Standardization of Laboratory Tests - How to do it

Size: px
Start display at page:

Download "Standardization of Laboratory Tests - How to do it"

Transcription

1 Standardization of Laboratory Tests - How to do it Greg Miller, PhD Virginia Commonwealth University Medical Center Richmond, Virginia, USA greg.miller@vcuhealth.org

2 Outline What is standardization Why is it important How is it accomplished How are challenges being addressed

3 What is standardization Equivalent results, within clinically meaningful limits, among different measurement procedures for the same laboratory test

4 Terminology Standardization: achieving equivalent results by having calibration traceable to a JCTLM listed reference system component Harmonization: achieving equivalent results among different measurement procedures Usually implies there is no JCTLM listed reference measurement procedure or certified reference material

5 Why is it important Patients will get the correct treatment Many medical decisions are informed by laboratory results Many clinical guidelines use a fixed laboratory test value for treatment decisions

6 How to achieve equivalent results 1. Calibration of all measurement procedures is traceable to a common reference system ISO 17511:2003 (under revision) 2. All measurement procedures measure the same quantity (the same molecular form) Analytical selectivity for the measurand

7 ISO Standards 17511:2003, Calibration Traceability 18153:2003, Traceability for Enzymes 15193:2009, Reference Measurement Procedures 15194:2009, Certified Reference Materials 15195:2003, Reference Measurement Laboratories

8 Joint Committee for Traceability in Laboratory Medicine Lists reference materials, reference measurement procedures and reference laboratories that conform to the ISO Standards

9 TRACEABILITY Panel of Patient Samples Primary Reference Material (pure substance) Pure Substance Calibrator Secondary Reference Material (matrix) Manufacturer s Product Calibrator Procedures for identity and mass balance Reference Measurement Procedure (e.g. Gravimetry) Reference Measurement Procedure (e.g. IDMS) Manufacturer s Internal Procedures Medical Laboratory Procedure Patient s Sample Patient s Result

10 Measurands for which reference procedures exist or can be developed Everything else we measure in the clinical laboratory

11 What happens when there is no reference measurement procedure

12 TRACEABILITY Traceability is established to a reference material Secondary Reference Material (matrix) Manufacturer s Product Calibrator Patient s Sample Manufacturer s Internal Procedures Medical Laboratory Procedure Patient s Result

13 TRACEABILITY Value assignment Commutability Secondary Reference Material (matrix) Manufacturer s Product Calibrator Patient s Sample Manufacturer s Internal Procedures Medical Laboratory Procedure Patient s Result

14 What happens when there is both: no reference measurement procedure no reference material

15 TRACEABILITY TRACEABILITY TRACEABILITY TRACEABILITY Traceability is established to to a material selected for a measurement procedure No coordination among producers (IVD or LDT) Manufacturer s Working Calibrator (master lot) Manufacturer s Product Calibrator Patient s Sample Manufacturer s Internal Procedure Medical Laboratory Procedure?????????? Patient s Result

16 The commutability challenge

17 TRACEAB ILITY Primary Reference Material (pure substance) Pure Substance Calibrator Secondary Reference Material (matrix) Manufacturer s Product Calibrator Patient s Sample Procedures for identity and mass balance Reference Measurement Procedure (e.g. Gravimetry) A non-commutable calibrator breaks the traceability chain Key step for commutability Reference Measurement Procedure (e.g. IDMS) Manufacturer s Internal Procedures Medical Laboratory Procedure Patient s Result

18 TRACEAB ILITY Primary Reference Material (pure substance) Pure Substance Calibrator Secondary Reference Material (matrix) Key step for commutability Manufacturer s Product Calibrator Patient s Sample Procedures for identity and mass balance Even though manufacturers show traceability, the Reference process fails to provide equivalent Measurement results for patient Procedure (e.g. Gravimetry) samples among different measurement procedures Reference Measurement Procedure (e.g. IDMS) Manufacturer s Internal Procedures Medical Laboratory Procedure Patient s Result

19 Commutability (Commutable) Property of a reference material demonstrated by the closeness of agreement between the relation among results for a reference material obtained from two measurement procedures and the relation among results for clinical samples from the same two measurement procedures (Rephrased from VIM 3: 2008)

20 Measurement Procedure 2 Commutable: same relationship for clinical samples and reference materials Clinical Samples Reference Materials Measurement Procedure 1

21 Measurement Procedure 2 Non-commutable: different relationship for clinical samples and reference materials Clinical Samples Reference Materials Measurement Procedure 1

22 Measurement Procedure 2 Calibration with non-commutable materials Clinical Samples RM as Calibrator Measurement Procedure 1 causes patient sample results to be different

23 Guidelines are available from CLSI: EP30-A Characterization and qualification of commutable reference materials for laboratory medicine (2010 as C53-A) EP14-A3 Evaluation of commutability of processed samples (2014)

24 Measurement Procedure 2 Result 50 Deming Regression Line 40 95% Prediction Interval Commutable 30 Commutable 20 Non-commutable 10 0 Native Clinical Samples Reference Materials Measurement Procedure 1 Result Adapted from CLSI EP30-A (used with permission)

25 Measurement Procedure 2 Result 50 Deming Regression Line 40 95% Prediction Interval Native Clinical Samples Reference Materials Measurement Procedure 1 Result Criteria for commutability near decision thresholds not resolved Adapted from CLSI EP30-A (used with permission)

26 Measurement Procedure 2 Result % Prediction Interval Deming Regression Line Native Clinical Samples Reference Materials Measurement Procedure 1 Result More scatter in the relationship should not make the RMs commutable Modified from CLSI EP30-A (used with permission)

27 IFCC Working Group on Commutability (established March 2013) Operating procedures for the formal assessment of commutability Criteria for commutability taking into account the intended use of a reference material

28 Criteria for commutability 1. Criteria based on statistical distribution of results for patient samples do not relate to the intended use of a RM o Criteria change among measurement procedures with different performance characteristics Method D Method B Method C Method A

29 Criteria for commutability 2. Recommend fixed criteria based on the intended use of the reference material Method D Method B Method C Method A

30 Intended use refers to the allowable total error (TE a ) for a medical decision Fraction of the uncertainty required for a RM s use in a calibration traceability hierarchy Fraction of the uncertainty required for assessment of performance using EQA

