Comparison of Trypan Blue and Fluorescence-Based Viability Detection Methods Via Morphological Observation

Size: px
Start display at page:

Download "Comparison of Trypan Blue and Fluorescence-Based Viability Detection Methods Via Morphological Observation"

Transcription

1 Comparison of Trypan Blue and Fluorescence-Based Viability Detection Methods Via Morphological Observation Nexcelom Bioscience LLC. 360 Merrimack Street, Building 9 Lawrence, MA T: F: E: info@nexcelom.com Rev. A

2 Morphological Analyses Comparing Trypan Blue and AOPI Viability Staining Methods Using Cellometer Image Cytometers Introduction All cell-based biological experiments, from standard cell culture to primary cells for evaluating pharmacological agents, require accurate determination of cell viability. The Trypan Blue (TB) exclusion assay is one of the oldest and most common methods by which to assess cell viability. (Altman et al. 1993; Louis and Siegel 2011). TB is cell membrane impermeable. Consequently, it enters cells only in the event of compromised membranes. Once inside a membrane compromised cell, TB turns the cells a blue color after it binds to intracellular proteins. The live cells within a population remain unstained. Although TB staining has long been the default method to evaluate cell viability, the method has some substantial shortcomings. Because TB itself is toxic, a short window of time exists between cell staining and counting (Tsaousis et al. 2012). Additionally, TB is prone to non-specific binding to cellular artifacts, particularly in clinical and primary cell samples, and, as of yet, there is no standardization of TB concentration to analyze cell viability. Lastly, TB staining requires manual counting under bright field illumination with a hemacytometer, a process which is laborious and introduces user-dependent errors. Fluorescent-based dyes now provide an alternative to the traditional TB method. Fluorescent viability dyes including DAPI, Hoechst 33342, ethidium bromide, propidium iodide (PI), SYTOX orange and red, acridine orange (AO), SYTO9/13, DRAQ5/7, calcein AM, and CFDA have all been used to determine viability measurements in flow or image-based cytometers (Wallen et al. 1980; Al-Rubeai et al. 1997; Foglieni et al. 2001; Chan et al. 2012). The AO and PI dual staining method is regularly implemented to assess viability of nucleated cells (Wallen et al. 1980; Darzynkiewicz et al. 1992; Mascotti et al. 2000; Foglieni et al. 2001; Ling et al. 2003; Solomon et al. 2010; Chan et al. 2012, 2013). AO can enter cell membranes, where it binds to DNA and RNA in live cells. PI is membrane impermeable and binds to DNA and RNA only when cell membranes are compromised in dying, dead, and necrotic cells. Cell viability within a given population is then determined by calculating the ratio of live cells to dead cells cells (Wallen et al. 1980). The past 20 years have seen many publications comparing TB exclusion and fluorescence-based cellular viability stains (Black and Berenbaum 1964; Jones and Senft 1985; Altman et al. 1993; Mascotti et al. 2000). 2

3 These studies reveal that over a time course analysis in standard cell culture, TB exclusion reports more viability in comparison to the fluorescence-based tools, although the studies did not speculate as to the reasoning behind the differences noted (Jones and Senft 1985; Mascotti et al. 2000). In this work, we perform a head-to-head comparison between cell viability as determined by TB and that determined by AO/PI using an automated imagebased cytometer, the Cellometer. Image-based cytometry allowed us to examine the morphological changes that occur in TB-stained cells, and these may explain the differences in viability counts reached by the two methods. Materials and Methods Preparation of Jurkat cells Jurkat cells were cultured in RPMI-1640 medium at 37ºC and 5% CO 2. For an analysis of naturally-dying cells, 5mL of cells (~4x10 6 cells/ml) were removed from the incubator and placed into a bench-top drawer. The sample remained at room temperature for the duration of the experiment. A small aliquot of these cells was taken and tested at the following time points: 0, 6, 12, 24, 48, 72, 96, and 168 hours. A similar experiment was run in parallel to analyze the impact of different TB concentrations on Jurkat cell viability, which was measured at time points 0, 3, 6, 9, 12, 27, and 33 hours. For the heat-killed viability comparison, 5mL of cells were placed into a tube and dropped into a boiling water bath for 15 minutes. After heat-killing, five samples were prepared at theoretical viabilities of 0, 25, 50, 75, and 100% by mixing the heat-killed with fresh Jurkat cells. Time course analysis of naturally-killed Jurkat cells Four detection methods were used to assess the naturally-dying cells: (1, 2) measuring viability with PI or AO/PI and counting with Cellometer Vision; (3) staining with 0.4% TB and counting with Cellometer Auto T4; (4) staining with 0.4% TB and manually counting with a hemacytometer. Samples of the naturally-dying cells at 0, 6, 12, 24, 48, 72, 96, and 168 hours were mixed with PI, AO/PI, or 0.4% TB and analyzed accordingly. Viability and concentration results were measured in quadruplicate for all time points. Time course analysis of naturally-killed fresh bulk murine splenocytes A lysing protocol was used to lyse the red blood cells in the splenocyte sample. Splenocyte viability was assessed at 0, 3, and 50 hours by staining with 0.4% TB, PI, or AO/PI to compare the various methods in these primary cells. TB-stained samples were measured manually by hemacytometer, while the AO/PI, and PI-stained samples were analyzed with Cellometer Vision. Again, all time point samples were assessed in quadruplicate. 3

4 Assessing impact of trypan blue concentration on viability assessment To measure the impact of TB concentration of viability of naturally-killed Jurkat cells, two methods were used: (1) AO/PI staining followed by analysis with Cellometer Vision; (2) staining with various concentrations of TB and manually counting with a hemacytometer. Samples at each time point (0, 3, 6, 9, 12, 27, and 33 hours) were stained by AO/PI or by 0.4, 0.2, 0.1, 0.05, 0.025, and % TB. All samples were analyzed in quadruplicate. Image cytometry protocol Immediately after staining, Jurkat cells were pipetted into a Nexcelom disposable counting chamber. AO was detected by optics module VB (pseudo-color green) and PI by VB (pseudo-color red). A color camera on the Cellometer Auto T4 was used to detect TB staining. In the case of TB-stained cells, live cells with bright centers and dark, blue-stained cells were counted. For PI analysis, total cells were determined in the bright field channel and dead cells in the fluorescent channel were counted. For AO/PI, AO-stained live cells (green) and PI-stained dead cells (orange) were counted. The viabilities were then calculated for each sample. Manual counting with the hemactyometer TB-stained Jurkat cells were pipetted into a Neubauer hemacytomter. Live and dead cells were counted under bright field from both heat-killed and naturally-dying Jurkat cells. Samples with approximately 50% viability were used for Figures 7 and 8. Phase contrast microscopy Jurkat cells stained with 0.4 and 0.1% TB were imaged under bright field using a digital camera. Both bright field and phase contrast images were captured to detect dimly diffused Jurkat cells. Results Time course analysis of naturally-killed Jurkat cells Figure 1A reports TB, PI, and AO/PI stained cells in the naturally-killed samples (0-168 hours) The bright field images report gradual morphological changes in the cell membranes. At 0-12 hours, the outer membranes were dark and quite defined. By 168 hours, cells were now grayish and undefined (green arrows in Figure 1b). TB-stained Jurkat cells showed similar morphological changes, from clearly defined to amorphous. Tight, dark cells became dim and diffuse shapes (red arrows Figure 4

