Intensity Modulated Arc Therapy: Technology and Clinical Implementation
|
|
- Rodger Lane
- 6 years ago
- Views:
Transcription
1 Intensity Modulated Arc Therapy: Technology and Clinical Implementation Cedric X. Yu University of Maryland School of Medicine
2 Abstract Intensity modulated arc therapy (IMAT) was proposed by Yu as an alternative to Tomotherapy. Over more than a decade, many progresses have been made. The advantages and limitations of IMAT technique have also been better understood. In recent years, single arc forms of IMAT has emerged and commercially adopted. The leading example was the Volumetric Modulated Art Therapy, a single arc form of IMAT that delivers apertures of varying weightings with a single arc rotation using dose rate variation of the treatment machine. With the commercial implementation of VMAT, wide clinical adoption is quickly taken root. However, there is a lack of a general understanding of how such arc treatments are planned, and what delivery limitations and compromises are made. Commercial promotions and competitions added further confusion for the end users. It is therefore necessary to provide a summary of this technology and some guidelines on its clinical implementation and quality assurance. The purpose of this review is to provide a summary of the works from the community leading to the wide clinical adaptation, and point to the issues still remaining, provide some perspectives on its further development in the context of increased clinical use of image guidance. Because there has been vast experience in IMRT using multiple intensity modulated fields, comparisons between IMAT and IMRT are also made in the review in the areas of planning, delivery and quality assurance.
3 I. HISTORICAL REVIEW 1, Early development leading to IMAT Although arc therapy can be traced back to the dawn of the 20 th century, arc therapy involving dynamic field shaping using a multileaf collimator was first described by Takahashi in 1965 [1]. He described a method of rotational therapy, which we now referred to as conformal arc therapy the beam aperture shaped by a multiple leaf collimator (MLC) dynamically varies to match the beam s-eye-view of the target. In 1982, Brahme et al solved an integral equation for a hypothetical target wrapped around a critical structure and treated with arc therapy [2]. They demonstrated that to deliver a uniform dose to the target while sparing the critical structure, the beam intensities have to be modulated. In 1983, Chin et al proposed and demonstrated that with computer optimization and the freedom of computer-controlled gantry rotation, collimator motion and dose-rate variation, highly conformal dose distribution can be achieved [3]. These initial developments on arc therapy were accompanied and followed by the development and wide adoption of three-dimensional radiation therapy (3DCRT) in the 1980s [4]. The need for more convenient field shaping brought the multileaf collimator (MLC) to radiotherapy practice. In 1985, Brahme et al published a paper which showed that if the intensities of radiation can be modulated across a radiation field, the increased freedom would afford a greater ability to shape the volume of high doses to better conform to the target than 3DCRT [5]. The motorized field shaping capabilities of MLC was quickly explored to modulate the intensities within a radiation field. Intensitymodulated radiation therapy (IMRT) aims to deliver a highly conformal dose to a tumor while sparing the surrounding normal tissues and sensitive structures. In 1992, Convery
4 and Rosenbloom derived the mathematical formula for realizing intensity modulation with the dynamic movement of collimator [6]. In 1994, more works were published to demonstrate the feasibility of using MLCs for intensity modulation [7-10]. The amount of work on this emerging technology quickly mushroomed, and clinical implementations of the IMRT technique immediately followed [11]. In 1993, another form of IMRT using rotational fan beams, called Tomotherapy, was proposed by Mackie et al [12]. The idea was quickly commercialized as the Peacock device by NOMOS Corporation [N/A]. Intensity modulation was achieved with a binary collimator, which opens and closes under computer control. As the fan beam continuously rotates around the patient, the exposure time of a small width of the fan beam, or a beamlet, can be adjusted with the opening and closing of the binary collimator, allowing the radiation to be delivered to the tumor through the most preferred directions and locations of the patient. The initial commercial system by NOMOS Corporation added the binary collimator on to a linear accelerator and delivered radiation treatments one slice at a time, and the treatment table had to be precisely indexed from one slice to the next. Helical tomotherapy was then developed by Tomotherapy, Inc. as a dedicated rotational IMRT system with a slip-ring rotating gantry. More automated delivery was achieved by continuous gantry rotation and treatment couch translation [same as 12]. The dosimetric advantages of rotational treatments are illustrated by Shepard et al [13], which summarizes results from an optimization series performed for a C-shaped target with a sensitive structure in the concavity of the C. For these simulations, all planning parameters such as percent dose constraints were held constant except for the number of beam angles. It was shown that each increase in the number of beam angles led to a more homogeneous dose in the tumor and a
5 lower dose to the sensitive structure. Significant dosimetric improvements continued well beyond the number of beam angles typically used for fixed field IMRT. It is also noteworthy that the total integral dose is nearly independent of the number of beam angles. # Angles Obj. Funct. Value Std. Dev. in target dose d 95 Mean dose to RAR Total integral dose Table 1: Impact of number of beam angles on plan quality. permission from [13] Reproduced with Intensity modulated arc therapy (IMAT) was introduced by the author in 1995[14]. Like tomotherapy, IMAT is delivered in an arc manner. Instead of using rotating fan beams as in tomotherapy, IMAT uses rotational cone beams of varying shapes to achieve intensity modulation. Starting from the same tomotherapy plan, which approximates the full arc with evenly spaced fixed fields, the strategy was to convert the intensity patterns into multiple segments and deliver with overlapping arcs. Based on the fact that numerous segment configurations can yield the same intensity pattern, it is possible to find such a segment configuration at each beam angle such that segments at successive angles are connected geometrically. The stacks of overlapping beam apertures can then be delivered with multiple overlapping arcs. Through the initial proof-of-principle study, it was shown that IMAT could be a valid alternative to tomotherapy in terms of treatment delivery. However, unlike tomotherapy, IMAT must account for restrictions on MLC movement as the gantry moves from one beam angle to the next. Because deliverability must take priority, an optimal field shape may have to be altered in order to have smooth delivery. As the result, plan quality should be adversely affected for some cases. This restriction does not apply to tomotherapy due to the use of a binary MLC. As the result,
