Abstract. Hematopathology / Counting in Thrombocytopenic Specimens

Size: px
Start display at page:

Download "Abstract. Hematopathology / Counting in Thrombocytopenic Specimens"

Transcription

1 Hematopathology / Counting in Thrombocytopenic Specimens The Accuracy of Platelet Counting in Thrombocytopenic Blood Samples Distributed by the UK National External Quality Assessment Scheme for General Haematology Barbara J. De la Salle, FIBMS, 1 Paul N. McTaggart, FIBMS, 1 Carol Briggs, FIBMS, 2 Paul Harrison, FRCPath, 3 Caroline J Doré, 4 Ian Longair, 2 Samuel J. Machin, FRCPath, 2 and Keith Hyde, FRCPath 1 Key Words: Proficiency testing; Thrombocytopenia; Platelet count; Laboratory hematology; Transfusion DOI: /AJCP86JMBFUCFCXA Abstract A knowledge of the limitations of automated platelet counting is essential for the effective care of thrombocytopenic patients and management of platelet stocks for transfusion. For this study, 29 external quality assessment specimen pools with platelet counts between 5 and /L were distributed to more than 1,100 users of 23 different hematology analyzer models. The same specimen pools were analyzed by the international reference method (IRM) for platelet counting at 3 reference centers. The IRM values were on average lower than the all-methods median values returned by the automated analyzers. The majority (~67%) of the automated analyzer results overestimated the platelet count compared with the IRM, with significant differences in 16.5% of cases. Performance differed between analyzer models. The observed differences may depend in part on the nature of the survey material and analyzer technology, but the findings have implications for the interpretation of platelet counts at levels of clinical decision making. Accurate and reproducible platelet counts are essential for the management of thrombocytopenic patients at risk of bleeding, such as patients undergoing cytotoxic therapy for hematologic malignancy. Current UK guidelines recommend a threshold of /L for prophylactic platelet transfusion and suggest that this threshold might be reduced further, to /L, for patients without risk factors. 1 This reduction would conserve valuable stocks of platelets for transfusion and reduce the exposure of patients to the risks associated with transfusion of blood components. The accuracy of platelet counts produced by routine automated hematology analyzers has been questioned, 2-4 and the limitations of platelet counting at these extremely low levels should be understood by people responsible for the care of patients. The methods used for automated platelet counting are impedance, optical scatter, optical fluorescence, and immunologic flow cytometry. The traditional gold standard method was manual phase contrast microscopy, 5,6 although this method is time-consuming and imprecise at low counts. 7 The introduction of the international reference method (IRM) for platelet counting by flow cytometry 7-9 has improved the precision and accuracy of platelet counting at thrombocytopenic levels and offers a suitable comparator for routine platelet counting methods. The UK National External Quality Assessment Scheme for General Haematology [UK NEQAS (H)] is uniquely able to undertake major state-of-the-art comparisons of equipment performance. Unlike many other external quality assessment (EQA) or proficiency testing providers, UK NEQAS (H) distributes the same in-house prepared survey material to all instruments for the performance assessment of full blood count (FBC) parameters, including the platelet Am J Clin Pathol 2012;137: DOI: /AJCP86JMBFUCFCXA 65

2 De la Salle et al / Counting in Thrombocytopenic Specimens count. UK NEQAS (H) has more than 2,000 analyzers from all of the major instrument manufacturers registered in the FBC scheme. The results of 29 EQA specimen pools with all-methods median (AMM) platelet counts of between 5 and /L, distributed since 2006, have been compared with the median IRM platelet count determined at 3 independent sites to assess the performance of automated hematology analyzers at these very low levels for clinical decision making. Materials and Methods Survey Material UK NEQAS (H) FBC survey material was prepared from leukocyte-depleted human whole blood, anticoagulated with CPD-A1, obtained from healthy blood donors who had consented to the use of their blood for EQA purposes, and supplied by the UK National Health Service Blood and Transplant Service. For each pool, 5 to 6 individual donations of ABO group compatible, leukocyte-depleted whole blood, less than 72 hours old, were used. ABO group compatible whole blood and buffy coat residues were added to each p ool to raise the platelet count to the desired level. During the preparation of each specimen pool, the platelet count was assessed using a Sysmex K-4500 hematology analyzer (Sysmex, Kobe, Japan). Specimen pools were partially fixed using glutaraldehyde and formaldehyde, and the antibiotics penicillin and gentamicin were added. 10 Specimen pools were dispensed and bottled using a bespoke Hook and Tucker Zenyx Beeline dispensing machine (Hook and Tucker Zenyx. Croydon, England). Specimen homogeneity was maintained during bottling using a bespoke mixing system. Homogeneity and Stability of Specimens Survey material homogeneity and stability was determined according to UK NEQAS (H) standard protocols for UK NEQAS (H) FBC survey material. For the study, 10 specimens selected from each pool were analyzed in triplicate for FBC parameters using a Sysmex K-4500 analyzer immediately after bottling, on survey distribution day, and on survey closing day. Within- and between-batch analysis was used to confirm the stability and the homogeneity of the batch for WBC count, hemoglobin concentration, and platelet count. Specimen Distribution This study includes data from 29 specimen pools with AMM values between 5 and /L, distributed as part of the UK NEQAS (H) FBC surveys 0603FB to 0904FB. Specimens were distributed to all registered participants in the UK NEQAS (H) FBC scheme by first class post within the United Kingdom and by door-to-door courier delivery outside the United Kingdom. Transit time varied from 1 to 4 days on average, and the closing date for return of results was 7 days from the distribution date. Data from a further 6 thrombocytopenic specimen pools distributed during the same period were excluded from the study because the data failed the statistical criterion for inclusion, which was that less than 50% of the instrument groups should show a statistically significant difference (P <.01) between the submethod (ie, instrument group) median platelet count for the specimen pool and the IRM. Instrument Grouping Within the UK NEQAS (H) FBC EQA scheme, participants are grouped into 15 instrument types for analysis according to analyzer technology. Each grouping may contain more than 1 analyzer model. For the purposes of this study, results were analyzed by individual analyzer model. The target minimum number of analyzers in each group was 20; data from 2 groups with 18 instruments were included for 2 specimen pools. Several analyzers have more than 1 platelet counting method available: optical, impedance, and, for 1 manufacturer, immunologic. For most analyzers, platelet counts by any of these modes may be reported to UK NEQAS (H). However for Sysmex XE-2100 instruments, previous UK NEQAS (H) FBC survey data has indicated a wider variation for optical compared with impedance platelet counts. Expert advice from the UK NEQAS (H) Steering Committee s General Scientific Advisory Group has suggested that this variation results from platelet activation during the preparation of survey material, making performance assessment difficult. On the advice of the scheme s scientific advisors, and in agreement with the instrument manufacturer, UK NEQAS (H) specifically requests that only the impedance count be reported from Sysmex XE-2100 instruments. Target Platelet Count and Acceptable Range The target platelet count on each specimen was assayed by the IRM 9 at 3 reference centers during the time that the survey was open. Each center assayed 2 specimens from each survey material pool in triplicate, giving a total of 6 assay values for each pool from each site. The 3 reference centers were UK NEQAS (H), University College Hospital London (UCLH, London, England), and the Oxford Haemophilia Centre (Oxford, England), equipped with the FACSCalibur [Becton Dickinson, Oxford, England; for UK NEQAS (H) and Oxford] or the Epics XL (Beckman Coulter, Miami, FL; for UCLH). The comparability of the methods at the 3 centers was validated at the start of the study by counting 10 fresh, whole blood, EDTA-anticoagulated specimens in triplicate at each site on the same day Table Am J Clin Pathol 2012;137: DOI: /AJCP86JMBFUCFCXA

3 Hematopathology / Original Article The acceptable range to ensure that platelet counts are clinically reliable in EQA may be set at 10% to 15% of the target or assigned value, 11 which would give an approximate range of ± /L in absolute terms for the accuracy of a target platelet count of between 35 and /L and an approximate absolute range of ± 3 for a target platelet count of 20 to /L. At target values less than /L, a percentage range is of less value because the range becomes so narrow in absolute terms (< /L) as to be meaningless in practice. For this reason, an acceptable EQA performance range in absolute terms was calculated for each specimen pool as the IRM ± /L for specimen pools with IRM platelet counts of /L or more and the IRM ± 3 for pools with IRM platelet counts of /L or less. The mean IRM counts from each reference center for the 29 survey material pools are shown in Figure 1 : 5 mean counts (of 80) fell outside the acceptable EQA performance Mean Platelet Count ( 10 9 /L) FB2 0605FB1 0606FB1 0607FB1 0608FB1 0609FB1 0611FB1 0701FB2 0704FB1 0706FB1 0707FB2 0708FB1 0709FB1 0711FB1 0712FB1 0801FB2 Table 1 Comparison of the IRM for Platelet Counting at Three Reference Sites * Specimen Pool Mean IRM Platelet Count ( 10 9 /L) Specimen Pool Oxford UCLH UK NEQAS (H) Clotted IRM, international reference method. * The Oxford Haemophilia Centre, Oxford, England; UCLH (University College Hospital London, London, England); and the UK NEQAS (H) (UK National External Quality Assessment Scheme for General Haematology). For the study, 10 fresh, whole blood specimen pools of varying platelet counts were tested in triplicate at each center on the same day. UK NEQAS (H) Oxford UCLH 0802FB1 0802FB2 0803FB1 0803FB2 0804FB2 0805FB1 0807FB1 0807FB2 0808FB1 0809FB1 0811FB1 0812FB1 0904FB2 Figure 1 Comparison of the mean international reference method platelet counts from the 3 reference centers: UK NEQAS (H) (UK National External Quality Assessment Scheme for General Haematology, Watford, England); the Oxford Haemophilia Centre, Oxford, England; and UCLH (University College Hospital London, London, England) for the 29 specimen pools distributed in the study. Counts were provided by all 3 centers for 22 of 29 specimen pools and by 2 of 3 centers for the remainder. The counts shown represent the mean of 6 counts on each specimen pool for each center. Am J Clin Pathol 2012;137: DOI: /AJCP86JMBFUCFCXA 67