31 Statistical models 1. Assess the closeness of agreement for the difference in bias for RM compared to clinical samples between measurement procedures 2. Assess harmonization effectiveness of a RM used for calibration traceability by a group of measurement procedures

32 Difference in bias model 1. Estimate the bias between 2 measurement procedures for the patient samples and for the reference material(s) 2. Estimate the random error components including sample specific effects 3. Calculate the difference in bias for reference material(s) vs. patient samples 4. Estimate the uncertainty of the difference in bias 5. Commutable if the difference in bias plus uncertainty are within a criterion that is suitable for the intended use of the reference material

33 Difference in bias vs. fixed criteria Bias ( y- x) CS RM Bias for CS Crit eria Crit eria Mean Concent ration ( x+ y) / 2

34 Difference in bias vs. fixed criteria Bias ( y- x) Mean Concent ration ( x+ y) / 2 Commutable: bias plus uncertainty is within criteria CS RM Bias for CS Crit eria Crit eria

35 Difference in bias vs. fixed criteria Non-commutable: bias plus uncertainty exceeds criteria Bias ( y- x) CS RM Bias for CS Crit eria Crit eria Mean Concent ration ( x+ y) / 2

36 Difference in bias vs. fixed criteria Indeterminate: bias plus uncertainty overlaps criteria Bias ( y- x) CS RM Bias for CS Crit eria Crit eria Mean Concent ration ( x+ y) / 2

37 Harmonization effectiveness model 1. Estimate the inter-measurement procedure CV for each clinical sample s results 2. Calculate an overall pooled inter-measurement procedure CV and its uncertainty for all clinical samples 3. Compare the pooled CV plus uncertainty to a fixed criterion that is suitable for the intended use of the reference material 4. Use the RM for calibration traceability and repeat steps 1-2-3

38 Harmonization effectiveness Current state Inter-Measurement Procedure CV (%) 20% 10% 5% Criterion 0%

39 Harmonization effectiveness Current state Inter-Measurement Procedure CV (%) 20% RM 1 as calibrator Not Commutable 10% 5% Criterion 0%

40 Harmonization effectiveness Current state Inter-Measurement Procedure CV (%) 20% RM 1 as calibrator Not Commutable 10% RM 2 as calibrator Indeterminate 5% Criterion 0%

41 Harmonization effectiveness Current state Inter-Measurement Procedure CV (%) 20% RM 1 as calibrator Not Commutable 10% RM 2 as calibrator RM 3 as calibrator Indeterminate Commutable 5% Criterion 0%

42 Other commutability topics 1. Qualify measurement procedures for inclusion 2. Individual samples vs. pools 3. Replacement of a RM with a new lot or batch 4. Correction to the assigned value of a noncommutable RM

43 TRACEABILITY TRACEABILITY TRACEABILITY TRACEABILITY The harmonization challenge Manufacturer s Working Calibrator (master lot) Manufacturer s Product Calibrator Patient s Sample Manufacturer s Internal Procedure Medical Laboratory Procedure?????????? Patient s Result

44 ISO TC 212 WG2 Revision of the traceability standard is expected to include traceability to a harmonization protocol as one type of calibration traceability hierarchy

45 A new standard is in development to support JCTLM listing of a harmonization protocol. ISO 20089: Requirements for international harmonization protocols intended to establish metrological traceability of values assigned to product (end user) calibrators and patient samples

46 TRACEABILITY Harmoniza on Reference Material Assign value Calibrate Harmoniza on Protocol Manufacturer s Working Calibrator (master lot) Manufacturer s Product Calibrator Assign value Calibrate Assign value Manufacturer s Selected Procedure Manufacturer s Standing Procedure Pa ent s Sample Calibrate Assign value Medical Laboratory Procedure Pa ent s Result

47 Critical components for harmonization 1. Calibration of all measurement procedures is traceable to a common reference protocol 2. All measurement procedures measure the same quantity (the same molecular form) 3. Traceability can be sustained over time and location

48 Example 1: harmonization protocol Clinical samples Process for value assignment Reserve set of clinical samples for sustainability End user calibrator End user calibrator End End user user calibrator product calibrator Equivalent results for clinical samples Assign the value of each end-user product calibrator to recover the values for clinical samples End user End user calibrator calibrator End Medical user lab measurement calibrator procedure Supported by producer (mfr or LDT) master lot(s) with appropriate target values EQA for surveillance

49 Example 2: harmonization protocol Clinical samples Process for value assignment Specification to recreate the clinical samples End user calibrator End user calibrator End End user user calibrator product calibrator Equivalent results for clinical samples Assign the value of each end-user product calibrator to recover the values for clinical samples End user End user calibrator calibrator End Clinical user lab measurement calibrator procedure Supported by producer (mfr or LDT) master lot(s) with appropriate target values EQA for surveillance

50 The prioritization and coordination challenge

51 International Forum organized by AACC in October, participants from 12 countries Representing 62 organizations & manufacturers

52 The Roadmap Develop an infrastructure to coordinate harmonization activities world wide: 1. Prioritize measurands by medical importance 2. Coordinate the work of different organizations 3. Develop technical processes to achieve harmonization when there is no reference measurement procedure or reference material 4. Promote surveillance of the success of harmonization

53

54 International Consortium for Harmonization of Clinical Laboratory Results Strategic Partners Group Council Organizational Member Harmonization Oversight Group Harmonization Implementation Groups Special Working Groups

55 ICHCLR: Council members

56

57

58

59 Perfect is the Enemy of Good

Harmonization: why you should care!

Harmonization: why you should care! Harmonization: why you should care! Greg Miller, PhD, DABCC Virginia Commonwealth University Richmond, Virginia gmiller@vcu.edu University of Utah, ARUP, April 20, 2017 Financial disclosures Siemens Healthcare

More information

What is the purpose of this website?

