1 EMPLOYING MICRO GAS CHROMATS FOR FLARE ANALYSIS TO MEET EMISSION REGULATIONS Copyright Material PCIC Europe Paper No. EUR17_34 Gavin Dewar Davie Plenderleith Eur Ing Steve Jackson MA Andrew Eady BEng(Hons) INEOS O&P UK INEOS O&P UK Siemens UK Siemens UK Abstract - Climate change is one of the most pressing worldwide challenges particularly following the 1997 Kyoto protocol. A major contribution to green house gases (GHG) originates from gas flaring: the World Bank estimates circa 150 bn cum gas was flared globally per year, producing 400m tons of CO2 which is equivalent to 25% of US gas consumption or 30% of EU consumption. In the EU the key regulation to limit flare GHG releases is the Emissions Trading Scheme (ETS) Carbon Credit Trading principle since 2005 for new plants (offshore flaring since 2009). Flaring is a controlled, open flame combustion process to burn unwanted or surplus hydrocarbon gases from different sources for over-pressure, maintenance, safety and other reasons. The INEOS Grangemouth polymer plants have combined high pressure and low pressure flare stacks. Since 2013 the CO2 emissions had to be declared. Prior to this INEOS installed ultrasonic flowmeters on the flare lines to measure the mass flow. These meters however require the volume of Nitrogen in the gas to be known, in order to calculate the mass flow. The remaining gas was then assumed to be hydrocarbons to be combusted. The Nitrogen Vol % at this point was calculated using the DCS via a complicated algorithm. This resulted in the potential to over-report emissions for EU ETS with significant financial impact. The existing flare seal oxygen analysers had become obsolete, therefore it was decided to replace these and install micro gas chromatographs (GC) to analyse actual hydrocarbon and H2 composition to the stacks as well as measuring flare N2 and O2 levels. This new measured approach would have a short payback, and improve the volumetric to mass flow calculation of the flare flow meters by using a measured, not calculated, Nitrogen value. This paper describes the implementation and operation of the new micro GCs at the INEOS Grangemouth chemical complex in Scotland during Index Terms Micro, Process Gas Chromatographs, Flare Stack, Emissions Regulations, Green House Gas (GHG), Emission Trading System (ETS). I. INTRODUCTION This paper will cover the carbon credit concept, an overview of the Grangemouth chemical plants, issues with the original set-up, the new analyser scope and lessons learnt. II. MICRO GAS CHROMATS FOR FLARE ANALYSIS Fig 1 INEOS Grangemouth A. The Carbon Credit Concept During flaring combustion hydrocarbons react with atmospheric O2 to form CO2 which is the most active GHG. Venting without combustion often releases methane the second main GHG directly to the atmosphere. Many countries in Europe and N America have established regulations to limit and reduce the amount of GHG emissions. Flares are specifically mentioned as a source of GHG and included in these regulations. The ETS is the central part of the EU s policy to reduce climate change and a key tool for reducing industrial GHG emissions. Introduced in 2005, it became the biggest international system for trading GHG emission allowances. It covers more than 11,000 facilities in 31 countries which mean around 45% of total GHG emissions from these countries. The EU ETS is now in its third phase, running from 2013 to 2020 with a major revision approved in 2009 to strengthen the system. This third phase also includes flaring from chemical production sites, and the allocation for routine and maintenance flaring. 1
2 The main legislative instrument to regulate the measurement of flare gas emission is the EU ETS Monitoring and Reporting Guideline. The EU allocates Carbon Credits (gas emission permits) as tradable allowances; emissions must be monitored and reported. In each country, this allocated limit ( cap ) is cascaded down over the various industrial sites, which have to report their individual emissions at the end of the year. Depending whether more or less CO2 than allocated have been emitted, sites sell or must buy credits. B. Introduction to the INEOS Grangemouth Chemical Plants INEOS have a fully integrated production complex with a range of chemical plants at the Grangemouth site including polypropylene, ethylene, polyethylene and ethanol. The site s ethane feedstock was taken predominately from the North Sea, but with declining volumes this is now being supplemented with imported US shale gas from a newly constructed ethane storage tank - fully commissioned in September commodities to the countries which limit the amount of CO2 that can be released to the atmosphere. To ensure compliance to each country s allocated C. Original Set-Up Material sent to each of the HFL and the LFL flares is metered by the recently installed, dedicated ultrasonic flow meters FI7021 and FI7022; these meters run an algorithm based on hydrocarbon flow to calculate the mass flow from a measured volumetric flow. The O2 content in the flare stacks was monitored by the online analysers AE7000 and AE7010. As of the beginning of 2013 the CO2, emissions from both polymer plants had to be declared under regulations related to the EU ETS. The accuracy requirement of these regulations meant that the method previously used was no longer sufficient and a new, conservative method had to be introduced. The new method assumed that all flow to flare is hydrocarbon to be combusted, unless it is known to be N2. D. Issues with Original Set-Up As a major portion of the flared material is N2 which is not metered, a conservative estimate leads to overreporting of carbon emissions. The difference between the old and new methods for the first two months of 2013 was substantial. Fig 2 INEOS Grangemouth The polymer plants I4 (polyethylene) and