1 Amino Acid Analysis Liquid Chromatography Column I. Introduction This manual provides the user with information necessary to obtain the maximum efficiency and life from the Waters Amino Acid Analysis Column. This column has been specifically designed for the separation and analysis of free amino acids. The column is a precision device. Waters Corporation uses very exacting manufacturing procedures to achieve a small average particle size, narrow range of particle size distribution and stable particle structure in the column. These procedures create the high efficiency column-to-column reproducibility and long separation life built into each amino acid column from Waters. However, care must be taken to maintain packing density and protect it from contamination by foreign material. The following information will aid you in obtaining the highest performance and longest life from your column. II. Background Information Although there are some hydrophobic interactions between the column and certain amino acids, the primary mode of separation is that of ion-exchange. The cation-exchange resin utilized has a large number of negatively charged (sulfonate) functional groups which are covalently bound to it. These negatively charged groups, in turn, readily associate with positively charged cations when they are present in solution. Before the amino acid sample is injected onto the column, the resin is equilibrated at a low ph with buffer containing a high concentration of sodium (Na+) ions. This places the column in the sodium form. As amino acids enter the column in this acidic buffer, they are strongly positively charged and readily exchange with the loosely bound sodium ions. As the buffer gradually changes to a higher ph during the gradient run, each amino acid becomes less positively charged reflecting its pk value and differentially elutes accordingly. Contents I. Introduction II. Background Information a. Elution System 1 b. Elution System 2 c. General Preparation Guide for One Liter of Eluent d. Use of Non-aqueous Solvents e. Backpressure f. Sample Considerations g. SEP-PAK C 18 Cartridge Clean Up for Amino Acid Samples h. Preparation Scheme III. Column Care & Use Considerations a. Column Care Considerations b. Column Storage c. Column Regeneration d. Troubleshooting e. Regeneration of Protein and Lipid Contaminated Column f. Regeneration of Heavy Metal Contaminated Column g. Available Service and Applications Information h. Chromatographic Troubleshooting i. Typical Separation IV. Column Performance Checks V. Column End Fittings a. End Fitting Removal Procedure b. Column Hookup c. Storage Appendix A. Amino Acid Analysis Column Specifications

2 Figure 1: Ion Exchange Mechanism of the halide ion requires immediate flushing of the system with distilled water containing 0.2% sodium azide or 0.1% trifluoroacetic acid before system shutdown. Allowing the halide buffers to sit static in the pumps overnight will lead to corrosion. a. Elution System 1: Waters Non-halide Buffer System Buffer Composition Buffer A-ES2 Sodium Citrate Dihydrate: 19.6 g Phenol (Preservative): 1.0 g HCI: to ph 3.00 Total Volume: 1 L Buffer B-ES1 Boric Acid: 1.5 g Sodium Nitrate: 21.0 g 6M NaOH: to ph 9.60 Total Volume: 1 L NOTE: Do not add phenol to Buffer B; it will decompose at this ph. CAUTION: Introduction of improperly prepared or contaminated buffers or samples onto the resin bed could irreversibly poison the column. Heavy metal ions can also poison the column. Column degradation will be indicated by band broadening loss in resolution and/or change in selectivity. All reagents to be used for high performance liquid chromatography (HPLC) analysis of amino acids must be of the highest purity. Water should be distilled and deionized with a purity equivalent to that obtained from a Milli-Q water purifica tion system. Impure buffers will lead to artifact peaks during analysis and shorten column lifetime. There are two elution systems available for analysis: 1) Elution System 1 is the Waters Non-halide Buffer System. It is recommended for use where possible since it employs no potentially corrosive halide ions. b. ELUTION SYSTEM 2: Halide Buffer System Buffer Composition Buffer Composition Buffer A-ES2 Sodium Citrate Dihydrate: 19.6 g Phenol (Preservative): 1.0 g HCI: to ph 3.00 Total Volume: 1 L Buffer B-ES2 Boric Acid: 1.5 g Sodium Chloride: g 6M NaOH: to ph 9.80 Total Volume: 1 L 2) Elution System 2 is a halide-containing buffer system and is recommended when maximum sensitivity is needed. Use of chloride enhances orthopthalaldehyde (OPA) fluorescence by approximately twofold as compared to the non halide elution system. The presence 2

3 c. General Preparation Guide for One Liter of Eluent Dissolve salts in approximately 900 ml of purified water. Adjust ph with appropriate acid or base. Adjust to final one liter volume with water. Filter through the Solvent Clarification Kit using an 0.45 µm aqueouscompatible filter. Transfer buffer to an amber glass bottle and purge for 15 minutes with pure nitrogen. Transfer buffer to analyzer. Buffers prepared in this manner will be stable for at least one week at room temperature. d. Use of Non-aqueous Solvents For additional resolution of the early-eluting amino acids, small amounts of methanol, ethanol, or propanol can be added to Buffer A. This alcohol concentration should not exceed 10% High concentrations of a non-aqueous modifier will cause the resin to swell resulting in high backpressure. e. Backpressure With a new column flowing at 0.4 ml/min in Buffer A at 62 C. typical system pressures will range from 800 psi psi. As the column ages, pressure will slowly increase, although this increase is not usually associated with decreased efficiency. The most common causes of high backpressure are: particulates in the sample or buffer, proteinaceous material coating the resin packing, mold formation in the buffers or in the column. Reduced temperature, higher flow rate and high solvent viscosity will all yield higher pressure. To prevent excessive pressure, do not flow through the column with the temperature control module off. f. Sample Considerations Acidic protein hydrolyzates should be lyophilized and reconstituted in Buffer A. Convenient sample concentration range: nmol amino acid/5 µl - 50 µl injection volume. Sample volume and sensitivity requirements should be optimized for individual needs. Samples that contain protein, lipid or heavy metals (e.g., tissue extracts, food samples, feed samples, etc.) should not be introduced directly on the column. Pre-purification on a strong exchange resin or a Sep-Pak C 18 cartridge is recommended. g. Sep-Pak C 18 Cartridge Cleanup for Amino Acid Samples Prepare the following solutions: Solution 1: 0.1 % trifluoroacetic acid (TFA) in water Solution 2: 0.1 % TFA in H20:MeOH, 80:20 Solution 3: 0.1 % TFA in H2O:MeOH, 70:30 h. Preparation Scheme 1. Activate a new Sep-Pak C 18 cartridge with two 10 ml volumes of methanol. 2. Wash with two 10 ml volumes of Solution Wash with 10 ml of Solution Mix 1 ml sample with 2 ml Solution 3. Sample should be water or acid solubilized and free of particulate matter. 5. Pass the sample through the Sep-Pak C 18 cartridge. 6. Discard the first 1 ml of eluent and collect the next 2 ml. This fraction will contain all amino acids. Lipids and high molecular weight pro teins will be retained on the Sep-Pak C 18 cartridge. 3