4 De la Salle et al / Counting in Thrombocytopenic Specimens range for the individual specimen pools. Of a total of 87 possible mean counts, 7 were not submitted owing to staffing shortages or reagent supply problems at the individual reference centers. Statistical Analysis Statistical analysis was performed using Stata 11 statistical analysis software (StataCorp, College Station, TX). Differences were evaluated by using the independent Student t test or Mann-Whitney U test, as appropriate. Median, estimated SD, 12 and coefficient of variation (CV) values were produced for all methods data and individual instrument group data. Cumulative Relative Frequency Distribution Curves Cumulative relative frequency distribution curves were drawn for each analyzer model and each specimen pool. The proportion of these curves that fell within the acceptable EQA performance range for each specimen pool was determined as an indicator of the accuracy of the results for each analyzer model, compared with the IRM, since the greater the proportion of the curve within the acceptable range, the closer the distribution of the results around the target IRM value. For the purpose of this study, instruments were assessed by the percentage of the curve that fell within the acceptable EQA performance range for each specimen. Reproducibility Study Usin g Individual Instruments A follow-up reproducibility study was undertaken with individual Abbott CELL-DYN 4000 (optical and impedance methods; Abbott Diagnostics, Santa Clara, CA), Siemens ADVIA 120 (Siemens Diagnostics, Tarrytown, NY), and Beckman Coulter LH 750 analyzers, using 4 specimen pools (designated A, B, C, and D) with platelet counts of approximately 5 to /L (pool A), 10 to /L (pool B), 20 to /L (pool C), and more than /L (pool D). Three specimens from each pool were tested 10 times on 3 occasions on the same day, with a 3-hour interval between each of the testing episodes. This approach provided 3 lots of 30 analyses from each of the 4 pools for overall reproducibility analysis for each analyzer. This exercise was repeated with a different set of 4 specimen pools (pools W, X, Y, and Z of approximately the same levels as pools A-D) using a Sysmex XE-2100 analyzer (impedance method) at a later date. The delay was due to a technical problem that made it impossible to complete the exercise on the Sysmex XE-2100 at the same time as the other 3 analyzers. Table 2 Make and Number of Each Analyzer Model and the Number of Specimen Pools Analyzed by Each * No. of Instruments Manufacturer/Model Minimum Maximum No. of Specimen Pools Analyzed (N = 611) Beckman Coulter (Miami, FL) Ac*T 5 Diff Ac*T Diff Gen-S HmX LH LH LH Abbott (Santa Clara, CA) CELL-DYN CELL-DYN CELL-DYN CELL-DYN Sapphire Horiba ABX (Montpellier, France) DX Pentra Pentra Pentra Siemens (Tarrytown, NY) ADVIA ADVIA Sysmex (Kobe, Japan) KX poch-100i SF XE XT-1800i XT-2000i * The number of analyzers in each group varied during the course of the project; the minimum and maximum numbers are given. 68 Am J Clin Pathol 2012;137: DOI: /AJCP86JMBFUCFCXA

5 Hematopathology / Original Article Results Analyzers The results of 23 analyzer models from 5 manufacturers were compared in this study, as listed in Table 2. Data were included for an analyzer model only if 20 (18 in 2 cases) or more instruments returned results for any specimen pool. Of the analyzer groups, 22 analyzed between 23 and 29 specimen pools each. The predominant instruments were the Beckman- Coulter LH 750, Siemens ADVIA 120 and 2120, and Sysmex XE One analyzer type, the Beckman Coulter LH 780, analyzed just 4 specimen pools, since there were insufficient numbers of these analyzers registered in the FBC scheme in the early distributions of the study. With the exception of the LH 780 group, all instruments analyzed specimen pools across the entire range of specimen values distributed. Platelet Counts From each survey material pool, 1,194 to 1,577 individual results were returned across the 23 instrument groups, giving a total of 611 median platelet counts. The AMM platelet count for each specimen pool ranged from 6 to /L. The median IRM count ranged from 5 to /L and was on average /L lower than the AMM result for each specimen pool. The median IRM counts and the acceptable EQA performance range for each specimen pool are shown in Table 3. The CV increased as the platelet count decreased Table 4, ranging from an all-methods value of about 32% at platelet counts of 5 to /L to about 10% at counts of 36 to /L. This same relationship was seen for all analyzer models, with CV values 3 to 4 times greater at counts of 5 to /L (CV range, 15%-43%) compared with counts of 36 to /L (CV range, 5%-11%). Bias Comparison of the 611 median platelet counts with the IRM values showed that 70 (11.5%) of the median counts were within ± /L of the IRM value; 304 (49.8%) showed a positive bias of more than /L higher than the IRM, with 101 (16.5%) showing a statistically significant positive bias (P <.01). 136 median counts were more than /L lower than the IRM, with just 2 (0.3%) showing a statistically significant negative bias (P <.01). Comparison of the median platelet counts with the IRM by instrument type showed that no instrument returned the same value as the IRM for more than 8 (28%) of the 29 specimen pools. Four Table 3 AMM and IRM Median Platelet Counts for Specimen Pools 0603FB to 0904FB and the Acceptable EQA Performance Range * Acceptable EQA Range ( 10 9 /L) Specimen Pool No. of Results AMM ( 10 9 /L) IRM Median ( 10 9 /L) Minimum Maximum 0603FB2 1, FB1 1, FB1 1, FB1 1, FB1 1, FB1 1, FB1 1, FB2 1, FB1 1, FB1 1, FB2 1, FB1 1, FB1 1, FB1 1, FB1 1, FB2 1, FB1 1, FB2 1, FB1 1, FB2 1, FB2 1, FB1 1, FB1 1, FB2 1, FB1 1, FB1 1, FB1 1, FB1 1, FB2 1, AMM, all-methods median; EQA, external quality assessment; IRM, international reference method. * Derived from the IRM ± /L where the IRM median was /L or less and the IRM ± /L where the IRM median was /L or more. Am J Clin Pathol 2012;137: DOI: /AJCP86JMBFUCFCXA 69

6 De la Salle et al / Counting in Thrombocytopenic Specimens instrument models, all from the same manufacturer, returned 63 of the 101 results showing a statistically significant positive bias. Six instrument models showed no significant bias in any of the specimen pools. In general, small bench-top or point-of-care instruments did not perform less well than large, top of the range analyzers Table 5. Accuracy Cumulative relative frequency distribution curves were constructed for each analyzer and specimen pool, and an example of these is shown in Figure 2 for the most common analyzer model from each manufacturer in the study. These curves were used to compare the results of each analyzer group with the IRM value and the acceptable EQA performance range. The cumulative relative frequency curves for each instrument model were analyzed to determine the proportion of each curve that fell within the acceptable range for each specimen pool. Figure 3 shows the mean percentage of the cumulative relative frequency distribution curves, Table 4 Mean CV for Each Instrument Group for Specimen Pools With IRM Values in Various Ranges IRM Median Platelet Count ( 10 9 /L) All methods CV (%) Mean CV (%) by instrument Beckman Coulter (Miami, FL) Ac*T 5 Diff Ac*T Diff Gen-S HmX LH LH LH Abbott (Santa Clara, CA) CELL-DYN CELL-DYN CELL-DYN CELL-DYN Sapphire Horiba ABX (Montpellier, France) DX Pentra Pentra Pentra Siemens (Tarrytown, NY) ADVIA ADVIA Sysmex (Kobe, Japan) KX poch-100i SF XE XT-1800i XT-2000i CV range (%) CV, coefficient of variation; IRM, international reference method. ie, the proportion of counts that fell within the acceptable EQA performance range for each instrument type. This value ranged from 9% to 80%. Of the predominant instrument groups, the Sysmex XE-2100 (impedance method) and the Beckman Coulter LH 750 achieved more than 70% of counts on average within the acceptable EQA performance range, and the Siemens ADVIA 120/2120 analyzers were within the 50% to 60% range seen with the majority of analyzer models. The 4 analyzer models that showed the greatest number of median counts with a statistically significant bias from the IRM (Abbott CELL-DYN 3200, 3700, 4000, and Sapphire) showed the lowest proportion of counts within the acceptable EQA performance range. Reproducibility Study Using Individual Instruments The results of the overall reproducibility study on individual analyzers Table 6 showed CV values up to 79% lower for 3 of the 4 analyzer models compared with the method group CV values (Table 4). These differences were Table 5 Statistical Significance of the Bias Observed in the Platelet Counts Compared With the International Reference Method Count by Instrument Type * No. With Significant Bias (P <.01) No. of Specimen Pools Positive Negative Beckman Coulter (Miami, FL) Ac*T 5DIFF Ac*T Diff Gen-S HmX LH LH LH Abbott (Santa Clara, CA) CELL-DYN CELL-DYN Cell-DYN CELL-DYN Sapphire Horiba ABX (Montpellier, France) DX Pentra Pentra Pentra Siemens (Tarrytown, NY) ADVIA ADVIA Sysmex (Kobe, Japan) KX poch-100i SF XE XT-1800i XT-2000i Total * In general, small bench-top analyzers did not perform less well than large, top of the range analyzers. 70 Am J Clin Pathol 2012;137: DOI: /AJCP86JMBFUCFCXA

