Joint Committee Issue Paper

Size: px
Start display at page:

Download "Joint Committee Issue Paper"

Transcription

1 NSF Standard(s) Impacted: Background: Provide a brief background statement indicating the cause and nature of concern, the impacts identified relevant to public health, public understanding, etc, and any other reason why the issue should be considered by the Committee. Reference as appropriate any specific section(s) of the standard(s) that are related to the issue. Currently NSF/ANSI 49 only allows HEPA/ULPA filter leak testing using an aerosol photometer and laskin nozzle aerosol generator. The European Standard for Biosafety cabinets (EN12469:2000) allows the use of either an aerosol photometer or discrete particle counter for filter testing. HEPA/ULPA filter leak testing with either an aerosol photometer or discrete particle counter have long been accepted in cleanroom testing as documented in ISO :2005 addressing Test Methods for Cleanrooms and associated controlled environments and IEST-RP-CC034.1 HEPA and ULPA Filter Leak Tests. Each method has its advantages and disadvantages, but both are commonly used in cleanroom and Biosafety cabinet applications globally. Adding the use of a discrete particle counter to the accepted methods for HEPA/ULPA filter leak testing for Class II biological safety cabinets in NSF/ANSI 49 would bring the standard into more alignment with other international standards. In addition, it could make the field certification of Biosafety cabinets less costly in regions where there are few BSCs or certification. Many of the companies or individuals servicing cleanrooms and clean air equipment already have a discrete particle counter to classify air cleanliness and to leak test filters in cleanrooms and other equipment. Requiring the purchase and maintenance of a costly piece of test equipment simply because NSF has not considered an accepted methodology hinders public safety by limiting the growth of proper field certification. Recommendation: Clearly state what action is needed: e.g., recommended changes to the standard(s) including the current text of the relevant section(s) indicating deletions by use of strike-out and additions by highlighting or underlining; e.g., reference of the issue to a Task Group for detailed consideration; etc. Using the recognized standards, for the consideration of the Joint Committee we have adapted the current language in Annex F to allow the use of a discrete particle counter to test HEPA/ULPA filters in a Class II BSC. F.5 HEPA/ULPA filter leak test (from NSF/ANSI ) F.5.1 Purpose This test determines the integrity of downflow and exhaust HEPA/ULPA filters, filter housings, and filter mounting frames. The cabinet shall be operated within ± 5 ft/min (0.025 m/s) of the nominal set point, with the exception of the downflow HEPA/ULPA filters on B1 cabinets. F.5.2 Apparatus

2 There are two approved methods for HEPA/ULPA filter leak testing; the aerosol photometer method and the discrete particle counter (DPC) method. The following devices may be used to carry out HEPA/ULPA filter leak testing: F For the aerosol photometer method, an aerosol photometer with linear or expanded logarithmic scale shall be used. The instrument shall be capable of indicating 100% upstream concentration with a minimum aerosol concentration of 10 μg/l of polydisperse dioctylphthalate (DOP) particles, or an equivalent fluid that provides the same particle size distribution (e.g., polyalpha olefin [PAO] di[2- ethylhexyl], sebecate, polyethylene glycol, and medicinal-grade light mineral oil)41 produced by the generator described in Annex A, section A or equivalent. It shall also be capable of detecting an aerosol concentration in the downstream equal to 10-5 of the upstream concentration of the same particles. The sampling rate of air shall be 1 ft3/min (5 x 10-4 m3/s) ± 10%. Probe area shall have a maximum open area of 1.7 in2 (11 cm2) and a minimum dimension of 0.50 in (1.3 cm). The photometer shall be set up in accordance with the photometer manufacturer's instructions or IEST-RP-CC-013 if instructions are not provided. F For the aerosol photometer method, an aerosol generator of the Laskin Nozzle type conforming to Annex A, figure A2 or equivalent shall be used to create an aerosol by flowing air through liquid DOP or equivalent substitute. When a Laskin nozzle generator is used, the compressed air supplied to the generator should be adjusted to a minimum of 20 psi (140 kpa), if using DOP or 23 psi (160 kpa) if using PAO, measured at the generator manufacturer s recommended location. The nozzles shall be covered with liquid to a depth not to exceed 1.25 in (31 mm). F For the aerosol photometer method, a pressure gauge for the generator having a maximum range of 0 to 80 psi (0 to 550 kpa) with resolution and accuracy of 1 psi (7 kpa) calibrated by the manufacturer or in accordance with the manufacturer's instructions shall be used. F For the DPC method, a discrete particle counter with the capability to detect particles 0.3 micrometers in diameter and greater and

3 compliant with ISO The discrete particle counter shall have an audible alarm that beeps for each particle that is counted and should have a counting mode that is continuous or at least as long as the time required to scan a filter. The sampling rate of air shall be 1 ft 3 /min (5 x 10-4 m3/s) ± 10%. Probe area shall have a maximum open area of 1.7 in 2 (11 cm 2 ) and a minimum dimension of 0.50 inches (1.3 cm). The particle counter shall be set up in accordance with the manufacturer's instructions. F For the DPC method, a suitable dilution system allowing the measurement of the upstream concentration. Given the challenge concentration specification of 12 to 24 million particles per ft 3 (420 to 840 million particles per cubic meter), this typically requires a dilution system of 1:100 or greater. F For the DPC method, a suitable challenge aerosol, detectable by the DPC and in the concentration 12 to 24 million particles per ft 3 (420 to 840 million particles per cubic meter) distributed uniformly upstream of the filter to be tested. F.5.3 Method of testing HEPA/ULPA filters F Filters that can be scanned with an aerosol photometer a) Turn on the cabinet blower and lights (types A1 and A2 and B2 downflow filter test). Remove the filter diffusers and protective covers if any are present. Place the generator so the aerosol is introduced into each cabinet fan upstream of the HEPA/ULPA filter(s). When the manufacturer has not identified the aerosol introduction point(s), introduce the aerosol in a manner to ensure thorough mixing in the cabinet airflow. For example, a T- connection can be fitted to the aerosol generator output to enable distribution of challenge into both entrances of a single blower or entrances of multiple blowers. The manufacturer shall determine the aerosol introduction point that provides the most uniform distribution. b) Turn on the photometer and adjust it in accordance with the manufacturer's instructions.

