IQCP Inspector Training Scenarios

Size: px
Start display at page:

Download "IQCP Inspector Training Scenarios"

Transcription

1 Scenario 1: You are reviewing the product insert for a non-waived point of care blood gas analyzer that uses a cartridge for each patient sample to provide ph, po2 and pco2 results. The manufacturer s instructions state that the electronic control simulator may be used as a daily control, and the liquid external controls should be run upon opening each new box of reagent cartridges and every 30 days that the same box of cartridges is in use. The laboratory has an IQCP in place for these tests on this analyzer, and the quality control plan indicates that the electronic simulator will be run every 8 hours, and the external liquid control will be run upon opening each new box of cartridges. The QC records for the electronic simulator and the external controls are reviewed monthly by the point of care coordinator, who is the director s designee for this function. The checklist requirement states: COM Quality Control Plan Elements Phase II The individualized quality control plan must define all aspects monitored based on the potential errors identified during the risk assessment, including the following parameters as applicable: The number, type(external and internal quality control systems), and frequency of quality control Criteria for acceptance performance Monitoring of the testing environment and reagents Specimen quality Instrument calibration, maintenance, and function checks Training and competency of testing personnel Provisions for multiple identical devices and variation for uses covered under one IQCP NOTE: The components of the quality control plan must meet regulatory and CAP accreditation requirements and be in compliance with the manufacturer instructions and recommendations, at minimum. The quality control plan must control the quality of the test process and ensure accurate and reliable test results. External control material samples must be analyzed at least every 31 days and with new lots and shipments of reagents or more frequently if indicated in the manufacturer's instructions. Based on the information provided, what should you do regarding compliance with COM.50500? A. Cite a deficiency for not following the manufacturer s instructions. B. Provide a recommendation to perform controls every 30 days. C. Request more information. D. Move on as the laboratory is in compliance College of American Pathologists. Page 1 of 18

2 The correct answer is C Request more information. Based on the information provided, it is not clear how often a new box of cartridges is opened. If a box is used for less than 30 days, running external controls upon opening of the box would be sufficient. Review several months of QC records to confirm this. If on the other hand, the same box of cartridges is used for more than 30 days, you should cite a deficiency. The laboratory s quality control plan is not following the manufacturer s instructions, which state that external controls must be run at least every 30 days if the same box of cartridges is in use. They would also not be in compliance with the checklist requirement because external control materials are not analyzed at least every 31 days as specified in the NOTE. If the laboratory is being inspected with the 2014 edition of the checklists: CITE: COM Documented QM/QC Plan AND Appropriate discipline specific checklist requirement for daily QC Suggested language for citation: The QCP for (point of care blood gas analyzer for ph, po2 and pco2) does not follow manufacturer's instructions for quality control. The QCP indicates that external liquid control will be run upon opening each new box of cartridges. The manufacturer s instructions say at least every 30 days if the same box of cartridges is in use College of American Pathologists. Page 2 of 18

3 Scenario 2: The laboratory you are inspecting is using the same non-waived activated clotting time instrument in their point of care program that you use in your laboratory s point of care program. They have implemented an IQCP for this test, and it meets the requirements for QC as specified by the manufacturer. As you review their quality control plan, you note that based on the risk assessment they conducted, they elected to run external controls less frequently than your laboratory determined was necessary based its own risk assessment. As you review the risk assessment, you find that they did not classify the risks the same way your laboratory did, and because of this, you disagree with their conclusions and feel they need to increase the frequency of the external controls. From the records reviewed on the test performance, you did not identify any quality issues or negative patient outcomes associated with the test. You are reviewing COM to determine if the lab is in compliance: COM Risk Assessment Phase II The IQCP for a test/device/instrument includes a risk assessment to evaluate potential sources of error to include all of the following: Preanalytic, analytic and postanalytic phases of the testing process Intended medical uses of the test and impact if inaccurate results are reported (clinical risk) Components of the tests including reagents, environment, specimen, testing personnel, and test system Variations in the components based on use of the tests (e.g. use in different environments, by different personnel, or multiple identical devices) Data from the laboratory's own environment, instrument/equipment performance, and testing personnel Manufacturer's instructions and recommendations Based on the information provided, what should you do? A. Cite COM.50300, indicating their risk assessment is not adequate. B. Cite COM because then lab performs QC less frequently than your lab. C. Move on as the laboratory is in compliance College of American Pathologists. Page 3 of 18

4 The correct answer is C Move on as the laboratory is in compliance. While you may disagree with the risk assessment the laboratory performed, as long as it encompasses all five required elements (specimen, reagents, test system, testing personnel and environment) and the quality control plan has been approved by the laboratory director, the laboratory is in compliance. Because some elements of the risk assessment will vary from laboratory to laboratory (eg, testing personnel, environment, specimen), it is possible for two laboratories using the same instrument/reagent system to evaluate the severity of a given risk differently and set different quality control frequencies accordingly. Inspectors should focus on evaluating whether the laboratory s IQCP includes all the required elements. Whether the inspector agrees with a laboratory s conclusions should have no bearing on the determination of compliance with checklist requirements. This is an opportunity for an educational exchange and discussion, and an inspector could include a recommendation for modifications, but recommendations are not required and a laboratory is not obligated to implement recommendations College of American Pathologists. Page 4 of 18

5 Scenario 3: Today s date is January 12, 2016, and you are inspecting the microbiology department when you notice that the laboratory is performing weekly QC for its microbiology susceptibility testing system following the CLSI guidelines. When you request to review the laboratory s IQCP, the supervisor tells you that they are still working on the data collection for the IQCP and it should be done in about 3 weeks. When you inquire why they are not performing daily QC in the interim, she replies that they are a small public health laboratory, and it would be cost prohibitive for them to run daily QC. Since they have always followed CLSI guidelines, and validated this years ago, they decided it wasn t necessary to perform QC daily in the interim because they hadn t changed any susceptibility reagents. Based on the information provided, what should you do? A. Cite the relevant IQCP requirements found in the All Common and Microbiology checklists. B. Recommend that the laboratory complete their IQCP in less than 3 weeks. C. Move on as microbiology is not eligible for IQCP College of American Pathologists. Page 5 of 18

6 The correct answer is A Cite the relevant IQCP requirements found in the All Common and Microbiology checklists. Providing a recommendation is not appropriate as the laboratory should have been in compliance by January 1, Susceptibility test systems are eligible and do require an IQCP in order to perform QC less frequently than daily or day of use. If the laboratory is being inspected with the 2014 edition of the checklists: Cite: COM.04000: Documented QM/QC Plan MIC.21910: Susceptibility Test QC Frequency Suggested language for citation: IQCP meeting 2015 checklist requirements not implemented for (list susceptibility test(s)) College of American Pathologists. Page 6 of 18

