Dosimetry of Small IMRT Fields

Size: px
Start display at page:

Download "Dosimetry of Small IMRT Fields"

Transcription

1 Dosimetry of Small IMRT Fields Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Symposium-October 10,

2 IMRT and Small Fields IMRT beams typically contain small fields or segments How small depends on the planner/planning system/institutional policy Minimum segment size used in planning should be equal or larger than that for which beam data was collected 2

3 Beam data for planning systems Without IMRT: 4x4 cm 2 40x40 cm 2 With IMRT: 2x2 cm 2 Or smaller 40x40 cm 2 3

4 Small field data challenges Detector size (affects profile and output factor measurements) Alignment of water tank, detector, etc. Jaw/MLC positioning errors Beam modeling in TPS IAEA and AAPM guidelines for absolute dosimetry of small fields TG-155: Small fields and non-equilibrium condition photon beam dosimetry 4

5 Focus on: Output Factors Output factors needed for treatment planning system commissioning Measurement field sizes/depths depend on the planning system used: Pinnacle: Range of field sizes (small->large), 10 cm depth, 100 cm SSD Eclipse: Range of field sizes (small->large), 10 cm depth, 100 cm SSD Xio: Range of field sizes (4x4-40x40), 10 cm depth Monaco: Range of field sizes (1x1-40x40), 10 cm depth, 100 cm SSD MasterPlan: Range of field sizes (5x5-20x20, smaller fields if needed), 10 cm depth, 100 cm SAD 5

6 TG 120 Recommendations Med. Phys. 38(3): (2011) 6

7 ASTRO Recommendations Practical Radiation Oncology (2011) 1,

8 8

9 9

10 Total scatter factor S cp versus field size measured with various available radiation detectors for (a) 6 MV and (b) 15 MV beams. 10 Das, Ding, and Ahnesjö: Small fields: Nonequilibrium radiation dosimetry Med. Phys. 35: (2008)

11 S. Dieterich and G. W. Sherouse: Comparison of seven commercial dosimetry diodes for SRS Med. Phys. 38: (2011) 11

12 Focus on: Output Factors and IMRT If the output factors measured for input into the treatment planning system are wrong What is the consequence for small field plans (IMRT, SBRT, SRS, )? 12

13 Small fields and IMRT A typical IMRT plan has 5-10 segments per beam There are at least ~2-3 large segments per beam So there may be ~ 0-7 small segments A fraction of these are < 4 cm 2 Assuming 4 cm 2 is the cut-off for potential erroneous measurements So the error introduced by inaccurate small field beam data should be minimal This also depends on the planning system and/or delivery system (linear accelerator) 13

14 Medical Physics, Vol. 39, No. 8, August 2012 pp Pinnacle treatment planning system Elekta and Varian accelerators Results may be different for other planning systems! 14

15 Method Two five-field IMRT plans created on TomoTherapy cheese phantom (one for each linac) Plans optimized to deliver 200 cgy to a cylindrical target of 2 cm diameter and 2 cm length Four structures used as organs-at-risk DMPO, minimum segment area: 2 cm 2, minimum segment MU: 1 Ion chamber location contoured 15

16 16

17 Output factor vs. field size 10 cm depth) Elekta Varian Relative Output Factors x2 3x3 4x4 5x5 6x6 8x8 10x10 Field Size (cm) 17

18 Method (Cont d) Small field output factors (2x2 and 3x3 cm 2 ) altered by ±5, 10, and 20% and machine re-commissioned 18

19 Altering small field output factors Relative Output Factors Measured -20% -10% -5% 5% 10% 20% x2 2x2 3x3 3x3 4x4 4x4 5x5 5x5 6x6 6x6 8x8 8x8 10x10 10x10 Field Size (cm) 19

20 Method (Cont d) Plan re-optimized every time Small change in segments between plans (0-2) All plans were delivered to the phantom with the ionization chamber positioned at isocenter In order to confirm the variations in output factors in the planning system, single static 2 2 cm 2 fields were created for each re-commissioned machine and delivered to the phantom 20

21

22

23 Summary of plans Elekta Varian Plan Name No. of segments Monitor Units 20% Decrease % Decrease 530 5% Decrease 515 Base % Increase % Increase % Increase 468 X: cm Y: cm 40% of segments 3x3 cm 2 or smaller Plan Name No. of segments Monitor Units 20% Decrease % Decrease 51 5% Decrease 50 Base 53 5% Increase 53 10% Increase 50 20% Increase 55 X: cm Y: 4.0 cm 55% of segments 3x3 cm 2 or smaller 23

24 Measured vs. Calculated dose-elekta Percentage Dose Difference (cgy) % decrease 10% decrease 5% decrease Base 5% increase 10% increase 20% increase Percent Output Factor Variation 24

25 Measured vs. Calculated dose-varian 5.00 Percentage Dose Difference (cgy) % decrease 10% decrease 5% decrease Base 5% increase 10% increase 20% increase Percent Output Factor Variation 25

26 Measured vs. Calculated dose-summary Elekta % Difference (Meas. vs. Calc.) Varian % Difference (Meas. vs. Calc.) Plan Name Plan Name 20% Decrease % Decrease 3.6 5% Decrease 0.8 Base 0 5% Increase % Increase % Increase % Decrease % Decrease % Decrease -0.1 Base 0 5% Increase % Increase % Increase % of segments 3x3 cm 2 or smaller 26

27 Why the difference? Elekta X-diaphragm traces MLCs, Y-jaws conform to each segment Varian X & Y jaws fixed for the delivery of all segments, encompassing the largest segment 27

28 Why the difference? Elekta Handled as a conformal jaw machine in Pinnacle MUs calculated by employing small field output factors as a consequence of conforming the collimator jaws to each of the segments Varian Handled as a non-conformal jaw machine in Pinnacle MUs calculated by employing output factors of the larger-encompassing jaw size, and modeling blockedfield effects Varian segment MUs not sensitive to small field OF changes 28