5 1b). The number of TB-stained large dim cells rose over time, while the morphological characteristics of AO or PI-stained cells remained constant. All stained dead cells (PI, AO/PI, and TB) increased over time. In the PI-stained samples, the later time points (96 and 168 hours) also showed less clearly defined cells under bright field, which may hamper the ability of obtaining accurate cell viability with PI (Figure 1a). AO/PI was the optimal method for measuring cell viability as the labeling of both live and dead cells eliminated the chance of counting membrane-poor cells, those late apoptotic and necrotic cells, as well as eliminating the counting of cellular debris under bright field. Figure 1a and b. Bright field and fluorescent images of PI, AO/PI, and TB-stained naturally-killed Jurkat cells. (a) Time points post incubation at room temperature (0-168 hours). (b) Higher magnification of time course images illustrating dim, diffused shapes (arrows). 5

6 Figure 2 compares the image cytometry and hemacytometer measurements. Clear differences between the fluorescence-based and TB assays were seen (2a). AO/PI and PI reported sharp declines in Jurkat viability at 12 hours. A 2-sample T-test at 12 hours comparing AO/PI to PI and AO/PI to TB showed that AO/PI was comparable to PI staining (p>0.05) but significantly different than TB counting (p<0.05). Even though AO/PI reported 70% viability, the TB method reported a measurement of 82.8% viability. These data confirm earlier studies which report that TB-stained cell samples lower than 80% typically overestimate cell viability (Mascotti et al. 2000). The fluorescence-based methods reported consistent total cell concentrations over the duration of the experiment, but the TB method showed a gradual decrease in cell concentration (Figure 2b) Figure 2c, d shows a plotting of measured live cell concentration (c) and measured dead cell population (d). All methods reported similar concentration measurements 6

7 over time for the live cells (c). The dead cell concentrations varied significantly, however. By the end of the time course, both the manual and TB-derived concentrations were greater than two times lower than the fluorescence-based detection methods. The initial differences in dead cell count by TB began at 12 hours which coincides with the time when AO/PI-measured cell viability decreased to ~70%. These data suggest that as viability declines, TB undercounts the number of dead cells, producing an inflated viability measurement. Figure 2a-d. Time course analysis (0-168 hours) of viability (a), total cell concentration (b), live cell count (c), and dead cell count (d) for TB, PI, and AO/PIstained, naturally-killed Jurkat cells. 7

8 Time course analysis of naturally-killed fresh bulk murine splenocytes Viability data in splenocytes were similar to those reported for Jurkat cells Figure 3 shows the AO/PI and PI viability data gradually decreasing from ~80% at time point 0 to ~65% at 50 hours post incubation. TB-stained samples at 0.4 and 0.1% concentrations showed a decrease in viability of only ~5%. Again, a 2-sample T-test revealed that there was a significantly difference in the viability between AO/PI and TB (p<0.05), whereas AO/PI was not significantly different than PI (p>0.05). Figure 3. Time course analysis (0, 33, 50 hours) of viability for TB, PI, and AO/PIstained primary murine splenocytes. Assessing impact of trypan blue concentration on viability assessment Figure 4 shows the images of cells stained with 0.4, 0.1, and 0.025% TB as well a those stained with AO/PI. High concentrations of TB (0.4%) stained more cells but also produced more large, dimly diffused shapes (red arrows). Lower TB concentrations (0.1 and 0.05%) stained fewer cells but those that were stained were tighter, dark shapes, unlike the ones seen at the higher TB concentrations, especially at 12 hours and beyond. AO/PI staining was consistent with previous results, reporting an increase in PIpositive cells over the time course. 8

9 Figure 4. Image from the time course analysis (0-33 hours) of varying concentrations of TB in naturally-killed Jurkat cells. Figure 5a shows that the viability results produced a trend similar to that scene in the time-course assay. Again, viabilities from each TB concentration were higher than from those derived from the AO/PI cells. Because this experiment had a lower initial viability (~80%), the differences were immediately noticeable with the TB viability counts always higher than the AO/PI counts. Again, a 2-sample T-test report that AO/PI viability counts were significantly different than all the TB concentrations (p<0.05). The measured concentration of dead cells for each parameter was significantly different (p<0.05) for AO/PI against all TB concentrations (Figure 5b, c), the largest difference occurring between AO/PI and 0.025% TB, likely due to the low concentration of TB. Even at higher concentrations of TB, lower than expected number of dead cells are reported, which confirms our previous findings that TB routinely overestimates viability and demonstrates that this phenomenon occurs at every TB concentration tested here. 9

10 Figure 5a-c. Time course analysis (0-33 hours) of concentration and viability with different concentrations of TB ( %). Viability is shown in a, live cell count in b, and dead cell count in c. Heat-killed viability comparison This experiment aimed to demonstrate that morphological differences between cells killed by heat and cells killed naturally impact how those cells are counted. Figure 6 shows clear morphological differences between heat-killed and non-heatkilled cells. Heat-killed cells did not undergo natural death by apoptosis or necrosis and so the cells are consistent in shape, color, circularity, and TB-staining pattern. The cells are tight and dark blue in color, and the viability counts of these cells between TB and PI were highly correlated (Figure 7). PI was used with these cells as heat-killed cells generate little debris. The non-heat-killed cells, however, did undergo apoptosis and necrosis. Therefore, they demonstrated a non-uniform morphology and TB-staining pattern. These cells did generate significant amounts of cellular debris and so AO/PI was required to accurately determine the viability without interference by the debris (Chan et al. 2013). 10

11 Figure 6. Bright field and fluorescent images of TB and PI-stained heat-killed Jurkat cells at various theoretical viability percentages (0-100%). 11

12 Figure 7. Comparison of viability measurements between manual, Cellometer Auto T4 with TB, and automated PI analysis of heat-killed Jurkat cells. The y-axis shows experimentally-derived viability percentages, while the x-axis indicates the 0-100% theoretical viability mixtures. Phase contrast microscopy Figure 8 shows the bright field and phase contrast images of TB-stained Jurkat cells. At a concentration of 0.4% TB, Jurkat cells were observed and identified as a dim, diffuse dead cell in the bright field image but was too dim to be detected in the PC image (red arrow). At 0.1% TB, dark cellular morphology in the bright field image was also detected in the PC image. These results underscore the importance of choosing appropriate TB dye concentrations and detection methods. 12

13 Figure 8. Phase contrast and bright field images at 0.4% and 0.1% TB. Red arrows indicate dim, diffused cell in TB-stained image that is not also seen under phase contrast. Green arrows indicate dark, dead cell seen in TB image that was also seen under phase contrast. Results/Discussion The aim of this study was to ascertain the variations in viability measurements between the TB exclusion assay and fluorescence-based viability labels. We uncovered two main reasons for variations in cellular morphology that contribute to variations in viability analysis: the concentration of TB used and the manner by which the cells died. Morphological differences occurred in cells stained with TB at different viability ranges. This is not surprising as naturally-killed Jurkat cells undergo extreme shape changes during cell death. High viability samples showed cells with uniform shape characteristics, suggesting that cell membranes remained whole enough to retain the TB dye. Low viability samples showed large, dim, non-uniform cells, suggesting a loss of membrane integrity. 13