7 Scenario 4: You are inspecting the automated chemistry department and notice that the laboratory has implemented an IQCP for their multi-channel automated instrument. As you review their quality control plan, you see that they have determined that all analytes tested on this analyzer require 2 or 3 levels of external quality control to be run daily. You note that this particular analyzer does not meet the CAP s eligibility requirement for IQCP, because it does not have any internal controls. When you bring this to the attention of the supervisor, he indicates that he is aware of this, but his director wanted IQCPs put in place throughout the laboratory. The director required all non-waived testing that did not meet IQCP eligibility to follow the CLIA-mandated QC frequency. Based on the information provided, what should you do? A. Cite a deficiency for implementing IQCP for an instrument that does not meet CAP eligibility criteria. B. Recommend that lab not refer to this as an IQCP as it is misleading. C. Move on as the laboratory is in compliance College of American Pathologists. Page 7 of 18

8 The correct answer is C Move on. While it is not necessary to implement an IQCP if the laboratory does not reduce the QC frequency to less than daily QC, it is not a reportable offense either College of American Pathologists. Page 8 of 18

9 Scenario 5: You are inspecting the microbiology department and note that they are performing weekly quality control for their Gram stains, but have not implemented an IQCP for this. MIC Gram Stain QC Phase II Quality control of Gram stain reagents is performed for intended reactivity and recorded for each new batch of stains and at least weekly against known gram-positive and gram-negative quality control organisms. Based on the information provided, what should you do? A. Cite the laboratory for COM.50500: Quality Control Plan Elements. B. Provide a recommendation to implement an IQCP. C. Move on as the laboratory is in compliance College of American Pathologists. Page 9 of 18

10 The correct answer is C Move on. Gram stains do not meet the eligibility criteria for IQCP, so they will fall under the specific quality control requirements listed in the Microbiology checklist (MIC.21540). The default CLIA requirement for Gram stains is weekly QC College of American Pathologists. Page 10 of 18

11 Scenario 6: Today s date is February 22, 2016 and you are inspecting the point of care testing at a large teaching institution. They have listed four tests for which they have implemented an IQCP on their List of Individualized Quality Control Plans form and have provided you with these IQCPs for your review. As you look through them, you note that one was not approved by the director until February 9, 2016, although it was approved by another pathologist in late December You review the quality control records for this test from January and February 2016 and find the practices are consistent with the QCP for this test and the manufacturer s requirements. When you ask the Point of Care Coordinator about the approval of the quality control plan, he explains that the director had been out on medical leave from November through the month of January. She signed the IQCP when she returned in early February. The pathologist who was the acting director during this time approved the IQCP; it is his signature with the December 2015 date. The requirement states: COM Quality Control Plan Approval Phase II The IQCP includes a written quality control plan approved by the laboratory director prior to implementation. NOTE: The quality control plan may be part of a test procedure or be a separate written plan. As an efficiency, a single plan may address multiple tests performed on one device. A separate, quality control plan approved by the laboratory director must be in place for each laboratory with a separate CAP and CLIA number. Based on the information provided, what should you do? A. Cite a deficiency as IQCP approval cannot be delegated. B. Cite a deficiency for COM but mark it corrected-onsite. C. Move on as the laboratory is in compliance College of American Pathologists. Page 11 of 18

12 The correct answer is C Move on. While it is correct that the director cannot delegate the approval of IQCPs, it is acceptable for an acting director to approve the IQCP in the absence of the director. In this case, the director acted appropriately by adding her approval when she resumed the directorship of the laboratory. Ideally, the acting director would indicate that they are approving the plan as the acting director College of American Pathologists. Page 12 of 18

13 Scenario 7: While inspecting the Immunology section, you run across an occurrence report with a physician complaint questioning the quality of results of a specific test. The physician was concerned that false positive results have been released for at least four of her patients. You recall that the Immunology section has implemented an IQCP for that test and ask for the records of the investigation. The complaint was received several months after the IQCP was implemented. The findings of the investigation consisted of a written summary which indicated: There was an error due to pipetting; the amount of diluent needed for control specimens was different than the amount needed for patient samples. The same technologist tested all of the patients in question. Retraining of the technologist involved was conducted. The laboratory did not go back and retest any patients, retrain other personnel, or have any records showing that the IQCP was reevaluated. The requirement states: COM Quality Assessment Monitoring Phase II Ongoing quality assessment monitoring is performed by the laboratory to ensure that the quality control plan is effective in mitigating the identified risks for the IQCP and includes the following: Review of quality control and instrument/equipment maintenance and function check data at least monthly Evaluation of errors relating to preanalytic, analytic and post analytic phases of the testing process Review of complaints from clinicians and other healthcare providers regarding the quality of testing to confirm the clinical efficacy of testing, and Evaluation of corrective actions taken if problems are identified Reapproval of the quality control plan by the laboratory director or designee at least annually Based on the information provided, what should you do? A. Cite a deficiency for COM B. Provide a recommendation that all staff review and retrain on pipetting procedures. C. Move on as the laboratory is incompliance by virtue of the investigation College of American Pathologists. Page 13 of 18

14 The correct answer is A-Cite a deficiency. The investigation is incomplete because it does not include: The evaluation of patient results, which was the basis of the physician complaint, The reevaluation of the risk assessment determination of dilution errors. If the laboratory is being inspected with the 2014 edition of the checklists: Cite: COM.04000: Documented QM/QC Plan Suggested language for citation: Ongoing quality assessment monitoring for (list test) is not performed as indicated by an incomplete investigation of a physician compliant and reevaluation of the risk assessment College of American Pathologists. Page 14 of 18

15 Scenario 8: You are inspecting the immunology section of the laboratory in early March 2016 and note on the IQCP Test List/Summary that an IQCP was developed for the nonwaived Influenza A/B test. A review of the procedure indicates that the laboratory is following manufacturer s instructions and has included much of this information in the risk assessment. The supervisor provides you a copy of the risk assessment that includes historical data collected over the past year. The data demonstrates the QC protocol developed has been in place for the last year and no issues related to QC or reagent storage occurred. The provided risk assessment also included information regarding the following topics: Specimen Reagents Test System Media Mixing kits Cross reactivity Poor collection Test at room temperature Test timing Transportation Expired reagents Measurements (Wrong Pipettes) Misidentified patient Storage temperature Misinterpretation of results Wrong swab type Shipping temperature Mucoid/specimen quality Mislabeled specimen Not following manufacturer s instructions Time of collection The requirement states: COM Risk Assessment Phase II The IQCP for a test/device/instrument includes a risk assessment to evaluate potential sources of error to include all of the following: Preanalytic, analytic and postanalytic phases of the testing process Intended medical uses of the test and impact if inaccurate results are reported (clinical risk) Components of the tests including reagents, environment, specimen, testing personnel, and test system Variations in the components based on use of the tests (e.g. use in different environments, by different personnel, or multiple identical devices) Data from the laboratory's own environment, instrument/equipment performance, and testing personnel Manufacturer's instructions and recommendations You explain that you will be citing a deficiency for COM because personnel and environment (required elements) are not included in the risk assessment. On your way to the Pre-summation Conference, the lab manager hands you a document signed by the lab director indicating that in the last ten years while he has been lab director there has never been an issue involving personnel or environmental factors. Based on the information provided, what should you do? A. Cite a deficiency for not including all the required elements. B. Change the deficiency to a recommendation that all elements should be included in the risk assessment. C. Recommend a review of personnel and environmental risks within 30 days. D. Remove the deficiency as there have been no issues in the past College of American Pathologists. Page 15 of 18