PP3 (polypropylene) share a common flare system which includes a high pressure flare (HFL) and a low pressure flare (LFL). The flares are designed to safely achieve complete combustion of any gaseous waste hydrocarbon emitted from the polymer plants. Inert N2 as well as being 80% of the composition of air is introduced into the polymer plants for blanketing and purging safety purposes but produces no GHGs so it is important to measure this gas to the flare line. Similarly H2 attracts no carbon credits but is injected to control the viscosity of the polymer. However, O2 has to be measured for safety reasons to ensure it has not seeped back down the flare stack and caused the O2 level in the flare line to be above 2% which would be potentially explosive if flare combustion is initiated. Under-estimating the mass of N2 in the flare stream not only costs money, it reduces the accuracy of the mass flow calculation performed by the flow computers FI7021 and FI7022. The computer calculates density based on the speed of sounds in the metered gas but because the algorithm used is designed for total hydrocarbon flow, so the presence of N2 in the line introduces an error into the calculation. The computer can compensate for the error if the volume fraction of N2 is known. As N2 to flare is under reported only a portion of this error is calculated. In addition, the O2 analysers, AE7000 and AE7010, were obsolete with limited spare parts available. Due to this a suitable replacement analyser had to be identified, ideally this would be installed during the August 2014 turnaround (TAR). 2
3 E. New Instrumentation/Analyser Scope The installation of a single analyser (on each flare) would provide analysis of both O2 and N2 content in the flare stream. The cost of installation would be minimised by allowing the use of existing sample lines and cabling for all of the required analysis. The proposed new analysers were Micro Gas Chromatographs (GCs). These would also provide basic composition data for the hydrocarbon content of the flare stream as shown in Figure 3. Normal Measured 8 Hr Repeatability Component Concentration Range Units % of FS HYDROGEN MOL% Calculated METHANE MOL% 0.50% ACETYLENE MOL% ETHYLENE MOL% ETHANE MOL% C2S 0-30 MOL% 0.50% C3S MOL% 0.50% C4S MOL% 0.50% C5S MOL% 0.50% 1-HEXENE MOL% HEXANE MOL% C6S 0-5 MOL% 0.50% NITROGEN MOL% 0.50% OXYGEN MOL% 0.50% CARBONDIOXIDE MOL% Stream Notes: Cycle time 2 min. Carrier gas He Group measurement of C2s, C3s, C4s, C5s and C6s without separation Fig 3 Proposed GC Measurement Ranges Existing, obsolete, paramagnetic O2 analysers AT and AT-7010 would be removed along with redundant sample handling equipment. There would also be a requirement to remove existing communications junction boxes that will no longer be required. This would in turn provide the space for mounting the new Micro GCs which would provide N2, O2, and hydrocarbon analysis of each flare stream, along with the necessary power transformer and communication junction boxes. The GCs would also require a new calibration system to be installed; this would make use of 24V dc relay outputs from the GCs to drive a solenoid valve arrangement. This would allow remote calibration to be carried out from the control room, minimising the requirement for technicians to be in the controlled flare area. The existing flowmeters would be reused but would be connected to the GCs and monitored as a digital input. These alarms would be handled by the analysers and communicated via the new communications system. Fig 4 Micro GC Overview Schematic Unfortunately the old cabling was insufficient to handle the signals from the new analysers so the existing Ethernet type analyser network on an adjacent plant was used to connect the GCs via new fibre optic runs. This gave the added benefit of providing a digital Modbus map of the new signals into the DCS for reporting, replacing the old 4-20mA analogue current loop signals. The equipment is installed in a Zone 1 hazardous area and the GCs have integral Ex d flameproof enclosures suitable for zone 1 operation. The Power Supply Unit (PSU) and Ethernet switches were combined and mounted in Ex d boxes with separate Ex e (increased safety) junction boxes for cable marshalling. Fig 5 Overview Wiring Diagram 3
4 F. Operational Results The GCs were installed during the August 2014 TAR. A remote calibration is run against a reference gas every 4 weeks. During 2015 the Scottish Environmental Protection Agency (SEPA) asked INEOS to have the GC system externally validated by an independent third party to receive Top Tier Reporting Accreditation. This verification was successfully carried out in September 2015 and Historical measured ranges were compared with the calibration gas mixture so that this could meet the requirements of ISO The whole set-up has been stable with very little drift against the 4 weekly calibrations. External validations will be carried out yearly to confirm continued performance. H. Lessons Learnt a) The original micro GC proposal was based on H2 carrier gas but this was changed to Helium before delivery. b) The liquid auto drain facility on the sampling system had lagging installed to provide frost protection and prevent blockage. c) After review, the regular health check interval was set at 4 weeks. During this procedure the full sample system including the filters and auto drains are checked and the GCs are calibrated. The facility for remote calibration and diagnostics from the control room is a very important safety feature that reduces the time technicians spend in the vicinity of the flare. The re-designed sampling systems had two parallel filters that guaranteed flow even if one filter was blinded, the set-up also utilised polymer drying tubes to eliminate all moisture before the micro GCs. The total installation has