7 Hematopathology / Original Article Cumulative Frequency % 80% 70% 60% 50% 40% 30% 20% 10% 0% Platelet Count ( 10 9 /L) ABX Pentra 120 ADVIA 120 CELL-DYN 4000 Coulter LH 750 Sysmex XE Figure 2 Cumulative relative frequency distribution curves for each analyzer type and specimen pool, as shown for the most common analyzer models from each manufacturer for specimen pool 0712FB1, were used to determine the proportion of counts for each analyzer group that fell within the acceptable external quality assessment (EQA) performance range for the specimen pool. The greater the proportion of the curve that fell within this range, the closer the observed results were to the international reference method (IRM). The IRM result and acceptable EQA performance range for specimen pool 0712FB1 was 7 ± /L. Note that the lines for the Sysmex XE-2100 and Coulter LH 750 instruments largely overlay each other. Manufacturer locations are as follows: ABX Pentra 120, Horiba ABX, Montpellier, France; ADVIA 120, Siemens, Tarrytown, NY; CELL-DYN 4000, Abbott, Santa Clara, CA; Coulter LH 750, Beckman Coulter, Miami, FL; and Sysmex XE-2100, Sysmex, Kobe, Japan. Plt-I Plt-O Plt-I, Plt-O, Plt-Imm Beckman Coulter Ac*T5 Diff Beckman Coulter Ac*T Diff Beckman Coulter Gen-S Beckman Coulter HmX Beckman Coulter LH 500 Beckman Coulter LH 750 Beckman Coulter LH 780 Abbott CELL-DYN 3200 Abbott CELL-DYN 3700 Abbott CELL-DYN 4000 Abbott CELL-DYN Sapphire Horiba ABX DX 120 Horiba ABX Pentra 120 Horiba ABX Pentra 60 Horiba ABX Pentra 80 Siemens ADVIA 120 Siemens ADVIA 2120 Sysmex KX-21 Sysmex poch-100i Sysmex SF-3000 Sysmex XE-2100 Sysmex XT-1800i Sysmex XT-2000i Figure 3 The mean percentage of counts for each specimen pool that fell within the acceptable external quality assessment performance range is shown for each analyzer type, together with the counting method offered by the analyzer platform: impedance (Plt-I), optical (Plt-O), or immunological (Plt-Imm). Sysmex XE-2100 instruments offer an optical and an impedance method but only impedance counts are collected in UK NEQAS (H) (UK National External Quality Assessment Scheme for General Haematology) surveys. Manufacturer locations are as follows: Abbott, Santa Clara, CA; Beckman Coulter, Miami, FL; Horiba ABX, Montpellier, France; Siemens, Tarrytown, NY; and Sysmex, Kobe, Japan. Am J Clin Pathol 2012;137: DOI: /AJCP86JMBFUCFCXA 71

8 De la Salle et al / Counting in Thrombocytopenic Specimens Table 6 Reproducibility Study Showing Mean Platelet Count and Between Replicate CV Values for Single Analyzers of the Most Commonly Represented Models * Abbott CELL-DYN 4000 Sysmex XE-2100 Beckman Coulter Siemens (Impedance) LH 750 ADVIA 120 Impedance Optical Specimen pool W A A A A Platelet count ( 10 9 /L) Between replicate CV (%) Specimen pool X B B B B Platelet count ( 10 9 /L) Between replicate CV (%) Specimen pool Y C C C C Platelet count ( 10 9 /L) Between replicate CV (%) Specimen pool Z D D D D Platelet count ( 10 9 /L) Between replicate CV (%) CV, coefficient of variation. * The reproducibility study undertaken with the Sysmex XE-2100 analyzer used specimen pools different from those used with other instruments as a result of technical problems at the XE-2100 testing site that prevented testing to the original schedule. Manufacturer locations are as follows: Abbott, Santa Clara, CA; Beckman Coulter, Miami, FL; Siemens, Tarrytown, NY; and Sysmex, Kobe, Japan. lowest in absolute terms for pool D (platelet count approximately /L). The individual analyzer CV values for these 3 analyzers approached the typical CVs of less than 3% that have been demonstrated in other evaluations, 13 indicating that the reproducibility of results is better on an individual analyzer than between analyzers of the same type at different laboratory locations. The CV values obtained from the Sysmex XE-2100 showed the same trend as the method group CV values but did not demonstrate the same reduction in CV seen with the other 3 analyzer types. The XE-2100 had been tested with a different set of specimen pools (W, X, Y, and Z) owing to an unforeseen technical problem; it is suggested that further work would be required to determine the reasons for the difference with this analyzer. Discussion In addition to the performance monitoring of individual laboratories, an important function of EQA is to improve the harmonization of laboratory results. UK NEQAS (H) is in a unique position to make a comparison of results from a substantial number of the analyzer models in most predominant use. This was a comprehensive study, covering a full range of thrombocytopenic specimen values, with a representative range and statistically robust number of analyzers. This study has shown that for thrombocytopenic EQA specimens, platelet counts using automated analyzers are overestimated in 66.3% of specimens (405/611) and significantly overestimated in 16.5% of specimens (101/611) when compared with the IRM platelet count. This observation is consistent with a multicenter study undertaken with EDTA blood samples from thrombocytopenic patients, demonstrating that the samples used in this study were valid. 7 No instrument matched the IRM for more than 28% (8/29) of specimen pools. Overestimation of thrombocytopenic platelet counts may result in the substantial undertransfusion of platelets in highrisk patients in need of platelet transfusion. The current suggested limit for administration of platelet concentrate is /L according to the British Committee for Standards in Haematology guidelines, and any consideration to reduce that limit to /L must take into account the performance limits of hematology analyzers. The CV increases as the platelet count decreases, particularly once the platelet count falls below /L (J. Parker-Williams, unpublished data, 2003). Previous EQA data 14 showed CV values of 22% to 66% at platelet counts of /L; CV results for specimens between 5 and /L in this study ranged from 15% to 43% for all analyzer types. Although variation was seen for different analyzers, the CV was inversely proportional to the platelet count for all analyzer models. The highest CV occurred at the lowest platelet counts, indicating that the poorest interanalyzer precision occurs with the most extreme thrombocytopenic specimens. The reproducibility of platelet counts in thrombocytopenic specimens analyzed on an individual analyzer is better, suggesting that within a single institution, less variability will be seen. This will allow a better demonstration of cumulative trends in a patient but does not indicate that results will be any more accurate. The reasons for inaccuracies in automated platelet counting have been well documented. Spuriously overestimated counts may arise from the failure of analyzers to discriminate red cell fragments and other nonplatelet particles from 72 Am J Clin Pathol 2012;137: DOI: /AJCP86JMBFUCFCXA

