CONCURRENT 10: THE COMPLETE BLOOD COUNT AND BEYOND: QUALITY ISSUES AND REFERENCE METHODS

Size: px
Start display at page:

Download "CONCURRENT 10: THE COMPLETE BLOOD COUNT AND BEYOND: QUALITY ISSUES AND REFERENCE METHODS"

Transcription

1 CONCURRENT 10: THE COMPLETE BLOOD COUNT AND BEYOND: QUALITY ISSUES AND REFERENCE METHODS Saturday May 14 th :30 3:00 PM ( hrs) International Society for Laboratory Hematology (ISLH)

2 CONCURRENT 10: THE COMPLETE BLOOD COUNT AND BEYOND: QUALITY ISSUES AND REFERENCE METHODS CHAIRS: Dr. Piero Cappelletti. SIPMEL, Pordenone, Italy Dr. Albert Huisman, UMC Utrecht, Utrecht, Netherlands International Society for Laboratory Hematology (ISLH)

3 Verification and Quality Control of automated Hematology Analyzers Dr. Albert Huisman, UMC Utrecht, Utrecht, Netherlands International Society for Laboratory Hematology (ISLH)

4 Verification and Quality Control of automated Hematology Analyzers Dr. Albert Huisman, UMC Utrecht, Utrecht, Netherlands International Society for Laboratory Hematology (ISLH)

5 Disclosure information: Dr Albert Huisman: Nothing to disclose The UMC Utrecht, department of Clinical Chemistry and Hematology has received funding for contract research from: Abbott Diagnostics Beckman-Coulter Diagnostic Grifols Mechatronics

6 Verification and quality control of automated Hematology Analyzers Automated Cell Analysis (Complete Blood Count (CBC))» Including 5 part Leukocyte differential count, reticulocyte count and new parameters (Reticulocyte Hemoglobin content Reticulated Platelets / Immature Platelet Fraction, ) Validation of a new analyzer Verification of a new analyzer ISO CLSI, ICSH, others Samples Precision Accuracy and comparability Sensitivity and specificity Reference intervals Quality control Daily / internal External Quality Assesment (EQA/Proficiency testing)

7 Why? Why bother about verification and quality control of a highly automated analyzer?

8 Because: CBC results form the starting point of numerous diagnostic schedules, treatments and interventions, for example: Erytrocyte / Platelet transfusions Work up of anemia (iron deficiency/ thalassemia / ) Hematological malignancies General work-up in various disease states Etctetera

9 Because: The ISO standard [Medical laboratories Particular requirements for quality and competence] requires a verification process

10 Advantages of Automated Cell Analysers: Excellent analyticalperformance Closed-tube analysis No inter-observer variability No slide distribution error Eliminate statistical variations Potential of reflex testing Availability of extra parameters e.g. MCV, RDW, %rp, More efficient (> 100 analyses/hour) and cost effective than manual method.

11 Automated cell counters: Complete Blood Count (CBC): Hemoglobin (Hb) concentration, RBC count & RBC indices (MCV, MCH, MCHC), WBC count, PLT count WBC differential count (5 normal WBC s) RDW, MPV Reticulocyte count Nucleated RBC count (NRBC s). Flagging, etc etc New parameters: Advanced RBC parameters: % microcytic RBC s, Reticulocyte Hemoglobin content,.. Reticulated Platelets / Immature Platelets.

12 Further advances in automation Multiple in line analysers Built-in Slide-Maker-Stainer Auto-validation in LIS Automated reflex/morphology etc. Including digital morphology or flow-cytometry solutions All Full blood analysis in line (ESR, HbA1c,..) Total Laboratory Automation Digital Morphology

13 Validation of a new analyzer: goal Provision of objective evidence that a hematologyanalyzer fulfills specified requirements (where the specified requirements are adequate for intended use). Validation is primarilya manufacturersresponsibilityto ensure that design goals are met and performance claims are stated (including safety and effectiveness). Validation study (manufacturer): collect data to support a regulatory submission and the registration of a hematology analyzer.

14 Validation of a new analyzer: goal Provision of objective evidence that a hematologyanalyzer fulfills specified requirements (where the specified requirements are adequate for intended use). Validation is primarilya manufacturersresponsibilityto ensure that design goals are met and performance claims are stated (including safety and effectiveness). Validation study (manufacturer): collect data to support a regulatory submission and the registration of a hematology analyzer.

15 Validation of a new analyzer (manufacturer) Objectives of a validation study: Generate data to assess safety and clinical efficacy from a medical perspective Develop performance information for labelingand marketing purposes Validate appropriate operationalperformance characteristics of the hematologyanalyzer in a typical end-user setting Develop data that are used to support submissionof the product for approval or clearance by regulatory bodies Source: CLSI H26A2

16 Validation of a new analyzer (manufacturer) Validation performance specifications: Limit of blank (LoB, background) Carryover Imprecision (reproducibility), short term and long-term Analytical measuring interval (AMI) (linearity) Lower limit of detection (LLoD) and lower limit of quantitation (LLoQ) Comparibility (correlation) Interferences Frequency and type of data invalidations Source: CLSI H26A2

17 Validation of a new analyzer (manufacturer) Performance specifications: current reference methods*: Selective microscopy for WBC differential RBC count and WBC count (impedance) Selective microhematocrit (PCV) for hematocrit (HCT) Hemoglobin by hemiglobincyanidemethod Selective PLT monoclonalantibody (contemporary reference method) Reticulocytesbyflow cytometry * Standardized and independent of manufacturer Source: CLSI H26A2

18 Validation of a new analyzer (manufacturer) Performance specifications: current reference methods: When no reference method is available: usuallycomparison with previous generation instrument Source: CLSI H26A2

19 Validation of a new analyzer (manufacturer) Currentreference methods, problems, need for improvement: Current reference methods are more or less outdated e.g.: microscopic 400 cell WBC differential count: elaborative and imprecise Huisman et al. Clin Lab Med 2015

20 Validation of a new analyzer (manufacturer) Currentreference methods, problems, need for improvement: Current reference methods are more or less outdated e.g.: microscopic 400 cell WBC differential count: elaborative and imprecise No reference methods available for clinically relevant parameters*: MCV Extended RBC parameters Reticulocyte Hbcontent MPV Reticulated platelets/immature platelet fraction. * Clinically relevant parameters are not standardized! This may lead to differences between different Hematology analyzers: Differences in clinical interpretation (confusion for clinicians) May prevent wider use Invalid aggregation of data Huisman et al. Clin Lab Med 2015

