Gastrointestinal Imaging Original Research

Size: px
Start display at page:

Download "Gastrointestinal Imaging Original Research"

Transcription

1 Gastrointestinal Imaging Original Research Mayer et al. Radiation Dose Savings in CT Gastrointestinal Imaging Original Research Caroline Mayer 1 Mathias Meyer 1 Christian Fink 1 Bernhard Schmidt 2 Martin Sedlmair 2 Stefan O. Schoenberg 1 Thomas Henzler 1 Mayer C, Meyer M, Fink C, et al. Keywords: automatic tube current modulation, automatic tube voltage selection, CT, image quality, radiation dose reduction DOI:.2214/AJR Received July 29, 13; accepted after revision January 3, 14. C. Mayer and M. Meyer contributed equally to this work. B. Schmidt and M. Sedlmair are employees of Siemens Healthcare. The other authors (who have no conflicts of interest to disclose) had control of the data and information submitted for publication. 1 Institute of Clinical Radiology and Nuclear Medicine, University Medical Center Mannheim, Medical Faculty Mannheim - Heidelberg University, Theodor-Kutzer-Ufer 1-3, D Mannheim, Germany. Address correspondence to T. Henzler (thomas.henzler@umm.de). 2 CT Division, Siemens Healthcare Sector, Forchheim, Germany. AJR 14; 3: X/14/ American Roentgen Ray Society Potential for Radiation Dose Savings in Abdominal and Chest CT Using Automatic Tube Voltage Selection in Combination With Automatic Tube Current Modulation OBJECTIVE. The purpose of this study was to evaluate the simultaneous use of automatic tube current modulation (ATCM) and automatic tube voltage selection (ATVS) for abdominal and chest CT examinations regarding radiation dose reduction and image quality. MATERIALS AND METHODS. We enrolled 617 patients who all underwent contrastenhanced chest or abdominal CT and divided them into two groups. In group A, 317 patients who underwent CT with only ATCM and a fixed body mass index adjusted tube voltage (1 kv or 0 kv) were enrolled. In group B, both ATCM and ATVS were used. Image attenuation and noise were measured in different anatomic regions. RESULTS. The mean contrast-to-noise ratio and the signal-to-noise ratio of abdomen and chest CT was higher in group B compared with group A (p < ). In total, the effective radiation doses for abdomen and chest CT examinations were significantly reduced in group B by 18% compared with group A (p < ). When only examining those who benefited from the ATVS tool, a dose reduction of 35% for chest CT and 42% for abdomen CT could be achieved (p < for each). CONCLUSION. The simultaneous use of ATVS and ATCM enables significant radiation dose reduction in abdominal and thoracic contrast-enhanced CT examinations compared with the use of ATCM alone. S ince the introduction of MDCT, the total number and clinical indications for MDCT examinations has grown steadily. CT is by far the largest contributor to medical radiation exposure among the U.S. and European population [1]. This has raised significant concerns about potential radiation hazards from diagnostic CT. Although the risks for a single person undergoing a single examination may be neglected, the increasing radiation exposure in the population is going to play a role in future public health issues [2]. Consequently, radiologists, physicists, and manufacturers are trying to decrease radiation exposure to fulfill the principle of keeping radiation dose as low as reasonably achievable [3]. Within the past years, various techniques for radiation dose reduction have been developed, such as automatic tube current modulation (ATCM), reduced tube voltage based on patient size, cone beam bowtie filters, iterative reconstruction algorithms, and decreased scanning length [4]. ATCM, which has been introduced into clinical routine imaging, enables automatic adjustment of the tube current in various planes (x, y, or z) on the basis of the size and attenuation of the body area scanned, allowing a constant image quality and improved radiation dose efficiency [5, 6]. However, most of these approaches lead to a simultaneous increase in image noise and a decrease in acceptable images of diagnostic quality, limiting their potential mainly because of the higher absorption of low-energy photons by the patient [7]. Recently, attention has been shifted toward the potential of low-tube-voltage imaging and newly introduced automatic tube voltage selection (ATVS) in addition to ATCM [8 ]. ATVS automatically maintains a constant contrast-to-noise ratio (CNR) depending on the examination type (e.g., angiographic, parenchymal with contrast administration, or soft tissue without contrast administration). This is accomplished by transforming the scanning parameters, e.g., tube voltage (kv) and tube current (mas), from a predefined reference patient to the parameters of the selected examination type also respecting the deviation from the reference patient. 292 AJR:3, August 14

2 Radiation Dose Savings in CT The deviation is calculated using the attenuation information that is based on the initial topogram scan. Transforming the scanning parameters in general results in a reduction of overall patient dose because in most cases the tube voltage is reduced while the tube current is increased, leading to an overall reduction in patient dose. If the examination requires more tube current to maintain the CNR, the patient dose might increase in some cases. The aim of this study was to evaluate the simultaneous use of ATCM and ATVS for abdomen and thorax contrast-enhanced CT examinations regarding radiation dose reduction and image quality. Materials and Methods Patients This retrospective single-center study was approved by our institutional review board and was in compliance with the Declaration of Helsinki and HIPAA. Because of the retrospective nature of our study, informed consent for the retrospective data analysis was waived. All CT examinations were clinically indicated for the assessment of abdominal and thoracic abnormalities. In total, 617 consecutive adult patients were enrolled in this study and divided into two groups. The first group (group A) included 317 patients (mean age [± SD] 62 ± 16 years; 146 women and 171 men) who underwent a standard protocol of the abdomen (n = 158) or the chest (n = 159) using ATCM (CARE Dose4D, Siemens Healthcare) and a fixed tube voltage of 1 kv. The second group (group B) consisted of 0 patients (mean age, 64 ± 12 years; 123 women and 177 men) who underwent a modified CT protocol of the abdomen (n = 135) or the chest (n = 165) using ATCM in combination with ATVS (CARE kv, Siemens Healthcare). For both groups, the topogram scan was obtained with a fixed tube voltage of 1 kv. For quality assurance of this new tool, the data acquisition was divided into two steps for group B. In an initial step, the ATVS tool was turned off, leaving only the ATCM tool as the dose modulator. The system then generated a topogram, calculating the estimated scanning and dose parameters, volume CT dose index ( ), dose-length product (DLP), effective tube current (mas), and kilovoltage. These parameters were recorded for all patients because they reflect the scanning parameters of each patient as if the patient had been scanned with the Care kv tool. This step did not include any additional patient scanning. In a second step, ATVS was then turned on and the patient was scanned using ATCM and ATVS. CT Technique All CT examinations were performed on a dual-source 64-MDCT system (Somatom Definition, Siemens Healthcare). CT examinations were performed with two protocols. The first protocol (group A) was performed with a standard 1 kv and ATCM. The second protocol (group B) was performed using ATCM and ATVS simultaneously. Detector collimation was set to mm and a pitch of 1.2 was used for all abdominal and thoracic examinations. For the abdominal portal venous examinations, contrast enhancement was achieved by injecting 5 ml of nonionic iodinated contrast material (0 mg I/mL iomeprol, Iomeron, Bracco) via an antecubital vein at a flow rate of 2.5 ml/s using a dual-syringe power injector (Stellant D CT Injection System, Medrad). The examination automatically started after a fixed delay of 70 seconds after the initiation of contrast material injection. Contrast enhancement for the arterial thoracic examinations was performed by injecting 80 ml of nonionic iodinated contrast material (0 mg I/mL iomeprol) via an antecubital vein at a flow rate of 3.5 ml/s using a dual-syringe power injector (Stellant D CT Injection System). Scanning initiation was determined individually using bolus tracking with a triggered threshold of 0 HU within an ROI that was placed in the descending aorta. CT raw data were reconstructed using filtered back projection with a slice thickness of 1.5 mm using a soft-tissue convolution kernel (Bf). Principle of the Automatic Tube Voltage Selection Tool The ATVS tool that was evaluated in this study was designed to work in conjunction with ATCM, which modulates tube current on the basis of the topogram after the optimal tube voltage is selected by ATVS. According to the attenuation along the z-axis obtained from the CT topogram, the software algorithm automatically calculates the tube current for all possible kilovoltage (70 1 kv) values at each scanning location that meets the user-prescribed image quality requirements. A tube current modulation curve is calculated over the entire scanning range, and the radiation dose is estimated on the basis of the kv / mas curves. Some tube voltages might result in lower doses but are rejected by ATVS because of tube current limits (i.e., the required tube current will exceed the scanner s maximum tube power capacity). If the required tube current at any location exceeds the system limits, the calculation is repeated automatically for the next tube potential that yields optimal image quality and the lowest individual dose. The kilovoltage selected by ATVS is a single discrete selection, so that the same kilovoltage is used throughout the CT examination. Before the examination is performed with ATVS, the optimized kilovoltage, optimized effective tube current, and that the system plans to deliver are displayed on the scanner interface [8]. Quantitative Image Analysis All CT measurements were performed on an offline communication workstation (Osirix, version 3.7.1, Aycan) by a third-year resident with 2 years of experience in whole-body CT who was blinded to the CT protocol. CT values (HU) and image noise (SD of measured CT values) were measured on the reconstructed 1.5-mm axial CT images using equal ROIs in various anatomic regions. On the abdominal images, ROIs were placed within the liver, spleen, portal vein, abdominal aorta, psoas muscle, visceral fat, and air (Fig.1). On the thoracic images ROIs were placed within the descending aorta; subscapularis muscle; and air within the trachea, lung parenchyma, and subcutaneous fat. The ROIs were drawn by hand in the form of ovals of various sizes depending on the anatomic region. However, the size in one anatomic region was kept equal throughout the patient evaluations. Each ROI was placed in the identical or nearly identical segment. SDs of the attenuation in these ROIs were recorded. Each anatomic area was measured three times, and the mean value of these three evaluations was used for further analysis. Subsequently, signal-to-noise ratio (SNR) and CNR were calculated for each image dataset according to the methods used by Szucs-Farkas et al. [11]. In detail, the following formulas were applied: SNR = mean HU measured in ROI / noise (SD from mean attenuation measured in ROI). For example, for the chest: mean attenuation in ROI of the descending aorta / SD from mean attenuation measured in ROI of the descending aorta, and CNR = mean HU measured in ROI mean HU measured in ROI of the surrounding tissue / SD from mean attenuation measured in ROI. For example, for the abdomen: mean HU measured in ROI of the liver mean HU measured in ROI of the subscapularis muscle) / SD from mean attenuation measured in ROI of the liver. Qualitative Image Analysis Subjective image quality was initially assessed separately by two radiologists with 2 and 6 years of experience and later in a consensus reading. Both radiologists rated the assessability of thoracic and abdominal structures, image noise, and diagnostic acceptability as described for wholebody staging CT examinations in the European Guidelines on Quality Criteria for CT [12]. Images were evaluated for overall diagnostic image quality on a 5-point Likert scale (1, poor; 2, fair; AJR:3, August