proved most robust with no flow blocks and no moisture reaching the GCs since startup. The new N2 measurement has been a great success. Originally because N2 was not measured the composition of the flare line had to be assumed to be all hydrocarbons. The measurements over two years have shown that the N2 level has always been above 60% (see Figure 8) and sometimes as much as 80% for long periods. This accurate measurement has made a massive saving on the ETS scheme and improved the volume flow to mass flow conversion calculation. G. Benefits of Micro GCs Field mountable with IP65 Ex d enclosure suitable for indoor or outdoor, hazardous area zone 1 operation without purging. No instrument air requirement, low carrier gas, low calibration gas and low power consumption Fast cycle times 2-3 minutes Very robust and stable measurements no sample pressure variations, no ambient temperature influences or memory effects for alternating contrast streams Remote diagnostics and calibration Exchange of modular standard parts instead of on-site repair 4
5 Fig 6 Sample System for new Micro GCs Fig 7 Micro GCs in Original Oxygen Analyser Cubicle 5
6 Fig 8 DCS Screenshots of Actual Measured Stream Compositions Fig 9 Micro GC Images 6
7 III. CONCLUSION The Micro GCs were compact enough to easily fit in the space left by the obsolete, unsupported O2 analysers and redundant sampling systems. Yet the new GCs provided N2, hydrocarbon and H2 analysis in addition to the original O2 measurement. All these signals were now presented digitally to the DCS and provided a much more accurate composition of the stream to the flare and hence the emissions after combustion. The re-design of the sample system increased reliability, reduced maintenance and improved safety with a remote calibration facility from the control room. ETS over-reporting ceased and the whole project had a conservative payback of less than one year. IV. ACKNOWLEDGEMENTS The authors acknowledge the contributions made to the chemical plants over the years by such UK analyser luminaries as Dave McCrae and Dave Davenport. V. REFERENCES [1] Applied Automation Analytical Training Department, 1998 Introduction to Gas Chromatography. [2] Siemens Brochure 2014 Gas Chromatography monitors Green House Gas (GHG) emission from flares. [3] Bryce and Jackson. Upgrading Process Gas Chromatographs at the BP Kinneil Terminal, Grangemouth, Scotland. PCIC Europe Conference Record, BER [4] EN/ISO/IEC 17025:2005 General Requirements for the competence of testing and calibration laboratories. VI. Gavin Dewar VITA Gavin graduated with a Higher National Certificate in Measurement, Control & Electrical Engineering from Falkirk College in 1995 and has spent over 20 years at INEOS and BP Grangemouth working in the instrument and electrical (I/E) department at the petrochemical plants handling maintenance, commissioning, design and installation. Gavin is currently the Analyser Engineer and I/E Specialist for a number of the chemical areas. Davie Plenderleith Davie graduated from Falkirk College in 1982 and has spent over 35 years at INEOS and BP Grangemouth with a break of 5 years overseas working on Middle East refineries. He has been the most senior analyser engineer on the petrochemical plants for the last 10 years. Davie is currently Analyser Subject Matter Expert (SME) for INEOS in Scotland. Eur Ing Steve Jackson MA BSc CEng Steve Jackson graduated from the University of Bath Chemical Engineering School with a BSc (Hons) in During 2016 he enhanced his academic qualifications with a Masters in Sales Management from the University of Portsmouth. In 1987 Steve became a Chartered Engineer and in 1994 a European Engineer. He has been a Member of the Institute of Measurement and Control for nearly 30 years. Over the course of his career Steve has worked for Fisher Controls, Elsag Bailey, ABB, Endress+Hauser and for the last 12 years Siemens in the fields of automation, instrumentation, process analytics and electrical engineering. Steve is also a committee member for the Energy Industries Council (EIC). Andrew Eady BEng (Hons) Andy graduated from Coventry University with a Manufacturing Engineering degree in 1991 and has spent 25 years working for the process analytics division of Siemens in various incarnations first as the Hartmann & Braun Company then as Elsag Bailey and briefly as ABB before becoming Siemens in Andy is currently the Process Analytics Technical Manager responsible for the design, quotation and implementation of gas chromatograph and continuous emissions systems. 7
Measurement & Analytics Measurement made easy. HP30 on-line process hydrogen analyzer Refinery hydrogen optimization
Measurement & Analytics Measurement made easy HP30 on-line process hydrogen analyzer Refinery hydrogen optimization A robust hydrogen analyzer package for refinery and petrochemical plants The HP30 process
More informationHILASE - QUALITY CONTROL INSTRUMENT
HILASE - QUALITY CONTROL INSTRUMENT The Hilase is a reliable and virtually maintenance free Process Analyser for the measurement of H 2 S, H 2 O, CO 2, CH 4, C 2 H 6, Methanol/Ethanol, COS, Ammonia, etc
More informationTechnical Guidance Note (Monitoring) Medium Combustion Plant Directive and Generator Controls: monitoring point source emissions
Technical Guidance Note (Monitoring) M5 Medium Combustion Plant Directive and Generator Controls: monitoring point source emissions Environment Agency XXXX 2018 Version X Draft for consultation Foreword
More informationADVANCES IN ON-LINE HYDROGEN MEASUREMENT
ADVANCES IN ON-LINE HYDROGEN MEASUREMENT Ulrich Gokeler Siemens Energy & Automation, Inc. Process Analyzers Houston, TX 77040, USA KEYWORDS Hydrogen, On-Line, Gas Analysis, Process Gas Chromatography ABSTRACT
More informationMach-RGA. The Fast Refinery Gas Analyzer. Analyzer