14 Furthermore, we noted that dim, diffuse shapes can vary greatly in brightness. This may complicate analysis when using a hemacytometer as TB staining may yield higher viability counts because of unintended exclusion of dead cells. The difference between the staining methods reached as high as 30%, confirming the findings of other studies (Mascotti et al. 2000). The overestimation of viability is also observed in primary murine bulk splenocytes. Between the AO/PI, PI, and TB methods there were significant differences (p 0.05), which further underlines the overestimation seen in TB manual counting, particularly when cultured or primary cell samples are at or below 70% viability. High concentrations of TB ( 0.4%) produced large, dim circles (Figure 9) that we termed balloons. These balloons can be seen under a light microscope and the Cellometer. The balloon cells are comprised of different sizes and brightness, with some more diffuse than others. Balloons formed within the first 30 seconds of staining at the high TB concentration. More balloons formed in the next 60 seconds, although at a slower rate. At low TB concentrations (0.025%), no balloon-like cells were seen, although this dye concentration may be too low to accurately stain all dead cells in the sample. The concentrations within each population are shown in Figure 10. At 33 hours, there were significantly more diffused balloon cells at 0.4 and 0.1% concentrations of TB (p 0.05). At 0.025% TB, there was a reversal in the number of counted balloons versus the number of counted dead cells (p 0.05), as well as a lower number of total counted cells, which we attribute to the low dye concentration. From these results, we conclude that at high TB concentrations, there is a morphological effect on the membranes of compromised dead cells, thereby producing the balloon-like cells. Lower concentrations of TB did not show the same morphological anomalies. The balloon structures are difficult to see and count, prompting an overestimation of viability. Other reasons for the differences between TB and fluorescence-based viability measurements may include: o Differences in the molecular sizes of the dyes which impacts how quickly the dyes enter the cells; o The fact that TB is a cytoplasmic dye and so when membranes degrade, the dye may be more easily lost, resulting in fewer dead cells being counted. As PI binds to nucleic acids, it is more readily retained regardless of membrane integrity which is why it s the dye of choice for late apoptotic and necrotic cell studies (Denecker et al. 2000; Yedjou et al. 2012). 14

15 Image cytometry allowed for the visualization of the morphological differences seen in high and low viability samples and in response to various concentrations of TB stain. Future studies will examine the impact of TB incubation time on cellular morphology, as well as a video analysis to show a time lapse of the morphological changes induced by TB Figure 9. Image cytometry analysis of morphological changes in TB concentration series. White arrows illustrate dead cells that have expanded to become dim, diffuse shapes. Black arrows illustrate dead cells that remain the dark and tight morphology. Figure 10a-d. Time course analysis (0-33 hours) of varying TB concentrations (0.4% a, 0.1% b, 0.025% c) quantifying the large, dim balloon-like Jurkat cells and the dark, tight Jurkat cells. Cell counts of balloon-like cells and dark, tight cells at 33 hours post incubation at the various concentrations of TB (d). 15

16 References 1. Al-Rubeai M, Welzenbach K, Lloyd DR, Emery AN (1997) A rapid method for evaluation of cell number and viability by flow cytometry. Cytotechnology 24: Altman SA, Randers L, Rao G (1993) Comparison of trypan blue dye exclusion and fluorometric assays for mammalian cell viability determinations. Biotechnol Prog 9: Black L, Berenbaum MC (1964) Factors affecting the dye exclusion test for cell viability. Exp Cell Res 35: Chan LL, Wilkinson AR, Paradis BD, Lai N (2012) Rapid image-based cytometry for comparison of fluorescent viability staining methods. J Fluoresc 22: Chan LLY, Laverty DJ, Smith T, Nejad P, Hei H, Gandhi R, Kuksin D, Qiu J (2013) Accurate measurement of peripheral blood mononuclear cell concentration using image cytometry to eliminate RBC-induced counting error. J Immunol Methods 388: Darzynkiewicz Z, Bruno S, Bino GD, Gorczyca W, Hotz MA, Lassota P, Traganos F (1992) Features of apoptotic cells measured by flow cytometry. Cytometry 13: Denecker G, Dooms H, Loo GV, Vercammen D, Grooten J, Fiers W, Declercq W, Vandenabeele P (2000) Phosphatidyl serine exposure during apoptosis precedes release of cytochrome c and decrease in mitochondrial transmembrane potential. FEBS Lett 465: Foglieni C, Meoni C, Davalli AM (2001) Fluorescent dyes for cell viability: an application on prefixed conditions. Histochem Cell Biol 115: Jones KH, Senft JA (1985) An improved method to determine cell viability by simultaneous staining with fluorescein diacetate-propidium iodide. J Histochem Cytochem 33: Ling E, Shirai K, Kanekatsu R, Kiguchi K (2003) Classification of larval circulating hemocytes of the silkworm Bombyx mori, by acridine orange and propidium iodide staining. Histochem Cell Biol 120: Louis KS, Siegel AC (2011) Cell viability analysis using trypan blue: manual and automated methods. In: Stoddart MJ (ed) Mammalian cell viability: methods and protocols, vol 740. Springer, Berlin, pp Mascotti K, McCullough J, Burger SR (2000) HPC viability measurement: trypan blue versus acridine orange and propidium iodide. Transfusion 40: Solomon M, Wofford J, Johnson C, Regan D, Creer MH (2010) Factors influencing cord blood viability assessment before cryopreservation. Transfusion 50: Tsaousis KT, Kopsachilis N, Tsinopoulos IT, Dimitrakos SA, Kruse FE, Welge-Luessen U (2012) Timedependent morphological alterations and viability of cultured human trabecular cells after exposure to trypan blue. Clin Exp Ophthalmol 41: Wallen CA, Higashikubo R, Dethlefsen LA (1980) Comparison of two flow cytometric assays for cellular RNA acridine orange and propidium iodide. Cytometry 3:

17 16. Yedjou CG, Saeed MA, Hossain MA, Dorsey W, Yu H, Tchounwou PB (2012) Basic apoptotic and necrotic cell death in human liver carcinoma (HepG(2)) cells induced by synthetic azamacrocycle. Environ Toxicol 1 7. doi: / tox

Comparing Fluorescence-Based Viability Detection Method using the Cellometer Vision

Comparing Fluorescence-Based Viability Detection Method using the Cellometer Vision Comparing Fluorescence-Based Viability Detection Method using the Cellometer Vision Nexcelom Bioscience LLC. 360 Merrimack Street, Building 9 Lawrence, MA 01843 T: 978.327.5340 F: 978.327.5341 E: info@nexcelom.com

More information

Cellometer Auto 2000 Cell Viability Counter Optimized Analysis of Primary Cells Cellometer Features of the Cellometer Auto 2000

Cellometer Auto 2000 Cell Viability Counter Optimized Analysis of Primary Cells Cellometer Features of the Cellometer Auto 2000 Cellometer 2000 Cell Counter Optimized Analysis of Primary Features of the Cellometer 2000 Dual Fluorescence and Bright Field Imaging: staining of both live and dead cells in heterogeneous samples The

More information

Cellometer K2. Cellometer K2 Image Cytometer Optimized Analysis of Primary Cells. Image Cytometers. for Cell Counting & Analysis.