4 III. Column Care & Use Considerations a. Column Care Considerations ALWAYS filter samples to prevent excessive pressure buildup due to particulate matter. Use a Sample Clarification Kit (Part No ) and, if necessary, centrifugation. For more details on the Waters Sample Clarification Kit call or write for Bulletin D97. ALWAYS filter buffers to remove microparticulate matter. This eliminates high pressures from plugged frits and also helps ensure long column life. Use a Solvent Clarification Kit (Part No ) to filter solvents. Vacuum filtration or sonification may be used to remove dissolved gasses which could affect your solvent delivery system. For more details on the Waters Solvent Clarification Kit call or write for Bulletin D97. ALWAYS use high purity solvents such as Waters solvents. For analytical work choose high quality Waters solvents. For more details on Waters solvents call or write for Bulletin M10. ALWAYS replace plugs in column fittings to prevent the column from drying out when not in use. b. Column Storage The column may be stored 1-2 days in Buffer A without flow. The presence of phenol in the buffer will prevent bacterial growth. For longer storage periods, 0.2 N NaOH is strongly recommended. c. Column Regeneration For protein hydrolyzate samples, regenerate the column at least once a week as follows: d. Troubleshooting This section discusses some possible problems and their solutions (refer to Table 1.) e. Regeneration of Protein and Lipid Contaminated Column Large proteins and some lipids can be removed following alkaline hydrolysis. Reverse direction of buffer flow through column. Elevate column heater temperature to C and pump 0.2 N NaOH solution through column for at least 2 hours. Particularly difficult contamination may require overnight treatment. Restore normal flow direction and temperature; reequilibrate column with Buffer A and test column with amino acid standard solution. If problem still exists, repeat the above treatment. Table 1: Typical Column Problems and Solutions Problem Cause Solution Excess pressure buildup Loss of resolution, broad peaks, low plate counts Column frits plugged with particulates or mold is growing in column or in buffers Sample and/or buffer precipitates on column Contaminated column Reverse flow on column. If this does not work, follow procedure for protein removal. If pressure is still high, carefully remove column inlet fitting being sure not to disturb resin. Clean in ultrasonic bath. Always filter solvents and samples. Follow procedure above. Follow procedures for metal and protein removal. Eluent: 0.2 N NaOH Flow: 0.5 ml/min Temperature: 62 C Flush Time: 30 min Crude samples may require more frequent regeneration. 4

5 f. Regeneration of Heavy Metal Contaminated Column Often metals can be removed by chelation under basic and acidic conditions. Reverse direction of buffer flow through column. Elevate column heater temperature to C and pump 0.2 N NaOH solution containing 0.25 g EDTA/liter through the column for minimum of 4 hours. Switch buffer tube to second container and pump 2 N HNO, solution containing 0.25 g EDTA/liter through the column for another 2-4 hours. The more advanced Waters HPLC System allows for the generation of a continuous gradient using only two buffers. This is equivalent to an infinite step gradient. Superior resolution is achieved without baseline disturbances. The constant ionic strength, ph gradient (0.2N Na +, ph 3.00 to ph 9.60) allows for rapid reequilibration without NaOH regeneration. i. Typical Separation Figure 2: Amino Acid Analysis Continuous Gradient Restore normal flow direction and temperature. Reequilibrate column with Buffer A and test column with amino acid standard solution; If problem still exists, repeat the above basic/acidic EDTA treatment. CAUTION: Never attempt to clean the column using hydrochloric acid (HCI) or sulphuric acid (H2SO,) solutions. Such acids are harmful to stainless steel. g. Available Service and Applications Information The Waters Corporation staff of trained and e perienced service specialists provides maintenance for Waters instruments on preventive and/or corrective levels. Contact Waters Corporation at or your local Waters representative for further details. For answers to specific chromatography questions in areas such as methods development, applications and quality control, contact your local Waters representative or Waters home office applications laboratory. As well as the standard hydrolyzate amino acids, unusual amino acids can be separated by the two-buffer, gradient system. Note cysteic acid, hydroxyproline, methionine sulfoxide, hydroxylysine, and tryptophan. Slight modifications of the ph and/or constituency of the A or B buffers can quickly be used to optimize new separations. h. Chromatographic Troubleshooting When a new column is installed, it is recommended that a separation of standard hydrolyzate amino acids be performed using Elution System 1. A typical separation is shown on the following page (Figures 3 & 4). 5

6 Figure 3: Amino Acid Analysis Special Amino Acids IV. Column Performance Checks 1. Cysteic Acid 2. Methionine Sulfoxide 3. Hydroxy Proline 4. Aspartic Acid 5. Threonine 6. Serine 7. Glutamic Acid 8. Proline 9. Glycine 10. Alanine 11. Cystine 12. Valine 13. Methionine 14. Isoleucine 15. Leucine 16. Tyrosine 17. Phenylalanine 18. Histidine 19. Hydroxy Lysine 20. Tryptophan 21. Lysine 22. Ammonia 23. Arginine Column: CAT EX Resin Column (0.4 cm x 25 cm) Sample: 1 NMOL AA STDS Injection Volume: 10 µl Flow Rate: 0.4 ml/min Mobile Phase: A. 0.2 N Na citrate, ph 3.0 B. 0.2 N Na borate, ph % B, curve 6, 48 min Detector: Model 420 ex: 338 nm em: 425 nm gain: 4 Table 2 provides performance check information. Use the table in conjunction with the following procedure to determine the condition of your column. If you feel it has become contaminated, refer to Troubleshooting section for appropriate column cleaning procedures. Table 2: Performance Check for Columns Column: Buffer: Flow rate: Waters Amino Acid Analysis Column (Part No ) A-ES1 Temperature: 62 C Chart Speed: Detection: 0.4 ml/min 0.25 cm/min Fluorescence (o-phthalaldehyde) Sample: Sample Amount: Results: Amino acid mixture contain ing equalmolar concentra tion aspartic acid, threonine, serine, and glutamic acid 1-2 n mole of each amino acid in 5-20 µl volume Chromatogram should show a minimum resolution of 85% for all amino acids as determined by the following equation: Figure 4: Resolution Variations in results as compared to the expected chromatogram can usually be explained by improper buffer preparation (weighing/volume error) or improper ph. The following guidelines describe the effect of ph: 1. If Buffer A is 0.20 ph units too high, there will be a loss of THR-SER resolution and OH-PRO will coelute with ASP. 2. If Buffer A is 0.30 ph units too low, there will be broadening of the neutral amino acid peaks and ASP will elute later than expected. 3. If Buffer B is 0.20 ph units too high, ARG will coelute with NH. If Buffer B is 0.20 units too low, ARG will elute minutes later than expected. Increased column temperature will lead to a decrease in the elution volume for all amino acids and a decrease in the THR-SER resolution. 6