16 The correct answer is A Cite a deficiency. The requirement states the required elements of the risk assessment. Risks associated with personnel and environment are very specific to the laboratory and must be evaluated. If the laboratory is being inspected with the 2014 edition of the checklists: Cite: COM.04000: Documented QM/QC Plan Suggested language for citation: The risk assessment for Influenza A/B test does not contain the required elements of personnel and environment College of American Pathologists. Page 16 of 18

17 Scenario 9: You are inspecting the Microbiology section and note that they have an IQCP in place for their molecular cartridge test, which includes an internal control. The IQCP risk assessment table includes categories of risk for the specimens, test system, reagents, environment, and personnel. There is also a detailed fishbone diagram that includes: temperature, humidity, quality of air, altitude, dust, adequate space, light, electrical, compressor, water quality, and air pressure. As you look for data addressing each of these items you realize some of these do not apply to this lab. Further investigation shows that this is the sample IQCP sent from the manufacturer. There are no specifics for this laboratory. COM Risk Assessment Phase II The IQCP for a test/device/instrument includes a risk assessment to evaluate potential sources of error to include all of the following: Preanalytic, analytic and postanalytic phases of the testing process Intended medical uses of the test and impact if inaccurate results are reported (clinical risk) Components of the tests including reagents, environment, specimen, testing personnel, and test system Variations in the components based on use of the tests (e.g. use in different environments, by different personnel, or multiple identical devices) Data from the laboratory's own environment, instrument/equipment performance, and testing personnel Manufacturer's instructions and recommendations NOTE: The risk assessment must include a process to identify the sources of potential failures and errors for a testing process, and evaluate frequency and impact of those failures and sources of error. The laboratory director must consider the laboratory's clinical and legal responsibilities for providing accurate and reliable patient test results. Published data and information may be used to supplement the risk assessment, but is not a substitute for the laboratory's own studies and evaluation. The laboratory must involve a representative sample of testing personnel in the process of conducting the risk assessment. It is not necessary for all personnel to be involved. The risk assessment for laboratories with multiple identical devices must show that an evaluation was performed if there are differences in testing personnel or environments where testing is performed, with customization of the quality control plan, as needed. The QC study performed to assess the performance and stability of the tests must support the QC frequency and elements defined in the laboratory's quality control plan. The study must include data representing, at a minimum the maximum interval between runs of external quality control. The laboratory may use historical data during the risk assessment for tests already in place. For affiliated laboratories (e.g. systems) with integrated procedures, each accredited laboratory must have its own IQCP approved by the laboratory director. There must be records demonstrating that risks specific to the site were evaluated involving a representative sample of local testing personnel to conduct the risk assessment and that laboratory-specific QC data were used in the study to support the defined frequency of quality control. Laboratories may use data from other sites to supplement risk assessments and to support their findings. Based on the information provided, what should you do? A. Cite a deficiency for not including lab-specific data. B. Cite a deficiency but mark it corrected-onsite if the evaluation is completed within 30 days. C. Move on as IQCP is just being introduced and the laboratory does have something in place College of American Pathologists. Page 17 of 18

18 The correct answer is A Cite a deficiency. The risk assessment and subsequent quality control plan must reflect laboratory specific risks (environment and personnel). The entire IQCP must be approved and implemented prior to reducing the frequency of QC. If the laboratory is being inspected with the 2014 edition of the checklists: Cite: COM.04000: Documented QM/QC Plan Suggested language for citation: The risk assessment for molecular cartridge test in Microbiology did not include the laboratory's own studies and evaluation (only used information provided by the manufacturer). The laboratory did not involve a representative sample of testing personnel in the process of conducting the risk assessment College of American Pathologists. Page 18 of 18

The All-Commons Checklist

The All-Commons Checklist The All-Commons Checklist All-Commons Checklist The All Common Checklist (COM) contains a core set of requirements that apply to all areas performing laboratory tests and procedures. In some instances,

More information

IQCP for the Akers Bio PIFA Heparin Platelet Factor 4 Rapid Assay

IQCP for the Akers Bio PIFA Heparin Platelet Factor 4 Rapid Assay On December 31, 2015 the ability to use Equivalent Quality Control (EQC) policies will expire. Laboratories will be required to either follow the quality control regulations as outlined in CLIA 88 or develop

More information

IQCP for Commercially Prepared CLSI-Exempt Media

IQCP for Commercially Prepared CLSI-Exempt Media Please note that some references to protocol, publications, performance data etc. are fictitious in this EXAMPLE. Please use your own DATA for your IQCP. The following represents one example of how you

More information

QUALITY CONTROL FOR AST: IQCP? Romney Humphries UCLA

QUALITY CONTROL FOR AST: IQCP? Romney Humphries UCLA QUALITY CONTROL FOR AST: IQCP? Romney Humphries UCLA WHAT IS QUALITY CONTROL? Procedures used to detect errors that occur due to: test system failure adverse environmental conditions variance in operator

More information

The Joint Commission's Most Challenging Standards and What you Can Do to Maintain Compliance

The Joint Commission's Most Challenging Standards and What you Can Do to Maintain Compliance The Joint Commission's Most Challenging Standards and What you Can Do to Maintain Compliance Stacy Olea, MT(ASCP), FACHE Executive Director, Laboratory Program September 6, 2016 Objectives Identify the

More information

Individualized Quality Control Plan IQCP Examples

Individualized Quality Control Plan IQCP Examples Individualized Quality Control Plan IQCP Examples Susan E. Sharp, Ph.D., ABMM, FAAM Director, Airport Way Regional Laboratory Director, Regional Microbiology and Molecular Infectious Diseases Laboratories

More information

The ABCs of IQCP 3/20/2017. Williamsport. UPMC Susquehanna. Katheryn Inglis, MS, MT(ASCP) B102: 3/27/2017

The ABCs of IQCP 3/20/2017. Williamsport. UPMC Susquehanna. Katheryn Inglis, MS, MT(ASCP) B102: 3/27/2017 The ABCs of IQCP Katheryn Inglis, MS, MT(ASCP) B02: 3/27/207 Williamsport UPMC Susquehanna Located in Williamsport, Pennsylvania Home of Little League World Series Located in central Pennsylvania, UPMC

More information

Master. Immunology Checklist. CAP Accreditation Program

Master. Immunology Checklist. CAP Accreditation Program Master Immunology Checklist CAP Accreditation Program College of American Pathologists 325 Waukegan Road Northfield, IL 60093-2750 www.cap.org 08.17.2016 2 of 33 Disclaimer and Copyright Notice On-site

More information

Leveraging ISO s Quality Management System to Achieve 24/7 Inspection Readiness. Randy Querry Accreditation Manager Clinical

Leveraging ISO s Quality Management System to Achieve 24/7 Inspection Readiness. Randy Querry Accreditation Manager Clinical Leveraging ISO 15189 s Quality Management System to Achieve 24/7 Inspection Readiness Randy Querry Accreditation Manager Clinical Goals Provide brief background on A2LA Differentiate between CLIA and ISO