Cellometer K2. Cellometer K2 Image Cytometer Optimized Analysis of Primary Cells. Image Cytometers. for Cell Counting & Analysis. Cellometer Optimized Analysis of Cells Features of the Cellometer Dual Fluorescence and Bright Field Imaging: staining of both live and dead cells in heterogeneous samples The Cellometer has drastically

More information

Cellometer. for Cell Counting & Analysis. Brewing Yeast Wine Yeast Platelets and Other Small Cells. Image Cytometer

Cellometer. for Cell Counting & Analysis. Brewing Yeast Wine Yeast Platelets and Other Small Cells. Image Cytometer Cellometer 2 Image Cytometer for Cell Counting & Analysis Brewing Yeast Wine Yeast and Other Small Cells Cellometer 2 Image Cytometer Optimized Analysis for Yeast and other Small Cells Features of the

More information

Cellometer Vision CBA

Cellometer Vision CBA Features of the Vision CBA Image Cytometry System All-in-One System Basic cell counting, primary cell viability, and cellbased assays. See for Yourself Why the Top Ten Pharmaceutical Companies Trust Cellometer

More information

Simply Counted. Importance of Accurate Cell Counting for Single Cell Sequencing Platforms Using the Cellometer K2 Fluorescent Viability Cell Counter

Simply Counted. Importance of Accurate Cell Counting for Single Cell Sequencing Platforms Using the Cellometer K2 Fluorescent Viability Cell Counter Cellometer Applications Importance of Accurate Cell Counting for Single Cell Sequencing Platforms Using the Cellometer K2 Fluorescent Viability Cell Counter Leo Li-Ying Chan 1, Belen Belete-Gilbert 1,

More information

Cellometer. Vision CBA. Image Cytometry System for 20µl Cell-Based Assays

Cellometer. Vision CBA. Image Cytometry System for 20µl Cell-Based Assays Cellometer Vision CBA Image Cytometry System for 2µl Cell-Based Assays Apoptosis Autophagy Cell Cycle Proliferation Transfection Viability and Others Features of the Vision CBA Image Cytometry System All-in-One

More information

Seeing is believing. Fast, accurate automated cell counting you can trust. Cellometer Auto T4. Cellometer Auto M10. Cellometer Auto X4

Seeing is believing. Fast, accurate automated cell counting you can trust. Cellometer Auto T4. Cellometer Auto M10. Cellometer Auto X4 Seeing is believing. Fast, accurate automated cell counting you can trust. Cellometer Auto T4 Cellometer Auto M10 Cellometer Auto X4 Cellometer Vision Simply Counted. Total cells: 810 Live cells circled

More information

Cellometer. Vision CBA. Image Cytometry System for 20µl Cell-Based Assays

Cellometer. Vision CBA. Image Cytometry System for 20µl Cell-Based Assays Cellometer Vision CBA Image Cytometry System for µl Cell-Based Apoptosis Cell Cycle and Others Features of the Vision CBA Image Cytometry System All-in-One System Basic cell counting, primary cell viability,

More information

Measuring Physiological and Metabolic Characteristics of Yeast Health for Beer Fermentation Samples using the Cellometer Vision

Measuring Physiological and Metabolic Characteristics of Yeast Health for Beer Fermentation Samples using the Cellometer Vision Measuring Physiological and Metabolic Characteristics of Yeast Health for Beer Fermentation Samples using the Cellometer Vision Nexcelom Bioscience LLC. 360 Merrimack Street, Building 9 Lawrence, MA 01843

More information

Multiplex Fluorescence Assays for Adherence Cells without Trypsinization

Multiplex Fluorescence Assays for Adherence Cells without Trypsinization Multiplex Fluorescence Assays for Adherence Cells without Trypsinization The combination of a bright field and three fluorescent channels allows the Celigo to perform many multiplexed assays. A gating

More information

ab Propidium Iodide Flow Cytometry Kit for Cell Cycle Analysis

ab Propidium Iodide Flow Cytometry Kit for Cell Cycle Analysis ab139418 Propidium Iodide Flow Cytometry Kit for Cell Cycle Analysis Instructions for Use To determine cell cycle status in tissue culture cell lines by measuring DNA content using a flow cytometer. This

More information

ab Propidium Iodide Flow Cytometry Kit for Cell Cycle Analysis

ab Propidium Iodide Flow Cytometry Kit for Cell Cycle Analysis ab139418 Propidium Iodide Flow Cytometry Kit for Cell Cycle Analysis Instructions for Use To determine cell cycle status in tissue culture cell lines by measuring DNA content using a flow cytometer. This

More information

Direct Cell Counting Assays for Immuno Therapy

Direct Cell Counting Assays for Immuno Therapy Direct Cell Counting Assays for Immuno Therapy Cytotoxicity assays play a central role in studying the function of immune effector cells such as cytolytic T lymphocytes (CTL) and natural killer (NK) cells.

More information

Cell Proliferation and Death

Cell Proliferation and Death Cell Proliferation and Death Derek Davies, Cancer Research UK http://www.london-research-institute.org.uk/technologies/120 Proliferation A cell Apoptosis Cell death Proliferation signals Senescence DNA

More information

Automated Method for Determination of Infectious Dose (TCID 50 ) using Celigo Imaging Cytometer

Automated Method for Determination of Infectious Dose (TCID 50 ) using Celigo Imaging Cytometer Automated Method for Determination of Infectious Dose (TCID 50 ) using Celigo Imaging Cytometer Nexcelom Bioscience LLC. 360 Merrimack Street, Building 9 Lawrence, MA 01843 T: 978.327.5340 F: 978.327.5341

More information

Defining viability in mammalian cell cultures

Defining viability in mammalian cell cultures Defining viability in mammalian cell cultures Susan M. Browne, Mohamed Al-Rubeai To cite this version: Susan M. Browne, Mohamed Al-Rubeai. Defining viability in mammalian cell cultures. Biotechnology Letters,

More information

MoxiCyte Viability Kit (#MXA055) User s Guide (Rev )

MoxiCyte Viability Kit (#MXA055) User s Guide (Rev ) MoxiCyte Viability Kit (#MXA055) User s Guide (Rev 20150406) FOR RESEARCH USE ONLY on the Moxi Flow or Zepi Flow Instruments Orflo Technologies, LLC Ketchum, ID Tech_support@orflo.com 1-855- 879-6694 Overview

More information

Evaluation of the CyFlow Cube 6 flow cytometer unit for the assessment of brewing yeast viability and vitality

Evaluation of the CyFlow Cube 6 flow cytometer unit for the assessment of brewing yeast viability and vitality Sysmex UK Ltd. Evaluation of the CyFlow Cube 6 flow cytometer unit for the assessment of brewing yeast viability and vitality 1 Evaluation of the Sysmex Cube 6 flow cytometer for the assessment of brewing

More information

Live and Dead Cell Assay

Live and Dead Cell Assay ab115347 Live and Dead Cell Assay Instructions for Use Differential fluorescent labeling of live and dead cells This product is for research use only and is not intended for diagnostic use. Last Updated

More information

Violet Chromatin Condensation/Dead Cell Apoptosis Kit with Vybrant DyeCycle Violet and SYTOX AADvanced for Flow Cytometry

Violet Chromatin Condensation/Dead Cell Apoptosis Kit with Vybrant DyeCycle Violet and SYTOX AADvanced for Flow Cytometry Violet Chromatin Condensation/Dead Cell poptosis Kit with Vybrant DyeCycle Violet and SYTOX Dvanced for Flow Cytometry Catalog no. 35135 Table 1. Contents and storage information. Material mount Concentration

More information

Autophagy/Cytotoxicity Dual Staining Kit

Autophagy/Cytotoxicity Dual Staining Kit Autophagy/Cytotoxicity Dual Staining Kit Catalog Number KA1299 96 assays Version: 05 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Background... 3 Principle of the Assay...

