RapidArc: Clinical Implementation

Size: px
Start display at page:

Download "RapidArc: Clinical Implementation"

Transcription

1 RapidArc: Clinical Implementation Fang-Fang Yin, PhD Q. Jackie Wu, PhD Acknowledgements Team efforts from staff at Duke Radiation Oncology, especially to Dr. J Chang, Dr. J O Daniel for providing slide information Technical and financial supports from Varian Medical Systems Duke University Medical Center Clinical Implementation of VMAT Fundamentals for RapidArc Infrastructure/Installation Acceptance testing/commissioning Planning Delivery Quality assurance Fundamentals for VMAT Intensity Modulated arc therapy (VMAT) An arc-based approach to IMRT To be delivered on a conventional linear accelerator with a conventional MLC During an arc, the leaves of the MLC move and dose rate changes continuously as the gantry rotates RapidArc is one format of VMAT 1

2 The Principle of IMRT: Dose Painting Conventional 3-field RT Expected 3-field IMRT Static and Rotational IMRT Static gantry IMRT VMAT Typical dose distribution OAR PTV Beam Profile OAR PTV Multiple apertures At each angle One aperture At each angle The Format of Cone-Beam IMRT Static Gantry IMRT Rotational IMRT Volumetric Rotational IMRT Options Existing Planning Systems Eclipse (Varian) Duke choice ERGO++/Monaco (Elekta) Pinnacle SmartArc (Philips) Prowess (Prowess) o One of 7 fields One arc from 179 o 181 o Count-clockwise Existing Delivery Systems RapidArc (Varian) Duke choice VMAT (Elekta) Cone-beam Therapy (Siemens) (WIP) Existing QA Systems Matrixx Duke choice for routine QA Film Duke choice for commission SunNuclear Delta 4 Duke choice for future QA device Optical Scanner 2

3 Where Are We (Duke)? Started investigation in June 2008 A research RapidArc planning station from Varian Clinical installation in October Acceptance testing, commissioning, QA programs Single arc, no couch rotation, partial arc First patient treatment December 2008 New versions in August 2009 and May 2011 Allow multiple arcs, couch rotation, partial blocking, etc. Clinical Implementation of RapidArc Fundamentals for RapidArc Infrastructure/Installation Acceptance testing/commissioning Planning Delivery Quality assurance Infrastructure/Installation Staff (dedicated and trained) Existing machines: 21EX machine with 120-leaf millennium MLC NovalisTx with 120-leaf HD MLC (SRS, SRT,SBRT) ARIA version 8.6 or above (v10 now) Eclipse planning station (hardware and software) QA equipment Clinical Implementation of RapidArc Fundamentals for RapidArc Infrastructure/Installation Acceptance testing/commissioning Planning Delivery Quality assurance 3

4 Acceptance Testing Machine readiness Verification of installation against items included in the purchase order Inspections of safety and quality of installation and components VMAT performance Testing of functionality of each component and system performance against specifications. End-to-end testing Dry-runs for a few test case from simulation to delivery Acceptance Testing Sample Test 1.1: Gantry Angle Calibration Tolerance: Test 1.2: Isocenter Calibration Tolerance: + 1 mm Test 1.3: General Arc Dosimetry Range: 0.2 MU/ to 5.0 MU/ Tolerance: + 1% Acceptance Testing Sample Test 1.4 dmlc Dosimetry 0.5cm MLC slit sliding over 4 cm range Gantry: 0, 90, 270, 180 Acceptance Testing Sample Test 2.1: Accuracy of dmlc position vs. gantry position Tolerance: + 1 mm Tolerance: + 2% (over mean value) 4

5 Acceptance Testing Sample Acceptance Testing Sample Test 2.2: Accuracy of dmlc position during arc Tolerance: + 1 mm 2.1: Picket Fence vs. Gantry Angle (static) Acceptance Testing Sample Acceptance Testing Sample 2.1: Picket Fence vs. Gantry Angle (static) 5

6 Acceptance Testing Sample Acceptance Testing Sample Test 2.3: Ability to accurately detect MLC position error Criteria: detect submillimeter error in position 2.3: Picket Fence with Errors Acceptance Testing Sample Test 2.4: Accuracy of dose rate and gantry speed control during RapidArc Tolerance: + 2% Acceptance Testing Sample Test 2.5: Ability to control leaf speed/position during RapidArc Tolerance: + 2% 6

7 Relative Dose Percent Deviation (%) Acceptance Testing Sample Off y-axis : -100 mm Off y-axis : 0 mm Off y-axis : 100 mm Gantry speed vs Dose rate 1.02 (Tolerance 2%) Off x-axis Position (mm) Acceptance Testing Sample Off y-axis: -100 mm Off y-axis: 0 mm Off y-axis: 100 mm Variation of gantry speed and dose rate (tolerance 2%) Off X-axis Position (mm) Commissioning Validate that VMAT is capable of delivering radiation beams as good as SG-IMRT could Define the limitations of planning optimization, gantry rotation, beam blocking, couch rotation, and leaf speed, collimator settings Develop treatment process and documentation Workflow for RapidArc Treatment Case selection Immobilization/simulation Planning/prescription Quality assurance Target localization Treatment/validation 7

8 Clinical Implementation of RapidArc Planning - Process Fundamentals for RapidArc Infrastructure/Installation Acceptance testing/commissioning Planning Delivery Quality assurance SG-IMRT RapidArc Planning - Preparation Site-specific, for each site 10 cases from previous IMRT Develop RapidArc plan with different options Comparison between IMRT vs. RapidArc Constraint/optimization Optimization algorithm Constraints sensitivity Planning - Strategy Selected site-specific planning strategy 1 arc: prostate only, prostate bed, prostate+sv, brain lesions 2 arcs: prostate+sv+ln, spinal, brain lesions Multiple arcs: Head and neck, brain lesions, analrectal Partial arcs: partial breast, liver, 8

9 Planning - Optimization Key challenge interconnectivity of the beam shapes within consecutive VMAT gantry positions Constraints Target and normal structures Mechanicals Optimization algorithms More parameters Aperture based objectives and algorithms Planning Quality and Efficiency MLC segments Number of segments: IMRT RapidArc Plan quality is comparable if same segments More segments: slow gantry rotation for RapidArc More segments: long treatment time Planner experience Understand algorithm Understand limitations Planning - Logistics Training multiple people involved in the planning Gaining experience Develop planning protocols Last ~ 3 months before patients start Clinical Implementation of RapidArc Fundamentals for RapidArc Infrastructure/Installation Acceptance testing/commissioning Planning Delivery Quality assurance 9

