Quality Control: Tips, Troubleshooting, and Tidbits

Size: px
Start display at page:

Download "Quality Control: Tips, Troubleshooting, and Tidbits"

Transcription

1 Quality Control: Tips, Troubleshooting, and Tidbits Nancy May, MBA, MT(ASCP) Technical Integration Specialist Agenda: 4 Tips 2 Troubleshootings Tools: Detector Parameters XbarM Tidbits 2 Sysmex America Inc. 1

2 Tip #1: Print your QC immediately when you submit to Insight. 3 Sysmex America Inc. 2

3 Reviewing QC Data Baso problem ( ) Unconnected Dots Sample Aspiration Error Missing Set of Dots Reviewing QC: Function Errors Sysmex America Inc. 3

4 Reviewing QC: Barcode Errors QC data is not plotted. QC must be re-analyzed to plot QC data. Sysmex America Inc. 4

5 Tip #2: If a point is just out of QC limits, have techs accept and rerun according to lab protocol. Tip #3: Do not use Login: admin for everyday use. 9 Delete Submitted Data Sysmex America Inc. 5

6 Managing QC in Insight Adding comments to any QC sample Removing samples from their Mean and SD calculations without deleting those samples from the analyzer L-J charts Sysmex America, Inc. All rights reserved. Not to be copied or distributed without expressed, written consent of Sysmex America, Inc. Insight Home Page Sysmex America Inc. 6

7 Definitions XA = Anomalous Data Data that falls outside Sysmex predetermined ranges for that level of control. These ranges are approx. 80% of the Assay Sheet Ranges Short samples, samples run in the wrong QC files samples that could skew the customer s true Mean and SD These samples are EXCLUDED from the Peer Group s Mean and SD, but are INCLUDED in the Customer s Mean and SD 14 Sysmex America Inc. 7

8 15 Comments Comments can be added to ANY QC sample Click on Manage Data for a particular sample Enter a comment, click on Save Comment Comments are printed on the Notes page of the Insight report Sysmex America Inc. 8

9 17 Manage QC Data Select Managed or Not Managed Select a comment from the drop down menu or free text a comment Click on Save Comment Sysmex America Inc. 9

10 Pre-Defined Comments Customers can choose to select a predefined comment Or they can enter a free text a comment Comments Comments are printed on the Notes page of the Insight report. They will be stored with the Insight data for a minimum of 2 ½ years Sysmex America Inc. 10

11 Questions?? Tip #4: Establish relevant/appropriate historical limits. Sysmex America Inc. 11

12 QC Chart without Limits QC Input without Limits Sysmex America Inc. 12

13 Calculating Historical Limits Average Historical CV s from Cumulative Insight reports on first 3 lot s of control material. For a 3SD graph, multiply average CV% X 3 for each parameter. Enter new limits in QC, Target/Limit. Use Spreadsheet that calculates 3CV, displays interim limits and adds drift factor. Control Levels Historical Cumulative CV s e-check Drifting Parameters Commercial controls cannot be completely stabilized over life of product and still retain sensitivity to reagents. Parameters which decrease over time: Mono - 4% to - 11% DIFF-Y - 15% to - 22% IRF - 12% to - 23% HFR - 13% to - 23% MFR - 11% to - 22% RBC-X - 5% to - 10% HPC - 2% to - 6% IG# - 3% to - 6% IG% - 2% to - 4% Source: Insight Report Guide (WI 2494 Rev 05, ) Sysmex America Inc. 13

14 Parameters which increase over time: MCV + 1% to + 2% RDW-SD + 1% to + 2% PLT-O + 2% to +16% RET% + 6% to +15% RET# + 6% to +15% IMI# + 2% to +10% IMI RF + 3% to + 4% LFR + 8% to +21% Source: Insight Report Guide (WI 2494 Rev 05, ) Historic Limits Spreadsheet Facility Name: Northern Nevada Medical Center Address: 2375 E. Prater Way City, State, Zip: Sparks, NV Instrument: XT-2000 Serial Number: Mode Open/Closed?: Closed XT Historical Limit Data e -CHECK LEVEL 3 LOT #1 LOT #2 LOT #3 3 Lot Limit% Limit% Parameter Average Average x 2 Average x 3 3 CV% Suggested CV% CV% CV% CV% (+ 2 CV) (+ 3 CV) Drift + Drift Limits RBC HGB HCT MCV MCH MCHC PLT RDW-SD RDW-CV WBC NEUT% LYMPH% MONO% EO% BASO% NEUT# LYMPH# MONO# EO# BASO# PLT-O RET# RET% IRF MPV DIFF-X 8.0 DIFF-Y 68.0 BASO-X Manufacturer Defined Limits for Detector Control/Service Parameters are 15.0 BASO-Y defined as a variation from target values and will not be changed lot to lot RBC-O These values are entered into the analyzer's control files HFR For further information please refer to: MFR Sysmex Customer Bulletin : Insight Program Modification LFR 56.0 RBC-X 64.0 RBC-Y 17.0 Sysmex America Inc. 14

15 Westgard website When should you use a multirule QC procedure? Not always! Sometimes a single rule QC procedure gives you all the error detection needed while at the same time maintaining low false rejections. This generally means eliminating the 1 2s rule because of its high false rejections and considering others such as 1 2.5s, 1 3s, and 1 3.5s which have acceptably low false rejection rates. The remaining issue is whether adequate error detection can be provided by these other single rule QC procedures. If medically important errorscan be detected 90% of the time (i.e., probability of error detection of 0.90 or greater), then a single rule QC procedure is adequate. 29 Westgard, cont. In general, you will find that single rule QC procedures are adequate for your highly automated and very precise chemistry and hematology analyzers, but you should avoid using 2s control limits or the 1 2s control rule to minimize waste and reduce costs.. To figure out exactly when to use single rule or multirule QC procedures, you will need to define the quality required for each test, look at the precision and accuracy being achieved by your method, then assess the probabilities for false rejection (P fr ) and error detection (P ed ) of the different candidate QC procedures. Aim for 90% error detection (P ed of 0.90 or greater) and 5% or less false rejections (P fr of 0.05 or less). With very stable analytical systems that seldom have problems, you may be able to settle for lower error detection,say 50%. 30 Sysmex America Inc. 15

16 WEBINAR: Control Rule Simplification of High Precision Sysmex Hematology Analyzers Fall 2010 Webinar by Dr George S. Cembrowski, MD, PhD Use 1 4.5S Rule for all parameters except HCT 1 3.5S 31 Questions?? Sysmex America Inc. 16

17 Troubleshooting Tools Detector Parameters Patient Moving Averages (XbarM) Significance of Detector Control Parameters Detector Control Parameters provide information on the performance of the analyzer but are not reported and have no inherent medical utility. Sysmex America Inc. 17