29 Verifying the results using static fields Single, 6 MV, 2 x 2 cm 2 field incident on phantom Same machines with modified small field OF used to compute the dose with the TPS Dose calculated and measured at 3 cm depth for all plans 29

30 Verifying the results using static fields Single jaw/mlc-defined 2x2 cm 2 field Elekta % Difference Plan Name Monitor Units Calculated Dose (Meas. vs. Calc.) 20% Decrease % Decrease % Decrease Base % Increase % Increase % Increase Single jaw-defined 2x2 cm 2 field Varian % Difference Plan Name Monitor Units Calculated Dose (Meas. vs. Calc.) 20% Decrease % Decrease % Decrease Base % Increase % Increase % Increase Correlation between OF change and dose/mu delivered

31 Verifying the results using static fields Single MLC-defined 2x2 cm 2 field (jaws at 10x10) Varian % Difference Plan Name Monitor Units Calculated Dose (Meas. vs. Calc.) 20% Decrease % Decrease 0.4 5% Decrease 0.0 Base 0.0 5% Increase % Increase % Increase 0.2 No correlation between OF change and dose/mu delivered 31

32 Results Incorrect output factors doaffect the dose delivered for a conformal jaw machine (Elekta, Siemens, Varian TrueBeam with jaw tracking) But they do not affect the dose delivered for a non-conformal jaw machine (Varian) As applicable to Pinnacle TPS! 32

33 Conclusion It is very important to have accurate output factor data in the TPS, as it is to have accurate beam profiles Inaccurate data could lead to miscalculation of dose If you don t have reliable data for small fields/segments, don t use them! 33

34 Questions? 34

7/31/2017. Implementing a Clinical Practice Guideline; Lessons From An Early Adopter of MPPG 5.a. Disclosures

7/31/2017. Implementing a Clinical Practice Guideline; Lessons From An Early Adopter of MPPG 5.a. Disclosures Implementing a Clinical Practice Guideline; Lessons From An Early Adopter of MPPG 5.a Dustin Jacqmin, PhD Assistant Professor, Human Oncology University of Wisconsin Madison None Disclosures 1 Outline

More information

RapidArc: Clinical Implementation

RapidArc: Clinical Implementation RapidArc: Clinical Implementation Fang-Fang Yin, PhD Q. Jackie Wu, PhD Acknowledgements Team efforts from staff at Duke Radiation Oncology, especially to Dr. J Chang, Dr. J O Daniel for providing slide

More information

} } 8/3/2016. Accounting for kv Imaging Dose. kilovoltage imaging devices/techniques. Current imaging dose determination methods

} } 8/3/2016. Accounting for kv Imaging Dose. kilovoltage imaging devices/techniques. Current imaging dose determination methods Accounting for kv Imaging Dose Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota 1 2016 AAPM Annual Meeting - SAM Therapy Educational Course, TU-B-201-2 Washington, DC, August

More information

Initial Certification

Initial Certification Initial Certification Therapeutic Medical Physics (TMP) Content Guide Part 2 Content Guide and Sample Questions PLEASE NOTE: List of Constants and Physical Values for Use on the Part 2 Physics Exams The

More information

VMAT Treatment Planning

VMAT Treatment Planning Swedish Medical Center VMAT Treatment Planning David Shepard Swedish Cancer Institute Seattle, WA Founded in 1910 by Dr. Nils Johanson and a group of Seattle s leading Swedish-born businessmen. We treat

More information

Uncertainties in treatment planning dose computation. Jeffrey V. Siebers

Uncertainties in treatment planning dose computation. Jeffrey V. Siebers Uncertainties in treatment planning dose computation Jeffrey V. Siebers Acknowledgements Dr David Followill, RPC for access to RPC data Drs. Catharine Clark, Pablo Carrasco, Warren Kilby, Guanghua Yan

More information

Approved [insert date]*

Approved [insert date]* Page 1 of 33 1 The American Association of Physicists in Medicine (AAPM) is a nonprofit professional society whose primary purposes are to advance the science, education and professional practice of medical

More information

Nathan Childress, Ph.D., DABR

Nathan Childress, Ph.D., DABR Nathan Childress, Ph.D., DABR Financial Disclaimer I own Mobius Medical Systems, LP, manufacturer of Mobius3D, MobiusFX, and DoseLab I used Mobius3D to generate results for this talk Introduction Dosimetrists

More information

Commissioning of the Delta 4 Discover An investigation of its influence on the beam quality and of its error detection capabilities Vikren Sarkar,

Commissioning of the Delta 4 Discover An investigation of its influence on the beam quality and of its error detection capabilities Vikren Sarkar, Commissioning of the Delta 4 Discover An investigation of its influence on the beam quality and of its error detection capabilities Vikren Sarkar, Adam Paxton, Bill Salter Questions we aimed to answer

More information

Characterization of interplay errors in step-and-shoot IMRT of the lung

Characterization of interplay errors in step-and-shoot IMRT of the lung Characterization of interplay errors in step-and-shoot IMRT of the lung Travis J. McCaw Under the supervision of Dr. Larry DeWerd Medical Radiation Research Center Department of Medical Physics University

More information

Clinical Evaluation of real-time, patient-centric QA using the Delta 4 Discover

Clinical Evaluation of real-time, patient-centric QA using the Delta 4 Discover Clinical Evaluation of real-time, patient-centric QA using the Delta 4 Discover Karl H Rasmussen, Ph.D.. khrasmussen@uthscsa.edu Nema Bassiri bassirin@livemail.uthscsa.edu Division of Medical Physics Cancer