10 Delivery - Tolerances Delivery Sample Parameters MLC leaf motion Leaf motion limit: 2.5 cm/s, 5 mm/degree Dose rate and gantry rotation speed One arc = one field MU, doserate, gantry speed linked Larger MU max doserate varying gantry speed Smaller MU max gantry speed varying doserate Middle varying gantry speed and doserate RapidArc Delivery Limits Variable gantry speed degree/sec Variable dose rate MU/min ( Novalis Tx) Variable dose per degree MU/degree Variable MLC speed cm/s Delivery - Efficiency Plan quality The quality of plan number of apertures Single arc 177 apertures Complexity structures: more apertures/arcs multiple arcs Delivery time Single arc less time (< 2 mins) but sometimes inferior quality Multiple arcs better quality but longer time 10

11 Delivery Treatment Time Treatment time = Patient set-up time + Delivery time Delivery time=beam-on time+between beam=on time Patient setup time: no reduction Beam-on time: 20-60% reduction (less MUs) Between beam-on time: 100% saving for single arc 20-60% saving for multiple arcs Delivery Partial Arc Mechanical collision If isocenter is close to center full arc If isocenter is close to peripheral partial arc Partial blocking is also available Partial RapidArc for Liver SBRT Multiple Arcs vs. Single Arc Single arc RapidArc MU = 714 IMRT Total MU = 1572 Multiple arcs Time: 1.3 min 11

12 VMAT For Large Size PTV VMAT For Large PTV Large field size -> sometimes IMRT beam has triple beam splits Multi-sections (parts) of the PTV, large variation of PTV shape, OARs and their constraints 3 Arcs, 1000 degree rotation 17cm Field Size << PTV in some directions Beam orientation selection is part of IMRT planning, is often not used for RapidArc (i.e. full arc) VMAT For Large Size PTV VMAT For Large PTV IMRT VMAT17cm IMRT VMAT17cm 12

13 D1% (%) VMAT For Large PTV IMRT VMAT17cm VMAT For Large PTV VMAT26CM VMAT 17cm VMAT For Large PTV Field Size Effect On VMAT Planning Quality VMAT17cm VMAT23cm 160 PTV Hot Spot (D1%) 150 Collimator 30 Collimator Projected Field Size X (cm) 13

14 VMAT For Head-and-Neck VMAT For Head-and-Neck VMAT IMRT Field Thru Shoulder Field Thru Shoulder VMAT For Head-and-Neck RapidArc For Head-and-Neck VMAT IMRT Rt Parotid Oral Cavity Cord Lt Parotid Pharynx 14

15 VMAT For Head-and-Neck Clinical Implementation of RapidArc Segments Thru Shoulder Fundamentals for RapidArc Infrastructure/Installation Acceptance testing/commissioning Planning Delivery Quality assurance Quality Assurance Quality Assurance The QA program for the VMAT is similar to SG-IMRT in principle but with different measurement approaches due to its dynamic nature that, during VMAT delivery, MLC leaves are moving Gantry is rotating Dose rate is changing The QA program: validate the functionality and performance of accepted features For each planned delivery Patient specific QA Machine specific QA 15

16 Quality Assurance Machine specific QA accuracy of the MLC leaf positions during VMAT delivery ability of the system to accurately vary the dose rate and gantry speed during VMAT delivery ability of the system to accurately vary the MLC leaf speed during VMAT delivery Tolerances: Acceptance baselines Machine QA Chart Daily Standard linac QA Standard MLC QA Rotational delivery of dose to an ion chamber phantom Monthly Leaf motion Gantry rotation Dose output leaf motion gantry motion dose rate changes Patient Specific QA Hybrid QA technique Plan to phantom Dose measurement to phantom Rotational nature Not to single plan Phantoms Instruments Ion chamber 2-D array (ion chamber, diodes, film, ) Patient Specific QA Data analysis Multiple planes (axial, coronal, sagittal) Profiles Points Gamma analysis Collision check Before patient on the couch When patient on the couch 16

17 RapidArc QA vs. IMRT QA More complex treatment delivery Varying gantry angle, gantry speed, dose rate, and MLC leaf motion ImSure does not calculate dose for RapidArc delivery Current IMRT technique: Portal dosimetry RapidArc technique: Ion chamber, film, and Matrixx QA Measurements Ion chamber Absolute dose: (meas calc) / calc < 3% Film (coronal, sagittal, and axial planes) Optional Matrixx (coronal and sagittal planes) 3-6 hrs/patient Gold Standard QA Tools Ion Chamber Film Ion Chamber Equipment 0.13cc or 0.01cc ion chamber Calibration 10x10cm 2, 100SSD, depth = d max, 200MU cgy/nc correction factor RA delivery Center of 30cm x 30cm x 20cm solid water phantom Compare to Eclipse calculation 39 VMAT plans 17

18 Film Equipment Kodak EDR2 film OmniPro ImRT Calibration 12 2cm x 2cm squares 0 300cGy RA delivery MultiCube (IBA dosimetry) Coronal, sagittal, and axial planes Compare to Eclipse calculation Gamma analysis 8 MAT plans Verification with Gold Standard QA Ion chamber: Median: +1.7% Range: -0.9% - 2.8% Film 24/24 > 93% passing rate 23/24 > 95% passing rate 20/24 > 97% passing rate Ion Chamber vs. Eclipse Result: Axial Film vs. Eclipse Ion chamber array vs. Eclipse Eclipse 3%, 3mm DTA, 5% threshold Film 18

19 Front Back % Pixels Passing Gamma (<1) % Pixels Passing Gamma (<1) 100 Film vs. Film vs. Eclipse 100 Film Film vs. Eclipse %, 3mm DTA, 5% threshold Plan # Coronal Sagittal Axial %, 3mm DTA, 5% threshold Sagittal and Coronal, 40% threshold Axial 3%, 3mm, axial thresholds up to 40% Plan # Coronal Sagittal Axial Validation of 2D Ion Chamber Array Angular Dependency: MatriXX Evolution (IBA Dosimetry) 1020 ionization chambers 0.07 cm 3 sensitive volume 0.4cm diameter 24 x 24 cm 2 grid 7.6mm spacing Automatic temperaturepressure correction Future solution from IBA: Gantry angle sensor Apply correction factor to each ion chamber based on angularity +2.2% 15x +3.0% 6x Measured/Calculated (%) +2.0% 15x +2.8% 6x -1.8% 15x -3.5% 6x Herzen et al., PMB 2007; 52: % 15x -1.7% 15x +3.7% 6x -2.0% 6x 19

20 %Difference from Eclipse Calculation %Pixels Passing Gamma (<1) %Pixels Passing (Gamma <= 1) Matrixx QA Results Patient Specific QA 100 Matrixx vs Eclipse MatrixxCoronal 3%, 3mm, axial thresholds up to 5% MatrixxSagittal RapidArc (Plan #) Ion Chamber Fig. vs. 1: Ion Eclipse Chamber vs. Eclipse & and ICA Matrixx vs. Eclipse Eclipse 6% IC IC Array 4% Film vs. Ion Film vs. Chamber Ion Array Array 2% 96 0% -2% % -6% Brain Prostate Bed Plan # Prostate Prostate + SV Spine 90 Coronal Sagittal Gamma = 3%, 3mm, 5% threshold Plan # 20