18 e-check QC Parameters Up to 48 Parameters (C + D + R + N) 12 CBC Parameters (RDW-CV, RDW-SD, PLT-O) 14 DIFF Parameters (%, #, NRBC #, %) 4 Retic Parameters (RET %, RET #, IRF, Ret-He) 16 Detector Control Parameters 2 HPC, IPF 16 Detector Control Parameters Parameter Detector Method Used MPV RBC/PLT (DC/Sheath flow) Diff-X, Diff-Y DIFF (Optical Flow) Baso-X, Baso-Y WBC/BASO (Optical Flow) NRBC-X, NRBC-Y NRBC (Optical Flow) IMI #, IMI-DC, IMI-RF IMI Channel (RF/DC) RBC-O, LFR, MFR, HFR, RBC-X, RBC-Y Retic (Optical Flow) * WBC Detector Control Parameters DIFF- Y BASO - Y DIFF - X (NEUT) BASO - X (WBC) NRBC- Y IMI - RF NRBC - X IMI DC 36 Sysmex America Inc. 18

19 RBC & Retic Detector Control Parameters RBC- Y RBC - X RBC-O PLT-O RBC-X RBC-Y LFR, MFR, HFR RBC-I, & PLT-I, MPV MPV 37 Detector Control Parameters MPV BASO-X, BASO-Y Measures platelet volume, very low MPV may indicate platelets below lower discriminator Distinguishes WBC s from RBC ghosts, and WBC s from Basophils Flags affected: Lyse Resistance WBC Abn Scattergram DIFF-X, DIFF-Y Monitors accuracy for diff % and # number Flags affected: WBC Abn Scattergram NRBC Left Shift Aty lymph Abn lymph/l-blast Plt Clumps Sysmex America Inc. 19

20 Detector Control Parameters NRBC-X, NRBC-Y IMI#, IMIDC, IMIRF RBC-X RBC-Y Ensures separation of NRBC population from RBC ghosts and WBC populations. Flags affected: NRBC, NRBC Abn Scattergram Monitors immature gran accuracy and precision of IMI detector. Flags affected: Left Shift, Blasts, Immature Grans Plt Clumps Mean fluorescence channel of mature RBC s in Retic scattergram on horizontal axis May affect IRF Mean channel for mature RBC s in Retic Scattergram on vertical axis. Questions?? Sysmex America Inc. 20

21 Patient Moving Average XbarM QC tool developed by Dr. Brian Bull in Algorithm is based on the inherent stability of patient results in a general hospital population. Monitors stability over time of: Instrumentation Reagents Technique Advantages of the Patient Moving Average Inexpensive, simple to use Utilizes unfixed patient blood Incorporated into current Hematology analyzers Real time QC Sysmex America Inc. 21

22 Xm - Modification of Bull s Xb Used to monitor Patient Moving Average Definition Bull s Algorithm N X M (i)=x M (i-1)+sgn{ SGN[X(j,i)-X M (i-1)* i=0 l X(j,i)-X M (i-1) l} *F N F={ SGN [X(j,i-X M (i-1)] l X(j,i)-X M (i-1) l} 2 i=0 N 34 Patient Moving Averages X-bar-M Effective in checking hour-by-hour changes in instrument and reagents Parameters remain stable once characteristics of population established Does not include: ID number zero Commercial QC data Calibration data Background counts Data that exceeds linearity range (@) Data that is unreliable data ( *) Data reported with analysis error Data displayed as **** or ---- Sysmex America Inc. 22

23 Installing X Series: Xm Settings Xm must be turned ON Set Xm batch: Sysmex recommends 20 During implementation: Interim limit or Assign Variable Target for Xm How to set XBarM limits QC Settings LJ, %, 3SD (only for auto-setting XbarM limits) Sysmex America Inc. 23

24 QC System Settings Control Method L-J Control Limit Setting Ratio (%) & 3SD 47 Establish historical limit & autoset target File holds 300 pts, each pt = mean of 20 pats = 6000 patients. Click and drag on all 300 pts. Select [Target / Limit] key. Click & drag to highlight all parameters. Select [Auto Setting] & Target and Limit and [OK] to auto set target and 3 CV limit. Only time you will have Limit selected. Sysmex America Inc. 24

25 QC Troubleshooting Tool Concern? Commercial QC - OK Xm - OK Commercial QC - OK Xm - Not OK Commercial QC - Not OK Xm - OK Commercial QC - Not OK Xm - Not OK No Problem Population Shift or Reagent Control problem Systematic Error XE-2100 Xm Flowchart - RBC parameters Each time Xm alerts as out, check Xm chart. Document each time Xm is out on QC log. Proceed after Xm points are out. Check if parameters that are out are related to patient population being run. i.e.) Dialysis run, pediatric run, oncology run, ER's Check maintenance log for any recent calibration or part replacement. If XE has been calibrated in the last 7 days, leave a message for supervisor. Targets may need adjusted. If YES, run XE Check reagent log. Has a reagent been changed that coincides with Xm going out. Check if reagent is running low or almost outdated. RBC/PLT HGB Ret % & #, RBC-O, RBC-X & Y IRF, LFR, MFR, HFR Cellpack Cellsheath Cellpack Sulfolyser Ret-Search II (Diluent & Dye) Sysmex America Inc. 25

26 XE-2100 Xm Flowchart - RBC parameters (cont) Check maintenance log. Has maintenance been done or not done recently? Perform maintenance. RBC/PLT HGB 1) Clog removal 2) Clean RBC Aperture 3) Clean SRV 1) Clean SRV Run the patient control from current and previous shift. Run commercial controls. Patient control & commercial control in. Patient control or commercial control out for same parameter. Run XE Monitor Xm Call Hotline If another point is out, call hotline SYSMEX Document all actions on action log. XE-2100 Xm Flowchart - WBC parameters (cont) Check reagent log. Has a reagent been changed that coincides with Xm going out? Check if reagnet is running low or almost outdated. nrbc #, nrbc-x & Y Diff parameters, Diff-X & Y IMI # WBC, Baso # & %, Baso-X & Y Stromatolyser-NR (Diluent & Dye) Stromatolyser-4DL Stromatolyser-4DS Stromatolyser-IM Stromatolyser-FB Document any reagent change on onboard reagent log. Check maintenance log. Has maintenance been done or not done recently? Perform maintenance. 1) Clog removal, 2) Clean IMI detector aperture, 3) Remove flow cell air bubbles, 4) Rinse Flowcell. Run the patient control from current and previous shift. Run commercial controls. Patient control & commercial control in. Patient control or commercial control out for same parameter. Run XE Monitor Xm Call Hotline If another point is out, call hotline SYSMEX Document all actions on action log. Sysmex America Inc. 26