21 Validation of a new analyzer (manufacturer) Currentreference methods, problems, need for improvement: Current reference methods are more or less outdated e.g.: microscopic 400 cell WBC differential count: elaborative and imprecise No reference methods available for clinically relevant parameters: MCV Extended RBC parameters Reticulocyte Hbcontent MPV Reticulated platelets/immature platelet fraction. Urgent need for improvement, role for professional societies Huisman et al. Clin Lab Med 2015

22 Verification

23 Instrument Verification by the end user laboratory The end user should asses whether the manufacturers claims on performance of the specific instrument also applyto the intended use criteria set by the laboratory: Verification Huisman et al. Clin Lab Med 2015

24 Instrument Verification by the end user laboratory The verification process includes performance analysis of: Accuracy Precision Reportable range of test results and reference intervals (normal ranges) Background (limit of blank) Carryover (sample) Lower limits of detection Quantitation Clinically reportable intervals (CRIs) Huisman et al. Clin Lab Med 2015

25 Instrument Verification by the end user laboratory According to ISO a certain level of verification of any new (hematology) analyzer has to be done (preferably) accordingto a professional standard. ISO also stimulates laboratories to take patient risk factors into consideration to meet these standards There are currenly 2 internationaldocuments available for verification of a hematologyanalyzer: Consensus documents with recommendations (not based on strong evidence but rather on expert opinion ). Huisman et al. Clin Lab Med 2015

26 Instrument Verification by the end user laboratory According to ISO a certain level of verification of any new (hematology) analyzer has to be done (preferably) accordingto a professional standard. ISO also stimulates laboratories to take patient risk factors into consideration to meet these standards There are currenly 2 internationaldocuments available for verification of a hematologyanalyzer: Clinical Laboratory Standards Institute (CLSI) (standard H26A2) published in 2010 International Council on Standardisation in Hematology (ICSH) guideline published in Huisman et al. Clin Lab Med 2015

27 Consensus documents with recommendations CLSI ICSH (published in the International Journal of Laboratory Hematology, the official ISLH journal

28 Instrument Verification by the end user laboratory Verification CLSI The laboratory (enduser) should follow the procedures of manufacturer validation, but the verification may be abbreviated; the goal is to verify that the manufacturer s stated performance is correct ICSH Verification is conformation of the evaluation performed by the manufacturer or published in the literature; the verification may be abbreviated (ie, focused to meet specific requirements at the test site) Precision/imprecision (within-run reproducibility, closeness of agreement between test results Should be performed with normal samples and samples at medical decision levels available in clinical laboratories. No specifications for number of measurements and reporting Single run of 10 measurements on the same sample (all reported parameters), 3 levels (normal, abnormal low, and abnormal high, around clinical decision points). Reported as SD and %CV

29 Instrument Verification by the end user laboratory Verification CLSI The laboratory (enduser) should follow the procedures of manufacturer validation, but the verification may be abbreviated; the goal is to verify that the manufacturer s stated performance is correct ICSH Verification is conformation of the evaluation performed by the manufacturer or published in the literature; the verification may be abbreviated (ie, focused to meet specific requirements at the test site) Precision/imprecision (within-run reproducibility, closeness of agreement between test results Should be performed with normal samples and samples at medical decision levels available in clinical laboratories. No specifications for number of measurements and reporting Single run of 10 measurements on the same sample (all reported parameters), 3 levels (normal, abnormal low, and abnormal high, around clinical decision points). Reported as SD and %CV

30 Instrument Verification by the end user laboratory Precision (between batch, longterm Comparability (comparison between evaluation HA and current HA) CLSI Should be performed with normal samples and samples at medical decision levels available in clinical laboratories. No specifications for time period and number of measurements Should be performed but not specified, if different modes are available, whether an extensive mode-to mode comparability should be performed ICSH Single sample, repeated daily, for a period of 20 30d. Three levels (all parameters, abnormal low and abnormal high, around clinical decision points). Fixed blood (control material) may be required At least samples, measured on both HA with samples with various disorders and with interfering substances

31 Instrument Verification by the end user laboratory Precision (between batch, longterm Comparability (comparison between evaluation HA and current HA) CLSI Should be performed with normal samples and samples at medical decision levels available in clinical laboratories. No specifications for time period and number of measurements Should be performed but not specified, if different modes are available, whether an extensive mode-to mode comparability should be performed ICSH Single sample, repeated daily, for a period of 20 30d. Three levels (all parameters, abnormal low and abnormal high, around clinical decision points). Fixed blood (control material) may be required At least samples, measured on both HA with samples with various disorders and with interfering substances

32 Instrument Verification by the end user laboratory Accuracy (closeness of agreement between measurement and true value) Reference intervals CLSI Should be performed; not otherwise specified Must be established or verified for all reportable parameters ICSH Depending on availability of reference method, often not applicable; in practice often compared with current HA Should be calculatedfor all components of the CBC; at least 120 samples from apparently healthy individuals (60 male, 60 female)

33 Instrument Verification by the end user laboratory Accuracy (closeness of agreement between measurement and true value) Reference intervals CLSI Should be performed; not otherwise specified Must be established or verified for all reportable parameters ICSH Depending on availability of reference method, often not applicable; in practice often compared with current HA Should be calculatedfor all components of the CBC; at least 120 samples from apparently healthy individuals (60 male, 60 female)

34 Instrument Verification by the end user laboratory Samples: Beware of pre-analytical variables! Time is one of the most important variables due to time dependent alterations of cells (volume and morphology) Usually anonymous surplus material Broad range of underlyingpathology(garantee that all cells are recognized ) Results should encompass the entire analyticalrange (very low (e.g. extreme trombocytopenia) to very high levels (e.g. extreme leucocytosis / CML). Age and gender (pediatricsamples?) All types of tubes that may enter the laboratory

35 Instrument Verification by the end user laboratory Huisman et al Clin Lab Med 2015

36 Instrument Verification by the end user laboratory: Precision Accuracy is the proximity of a measurement result to the true value and mainly dependent on systematic error (the term bias should be avoided); precision is the reproducibility of the measurements and mainly dependent on random error. Huisman et al IJKH 2016

37 Precision The desired precision of a CBC parameter is dependenton the biological variation of this parameter Work of George Cembrowski et al (Clin Lab Med 2015; IJLH 2016).