More information

COMMISSION ON LABORATORY ACCREDITATION. Laboratory Accreditation Program CHEMISTRY AND TOXICOLOGY CHECKLIST

COMMISSION ON LABORATORY ACCREDITATION. Laboratory Accreditation Program CHEMISTRY AND TOXICOLOGY CHECKLIST Revised: 04/06/2006 COMMISSION ON LABORATORY ACCREDITATION Laboratory Accreditation Program CHEMISTRY AND TOXICOLOGY CHECKLIST Disclaimer and Copyright Notice The College of American Pathologists (CAP)

More information

June 2013 Clinical Laboratory News: Volume 39, Number 6

June 2013 Clinical Laboratory News: Volume 39, Number 6 June 2013 Clinical Laboratory News: Volume 39, Number 6 Competency Assessment Does the New CLIA Guidance Mean Big Changes? By Bill Malone Laboratory professionals work in one of the most highly regulated

More information

American Society of Cytopathology Core Curriculum in Molecular Biology

American Society of Cytopathology Core Curriculum in Molecular Biology American Society of Cytopathology Core Curriculum in Molecular Biology American Society of Cytopathology Core Curriculum in Molecular Biology Chapter 4 Stephanie A. Hamilton, EdD, SCT, MB(ASCP) CM MD Anderson

More information

A New Era in Quality Control: Developing an IQCP for Your Laboratory

A New Era in Quality Control: Developing an IQCP for Your Laboratory A New Era in Quality Control: Developing an IQCP for Your Laboratory Gerald A. Capraro, Ph.D., D(ABMM) Medical Director, Clinical Microbiology & Diagnostic Virology Laboratories LSU Health Sciences Center

More information

By Ginger Bailly 18 Feb For Official Use Only

By Ginger Bailly 18 Feb For Official Use Only By Ginger Bailly 18 Feb 2014 For Official Use Only 1 Discuss: Why Quality Assurance is Important How to Implement a Quality Management System Explore the Following Key Elements: Traceability Document Control

More information

College of Physicians and Surgeons of Saskatchewan Laboratory Quality Assurance Program. Policy Manual Edition

College of Physicians and Surgeons of Saskatchewan Laboratory Quality Assurance Program. Policy Manual Edition College of Physicians and Surgeons of Saskatchewan Laboratory Quality Assurance Program Policy Manual 2016 Edition LABORATORY QUALITY ASSURANCE POLICY MANUAL SUMMARY OF POLICY MANUAL CHANGES The following

More information

QUALITY CONTROL/QUALITY ASSURANCE IN THE MOLECULAR MICROBIOLOGY LABORATORY

QUALITY CONTROL/QUALITY ASSURANCE IN THE MOLECULAR MICROBIOLOGY LABORATORY QUALITY CONTROL/QUALITY ASSURANCE IN THE MOLECULAR MICROBIOLOGY LABORATORY Richard L. Hodinka, Ph.D. University of South Carolina School of Medicine Greenville Greenville Health System, Greenville, SC

More information

Laboratory Internal Audit Plan MAY 2017

Laboratory Internal Audit Plan MAY 2017 Laboratory Internal Audit Plan MAY 2017 Acknowledgements The following APHL workgroup members contributed to this document: Charlene Thomas, BS, MT(ASCP), manager, microbiologist, Alabama Bureau of Clinical

More information

e-lab Solutions Connect

e-lab Solutions Connect datainnovations.com e-lab Solutions Connect Automated Proficiency Test Reporting October 7, 2015 2013 College of American Pathologists. All rights reserved V 1.0 102013 Agenda e-lab Solutions Connect 1

More information

Competency Assessment: Is Your Program Competent? Judy Sullivan, MS, MT(ASCP)SBB HAABB Meeting April 30, 2013

Competency Assessment: Is Your Program Competent? Judy Sullivan, MS, MT(ASCP)SBB HAABB Meeting April 30, 2013 Competency Assessment: Is Your Program Competent? Judy Sullivan, MS, MT(ASCP)SBB HAABB Meeting April 30, 2013 Objectives Discuss requirements related to competency assessment Describe what assessors, inspectors,

More information

Strengthening Laboratory Management Toward Accreditation. Module 6: Quality Assurance

Strengthening Laboratory Management Toward Accreditation. Module 6: Quality Assurance Strengthening Laboratory Management Toward Accreditation Module 6: Quality Assurance Key Message My lab assures accurate and reliable testing processes. Desired Outcome Consistently accurate and reliable

More information

Case Study College of American Pathologists. All rights reserved. cap.org

Case Study College of American Pathologists. All rights reserved. cap.org Case Study How to Reduce Clerical Errors Mark Shearer, MCLT, MT(ASCP) Director of Chemistry, CompuNet Clinical Laboratories Clerical errors represent the single largest source of errors on proficiency

More information

Quality Control for Chemistry Laboratory Dynacare Kasper Laboratories Procedures

Quality Control for Chemistry Laboratory Dynacare Kasper Laboratories Procedures 54 Quality Control for Chemistry Laboratory Dynacare Kasper Laboratories Procedures Lorena Ivona Stefan 1*, Deborah Reid 2 1. Emergency Clinical Hospital, Laboratory Medical Analysis, Craiova, Romania

More information

CLIA Update Judith Yost, M.A., M.T.(ASCP) Director Division of Laboratory Services CLIA

CLIA Update Judith Yost, M.A., M.T.(ASCP) Director Division of Laboratory Services CLIA Update 2013 Judith Yost, M.A., M.T.(ASCP) Director Division of Laboratory Services Topics For Discussion CMS/ Laboratory Enrollment Data Enforcement Actions Summary Data Regulations Update PT Revisions

More information

Avoiding the Most Frequent Joint Commission Laboratory Survey Findings using Robust Process Improvement Tools (RPI) AVOIDING THE MOST

Avoiding the Most Frequent Joint Commission Laboratory Survey Findings using Robust Process Improvement Tools (RPI) AVOIDING THE MOST Avoiding the Most Frequent Joint Commission Laboratory Survey Findings using Robust Process Improvement Tools (RPI) AVOIDING THE MOST FREQUENT JOINT COMMISSION LABORATORY SURVEY FINDINGS USING ROBUST PROCESS

More information

LabGuide 13 How to Verify Performance Specifications

LabGuide 13 How to Verify Performance Specifications 1 LG 13 Comments? Feedback? Questions? Email us at info@cola.org or call us at 800 981-9883 LabGuide 13 How to Verify Performance Specifications The verification of performance specifications confirms

More information

Master. Microbiology Checklist. CAP Accreditation Program

Master. Microbiology Checklist. CAP Accreditation Program Master Microbiology Checklist CAP Accreditation Program College of American Pathologists 325 Waukegan Road Northfield, IL 60093-2750 www.cap.org 08.17.2016 2 of 105 Disclaimer and Copyright Notice On-site

More information

Today s Featured Speaker

Today s Featured Speaker Today s Featured Speaker Judy Yost, M.A., M.T.(ASCP) received her B.S. Degree at Wilkes College and her M.A. in hospital management from Central Michigan University. She is an American Society for Clinical

