INSTRUMENT CALIBRATION Section 3 Status & Trend Networks Sampling Manual

Size: px
Start display at page:

Download "INSTRUMENT CALIBRATION Section 3 Status & Trend Networks Sampling Manual"

Transcription

1 INSTRUMENT CALIBRATION Section 3 Status & Trend Networks Sampling Manual

2 Initial Calibration (IC): DEFINITIONS (page 15 of Sampling Manual) The instrument or meter electronics are adjusted to a theoretical value or a known value of a calibration standard. Initial Calibration Verification (ICV): The instrument or meter calibration is checked, directly following the initial calibration, by measuring a calibration standard of known value as if it were a sample. The measured result is compared to the calibration acceptance criteria.

3 DEFINITIONS Continuing Calibration Verification (CCV): The instrument or meter calibration is checked or verified by measuring a calibration standard of known value as if it were a sample. The measured result is compared to the calibration acceptance criteria. Acceptance Criteria: The numerical limits within which calibration verifications are acceptable.

4 Bracketing Requirements All field data must be bracketed by passing calibration or verification data. Chronologically Need verification before collecting field data AND verification after collecting field data. Quantitatively Standards used for calibration / verification must encompass range of values observed in field.

5 Specific Conductance Example Bracket samples by standard range and time! 7:30 Calibrate (1000 umhos/cm) 7:35 ICV (100 umhos/cm) 10:00 Sample Collection (1500 umhos/cm) 15:30 End of Day CCV (5000 umhos/cm)

6 Documentation Calibration log required for all calibrations and verifications Meter ID Date / Time of each calibration or verification Info. about standards Instrument reading Result (Pass / Fail) Name of person performing each activity

7 Standards Do not use expired standards unless they have been previously verified against other (non-expired) standards. Documentation must be maintained for all verification events. (List verification date & person performing verification). If using electronic (SharePoint) logs for standards, do not overwrite information about previous verification events. Do not reuse standards.

8 Parameter Number of Decimal Places To Record Calibration / Verification Frequency Acceptance Criteria ph All Displayed Daily IC, ICV, CCV ± 0.2 SU Specific Conductance All Displayed Daily IC, ICV, CCV ± 5% Dissolved Oxygen (mg/l) All Displayed Daily IC, ICV, CCV ± 0.3 mg/l Dissolved Oxygen (% Sat.) All Displayed N/A N/A Temperature All Displayed Quarterly CCV ± 0.5 C Turbidity All Displayed Daily CCV; Quarterly IC, ICV, secondary standard verification; NTU: ± 10%; NTU: ± 8%; NTU: ± 6.5%; > 100 NTU: ± 5% Depth Cal. / Ver.: 2 for electronic devices; 1 for manual devices Field Meas.: 2 if total depth < 0.6 m; 1 if total depth 0.6 m Daily: IC, ICV for Sondes. Quarterly: Verify Sondes & Electronic Devices. Every 6 months: Inspect Manual Devices. ICV: ± 5% or ± 0.05 m, whichever is greater; Electronic Device Check: ± 10%; Line Increments: ± 10%; Total Length of Lines: ± 5%;

9 Daily Calibration Procedures: Dissolved Oxygen (DO), Specific Conductance, and ph Required: Initial Calibration (IC) at beginning of day. Required: Initial Calibration Verification (ICV) immediately following IC. Required: Continuing Calibration Verification (CCV) at end of day. Recommended: Mid-day CCV at sample site. Mid-day CCV required if end of day CCVs routinely fail. Recommended: Calibration in the field

10 ph Perform 3-pt calibration if meter is capable. Always start with ph 7 buffer Rinse well with DI and buffer before each calibration point or reading. Use different buffers for ICV & CCV. Select buffers to bracket values measured in the field. May need buffers <4.0 or >10.0. Check ph millivolts at least weekly (daily check recommended) ph mv acceptable ranges: ph 4.0 = +180 ± 50; ph 7.0 = 0 ± 50; ph 10.0 = -180 ± 50

11 SPECIFIC CONDUCTANCE Always use Potassium Chloride (KCl) standards. Rinse well with DI and standard before each calibration point or reading. Use different standards for ICV & CCV. Select standards to bracket values measured in the field. If observed field values are all lower than 100 µmhos/cm, use the 100 µmhos/cm standard for CCV. You do not need to use standards lower than this value.

12 Examples If all field meas. < 100 umhos/cm: If field meas. have wide range. Some > 1000 umhos/cm and some < 100 umhos/cm:

13 DISSOLVED OXYGEN Inspect probe & membrane daily Calibrate with water-saturated air or air-saturated water Temperature sensitive Use room temperature water Allow meter to warm up and stabilize before calibrating Compare to Solubility of Oxygen in Water Table (Table 5) Check DO gain and charge at least weekly DO charge acceptable range = 25 to 75 DO gain acceptable range = +0.7 to +1.4 ODO gain acceptable range = to TEMP D.O. TEMP D.O. TEMP D.O. TEMP D.O. TEMP D.O. deg C mg/l deg C mg/l deg C mg/l deg C mg/l deg C mg/l

14 Parameter Number of Decimal Places To Record Calibration / Verification Frequency Acceptance Criteria ph All Displayed Daily IC, ICV, CCV ± 0.2 SU Specific Conductance All Displayed Daily IC, ICV, CCV ± 5% Dissolved Oxygen (mg/l) All Displayed Daily IC, ICV, CCV ± 0.3 mg/l Dissolved Oxygen (% Sat.) All Displayed N/A N/A Temperature All Displayed Quarterly CCV ± 0.5 C Turbidity All Displayed Daily CCV; Quarterly IC, ICV, secondary standard verification; NTU: ± 10%; NTU: ± 8%; NTU: ± 6.5%; > 100 NTU: ± 5% Depth Cal. / Ver.: 2 for electronic devices; 1 for manual devices Field Meas.: 2 if total depth < 0.6 m; 1 if total depth 0.6 m Daily: IC, ICV for Sondes. Quarterly: Verify Sondes & Electronic Devices. Every 6 months: Inspect Manual Devices. ICV: ± 5% or ± 0.05 m, whichever is greater; Electronic Device Check: ± 10%; Line Increments: ± 10%; Total Length of Lines: ± 5%;