21 Effective vs. Efficient Stage 1: Intensive QA Ion chamber, film in 3 planes, ion chamber array in 2 planes, 3D polymer gel dosimetry Stage 2: Rigorous QA Ion chamber, ion chamber array in 2 planes Stage 3: Effective and Efficient QA Ion chamber, ion chamber array in 1 plane Ion chamber array only Effective vs. Efficient Preparation Delivery Analysis Ion chamber 1 st Film 1 st Ion chamber array 1 st Ion chamber additional Film additional Ion chamber array additional IC + 3 Film + 2 ICA IC + 2 ICA IC + ICA Effective vs. Efficient When implementing a new technology Perform intensive patient-specific QA for the first group of patients Rely on QA gold standards 3D QA very useful Use gold standard QA technology to transition to newer QA devices Goal: Effective QA Conclusion RapidArc is one format of rotational IMRT for dose painting Implementation of RapidArc requires careful planning, testing, and verifications. Thoroughly testing and commissioning are necessary prior to patient treatment QA is a critical step, always compare with static field IMRT plan in the early phase RapidArc should be judged by its accuracy, safety, efficiency, applicability, integration, and adaptation 21

22 Thank you for your attention 22

VMAT Treatment Planning

VMAT Treatment Planning Swedish Medical Center VMAT Treatment Planning David Shepard Swedish Cancer Institute Seattle, WA Founded in 1910 by Dr. Nils Johanson and a group of Seattle s leading Swedish-born businessmen. We treat

More information

8/17/2011. Implementation and Quality Assurance Considerations for Volumetric Modulated Arc Therapy (VMAT)*

8/17/2011. Implementation and Quality Assurance Considerations for Volumetric Modulated Arc Therapy (VMAT)* and Quality Assurance Considerations for Volumetric Modulated Arc Therapy * What is? Report of writing group from TETAWG under Therapy Physics Committee (TPC) of AAPM Writing Group members: James M. Galvin,

More information

Clinical Evaluation of real-time, patient-centric QA using the Delta 4 Discover

Clinical Evaluation of real-time, patient-centric QA using the Delta 4 Discover Clinical Evaluation of real-time, patient-centric QA using the Delta 4 Discover Karl H Rasmussen, Ph.D.. khrasmussen@uthscsa.edu Nema Bassiri bassirin@livemail.uthscsa.edu Division of Medical Physics Cancer

More information

Dosimetry of Small IMRT Fields

Dosimetry of Small IMRT Fields Dosimetry of Small IMRT Fields Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Symposium-October 10, 2013 1 IMRT and Small Fields IMRT beams typically contain small

More information

Evaluation of Dynamic Conformal Arc Therapy for Treatment of Lung and Liver

Evaluation of Dynamic Conformal Arc Therapy for Treatment of Lung and Liver Evaluation of Dynamic Conformal Arc Therapy for Treatment of Lung and Liver Sotiri Stathakis, PhD, DABR Disclosure ELEKTA Monaco research support I do not have a financial interest in any company or product

More information

Study of 2D ion chamber array for angular response and QA of dynamic MLC and pretreatment

Study of 2D ion chamber array for angular response and QA of dynamic MLC and pretreatment Anup Kumar Bhardwaj et al Study of 2D ion chamber array for angular response and QA of dynamic MLC... Study of 2D ion chamber array for angular response and QA of dynamic MLC and pretreatment IMRT plans

More information

Appreciation for discussions with Daniel Low, Kyle Antes, Stephen Kry, and the faculty of Duke University.

Appreciation for discussions with Daniel Low, Kyle Antes, Stephen Kry, and the faculty of Duke University. Jennifer O Daniel, Ph.D. Duke University Medical Center Spring Clinical AAPM Meeting, 2016 Appreciation for discussions with Daniel Low, Kyle Antes, Stephen Kry, and the faculty of Duke University. The

More information

7/31/2017. Implementing a Clinical Practice Guideline; Lessons From An Early Adopter of MPPG 5.a. Disclosures

7/31/2017. Implementing a Clinical Practice Guideline; Lessons From An Early Adopter of MPPG 5.a. Disclosures Implementing a Clinical Practice Guideline; Lessons From An Early Adopter of MPPG 5.a Dustin Jacqmin, PhD Assistant Professor, Human Oncology University of Wisconsin Madison None Disclosures 1 Outline

More information

ArcCHECK & 3DVH. The Ultimate 4D Patient QA Solution. Your Most Valuable QA and Dosimetry Tools

ArcCHECK & 3DVH. The Ultimate 4D Patient QA Solution. Your Most Valuable QA and Dosimetry Tools ArcCHECK & 3DVH The Ultimate 4D Patient QA Solution Your Most Valuable QA and Dosimetry Tools THE 4D QA GOLD STANDARD The ArcCHECK is the world s most selected independent 4D measurement array. Simply

More information

VMAT monthly QA using two techniques: 2D ion chamber array with an isocentric gantry mount and an in vivo dosimetric device attached to gantry

VMAT monthly QA using two techniques: 2D ion chamber array with an isocentric gantry mount and an in vivo dosimetric device attached to gantry Journal of Radiotherapy in Practice Journal of Radiotherapy in Practice (2013) Page 1 of 7 & Cambridge University Press 2013 doi:10.1017/s1460396912000556 Technical Note VMAT monthly QA using two techniques:

More information

Approved [insert date]*

Approved [insert date]* Page 1 of 33 1 The American Association of Physicists in Medicine (AAPM) is a nonprofit professional society whose primary purposes are to advance the science, education and professional practice of medical

More information

Going Clinical with the Accuray Radixact System. The Montefiore Experience

Going Clinical with the Accuray Radixact System. The Montefiore Experience Going Clinical with the Accuray Radixact System The Montefiore Experience Disclaimer The views expressed in the presentation are those of the presenters and do not necessarily reflect the views or policies

More information

Initial Certification

Initial Certification Initial Certification Therapeutic Medical Physics (TMP) Content Guide Part 2 Content Guide and Sample Questions PLEASE NOTE: List of Constants and Physical Values for Use on the Part 2 Physics Exams The

More information

An improvement in IMRT QA results and beam matching in linacs using statistical process control

An improvement in IMRT QA results and beam matching in linacs using statistical process control JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 5, 2014 An improvement in IMRT QA results and beam matching in linacs using statistical process control Justin D. Gagneur a and Gary A. Ezzell

More information

Patient DVH based QA metrics using the Planned Dose Perturbation Algorithm

Patient DVH based QA metrics using the Planned Dose Perturbation Algorithm Session V: New Therapy Technologies Patient DVH based QA metrics using the Planned Dose Perturbation Algorithm March 31, 2012 Hosang Jin, Ph.D. Assistant Professor University of Oklahoma Conflict of Interest