More information

CLIA Update 2012! CLIA. Judith Yost. Judy Yost MA, MT(ASCP) Director, Division of Laboratory Services

CLIA Update 2012! CLIA. Judith Yost. Judy Yost MA, MT(ASCP) Director, Division of Laboratory Services Update 2012! Judith Yost Judy Yost MA, MT(ASCP) Director, Division of Laboratory Services Topics for Discussion Statistics/CMS Survey Deficiency Data Compliance Tips Personnel & Competency Assessment Guidance

More information

August 2017 Changes. Flow Cytometry Checklist. CAP Accreditation Program

August 2017 Changes. Flow Cytometry Checklist. CAP Accreditation Program August 2017 Changes Flow Cytometry Checklist CAP Accreditation Program College of American Pathologists 325 Waukegan Road Northfield, IL 60093-2750 www.cap.org 08.21.2017 Disclaimer and Copyright Notice

More information

APPENDIX C SURVEY PROCEDURES AND INTERPRETIVE GUIDELINES FOR LABORATORIES AND LABORATORY SERVICES

APPENDIX C SURVEY PROCEDURES AND INTERPRETIVE GUIDELINES FOR LABORATORIES AND LABORATORY SERVICES APPENDIX C SURVEY PROCEDURES AND INTERPRETIVE GUIDELINES FOR LABORATORIES AND LABORATORY SERVICES SURVEY PROCEDURES AND INTERPRETIVE GUIDELINES FOR LABORATORIES AND LABORATORY SERVICES Part 1 Policy for

More information

laboratorymedicine> august 2001> number 8> volume 32 on 24 February 2018

laboratorymedicine> august 2001> number 8> volume 32 on 24 February 2018 interpretation [management/administration and training generalist] A Four-Part Approach to Competency Assessment Jennifer Schiffgens, MBA, MT(ASCP), 1 and Valerie Bush, PhD, MT(ASCP) 2 From 1 Clinical

More information

Copyright. Jeremiah J. Kelly (2015). All rights reserved. Further dissemination without express written consent strictly prohibited.

Copyright. Jeremiah J. Kelly (2015). All rights reserved. Further dissemination without express written consent strictly prohibited. Statutory Framework for Devices Medical Devices Investigational Use Application IDE (21 CFR 812) Abbreviated IDE Exempt Pre-Market Approval Applications 510(k) Pre-marketing Notification (21 CFR 807(e))

More information

MAINTENANCE OF QUALITY IN THE LABORATORY Dr Tom Hartley Quality Manager : Royal Hobart Hospital Senior Research Fellow : University of Tasmania

MAINTENANCE OF QUALITY IN THE LABORATORY Dr Tom Hartley Quality Manager : Royal Hobart Hospital Senior Research Fellow : University of Tasmania MAINTENANCE OF QUALITY IN THE LABORATORY Dr Tom Hartley Quality Manager : Royal Hobart Hospital Senior Research Fellow : University of Tasmania AUSTRALIA Sponsor : Nancy Dale Scholarship, AACB Objectives

More information

QUICK REFERENCE INSTRUCTIONS For use with Sofia and Sofia 2. Rx only

QUICK REFERENCE INSTRUCTIONS For use with Sofia and Sofia 2. Rx only QUICK REFERENCE INSTRUCTIONS For use with Sofia and Sofia 2. Rx only Test Procedure Study the Package Insert and User Manual thoroughly before using Quick Reference Instructions. This is not a complete

More information

Purpose: Career Ladders formally advance employees through higher levels of job proficiency and responsibility.

Purpose: Career Ladders formally advance employees through higher levels of job proficiency and responsibility. Career Ladder Editable Template This template is meant to be a guideline only, and can be edited to fit your organizational requirements or limitations. Job Classification: Laboratory Scientist Brief Description

More information

What is an LDT? The Do s s And Don ts of Validating Laboratory Developed Tests

What is an LDT? The Do s s And Don ts of Validating Laboratory Developed Tests The Do s s And Don ts of Validating Laboratory Developed Tests Deirdre Astin, MS, MT (ASCP) Lab Quality Confab 2011 November 2011 San Antonio Texas What is an LDT? Laboratory developed tests (LDTs) are

More information

Asian Journal of Research in Biological and Pharmaceutical Sciences Journal home page:

Asian Journal of Research in Biological and Pharmaceutical Sciences Journal home page: Research Article ISSN: 2349 4492 Asian Journal of Research in Biological and Pharmaceutical Sciences Journal home page: www.ajrbps.com IDENTIFICATION OF THE PRE-ANALYTICAL ERRORS AMONG GOVERNMENT HOSPITALS

More information

Internal Quality Control Practices in Coagulation Laboratories: recommendations based on a patterns-of-practice survey

Internal Quality Control Practices in Coagulation Laboratories: recommendations based on a patterns-of-practice survey International Journal of Laboratory Hematology ORIGINAL ARTICLE The Official journal of the International Society for Laboratory Hematology INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY Internal Quality

More information

Table of Contents. Proficiency Testing (PT) Manual. 4 How to Interpret Your Evaluation General Customer Information...

Table of Contents. Proficiency Testing (PT) Manual. 4 How to Interpret Your Evaluation General Customer Information... 2017Pr o ci encytest i ngmanual Per f or manceyoucanmeasur e Accur acyyoucant r ust Table of Contents Proficiency Testing (PT) Manual 1 General Customer Information... 3 10 Order Confirmation...3 Regulatory

More information

Guidance for Industry and FDA Staff. Replacement Reagent and Instrument Family Policy

Guidance for Industry and FDA Staff. Replacement Reagent and Instrument Family Policy Guidance for Industry and FDA Staff Replacement Reagent and Instrument Family Policy Document issued on: December 11, 2003 This document supersedes Data for Commercialization of Original Equipment Manufacturers

More information

Inspecting Cytopathology

Inspecting Cytopathology Inspecting Cytopathology Managing the Inspection Process I. Introduction Conduct a brief interview with the Lab director and/or Cytology supervisor to pose some QM related questions. Give your host a prepared

More information

AAG PRINCIPLE SPECIMEN. REF (150 tests) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

AAG PRINCIPLE SPECIMEN. REF (150 tests) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE IMMAGE Immunochemistry Systems Chemistry Information Sheet Copyright 2010 Beckman Coulter, Inc. Alpha 1 -Acid Glycoprotein REF 447780 (150 tests) For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by:

More information

Virginia s Health Competencies for Direct Support Professionals and Supervisors who support individuals with Developmental Disabilities

Virginia s Health Competencies for Direct Support Professionals and Supervisors who support individuals with Developmental Disabilities Health Competencies Checklist (rev. 6.10.16) Virginia s Health Competencies for Direct Support Professionals and Supervisors who support individuals with Developmental Disabilities The Direct Support Professional

More information

APFCB. Improving Laboratory Performance Through Quality Control. Designing an Appropriate QC Procedure for Your Laboratory QUALITY CONTROL

