 3 HOMOSIL, HERASIL 1, 2 and 3 1. GENERAL PRODUCT DESCRIPTION Heraeus HOMOSIL, HERASIL 1, 2, and 3 are optical quartz glass grades manufactured by flame fusion of natural quartz crystals. They combine excellent

More information

QUALITY MANUAL. Revision K. JADE PRECISION MEDICAL COMPONENTS, LLC 3063 B Philmont Avenue Huntingdon Valley, PA 19006

QUALITY MANUAL. Revision K. JADE PRECISION MEDICAL COMPONENTS, LLC 3063 B Philmont Avenue Huntingdon Valley, PA 19006 QUALITY MANUAL Revision K JADE PRECISION MEDICAL COMPONENTS, LLC 3063 B Philmont Avenue Huntingdon Valley, PA 19006 Page: 1 of 14 Table of Contents 1. Purpose & Scope... 2 2. Applicable Standards... 2

More information

Standards Australia LICENCE

Standards Australia LICENCE Standards Australia LICENCE Title: Licensee: Date: Conditions of use (Click here for full conditions of Licence) Check if current WEB SEARCH Find similar documents StandardsWatch (Info and Login) Visit

More information

CAPITAL AVIONICS, INC. Quality Manual

CAPITAL AVIONICS, INC. Quality Manual CAPITAL AVIONICS, INC. Issued 31 July 2018 Conforms to ISO 9001:2015 2018 ; all rights reserved. This document may contain proprietary information and may only be released to third parties with approval

More information

LA LM56UXS1. Near Ultraviolet Laser Diode (406 nm), Multi Mode. Important Safety Advice

LA LM56UXS1. Near Ultraviolet Laser Diode (406 nm), Multi Mode. Important Safety Advice Important Safety Advice Depending on the mode of operation, these devices emit highly concentrated visible light which can be hazardous to the human eye. Products which incorporate these devices have to

More information

Engineering Tender Specification

Engineering Tender Specification PELICAN INSTRUMENT PROJECT BRAGG INSTITUTE Document Number: NBIP-ES-407-2018 Job No 0287P-1 Engineering Tender Specification Design, Manufacture and Delivery of The Slit System for the Pelican Instrument

More information

Application Note 6: On-line Reaction Monitoring of Alcoholic Fermentation

Application Note 6: On-line Reaction Monitoring of Alcoholic Fermentation Application Note 6: On-line Reaction Monitoring of Alcoholic Fermentation NMR spectroscopy is an excellent technique for monitoring organic reactions due to its high degree of functional group specificity.

More information

Prometheus-L Reactor Building Layout

Prometheus-L Reactor Building Layout Prometheus-L Reactor Building Layout Two Main Options for the Final Optic (1) SiO 2 or CaF 2 wedges 85 40 cm stiff, lightweight, actively cooled, neutron transparent substrate 4.6 m (2) Grazing incidence

More information

Production Part Approval Process (PPAP)

Production Part Approval Process (PPAP) Document ID: QN-M04 Page 1 of 15 Aerojet Rocketdyne Quality Note M04 Date: 12/23/2016 Table of Contents 1. PURPOSE... 2 2. APPLICATION... 2 2.1. Resubmissions... 3 2.2. Applicable Documents... 3 3. CONVENTIONS...

More information

SMC Corporation of America Supplier Quality Assurance Manual

SMC Corporation of America Supplier Quality Assurance Manual QAD 7.4.1B Revision # 15 SMC Corporation of America Supplier Quality Assurance Manual Approval Date Revision Mark Miller 2/19/16 15 Confidentiality Statement All SMC information including but not limited

More information

Nonconforming Material System Procedure

Nonconforming Material System Procedure Page: 1 of 13 Nonconforming Material System Procedure Page: 2 of 13 1.0 PURPOSE AND SCOPE 1.1 Purpose a. To define the system for identifying, segregating and controlling, and disposing of material, identified

More information

FAIRFIELD GLOBAL SUPPLIER QUALITY PROGRAM

FAIRFIELD GLOBAL SUPPLIER QUALITY PROGRAM FAIRFIELD GLOBAL SUPPLIER QUALITY PROGRAM 1 QA.101 rev. 3-2012 PURPOSE OF THIS STANDARD Fairfield is committed to continuous product quality improvement. Our management team is convinced that only with

More information

GURU JAMBHESHWAR, UNIVERSITY OF SCIENCE & TECHNOLOGY, HISAR Department of Bio & Nano Tech. GJUS&T, Hisar.

GURU JAMBHESHWAR, UNIVERSITY OF SCIENCE & TECHNOLOGY, HISAR Department of Bio & Nano Tech. GJUS&T, Hisar. GURU JAMBHESHWAR, UNIVERSITY OF SCIENCE & TECHNOLOGY, HISAR INVITATION FOR QUOTATION You are invited to submit your most competitive quotation for the following goods with item wise detailed specifications

More information

LA LS56HRC1. Hyper red Laser Diode (660 nm), Single Mode. Important Safety Advice

LA LS56HRC1. Hyper red Laser Diode (660 nm), Single Mode. Important Safety Advice Important Safety Advice Depending on the mode of operation, these devices emit highly concentrated visible light which can be hazardous to the human eye. Products which incorporate these devices have to

More information

High-Speed Infrared Imaging for Characterization of the Additive Manufacturing Process

High-Speed Infrared Imaging for Characterization of the Additive Manufacturing Process 14 th Quantitative InfraRed Thermography Conference High-Speed Infrared Imaging for Characterization of the Additive Manufacturing Process by M.A. Langevin*, A. Huot*, S. Boubanga*, P. Lagueux*, E. Guyot*

More information

Supplier. Quality. Manual

Supplier. Quality. Manual DEPARTMENT: Quality Assurance Title: Supplier - Quality Manual ISSUED ON: 29-Jul-09 PAGE 1 of 12 Supplier Quality Manual Distribution to: Electronically controlled. If this document is in hard copy it

More information