7 All peaks should be sharp and symmetrical. Backpressure should be less than 2,000 psi. If you suspect that the quality of your column has decreased, the above test is a rapid method of determining the efficiency of your separation. Peaks which are not symmetrical could in dicate partially plugged frits are at fault or foreign material has accumulated on the resin packing. Generally this phenomenon is also ac companied by increased backpressure (refer to Table 2 for corrective action). Alternatively, resin packing material may have shifted, voided or become uneven. Careful removal of top column packing nut and visual inspection of the resin bed is the best diagnostic procedure to determine whether this is the cause. Since removal of the nut could, itself, disturb the packing, this should be a last resort measure. If resin bed is not smooth and completely full, please contact your nearest Waters representative for instructions. If backpressure is normal but resolution is poor or peaks are very broad to nonexistent, the problem could be heavy metal contamination of the resin (refer to Table 1 for corrective action). It should always be kept in mind that column efficiency is affected by the liquid chromatography (LC) system as well as buffers and reagents. Care should be taken to make sure the LC system is working normally, that tubing and post-column reaction coil(s) are properly connected and leak free, and that buffers and reagents are fresh, filtered and tree from contamination. Whenever problems persist after making the above tests and checks on your system, con tact your local Waters representative for further assistance. V. Column End Fittings The column is symmetrical, having the same kind of fittings and frits at both ends. Flow of buffer can occur in either direction without harmful effects. However, the end bearing and column label should be considered the inlet since this was the direction of flow during packing and testing procedures. Use of the same flow direction for analysis is encouraged, and it will prove helpful during any necessary cleaning procedures. Both end fittings are made up of a smaller and a larger hex nut. The larger hex nuts contain a non-removable stainless steel frit to both retain the resin packing and to catch particulates before they reach the resin. In normal operation particulates from the sample or buffer should be filtered out using a 0.45 µm membrane filter (refer to Section III d.). If the backpressure of the column should rise above 2500 psi, reversing the direction of flow on the column for a few minutes will usually flush particulates away. If flushing and cleaning procedures (see Troubleshooting, Section III d.) do not work, the frits may be cleaned ultrasonically following their removal from the column. a. End Fitting Removal Procedure CAUTION: Removal of end fittings may disturb the resin packing material and cause skewed or broad peaks. Therefore, this procedure should be used only as a last resort measure. 1. Work only in a clean area where dust, dirt and debris will not contaminate the exposed packing material. 2. Secure column in a vice. Hold in an upright position by clamping lightly (do not over tighten) onto the smaller hex fitting. 3. Remove only one fitting at a time to avoid loss of packing material. Always begin by removing column inlet (end with column label) first. 4. Use 5/8 inch (16 mm) wrench, turning counterclockwise slowly until fitting is removed. 5. With a clean spatula or similar tool, remove any resin packing which has adhered to the frit and place smoothly and carefully back on to resin bed at end of column tube. 6. Fritted fitting may now be cleaned in ultrasonic cleaner using appropriate solvent. 6 N nitric acid (HNO,) often works well but be very cautious to avoid contact with skin. 7. Column fittings should be reassembled as soon as possible after inspection and cleaning to avoid drying and contamination of resin. 7

8 b. Column Hookup 1. Direction of flow is indicated on the column. 2. Connect one adaptor to the inlet and one adaptor to the outlet of the column. 3. Connect the tubing from the injector to the column inlet union. Be sure to use inch (0.23 mm) i.d. tubing for this purpose. 4. Connect the outlet adaptor to Mixing Tee #1 of the post column reaction system. c. Storage If you do not intend to use this column for a period of a week or longer, it is best to store it in 0.2 N sodium hydroxide (NaOH). Using your HPLC, pump the filtered NaOH solution through the column at 0.4 ml/min for at least 30 min. To reduce solution viscosity, column heater should be on at normal analysis temperature (60-70 C). At the end of this period turn off pump and disconnect inlet and outlet tubing. Replace the plug fittings on the column, screwing them in tightly to prevent evaporation and drying out of the resin. Appendix A: Amino Acid Analysis Column Specifications Column Size Length: 25 cm Diameter: 0.46 cm i.d. Packing Material Styrene - divinyl benzene copolymer with sulfonic acid functionality in Na+ form Particle Size 9 ± 0.5 Nm in diameter Separation Method Strong Cation Exchange ph Range 1-14 Pressure Limit 3000 psi Shipping Solvent 0.2 M NaOH (Na + form) The HPLC should now be flushed out with filtered water before it is stored or used for any other pur pose. A stainless steel jumper tube which takes the place of the column will prove useful for this purpose. 8

9 Sales Offices: Austria and European Export (Central South Eastern Europe, CIS and Middle East) Australia Belgium Brazil Canada x2205 China CIS/Russia Czech Republic Denmark Finland France Germany Hong Kong Hungary India and India Subcontinent Ireland Italy Japan Korea Mexico The Netherlands Norway Poland Puerto Rico Singapore Spain Sweden Switzerland Taiwan United Kingdom Waters Corporation. Waters and Sep-Pak are registered trademarks of Waters Corporation. Milli-Q is a registered trademark of Millipore Corporation. October 2007 WAT Rev 02 VW-PDF Waters Corporation 34 Maple Street Milford, MA U.S.A. T: F:

Protein-Pak Hi Res HIC Column and HIC Protein Standard

Protein-Pak Hi Res HIC Column and HIC Protein Standard Protein-Pak Hi Res HIC Column and HIC Protein Standard CONTENTS I. INTRODUCTION II. a. Mobile Phase b. Flow Direction CONNECTING COLUMN TO LC SYSTEM I. INTRODUCTION This offering contains non-porous, polymethacrylate-based

More information

Hillary B. Hewitson, Thomas E. Wheat, Diane M. Diehl INTRODUCTION

Hillary B. Hewitson, Thomas E. Wheat, Diane M. Diehl INTRODUCTION Enhanc em ent o f t h e U P L C Am ino Ac i d Ana lysis So lu t io n w it h F l e x ibl e Detector Options Hillary B. Hewitson, Thomas E. Wheat, Diane M. Diehl INTRODUCTION The measurement of amino acids

More information

Ultrahydrogel Columns

Ultrahydrogel Columns CONTENTS I. INTRODUCTION II. INSTALLATION a. Solvents b. Column c. Organic solvents d. Aqueous salt and buffer solution e. Flow rate f. Solvent and sample preparation g. Guard columns h. Column installation

More information

Agilent MetaCarb Ca PLUS Carbohydrate Column Ca ++ Form User Manual Part No. A5205

Agilent MetaCarb Ca PLUS Carbohydrate Column Ca ++ Form User Manual Part No. A5205 Agilent MetaCarb Ca PLUS Carbohydrate Column Ca ++ Form User Manual Part No. A5205 WARNING: The Agilent MetaCarb Ca PLUS calcium column is packed with a polymeric material that requires special care. Introduction

More information

Instructions for Capillary Electrophoresis Peptide Analysis Kit

Instructions for Capillary Electrophoresis Peptide Analysis Kit Instructions for Capillary Electrophoresis Peptide Analysis Kit Catalog Number 148-4110 For Technical Service Call Your Local Bio-Rad Office or in the U.S. Call 1-800-4BIORAD (1-800-424-6723) Table of

More information

Chromatography Column Performance and Data Analysis Success Guide. Hints and Tips for Better Purifications

Chromatography Column Performance and Data Analysis Success Guide. Hints and Tips for Better Purifications Chromatography Column Performance and Data Analysis Success Guide Hints and Tips for Better Purifications This Chromatography Success Guide provides practical advice on preparative chromatography and protein

More information

EASY-Spray Columns. Guidance for column set up and installation Tips to maximize column lifetime

EASY-Spray Columns. Guidance for column set up and installation Tips to maximize column lifetime EASY-Spray Columns Guidance for column set up and installation Tips to maximize column lifetime EASY-Spray Column Tips and Tricks This document provides guidance for Thermo Scientific EASY-Spray column

More information

Chromatography Success Guide. Tips and Tricks for Successful Preparative Chromatography and Protein Purification

Chromatography Success Guide. Tips and Tricks for Successful Preparative Chromatography and Protein Purification Chromatography Success Guide Tips and Tricks for Successful Preparative Chromatography and Protein Purification This Chromatography Success Guide provides practical advice on preparative chromatography

More information

AFFINITY HIS-TAG PURIFICATION

AFFINITY HIS-TAG PURIFICATION DESCRIPTION Nickel NTA Agarose Cartridges 5ml are used for purification of histidine-tagged proteins in native or denaturing conditions. This cartridge can be used with an automated chromatography system,

More information

Fast mass transfer Fast separations High throughput and improved productivity Long column lifetime Outstanding reproducibility Low carryover

Fast mass transfer Fast separations High throughput and improved productivity Long column lifetime Outstanding reproducibility Low carryover columns ProSwift Reversed-Phase Monolith Columns for Protein Analysis ProSwift reversed-phase columns use a unique monolith technology for fast, high-resolution HPLC and LC/MS separations of proteins.