More information

To the Graduate Council:

To the Graduate Council: To the Graduate Council: I am submitting herewith a thesis written by Michelle Lynn Neeley entitled Patient- Specific Quality Assurance for IMRT. I have examined the final electronic copy of this thesis

More information

TG142 Quality Assurance of Medical Accelerators

TG142 Quality Assurance of Medical Accelerators TG142 Quality Assurance of Medical Accelerators Joe Hanley, Ph.D. Hackensack University Medical Center, Hackensack NJ TG142 MEMBERS Chair: Eric E. Klein, Ph.D., Washington University Joe Hanley, Ph.D.,

More information

Varian MCO Benefits in Head and Neck Planning

Varian MCO Benefits in Head and Neck Planning Varian MCO Benefits in Head and Neck Planning Dominic DiCostanzo, MS, DABR November 3, 2017 The Ohio State University Comprehensive Cancer Center Arthur G. James Cancer Hospital and Richard J. Solove Research

More information

Tra n s m is s ion dete c to r fo r a t-t re a t m e n t Q A

Tra n s m is s ion dete c to r fo r a t-t re a t m e n t Q A Tra n s m is s ion dete c to r fo r a t-t re a t m e n t Q A Confidence Safety Accuracy Efficiency The innovative Delta4 Discover transmission detector provides confidence and patient safety based on real

More information

Integrating the MR-Linac into Radiation Therapy Practice

Integrating the MR-Linac into Radiation Therapy Practice Integrating the MR-Linac into Radiation Therapy Practice Mikki Campbell MRT(T), MHE Winnie Li MRT(T), MSc UTDRO Evening Journal Club MR-Linac: From Prototype to Clinical January 24, 2019 Objectives To

More information

Experience and Suggestions on Commissioning and QA Procedures for Halcyon Version 1.2

Experience and Suggestions on Commissioning and QA Procedures for Halcyon Version 1.2 Experience and Suggestions on Commissioning and QA Procedures for Halcyon Version 1.2 Contributors: MD Anderson Cancer Center University of Pennsylvania Laurence Court Dimitris Mihailidis Peter Balter

More information

Benchmark & FAQ for NRG BR001: A Phase 1 Study of SBRT for the Treatment of Multiple Metastases

Benchmark & FAQ for NRG BR001: A Phase 1 Study of SBRT for the Treatment of Multiple Metastases Benchmark & FAQ for NRG BR001: A Phase 1 Study of SBRT for the Treatment of Multiple Metastases Credentialing Requirements (SECTION 5) Facility Questionnaire Benchmark planning 3DCRT credentialing IMRT

More information

Reducing VMAT Patient Specific Machine The Christie

Reducing VMAT Patient Specific Machine The Christie Reducing VMAT Patient Specific Machine Verification @ The Christie Phil Whitehurst Christie Medical Physics & Engineering Wilmslow Road Manchester M20 4BX Background Outline Christie IMRT/VMAT Techniques

More information

NEW. Pre-Treatment Plan Verification. Efficient, Intuitive, Connected. peace o f min d IBA DOSIMETRY // 1

NEW. Pre-Treatment Plan Verification. Efficient, Intuitive, Connected. peace o f min d IBA DOSIMETRY // 1 NEW Pre-Treatment Plan Verification dose safety Efficient, Intuitive, Connected IBA DOSIMETRY // 1 Efficient, Intuitive & Connected Plan Verification through the Central Database Network dose safety What

More information

This is probably the kind of radiotherapy that you are used to delivering in your country.

This is probably the kind of radiotherapy that you are used to delivering in your country. This is probably the kind of radiotherapy that you are used to delivering in your country. This slide shows images two patients, one with a T3 N2c nasopharyngeal CA and the other with a T1/2 N1/2 lung

More information

NEW. Pre-Treatment Plan Verification. Efficient, Intuitive, Connected. peace o f min d IBA DOSIMETRY // 1

NEW. Pre-Treatment Plan Verification. Efficient, Intuitive, Connected. peace o f min d IBA DOSIMETRY // 1 NEW Pre-Treatment Plan Verification dose safety Efficient, Intuitive, Connected IBA DOSIMETRY // 1 Efficient, Intuitive & Connected Plan Verification through the Central Database Network dose safety What

More information

Sensitivity of array detector measurements in determining shifts of MLC leaf positions

Sensitivity of array detector measurements in determining shifts of MLC leaf positions Received: 6 January 2017 Revised: 1 May 2017 Accepted: 1 June 2017 DOI: 10.1002/acm2.12148 RADIATION ONCOLOGY PHYSICS Sensitivity of array detector measurements in determining shifts of MLC leaf positions

More information

Elekta VMAT. Freedom to Treat. Enhanced dose conformance for patient-specific treatment

Elekta VMAT. Freedom to Treat. Enhanced dose conformance for patient-specific treatment Elekta VMAT Freedom to Treat Enhanced dose conformance for patient-specific treatment Speed or Accuracy Why choose? Clinical data has shown that around 50% of patients would benefit from IMRT treatment

More information

ViewRay Experience. Sasa Mutic, Ph.D. Washington University School of Medicine

ViewRay Experience. Sasa Mutic, Ph.D. Washington University School of Medicine ViewRay Experience Sasa Mutic, Ph.D. Washington University School of Medicine You can observe a lot, just by watching -Yogi Berra Does organ motion matter? A slide often used to illustrate RT capabilities

More information

Disclosures I have provided Physics Consultation Services to Varian and Vision RT

Disclosures I have provided Physics Consultation Services to Varian and Vision RT Surface Imaging to Inform Statistical Quality Control for SRS Treatments Michael J Tallhamer M.Sc. DABR Chief of Radiation Physics Centura Health MichaelTallhamer@Centura.org Quality Improvement and Safety

More information

Stereotactic Body Radiation Therapy: Planning and Delivery

Stereotactic Body Radiation Therapy: Planning and Delivery Stereotactic Body Radiation Therapy: Planning and Delivery Yong Yang, Ph.D. Department of Radiation Oncology Stanford University 6 AAPM Therapy Educational Course Stanford Radiation Physics Lei Xing, Ph.D.