9 Hematopathology / Original Article platelets 2,15 ; conversely, underestimation may occur when large or giant platelets are not discriminated from red cells and are excluded from the platelet count. 16 The study showed a range of EQA performance between analyzers, with some demonstrating apparently greater accuracy than others. When assessed in terms of the proportion of counts within the acceptable EQA performance range, the Beckman Coulter LH 750 and Sysmex XE-2100 (impedance method) analyzers had more than 70% of counts within the acceptable EQA performance range, compared to fewer than 10% for the Abbott CELL-DYN 3200 and 4000 analyzers. However, the influence of the nature of the survey material may be a factor in this observation, even though results were similar to previous studies in EDTA-anticoagulated blood samples. 4 Several analyzer platforms offer more than 1 platelet counting method, depending on (among other factors) the platelet count. Abbott CELL-DYN 4000 and Sapphire analyzers have 3 methods available: impedance, optical, and immunologic; Sysmex XE-2100 instruments offer optical and impedance counting; Abbott CELL-DYN 3200 and Siemens ADVIA 120 and 2120 analyzers offer optical platelet counting only. The preparation of UK NEQAS (H) FBC survey material requires manipulation and fixation of donor blood. This in vitro manipulation may cause platelet activation, which has been cited as a cause of intermethod variation in platelet counts with greater impact on optical than impedance or immunologic methods. 17,18 UK NEQAS (H) has for some time requested the return of impedance counts only for Sysmex XE-2100 instruments, after consultation with the scheme scientific advisors and the instrument manufacturers. This instruction was in place for the entire period of this study. Although the platelet counting method used by Abbott CELL-DYN users was not gathered for every specimen in this study, the question was asked for surveys 0711FB to 0904FB. Not all participants responded to the question; however, 92% of Abbott CELL-DYN 3200, 4000, and Sapphire users who responded indicated that they used the optical platelet counting method for specimens in the study. By contrast, 89% of Abbott CELL-DYN 3700 users reported using the impedance method. No Abbott CELL- DYN user indicated use of the immunocount. These observations could account for the very wide variability in results and the apparent inaccuracy of the results returned by these instruments. An evaluation of the methods used by these analyzers using fresh, EDTA-anticoagulated blood specimens 19 showed all 3 methods available on the Abbott CELL-DYN platforms (optical, impedance, and immunologic) correlated well with the IRM count, although the immunologic method, which showed the best correlation, is not used routinely. Although the manipulation of the survey material used in this study may have been an artifactual cause of intermethod variation, the potential for discrepancy between platelet counting methods in the presence of marked platelet activation or platelet size differences has implications for clinical decision making in conditions such as disseminated intravascular coagulation 18 and idiopathic thrombocytopenic purpura. 20,21 This study highlights the need for a single material that can be used to assess performance across all technologies and to challenge automated cell counters at the difficult levels of critical clinical decision making. It is essential that pathology service providers and clinicians who rely on these services understand the limitations of the instrumentation in use and the measurement uncertainty of automated platelet counters. We hope that this analysis of performance of platelet counts at highly challenging levels will encourage dialogue among persons responsible for the development of policy, the service providers, the bodies responsible for performance monitoring, and the instrument manufacturers. The overestimation of platelet counts shown in previous work with fresh EDTA specimens 4 and confirmed using stabilized specimens in this study could indicate problems with instrument calibration at thrombocytopenic levels. Addressing this issue may improve platelet counts in thrombocytopenia in the future. From the 1 UK National External Quality Assessment Scheme for General Haematology, Watford, England; 2 Department of Haematology, University College Hospital, London, England; 3 The Oxford Haemophilia & Thrombosis Centre, The Churchill Hospital, Oxford, England; and 4 Medical Research Council Clinical Trials Unit, London. Address reprint requests to Barbara De la Salle: UK NEQAS (H), PO Box 14, Watford, WD18 0FJ, United Kingdom. Paul Harrison is a consultant for Sysmex UK. The University College Hospital London has received an unrestricted educational grant from Sysmex Europe in the last 12 months. Acknowledgment: We thank Anne Mahon for expert technical assistance throughout the study. References 1. British Committee for Standards in Haematology. Guidelines for the use of platelet transfusions. Br J Haematol. 2003;122: Ault KA. Platelet counting: is there room for improvement? Lab Haematol. 1996;2: Norris S, Pantelidou D, Smith D, et al. Immunoplatelet counting: potential for reducing the use of platelet transfusions through more accurate platelet counting. Br J Haematol. 2003;121: Segal HC, Briggs C, Kunka S, et al. Accuracy of platelet counting haematology analysers in severe thrombocytopenia and potential impact of platelet transfusion. Br J Haematol. 2005;128: Brecher G, Schneidermann M, Cronkite EP. The reproducibility of the platelet count. Am J Clin Pathol. 1953;23: Am J Clin Pathol 2012;137: DOI: /AJCP86JMBFUCFCXA 73

10 De la Salle et al / Counting in Thrombocytopenic Specimens 6. International Council for Standardization in Hematology. Recommended methods for the visual determination of white cell and platelet counts. WHO LAB :3. 7. Harrison P, Horton A, Grant D, et al. Immunoplatelet counting: a proposed new reference procedure. Br J Haematol. 2000;108: Harrison P, Ault KA, Chapman S, et al. An interlaboratory study of a candidate reference method for platelet counting. Am J Clin Pathol. 2001;115: International Council for Standardization in Haematology. Platelet counting by the RBC/platelet ratio method: a reference method. Am J Clin Pathol. 2001;115: Reardon DM, Mack D, Warner B. A whole blood control for blood count analysers and a source material for an EQA scheme. Med Lab Sci. 1991;48: Lewis SJ, De la Salle BJ. Quality assurance. In: Lewis SM, Bain BJ, Bates I, eds. Dacie and Lewis Practical Haematology. 10th ed. Edinburgh, Scotland: Churchill Livingstone; 2006: Huber PJ. Robust Statistics. New York, NY: Wiley; Bentley SA, Johnson A, Bishop CA. A parallel evaluation of four automated hematology analyzers. Am J Clin Pathol. 1993;100: Parker-Williams J. Immunoplatelet counting: platelet transfusions [letter]. Br J Haematol. 2003;123: Zandecki M, Genevieve F, Gerard J, et al. Spurious counts and spurious results on haematology analysers: a review, part I: platelets. Int J Lab Hematol. 2007;29: Kunicka JE, Fischer G, Murphy J, et al. Improved platelet counting using two dimensional laser light scatter. Am J Clin Pathol. 2000;114: Hervig T, Haugen T, Liseth K, et al. The platelet count accuracy of platelet concentrates obtained by using automated analysis is influenced by instrument bias and activated platelet components. Vox Sang. 2004;87: Kim SY, Kim JE, Kim HK, et al. Accuracy of platelet counting by automated hematologic analyzers in acute leukemia and disseminated intravascular coagulation: potential effects of platelet activation. Am J Clin Pathol. 2010;134: Grimaldi E, Del Vecchio L, Scopacasa F, et al. Evaluation of the platelet counting by Abbott CELL-DYN SAPPHIRE haematology analysers compared with flow cytometry. Int J Lab Hematol. 2007;31: Bowles KM, Bloxham DM, Perry DJ, et al. Discrepancy between impedance and immunofluorescence platelet counting has implications for clinical decision making in patients with idiopathic thrombocytopenic purpura. Br J Haematol. 2006;134: DiQuattro M, Gagliano F, Calabro GM, et al. Relationship between platelet counts, platelet volumes, and reticulated platelets in patients with ITP: evidence for significant platelet inaccuracies with conventional instrument methods. Int J Lab Hematol. 2009;31: Am J Clin Pathol 2012;137: DOI: /AJCP86JMBFUCFCXA

COMPARISON OF PLATELET COUNTS BY SYSMEX XE 2100 AND LH 750 WITH THE INTERNATIONAL FLOW REFERENCE METHOD IN THROMBOCYTOPENIC PATIENTS

COMPARISON OF PLATELET COUNTS BY SYSMEX XE 2100 AND LH 750 WITH THE INTERNATIONAL FLOW REFERENCE METHOD IN THROMBOCYTOPENIC PATIENTS COMPARISON OF PLATELET COUNTS BY SYSMEX XE 2100 AND LH 750 WITH THE INTERNATIONAL FLOW REFERENCE METHOD IN THROMBOCYTOPENIC PATIENTS TINA DADU 1, KUNAL SEHGAL 2, ANJUM SHAIKH 2, SHANAZ KHODAIJI 2 1. BLK

More information

Reduction of the Platelet Review Rate Using the Two-Dimensional Platelet Method

Reduction of the Platelet Review Rate Using the Two-Dimensional Platelet Method Coagulation and Transfusion Medicine / REDUCTION OF THE PLTELET REVIEW RTE Reduction of the Platelet Review Rate Using the Two-Dimensional Platelet Method Deanne H. Chapman, MT(SCP)SH, 1 Jonn Hardin, MT,

More information

Improved Platelet Counting Using Two-Dimensional Laser Light Scatter

Improved Platelet Counting Using Two-Dimensional Laser Light Scatter Hematopathology / A NEW APPROACH TO PLATELET COUNTING Improved Platelet Counting Using Two-Dimensional Laser Light Scatter Jolanta E. Kunicka, PhD, Gena Fischer, MT(ASCP), Justin Murphy, and David Zelmanovic,

More information

Validation of Platelet Counting Accuracy With the Celltac F Automated Hematology Analyzer

Validation of Platelet Counting Accuracy With the Celltac F Automated Hematology Analyzer Journal of Automated Methods & Management in Chemistry, 5 (5), no. 4, 35 39 c 5 Hindawi Publishing Corporation Validation of Platelet ing Accuracy With the Celltac F Automated Hematology Analyzer Yutaka

More information

Evaluation of the Sysmex XN-550, a Novel Compact Haematology analyser from the XN-L series, compared to the XN-20 system.

Evaluation of the Sysmex XN-550, a Novel Compact Haematology analyser from the XN-L series, compared to the XN-20 system. Evaluation of the Sysmex XN-550, a Novel Compact Haematology analyser from the XN-L series, compared to the XN-20 system. H Tailor 1, I Mackie 2, A Mellick 3, S Machin 1, 2 1 Haematology Evaluations Unit,

More information

Validation of Platelet Counting Accuracy With the Celltac F Automated Hematology Analyzer

Validation of Platelet Counting Accuracy With the Celltac F Automated Hematology Analyzer Automated Methods & Management in Chemistry, 25 (25), no. 4, 235 239 c 25 Validation of Platelet Counting Accuracy With the Celltac F Automated Hematology Analyzer Yutaka Nagai, 1 Hiroshi Kondo, 2 and

More information

Platelet Counting by the RBC/Platelet Ratio Method A Reference Method

Platelet Counting by the RBC/Platelet Ratio Method A Reference Method Coagulation and Transfusion Medicine / REFERENCE METHOD FOR PLATELET COUNTING Platelet Counting by the RBC/Platelet Ratio Method A Reference Method International Council for Standardization in Haematology

More information

The clinical utility of IPF in thrombocytopenia

The clinical utility of IPF in thrombocytopenia The clinical utility of IPF in thrombocytopenia Platelets platelets Platelets are the smallest blood cells which are released by cytoplasmic blebbing of megakaryocytes with an average size of 2um. Normal

More information

ICSH guidelines for the verification and performance of automated cell counters for body fluids