4 c) Determine the aerosol concentration upstream of the HEPA/ULPA filter. When the challenged airflow is not contaminated, sample the aerosol concentration upstream of the HEPA/ULPA filter. When the challenged airflow is contaminated or when measuring the upstream concentration is not practical, the upstream concentration can be calculated. For example, when DOP is used as the challenge aerosol with a Laskin nozzle aerosol generator at 20 psi (140 kpa), the following formula applies: μg/l = 13,500 x number of nozzles / ft3/min of challenged air NOTE Use of DOP substitutes will require modification of this formula, unless the photometer is calibrated with the substitutes to yield results equivalent to those of DOP. Use of DOP substitutes will also require pressures different from 20 psig. Use an aerosol concentration that is at least equal to the photometric equivalent of 10 μg/l of DOP. d) Set up the photometer to the upstream challenge in accordance with the photometer manufacturer's instructions to detect leaks greater than or equal to 0.01% of the upstream concentration. e) With the nozzle of the probe held not more than 1.0 in (2.5 cm) from the area being tested, scan the entire downstream side of the HEPA/ULPA filter(s) and the perimeter of each filter pack by passing the photometer probe in slightly overlapping strokes at a traverse rate of not more than 2 in/s (5 cm/s). Separate passes shall be made around the entire periphery of the filter, along the bond between the filter pack and frame, and around the seal between the filter and the device. f) Reported values shall be: upstream aerosol challenge concentration;

5 method used to report concentration (measured or calculated); maximum leak penetration in percent; method used (scanned or probe tested); and name of test (HEPA/ULPA filter leak test). F Filters that cannot be scanned with an aerosol photometer a) When a cabinet is ducted so that the exhaust filter cannot be scanned, it may be leak tested by drilling a hole approximately 0.3 inches (1 cm) in diameter in the duct at a downstream location that will produce a well-mixed aerosol and inserting the photometer sampling probe with rigid extension tubing through the hole. b) Reported values shall be: upstream aerosol challenge concentration; method used to report concentration (measured or calculated); maximum leak penetration in percent; method used (scanned or probe tested); and name of test (HEPA/ULPA filter leak test). F Filters that can be scanned with a discrete particle counter a) Turn on the cabinet blower and lights (types A1 and A2 and B2 downflow filter test). Remove the filter diffusers and protective covers if any are present. Place the generator so the aerosol is introduced into each cabinet fan upstream of the HEPA/ULPA filter(s). When the manufacturer has not identified the aerosol introduction point(s), introduce the aerosol in a manner to ensure thorough mixing in the cabinet airflow. For example, a T- connection can be fitted to the aerosol generator output to enable distribution of challenge into both entrances of a single blower or entrances of multiple blowers. The manufacturer shall determine the aerosol introduction point that provides the most uniform distribution.

6 b) Turn on the particle counter and adjust it in accordance with the manufacturer's instructions. c) Determine the particulate concentration upstream of the HEPA/ULPA filter. When the challenged airflow is not contaminated, sample the particle concentration upstream of the HEPA/ULPA filter using a suitable dilution system to verify a concentration of detectable particles. When the challenged airflow is contaminated or when measuring the upstream concentration is not practical, the upstream concentration can be calculated from a previous measurement of volumetric flow and particulate challenge concentration using the same generator and particle counter. For example, if C 1 (particles per ft 3 or m 3 ) was measured in a flow of Q 1 (cfm or m 3 /s), the concentration of C 2 in a flow of Q 2 (cfm or m 3 /s) can be calculated with the following formula: C 2 = C 1 Q 1 Q 2 Use a detectable particle concentration that is 12 to 24 million particles per ft 3 (420 to 840 million particles per cubic meter). NOTE: The DPC test parameters identified are derived from ISO Test Methods, Section B.6.3 Procedure for installed filter system leakage scan test with a DPC. S r C c P L 472 D p N p Note: 472 above from ISO represents the sample flow rate of 472 cm3/second or 1 cfm Using the formula for the scan rate above from ISO and IEST-RP-CC034 and inserting the recommended values specified for challenge concentration, probe width and scan rate, we can relate N p (the number of particles characterizing a leak while scanning) directly to the standard leak penetration and upstream challenge concentration.

7 S r = Scan rate shall be 2 inches/sec (5 cm/sec) C c = Challenge aerosol concentration upstream of filter in particles/cm 3 shall be 12 to 24 million particles per ft 3 (420 to 840 million particles per cubic meter) which yields a minimum concentration of 420 particles/cm 3 P L = standard leak penetration of filter to be tested. The leak criteria for the photometer test and for a BSC filter leak test using a particle counter as specified in EN12469 is 0.01% or D p = probe dimension parallel to the scan direction in cm. Our specification provides a minimum dimension of 0.5 in (1.3 cm). N p = expected number of particle counts which characterize the designated leak while scanning which shall be 5. S r C c P L 472 D p N p 5 cm/sec 420 particles/cm cm3/sec 1.3 cm 5 particles d) Set up the particle counter in accordance with the manufacturer's instructions to detect particles of the required size (0.3 micrometers) and larger. e) With the nozzle of the probe held not more than 1.0 inch (2.5 cm) from the area being tested, scan the entire downstream side of the HEPA/ULPA filter(s) and the perimeter of each filter pack by passing the photometer probe in slightly overlapping strokes at a traverse rate of not more than 2 inches/second (5 cm/s). Separate passes shall be made around the entire periphery of the filter, along the bond between the filter pack and frame, and around the seal between the filter and the device. In environments with a high background of airborne particles care must be taken to assure accurate results. A potential leak is identified by an observed particle count N p of 5 particles during the time the probe passes over the defect.

8 f) When a potential leak is identified by a particle count during scanning, the potential leak should be verified by a remeasurement with the probe stationary over the suspect location. While scanning, any abrupt increase in particle counts should be cause for investigation. With the minimum concentrations specified above, a potential leak would be verified with a particle count of over 290 in a 15 second interval. NOTE: The DPC test parameters identified are derived from ISO Test Methods, Section B.6.3 Procedure for installed filter system leakage scan test with a DPC. N pa = C c P L 472 T r N pa = Actual number of particle counts which characterize the designated leak T r = Sustained residence time in seconds. Using the formulas from above and inserting the values specified, we derive our leak verification value of particles cm3 N pa = C c P L 472 T r f) Reported values shall be: cm3 15 seconds sec upstream challenge particle concentration; method used to report concentration (measured or calculated); maximum leak penetration in percent; method used (scanned or probe tested); and name of test (HEPA/ULPA filter leak test). F Filters that cannot be scanned with a discrete particle counter a) When a cabinet is ducted so that the exhaust filter cannot be scanned, it may be leak tested by drilling a hole approximately 0.3

9 inches (1 cm) in diameter in the duct at a downstream location that will produce a well-mixed aerosol and inserting the particle counter sampling probe with rigid extension tubing through the hole. b) With the minimum concentrations specified above, a potential leak of 0.005% would be verified with a particle count of over 145 in a 15 second interval. NOTE: The DPC test parameters identified are derived from ISO Test Methods, Section B.6.3 Procedure for installed filter system leakage scan test with a DPC. N pa = Actual number of particle counts which characterize the designated leak T r = Sustained residence time in seconds. Using the formulas from above and inserting the values specified, we derive our leak verification value of particles cm3 N pa = C c P L 472 T r c) Reported values shall be: cm3 15 seconds sec upstream challenge particle concentration; method used to report concentration (measured or calculated); maximum leak penetration in percent; method used (scanned or probe tested); and name of test (HEPA/ULPA filter leak test). F.5.4 Acceptance F Filters that can be scanned Sustained aerosol or particle penetration shall not exceed 0.01% of the upstream concentration at any point. F Filters that cannot be scanned