APFCB. Improving Laboratory Performance Through Quality Control. Designing an Appropriate QC Procedure for Your Laboratory QUALITY CONTROL APFCB Improving Laboratory Performance Through Quality Control Designing an Appropriate QC Procedure for Your Laboratory - Five simple steps for QC success By Sarah Kee, BSc PGCE QC Scientific Consultant,

More information

National External Quality Assurance Program Pakistan (NEQAPP) a milestone in proficiency testing in Pakistan

National External Quality Assurance Program Pakistan (NEQAPP) a milestone in proficiency testing in Pakistan National External Quality Assurance Program Pakistan (NEQAPP) a milestone in proficiency testing in Pakistan Muhammad Usman Munir 1, Aamir Ijaz 2 1 Department of Pathology, Combined Military Hospital,

More information

EUROPEAN COMMISSION DIRECTORATE-GENERAL FOR HEALTH AND FOOD SAFETY. EudraLex. Volume 4

EUROPEAN COMMISSION DIRECTORATE-GENERAL FOR HEALTH AND FOOD SAFETY. EudraLex. Volume 4 EUROPEAN COMMISSION DIRECTORATE-GENERAL FOR HEALTH AND FOOD SAFETY Ref. Ares(2015)4234460-12/10/2015 Medicinal products quality, safety and efficacy Brussels, 12 October 2015 EudraLex Volume 4 EU Guidelines

More information

Quality Control and Quality Assurance for Antibiotic Testing

Quality Control and Quality Assurance for Antibiotic Testing Quality assurance Quality Control and Quality Assurance for Antibiotic Testing 26 Sep 2013 Microbiology Technical Workshop Lily Ng Siew Yong practice of assessing performance in all steps of the process

More information

The integrity of your research depends on the integrity of your cell line.

The integrity of your research depends on the integrity of your cell line. The integrity of your research depends on the integrity of your cell line. How well do you know your cell line? Despite the best efforts of researchers, a signifi cant number of cell lines are contaminated

More information

SAMPLE PROCEDURE , 11/07

SAMPLE PROCEDURE , 11/07 SAMPLE PROCEDURE This Sample Procedure is not intended as a substitute for your facility s Procedure Manual or reagent labeling, but rather as a model for your use in customizing for your laboratory s

More information

Medical Microbiology Laboratory. Quality Management for the. Roberta B. Carey, Ph.D., D(ABMM), CMQ/OE(ASQ)

Medical Microbiology Laboratory. Quality Management for the. Roberta B. Carey, Ph.D., D(ABMM), CMQ/OE(ASQ) Analysis. Answers. Action. www.aphl.org Quality Management for the Medical Microbiology Laboratory Roberta B. Carey, Ph.D., D(ABMM), CMQ/OE(ASQ) Director, Laboratory Quality Management Program Centers

More information

Quality Manual. Index

Quality Manual. Index This quality policy manual is for the use of, its Clients, Vendors, and the appropriate Regulatory Agencies. Unauthorized duplication or transfer to others is prohibited. Index Authorization... 3 Introduction...

More information

DEPARTMENT OF CLINICAL LABORATORY SCIENCES SCHOOL OF HEALTH TECHNOLOGY AND MANAGEMENT THE UNIVERSITY AT STONY BROOK STONY BROOK, NEW YORK

DEPARTMENT OF CLINICAL LABORATORY SCIENCES SCHOOL OF HEALTH TECHNOLOGY AND MANAGEMENT THE UNIVERSITY AT STONY BROOK STONY BROOK, NEW YORK DEPARTMENT OF CLINICAL LABORATORY SCIENCES SCHOOL OF HEALTH TECHNOLOGY AND MANAGEMENT THE UNIVERSITY AT STONY BROOK STONY BROOK, NEW YORK 11794-8205 MICROBIOLOGY COMPETENCY EVALUATION FORM STUDENT NAME:

More information

Note: The taller 100 x 13 mm tube is not designed to run on the ACL TOP. Plasma after centrifugation of these tubes must be placed in sample cups.

Note: The taller 100 x 13 mm tube is not designed to run on the ACL TOP. Plasma after centrifugation of these tubes must be placed in sample cups. Created Revised Reviewed Approved Lupus Insensitive aptt on the ACL TOP HEM 4.15.1 7/13/2012 4/11/2013 4/11/2013 4/11/2013 I. PRINCIPLE The activated partial thromboplastin time is a global screening procedure

More information

Herpes Simplex Virus Type 1 (HSV-1) IgM ELISA Kit Protocol

Herpes Simplex Virus Type 1 (HSV-1) IgM ELISA Kit Protocol Herpes Simplex Virus Type 1 (HSV-1) IgM ELISA Kit Protocol (Cat. No.:EK-310-93) 330 Beach Road, Burlingame CA Tel: 650-558-8898 Fax: 650-558-1686 E-Mail: info@phoenixpeptide.com www.phoenixpeptide.com

More information

CSA Group Products, Processes and Services Scheme

CSA Group Products, Processes and Services Scheme 1. CONTACT CSA CSA Group Products, Processes and Services Scheme 1.1 Getting started is easy. There are three ways to contact us: Call 866 797 4272 Email client.services@csagroup.org Complete a request

More information

'How-To' Implement an Effective Proactive Risk Management Strategy

'How-To' Implement an Effective Proactive Risk Management Strategy 'How-To' Implement an Effective Proactive Management Strategy Sedef YENICE Satellite Educational Workshop on Intelligent Clinical Laboratory Management: Impacts on Quality System Improvement Hilton Durban

More information

Evolve your lab in 4 steps

Evolve your lab in 4 steps presents Evolve your lab in 4 steps Take Sample Management from Good to Great 1 ebook Contents Here s a quick preview of what you can expect in the pages ahead: Part 1: Rethinking Your Sample-Handling

More information

Sysmex Introduction and Africa Overview. Eric Osei General Manager West & Central Africa

Sysmex Introduction and Africa Overview. Eric Osei General Manager West & Central Africa Sysmex Introduction and Africa Overview Eric Osei General Manager West & Central Africa Sysmex Corporation, Japan Head Office Established 1968 Business Field Position in the IVD Kobe, Japan IVD, IT, Scientific

More information

Responding to an FDA 483

Responding to an FDA 483 Responding to an FDA 483 Jim Melancon VP, Associate General Counsel BioScrip, Inc. Marc Stranz, PharmD Healthcare Consultant Disclosure The speakers declare no conflicts of interest or financial interest

More information

Implementing 7 Day Platelet Dating with the Platelet PGD Test

Implementing 7 Day Platelet Dating with the Platelet PGD Test Implementing 7 Day Platelet Dating with the Platelet PGD Test The Verax Biomedical Platelet Pan Genera Detection (PGD ) Test is a rapid, qualitative immunoassay for the detection of aerobic and anaerobic

More information

Background Verification Supplement

Background Verification Supplement Background Verification Supplement Dartmouth s largest investment is in human capital its workforce. As with any significant investment, it is important to identify the information Dartmouth needs to make

More information

Metrics in Microbiology Monitoring Practices

Metrics in Microbiology Monitoring Practices Metrics in Microbiology Monitoring Practices Crystal Booth, M.M. Masters of Microbiology from North Carolina State University Former Associate Director of Microbiology at Novartis Metrics in Microbiology