More information

Michael Gillin, PhD Professor, Chief of Clinical Services, Dept. of Radiation Physics

Michael Gillin, PhD Professor, Chief of Clinical Services, Dept. of Radiation Physics Michael Gillin, PhD Professor, Chief of Clinical Services, Dept. of Radiation Physics Michael Gillin - Grandfather My twin 12 month old grandsons, Arthur and Charles, live in Minneapolis. UT MDACC > 8,000

More information

Clinically relevant investigation of flattening filter-free skin dose

Clinically relevant investigation of flattening filter-free skin dose JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 6, 2016 Clinically relevant investigation of flattening filter-free skin dose Christopher L. Guy, 1 Kishor Karki, 1 Manju Sharma, 1,2 and

More information

Patient DVH based QA metrics using the Planned Dose Perturbation Algorithm

Patient DVH based QA metrics using the Planned Dose Perturbation Algorithm Session V: New Therapy Technologies Patient DVH based QA metrics using the Planned Dose Perturbation Algorithm March 31, 2012 Hosang Jin, Ph.D. Assistant Professor University of Oklahoma Conflict of Interest

More information

TG142 Quality Assurance of Medical Accelerators

TG142 Quality Assurance of Medical Accelerators TG142 Quality Assurance of Medical Accelerators Joe Hanley, Ph.D. Hackensack University Medical Center, Hackensack NJ TG142 MEMBERS Chair: Eric E. Klein, Ph.D., Washington University Joe Hanley, Ph.D.,

More information

Recommendations of MPPG #5 and Practical Implementation Strategies

Recommendations of MPPG #5 and Practical Implementation Strategies Recommendations of MPPG #5 and Practical Implementation Strategies (MPPG #5.a.: Commissioning and QA of Treatment Planning Dose Calculations: Megavoltage Photon and Electron Beams) TG 244 Subcommittee

More information

The place of TomoDose in the tomotherapy QA schedule. Koen Tournel, Medical Physics Radiotherapy UZ Brussel

The place of TomoDose in the tomotherapy QA schedule. Koen Tournel, Medical Physics Radiotherapy UZ Brussel The place of TomoDose in the tomotherapy QA schedule Koen Tournel, Medical Physics Radiotherapy UZ Brussel Introduction Tomotherapy and QA Importance of beam geometry QA Importance of constancy and statistics

More information

7/30/2018. Script-Aided Automated TPS Testing. Outline. Background TPS Commissioning and QA

7/30/2018. Script-Aided Automated TPS Testing. Outline. Background TPS Commissioning and QA Script-Aided Automated TPS Testing Bin Cai, PhD Division of Medical physics 0 Outline Motivation of automation and standardization of TPS testing General concept and proposed approach Introduction of automated

More information

AAPM IGRT Recommendations

AAPM IGRT Recommendations AAPM IGRT Recommendations Joint ICTP-IAEA International Workshop on the Implementation of Image Guided Radiotherapy (IGRT) 8-12 May 2017 J. Daniel Bourland, PhD, DABR Professor, Departments of Radiation

More information

Stereotactic Body Radiation Therapy: Planning and Delivery

Stereotactic Body Radiation Therapy: Planning and Delivery Stereotactic Body Radiation Therapy: Planning and Delivery Yong Yang, Ph.D. Department of Radiation Oncology Stanford University 6 AAPM Therapy Educational Course Stanford Radiation Physics Lei Xing, Ph.D.

More information

Acceptance testing, commissioning and routine QA of the Varian TrueBeam Richard Popple

Acceptance testing, commissioning and routine QA of the Varian TrueBeam Richard Popple Acceptance testing, commissioning and routine QA of the Varian TrueBeam Richard Popple Disclosure UAB has Varian research contracts Honorarium from Varian Not your father s Chevy Not your father s Chevy

More information

RT Workspace from Standard Imaging

RT Workspace from Standard Imaging RT Workspace from Standard Imaging In 1895 Wilhelm Röntgen discovers the X-ray (receives Nobel prize in 1901) From 1900 1950 radium was chiefly used for radiation treatment (based on work of Marie Curie

More information

Shielding evaluation for IMRT implementation in an existing accelerator vault

Shielding evaluation for IMRT implementation in an existing accelerator vault JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 3, SUMMER 2003 Shielding evaluation for IMRT implementation in an existing accelerator vault R. A. Price,* O. Chibani, and C.-M. Ma Department

More information

AAPM Medical Physics Practice Guideline 8.a.: Linear accelerator performance tests

AAPM Medical Physics Practice Guideline 8.a.: Linear accelerator performance tests Received: 30 January 2017 Revised: 28 February 2017 Accepted: 6 March 2017 DOI: 10.1002/acm2.12080 AAPM REPORTS & DOCUMENTS AAPM Medical Physics Practice Guideline 8.a.: Linear accelerator performance

More information

Evaluation of Dynamic Conformal Arc Therapy for Treatment of Lung and Liver

Evaluation of Dynamic Conformal Arc Therapy for Treatment of Lung and Liver Evaluation of Dynamic Conformal Arc Therapy for Treatment of Lung and Liver Sotiri Stathakis, PhD, DABR Disclosure ELEKTA Monaco research support I do not have a financial interest in any company or product

More information

Going Clinical with the Accuray Radixact System. The Montefiore Experience

Going Clinical with the Accuray Radixact System. The Montefiore Experience Going Clinical with the Accuray Radixact System The Montefiore Experience Disclaimer The views expressed in the presentation are those of the presenters and do not necessarily reflect the views or policies

More information

ArcCHECK & 3DVH. The Ultimate 4D Patient QA Solution. Your Most Valuable QA and Dosimetry Tools