38 Cembrowski et al IJLH 2016

39 Precision: In order to generate (clinically) relevant results the analytical precision (%CV reproducibility) of a CBC result should be less than half and preferableless than ¼ of the biological variation. State of the art possibilities of current generation of HA

40 Instrument Verification by the end user laboratory

41 Instrument Verification by the end user laboratory De analytical %CV does add to the total %CV

42 Instrument Verification by the end user laboratory Room for improvement: Reticulocytes WBC WBC differential

43 Instrument Verification by the end user laboratory Carry-over, is it important? Background, is it important?

44 Instrument Verification by the end user laboratory Carry-over, is it important? Example: carry over 1%, Sample 1: platelet count 1000 x 10 9 /L Sample 2: platelet count 10 x 10exp 9 /L (subsequent sample) A carry-over of 1% will result in a falsely increased platelet count of 20 x 10 9 /L (100% increase)

45 Instrument Verification by the end user laboratory Background, is it important? A background (for example in the WBC channel) can lead to false positive results in CSF samples

46 Instrument Verification by the end user laboratory Reference: Tracy George et al.

47 Quality Control Internal Quality Control Performed on a daily basis Electronic built in quality control flags Are there mechanical / electronical problems? Control materials often provided by manufacturer Usually 3-control levels (low-/medium-/high- level) ~Wide range Often manipulated blood 1 versus multiple times/day depending on stability of analyser Comparison between different analysers possible Expensive 3.5 SD (Cembrowski et al)

48 Quality Control Internal Quality Control Statistical QC based on generated results Moving Average Built in software or LIS Large workload required Always Available Cheap

49 Quality Control External Quality Control / Proficiency testing Several times/year How do you perform in comparison with other labs and other types of equipment? ¹Calibrator

50 Acknowledgements: Jolande Vis Sue Ellen Verbrugge

51 Thanks!

52 Contact: Dr Albert Huisman University Medical Center Utrecht Department of Clinical Chemistry and Haematology G Heidelberglaan CX Utrecht Netherlands

53 Thank you for being with us!

54 See you at ISLH 2017 in Honolulu, Hawai ISLH 2017

55 See you at ISLH 2017 in Honolulu, Hawai ISLH 2017 May

Automated Hematology System. Your Complete Choice

Automated Hematology System. Your Complete Choice XE-5000 Automated Hematology System Your Complete Choice Advanced Technology Solutions to Meet Your Lab s Needs Even with these challenges, the need for hematology testing has remained steady or continued

More information

Automated Hematology Analyzer. Differentiate with the XT-4000i

Automated Hematology Analyzer. Differentiate with the XT-4000i XT-4000i Automated Hematology Analyzer Differentiate with the XT-4000i Advanced Technology Solutions to Meet Your Lab s Needs Even with these challenges, the need for hematology testing has remained steady

More information

Performance Evaluation of the Coulter LH 750 Hematology Analyzer

Performance Evaluation of the Coulter LH 750 Hematology Analyzer ISLH Laboratory Hematology 7:217 228 2001 Carden Jennings Publishing Co., Ltd. Official Publication Performance Evaluation of the Coulter LH 750 Hematology Analyzer TRACEY FERNANDEZ, LYNN BESSERT DOMACK,

More information

SF Cube. Exclusive Technology, Inclusive Approach. BC-6800 Auto Hematology Analyzer. Cell Analysis Technology

SF Cube. Exclusive Technology, Inclusive Approach. BC-6800 Auto Hematology Analyzer. Cell Analysis Technology Cell Analysis Technology Exclusive Technology, Inclusive Approach BC-6800 Auto Hematology Analyzer Cell Analysis Technology SF Cube SF Cube is a pathbreaking technology for reliable blood cell analysis,

More information

Greater Possibilities By Design

Greater Possibilities By Design Greater Possibilities By Design January 23, 2013 Nilam Patel MT(ASCP)SH Sr. Product Manager, Automation Solutions Objectives Review current and new portfolio Technology and benefits Operational benefits

More information

Ruby. A shining example of. Put science on your side. Intended Use:

Ruby. A shining example of. Put science on your side. Intended Use: High efficiency hematology A shining example of advanced technology Intended Use: The CELL-DYN Ruby is a multi-parameter automated Hematology analyzer designed for in vitro use in clinical laboratories.

More information

Automated Hematology Analyzer. Small Footprint. Big Difference.

Automated Hematology Analyzer. Small Footprint. Big Difference. XS-1000i Automated Hematology Analyzer Small Footprint. Big Difference. Small But Powerful: The XS-1000i Even with these challenges, the need for hematology testing has remained steady or continues to

More information

SAMPLE. Validation, Verification, and Quality Assurance of Automated Hematology Analyzers; Approved Standard Second Edition

SAMPLE. Validation, Verification, and Quality Assurance of Automated Hematology Analyzers; Approved Standard Second Edition June 2010 Validation, Verification, and Quality Assurance of Automated Hematology Analyzers; Approved Standard Second Edition This document provides guidance for the validation, verification, calibration,

More information

XN Series. What s in it for me? Krista Curcio, MT(ASCP), MBA Product Manager, Hematology Systems Sysmex America, Inc. All rights reserved.

XN Series. What s in it for me? Krista Curcio, MT(ASCP), MBA Product Manager, Hematology Systems Sysmex America, Inc. All rights reserved. XN Series What s in it for me? Krista Curcio, MT(ASCP), MBA Product Manager, Hematology Systems 2012 Sysmex America, Inc. All rights reserved. Agenda XN Series Product Overview A system to fit every lab

More information

Department of Pathology

Department of Pathology KASTURBA HEALTH SOCIETY'S MAHATMA GANDHI INSTITUTE OF MEDICAL SCIENCES, P.O. SEVAGRAM, WARDHA (DISTT.) MAHARASHTRA STATE - 442 102 Specifications for Equipment Department of Pathology 1) 3 Part Hematology

More information

BC-6800 Auto Hematology Analyzer. Small Cube, Big Difference

BC-6800 Auto Hematology Analyzer. Small Cube, Big Difference BC-6800 Auto Hematology Analyzer Small Cube, Big Difference Small Cube, Big Difference 2D Forward scatter EOS NEU + Side scatter MON LYM 3D Forward scatter MON NEU LYM EOS + Side scatter + Fluorescence

More information

Reshaping compact automation

Reshaping compact automation Automated Hematology Analyzer XN-SERIES Reshaping compact automation www.sysmex.com/us Optimize your laboratory As analyzers become increasingly sophisticated, the laboratory s contribution to the continuum

More information

Reshaping compact automation

Reshaping compact automation Automated Hematology Analyzer XN-SERIES Reshaping compact automation www.sysmex.com/us Optimize your laboratory As analyzers become increasingly sophisticated, the laboratory s contribution to the continuum

More information

It Should have five discrete analysis modes CBC, CBC+ DIFF, CBC + Retic, CBC+Retic+Diff & Retic only.