More information

Comprehensive study of parameters for volumetric modulated arc therapy (VMAT) treatment planning

Comprehensive study of parameters for volumetric modulated arc therapy (VMAT) treatment planning Louisiana State University LSU Digital Commons LSU Master's Theses Graduate School 2010 Comprehensive study of parameters for volumetric modulated arc therapy (VMAT) treatment planning Catharine Elizabeth

More information

PerFRACTION 3D. 3D Pre-Treatment QA and In-Vivo Monitoring. Your Most Valuable QA and Dosimetry Tools. SunCHECK

PerFRACTION 3D. 3D Pre-Treatment QA and In-Vivo Monitoring. Your Most Valuable QA and Dosimetry Tools. SunCHECK PerFRACTION 3D 3D Pre-Treatment QA and In-Vivo Monitoring SunCHECK Your Most Valuable QA and Dosimetry Tools AUTOMATE YOUR QA PerFRACTION listens for and captures your pre-treatment and in-vivo QA files

More information

Commissioning of the Delta 4 Discover An investigation of its influence on the beam quality and of its error detection capabilities Vikren Sarkar,

Commissioning of the Delta 4 Discover An investigation of its influence on the beam quality and of its error detection capabilities Vikren Sarkar, Commissioning of the Delta 4 Discover An investigation of its influence on the beam quality and of its error detection capabilities Vikren Sarkar, Adam Paxton, Bill Salter Questions we aimed to answer

More information

The Role of In-Room kv X-Ray Imaging for Patient Setup and Target Localization (TG104)

The Role of In-Room kv X-Ray Imaging for Patient Setup and Target Localization (TG104) The Role of In-Room kv X-Ray Imaging for Patient Setup and Target Localization (TG104) John Wong (jwong35@jhmi.edu) David Jaffray (David.Jaffray@rmp.uhn.on.ca) Fang-Fang Yin (fangfang.yin@duke.edu) AAPM

More information

DESIGN OF AN ANTHROPOMORPHIC INTENSITY MODULATED RADIATION THERAPY QUALITY ASSURANCE PHANTOM

DESIGN OF AN ANTHROPOMORPHIC INTENSITY MODULATED RADIATION THERAPY QUALITY ASSURANCE PHANTOM DESIGN OF AN ANTHROPOMORPHIC INTENSITY MODULATED RADIATION THERAPY QUALITY ASSURANCE PHANTOM Dee-Ann Radford, David S. Followill, Peter A. Balter, William F. Hanson Department of Radiation Physics The

More information

RT Workspace from Standard Imaging

RT Workspace from Standard Imaging RT Workspace from Standard Imaging In 1895 Wilhelm Röntgen discovers the X-ray (receives Nobel prize in 1901) From 1900 1950 radium was chiefly used for radiation treatment (based on work of Marie Curie

More information

Three-dimensional dose prediction based on two-dimensional verification measurements for IMRT

Three-dimensional dose prediction based on two-dimensional verification measurements for IMRT JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 5, 2014 Three-dimensional dose prediction based on two-dimensional verification measurements for IMRT Iori Sumida, 1a Hajime Yamaguchi, 2

More information

Elekta VMAT. Choose freedom

Elekta VMAT. Choose freedom Elekta VMAT Choose freedom Freedom... Elekta VMAT is an advanced radiotherapy solution which has been developed to combine dramatically reduced treatment times with a removal of traditional restrictions

More information

MRI-Guided On-line Adaptive Radiotherapy The UCLA Physics Experience. Disclosures 8/3/2016

MRI-Guided On-line Adaptive Radiotherapy The UCLA Physics Experience. Disclosures 8/3/2016 MRI-Guided On-line Adaptive Radiotherapy The UCLA Physics Experience James Lamb, Nzhde Agazaryan, Minsong Cao, Mitch Kamrava, Percy Lee, Daniel Low, David Thomas, Yingli Yang Department of Radiation Oncology

More information

AAPM Medical Physics Practice Guideline 8.a.: Linear accelerator performance tests

AAPM Medical Physics Practice Guideline 8.a.: Linear accelerator performance tests Received: 30 January 2017 Revised: 28 February 2017 Accepted: 6 March 2017 DOI: 10.1002/acm2.12080 AAPM REPORTS & DOCUMENTS AAPM Medical Physics Practice Guideline 8.a.: Linear accelerator performance

More information

The impact of plan complexity on the accuracy of VMAT for the treatment of head and neck cancer

The impact of plan complexity on the accuracy of VMAT for the treatment of head and neck cancer The impact of plan complexity on the accuracy of VMAT for the treatment of head and neck cancer A thesis submitted in partial fulfilment of the requirements for the Degree of Master of Science in Medical

More information

Michael Gillin, PhD Professor, Chief of Clinical Services, Dept. of Radiation Physics

Michael Gillin, PhD Professor, Chief of Clinical Services, Dept. of Radiation Physics Michael Gillin, PhD Professor, Chief of Clinical Services, Dept. of Radiation Physics Michael Gillin - Grandfather My twin 12 month old grandsons, Arthur and Charles, live in Minneapolis. UT MDACC > 8,000

More information

Acceptance testing, commissioning and routine QA of the Varian TrueBeam Richard Popple

Acceptance testing, commissioning and routine QA of the Varian TrueBeam Richard Popple Acceptance testing, commissioning and routine QA of the Varian TrueBeam Richard Popple Disclosure UAB has Varian research contracts Honorarium from Varian Not your father s Chevy Not your father s Chevy

More information

RayStation Automation is the Future of Planning. Carmen Sawyers, MS, DABR Ackerman Cancer Center

RayStation Automation is the Future of Planning. Carmen Sawyers, MS, DABR Ackerman Cancer Center RayStation Automation is the Future of Planning Carmen Sawyers, MS, DABR Ackerman Cancer Center Overview Why Automation? o RayStation Automation Tools A Different Approach o RayStation Advanced Automation

More information

Multiple Metastasis Insert for the Lucy 3D QA Phantom

Multiple Metastasis Insert for the Lucy 3D QA Phantom Multiple Metastasis Insert for the Lucy 3D QA Phantom Tobin Hyman, MS DABR Department of Radiation Oncology, McLeod Health, Florence, SC 3120 Deming Way Middleton WI 53562-1461 USA 800-261-4446. ph 608-831-0025.

More information

Computational Verification in Interventional Radiation Oncology (Brachytherapy)

Computational Verification in Interventional Radiation Oncology (Brachytherapy) Computational Verification in Interventional Radiation Oncology (Brachytherapy) Prof. Dr. Dimos Baltas Abteilung Medizinische Physik & Engineering Strahlenklinik Brussels, 5- December- 2014 Starkenburgring

More information

Clinical Implementation of Auto-Planning

Clinical Implementation of Auto-Planning Clinical Implementation of Auto-Planning Ping Xia, Ph.D. Professor of Molecular Medicine Cleveland Clinic, Cleveland Disclosures Philips: Research Grant Siemens: Research Grant Why Auto-Planning? IMRT

More information

Treatment Quality Assurance Cone Beam Image Guided Radiation Therapy. Jean-Pierre Bissonnette, PhD, MCCPM

Treatment Quality Assurance Cone Beam Image Guided Radiation Therapy. Jean-Pierre Bissonnette, PhD, MCCPM Treatment Quality Assurance Cone Beam Image Guided Radiation Therapy Jean-Pierre Bissonnette, PhD, MCCPM Disclosure Work supported, in part, by Elekta Oncology Systems Commercial Interest in Penta-Guide