6 tomotherapy should theoretically have the best plan quality. As compared with tomotherapy, IMAT also has some advantages: (1) IMAT does not need to move the patient during treatment and avoids abutment issues seen with serial tomotherapy; (2) IMAT retains the ability of using non-coplanar beams and arcs, which has great value for brain and head/neck tumors; (3) IMAT uses a conventional linac, thus complex rotational IMRT treatments and simple palliative treatments can be delivered with the same treatment unit. 2. Continued Efforts in Planning and Clinical Implementation The wide availability of linear accelerators and the advantages if IMAT propelled continued progress of the IMAT technique. In 2000, a Phase I clinical trial of IMAT using forward planning was conducted at the University of Maryland Error! Reference source not found., to test the safety and feasibility of changing the field shape during gantry rotation. Fifty patients with cancers of various sites were treated using IMAT. Due to the lack of an IMAT inverse planning system, forward planning was used for determining the arc range and aperture shapes. Arcs were approximated as multiple shaped fields spaced every 5 to 10 degrees around the patient. Multiple coplanar or non-coplanar arcs were allowed. At each beam angle, irregular field shapes were defined based on the beams-eye-view (BEV) of the planning target volume and normal critical structures. Typically, at a given angle, one field shape conformed to the beam seye-view of the target, and additional field shapes were set to shield individual critical structures that overlap the target in the BEV. Because beam rotation and irradiation are concurrent, the delivery is very efficient. A typical treatment with 3 to 5 arcs takes less than 10 minutes from start to finish, comparable to conventional techniques. As is the case with most forward planning techniques, these approaches rely heavily on the experience of the planner and could result in sub-optimal plans and prolonged planning times. Another clinical study was conducted at Ghent University Hospital. The aperture shapes were first determined based on the BEV of the target and critical structures similar to the
7 approach adopted by University of Maryland. However, the anatomy based apertures were further refined by allowing the leaves to move slightly using a greedy search optimization scheme (DeGersem 2004). Treatment planning studies were published for rectal cancer and whole abdominopelvic radiation therapy [17-18]. Wong et al formulated the forward planning into a practical approach which they termed Simplified Intensity Modulated Arc Therapy (SIMAT) [19]. SIMAT starts by creating multiple arcs based on the BEV of the anatomy with each arc serving a distinct planning goal, such as covering the whole target, or protecting one critical structure. The weightings of the arcs were subsequently optimized assuming a constant dose rate delivery. The SIMAT strategy was applied to various sites including prostate and highrisk endometrial cancer [20-21]. Cortutz et al also demonstrated that IMAT strategy could be used to treat a target that wrapped around a critical structure [22]. Under the assumption that arcs can only be delivered with constant dose rates, the author also proposed hybrid IMRT and IMAT to purther increase efficiency [] 3, Works Led to Single Arc IMAT In proposing IMAT as an alternative to tomotherapy, Yu predicted that with the increase in the number of gantry angles, the number of intensity levels at each gantry angle can be reduced without degrading plan quality [14]. He argued that the plan quality is a function of the total number of quanta defined as the product of the number of beam angles and the number of intensity levels. In another word, it is the total number of aperture shape variations that determines the plan quality. If this is true, a single arc with sufficient number of aperture shape variations would be able to create the optimal treatment plans.
8 Many works that followed have attempted to use a single arc for IMAT. Because linear accelerators at the time cannot vary the dose rate dynamicaaly during gantry rotation. Most previous works on single arc IMAT was under the assumption that the machine dose rate has to be constant during arc rotation. As illustrated by Jiang et al [], a single arc with 36 beam aperture variations under constant dose rate cannot realize the optimal plan quality. To reach the desired plan quality, one must increase the number of field segments or apertures, or allow the dose rate to vary during gantry rotation, or both. MacKenzie et al proposed a technique whereby 24 equally spaced beam orientations are optimized for sliding window IMRT and arc delivery is performed by allowing the gantry of the linear accelerator to rotate to static gantry orientations and deliver the optimized sliding window IMRT deliveries [27]. Crooks et al developed an IMAT planning algorithm that is based on the observation that the dose error for beam apertures being off by a small angular deviation is very small [28]. In their algorithm, IMRT fields were created approximately 30 apart with segments per beam direction. The segments were spread out with the previous observation regarding angular deviations and the plan was simulated and delivered in a single arc with deviations from the original IMRT plan on the order of 5-10%. Cameron developed a technique that uses a sweeping window arc therapy (SWAT) technique that initialized the collimator to 90 so that the leaf positions are normal to the axial plane of the patient [29]. Shapes of the MLC apertures prior to optimization are initialized so that the MLC leaf positions sweep across the PTV with rotation. Optimization of MLC leaf positions is then performed by simulated annealing and arc weight optimization that can be performed for a constant or variable angular dose rate.
9 Ulrich et al developed an optimization technique whereby arc therapy plans are optimized for a single arc delivery [30]. In their algorithm, arc shapes are optimized by a tabu search optimization algorithm and arc weights are optimized by a gradient search. The algorithm demonstrates better treatment plans than an in-house IMRT optimization technique and requires a variable dose rate delivery with gantry rotation. By assuming that the machine dose rate can vary as needed, Otto developed a single arc IMAT algorithm that he referred to as Volumetric Modulated Arc Therapy (VMAT) [31]. In addition to allowing dose rate variation, VMAT algorithm uses progressive beam angle sampling to optimize large number (> 100) of apertures using direct aperture optimization. The apertures shapes and weights are optimized initially for a number of coarsely spaced gantry angles with little consideration of aperture connectivity. Once the solution converges, additional gantry angles are inserted. As the angular spacing becomes smaller, the optimizer considers aperture shape connectivity both in the initialization of aperture shapes and during the optimization. The shapes of the newly inserted apertures are linearly interpolated from their angular neighbors. Such coarse-tofine sampling has been termed progressive sampling, and allows the optimization to converge faster. Because the aperture shape connectivity is ignored initially, the optimizer is give the freedom to aim for optimal dose distribution. Because the final plan ensures aperture connectivity, the optimized single arc can be delivered within 2 minutes. 4. Commercial Development and nomenclatures Although IMAT has been proposed since 1995 and many researchers have developed different planning methods and demonstrate that IMAT is capable of creating highly