It Should have five discrete analysis modes CBC, CBC+ DIFF, CBC + Retic, CBC+Retic+Diff & Retic only. Specification of 5 Part Differential Haematology Analyser Specification as per the Tender Amended Specifications as per the Notification in Chapter 4 Pre Bid meeting held on 22-02-2018 in Chapter 4 It

More information

See Brilliant Results

See Brilliant Results See Brilliant Results without all the reviews. * In Development Product not available in the U.S. First Pass Efficiency. Getting it right the first time. GOALS: More reportable WBC and WBC differential

More information

BEYOND A BETTER BOX XN-V SERIES TM MULTISPECIES HEMATOLOGY ANALYZERS. Accelerate Your Multispecies Research Laboratory XN-V IS FOR ANIMAL USE ONLY

BEYOND A BETTER BOX XN-V SERIES TM MULTISPECIES HEMATOLOGY ANALYZERS. Accelerate Your Multispecies Research Laboratory XN-V IS FOR ANIMAL USE ONLY BEYOND A BETTER BOX XN-V SERIES TM MULTISPECIES HEMATOLOGY ANALYZERS Accelerate Your Multispecies Research Laboratory XN-V IS FOR ANIMAL USE ONLY MOVE YOUR MULTI-SPECIES LABORATORY BEYOND A BETTER BOX*

More information

XN-SERIES AUTOMATED HEMATOLOGY SYSTEMS

XN-SERIES AUTOMATED HEMATOLOGY SYSTEMS XN-SERIES AUTOMATED HEMATOLOGY SYSTEMS Reshaping Compact Automation BEYOND A BETTER BOX MOVE YOUR HEMATOLOGY LABORATORY BEYOND A BETTER BOX It s what makes Sysmex the smarter choice and sets our integrated

More information

BC Auto Hematology Analyzer. A CUTE 5-part

BC Auto Hematology Analyzer. A CUTE 5-part BC-5000 Auto Hematology Analyzer A CUTE 5-part Why do we need 5-part hematology analyzers? WBC differential: 3-part WBC differential: 5-part Lymphocyte Mid-size cell? Eosinophil Basophil Neutrophil DIFF

More information

UNRIVALED SIMPLICITY. Exceptional Ease of Use. The STAR brings new meaning to the word simple

UNRIVALED SIMPLICITY. Exceptional Ease of Use. The STAR brings new meaning to the word simple UNRIVALED SIMPLICITY The QBC STAR Centrifugal Hematology System is the simplest solution for in-office CBC testing. An innovative approach employing dry hematology reagents instead of the bulky liquid

More information

BC Auto Hematology Analyzer. A CUTE 5-part

BC Auto Hematology Analyzer. A CUTE 5-part BC-5150 Auto Hematology Analyzer A CUTE 5-part Why do we need 5-part hematology analyzers? WBC differential: 3-part WBC differential: 5-part Lymphocyte Mid-size cell? Eosinophil Basophil Neutrophil DIFF

More information

Verification of abnormal results from Coulter instrumentation

Verification of abnormal results from Coulter instrumentation Created Lisa Senzel Asst. Prof. - Clinical 5/9/2008 Path Revised Bruce Kube Day Heme Tech 1 10/24/2012 Reviewed Jay Bock M.D. 4/10/2012 Procedure: Approved Lisa Senzel Asst. Prof. - Clinical Path Replaces

More information

XN-9100 AUTOMATED HEMATOLOGY SYSTEMS

XN-9100 AUTOMATED HEMATOLOGY SYSTEMS XN-9100 AUTOMATED HEMATOLOGY SYSTEMS Reshaping Scalable Automation BEYOND A BETTER BOX MOVE YOUR HEMATOLOGY LABORATORY BEYOND A BETTER BOX It s what makes Sysmex the smarter choice and sets our integrated

More information

ICSH guidelines for the verification and performance of automated cell counters for body fluids

ICSH guidelines for the verification and performance of automated cell counters for body fluids International Journal of Laboratory Hematology ORIGINAL ARTICLE The Official journal of the International Society for Laboratory Hematology INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY ICSH guidelines

More information

The Simple Solution for CBC Testing

The Simple Solution for CBC Testing QBC STAR Centrifugal Hematology Analyzer The Simple Solution for CBC Testing Diagnostics Innovative Solutions for a Healthier World UNRIVALED SIMPLICITY The QBC STAR Centrifugal Hematology System is the

More information

BEYOND A BETTER BOX XN-SERIES TM AUTOMATED HEMATOLOGY SYSTEMS. Reshaping Compact Automation

BEYOND A BETTER BOX XN-SERIES TM AUTOMATED HEMATOLOGY SYSTEMS. Reshaping Compact Automation BEYOND A BETTER BOX XN-SERIES TM AUTOMATED HEMATOLOGY SYSTEMS Reshaping Compact Automation SMALL, SMART, COMPACT AUTOMATION It s an approach to hematology testing that s never been done before. An approach

More information

Evaluation of the Sysmex XN-550, a Novel Compact Haematology analyser from the XN-L series, compared to the XN-20 system.

Evaluation of the Sysmex XN-550, a Novel Compact Haematology analyser from the XN-L series, compared to the XN-20 system. Evaluation of the Sysmex XN-550, a Novel Compact Haematology analyser from the XN-L series, compared to the XN-20 system. H Tailor 1, I Mackie 2, A Mellick 3, S Machin 1, 2 1 Haematology Evaluations Unit,

More information

BEYOND A BETTER BOX XN-9000 AUTOMATED HEMATOLOGY SYSTEMS. Reshaping Scalable Automation

BEYOND A BETTER BOX XN-9000 AUTOMATED HEMATOLOGY SYSTEMS. Reshaping Scalable Automation BEYOND A BETTER BOX XN-9000 AUTOMATED HEMATOLOGY SYSTEMS Reshaping Scalable Automation 2 Sysmex XN-Series Automated Hematology Analyzers are designed to exceed the expectations of real people who work

More information

11/2/2015. Describe the technologies available on the XN-Series and the reason(s) for selection

11/2/2015. Describe the technologies available on the XN-Series and the reason(s) for selection XN-Series Technology What s Inside the Box? OBJECTIVES Describe the technologies available on the XN-Series and the reason(s) for selection Utilize example case studies to illustrate the new capabilities

More information

Immature Granulocyte Enumeration Our Journey from Manual to Automated Reporting

Immature Granulocyte Enumeration Our Journey from Manual to Automated Reporting Immature Granulocyte Enumeration Our Journey from Manual to Automated Reporting Mary Capper MT(ASCP)SH Supervisor, Hematology/Hemostasis University of Iowa Hospitals and Clinics Objectives Describe the

More information

Quality Control: Tips, Troubleshooting, and Tidbits

Quality Control: Tips, Troubleshooting, and Tidbits Quality Control: Tips, Troubleshooting, and Tidbits Nancy May, MBA, MT(ASCP) Technical Integration Specialist Agenda: 4 Tips 2 Troubleshootings Tools: Detector Parameters XbarM Tidbits 2 Sysmex America

More information

WADA Technical Document TD2019BAR. Blood Analytical Requirements for the Athlete Biological Passport