15 Temperature Required: quarterly CCV. Compare meter reading to NIST traceable thermometer reading. Perform verification at 2 different temperatures (e.g. ice bath & warm water bath)

16 Parameter Number of Decimal Places To Record Calibration / Verification Frequency Acceptance Criteria ph All Displayed Daily IC, ICV, CCV ± 0.2 SU Specific Conductance All Displayed Daily IC, ICV, CCV ± 5% Dissolved Oxygen (mg/l) All Displayed Daily IC, ICV, CCV ± 0.3 mg/l Dissolved Oxygen (% Sat.) All Displayed N/A N/A Temperature All Displayed Quarterly CCV ± 0.5 C Turbidity All Displayed Daily CCV; Quarterly IC, ICV, secondary standard verification; NTU: ± 10%; NTU: ± 8%; NTU: ± 6.5%; > 100 NTU: ± 5% Depth Cal. / Ver.: 2 for electronic devices; 1 for manual devices Field Meas.: 2 if total depth < 0.6 m; 1 if total depth 0.6 m Daily: IC, ICV for Sondes. Quarterly: Verify Sondes & Electronic Devices. Every 6 months: Inspect Manual Devices. ICV: ± 5% or ± 0.05 m, whichever is greater; Electronic Device Check: ± 10%; Line Increments: ± 10%; Total Length of Lines: ± 5%;

17 Turbidity (GW Projects Only) Required: Quarterly IC using at least 2 primary formazin standards. Required: ICV immediately following IC using at least 1 primary formazin standard. Required (if using secondary standards): Immediately following ICV, verify all secondary gel standards. Reassign current reading as the standard s value. Required: Daily CCV using at least 1 secondary gel standard or factory-sealed primary formazin standard. Do not use turbidity-free water or <0.1 NTU standard for ICVs or CCVs.

18 Turbidity Calibration Log

19 Turbidity (GW Projects Only) Reading a standard or sample: Place meter on level surface. Check cuvette for scratches. Wipe cuvette dry with lint-free cloth. Handle cuvette carefully (no fingerprints or dirt on glass). Ensure cuvette orientation is correct when inserting into meter.

20 Parameter Number of Decimal Places To Record Calibration / Verification Frequency Acceptance Criteria ph All Displayed Daily IC, ICV, CCV ± 0.2 SU Specific Conductance All Displayed Daily IC, ICV, CCV ± 5% Dissolved Oxygen (mg/l) All Displayed Daily IC, ICV, CCV ± 0.3 mg/l Dissolved Oxygen (% Sat.) All Displayed N/A N/A Temperature All Displayed Quarterly CCV ± 0.5 C Turbidity All Displayed Daily CCV; Quarterly IC, ICV, secondary standard verification; NTU: ± 10%; NTU: ± 8%; NTU: ± 6.5%; > 100 NTU: ± 5% Depth Cal. / Ver.: 2 for electronic devices; 1 for manual devices Field Meas.: 2 if total depth < 0.6 m; 1 if total depth 0.6 m Daily: IC, ICV for Sondes. Quarterly: Verify Sondes & Electronic Devices. Every 6 months: Inspect Manual Devices. ICV: ± 5% or ± 0.05 m, whichever is greater; Electronic Device Check: ± 10%; Line Increments: ± 10%; Total Length of Lines: ± 5%;

21 Depth (SW Projects Only) Parameter Number of Decimal Places To Record Calibration / Verification Frequency Acceptance Criteria Depth Cal. / Ver.: 2 for electronic devices; 1 for manual devices Field Meas.: 2 if total depth < 0.6 m; 1 if total depth 0.6 m Daily: IC, ICV for Sondes. Quarterly: Verify Sondes & Electronic Devices. Every 6 months: Inspect Manual Devices. ICV: ± 5% or ± 0.05 m, whichever is greater; Electronic Device Check: ± 10%; Line Increments: ± 10%; Total Length of Lines: ± 5%; Rounding Rule: Numbers 4 are rounded down (e.g becomes 5.1). Numbers 5 are rounded up (e.g becomes 5.2).

22 Depth (SW Projects Only) Daily Requirement: Zero depth sensor on field multiparameter meters (if applicable) and perform ICV. If reading not within 5% of zeroed reading, redo zeroing or conduct maintenance. photo source:

23 Depth (SW Projects Only) Quarterly Requirements: Check all electronic devices used for depth measurement by comparing reading to a reference device (graduated bucket, metal measuring tape, verified secchi disk line). Service device if reading not within 10% of reference.

24 Every 6 Months: Depth (SW Projects Only) Compare all lines used for depth measurement to a reference device (meter stick or metal measuring tape). Redo markings if acceptance criteria not met. 1. Check incremental markings. Acceptance criteria ± 10%. 2. Check total length (up to greatest anticipated depth in field). Acceptance criteria ± 5%.

25 Quarterly Depth Verification Log

26 IC or ICV fails Troubleshooting Perform maintenance / cleaning and attempt IC again. Do not use meter for data collection until issue is resolved. CCV fails Rinse probe with standard and attempt CCV again. If failure still present, affected field data must be J qualified w/ a comment describing the problem. Bracketing requirements not met Affected field data must be J qualified w/ a comment describing the problem.

27 QUESTIONS? Stephanie Sunderman-Barnes

Instrument Calibration. Section 3 Status & Trend Networks Sampling Manual Stephanie Sunderman-Barnes DEAR / WQAP / WMS September 2018

Instrument Calibration. Section 3 Status & Trend Networks Sampling Manual Stephanie Sunderman-Barnes DEAR / WQAP / WMS September 2018 Instrument Calibration Section 3 Status & Trend Networks Sampling Manual Stephanie Sunderman-Barnes DEAR / WQAP / WMS September 2018 1 Initial Calibration (IC): DEFINITIONS (page 15 of Sampling Manual)

More information

Quality Assurance / Quality Control (QA / QC)

Quality Assurance / Quality Control (QA / QC) Quality Assurance / Quality Control (QA / QC) DOCUMENTATION DOCUMENTATION Provides a complete history of any sample from initiation to project completion Includes all associated activities Documentation

More information

Quality Assurance / Quality Control. Stephanie Sunderman-Barnes DEAR / WQAP / WMS September 2018

Quality Assurance / Quality Control. Stephanie Sunderman-Barnes DEAR / WQAP / WMS September 2018 Quality Assurance / Quality Control Stephanie Sunderman-Barnes DEAR / WQAP / WMS September 2018 DOCUMENTATION DOCUMENTATION Provides a complete history of any sample from initiation to project completion