10 conformal treatment plans that can be efficient delivered, large scale clinical implementation did not start until Varian adopted Otto s VMAT algorithm and marketed with the trade name of RapidArc TM in The linac control also updated to allow the dose rate variation during gantry rotation. Not long after Varian s announcement, Elekta also started to market their IMAT solution with the trade name VMAT TM. Recently, Philips Medical Systems, Inc. has announced plans to provide a rotational IMRT solution with the trade name of SmartArc TM. In describing their two-step planning method for single arc IMAT, Wang et al named their method as arc-modulated radiation therapy or AMRT [32]. Since all these variations subscribe to the same principle of intensity modulated arc therapy, they are just different implementations of IMAT. Therefore, in this review, no different trade names will be used. Instead, the original acronym IMAT will be used throughout this paper. Conclusion Rotational IMRT using conventional linear accelerators has after 12 years since the initially proposed in Significant technological developments in treatment planning by numerous researchers and in linear accelerator design led to its eventual clinical adoption. With on-board CT capabilities now available with linear accelerators from all 3 major vendors, IMAT promises a great potential for widespread clinical application.. With on-board CT capabilities now available with linear accelerators from all 3 major vendors, IMAT promises a great potential for widespread clinical application. References 1. Takahashi S Conformation radiotherapy. Rotation techniques as applied to radiography and radiotherapy of cancer. Acta Radiol Diagn (Stockh). Suppl
11 2. Brahme A, Roos JE and Lax I Solution of an integral equation encountered in radiation therapy Phys. Med. Biol Chin LM, Kijewski PK, Svensson GK, Bjärngard BE.1983 Dose optimization with computer-controlled gantry rotation, collimator motion and dose-rate variation. Int J Radiat Oncol Biol Phys Perez CA, Purdy JA, Harms W, Gerber R, Graham MV, Matthews JW, Bosch W, Drzymala R, Emami B, Fox S, et al. Three-dimensional treatment planning and conformal radiation therapy: preliminary evaluation, Radiother Oncol (1): A Brahme, Optimization of stationary and moving beam radiation therapy techniques, Radiother Oncol. 12, (1988). 6. Convery DJ and Rosenbloom ME The generation of intensity-modulated fields for conformal radiotherapy by dynamic collimation Phys. Med. Biol Bortfeld T, Boyer A L, Schlegel W, Kahler D L and Waldron T J 1994a Realization and verification of three-dimensional conformal radiotherapy with modulated fields Int. J. Radiat. Oncol. Biol. Phys Bortfeld T, Kahler D L, Waldron T J and Boyer A L 1994b X-ray field compensation with multileaf collimators Int. J. Radiat. Oncol. Biol. Phys C.X. Yu, M.J. Symons, M.N. Du et al : A method for implementing dynamic photon beam intensity modulation using independent jaws and multileaf collimator, Phys. Med. Biol. 40, (1995). 10. Spirou S V and Chui C S 1994 Generation of arbitrary fluence profiles by dynamic jaws or multileaf collimators Med. Phys Stein J, Bortfeld T, Doerschel B and Schlegel W 1994 Dynamic x-ray compensation for conformal radiotherapy by means of multi-leaf collimation Radiother. Oncol Ling C C et al 1996 Conformal radiation treatment of prostate cancer using inversely-planned intensity-modulated photon beams produced with dynamic multileaf collimation Int. J. Radiat. Oncol. Biol. Phys
12 13. C. Burman, C. Chui, G. Kutcher, S. Leibel, M. Zelefsky, T. LoSasso, S. Spirou, Q. Wu, J. Yang, J. Stein, R. Mohan, Z. Fuks, C. Ling Planning, delivery, and quality assurance of intensity-modulated radiotherapy using dynamic multileaf collimator: a strategy for large-scale implementation for the treatment of carcinoma of the prostate, Int J Radiat Oncol Biol Phys., 39(4), (1997). 14. Mackie TR, Holmes T, Swerdloff S, Reckwerdt P, Deasy JO, Yang J, Paliwal B, Kinsella T.1993 Tomotherapy: a new concept for the delivery of conformal radiotherapy Med. Phys M. Carol Integrated 3D conformal planning/multivane intensity modulating delivery system for radiotherapy, in 3D Radiation Treatment Planning and Conformal Therapy, edited by J.A. Purdy and B. Emami (Medical Physics Publishing, Madison, WI), pp (1995) 16. Shepard D M, Olivera G, Angelos L, Sauer O, Reckwerdt P and Mackie T R 1999 A simple model for examining issues in radiotherapy optimization Med. Phys Yu CX 1995 Intensity-modulated arc therapy with dynamic multileaf collimation: an alternative to tomotherapy Phys. Med. Biol Yu CX, Li XA, Ma L, Chen D, Naqvi SA, Shepard DM, Sarfaraz M, Holmes TW, Suntharalingam M and Mansfield CM 2002 Clinical implementation of intensitymodulated arc therapy. Int J Radiat Oncol Biol Phys Claus F, De Gersem W, Vanhoutte I, Duthoy W, Remouchamps V, De Wagter C, De Neve W Evaluation of a leaf position optimization tool for intensity modulated radiation therapy of head and neck cancer. Radiother Oncol DeGersem W. Intensity Modulated Arc Therapy. Proc of ICCR, Duthoy W, De Gersem W, Vergote K, Coghe M, Boterberg T, De Deene Y, De Wagter C, Van Belle S, De Neve W Whole abdominopelvic radiotherapy (WAPRT) using intensity-modulated arc therapy (IMAT): first clinical experience. Int J Radiat Oncol Biol Phys Duthoy W, De Gersem W, Vergote K, Boterberg T, Derie C, Smeets P, De Wagter C, De Neve W 2004 Clinical implementation of intensity-modulated arc therapy (IMAT) for rectal cancer. Int J Radiat Oncol Biol Phys
13 23. Wong E, Chen JZ, Greenland J 2002 Intensity-modulated arc therapy simplified. Int J Radiat Oncol Biol Phys Bauman G, Gete E, Chen JZ and Wong E Simplified intensity-modulated arc therapy for dose escalated prostate cancer radiotherapy. Med Dosim Wong E, D'Souza DP, Chen JZ, Lock M, Rodrigues G, Coad T, Trenka K, Mulligan M, Bauman GS 2005 Intensity-modulated arc therapy for treatment of high-risk endometrial malignancies. Int J Radiat Oncol Biol Phys Cotrutz C, Kappas C, Webb S 2000 Intensity modulated arc therapy (IMAT) with centrally blocked rotational fields. Phys. Med. Biol G. Tang, M. Earl, S. Luan, S. Naqvi and C.X. Yu, Converting multiple-arc Intensitymodulated Arc Therapy into a single arc for efficient delivery, Int. J. Rad. Oncol. Biol. Phys 69(3,) Supplement, S673 (2007). 28. Ma L, Yu CX, Earl M, Holmes T, Sarfaraz M, Li XA, Shepard D, Amin P, DiBiase S, Suntharalingam M and Mansfield C Optimized intensitymodulated arc therapy for prostate cancer treatment. Int J Cancer Li XA, Ma L, Naqvi SA, Shih R, Yu CX 2004 Monte Carlo dose verification for intensity-modulated arc therapy. Phys. Med. Biol Tang G, Earl MA, Luan S, Wang C, Cao D, Yu CX, Naqvi SA 2008 Stochastic versus deterministic kernel-based superposition approaches for dose calculation of intensity-modulated arcs. Phys. Med. Biol MacKenzie MA, Robinson DM Intensity modulated arc deliveries approximated by a large number of fixed gantry position sliding window dynamic multileaf collimator fields. Med Phys Crooks SM, Wu X, Takita C, Watzich M, Xing L 2003 Aperture modulated arc therapy. Phys. Med. Biol Cameron C Sweeping-window arc therapy: an implementation of rotational IMRT with automatic beam-weight calculation. Phys Med Biol Ulrich S, Nill S, and Oelfke U 2007 Development of an optimization concept for arc-modulated cone beam therapy. Phys. Med. Biol Otto K 2008 Volumetric modulated arc therapy: IMRT in a single gantry arc. Med Phys Wang C, Luan S, Tang G, Chen DZ, Earl MA, Yu CX 2008 Arc-Modulated Radiation Therapy (AMRT): A Single-Arc Form of Intensity-Modulated Arc Therapy. Phys. Med. Biol
14 37. Earl MA, Shepard DM, Naqvi S, Li XA, and Yu CX, 2003 Inverse planning for intensity-modulated arc therapy using direct aperture optimization Phys. Med. Biol. 48, Shepard DM, Cao D, Afghan MK, Earl MA 2007 An arc-sequencing algorithm for intensity modulated arc therapy. Med Phys Gladwish A, Oliver M, Craig J, Chen J, Bauman G, Fisher B and Wong E Segmentation and leaf sequencing for intensity modulated arc therapy. Med Phys Luan S, Wang C, Cao D, Chen DZ, Shepard DM, Yu CX Leaf-sequencing for intensity-modulated arc therapy using graph algorithms. Med Phys Cao D, Earl MA, Luan S, Shepard DM 2006 Continuous intensity map optimization (CIMO): a novel approach to leaf sequencing in step and shoot IMRT. Med Phys Shepard DM, Earl MA, Li XA, Naqvi SA and Yu CX 2002 Direct aperture optimization: a turnkey solution for step-and-shoot IMRT. Med Phys Tang G, Earl MA, Luan S, Naqvi SA and Yu CX 2007 Converting multiple-arc Intensity-modulated Arc Therapy into a single arc for efficient delivery. Int. J. Rad. Oncol. Biol. Phys 69 S673.