More information

Quality Management. Carlos Bachier, MD

Quality Management. Carlos Bachier, MD Quality Management Carlos Bachier, MD Goals of Quality Management Ensure credibility of outcomes Improve patient safety and quality of processes Significantly reduce errors Quality Medical and Laboratory

More information

Questions and Answers for UC Faculty Regarding UC's Labor Contract with Its Teaching Assistants

Questions and Answers for UC Faculty Regarding UC's Labor Contract with Its Teaching Assistants Questions and Answers for UC Faculty Regarding UC's Labor Contract with Its Teaching Assistants In the spring of 2000, University of California academic student employees (ASEs) ratified a labor contract

More information

Building and Validating an Autoverification System in the Clinical Chemistry Laboratory

Building and Validating an Autoverification System in the Clinical Chemistry Laboratory Building and Validating an Autoverification System in the Clinical Chemistry Laboratory Mu-Chin Shih, MT, 1,2 Huey-Mei Chang, MS, 6 Ni Tien, MS, 1,2 Chiung-Tzu Hsiao, MS, 1,5 Ching-Tien Peng, MD 3,4 (

More information

Judith Yost, MA, MT(ASCP) Director, Division of Laboratory Services. Laboratory Certification Intersection w/ Quality Management Systems

Judith Yost, MA, MT(ASCP) Director, Division of Laboratory Services. Laboratory Certification Intersection w/ Quality Management Systems Judith Yost, MA, MT(ASCP) Director, Division of Laboratory Services Laboratory Certification Intersection w/ Quality Management Systems Topics for Discussion Concepts, Key Standards & Intent of Laboratory

More information

I-STAT POCT System PRINCIPLE:

I-STAT POCT System PRINCIPLE: PRINCIPLE: I-STAT POCT System The i-stat system uses a single disposable cartridge. The cartridge, which has been filled with fresh blood, is placed into a hand-held analyzer for analysis. The results

More information

WHO Prequalification of In Vitro Diagnostics Programme

WHO Prequalification of In Vitro Diagnostics Programme P r e q u a l i f i c a t i o n T e a m - D i a g n o s t i c s Information for Manufacturers on the Manufacturing Site(s) Inspection (Assessment of the Quality Management System) WHO Prequalification

More information

A Risk-based Approach for In Vitro Companion Diagnostics Device FDA Approval Process Associated with Therapies that have Breakthrough Designation

A Risk-based Approach for In Vitro Companion Diagnostics Device FDA Approval Process Associated with Therapies that have Breakthrough Designation A Risk-based Approach for In Vitro Companion Diagnostics Device FDA Approval Process Associated with Therapies that have Breakthrough Designation A Risk-based Approach for In Vitro Companion Diagnostics

More information

A Verification Study for Implementing the Revised CLSI Breakpoints. Summary. Breakpoint Differences Cephalosporin Breakpoints for Enterobacteriaceae

A Verification Study for Implementing the Revised CLSI Breakpoints. Summary. Breakpoint Differences Cephalosporin Breakpoints for Enterobacteriaceae A Verification Study for Implementing the Revised CLSI Breakpoints Jean B. Patel, PhD, D(ABMM) Deputy Director, Office of Antimicrobial Resistance National Center for Emerging and Zoonotic Infectious Disease

More information

A Verification Study for Implementing the Revised CLSI Breakpoints. Summary. Glossary CDC 1

A Verification Study for Implementing the Revised CLSI Breakpoints. Summary. Glossary CDC 1 A Verification Study for Implementing the Revised CLSI Breakpoints Jean B. Patel, PhD, D(ABMM) Deputy Director, Office of Antimicrobial Resistance National Center for Emerging and Zoonotic Infectious Disease

More information

TEST REQUISITION, PATIENT INFORMATION, AND CONSENT HUDSONALPHA CLINICAL SERVICES LABORATORY

TEST REQUISITION, PATIENT INFORMATION, AND CONSENT HUDSONALPHA CLINICAL SERVICES LABORATORY 1. Patient Information Patient Name TEST REQUISITION, PATIENT INFORMATION, AND CONSENT HUDSONALPHA CLINICAL SERVICES LABORATORY HudsonAlpha Clinical Services Lab, LLC HUDSONALPHA CLINICAL SERVICES LABORATORY

More information

Directions: OK, 1. Employment specialists may have caseloads: 2. Employment specialists: 3. In the IPS model, benefits counseling is considered:

Directions: OK, 1. Employment specialists may have caseloads: 2. Employment specialists: 3. In the IPS model, benefits counseling is considered: IPS-Q Directions: The Individual Placement and Support (IPS) model of supported employment is a program that provides employment services to persons with a mental illness. For each question, please select

More information

CRI Symposium for Clinical Laboratories

CRI Symposium for Clinical Laboratories CRI Symposium for Clinical Laboratories Where education meets imagination! OCTOBER 15-18, 2014 Buena Vista Palace Hotel & Spa Orlando, FL This was the best educational experience I have ever had. Page

More information

QUALITY ASSURANCE IN MICROBIOLOGY

QUALITY ASSURANCE IN MICROBIOLOGY Indian Journal of Medical Microbiology, (2004) 22 (2):81-86 Special Article QUALITY ASSURANCE IN MICROBIOLOGY DR Arora Abstract Quality assurance (QA) is the total process whereby the quality of laboratory

More information

The Advantages of Automatic Quality Control (AQC) for Blood Gas Testing QC

The Advantages of Automatic Quality Control (AQC) for Blood Gas Testing QC Siemens Healthcare Diagnostics, the leading clinical diagnostics company, is committed to providing clinicians with the vital information they need for the accurate diagnosis, treatment and monitoring

More information

9 February CC: Dr. Christopher E. Urbina, MD, MPH Executive Director and Chief Medical Officer. Dear Mr. Butcher:

9 February CC: Dr. Christopher E. Urbina, MD, MPH Executive Director and Chief Medical Officer. Dear Mr. Butcher: James A. Ruth, Ph.D., D-ABFT Professor Emeritus and Adjunct Professor University of Colorado School of Pharmacy, Anschutz Medical Campus Director, Forensic Laboratories, Denver 9 February 2013 Mr. David

More information

Replacing Analytical Methods for Release and Stability Testing CBER Perspective

Replacing Analytical Methods for Release and Stability Testing CBER Perspective Replacing Analytical Methods for Release and Stability Testing CBER Perspective Presentation at the CMC Strategy Forum January 27, 2014 Lokesh Bhattacharyya Chief, Lab of Analytical Chemistry and Blood

More information

POLICIES AND PROCEDURES MANUAL CLIA #01D Sharon P. Massingale, Ph.D., HCLD (ABB) Laboratory Director

POLICIES AND PROCEDURES MANUAL CLIA #01D Sharon P. Massingale, Ph.D., HCLD (ABB) Laboratory Director POLICIES AND PROCEDURES MANUAL CLIA #01D0665512 Sharon P. Massingale, Ph.D., HCLD (ABB) Laboratory Director Prepared by: Bureau of Clinical Laboratories Quality Management Division December 2014 ALABAMA