WADA Technical Document TD2019BAR. Blood Analytical Requirements for the Athlete Biological Passport 1. Introduction Blood Analytical Requirements for the Athlete Biological Passport This Technical Document (TD) has been established to harmonize the analysis of blood Samples collected, both In-Competition

More information

Features & Benefits Five-part differential on cat, dog, horse, cow, alpaca and llama; three-part differential on the nine other species

Features & Benefits Five-part differential on cat, dog, horse, cow, alpaca and llama; three-part differential on the nine other species Telephone: Facsimile: e-mail: + 27 (0)11 791 0025 + 27 (0)86 551 0832 johan@anjotech.co.za VetScan HM5 Hematology System Advanced Hematology Five-Part Differential The VetScan HM5 is a fully automated

More information

norma Hematology family norma Access norma norma Vision Reagents

norma Hematology family norma Access norma norma Vision Reagents norma Hematology family norma Access norma Reagents norma Vision Smallest reagent consumption in the world Smallest processed sample volume, ideal for pediatric use Microfluidic technology 3 A smart hematology

More information

Practical Guide to Running Controls

Practical Guide to Running Controls Practical Guide to Running Controls Internal quality control Introduction In addition to the competenceof the lab managerand the technical validation, the use and proper management of quality control (IQC,

More information

The Future of Hematological Diagnosis

The Future of Hematological Diagnosis Interview The Future of Hematological Diagnosis Bruce H. Davis, M.D. Bruce H. Davis, M.D. is the Director of Analytical Cytometry in the Department of Clinical Pathology at the William Beaumont Hospital

More information

Original Article. A Novel Automated Slide-Based Technology for Visualization, Counting, and Characterization of the Formed Elements of Blood

Original Article. A Novel Automated Slide-Based Technology for Visualization, Counting, and Characterization of the Formed Elements of Blood Original Article A Novel Automated Slide-Based Technology for Visualization, Counting, and Characterization of the Formed Elements of Blood A Proof of Concept Study James W. Winkelman, MD; Milenko J. Tanasijevic,

More information

Iron Deficiency Anemia

Iron Deficiency Anemia Diagnosis of Iron Deficiency Anemia Using the Abaxis VetScan HM2 Hematology System Contributing Authors: Mary Anna Thrall, BA, DVM, MS, DACVP; Barton Gillespie, BS, DVM Ross University School of Veterinary

More information

Upon completion of the Clinical Hematology rotation, the MLS student will be able to:

Upon completion of the Clinical Hematology rotation, the MLS student will be able to: Clinical Performance Objectives in Clinical Hematology Department of Medical and Research Technology University of Maryland School of Medicine Spring 2015 Upon completion of the Clinical Hematology rotation,

More information

Dry Hematology Analyzer Fast, Simple CBC Analysis Designed for the Point of Care

Dry Hematology Analyzer Fast, Simple CBC Analysis Designed for the Point of Care QBC STAR TM Dry Hematology Analyzer Fast, Simple CBC Analysis Designed for the Point of Care Reach for the STAR The QBC STAR hematology analyzer expands and advances the unique capabilities of previous

More information

Pushing limits in routine laboratory haematology with the XT-4000i

Pushing limits in routine laboratory haematology with the XT-4000i Pushing limits in routine laboratory haematology with the XT-4000i Sysmex Xtra Online December 2013 Fig. 1 The XT-4000i haematology system With the launch of the XE-5000, Sysmex set new standards in haematology.

More information

2015 Version MODULE 8. Laboratory Testing. SLMTA Trainer s Guide

2015 Version MODULE 8. Laboratory Testing. SLMTA Trainer s Guide 2015 Version MODULE 8 Laboratory Testing SLMTA Trainer s Guide MODULE 8. LABORATORY TESTING Performance Outcome Overview With satisfactory participation in the training and successful implementation of

More information

SCBM343- Complete Blood Count

SCBM343- Complete Blood Count SCBM343- Complete Blood Count Associate Professor Dr. Wannee Jiraungkoorskul Department of Pathobiology, Faculty of Science, Mahidol University Tel: 02-201-5563, E-mail: wannee.jir@mahidol.ac.th 1 Objectives

More information

CLSI C60: Assay Validation & Post-Validation Monitoring

CLSI C60: Assay Validation & Post-Validation Monitoring CLSI C60: Assay Validation & Post-Validation Monitoring Ross J. Molinaro, MT(ASCP), PhD, DABCC, FACB Medical Director Core Laboratory, Emory University Hospital Midtown Emory Clinical Translational Research

More information

Guide to Fulfillment of Validation and Verification of Examination Requirements

Guide to Fulfillment of Validation and Verification of Examination Requirements DIAGNOSTIC ACCREDITATION PROGRAM College of Physicians and Surgeons of British Columbia 300 669 Howe Street Telephone: 604-733-7758 ext. 2635 Vancouver BC V6C 0B4 Toll Free: 1-800-461-3008 (in BC) www.cpsbc.ca

More information

norma Hematology family norma norma Vision Reagents

norma Hematology family norma norma Vision Reagents norma Hematology family norma Reagents norma Vision Smallest reagent consumption in the world Smallest processed sample volume, ideal for pediatric use Microfluidic technology 3 The Icon 3 is a smart,

More information

ABX Pentra DX120. Hematology analyzer. 45 parameters Integrated slide-maker Expert validation system

ABX Pentra DX120. Hematology analyzer. 45 parameters Integrated slide-maker Expert validation system ABX Pentra DX120 Hematology analyzer 45 parameters Integrated slide-maker Expert validation system ABX Pentra DX 120 In the heart of hematopoiesis Complete cytology platform 120 samples per hour Continuous

More information

Diagnostic Medical Equipment New Revenue for Your Medical Practice

Diagnostic Medical Equipment New Revenue for Your Medical Practice Diagnostic Medical Equipment New Revenue for Your Medical Practice Consultant Libby Knollmeyer s Lab Machine Primer Series Part 1: Hematology Analyzers When laboratories in physician offices move out of

More information

9/29/2015. Describe the technologies available on the XN-Series and the reason(s) for selection

9/29/2015. Describe the technologies available on the XN-Series and the reason(s) for selection XN-Series Technology What s Inside the Box? OBJECTIVES Describe the technologies available on the XN-Series and the reason(s) for selection Utilize example case studies to illustrate the new capabilities

More information

Verification of Method XXXXXX

Verification of Method XXXXXX Reviewed By: Date: Technical Supervisor Reviewed By: Date: General Supervisor Reviewed By: Date: Quality Assurance Coordinator Approved By: Date: Director Page 1 of 11 Table of Contents Purpose and Principle:...