10 Sustained aerosol or particle penetration shall not exceed 0.005% of the upstream concentration. Supplementary Materials (photographs, diagrams, reports, etc.): If not provided electronically, the submitter will be responsible to have sufficient copies to distribute to committee members. I hereby grant NSF International the non-exclusive, royalty free rights, including non-exclusive, royalty free rights in copyright; in this item and I understand that I acquire no rights in any publication of NSF International in which this item in this or another similar or analogous form is used. Signature*: David S. Phillips Company: Thermo Fisher Scientific Telephone Number: david.phillips@thermofisher.com Submission Date: 5/5/2016 Please submit to: Al Rose, arose@nsf.org *Type written name will suffice as signature

AUTOMATED HEPA/ULPA FILTER SCANNING SYSTEM FOR LEAK DETECTION AND FILTER PERFORMANCE TESTING

AUTOMATED HEPA/ULPA FILTER SCANNING SYSTEM FOR LEAK DETECTION AND FILTER PERFORMANCE TESTING AUTOMATED HEPA/ULPA FILTER SCANNING SYSTEM FOR LEAK DETECTION AND FILTER PERFORMANCE TESTING Francisco J. Romay and James J. Sun, MSP Corporation, 1313 Fifth St. SE, Minneapolis, MN, 55414 and Benjamin

More information

NBC SPECIFICATIONS FOR CLEAN ROOM CERTIFICATION TESTING

NBC SPECIFICATIONS FOR CLEAN ROOM CERTIFICATION TESTING The Narragansett Bay Commission Vincent J. Mesolella Corporate Office Building Chairman One Service Road Providence, RI 02905 Raymond J. Marshall, P.E. 401 461 8848 Executive Director 401 461 6540 FAX

More information

Future Directions and Historical Development of NSF/ANSI 49 Biosafety Cabinetry: Design, Construction, Performance and Field Certification

Future Directions and Historical Development of NSF/ANSI 49 Biosafety Cabinetry: Design, Construction, Performance and Field Certification Future Directions and Historical Development of NSF/ANSI 49 Biosafety Cabinetry: Design, Construction, Performance and Field Certification David S. Phillips, DM The world leader in serving science NSF/ANSI

More information

THE WORLD LEADER IN CLEAN AIR SOLUTIONS. Pharmaceutical Clean Air Solutions PARTICULATE AND GASEOUS FILTRATION

THE WORLD LEADER IN CLEAN AIR SOLUTIONS. Pharmaceutical Clean Air Solutions PARTICULATE AND GASEOUS FILTRATION THE WORLD LEADER IN CLEAN AIR SOLUTIONS Pharmaceutical Clean Air Solutions PARTICULATE AND GASEOUS FILTRATION AAF Flanders has an in-depth understanding of the challenges and opportunities for pharmaceutical

More information

PAO Compatible eptfe Technology HEPA Filters for Cleanroom Pharmaceutical Applications

PAO Compatible eptfe Technology HEPA Filters for Cleanroom Pharmaceutical Applications MARCH 2016 WHITE PAPER PAO Compatible Technology HEPA Filters for Cleanroom Pharmaceutical Applications Rahul Bharadwaj, Nathaniel Nance and Michael Osborne AAF International Louisville KY, USA Abstract

More information

Particle Sample Tube Lengths for Pharmaceutical Monitoring. Version 01. Rev 01.

Particle Sample Tube Lengths for Pharmaceutical Monitoring. Version 01. Rev 01. T E C H N I C A L P A P E R Particle Sample Tube Lengths for Pharmaceutical Monitoring Version 01. Rev 01. Particle Sample Tube Lengths for Pharmaceutical Monitoring Lighthouse Worldwide Solutions Airborne

More information

Whitepaper HVAC Pre-filtration

Whitepaper HVAC Pre-filtration Whitepaper HVAC Pre-filtration In Pharmaceutical Facilities I N D U S T R Y I N S I G H T S Norman A. Goldschmidt September 11, 2011 Principal, Engineering www.geieng.com I N D U S T R Y I N S I G H T

More information

This document is a preview generated by EVS

This document is a preview generated by EVS INTERNATIONAL STANDARD ISO 16170 First edition 2016-07-01 Corrected version 2017-04 In situ test methods for high efficiency filter systems in industrial facilities Méthodes d essai in situ pour les systèmes

More information

This document is a preview generated by EVS

This document is a preview generated by EVS INTERNATIONAL STANDARD ISO 16890-2 First edition 2016-12-01 Air filters for general ventilation Part 2: Measurement of fractional efficiency and air flow resistance Filtres à air de ventilation générale

More information

Indoor Air Quality Solutions

Indoor Air Quality Solutions Indoor Air Quality Solutions Ultimate Solutions for Indoor Air Quality Investigation Demand for quality indoor air continues to increase as we are spending more time indoors than ever before. Indoor air

More information

Airborne Particle Detection

Airborne Particle Detection Airborne Particle Detection Detects airborne particles from 0.3 to 5 µm Monitor 6 particle sizes simultaneously Choose from two models: Pro for normal environments CR for cleanrooms and clean areas Data

More information

Supplementary Training Modules on Good Manufacturing Practice. Validation. WHO Technical Report Series, No. 937, Annex 4.

Supplementary Training Modules on Good Manufacturing Practice. Validation. WHO Technical Report Series, No. 937, Annex 4. Supplementary Training Modules on Good Manufacturing Practice Validation WHO Technical Report Series, No. 937, 2006. Annex 4. Validation Slide 1 of 48 August 2006 Validation! Part 1. General overview on

More information

MODEL 3340 LASER AEROSOL SPECTROMETER

MODEL 3340 LASER AEROSOL SPECTROMETER MODEL 3340 LASER AEROSOL SPECTROMETER Application ote 3340-001 Use of the TSI LAS Model 3340 for HEPA Filter Testing Hans-Georg Horn TSI Incorporated Basic Setup for Filter Testing Measuring filter efficiency

More information

The benefits of expanded polytetrafluoroethylene

The benefits of expanded polytetrafluoroethylene Reprinted from PHARMACEUTICAL ENGINEERING The Official Magazine of ISPE November/December 2011, Vol. 31 No. 6 www.ispe.org Copyright ISPE 2011 This article presents the study results of using an ultra

More information

Test Report. Room: Production Room. Serial Number Air Particle Counter Aerosol Generator Sound Level Meter + Calibrator Termohygrometer

Test Report. Room: Production Room. Serial Number Air Particle Counter Aerosol Generator Sound Level Meter + Calibrator Termohygrometer Report Number: 37AR17 Report Date: 14-11-2017 250-351 Amadora - Portugal Department: Phone: 21 47 77 70 Room: Production Room Room Conditions: Resting Number of people: 1 Room Area (m²): 10.4 Equipment

More information

Steven L. Alderman, Michael S. Parsons, and Charles A. Waggoner. Institute for Clean Energy Technology, Mississippi State University

Steven L. Alderman, Michael S. Parsons, and Charles A. Waggoner. Institute for Clean Energy Technology, Mississippi State University The Effects of Media Velocity and Particle Size Distribution on Most Penetrating Particle Size and Filter Loading Capacity of 12 x12 x11.5 AG 1 HEPA Filters Steven L. Alderman, Michael S. Parsons, and

More information

STANDARDS FOR COMPOUNDING PHARMACIES

STANDARDS FOR COMPOUNDING PHARMACIES STANDARDS FOR COMPOUNDING PHARMACIES APPLICATION NOTE LC-139 (US) Introduction This publication provides excerpts from some of the many guidelines and standards that pertain to compounding pharmacies.