More information

SolarBee/P.A.M Floc Log Test At Hilaman Park Golf Course. Final Monitoring Plan

SolarBee/P.A.M Floc Log Test At Hilaman Park Golf Course. Final Monitoring Plan SolarBee/P.A.M Floc Log Test At Hilaman Park Golf Course Final Monitoring Plan Prepared by Borja Crane-Amores May 21, 2007 1.0 Study Objectives Florida s nonpoint source and watershed restoration programs,

More information

QAM-I-113 Operation and Calibration of the Dissolved Oxygen Meter

QAM-I-113 Operation and Calibration of the Dissolved Oxygen Meter 1.0 Applicability and Purpose i. This procedure applies to the operation and calibration of the YSI Model 5100 dissolved oxygen meter or equivalent and associated probe. By performing the calibration procedure

More information

z /zr/r/ lnitials: tv4'wl z"z{-t L 3 /, /tt Operation and Calibration of the Dissolved Oxygen Meter Revision 5 Approval: Laboratory Man ager/lqao/rso

z /zr/r/ lnitials: tv4'wl zz{-t L 3 /, /tt Operation and Calibration of the Dissolved Oxygen Meter Revision 5 Approval: Laboratory Man ager/lqao/rso QAM.I-1 13 Operation and Calibration of the Dissolved Oxygen Meter Revision 5 Approval: Laboratory Man ager/lqao/rso Date z /zr/r/ Date z"z{-t L Effective date: Renewal date: * 1r i l: 3 /, /tt lnitials:

More information

F U C H S L U B R I C A N T S C O.

F U C H S L U B R I C A N T S C O. F U C H S L U B R I C A N T S C O. SENSOR INFORMATION, CARE AND WARRANTY RM CONCENTRATION / BRIX SENSOR RANGE: 0-35 BRIX GENERAL RM CONCENTRATION SENSOR CARE AND MAINTENANCE RECOMMENDATIONS BY MANUFACTURER

More information

Sampling Manual Revision Oct /15/2017 Page A-1 of 37 Revision to the Status & Trend Networks Sampling Manual Effective October 1, 2017

Sampling Manual Revision Oct /15/2017 Page A-1 of 37 Revision to the Status & Trend Networks Sampling Manual Effective October 1, 2017 Page A-1 of 37 Revision to the Status & Trend Networks Sampling Manual Effective October 1, 2017 Updates to Sampling Manual Section 1: Introduction, Description of Networks, Training The number of Ground

More information

OPERATING INSTRUCTIONS. Portable Turbidity Meter

OPERATING INSTRUCTIONS. Portable Turbidity Meter OPERATING INSTRUCTIONS Portable Turbidity Meter ELE International Chartmoor Road, Chartwell Business Park, Leighton Buzzard Bedfordshire, LU7 4WG England Sales: Civil +44 (0) 870 777 7706 Sales: Env. +44

More information

BOSQUE ECOSYSTEM MONITORING PROGRAM River or Ditch Water Quality Sampling

BOSQUE ECOSYSTEM MONITORING PROGRAM River or Ditch Water Quality Sampling River or Ditch Sampling Materials lifejackets/personal flotation devices (PDFs) 1 per person, always the brown/gray one because of pockets ph meter and calibration solutions waders 1 per person conductivity

More information

SURFACE WATER SAMPLING PROCEDURES

SURFACE WATER SAMPLING PROCEDURES SURFACE WATER SAMPLING PROCEDURES Section 5 February 2017 Status & Trend Networks Sampling Manual Carolyn Voyles WMS Project Manager Major Topics Determination of When / When Not to Collect Samples Sample

More information

DEP Requirements for Sample Collection

DEP Requirements for Sample Collection South District DEP Requirements for Sample Collection AUGUST 2016 What we will cover today: General sampling considerations Meter calibration and verification Field testing and measurement General sampling

More information

STEPS TO DMR-QA SUCCESS. BY: Marcy Bolek

STEPS TO DMR-QA SUCCESS. BY: Marcy Bolek STEPS TO DMR-QA SUCCESS BY: Marcy Bolek marcy@alloway.com Presentation Overview Purpose of DMR-QA Preparation DMR-QA Ethics Analysis Ammonia BOD Residual Chlorine ph TSS Quality Control Reporting What

More information

VAN LONDON COMPANY COMBINATION ph ELECTRODE INSTRUCTION MANUAL

VAN LONDON COMPANY COMBINATION ph ELECTRODE INSTRUCTION MANUAL VAN LONDON COMPANY COMBINATION ph ELECTRODE INSTRUCTION MANUAL GENERAL INFORMATION Van London combination electrodes offer the convenience of having the reference and measuring electrodes combined in a

More information

TN100 PortableTurbidity Meter. Instruction Manual.

TN100 PortableTurbidity Meter. Instruction Manual. TN100 PortableTurbidity Meter Instruction Manual. Table of Contents 1. Overview -------------------------------------------------------------------------------------- 1 2. Technical Specifications --------------------------------------------------------------

More information

Lead & Copper Water Quality Parameters

Lead & Copper Water Quality Parameters Lead & Copper Water Quality Parameters Routine Quality Control Requirements 141.89 Analytical methods. a) Analyses for lead, copper, ph, conductivity, calcium, alkalinity, orthophosphate, silica, and temperature

More information

NPDES Lab Compliance Inspections General Lab Criteria (GLC)

NPDES Lab Compliance Inspections General Lab Criteria (GLC) NPDES Lab Compliance Inspections General Lab Criteria (GLC) Why? How? What? Why is accurate data important? Water Quality Based Effluent Limits Facility demonstrate compliance with effluent limits Design

More information

Turbidity by Nephelometric Determination

Turbidity by Nephelometric Determination SOP AMBL-106-A Page 1 of 5 Standard Operating Procedure AMBL-106-A Prepared: 1/2/2018 Revised: Prepared by: Terry E. Baxter Reviewed by: Turbidity by Nephelometric Determination METHOD SUMMARY This SOP

More information

MAINTENANCE & CALIBRATION: HF TURBIDIMETERS Rev. 4/5/16

MAINTENANCE & CALIBRATION: HF TURBIDIMETERS Rev. 4/5/16 MAINTENANCE & CALIBRATION: HF TURBIDIMETERS Rev. 4/5/16 These procedures are to be performed by Streamkeeper staff or trained volunteers. Calibrate the turbidimeters before every quarterly monitoring season.