More information

DxH part differential closed tube hematology analyzer

DxH part differential closed tube hematology analyzer DxH 520 5-part differential closed tube hematology analyzer DxH 520 HEMATOLOGY ANALYZER Minimum footprint. Maximum productivity. The new closed tube 5-part differential analyzer because great things come

More information

Establishing Quality Control Target Values and Standard Deviations for Hematology Instrumentation

Establishing Quality Control Target Values and Standard Deviations for Hematology Instrumentation Establishing Quality Control Target Values and Standard Deviations for Hematology Instrumentation For informational purposes only. WHO REQUIRES IT AND WHAT THEY REQUIRE Regulatory agencies require the

More information

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK Haematology Department Contact: Ian Diton Conquest Hospital Tel: +44 (0) 1323 417400 ex 4127 The Ridge E-Mail: ian.diton@nhs.net St Leonards

More information

Guide to Fulfillment of Laboratory Results Comparability Requirements

Guide to Fulfillment of Laboratory Results Comparability Requirements DIAGNOSTIC ACCREDITATION PROGRAM College of Physicians and Surgeons of British Columbia 300 669 Howe Street Telephone: 604-733-7758 ext. 2635 Vancouver BC V6C 0B4 Toll Free: 1-800-461-3008 (in BC) www.cpsbc.ca

More information

Reduction of the Platelet Review Rate Using the Two-Dimensional Platelet Method

Reduction of the Platelet Review Rate Using the Two-Dimensional Platelet Method Coagulation and Transfusion Medicine / REDUCTION OF THE PLTELET REVIEW RTE Reduction of the Platelet Review Rate Using the Two-Dimensional Platelet Method Deanne H. Chapman, MT(SCP)SH, 1 Jonn Hardin, MT,

More information

Validation of Platelet Counting Accuracy With the Celltac F Automated Hematology Analyzer

Validation of Platelet Counting Accuracy With the Celltac F Automated Hematology Analyzer Journal of Automated Methods & Management in Chemistry, 5 (5), no. 4, 35 39 c 5 Hindawi Publishing Corporation Validation of Platelet ing Accuracy With the Celltac F Automated Hematology Analyzer Yutaka

More information

ICSH guidelines for the evaluation of blood cell analysers including those used for differential leucocyte and reticulocyte counting

ICSH guidelines for the evaluation of blood cell analysers including those used for differential leucocyte and reticulocyte counting International Journal of Laboratory Hematology ORIGINAL ARTICLE The Official journal of the International Society for Laboratory Hematology INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY ICSH guidelines

More information

HST-330 TM and Total Laboratory Automation

HST-330 TM and Total Laboratory Automation HST-330 TM and Total Laboratory Automation Hyun-Sook CHI, MD* * Department of Clinical Pathology, University of Ulsan College of Medicine and Asan Medical Center, Seoul, Korea. Total laboratory automation

More information

Scholars Journal of Applied Medical Sciences (SJAMS)

Scholars Journal of Applied Medical Sciences (SJAMS) Scholars Journal of Applied Medical Sciences (SJAMS) Abbreviated Key Title: Sch. J. App. Med. Sci. Scholars Academic and Scientific Publisher A Unit of Scholars Academic and Scientific Society, India www.saspublisher.com

More information

Ready, Set, Test! AACC Conference Mass Spectrometry in the Clinical Lab: Best Practice and Current Applications September 17-18, 2013 St.

Ready, Set, Test! AACC Conference Mass Spectrometry in the Clinical Lab: Best Practice and Current Applications September 17-18, 2013 St. Ready, Set, Test! Ross Molinaro, PhD, MLS(ASCP) CM, DABCC, FACB Medical Director, Clinical Laboratories Emory University Hospital Midtown Emory Clinical Translational Research Laboratory AACC Conference

More information

XN Hematology Case Studies Every Picture Tells a Story

XN Hematology Case Studies Every Picture Tells a Story XN Hematology Case Studies Every Picture Tells a Story Barbara J Connell MS, MT(ASCP)SH Senior Product Manager Technical Marketing & Communications Sysmex America, Inc Objectives Describe the advantages

More information

WAIT! Ready, Set, Test! Financial Disclosure. Research/Educational grants/consulting/salary support

WAIT! Ready, Set, Test! Financial Disclosure. Research/Educational grants/consulting/salary support Ready, Set, Test! Ross Molinaro, PhD, MLS(ASCP) CM, DABCC, FACB Medical Director, Clinical Laboratories Emory University Hospital Midtown Emory Clinical Translational Research Laboratory AACC Conference

More information

Validation of Platelet Counting Accuracy With the Celltac F Automated Hematology Analyzer

Validation of Platelet Counting Accuracy With the Celltac F Automated Hematology Analyzer Automated Methods & Management in Chemistry, 25 (25), no. 4, 235 239 c 25 Validation of Platelet Counting Accuracy With the Celltac F Automated Hematology Analyzer Yutaka Nagai, 1 Hiroshi Kondo, 2 and

More information

Total Analytic Error From Concept to Application

Total Analytic Error From Concept to Application Página 1 de 5 Clinical Laboratory News Subscribe CLN Daily CLN Stat CLN Articles Total Analytic Error From Concept to Application Author: James O. Westgard, PhD, and Sten A. Westgard, MS // Date: SEP.1.2013

More information

XN Series. Greater Possibilities by Design What s in it for me? Krista Curcio, MT(ASCP), MBA Product Manager, Hematology Systems

XN Series. Greater Possibilities by Design What s in it for me? Krista Curcio, MT(ASCP), MBA Product Manager, Hematology Systems XN Series Greater Possibilities by Design What s in it for me? Krista Curcio, MT(ASCP), MBA Product Manager, Hematology Systems 2012 Sysmex America, Inc. All rights reserved. Agenda XN Series Product Overview

More information

AUTOMATED HEMATOLOGY IN THE CLINICAL LAB

AUTOMATED HEMATOLOGY IN THE CLINICAL LAB AUTOMATED HEMATOLOGY IN THE CLINICAL LAB Diane Gaspari, SH(ASCP) Core Lab Manager York Hospital, York PA ggaspari@wellspan.org Objectives of Automation Increase efficiency; streamline workflow Increase

More information

Test Name Results Units Bio. Ref. Interval. Packed Cell Volume (PCV) %

Test Name Results Units Bio. Ref. Interval. Packed Cell Volume (PCV) % Lab No Mr DUMMY-----Z289 LLT12448 Age 15 Years Gender 16/4/2018 22600M 16/4/2018 23529M 27/6/2018 73049M Ref By Test Results Units Bio Ref Interval ANEMIA ANEL 1 COMLETE BLOOD COUNT (CBC) (Electrical Impedance