More information

B. ASHRAE 1991 HVAC Applications Handbook: Chapter 34--Testing, Adjusting and Balancing.

B. ASHRAE 1991 HVAC Applications Handbook: Chapter 34--Testing, Adjusting and Balancing. PART 1 GENERAL 1.1 RELATED DOCUMENTS SECTION 230593 TESTING, ADJUSTING, AND BALANCING A. Drawings and general provisions of Contract, including General and Supplementary Conditions and Division 1 Specification

More information

VALIDATION AS A KEY TOOL IN SUCCESSFUL TRANSFER OR ESTABLISHMENT OF A MEDICAL DEVICE COMPANY IN A NEW FACILITY

VALIDATION AS A KEY TOOL IN SUCCESSFUL TRANSFER OR ESTABLISHMENT OF A MEDICAL DEVICE COMPANY IN A NEW FACILITY VALIDATION AS A KEY TOOL IN SUCCESSFUL TRANSFER OR ESTABLISHMENT OF A MEDICAL DEVICE COMPANY IN A NEW FACILITY Oren Levin, Diane Gordon VisionCare Ophthalmic Technologies, POB 3879, Petah Tikva 49130,

More information

F-Series 1 Elements Performance Validation

F-Series 1 Elements Performance Validation F-Series 1 Elements Performance Validation An independent test report validating the performance of the nano F-Series 1 Compressed Air & Gas Filter Elements to ISO 12500-1:2007 air quality standards. Conducted

More information

BIOTRAK REAL-TIME VIABLE PARTICLE COUNTER ROOT CAUSE INVESTIGATION GUIDANCE

BIOTRAK REAL-TIME VIABLE PARTICLE COUNTER ROOT CAUSE INVESTIGATION GUIDANCE BIOTRAK REAL-TIME VIABLE PARTICLE COUNTER ROOT CAUSE INVESTIGATION GUIDANCE APPLICATION NOTE CC-123 (US) Introduction This guidance provides recommendations for how the BioTrak Real-Time Viable Particle

More information

METHOD 1A - SAMPLE AND VELOCITY TRAVERSES FOR STATIONARY SOURCES WITH SMALL STACKS OR DUCTS. NOTE: This method does not include all of the

METHOD 1A - SAMPLE AND VELOCITY TRAVERSES FOR STATIONARY SOURCES WITH SMALL STACKS OR DUCTS. NOTE: This method does not include all of the 206 METHOD 1A - SAMPLE AND VELOCITY TRAVERSES FOR STATIONARY SOURCES WITH SMALL STACKS OR DUCTS NOTE: This method does not include all of the specifications (e.g., equipment and supplies) and procedures

More information

Assessment of Containment and Product Protection In a Class II Type B3 Biological Safety Cabinet Under As-used Laboratory Conditions

Assessment of Containment and Product Protection In a Class II Type B3 Biological Safety Cabinet Under As-used Laboratory Conditions Assessment of Containment and Product Protection In a Class II Type B3 Biological Safety Cabinet Under As-used Laboratory Conditions Peter Harris and Dave Lupo B & V Testing, Inc. ABSA, 1999 Study Objective

More information

To: PMP Working Group From: Chris Parkin

To: PMP Working Group From: Chris Parkin To: PMP Working Group From: Chris Parkin MINUTES OF PMP 8th & 9th OCTOBER MEETINGS Cleaner Fuels & Vehicles 4 Location: Tel: +44 20 7944 2958 Fax: +44 20 7944 2605 Ref: (Paper) Date: 15 October 2007 Copies:

More information

TEST REPORT. Purpose Personal Home Sterilizer in Each Use. January 2005

TEST REPORT. Purpose Personal Home Sterilizer in Each Use. January 2005 TEST REPORT Measurement of Ozone (O 3 ) Leakage of Multi- Purpose Personal Home Sterilizer in Each Use January 2005 Air Resources Research Center Korea Institute of Science and Technology (KIST) KLENZ

More information

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 14644-2 First edition 2000-09-15 Cleanrooms and associated controlled environments Part 2: Specifications for testing and monitoring to prove continued compliance with ISO 14644-1

More information

COMPRESSED AIR SYSTEM DESIGNS BEGIN WITH THE DEMANDS

COMPRESSED AIR SYSTEM DESIGNS BEGIN WITH THE DEMANDS COMPRESSED AIR SYSTEM DESIGNS BEGIN WITH THE DEMANDS BY: BILL SCALES, P.E. SCALES INDUSTRIAL TECHNOLOGIES, INC. The purpose of an assessment of an existing compressed air system, or design of a new one,

More information

TEST REPORT SCOPE: EMISSIONS, EFFICIENCY AND OUTPUT FUEL: EPA TEST FUEL (CRIBS) TEST STANDARD: EPA MODEL: SOLUTION 2.

TEST REPORT SCOPE: EMISSIONS, EFFICIENCY AND OUTPUT FUEL: EPA TEST FUEL (CRIBS) TEST STANDARD: EPA MODEL: SOLUTION 2. TEST REPORT SCOPE: EMISSIONS, EFFICIENCY AND OUTPUT FUEL: EPA TEST FUEL (CRIBS) TEST STANDARD: EPA MODEL: SOLUTION 2.3-I WOOD INSERT Notice to reader: Our Solution 2.3-I wood insert was tested as part

More information

CHAPTER 6 : GAS SAMPLING SYSTEMS

CHAPTER 6 : GAS SAMPLING SYSTEMS CHAPTER 6 : GAS SAMPLING SYSTEMS 1 Scope : 1.1 This Chapter describes two types of gas sampling systems in paragraphs 2.1 and 2.2 meeting the requirements specified in para 4.2 of Chapter 3 of this Part.

More information

24th DOE/NRC NUCLEAR AIR CLEANING AND TREATMENT CONFERENCE

24th DOE/NRC NUCLEAR AIR CLEANING AND TREATMENT CONFERENCE 24th DOE/NRC NUCLEAR AIR CLEANING AND TREATMENT CONFERENCE A GENERAL CORRELATION OF MPPS PENETRATION AS A FUNCTION OF FACE VELOCITY WITH THE MODEL 8140 USING THE CERTITEST 8160 Norman Lifshutz and Mary

More information

ISO INTERNATIONAL STANDARD. Cleanrooms and associated controlled environments Part 1: Classification of air cleanliness

ISO INTERNATIONAL STANDARD. Cleanrooms and associated controlled environments Part 1: Classification of air cleanliness INTERNATIONAL STANDARD ISO 14644-1 First edition 1999-05-01 Cleanrooms and associated controlled environments Part 1: Classification of air cleanliness Salles propres et environnements maîtrisés apparentés