VMAT monthly QA using two techniques: 2D ion chamber array with an isocentric gantry mount and an in vivo dosimetric device attached to gantry
Journal of Radiotherapy in Practice Journal of Radiotherapy in Practice (2013) Page 1 of 7 & Cambridge University Press 2013 doi:10.1017/s1460396912000556 Technical Note VMAT monthly QA using two techniques:
More informationIMRT and Inverse Planning: From the Art to the State-of-the-Art
IMRT and Inverse Planning: From the Art to the State-of-the-Art Thomas Bortfeld Massachusetts General Hospital Department of Radiation Oncology TBortfeld@mgh.harvard.edu Evolution of the art of radiation
More informationVMAT Treatment Planning
Swedish Medical Center VMAT Treatment Planning David Shepard Swedish Cancer Institute Seattle, WA Founded in 1910 by Dr. Nils Johanson and a group of Seattle s leading Swedish-born businessmen. We treat
More informationElekta VMAT. Freedom to Treat. Enhanced dose conformance for patient-specific treatment
Elekta VMAT Freedom to Treat Enhanced dose conformance for patient-specific treatment Speed or Accuracy Why choose? Clinical data has shown that around 50% of patients would benefit from IMRT treatment
More informationFigure 1: Radiation beams through a multi-leaf collimator(mlc)
Sabbatical Proposal Athula Gunawardena Department of Mathematical and Computer Sciences Algorithms and Treatment Quality Comparisons in Intensity Modulated Radiation Therapy 1. Objectives and Project Plan
More informationThe Optimization of Inverse Planning and IMRT
The Optimization of Inverse Planning and IMRT James M. Galvin, D.Sc. Ying Xiao, Ph.D. Yan Chen, Ph.D. Greg Bednarz, Ph.D. Darek Michalski, Ph.D. Yair Censor, Ph.D. Chris Houser, B.S. Murshed Hossain, PhD
More informationVOLUMETRIC MODULATED ARC THERAPY (VMAT): ADVANCED DELIVERY TECHNIQUES FOR STATIC AND MOVING TARGETS. A Dissertation. presented to
VOLUMETRIC MODULATED ARC THERAPY (VMAT): ADVANCED DELIVERY TECHNIQUES FOR STATIC AND MOVING TARGETS A Dissertation presented to the Faculty of the Graduate School at the University of Missouri-Columbia
More informationEvaluation of Dynamic Conformal Arc Therapy for Treatment of Lung and Liver
Evaluation of Dynamic Conformal Arc Therapy for Treatment of Lung and Liver Sotiri Stathakis, PhD, DABR Disclosure ELEKTA Monaco research support I do not have a financial interest in any company or product
More informationCommissioning of a commercially available system for intensitymodulated radiotherapy dose delivery with dynamic multileaf collimation
Radiotherapy and Oncology 60 (2001) 215±224 www.elsevier.com/locate/radonline Commissioning of a commercially available system for intensitymodulated radiotherapy dose delivery with dynamic multileaf collimation
More informationComprehensive study of parameters for volumetric modulated arc therapy (VMAT) treatment planning
Louisiana State University LSU Digital Commons LSU Master's Theses Graduate School 2010 Comprehensive study of parameters for volumetric modulated arc therapy (VMAT) treatment planning Catharine Elizabeth
More informationIncorporating Prior Knowledge into IMRT Beam Orientation Optimization
Incorporating Prior Knowledge into IMRT Beam Orientation Optimization Lei Xing, Ph.D. and Andrei Pugachev, M.S. Department of Radiation Oncology Stanford University School of Medicine Stanford, CA 94305-5304
More informationShielding evaluation for IMRT implementation in an existing accelerator vault
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 3, SUMMER 2003 Shielding evaluation for IMRT implementation in an existing accelerator vault R. A. Price,* O. Chibani, and C.-M. Ma Department
More informationStudy of 2D ion chamber array for angular response and QA of dynamic MLC and pretreatment
Anup Kumar Bhardwaj et al Study of 2D ion chamber array for angular response and QA of dynamic MLC... Study of 2D ion chamber array for angular response and QA of dynamic MLC and pretreatment IMRT plans
More informationIntensity Modulated Radiation Therapy: The good, the bad, and the misconceptions. Indications for IMRT. Indications for IMRT Billing
1 Intensity Modulated Radiation Therapy: The good, the bad, and the misconceptions. J. M. Galvin, D.Sc. Department of Radiation Oncology Jefferson Medical College, Thomas Jefferson University Philadelphia,
More informationNote RADIATION QUALITY OF A TOMOTHERAPY PHOTON FAN BEAM
Note RADIATION QUALITY OF A TOMOTHERAPY PHOTON FAN BEAM Abstract Tomotherapy, a novel radiotherapy technique, uses narrow fan beams for cancer patient treatment. Photon energy spectra for a rectangular
More informationSampling Issues for Optimization in Radiotherapy. Michael C. Ferris R. Einarsson Z. Jiang D. Shepard
Sampling Issues for Optimization in Radiotherapy Michael C. Ferris R. Einarsson Z. Jiang D. Shepard Radiation Treatment Planning Cancer is the 2nd leading cause of death in U.S. Only heart disease kills
More informationData Mining to Aid Beam Angle Selection for IMRT
Data Mining to Aid Beam Angle Selection for IMRT Stuart Price-University of Maryland Howard Zhang- University of Maryland School of Medicine Bruce Golden- University of Maryland Edward Wasil- American
More informationIMRT: delivery techniques and quality assurance
The British Journal of Radiology, 76 (2003), 766 776 DOI: 10.1259/bjr/12907222 E 2003 The British Institute of Radiology Review article IMRT: delivery techniques and quality assurance P C WILLIAMS, BSc,
More informationRapidArc: Clinical Implementation
RapidArc: Clinical Implementation Fang-Fang Yin, PhD Q. Jackie Wu, PhD Acknowledgements Team efforts from staff at Duke Radiation Oncology, especially to Dr. J Chang, Dr. J O Daniel for providing slide