More information

COMPARISON OF PLATELET COUNTS BY SYSMEX XE 2100 AND LH 750 WITH THE INTERNATIONAL FLOW REFERENCE METHOD IN THROMBOCYTOPENIC PATIENTS

COMPARISON OF PLATELET COUNTS BY SYSMEX XE 2100 AND LH 750 WITH THE INTERNATIONAL FLOW REFERENCE METHOD IN THROMBOCYTOPENIC PATIENTS COMPARISON OF PLATELET COUNTS BY SYSMEX XE 2100 AND LH 750 WITH THE INTERNATIONAL FLOW REFERENCE METHOD IN THROMBOCYTOPENIC PATIENTS TINA DADU 1, KUNAL SEHGAL 2, ANJUM SHAIKH 2, SHANAZ KHODAIJI 2 1. BLK

More information

Business Case for Haematology Analyser Replacement Recommendation

Business Case for Haematology Analyser Replacement Recommendation Business Case for Haematology Analyser Replacement Recommendation Agenda Item 108/11 1. Executive Summary The Business Case [BC] identifying the need to replace the analytical platform in the Haematology

More information

LabGuide 13 How to Verify Performance Specifications

LabGuide 13 How to Verify Performance Specifications 1 LG 13 Comments? Feedback? Questions? Email us at info@cola.org or call us at 800 981-9883 LabGuide 13 How to Verify Performance Specifications The verification of performance specifications confirms

More information

Verification or validation, that is the question

Verification or validation, that is the question Perspective Page 1 of 7 Verification or validation, that is the question Giorgia Antonelli 1, Andrea Padoan 1, Ada Aita 1, Laura Sciacovelli 2, Mario Plebani 1,2 1 U.O.C. of Laboratory Medicine, Department

More information

Detection Capability of Faecal Haemoglobin Examinations

Detection Capability of Faecal Haemoglobin Examinations Detection Capability of Faecal Haemoglobin Examinations Callum G Fraser* and Sally C Benton** *Centre for Research into Cancer Prevention and Screening, University of Dundee, Scotland, and **Bowel Cancer

More information

Designing QC Activities to More Precisely Manage Analytical Accuracy and Patient Risk

Designing QC Activities to More Precisely Manage Analytical Accuracy and Patient Risk Designing QC Activities to More Precisely Manage Analytical Accuracy and Patient Risk Curtis A. Parvin, Ph.D. Manager of Advanced Statistical Research Bio-Rad Laboratories, QSD curtis_parvin@bio-rad.com

More information

Initial performance evaluation of the UniCel â DxH 800 Coulter â cellular analysis system

Initial performance evaluation of the UniCel â DxH 800 Coulter â cellular analysis system International Journal of Laboratory Hematology ORIGINAL ARTICLE The Official journal of the International Society for Laboratory Hematology INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY Initial performance

More information

The Journey Towards Standardization in Validation of Fluorescent Cell-Based Assays

The Journey Towards Standardization in Validation of Fluorescent Cell-Based Assays The Journey Towards Standardization in Validation of Fluorescent Cell-Based Assays Teri Oldaker, CLS, Ccy Genoptix Medical Laboratory Flow Cytometry Workshop ISCT 2015, Las Vegas, NV Objectives Background

More information

AAG PRINCIPLE SPECIMEN. REF (150 tests) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

AAG PRINCIPLE SPECIMEN. REF (150 tests) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE IMMAGE Immunochemistry Systems Chemistry Information Sheet Copyright 2010 Beckman Coulter, Inc. Alpha 1 -Acid Glycoprotein REF 447780 (150 tests) For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by:

More information

Prepublication Requirements

Prepublication Requirements Issued Prepublication Requirements The Joint Commission has approved the following revisions for prepublication. While revised requirements are published in the semiannual updates to the print manuals

More information

New Frontiers in the UK Hematology and Flow Cytometry Market: Business Challenges, Emerging Technologies, Competitive Landscape.

New Frontiers in the UK Hematology and Flow Cytometry Market: Business Challenges, Emerging Technologies, Competitive Landscape. New Frontiers in the UK Hematology and Flow Cytometry Market: Business Challenges, Emerging Technologies, Competitive Landscape Table of Contents I. Introduction II. Worldwide Market and Technology Overview

More information

Quintus 5-part hematology analyzer In-depth, quality-controlled 26-parameter results

Quintus 5-part hematology analyzer In-depth, quality-controlled 26-parameter results Outstanding user-interface Only three reagents needed 100-sample Autoloader option Quintus 5-part hematology analyzer In-depth, quality-controlled 26-parameter results Premium system solution boosted by

More information

MOLECULAR TESTING: VERIFYING/VALIDATING INSTRUMENTS, REAGENTS AND ASSAYS. Richard L. Hodinka, Ph.D.

MOLECULAR TESTING: VERIFYING/VALIDATING INSTRUMENTS, REAGENTS AND ASSAYS. Richard L. Hodinka, Ph.D. MOLECULAR TESTING: VERIFYING/VALIDATING INSTRUMENTS, REAGENTS AND ASSAYS Richard L. Hodinka, Ph.D. University of South Carolina School of Medicine Greenville Greenville Health System, Greenville, SC hodinka@greenvillemed.sc.edu

More information

Haematology. Reagents and analyzers SWISS HAEMATOLOGY S O L U T I O N

Haematology. Reagents and analyzers SWISS HAEMATOLOGY S O L U T I O N Haematology Catalogue Reagents and analyzers SWISS HAEMATOLOGY S O L U T I O N 24 parameters 5 DIFF analyzer with true optical differential measurement Optical laser method for 5-part WBC differential

More information

New Frontiers in the US Hematology and Flow Cytometry Market: Business Challenges, Emerging Technologies, Competitive Landscape.

New Frontiers in the US Hematology and Flow Cytometry Market: Business Challenges, Emerging Technologies, Competitive Landscape. New Frontiers in the US Hematology and Flow Cytometry Market: Business Challenges, Emerging Technologies, Competitive Landscape Table of Contents I. Introduction II. Worldwide Market and Technology Overview

More information

SAMPLE. Establishing and Verifying an Extended Measuring Interval Through Specimen Dilution and Spiking

SAMPLE. Establishing and Verifying an Extended Measuring Interval Through Specimen Dilution and Spiking 1st Edition EP34 Establishing and Verifying an Extended Measuring Interval Through Specimen Dilution and Spiking It is often medically necessary to provide results for specimens with concentrations above

More information

Starrsed RS. Gold Standard ESR

Starrsed RS. Gold Standard ESR Starrsed RS Gold Standard ESR Starrsed RS Does your busy laboratory perform up to 135 erythrocyte sedimentation rate (ESR) tests per hour and are you looking to automate this process? Or did you already

More information

Common Deficiencies

Common Deficiencies Question text Deficiency #12: Does the laboratory have a documented system to ensure consistency of morphologic observations among all personnel performing (U/A, Micro, Hem) (sample) Suggested methods

More information

Risk Management in IVD Producer Relation between manufacturer and user. S.M.Boutorabi DCLS, PhD

Risk Management in IVD Producer Relation between manufacturer and user. S.M.Boutorabi DCLS, PhD Risk Management in IVD Producer Relation between manufacturer and user S.M.Boutorabi DCLS, PhD IVD Manufacturer Requirements International Standard ISO 13485-2016 Medical Devices Quality Management Systems

More information

Auto-Validation Don t Rule It Out!