ArcCHECK & 3DVH. The Ultimate 4D Patient QA Solution. Your Most Valuable QA and Dosimetry Tools ArcCHECK & 3DVH The Ultimate 4D Patient QA Solution Your Most Valuable QA and Dosimetry Tools THE 4D QA GOLD STANDARD The ArcCHECK is the world s most selected independent 4D measurement array. Simply

More information

An improvement in IMRT QA results and beam matching in linacs using statistical process control

An improvement in IMRT QA results and beam matching in linacs using statistical process control JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 5, 2014 An improvement in IMRT QA results and beam matching in linacs using statistical process control Justin D. Gagneur a and Gary A. Ezzell

More information

Technical White Paper: An Educational Report On Helical IMRT EXECUTIVE SUMMARY. CRCPD Publication #E-11-3

Technical White Paper: An Educational Report On Helical IMRT EXECUTIVE SUMMARY. CRCPD Publication #E-11-3 CRCPD Publication #E-11-3 Technical White Paper: An Educational Report On Helical IMRT Developed by CRCPD s H-34 Task Force for Helical IMRT EXECUTIVE SUMMARY Helical IMRT (Intensity Modulated Radiation

More information

Complex Treatments in Proton Therapy - Selected Topics. Martin Bues, PhD Proton Therapy M.D. Anderson Cancer Center

Complex Treatments in Proton Therapy - Selected Topics. Martin Bues, PhD Proton Therapy M.D. Anderson Cancer Center Complex Treatments in Proton Therapy - Selected Topics Educational Workshop PTCOG 46, May 18-20, Wanjie, China Martin Bues, PhD Proton Therapy M.D. Anderson Cancer Center Outline Topic 1: Topic 2: Topic

More information

Proposal on Guideline for Quality Assurance of Radiation Treatment Planning System

Proposal on Guideline for Quality Assurance of Radiation Treatment Planning System Original Article PMP Progress in Medical Physics 28(4), December 2017 https://doi.org/10.14316/pmp.2017.28.4.197 pissn 2508-4445, eissn 2508-4453 Proposal on Guideline for Quality Assurance of Radiation

More information

Study of 2D ion chamber array for angular response and QA of dynamic MLC and pretreatment

Study of 2D ion chamber array for angular response and QA of dynamic MLC and pretreatment Anup Kumar Bhardwaj et al Study of 2D ion chamber array for angular response and QA of dynamic MLC... Study of 2D ion chamber array for angular response and QA of dynamic MLC and pretreatment IMRT plans

More information

VMAT monthly QA using two techniques: 2D ion chamber array with an isocentric gantry mount and an in vivo dosimetric device attached to gantry

VMAT monthly QA using two techniques: 2D ion chamber array with an isocentric gantry mount and an in vivo dosimetric device attached to gantry Journal of Radiotherapy in Practice Journal of Radiotherapy in Practice (2013) Page 1 of 7 & Cambridge University Press 2013 doi:10.1017/s1460396912000556 Technical Note VMAT monthly QA using two techniques:

More information

PerFRACTION 3D. 3D Pre-Treatment QA and In-Vivo Monitoring. Your Most Valuable QA and Dosimetry Tools. SunCHECK

PerFRACTION 3D. 3D Pre-Treatment QA and In-Vivo Monitoring. Your Most Valuable QA and Dosimetry Tools. SunCHECK PerFRACTION 3D 3D Pre-Treatment QA and In-Vivo Monitoring SunCHECK Your Most Valuable QA and Dosimetry Tools AUTOMATE YOUR QA PerFRACTION listens for and captures your pre-treatment and in-vivo QA files

More information

Eric E. Klein, Ph.D. Chair of TG-142

Eric E. Klein, Ph.D. Chair of TG-142 Eric E. Klein, Ph.D. Chair of TG-142 Professor of Radiation Oncology Washington University St. Louis, MO 04 October, 2010 NE AAPM Chapter American Association of Physicists in Medicine Establishes standards

More information

SunCHECK * it. *2.0: INDEPENDENT QA. YOUR WAY.

SunCHECK * it. *2.0: INDEPENDENT QA. YOUR WAY. SunCHECK * it. *2.0: INDEPENDENT QA. YOUR WAY. SunCHECK it. // 2.0: INDEPENDENT QA. YOUR WAY. Integrated. Independent. Radiation Oncology departments are busy. In pursuit of Patient Safety, Medical Physicists

More information

Validation of a modern second-check dosimetry system using a novel verification phantom

Validation of a modern second-check dosimetry system using a novel verification phantom Received: 7 April 2016 Accepted: 17 November 2016 DOI: 10.1002/acm2.12025 RADIATION ONCOLOGY PHYSICS Validation of a modern second-check dosimetry system using a novel verification phantom Daniel G. McDonald

More information

8/17/2011. Implementation and Quality Assurance Considerations for Volumetric Modulated Arc Therapy (VMAT)*

8/17/2011. Implementation and Quality Assurance Considerations for Volumetric Modulated Arc Therapy (VMAT)* and Quality Assurance Considerations for Volumetric Modulated Arc Therapy * What is? Report of writing group from TETAWG under Therapy Physics Committee (TPC) of AAPM Writing Group members: James M. Galvin,

More information

The step-and-shoot IMRT overshooting phenomenon: a novel method to mitigate patient overdosage

The step-and-shoot IMRT overshooting phenomenon: a novel method to mitigate patient overdosage JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 4, 2016 The step-and-shoot IMRT overshooting phenomenon: a novel method to mitigate patient overdosage Heming Zhen, 1 Luo Ouyang, 2 Qinan

More information

ViewRay Experience. Sasa Mutic, Ph.D. Washington University School of Medicine

ViewRay Experience. Sasa Mutic, Ph.D. Washington University School of Medicine ViewRay Experience Sasa Mutic, Ph.D. Washington University School of Medicine You can observe a lot, just by watching -Yogi Berra Does organ motion matter? A slide often used to illustrate RT capabilities