More information8/17/2011. Implementation and Quality Assurance Considerations for Volumetric Modulated Arc Therapy (VMAT)*
and Quality Assurance Considerations for Volumetric Modulated Arc Therapy * What is? Report of writing group from TETAWG under Therapy Physics Committee (TPC) of AAPM Writing Group members: James M. Galvin,
More informationComparison of the KonRad IMRT and XiO treatment planning systems
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 9, NUMBER 3, SUMMER 2008 Comparison of the KonRad IMRT and XiO treatment planning systems Bodo Reitz, 1, 2,a and Moyed Miften 1, 2 Department of Radiation
More informationROLE OF BEAM ORIENTATION OPTIMIZATION IN INTENSITY- MODULATED RADIATION THERAPY
PII S0360-3016(01)01502-4 Int. J. Radiation Oncology Biol. Phys., Vol. 50, No. 2, pp. 551 560, 2001 Copyright 2001 Elsevier Science Inc. Printed in the USA. All rights reserved 0360-3016/01/$ see front
More informationElekta VMAT. Choose freedom
Elekta VMAT Choose freedom Freedom... Elekta VMAT is an advanced radiotherapy solution which has been developed to combine dramatically reduced treatment times with a removal of traditional restrictions
More informationGuidance document on Delivery, Treatment Planning, and Clinical Implementation of IMRT
Guidance document on Delivery, Treatment Planning, and Clinical Implementation of IMRT IMRT subcommittee of the Radiation Therapy Committee Contents 1.0 Introduction 2.0 Delivery systems for IMRT 3.0 Treatment
More informationEffects of changing modulation and pitch parameters on tomotherapy delivery quality assurance plans
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 5, 2015 Effects of changing modulation and pitch parameters on tomotherapy delivery quality assurance plans Diana Binny, 1a Craig M. Lancaster,
More informationRESHAPING RADIATION THERAPY AN INTRODUCTION TO TOMOTHERAPY
RESHAPING RADIATION THERAPY AN INTRODUCTION TO TOMOTHERAPY RESHAPING RADIATION THERAPY AN INTRODUCTION TO TOMOTHERAPY SIMPLIFYING CLINICAL WORKFLOW ATTRACTING AND RETAINING PATIENTS ESTABLISHING A CENTER
More informationClinical Evaluation of real-time, patient-centric QA using the Delta 4 Discover
Clinical Evaluation of real-time, patient-centric QA using the Delta 4 Discover Karl H Rasmussen, Ph.D.. khrasmussen@uthscsa.edu Nema Bassiri bassirin@livemail.uthscsa.edu Division of Medical Physics Cancer
More informationA comparison of two commercial treatment-planning systems for IMRT
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 6, NUMBER 3, SUMMER 2005 A comparison of two commercial treatment-planning systems for IMRT M. Peter Petric, 1 Brenda G. Clark, 1 and James L. Robar
More informationTHE EFFECTIVE APPLICATION OF DYNAMIC ARC RADIOTHERAPY. A thesis submitted to The University of Manchester for the degree of. PhD
THE EFFECTIVE APPLICATION OF DYNAMIC ARC RADIOTHERAPY A thesis submitted to The University of Manchester for the degree of PhD In the Faculty of Medical and Human Sciences 2013 CHRISTOPHER JAMES BOYLAN
More informationThis is probably the kind of radiotherapy that you are used to delivering in your country.
This is probably the kind of radiotherapy that you are used to delivering in your country. This slide shows images two patients, one with a T3 N2c nasopharyngeal CA and the other with a T1/2 N1/2 lung
More informationFusionArc optimization: A hybrid volumetric modulated arc therapy (VMAT) and intensity modulated radiation therapy (IMRT) planning strategy
optimization: A hybrid volumetric modulated arc therapy (VMAT) and intensity modulated radiation therapy (IMRT) planning strategy Martha M. Matuszak a) Department of Radiation Oncology, University of Michigan,
More informationImpact of patient rotational errors on target and critical structure dose in IMRT: A 3D simulation study
University of Wollongong Research Online Faculty of Engineering and Information Sciences - Papers: Part A Faculty of Engineering and Information Sciences 2013 Impact of patient rotational errors on target
More informationTransitioning to PBS Proton Therapy From IMRT or Passive Scattered Based Proton Therapy
Transitioning to PBS Proton Therapy From IMRT or Passive Scattered Based Proton Therapy Mark Pankuch, PhD Northwestern Medicine Chicago Proton Center July 31, 2017 Progression to better technology 1 Adoption
More informationSoftware for plan quality control
Software for plan quality control D La Russa *1, R Taylor 1, E Henderson 1, K Lekx 1, D Mason 2, (1) The Ottawa Hospital Cancer Centre, Ottawa, ON, (2) R. S. McLaughlin Durham Regional Cancer Centre, Oshawa,
More informationOn the geud biological optimization objective for organs at risk in Photon Optimizer of Eclipse treatment planning system
Received: 12 May 2017 Revised: 13 October 2017 Accepted: 16 October 2017 DOI: 10.1002/acm2.12224 RADIATION ONCOLOGY PHYSICS On the geud biological optimization objective for organs at risk in Photon Optimizer
More informationArcCHECK & 3DVH. The Ultimate 4D Patient QA Solution. Your Most Valuable QA and Dosimetry Tools
ArcCHECK & 3DVH The Ultimate 4D Patient QA Solution Your Most Valuable QA and Dosimetry Tools THE 4D QA GOLD STANDARD The ArcCHECK is the world s most selected independent 4D measurement array. Simply
More informationCharacterization of interplay errors in step-and-shoot IMRT of the lung
Characterization of interplay errors in step-and-shoot IMRT of the lung Travis J. McCaw Under the supervision of Dr. Larry DeWerd Medical Radiation Research Center Department of Medical Physics University
More informationTitle. CitationPhysica Medica, 30(5): Issue Date Doc URL. Type. File Information.