More information

Clinical trials with hypo-fractionation for breast conserving therapy (BCT): Shorter overall treatment time Effective radiation treatment

Clinical trials with hypo-fractionation for breast conserving therapy (BCT): Shorter overall treatment time Effective radiation treatment AAMD Regional Meeting, March 21st 2015 Clinical trials with hypo-fractionation for breast conserving therapy (BCT): Shorter overall treatment time Effective radiation treatment Whelan et al 2013 & Yu et

More information

Optimizing Treatment Planning Process in Clinical Enviroment

Optimizing Treatment Planning Process in Clinical Enviroment Optimizing Treatment Planning Process in Clinical Enviroment Wenzheng Feng 12 (1) Columbia University Medical Center, New York, NY (2) Bayhealth Medical Center, Dover, DE Objectives: 1. Be familiar with

More information

Ionisation chamber array, for setting up of a linac, and comparison to a 3D Water Tank. Presented by Neil McCann Systems Engineer for Elekta

Ionisation chamber array, for setting up of a linac, and comparison to a 3D Water Tank. Presented by Neil McCann Systems Engineer for Elekta Ionisation chamber array, for setting up of a linac, and comparison to a 3D Water Tank Presented by Neil McCann Systems Engineer for Elekta Why new dosimetry equipment? Replace our ageing IBA Buddelship

More information

Optimal Use of Emerging g Technologies

Optimal Use of Emerging g Technologies Optimal Use of Emerging g Technologies Radhe Mohan, PhD Professor and McNeil Endowed Chair Department of Radiation Physics 1 Optimal Use of Emerging Technologies A perspective Complexity and sophistication

More information

Thanks to Jeff Williamson, Mike Steinberg, James Purdy

Thanks to Jeff Williamson, Mike Steinberg, James Purdy Thanks to Jeff Williamson, Mike Steinberg, James Purdy Radiation Therapy 1.6M new cancer cases this year in US Approximately 60% of cancer patients receive radiation therapy during the course of their

More information

Controlli di qualità in SBRT

Controlli di qualità in SBRT Controlli di qualità in SBRT Carmelo Marino 24/25 OTTOBRE 2014 Università degli studi di Milano Each step in the integrated process of RT needs quality control and quality assurance (QA) to prevent errors

More information

On the sensitivity of TG-119 and IROC credentialing to TPS commissioning errors

On the sensitivity of TG-119 and IROC credentialing to TPS commissioning errors JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 1, 2016 On the sensitivity of TG-119 and IROC credentialing to TPS commissioning errors Drew McVicker, 1 Fang-Fang Yin, 2 Justus D. Adamson

More information

Proton Treatment Planning Systems Daniel Yeung

Proton Treatment Planning Systems Daniel Yeung New Developments in Proton Treatment Planning Systems Daniel Yeung Statement of Disclosure Funded Research & Development: Philips Medical Systems IBA Proton Planning Systems Commercial Systems Academic

More information

Real-Time Monitoring of Treatment Delivery. Sonja Dieterich, Ph.D., FAAPM Professor

Real-Time Monitoring of Treatment Delivery. Sonja Dieterich, Ph.D., FAAPM Professor Real-Time Monitoring of Treatment Delivery Sonja Dieterich, Ph.D., FAAPM Professor Disclosures UC Davis has a contract with Sun Nuclear for beta-testing PerFraction Scientific Advisor, MGS Research Patients

More information

The step-and-shoot IMRT overshooting phenomenon: a novel method to mitigate patient overdosage

The step-and-shoot IMRT overshooting phenomenon: a novel method to mitigate patient overdosage JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 4, 2016 The step-and-shoot IMRT overshooting phenomenon: a novel method to mitigate patient overdosage Heming Zhen, 1 Luo Ouyang, 2 Qinan

More information

Practical Workflow and the Cost of Adaptive Therapy. Rojano Kashani, Ph.D., DABR Washington University School of Medicine March 7, 2015

Practical Workflow and the Cost of Adaptive Therapy. Rojano Kashani, Ph.D., DABR Washington University School of Medicine March 7, 2015 Practical Workflow and the Cost of Adaptive Therapy Rojano Kashani, Ph.D., DABR Washington University School of Medicine March 7, 2015 Disclosures Travel expenses paid by ViewRay Inc. for on-site software

More information

Adapting VMAT plans optimized for an HD120 MLC for delivery with a Millennium MLC

Adapting VMAT plans optimized for an HD120 MLC for delivery with a Millennium MLC Received: 31 January 2017 Revised: 15 May 2017 Accepted: 5 June 2017 DOI: 10.1002/acm2.12134 RADIATION ONCOLOGY PHYSICS Adapting VMAT plans optimized for an HD120 MLC for delivery with a Millennium MLC

More information

MR Guided Radiation Therapy: ViewRay System Sasa Mutic, Ph.D.

MR Guided Radiation Therapy: ViewRay System Sasa Mutic, Ph.D. MR Guided Radiation Therapy: ViewRay System Sasa Mutic, Ph.D. Conflict of Interest Statement Shareholder - ViewRay Inc. Clinical advisory board - ViewRay Inc. Research and service - ViewRay Inc. Learning

More information

Adaptive Re-planning for Lung RT with Multi-targets

Adaptive Re-planning for Lung RT with Multi-targets Adaptive Re-planning for Lung RT with Multi-targets Feng Liu, An Tai, Ergun Ahunbay, Elizabeth Gore, Candice Johnstone, X. Allen Li Medical College of Wisconsin NCC AAMP, Madison, Oct. 11, 2013 Motivation:

More information

Data Mining to Aid Beam Angle Selection for IMRT

Data Mining to Aid Beam Angle Selection for IMRT Data Mining to Aid Beam Angle Selection for IMRT Stuart Price-University of Maryland Howard Zhang- University of Maryland School of Medicine Bruce Golden- University of Maryland Edward Wasil- American

More information

Nathan Childress, Ph.D., DABR

Nathan Childress, Ph.D., DABR Nathan Childress, Ph.D., DABR Financial Disclaimer I own Mobius Medical Systems, LP, manufacturer of Mobius3D, MobiusFX, and DoseLab I used Mobius3D to generate results for this talk Introduction Dosimetrists

More information

VOLUMETRIC MODULATED ARC THERAPY (VMAT): ADVANCED DELIVERY TECHNIQUES FOR STATIC AND MOVING TARGETS. A Dissertation. presented to

VOLUMETRIC MODULATED ARC THERAPY (VMAT): ADVANCED DELIVERY TECHNIQUES FOR STATIC AND MOVING TARGETS. A Dissertation. presented to VOLUMETRIC MODULATED ARC THERAPY (VMAT): ADVANCED DELIVERY TECHNIQUES FOR STATIC AND MOVING TARGETS A Dissertation presented to the Faculty of the Graduate School at the University of Missouri-Columbia