More information

biij Dosimetric evaluation of radiation dose rate effect in respiratory gated intensity modulated radiation therapy

biij Dosimetric evaluation of radiation dose rate effect in respiratory gated intensity modulated radiation therapy Available online at http://www.biij.org/2012/1/e5 doi: 10.2349/biij.8.1.e5 biij Biomedical Imaging and Intervention Journal ORIGINAL ARTICLE Dosimetric evaluation of radiation dose rate effect in respiratory

More information

Experience and Suggestions on Commissioning and QA Procedures for Halcyon Version 1.2

Experience and Suggestions on Commissioning and QA Procedures for Halcyon Version 1.2 Experience and Suggestions on Commissioning and QA Procedures for Halcyon Version 1.2 Contributors: MD Anderson Cancer Center University of Pennsylvania Laurence Court Dimitris Mihailidis Peter Balter

More information

Real-Time Monitoring of Treatment Delivery. Sonja Dieterich, Ph.D., FAAPM Professor

Real-Time Monitoring of Treatment Delivery. Sonja Dieterich, Ph.D., FAAPM Professor Real-Time Monitoring of Treatment Delivery Sonja Dieterich, Ph.D., FAAPM Professor Disclosures UC Davis has a contract with Sun Nuclear for beta-testing PerFraction Scientific Advisor, MGS Research Patients

More information

Follow this and additional works at:

Follow this and additional works at: The University of Toledo The University of Toledo Digital Repository Theses and Dissertations 2013 Dosimetric verification of respiratory-gated radiation therapy using a dynamic phantom for commissioning

More information

The Function of the Radiation Team - Physicist

The Function of the Radiation Team - Physicist The Function of the Radiation Team - Physicist R. Alfredo C. Siochi, PhD University of Iowa Hospitals and Clinics What has changed Functions are pretty much the same now as what is outlined in Asp et al,

More information

Implementation of a quality assurance program for computerized treatment planning systems

Implementation of a quality assurance program for computerized treatment planning systems Implementation of a quality assurance program for computerized treatment planning systems Priscilla R. T. L. Camargo and Laura N. Rodrigues Centro de Metrologia das Radiações, Instituto de Pesquisas Energéticas

More information

Controlli di qualità in SBRT

Controlli di qualità in SBRT Controlli di qualità in SBRT Carmelo Marino 24/25 OTTOBRE 2014 Università degli studi di Milano Each step in the integrated process of RT needs quality control and quality assurance (QA) to prevent errors

More information

@01-258_via99-011_hepatoma_onscreen

@01-258_via99-011_hepatoma_onscreen Lei Xing, Ph.D. & Jacob Haimson Professor Department of Radiation Oncology & MIPS Stanford University School of Medicine Case study TrueBeam 6 MV FFF RapidArc Tx of a Lung SBRT 3D modeling Treatment planning

More information

Experimental evaluation of the accuracy of skin dose calculation for a commercial treatment planning system

Experimental evaluation of the accuracy of skin dose calculation for a commercial treatment planning system JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 9, NUMBER 1, WINTER 2008 Experimental evaluation of the accuracy of skin dose calculation for a commercial treatment planning system Laurence E. Court,

More information

Elekta VMAT. Freedom to Treat. Enhanced dose conformance for patient-specific treatment

Elekta VMAT. Freedom to Treat. Enhanced dose conformance for patient-specific treatment Elekta VMAT Freedom to Treat Enhanced dose conformance for patient-specific treatment Speed or Accuracy Why choose? Clinical data has shown that around 50% of patients would benefit from IMRT treatment

More information

Design and dosimetry characteristics of a commercial applicator system for intra-operative electron beam therapy utilizing ELEKTA Precise accelerator

Design and dosimetry characteristics of a commercial applicator system for intra-operative electron beam therapy utilizing ELEKTA Precise accelerator JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 11, NUMBER 4, fall 2010 Design and dosimetry characteristics of a commercial applicator system for intra-operative electron beam therapy utilizing ELEKTA

More information

Quality Assurance for particle beam therapy

Quality Assurance for particle beam therapy Quality Assurance for particle beam therapy PTCOG Educational Workshop, Essen 2013 Heidelberg Ion Beam Therapy Center at the University Hospital German Cancer Research Center, Heidelberg, Germany Outline

More information

Comprehensive study of parameters for volumetric modulated arc therapy (VMAT) treatment planning

Comprehensive study of parameters for volumetric modulated arc therapy (VMAT) treatment planning Louisiana State University LSU Digital Commons LSU Master's Theses Graduate School 2010 Comprehensive study of parameters for volumetric modulated arc therapy (VMAT) treatment planning Catharine Elizabeth

More information

Three-dimensional dose prediction based on two-dimensional verification measurements for IMRT

Three-dimensional dose prediction based on two-dimensional verification measurements for IMRT JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 5, 2014 Three-dimensional dose prediction based on two-dimensional verification measurements for IMRT Iori Sumida, 1a Hajime Yamaguchi, 2

More information

Reducing VMAT Patient Specific Machine The Christie

Reducing VMAT Patient Specific Machine The Christie Reducing VMAT Patient Specific Machine Verification @ The Christie Phil Whitehurst Christie Medical Physics & Engineering Wilmslow Road Manchester M20 4BX Background Outline Christie IMRT/VMAT Techniques

More information

Tra n s m is s ion dete c to r fo r a t-t re a t m e n t Q A

Tra n s m is s ion dete c to r fo r a t-t re a t m e n t Q A Tra n s m is s ion dete c to r fo r a t-t re a t m e n t Q A Confidence Safety Accuracy Efficiency The innovative Delta4 Discover transmission detector provides confidence and patient safety based on real