Auto-Validation Don t Rule It Out! Auto-Validation Don t Rule It Out! Elinore Craig Product Manager, IT Integration Agenda Barriers to auto-validation Success stories Overcoming the barriers 1 Barriers AACC Auto-validation Survey About

More information

True Walkaway Automated Urinalysis, Chemistry to Microscopy The CLINITEK AUWi System Answers for life.

True Walkaway Automated Urinalysis, Chemistry to Microscopy The CLINITEK AUWi System Answers for life. www.siemens.com/diagnostics True Walkaway Automated Urinalysis, Chemistry to Microscopy The CLINITEK AUWi System Answers for life. Can I achieve true unattended urinalysis testing in my laboratory? With

More information

DIAGNOSTIC ACCREDITATION PROGRAM College of Physicians and Surgeons of British Columbia

DIAGNOSTIC ACCREDITATION PROGRAM College of Physicians and Surgeons of British Columbia DIAGNOSTIC ACCREDITATION PROGRAM 300 669 Howe Street Telephone: 604-733-7758 Vancouver BC V6C 0B4 Toll Free: 1-800-461-3008 (in BC) www.cpsbc.ca Fax: 604-733-3503 Proficiency Testing Investigation: Sources

More information

Iron Status Assessment <DUE: 5/15/14>

Iron Status Assessment <DUE: 5/15/14> Iron Status Assessment Note: Clinical values are reported to one decimal place; please report your final calculations accordingly (HgB, Hct, MCV, MCH, MCHC). Must show calculations and units

More information

Feature. Hematology Analyzer: From Workhorse to Thoroughbred. Ellen Sullivan

Feature. Hematology Analyzer: From Workhorse to Thoroughbred. Ellen Sullivan Hematology Analyzer: From Workhorse to Thoroughbred Ellen Sullivan DOI: 10.1309/TMQ6T4CBCG408141 labmedicine.com May 2006 Volume 37 Number 5 LABMEDICINE 273 Hematology analyzers are the workhorses of the

More information

APB PRINCIPLE SPECIMEN. REF (300 tests) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

APB PRINCIPLE SPECIMEN. REF (300 tests) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE IMMAGE Immunochemistry Systems Chemistry Information Sheet Copyright 2010 Beckman Coulter, Inc. Apolipoprotein B REF 447730 (300 tests) For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Date Reviewed

More information

Harmonization of red blood cell distribution width (RDW): an attainable target?

Harmonization of red blood cell distribution width (RDW): an attainable target? Original Article Page 1 of 5 Harmonization of red blood cell distribution width (RDW): an attainable target? Giuseppe Lippi 1, Silvia Pipitone 2, Emmanuel J. Favaloro 3,4 1 Section of Clinical Biochemistry,

More information

A Division of Woodley Equipment Company Ltd

A Division of Woodley Equipment Company Ltd LEADING EDGE TECHNOLOGY FOR POINT OF CARE HAEMATOLOGY A Division of Woodley Equipment Company Ltd Contents The QBC STAR POC Haematology System...................4 Operating Ranges.......................................7

More information

2008 HFC-A PARTICIPANT SUMMARY

2008 HFC-A PARTICIPANT SUMMARY 2008 HFC-A PARTICIPANT SUMMARY Evaluation Criteria Analyte Target Value Evaluation Criteria Red Blood Cell Peer Group ± 3 SD White Blood Cell Peer Group ± 3 SD Cytopreparation Differential Educational

More information

MANAGEMENT BRIEF. How Automation is Revolutionizing White Cell Differential Analysis.

MANAGEMENT BRIEF. How Automation is Revolutionizing White Cell Differential Analysis. MANAGEMENT BRIEF How Automation is Revolutionizing White Cell Differential Analysis Introduction Confidence building becomes a real issue when a laboratory technologist generalist must classify abnormal

More information

RUO Addendum. UniCel DxH Series with System Manager Software. Coulter Cellular Analysis System. B26673AC July 2015

RUO Addendum. UniCel DxH Series with System Manager Software. Coulter Cellular Analysis System. B26673AC July 2015 RUO Addendum UniCel DxH Series with System Manager Software Coulter Cellular Analysis System Slidemaker Stainer July 2015 Manufactured by Beckman Coulter, Inc. 250 S. Kraemer Blvd. Brea, CA 92821 U.S.A.

More information

S1725 Method Validation and Verification: Case Studies and Laboratory Challenges

S1725 Method Validation and Verification: Case Studies and Laboratory Challenges S1725 Method Validation and Verification: Case Studies and Laboratory Challenges James H. Nichols, PhD, D(ABCC) Lauren N. Person, DO, MPH, FCAP Objectives Identify the difference between method validation

More information

Evaluation of the Automated Immature Granulocyte Count (IG) on Sysmex XE-2100 Automated Haematology Analyser vs. Visual Microscopy (NCCLS H20-A)

Evaluation of the Automated Immature Granulocyte Count (IG) on Sysmex XE-2100 Automated Haematology Analyser vs. Visual Microscopy (NCCLS H20-A) Evaluation of the Automated Immature Granulocyte Count (IG) on Sysmex XE-21 Automated Haematology Analyser vs. Visual Microscopy (NCCLS H2-A) Th WEILAND *1, H KALKMAN *2, and H HEIHN *1 *1 Central Laboratory,

More information

PRINCIPLE SPECIMEN. REF (300 tests) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

PRINCIPLE SPECIMEN. REF (300 tests) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE IMMAGE Immunochemistry Systems Chemistry Information Sheet Copyright 2010 Beckman Coulter, Inc. Complement REF 446450 (300 tests) For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Date Reviewed by:

More information

Prepared by Date Adopted Supersedes Procedure # Adapted from HPTN Policy. Review Date Revision Date Signature

Prepared by Date Adopted Supersedes Procedure # Adapted from HPTN Policy. Review Date Revision Date Signature APPENDIX IV: LABORATORY QUALITY CONTROL POLICY Prepared by Date Adopted Supersedes Procedure # Adapted from HPTN Policy N/A Review Date Revision Date Signature Distributed to # of Copies Distributed to

More information