3 Mayer et al. 3, moderate; 4, good; and 5, excellent). Additionally, image noise was subjectively evaluated using the following 5-point Likert scale (1, unacceptable image noise; 2, above average noise; 3, average image noise; 4, less than average noise; and 5, minimal image noise). A C Fig. 1 Comparison of patient groups. A D, Axial contrast-enhanced CT images of 67-year-old man from group B (A and B) and 55-year-old man from group A (C and D). During atrial phase (A) and portal venous phase (B), images were obtained for combined approach in group B at tube voltage of 0 kv. Mean image noise exemplarily measured in portal vein (white circle) and latissimus dorsi muscle (black circle) was 24 and 23 HU (A) and 25 and 22 HU (B). In group A with fixed tube voltage of 1 kv exemplarily measured mean image noise was 24 and 18 HU during atrial phase (C) and 24 and 21 HU during portal venous phase (D). Radiation Dose Assessment The, DLP, effective diameter, and size-specific estimate were obtained from the CT dose report for each patient for the abdomen and the chest acquisitions [13]. The effective dose was estimated by multiplying the DLP by body-specific conversion coefficients according to the generic method presented in the European Guidelines for MDCT (dose conversion coefficients in msv Gy 1 cm 1 : thorax, and abdomen-pelvis, 0.017) [14]. To detect any relationship between body constitution and radiation dose, we evaluated body habitus by taking the maximum anteroposterior and the transverse diameters of the abdomen and the chest. Statistical Analysis All statistical analyses were performed using SAS 9.0 software (SAS Institute) at a significance level of Continuous variables are expressed as mean ± SD and categoric variables as percentages. To compare the CT density and image noise, SNR, CNR, and radiation parameters between both groups one-way ANOVA was performed. To compare the relationship between body constitution and radiation dose, Pearson correlation was performed. Results Body Constitution No statistically significant differences within the measured body constitution parameters were found between both groups, in particular, for the chest anteroposterior diameter (group A, 23.2 ± 2.9 cm vs group B, 23.7 ± 2.6 cm; p > 0.05) and transverse diameter (group A, 33.0 ± 4.3 cm vs group B, 34.7 ± 4.3 cm; p > 0.05) and for the abdomen anteroposterior diameter (group A, B D 294 AJR:3, August 14

4 Radiation Dose Savings in CT TABLE 1: Radiation Dose Parameters of Group A and Group B Group A (n = 317) Group B (n = 0) p (ATCM vs Simultaneous Use of ATCM and ATVS) Radiation Dose Parameters 26.2 ± 4.1 cm vs group B, 25.3 ± 3.5 cm) and transverse diameter (group A, 33.2 ± 3.5 cm vs group B, 32.6 ± 3.1 cm; p > 0.05 for both). Radiation Dose The overall comparison of both groups showed a significant dose reduction of 18% in group B compared with group A (p < ). No significant differences in tube current modulation were observed in group A and group B. Radiation dose for groups A and B as well as the estimated dose parameters for the single use of ATCM in group B are displayed in Table 1. Chest CT Among the 165 chest examinations in group B, tube voltage was reduced in 96 patients from 1 to 0 kv and in three patients to 80 kv (99 patients; 61%) (Table 2). In total an overall dose reduction of 16.8% was found for group B compared with group A (p = ) (Table 1). In a second step, we performed a subgroup analysis by separating patients of group B who benefited from the ATVS tool because the tube voltage of their examination was reduced because of this program. This was the case for 99 patients undergoing chest CT and for 48 patients undergoing abdominal CT. The radiation dose parameters were assessed and compared with the parameters of group A. Comparing this subgroup to group A, a dose reduction of 34% could be achieved (mean DLP, ± ATCM Only ATCM Only ATCM and ATVS Within Group B Group A vs Group B Chest DLP (mgy cm ) ± ± ± (mgy) 7.9 ± ± ± Effective current time product per tube (mas) ± ± ± Effective patient diameter (cm) 27.6 ± ± Size-specific estimate (mgy).4 ± ± ± Effective dose (msv) 6.0 ± ± ± Abdomen DLP (mgy cm ) ± ± ± (mgy) 12.3 ± ± ± Effective current time product per tube (mas) ± ± ± Effective patient diameter (cm) 29.5 ± ± Size-specific estimate (mgy) 15.1 ± ± ± Effective dose (msv).7 ± ± ± Note Except for p values, data are mean ± SD. ATCM = automatic tube current modulation, ATVS = automatic tube voltage selection, DLP = dose-length product, = volume CT dose index. Dash indicates not applicable. mgy for group A and mean DLP, 6.5 ± 49 mgy for subgroup B; p < ). Furthermore, no differences in tube current modulation were observed between both groups (group A, ± 66.6 mas vs subgroup B, ± 26.7 mas; p = ). To reduce bias due to intergroup differences, we also compared the estimated data of an ATCM single use in group B with the true Care kv data for group B. By comparing the dose parameters of these two groups, a dose reduction of 19% could be achieved (mean DLP, ± 0.3 mgy vs ± 97.8 mgy; p < ) (Table 1). The quantitative image analysis revealed no significant differences in image noise for chest CT between both groups, with the exception of the lung (p = ) (Fig. 2). Mean density was significantly higher in fat tissue and the ascending aorta for group B compared with group A (all p < ) (Table 3). In addition, the CNR and the SNR of the ascending aorta were significantly higher in group B (both p < ). A detailed list of the quantitative values is displayed in Table 3. Independent of the approach chosen, all images were rated as diagnostically sufficient, and the level of interobserver agreement was excellent (for overall image quality, κ = 0.9, and for image noise, κ = 0.84). The median range image quality scores (5 [4 5] vs 5 [4 5]; p = 0.816) as well as the TABLE 2: Recommended Kilovoltage Settings for Group B Tube Voltage (kv) ATCM Only ATCM and ATVS Chest Abdomen Note Data are number of examinations. ATCM = automatic tube current modulation, ATVS = automatic tube voltage selection. image noise scores (5 [4 5] vs 5 [4 5]; p = 0.721) assigned by the two radiologists did not differ significantly between the two groups for chest CT. Abdominal CT Among the 135 abdominal CT examinations in group B, tube potential was reduced to 0 kv in 48, increased to 1 kv in four, and remained unchanged in 83 (Table 2). This resulted in a dose reduction of 18.4% for group B compared with the dose parameters for group AJR:3, August