More information

Econo-Pac High Capacity Ion Exchange Cartridges, 5 ml Instruction Manual. Catalog Numbers , ,

Econo-Pac High Capacity Ion Exchange Cartridges, 5 ml Instruction Manual. Catalog Numbers , , Econo-Pac High Capacity Ion Exchange Cartridges, 5 ml Instruction Manual Catalog Numbers 732-0026, 732-0027 732-0066, 732-0067 Table of Contents Section 1 Introduction... 1 Section 2 Connection to Bio-Rad

More information

Nuvia cprime Hydrophobic Cation Exchange Media

Nuvia cprime Hydrophobic Cation Exchange Media Nuvia cprime Hydrophobic Cation Exchange Media Instruction Manual Catalog numbers 156-3401 156-3404 156-3402 156-3405 156-3403 156-3406 Please read the instructions in this manual prior to using Nuvia

More information

G Sep Size Exclusion Columns

G Sep Size Exclusion Columns G-Biosciences 1-800-628-7730 1-314-991-6034 technical@gbiosciences.com A Geno Technology, Inc. (USA) brand name G Sep Size Exclusion Columns (Cat. # 786 1297, 786 1298, 786 1300, 786 1301, 786 1302, 786

More information

3. Close the bottom end of the column and apply the packing device on top. Pump water through the upper adaptor to remove air.

3. Close the bottom end of the column and apply the packing device on top. Pump water through the upper adaptor to remove air. INSTRUCTIONS FOR USE WorkBeads Protein A Product name Pack size Article number WorkBeads Protein A Bulk Media 1.5 ml 40605001 Bulk Media 5 ml 40605002 Bulk Media 10 ml 40605003 Bulk Media 100 ml 40605004

More information

Determination of Oxalate and Other Anions in Bayer Liquor Using Ion Chromatography

Determination of Oxalate and Other Anions in Bayer Liquor Using Ion Chromatography Application Note 206 Determination of Oxalate and Other Anions in Bayer Liquor Using Ion Chromatography Introduction The Bayer Process is used to isolate alumina (Al 2 O ) from bauxite. In this process,

More information

Application Note USD Purification of Mouse IgM from Cell Culture Supernatant by Cation Exchange Chromatography on CM Ceramic HyperD F Sorbent

Application Note USD Purification of Mouse IgM from Cell Culture Supernatant by Cation Exchange Chromatography on CM Ceramic HyperD F Sorbent Application Note USD 241 Purification of Mouse IgM from Cell Culture Supernatant by Cation Exchange Chromatography on CM Ceramic HyperD F Sorbent What this Study Demonstrates T h i s s t u d y o n C a

More information

Affi-Gel Protein A MAPS II Kit Instruction Manual

Affi-Gel Protein A MAPS II Kit Instruction Manual Affi-Gel Protein A MAPS II Kit Instruction Manual Catalog Number 153-6159 For Technical Service Call Your Local Bio-Rad Office or in the U.S. Call 1-800-4BIORAD (1-800-424-6723) Table of Contents Introduction...1

More information

AFFINITY HIS-TAG PURIFICATION

AFFINITY HIS-TAG PURIFICATION DESCRIPTION Resins are products that allow batch or column purifications. This product is supplied as a suspension in 50% aqueous suspension containing 30 vol % ethanol. INSTRUCTIONS The resins are adapted

More information

PROCEDURE FOR USE NICKEL NTA Magnetic Agarose Beads (5%)

PROCEDURE FOR USE NICKEL NTA Magnetic Agarose Beads (5%) 1 AFFINITY HIS-TAG PURIFICATION PROCEDURE FOR USE NICKEL NTA Magnetic Agarose Beads (5%) DESCRIPTION Nickel NTA Magnetic Agarose Beads are products that allow rapid and easy small-scale purification of

More information

Chem 321 Lecture 23 - Liquid Chromatography 11/19/13

Chem 321 Lecture 23 - Liquid Chromatography 11/19/13 Chem 321 Lecture 23 - Liquid Chromatography 11/19/13 Student Learning Objectives High Performance Liquid Chromatography With the advent of relatively inexpensive and reliable pumps, the development of

More information

Nickel-NTA Agarose Suspension

Nickel-NTA Agarose Suspension Nickel-NTA Agarose Suspension Agarose beads for purification of His-tagged proteins Product No. A9735 Description Nickel-NTA Agarose Suspension is an agarose-based affinity chromatography resin allowing

More information

MassPREP On-Line Desalting Cartridge

MassPREP On-Line Desalting Cartridge CONTENTS I. INTRODUCTION II. INSTALLING THE MASSPREP ON-LINE DESALTING CARTRIDGE INTO THE SENTRY.1 X 1 MM GUARD COLUMN HOLDER III. RECOMMENDED LC/MS SYSTEM CONFIGURATION TO MINIMIZE MS SOURCE CONTAMINATION

More information

ANALYTICAL STANDARDS AND REAGENTS

ANALYTICAL STANDARDS AND REAGENTS ANALYTICAL STANDARDS AND REAGENTS productivity future growth globalization compliance Waters, the industry leader in liquid chromatography and mass spectrometry, understands that standards and reagents

More information

INSTRUCTIONS The resins are adapted to work mainly in native conditions like denaturing.

INSTRUCTIONS The resins are adapted to work mainly in native conditions like denaturing. 1 AFFINITY HIS-TAG PURIFICATION PROCEDURE FOR USE Nickel NTA Agarose Beads DESCRIPTION Resins are products that allow batch or column purifications. This product is supplied as a suspension in 50% aqueous

More information

[ product solution ] Waters Oasis µ Elution PlateS. Patented Innovation. Elution volume as low as 25 μl. No evaporation and reconstitution

[ product solution ] Waters Oasis µ Elution PlateS. Patented Innovation. Elution volume as low as 25 μl. No evaporation and reconstitution [ product solution ] Waters µ Elution PlateS Elution volume as low as 25 μl No evaporation and reconstitution Ideal for small sample volumes Up to a 15x increase in concentration Patented Innovation Now

More information

ACQUITY UPC 2 BEH, CSH, and HSS Columns

ACQUITY UPC 2 BEH, CSH, and HSS Columns ACQUITY UPC 2 BEH, CSH, and HSS Columns CONTENTS I. INTRODUCTION II. GETTING STARTED a. Safety Considerations b. Column Connectors and Installation c. ecord Installation d. Column Equilibration e. Initial

More information

columns IonPac ICE-AS1 Ion-Exclusion Column Superior Chromatographic Performance Separates weakly ionized acids by differences in pk a

columns IonPac ICE-AS1 Ion-Exclusion Column Superior Chromatographic Performance Separates weakly ionized acids by differences in pk a columns IonPac ICE-AS Ion-Exclusion Column The IonPac ICE-AS ion-exclusion column provides fast analysis of aliphatic organic acids and alcohols in complex- or high-ionic strength samples using a moderately