More information

Calcein AM Cell Viability Kit

Calcein AM Cell Viability Kit Instructions For Research Use Only. Not For Use In Diagnostic Procedures Calcein AM Cell Viability Kit Catalog# 4892-010-K 1000 Tests* * Calculated based on using 1 μm final concentration of Calcein AM;

More information

Performance of cell viability and cytotoxicity assays on the IN Cell Analyzer 3000

Performance of cell viability and cytotoxicity assays on the IN Cell Analyzer 3000 GE Healthcare Application Note 28-4070-51 AA IN Cell Analyzer 3000 Performance of cell viability and cytotoxicity assays on the IN Cell Analyzer 3000 Key words: cell-based assay viability cytotoxicity

More information

Hematopoietic Progenitor Cell Product Characterization

Hematopoietic Progenitor Cell Product Characterization Hematopoietic Progenitor Cell Product Characterization Carolyn A. Taylor, Ph.D. Professor of Medicine Director of BMT Program Cell Processing Laboratory Product Testing and Characterization Goals Required

More information

ab CFSE Fluorescent Cell Labeling Kit

ab CFSE Fluorescent Cell Labeling Kit ab113853 CFSE Fluorescent Cell Labeling Kit Instructions for Use For the durable fluorescent labeling of live cells for fluorescent microscopy and flow cytometry, population growth studies and within sample

More information

Propidium Iodide. Catalog Number: Structure: Molecular Formula: C 27H 34I 2N 4. Molecular Weight: CAS #

Propidium Iodide. Catalog Number: Structure: Molecular Formula: C 27H 34I 2N 4. Molecular Weight: CAS # Catalog Number: 195458 Propidium Iodide Structure: Molecular Formula: C 27H 34I 2N 4 Molecular Weight: 668.45 CAS # 25535-16-4 Physical Description: Dark red crystals Description: Reagent used for the

More information

NucleoCounter NC-3000

NucleoCounter NC-3000 Application note No. 3002. Rev. 1.5 NucleoCounter NC-3000 Cell cycle analysis of fixed cells Product description The NucleoCounter NC-3000 system enables the user to perform automated cell counting and

More information

Phagocytosis Assay Kit (IgG PE)

Phagocytosis Assay Kit (IgG PE) Phagocytosis Assay Kit (IgG PE) Item No. 600540 www.caymanchem.com Customer Service 800.364.9897 Technical Support 888.526.5351 1180 E. Ellsworth Rd Ann Arbor, MI USA TABLE OF CONTENTS GENERAL INFORMATION

More information

11/19/2013. Janine Zankl FACS Core Facility 13. November Cellular Parameters. Cellular Parameters. Monocytes. Granulocytes.

11/19/2013. Janine Zankl FACS Core Facility 13. November Cellular Parameters. Cellular Parameters. Monocytes. Granulocytes. DEPARTEMENT BIOZENTRUM Janine Zankl FACS Core Facility 13. November 2013 Cellular Parameters Granulocytes Monocytes Basophils Neutrophils Lymphocytes Eosinophils Cellular Parameters 1 What Is Flow Cytometry?

More information

Apoptosis detection. Apoptosis assays for the Attune Acoustic Focusing Cytometer. APOPTOSIS DETECTION Attune Acoustic Focusing Cytometer

Apoptosis detection. Apoptosis assays for the Attune Acoustic Focusing Cytometer. APOPTOSIS DETECTION Attune Acoustic Focusing Cytometer POPTOSIS ETECTION ttune coustic Focusing Cytometer poptosis detection poptosis assays for the ttune coustic Focusing Cytometer poptosis is a carefully regulated process of cell death that occurs as a normal

More information

ab CFSE Fluorescent Cell Labeling Kit

ab CFSE Fluorescent Cell Labeling Kit ab113853 CFSE Fluorescent Cell Labeling Kit Instructions for Use For the durable fluorescent labeling of live cells for fluorescent microscopy and flow cytometry, population growth studies and within sample

More information

Cell Staining. Introduction

Cell Staining. Introduction Cell Staining Introduction Visualization of a cell with fluorescent compounds provides a wide variety of information for the analysis of cell functions. Various activities and structures of a cell can

More information

Technical Bulletin. Multiple Methods for Detecting Apoptosis on the BD Accuri C6 Flow Cytometer. Introduction

Technical Bulletin. Multiple Methods for Detecting Apoptosis on the BD Accuri C6 Flow Cytometer. Introduction March 212 Multiple Methods for Detecting Apoptosis on the BD Accuri C6 Flow Cytometer Contents 1 Introduction 2 Annexin V 4 JC-1 5 Caspase-3 6 APO-BrdU and APO-Direct Introduction Apoptosis (programmed

More information

For detecting mitochondrial membrane potential changes in cells using our proprietary fluorescence probe.

For detecting mitochondrial membrane potential changes in cells using our proprietary fluorescence probe. ab112133 JC-10 Mitochondrial Membrane Potential Assay Kit Flow Cytometry Instructions for Use For detecting mitochondrial membrane potential changes in cells using our proprietary fluorescence probe. This

More information

Flow Cytometry Support Reagents

Flow Cytometry Support Reagents Excite and inspire Flow Cytometry Support Reagents Introduction Miltenyi Biotec is a leading supplier of flow cytometry products, offering one of the broadest ranges of antibodies, kits, assays, and support

More information

CFSE Cell Division Assay Kit

CFSE Cell Division Assay Kit CFSE Cell Division Assay Kit Item No. 10009853 www.caymanchem.com Customer Service 800.364.9897 Technical Support 888.526.5351 1180 E. Ellsworth Rd Ann Arbor, MI USA TABLE OF CONTENTS GENERAL INFORMATION

More information

Choose the right assay to fit your experimental needs:

Choose the right assay to fit your experimental needs: The analysis of cell viability, cytotoxicity, cell cycle state, cell proliferation, and cell death are critical to most cell-based studies. Use this guide to understand your options when you need to assay

More information

CFSE Cell Division Assay Kit

CFSE Cell Division Assay Kit CFSE Cell Division Assay Kit Catalog Number KA1302 100 assays Version: 02 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Background... 3 General Information...

More information

ab JC-10 Mitochondrial Membrane Potential Assay Kit Flow Cytometry

ab JC-10 Mitochondrial Membrane Potential Assay Kit Flow Cytometry ab112133 JC-10 Mitochondrial Membrane Potential Assay Kit Flow Instructions for Use For detecting mitochondrial membrane potential changes in cells using our proprietary fluorescence probe. This product

More information

Conquer cell counting

Conquer cell counting Conquer cell counting Cell Countess II Automated Cell Counters Fast Accurate Affordable Countess II Automated Cell Counters Advanced technology at an affordable price Accurate counts in as little as 10

More information

APO-DIRECT Kit. A Complete Kit for Measuring Apoptosis by Flow Cytometry. Cat. No. TNB-6611-KIT. Apoptotic Cells. Gate. S-phase. Apoptosis.