5 Mayer et al. Mean Image Noise per Region (HU) Group A Abdominal aorta Air Fat Liver Group B Psoas muscle Portal vein Spleen Group A TABLE 3: Quantitative Analysis Results in Group A and Group B A Parameter Group A (n = 317) Group B (n = 0) p Chest and abdomen density values (HU) Descending aorta ± ± Subscapularis muscle 45.8 ± ± Trachea 925 ± ± Lung 842 ± ± Chest fat 96.6 ± ± Liver 84.4 ± ± Spleen 94.7 ± ± Portal vein ± ± Abdominal aorta ± ± Psoas muscle 48 ± ± Abdominal fat 98.8 ± ± Air ± ± Chest and abdomen CNR values Ascending aorta 5.8 ± ± Liver 2.0 ± ± Portal vein 3.6 ± ± Abdominal aorta 3.2 ± ± Chest and abdomen SNR values Descending aorta 8.1 ± ± Trachea 28.6 ± ± Lung.7 ± ± Chest fat 5.1 ± ± Liver 4.6 ± ± Portal vein 5.6 ± ± Abdominal aorta 5.1 ± ± Abdominal fat 6.4 ± ± Note CNR = contrast-to-noise ratio, SNR = signal-to-noise ratio. Mean Image Noise per Region (HU) Ascending aorta Fat Lung Group B Subscapularis muscle Trachea B Fig. 2 Mean noise levels for soft-tissue regions. A and B, Graphs show plots of mean noise levels for soft-tissue regions determined for two different groups investigated in this study (group A, automatic tube current modulation only and group B, automatic tube current modulation and tube voltage selection) for abdomen (A) and chest (B). Note that some data points overlap. Horizontal lines represent mean image noise per reconstruction technique. A (p = 0.016) (Table 1). Comparing the parameters of group A with those who benefited from Care kv in group B (48 patients), as the tube voltage was reduced, an even higher dose reduction of 42% could be achieved (mean DLP group A, ± mgy vs group B, 364 ± 79.7 mgy; p < ). When again looking at the estimated dose parameters of the ATCM single examination in group B and comparing it with the true Care kv data for group B, we found a dose reduction of 17% (mean DLP, ± mgy vs ± mgy; p < ) (Table 1). A significant moderate linear correlation was observed between body constitution and for chest CT (anteroposterior diameter r = 0.46 and transverse diameter r = 0.55, p < ) and for the DLP (anteroposterior diameter r = 0.45 and transverse diameter r = 0.51; p < ). For the abdominal CT examinations, the correlation between body constitution parameters and / DLP was lower when compared with the chest examinations but still statistically significant ( : anteroposterior diameter r = 0.32 and transverse diameter r = 0.36, p < ; DLP: anteroposterior diameter r = 0.28 and transverse diameter r = 0.29, p < ). The results are shown in Figure 3. The quantitative image analysis revealed that image noise was significantly higher in group B within the spleen, liver, and fat tissue compared with group A (p < ) (Fig. 1). The mean density values for group B were only significantly higher in the abdominal muscle tissue compared with group A. In addition, the liver 296 AJR:3, August 14

6 Radiation Dose Savings in CT (mgy) (mgy) 50 r = 0.46 p < Thorax Anteroposterior (cm) r = 0.32 p < Thorax Anteroposterior (cm) CNR was significantly higher in group B (p = ), whereas no significant difference in CNR was observed for all other organs. Furthermore, the SNR was significantly higher in the abdominal fat tissue of group B (p < ), whereas significantly lower SNR values were found in the liver parenchyma compared with group A (p = ). The values for objective image quality for both groups are summarized in Table 3. Independent of the approach chosen, all images were rated as diagnostically sufficient, and the level of interobserver agreement was excellent (for overall image quality, κ = 0.89 and for image noise, κ = 0.81). The median range image quality scores (5 [4 5] vs 5 [4 5]; p = 0.645) as well as the image noise scores (5 [4 5] vs 5 [4 5]; p = 0.567) assigned by the two radiologists did not differ significantly between the two groups for abdominal CT. A (mgy) r = 0.55 p < Thorax Transverse (cm) Thorax Transverse (cm) C D Fig. 3 Correlation between patient measurement and volume CT dose index ( ). A D, Graphs show linear statistical correlation between patient anteroposterior and transverse body diameter and volume in abdominal and chest CT: A, anteroposterior chest CT; B, transverse chest CT; C, anteroposterior abdominal CT; and D, transverse abdominal CT. (mgy) 50 r = 0.36 p < Discussion There are many CT technical parameters that can be adjusted to reduce radiation dose on CT, including tube potential; tube current; iterative reconstruction algorithms; and, with scanners from some manufactures, the pitch [15 17]. However, adjusting one parameter independently of all others will increase image noise and may compromise image quality, with the exception of iterative reconstruction algorithms, which make it possible to diminish this problem [5, 18]. Among the various dose-reduction techniques, ATCM, which automatically adjusts the tube current on the basis of patient size and attenuation, has become one of the most comprehensive approaches and is therefore used in clinical routine [6]. Several studies have evaluated this technique in different body regions, including the head, neck, chest, and abdomen. All of these studies concluded that a significant dose reduction between 15% and 50% could be achieved while maintaining sufficient image quality by using ATCM techniques [5, 6, 19]. Given that tube voltage has an exponential relationship with radiation dose [], lowering the tube voltage may also be an effective approach to reduce radiation exposure because of the introduction of ATVS, which recommends the optimal B tube current and tube voltage settings for each patient on the basis of body constitution and body attenuation obtained from the individual CT topogram. Hence, the optimal combination of tube voltage and tube current modulation is calculated, resulting in the lowest achievable radiation dose for each patient. Several studies have evaluated ATVS in phantom- and patientbased CT angiography (CTA) and abdominal CT studies [8,, 21]. Lee et al. [8] assessed dose reduction and image quality of contrast-enhanced liver CT using ATVS. The authors observed an overall dose reduction of 31% depending on the contrast gain settings using ATVS, which is in accordance to the level of dose reduction found in our study for abdominal CT. The overall image quality was acceptable despite higher image noise in ATVS studies. They also reported a high correlation between the patient s BMI and dose reduction, noting that underweight patients had more benefits of dose reduction (33%) than normal weight patients [8]. Goetti et al. [] examined the impact of automatic tube potential selection on image quality and radiation dose in 35 patients undergoing whole-body CTA after endovascular aneurysm repair of the abdominal aorta. They showed that ATVS led to a decrease of tube voltage to 0 and 80 kv in 51% of their patients and an overall radiation dose reduction of 16% compared with imaging with a fixed tube voltage of 1 kv. Image noise was higher in the ATVS group as well but still allowed sufficient diagnostic quality []. Hough et al. [21] reported similar results. This group compared ATVS in 36 patients undergoing contrast-enhanced abdominopelvic CT with a size-matched control group. In this retrospective study, a dose reduction between 18% and 25% depending on the tube voltage setting (1 or 0 kv) was achieved [21]. Our study showed that by using ATVS an overall decrease in radiation dose of 18% is feasible. In particular for chest and abdomen CT, dose reductions of 16.8% and 18.4%, respectively, were observed compared with the ATCM alone. These results are in accordance with the previously mentioned results showing the feasibility of this tool for chest and abdominal CT. In comparison with previous studies, we further analyzed those patients who benefited from ATVS and preformed a subgroup analysis. In 35% of the abdomen examinations and in 60% of the chest examinations, tube voltage was reduced to AJR:3, August