More information

Typical Experiments. How it Works

Typical Experiments. How it Works ph Sensor DT016A The ph sensor can be connected to the Nova5000, MultiLogPRO or TriLink data loggers. The ph sensor is capable of measuring the entire range of 0-14 ph and is used for various experiments

More information

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment FAST SHIPPING AND DELIVERY TENS OF THOUSANDS OF IN-STOCK ITEMS EQUIPMENT DEMOS HUNDREDS OF MANUFACTURERS SUPPORTED

More information

Elsinore Valley Municipal Water District. Lake Elsinore Phase 2 Water Quality Monitoring Plan

Elsinore Valley Municipal Water District. Lake Elsinore Phase 2 Water Quality Monitoring Plan Elsinore Valley Municipal Water District Lake Elsinore Phase 2 Water Quality Monitoring Plan to Evaluate the Efficacy of the In-Lake Nutrient Reduction Facilities (Aeration and Mixing) for Lake Elsinore

More information

FIELD PROCEDURE: TURBIDITY

FIELD PROCEDURE: TURBIDITY FIELD PROCEDURE: TURBIDITY EQUIPMENT NEEDED: Note: Don t leave this equipment out in freezing temperatures unless you re using it. If air temperatures are just below freezing, keep the turbidity standards

More information

Instruction Manual HI Portable Microprocessor Turbidity Meter

Instruction Manual HI Portable Microprocessor Turbidity Meter Instruction Manual HI 93703 Portable Microprocessor Turbidity Meter 1 PRELIMINARY EXAMINATION Dear Customer, Thank you for choosing a HANNA instruments product. Please read this instruction manual carefully

More information

Eureka Sub3 Water-Quality Multiprobe

Eureka Sub3 Water-Quality Multiprobe Highly reliable water-quality data for temperature, optical dissolved oxygen, conductivity, ph, depth or level, ORP, and turbidity Learn the software in 10 minutes Industry-best customer service Use in

More information

HACH TU 5 SERIES TURBIDIMETERS SOP: COMPARISON LAB AND PROCESS. The next standard in the evolution of turbidity

HACH TU 5 SERIES TURBIDIMETERS SOP: COMPARISON LAB AND PROCESS. The next standard in the evolution of turbidity HACH TU 5 SERIES TURBIDIMETERS SOP: COMPARISON LAB AND PROCESS The next standard in the evolution of turbidity COMPARE TU5 LAB AND PROCESS GUIDE & TROUBLESHOOT To guarantee that lab and process readings

More information

Sampling Day Activities

Sampling Day Activities Sampling Day Activities Approved by: WWA Director Reviewed by: Volunteer Monitoring Coordinator On the day of sampling, the following activities are to be conducted at the sampling site in the order presented

More information

SOP-C-102 Determination of Chemical Oxygen Demand

SOP-C-102 Determination of Chemical Oxygen Demand i. Identification of the method a. Hach 8000, Chemical Oxygen Demand, with additions to meet project requirements ii. Applicable matrix or matrices a. Nonpotable water and aqueous solutions iii. iv. Limits

More information

With Discovery Incubator.

With Discovery Incubator. TEST KIT POTABLE WATER With Discovery Incubator. FEATURES & BENEFITS 8 dip slide capacity Compact and convenient design Weighs only 1.2 kg Temperature range: ambient +5 to 45 C Accurate temperature control

More information

Using Real-Time Monitoring and Phycocyanin Probes to Track Cyanobacteria Blooms. Tom Faber US EPA New England Regional Laboratory

Using Real-Time Monitoring and Phycocyanin Probes to Track Cyanobacteria Blooms. Tom Faber US EPA New England Regional Laboratory Using Real-Time Monitoring and Phycocyanin Probes to Track Cyanobacteria Blooms Tom Faber US EPA New England Regional Laboratory Measuring Cyanobacteria EPA buoy monitoring program Cyanobacteria cell counts

More information

Development of the U-50 Series Multi-Parameter Water Quality Checker

Development of the U-50 Series Multi-Parameter Water Quality Checker S e l e c t e d A r t i c l e Selected Article Development of the U-0 Series Multi-Parameter Water Quality Checker Yuichiro Komatsu, Katsunobu Ehara, Katsuaki Ogura Multi-parameter water quality checker

More information

Mercury. App K APEX INSTRUMENTS, INC. Appendix K Automated Mercury Source Sampler Model XC-6000EPC. Console Audit Manual

Mercury. App K APEX INSTRUMENTS, INC. Appendix K Automated Mercury Source Sampler Model XC-6000EPC. Console Audit Manual Mercury App K APEX INSTRUMENTS, INC. Appendix K Automated Mercury Source Sampler Model XC-6000EPC Console Audit Manual XC6000 Quarterly Audit Procedure... 3 Audit Setup:... 5 Starting the Console Audit:...

More information

HI EPA Compliant Turbidity Meter HI EPA Compliant Turbidity and Chlorine Meter

HI EPA Compliant Turbidity Meter HI EPA Compliant Turbidity and Chlorine Meter HI 98703 EPA Compliant Turbidity Meter HI 93414 EPA Compliant Turbidity and Chlorine Meter SPECIFICATIONS Turbidity (HI 93414 & HI 98703) Range Range Selection ACCESSORIES HI 710005 Voltage adapter HI

More information

EPA Compliant Turbidity & Chlorine Meter

EPA Compliant Turbidity & Chlorine Meter EPA Compliant Turbidity & Chlorine Meter New Turbidity Technology from HANNA HANNA s new HI 98703 and HI 93414 utilize an EPA compliant tungsten light source and state-of-the-art optical system that allow

More information

July 27, A NEW EPA APPROVED METHOD FOR TURBIDITY ANALYSIS 360 x 90, Turbidity Method TURBIDITY MEASUREMENT HISTORY

July 27, A NEW EPA APPROVED METHOD FOR TURBIDITY ANALYSIS 360 x 90, Turbidity Method TURBIDITY MEASUREMENT HISTORY A NEW EPA APPROVED METHOD FOR TURBIDITY ANALYSIS 360 x 90, Turbidity Method 10258 TURBIDITY MEASUREMENT HISTORY 2 1 TURBIDITY MEASUREMENT OVER THE LAST 50 YEARS 3 TURBIDITY MEASUREMENT Turbidity is the