Mach-RGA The Fast JAS Mach-RGA Refinery gas is used as a fuel gas or as feedstock for further chemical processing. US EPA Title V Permit requires operators to monitor and report the composition of refinery
More informationHeat Value Gas Chromatograph ISO standard version
No. SS2-HGC303-0100 OVERVIEW Heat Value Gas Chromatograph ISO standard version The model HGC303 is the world smallest gas chromatograph, which is capable of analyzing 11 different components of natural
More informationFast Refinery Gas Analysis Using the Agilent 490 Micro GC QUAD
Fast Refinery Gas Analysis Using the Agilent 0 Micro GC QUAD Application Note Energy & Fuels Authors Coen Duvekot Agilent Technologies, Inc. Introduction There is a large variation in the composition and
More informationThe Process TDLS Solution
The Process TDLS Solution TDLS 200 Tunable Diode Laser Spectroscopy (TDLS) Analyzer Bulletin 11Y01B01-E-A www.yokogawa.com/an/index.html The first Tunable Diode Laser analyzer, designed specifically for
More informationSentinel LNG. Panametrics Ultrasonic Flowmeter for Cryogenic Liquids. GE Sensing & Inspection Technologies. Benefits. Applications
GE Sensing & Inspection Technologies Sentinel LNG Panametrics Ultrasonic Flowmeter for Cryogenic Liquids Benefits Improved performance, reduced maintenance and dynamic flow measurement is now available
More informationImproving Natural Gas Liquefaction Plant Performance with Process Analyzers
Process Analytics Improving Natural Gas Liquefaction Plant Performance with Process Analyzers LNG is natural gas in its liquid state with high energy density, which makes it useful for storage and transportation
More informationHAZARDOUS AREA MOTOR/DRIVE PROTECTION PRINCIPLES AND OPTIMUM SELECTION CRITERIA
HAZARDOUS AREA MOTOR/DRIVE PROTECTION PRINCIPLES AND OPTIMUM SELECTION CRITERIA Copyright Material PCIC Europe Paper No. PCIC Middle East ME17-4 Dip.-Ing Hugo Stadler Siemens Germany Eur Ing Steve Jackson
More informationSentinel LNG. Panametrics Ultrasonic Flowmeter for Cryogenic Liquids. GE Sensing & Inspection Technologies. Benefits. Applications
GE Sensing & Inspection Technologies Sentinel LNG Panametrics Ultrasonic Flowmeter for Cryogenic Liquids Benefits Improved performance, reduced maintenance and dynamic flow measurement is now available
More informationSIMPLIFY AND ACCELERATE GAS ANALYSIS. Micro GC Fusion GAS ANALYZER
SIMPLIFY AND ACCELERATE GAS ANALYSIS Micro GC Fusion GAS ANALYZER Maximum throughput and easy operation for fast decisions when they matter most Micro GC Fusion offers significant throughput gain through
More informationOnline dissolved gas analysis sensor
GAS Guard 8 Online dissolved gas analysis sensor Answers for energy. Dissolved Gas Analysis (DGA) Dissolved gas analysis is the single-most comprehensive tool for transformer condition assessment. The
More informationGlobal Calibration Gases LLC
Global Calibration Gases LLC 1090 Commerce Blvd N Sarasota, FL 34243 www.globalcalibrationgases.com ISO 17025:2005 Accredited For Calibration and Testing PGVP: N22016 GLOBAL CALIBRATION GASES IS A WORLD
More informationMicroSAM / Sitrans CV
MicroSAM / Sitrans CV Product for standard process applications and natural gas CV/BTU analysis field mountable technology leader (MEMS) Industry 09/2007, Page 1 MicroSAM / Sitrans CV & MEMS Technology
More informationAnalyser systems in 2030 a vision
Analyser systems in 2030 a vision A.L.M. van Adrichem November 5 th 2012 CPAC Seattle This presentation includes forward-looking statements. Actual future conditions (including economic conditions, energy
More informationWI/CV Multiparameter Gas Analyzer System SG12WMTA-Series
WI/CV Multiparameter Gas Analyzer System SG12WMTA-Series The Process & Fuel Gas Analysis In all industry combustion process where combustible gas are used there is a particular attention for Solutions
More information8 stream capability with dual mode (H 2 S & total Sulphur)
The model 902 H 2 S / Total Sulphur analyzer from Galvanic Applied Sciences Inc. is designed to measure H 2 S in gaseous streams. The 902 detection system is based on the lead acetate tape method. With
More informationSECTION ULTRASONIC BIOGAS FLOW MEASURING SYSTEM
SECTION 40 71 66 ULTRASONIC BIOGAS FLOW MEASURING SYSTEM PART 1 GENERAL 1.01 SUMMARY A. Provide an ultrasonic loop powered flow meter for permanent installation on the anaerobic digestion wet biogas. The
More informationHYDROGEN MEMBRANE OVERVIEW
HYDROGEN MEMBRANE OVERVIEW Advanced membrane technology for hydrogen purification and recovery www.medal.airliquide.com PROCESS PRINCIPLE Air Liquide Advanced Separations (ALaS) hollow fiber membranes
More informationExpanding the Versatility of Micro GC in the Shale Gas Play; The Agilent 490 Micro GC offers unprecedented reliability and flexibility
Expanding the Versatility of Micro GC in the Shale Gas Play; The Agilent 490 Micro GC offers unprecedented reliability and flexibility Presenter: Chuck Runkle Agilent Technologies Measure Mitigate What
More informationBest practices for flow calibration management
Best practices for flow calibration management Understanding when and why your flowmeters need to be calibrated By: Kyle Shipps, Calibration Manager & Nathan Hedrick, National Flow Product Manager, Endress+Hauser
More informationETHYLENE PLANT DRY CRACKED GAS Product Code 54801
SS2100 WATER IN ETHYLENE PLANT DRY CRACKED GAS Product Code 54801 KEY FEATURES Extremely fast analysis time 1 second updates possible Non-contact laser and detector - Reliable Tunable Diode Laser lasts
More informationPrima PRO process mass spectrometer
APPLICATION NOTE Prima PRO process mass spectrometer Accurate multi-component blast furnace gas analysis maximizes iron production and minimizes coke consumption No.?????? Author: Graham Lewis, Thermo
More informationDECENTRALIZING PROCESS GAS CHROMATOGRAPHS WOULD THERE BE ANY BENEFITS?