7 Mayer et al. the next lower level. For these particular patients, an overall dose reduction of 42% in abdomen CT and 34% in chest CT could be achieved compared with a single ATCM approach. Surprisingly, tube current modulation did not increase simultaneously with a decrease in tube voltage on the ATVS study as we expected. The tube current modulation was similar in both groups. Usually lowering the tube voltage leads to a compensatory increase in tube current because both parameters depend on each other. In contrast to the previously mentioned studies, we did not use BMI but measured body anteroposterior and transverse diameters instead. As shown in several other studies, BMI is not a reliable parameter to assess patient habitus. Over- or underdosing can occur when BMI is used to select tube potential [22]. In addition, reduced radiation dose and substantially improved image quality could be achieved when accounting for differences in body habitus and size according to anteroposterior and transverse diameters instead of BMI [23]. A significant weak to moderate linear correlation was observed between body diameter and radiation dose parameters. As shown in previous studies, patients with a small body diameter received lower radiation doses than those of with large body diameters [5, 8]. Regarding the quantitative analysis of our study, image noise and CNR increased significantly in some anatomic regions with ATVS, but the range of image noise was rated diagnostically acceptable. Whether the combined use of Care kv and newly reintroduced iterative reconstructions may diminish the increase in image noise has to be assessed in further studies. On the other hand, low-kilovoltage examinations provide higher vessel lumen attenuation of iodinated contrast media, which may be beneficial in detecting smaller differences in the vascularization of lesions because the lesion-to-background contrast increases [24]. Different studies have shown the benefit of this relation in the detection of small hypervascular lesions of the liver or pancreas [25, 26]. Because subjective image quality is particularly important to maintain diagnostic image quality when reducing the tube voltage setting, our results displayed no differences in subjective image quality or image noise regardless of the approach chosen. These findings are in accordance with previous studies that reported high diagnostic quality although tube voltage and tube modulation were reduced [, 21] Several limitations have to be acknowledged. First, our data may be influenced by an intergroup bias because patients and conditions were different between the two groups. To minimize this bias, we further compared the estimated data of using ATCM alone in group B with the true ATVS data for group B so that a comparison of both programs in the same patients would be possible. These results revealed findings similar to the comparison between group A and group B. Finally, the ATVS technology is vendor specific, which may limit its use in clinical practice. In conclusion, the use of ATVS for chest and abdominal CT results in a significant dose reduction while maintaining adequate image quality and diagnostic confidence without user interaction. The ATVS tool reduced effective tube voltage in many patients (49%), resulting in a dose reduction of 18%, showing the potential of this new dose modulation tool. Acknowledgments We thank Thomas Flohr and Paul Apfaltrer for their support. References 1. Mettler FA Jr, Bhargavan M, Faulkner K, et al. Radiologic and nuclear medicine studies in the United States and worldwide: frequency, radiation dose, and comparison with other radiation sources Radiology 09; 253: Brenner DJ, Hall EJ. Computed tomography: an increasing source of radiation exposure. N Engl J Med 07; 357: Mountford PJ, Temperton DH. Recommendations of the International Commission on Radiological Protection (ICRP) Eur J Nucl Med 1992; 19: May MS, Wust W, Brand M, et al. Dose reduction in abdominal computed tomography: intraindividual comparison of image quality of full-dose standard and half-dose iterative reconstructions with dual-source computed tomography. Invest Radiol 11; 46: Lee EJ, Lee SK, Agid R, Howard P, Bae JM, ter- Brugge K. Comparison of image quality and radiation dose between fixed tube current and combined automatic tube current modulation in craniocervical CT angiography. AJNR 09; : Lee S, Yoon SW, Yoo SM, et al. Comparison of image quality and radiation dose between combined automatic tube current modulation and fixed tube current technique in CT of abdomen and pelvis. Acta Radiol 11; 52: Yu L, Li H, Fletcher JG, McCollough CH. Automatic selection of tube potential for radiation dose reduction in CT: a general strategy. Med Phys ; 37: Lee KH, Lee JM, Moon SK, et al. Attenuationbased automatic tube voltage selection and tube current modulation for dose reduction at contrastenhanced liver CT. Radiology 12; 265: Winklehner A, Goetti R, Baumueller S, et al. Automated attenuation-based tube potential selection for thoracoabdominal computed tomography angiography: improved dose effectiveness. Invest Radiol 11; 46: Goetti R, Winklehner A, Gordic S, et al. Automated attenuation-based kilovoltage selection: preliminary observations in patients after endovascular aneurysm repair of the abdominal aorta. AJR 12; 199:[web]W380 W Szucs-Farkas Z, Strautz T, Patak MA, Kurmann L, Vock P, Schidera ST. Is body weight the most appropriate criterion to select patients eligible for low-dose pulmonary CT angiography? Analysis of objective and subjective image quality at 80 kvp in 0 patients. Eur Radiol 09; 19: Danish Society of Radiology website. European guidelines on quality criteria for computed tomography. Published Accessed April 15, American Association of Physicists in Medicine website. Size-specific dose estimates (SSDE) in pediatric and adult body CT examinations. Report 4. Published 11. Accessed April 15, [No authors listed]. The 07 Recommendations of the International Commission on Radiological Protection. ICRP publication 3. Ann ICRP 07; 37: McCollough CH, Bruesewitz MR, Kofler JM Jr. CT dose reduction and dose management tools: overview of available options. RadioGraphics 06; 26: Cody DD, Moxley DM, Krugh KT, O Daniel JC, Wagner LK, Eftekhari F. Strategies for formulating appropriate MDCT techniques when imaging the chest, abdomen, and pelvis in pediatric patients. AJR 04; 182: Huda W, Lieberman KA, Chang J, Roskopf ML. Patient size and x-ray technique factors in head computed tomography examinations. I. Radiation doses. Med Phys 04; 31: Kalra MK, Maher MM, Toth TL, et al. Techniques and applications of automatic tube current modulation for CT. Radiology 04; 233: Greess H, Normayr A, Wolf H, et al. Dose reduc- 298 AJR:3, August 14

8 Radiation Dose Savings in CT tion in CT examination of children by an attenuation-based on-line modulation of tube current (CARE Dose). Eur Radiol 02; 12: Gunn ML, Kohr JR. State of the art: technologies for computed tomography dose reduction. Emerg Radiol ; 17: Hough DM, Fletcher JG, Grant KL, et al. Lowering kilovoltage to reduce radiation dose in contrast-enhanced abdominal CT: initial assessment of a prototype automated kilovoltage selection tool. AJR 12; 199: Ghoshhajra BB, Engel LC, Major GP, et al. Direct chest area measurement: a potential anthropometric replacement for BMI to inform cardiac CT dose parameters? J Cardiovasc Comput Tomogr 11; 5: van der Wall EE, Schuijf JD, Bax JJ. Use of the anterior-posterior chest diameter in CT: reduction in radiation dose? Int J Cardiovasc Imaging ; 26: Eller A, May MS, Scharf M, et al. Attenuationbased automatic kilovolt selection in abdominal computed tomography: effects on radiation exposure and image quality. Invest Radiol 12; FOR YOUR INFORMATION Mark your calendar for the following ARRS annual meetings: April 19 24, 15 Toronto Convention Centre, Toronto, ON, Canada April 17 22, 16 Los Angeles Convention Center, Los Angeles, CA April May 5, 17 Hyatt Regency New Orleans, New Orleans, LA April 22 27, 18 Marriott Wardman Park Hotel, Washington DC 47: Marin D, Nelson RC, Samei E, et al. Hypervascular liver tumors: low tube voltage, high tube current multidetector CT during late hepatic arterial phase for detection initial clinical experience. Radiology 09; 251: Marin D, Nelson RC, Barnhart H, et al. Detection of pancreatic tumors, image quality, and radiation dose during the pancreatic parenchymal phase: effect of a low-tube-voltage, high-tube-current CT technique preliminary results. Radiology ; 256: AJR:3, August

Size-based protocol optimization using automatic tube current modulation and automatic kv selection in computed tomography

Size-based protocol optimization using automatic tube current modulation and automatic kv selection in computed tomography JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 1, 2016 Size-based protocol optimization using automatic tube current modulation and automatic kv selection in computed tomography Robert

More information

The Latest in Radiation Dose Reduction Techniques in CT

The Latest in Radiation Dose Reduction Techniques in CT The Latest in Radiation Dose Reduction Techniques in CT Cynthia H. McCollough, PhD, FAAPM, FACR Professor of Radiologic Physics Director, CT Clinical Innovation Center Department of Radiology Mayo Clinic,

More information

Abdominal CT with Single-Energy Metal Artifact Reduction (SEMAR): Initial Experiences

Abdominal CT with Single-Energy Metal Artifact Reduction (SEMAR): Initial Experiences Abdominal CT with Single-Energy Metal Artifact Reduction (SEMAR): Initial Experiences Poster No.: C-0674 Congress: ECR 2014 Type: Scientific Exhibit Authors: K. Sofue 1, T. Yoshikawa 1, N. Negi 1, Y. Ohno

More information

40TH ANNUAL MEETING. CTA Dose Reduction: Special Considerations in Children. Jeffrey C. Hellinger, MD FACC. October 13 16, 2012 Pasadena, CA

40TH ANNUAL MEETING. CTA Dose Reduction: Special Considerations in Children. Jeffrey C. Hellinger, MD FACC. October 13 16, 2012 Pasadena, CA 40TH ANNUAL MEETING October 13 16, 2012 Pasadena, CA CTA Dose Reduction: Special Considerations in Children Jeffrey C. Hellinger, MD FACC New York Cardiovascular Institute Lenox Hill Radiology and Medical

More information

Low-dose and High-resolution Cardiovascular Imaging with Revolution* CT

Low-dose and High-resolution Cardiovascular Imaging with Revolution* CT GE Healthcare Case study Low-dose and High-resolution Cardiovascular Imaging with Revolution* CT Jean-Louis Sablayrolles, M.D. Laurent Macron, M.D. Jacques Feignoux, M.D. Centre Cardiologique du Nord,

More information

8/1/2018. Optimization Strategies for Pediatric CT Imaging. Educational Objectives. Brain CT

8/1/2018. Optimization Strategies for Pediatric CT Imaging. Educational Objectives. Brain CT Image Quality Radiation Dose Contrast Dose Sedation Optimization Strategies for Pediatric CT Imaging Samuel Brady, M.S. Ph.D. DABR samuel.brady@cchmc.org 8-1-218 @CincyKidsRad facebook.com/cincykidsrad

More information

Computed Tomography: Optimization of acquisition protocols & Justification of clinical referrals. Koos Geleijns, medical physicist

Computed Tomography: Optimization of acquisition protocols & Justification of clinical referrals. Koos Geleijns, medical physicist Computed Tomography: Optimization of acquisition protocols & Justification of clinical referrals Koos Geleijns, medical physicist CT delivers excellent 3D image quality CT delivers excellent 3D image quality

More information

Dual-Energy MDCT in Hypervascular Liver Tumors: Effect of Body Size on Selection of the Optimal Monochromatic Energy Level

Dual-Energy MDCT in Hypervascular Liver Tumors: Effect of Body Size on Selection of the Optimal Monochromatic Energy Level Medical Physics and Informatics Original Research Mileto et al. Effect of Body Size on Dual-Energy MDCT of Hypervascular Liver Tumors Medical Physics and Informatics Original Research Achille Mileto 1