APO-DIRECT Kit. A Complete Kit for Measuring Apoptosis by Flow Cytometry. Cat. No. TNB-6611-KIT. Apoptotic Cells. Gate. S-phase. Apoptosis. APO-DIRECT Kit Cat. No. TNB-6611-KIT A Complete Kit for Measuring Apoptosis by Flow Cytometry Gate S-phase DNA Width Apoptosis G1 G2 DNA Area DNA Content Apoptotic Cells G1 S G2 DNA Content Apoptosis Technical

More information

INTRODUCTION TO FLOW CYTOMETRY

INTRODUCTION TO FLOW CYTOMETRY DEPARTEMENT BIOZENTRUM INTRODUCTION TO FLOW CYTOMETRY F ACS C ore F acility Janine Zankl FACS Core Facility 3. Dezember 2015, 4pm Cellular Parameters Granulocytes Monocytes Basophils Lymphocytes Neutrophils

More information

ab Cell Viability Assay Kit Fluorometric Dual Green/Red

ab Cell Viability Assay Kit Fluorometric Dual Green/Red ab112121 Cell Viability Assay Kit Fluorometric Dual Green/Red Instructions for Use For detecting cell viability in suspension and adherent cells by using dual proprietary green and red fluorescence probes.

More information

Use of Phase Contrast Imaging to Track Morphological Cellular Changes due to Apoptotic Activity

Use of Phase Contrast Imaging to Track Morphological Cellular Changes due to Apoptotic Activity A p p l i c a t i o n N o t e Use of Phase Contrast Imaging to Track Morphological Cellular Changes due to Apoptotic Activity Brad Larson and Peter Banks, Applications Department, BioTek Instruments, Inc.,

More information

In situ detection kit for programmed Cell Death MEBSTAIN Apoptosis Kit Direct

In situ detection kit for programmed Cell Death MEBSTAIN Apoptosis Kit Direct For research use only In situ detection kit for programmed Cell Death MEBSTAIN Apoptosis Kit Direct CODE No. 8445 In situ detection kit for Programmed Cell Death MEBSTAIN Apoptosis Kit Direct Cat. No.

More information

Celigo Assays.

Celigo Assays. Celigo Assays April 2017 www.nexcelom.com/celigo Nexcelom's team of Field Applications Scientists, R&D Specialists and Product Managers are in frequent contact with researchers in the field, developing

More information

Propidium Iodide Solution

Propidium Iodide Solution G-Biosciences 1-800-628-7730 1-314-991-6034 technical@gbiosciences.com A Geno Technology, Inc. (USA) brand name Propidium Iodide Solution (1mg/ml in deionized water) (Cat. # 786 1272, 786 1273) think proteins!

More information

Add Live, Dead and Total Dyes. Overnight Variable 30 min. 10 min 5 min. 30 min

Add Live, Dead and Total Dyes. Overnight Variable 30 min. 10 min 5 min. 30 min Cell Viability Analysis using Calcein AM, Propidium Iodide and Hoechst Application Description Celigo Application Plate Type Major Steps Cell viability analysis using Calcein AM, Propidium Iodide and Hoechst

More information

NM interference in the flow cytometric Annexin V/Propidium Iodide measurement

NM interference in the flow cytometric Annexin V/Propidium Iodide measurement Project: VIGO NM interference in the flow cytometric Annexin V/Propidium Iodide measurement AUTHORED BY: DATE: Cordula Hirsch 20.01.2014 REVIEWED BY: DATE: Harald Krug 10.04.2014 APPROVED BY: DATE: DOCUMENT

More information

Total Cytotoxicity & Apoptosis Detection Kit

Total Cytotoxicity & Apoptosis Detection Kit Total Cytotoxicity & Apoptosis Detection Kit catalog #971, trial size (125 tests) catalog #972, regular size (250 tests) RESEARCH USE ONLY 1. INTRODUCTION In a single tube, scientists can now differentiate

More information

ab Apoptosis/Necrosis Detection Kit (blue, red, green)

ab Apoptosis/Necrosis Detection Kit (blue, red, green) ab176750 Apoptosis/Necrosis Detection Kit (blue, red, green) Instructions for use: For detection of apoptosis and necrosis in adherent or suspension cells. This product is for research use only and is

More information

Titration of Fluorochrome-Conjugated Antibodies for Labeling Cell Surface Markers on Live Cells

Titration of Fluorochrome-Conjugated Antibodies for Labeling Cell Surface Markers on Live Cells Titration of Fluorochrome-Conjugated Antibodies for Labeling Cell Surface Markers on Live Cells Ruud Hulspas 1 UNIT 6.29 1 Cytonome/ST, Boston, Massachusetts ABSTRACT Nonspecific antibody binding is best

More information

LUNA-FL Automated Fluorescence Cell Counter

LUNA-FL Automated Fluorescence Cell Counter LUNA-FL Automated Fluorescence Cell Counter User Manual Document #: LBSM-0003 Version: VL1706-01 1 IMPORTANT LUNA-FL Automated Fluorescence Cell Counter is a Laboratory Electrical Instrument for Scientific

More information

Tali Viability Kit Dead Cell Green

Tali Viability Kit Dead Cell Green Tali Viability Kit Dead Cell Green *for use with Tali Assays: Green, Green + Red* Catalog no. A10787 Table 1 Contents and storage Material Amount Concentration Storage* Stability Tali Dead Cell Green (SYTOX

More information

RayBio Apoptosis Detection Kit, MitoCapture

RayBio Apoptosis Detection Kit, MitoCapture RayBio Apoptosis Detection Kit, MitoCapture User Manual Version 1.0 Mar 17, 2015 RayBio Apoptosis Detection Kit, MitroCapture Protocol (Cat#: 68MC-Apo-S100) RayBiotech, Inc. We Provide You With Excellent

More information

Multidrug Resistance Assay Kit (Calcein AM)

Multidrug Resistance Assay Kit (Calcein AM) Multidrug Resistance Assay Kit (Calcein AM) Item No. 600370 www.caymanchem.com Customer Service 800.364.9897 Technical Support 888.526.5351 1180 E. Ellsworth Rd Ann Arbor, MI USA TABLE OF CONTENTS GENERAL

More information

Advanced Calcein AM Cell Viability Kit

Advanced Calcein AM Cell Viability Kit Advanced Calcein AM Cell Viability Kit 25-250 tests, catalog#9154 FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES. 1. INTRODUCTION The Advanced Calcein AM Cell Viability kit developed by ImmunoChemistry

More information

Advanced Calcein AM Cell Viability Kit

Advanced Calcein AM Cell Viability Kit Advanced Calcein AM Cell Viability Kit 25-250 tests, catalog#9154 FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES. 1. INTRODUCTION The Advanced Calcein AM Cell Viability kit developed by ImmunoChemistry

More information

NucView TM 488 Caspase-3 Assay Kit for Live Cells

NucView TM 488 Caspase-3 Assay Kit for Live Cells NucView TM 488 Caspase-3 Assay Kit for Live Cells Catalog Number: 30029 (100-500 assays) Contact Information Address: Biotium, Inc. 3423 Investment Blvd. Suite 8 Hayward, CA 94545 USA Telephone: (510)