Title Preliminary analysis for integration of spot-scannin Author(s)Shimizu, S.; Matsuura, T.; Umezawa, M.; Hiramoto, K. CitationPhysica Medica, 30(5): 555-558 Issue Date 2014-07 Doc URL http://hdl.handle.net/2115/56596
More informationOptimal Use of Emerging g Technologies
Optimal Use of Emerging g Technologies Radhe Mohan, PhD Professor and McNeil Endowed Chair Department of Radiation Physics 1 Optimal Use of Emerging Technologies A perspective Complexity and sophistication
More informationRadiation Oncology. Open Access. Abstract. BioMed Central
Radiation Oncology BioMed Central Methodology The GLAaS algorithm for portal dosimetry and quality assurance of RapidArc, an intensity modulated rotational therapy Giorgia Nicolini 1, Eugenio Vanetti 1,
More informationAutomated IMRT planning with regional optimization using planning scripts
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 14, NUMBER 1, 2013 Automated IMRT planning with regional optimization using planning scripts Ilma Xhaferllari, 1,3a Eugene Wong, 1,2,5 Karl Bzdusek,
More informationPreliminary Study for Dosimetric Characteristics of 3Dprinted. Materials with Megavoltage Photons
Preliminary Study for Dosimetric Characteristics of 3Dprinted Materials with Megavoltage Photons Seonghoon Jeong and Myonggeun Yoon Department of Bio-Convergence Engineering, Korea University, Seoul, Korea
More informationChristie Medical Physics and Engineering (CMPE), The Christie NHS Foundation Trust, Wilmslow Road, Manchester M20 4BX, UK
A megavoltage scatter correction technique for cone-beam CT images acquired during VMAT delivery C. J. Boylan 1,2, T. E. Marchant 1,2, J. Stratford 3, J. Malik 4, A. Choudhury 4, R. Shrimali 4, J. Rodgers
More informationClinical trials with hypo-fractionation for breast conserving therapy (BCT): Shorter overall treatment time Effective radiation treatment
AAMD Regional Meeting, March 21st 2015 Clinical trials with hypo-fractionation for breast conserving therapy (BCT): Shorter overall treatment time Effective radiation treatment Whelan et al 2013 & Yu et
More informationAn improvement in IMRT QA results and beam matching in linacs using statistical process control
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 5, 2014 An improvement in IMRT QA results and beam matching in linacs using statistical process control Justin D. Gagneur a and Gary A. Ezzell
More informationRayStation Automation is the Future of Planning. Carmen Sawyers, MS, DABR Ackerman Cancer Center
RayStation Automation is the Future of Planning Carmen Sawyers, MS, DABR Ackerman Cancer Center Overview Why Automation? o RayStation Automation Tools A Different Approach o RayStation Advanced Automation
More informationRapidArc Inverse Planning, Dose Calculation and Clinical Application
RapidArc Inverse Planning, Dose Calculation and Clinical Application A thesis submitted in partial fulfilment of the requirements for the Master s Degree in Medical Physics by David Jolly Department of
More informationIMAGE GUIDANCE DOSES IN RADIOTHERAPY. Scott Crowe
IMAGE GUIDANCE DOSES IN RADIOTHERAPY Scott Crowe IMAGING DOSE ALARA requires that imaging dose is managed and optimised Imaging dose presents an increased risk for paediatric patients Increased effective
More informationIMRT: Progress in Technology and Reimbursement
Article reprinted from Radiology Management November/December 2001. IMRT: Progress in Technology and Reimbursement by Ralph Young, M.S.; Bette Snyder, M.B.A. Images provided by Varian Medical Systems Copyright
More informationImaging/Imagine Needs for Proton Therapy: Treatment Planning. Lei Dong, Ph.D. Scripps Proton Therapy Center San Diego, CA
Imaging/Imagine Needs for Proton Therapy: Treatment Planning Lei Dong, Ph.D. Scripps Proton Therapy Center San Diego, CA AAPM Annual Meeting Indianapolis, Aug. 07, 2013 Disclosure Software licensing agreement
More informationPractical Workflow and the Cost of Adaptive Therapy. Rojano Kashani, Ph.D., DABR Washington University School of Medicine March 7, 2015
Practical Workflow and the Cost of Adaptive Therapy Rojano Kashani, Ph.D., DABR Washington University School of Medicine March 7, 2015 Disclosures Travel expenses paid by ViewRay Inc. for on-site software
More informationClinical Implementation of the PEREGRINE Monte Carlo Dose Calculations System for Photon Beam Therapy
UCRL-JC-133076 PREPRINT Clinical Implementation of the PEREGRINE Monte Carlo Dose Calculations System for Photon Beam Therapy C.L. Hartmann Siantar, P.M. Bergstrom T.P. Daly, M. Descalle D. Garrett, R.K.
More informationAssessment of the modulation degrees of intensity-modulated radiation therapy plans
Park et al. Radiation Oncology (2018) 13:244 https://doi.org/10.1186/s13014-018-1193-9 RESEARCH Assessment of the modulation degrees of intensity-modulated radiation therapy plans So-Yeon Park 1,2, Jung-in
More informationN.MAFFEI, G.GUIDI, C.VECCHI, G.BALDAZZI Physics Department, University of Bologna, via Irnerio Bologna, Italy
AN ARTIFICIAL NEURAL NETWORK TO PREDICT TIME OF REPLANNING FOR TOMOTHERAPY TREATMENTS N.MAFFEI, G.GUIDI, C.VECCHI, G.BALDAZZI Physics Department, University of Bologna, via Irnerio 40 40138 Bologna, Italy
More informationFocusing Energy on Saving Lives. Year End Review Investor Relations Meeting ASTRO Oct. 30, 2007
Focusing Energy on Saving Lives Year End Review Investor Relations Meeting ASTRO Oct. 30, 2007 ASTRO IR Day & FY2007 Year-End Review Tim Guertin President and CEO ASTRO IR Day Oct. 30, 2007 Forward-Looking
More informationImproving IMRT delivery efficiency using intensity limits during inverse planning
Improving IMRT delivery efficiency using intensity limits during inverse planning Martha M. Coselmon, a Jean M. Moran, Jeffrey D. Radawski, and Benedick A. Fraass Department of Radiation Oncology, University
More informationViewRay Experience. Sasa Mutic, Ph.D. Washington University School of Medicine
ViewRay Experience Sasa Mutic, Ph.D. Washington University School of Medicine You can observe a lot, just by watching -Yogi Berra Does organ motion matter? A slide often used to illustrate RT capabilities
More informationRT Workspace from Standard Imaging
RT Workspace from Standard Imaging In 1895 Wilhelm Röntgen discovers the X-ray (receives Nobel prize in 1901) From 1900 1950 radium was chiefly used for radiation treatment (based on work of Marie Curie