More information

Focusing Energy on Saving Lives. Year End Review Investor Relations Meeting ASTRO Oct. 30, 2007

Focusing Energy on Saving Lives. Year End Review Investor Relations Meeting ASTRO Oct. 30, 2007 Focusing Energy on Saving Lives Year End Review Investor Relations Meeting ASTRO Oct. 30, 2007 ASTRO IR Day & FY2007 Year-End Review Tim Guertin President and CEO ASTRO IR Day Oct. 30, 2007 Forward-Looking

More information

Shielding evaluation for IMRT implementation in an existing accelerator vault

Shielding evaluation for IMRT implementation in an existing accelerator vault JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 3, SUMMER 2003 Shielding evaluation for IMRT implementation in an existing accelerator vault R. A. Price,* O. Chibani, and C.-M. Ma Department

More information

AAPM IGRT Recommendations

AAPM IGRT Recommendations AAPM IGRT Recommendations Joint ICTP-IAEA International Workshop on the Implementation of Image Guided Radiotherapy (IGRT) 8-12 May 2017 J. Daniel Bourland, PhD, DABR Professor, Departments of Radiation

More information

03/08/2017. Advances and Challenges in Contour QA for Adaptive RT. Objective. Disclosures

03/08/2017. Advances and Challenges in Contour QA for Adaptive RT. Objective. Disclosures Advances and Challenges in Contour QA for Adaptive RT Kristy K Brock, PhD, DABR, FAAPM Professor, Department of Imaging Physics Director, Image Guided Cancer Therapy Program University of Texas MD Anderson

More information

Validation of a method for in vivo 3D dose reconstruction in SBRT using a new transmission detector

Validation of a method for in vivo 3D dose reconstruction in SBRT using a new transmission detector Received: 9 February 17 Revised: 9 February 17 Accepted: 27 March 17 DOI: 1.12/acm2.1213 RADIATION ONCOLOGY PHYSICS Validation of a method for in vivo 3D dose reconstruction in SBRT using a new transmission

More information

Maritimes an der Waterkant: Neues von Dolphin & COMPASS Hamburg, Mai 2018

Maritimes an der Waterkant: Neues von Dolphin & COMPASS Hamburg, Mai 2018 Maritimes an der Waterkant: Neues von Dolphin & COMPASS Hamburg, Mai 2018 Dr. Lutz Müller Lutz.Mueller@iba-group.com 2017 IBA SA a short History of Intensity Modulated Treatments 2000-2005 Studies 2005-2010

More information

Technical White Paper: An Educational Report On Helical IMRT EXECUTIVE SUMMARY. CRCPD Publication #E-11-3

Technical White Paper: An Educational Report On Helical IMRT EXECUTIVE SUMMARY. CRCPD Publication #E-11-3 CRCPD Publication #E-11-3 Technical White Paper: An Educational Report On Helical IMRT Developed by CRCPD s H-34 Task Force for Helical IMRT EXECUTIVE SUMMARY Helical IMRT (Intensity Modulated Radiation

More information

IMRT and Inverse Planning: From the Art to the State-of-the-Art

IMRT and Inverse Planning: From the Art to the State-of-the-Art IMRT and Inverse Planning: From the Art to the State-of-the-Art Thomas Bortfeld Massachusetts General Hospital Department of Radiation Oncology TBortfeld@mgh.harvard.edu Evolution of the art of radiation

More information

THE EFFECTIVE APPLICATION OF DYNAMIC ARC RADIOTHERAPY. A thesis submitted to The University of Manchester for the degree of. PhD

THE EFFECTIVE APPLICATION OF DYNAMIC ARC RADIOTHERAPY. A thesis submitted to The University of Manchester for the degree of. PhD THE EFFECTIVE APPLICATION OF DYNAMIC ARC RADIOTHERAPY A thesis submitted to The University of Manchester for the degree of PhD In the Faculty of Medical and Human Sciences 2013 CHRISTOPHER JAMES BOYLAN

More information

X. Allen Li. Why use outcome models? Biologically based treatment planning. Practical issues for biologically based treatment planning

X. Allen Li. Why use outcome models? Biologically based treatment planning. Practical issues for biologically based treatment planning Practical issues for biologically based treatment planning X. Allen Li Medical College of Wisconsin Limitations of dose-volume based treatment planning DV metrics are merely surrogate measures of radiation

More information

A comparison of two commercial treatment-planning systems for IMRT

A comparison of two commercial treatment-planning systems for IMRT JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 6, NUMBER 3, SUMMER 2005 A comparison of two commercial treatment-planning systems for IMRT M. Peter Petric, 1 Brenda G. Clark, 1 and James L. Robar

More information

Commissioning of the Heidelberg Ion Beam Therapy Centre

Commissioning of the Heidelberg Ion Beam Therapy Centre Commissioning of the Heidelberg Ion Beam Therapy Centre Jacksonville, May 2008 O. Jäkel 1,2, M. Ellerbrock 1, P. Heeg 1, B. Ackermann 1, M. Winter 1, K. Parodi 1, S. Brons 1, A. Mairani 1,2, T. Haberer

More information

Does gated beam delivery impact delivery accuracy on an Elekta linac?

Does gated beam delivery impact delivery accuracy on an Elekta linac? Received: 7 March 2016 Accepted: 4 October 2016 DOI: 10.1002/acm2.12020 RADIATION ONCOLOGY PHYSICS Does gated beam delivery impact delivery accuracy on an Elekta linac? Mohammed Jermoumi 1 Roger Xie 2

More information

RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident

RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident Resident s Name: RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident Rotation: PHYS 705: Clinical Rotation 3 Inclusive dates of rotation: Aug. 25, 2015 Feb. 25, 2016 Director or Associate

More information

RPC Prostate Phantom. Guidelines for Planning and Treating the RPC IMRT Prostate Phantom. Revised November 2010

RPC Prostate Phantom. Guidelines for Planning and Treating the RPC IMRT Prostate Phantom. Revised November 2010 RPC Prostate Phantom Guidelines for Planning and Treating the RPC IMRT Prostate Phantom. Revised November 2010 Credentialing for this protocol requires three steps: (1) submission of the Facility Questionaire

More information

An updates on ART for prostate, pancreas and breast

An updates on ART for prostate, pancreas and breast An updates on ART for prostate, pancreas and breast X. Allen Li Medical College of Wisconsin TH-E-137, AAPM Edu Course, Aug. 8, 2013 Acknowledgement Ergun Ahunbay, Ph.D Guangpei Chen, Ph.D Feng Liu, Ph.D

More information

Quality Assurance for particle beam therapy

Quality Assurance for particle beam therapy Quality Assurance for particle beam therapy PTCOG Educational Workshop, Essen 2013 Heidelberg Ion Beam Therapy Center at the University Hospital German Cancer Research Center, Heidelberg, Germany Outline