More information

SUMMARY AND EXPLANATION

SUMMARY AND EXPLANATION This Procedural Bulletin is intended to provide a ready outline reference for performance of the assay. These abbreviated directions for use are not intended to replace the complete package insert. It

More information

Facility and Equipment Maintenance

Facility and Equipment Maintenance Facility and Equipment Maintenance Deborah L. Griffin Manager, Quality Assurance for Cellular Therapies Hillman Cancer Center University of Pittsburgh Cancer Institute University of Pittsburgh Medical

More information

Association of American Railroads Quality Assurance System Evaluation (QASE) Checklist Rev. 1/12/2017

Association of American Railroads Quality Assurance System Evaluation (QASE) Checklist Rev. 1/12/2017 Company: Prepared By: Date: Changes from previous version highlighted in yellow. Paragraph Element Objective Evidence 2.1 Objective of Quality Assurance Program 2.2 Applicability and Scope 2.3 QA Program

More information

The Risk Management + Design Controls Connection: What Device Makers Need to Know

The Risk Management + Design Controls Connection: What Device Makers Need to Know !!! The Risk Management + Design Controls Connection: What Device Makers Need to Know Jon Speer Founder & VP of QA/RA greenlight.guru Table of Contents 1 Intended Use & User Needs 6 Verification, Validation,

More information

Good Clinical Practice (GCP): how does a laboratory that supports a clinical trial ensure compliance with these Regulations

Good Clinical Practice (GCP): how does a laboratory that supports a clinical trial ensure compliance with these Regulations Good Clinical Practice (GCP): how does a laboratory that supports a clinical trial ensure compliance with these Regulations 9th edition of Synevo Clinical Research Symposium. 17 th November 2017 Bucharest.

More information

Bill Kabat and Carmen Irizarry of the ACTG/IMPAACT Lab Tech Committee. Network Name, Title Signature Date

Bill Kabat and Carmen Irizarry of the ACTG/IMPAACT Lab Tech Committee. Network Name, Title Signature Date Title: Dried PBMC Pellet Processing Standard Operating Procedure (LDMS specimen code = BLD/EDTA or ACD/PEL) Origination Date: 01 Jun 2004 Total Pages: 6 Effective Date: 29 Nov 2012 SOP Number LTC-SOP-59

More information

Immunoassay Kit Catalog # KCA0021. Canine. C-Reactive Protein

Immunoassay Kit Catalog # KCA0021. Canine. C-Reactive Protein Immunoassay Kit Catalog # KCA0021 Canine C-Reactive Protein BioSource International, Inc. 542 Flynn Road Camarillo, California 93012 USA Tel: 805-987-0086 800-242-0607 FAX: 805-987-3385 email: tech.support@biosource.com

More information

IgG Antibodies To Toxoplasma. Gondii ELISA Kit

IgG Antibodies To Toxoplasma. Gondii ELISA Kit IgG Antibodies To Toxoplasma Catalog No: IRAPKT1405 Gondii ELISA Kit Lot No: SAMPLE INTENDED USE The Toxoplasma IgG ELISA is intended for use in evaluating a patient s serologic status to Toxoplasma gondii

More information

A24 CALA Checklist for Microbiology Revision 4.1 February 16, 2018

A24 CALA Checklist for Microbiology Revision 4.1 February 16, 2018 A24 CALA Checklist for Microbiology Revision 4.1 February 16, 2018 Laboratory Name: Appendix Name: Appendix Number: Assessor: Date: Rev. 4.1 CALA CHECKLIST FOR MICROBIOLOGY Item 01 SAMPLING AND SAMPLE

More information

BULLETIN 15 (Publication 35) Qualified Products List for Construction

BULLETIN 15 (Publication 35) Qualified Products List for Construction Bulletin 15 Description Bulletin 15 is a listing of prequalified materials that are eligible for use on Department construction projects. The purpose of Bulletin 15 is to provide contractors, consultants,

More information

How to build a Quality & Safety System

How to build a Quality & Safety System How to build a Quality & Safety System 29 November 2011 How to build a Quality & Safety System Helpful stuff for operators who have not previously run a formal management system In many cases organisations

More information

IRVINE UNIFIED SCHOOL DISTRICT CONTRACTOR S PREQUALIFICATION QUESTIONNAIRE, 20

IRVINE UNIFIED SCHOOL DISTRICT CONTRACTOR S PREQUALIFICATION QUESTIONNAIRE, 20 CONTRACTOR S PREQUALIFICATION QUESTIONNAIRE, 20 The Irvine Unified School District ( District ) has determined that bidders on an annual basis must be prequalified prior to submitting a bid on a project

More information

Verifying New Reagent Lot Performance

Verifying New Reagent Lot Performance Verifying New Reagent Lot Performance Julianne Addison Application Consultant, Siemens RSC Answers for life. Objectives Describe issues relating to changing reagent lots Discuss the reasons QC material

More information

IgG Antibodies TO Rubella Virus ELISA Kit Protocol

IgG Antibodies TO Rubella Virus ELISA Kit Protocol IgG Antibodies TO Rubella Virus ELISA Kit Protocol (Cat. No.:EK-310-81) 330 Beach Road, Burlingame CA Tel: 650-558-8898 Fax: 650-558-1686 E-Mail: info@phoenixpeptide.com www.phoenixpeptide.com INTENDED

More information

Illumina Clinical Services Laboratory

Illumina Clinical Services Laboratory Illumina Clinical Services Laboratory CLIA Certificate No.: 05D1092911 Illumina Clinical Services Laboratory TruGenome Technical Sequence Data Test Requisition Form The Illumina Clinical Services Laboratory

More information

Medical Device Regulatory Framework 9 SEPTEMBER 2015 FUNDISA CONFERENCE JANE ROGERS

Medical Device Regulatory Framework 9 SEPTEMBER 2015 FUNDISA CONFERENCE JANE ROGERS Medical Device Regulatory Framework 9 SEPTEMBER 2015 FUNDISA CONFERENCE JANE ROGERS Key Topics Definitions Essential Principles Classification Conformity Assessment Framework License to Manufacture, Import,

More information

CLIA Complexity: MODERATE

CLIA Complexity: MODERATE CLIA Complexity: MODERATE INTENDED USE The is intended for the rapid detection of Group A Streptococcal antigen directly from throat swabs and beta-hemolytic colonies recovered from culture. This test

More information

DESIGN & CONSTRUCTION PHASE COMMISSIONING PLAN TEMPLATE

DESIGN & CONSTRUCTION PHASE COMMISSIONING PLAN TEMPLATE DESIGN & CONSTRUCTION PHASE COMMISSIONING PLAN TEMPLATE Based upon B3 Minnesota Sustainable Building Guidelines VERSION 2.1 Notes to the reader have been added to this document within numerous text boxes

More information

NBS Molecular Training Workshop

NBS Molecular Training Workshop Rachel Lee, Ph.D. Texas Department of State Health Services NBS Molecular Training Workshop March 12, 2015 Quality Management Components Organization Personnel Documents and Records Advisory Services Laboratory

More information