More informationAn updates on ART for prostate, pancreas and breast
An updates on ART for prostate, pancreas and breast X. Allen Li Medical College of Wisconsin TH-E-137, AAPM Edu Course, Aug. 8, 2013 Acknowledgement Ergun Ahunbay, Ph.D Guangpei Chen, Ph.D Feng Liu, Ph.D
More informationDose Prediction with U-net: A Feasibility Study for Predicting Dose Distributions from Contours using Deep Learning on Prostate IMRT Patients
Dose Prediction with U-net: A Feasibility Study for Predicting Dose Distributions from Contours using Deep Learning on Prostate IMRT Patients Dan Nguyen, Troy Long, Xun Jia, Weiguo Lu, Xuejun Gu, Zohaib
More informationThe impact of plan complexity on the accuracy of VMAT for the treatment of head and neck cancer
The impact of plan complexity on the accuracy of VMAT for the treatment of head and neck cancer A thesis submitted in partial fulfilment of the requirements for the Degree of Master of Science in Medical
More informationAbstract Background: Aim: Materials & Methods Results: & Discussion: Conclusions:
Abstract Background: Radiation therapy is an important modality in the treatment of cancer. In the past decades radiation therapy has been developing towards more dynamic delivery techniques and Intensity-Modulated
More informationIMRT AND VMAT: CURRENT AND FUTURE BEST PRACTICES (OR: DON T FORGET ABOUT IMRT AND HOW TO SETUP STATIC FIELDS)
IMRT AND VMAT: CURRENT AND FUTURE BEST PRACTICES (OR: DON T FORGET ABOUT IMRT AND HOW TO SETUP STATIC FIELDS) Anthony Magliari MS, CMD DISCLAIMER: I am employed by Varian Medical Systems on the Medical
More informationImpact of very long time output variation in the treatment of total marrow irradiation with helical tomotherapy
Takahashi and Hui Radiation Oncology 2013, 8:123 RESEARCH Open Access Impact of very long time output variation in the treatment of total marrow irradiation with helical tomotherapy Yutaka Takahashi 1*
More information} } 8/3/2016. Accounting for kv Imaging Dose. kilovoltage imaging devices/techniques. Current imaging dose determination methods
Accounting for kv Imaging Dose Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota 1 2016 AAPM Annual Meeting - SAM Therapy Educational Course, TU-B-201-2 Washington, DC, August
More informationClinical Experience with Philips Auto-Planning
Christian Rønn Hansen Odense University Hospital. Laboratory of Radiation Physics. Odense. Denmark. University of Southern Denmark. Institute of Clinical Research. Odense. Denmark. School of Physics -
More informationPinnacle 3. Treatment Planning. Flexible, intuitive planning environment
Pinnacle 3 Treatment Planning Flexible, intuitive planning environment Pinnacle 3 Flexible, intuitive planning environment Planning plays a central role in influencing treatment confidence as well as departmental
More informationbiij Dosimetric evaluation of radiation dose rate effect in respiratory gated intensity modulated radiation therapy
Available online at http://www.biij.org/2012/1/e5 doi: 10.2349/biij.8.1.e5 biij Biomedical Imaging and Intervention Journal ORIGINAL ARTICLE Dosimetric evaluation of radiation dose rate effect in respiratory
More informationAutoplanning in the IGRT era. Ben Heijmen
Autoplanning in the IGRT era Ben Heijmen ICARO 2, Vienna, 2017 Disclosures Work in part funded by research grants from Elekta AB, and Accuray Inc. Elekta AB is preparing commercialization of the Erasmus-iCycle
More informationIntroduction to Radiation Therapy : Basics, and Recent Technologies
Introduction to Radiation Therapy : Basics, and Recent Technologies Pusan National University Undergraduate Student Seminar Justin. C. Park (Chun Joo Park) Dept. of Electrical and Computer Engineering
More informationTechnical White Paper: An Educational Report On Helical IMRT EXECUTIVE SUMMARY. CRCPD Publication #E-11-3
CRCPD Publication #E-11-3 Technical White Paper: An Educational Report On Helical IMRT Developed by CRCPD s H-34 Task Force for Helical IMRT EXECUTIVE SUMMARY Helical IMRT (Intensity Modulated Radiation
More informationAdapting VMAT plans optimized for an HD120 MLC for delivery with a Millennium MLC
Received: 31 January 2017 Revised: 15 May 2017 Accepted: 5 June 2017 DOI: 10.1002/acm2.12134 RADIATION ONCOLOGY PHYSICS Adapting VMAT plans optimized for an HD120 MLC for delivery with a Millennium MLC
More informationDisclosures I have provided Physics Consultation Services to Varian and Vision RT
Surface Imaging to Inform Statistical Quality Control for SRS Treatments Michael J Tallhamer M.Sc. DABR Chief of Radiation Physics Centura Health MichaelTallhamer@Centura.org Quality Improvement and Safety
More informationGoing Clinical with the Accuray Radixact System. The Montefiore Experience
Going Clinical with the Accuray Radixact System The Montefiore Experience Disclaimer The views expressed in the presentation are those of the presenters and do not necessarily reflect the views or policies
More informationClinical feasibility of dual tube floor based kilo voltage X-ray localization system for image guided radiation therapy
2018; 4(3): 384-391 ISSN Print: 2394-7500 ISSN Online: 2394-5869 Impact Factor: 5.2 IJAR 2018; 4(3): 384-391 www.allresearchjournal.com Received: 26-01-2018 Accepted: 27-02-2018 Suresh Rao Department of
More informationDESIGN OF AN ANTHROPOMORPHIC INTENSITY MODULATED RADIATION THERAPY QUALITY ASSURANCE PHANTOM
DESIGN OF AN ANTHROPOMORPHIC INTENSITY MODULATED RADIATION THERAPY QUALITY ASSURANCE PHANTOM Dee-Ann Radford, David S. Followill, Peter A. Balter, William F. Hanson Department of Radiation Physics The
More informationTo the Graduate Council:
To the Graduate Council: I am submitting herewith a thesis written by Michelle Lynn Neeley entitled Patient- Specific Quality Assurance for IMRT. I have examined the final electronic copy of this thesis
More informationDisclosures 7/31/2017. Clinical Impact and Applications of 4D Imaging (in RT)
Clinical Impact and Applications of 4D Imaging (in RT) Geoff Hugo, Ph.D. Virginia Commonwealth University Washington University School of Medicine gdhugo@wustl.edu Disclosures Employee of Virginia Commonwealth
More informationIntegrating the MR-Linac into Radiation Therapy Practice
Integrating the MR-Linac into Radiation Therapy Practice Mikki Campbell MRT(T), MHE Winnie Li MRT(T), MSc UTDRO Evening Journal Club MR-Linac: From Prototype to Clinical January 24, 2019 Objectives To