What is the purpose of this website? What is the purpose of this website? This website aims to explain: How it is possible for different methods to give different results for the same test on the same patient sample Why it is important to

More information

Examples of reference materials

Examples of reference materials Examples of reference materials JCTLM Working Group for Traceability: Education and Promotion (WG-TEP) Area 2: Mini-presentations to explain scientific concepts What are reference materials? Reference

More information

Clinical Chemistry Approach to Evaluation of Commutability

Clinical Chemistry Approach to Evaluation of Commutability Clinical Chemistry Approach to Evaluation of Commutability Hubert W. Vesper, Ph.D. Director, Clinical Standardization Programs Division of Laboratory Sciences Centers for Disease Control and Prevention,

More information

Metrological Traceability and Assay Standardization in Laboratory Medicine. STRESA, ITALY May 24, 2013

Metrological Traceability and Assay Standardization in Laboratory Medicine. STRESA, ITALY May 24, 2013 Metrological Traceability and Assay Standardization in Laboratory Medicine STRESA, ITALY May 24, 2013 Standardization of ctni: Is there a light at the end of the tunnel? Jill Tate Pathology Queensland

More information

International Consortium for Harmonization of Clinical Laboratory Results

International Consortium for Harmonization of Clinical Laboratory Results Toolbox of technical procedures for harmonization Page 1 of 33 International Consortium for Harmonization of Clinical Laboratory Results Toolbox of technical procedures to be considered when developing

More information

JCTLM Member Activities

JCTLM Member Activities JCTLM Members' and Stakeholders' Meeting 30 November -1 December 2015 BIPM, Sevres JCTLM Member Activities Prof Mauro Panteghini CIRME Scientific Coordinator Research Centre for Metrological Traceability

More information

SAMPLE. Evaluation of Total Analytical Error for Quantitative Medical Laboratory Measurement Procedures

SAMPLE. Evaluation of Total Analytical Error for Quantitative Medical Laboratory Measurement Procedures 2nd Edition EP21 Evaluation of Total Analytical Error for Quantitative Medical Laboratory Measurement Procedures This guideline provides manufacturers and end users with an understanding of concepts related

More information

Reference systems and international standardisation of measurements for clinically important enzymes

Reference systems and international standardisation of measurements for clinically important enzymes Reference systems and international standardisation of measurements for clinically important enzymes VII th Czech National Congress of Clinical Biochemistry (11. 13.9.2005, Olomouc) Gerhard Schumann, Klinische

More information

A2LA. R231 Specific Requirements: Threat Agent Testing Laboratory Accreditation Program. December 6, 2017

A2LA. R231 Specific Requirements: Threat Agent Testing Laboratory Accreditation Program. December 6, 2017 Laboratory Page 1 of 17 Laboratory December 6, 2017 2017 by A2LA All rights reserved. No part of this document may be reproduced in any form or by any means without the prior written permission of A2LA.

More information

NATA Policy on Calibration and Measurement Traceability

NATA Policy on Calibration and Measurement Traceability NATA Policy on Calibration and Measurement Traceability Paul McMullen Sector Manager, Calibration This Session 1. VIM Definitions Calibration and Metrological Traceability 2. Previous Policy on Metrological

More information

Standardisation why bother?

Standardisation why bother? Standardization in Clinical Enzymology: the IFCC efforts Mauro Panteghini University of Milan Medical School Director, Centre for Metrological Traceability in Laboratory Medicine (CIRME) Milano,, IT 1

More information

American Association for Laboratory Accreditation

American Association for Laboratory Accreditation ISO 15189 Page 1 of 17 Accredited clinical laboratories are required to meet the following additional requirements as contained in P905 - A2LA Metrological Traceability Policy for ISO 15189 Laboratory

More information

Diagnostic Accreditation Program 2015 Laboratory Medicine Accreditation Standards FAQ

Diagnostic Accreditation Program 2015 Laboratory Medicine Accreditation Standards FAQ Diagnostic Accreditation Program 2015 Laboratory Medicine Accreditation Standards FAQ If you have questions about items in the standards please email them to: laboratorymedicine@cpsbc.ca Click on an item

More information

Verifying New Reagent Lot Performance

Verifying New Reagent Lot Performance Verifying New Reagent Lot Performance Julianne Addison Application Consultant, Siemens RSC Answers for life. Objectives Describe issues relating to changing reagent lots Discuss the reasons QC material

More information

Post-market surveillance of the implementation of result traceability in clinical enzymology

Post-market surveillance of the implementation of result traceability in clinical enzymology Post-market surveillance of the implementation of result traceability in clinical enzymology Dr. Rob Jansen St Anna Ziekenhuis Geldrop, NL November 25th 2008 Dr. Rob Jansen 1 Which criteria should the

More information

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 17511 First edition 2003-08-15 In vitro diagnostic medical devices Measurement of quantities in biological samples Metrological traceability of values assigned to calibrators

More information

Performance requirements for ctni measurement in clinical setting: need to standardize as well

Performance requirements for ctni measurement in clinical setting: need to standardize as well 5 th International CIRME Meeting STANDARDIZATION OF CARDIAC TROPONIN I: THE ONGOING IINTERNATIONAL EFFORTS NOVEMBER 30 th, 2011 Centre for Metrological Traceability in Laboratory Medicine (CIRME) Director:

More information

Traceability and the IVD Industry: The Manufacturer s Role

Traceability and the IVD Industry: The Manufacturer s Role JCTLM Members & Stakeholders Meeting: Dec 2017 Traceability and the IVD Industry: The Manufacturer s Role Dave Armbruster, PhD, DABCC, FACB & Graham Beastall 1 1 Global harmonization: The World is Flat

More information

JOINT Reference Methods & Materials Teleconference Minutes

JOINT Reference Methods & Materials Teleconference Minutes Global Biomarkers Standardization Consortium JOINT Reference Methods & Materials Teleconference Minutes Friday April 20, 2012 Co-Chairs: Facilitator: Henrik Zetterberg & Kaj Blennow Maria Carrillo, Alzheimer

More information

Clinica Chimica Acta

Clinica Chimica Acta CCA-13295; No of Pages 7 Clinica Chimica Acta xxx (2013) xxx xxx Contents lists available at ScienceDirect Clinica Chimica Acta journal homepage: www.elsevier.com/locate/clinchim Verification of in vitro

More information

The IVD Directive and Availability of Reference Systems for IVD Medical Devices: A View from Industry

The IVD Directive and Availability of Reference Systems for IVD Medical Devices: A View from Industry How to Cite this article: The IVD Directive and Availability of Reference Systems for IVD Medical Devices: A View from Industry, Greenberg N, ejifcc vol 13 no 3: http://www.ifcc.org/ejifcc/vol13no3/130301006.htm

More information

Rules for accreditation of providers of proficiency testing/interlaboratory

Rules for accreditation of providers of proficiency testing/interlaboratory Rules for accreditation of providers of proficiency testing/interlaboratory tests Scope: With this regulation the requirements for accreditations of providers of proficiency testing/interlaboratory tests/ring