DECENTRALIZING PROCESS GAS CHROMATOGRAPHS WOULD THERE BE ANY BENEFITS? Ulrich Gokeler, Bob Farmer Friedhelm Mueller Siemens Applied Automation Siemens AG 500 West Highway 60 A&D PA25 Bartlesville, OK 74003,
More informationProsonic Flow B 200. For reliable flow measurement of biogas without limitations
Prosonic Flow B 200 For reliable flow measurement of biogas without limitations Specially designed ultrasonic flowmeter for measuring biogas, landfill or digester gas One-of-a-kind direct measurement of
More informationPROCESS ANALYZERS FOR PURIFIED BULK GAS QC/QA: ENTERING THE PPT RANGE
JPACSM 27 PROCESS ANALYZERS FOR PURIFIED BULK GAS QC/QA: ENTERING THE PPT RANGE Marco Succi SAES Getters S.p.A. Milano, Italy James G. Moncur and William L. Robinson Independent Consultants KEYWORDS: TRACE
More informationONLINE TOC ANALYZER B-Series
ONLINE TOC ANALYZER B-Series Features of Online TOC Analyzers Unique, Patented Two Zone Furnace Design Having two temperature zones assure complete combustion of carbon and nitrogen containing compounds
More informationCALDON LEFM Ultrasonic Flowmeters for Gas Applications. Integrating experience, proven technology, and innovation
CALDON LEFM Ultrasonic Flowmeters for Gas Applications Integrating experience, proven technology, and innovation Four-path CALDON LEFM 340Ci ultrasonic flowmeter. Contents Count on Cameron........................................................................3
More informationTDLAS. Application Note. Tunable Diode Laser Absorption Spectroscopy (TDLAS) for Carbon Dioxide Measurements In Natural Gas Processing
PROCESS INSTRUMENTS Application Note Tunable Diode Laser Absorption Spectroscopy () for Carbon Dioxide Measurements In Natural Gas Processing Introduction Natural gas is one of the world s most important
More informationGC8000 Process Gas Chromatograph. Bulletin 11B08A01-01E
GC8000 Process Gas Chromatograph Bulletin 11B08A01-01E Solutions Long History of GC Technical Excellence okogawa has a long history of supplying process gas chromatographs to the oil & gas, refining and
More informationSERVOMEX PRODUCT OVERVIEW
SERVOMEX PRODUCT OVERVIEW KILN GAS ANALYSIS SYSTEM Designed to solve production problems, but with maintenance teams in mind - simple, fast and inexpensive for low cost of ownership and long durability.
More informationNatural Gas Monitoring
Natural Gas Monitoring Natural Gas properties measurement Overview GasPT Changing the standards for gas quality measurement An extractive chromatograph sampling system can, when combined with the actual
More informationSentinel LNG. Panametrics Ultrasonic Flowmeter for Cryogenic Liquids. GE Sensing & Inspection Technologies. Benefits. Applications
GE Sensing & Inspection Technologies Sentinel LNG Panametrics Ultrasonic Flowmeter for Cryogenic Liquids Benefits - Improved performance, reduced maintenance and dynamic flow measurement is now available
More informationSolutions. Integrated Analyzer Systems & Services. Your Global Partner For Customized Analytical Solutions
Solutions Integrated Analyzer Systems & Services Your Global Partner For Customized Analytical Solutions BU 10A02A01-01E-A www.yokogawa.com/us/ia/analytical/measurementation/home.htm Core Expertise Measurementation
More informationUse of Process Analytics in Cryogenic Air Separation Plants
Use of Process Analytics in Cryogenic Air Separation Plants Chemical Industry Case Study For further treatment of primary and intermediate materials, large amounts of oxygen and nitrogen are required by
More informationWhat ensures measurement to accurate standards?
What ensures measurement to accurate standards? The SITRANS FUT1010 calibrated ultrasonic flow meter tells you exactly what s flowing Answers for Industry. SITRANS FUT1010 Custody transfer accuracy with
More informationBiogas Conference Kolding, DK, 04 April 2017 BINDER. Biogas Monitoring Why and How? Jan Talkenberger, Manager International Sales, Binder GmbH
@ Biogas Conference Kolding, DK, 04 April 2017 Biogas Monitoring Why and How? Jan Talkenberger, Manager International Sales, Binder GmbH 1 Binder Group AG Finance Holding Binder Engineering GmbH Sales
More information50 Years of PSA Technology for H2 Purification
50 Years of PSA Technology for H 2 Purification 1 50 Years of PSA Technology for H2 Purification The first industrial application of Pressure Swing Adsorption (PSA) went on stream in 1966 at a Union Carbide
More informationStationary Gas Analysis System Goliath for bio methane, landfill gas, sewage gas, mine gas and biogas
ressure Stationary Gas Analysis System Goliath for bio methane, landfill gas, sewage gas, mine gas and biogas Datasheet DS 320 E age 1 hone: +49 60 21-45 807-0 ressure Table of Contents Overview 3 General
More informationMexico Gas Flaring Reduction. EU Flaring Regulatory Measurement Requirements & Best Practice Lynn Hunter
Mexico Gas Flaring Reduction EU Flaring Regulatory Measurement Requirements & Best Practice Lynn Hunter TÜV SÜD International AMERICAS: Headquarters TÜV SÜD America Inc. Peabody, Massachusetts AMERICAS
More informationGene Beebe. September 20 th 2016
BTU Measurement from Fuel Gas to Flare Gene Beebe Vector Controls and Automation Group September 20 th 2016 2/16/2017 Vector Controls and Automation Group 2 What is BTU? Amountof heat requiredto raise
More informationSafer, More Efficient Combustion Control for Fired Heaters
Safer, More Efficient Combustion Control for Fired Heaters By Julie Valentine, Emerson Electric, Inc. Introduction Fired Heaters are devices used for high-temperature heating. A fired heater is used in
More informationUV300 LINE WATER ANALYSER TETHYS INSTRUMENTS 57, CHEMIN DU VIEUX CHÊNE, MEYLAN -FRANCE-
ON LINE WATER ANALYSER TETHYS INSTRUMENTS 57, CHEMIN DU VIEUX CHÊNE, 38240 MEYLAN -FRANCE- TEL : +33 4 76 41 86 39 - FAX : +33 4 76 41 92 27 MAIL : SALES@TETHYS-INSTRUMENTS.COM - WEB : WWW.TETHYS-INSTRUMENTS.COM
More informationKnowledge is Power /SAFEGUARDING ASSETS AND PERFORMANCE THROUGH ONLINE MEASUREMENTS
Knowledge is Power /SAFEGUARDING ASSETS AND PERFORMANCE THROUGH ONLINE MEASUREMENTS A Small Investment to Generate Measurable Benefits Sometimes, solving challenges is surprisingly easy and cost effective.