27 Questions?? Tidbits PLT Clumping? flag ISLH Consensus Ruleswww.islh.org PLT rule: <100 or >1000 and first time Pediatric Normal Ranges: Pediatric Reference Intervals, 7 th edition, Sysmex America Inc. 27

28 Thank you! Questions? Sysmex America Inc. 28

QC and Changing Lots

QC and Changing Lots QC and Changing Lots Back to the Basics Karen Hoffman MT(ASCP) Senior Technical Integration Specialist Laboratory Solutions Services Sysmex America, Inc. Objectives On completion of this training you should

More information

BEYOND A BETTER BOX XN-V SERIES TM MULTISPECIES HEMATOLOGY ANALYZERS. Accelerate Your Multispecies Research Laboratory XN-V IS FOR ANIMAL USE ONLY

BEYOND A BETTER BOX XN-V SERIES TM MULTISPECIES HEMATOLOGY ANALYZERS. Accelerate Your Multispecies Research Laboratory XN-V IS FOR ANIMAL USE ONLY BEYOND A BETTER BOX XN-V SERIES TM MULTISPECIES HEMATOLOGY ANALYZERS Accelerate Your Multispecies Research Laboratory XN-V IS FOR ANIMAL USE ONLY MOVE YOUR MULTI-SPECIES LABORATORY BEYOND A BETTER BOX*

More information

XN Series. What s in it for me? Krista Curcio, MT(ASCP), MBA Product Manager, Hematology Systems Sysmex America, Inc. All rights reserved.

XN Series. What s in it for me? Krista Curcio, MT(ASCP), MBA Product Manager, Hematology Systems Sysmex America, Inc. All rights reserved. XN Series What s in it for me? Krista Curcio, MT(ASCP), MBA Product Manager, Hematology Systems 2012 Sysmex America, Inc. All rights reserved. Agenda XN Series Product Overview A system to fit every lab

More information

Automated Hematology Analyzer. Differentiate with the XT-4000i

Automated Hematology Analyzer. Differentiate with the XT-4000i XT-4000i Automated Hematology Analyzer Differentiate with the XT-4000i Advanced Technology Solutions to Meet Your Lab s Needs Even with these challenges, the need for hematology testing has remained steady

More information

Automated Hematology System. Your Complete Choice

Automated Hematology System. Your Complete Choice XE-5000 Automated Hematology System Your Complete Choice Advanced Technology Solutions to Meet Your Lab s Needs Even with these challenges, the need for hematology testing has remained steady or continued

More information

SF Cube. Exclusive Technology, Inclusive Approach. BC-6800 Auto Hematology Analyzer. Cell Analysis Technology

SF Cube. Exclusive Technology, Inclusive Approach. BC-6800 Auto Hematology Analyzer. Cell Analysis Technology Cell Analysis Technology Exclusive Technology, Inclusive Approach BC-6800 Auto Hematology Analyzer Cell Analysis Technology SF Cube SF Cube is a pathbreaking technology for reliable blood cell analysis,

More information

Greater Possibilities By Design

Greater Possibilities By Design Greater Possibilities By Design January 23, 2013 Nilam Patel MT(ASCP)SH Sr. Product Manager, Automation Solutions Objectives Review current and new portfolio Technology and benefits Operational benefits

More information

Automated Hematology Analyzer. Small Footprint. Big Difference.

Automated Hematology Analyzer. Small Footprint. Big Difference. XS-1000i Automated Hematology Analyzer Small Footprint. Big Difference. Small But Powerful: The XS-1000i Even with these challenges, the need for hematology testing has remained steady or continues to

More information

BC-6800 Auto Hematology Analyzer. Small Cube, Big Difference

BC-6800 Auto Hematology Analyzer. Small Cube, Big Difference BC-6800 Auto Hematology Analyzer Small Cube, Big Difference Small Cube, Big Difference 2D Forward scatter EOS NEU + Side scatter MON LYM 3D Forward scatter MON NEU LYM EOS + Side scatter + Fluorescence

More information

Reshaping compact automation

Reshaping compact automation Automated Hematology Analyzer XN-SERIES Reshaping compact automation www.sysmex.com/us Optimize your laboratory As analyzers become increasingly sophisticated, the laboratory s contribution to the continuum

More information

Reshaping compact automation

Reshaping compact automation Automated Hematology Analyzer XN-SERIES Reshaping compact automation www.sysmex.com/us Optimize your laboratory As analyzers become increasingly sophisticated, the laboratory s contribution to the continuum

More information

XN-SERIES AUTOMATED HEMATOLOGY SYSTEMS

XN-SERIES AUTOMATED HEMATOLOGY SYSTEMS XN-SERIES AUTOMATED HEMATOLOGY SYSTEMS Reshaping Compact Automation BEYOND A BETTER BOX MOVE YOUR HEMATOLOGY LABORATORY BEYOND A BETTER BOX It s what makes Sysmex the smarter choice and sets our integrated

More information

BEYOND A BETTER BOX XN-SERIES TM AUTOMATED HEMATOLOGY SYSTEMS. Reshaping Compact Automation

BEYOND A BETTER BOX XN-SERIES TM AUTOMATED HEMATOLOGY SYSTEMS. Reshaping Compact Automation BEYOND A BETTER BOX XN-SERIES TM AUTOMATED HEMATOLOGY SYSTEMS Reshaping Compact Automation SMALL, SMART, COMPACT AUTOMATION It s an approach to hematology testing that s never been done before. An approach

More information

Note: Please complete the report and submit it to SMILE and your PNL within 30 days. Site/Laboratory Name: New York EQA Provider and #: CAP

Note: Please complete the report and submit it to SMILE and your PNL within 30 days. Site/Laboratory Name: New York EQA Provider and #: CAP SMLE GENERAL NVESTGATON CHECKLST/FORM SURVEY NFORMATON EXTERNAL QUALTY ASSURANCE (EQA) Note: Please complete the report and submit it to SMLE and your PNL within 30 days. Site/Laboratory Name: New York

More information

11/2/2015. Describe the technologies available on the XN-Series and the reason(s) for selection

11/2/2015. Describe the technologies available on the XN-Series and the reason(s) for selection XN-Series Technology What s Inside the Box? OBJECTIVES Describe the technologies available on the XN-Series and the reason(s) for selection Utilize example case studies to illustrate the new capabilities

More information

CLINICAL SIGNIFICANCE

CLINICAL SIGNIFICANCE Subject Sysmex XS-1000i Procedure Index Number Lab-1501 Section Laboratory Subsection Regional Clinic/Affiliate Hospital Laboratories Category Departmental Contact Kamprud, Elizabeth A Last Revised 12/18/2017