ICSH guidelines for the verification and performance of automated cell counters for body fluids International Journal of Laboratory Hematology ORIGINAL ARTICLE The Official journal of the International Society for Laboratory Hematology INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY ICSH guidelines

More information

Platelet Counts. Platelet Counts in Severely Thrombopenic Samples Samuel J Machin and Carol Briggs

Platelet Counts. Platelet Counts in Severely Thrombopenic Samples Samuel J Machin and Carol Briggs 10. Platelet Counts Platelet Counts in Severely Thrombopenic Samples Samuel J Machin and Carol Briggs Introduction Platelets play a major role in haemostasis and therefore numerical deficiency or defects

More information

CONCURRENT 10: THE COMPLETE BLOOD COUNT AND BEYOND: QUALITY ISSUES AND REFERENCE METHODS

CONCURRENT 10: THE COMPLETE BLOOD COUNT AND BEYOND: QUALITY ISSUES AND REFERENCE METHODS CONCURRENT 10: THE COMPLETE BLOOD COUNT AND BEYOND: QUALITY ISSUES AND REFERENCE METHODS Saturday May 14 th 2016 1:30 3:00 PM (13.30 15.00hrs) International Society for Laboratory Hematology (ISLH) CONCURRENT

More information

The Simple Solution for CBC Testing

The Simple Solution for CBC Testing QBC STAR Centrifugal Hematology Analyzer The Simple Solution for CBC Testing Diagnostics Innovative Solutions for a Healthier World UNRIVALED SIMPLICITY The QBC STAR Centrifugal Hematology System is the

More information

CELL-DYN ImmunoPlt (CD61)

CELL-DYN ImmunoPlt (CD61) CELL-DYN ImmunoPlt (CD61) REF REAGENT 03H05-01 40 tests 1. NAME CELL-DYN ImmunoPlt (CD61) Reagent 2. INTENDED USE CELL-DYN ImmunoPlt (CD61) Reagent is a one-color direct immunofluorescence reagent for

More information

UNRIVALED SIMPLICITY. Exceptional Ease of Use. The STAR brings new meaning to the word simple

UNRIVALED SIMPLICITY. Exceptional Ease of Use. The STAR brings new meaning to the word simple UNRIVALED SIMPLICITY The QBC STAR Centrifugal Hematology System is the simplest solution for in-office CBC testing. An innovative approach employing dry hematology reagents instead of the bulky liquid

More information

Automated Platelet Counts

Automated Platelet Counts Automated Platelet Counts Accuracy, Precision, and Range DENNIS W. ROSS, M.D., PH.D., LANIER AYSCUE, M.T.(ASCP), AND MARGARET GULLEY, B.S. Ross, Dennis W., Ayscue, Lanier, and Gulley, Margaret: Automated

More information

Automated Hematology Analyzer. Small Footprint. Big Difference.

Automated Hematology Analyzer. Small Footprint. Big Difference. XS-1000i Automated Hematology Analyzer Small Footprint. Big Difference. Small But Powerful: The XS-1000i Even with these challenges, the need for hematology testing has remained steady or continues to

More information

Pushing limits in routine laboratory haematology with the XT-4000i

Pushing limits in routine laboratory haematology with the XT-4000i Pushing limits in routine laboratory haematology with the XT-4000i Sysmex Xtra Online December 2013 Fig. 1 The XT-4000i haematology system With the launch of the XE-5000, Sysmex set new standards in haematology.

More information

The Future of Hematological Diagnosis

The Future of Hematological Diagnosis Interview The Future of Hematological Diagnosis Bruce H. Davis, M.D. Bruce H. Davis, M.D. is the Director of Analytical Cytometry in the Department of Clinical Pathology at the William Beaumont Hospital

More information

Business Case for Haematology Analyser Replacement Recommendation

Business Case for Haematology Analyser Replacement Recommendation Business Case for Haematology Analyser Replacement Recommendation Agenda Item 108/11 1. Executive Summary The Business Case [BC] identifying the need to replace the analytical platform in the Haematology

More information

Automated Hematology System. Your Complete Choice

Automated Hematology System. Your Complete Choice XE-5000 Automated Hematology System Your Complete Choice Advanced Technology Solutions to Meet Your Lab s Needs Even with these challenges, the need for hematology testing has remained steady or continued

More information

Novel POC analysis for determination of total and 5-part differential WBC count among a US population, in comparison to Beckman Coulter LH750

Novel POC analysis for determination of total and 5-part differential WBC count among a US population, in comparison to Beckman Coulter LH750 Novel POC analysis for determination of total and -part differential WBC count among a US population, in comparison to Beckman Coulter LH7 Reed, Jeremy*, Johnsson, Elin**, Hockum, Sandra** * Quest Diagnostics,

More information

Greater Possibilities By Design

Greater Possibilities By Design Greater Possibilities By Design January 23, 2013 Nilam Patel MT(ASCP)SH Sr. Product Manager, Automation Solutions Objectives Review current and new portfolio Technology and benefits Operational benefits

More information

Harmonization of red blood cell distribution width (RDW): an attainable target?

Harmonization of red blood cell distribution width (RDW): an attainable target? Original Article Page 1 of 5 Harmonization of red blood cell distribution width (RDW): an attainable target? Giuseppe Lippi 1, Silvia Pipitone 2, Emmanuel J. Favaloro 3,4 1 Section of Clinical Biochemistry,

More information

Original Article. A Novel Automated Slide-Based Technology for Visualization, Counting, and Characterization of the Formed Elements of Blood

Original Article. A Novel Automated Slide-Based Technology for Visualization, Counting, and Characterization of the Formed Elements of Blood Original Article A Novel Automated Slide-Based Technology for Visualization, Counting, and Characterization of the Formed Elements of Blood A Proof of Concept Study James W. Winkelman, MD; Milenko J. Tanasijevic,

More information

Transfusion effect of random donor platelet and single donor platelet in thrombocytopenic patients at tertiary care hospital of South Gujarat

Transfusion effect of random donor platelet and single donor platelet in thrombocytopenic patients at tertiary care hospital of South Gujarat International Journal of Research in Medical Sciences Unagar CA et al. Int J Res Med Sci. 2017 Jul;5(7):3033-3037 www.msjonline.org pissn 2320-6071 eissn 2320-6012 Original Research Article DOI: http://dx.doi.org/10.18203/2320-6012.ijrms20172983

More information

Evaluation of the Automated Immature Granulocyte Count (IG) on Sysmex XE-2100 Automated Haematology Analyser vs. Visual Microscopy (NCCLS H20-A)

Evaluation of the Automated Immature Granulocyte Count (IG) on Sysmex XE-2100 Automated Haematology Analyser vs. Visual Microscopy (NCCLS H20-A) Evaluation of the Automated Immature Granulocyte Count (IG) on Sysmex XE-21 Automated Haematology Analyser vs. Visual Microscopy (NCCLS H2-A) Th WEILAND *1, H KALKMAN *2, and H HEIHN *1 *1 Central Laboratory,

More information

Performance Evaluation of the Coulter LH 750 Hematology Analyzer

Performance Evaluation of the Coulter LH 750 Hematology Analyzer ISLH Laboratory Hematology 7:217 228 2001 Carden Jennings Publishing Co., Ltd. Official Publication Performance Evaluation of the Coulter LH 750 Hematology Analyzer TRACEY FERNANDEZ, LYNN BESSERT DOMACK,

More information

Platelet Refractoriness: The Basics. Martin H. Bluth, MD, PhD

Platelet Refractoriness: The Basics. Martin H. Bluth, MD, PhD Platelet Refractoriness: The Basics Martin H. Bluth, MD, PhD Complete Toxicology Laboratories, LLC Objectives Define platelet refractoriness and associated conditions that may cause platelet refractoriness.

More information

ThromboLUX for determining platelet quality

ThromboLUX for determining platelet quality Updating information provided by the company October 2012 Horizon Scanning Centre October 2011 ThromboLUX for determining platelet quality SUMMARY Lay summary click here This briefing is based on information

More information

Immunophenotyping of Peripheral Blood and Bone Marrow Cells by Flow Cytometry *Akanni EO and # Palini A.