More information

TECHNICAL AND PRODUCTION REQUIREMENTS of ISO Guide 34:2009

TECHNICAL AND PRODUCTION REQUIREMENTS of ISO Guide 34:2009 SANAS Accr. No/s. TECHNICAL AND PRODUCTION REQUIREMENTS of ISO Guide 34:2009 Date/s of evaluation Assessor/s & Observers Facility Area / field of Representative operation This report covers the following:

More information

GLP/SC/01 Basic statistical tools for analytical chemistry (2 days)

GLP/SC/01 Basic statistical tools for analytical chemistry (2 days) GLP Consulting http://consultglp.com Courses on offer (See outlines below) GLP/SC/01 Basic statistical tools for analytical chemistry (2 days) GLP/SC/02 The GUM bottom-up evaluation technique of measurement

More information

NIST Special Publication Standard Reference Materials

NIST Special Publication Standard Reference Materials NIST Special Publication 260-136 Standard Reference Materials Definitions of Terms and Modes Used at NIST for Value-Assignment of Reference Materials for Chemical Measurements W. May, R. Parris, C. Beck,

More information

ILAC-P10:01/2013 -Calibration-

ILAC-P10:01/2013 -Calibration- ACTIVITY 1.4 TRAININGS OF MOLDAC PERSONNEL INVOLVED IN ACCREDITATION PROCESS INCLUDING THE ISSUE REGARDING THE REQUIREMENTS OF THE NEW VERSIONS OF THE EA, IAF, ILAC DOCUMENTS ILAC-P10:01/2013 -Calibration-

More information

International Standards for Biotechnology

International Standards for Biotechnology Bureau International des Poids et Mesures International Standards for Biotechnology Dr R I Wielgosz BIPM BIPM - 2011 1 International Compatibility of Measurement Standards The task of the BIPM is to ensure

More information

STANDARDIZATION IN CLINICAL ENZYMOLOGY

STANDARDIZATION IN CLINICAL ENZYMOLOGY STANDARDIZATION IN CLINICAL ENZYMOLOGY Ilenia Infusino, Mauro Panteghini* Enzyme Reference Laboratory Centre for Metrological Traceability in Laboratory Medicine (CIRME) University of Milan, Milano, Italy

More information

Getting Started with ISO in Your Lab: Introduction to Resources, Training, Implementation, and Timelines for Earning Accreditation

Getting Started with ISO in Your Lab: Introduction to Resources, Training, Implementation, and Timelines for Earning Accreditation Getting Started with ISO 15189 in Your Lab: Introduction to Resources, Training, Implementation, and Timelines for Earning Accreditation Topics What is ISO 15189? What is Accreditation? Preparing for Assessment

More information

Validating, Verifying, and Evaluating Your Test Methods: It s NOT a Regulatory Exercise!

Validating, Verifying, and Evaluating Your Test Methods: It s NOT a Regulatory Exercise! Validating, Verifying, and Evaluating Your Test Methods: It s NOT a Regulatory Exercise! Pat Garrett, Ph.D., DABCC Renee Howell, Ph.D., MT(ASCP) SeraCare Life Sciences, Inc. AACC Annual Meeting July 29,

More information

Available online at ScienceDirect. Procedia Engineering 132 (2015 )

Available online at  ScienceDirect. Procedia Engineering 132 (2015 ) Available online at www.sciencedirect.com ScienceDirect Procedia Engineering 132 (2015 ) 811 815 The Manufacturing Engineering Society International Conference, MESIC 2015 Metrological Regulations for

More information

The role of accreditation bodies and EQA organisers in the. specifications

The role of accreditation bodies and EQA organisers in the. specifications The role of accreditation bodies and EQA organisers in the assessment of analytical quality specifications Ian Mann Swiss Accreditation Service ian.mann@sas.ch Content Important partners and aspects of

More information

Accreditation of Reference Material Producers

Accreditation of Reference Material Producers Wilton Park House, Wilton Place, Dublin 2, Ireland Tel +353 1 607 3003 Fax +353 1 607 3109 E-mail inab@inab.ie Web www.inab.ie Accreditation of Reference Material Producers PS21 1) Purpose This statement

More information

Chinese regulations on IVDs and traceability requirements

Chinese regulations on IVDs and traceability requirements Chinese regulations on IVDs and traceability requirements Hongmei Li CMD/NRCCRM NIM China NOV 30,2015 at BIPM Outline 1. IVD In China 2. China s Efforts on IVD traceability 3. Future Demands and Tendency

More information

PRINCIPLES FOR THE USE OF SAMPLING AND TESTING IN INTERNATIONAL FOOD TRADE CAC/GL Adopted in Revision: 2015.

PRINCIPLES FOR THE USE OF SAMPLING AND TESTING IN INTERNATIONAL FOOD TRADE CAC/GL Adopted in Revision: 2015. PRINCIPLES FOR THE USE OF SAMPLING AND TESTING IN INTERNATIONAL FOOD TRADE CAC/GL 83-2013 Adopted in 2013. Revision: 2015. CAC/GL 83-2013 2 1. INTRODUCTION Sampling and testing are, among others, procedures

More information

CCQM P58.1: Metrology for Clinical Sciences. Diagnostic Formulation Meeting, 24 th Feb James Noble

CCQM P58.1: Metrology for Clinical Sciences. Diagnostic Formulation Meeting, 24 th Feb James Noble CCQM P58.1: Metrology for Clinical Sciences Diagnostic Formulation Meeting, 24 th Feb James Noble Development of a Reference Immunoassay for ctni Cardiac Troponin I (ctni) is an important marker for the

More information

Meeting the ISO/IEC Requirements for Traceability and Measurement Uncertainty. APLAC Approaches. Dr Bernard King NARL, Australia

Meeting the ISO/IEC Requirements for Traceability and Measurement Uncertainty. APLAC Approaches. Dr Bernard King NARL, Australia Meeting the ISO/IEC 17025 Requirements for Traceability and Measurement Uncertainty APLAC Approaches Dr Bernard King NARL, Australia Interpretation of ISO/IEC 17025 Requirements Extensive consultation