More information

AFFINITY HIS-TAG PURIFICATION

AFFINITY HIS-TAG PURIFICATION DESCRIPTION Resins are products that allow batch or column purifications. This product is supplied as a suspension in 50% aqueous suspension containing 30 vol % ethanol. INSTRUCTIONS The resins are adapted

More information

Rev. CO Packed Column for Ion Exchange Chromatography. TSKgel STAT Series INSTRUCTION MANUAL TOSOH CORPORATION

Rev. CO Packed Column for Ion Exchange Chromatography. TSKgel STAT Series INSTRUCTION MANUAL TOSOH CORPORATION Rev. CO065000 Packed Column for Ion Exchange Chromatography R TSKgel STAT Series INSTRUCTION MANUAL TOSOH CORPORATION To help protect you and/or your property from potential damage, please read this manual

More information

Userguide: SepFast Gel Filtration columns

Userguide: SepFast Gel Filtration columns Userguide: SepFast Gel Filtration columns 1. Introduction Gel Filtration is a proven technique which is widely used for size-based molecular separation. SepFast Gel Filtration resin is specially designed

More information

1 ml gel corresponds to ml of 75% (v/v) Glutathione Agarose suspension.

1 ml gel corresponds to ml of 75% (v/v) Glutathione Agarose suspension. 1 AFFINITY GST PURIFICATION Procedure for Use Glutathione Agarose 4 Resin DESCRIPTION Glutathione Agarose Resin is used to purify recombinant derivatives of glutathione S-transferases or glutathione binding

More information

Phenyl Sepharose CL-4B

Phenyl Sepharose CL-4B GE Healthcare Instructions 71-7080-00 AE Hydrophobic interaction chromatography Phenyl Sepharose CL-4B Phenyl Sepharose TM CL-4B is a separation medium for hydrophobic interaction chromatography (HIC).

More information

On-Line. User s Guide SPE CARTRIDGES. for Rapid Cleanup and Extraction of Analytes

On-Line. User s Guide SPE CARTRIDGES. for Rapid Cleanup and Extraction of Analytes On-Line SPE CARTRIDGES for Rapid Cleanup and Extraction of Analytes User s Guide What is the strata-x on-line SPE cartridge? The strata-x on-line cartridge combines the revolutionary benefits of the patent

More information

An Automated System for At-Line Process Analysis of Biopharmaceutical Fermentation Reactions

An Automated System for At-Line Process Analysis of Biopharmaceutical Fermentation Reactions An Automated System for At-Line cess Analysis of Biopharmaceutical Fermentation Reactions Frank Steiner, Susanne Fabel, Fraser McLeod, Dionex Corporation, Germering, Germany Abstract Monitoring amino acid

More information

R R Innovation Way P/N SECKIT-7830 Newark, DE 19711, USA Tel: Fax: Website: Published in November 2013

R R Innovation Way P/N SECKIT-7830 Newark, DE 19711, USA Tel: Fax: Website:  Published in November 2013 5-100 Innovation Way Newark, DE 19711, USA Tel:302-3661101 Fax:302-3661151 Website: www.sepax-tech.com Published in November 2013 P/N SECKIT-7830 These Phases are developed based on innovative surface

More information

1. Connect the column to the chromatography system, syringe or pump.

1. Connect the column to the chromatography system, syringe or pump. BabyBio TREN INSTRUCTIONS IN 45 655 030 The ready-to-use BabyBio TREN columns are prepacked with WorkBeads 40 TREN resin and are available in two column sizes, 1 ml and 5 ml. WorkBeads 40 TREN resin for

More information

PESTICIDE SCREENING APPLICATION SOLUTION

PESTICIDE SCREENING APPLICATION SOLUTION PESTICIDE SCREENING APPLICATION SOLUTION The most innovative pesticide screening solution ever built for food, beverage, and water testing. As the global food trade continues to expand, so does the challenge

More information

5 AMP Sepharose 4B. instructions

5 AMP Sepharose 4B. instructions instructions 5 AMP Sepharose 4B 5' AMP Sepharose 4B interacts strongly with NAD + -dependent dehydrogenases and ATP-dependent enzymes. Selective elution with gradients of NAD + or NADP + has allowed the

More information

Improving Resolution and Column Loading Systematically in Preparative Liquid Chromatography for Isolating a Minor Component from Peppermint Extract

Improving Resolution and Column Loading Systematically in Preparative Liquid Chromatography for Isolating a Minor Component from Peppermint Extract Improving Resolution and Column Loading Systematically in Preparative Liquid Chromatography for Isolating a Minor Component from Peppermint Extract Jo-Ann M. Jablonski and Rui Chen Waters Corporation,

More information

ACQUITY UPLC Protein BEH SEC Columns and Standards

ACQUITY UPLC Protein BEH SEC Columns and Standards and Standards CONTENTS I. INTRODUCTION II. CONFIGURING AN ACQUITY UPLC SYSTEM FOR USE IN SEC PROTEIN SEPARATIONS a. Calibrators III. GETTING STARTED a. ecord installation b. Column connectors c. Column

More information

New Core-Shell Technology

New Core-Shell Technology New Core-Shell Technology Fortis Speedcore columns are the very latest in core-shell technology. Incorporating our optimised bonding and packing practices with a core-shell particle provides the analyst

More information

ISOLATE II Blood DNA Kit. Product Manual

ISOLATE II Blood DNA Kit. Product Manual ISOLATE II Blood DNA Kit Product Manual 2 Product Manual www.bioline.com/isolate Blood DNA Kit ISOLATE II Blood DNA Kit ISOLATE II Blood DNA Kit 1 Kit contents 04 2 Description 04 3 Storage 05 4 Safety

More information

Size-Select Kit for NGS Library Preparation Product # 53600

Size-Select Kit for NGS Library Preparation Product # 53600 3430 Schmon Parkway Thorold, ON, Canada L2V 4Y6 Phone: 866-667-4362 (905) 227-8848 Fax: (905) 227-1061 Email: techsupport@norgenbiotek.com Size-Select Kit for NGS Library Preparation Product # 53600 Product

More information

Pulsed Field Gel Electrophoresis

Pulsed Field Gel Electrophoresis Pulsed Field Gel Electrophoresis Protocol Bulletin 6225 Restriction Enzyme Digestion Restriction enzyme digestion is used to prepare DN for analysis or other processing steps. Guidelines about preparing

More information

Electrophoresis. Ready-to-Run Buffers and Solutions

Electrophoresis. Ready-to-Run Buffers and Solutions Electrophoresis Ready-to-Run Buffers and Solutions Bio-Rad is a premier provider of buffers and premixed reagents for life science research. We offer a variety of different products for all your protein

More information

ProteoMiner Protein Enrichment Technology

ProteoMiner Protein Enrichment Technology Sample Preparation ProteoMiner Protein Enrichment Technology Digging Deeper in the Proteome Detect More Proteins With ProteoMiner Technology Than With Immunodepletion ProteoMiner Technology vs. an Agilent

More information

AFFINITY GST PURIFICATION

AFFINITY GST PURIFICATION DESCRIPTION Glutathione Agarose Resin is used to purify recombinant derivatives of glutathione S-transferases or glutathione binding proteins. are products that allow batch or column purifications. Purification

More information

ENSURING QUALITY THROUGH COMPLIANCE [ COMPLIANCE ]