Immunophenotyping of Peripheral Blood and Bone Marrow Cells by Flow Cytometry *Akanni EO and # Palini A. Immunophenotyping of Peripheral Blood and Bone Marrow Cells by Flow Cytometry *Akanni EO and # Palini A. * Department of Haematology & Blood Transfusion,College of Health Science, Ladoke Akintola University

More information

Plateletcrit is superior to platelet count for assessing platelet status in Cavalier King Charles Spaniels

Plateletcrit is superior to platelet count for assessing platelet status in Cavalier King Charles Spaniels Veterinary Clinical Pathology ISSN 0275-6382 Plateletcrit is superior to platelet count for assessing platelet status in Cavalier King Charles Spaniels Harold Tvedten 1, Inger Lilliehöök 2, Anna Hillström

More information

A Division of Woodley Equipment Company Ltd

A Division of Woodley Equipment Company Ltd LEADING EDGE TECHNOLOGY FOR POINT OF CARE HAEMATOLOGY A Division of Woodley Equipment Company Ltd Contents The QBC STAR POC Haematology System...................4 Operating Ranges.......................................7

More information

Application Note for Gentian Cystatin C Immunoassay on. ABX Pentra 400 1, Horiba Medical 1 v03en-june 2011

Application Note for Gentian Cystatin C Immunoassay on. ABX Pentra 400 1, Horiba Medical 1 v03en-june 2011 Page 1 of 5 Application Note for Gentian Cystatin C Immunoassay on For in vitro diagnostic use only ABX Pentra 400 1, Horiba Medical 1 v03en-june 2011 Intended Use The Gentian Cystatin C Immunoassay is

More information

AAG PRINCIPLE SPECIMEN. REF (150 tests) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

AAG PRINCIPLE SPECIMEN. REF (150 tests) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE IMMAGE Immunochemistry Systems Chemistry Information Sheet Copyright 2010 Beckman Coulter, Inc. Alpha 1 -Acid Glycoprotein REF 447780 (150 tests) For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by:

More information

Dry Hematology Analyzer Fast, Simple CBC Analysis Designed for the Point of Care

Dry Hematology Analyzer Fast, Simple CBC Analysis Designed for the Point of Care QBC STAR TM Dry Hematology Analyzer Fast, Simple CBC Analysis Designed for the Point of Care Reach for the STAR The QBC STAR hematology analyzer expands and advances the unique capabilities of previous

More information

The International Haemostasis External Quality Control Program

The International Haemostasis External Quality Control Program The International Haemostasis External Quality Control Program Intended use of Quality Control Primary Purpose of the Clinical Laboratory To produce accurate results that will correctly diagnose and interpret

More information

PERFECT-COUNT MICROSPHERES

PERFECT-COUNT MICROSPHERES PERFECT-COUNT MICROSPHERES Perfect-Count Microspheres-Product code PCB-100 for 100 tests Introduction In recent years, the determination of absolute cell counts has been shown to be relevant in different

More information

SAMPLE. Validation, Verification, and Quality Assurance of Automated Hematology Analyzers; Approved Standard Second Edition

SAMPLE. Validation, Verification, and Quality Assurance of Automated Hematology Analyzers; Approved Standard Second Edition June 2010 Validation, Verification, and Quality Assurance of Automated Hematology Analyzers; Approved Standard Second Edition This document provides guidance for the validation, verification, calibration,

More information

WADA Technical Document TD2019BAR. Blood Analytical Requirements for the Athlete Biological Passport

WADA Technical Document TD2019BAR. Blood Analytical Requirements for the Athlete Biological Passport 1. Introduction Blood Analytical Requirements for the Athlete Biological Passport This Technical Document (TD) has been established to harmonize the analysis of blood Samples collected, both In-Competition

More information

Automated Hematology Analyzer. Differentiate with the XT-4000i

Automated Hematology Analyzer. Differentiate with the XT-4000i XT-4000i Automated Hematology Analyzer Differentiate with the XT-4000i Advanced Technology Solutions to Meet Your Lab s Needs Even with these challenges, the need for hematology testing has remained steady

More information

9/7/17. Implementing the ARK Immunoassay for Therapeutic Drug Monitoring of High-Dose Methotrexate. Disclosure. Learning Objectives

9/7/17. Implementing the ARK Immunoassay for Therapeutic Drug Monitoring of High-Dose Methotrexate. Disclosure. Learning Objectives Implementing the ARK Immunoassay for Therapeutic Drug Monitoring of High-Dose Methotrexate Kristine R. Crews, PharmD, FCCP, BCPS St. Jude Children s Research Hospital September 14, 2017 Disclosure I declare

More information

Date. Albumin measurements are used in the diagnosis and treatment of numerous diseases primarily involving the liver and/or kidneys.

Date. Albumin measurements are used in the diagnosis and treatment of numerous diseases primarily involving the liver and/or kidneys. SYNCHRON System(s) Chemistry Information Sheet Copyright 2013 Beckman Coulter, Inc. Albumin REF 442765 For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by Date Reviewed by Date PRINCIPLE INTENDED USE

More information

Establishing Quality Control Target Values and Standard Deviations for Hematology Instrumentation

Establishing Quality Control Target Values and Standard Deviations for Hematology Instrumentation Establishing Quality Control Target Values and Standard Deviations for Hematology Instrumentation For informational purposes only. WHO REQUIRES IT AND WHAT THEY REQUIRE Regulatory agencies require the

More information

ABX Pentra DX120. Hematology analyzer. 45 parameters Integrated slide-maker Expert validation system

ABX Pentra DX120. Hematology analyzer. 45 parameters Integrated slide-maker Expert validation system ABX Pentra DX120 Hematology analyzer 45 parameters Integrated slide-maker Expert validation system ABX Pentra DX 120 In the heart of hematopoiesis Complete cytology platform 120 samples per hour Continuous

More information

Automation In Life Sciences and Diagnostics Applications

Automation In Life Sciences and Diagnostics Applications Automation In Life Sciences and Diagnostics Applications Pedro Diaz Director of Research Beckman Coulter Life Sciences February 20-22, 2013 Orlando World Marriott Center Orlando, Florida USA Beckman Coulter

More information

norma Hematology family norma Access norma norma Vision Reagents

norma Hematology family norma Access norma norma Vision Reagents norma Hematology family norma Access norma Reagents norma Vision Smallest reagent consumption in the world Smallest processed sample volume, ideal for pediatric use Microfluidic technology 3 A smart hematology

More information

BEYOND A BETTER BOX XN-V SERIES TM MULTISPECIES HEMATOLOGY ANALYZERS. Accelerate Your Multispecies Research Laboratory XN-V IS FOR ANIMAL USE ONLY

BEYOND A BETTER BOX XN-V SERIES TM MULTISPECIES HEMATOLOGY ANALYZERS. Accelerate Your Multispecies Research Laboratory XN-V IS FOR ANIMAL USE ONLY BEYOND A BETTER BOX XN-V SERIES TM MULTISPECIES HEMATOLOGY ANALYZERS Accelerate Your Multispecies Research Laboratory XN-V IS FOR ANIMAL USE ONLY MOVE YOUR MULTI-SPECIES LABORATORY BEYOND A BETTER BOX*

More information

Ready, Set, Test! AACC Conference Mass Spectrometry in the Clinical Lab: Best Practice and Current Applications September 17-18, 2013 St.

Ready, Set, Test! AACC Conference Mass Spectrometry in the Clinical Lab: Best Practice and Current Applications September 17-18, 2013 St. Ready, Set, Test! Ross Molinaro, PhD, MLS(ASCP) CM, DABCC, FACB Medical Director, Clinical Laboratories Emory University Hospital Midtown Emory Clinical Translational Research Laboratory AACC Conference

More information

WAIT! Ready, Set, Test! Financial Disclosure. Research/Educational grants/consulting/salary support

WAIT! Ready, Set, Test! Financial Disclosure. Research/Educational grants/consulting/salary support Ready, Set, Test! Ross Molinaro, PhD, MLS(ASCP) CM, DABCC, FACB Medical Director, Clinical Laboratories Emory University Hospital Midtown Emory Clinical Translational Research Laboratory AACC Conference

More information

NEW PRODUCT DEVELOPMENT OPPORTUNITES AND BUSINESS EXPANSION STRATEGIES FOR INSTRUMENT AND REAGENT SUPPLIERS FRANCE GERMANY ITALY SPAIN UK USA

NEW PRODUCT DEVELOPMENT OPPORTUNITES AND BUSINESS EXPANSION STRATEGIES FOR INSTRUMENT AND REAGENT SUPPLIERS FRANCE GERMANY ITALY SPAIN UK USA MEDICAL DEVICES PHARMACEUTICALS CHEMICALS FOOD & BEVERAGE ELECTRONICS This study is being updated. Subscribers will receive complimentary copies of the previous edition while the new report is in preparation.

More information

Ruby. A shining example of. Put science on your side. Intended Use:

Ruby. A shining example of. Put science on your side. Intended Use: High efficiency hematology A shining example of advanced technology Intended Use: The CELL-DYN Ruby is a multi-parameter automated Hematology analyzer designed for in vitro use in clinical laboratories.

More information

Pre-analytics in coagulation lab: why struggle for better results?

Pre-analytics in coagulation lab: why struggle for better results? Pre-analytics in coagulation lab: why struggle for better results? Valdas Banys, MD, PhD Vilnius University Faculty of Medicine Vilnius University Hospital Santariskiu Klinikos, Center of Laboratory Medicine

More information

Biological Variations of Leukocyte Numerical and Morphologic Parameters Determined by UniCel DxH 800 Hematology Analyzer

Biological Variations of Leukocyte Numerical and Morphologic Parameters Determined by UniCel DxH 800 Hematology Analyzer Biological Variations of Leukocyte Numerical and Morphologic Parameters Determined by UniCel DxH 800 Hematology Analyzer Huqiang Tang, MS, MPH; Jiyong Jing, MS; Dandan Bo, BS; Dongsheng Xu, MD, PhD N Context.