More information

VALIDATION TRACEABILITY MEASUREMENT UNCERTAINTY CHALLENGES FOR THE 21 ST CENTURY S ANALYSTS

VALIDATION TRACEABILITY MEASUREMENT UNCERTAINTY CHALLENGES FOR THE 21 ST CENTURY S ANALYSTS VALIDATION TRACEABILITY MEASUREMENT UNCERTAINTY CHALLENGES FOR THE 21 ST CENTURY S ANALYSTS Workshop group 2.3: Convener: Steve Ellison Rapporteur: Workshop group 2.3 participants: Kreimeyer, Maria (D)

More information

Quality Assurance by Liz McChlery

Quality Assurance by Liz McChlery Quality Assurance by Liz McChlery Define QA and Total Quality Management Describe Quality Control Define and distinguish Accuracy and Precision Define a Biological Reference Interval Describe ISO15189

More information

A Risk-based Approach for In Vitro Companion Diagnostics Device FDA Approval Process Associated with Therapies that have Breakthrough Designation

A Risk-based Approach for In Vitro Companion Diagnostics Device FDA Approval Process Associated with Therapies that have Breakthrough Designation A Risk-based Approach for In Vitro Companion Diagnostics Device FDA Approval Process Associated with Therapies that have Breakthrough Designation A Risk-based Approach for In Vitro Companion Diagnostics

More information

A2LA. P113a Policy on Reference Material Traceability for Life Sciences Testing Laboratories

A2LA. P113a Policy on Reference Material Traceability for Life Sciences Testing Laboratories Page 1 of 6 Introduction The Life Sciences Advisory Committee realizes that the traceability of reference materials (RMs) per the requirements of ISO/IEC 17025:2005 (Section 5.6.3) or ISO/IEC 17025:2017(Section

More information

Exposing Myths, Legends and Tales About Reference Materials. Presented by: Matt Sica - ANAB Steve Arpie Absolute Joe Konschnik - RESTEK

Exposing Myths, Legends and Tales About Reference Materials. Presented by: Matt Sica - ANAB Steve Arpie Absolute Joe Konschnik - RESTEK Exposing Myths, Legends and Tales About Reference Materials Presented by: Matt Sica - ANAB Steve Arpie Absolute Joe Konschnik - RESTEK Overview Reference Materials Accreditation Reference Materials Production

More information

A016 Traceability of measurement results compared to national and international measurement standards

A016 Traceability of measurement results compared to national and international measurement standards 08.04.2016 Version 08 Page 1 de 9 A016 Traceability of measurement results compared to national and international measurement standards Modifications: p 2 South Lane Tower I 1, avenue du Swing L-4367 Belvaux

More information

Risk Based Approach To Complaint Handling

Risk Based Approach To Complaint Handling Risk Based Approach To Complaint Handling Khaudeja Bano, M.D. Senior Medical Director, Medical Device Safety Head, AbbVie Inc. Life Sciences Product Complaints Congress Europe Dublin, Ireland December

More information

ENERGY PERFORMANCE PROTOCOL QUALITY ASSURANCE SPECIFICATION

ENERGY PERFORMANCE PROTOCOL QUALITY ASSURANCE SPECIFICATION ENERGY PERFORMANCE PROTOCOL QUALITY ASSURANCE SPECIFICATION Version 1.0 April 2015 Table of Contents 1.0 INVESTOR CONFIDENCE PROJECT 1.1 ENERGY EFFICIENCY PERFORMANCE QUALITY ASSURANCE SPECIFICATION 1.2

More information

Traceability and heterogenous analytes whatmore?

Traceability and heterogenous analytes whatmore? CIRME 1 Traceability and heterogenous analytes whatmore? IRMM - Institute for Reference Materials and Measurements Geel - Belgium http://irmm.jrc.ec.europa.eu/ http://www.jrc.ec.europa.eu/ Overview 2 1.

More information

Vendor Qualification Survey

Vendor Qualification Survey 1200 West 96 th St Minneapolis, MN 55431 Ph: 952-888-7900 Fax: 952-888-2719 Vendor Qualification Survey Vendor Information Company Name: Date: Address: City: Phone Number: email address: Product or Service

More information

GUIDE Uncertainty of measurement Part 1: Introduction to the expression of uncertainty in measurement

GUIDE Uncertainty of measurement Part 1: Introduction to the expression of uncertainty in measurement GUIDE 98-1 Uncertainty of measurement Part 1: Introduction to the expression of uncertainty in measurement First edition 2009 ISO/IEC 2009 This is a preview - click here to buy the full publication PDF

More information

GxP Auditing, Remediation, and Staff Augmentation

GxP Auditing, Remediation, and Staff Augmentation GxP Auditing, Remediation, and Staff Augmentation TABLE OF CONTENTS 3 Introduction 4 GxP Auditing 4 GMP Auditing 5 GCP Auditing 6 GLP Auditing 7 Pharmacovigilance Auditing 7 Vendor/Supplier Auditing 8

More information

GxP Auditing, Remediation, and Staff Augmentation

GxP Auditing, Remediation, and Staff Augmentation GxP Auditing, Remediation, and Staff Augmentation TABLE OF CONTENTS 3 Introduction 4 GxP Auditing 4 GMP Auditing 5 GCP Auditing 5 GLP Auditing 6 Pharmacovigilance Auditing 6 Vendor/Supplier Auditing 7

More information

INTERPRETATION AND GUIDANCE ON THE ESTIMATION OF UNCERTAINTY OF MEASUREMENT IN TESTING

INTERPRETATION AND GUIDANCE ON THE ESTIMATION OF UNCERTAINTY OF MEASUREMENT IN TESTING KAN-G-20 INTERPRETATION AND GUIDANCE ON THE ESTIMATION OF UNCERTAINTY OF MEASUREMENT IN TESTING Issue number : 1 April 2016 Komite Akreditasi Nasional National Accreditation Body of Indonesia Gedung I

More information

DRAFT MEDICAL DEVICE GUIDANCE DOCUMENT

DRAFT MEDICAL DEVICE GUIDANCE DOCUMENT November 2015 DRAFT DRAFT MEDICAL DEVICE GUIDANCE DOCUMENT REQUIREMENTS FOR LABELLING OF MEDICAL DEVICES Medical Device Authority MINISTRY OF HEALTH MALAYSIA Contents Page Preface... iii 1 Introduction.