More information

Handzettel 1. CARE Right Computed Tomography, committed to the right dose Ivo Driesser Austin, July 22 nd 2014

Handzettel 1. CARE Right Computed Tomography, committed to the right dose Ivo Driesser Austin, July 22 nd 2014 CARE Right Computed Tomography, committed to the right dose Ivo Driesser Austin, July 22 nd 2014 Answers for life. Answers for life. Is 1 msv the right dose for every patient? Female, 54 Abdominal CT Male,

More information

Patient doses from CT examinations in region of Prishtina, Kosovo

Patient doses from CT examinations in region of Prishtina, Kosovo Patient doses from CT examinations in region of Prishtina, Kosovo Sehad KADIRI, Gëzim HODOLLI, Kostandin DOLLANI Institute of Occupational Medicine, Radiation Protection Service, Obiliq, Kosovo Institute

More information

Leading Canadian Pediatric Hospital Continues to Drive Dose Down for Sick Children

Leading Canadian Pediatric Hospital Continues to Drive Dose Down for Sick Children Leading Canadian Pediatric Hospital Continues to Drive Dose Down for Sick Children At Canada s premier pediatric hospital also recognized as one of the world s foremost pediatric healthcare institutions

More information

Image Gently Development of Pediatric CT Protocols 2014

Image Gently Development of Pediatric CT Protocols 2014 Image Gently Development of Pediatric CT Protocols 2014 Introduction In 2001, several scientific articles[1-3] in the radiology literature received considerable media attention by pointing out that using

More information

Introduction of an Effective Method for the Optimization of CT Protocols Using Iterative Reconstruction Algorithms: Comparison With Patient Data

Introduction of an Effective Method for the Optimization of CT Protocols Using Iterative Reconstruction Algorithms: Comparison With Patient Data Medical Physics and Informatics Original Research Kordolaimi et al. Medical Physics and Informatics Original Research Sofia D. Kordolaimi 1 Ioannis Saradeas 1 Agapi Ploussi 1 Ioannis Pantos 2 Stylianos

More information

Computed tomography. High. performance. no trade-offs. Philips Ingenuity CT family

Computed tomography. High. performance. no trade-offs. Philips Ingenuity CT family Computed tomography High performance no trade-offs Philips Ingenuity CT family Keeping you ahead Until now, CT scanning has too often been about trade-offs. You ve been forced to choose between high image

More information

Gastrointestinal Imaging Original Research

Gastrointestinal Imaging Original Research Gastrointestinal Imaging Original Research Schindera et al. Abdominal CT for Obese Patients Gastrointestinal Imaging Original Research Sebastian T. Schindera 1,2 Devang Odedra 1 Diego Mercer 1 Seng Thipphavong

More information

Patient-centered CT imaging: New methods for patient-specific optimization 1 of image quality and radiation dose

Patient-centered CT imaging: New methods for patient-specific optimization 1 of image quality and radiation dose Patient-centered CT imaging: New methods for patient-specific optimization 1 of image quality and radiation dose ipatient is an advanced platform that delivers focused innovations to facilitate patient-centered

More information

Study of Image Quality, Radiation Dose and Low Contrast Resolution from MSCT Head by Using Low Tube Voltage

Study of Image Quality, Radiation Dose and Low Contrast Resolution from MSCT Head by Using Low Tube Voltage Journal of Physics: Conference Series PAPER OPEN ACCESS Study of Image Quality, Radiation Dose and Low Contrast Resolution from MSCT Head by Using Low Tube Voltage To cite this article: Jumriah et al 2018

More information

Automatic Bone Removal Technique in Whole-Body Dual-Energy CT Angiography: Performance and Image Quality

Automatic Bone Removal Technique in Whole-Body Dual-Energy CT Angiography: Performance and Image Quality Vascular and Interventional Radiology Original Research Schulz et al. Bone Removal in CT Angiography Vascular and Interventional Radiology Original Research Boris Schulz 1 Katharina Kuehling Wolfgang Kromen

More information

AAPM ACTIVITIES WITH RESPECT TO CT IMAGING Cynthia McCollough, PhD, DABR, FAAPM, FACR, FAIMBE

AAPM ACTIVITIES WITH RESPECT TO CT IMAGING Cynthia McCollough, PhD, DABR, FAAPM, FACR, FAIMBE AAPM ACTIVITIES WITH RESPECT TO CT IMAGING Cynthia McCollough, PhD, DABR, FAAPM, FACR, FAIMBE 1631 Prince Street, Alexandria, VA 22314 571-298-1300 www.aapm.org DISCLOSURES President-elect designate, AAPM

More information

Clarity CT Technology

Clarity CT Technology Clarity CT Technology WHITE PAPER January 2013 Using state of the art algorithms Sapheneia Clarity CT allows physicians to lower radiation dose when acquiring CT data while maintaining image quality. The

More information

Not for publication in the USA Erlangen, November 26, 2017

Not for publication in the USA Erlangen, November 26, 2017 Press Not for publication in the USA Erlangen, November 26, 2017 RSNA 2017 in Chicago: South Building, Hall A, Booth 1937 strengthens its CT portfolio by improving patient experience and expanding precision

More information

Optimization of Parameters in 16-slice CT-scan Protocols for Reduction of the Absorbed Dose

Optimization of Parameters in 16-slice CT-scan Protocols for Reduction of the Absorbed Dose Iranian Journal of Medical Physics Vol. 11, No. 2 & 3, Spring & Summer 2014, 270-275 Received: March 10, 2014; Accepted: April 27, 2014 Original Article Optimization of Parameters in 16-slice CT-scan Protocols

More information

CTA Throughout the Ages

CTA Throughout the Ages CTA Throughout the Ages Suhny Abbara, MD Associate Professor, Harvard Medical School Director of Education, Cardiac MRCT Program Director Cardiovascular Imaging Fellowship, Massachusetts General Hospital

More information

GE Healthcare. Introducing Veo * on Discovery * CT750 HD Great care by design

GE Healthcare. Introducing Veo * on Discovery * CT750 HD Great care by design GE Healthcare Introducing Veo * on Discovery * CT750 HD Great care by design Radiologists and clinicians demand the ultimate levels of clarity and detail in their images. In making an informed and confident

More information

Optimization of Low-Dose CT Protocol in Pediatric Nuclear Medicine Imaging

Optimization of Low-Dose CT Protocol in Pediatric Nuclear Medicine Imaging Optimization of Low-Dose CT Protocol in Pediatric Nuclear Medicine Imaging Hanna Piwowarska-Bilska 1, Leszek J. Hahn 2, Bozena Birkenfeld 1, Katarzyna Cichon-Bankowska 1, Maria H. Listewnik 1, and Piotr

More information

CALIFORNIA LAW AND RESPONSE + FIVE QUICK LESSONS ON CT DOSE OPTIMIZATION

CALIFORNIA LAW AND RESPONSE + FIVE QUICK LESSONS ON CT DOSE OPTIMIZATION CALIFORNIA LAW AND RESPONSE + FIVE QUICK LESSONS ON CT DOSE OPTIMIZATION John M. Boone, PhD Professor and Vice Chair (Research) of Radiology Professor of Biomedical Engineering Department of Radiology

More information

CT post processing and low dose scanning

CT post processing and low dose scanning CT post processing and low dose scanning Gabor Szell GE Healthcare CT Modality Manager EE Annual Scientific and Educational Meeting Innovations in Cardiothoracic Imaging 201 13-14 May 2011, Tokuda Hospital

More information

Toward Large-Scale Process Control to Enable Consistent CT Radiation Dose Optimization

Toward Large-Scale Process Control to Enable Consistent CT Radiation Dose Optimization Pediatric Imaging Review Larson et al. CT Radiation Dose Optimization Pediatric Imaging Review FOCUS ON: David B. Larson 1 Keith J. Strauss 2 Daniel J. Podberesky 3 Larson DB, Strauss KJ, Podberesky DJ

More information

Dual Source Flash Spiral for a scan speed of 458 mm/s

Dual Source Flash Spiral for a scan speed of 458 mm/s Core Technologies Dual Source Flash Spiral for a scan speed of 458 mm/s With Flash Spiral, dose values under 1 msv especially important for the most dose-sensitive patients are so frequent they can be

More information

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method Iran. J. Radiat. Res., 2004; 1(4): 187-194 Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method M.R. Ay 1, M. Shahriari 2, S. Sarkar 3, P.