ENSURING QUALITY THROUGH COMPLIANCE [ COMPLIANCE ] ENSURING QUALITY THROUGH COMPLIANCE [ COMPLIANCE ] BEING IN COMPLIANCE HAS NEVER BEEN MORE IMPORTANT Quality encompasses more than meeting regulatory requirements. It extends throughout your organization

More information

A Fast and Robust Linear ph Gradient Separation Platform for Monoclonal Antibody (MAb) Charge Variant Analysis

A Fast and Robust Linear ph Gradient Separation Platform for Monoclonal Antibody (MAb) Charge Variant Analysis A Fast and Robust Linear ph Gradient Separation Platform for Monoclonal Antibody (MAb) Charge Variant Analysis Shanhua Lin, Julia Baek, and Chris Pohl Thermo Fisher Scientific, Sunnyvale, CA, USA Overview

More information

TechnicalNOTE METABOLYNX EXACT MASS DEFECT FILTER: RAPIDLY DISCRIMINATE BETWEEN TRUE METABOLITES AND FALSE POSITIVES INTRODUCTION

TechnicalNOTE METABOLYNX EXACT MASS DEFECT FILTER: RAPIDLY DISCRIMINATE BETWEEN TRUE METABOLITES AND FALSE POSITIVES INTRODUCTION METABOLYNX EXACT MASS DEFECT FILTER: RAPIDLY DISCRIMINATE BETWEEN TRUE METABOLITES AND FALSE POSITIVES Jose M. Castro-Perez 1, Jane Kirby 1, Laurent Leclercq 2, Claude Delatour 2, Jean-Michel Mineur 2,

More information

Analysis of biomolecules by SEC and Ion-Exchange UPLC

Analysis of biomolecules by SEC and Ion-Exchange UPLC Analysis of biomolecules by SEC and Ion-Exchange UPLC Anders Feldthus, Waters Nordic 2011 Waters Corporation 1 Waters Commitment To develop, commercialize and market columns that when used on Waters ACQUITY

More information

PRODUCT MANUAL FOR THE. ELECTROLYTIC ph MODIFIER (EPM) EPM, P/N

PRODUCT MANUAL FOR THE. ELECTROLYTIC ph MODIFIER (EPM) EPM, P/N PRODUCT MANUAL FOR THE EGC II K 2 CO 3 CARTRIDGE EGC II K 2 CO 3 CARTRIDGE, P/N 058904 ELECTROLYTIC ph MODIFIER (EPM) EPM, P/N 063175 EGC CARBONATE MIXER EGC CARBONATE MIXER, 2-mm, P/N 063443 EGC CARBONATE

More information

E.Z.N.A. Water DNA Kit. D preps D preps D preps

E.Z.N.A. Water DNA Kit. D preps D preps D preps E.Z.N.A. Water DNA Kit D5525-00 5 preps D5525-01 50 preps D5525-02 200 preps April 2017 E.Z.N.A. Water DNA Kit Table of Contents Introduction and Overview...2 Kit Contents/Storage and Stability...3 Preparing

More information

Kit Components Product # (50 samples) Wash Solution A Elution Buffer B

Kit Components Product # (50 samples) Wash Solution A Elution Buffer B 3430 Schmon Parkway Thorold, ON, Canada L2V 4Y6 Phone: 866-667-4362 (905) 227-8848 Fax: (905) 227-1061 Email: techsupport@norgenbiotek.com Cells and Tissue DNA Isolation Kit Product # 53100 Product Insert

More information

E.Z.N.A. Blood DNA Maxi Kit. D preps D preps

E.Z.N.A. Blood DNA Maxi Kit. D preps D preps E.Z.N.A. Blood DNA Maxi Kit D2492-00 2 preps D2492-03 50 preps April 2014 E.Z.N.A. Blood DNA Maxi Kit Table of Contents Introduction and Overview...2 Kit Contents/Storage and Stability...3 Preparing Reagents...4

More information

QIAcube HT Your Purification Expert

QIAcube HT Your Purification Expert HT Your Purification Expert Fast and reliable 96-well nucleic acid purification Sample & Assay Technologies Economical, high-throughput nucleic acid purification from virtually all sample types enables

More information

Product # Specifications. Kit Specifications Maximum Column Binding Capacity (gdna) Maximum Column Loading Volume.

Product # Specifications. Kit Specifications Maximum Column Binding Capacity (gdna) Maximum Column Loading Volume. 3430 Schmon Parkway Thorold, ON, Canada L2V 4Y6 Phone: 866-667-4362 (905) 227-8848 Fax: (905) 227-1061 Email: techsupport@norgenbiotek.com FFPE DNA Purification Kit Product # 47400 Product Insert Norgen

More information

Fast and reliable purification of up to 100 µg of transfection-grade plasmid DNA using a spin-column.

Fast and reliable purification of up to 100 µg of transfection-grade plasmid DNA using a spin-column. INSTRUCTION MANUAL ZymoPURE Plasmid Miniprep Kit Catalog Nos. D4209, D4210, D4211 & D4212 (Patent Pending) Highlights Fast and reliable purification of up to 100 µg of transfection-grade plasmid DNA using

More information

CORESTA RECOMMENDED METHOD N 61

CORESTA RECOMMENDED METHOD N 61 CORESTA RECOMMENDED METHOD N 61 DETERMINATION OF 1,2-PROPYLENE GLYCOL, GLYCEROL AND SORBITOL IN TOBACCO AND TOBACCO PRODUCTS BY HIGH PERFORMANCE LIQUID CHROMATOGRAPHY (HPLC) (July 2015) 0. INTRODUCTION

More information

Average Yields* Yeast DNA Yeast RNA Time to Complete 10 Purifications * Yield will vary depending on the type of sample processed

Average Yields* Yeast DNA Yeast RNA Time to Complete 10 Purifications * Yield will vary depending on the type of sample processed 3430 Schmon Parkway Thorold, ON, Canada L2V 4Y6 Phone: 866-667-4362 (905) 227-8848 Fax: (905) 227-1061 Email: techsupport@norgenbiotek.com Fungi/Yeast RNA/DNA Purification Kit Product # 35800 Product Insert

More information

Fraction Analysis of Cysteine Linked Antibody-Drug Conjugates Using Hydrophobic Interaction. chromatography. Agilent 1260 Infinity II Bio-Inert System

Fraction Analysis of Cysteine Linked Antibody-Drug Conjugates Using Hydrophobic Interaction. chromatography. Agilent 1260 Infinity II Bio-Inert System Application Note Biologics & Biosimilars Fraction Analysis of Cysteine Linked Antibody-Drug Conjugates Using Hydrophobic Interaction Chromatography Agilent 126 Infinity II Bio-Inert System 7 6 5 4 5. 7.5

More information

BabyBio S BabyBio Q BabyBio DEAE

BabyBio S BabyBio Q BabyBio DEAE DATA SHEET BabyBio S BabyBio Q BabyBio DEAE BabyBio S, BabyBio Q and BabyBio DEAE are pre-packed, ready to use ion exchange columns for easy and convenient purification of proteins. Rapid method screening

More information

ENHANCED MS WITH THE TURN OF A KEY

ENHANCED MS WITH THE TURN OF A KEY [ ionkey/ms ] ENHANCED MS WITH THE TURN OF A KEY EXPERIENCE THE FUTURE OF MS, EXPERIENCE ionkey/ms THE LC-MS YOU KNOW, LIKE YOU VE NEVER SEEN BEFORE ionkey/ms Systems integrate the UPLC separation into