More information

Standard Operating Procedure for: Horiba U-22XD Multi-Parameter Water Quality Monitoring System (1050R01 Horiba U-22.doc) Missouri State University

Standard Operating Procedure for: Horiba U-22XD Multi-Parameter Water Quality Monitoring System (1050R01 Horiba U-22.doc) Missouri State University Standard Operating Procedure for: Horiba U-22XD Multi-Parameter Water Quality Monitoring System (1050R01 Horiba U-22.doc) Missouri State University and Ozarks Environmental and Water Resources Institute

More information

OWNERS MANUAL. Micro TPI Field Portable Turbidimeter

OWNERS MANUAL. Micro TPI Field Portable Turbidimeter OWNERS MANUAL Micro TPI Field Portable Turbidimeter HF scientific, inc. 3170 Metro Parkway Ft. Myers, FL 33916 Phone: 239-337-2116 Fax: 239-332-7643 Email: Info@hfscientific.com Manual Part No. 22666 (12/04)

More information

EPA APPROVED METHOD FOR TURBIDITY ANALYSIS 360 x 90, Turbidity Method 10258

EPA APPROVED METHOD FOR TURBIDITY ANALYSIS 360 x 90, Turbidity Method 10258 EPA APPROVED METHOD FOR TURBIDITY ANALYSIS 360 x 90, Turbidity Method 10258 TURBIDITY MEASUREMENT HISTORY 3 TURBIDITY MEASUREMENT OVER THE LAST 50 YEARS TURBIDITY MEASUREMENT Turbidity is the interaction

More information

Self-cleaning Turbidity Sensor

Self-cleaning Turbidity Sensor 00713589 Self-cleaning Turbidity Sensor Safety Precautions Please read this entire instruction sheet before operating this sensor. Pay particular attention to all danger and caution statements. Failure

More information

Turbidity Meter Instruction Manual

Turbidity Meter Instruction Manual Turbidity Meter 860040 Instruction Manual TABLE OF CONTENTS INTRODUCTION...2 FEATURES...3 MATERIALS SUPPLIED...3 FRONT PANEL DESCRIPTION...4 SET UP...5 CALIBRATION...6 MEASUREMENT PROCEDURES...8 ZERO ADJUSTMENT...9

More information

Watershed Monitoring January 2012 Q-Meeting QA Update: January 2012 Sampling Manual

Watershed Monitoring January 2012 Q-Meeting QA Update: January 2012 Sampling Manual Watershed Monitoring January 2012 Q-Meeting QA Update: January 2012 Sampling Manual Shannon Gerardi QA Officer Overview January 2012 Sampling Manual updates per section since last Q-Meeting (February 2011)

More information

Dissolved Oxygen Probe (Order Code: DO-BTA)

Dissolved Oxygen Probe (Order Code: DO-BTA) Dissolved Oxygen Probe (Order Code: DO-BTA) The Vernier Dissolved Oxygen Probe uses a Clark-type polarographic electrode to measure the concentration of dissolved oxygen in water. You can use this sensor

More information

Standard Operating Procedure for:

Standard Operating Procedure for: Standard Operating Procedure for: Eureka Amphibian and Manta Water Quality Multiprobe for Multiple Location Parameter Measurement (1200R02 Eureka Snapshot.doc) Missouri State University and Ozarks Environmental

More information

TB100 Portable Turbidity Meter Instruction Manual

TB100 Portable Turbidity Meter Instruction Manual TB100 Portable Turbidity Meter Instruction Manual BANTE INSTRUMENTS CO., LTD TB100 Portable Turbidity Meter 1 Introduction Thank you for selecting the TB100 portable turbidity meter. This manual provides

More information

Required Documentation for PA State (Chapter 252) Accreditation

Required Documentation for PA State (Chapter 252) Accreditation Required Documentation for PA State (Chapter 252) Accreditation Disclaimer: The information in this guidance document does not supplant the provisions of the Environmental Laboratory Accreditation Regulations,

More information

Instruction Manual HI RH/ C Transmitter. This Instrument is in Compliance with the CE Directives.

Instruction Manual HI RH/ C Transmitter. This Instrument is in Compliance with the CE Directives. Instruction Manual HI 8666 RH/ C Transmitter This Instrument is in Compliance with the CE Directives www.hannainst.com Dear Customer, Thank you for choosing a Hanna Instruments Product. Please read this

More information

Engineered Systems & Designs TURBIDITY METER

Engineered Systems & Designs TURBIDITY METER Engineered Systems & Designs TURBIDITY METER Operator s Manual for Model 800 Turbidimeter Model 800 with Optional Battery Pack and Carrying Case INDEX Specifications for Turbidity.................. 1 Introduction..............

More information

Portable Multi-Parameter Water Quality Meter LMPWM-A1 Series

Portable Multi-Parameter Water Quality Meter LMPWM-A1 Series Portable MultiParameter Water Quality Meter LMPWMA1 Series www.labtron.com info@labtron.com Portable Multiparameter Water Quality Meter LMPWMA1 series Portable Multiparameter Water Quality Meter LMPWMA1

More information

Warm Mineral Springs Sampling by Sarasota County

Warm Mineral Springs Sampling by Sarasota County Warm Mineral Springs Sampling by Sarasota County John Ryan, Kathryn Meaux, Rene Janneman and Jon S. Perry Sarasota County Environmental Services Sarasota, Florida September 11 Warm Mineral Springs is a

More information

2007 Great Bay Water Quality (DataSonde) Monitoring Program

2007 Great Bay Water Quality (DataSonde) Monitoring Program University of New Hampshire University of New Hampshire Scholars' Repository PREP Reports & Publications Institute for the Study of Earth, Oceans, and Space (EOS) 3-31-2008 2007 Great Bay Water Quality

More information

Lake Elsinore and San Jacinto Watersheds Authority. Canyon Lake Phase 2 Water Quality Monitoring Plan

Lake Elsinore and San Jacinto Watersheds Authority. Canyon Lake Phase 2 Water Quality Monitoring Plan Lake Elsinore and San Jacinto Watersheds Authority December 2010 Background Canyon Lake was formed in 1928 when the Canyon Lake (Railroad Canyon) Dam was constructed; the lake has three main sections the