More information

Enhanced dynamic wedge output factors for Varian 2300CD and the case for a reference database

Enhanced dynamic wedge output factors for Varian 2300CD and the case for a reference database JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 5, 2015 Enhanced dynamic wedge output factors for Varian 2300CD and the case for a reference database Christopher F. Njeh Department of Radiation

More information

Thanks to Jeff Williamson, Mike Steinberg, James Purdy

Thanks to Jeff Williamson, Mike Steinberg, James Purdy Thanks to Jeff Williamson, Mike Steinberg, James Purdy Radiation Therapy 1.6M new cancer cases this year in US Approximately 60% of cancer patients receive radiation therapy during the course of their

More information

Treatment Quality Assurance Cone Beam Image Guided Radiation Therapy. Jean-Pierre Bissonnette, PhD, MCCPM

Treatment Quality Assurance Cone Beam Image Guided Radiation Therapy. Jean-Pierre Bissonnette, PhD, MCCPM Treatment Quality Assurance Cone Beam Image Guided Radiation Therapy Jean-Pierre Bissonnette, PhD, MCCPM Disclosure Work supported, in part, by Elekta Oncology Systems Commercial Interest in Penta-Guide

More information

7/22/2014. The Necessity of Innovation in Medical Physics. Disclosures. AAPM - Medical Physicists. John Bayouth, Ph.D.

7/22/2014. The Necessity of Innovation in Medical Physics. Disclosures. AAPM - Medical Physicists. John Bayouth, Ph.D. The Necessity of Innovation in Medical Physics John Bayouth, Ph.D. President, AAPM Professor and Chief of Physics Department of Human Oncology University of Wisconsin Disclosures Principal employer: University

More information

SPECIFICATION STATEMENT OF WORK. Performance Specification for a radiotherapy linear accelerator

SPECIFICATION STATEMENT OF WORK. Performance Specification for a radiotherapy linear accelerator Specifications / Statement of Work / Terms of Reference SPECIFICATION STATEMENT OF WORK Performance Specification for a radiotherapy linear accelerator 1. Scope These specifications describe the requirements

More information

DESIGN OF AN ANTHROPOMORPHIC INTENSITY MODULATED RADIATION THERAPY QUALITY ASSURANCE PHANTOM

DESIGN OF AN ANTHROPOMORPHIC INTENSITY MODULATED RADIATION THERAPY QUALITY ASSURANCE PHANTOM DESIGN OF AN ANTHROPOMORPHIC INTENSITY MODULATED RADIATION THERAPY QUALITY ASSURANCE PHANTOM Dee-Ann Radford, David S. Followill, Peter A. Balter, William F. Hanson Department of Radiation Physics The

More information

To the Graduate Council:

To the Graduate Council: To the Graduate Council: I am submitting herewith a thesis written by Michelle Lynn Neeley entitled Patient- Specific Quality Assurance for IMRT. I have examined the final electronic copy of this thesis

More information

Computational Verification in Interventional Radiation Oncology (Brachytherapy)

Computational Verification in Interventional Radiation Oncology (Brachytherapy) Computational Verification in Interventional Radiation Oncology (Brachytherapy) Prof. Dr. Dimos Baltas Abteilung Medizinische Physik & Engineering Strahlenklinik Brussels, 5- December- 2014 Starkenburgring

More information

Disclosures. Outline 8/2/2017. Emerging Technologies for IORT: Unidirectional Planar Brachytherapy Sources. Julius V Turian PhD 8/2/2017

Disclosures. Outline 8/2/2017. Emerging Technologies for IORT: Unidirectional Planar Brachytherapy Sources. Julius V Turian PhD 8/2/2017 Emerging Technologies for IORT: Unidirectional Planar Brachytherapy Sources 8/2/2017 59 th AAPM Annual Meeting Denver CO Session Number: WE-B-702-04 Disclosures Rush University is participating in a pancreatic

More information

Evaluation of water equivalency of Plastic Water for high-energy electron beams using IAEA TRS-398 Code of Practice

Evaluation of water equivalency of Plastic Water for high-energy electron beams using IAEA TRS-398 Code of Practice Radiol Oncol 2004; 38(1): 55-60. Evaluation of water equivalency of Plastic Water for high-energy electron beams using IAEA TRS-398 Code of Practice Božidar Casar 1, Urban Zdešar 2 and Vlado Robar 1 1

More information

IMAGE GUIDANCE DOSES IN RADIOTHERAPY. Scott Crowe

IMAGE GUIDANCE DOSES IN RADIOTHERAPY. Scott Crowe IMAGE GUIDANCE DOSES IN RADIOTHERAPY Scott Crowe IMAGING DOSE ALARA requires that imaging dose is managed and optimised Imaging dose presents an increased risk for paediatric patients Increased effective

More information

This is probably the kind of radiotherapy that you are used to delivering in your country.

This is probably the kind of radiotherapy that you are used to delivering in your country. This is probably the kind of radiotherapy that you are used to delivering in your country. This slide shows images two patients, one with a T3 N2c nasopharyngeal CA and the other with a T1/2 N1/2 lung

More information

Quality assurance: Fundamental reproducibility tests for 3D treatment-planning systems

Quality assurance: Fundamental reproducibility tests for 3D treatment-planning systems JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 6, NUMBER 3, SUMMER 2005 Quality assurance: Fundamental reproducibility tests for 3D treatment-planning systems Charles M. Able and Michael D. Thomas

More information

Assessing dose variance from immobilization devices in VMAT head and neck treatment planning: A retrospective case study analysis

Assessing dose variance from immobilization devices in VMAT head and neck treatment planning: A retrospective case study analysis Assessing dose variance from immobilization devices in VMAT head and neck treatment planning: A retrospective case study analysis Alyssa Olson, B.S., R.T.(T), Kristine Phillips, B.S., R.T.(T), Tamara Eng

More information

IMRT and Inverse Planning: From the Art to the State-of-the-Art

IMRT and Inverse Planning: From the Art to the State-of-the-Art IMRT and Inverse Planning: From the Art to the State-of-the-Art Thomas Bortfeld Massachusetts General Hospital Department of Radiation Oncology TBortfeld@mgh.harvard.edu Evolution of the art of radiation

More information

On the sensitivity of TG-119 and IROC credentialing to TPS commissioning errors

On the sensitivity of TG-119 and IROC credentialing to TPS commissioning errors JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 1, 2016 On the sensitivity of TG-119 and IROC credentialing to TPS commissioning errors Drew McVicker, 1 Fang-Fang Yin, 2 Justus D. Adamson

More information

Does gated beam delivery impact delivery accuracy on an Elekta linac?