More information

SPECIFICATION. Ir-192 hot cell

SPECIFICATION. Ir-192 hot cell SPECIFICATION Ir-192 hot cell 1. Scope The purpose of this document is to define the specifications for the construction of a class C hot cell with two chambers (hereinafter referred as the System ) intended

More information

Code for Construction and Acceptance of Cleanroom

Code for Construction and Acceptance of Cleanroom UDC NATIONAL STANDARD OF THE PEOPLE'S REPUBLIC OF CHINA p GB 50591-201() Code for Construction and Acceptance of Cleanroom Issued on: July 15, 2010 Implemented on: February 1, 2011 Jointly Issued by Ministry

More information

Efficacy of Novaerus NV 1050 System against NO 2 and Formaldehyde Gases

Efficacy of Novaerus NV 1050 System against NO 2 and Formaldehyde Gases Efficacy of Novaerus NV 1050 System against NO 2 and Formaldehyde Gases Jamie Balarashti a, Dan Merchant a, Zach Conley a a Aerosol Research and Engineering Laboratories Inc. Olathe KS Background: This

More information

Air Purification Fine Dust - HEPA/ULPA Sterionizer - Ion Technology

Air Purification Fine Dust - HEPA/ULPA Sterionizer - Ion Technology Air Purification Fine Dust - HEPA/ULPA Sterionizer - Ion Technology Fine Dust Filter Final Filter for Fan Filter Units Final Filter for Clean Rooms Final Filter for Channel Installations Highly Active

More information

Indoor Air Quality and Mechanical Ventilation (Page 1 of 5)

Indoor Air Quality and Mechanical Ventilation (Page 1 of 5) Indoor Air Quality and Mechanical Ventilation (Page 1 of 5) Title 24, Part 6, Section 150.0(o) Ventilation for Indoor Air Quality. All dwelling units shall meet the requirements of ANSI/ASHRAE Standard62.2.

More information

Blooming prospects. Ways to improve your painting results

Blooming prospects. Ways to improve your painting results Blooming prospects Ways to improve your painting results Reduce paint damage caused by dust particles but how? Excellent paint jobs are a reason to be happy. On the contrary, paint damage is aggravating.

More information

Capital Projects Group Air Filters and Accessories Specification

Capital Projects Group Air Filters and Accessories Specification Capital Projects Group Air Filters and Accessories Specification Specification 23 41 00 Revision 0 Date: August 2018 Air Filters and Accessories Specification Specification 23 41 00 Publication Date: August

More information

The Best Recipe for Germ-Free Air: Viledon Filter Systems for Cleanroom Technology

The Best Recipe for Germ-Free Air: Viledon Filter Systems for Cleanroom Technology The Best Recipe for Germ-Free Air: Viledon Filter Systems for Cleanroom Technology Where every single germ counts... The pharmaceutical industry bears an unusually high degree of responsibility to both

More information

CASE STUDIES OF DUST AMS CALIBRATION AT EXTREME CONDITIONS

CASE STUDIES OF DUST AMS CALIBRATION AT EXTREME CONDITIONS CASE STUDIES OF DUST AMS CALIBRATION AT EXTREME CONDITIONS Jurij ČRETNIK RACI d.o.o., Tehnološki park 24, SI-1000 LJUBLJANA, SLOVENIA jurij.cretnik@raci.si Matjaž ŠLIBAR RACI d.o.o., Tehnološki park 24,

More information

Specific Accreditation Criteria Materials ISO/IEC Annex. Cement testing

Specific Accreditation Criteria Materials ISO/IEC Annex. Cement testing Specific Accreditation Criteria Materials ISO/IEC 17025 Annex Cement testing January 2018 Copyright National Association of Testing Authorities, Australia 2013 This publication is protected by copyright

More information

IN-PLACE HEPA FILTER PENETRATION TEST*

IN-PLACE HEPA FILTER PENETRATION TEST* IN-PLACE HEPA FILTER PENETRATION TEST* W. Bergman, K. Wilson, J. Elliott, B. Bettencourt and J. W. Slawskil Lawrence Livermore National Laboratory Livermore, CA 94550 Abstract We have demonstrated the

More information

SECTION HVAC AIR CLEANING DEVICES

SECTION HVAC AIR CLEANING DEVICES SECTION 23 40 00 PART 1 - GENERAL 1.1 DESCRIPTION SPEC WRITER NOTES: 1. Use this section only for NCA projects. 2. Delete between // // if not applicable to project. Also delete any other item or paragraph

More information

Clean air solutions CAMSAFE 2

Clean air solutions CAMSAFE 2 Clean air solutions CAMSAE 2 CamSafe 2 CAMIL MORE THAN A CLEAN AIR SOLUTIONS PROVIDER CAMSAE 2 NOT ONLY A TIGHT HOUSING Camfil is a world leader in clean air technology and air filter production. Our organization

More information

ITEM 431 PNEUMATICALLY PLACED CONCRETE. All material shall conform to the pertinent requirements of the following items:

ITEM 431 PNEUMATICALLY PLACED CONCRETE. All material shall conform to the pertinent requirements of the following items: ITEM 431 PNEUMATICALLY PLACED CONCRETE 431.1. Description. This Item shall govern for furnishing and placing of "Pneumatically Placed Concrete" for encasement of designated structural steel members, the

More information

PARTICULATE. Massachusetts. We are pleased to announce that we have expanded to THE. Our Services Testing & Certification

PARTICULATE. Massachusetts. We are pleased to announce that we have expanded to THE. Our Services Testing & Certification S U M M E R 2 0 1 2, V O L U M E 3. THE PARTICULATE I N S I D E T H I S I S S U E : MCI ACHIEVEMENTS 1 We are pleased to announce that we have expanded to Massachusetts MEET & GREET 2 Demand for quality

More information

Louvers and Dampers April Product Guide Specification SECTION SMOKE DAMPERS

Louvers and Dampers April Product Guide Specification SECTION SMOKE DAMPERS Louvers and Dampers April 2008 P.O. Box 606 7435 Industrial Road Florence, Kentucky, 41042 Phone: (859) 647-2299 Fax: (859) 647-7810 Web Site: www.louvers-dampers.com Product Guide Specification SECTION

More information

A guideline to good manufacturing practice (GMP) requirements

A guideline to good manufacturing practice (GMP) requirements A guideline to good manufacturing practice (GMP) requirements Part 2: Validation Get all guidelines on www.pharmaguideline.com Email- info@pharmaguideline.com Page 1 of 96 1. Abbreviations EP: GMP: MF:

More information

INSTALLATION QUALIFICATIONPROTOCOL FOR DYNAMIC PASS BOX

INSTALLATION QUALIFICATIONPROTOCOL FOR DYNAMIC PASS BOX Page 1 of 10 INSTALLATION QUALIFICATIONPROTOCOL FOR DYNAMIC PASS BOX CUSTOMER: M/s. NATCO PHARMA LIMITED. K O T H U R EQUIPMENT: DYNAMIC PASS BOX (600 W x 600 D x 1000 H mm) SUBMITTED BY: P H ARMA ENGINEERS

More information

ISO INTERNATIONAL STANDARD. Cleanrooms and associated controlled environments Part 4: Design, construction and start-up

ISO INTERNATIONAL STANDARD. Cleanrooms and associated controlled environments Part 4: Design, construction and start-up INTERNATIONAL STANDARD ISO 14644-4 First edition 2001-04-01 Cleanrooms and associated controlled environments Part 4: Design, construction and start-up Salles propres et environnements maîtrisés apparentés

More information

Heating, ventilation, and air system encompasses heating, ventilation, and air conditioning, which is integral component of

Heating, ventilation, and air system encompasses heating, ventilation, and air conditioning, which is integral component of REVIEW ARTICLE Demand of pharmaceutical facility functionality: Validation and qualification of HVAC system Anamika Singh, Sapna Malviya, Anil Kharia Department of Quality Assurance of Modern Institute

More information

Ricardo-AEA. In-Situ Filter Testing & Stack Sampler Characterization. John Osbourn. Particle Measurement Centre.