More information

Immature Granulocyte Enumeration Our Journey from Manual to Automated Reporting

Immature Granulocyte Enumeration Our Journey from Manual to Automated Reporting Immature Granulocyte Enumeration Our Journey from Manual to Automated Reporting Mary Capper MT(ASCP)SH Supervisor, Hematology/Hemostasis University of Iowa Hospitals and Clinics Objectives Describe the

More information

EDTA-induced pseudothrombocytosis and citrate-induced platelet agglutination in a patient with Waldenstrom macroglobulinemia

EDTA-induced pseudothrombocytosis and citrate-induced platelet agglutination in a patient with Waldenstrom macroglobulinemia CASE REPORT EDTA-induced pseudothrombocytosis and citrate-induced platelet agglutination in a patient with Waldenstrom macroglobulinemia Pia B ukmann Larsen 1,, Jonas Vikesa 2, & Lennart Friis-Hansen 3

More information

PAB PRINCIPLE. Kit Reorder # ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

PAB PRINCIPLE. Kit Reorder # ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE SYNCHRON CX System(s) Chemistry Information Sheet Copyright 2008 Beckman Coulter, Inc. Prealbumin Kit Reorder # 475106 For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Date Reviewed by: Date PRINCIPLE

More information

New Strategies for the Optimal Use of Platelet Transfusions

New Strategies for the Optimal Use of Platelet Transfusions TRANSFUSION MEDICINE New Strategies for the Optimal Use of Platelet Transfusions Morris A. Blajchman, 1 Sherrill J. Slichter, 2 Nancy M. Heddle, 3 and Michael F. Murphy 4 1 Departments of Medicine and

More information

Quality Parameters of Platelet Concentrate at Kenyatta National Hospital s Blood Transfusion Unit

Quality Parameters of Platelet Concentrate at Kenyatta National Hospital s Blood Transfusion Unit International Journal of Hematological Disorders, 2014, Vol. 1, No. 1, 35-40 Available online at http://pubs.sciepub.com/ijhd/1/1/7 Science and Education Publishing DOI:10.12691/ijhd-1-1-7 Quality Parameters

More information

Total Analytic Error From Concept to Application

Total Analytic Error From Concept to Application Página 1 de 5 Clinical Laboratory News Subscribe CLN Daily CLN Stat CLN Articles Total Analytic Error From Concept to Application Author: James O. Westgard, PhD, and Sten A. Westgard, MS // Date: SEP.1.2013

More information

Annual meeting of the Belgian Society for Advancement of Cytometry. Platelet counting: current issues and solutions

Annual meeting of the Belgian Society for Advancement of Cytometry. Platelet counting: current issues and solutions Annual meeting of the Belgian Society for Advancement of Cytometry Platelet counting: current issues and solutions Prof. François Mullier, Prof. Bernard Chatelain Université catholique de Louvain, CHU

More information

11/2/2015. Describe the technologies available on the XN-Series and the reason(s) for selection

11/2/2015. Describe the technologies available on the XN-Series and the reason(s) for selection XN-Series Technology What s Inside the Box? OBJECTIVES Describe the technologies available on the XN-Series and the reason(s) for selection Utilize example case studies to illustrate the new capabilities

More information

FEDERAL PUBLIC SERVICE, HEALTH, FOOD CHAIN SECURITY AND ENVIRONMENT CLINICAL BIOLOGY COMMISSION COMMITTEE OF EXPERTS GLOBAL REPORT

FEDERAL PUBLIC SERVICE, HEALTH, FOOD CHAIN SECURITY AND ENVIRONMENT CLINICAL BIOLOGY COMMISSION COMMITTEE OF EXPERTS GLOBAL REPORT FEDERAL PUBLIC SERVICE, HEALTH, FOOD CHAIN SECURITY AND ENVIRONMENT CLINICAL BIOLOGY COMMISSION COMMITTEE OF EXPERTS GLOBAL REPORT EXTERNAL QUALITY ASSESSMENT IN CLINICAL BIOLOGY FLOW CYTOMETRY CD34+ STEM

More information

Stability of Plasma for Add-On PT and APTT Tests

Stability of Plasma for Add-On PT and APTT Tests COAGULATION AND TRANSFUSION MEDICINE Original Article Stability of Plasma for Add-On PT and APTT Tests DEMETRA NEOFOTISTOS, MT, MARIA OROPEZA, MT(ASCP), AND CHUNG-HSIN TS'AO, PhD We conducted studies to

More information

Improving Outcomes of Cardiac Surgery with a Point-of-Care Based Transfusion Algorithm

Improving Outcomes of Cardiac Surgery with a Point-of-Care Based Transfusion Algorithm Improving Outcomes of Cardiac Surgery with a Point-of-Care Based Transfusion Algorithm Summer Syed (local PI) Keyvan Karkouti Deputy Anesthesiologist-in-Chief, Toronto General Hospital, University Health

More information

Standardization of Reticulocyte Values in an Antidoping Context

Standardization of Reticulocyte Values in an Antidoping Context Hematopathology / STANDARDIZATION OF RETICULOCYTE VALUES Standardization of Reticulocyte Values in an Antidoping Context Michael J. Ashenden, PhD, Ken Sharpe, PhD, 2 Rasmus Damsgaard, MD, 3 and Lisa Jarvis,

More information

New Frontiers in the UK Hematology and Flow Cytometry Market: Business Challenges, Emerging Technologies, Competitive Landscape.

New Frontiers in the UK Hematology and Flow Cytometry Market: Business Challenges, Emerging Technologies, Competitive Landscape. New Frontiers in the UK Hematology and Flow Cytometry Market: Business Challenges, Emerging Technologies, Competitive Landscape Table of Contents I. Introduction II. Worldwide Market and Technology Overview

More information

SCIENTIFIC INSTITUTE OF PUBLIC HEALTH QUALITY OF MEDICAL LABORATORIES CLINICAL BIOLOGY COMMISSION COMMITTEE OF EXPERTS GLOBAL REPORT

SCIENTIFIC INSTITUTE OF PUBLIC HEALTH QUALITY OF MEDICAL LABORATORIES CLINICAL BIOLOGY COMMISSION COMMITTEE OF EXPERTS GLOBAL REPORT SCIENTIFIC INSTITUTE OF PUBLIC HEALTH QUALITY OF MEDICAL LABORATORIES CLINICAL BIOLOGY COMMISSION COMMITTEE OF EXPERTS GLOBAL REPORT EXTERNAL QUALITY ASSESSMENT IN CLINICAL BIOLOGY FLOW CYTOMETRY CD34+

More information

Falsely Elevated INR Results Due to the Sensitivity of a Thromboplastin Reagent to Heparin

Falsely Elevated INR Results Due to the Sensitivity of a Thromboplastin Reagent to Heparin COAGULATION AND TRANSFUSION MEDICINE Falsely Elevated INR Results Due to the Sensitivity of a Thromboplastin Reagent to Heparin BRENDA F. LEECH, RT, 1 AND CEDRIC J. CARTER, MB, FRCP 2 The aim of this study

More information

CLSI C60: Assay Validation & Post-Validation Monitoring

CLSI C60: Assay Validation & Post-Validation Monitoring CLSI C60: Assay Validation & Post-Validation Monitoring Ross J. Molinaro, MT(ASCP), PhD, DABCC, FACB Medical Director Core Laboratory, Emory University Hospital Midtown Emory Clinical Translational Research

More information

XN Series. What s in it for me? Krista Curcio, MT(ASCP), MBA Product Manager, Hematology Systems Sysmex America, Inc. All rights reserved.

XN Series. What s in it for me? Krista Curcio, MT(ASCP), MBA Product Manager, Hematology Systems Sysmex America, Inc. All rights reserved. XN Series What s in it for me? Krista Curcio, MT(ASCP), MBA Product Manager, Hematology Systems 2012 Sysmex America, Inc. All rights reserved. Agenda XN Series Product Overview A system to fit every lab

More information

New Frontiers in the US Hematology and Flow Cytometry Market: Business Challenges, Emerging Technologies, Competitive Landscape.

New Frontiers in the US Hematology and Flow Cytometry Market: Business Challenges, Emerging Technologies, Competitive Landscape. New Frontiers in the US Hematology and Flow Cytometry Market: Business Challenges, Emerging Technologies, Competitive Landscape Table of Contents I. Introduction II. Worldwide Market and Technology Overview

More information

XN-SERIES AUTOMATED HEMATOLOGY SYSTEMS

XN-SERIES AUTOMATED HEMATOLOGY SYSTEMS XN-SERIES AUTOMATED HEMATOLOGY SYSTEMS Reshaping Compact Automation BEYOND A BETTER BOX MOVE YOUR HEMATOLOGY LABORATORY BEYOND A BETTER BOX It s what makes Sysmex the smarter choice and sets our integrated

More information

Navios EX FLOW CYTOMETER POWERFUL, DEPENDABLE CLINICAL FLOW CYTOMETRY

Navios EX FLOW CYTOMETER POWERFUL, DEPENDABLE CLINICAL FLOW CYTOMETRY Navios EX FLOW CYTOMETER POWERFUL, DEPENDABLE CLINICAL FLOW CYTOMETRY BECAUSE EVERY EVENT MATTERS The Navios EX flow cytometer offers a solution for advanced cytometry applications optimized for the clinical

More information

norma Hematology family norma norma Vision Reagents

norma Hematology family norma norma Vision Reagents norma Hematology family norma Reagents norma Vision Smallest reagent consumption in the world Smallest processed sample volume, ideal for pediatric use Microfluidic technology 3 The Icon 3 is a smart,

More information

Clinical Chemistry Approach to Evaluation of Commutability

Clinical Chemistry Approach to Evaluation of Commutability Clinical Chemistry Approach to Evaluation of Commutability Hubert W. Vesper, Ph.D. Director, Clinical Standardization Programs Division of Laboratory Sciences Centers for Disease Control and Prevention,

More information

APB PRINCIPLE SPECIMEN. REF (300 tests) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