More information

Immature Granulocyte Enumeration Our Journey from Manual to Automated Reporting

Immature Granulocyte Enumeration Our Journey from Manual to Automated Reporting Immature Granulocyte Enumeration Our Journey from Manual to Automated Reporting Mary Capper MT(ASCP)SH Supervisor, Hematology/Hemostasis University of Iowa Hospitals and Clinics Objectives Describe the

More information

See Brilliant Results

See Brilliant Results See Brilliant Results without all the reviews. * In Development Product not available in the U.S. First Pass Efficiency. Getting it right the first time. GOALS: More reportable WBC and WBC differential

More information

XN Series. Greater Possibilities by Design What s in it for me? Krista Curcio, MT(ASCP), MBA Product Manager, Hematology Systems

XN Series. Greater Possibilities by Design What s in it for me? Krista Curcio, MT(ASCP), MBA Product Manager, Hematology Systems XN Series Greater Possibilities by Design What s in it for me? Krista Curcio, MT(ASCP), MBA Product Manager, Hematology Systems 2012 Sysmex America, Inc. All rights reserved. Agenda XN Series Product Overview

More information

Ruby. A shining example of. Put science on your side. Intended Use:

Ruby. A shining example of. Put science on your side. Intended Use: High efficiency hematology A shining example of advanced technology Intended Use: The CELL-DYN Ruby is a multi-parameter automated Hematology analyzer designed for in vitro use in clinical laboratories.

More information

BC Auto Hematology Analyzer. A CUTE 5-part

BC Auto Hematology Analyzer. A CUTE 5-part BC-5000 Auto Hematology Analyzer A CUTE 5-part Why do we need 5-part hematology analyzers? WBC differential: 3-part WBC differential: 5-part Lymphocyte Mid-size cell? Eosinophil Basophil Neutrophil DIFF

More information

BC Auto Hematology Analyzer. A CUTE 5-part

BC Auto Hematology Analyzer. A CUTE 5-part BC-5150 Auto Hematology Analyzer A CUTE 5-part Why do we need 5-part hematology analyzers? WBC differential: 3-part WBC differential: 5-part Lymphocyte Mid-size cell? Eosinophil Basophil Neutrophil DIFF

More information

9/29/2015. Describe the technologies available on the XN-Series and the reason(s) for selection

9/29/2015. Describe the technologies available on the XN-Series and the reason(s) for selection XN-Series Technology What s Inside the Box? OBJECTIVES Describe the technologies available on the XN-Series and the reason(s) for selection Utilize example case studies to illustrate the new capabilities

More information

XN-9100 AUTOMATED HEMATOLOGY SYSTEMS

XN-9100 AUTOMATED HEMATOLOGY SYSTEMS XN-9100 AUTOMATED HEMATOLOGY SYSTEMS Reshaping Scalable Automation BEYOND A BETTER BOX MOVE YOUR HEMATOLOGY LABORATORY BEYOND A BETTER BOX It s what makes Sysmex the smarter choice and sets our integrated

More information

BEYOND A BETTER BOX XN-9000 AUTOMATED HEMATOLOGY SYSTEMS. Reshaping Scalable Automation

BEYOND A BETTER BOX XN-9000 AUTOMATED HEMATOLOGY SYSTEMS. Reshaping Scalable Automation BEYOND A BETTER BOX XN-9000 AUTOMATED HEMATOLOGY SYSTEMS Reshaping Scalable Automation 2 Sysmex XN-Series Automated Hematology Analyzers are designed to exceed the expectations of real people who work

More information

Quintus 5-part hematology analyzer In-depth, quality-controlled 26-parameter results

Quintus 5-part hematology analyzer In-depth, quality-controlled 26-parameter results Outstanding user-interface Only three reagents needed 100-sample Autoloader option Quintus 5-part hematology analyzer In-depth, quality-controlled 26-parameter results Premium system solution boosted by

More information

It Should have five discrete analysis modes CBC, CBC+ DIFF, CBC + Retic, CBC+Retic+Diff & Retic only.

It Should have five discrete analysis modes CBC, CBC+ DIFF, CBC + Retic, CBC+Retic+Diff & Retic only. Specification of 5 Part Differential Haematology Analyser Specification as per the Tender Amended Specifications as per the Notification in Chapter 4 Pre Bid meeting held on 22-02-2018 in Chapter 4 It

More information

Performance Evaluation of the Coulter LH 750 Hematology Analyzer

Performance Evaluation of the Coulter LH 750 Hematology Analyzer ISLH Laboratory Hematology 7:217 228 2001 Carden Jennings Publishing Co., Ltd. Official Publication Performance Evaluation of the Coulter LH 750 Hematology Analyzer TRACEY FERNANDEZ, LYNN BESSERT DOMACK,

More information

We Believe the Possibilities. Automated hematology Analyzer. XN-L Series. Leading Hematology for Better Patient Care

We Believe the Possibilities. Automated hematology Analyzer. XN-L Series. Leading Hematology for Better Patient Care We Believe the Possibilities. Automated hematology Analyzer XN-L Series Leading Hematology for Better Patient Care XN Introducing XN-L Series XN-L Series is the latest compact fully-automated 6-part differential

More information

XN Hematology Case Studies Every Picture Tells a Story

XN Hematology Case Studies Every Picture Tells a Story XN Hematology Case Studies Every Picture Tells a Story Barbara J Connell MS, MT(ASCP)SH Senior Product Manager Technical Marketing & Communications Sysmex America, Inc Objectives Describe the advantages

More information

Auto-Validation Don t Rule It Out!

Auto-Validation Don t Rule It Out! Auto-Validation Don t Rule It Out! Elinore Craig Product Manager, IT Integration Agenda Barriers to auto-validation Success stories Overcoming the barriers 1 Barriers AACC Auto-validation Survey About

More information

Features & Benefits Five-part differential on cat, dog, horse, cow, alpaca and llama; three-part differential on the nine other species

Features & Benefits Five-part differential on cat, dog, horse, cow, alpaca and llama; three-part differential on the nine other species Telephone: Facsimile: e-mail: + 27 (0)11 791 0025 + 27 (0)86 551 0832 johan@anjotech.co.za VetScan HM5 Hematology System Advanced Hematology Five-Part Differential The VetScan HM5 is a fully automated

More information

CONCURRENT 10: THE COMPLETE BLOOD COUNT AND BEYOND: QUALITY ISSUES AND REFERENCE METHODS

CONCURRENT 10: THE COMPLETE BLOOD COUNT AND BEYOND: QUALITY ISSUES AND REFERENCE METHODS CONCURRENT 10: THE COMPLETE BLOOD COUNT AND BEYOND: QUALITY ISSUES AND REFERENCE METHODS Saturday May 14 th 2016 1:30 3:00 PM (13.30 15.00hrs) International Society for Laboratory Hematology (ISLH) CONCURRENT