Does gated beam delivery impact delivery accuracy on an Elekta linac? Received: 7 March 2016 Accepted: 4 October 2016 DOI: 10.1002/acm2.12020 RADIATION ONCOLOGY PHYSICS Does gated beam delivery impact delivery accuracy on an Elekta linac? Mohammed Jermoumi 1 Roger Xie 2

More information

Appreciation for discussions with Daniel Low, Kyle Antes, Stephen Kry, and the faculty of Duke University.

Appreciation for discussions with Daniel Low, Kyle Antes, Stephen Kry, and the faculty of Duke University. Jennifer O Daniel, Ph.D. Duke University Medical Center Spring Clinical AAPM Meeting, 2016 Appreciation for discussions with Daniel Low, Kyle Antes, Stephen Kry, and the faculty of Duke University. The

More information

Disclosures I have provided Physics Consultation Services to Varian and Vision RT

Disclosures I have provided Physics Consultation Services to Varian and Vision RT Surface Imaging to Inform Statistical Quality Control for SRS Treatments Michael J Tallhamer M.Sc. DABR Chief of Radiation Physics Centura Health MichaelTallhamer@Centura.org Quality Improvement and Safety

More information

Ionisation chamber array, for setting up of a linac, and comparison to a 3D Water Tank. Presented by Neil McCann Systems Engineer for Elekta

Ionisation chamber array, for setting up of a linac, and comparison to a 3D Water Tank. Presented by Neil McCann Systems Engineer for Elekta Ionisation chamber array, for setting up of a linac, and comparison to a 3D Water Tank Presented by Neil McCann Systems Engineer for Elekta Why new dosimetry equipment? Replace our ageing IBA Buddelship

More information

TG-104: In-room kv Computed Tomography for Image-guidance

TG-104: In-room kv Computed Tomography for Image-guidance TG-104: In-room kv Computed Tomography for Image-guidance D.A. Jaffray, Ph.D. Radiation Medicine Program Princess Margaret Hospital/Ontario Cancer Institute Professor Departments of Radiation Oncology

More information

Impact of very long time output variation in the treatment of total marrow irradiation with helical tomotherapy

Impact of very long time output variation in the treatment of total marrow irradiation with helical tomotherapy Takahashi and Hui Radiation Oncology 2013, 8:123 RESEARCH Open Access Impact of very long time output variation in the treatment of total marrow irradiation with helical tomotherapy Yutaka Takahashi 1*

More information

Part 3 Oral Exam Content Guide

Part 3 Oral Exam Content Guide Initial Certification in Medical Physics Part 3 Oral Exam Content Guide The oral examination is designed to test your knowledge and fitness to practice applied medical physics in the specified specialty(ies).

More information

X. Allen Li. Why use outcome models? Biologically based treatment planning. Practical issues for biologically based treatment planning

X. Allen Li. Why use outcome models? Biologically based treatment planning. Practical issues for biologically based treatment planning Practical issues for biologically based treatment planning X. Allen Li Medical College of Wisconsin Limitations of dose-volume based treatment planning DV metrics are merely surrogate measures of radiation

More information

Clinical Implementation of the PEREGRINE Monte Carlo Dose Calculations System for Photon Beam Therapy

Clinical Implementation of the PEREGRINE Monte Carlo Dose Calculations System for Photon Beam Therapy UCRL-JC-133076 PREPRINT Clinical Implementation of the PEREGRINE Monte Carlo Dose Calculations System for Photon Beam Therapy C.L. Hartmann Siantar, P.M. Bergstrom T.P. Daly, M. Descalle D. Garrett, R.K.

More information

Validation of a method for in vivo 3D dose reconstruction in SBRT using a new transmission detector

Validation of a method for in vivo 3D dose reconstruction in SBRT using a new transmission detector Received: 9 February 17 Revised: 9 February 17 Accepted: 27 March 17 DOI: 1.12/acm2.1213 RADIATION ONCOLOGY PHYSICS Validation of a method for in vivo 3D dose reconstruction in SBRT using a new transmission

More information

Sensitivity of array detector measurements in determining shifts of MLC leaf positions

Sensitivity of array detector measurements in determining shifts of MLC leaf positions Received: 6 January 2017 Revised: 1 May 2017 Accepted: 1 June 2017 DOI: 10.1002/acm2.12148 RADIATION ONCOLOGY PHYSICS Sensitivity of array detector measurements in determining shifts of MLC leaf positions

More information

MR Guided Radiation Therapy: ViewRay System Sasa Mutic, Ph.D.