More information

Developing Robust and Efficient IEX Methods for Charge Variant Analysis of Biotherapeutics Using ACQUITY UPLC H-Class System and Auto Blend Plus

Developing Robust and Efficient IEX Methods for Charge Variant Analysis of Biotherapeutics Using ACQUITY UPLC H-Class System and Auto Blend Plus Developing Robust and Efficient IEX Methods for Charge Variant Analysis of Biotherapeutics Using ACQUITY UPLC H-Class System and Auto Blend Plus Robert Birdsall, Thomas Wheat, and Weibin Chen Waters Corporation,

More information

KOBRA CELL Art. No. K01

KOBRA CELL Art. No. K01 R-Biopharm Rhône Ltd KOBRA CELL Art. No. K01 Instruction Manual R-Biopharm for the safety of your analysis. R-Biopharm Rhone Ltd 2 Contents 1. Introduction to the KOBRA CELL... 4 1.1 Intended Use... 4

More information

Application Note. Author. Abstract. Biopharmaceuticals. Verified for Agilent 1260 Infinity II LC Bio-inert System. Sonja Schneider

Application Note. Author. Abstract. Biopharmaceuticals. Verified for Agilent 1260 Infinity II LC Bio-inert System. Sonja Schneider Combining small-scale purification and analysis of monoclonal antibodies on one instrument Protein purification with high-volume injection using the Agilent 126 Infinity Bio-inert Quaternary LC System

More information

Module 16: Gel filtration: Principle, Methodology & applications. Dr. Savita Yadav Professor Department of Biophysics AIIMS, New Delhi

Module 16: Gel filtration: Principle, Methodology & applications. Dr. Savita Yadav Professor Department of Biophysics AIIMS, New Delhi PAPER 9: TECHNIQUES USED IN MOLECULAR BIOPHYSICS I Module 16: Gel filtration: Principle, Methodology & applications Dr. Savita Yadav Professor Department of Biophysics AIIMS, New Delhi Module 16 Gel filtration:

More information

E.Z.N.A. Yeast Plasmid Mini Kit. D preps D preps

E.Z.N.A. Yeast Plasmid Mini Kit. D preps D preps E.Z.N.A. Yeast Plasmid Mini Kit D3376-00 5 preps D3376-01 50 preps November 2015 E.Z.N.A. Yeast Plasmid Mini Kit Table of Contents Introduction and Overview...2 Kit Contents/Storage and Stability...3 Preparing

More information

SYKAM Amino Acid Analyzer

SYKAM Amino Acid Analyzer C H R O M AT O G R A P H Y SYKAM Amino Acid Analyzer Automatic Amino Acid Analyzer S 433 The innovative automatic Amino Acid Analyzer S 433 combines the advantages of the classical ion exchange separation

More information

Tissue & Cell Genomic DNA Purification Kit. Cat. #:DP021/ DP Size:50/150 reactions Store at RT For research use only

Tissue & Cell Genomic DNA Purification Kit. Cat. #:DP021/ DP Size:50/150 reactions Store at RT For research use only Tissue & Cell Genomic DNA Purification Kit Cat. #:DP021/ DP021-150 Size:50/150 reactions Store at RT For research use only 1 Description: The Tissue & Cell Genomic DNA Purification Kit provides a rapid,

More information

Product # 24700, 24750

Product # 24700, 24750 3430 Schmon Parkway Thorold, ON, Canada L2V 4Y6 Phone: 866-667-4362 (905) 227-8848 Fax: (905) 227-1061 Email: techsupport@norgenbiotek.com Genomic DNA Isolation Kit Product # 24700, 24750 Product Insert

More information

MAbPac SCX-10 Columns

MAbPac SCX-10 Columns User Manual MAbPac SCX-10 Columns 065393 Revision 09 October 2015 For Research Use Only. Not for use in diagnostic procedures. Product Manual for MAbPac SCX-10, 10µm Analytical Column (10μm, 9 x 250 mm,

More information

AnaTag HiLyte Fluor 647 Protein Labeling Kit

AnaTag HiLyte Fluor 647 Protein Labeling Kit AnaTag HiLyte Fluor 647 Protein Labeling Kit Catalog # 72049 Kit Size 3 Conjugation Reactions This kit is optimized to conjugate HiLyte Fluor 647 SE to proteins (e.g., IgG). It provides ample materials

More information

ProteoSpin Urine Protein Concentration Maxi Kit Product # 21600

ProteoSpin Urine Protein Concentration Maxi Kit Product # 21600 3430 Schmon Parkway Thorold, ON, Canada L2V 4Y6 Phone: 866-667-4362 (905) 227-8848 Fax: (905) 227-1061 Email: techsupport@norgenbiotek.com ProteoSpin Urine Protein Concentration Maxi Kit Product # 21600

More information

Please read all the information in booklet before using the unit

Please read all the information in booklet before using the unit 100 Please read all the information in booklet before using the unit Bioneer Corporation 8-11,Munpyeongseo-ro, Daedeok-gu, Daejeon 34302, Republic of Korea Tel: +82-42-930-8777 Fax: +82-42-930-8688 Email:

More information

QIAcube Pure Efficiency

QIAcube Pure Efficiency QIAcube Pure Efficiency Sample & Assay Technologies Walkaway spin-column processing The QIAcube automates your spin preps The revolutionary QIAcube makes automated sample prep available to all labs. The

More information

This User Bulletin does not apply to PROCISE clc instruments.

This User Bulletin does not apply to PROCISE clc instruments. User Bulletin PROCISE Protein Sequencing System (491/492/494) 1 July 1997 (updated 06/2001) SUBJECT: PROCISE Feature Upgrade Kit Summary This User Bulletin describes the hardware, firmware, and software

More information

MD60002 MD MD62002

MD60002 MD MD62002 MagSi-DNA saliva Art.No. MD60002 MD61002 - MD62002 Product Manual Version 1.0 22/01/2015 Table of Contents 1. General Information...3 1.1 Intended Use...3 1.2 Kit specifications...3 1.3 Basic principle...3

More information

Purification of Recombinant Proteins on Nuvia TM cprime TM Hydrophobic Cation Exchange Media: A Simple Approach to Method Development

Purification of Recombinant Proteins on Nuvia TM cprime TM Hydrophobic Cation Exchange Media: A Simple Approach to Method Development Purification of Recombinant Proteins on Nuvia TM cprime TM Hydrophobic Cation Exchange Media: A Simple Approach to Method Development Xuemei M. He, Sherif Hanala, and Mark Snyder, Bio-Rad Laboratories,

More information

Note: for laboratory research use only. RNA High-purity Total RNA Rapid Extraction Kit (Spin-column) Signalway Biotechnology

Note: for laboratory research use only. RNA High-purity Total RNA Rapid Extraction Kit (Spin-column) Signalway Biotechnology Note: for laboratory research use only RNA High-purity Total RNA Rapid Extraction Kit (Spin-column) Cat. #: RP1202 (50preps) Signalway Biotechnology I. Kit Content, Storage Condition and Stability Content

More information

Cytoplasmic & Nuclear RNA Purification Kit Product # 21000, 37400

Cytoplasmic & Nuclear RNA Purification Kit Product # 21000, 37400 3430 Schmon Parkway Thorold, ON, Canada L2V 4Y6 Phone: 866-667-4362 (905) 227-8848 Fax: (905) 227-1061 Email: techsupport@norgenbiotek.com Cytoplasmic & Nuclear RNA Purification Kit Product # 21000, 37400