More information

Laboratory Quality Assurance Manager & Laboratory Assessor RULES & HANDBOOK

Laboratory Quality Assurance Manager & Laboratory Assessor RULES & HANDBOOK EOQ Personnel Registration Scheme Laboratory Quality Assurance Manager & RULES & HANDBOOK Prepared by: Dr. Eugenia Soboleva, Quality Austria In accordance with the working group on EOQ product development

More information

Development of an immunoassay-based measurement procedure of higher metrological order for ctni. James Noble Date: 30/11/11

Development of an immunoassay-based measurement procedure of higher metrological order for ctni. James Noble Date: 30/11/11 Development of an immunoassay-based measurement procedure of higher metrological order for ctni James Noble Date: 3/11/11 Analytical requirements of a RMP 1. Comparable ctni specificity to commercial assays

More information

What is an LDT? The Do s s And Don ts of Validating Laboratory Developed Tests

What is an LDT? The Do s s And Don ts of Validating Laboratory Developed Tests The Do s s And Don ts of Validating Laboratory Developed Tests Deirdre Astin, MS, MT (ASCP) Lab Quality Confab 2011 November 2011 San Antonio Texas What is an LDT? Laboratory developed tests (LDTs) are

More information

PAB PRINCIPLE. Kit Reorder # ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

PAB PRINCIPLE. Kit Reorder # ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE SYNCHRON CX System(s) Chemistry Information Sheet Copyright 2008 Beckman Coulter, Inc. Prealbumin Kit Reorder # 475106 For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Date Reviewed by: Date PRINCIPLE

More information

TL 9000 R6.0 REQUIREMENTS

TL 9000 R6.0 REQUIREMENTS Logo or heading here TL 9000 R6.0 REQUIREMENTS Nancy Patterson Nokia Tom Yohe - Consultant TL 9000 CONTAINS ALL OF THE AUDITABLE CLAUSES OF ISO 9001:2008 http://tl9000.org/alerts/alerts.html#10 QUEST FORUM

More information

Perspective: Convergence of CLIA and FDA Requirements A Rational Shift in the Regulatory Paradigm

Perspective: Convergence of CLIA and FDA Requirements A Rational Shift in the Regulatory Paradigm Perspective: Convergence of CLIA and FDA Requirements A Rational Shift in the Regulatory Paradigm Planning for Efficiencies of Data, Resources, and Timelines A PRECISION BRIEF Introduction As precision

More information

MEDICAL DEVICE CLINICAL INVESTIGATIONS AND ISO 14155

MEDICAL DEVICE CLINICAL INVESTIGATIONS AND ISO 14155 MEDICAL DEVICE CLINICAL INVESTIGATIONS AND ISO 14155 EXECUTIVE SUMMARY Medical device regulations around the world generally require manufacturers of most types of medical devices to supply data as part

More information

PREAMBLE. accreditation of External Quality Assessment schemes in the UK.

PREAMBLE. accreditation of External Quality Assessment schemes in the UK. IFCC/EMD/CAQ in medical laboratories PREAMBLE This document is directed to External Quality Assurance Programmes (EQAP) organizers who wish, on a voluntary basis, to demonstrate their competence, for the

More information

Session 10: The UNFPA Prequalification Site Inspection

Session 10: The UNFPA Prequalification Site Inspection Session 10: The UNFPA Prequalification Site Inspection Group Brainstorm: Site Inspection In your experience, in what areas have you disagreed with the inspectors? Objectives Provide overview of what inspectors

More information

CRITERIA FOR PERFORMING CALIBRATION AND INTERMEDIATE CHECKS ON EQUIPMENT USED IN ACCREDITED FACILITIES

CRITERIA FOR PERFORMING CALIBRATION AND INTERMEDIATE CHECKS ON EQUIPMENT USED IN ACCREDITED FACILITIES CRITERIA FOR PERFORMING CALIBRATION AND INTERMEDIATE CHECKS ON EQUIPMENT USED IN ACCREDITED FACILITIES Prepared by: Technical Manager Approved by: Chief Executive Officer Approval Date: 2017-08-10 Effective

More information

The Role of Proficiency Testing in Laboratory Capacity Building in Chile

The Role of Proficiency Testing in Laboratory Capacity Building in Chile 2014/SCSC/WKSP2/017 Session: 7.5 The Role of Proficiency Testing in Laboratory Capacity Building in Chile Submitted by: Chile Food Safety Cooperation Forum Partnership Training Institute Network Proficiency

More information

Overview of NIST s JCTLM Activities: Past, Present, and Future. Karen W. Phinney Biomolecular Measurement Division

Overview of NIST s JCTLM Activities: Past, Present, and Future. Karen W. Phinney Biomolecular Measurement Division Overview of NIST s JCTLM Activities: Past, Present, and Future Karen W. Phinney Biomolecular Measurement Division Structure of NIST s Laboratory Programs Undersecretary of Commerce for Standards and Technology

More information

Role of the WHO IVD Prequalification Programme in Light of National Regulatory Authority Approval

Role of the WHO IVD Prequalification Programme in Light of National Regulatory Authority Approval Role of the WHO IVD Prequalification Programme in Light of National Regulatory Authority Approval UN Prequalification of Medicines, Diagnostics and Vaccines 6 th Consultative Stakeholder Meeting 4 April

More information

WELMEC European cooperation in legal metrology

WELMEC European cooperation in legal metrology WELMEC 8.6 Issue 1 WELMEC European cooperation in legal metrology Measuring Instruments Directive 2004/22/EC Presumption of Conformity of the Quality System of Manufacturers with Module D or H 1 when EN

More information

SAMPLE. Liquid Chromatography-Mass Spectrometry Methods; Approved Guideline

SAMPLE. Liquid Chromatography-Mass Spectrometry Methods; Approved Guideline October 2014 Liquid Chromatography-Mass Spectrometry Methods; Approved Guideline This document provides guidance to the clinical laboratorian for the reduction of interlaboratory variance and the evaluation

More information

DELTA tool: Updates. A tool to evaluate and benchmark AQ model performances. P. Thunis, K. Cuvelier* JRC-IES, (*) Ext JRC-IES.