More information

Standardizing CT-Guided Biopsy Procedures: Patient Dose and Image Noise

Standardizing CT-Guided Biopsy Procedures: Patient Dose and Image Noise Vascular and Interventional Radiology Original Research Tam et al. Standardization of Image Acquisition Protocol for CT-Guided Biopsy Vascular and Interventional Radiology Original Research FOCUS ON: Alda

More information

Submillisievert Chest CT With Filtered Back Projection and Iterative Reconstruction Techniques

Submillisievert Chest CT With Filtered Back Projection and Iterative Reconstruction Techniques Cardiopulmonary Imaging Original Research Padole et al. Submillisievert Chest CT Cardiopulmonary Imaging Original Research Atul Padole 1 Sarabjeet Singh Jeanne B. Ackman Carol Wu Synho Do Sarvenaz Pourjabbar

More information

Minimum Detectability and Dose Analysis for Size- based Optimization of CT Protocols. Christopher Craig Smitherman

Minimum Detectability and Dose Analysis for Size- based Optimization of CT Protocols. Christopher Craig Smitherman Minimum Detectability and Dose Analysis for Size- based Optimization of CT Protocols by Christopher Craig Smitherman Graduate Program in Medical Physics Duke University Date: Approved: Ehsan Samei, Supervisor

More information

Optimal Tube Potential for Radiation Dose Reduction

Optimal Tube Potential for Radiation Dose Reduction Note: This copy is for your personal non-commercial use only. To order presentation-ready copies for distribution to your colleagues or clients, contact us at www.rsna.org/rsnarights. IMAGING PHYSICS Optimal

More information

IAEA-TECDOC Dose Reduction in CT while Maintaining Diagnostic Confidence: A Feasibility/Demonstration Study

IAEA-TECDOC Dose Reduction in CT while Maintaining Diagnostic Confidence: A Feasibility/Demonstration Study IAEA-TECDOC-1621 Dose Reduction in CT while Maintaining Diagnostic Confidence: A Feasibility/Demonstration Study IAEA-TECDOC-1621 Dose Reduction in CT while Maintaining Diagnostic Confidence: A Feasibility/Demonstration

More information

Effect of Thin Overlapping Reconstruction on the Attenuation of Small ( 3 cm) Renal Cysts in the Nephrographic Phase of MDCT: a Phantom Study

Effect of Thin Overlapping Reconstruction on the Attenuation of Small ( 3 cm) Renal Cysts in the Nephrographic Phase of MDCT: a Phantom Study Effect of Thin Overlapping Reconstruction on the Attenuation of Small ( 3 cm) Renal Cysts in the Nephrographic Phase of MDCT: a Phantom Study Sun Ho Kim, MD 1, 2 Seung Hyup Kim, MD 1 Index terms: Kidney

More information

A post-processing system uses these differential data to generate:

A post-processing system uses these differential data to generate: Iterative reconstruction in single source dual-energy CT pulmonary angiography: Is sufficient to achieve a radiation dose as low as state-of-the-art single-energy CTPA? Poster No.: B-0300 Congress: ECR

More information

The use of CT guidance for spine steroid injections has increased

The use of CT guidance for spine steroid injections has increased ORIGINAL RESEARCH J.S. Schauberger P.G. Kranz K.R. Choudhury J.D. Eastwood L. Gray J.K. Hoang CT-Guided Lumbar Nerve Root Injections: Are We Using the Correct Radiation Dose Settings? BACKGROUND AND PURPOSE:

More information

Medical Physics and Informatics Original Research

Medical Physics and Informatics Original Research Medical Physics and Informatics Original Research Jaffe et al. Radiation Dose for Body CT Medical Physics and Informatics Original Research Tracy A. Jaffe 1 Terry T. Yoshizumi Greta Toncheva Colin Anderson-Evans

More information

2017 ACR Computed Tomography Quality Control Manual FAQS

2017 ACR Computed Tomography Quality Control Manual FAQS Updated 11-15-2017 2017 ACR Computed Tomography Quality Control Manual FAQS Q. The updated 2017 ACR Computed Tomography Quality Control Manual has been released. (Visit www.acr.org/education/education-catalog.)

More information

Low radiation dose subtraction CT angiography (CTA) for diagnosis of peripheral arterial occlusive disease

Low radiation dose subtraction CT angiography (CTA) for diagnosis of peripheral arterial occlusive disease Low radiation dose subtraction CT angiography (CTA) for diagnosis of peripheral arterial occlusive disease Poster No.: C-1978 Congress: ECR 2013 Type: Authors: Scientific Exhibit M. Suzuki 1, R. Tanaka

More information

CT of the Abdomen with Reduced Tube Voltage in Adults: A Practical Approach 1

CT of the Abdomen with Reduced Tube Voltage in Adults: A Practical Approach 1 GASTROINTESTINAL IMAGING 1922 CT of the Abdomen with Reduced Tube Voltage in Adults: A Practical Approach 1 Adeel R. Seyal, MD Atilla Arslanoglu, MD Samir F. Abboud, MD Azize Sahin, MD Jeanne M. Horowitz,

More information

Future bound. Philips Ingenuity Core

Future bound. Philips Ingenuity Core Future bound Philips Ingenuity Core High reliability Low-dose, high-quality imaging and coverage, and the ability to personalize image quality* patient by patient. Expect excellence in routine imaging,

More information

QIBA Profile: Small Lung Nodule Volume Assessment and Monitoring in Low Dose CT Screening

QIBA Profile: Small Lung Nodule Volume Assessment and Monitoring in Low Dose CT Screening QIBA Profile: Small Lung Nodule Assessment in CT Screening Profile - 2017 1 2 3 4 5 6 QIBA Profile: Small Lung Nodule Volume Assessment and Monitoring in Low Dose CT Screening 7 8 Stage: Publicly Reviewed

More information

Optimisation in CT. A case for shared approach. Stephen J. Golding University of Oxford

Optimisation in CT. A case for shared approach. Stephen J. Golding University of Oxford Optimisation in CT A case for shared approach Stephen J. Golding University of Oxford Body CT 1979: 10 mm sections 20 second exposure 60 second reconstruction Body CT 2007 Submillimetre sections instant

More information

Breast Reconstruction Using Patients Own Tissue Based on CT Angiography and 3-D Surface Scanning

Breast Reconstruction Using Patients Own Tissue Based on CT Angiography and 3-D Surface Scanning Breast Reconstruction Using Patients Own Tissue Based on CT Angiography and 3-D Surface Scanning Jalil JALALI 1, Maximilian EDER 2, Stefan RAITH 2, Alexander VOLF 2, Fee von WALDENFELS 2, Laszlo KOVACS

More information

Accuracy of Patient-Specific Organ Dose Estimates Obtained Using an Automated Image Segmentation Algorithm

Accuracy of Patient-Specific Organ Dose Estimates Obtained Using an Automated Image Segmentation Algorithm Marquette University e-publications@marquette Biomedical Engineering Faculty Research and Publications Biomedical Engineering, Department of 11-29-2016 Accuracy of Patient-Specific Organ Dose Estimates

More information

Radiology Rounds A Newsletter for Referring Physicians Massachusetts General Hospital Department of Radiology

Radiology Rounds A Newsletter for Referring Physicians Massachusetts General Hospital Department of Radiology Radiology Rounds A Newsletter for Referring Physicians Massachusetts General Hospital Department of Radiology Multi-Energy Computed Tomography - New Opportunities In Imaging the Abdomen Materials can be

More information

By completing this questionnaire, you are indicating your consent to participate in this research.

By completing this questionnaire, you are indicating your consent to participate in this research. Introduction "Is your Imaging EuroSafe?" comprises a series of monthly surveys on CT DRLs for different indications, which are being carried out by the EuroSafe Imaging campaign, coordinated by the European

More information

Patient radiation dose and image quality optimization in CT. Optimiser la dose patient et la qualité image en scanner

Patient radiation dose and image quality optimization in CT. Optimiser la dose patient et la qualité image en scanner Conférence SPIMED - 8 juin 2012 IQ Patient safety Dose Reduction Patient radiation dose and image quality optimization in CT Optimiser la dose patient et la qualité image en scanner Philippe Coulon, PhD.

More information

The Consistent Quality of Connected Radiology

The Consistent Quality of Connected Radiology The Consistent Quality of Connected Radiology Radiology is Evolving New compliance requirements, heightened radiation awareness and increasing technical complexity present new challenges to radiology professionals.