More information

Proton Therapy Workflow Martijn Engelsman HollandPTC and TU Delft

Proton Therapy Workflow Martijn Engelsman HollandPTC and TU Delft Proton Therapy Workflow Martijn Engelsman HollandPTC and TU Delft Disclaimer: I have no financial interest in any vendor of radiotherapy hardware, software, or otherwise PSI Winterschool, January 2012

More information

Clinical Background of Knowledgebased Models In IMRT/VMAT Planning

Clinical Background of Knowledgebased Models In IMRT/VMAT Planning Clinical Background of Knowledgebased Models In IMRT/VMAT Planning Jackie Wu, PhD, FAAPM Professor Duke University Medical Center Department of Radiation Oncology Disclosure Research Grant: NIH/NCI Master

More information

Clinical Experience with Philips Auto-Planning

Clinical Experience with Philips Auto-Planning Christian Rønn Hansen Odense University Hospital. Laboratory of Radiation Physics. Odense. Denmark. University of Southern Denmark. Institute of Clinical Research. Odense. Denmark. School of Physics -

More information

TG-104: In-room kv Computed Tomography for Image-guidance

TG-104: In-room kv Computed Tomography for Image-guidance TG-104: In-room kv Computed Tomography for Image-guidance D.A. Jaffray, Ph.D. Radiation Medicine Program Princess Margaret Hospital/Ontario Cancer Institute Professor Departments of Radiation Oncology

More information

8/2/2011. Elements of a Process Flow Trying use good planning to eliminate iterations. Thanks for contributions to this talk from

8/2/2011. Elements of a Process Flow Trying use good planning to eliminate iterations. Thanks for contributions to this talk from Head and Neck Treatment Planning Disclosure Grants from Varian Medical System Thanks for contributions to this talk from Robert Foote, M.D. From M.D. Anderson Yolanda Garces, M.D. Lei Dong, Ph.D. Shelley

More information

Abstract Background: Aim: Materials & Methods Results: & Discussion: Conclusions:

Abstract Background: Aim: Materials & Methods Results: & Discussion: Conclusions: Abstract Background: Radiation therapy is an important modality in the treatment of cancer. In the past decades radiation therapy has been developing towards more dynamic delivery techniques and Intensity-Modulated

More information

Clinical commissioning and operation of IBA Proteus ONE: the Royal Oak's experience

Clinical commissioning and operation of IBA Proteus ONE: the Royal Oak's experience Clinical commissioning and operation of IBA Proteus ONE: the Royal Oak's experience Xuanfeng (Leo) Ding, Ph.D., DABR Lead Medical Physicist, Proton Beam Therapy Adjunct Professor, Oakland University Sept.

More information

N.MAFFEI, G.GUIDI, C.VECCHI, G.BALDAZZI Physics Department, University of Bologna, via Irnerio Bologna, Italy

N.MAFFEI, G.GUIDI, C.VECCHI, G.BALDAZZI Physics Department, University of Bologna, via Irnerio Bologna, Italy AN ARTIFICIAL NEURAL NETWORK TO PREDICT TIME OF REPLANNING FOR TOMOTHERAPY TREATMENTS N.MAFFEI, G.GUIDI, C.VECCHI, G.BALDAZZI Physics Department, University of Bologna, via Irnerio 40 40138 Bologna, Italy

More information

Part 3 Oral Exam Content Guide

Part 3 Oral Exam Content Guide Initial Certification in Medical Physics Part 3 Oral Exam Content Guide The oral examination is designed to test your knowledge and fitness to practice applied medical physics in the specified specialty(ies).

More information

Focusing Energy on Saving Lives. Tim Guertin, CEO Elisha Finney, CFO JP Morgan Health Care Conference San Francisco January 7, 2008

Focusing Energy on Saving Lives. Tim Guertin, CEO Elisha Finney, CFO JP Morgan Health Care Conference San Francisco January 7, 2008 Focusing Energy on Saving Lives Tim Guertin, CEO Elisha Finney, CFO JP Morgan Health Care Conference San Francisco January 7, 2008 Forward-Looking Statements Except for historical information, the information

More information

SunCHECK * it. *2.0: INDEPENDENT QA. YOUR WAY.

SunCHECK * it. *2.0: INDEPENDENT QA. YOUR WAY. SunCHECK * it. *2.0: INDEPENDENT QA. YOUR WAY. SunCHECK it. // 2.0: INDEPENDENT QA. YOUR WAY. Integrated. Independent. Radiation Oncology departments are busy. In pursuit of Patient Safety, Medical Physicists

More information

biij Dosimetric evaluation of radiation dose rate effect in respiratory gated intensity modulated radiation therapy

biij Dosimetric evaluation of radiation dose rate effect in respiratory gated intensity modulated radiation therapy Available online at http://www.biij.org/2012/1/e5 doi: 10.2349/biij.8.1.e5 biij Biomedical Imaging and Intervention Journal ORIGINAL ARTICLE Dosimetric evaluation of radiation dose rate effect in respiratory

More information

Guidance document on Delivery, Treatment Planning, and Clinical Implementation of IMRT

Guidance document on Delivery, Treatment Planning, and Clinical Implementation of IMRT Guidance document on Delivery, Treatment Planning, and Clinical Implementation of IMRT IMRT subcommittee of the Radiation Therapy Committee Contents 1.0 Introduction 2.0 Delivery systems for IMRT 3.0 Treatment

More information

Validation of a modern second-check dosimetry system using a novel verification phantom

Validation of a modern second-check dosimetry system using a novel verification phantom Received: 7 April 2016 Accepted: 17 November 2016 DOI: 10.1002/acm2.12025 RADIATION ONCOLOGY PHYSICS Validation of a modern second-check dosimetry system using a novel verification phantom Daniel G. McDonald

More information

Effects of changing modulation and pitch parameters on tomotherapy delivery quality assurance plans

Effects of changing modulation and pitch parameters on tomotherapy delivery quality assurance plans JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 5, 2015 Effects of changing modulation and pitch parameters on tomotherapy delivery quality assurance plans Diana Binny, 1a Craig M. Lancaster,

More information

MV Cone Beam CT Imaging for daily localiza6on: (part II)

MV Cone Beam CT Imaging for daily localiza6on: (part II) MV Cone Beam CT Imaging for daily localiza6on: (part II) Jean Pouliot, Ph.D. Professor UCSF Helen Diller Family Comprehensive Cancer Center AAPM CE-Therapy Series Panel Session 7h30-9h25 am, July 28th,

More information

2018 Fall Meeting November 8, 2018 ~ 4:00 8:00

2018 Fall Meeting November 8, 2018 ~ 4:00 8:00 Danielle Lack Stephen Gardner Michael Huberts Choonik Lee Chadd Smith President President-Elect Treasurer Secretary Representative 2018 Fall Meeting November 8, 2018 ~ 4:00 8:00 University of Michigan

More information