More informationCompensator-intensity-modulated Radiotherapy A Traditional Tool for Modern Application
Radiotherapy & Imaging Compensator-intensity-modulated Radiotherapy A Traditional Tool for Modern Application a report by Sha Chang Associate Professor and Chief of Physics & Computing, Department of Radiation
More informationimportant role in achieving optimal dose distribution for patients. Geometrical
White Paper Agility * Intelligent Design Kevin Brown, Global Vice President Scientific Research Giulia Thompson, Senior Research Physicist Elekta, Ltd Abstract Multileaf collimators (MLCs) have a variety
More informationPinnacle 3. Treatment Planning. Flexible, intuitive planning environment
Pinnacle 3 Treatment Planning Flexible, intuitive planning environment Pinnacle 3 Flexible, intuitive planning environment Planning plays a central role in influencing treatment confidence as well as departmental
More informationIMRT Planning: Concepts and Recommendations of the ICRU report n. 83
School on Medical Physics for Radiation Therapy: Dosimetry and Treatment Planning for Basic and Advanced Applications Trieste - Italy, 27 March - 7 April 2017 IMRT Planning: Concepts and Recommendations
More informationRPC Prostate Phantom. Guidelines for Planning and Treating the RPC IMRT Prostate Phantom. Revised November 2010
RPC Prostate Phantom Guidelines for Planning and Treating the RPC IMRT Prostate Phantom. Revised November 2010 Credentialing for this protocol requires three steps: (1) submission of the Facility Questionaire
More informationAutomated population-based planning for whole brain radiation therapy
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 5, 2015 Automated population-based planning for whole brain radiation therapy Eduard Schreibmann, a Tim Fox, Walter Curran, Hui-Kuo Shu, Ian
More informationDigital Imaging and Communications in Medicine (DICOM) Supplement 175: Second Generation Radiotherapy. C-Arm RT Treatment Modalities
sup175_dlb_10_carmradiations.doc: Sup 175: 2 nd Generation RT C-Arm RT Treatment Modalities Page 1 2 4 Digital Imaging and Communications in Medicine (DICOM) 6 Supplement 175: Second Generation Radiotherapy
More informationAcknowledgements. Obstacles and advances in IMRT treatment planning. Outline. Where we need to go: IMRT software design goals.
Obstacles and advances in IMRT treatment planning Joe Deasy, PhD Acknowledgements Eva Lee, Ga Tech/Emory Jong Lee, Univ Mich Jim Dempsey, U. Florida Thomas Bortfeld, MGH James Alaly, WUSTL Konstantin Zakarian,
More informationINVERSE PLANUNG BEI 3 VERSCHIEDENEN TPS
Division Medical Radiation Physics Dept. of Radiotherapy & Radiobiology Medical University of Vienna INVERSE PLANUNG BEI 3 VERSCHIEDENEN TPS DER VERSUCH EINES VERGLEICHS D. Georg, B. Kroupa, K. Dieckmann,
More informationVarian MCO Benefits in Head and Neck Planning
Varian MCO Benefits in Head and Neck Planning Dominic DiCostanzo, MS, DABR November 3, 2017 The Ohio State University Comprehensive Cancer Center Arthur G. James Cancer Hospital and Richard J. Solove Research
More informationDoes gated beam delivery impact delivery accuracy on an Elekta linac?
Received: 7 March 2016 Accepted: 4 October 2016 DOI: 10.1002/acm2.12020 RADIATION ONCOLOGY PHYSICS Does gated beam delivery impact delivery accuracy on an Elekta linac? Mohammed Jermoumi 1 Roger Xie 2
More informationNathan Childress, Ph.D., DABR
Nathan Childress, Ph.D., DABR Financial Disclaimer I own Mobius Medical Systems, LP, manufacturer of Mobius3D, MobiusFX, and DoseLab I used Mobius3D to generate results for this talk Introduction Dosimetrists
More informationTreatment Quality Assurance Cone Beam Image Guided Radiation Therapy. Jean-Pierre Bissonnette, PhD, MCCPM
Treatment Quality Assurance Cone Beam Image Guided Radiation Therapy Jean-Pierre Bissonnette, PhD, MCCPM Disclosure Work supported, in part, by Elekta Oncology Systems Commercial Interest in Penta-Guide
More informationDevelopmentofanovel treatment planning test for credentialing rotational intensity-modulated radiotherapy techniques in the UK
ARTICLE INFORMATION 13 The Authors Received: 13 June 12 Revised: 29 October 12 Accepted: 12 November 12 doi: 10.1259/bjr.1315 Cite this article as: Tsang Y, Ciurlionis L, Clark C, Venables K, on behalf
More informationComputational Verification in Interventional Radiation Oncology (Brachytherapy)
Computational Verification in Interventional Radiation Oncology (Brachytherapy) Prof. Dr. Dimos Baltas Abteilung Medizinische Physik & Engineering Strahlenklinik Brussels, 5- December- 2014 Starkenburgring
More informationMV Cone Beam CT Imaging for daily localiza6on: (part II)
MV Cone Beam CT Imaging for daily localiza6on: (part II) Jean Pouliot, Ph.D. Professor UCSF Helen Diller Family Comprehensive Cancer Center AAPM CE-Therapy Series Panel Session 7h30-9h25 am, July 28th,
More informationCostlets: A Generalized Approach to Cost Functions for Automated Optimization of IMRT Treatment Plans
Optimization and Engineering, 6, 421 448, 2005 c 2005 Springer Science + Business Media, Inc. Manufactured in The Netherlands Costlets: A Generalized Approach to Cost Functions for Automated Optimization
More informationObioma Nwankwo 1,2*, Hana Mekdash 1, Dwi Seno Kuncoro Sihono 1, Frederik Wenz 1 and Gerhard Glatting 2
Nwankwo et al. Radiation Oncology (2015) 10:111 DOI 10.1186/s13014-015-0416-6 RESEARCH Open Access Knowledge-based radiation therapy (KBRT) treatment planning versus planning by experts: validation of
More informationAssessing dose variance from immobilization devices in VMAT head and neck treatment planning: A retrospective case study analysis
Assessing dose variance from immobilization devices in VMAT head and neck treatment planning: A retrospective case study analysis Alyssa Olson, B.S., R.T.(T), Kristine Phillips, B.S., R.T.(T), Tamara Eng
More informationDual-gated volumetric modulated arc therapy
Fahimian et al. Radiation Oncology 2014, 9:209 RESEARCH Open Access Dual-gated volumetric modulated arc therapy Benjamin Fahimian 1, Junqing Wu 1,2, Huanmei Wu 3, Sarah Geneser 1 and Lei Xing 1* Abstract
More informationQuality assurance: Fundamental reproducibility tests for 3D treatment-planning systems
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 6, NUMBER 3, SUMMER 2005 Quality assurance: Fundamental reproducibility tests for 3D treatment-planning systems Charles M. Able and Michael D. Thomas
More information