More information

Des Moines River Water Quality Network

Des Moines River Water Quality Network Des Moines River Water Quality Network Monthly Report for December 2014 Donna S Lutz Assistant Scientist/Project Manager 394 Town Eng Bldg Iowa State University (515) 294-9720 dslutz@iastate.edu http://www.ccee.iastate.edu/dmrwqn

More information

Water Quality Physical / Chemical Data Report Page 1. United States Army Corps of Engineers Nashville District CELRN-EC-H / Water Management Section

Water Quality Physical / Chemical Data Report Page 1. United States Army Corps of Engineers Nashville District CELRN-EC-H / Water Management Section Water Quality Physical / Chemical Data Report Page 1 Station: 3LAU10001 Project: Laurel Location: Laurel River, Mile 2.2 State: Kentucky Latitude: 36º 57" 41' Quad: Sawyer, KY County: Laurel Longitude:

More information

Sulfate. USEPA 1 SulfaVer 4 Method 2 Method to 70 mg/l SO 4. Powder Pillows. Test preparation. Before starting.

Sulfate. USEPA 1 SulfaVer 4 Method 2 Method to 70 mg/l SO 4. Powder Pillows. Test preparation. Before starting. Sulfate DOC316.53.01473 USEPA 1 SulfaVer 4 Method 2 Method 8051 2 to 70 mg/l SO 4 2 Powder Pillows Scope and application: For water, wastewater and seawater. 1 USEPA accepted for reporting wastewater analyses.

More information

SUMMARY OF STANDARDS FOR WATER ANALYSIS ON COURSE WAT-E2120 SPRING 2017

SUMMARY OF STANDARDS FOR WATER ANALYSIS ON COURSE WAT-E2120 SPRING 2017 SUMMARY OF STANDARDS FOR WATER ANALYSIS ON COURSE WAT-E2120 SPRING 2017 - electrical conductivity - turbidity - NPOC - color - COD Mn - UV-absorbance - bacteria - total iron Edited by Aino 1 Determination

More information

ANALYTICAL SUMMARY REPORT

ANALYTICAL SUMMARY REPORT ANALYTICAL SUMMARY REPORT August 11, 2014 801 Washington Cottage Grove, OR 97424 Work Order: G14070537 Quote ID: G433 - Soil/Product Mix Project Name: Energy Laboratories Inc. Gillette WY received the

More information

SURFACE WATER SAMPLING PROCEDURES

SURFACE WATER SAMPLING PROCEDURES SURFACE WATER SAMPLING PROCEDURES Section 5 September 20, 2017 Status & Trend Networks Sampling Manual Carolyn Voyles WMS Project Manager Major Topics Determination of When / When Not to Collect Samples

More information

Odyssey Conductivity & Temperature Logger.

Odyssey Conductivity & Temperature Logger. Odyssey Conductivity & Temperature Logger. Contents. Salinity Sensor & Software.... 2 Conductivity Cell Calibration.... 2 Standard Solutions.... 3 Temperature Coefficient... 5 Positioning the Cell....

More information

BOD/CBOD, TSS ph OPERATOR BASICS. Presented by: Marcy Bolek - Alloway

BOD/CBOD, TSS ph OPERATOR BASICS. Presented by: Marcy Bolek - Alloway BOD/CBOD, TSS ph OPERATOR BASICS Presented by: Marcy Bolek - Alloway CONTENTS BOD/CBOD ANALYSIS TSS ANALYSIS ph ANALYSIS BIOCHEMICAL OXYGEN DEMAND (BOD) mg/l oxygen bacteria use to oxidize organic matter

More information

Chemical Oxygen Demand

Chemical Oxygen Demand HI83314 Multiparameter Photometer with COD for Wastewater with Digital ph Electrode Input HI83314 photometer measures 10 different key wastewater quality parameters using 20 different methods that allow

More information

Model 435A ph Monitor/Recorder Manual

Model 435A ph Monitor/Recorder Manual Model 435A ph Monitor/Recorder Manual A digital version of this manual is available online at www.krugerandeckels.com 2 Contents Description... 4 Operation... 4 Electrodes... 5 Electrode Temperature...

More information

FIOLAX. Solutions. Buffer solutions in the unique double-end ampoules offer a particularly high degree of reliability and measuring accuracy.

FIOLAX. Solutions. Buffer solutions in the unique double-end ampoules offer a particularly high degree of reliability and measuring accuracy. Solutions Buffer solutions in the unique double-end ampoules offer a particularly high degree of reliability and measuring accuracy. The exactness of the ph measurement is mainly dependent on the accuracy

More information

Solutions. Standard buffer solutions according to DIN Hot steam sterilized for longer stability, no preservation agents used.

Solutions. Standard buffer solutions according to DIN Hot steam sterilized for longer stability, no preservation agents used. Solutions Standard buffer solutions according to DIN 19 266 Hot steam sterilized for longer stability, no preservation agents used. Order No. Type No. ph value at 25 C Contents 285137977 L 4791 1.68 60

More information

Centre d expertise en analyse environnementale du Québec

Centre d expertise en analyse environnementale du Québec Centre d expertise en analyse environnementale du Québec Methods for taking, preserving and analyzing samples to monitor the water quality of pools and other artificial reservoirs DR-09-05 Edition 2007-06-20

More information

Dissolved Oxygen. LabQuest INTRODUCTION

Dissolved Oxygen. LabQuest INTRODUCTION Dissolved Oxygen LabQuest 5 INTRODUCTION Oxygen gas dissolved in water is vital to the existence of most aquatic organisms. Oxygen is a key component in cellular respiration for both aquatic and terrestrial

More information

MultiLab BROCHURE MULTIPARAMETER LABORATORY INSTRUMENTS W61-02

MultiLab BROCHURE MULTIPARAMETER LABORATORY INSTRUMENTS W61-02 MultiLab MULTIPARAMETER LABORATORY INSTRUMENTS BROCHURE W61-02 MultiLab YSI.com/MultiLab Customize your MultiLab system to meet the demands of your laboratory application. With the ability to measure up

More information

A natural impurity in water. The test will measure the amount of free oxygen "gas" dissolved in your water sample in ppm (mg/l).