More information

Haematology. Reagents and analyzers SWISS HAEMATOLOGY S O L U T I O N

Haematology. Reagents and analyzers SWISS HAEMATOLOGY S O L U T I O N Haematology Catalogue Reagents and analyzers SWISS HAEMATOLOGY S O L U T I O N 24 parameters 5 DIFF analyzer with true optical differential measurement Optical laser method for 5-part WBC differential

More information

Department of Pathology

Department of Pathology KASTURBA HEALTH SOCIETY'S MAHATMA GANDHI INSTITUTE OF MEDICAL SCIENCES, P.O. SEVAGRAM, WARDHA (DISTT.) MAHARASHTRA STATE - 442 102 Specifications for Equipment Department of Pathology 1) 3 Part Hematology

More information

Diagnostic Medical Equipment New Revenue for Your Medical Practice

Diagnostic Medical Equipment New Revenue for Your Medical Practice Diagnostic Medical Equipment New Revenue for Your Medical Practice Consultant Libby Knollmeyer s Lab Machine Primer Series Part 1: Hematology Analyzers When laboratories in physician offices move out of

More information

Evaluation of the Automated Immature Granulocyte Count (IG) on Sysmex XE-2100 Automated Haematology Analyser vs. Visual Microscopy (NCCLS H20-A)

Evaluation of the Automated Immature Granulocyte Count (IG) on Sysmex XE-2100 Automated Haematology Analyser vs. Visual Microscopy (NCCLS H20-A) Evaluation of the Automated Immature Granulocyte Count (IG) on Sysmex XE-21 Automated Haematology Analyser vs. Visual Microscopy (NCCLS H2-A) Th WEILAND *1, H KALKMAN *2, and H HEIHN *1 *1 Central Laboratory,

More information

HST-330 TM and Total Laboratory Automation

HST-330 TM and Total Laboratory Automation HST-330 TM and Total Laboratory Automation Hyun-Sook CHI, MD* * Department of Clinical Pathology, University of Ulsan College of Medicine and Asan Medical Center, Seoul, Korea. Total laboratory automation

More information

Product Fact Sheet Added values XN-CBC and XN-DIFF

Product Fact Sheet Added values XN-CBC and XN-DIFF Added values XN-CBC and XN-DIFF Clinical values Product name Reliable NRBC counts for all neonate and adult samples of low and high counting ranges enable immediate therapeutic action, e.g. for patients

More information

norma Hematology family norma norma Vision Reagents

norma Hematology family norma norma Vision Reagents norma Hematology family norma Reagents norma Vision Smallest reagent consumption in the world Smallest processed sample volume, ideal for pediatric use Microfluidic technology 3 The Icon 3 is a smart,

More information

norma Hematology family norma Access norma norma Vision Reagents

norma Hematology family norma Access norma norma Vision Reagents norma Hematology family norma Access norma Reagents norma Vision Smallest reagent consumption in the world Smallest processed sample volume, ideal for pediatric use Microfluidic technology 3 A smart hematology

More information

DxH part differential closed tube hematology analyzer

DxH part differential closed tube hematology analyzer DxH 520 5-part differential closed tube hematology analyzer DxH 520 HEMATOLOGY ANALYZER Minimum footprint. Maximum productivity. The new closed tube 5-part differential analyzer because great things come

More information

WADA Technical Document TD2019BAR. Blood Analytical Requirements for the Athlete Biological Passport

WADA Technical Document TD2019BAR. Blood Analytical Requirements for the Athlete Biological Passport 1. Introduction Blood Analytical Requirements for the Athlete Biological Passport This Technical Document (TD) has been established to harmonize the analysis of blood Samples collected, both In-Competition

More information

Verification of abnormal results from Coulter instrumentation

Verification of abnormal results from Coulter instrumentation Created Lisa Senzel Asst. Prof. - Clinical 5/9/2008 Path Revised Bruce Kube Day Heme Tech 1 10/24/2012 Reviewed Jay Bock M.D. 4/10/2012 Procedure: Approved Lisa Senzel Asst. Prof. - Clinical Path Replaces

More information

AUTOMATED HEMATOLOGY IN THE CLINICAL LAB

AUTOMATED HEMATOLOGY IN THE CLINICAL LAB AUTOMATED HEMATOLOGY IN THE CLINICAL LAB Diane Gaspari, SH(ASCP) Core Lab Manager York Hospital, York PA ggaspari@wellspan.org Objectives of Automation Increase efficiency; streamline workflow Increase

More information

RUO Addendum. UniCel DxH Series with System Manager Software. Coulter Cellular Analysis System. B26673AC July 2015

RUO Addendum. UniCel DxH Series with System Manager Software. Coulter Cellular Analysis System. B26673AC July 2015 RUO Addendum UniCel DxH Series with System Manager Software Coulter Cellular Analysis System Slidemaker Stainer July 2015 Manufactured by Beckman Coulter, Inc. 250 S. Kraemer Blvd. Brea, CA 92821 U.S.A.

More information

ABX Pentra XL 80. Hematology analyzer. Auto-sampling 5-part Differential Customized Dilution Ratio Integrated Validation Station

ABX Pentra XL 80. Hematology analyzer. Auto-sampling 5-part Differential Customized Dilution Ratio Integrated Validation Station ABX Pentra XL 80 Hematology analyzer Auto-sampling 5-part Differential Customized Dilution Ratio Integrated Validation Station Pentra XL 80 A new way to manage your hematology workflow With Pentra XL 80

More information

Examples of body fluid analysis with the XE-5000

Examples of body fluid analysis with the XE-5000 Examples of body fluid analysis with the XE-5000 Sysmex Xtra Online March 2011 If patients are to be treated rapidly and effectively, laboratory results must be both rapid and reliable if at all possible,

More information

Establishing Quality Control Target Values and Standard Deviations for Hematology Instrumentation

Establishing Quality Control Target Values and Standard Deviations for Hematology Instrumentation Establishing Quality Control Target Values and Standard Deviations for Hematology Instrumentation For informational purposes only. WHO REQUIRES IT AND WHAT THEY REQUIRE Regulatory agencies require the

More information

Practical Guide to Running Controls

Practical Guide to Running Controls Practical Guide to Running Controls Internal quality control Introduction In addition to the competenceof the lab managerand the technical validation, the use and proper management of quality control (IQC,

More information

Sysmex Educational Enhancement and Development No

Sysmex Educational Enhancement and Development No SEED Haematology Sysmex Educational Enhancement and Development No 2 2015 SNCS the quality control solution for the haematology laboratory The purpose of this newsletter is to provide an overview of the

More information

XS-1000i New Sysmex 5-part diff haematology analyser with fluorescence technology

XS-1000i New Sysmex 5-part diff haematology analyser with fluorescence technology XS-1000i New Sysmex 5-part diff haematology analyser Fig. 1: sysmex xs-1000i Since the introduction of fluorescence technology for leukocyte differentiation by sysmex in the year 1999, thousands of laboratories

More information

Medonic M-series M32. User s Manual

Medonic M-series M32. User s Manual Medonic M-series M32 User s Manual Page ii Table of Contents TABLE OF CONTENTS Section 1. Introduction 5 Medonic M-series M32 Systems... 5 Contact Details... 5 Analyzer Overview... 6 Consumable Overview...