DELTA tool: Updates. A tool to evaluate and benchmark AQ model performances. P. Thunis, K. Cuvelier* JRC-IES, (*) Ext JRC-IES. DELTA tool: Updates A tool to evaluate and benchmark AQ model performances P. Thunis, K. Cuvelier* JRC-IES, (*) Ext JRC-IES www.jrc.ec.europa.eu Background & Outline FAIRMODE is a Forum for Air Quality

More information

How to Choose the Right Equipment/Platforms for your Laboratory

How to Choose the Right Equipment/Platforms for your Laboratory How to Choose the Right Equipment/Platforms for your Laboratory (A Public Hospital Laboratory Perspective) Michelle J. Francis Overview Public Hospital Pressures / Challenges Commercial vs In-house assays

More information

European Pharmacopoeia Reference Standards A Lodi, Head of the Laboratory Department, EDQM, Council of Europe

European Pharmacopoeia Reference Standards A Lodi, Head of the Laboratory Department, EDQM, Council of Europe European Pharmacopoeia Reference Standards A Lodi, Head of the Laboratory Department, EDQM, Council of Europe Content Ph. Eur. Reference Standards General notices, terms and definitions Key attributes

More information

Recast Medical Device directives Impacts on materiovigilance

Recast Medical Device directives Impacts on materiovigilance Recast Medical Device directives Impacts on materiovigilance Journée Vigilance 23.03.2017 Valérie Nys Revision of the EU Medical Devices Legislation Directive 90/385/EEC on active implantable medical devices

More information

General requirements for the competence of testing and calibration laboratories. In this presentation:

General requirements for the competence of testing and calibration laboratories. In this presentation: General requirements for the competence of testing and calibration laboratories ISO/IEC 17025:2017 In this presentation: Explanation New requirement Interpretation / Examples / Questions Agenda Welcome

More information

V&V Best Practices. CASSS, CMC Strategy Forum Steven W. Badelt, PhD Managing Partner Suttons Creek, Inc. SUTTONSCREEK.COM

V&V Best Practices. CASSS, CMC Strategy Forum Steven W. Badelt, PhD Managing Partner Suttons Creek, Inc. SUTTONSCREEK.COM V&V Best Practices CASSS, CMC Strategy Forum Steven W. Badelt, PhD Managing Partner Suttons Creek, Inc. Solving the problem of Complexity. 21CFR820.30 & FDA Guidance INCOSE System Engineering Handbook

More information

consent. With the exception, such as impossible to obtain informed consent objectively or there is no risk of the clinical trial for subjects, the res

consent. With the exception, such as impossible to obtain informed consent objectively or there is no risk of the clinical trial for subjects, the res The Technical Guidelines for In Vitro Diagnostic Reagents Clinical Trial Article 1 In vitro diagnostic reagents clinical trial (include comparison experiment with marketed products) refers to the systematically

More information

ILAC-P10:2002. ILAC Policy on Traceability of Measurement Results

ILAC-P10:2002. ILAC Policy on Traceability of Measurement Results ILAC-P10:2002 ILAC Policy on Traceability of Measurement Results Copyright ILAC 2002 ILAC encourages the authorised reproduction of its publications, or parts thereof, by organisations wishing to use such

More information

Report on the BIPM-WADA Workshop: Standards and Metrology in support of Anti-Doping Analysis

Report on the BIPM-WADA Workshop: Standards and Metrology in support of Anti-Doping Analysis Rapport BIPM-2016/01 Report on the BIPM-WADA Workshop: Standards and Metrology in support of Anti-Doping Analysis 2016 Pavillon de Breteuil, F-92312 SÈVRES Cedex, France Introduction BIPM-WADA Workshop:

More information

MEDICINES CONTROL COUNCIL

MEDICINES CONTROL COUNCIL PRE-REGISTRATION CONSULTATION MEETING MATERIAL CHECKLIST MEDICINES CONTROL COUNCIL PRE-REGISTRATION CONSULTATION MEETING MATERIAL The CHECKLIST is to be used for the preparation of a formal request to

More information

Establishing Quality Control Target Values and Standard Deviations for Hematology Instrumentation

Establishing Quality Control Target Values and Standard Deviations for Hematology Instrumentation Establishing Quality Control Target Values and Standard Deviations for Hematology Instrumentation For informational purposes only. WHO REQUIRES IT AND WHAT THEY REQUIRE Regulatory agencies require the

More information

The new ISO/IEC 17025:2017

The new ISO/IEC 17025:2017 The new ISO/IEC 17025:2017 By Dr. George Anastasopoulos Director, Conformity assessment, IAS Email: ganastasopoulos@iasonline.org Introduction - Background information ISO/IEC 17025 was first issued in

More information

Transfer of Methods Supporting Biologics and Vaccines

Transfer of Methods Supporting Biologics and Vaccines Transfer of Methods Supporting Biologics and Vaccines Timothy Schofield Arlenda Inc. tim.schofield@arlenda.com Presented at the BPD event on Case Studies in Tech Transfer esearch Triangle Park, NC, December

More information

18.H Questionnaire for preparing GMP-inspections

18.H Questionnaire for preparing GMP-inspections Questionnaire for preparing -inspections Questionnaire for preparing -inspections Here you will find answers to the following questions: What questions are typically asked during inspections based on current

More information

INDUSTRY S VIEW ON THE FUTURE OF IN VITRO DIAGNOSTIC (IVD) LEGISLATION IN EUROPE. May 2013 SPECIAL REPRINT. By Jesús Rueda Rodríguez, EDMA

INDUSTRY S VIEW ON THE FUTURE OF IN VITRO DIAGNOSTIC (IVD) LEGISLATION IN EUROPE. May 2013 SPECIAL REPRINT. By Jesús Rueda Rodríguez, EDMA May 2013 SPECIAL REPRINT INDUSTRY S VIEW ON THE FUTURE OF IN VITRO DIAGNOSTIC (IVD) LEGISLATION IN EUROPE By Jesús Rueda Rodríguez, EDMA Reproduced with the kind permission of Global Regulatory Press from

More information

Terms and conditions for laboratories appointed to undertake the testing of compost under the Compost Certification Scheme aligned to PAS 100 and the

Terms and conditions for laboratories appointed to undertake the testing of compost under the Compost Certification Scheme aligned to PAS 100 and the Terms and conditions for laboratories appointed to undertake the testing of compost under the Compost Certification Scheme aligned to PAS 100 and the Compost Quality Protocol Renewable Energy Assurance