More informationHeated FID OVF Portable THC/TVOC Analyzer
Heated FID OVF-3000 Portable THC/TVOC Analyzer The OVF-3000 is light weight and portable, competitively priced compact and fully heated FID (HFID) based total hydrocarbon analyzer for high accuracy, sensitivity
More informationAuthor: Donald Pretty, Project Manager, Presenter: John Downs, Environmental Manager, B&W PGG, KVB-Enertec Products January 29, 2015
Author: Donald Pretty, Project Manager, Presenter: John Downs, Environmental Manager, B&W PGG, KVB-Enertec Products January 29, 2015 May 2007 Rule was proposed as an amendment to the NSPS Subpart J Rule
More informationDifficult CEMS Applications at Refineries & Chemical Plants
Difficult CEMS Applications at Refineries & Chemical Plants Bart Leininger, P.E. Principal Ashworth Leininger Group Camarillo, CA 4/3/18 Page 1 Objectives Discussion of CEMS and Process Measurement Sample
More informationOptiPEAK TDL600 Moisture in Natural Gas Analyzer
OptiPEAK TDL600 Moisture in Natural Gas Analyzer Process Analyzers The next generation TDLAS Analyzer for automatic online measurement of moisture in variable compositions of natural gas and biomethane.
More informationoil standards TRUE NORTH AND ATLANTIS
oil standards TRUE NORTH AND ATLANTIS Accuracy Matters Accurate lab data is critical to sound decision making. Running certified oil standards is the ideal way to ensure the accuracy of your DGA results.
More informationWhite Paper. Solutions to Contain VOC Emissions and Comply with Evolving EPA Regulations.
White Paper. Solutions to Contain VOC Emissions and Comply with Evolving EPA Regulations. 02 Solutions to Contain VOC Emissions and Comply with Evolving EPA Regulations. Introduction Terminal operators,
More informationMeet Ja, H2S, and BTU flare regulations while reducing your total cost of ownership
Process Analytics Meet Ja, H2S, and BTU flare regulations while reducing your total cost of ownership As environmental regulations for flare emissions become more stringent, companies are facing immense
More informationPROCESS MOISTURE ANALYZERS Measuring moisture in gas or HC liquids in hazardous areas
PROCESS MOISTURE ANALYZERS Measuring moisture in gas or HC liquids in hazardous areas EExd Construction Safety by containment PROCESS MOISTURE ANALYZER Channel 1 dew-point & pressure sensor Through-glass
More informationSG06UNINC-Series BGA/IP55/P Series in Outdoor Designed
SG06UNINC-Series BGA/IP55/P Series in Outdoor Designed BIOGAS measurement and analysis in continuous of methane, carbon dioxide, hydrogen sulphide, oxygen and hydrogen. BGA/INCA as INnovative Case Analyzer
More informationAn Industry Perspective: Significant Challenges and Potential Costs of CCS
An Industry Perspective: Significant Challenges and Potential Costs of CCS Slide 1 30 October, 2013 An Introduction to INEOS INEOS is a global manufacturer of petrochemicals, specialty chemicals and oil
More informationStack Gas Analyzer ENDA-7000 ULTIMATE. Ultimate Reliability. Ultimate. Ultimate. Dependability. Accessibility
Stack Gas Analyzer ENDA-7000 ULTIMATE Ultimate Reliability Ultimate Dependability Ultimate Accessibility ENDA-7000 ULTIMATE Experience HORIBA, the global Single Source Solutions Provider with over 50 years
More informationFLUXUS G. Non-Invasive Gas Flow Measurement. External Measurement of Internal Flow. Natural Gas Extraction. Offshore. Gas Transmission.
FLUXUS G Non-Invasive Gas Flow Measurement Natural Gas Extraction Offshore Gas Transmission Storage Chemical Industry Petrochemical Industry Manufacturing Industry External Measurement of Internal Flow
More informationABB MEASUREMENT & ANALYTICS ANALYTICAL MEASUREMENT. Ethylene plant production excels with the PGC1000
ABB MEASUREMENT & ANALYTICS ANALYTICAL MEASUREMENT Ethylene plant production excels with the PGC1000 2 P G C 1 0 0 0 E T H Y L E N E P L A N T P R O D U C T I O N P B/ 2 1 0 2 74 2 - E N Measurement made
More informationGas Analyzer Spare Part Recommendations
Gas Analyzer Spare Part Recommendations Reliable gas analysis data can play an important role in a wide range of industries. Failure of gas analyzers can interrupt that data flow and contribute to costly
More informationRoad Tunnel Atmosphere Monitoring Systems TunnelCraft 3 - Air Quality Monitor Continuous measurement of NO2 and Visibility in road tunnels
TunnelTech 404 CODEL NO2, and Visibility Air Quality Monitor Road Systems TunnelCraft 3 - Air Quality Monitor Continuous measurement of NO2 and Visibility in road tunnels High accuracy determination of
More informationBINDER. 3rd NATIONAL BIOGAS CONFERENCE Midrand, South Africa, 03 Nov Biogas Monitoring Technologies, Challenges, Solutions
@ 3rd NATIONAL BIOGAS CONFERENCE Midrand, South Africa, 03 Nov 2017 Biogas Monitoring Technologies, Challenges, Solutions Jan Talkenberger, Manager International Sales, Binder GmbH 1 Introduction Binder
More informationSYNENERGY. Unifying hydrocarbon management in a single system. Oil & Gas flow metering and calibration supervisory. Analyser management
SYNENERGY Unifying hydrocarbon management in a single system Oil & Gas flow metering and calibration supervisory Analyser management Pipeline Leak Detection and Localisation SynEnergy Hydrocarbon Management
More informationSubmerged Motor LNG Pumps In Send-Out System Service. Unique pumps for unique duty...