More information

Hematology and Coagulation Controls

Hematology and Coagulation Controls Bio-Rad Laboratories QUALITY CONTROL SOLUTIONS FOR HEMATOLOGY AND COAGULATION Hematology and Coagulation Controls Your Quality Goals Realized Comprehensive QC solutions for building an effective quality

More information

SEED HAEMATOLOGY. SNCS IQAS Online the quality control solution for the haematology laboratory

SEED HAEMATOLOGY. SNCS IQAS Online the quality control solution for the haematology laboratory SYSMEX EDUCATIONAL ENHANCEMENT AND DEVELOPMENT NOVEMBER 2015 SEED HAEMATOLOGY SNCS IQAS Online the quality control solution for the haematology laboratory Introduction It is standard practice for laboratories

More information

Successful performance in proficiency testing (PT) is required

Successful performance in proficiency testing (PT) is required Performance Characteristics of Several Rules for Self-interpretation of Proficiency Testing Data R. Neill Carey, PhD; George S. Cembrowski, MD; Carl C. Garber, PhD; Zohreh Zaki, MD Context. Proficiency

More information

TENDER DOCUMENT. NIT No.: NRO/CON/738/672 Date: FOR VOLUME III BILL OF QUANTITY / PRICE BID ENGINEERING PROJECTS (INDIA) LIMITED

TENDER DOCUMENT. NIT No.: NRO/CON/738/672 Date: FOR VOLUME III BILL OF QUANTITY / PRICE BID ENGINEERING PROJECTS (INDIA) LIMITED TENDER DOCUMENT NIT : NRO/CON/738/672 Date: 12.10.2018 FOR Supply, Installation, Testing, Commissioning & Handingover of Medical equipments (Group-7E) for Government Medical College, Barmer (Rajasthan).

More information

Dry Hematology Analyzer Fast, Simple CBC Analysis Designed for the Point of Care

Dry Hematology Analyzer Fast, Simple CBC Analysis Designed for the Point of Care QBC STAR TM Dry Hematology Analyzer Fast, Simple CBC Analysis Designed for the Point of Care Reach for the STAR The QBC STAR hematology analyzer expands and advances the unique capabilities of previous

More information

UNRIVALED SIMPLICITY. Exceptional Ease of Use. The STAR brings new meaning to the word simple

UNRIVALED SIMPLICITY. Exceptional Ease of Use. The STAR brings new meaning to the word simple UNRIVALED SIMPLICITY The QBC STAR Centrifugal Hematology System is the simplest solution for in-office CBC testing. An innovative approach employing dry hematology reagents instead of the bulky liquid

More information

The Simple Solution for CBC Testing

The Simple Solution for CBC Testing QBC STAR Centrifugal Hematology Analyzer The Simple Solution for CBC Testing Diagnostics Innovative Solutions for a Healthier World UNRIVALED SIMPLICITY The QBC STAR Centrifugal Hematology System is the

More information

QUICK REFERENCE GUIDE. Easy operation fast results. Closer to Care.

QUICK REFERENCE GUIDE. Easy operation fast results. Closer to Care. QUICK REFERENCE GUIDE Easy operation fast results Unpacking and Setup Unpack the VetScan HM5 analyzer from its shipping container and put out the accesory box. Note: Store the shipping box, accessory box,

More information

ABX Pentra DX120. Hematology analyzer. 45 parameters Integrated slide-maker Expert validation system

ABX Pentra DX120. Hematology analyzer. 45 parameters Integrated slide-maker Expert validation system ABX Pentra DX120 Hematology analyzer 45 parameters Integrated slide-maker Expert validation system ABX Pentra DX 120 In the heart of hematopoiesis Complete cytology platform 120 samples per hour Continuous

More information

Designing QC Activities to More Precisely Manage Analytical Accuracy and Patient Risk

Designing QC Activities to More Precisely Manage Analytical Accuracy and Patient Risk Designing QC Activities to More Precisely Manage Analytical Accuracy and Patient Risk Curtis A. Parvin, Ph.D. Manager of Advanced Statistical Research Bio-Rad Laboratories, QSD curtis_parvin@bio-rad.com

More information

Measurement Uncertainty Guide. ISO Accreditation Program

Measurement Uncertainty Guide. ISO Accreditation Program Measurement Uncertainty Guide ISO 15189 Accreditation Program Background Why This is Necessary The ISO 15189:2012 standard contains enhanced expectations regarding measurement uncertainty (MU) in clause

More information

Fully Automated EIA System. True open flexibility.

Fully Automated EIA System. True open flexibility. True open flexibility www.triturus.com The immunoassay analyzer of choice: completely open, fully automated, multi-test and multi-batch. TRUE OPEN FLEXIBILITY What is Triturus? Triturus is a completely

More information

Instruction Manual. Catalog # For use with Bio-Plex Manager Software Version 4.0 and MCV plate III

Instruction Manual. Catalog # For use with Bio-Plex Manager Software Version 4.0 and MCV plate III Instruction Manual Catalog # 171-203001 For use with Bio-Plex Manager Software Version 4.0 and MCV plate III For technical service, call your local Bio-Rad office, or in the US, call 1-800-4BIORAD (1-800-424-6723)

More information

TEPZZ 4Z 765B_T EP B1 (19) (11) EP B1 (12) EUROPEAN PATENT SPECIFICATION

TEPZZ 4Z 765B_T EP B1 (19) (11) EP B1 (12) EUROPEAN PATENT SPECIFICATION (19) TEPZZ 4Z 765B_T (11) EP 2 402 765 B1 (12) EUROPEAN PATENT SPECIFICATION (45) Date of publication and mention of the grant of the patent: 14.06.2017 Bulletin 2017/24 (51) Int Cl.: G01N 35/00 (2006.01)

More information

Note: The taller 100 x 13 mm tube is not designed to run on the ACL TOP. Plasma after centrifugation of these tubes must be placed in sample cups.