More information

Violet Annexin V/Dead Cell Apoptosis Kit with Pacific Blue Annexin V/SYTOX AADvanced for Flow Cytometry

Violet Annexin V/Dead Cell Apoptosis Kit with Pacific Blue Annexin V/SYTOX AADvanced for Flow Cytometry Violet nnexin V/Dead Cell poptosis Kit with Pacific Blue nnexin V/SYTOX Dvanced for Flow Cytometry Catalog no. 35136 Table 1. Contents and storage information. Material mount Storage Stability Pacific

More information

Cell Cycle Phase Determination Kit

Cell Cycle Phase Determination Kit Cell Cycle Phase Determination Kit Catalog Number KA1301 100 assays Version: 04 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Background... 3 General

More information

ab MDR Assay Kit - flow cytometry (Fluorometric - gold)

ab MDR Assay Kit - flow cytometry (Fluorometric - gold) ab204535 MDR Assay Kit - flow cytometry (Fluorometric - gold) Instructions for Use For functional detection and profiling of multidrug resistant phenotypes in live cells (both suspension and adherent).

More information

Institute of Science & Technology in Medicine Control of Substance Hazardous to Health COSHH assessment

Institute of Science & Technology in Medicine Control of Substance Hazardous to Health COSHH assessment Institute of Science & Technology in Medicine Control of Substance Hazardous to Health COSHH assessment Title of Experiment/Procedure: Cell counting and viability using FACS, counting beads and propidium

More information

The Cell Cycle, DNA Replication, and Mitosis

The Cell Cycle, DNA Replication, and Mitosis Chapter 19 The Cell Cycle, DNA Replication, and Mitosis 首頁 OUTLINE Overview of the Cell Cycle DNA Replication DNA Damage and Repair Nuclear and Cell Division Regulation of the Cell Cycle Growth Factors

More information

EarlyTox Cell Integrity Kit

EarlyTox Cell Integrity Kit EarlyTox Cell Integrity Kit The EarlyTox Cell Integrity Kit from Molecular Devices is an optimized set of reagents that simplifies the measurement of live and dead cells in a single well. The assay uses

More information

Challenges for Product Evaluation by Flow Cytometry for Cellular Therapy Product Processing Laboratories

Challenges for Product Evaluation by Flow Cytometry for Cellular Therapy Product Processing Laboratories Challenges for Product Evaluation by Flow Cytometry for Cellular Therapy Product Processing Laboratories C A R O LY N A. K E E V E R - TAY L O R, P H D M E D I C A L C O L L E G E O F W I S C O N S I N

More information

Transforming Cell Science

Transforming Cell Science Transforming Cell Science T E C H N OT E 1 1 / 2 0 1 1 Comparison of Moxi Z Cell Counter with an Imaging Cell Counter and a High-End Coulter System Introduction Cell counts are routinely performed in life

More information

The Nuclear Area Factor (NAF): a measure for cell apoptosis using microscopy and image analysis

The Nuclear Area Factor (NAF): a measure for cell apoptosis using microscopy and image analysis The Nuclear Area Factor (NAF): a measure for cell apoptosis using microscopy and image analysis Mark A. DeCoster Department of Biomedical Engineering and Institute for Micromanufacturing, Louisiana Tech

More information

Orflo Application Note 8 /2017 Rapid Apoptosis Monitoring Using Annexin V and Orflo s Moxi GO II Next Generation Flow Cytometer

Orflo Application Note 8 /2017 Rapid Apoptosis Monitoring Using Annexin V and Orflo s Moxi GO II Next Generation Flow Cytometer Introduction/Background Orflo Application Note 8 /2017 Cellular apoptosis is a sophisticated mechanism employed by cells to carefully control death in response to cell injury. Commonly referred to as programmed

More information

APO-BrdU Kit. A Complete TUNEL Kit for Measuring Apoptosis by Dual Color Flow Cytometry. Cat. No. TNB-6671-KIT. Apoptotic Cells. Gate.

APO-BrdU Kit. A Complete TUNEL Kit for Measuring Apoptosis by Dual Color Flow Cytometry. Cat. No. TNB-6671-KIT. Apoptotic Cells. Gate. APO-BrdU Kit Cat. No. TNB-6671-KIT A Complete TUNEL Kit for Measuring Apoptosis by Dual Color Flow Cytometry DNA Width Gate S-phase Apoptosis G1 G2 DNA Area DNA Content Apoptotic Cells S G2 DNA Content

More information

Amnis ImageStream : Technical Reports & Applications

Amnis ImageStream : Technical Reports & Applications Amnis ImageStream : Technical Reports & Applications ImageStream : Flow Cytometry and Microscopy in a Single Platform The ImageStream achieves true multispectral Imaging in Flow by combining microscopy

More information

ab Apoptosis/Necrosis Detection Kit (blue, green, red)

ab Apoptosis/Necrosis Detection Kit (blue, green, red) ab176749 Apoptosis/Necrosis Detection Kit (blue, green, red) Instructions for use: For detection of apoptosis and necrosis in adherent or suspension cells. This product is for research use only and is

More information

TITLE: LIVE/DEAD VIABILITY FOR CLINICAL SAMPLES

TITLE: LIVE/DEAD VIABILITY FOR CLINICAL SAMPLES Paul K. Wallace, Ph.D. Director Roswell Park Cancer Institute Elm & Carlton Streets Voice:(716) 845-8471 Buffalo, NY 14263 Fax:(716) 845-8806 FILE NAME FL-SRP-2090.00 Live/Dead Viability for Clinical Samples

More information

Murine in vivo CD8 + T Cell Killing Assay Myoungjoo V. Kim 1*, Weiming Ouyang 2, Will Liao 3, Michael Q. Zhang 4 and Ming O. Li 5

Murine in vivo CD8 + T Cell Killing Assay Myoungjoo V. Kim 1*, Weiming Ouyang 2, Will Liao 3, Michael Q. Zhang 4 and Ming O. Li 5 Murine in vivo CD8 + T Cell Killing Assay Myoungjoo V. Kim 1*, Weiming Ouyang 2, Will Liao 3, Michael Q. Zhang 4 and Ming O. Li 5 1 Department of Immunobiology, Yale University School of Medicine, New

More information

INDIAN INSTITUTE OF TECHNOLOGY ROORKEE NPTEL NPTEL ONLINE CERTIFICATION COURSE. Biomedical Nanotechnology

INDIAN INSTITUTE OF TECHNOLOGY ROORKEE NPTEL NPTEL ONLINE CERTIFICATION COURSE. Biomedical Nanotechnology INDIAN INSTITUTE OF TECHNOLOGY ROORKEE NPTEL NPTEL ONLINE CERTIFICATION COURSE Biomedical Nanotechnology Lec-19 In Vitro Methods of Study Antibacterial and Anticancer Properties of Nanomaterials Dr. P.