MR Guided Radiation Therapy: ViewRay System Sasa Mutic, Ph.D. MR Guided Radiation Therapy: ViewRay System Sasa Mutic, Ph.D. Conflict of Interest Statement Shareholder - ViewRay Inc. Clinical advisory board - ViewRay Inc. Research and service - ViewRay Inc. Learning

More information

important role in achieving optimal dose distribution for patients. Geometrical

important role in achieving optimal dose distribution for patients. Geometrical White Paper Agility * Intelligent Design Kevin Brown, Global Vice President Scientific Research Giulia Thompson, Senior Research Physicist Elekta, Ltd Abstract Multileaf collimators (MLCs) have a variety

More information

The application of error reduction QA philosophy in HDR brachytherapy

The application of error reduction QA philosophy in HDR brachytherapy The application of error reduction QA philosophy in HDR brachytherapy Bruce Thomadsen University of Wisconsin - Madison Learning Objectives Learning objectives: To understand 1. The problem with the current

More information

A comparison of two commercial treatment-planning systems for IMRT

A comparison of two commercial treatment-planning systems for IMRT JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 6, NUMBER 3, SUMMER 2005 A comparison of two commercial treatment-planning systems for IMRT M. Peter Petric, 1 Brenda G. Clark, 1 and James L. Robar

More information

The Roadmap Ahead for Big Data in? Radiation Oncology. Conflict of Interest Statement. Oncology Radiation Oncology Medical Physics

The Roadmap Ahead for Big Data in? Radiation Oncology. Conflict of Interest Statement. Oncology Radiation Oncology Medical Physics The Roadmap Ahead for Big Data in Radiation Oncology Sasa Mutic, PhD Professor Vice-chair of Medical Physics and Clinical Strategy Conflict of Interest Statement TreatSafely founder and partner Radialogica

More information

RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident

RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident Resident s Name: RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident Rotation: PHYS 705: Clinical Rotation 3 Inclusive dates of rotation: Aug. 25, 2015 Feb. 25, 2016 Director or Associate

More information

Acknowledgments. DEBATE IMRT Verification QA: TO MEASURE. Leah Schubert Jenia Vinogradskiy John Lucido AAPM TG218

Acknowledgments. DEBATE IMRT Verification QA: TO MEASURE. Leah Schubert Jenia Vinogradskiy John Lucido AAPM TG218 DEBATE IMRT Verification QA: TO MEASURE Moyed Miften, PhD University of Colorado Acknowledgments Leah Schubert Jenia Vinogradskiy John Lucido AAPM TG218 1 My Proposition is EASY to Argue for 3 Words: Measurement

More information

Product Solutions YOUR MOST VALUABLE QA & DOSIMETRY TOOLS

Product Solutions YOUR MOST VALUABLE QA & DOSIMETRY TOOLS Product Solutions YOUR MOST VALUABLE QA & DOSIMETRY TOOLS PRODUCT SOLUTIONS // Table of Contents Contents Better Outcomes Through Technology...5 Innovation That Improves Outcomes While Saving Time...6

More information

Proton Treatment Planning Systems Daniel Yeung

Proton Treatment Planning Systems Daniel Yeung New Developments in Proton Treatment Planning Systems Daniel Yeung Statement of Disclosure Funded Research & Development: Philips Medical Systems IBA Proton Planning Systems Commercial Systems Academic

More information

Comparison of measured and computed portal dose for IMRT treatment

Comparison of measured and computed portal dose for IMRT treatment JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 3, SUMMER 2006 Comparison of measured and computed portal dose for IMRT treatment Savino Cilla, 1 Pietro Viola, 1 Luigi Azario, 3 Luca Grimaldi,

More information

2018 Fall Meeting November 8, 2018 ~ 4:00 8:00

2018 Fall Meeting November 8, 2018 ~ 4:00 8:00 Danielle Lack Stephen Gardner Michael Huberts Choonik Lee Chadd Smith President President-Elect Treasurer Secretary Representative 2018 Fall Meeting November 8, 2018 ~ 4:00 8:00 University of Michigan

More information

RADIATION SAFETY ISSUES LINKED TO THE OMNIPRESENCE OF COMPUTERS

RADIATION SAFETY ISSUES LINKED TO THE OMNIPRESENCE OF COMPUTERS RADIATION SAFETY ISSUES LINKED TO THE OMNIPRESENCE OF COMPUTERS Jean-Claude ROSENWALD Institut Curie, Paris Consultant DOSIsoft Earlier use of computers in radiotherapy 1950 1960 : replacement of tedious

More information

RPC Prostate Phantom. Guidelines for Planning and Treating the RPC IMRT Prostate Phantom. Revised November 2010

RPC Prostate Phantom. Guidelines for Planning and Treating the RPC IMRT Prostate Phantom. Revised November 2010 RPC Prostate Phantom Guidelines for Planning and Treating the RPC IMRT Prostate Phantom. Revised November 2010 Credentialing for this protocol requires three steps: (1) submission of the Facility Questionaire

More information

Clinical commissioning and operation of IBA Proteus ONE: the Royal Oak's experience

Clinical commissioning and operation of IBA Proteus ONE: the Royal Oak's experience Clinical commissioning and operation of IBA Proteus ONE: the Royal Oak's experience Xuanfeng (Leo) Ding, Ph.D., DABR Lead Medical Physicist, Proton Beam Therapy Adjunct Professor, Oakland University Sept.

More information

Medical Physics and the Evolving Administrative Role

Medical Physics and the Evolving Administrative Role Medical Physics and the Evolving Administrative Role Michael Howard, PhD DABMP Director Sarah Cannon Cancer Center Chief of Medical Physics AAPM Spring Clinical Meeting 2015 Overview The role of the medical

More information

Effect of Change in Orientation of Enhanced Dynamic Wedges on Radiotherapy Treatment Dose

Effect of Change in Orientation of Enhanced Dynamic Wedges on Radiotherapy Treatment Dose International Journal of Advancements in Research & Technology, Volume 2, Issue 5, May-2013 496 Effect of Change in Orientation of Enhanced Dynamic Wedges on Radiotherapy Treatment Dose Saeed Ahmad Buzdar

More information