APB PRINCIPLE SPECIMEN. REF (300 tests) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE IMMAGE Immunochemistry Systems Chemistry Information Sheet Copyright 2010 Beckman Coulter, Inc. Apolipoprotein B REF 447730 (300 tests) For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Date Reviewed

More information

Reshaping compact automation

Reshaping compact automation Automated Hematology Analyzer XN-SERIES Reshaping compact automation www.sysmex.com/us Optimize your laboratory As analyzers become increasingly sophisticated, the laboratory s contribution to the continuum

More information

Quality Control in Flow. Dr David Westerman Head of Haematopathology Peter MacCallum Cancer Centre

Quality Control in Flow. Dr David Westerman Head of Haematopathology Peter MacCallum Cancer Centre Quality Control in Flow Dr David Westerman Head of Haematopathology Peter MacCallum Cancer Centre Aims Quality Assurance Quality Control Literature In house competencies SHOT DATA 1996-2009 Ref: SHOT Annual

More information

Advances in Biochemistry & Applications in Medicine

Advances in Biochemistry & Applications in Medicine Advances in Biochemistry & Applications in Medicine Chapter 1 ISBN: 978-93-87500-49-5 Quality Control in a Clinical Laboratory Vaneet Kaur 1 ; Pawan Kumar Kare 1* ; Himanshu Madaan 1 1 Department of Biochemistry,

More information

Reshaping compact automation

Reshaping compact automation Automated Hematology Analyzer XN-SERIES Reshaping compact automation www.sysmex.com/us Optimize your laboratory As analyzers become increasingly sophisticated, the laboratory s contribution to the continuum

More information

DIAGNOSTICS ASSESSMENT PROGRAMME

DIAGNOSTICS ASSESSMENT PROGRAMME Diagnostics Consultation Document s THEME: LEVEL OF EVIDENCE 1. Section Response 1. Provisional Recommendations (page 2) Biomedical, Inc., the manufacturer of the My5-FU assay, is disappointed in the draft

More information

Molecular Diagnosis Challenges & Solutions. Using Molecular Kits or Laboratory Developed Tests (Home Brew), Emphasis on Validation

Molecular Diagnosis Challenges & Solutions. Using Molecular Kits or Laboratory Developed Tests (Home Brew), Emphasis on Validation Using Molecular Kits or Laboratory Developed Tests (Home Brew), Emphasis on Validation Molecular Diagnosis Challenges & Solutions Behzad Poopak, DCLS PhD Tehran Medical Branch- Islamic Azad University

More information

XN Hematology Case Studies Every Picture Tells a Story

XN Hematology Case Studies Every Picture Tells a Story XN Hematology Case Studies Every Picture Tells a Story Barbara J Connell MS, MT(ASCP)SH Senior Product Manager Technical Marketing & Communications Sysmex America, Inc Objectives Describe the advantages

More information

Platelet Transfusion Therapy

Platelet Transfusion Therapy REVIEW ARTICLE Platelet Transfusion Therapy *S Karim 1, E Hoque 2, MM Hoque 3, MMR Siddiqui 4 1 Dr. Shanaz Karim, Assistent Professor of Transfusion Medicine, Dhaka Medical College Hospital 2 Dr. Ehteshamul

More information

AIMS FELLOWSHIP CURRICULUM GENERAL/CORE LABORATORY IV

AIMS FELLOWSHIP CURRICULUM GENERAL/CORE LABORATORY IV AIMS FELLOWSHIP CURRICULUM GENERAL/CORE LABORATORY IV Module Code: GENERAL IV (Compulsory Module) Module Title: Haematology Module Convenor: Mr David Condie, B. App.Sci (QUT), FAIMS,MASM, MBA Senior Scientist

More information

9/29/2015. Describe the technologies available on the XN-Series and the reason(s) for selection

9/29/2015. Describe the technologies available on the XN-Series and the reason(s) for selection XN-Series Technology What s Inside the Box? OBJECTIVES Describe the technologies available on the XN-Series and the reason(s) for selection Utilize example case studies to illustrate the new capabilities

More information

VALIDATION OF AN AUTOMATED ANTIGLOBULIN TEST MACHINE THE LORNE ACT-24. Zs. SZABÓ, É.K. BENKÖ, J. BUCZKO, Á. FRISS* D.M. DOWNIE+ AND D.

VALIDATION OF AN AUTOMATED ANTIGLOBULIN TEST MACHINE THE LORNE ACT-24. Zs. SZABÓ, É.K. BENKÖ, J. BUCZKO, Á. FRISS* D.M. DOWNIE+ AND D. VALIDATION OF AN AUTOMATED ANTIGLOBULIN TEST MACHINE THE LORNE ACT-24 Zs. SZABÓ, É.K. BENKÖ, J. BUCZKO, Á. FRISS* D.M. DOWNIE+ AND D. VOAK + CYTOTECH INSTRUMENTS LTD., BUDAPEST, HUNGARY. *CENTRAL MILITARY

More information

Challenges for Flow Cytometry in Regulated Bioanalysis

Challenges for Flow Cytometry in Regulated Bioanalysis Challenges for Flow Cytometry in Regulated Bioanalysis Minesh Patel Merck Millipore Discovery & Development Solutions. Oxford, UK. Overview Flow cytometry principles Current uses and regulatory environments

More information

Automated Optical Counting of Blood Platelets

Automated Optical Counting of Blood Platelets Automated Optical Counting of Blood Platelets By GEOFFREY M. BRrrnx, SHIRLEY A. DEW, AND ELvI K. FEWELL We have evaluated the use of an optical particle counter to perform automated platelet counts on

More information

Navios EX FLOW CYTOMETER POWERFUL, DEPENDABLE FLOW CYTOMETRY

Navios EX FLOW CYTOMETER POWERFUL, DEPENDABLE FLOW CYTOMETRY Navios EX FLOW CYTOMETER POWERFUL, DEPENDABLE FLOW CYTOMETRY BECAUSE EVERY EVENT MATTERS The Navios EX flow cytometer offers a solution for advanced cytometry applications with optimized workflows for

More information

Confirm assay performance Identify poor assay performance. Confirm correction of poor performance

Confirm assay performance Identify poor assay performance. Confirm correction of poor performance Performance criteria for EQA schemes need for harmonization Graham Jones Department of Chemical Pathology St Vincent s Hospital, Sydney Milan November 2014 The Role of External Quality Assurance Confirm

More information

Aim. To provide basic knowledge on the processing, testing and issue of blood donations by NHSBT

Aim. To provide basic knowledge on the processing, testing and issue of blood donations by NHSBT From Donor To Door Aim To provide basic knowledge on the processing, testing and issue of blood donations by NHSBT At the end of this session you will be aware of: The stages of processing, testing and

More information

BC Auto Hematology Analyzer. A CUTE 5-part

BC Auto Hematology Analyzer. A CUTE 5-part BC-5000 Auto Hematology Analyzer A CUTE 5-part Why do we need 5-part hematology analyzers? WBC differential: 3-part WBC differential: 5-part Lymphocyte Mid-size cell? Eosinophil Basophil Neutrophil DIFF

More information

Scholars Journal of Applied Medical Sciences (SJAMS)

Scholars Journal of Applied Medical Sciences (SJAMS) Scholars Journal of Applied Medical Sciences (SJAMS) Abbreviated Key Title: Sch. J. App. Med. Sci. Scholars Academic and Scientific Publisher A Unit of Scholars Academic and Scientific Society, India www.saspublisher.com

More information

WEQAS: what is it and what does it mean?. Celia Critchley Point of Care Co-ordinator Warrington and Halton Hospitals NHS Foundation Trust

WEQAS: what is it and what does it mean?. Celia Critchley Point of Care Co-ordinator Warrington and Halton Hospitals NHS Foundation Trust WEQAS: what is it and what does it mean?. Celia Critchley Point of Care Co-ordinator Warrington and Halton Hospitals NHS Foundation Trust Objectives Overview of External Quality Assessment and its relevance

More information

CASE-STUDY- VALIDATION of PCR based methodology. Beata Surmacz-Cordle Senior Analytical Development Scientist

CASE-STUDY- VALIDATION of PCR based methodology. Beata Surmacz-Cordle Senior Analytical Development Scientist CASE-STUDY- VALIDATION of PCR based methodology Beata Surmacz-Cordle Senior Analytical Development Scientist UK RMP Pluripotent Stem Cell Platform Validation Workshop 2 nd June 2016 RT-qPCR assay for detection

More information

See Brilliant Results

See Brilliant Results See Brilliant Results without all the reviews. * In Development Product not available in the U.S. First Pass Efficiency. Getting it right the first time. GOALS: More reportable WBC and WBC differential

More information

BC Auto Hematology Analyzer. A CUTE 5-part

BC Auto Hematology Analyzer. A CUTE 5-part BC-5150 Auto Hematology Analyzer A CUTE 5-part Why do we need 5-part hematology analyzers? WBC differential: 3-part WBC differential: 5-part Lymphocyte Mid-size cell? Eosinophil Basophil Neutrophil DIFF

More information

- - - Backup_Lab_Guidelines v

- - - Backup_Lab_Guidelines v - - - Backup_Lab_Guidelines v2.0 2010-11-10 Table of Contents Introduction... 1 Back-Up Guidelines for Safety Testing Laboratories... 2 Introduction... 2 Development and review of the back-up plan... 2

More information