More information

HIC as a Complementary, Confirmatory Tool to SEC for the Analysis of mab Aggregates

HIC as a Complementary, Confirmatory Tool to SEC for the Analysis of mab Aggregates HIC as a Complementary, Confirmatory Tool to SEC for the Analysis of mab Aggregates Julia Baek, Shanhua Lin, Xiaodong Liu, Thermo Fisher Scientific, Sunnyvale, CA Application Note 2126 Key Words Size exclusion

More information

Method Development Considerations for Reversed-Phase Protein Separations

Method Development Considerations for Reversed-Phase Protein Separations Method Development Considerations for Reversed-Phase Protein Separations Hillary B. Hewitson, Thomas E. Wheat, Paula Hong, Kenneth J. Fountain APPLICATION BENEFITS n The BEH00 C 4 chemistry is available

More information

Kit Specifications 45 g. 45 g of RNA 8 L

Kit Specifications 45 g. 45 g of RNA 8 L RNA Clean-Up and Concentration Micro-Elute Kit Product # 61000 Product Insert Norgen s RNA Clean-Up and Concentration Micro-Elution Kit provides a rapid method for the purification, cleanup and concentration

More information

AxyPrep PCR Clean-up Kit AxyPrep-96 PCR Clean-up Kit

AxyPrep PCR Clean-up Kit AxyPrep-96 PCR Clean-up Kit AxyPrep PCR Clean-up Kit AxyPrep-96 PCR Clean-up Kit For the purification of amplicons from PCRs Kit contents, storage and stability AxyPrep PCR Clean-up Kit Cat. No. AP-PCR-4 AP-PCR-50 AP-PCR-250 Kit

More information

[ VION IMS QTOF ] BEYOND RESOLUTION

[ VION IMS QTOF ] BEYOND RESOLUTION [ VION IMS QTOF ] BEYOND RESOLUTION THE BENEFITS ARE CLEAR and routinely available, Complex samples give complex data with overlapping spectra and background interferences, making compound identification

More information

A RAPID METHOD FOR THE DETERMINATION OF CHLORAMPHENICOL RESIDUES IN BLACK TIGER SHRIMP

A RAPID METHOD FOR THE DETERMINATION OF CHLORAMPHENICOL RESIDUES IN BLACK TIGER SHRIMP INTRODUCTION The global production of fish, as a food item, has greatly increased over the last few decades and is now at the highest levels on record. Currently, more than 5 of world protein consumption

More information

RNA Clean-Up and Concentration Kit Product # 23600, 43200

RNA Clean-Up and Concentration Kit Product # 23600, 43200 3430 Schmon Parkway Thorold, ON, Canada L2V 4Y6 Phone: 866-667-4362 (905) 227-8848 Fax: (905) 227-1061 Email: techsupport@norgenbiotek.com RNA Clean-Up and Concentration Kit Product # 23600, 43200 Product

More information

TaKaRa MiniBEST Agarose Gel DNA Extraction Kit Ver.4.0

TaKaRa MiniBEST Agarose Gel DNA Extraction Kit Ver.4.0 Cat. # 9762 For Research Use TaKaRa MiniBEST Agarose Gel Product Manual Table of Contents I. Description... 3 II. Kit Components... 3 III. Shipping and Storage... 3 IV. Preparation before Use... 4 V. Protocol...

More information

AccuPrep GMO DNA Extraction Kit Cat. No.: K-3031

AccuPrep GMO DNA Extraction Kit Cat. No.: K-3031 AccuPrep GMO DNA Extraction Kit Cat. No.: K-3031 V109C3 Safety Warnings and Precautions. For research use only. It is not recommended to diagnose disease in humans or animals. DO NOT use to humans or animals.

More information

Histology FISH Accessory Kit Step-by-Step Procedure

Histology FISH Accessory Kit Step-by-Step Procedure PROCEDURE Histology FISH Accessory Kit Step-by-Step Procedure Code K5799 For probes diluted in Formamide based hybridization buffer Reagent Preparation Equilibration Deparaffinization/Rehydration Prepare

More information

Operating Instructions

Operating Instructions POROS 50 HQ, DEAE, and PI Perfusion Chromatography Bulk Media for Anion Exchange Chromatography Operating Instructions Note: These instructions cover the specific operational characteristics of POROS 50

More information

MicroElute Cycle-Pure Kit

MicroElute Cycle-Pure Kit MicroElute Cycle-Pure Kit D6293-00 5 preps D6293-01 50 preps D6293-02 200 preps MicroElute Gel Extraction Kit D6294-00 5 preps D6294-01 50 preps D6294-02 200 preps MicroElute DNA Clean Up Kit D6296-00

More information

QIAsymphony SP Protocol Sheet

QIAsymphony SP Protocol Sheet QIAsymphony SP Protocol Sheet Reference_200_V5 protocol General information These protocols are for purification of total DNA from samples encountered in forensic, human identity, and biosecurity applications

More information

Fast, easy, reliable, ultra-pure transfection grade plasmid DNA Maxiprep using a microcentrifuge spin-column.

Fast, easy, reliable, ultra-pure transfection grade plasmid DNA Maxiprep using a microcentrifuge spin-column. INSTRUCTION MANUAL ZymoPURE II Plasmid Maxiprep Kit Catalog Nos. D4202 & D4203 (Patent Pending) Highlights Fast, easy, reliable, ultra-pure transfection grade plasmid DNA Maxiprep using a microcentrifuge

More information

1. Bloomsbury BBSRC Centre for Structural Biology, Birkbeck College and University College London.

1. Bloomsbury BBSRC Centre for Structural Biology, Birkbeck College and University College London. Purification/Polishing of His-tagged proteins - Application of Centrifugal Vivapure Ion-exchange Membrane Devices to the Purification/Polishing of Histagged Background Multi-milligram quantities of highly

More information

HiTrap Heparin HP, 1 ml and 5 ml

HiTrap Heparin HP, 1 ml and 5 ml GE Healthcare Instructions 71-7004-00 AS HiTrap affinity columns HiTrap Heparin HP, 1 ml and 5 ml HiTrap Heparin HP is a prepacked ready to use, column for preparative affinity chromatography. The special

More information

Polyvidone Polyvinylpyrrolidone H 2 C H C N

Polyvidone Polyvinylpyrrolidone H 2 C H C N 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 (C 6 H 9 NO)n [9003-39-8] Poly [(2-oxo-1-pyrrolidinyl) ethylene] Povidone (Rev. 1, Stage 4)

More information

Bivalirudin Purification:

Bivalirudin Purification: Bivalirudin Purification: Sorbent Screening and Overload Experiments Marc Jacob, Joshua Heng, and Tivadar Farkas Phenomenex, Inc., 411 Madrid Ave., Torrance, CA 90501 USA PO94190412_W Abstract In this

More information

NPTEL VIDEO COURSE PROTEOMICS PROF. SANJEEVA SRIVASTAVA

NPTEL VIDEO COURSE PROTEOMICS PROF. SANJEEVA SRIVASTAVA LECTURE-06 PROTEIN PURIFICATION AND PEPTIDE ISOLATION USING CHROMATOGRAPHY TRANSCRIPT Welcome to the proteomics course. Today, we will talk about protein purification and peptide isolation using chromatography

More information