More information

Supplementary Materials and Methods

Supplementary Materials and Methods Supplementary Materials and Methods Reagents Supplementary Material (ESI) for Lab on a Chip RPMI medium, FBS, HEPES buffer solution, sodium pyruvate, penicillin, and streptomycin were obtained from Biological

More information

Measuring Wound Healing and Cell Migration using Celigo Imaging Cytometer

Measuring Wound Healing and Cell Migration using Celigo Imaging Cytometer Measuring Wound Healing and Cell Migration using Celigo Imaging Cytometer Nexcelom Bioscience LLC. 360 Merrimack Street, Building 9 Lawrence, MA 01843 T: 978.327.5340 F: 978.327.5341 E: info@nexcelom.com

More information

Nucleic Acid Staining. Fluorophores & Applica6ons

Nucleic Acid Staining. Fluorophores & Applica6ons Nucleic Acid Staining Fluorophores & Applica6ons Types of Nucleic Acid Stains Intercala)ng dyes- - ethidium bromide and propidium iodide Minor- groove binders- - DAPI and the Hoechst dyes Miscellaneous-

More information

Living and Dead Cells Staining: -Cellstain- Double Staining Kit

Living and Dead Cells Staining: -Cellstain- Double Staining Kit Introduction -Cellstain - Double Staining Kit combines Calcein-AM (used for fluorescent staining the living cells) and Propidium Iodide (used for a fluorescent staining of the dead cells) for simultaneous

More information

JC 1 Mitochondrial Membrane Potential Assay

JC 1 Mitochondrial Membrane Potential Assay G-Biosciences 1-800-628-7730 1-314-991-6034 technical@gbiosciences.com A Geno Technology, Inc. (USA) brand name JC 1 Mitochondrial Membrane Potential Assay (Cat. # 786 1321, 786 1322) think proteins! think

More information

Detection of bacterial aggregation by flow cytometry

Detection of bacterial aggregation by flow cytometry Detection of bacterial aggregation by flow cytometry Jack Coleman 1 and Melis McHenry 2 1 Enzo Life Sciences, Farmingdale NY, USA, 2 Beckman Coulter Inc., Miami FL, USA PROTEOSTAT Aggresome Detection Kit

More information

Technical Bulletin. Caspase 10 Fluorescein (FLICA) Assay. Catalog Number CS0310 Storage Temperature 2-8 C

Technical Bulletin. Caspase 10 Fluorescein (FLICA) Assay. Catalog Number CS0310 Storage Temperature 2-8 C Caspase 10 Fluorescein (FLICA) Assay Catalog Number CS0310 Storage Temperature 2-8 C Technical Bulletin Product Description Caspases are detected by immunoprecipitation, immunoblotting with caspase specific

More information

Everything counts. But nothing counts like Countess II automated cell counters. Fast Accurate Affordable

Everything counts. But nothing counts like Countess II automated cell counters. Fast Accurate Affordable Everything counts But nothing counts like Countess II automated cell counters Fast Accurate Affordable Countess II automated cell counters Advanced technology at an affordable price Precise, accurate counts

More information

Everything counts. But nothing counts like the

Everything counts. But nothing counts like the Everything counts But nothing counts like the Countess II FL Automated Cell Counter The new Countess II FL Automated Cell Counter is a benchtop cell assay platform equipped with state-of-the-art optics

More information

Assessment of Cell Viability

Assessment of Cell Viability Assessment of Cell Viability Simon Johnson, 1 Vy Nguyen, 1 and David Coder 2 UNIT 9.2 1 University of Washington School of Medicine, Seattle, Washington 2 Baybioscience, Irvine, California ABSTRACT Cell

More information

Fundamentals and Applications of Biofilms Analysis, Structure and Physiology of Bacterial Biofilms Ching-Tsan Huang ( 黃慶璨 ) Office: Agronomy

Fundamentals and Applications of Biofilms Analysis, Structure and Physiology of Bacterial Biofilms Ching-Tsan Huang ( 黃慶璨 ) Office: Agronomy 1 Fundamentals and Applications of Biofilms Analysis, Structure and Physiology of Bacterial Biofilms Ching-Tsan Huang ( 黃慶璨 ) Office: Agronomy Building, Room 111 Tel: (02) 33664454 E-mail: cthuang@ntu.edu.tw

More information

Mitochondrial Membrane Potential Detection Kit

Mitochondrial Membrane Potential Detection Kit Mitochondrial Membrane Potential Detection Kit Instruction Manual Catalog #280002 Mitochondrial Membrane Potential Detection Kit Revision B Research Use Only. Not for Use in Diagnostic Procedures. 280002-12

More information

Autophagy/Cytotoxicity Dual Staining Kit

Autophagy/Cytotoxicity Dual Staining Kit Autophagy/Cytotoxicity Dual Staining Kit Catalog Number KA1299 96 assays Version: 03 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Background... 3 Principle of the Assay...

More information

A Rapid Single Cell Sorting Verification Method Using Plate-Based Image Cytometry

A Rapid Single Cell Sorting Verification Method Using Plate-Based Image Cytometry A Rapid Single Cell Sorting Verification Method Using Plate-Based Image Cytometry Eric S. Zigon, 1 Suzanne M. Purseglove, 2 Vasilis Toxavidis, 1 * William Rice, 2 John Tigges, 1 * Leo Li-Ying Chan 2 *

More information

ab TMRE Mitochondrial Membrane Potential Assay Kit

ab TMRE Mitochondrial Membrane Potential Assay Kit ab113852 TMRE Mitochondrial Membrane Potential Assay Kit Instructions for Use For the measurement of mitochondrial membrane potential by flow cytometry, fluorescence plate reader and fluorescence microscopy

More information

Autologous Infusions Alisha Mussetter Clinical Research Coordinator II Wednesday February 17 th, 2016

Autologous Infusions Alisha Mussetter Clinical Research Coordinator II Wednesday February 17 th, 2016 Autologous Infusions Alisha Mussetter Clinical Research Coordinator II Wednesday February 17 th, 2016 How to participate in a live poll via the mobile app Ensure your phone is connected to WiFi or cellular

More information

DETECTION OF APOPTOSIS BY 7-AMINO-ACTINOMYCIN D STAINING

DETECTION OF APOPTOSIS BY 7-AMINO-ACTINOMYCIN D STAINING MATERIALS: DETECTION OF APOPTOSIS BY 7-AMINO-ACTINOMYCIN D STAINING 1. 1 X PBS (PBSAz, 1 X PBS, e.g., Irvine Scientific, CA, containing 2% newborn calf serum and 0.1% sodium azide) 2. 7-Amino-actinomycin

More information

EZ Vision DNA Dye as Loading Buffer

EZ Vision DNA Dye as Loading Buffer EZ Vision DNA Dye as Loading Buffer Code Description Size N472-1ML N472-KIT N650-1ML N650-KIT N313-1ML N313-KIT N473-2PK N473-3PK EZ Vision One DNA Dye as Loading Buffer, 6X EZ Vision Two, DNA Dye as Loading

More information

Total Cytotoxicity & Apoptosis Detection Kit A direct assay to accurately quantify cytotoxicity including cells in early apoptosis.

Total Cytotoxicity & Apoptosis Detection Kit A direct assay to accurately quantify cytotoxicity including cells in early apoptosis. Product Ordering Information: Total Cytotoxicity Kit 125 tests catalog# 971 250 tests catalog# 972 Total Cytotoxicity & Apoptosis Detection Kit A direct assay to accurately quantify cytotoxicity including

More information