More information

Customer Experience Insight

Customer Experience Insight Customer Experience Insight Exceeded Contrast Dose Threshold - CT Key Takeaways Using anonymized data collected by the Radimetrics Enterprise Platform we identified a customer with a contrast dose outlier

More information

Precision in Quantitative Imaging: Trial Development and Quality Assurance

Precision in Quantitative Imaging: Trial Development and Quality Assurance Precision in Quantitative Imaging: Trial Development and Quality Assurance Susanna I Lee MD, PhD Thanks to: Mitchell Schnall, Mark Rosen. Dan Sullivan, Patrick Bossuyt Imaging Chain: Patient Data Raw data

More information

Optimization of image quality and radiation dose in neuroradiological computed tomography

Optimization of image quality and radiation dose in neuroradiological computed tomography Optimization of image quality and radiation dose in neuroradiological computed tomography Löve, Askell 2013 Link to publication Citation for published version (APA): Löve, A. (2013). Optimization of image

More information

CT QA SOLUTIONS. Ensure Accurate Screening, Diagnosis and Monitoring

CT QA SOLUTIONS. Ensure Accurate Screening, Diagnosis and Monitoring CT QA SOLUTIONS Ensure Accurate Screening, Diagnosis and Monitoring ACCREDITATION ADVANCED CT IMAGE QUALITY QA AUTOMATIC EXPOSURE CONTROL CT CHARACTERIZATION MULTI-ENERGY CT CT PERFUSION QA COMPLETE CT

More information

QIBA Profile: Small Lung Nodule Volume Assessment and Monitoring in Low Dose CT Screening

QIBA Profile: Small Lung Nodule Volume Assessment and Monitoring in Low Dose CT Screening QIBA Profile: Lung Nodule Assessment in CT Screening Profile - 2017 1 2 3 4 5 6 QIBA Profile: Small Lung Nodule Volume Assessment and Monitoring in Low Dose CT Screening 7 8 Stage: Publicly Reviewed (draft)

More information

Making Reliable Low-dose CT a Clinical Reality

Making Reliable Low-dose CT a Clinical Reality Making Reliable Low-dose CT a Clinical Reality By Ramakrishnan RS, MD, DNB, Consultant Radiologist 1 ; George Joseph, MD, DMRD, Consultant Radiologist 2 ; and Sundar RK, BSc, DRT, DAMIT, Clinical Applications

More information

Adaptive Iterative Dose Reduction in 3D

Adaptive Iterative Dose Reduction in 3D technology history For over 130 years, Toshiba has been a world leader in developing technology to improve the quality of life. Our 50,000 global patents demonstrate a long, rich history of leading innovation.

More information

The FUTURE has ARRIVED NeuViz 128

The FUTURE has ARRIVED NeuViz 128 Contacts Middle East & North Africa No. 705/706, Building 26,Al-Baker Building, Dubai Healthcare City, UAE Tel: +971 4 4404885 North America 14425 Torrey Chase, Suite 100, Houston, TX 77014 Tel: +1 281

More information

Progress Towards an International Image Quality Monitoring Framework for Quantitative Imaging

Progress Towards an International Image Quality Monitoring Framework for Quantitative Imaging Progress Towards an International Image Quality Monitoring Framework for Quantitative Imaging Ricardo S. Avila rick.avila@accumetra.com October 2, 2017 Quantitative Imaging Workshop XIV Hubble Space Telescope

More information

CT QA SOLUTIONS. Ensure Accurate Screening, Diagnosis and Monitoring ACCREDITATION CT IMAGE QUALITY QA CTDI CT CHARACTERIZATION CT PERFUSION QA

CT QA SOLUTIONS. Ensure Accurate Screening, Diagnosis and Monitoring ACCREDITATION CT IMAGE QUALITY QA CTDI CT CHARACTERIZATION CT PERFUSION QA CT QA SOLUTIONS Ensure Accurate Screening, Diagnosis and Monitoring ACCREDITATION CT IMAGE QUALITY QA CTDI CT CHARACTERIZATION CT PERFUSION QA COMPLETE CT QA With more than 40 years of experience in the

More information

How Tracking Radiologic Procedures and Dose Helps: Experience From Finland

How Tracking Radiologic Procedures and Dose Helps: Experience From Finland Medical Physics and Informatics Original Research Seuri et al. Tracking Radiologic Procedures and Dose Medical Physics and Informatics Original Research FOCUS ON: Raija Seuri 1 Madan M. Rehani 2 Mika Kortesniemi

More information

Jonathon A. Nye, Ph.D Emory University School of Medicine Atlanta, GA

Jonathon A. Nye, Ph.D Emory University School of Medicine Atlanta, GA Artifacts related to CT Jonathon A. Nye, Ph.D Emory University School of Medicine Atlanta, GA Acknowledgements: Fabio Esteves MD 1, James Galt PhD 1, Perry Sprawls PhD 1, Robert Eisner PhD 2, James Hamill

More information

IMAGE GUIDANCE DOSES IN RADIOTHERAPY. Scott Crowe

IMAGE GUIDANCE DOSES IN RADIOTHERAPY. Scott Crowe IMAGE GUIDANCE DOSES IN RADIOTHERAPY Scott Crowe IMAGING DOSE ALARA requires that imaging dose is managed and optimised Imaging dose presents an increased risk for paediatric patients Increased effective

More information

Low Dose, Lightning Speed, Latest Applications

Low Dose, Lightning Speed, Latest Applications RXL EDITION Low Dose, Lightning Speed, Latest Applications The New Standard for Radiology Toshiba Medical Systems is committed to the development of new technologies to minimize radiation dose while maintaining

More information

2018/7/31. Primary Objective

2018/7/31. Primary Objective Andrew Kuhls-Gilcrist, PhD, DABR July 31, 2018 Primary Objective The purpose of this session is to allow vendors to present their solutions and describe how their product can empower the clinician and

More information

Jessop M*(1), Thompson JD(2), Coward J(2), Sanderud A(3), Jorge J(4), Groot M de(5), Lança L(6), Hogg P(2).

Jessop M*(1), Thompson JD(2), Coward J(2), Sanderud A(3), Jorge J(4), Groot M de(5), Lança L(6), Hogg P(2). Jessop M*(1), Thompson JD(2), Coward J(2), Sanderud A(3), Jorge J(4), Groot M de(5), Lança L(6), Hogg P(2). *Corresponding author 1.University of Salford; Brighton and Sussex University Hospitals NHS Trust

More information

1. Executive Summary

1. Executive Summary 1. Executive Summary 1.1 General The fluoroscope is defined as an instrument used chiefly in industry and in the practice of medicine for observing the internal structure of objects (such as the living

More information

AAPM Scientific Meeting Imaging Symposium. State of the Art in Quantitative Imaging CT, PET and MRI. Which Imaging Modality is the Most Quantitative

AAPM Scientific Meeting Imaging Symposium. State of the Art in Quantitative Imaging CT, PET and MRI. Which Imaging Modality is the Most Quantitative AAPM Scientific Meeting Imaging Symposium State of the Art in Quantitative Imaging CT, PET and MRI Michael McNitt-Gray, PhD, FAAPM; UCLA Paul Kinahan, PhD, U. Washington Ed Jackson, PhD, FAAPM, UT-MD Anderson

More information

GE Healthcare. Optima * CT660 Second Edition

GE Healthcare. Optima * CT660 Second Edition GE Healthcare Optima * CT660 Second Edition Your vision of quality care made real The Optima CT660 system helps you deliver competent, personalized care that helps fulfill your mission and please your

More information

More flexibility More performance Less dose

More flexibility More performance Less dose 2010 General Electric Company All rights reserved. GE and GE Monogram are trademarks of General Electric Company. GE Healthcare ASiR, Performix, microvoxel, Volara, and SmartView are trademarks of General

More information

} } 8/3/2016. Accounting for kv Imaging Dose. kilovoltage imaging devices/techniques. Current imaging dose determination methods

} } 8/3/2016. Accounting for kv Imaging Dose. kilovoltage imaging devices/techniques. Current imaging dose determination methods Accounting for kv Imaging Dose Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota 1 2016 AAPM Annual Meeting - SAM Therapy Educational Course, TU-B-201-2 Washington, DC, August

More information

Dose comparison between Ziehm Vision RFD 3D, Medtronic O-arm O2 and Samsung NeuroLogica BodyTom

Dose comparison between Ziehm Vision RFD 3D, Medtronic O-arm O2 and Samsung NeuroLogica BodyTom 01 Dose comparison between RFD 3D, Medtronic O-arm O2 and Samsung NeuroLogica BodyTom Discover significantly lower dose levels for the RFD 3D. Today, intraoperative 3D imaging is regarded as the gold standard

More information

Optimizing Radiation Dose for Samsung GC85A

Optimizing Radiation Dose for Samsung GC85A White Paper Article # WP201602-DR-Dose/ Issue Date 23 Feb., 2016 Optimizing Radiation Dose for Samsung GC85A Hong Eo, M.D. / Hyesun Hwang, M.D. / Samsung Medical Center Introduction ALARA (As Low As Reasonably

More information

COCIR CT MANUFACTURERS VOLUNTARY COMMITMENT REGARDING CT DOSE OPTIMIZATION Annual Report

COCIR CT MANUFACTURERS VOLUNTARY COMMITMENT REGARDING CT DOSE OPTIMIZATION Annual Report COCIR CT MANUFACTURERS VOLUNTARY COMMITMENT REGARDING CT DOSE OPTIMIZATION 2015 Annual Report Preamble In 2010 discussions took place between COCIR and the Heads of European Radiation Competent Authorities

More information

On the Importance of Computation in Clinical Radiology

On the Importance of Computation in Clinical Radiology On the Importance of Computation in Clinical Radiology Manuel Arreola, Ph.D., DABR Director of Radiological Physics Radiology Department University of Florida/Shands Healthcare My Disclaimer I ve received