Ricardo-AEA. In-Situ Filter Testing & Stack Sampler Characterization. John Osbourn. Particle Measurement Centre. Ricardo-AEA In-Situ Filter Testing & Stack Sampler Characterization John Osbourn Particle Measurement Centre www.ricardo-aea.com Outline In-Situ Filter Testing Stack Sampler Characterization Filtration

More information

SAMPLING ISSUES AND THE IN-PLACE LEAK TEST. John R. Hunt NCS Corporation. Abstract

SAMPLING ISSUES AND THE IN-PLACE LEAK TEST. John R. Hunt NCS Corporation. Abstract SAMPLING ISSUES AND THE IN-PLACE LEAK TEST John R. Hunt NCS Corporation Abstract Sampling manifolds have been accepted as a means of overcoming difficulties with non-homogeneity of the tracer-air stream

More information

NetfilTM. World-class Solutuions in. Proven Technology of High Temp HEPA Filters for Sterilizing Zone. Clean Room Equipment & Filters

NetfilTM. World-class Solutuions in. Proven Technology of High Temp HEPA Filters for Sterilizing Zone. Clean Room Equipment & Filters NetfilTM World-class Solutuions in Clean Room Equipment & Filters Proven Technology of High Temp HEPA Filters for Sterilizing Zone COMPANY PROFILE Netfil Technik Pvt. Ltd. is an organization with absolute

More information

Space Conditioning System Airflow Rate

Space Conditioning System Airflow Rate (Page 1 of 3) A. Ducted Cooling System Information System Identification or Name 02 System Location or Area Served 03 System Installation Type 04 Nominal Cooling Capacity (tons) of Condenser 05 Condenser

More information

Main Steps for Radiopharmaceuticals Hot Cells Validation in Accordance with GMP Requirements: Methodology and Practical Guide

Main Steps for Radiopharmaceuticals Hot Cells Validation in Accordance with GMP Requirements: Methodology and Practical Guide Journal of Environmental Science and Engineering A 7 (2018) 132-139 doi:10.17265/2162-5298/2018.03.004 D DAVID PUBLISHING Main Steps for Radiopharmaceuticals Hot Cells Validation in Accordance with GMP

More information

GENERAL TESTING AND REPORTING REQUIREMENTS

GENERAL TESTING AND REPORTING REQUIREMENTS GENERAL TESTING AND REPORTING REQUIREMENTS Ohio EPA requires that an Ohio Intent to Test ( ITT ) form be filed with the appropriate Ohio EPA representative at least 30 days (or more if required by regulation)

More information

2015 International Mechanical Code Errata (Portions of text and tables not shown are unaffected by the errata)

2015 International Mechanical Code Errata (Portions of text and tables not shown are unaffected by the errata) PIPING MATERIAL 1 ST PRINTING (September 1, 2015) CHAPTER 3 GENERAL REGULATIONS TABLE 305.4 PIPING SUPPORT SPACING a MAXIMUM HORIZONTAL SPACING (feet) MAXIMUM VERTICAL SPACING (feet) ABS pipe 4 10 c Aluminum

More information

INSPECTOR S OBSERVATIONS AND WARNING LETTER

INSPECTOR S OBSERVATIONS AND WARNING LETTER Microrite, Inc. San Jose, CA, USA, http://www.microrite.com AIR VELOCITY MEASUREMENTS AND CORRELATION TO SMOKE STUDIES FOR ASEPTIC OPERATIONS KEYWORDS Air Velocity, Air Flows, Smoke Study, Air Pattern

More information

When Quality Matters. Microbial Air Samplers: Selection Criteria in Biopharma Manufacturing. Celebrating Over 50 Years In Business!

When Quality Matters. Microbial Air Samplers: Selection Criteria in Biopharma Manufacturing. Celebrating Over 50 Years In Business! When Quality Matters Celebrating Over 50 Years In Business! Microbial Air Samplers: Selection Criteria in Biopharma Manufacturing Rev. 1.0 (Jan 2016) 1 Doc No.: TN-160104-1 Printed: January 4, 2016 Copyright

More information

METHOD BERYLLIUM SCREENING METHOD. Analyte CAS No. Sensitivity Beryllium (Be) Dependent upon analytical procedure used

METHOD BERYLLIUM SCREENING METHOD. Analyte CAS No. Sensitivity Beryllium (Be) Dependent upon analytical procedure used 1764 METHOD 103 - BERYLLIUM SCREENING METHOD 1.0 Scope and Application. 1.1 Analytes. Analyte CAS No. Sensitivity Beryllium (Be) 7440-41-7 Dependent upon analytical procedure used 1.2 Applicability. This

More information

Indoor Air Quality Tools

Indoor Air Quality Tools Indoor Air Quality Tools In response to the growing importance of air quality in buildings, workplaces and homes, Fluke offers you a range of tools that monitor temperature, humidity, air velocity, particulate,

More information

CSC SPECIFICATIONS DRY SHAKE CRYSTALLINE WATERPROOFING Section

CSC SPECIFICATIONS DRY SHAKE CRYSTALLINE WATERPROOFING Section DRY SHAKE CRYSTALLINE WATERPROOFING Section 03 35 14 This section includes Krystol Broadcast, a crystalline waterproofing powder applied to the surface of freshly placed concrete during the final stages

More information

CSI SPECIFICATIONS DRY SHAKE CRYSTALLINE WATERPROOFING Section

CSI SPECIFICATIONS DRY SHAKE CRYSTALLINE WATERPROOFING Section DRY SHAKE CRYSTALLINE WATERPROOFING Section 03 35 14 This section includes Krystol Broadcast, a crystalline waterproofing powder applied to the surface of freshly placed concrete during the final stages

More information

http://www.sis.se http://www.sis.se http://www.sis.se http://www.sis.se http://www.sis.se Provläsningsexemplar / Preview SWEDISH STANDARD SS 2680:2004 Approved 2004-10-08 Edition 2 Contamination control

More information

QUALITY ASSURANCE PROTOCOL (QAP)

QUALITY ASSURANCE PROTOCOL (QAP) PREFACE This Standard details the level of documentation that PVT.LTD (IET) will provide during a contract for fabricating tank(s) or portable equipment s. Extensive documentation is required for this