A pump is a pump, right? Not exactly, given all the variations on the theme. Take, for example those pumps primarily used for transfer of liquefied natural gas (LNG) and other liquefied gases. They re
More informationOxygen & Other Gases Measurement and Analysis
Oxygen & Other Gases Measurement and Analysis Protect processes, reduce costs and maintain quality Oxygen & Binary Gas Measurement and Analysis A range of quality process analyzers from experts in control
More informationH2scan Market Overview
H2scan Market Overview H2scan Corporation Valencia, California WRS Sensor Cap Hydrogen Sensor Systems H2scan is the leading provider of high performance hydrogen specific sensor systems, which provide
More informationREPORT. Hobré HIGAS, a Portable Natural Gas Quality Meter. Wobbe Index Analyzer. Project report September 2015
Hobré HIGAS, a Portable Natural Gas Quality Meter Wobbe Index Analyzer Project report September 2015 REPORT Danish Gas Technology Centre Dr. Neergaards Vej 5B DK-2970 Hørsholm Tlf. +45 2016 9600 www.dgc.dk
More informationMIDSTREAM GAS QUALITY
Pellmann 20 th June 2017 MIDSTREAM GAS QUALITY Measurement Solutions Honeywell Internal What we measure and why: 1 Heating Value Density Compressibility Composition Volume to energy to $value About Our
More informationTVOC Heated FID VE 7
Rack Mount/Table Top THC Analyzer TVOC Heated FID VE 7 The VE7 19"/5PU rack mount and table top heated emission analyzer for the continuous determination of the mass concentration of total gaseous organic
More informationRosemount Analytical 700XA Process Gas Chromatograph
Rosemount Analytical 700XA Process Gas Chromatograph Our leading-edge instruments and applications expertise based on a rich heritage of more than 70 years of online process analysis experience make Emerson
More informationSTRENGTHENING PLANT SAFETY IN COAL GRINDING MILLS AND STORAGE SILOS WITH CARBON MONOXIDE MONITORING
COAL HANDLING SAFETY STRENGTHENING PLANT SAFETY IN COAL GRINDING MILLS AND STORAGE SILOS WITH CARBON MONOXIDE MONITORING Risks of unwanted combustion potentially causing injury, damage and downtime occur
More informationCritical Ethylene Level and Flow Applications and Case Studies
Special Applications Series: Critical Ethylene Level and Flow Applications and Case Studies Maintaining Safety, Reliability and Productivity In ethylene plants, balancing environment, health, and safety
More informationPerformance and innovation a team effort. Our MicroSAM and MAXUM edition II gas chromatographs. Process Analytics. Answers for industry.
Performance and innovation a team effort Our MicroSAM and MAXUM edition II gas chromatographs. Process Analytics Answers for industry. Contents Why choose gas chromatography? We know what s in it we ve
More informationEuroFID Total Hydrocarbon Analyzer
Product information EuroFID Total Hydrocarbon Analyzer Precise Determination of Total Hydrocarbons in Air for Corrosive as well as Condensing Gases Analyzers and Process Instrumentation Proven Analyzer
More informationContinuous Flow Monitoring. In-situ, low maintenance devices for continuous measurement of hot gas flow
Product Data Sheet CODEL Continuous Flow Monitoring VCEM5000/5100 Flue Gas Flow Monitor In-situ, low maintenance devices for continuous measurement of hot gas flow Non-contact infra-red sensing Continuous
More informationAPPLICATION NOTE. UOP : Analysis of Trace CO and CO 2 in bulk H 2 and Light Gaseous Hydrocarbons by GC
UOP 63-13: Analysis of Trace CO and CO 2 in bulk H 2 and Light Gaseous Hydrocarbons by GC Fast Analysis in
More informationPRECISE CONTROL OF PROTECTIVE ATMOSPHERES
The zirconia carbon sensor has been used for nearly three decades to control the carbon potential in many carburizing applications. More recently, the same sensor has been used very effectively in a variety
More informationNew advances in mine site gas analysis using gas chromatographs
University of Wollongong Research Online Coal Operators' Conference Faculty of Engineering and Information Sciences 2017 New advances in mine site gas analysis using gas chromatographs Lauren Forrester
More informationAsset integrity solutions. Improve integrity and reduce costs through more effective asset management. prime. Risk. Assessment.
Opportunity Asset integrity solutions Improve integrity and reduce costs through more effective asset management Risk Competency prime Training & Development Value Creation Risk Management 2 ASSET INTEGRITY
More informationSECTION PRESSURE AND DIFFERENTIAL PRESSURE GAUGES
SECTION 40 72 43 PRESSURE AND DIFFERENTIAL PRESSURE GAUGES PART 1- GENERAL A. The Waterpilot FMX21 is designed for hydrostatic level measurement. It has a robust ceramic sensor and integrated temperature
More informationFast On-Site Mine Safety Analysis by the Agilent 490 Micro GC
Fast On-Site Mine Safety Analysis by the Agilent 490 Micro GC Application Note Micro Gas Chromatography, Mine gas analysis, Mine safety Authors Darren Brady Safety in Mines Testing and Research Station
More informationPaper TO7-2 DEVELOPMENT OF REAL-TIME GAS QUALITY MEASUREMENT
International Gas Union Research Conference 2014 Paper TO7-2 DEVELOPMENT OF REAL-TIME GAS QUALITY MEASUREMENT Terry Williams Orbital Global Solutions United Kingdom 1 1. INTRODUCTION This paper maps the
More informationContinuous Emissions Monitoring Systems. Engineering & Design. For PetroChem Processes. a technical solution to meet every need
Continuous Emissions Monitoring Systems Engineering & Design For PetroChem Processes a technical solution to meet every need Cemtek Environmental Inc. 3041 S. Orange Ave. Santa Ana, CA 92707 800-400-0200
More informationPRODUCT INFORMATION MCS100E HW/PD/CD MULTICOMPONENT ANALYZER SYSTEMS EMISSION AND RAW GAS MONITORING WITH HOT MEASUREMENT.