A natural impurity in water. The test will measure the amount of free oxygen gas dissolved in your water sample in ppm (mg/l). What is? A natural impurity in water. The test will measure the amount of free oxygen "gas" dissolved in your water sample in ppm (mg/l). Aquatic animals such as fish and zooplankton, can not use the oxygen

More information

METHOD DETERMINATION OF TURBIDITY BY NEPHELOMETRY. Edited by James W. O'Dell Inorganic Chemistry Branch Chemistry Research Division

METHOD DETERMINATION OF TURBIDITY BY NEPHELOMETRY. Edited by James W. O'Dell Inorganic Chemistry Branch Chemistry Research Division METHOD 180.1 DETERMINATION OF TURBIDITY BY NEPHELOMETRY Edited by James W. O'Dell Inorganic Chemistry Branch Chemistry Research Division Revision 2.0 August 1993 ENVIRONMENTAL MONITORING SYSTEMS LABORATORY

More information

Standard Operating Procedure for GLNPO Turbidity: Nephelometric Method

Standard Operating Procedure for GLNPO Turbidity: Nephelometric Method Standard Operating Procedure for GLNPO Turbidity: Nephelometric Method Marvin Palmer United States Environmental Protection Agency Great Lakes National Program Office Region 5 77 West Jackson 9th Floor

More information

Turbidity. LabQuest INTRODUCTION

Turbidity. LabQuest INTRODUCTION INTRODUCTION is a measure of water s lack of clarity. Water with high turbidity is cloudy, while water with low turbidity is clear. The cloudiness is produced by light reflecting off of particles in the

More information

Oyster ph - Conductivity - TDS - Temperature Meters

Oyster ph - Conductivity - TDS - Temperature Meters User's Guide Oyster ph - Conductivity - TDS - Temperature Meters Models 34135-A 341350 341350-P 34145A 341450 341450-P INTRODUCTION Congratulations on your purchase of the Extech's Oyster Series meter

More information

Certificate of Accreditation

Certificate of Accreditation PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of: (Hereinafter called the Organization) and hereby declares

More information

:00 am 6:00 pm Mon-Thu (MT) 6:00 am 5:00 pm Fri (MT)

:00 am 6:00 pm Mon-Thu (MT) 6:00 am 5:00 pm Fri (MT) 800.372.0122 303.431.8454 6:00 am 6:00 pm Mon-Thu (MT) 6:00 am 5:00 pm Fri (MT) info@eraqc.com www.eraqc.com www.facebook.com/eraqc www.youtube.com/eraqc 2013 Waters Corporation 1 Top Tips for Conducting

More information

Chloroplast photosynthesis. Effects of varying light intensities and wavelengths on the activity of the oxygen evolving complex

Chloroplast photosynthesis. Effects of varying light intensities and wavelengths on the activity of the oxygen evolving complex BISC 367 Plant Physiology Laboratory Simon Fraser University 1 Chloroplast photosynthesis Effects of varying light intensities and wavelengths on the activity of the oxygen evolving complex Photosynthesis

More information

PORTABLE ph METER PRICE CODE NUMBER MODEL 100

PORTABLE ph METER PRICE CODE NUMBER MODEL 100 PORTABLE METER MODEL 100 PRICE CODE NUMBER 56-0032 Refer to Bulletin A-301. OPERATION AND SERVICE GUIDE O-670A DEC. 1997 GENERAL INFORMATION Model 100 meter is designed for measuring over the entire 0

More information

Table 1: QA/QC Data Comparison Rankings for deployment between June 24 and July 21, 2009.

Table 1: QA/QC Data Comparison Rankings for deployment between June 24 and July 21, 2009. General The Water Resources Management Division staff monitors the real-time web page on a daily basis. This monthly deployment report interprets the data from a water quality monitoring station on Minipi

More information

3. Add 0.4 ml of. 7. Use a TenSette

3. Add 0.4 ml of. 7. Use a TenSette Method 10173 ORGANIC CARBON, TOTAL, Mid Range (15 150 mg/l C) Direct Method * For wastewater and industrial waters HRS MIN SEC 1. Turn on the DRB 200 reactor. Heat to 103 105 C. Note: See DRB 200 user

More information

SAMPLE PRESERVATION Liz Miller February 21, 2017

SAMPLE PRESERVATION Liz Miller February 21, 2017 SAMPLE PRESERVATION Liz Miller February 21, 2017 Most recent custody sheet versions: Surface Water = January 2017 Ground Water = October 2016 Can be found at https://fldeploc.dep.state.fl.us/status/ Pay

More information

Utilization of Geochemical Parameters in Developing a Conceptual Site Model

Utilization of Geochemical Parameters in Developing a Conceptual Site Model Utilization of Geochemical Parameters in Developing a Conceptual Site Model Annual Risk Assessment & Remediation Seminar MSECA, Indianapolis, IN October 29 2012 µg/l Steven R. Irvin & Jim V. Rouse Acuity

More information

Let s take a look several devices and methods for obtaining measurements of water quality in the field.

Let s take a look several devices and methods for obtaining measurements of water quality in the field. EHST 2361 Environmental Sampling and Analysis Calibration and Use of Dissolved Oxygen and ph/conductivity Field Meters and Colorimeters in Water Quality Investigations Introduction We are taught that samples

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 QUALITY CONTROL SERVICES, INC. 2340 SE 11 th Avenue Portland, OR 97214 George Cahill Phone: 503 236 2712 CALIBRATION Valid To: September

More information

Thermo Scientific Orion Versa Star Pro Benchtop Meters

Thermo Scientific Orion Versa Star Pro Benchtop Meters LABORATORY PRODUCTS Thermo Scientific Orion Versa Star Pro Benchtop Meters Versatile, high-performance lab measurement systems Product Specifications Thermo Scientific Orion Versa Star Pro premium electrochemistry

More information

Instructors/Trainers. Groundwater Sampling Training Workshop. Objectives of GW Sampling. Agenda Estimated time (minutes) Course Objectives

Instructors/Trainers. Groundwater Sampling Training Workshop. Objectives of GW Sampling. Agenda Estimated time (minutes) Course Objectives Instructors/Trainers Groundwater Sampling Training Workshop Florida Society of Environmental Analysts Spring Meeting Sirata Beach Resort May 21, 2008 Delia Ivanoff, South FL Water Management District Tom