More information

Thermo Scientific Biological Safety Cabinets Product Selection Guide

Thermo Scientific Biological Safety Cabinets Product Selection Guide Thermo Scientific Biological Safety Cabinets Product Selection Guide New lab construction or renovations require significant time, resources and planning. Early in your planning process, it is important

More information

10 Thorough examination and test

10 Thorough examination and test 10 Thorough examination and test Key points Every employer s LEV system requires statutory thorough examination and testing by a competent person. The examination and testing report should have a prioritised

More information

Element B3 LEV Systems

Element B3 LEV Systems LEV Systems Ventilation Natural or Mechanical COSHH Extract Ventilation Important control strategy Exposure to dust, fumes, mists, vapours, gases LEV Remove contaminant at source Local Exhaust Ventilation

More information

Contact person Date Reference Page Christian Mossberg PX23030A 1 (2) Energy Technology

Contact person Date Reference Page Christian Mossberg PX23030A 1 (2) Energy Technology issued by an Accredited Testing Laboratory Contact person Christian Mossberg 2012-10-25 PX23030A 1 (2) Energy Technology +46 10 516 59 13 Christian.Mossberg@sp.se 1002 ISO/IEC 17025 Lindbergs Ventilation

More information

Designing Efficient Sub Slab Collection and Vapor Barrier Systems. Presented By: Tom Hatton Clean Vapor, LLC cleanvapor.com

Designing Efficient Sub Slab Collection and Vapor Barrier Systems. Presented By: Tom Hatton Clean Vapor, LLC cleanvapor.com Designing Efficient Sub Slab Collection and Vapor Barrier Systems Presented By: Tom Hatton Clean Vapor, LLC cleanvapor.com thatton@cleanvapor.com 1 Effective and Efficient Vent Systems Start in the Planning

More information

Aerosol Short Course: Physics, Measurement, and Sampling

Aerosol Short Course: Physics, Measurement, and Sampling Aerosol Short Course: Physics, Measurement, and Sampling Boeing Ed Wells Course Tom Peters The University of Iowa Department of Occupational and Environmental Health Iowa City, IA III. Sampling Overview

More information

INTERNATIONAL STANDARD

INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 16431 First edition 2012-01-15 Corrected version 2012-03-15 Hydraulic fluid power System cleanup procedures and verification of cleanliness of assembled systems Transmissions

More information

FIVE STAR PRODUCTS, INC. (800)

FIVE STAR PRODUCTS, INC.  (800) FIVE STAR PRODUCTS, INC. www.fivestarproducts.com (800) 243-2206 Page 1 of 11 DESIGN-A-SPEC GUIDELINES FIVE STAR RS ANCHOR GEL CONTENTS PART A - GENERAL CONDITIONS PART B - MATERIAL SPECIFICATIONS PART

More information

FLORIDA SOCIETY OF HEALTH SYSTEM PHARMACISTS NOVEMBER 4, 2016 JAMES T WAGNER CONTROLLED ENVIRONMENT CONSULTING

FLORIDA SOCIETY OF HEALTH SYSTEM PHARMACISTS NOVEMBER 4, 2016 JAMES T WAGNER CONTROLLED ENVIRONMENT CONSULTING Engineering Controls for Compliance with USP Chapters and FLORIDA SOCIETY OF HEALTH SYSTEM PHARMACISTS NOVEMBER 4, 2016 JAMES T WAGNER CONTROLLED ENVIRONMENT CONSULTING Learning and Performance

More information

CERTIFICATE OF FIELD VERIFICATION & DIAGNOSTIC TESTING (Page 1 of 8)

CERTIFICATE OF FIELD VERIFICATION & DIAGNOSTIC TESTING (Page 1 of 8) CERTIFICATE OF FIELD VERIFICATION & DIAGNOSTIC TESTING (Page 1 of 8) Builder or Installer Name Builder or Installer Contact Telephone Plan/Permit (Additions or Alterations) Number Compliance Method (Prescriptive)

More information

Gas Industry Standard

Gas Industry Standard Gas Industry Standard GIS/L2:2006 Specification for Steel pipe 21.3 mm to 1 219 mm outside diameter for operating pressures up to 7 bar (supplementary to BS EN 10208-1) Contents Foreword Mandatory and

More information

- 1 - Agenda Items. atmospheric dust the invisible cesspool. 1 micrometer = 1/25,400 of an inch. Fiber Size. Lab v Real life conditions

- 1 - Agenda Items. atmospheric dust the invisible cesspool. 1 micrometer = 1/25,400 of an inch. Fiber Size. Lab v Real life conditions Count.01 -.02.02 -.05.05 -.1.1 -.22.22 -.46.46-1 1-2.15 2.15-4.64 4.64-10 10-21.54 21.54 + Agenda Items Industry Hot Topics TCO-LCC - the Science based approach Update on Industry Standards Common Products

More information

ROAD COMMISSION FOR OAKLAND COUNTY SPECIAL PROVISION FOR QUALITY CONTROL AND ACCEPTANCE OF PORTLAND CEMENT CONCRETE

ROAD COMMISSION FOR OAKLAND COUNTY SPECIAL PROVISION FOR QUALITY CONTROL AND ACCEPTANCE OF PORTLAND CEMENT CONCRETE ROAD COMMISSION FOR OAKLAND COUNTY SPECIAL PROVISION FOR QUALITY CONTROL AND ACCEPTANCE OF PORTLAND CEMENT CONCRETE RCOC/CONST:SH/JO PAGE 1 OF 9 a. Description. This special provision establishes Portland

More information

Medical and Pharmaceutical Packaging Styles

Medical and Pharmaceutical Packaging Styles Aligned Test Methods and Sampling Plans for DuPont Tyvek Medical and Pharmaceutical Packaging Styles June 2017 Aligned Test Methods The Medical Packaging Transition Project (MPTP) includes manufacturing

More information

Method Implementation Document for BS EN * - Application to manual stack emissions monitoring

Method Implementation Document for BS EN * - Application to manual stack emissions monitoring Method Implementation Document for BS EN 16911-1* - Application to manual stack emissions monitoring *BS EN ISO 16911-1:2013 Stationary source emissions - Manual and automatic determination of velocity

More information

Packaging Operations Technical Subcommittee

Packaging Operations Technical Subcommittee and Sterile About the Presenter John Derek Thompson Lead Sterile Engineer, DePuy Synthes Certified Professional with the IOPP Member of the Medical Device Technical Committee Chairman of the Member of

More information

Monday 8 th July Symposium 1 - Environmental Monitoring of Production Areas

Monday 8 th July Symposium 1 - Environmental Monitoring of Production Areas Monday 8 th July 2013 Symposium 1 - Environmental Monitoring of Production Areas Dealing with Environmental Monitoring Out of Specification Results and Changes to ISO 14644 David Buckley Rev 2 1 Kaoru

More information

CLASS I MICROBIOLOGICAL SAFETY CABINET - BIO 1

CLASS I MICROBIOLOGICAL SAFETY CABINET - BIO 1 CLASS I MICROBIOLOGICAL SAFETY CABINET - BIO 1 PRODUCT SHEET NO. 101 VERSION 2.0 COMPACT HIGHLY SPECIFIED, HIGH PERFORMANCE CLASS I MICROBIOLOGICAL SAFETY CABINETS Applications Handling of biological agents