Note: The taller 100 x 13 mm tube is not designed to run on the ACL TOP. Plasma after centrifugation of these tubes must be placed in sample cups. Created Revised Reviewed Approved Lupus Insensitive aptt on the ACL TOP HEM 4.15.1 7/13/2012 4/11/2013 4/11/2013 4/11/2013 I. PRINCIPLE The activated partial thromboplastin time is a global screening procedure

More information

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK Haematology Department Contact: Ian Diton Conquest Hospital Tel: +44 (0) 1323 417400 ex 4127 The Ridge E-Mail: ian.diton@nhs.net St Leonards

More information

Scholars Journal of Applied Medical Sciences (SJAMS)

Scholars Journal of Applied Medical Sciences (SJAMS) Scholars Journal of Applied Medical Sciences (SJAMS) Abbreviated Key Title: Sch. J. App. Med. Sci. Scholars Academic and Scientific Publisher A Unit of Scholars Academic and Scientific Society, India www.saspublisher.com

More information

True Walkaway Automated Urinalysis, Chemistry to Microscopy The CLINITEK AUWi System Answers for life.

True Walkaway Automated Urinalysis, Chemistry to Microscopy The CLINITEK AUWi System Answers for life. www.siemens.com/diagnostics True Walkaway Automated Urinalysis, Chemistry to Microscopy The CLINITEK AUWi System Answers for life. Can I achieve true unattended urinalysis testing in my laboratory? With

More information

Pushing limits in routine laboratory haematology with the XT-4000i

Pushing limits in routine laboratory haematology with the XT-4000i Pushing limits in routine laboratory haematology with the XT-4000i Sysmex Xtra Online December 2013 Fig. 1 The XT-4000i haematology system With the launch of the XE-5000, Sysmex set new standards in haematology.

More information

AU680 Chemistry Analyzer In-Lab Training Manual

AU680 Chemistry Analyzer In-Lab Training Manual AU680 Chemistry Analyzer In-Lab Training Manual AU680 In-Lab Training Manual Page 1 of 152 AU680 In-Lab Training Manual Page 2 of 152 AU680 Chemistry Analyzer WARNINGS AND PRECAUTIONS Read all product

More information

Novel POC analysis for determination of total and 5-part differential WBC count among a US population, in comparison to Beckman Coulter LH750

Novel POC analysis for determination of total and 5-part differential WBC count among a US population, in comparison to Beckman Coulter LH750 Novel POC analysis for determination of total and -part differential WBC count among a US population, in comparison to Beckman Coulter LH7 Reed, Jeremy*, Johnsson, Elin**, Hockum, Sandra** * Quest Diagnostics,

More information

Sysmex Services. NE User Group Meeting June 20 th, Copyright 2013 by Sysmex America, Inc. All rights reserved.

Sysmex Services. NE User Group Meeting June 20 th, Copyright 2013 by Sysmex America, Inc. All rights reserved. Sysmex Services NE User Group Meeting June 20 th, 2013 1 Hematology Specialists 2 Laboratory Generalists Generalist: Broad knowledge of patient care Holistic view of patient results Flexible and marketable

More information

Establishing Chemistry QC Ranges

Establishing Chemistry QC Ranges Chemistry Guideline for Establishing New Control Lot Means and Quality Control (QC) Ranges Through Parallel Testing and Historic Coefficient of Variation (%CV h ) Authored by Kurt Michael and Paul Richardson

More information

Laboratory Quality Control Management System. A New Approach to Quality Control

Laboratory Quality Control Management System. A New Approach to Quality Control Laboratory Quality Control Management System A New Approach to Quality Control Overview Provide hospitals and reference laboratories with tools to capture Quality Control (QC)/Proficiency Testing (PT)

More information

Upon completion of the Clinical Hematology rotation, the MLS student will be able to:

Upon completion of the Clinical Hematology rotation, the MLS student will be able to: Clinical Performance Objectives in Clinical Hematology Department of Medical and Research Technology University of Maryland School of Medicine Spring 2015 Upon completion of the Clinical Hematology rotation,

More information

Original Article. A Novel Automated Slide-Based Technology for Visualization, Counting, and Characterization of the Formed Elements of Blood

Original Article. A Novel Automated Slide-Based Technology for Visualization, Counting, and Characterization of the Formed Elements of Blood Original Article A Novel Automated Slide-Based Technology for Visualization, Counting, and Characterization of the Formed Elements of Blood A Proof of Concept Study James W. Winkelman, MD; Milenko J. Tanasijevic,

More information

From routine to specific testing. Now, pass your Analysis with flying Colours. High-precision Optical system. High quality analytical technology

From routine to specific testing. Now, pass your Analysis with flying Colours. High-precision Optical system. High quality analytical technology From routine to specific testing Now, pass your Analysis with flying Colours Up to 420 tests per hour with the ISE module 55 on-board parameters with back-up availability Innovative cassette format Integrated

More information

2015 Version MODULE 8. Laboratory Testing. SLMTA Trainer s Guide

2015 Version MODULE 8. Laboratory Testing. SLMTA Trainer s Guide 2015 Version MODULE 8 Laboratory Testing SLMTA Trainer s Guide MODULE 8. LABORATORY TESTING Performance Outcome Overview With satisfactory participation in the training and successful implementation of

More information

Applied Biosystems 7900HT Fast Real-Time PCR System. Performing Fast Gene Quantitation with 384-Well Plates DRAFT

Applied Biosystems 7900HT Fast Real-Time PCR System. Performing Fast Gene Quantitation with 384-Well Plates DRAFT User Bulletin Applied Biosystems 7900HT Fast Real-Time PCR System October 3, 2005 SUBJECT: Performing Fast Gene Quantitation with 384-Well Plates In This User Bulletin System Overview System Requirements

More information

Common Deficiencies

Common Deficiencies Question text Deficiency #12: Does the laboratory have a documented system to ensure consistency of morphologic observations among all personnel performing (U/A, Micro, Hem) (sample) Suggested methods

More information

SPECIFIC DEPARTMENT INSTRUCTIONS:

SPECIFIC DEPARTMENT INSTRUCTIONS: Subject LIS Downtime Instructions - Regional Clinic / Affiliate Hospital Laboratories Index Number Lab-1520 Section Laboratory Subsection Regional Clinic / Affiliate Hospital Laboratories Category Departmental

More information

Implementation and evaluation of the Extended IPU GFHC sample validation rule set

Implementation and evaluation of the Extended IPU GFHC sample validation rule set Implementation and evaluation of the Extended IPU GFHC sample validation rule set JANE CULLIGAN & LAURA KELLY M AT E R M I S E R I C O R D I A E U N I V E R S I T Y H O S P I TA L D U B L I N, I R E L

More information

-ah. Huntsville Hospital Huntsville, AL. Connecting our Lab Vision to Lab Success: Our Upgrage Journey and Improvements Achieved