PRODUCT INFORMATION MCS100E HW/PD/CD MULTICOMPONENT ANALYZER SYSTEMS EMISSION AND RAW GAS MONITORING WITH HOT MEASUREMENT CEMS Solution CONTINUOUS, EXTRACTIVE FLUE GAS MONITORING On MCS100E HW, from sampling
More informationHeated FID 109A. Rack Mount/Table Top NMGOC Analyzer
Rack Mount/Table Top NMGOC Analyzer Heated FID 109A J.U.M. Engineering 2014/2016 The 109A is a rack mount and table top heated emission analyzer for the continuous / simultaneous determination of the mass
More informationAdvanced Automation Lowers Labor Cost and Improves Performance at the Las Palmas, California Wastewater Treatment Plant
Advanced Automation Lowers Labor Cost and Improves Performance at the Las Palmas, California Wastewater Treatment Plant By Nate Ventress, Process Automation Specialist, Western Region, Festo The Salinas
More informationExperience with Ultrasonic meters for Biogas
Products Solutions Services Experience with Ultrasonic meters for Biogas Gastekniske Dage 2017 - Billund Slide 1 What is Biogas? Biogas typically refers to a gas produced by the biological breakdown of
More informationIntrinsic Safety & Hazardous Areas
Intrinsic Safety & Hazardous Areas by Steve Mackay 1 EIT Micro-Course Series Every two weeks we present a 35 to 45 minute interactive course Practical, useful with Q & A throughout PID loop Tuning / Arc
More informationIII.P. Recovery and utilization of waste gas in refinery facilities
TYPE III - OTHER PROJECT ACTIVITIES Project participants shall take into account the general guidance to the methodologies, information on additionality, abbreviations and general guidance on leakage provided
More informationP r o d u C T I N f o r m at I o n. The atmosphere must be just right
P r o d u C T I N f o r m at I o n INESIC700 TBS The atmosphere must be just right INESIC700 TBS The atmosphere must be just right onitoring gas levels in underground coal mines is a challenging task and
More informationELECTRONIC EARTHING EQUIPMENT SERIES GRD 4200
ELECTRONIC EARTHING EQUIPMENT SERIES GRD 4200 The GRD 4200 is an electronic earthing equipment for tankers during loading and unloading operations of flammable liquids in places with danger of explosion.
More informationHYDROSTAB SANERINGSTECHNIEK VOOR STORTPLAATSEN TECHNIQUE D ASSAINISSEMENT POUR CET
Annex 4 MONITORING PLAN HUNGARIAN PROGRAMME FOR REHABILITATION OF LANDFILLS THROUGH MITIGATION OF METHANE GAS EMISSIONS (9 JANUARY 2006) 1 CONTENTS 1 INTRODUCTION 3 2 PROJECT-SPECIFIC INDICATORS 4 2.1
More informationWell Pad Automation Improves Capital Efficiency and Reduces Fiscal Risk Michael Machuca* Emerson Process Management
URTeC: 1874921 Well Pad Automation Improves Capital Efficiency and Reduces Fiscal Risk Michael Machuca* Emerson Process Management Copyright 2014, Unconventional Resources Technology Conference (URTeC)
More informationHydrocarbon Processing Industry Gas Chromatographs. Standard GC Analyzer Solutions
Hydrocarbon Processing Industry Gas Chromatographs Standard GC Analyzer Solutions Scion Instrument s GC Analyzer specialists have the knowledge and experience to provide pre-configured gas chromatographic
More informationSiemens Total Solution for Greenhouse Gas (GHG) Emissions from Fuel Gas
Siemens Total Solution for Greenhouse Gas (GHG) Emissions from Fuel Gas 40 CFR Part 98 MANDATORY REPORTING OF GREENHOUSE GAS EMISSIONS www.usa.siemens.com/processanalytics 1 Regulation Intent On September
More informationThe Proper Mechanism for Field Validating Moisture Analyzers. GE Measurement & Sensing GE Oil & Gas
The Proper Mechanism for Field Validating Moisture Analyzers Frank Gerritse Narge Sparages GE Measurement & Sensing GE Oil & Gas Abstract Field verification of moisture analyzer calibrations mitigates
More informationMulti-component gas analysis with speed, repeatability, and reliability.
Multi-component gas analysis with speed, repeatability, and reliability. Rosemount Quantum Cascade Laser Analyzers Achieve unmatched gas analysis performance and realize cost savings with the most advanced
More informationX-STREAM Flameproof Gas Analyzers
Product Data Sheet Product Data X-STREAM Flameproof Gas Analyzers Rosemount Analytical s X-STREAM flameproof process gas analyzers are now available in the standard model (right) and the enhanced model
More informationNon-Invasive Gas Flow Measurement
www.flexim.com FLUXUS G Non-Invasive Gas Flow Measurement Gas Storage Facilities Gas Transmission Natural Gas Extraction Chemical Industry Petrochemical Industry Offshore Manufacturing Industry Measure
More informationCOMBIMASS Thermal Gas Flow
COMBIMASS Thermal Gas Flow The COMBIMASS series enables gas flow measurement for various applications in all industry sectors. COMBIMASS works according to the principle of thermal dispersion and determines
More information