More information

SAMPLE SPECIFICATION for CROSSHOLE SONIC LOGGING (CSL) September 2015

SAMPLE SPECIFICATION for CROSSHOLE SONIC LOGGING (CSL) September 2015 SAMPLE SPECIFICATION for CROSSHOLE SONIC LOGGING (CSL) September 2015 Note: This sample specification contains recommended or typical quantities in parenthesis, in the format (quantity); the specifying

More information

CSI SPECIFICATIONS SURFACE-APPLIED CRYSTALLINE WATERPROOFING Section

CSI SPECIFICATIONS SURFACE-APPLIED CRYSTALLINE WATERPROOFING Section SURFACE-APPLIED CRYSTALLINE WATERPROOFING Section 07 16 16 This section includes Krystol T1 crystalline waterproofing, applied to new or existing concrete substrates, above or below grade, on either dry

More information

Section Asbestos Containment Procedures

Section Asbestos Containment Procedures Section Cover Page Section 02 82 05 2017-10-01 Use this Section where full containment will be required, or if a Contractor option, may be required, depending on the removal method(s) selected by the Contractor.

More information

Field Accuracy of Temperature Measurements in TAB Work Andrew P. Nolfo, PE

Field Accuracy of Temperature Measurements in TAB Work Andrew P. Nolfo, PE Field Accuracy of Temperature Measurements in TAB Work Andrew P. Nolfo, PE The purpose of this article is to discuss the measuring and reporting of temperatures at heat transfer equipment during the Testing

More information

METHOD 2C - DETERMINATION OF GAS VELOCITY AND VOLUMETRIC FLOW RATE IN SMALL STACKS OR DUCTS (STANDARD PITOT TUBE)

METHOD 2C - DETERMINATION OF GAS VELOCITY AND VOLUMETRIC FLOW RATE IN SMALL STACKS OR DUCTS (STANDARD PITOT TUBE) 274 METHOD 2C - DETERMINATION OF GAS VELOCITY AND VOLUMETRIC FLOW RATE IN SMALL STACKS OR DUCTS (STANDARD PITOT TUBE) NOTE: This method does not include all of the specifications (e.g., equipment and supplies)

More information

Course: Quality Assurance Module 5 Welders/Welding personnel

Course: Quality Assurance Module 5 Welders/Welding personnel Version 1.0 2010.11.02 1 of 7 Course: Quality Assurance Module 5 Welders/Welding personnel Version 1.0 2010.11.02 2 of 7 Table of Contents MODULE 5...3 Surface inspection on cracks and other surface imperfections

More information

Selecting Airborne-Particle Counters for ISO Certification of Cleanrooms

Selecting Airborne-Particle Counters for ISO Certification of Cleanrooms Application Note Selecting Airborne-Particle Counters for ISO Certification of Cleanrooms Basics Certain factors are routinely considered for any particle counter purchase, such as price, warranty, quality

More information

Environmental Control Requirements. Gordon Farquharson July 2017

Environmental Control Requirements. Gordon Farquharson July 2017 Environmental Control Requirements Gordon Farquharson July 2017 Learning Objectives Learn how contamination control standards and guidance sets requirements for Critical Parameters ISO-14644-1 Classification

More information

EHH-601D. Miami-Dade Qualified Wind-Driven Rain Louver

EHH-601D. Miami-Dade Qualified Wind-Driven Rain Louver Application and Design is a Florida Product Approved and Miami-Dade Qualified stationary horizontal blade wind driven rain extruded aluminum louver designed to protect air intake and exhaust openings in

More information

Standard Laboratory Test Practice For Measurement Of Surface Appearance Change Of Textile Floor Covering As A Result Of Interim Maintenance

Standard Laboratory Test Practice For Measurement Of Surface Appearance Change Of Textile Floor Covering As A Result Of Interim Maintenance P.O. Box 2048 Dalton Georgia 30722-2048 706.278.3176 carpet-rug.org CRI TM 121 Standard Laboratory Test Practice for Measuring Surface Appearance Change of Textile Floor Covering As a Result of Interim

More information

ISO 2889 INTERNATIONAL STANDARD. Sampling airborne radioactive materials from the stacks and ducts of nuclear facilities

ISO 2889 INTERNATIONAL STANDARD. Sampling airborne radioactive materials from the stacks and ducts of nuclear facilities INTERNATIONAL STANDARD ISO 2889 Second edition 2010-03-15 Sampling airborne radioactive materials from the stacks and ducts of nuclear facilities Échantillonnage des substances radioactives contenues dans

More information

HEPAfine. ULPAfine. HEPA Filtration Products. High Efficiency Particulate Air Filters for your most demanding applications

HEPAfine. ULPAfine. HEPA Filtration Products. High Efficiency Particulate Air Filters for your most demanding applications HEPA Filtration Products HEPAfine High Efficiency Particulate Air Filters for your most demanding applications ULPAfine Ultra Low Penetration Air Filters for your most critical applications Reliable HEPAfine

More information

ISO INTERNATIONAL STANDARD. Stationary Source emissions - Determination of asbestos plant emissio.ns - Method by fibre count measurement

ISO INTERNATIONAL STANDARD. Stationary Source emissions - Determination of asbestos plant emissio.ns - Method by fibre count measurement INTERNATIONAL STANDARD ISO 10397 First edition 93-08-01 Stationary Source emissions - Determination of asbestos plant emissio.ns - Method by fibre count measurement Emissions de sources fixes - Determination

More information

ASSESSMENT, CLEANING & RESTORATION OF HVAC SYSTEMS (ACR-2002)

ASSESSMENT, CLEANING & RESTORATION OF HVAC SYSTEMS (ACR-2002) ASSESSMENT, CLEANING & RESTORATION OF HVAC SYSTEMS (ACR-2002) C Cochrane * and T Yacobellis National Air Duct Cleaners Association ACR 2002 Committee, Washington, DC, USA ABSTRACT In the fall of 2001 NADCA

More information

CRITERIA FOR TESTING IMPACT & NON IMPACT RESISTANT BUILDING ENVELOPE COMPONENTS USING UNIFORM STATIC AIR PRESSURE

CRITERIA FOR TESTING IMPACT & NON IMPACT RESISTANT BUILDING ENVELOPE COMPONENTS USING UNIFORM STATIC AIR PRESSURE TESTING APPLICATION STANDARD (TAS) 202-94 CRITERIA FOR TESTING IMPACT & NON IMPACT RESISTANT BUILDING ENVELOPE COMPONENTS USING UNIFORM STATIC AIR PRESSURE 1. Scope 1.1 This protocol covers procedures

More information

Ductless Enclosures, Low-Profile Horizontal Design

Ductless Enclosures, Low-Profile Horizontal Design Ductless Enclosures, Low-Profile Horizontal Design Safety, Convenience and Performance. Fume Box AP60H An Affordable, Effective Air Filtration System Solution p:2 Product Overview Ductless Enclosures,

More information