More information

SPECIFICATION CT SIMULATOR FOR BLACK LION HOSPITAL, ADDIS ABEBA, ETHIOPIA

SPECIFICATION CT SIMULATOR FOR BLACK LION HOSPITAL, ADDIS ABEBA, ETHIOPIA SPECIFICATION CT SIMULATOR FOR BLACK LION HOSPITAL, ADDIS ABEBA, ETHIOPIA 1. Scope This specification describes the requirements for the supply, delivery, installation, and acceptance testing of a CT Simulator

More information

Purpose: Materials and Methods: Results: Conclusion: Original Research n Gastrointestinal Imaging

Purpose: Materials and Methods: Results: Conclusion: Original Research n Gastrointestinal Imaging Note: This copy is for your personal non-commercial use only. To order presentation-ready copies for distribution to your colleagues or clients, contact us at www.rsna.org/rsnarights. Low Contrast Agent

More information

Computed Tomography. Printed in Japan

Computed Tomography. Printed in Japan Computed Tomography WARNING: Any reference to x-ray exposure, intravenous contrast dosage, and other medication is intended as a reference guideline only. The guidelines in this document do not substitute

More information

Quality control of CT systems by automated monitoring of key performance indicators: a two-year study

Quality control of CT systems by automated monitoring of key performance indicators: a two-year study JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 4, 2015 Quality control of CT systems by automated monitoring of key performance indicators: a two-year study Patrik Nowik, a Robert Bujila,

More information

Time-resolved 4D TWIST MR angiography of congenital heart and vessel anomalies: image quality and added diagnostic value compared to 3D MRA

Time-resolved 4D TWIST MR angiography of congenital heart and vessel anomalies: image quality and added diagnostic value compared to 3D MRA Time-resolved 4D TWIST MR angiography of congenital heart and vessel anomalies: image quality and added diagnostic value compared to 3D MRA Poster No.: C-1663 Congress: ECR 2012 Type: Scientific Paper

More information

Image quality improvement using modelbased iterative reconstruction in low dose chest CT for children with necrotizing pneumonia

Image quality improvement using modelbased iterative reconstruction in low dose chest CT for children with necrotizing pneumonia Sun et al. BMC Medical Imaging (2017) 17:24 DOI 10.1186/s12880-017-0177-9 RESEARCH ARTICLE Open Access Image quality improvement using modelbased iterative reconstruction in low dose chest CT for children

More information

PERFORM Operating Document. Radiation Dose Estimates, Calculations and Tracking Methodology

PERFORM Operating Document. Radiation Dose Estimates, Calculations and Tracking Methodology PERFORM Operating Document Radiation Dose Estimates, Calculations and Tracking Methodology PC-POD-IM-003-v03 Revision History Version Reason for Revision Date 01 New POD Jul/27/2015 02 Update section 2.2

More information

7/28/2017. Michael Speidel. University of Wisconsin - Madison

7/28/2017. Michael Speidel. University of Wisconsin - Madison Scanning-Beam Digital X-ray (SBDX) is a technology for real-time fluoroscopy and angiography in the cardiac cath lab Michael Speidel Characterized by high speed beam scanning with a 2D array of focal spots

More information

Power of BRANSIST safire in Neuroendovascular Therapy

Power of BRANSIST safire in Neuroendovascular Therapy Vascular Power of BRANSIST safire in Neuroendovascular Therapy Department of Radiology, Kinki University Hospital Suguru Ueda Mr. Suguru Ueda 1. Introduction Kinki University Hospital is located in the

More information

NIST Medical Imaging Informatics Activities. Medical Imaging

NIST Medical Imaging Informatics Activities. Medical Imaging NIST Medical Imaging Informatics Activities Ram Sriram Mary Brady Alden Dima Medical Imaging Biomarker Testing Improving Change Analysis in Lung Cancer Statistically Valid and Clinically Meaningful Biomarkers

More information

COCIR CT Manufacturers Voluntary Commitment Regarding CT Dose 2014 Annual Report

COCIR CT Manufacturers Voluntary Commitment Regarding CT Dose 2014 Annual Report Preamble COCIR CT Manufacturers Voluntary Commitment Regarding CT Dose 2014 Annual Report This 2 nd Annual Report defines the COCIR CT manufacturers voluntary commitment to HERCA as a result of the meeting

More information

Extending quality care to more people. Brivo CT315

Extending quality care to more people. Brivo CT315 GE Healthcare Extending quality care to more people. Brivo CT315 Big impact. Improving the quality of healthcare around the world is a goal we all share. But this can be a daunting challenge, with patients

More information

The availability of flat panel detector angiographic equipment

The availability of flat panel detector angiographic equipment ORIGINAL RESEARCH A.S. Ahmed Y. Deuerling-Zheng C.M. Strother K.A. Pulfer M. Zellerhoff T. Redel K. Royalty D. Consigny M.J. Lindstrom D.B. Niemann Impact of Intra- Injection Parameters on,, and Time-Concentration

More information

Table of Contents. Adaptive Diagnostics...7 Integrated Dose Reduction Streamlined Workflow Clinical Images... 26

Table of Contents. Adaptive Diagnostics...7 Integrated Dose Reduction Streamlined Workflow Clinical Images... 26 2 Table of Contents Adaptive Diagnostics...7 Integrated Dose Reduction... 17 Streamlined Workflow... 21 Clinical Images... 26 3 AQUILION TM PRIME PROVIDES CLINICAL FLEXIBILITY, ENHANCED WORKFLOW FEATURES

More information

Experience With Patient Dosimetry and Quality Control Online for Diagnostic and Interventional Radiology Using DICOM Services

Experience With Patient Dosimetry and Quality Control Online for Diagnostic and Interventional Radiology Using DICOM Services Medical Physics and Informatics Review Vano et al. Online Quality Control for Diagnostic and Interventional Radiology Medical Physics and Informatics Review FOCUS ON: Eliseo Vano 1,2 Jose I. Ten 2,3 Jose

More information

3D Filtered Backprojection Fundamentals, Practicalities, and Applications

3D Filtered Backprojection Fundamentals, Practicalities, and Applications . H. Siewerdsen et al. (AAPM 2011, Vancouver BC) For PPT slides, contact: jeff.siewerdsen@jhu.edu Education Image Reconstruction I 3D Filtered Backprojection Fundamentals, Practicalities, and Applications

More information

Fast enough to stop the Capable of delineating Unprecedented imaging power for the. Virtual endoscopy. The gatewa

Fast enough to stop the Capable of delineating Unprecedented imaging power for the. Virtual endoscopy. The gatewa Fast enough to stop the Capable of delineating Unprecedented imaging power for the M U L T I S L I C E Virtual endoscopy The gatewa motion of a beating heart. anatomic structures as small as 0.25mm. earliest,

More information

A New Standard for the Future

A New Standard for the Future technology history For over 130 years, Toshiba has been a world leader in developing technology to improve the quality of life. Our 50,000 global patents demonstrate a long, rich history of leading innovation.

More information

Your diagnostic partner. Every day. Optima* CT540

Your diagnostic partner. Every day. Optima* CT540 GE Healthcare Your diagnostic partner. Every day. Optima* CT540 At GE Healthcare, the goal was to design a CT system that would provide our clinical partners with the imaging capabilities needed every

More information

RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident

RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident Resident s Name: RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident Rotation: PHYS 705: Clinical Rotation 3 Inclusive dates of rotation: Aug. 25, 2015 Feb. 25, 2016 Director or Associate

More information

CQIE PET PROCEDURES. American College of Radiology Clinical Research Center. Centers for Quantitative Imaging Excellence LEARNING MODULE

CQIE PET PROCEDURES. American College of Radiology Clinical Research Center. Centers for Quantitative Imaging Excellence LEARNING MODULE Centers for Quantitative Imaging Excellence LEARNING MODULE CQIE PET PROCEDURES American College of Radiology Clinical Research Center Imaging Core Laboratory v2.1 Centers for Quantitative Imaging Excellence

More information

Medical Radiation Reporting: A Manufacturer s Perspective

Medical Radiation Reporting: A Manufacturer s Perspective Medical Radiation Reporting: A Manufacturer s Perspective Rich Mather, PhD Senior Manager, Clinical Programs Toshiba Medical Research Institute, USA Vendor Responsibilities Build a Dose Conscious Machine

More information

Optimizing CT Dose and Image Quality for Different Patient Sizes

Optimizing CT Dose and Image Quality for Different Patient Sizes Optimizing CT Dose and Image Quality for Different Patient Sizes Dianna D. Cody, Ph.D., DABR, FAAPM UT MD Anderson Cancer Center Michael F. McNitt-Gray, Ph.D., DABR David Geffen School of Medicine at UCLA

More information

CTV using 80 kvp with MBIR will be optimal protocol for diagnosis of DVT with lowest radiation dose

CTV using 80 kvp with MBIR will be optimal protocol for diagnosis of DVT with lowest radiation dose VSVA51 Vascular Imaging Series: CT Angiography New Techniques and Their Application Series Courses CT VA AMA PRA Category 1 Credits : 3.25 ARRT Category A+ Credits: 4.00 Thu, Dec 4 8:30 AM - 12:00 PM Location:

More information