More information

Module 2, Add on Lesson Turbidity Sensor. Teacher. 90 minutes

Module 2, Add on Lesson Turbidity Sensor. Teacher. 90 minutes Module 2, Add on Lesson Turbidity Sensor Teacher 90 minutes Purpose Construct a sensor to measure the turbidity of water Graph data and reason about curves and linear relationships Calibrate the turbidity

More information

Hach Method Luminescence Measurement of Dissolved Oxygen (LDO ) in Water and Wastewater

Hach Method Luminescence Measurement of Dissolved Oxygen (LDO ) in Water and Wastewater Hach Method 10360 Luminescence Measurement of Dissolved Oxygen (LDO ) in Water and Wastewater LED Photo Diode LED Hach Method 10360 Luminescence Measurement of Dissolved Oxygen in Water and Wastewater

More information

Total Dissolved Solids Dried at 180ºC. Total Volatile Dissolved Solids at 550 C

Total Dissolved Solids Dried at 180ºC. Total Volatile Dissolved Solids at 550 C Title: Total Dissolved Solids Dried at 180ºC and Total Volatile Dissolved Solids at 500 C Page 1 of 9 Total Dissolved Solids Dried at 180ºC Total Volatile Dissolved Solids at 550 C References: SM 2540

More information

TSS The TSS analyser was operational during the reporting period. The composite sampler is in place to cover any shortfalls in the TSS analyser.

TSS The TSS analyser was operational during the reporting period. The composite sampler is in place to cover any shortfalls in the TSS analyser. Interim Environmental Report Period Ending: 3 th April 28 Compiled By: Siobhan Quinn, Aoife Reynolds & Thomas McGlynn Approved By: Tony Doyle 1 Monitoring Data 1.1 Monitoring Equipment Axonics Axonics

More information

Method 0040 for Volatile Total Organics as Unspeciated Mass in Stack Gas Emissions

Method 0040 for Volatile Total Organics as Unspeciated Mass in Stack Gas Emissions Method 0040 for Volatile Total Organics as Unspeciated Mass in Stack Gas Emissions Sampling and Field Procedure for Volatile Unspeciated Mass from Method 0040 Samples in Stack Gas Sample Name: Sampler:

More information

Collecting a Surface Water Sample

Collecting a Surface Water Sample Collecting a Surface Water Sample CAUTION: Never carry or lift the pole above your head, as touching power lines could cause electrocution. Notes: Students (

More information

Module 2, Add on Lesson Turbidity Sensor. Student. 90 minutes

Module 2, Add on Lesson Turbidity Sensor. Student. 90 minutes Module 2, Add on Lesson Turbidity Sensor Student 90 minutes Purpose Construct a sensor to measure the turbidity of water Graph data and reason about curves and linear relationships Calibrate the turbidity

More information

ProODO Optical Dissolved Oxygen Meter for the Lab or Field SPECIFICATIONS

ProODO Optical Dissolved Oxygen Meter for the Lab or Field SPECIFICATIONS ProODO Optical Dissolved Oxygen Meter for the Lab or Field SPECIFICATIONS ProODO Digital Optical Dissolved Oxygen Meter YSI ODO technology reduces the time required to calibrate and maintain your sensors.

More information

3. Add 0.4 ml of. 7. Use a TenSette

3. Add 0.4 ml of. 7. Use a TenSette Method 10129 ORGANIC CARBON, TOTAL, Low Range (0.0 20.0 mg/l C) Direct Method * For water, drinking water, and wastewater HRS MIN SEC 1. Turn on the DRB 200 reactor. Heat to 103-105 C. Note: See DRB 200

More information

Waterford River Kilbride NF02ZM0009 December 2013 to January 2014

Waterford River Kilbride NF02ZM0009 December 2013 to January 2014 Waterford River @ Kilbride NF02ZM0009 December 2013 to January 2014 Government of Newfoundland & Labrador Department of Environment and Conservation Real Time Water Quality Monthly Report Waterford River

More information

Department of Laboratories St. Louis, MO ISSUE DATE: June 2007 REVISION DATE: May 2012 REVIEWED DATE: May 2018 PRINCIPLE:

Department of Laboratories St. Louis, MO ISSUE DATE: June 2007 REVISION DATE: May 2012 REVIEWED DATE: May 2018 PRINCIPLE: Page 1 of 7 PROCEDURE: LEAD ESA LEADCARE II / HEALTHY KIDS EXPRESS ISSUE DATE: June 2007 REVISION DATE: May 2012 REVIEWED DATE: May 2018 PRINCIPLE: The LEADCARE II System relies on electrochemistry and

More information

CEE 311 Environmental Engineering I Fall 2015 Dr. Jagadish Torlapati BIOCHEMICAL OXYGEN DEMAND (BOD)

CEE 311 Environmental Engineering I Fall 2015 Dr. Jagadish Torlapati BIOCHEMICAL OXYGEN DEMAND (BOD) CEE 311 Environmental Engineering I Fall 2015 Dr. Jagadish Torlapati BIOCHEMICAL OXYGEN DEMAND (BOD) Objectives: samples. To measure the Biochemical Oxygen Demand of given wastewater Sample Source: Rowan

More information

Single-Use Sensors. TruFluor and TruTorr Excellence in bio-processing measurement

Single-Use Sensors. TruFluor and TruTorr Excellence in bio-processing measurement Data File #808-DST-001-Rev1 Single-Use Sensors TruFluor and TruTorr Excellence in bio-processing measurement Secure Minimize contamination, maximize single-use container integrity Measure Obtain real-time

More information

Waterford River at Kilbride

Waterford River at Kilbride Real Time Water Quality Report Waterford River at Kilbride Deployment Period June 2, 2015 to August 20, 2015 Government of Newfoundland & Labrador Department of Environment and Conservation Water Resources

More information

Calibration and Standards. AMCO Standards for use in Drinking Water Analysis

Calibration and Standards. AMCO Standards for use in Drinking Water Analysis Quality Control using Measurements measurements are of extreme importance in quality monitoring in water, wastewater, beverage production, electroplating and petrochemical applications. Typical turbidity

More information

Quick Select Guide. 6-Series. products

Quick Select Guide. 6-Series. products Quick Select Guide 6-Series products 6-Series instruments Premier Instruments The YSI 6-Series is a full line of multiparameter instruments and accessories for most unattended monitoring, water quality

More information