-ah. Huntsville Hospital Huntsville, AL. Connecting our Lab Vision to Lab Success: Our Upgrage Journey and Improvements Achieved Connecting our Lab Vision to Lab Success: Our Upgrage Journey and Improvements Achieved Huntsville Hospital Huntsville, AL Nichola Marcus, Automated Laboratory Manager -ah About Huntsville Alabama 2 3

More information

USE OF INFORMATICS IN ANALYTICAL QUALITY

USE OF INFORMATICS IN ANALYTICAL QUALITY USE OF INFORMATICS IN ANALYTICAL QUALITY DR SUVIN SHETTY MD HEAD, DEPT OF LABORATORY MEDICINE DR L H HIRANANDANI HOSPITAL, MUMBAI 4/14/2017 Dr Suvin Shetty, Informatics In Analytical Quality 1 DR L H HIRANANDANI

More information

Document Review 1 2 3

Document Review 1 2 3 01 DOCUMENT CONTROL 01 4.3 The current authorized test method and (7.2.1.2) supporting work instructions are available to the analyst. 02 01 5.4 (7.2.1.5) TEST METHOD VERIFICATION/VALIDATION For standard

More information

COMPARISON OF PLATELET COUNTS BY SYSMEX XE 2100 AND LH 750 WITH THE INTERNATIONAL FLOW REFERENCE METHOD IN THROMBOCYTOPENIC PATIENTS

COMPARISON OF PLATELET COUNTS BY SYSMEX XE 2100 AND LH 750 WITH THE INTERNATIONAL FLOW REFERENCE METHOD IN THROMBOCYTOPENIC PATIENTS COMPARISON OF PLATELET COUNTS BY SYSMEX XE 2100 AND LH 750 WITH THE INTERNATIONAL FLOW REFERENCE METHOD IN THROMBOCYTOPENIC PATIENTS TINA DADU 1, KUNAL SEHGAL 2, ANJUM SHAIKH 2, SHANAZ KHODAIJI 2 1. BLK

More information

Medical Electronic Systems QwikCheck Beads Validation and Training Kit

Medical Electronic Systems QwikCheck Beads Validation and Training Kit Medical Electronic Systems www.mes-global.com Medical Electronic Systems QwikCheck Beads Validation and Training Kit OVERVIEW The QwikCheck Beads Validation and Training Kit is designed as a proficiency,

More information

Advances in Biochemistry & Applications in Medicine

Advances in Biochemistry & Applications in Medicine Advances in Biochemistry & Applications in Medicine Chapter 1 ISBN: 978-93-87500-49-5 Quality Control in a Clinical Laboratory Vaneet Kaur 1 ; Pawan Kumar Kare 1* ; Himanshu Madaan 1 1 Department of Biochemistry,

More information

Hematology analyzers. Test menu: Chartable (standard menu: WBC, RBC, Hb, Hct, MCV, MCH, MCHC, PLT, %&# neut, mono, lymph, eos, baso):

Hematology analyzers. Test menu: Chartable (standard menu: WBC, RBC, Hb, Hct, MCV, MCH, MCHC, PLT, %&# neut, mono, lymph, eos, baso): december 2014 cap today 21 32 analyzers in focus, from menu to special features Two new analyzers and a new workcell automation solution are in this year s product guide to hematology analyzers. Sysmex

More information

BEYOND A BETTER BOX DI-60 INTEGRATED SLIDE PROCESSING SYSTEM. Seamless Integration

BEYOND A BETTER BOX DI-60 INTEGRATED SLIDE PROCESSING SYSTEM. Seamless Integration BEYOND A BETTER BOX DI-60 INTEGRATED SLIDE PROCESSING SYSTEM Seamless Integration RESHAPING HEMATOLOGY Scalable Automation and On-board Slide Processing The Sysmex DI-60 Integrated Cell Image Analyzer

More information

SMILE Johns Hopkins University Baltimore, MD USA. Guidelines for Manual Evaluation of CAP Hemocytometer Fluid Count

SMILE Johns Hopkins University Baltimore, MD USA. Guidelines for Manual Evaluation of CAP Hemocytometer Fluid Count SMILE Johns Hopkins University Baltimore, MD USA Guidelines for Manual Evaluation of CAP Hemocytometer Fluid Count Author: Heidi Hanes, BS, MT (ASCP) SH Document Number: Pro22-25 Effective (or Post) Date:

More information

Validation of a quality assessment system for blood gas and electrolyte testing

Validation of a quality assessment system for blood gas and electrolyte testing Clinica Chimica Acta 382 (2007) 65 70 www.elsevier.com/locate/clinchim Validation of a quality assessment system for blood gas and electrolyte testing John G. Toffaletti a,, Elizabeth H. McDonnell a, Lakshmi

More information

Objectives. Martin Health System 5/10/2016

Objectives. Martin Health System 5/10/2016 Immucor User Group Meeting Boca Raton, Florida May 12, 2016 Patricia Houtz BS, MT (ASCP) Martin Health System Blood Bank Supervisor Objectives Describe Martin s antibody resolution workflow Illustrate

More information

OFFICE OF THE DIRECTOR (MEDICAL SERVICES) NEW DELHI MUNICIPAL COUNCIL CHARAK PALIKA HOSPITAL MOTI BAGH : NEW DELHI

OFFICE OF THE DIRECTOR (MEDICAL SERVICES) NEW DELHI MUNICIPAL COUNCIL CHARAK PALIKA HOSPITAL MOTI BAGH : NEW DELHI OFFICE OF THE DIRECTOR (MEDICAL SERVICES) NEW DELHI MUNICIPAL COUNCIL CHARAK PALIKA HOSPITAL MOTI BAGH : NEW DELHI Name of the firm to whom issued --------------------------------------------------------

More information

A Division of Woodley Equipment Company Ltd

A Division of Woodley Equipment Company Ltd LEADING EDGE TECHNOLOGY FOR POINT OF CARE HAEMATOLOGY A Division of Woodley Equipment Company Ltd Contents The QBC STAR POC Haematology System...................4 Operating Ranges.......................................7

More information

Simply stated, hematology analyzers determine the

Simply stated, hematology analyzers determine the Simply stated, hematology analyzers determine the quantity of each type of cell that makes up the patient s blood. Only recently has the role and importance of each blood component been understood. As

More information

2008 HFC-A PARTICIPANT SUMMARY

2008 HFC-A PARTICIPANT SUMMARY 2008 HFC-A PARTICIPANT SUMMARY Evaluation Criteria Analyte Target Value Evaluation Criteria Red Blood Cell Peer Group ± 3 SD White Blood Cell Peer Group ± 3 SD Cytopreparation Differential Educational

More information