More information

TRACEABILITY IN EXTERNAL QUALITY ASSESSMENT

TRACEABILITY IN EXTERNAL QUALITY ASSESSMENT TRACEABILITY IN EXTERNAL QUALITY ASSESSMENT HOW WE ENSURE TRACEABILITY IN EQA AND STRESS ITS IMPORTANCE TO USERS PROGBA (BUENOS AIRES EXTERNAL QUALITY ASSESSMENT SCHEME) CEMIC UNIVERSITY HOSPITAL BUENOS

More information

Using Guard Bands to Accommodate Uncertainty in the Spark AES Analysis of Aluminum or Aluminum Alloys When Determining Compliance with Specified

Using Guard Bands to Accommodate Uncertainty in the Spark AES Analysis of Aluminum or Aluminum Alloys When Determining Compliance with Specified Using Guard Bands to Accommodate Uncertainty in the Spark AES Analysis of Aluminum or Aluminum Alloys When Determining Compliance with Specified Composition s Acknowledgement The Aluminum Association gratefully

More information

Raad voor Accreditatie (Dutch Accreditation Council RvA) Accreditation of Reference Material Producers

Raad voor Accreditatie (Dutch Accreditation Council RvA) Accreditation of Reference Material Producers Raad voor Accreditatie (Dutch Accreditation Council RvA) Accreditation of Reference Material Producers Document code: RvA-T034-UK Version 4, 01 mei 2013 A RvA-Explanatory note describes the policy and/or

More information

in touch Greetings from the Chairman: Greetings from the Chairman New ISO and new 9004 to come Page 1

in touch Greetings from the Chairman: Greetings from the Chairman New ISO and new 9004 to come Page 1 Greetings from the Chairman: Dear IPC InTouch Followers, Greetings This year Aegina Island, Greece hosted to the 20th General Assembly Meeting of IPC. We have again successfully completed an intensive

More information

Preparation Approval Authorisation Application date. Directive Council The President

Preparation Approval Authorisation Application date. Directive Council The President Title Proficiency Testing Reference RT-24 Revision 02 Date 2015-10-27 NOTE: The present document represent the English version of document under reference at the specified revision. In case of conflict

More information

FINAL DOCUMENT. International Medical Device Regulators Forum. Medical Device Regulatory Audit Reports

FINAL DOCUMENT. International Medical Device Regulators Forum. Medical Device Regulatory Audit Reports FINAL DOCUMENT International Medical Device Regulators Forum Title: Authoring Group: Medical Device Regulatory Audit Reports IMDRF MDSAP Working Group Date: 2 October 2015 Toshiyoshi Tominaga, IMDRF Chair

More information

Good Automated Manufacturing Practices (GAMP)

Good Automated Manufacturing Practices (GAMP) Good Automated Manufacturing Practices (GAMP) Klaus Krause, Amgen ISPE/GAMP Americas Steering Committee ISPE San Francisco/Bay Area Chapter Meeting, October 7, 2004 Presentation Overview I. GAMP - Organization

More information

Measurement Canada s role in Timber Measurement

Measurement Canada s role in Timber Measurement Measurement Canada s role in Timber Measurement TMS Meeting, Ferndale, WA - April 2013 Fair Measure For All Measurement Canada is responsible for ensuring the integrity and accuracy of measurement in the

More information

QUALITY CONTROL/QUALITY ASSURANCE IN THE MOLECULAR MICROBIOLOGY LABORATORY

QUALITY CONTROL/QUALITY ASSURANCE IN THE MOLECULAR MICROBIOLOGY LABORATORY QUALITY CONTROL/QUALITY ASSURANCE IN THE MOLECULAR MICROBIOLOGY LABORATORY Richard L. Hodinka, Ph.D. University of South Carolina School of Medicine Greenville Greenville Health System, Greenville, SC

More information

Date. Alkaline phosphatase measurements are used in the diagnosis and treatment of liver, bone, parathyroid, and intestinal diseases.

Date. Alkaline phosphatase measurements are used in the diagnosis and treatment of liver, bone, parathyroid, and intestinal diseases. SYNCHRON System(s) Chemistry Information Sheet Copyright 2013 Beckman Coulter, Inc. Alkaline Phosphatase REF 442670 (200 tests/cartridge) REF 476821 (400 tests/cartridge) For In Vitro Diagnostic Use ANNUAL

More information

Data Analytics in Healthcare

Data Analytics in Healthcare Data Analytics in Healthcare Insights from leveraging Data Analytics for Population Health Management Purush Yeluripati Getting Your Transformation Right SYMPOSIUM March 22-23, 2018 5th Annual Big Data

More information

Measuring Instruments Directive 2014/32/EU Assessment of Notified Bodies in Charge of Type Examination Presumption of Conformity based on EN 17065

Measuring Instruments Directive 2014/32/EU Assessment of Notified Bodies in Charge of Type Examination Presumption of Conformity based on EN 17065 WELMEC 8.5, 2017 Measuring Instruments Directive 2014/32/EU Assessment of Notified Bodies in Charge of Type Examination Presumption of Conformity based on EN 17065 For information: This guide is available

More information

By Ginger Bailly 18 Feb For Official Use Only

By Ginger Bailly 18 Feb For Official Use Only By Ginger Bailly 18 Feb 2014 For Official Use Only 1 Discuss: Why Quality Assurance is Important How to Implement a Quality Management System Explore the Following Key Elements: Traceability Document Control

More information

Clinical Trials Management for Molecular Diagnostics. April 2016

Clinical Trials Management for Molecular Diagnostics. April 2016 Clinical Trials Management for Molecular Diagnostics April 2016 Clinical Operations Responsibilities Accrue samples that have proper informed consent for use Retrospective cohorts Remnant samples Prospective

More information

The EFLM Committee on Education and Training. Opinion Paper

The EFLM Committee on Education and Training. Opinion Paper Clin Chem Lab Med 2015; aop Opinion Paper Elizabeta Topic*, Nora Nikolac, Mauro Panteghini, Elvar Theodorsson, Gian Luca Salvagno, Marijana Miler, Ana-Maria Simundic, Ilenia Infusino, Gunnar Nordin and

More information

Proposed New USP General Chapter: The Analytical Procedure Lifecycle 1220

Proposed New USP General Chapter: The Analytical Procedure Lifecycle 1220 Page 1 of 9 STIMULI TO THE REVISION PROCESS Stimuli articles do not necessarily reflect the policies of the USPC or the USP Council of Experts Proposed New USP General Chapter: The Analytical Procedure

More information